BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO.

2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 1 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Permanent Facility
ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
1 AC current (@50 Hz) Multimeter by Direct 1 A to 10 A 0.49%
Current (< 1
Method.
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
2 AC current @50Hz Multimeter by Direct 1 mA to 100 mA 0.38 % to 0.12 %
Current (< 1
Method.
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
3 AC current @50Hz Multimeter by Direct 100 mA to 1 A 0.19 % to 0.20 %
Current (< 1
Method.
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
4 AC Voltage @50Hz Multimeter by Direct 100 mV to 100 V 0.12 % to 0.11 %
Current (< 1
Method.
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 2 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
5 AC Voltage @50Hz Multimeter by Direct 100 V to 1000 V 0.11 %
Current (< 1
Method.
GHz)
(Measure)
ELECTRO-
Using 5½ Digital
TECHNICAL-
AC current (Source) Multifunction
6 Alternating 1 mA to 2 A 0.30 % to 0.67 %
@50 Hz Calibrator by Direct
Current (< 1
Method.
GHz) (Source)
ELECTRO- Using 5½ Digital
TECHNICAL- Multifunction
AC current (Source)
7 Alternating Calibrator with 2 A to 10 A 0.67 % to 0.38 %
@50 Hz
Current (< 1 Current Coil by
GHz) (Source) Direct Method
ELECTRO- Using 5½ Digital
TECHNICAL- Multifunction
AC current (Source)
8 Alternating Calibrator with 10 A to 900 A 1.5%
with CC @50 Hz
Current (< 1 Current Coil by
GHz) (Source) Direct Method.
ELECTRO-
Using 5½ Digital
TECHNICAL-
AC Voltage (Source) Multifunction
9 Alternating 200 mV to 200 V 0.30 % to 0.27 %
@50Hz Calibrator by Direct
Current (< 1
Method.
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 3 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using 5½ Digital
TECHNICAL-
AC Voltage (Source) Multifunction
10 Alternating 200 V to 1000 V 0.27 % to 0.22 %
@50Hz Calibrator by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
AC Voltage (Source) Multifunction
11 Alternating 5 mV to 200 mV 1.12 % to 0.30 %
@50Hz Calibrator by Direct
Current (< 1
Method.
GHz) (Source)
ELECTRO-
TECHNICAL- Using 6½ Digital
12 DIRECT DC current Multimeter by Direct 1 A to 10 A 0.1 % to 0.19 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
13 DIRECT DC Current Multimeter Direct 1 mA to 100 mA 0.12%
CURRENT Method.
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
14 DIRECT DC Current Multimeter by Direct 100 mA to 1 A 0.12 % to 0.1 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL-
Using 6 1/2 DMM
15 DIRECT DC Voltage 1 mV to 100 mV 0.2 % to 0.1 %
,Direct method.
CURRENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 4 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digital
16 DIRECT DC Voltage Multimeter by Direct 100 mV to 100 V 0.01%
CURRENT Method.
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
17 DIRECT DC Voltage Multimeter by Direct 100 V to 1000 V 0.01 % to 0.13 %
CURRENT Method
(Measure)
ELECTRO- Using 5½ Digital
TECHNICAL- Multifunction
DC current (Source
18 DIRECT Calibrator with 10 A to 900 A 1.3%
with CC)
CURRENT Current Coil by
(Source) Direct Method
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
19 DIRECT DC current (Source) 1 mA to 200 mA 0.19 % to 0.21 %
Calibrator by Direct
CURRENT
Method.
(Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
20 DIRECT DC current (Source) 2 A to 10 A 0.22 % to 0.23 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
21 DIRECT DC current (Source) 200 mA to 2 A 0.21 % to 0.23 %
Calibrator by Direct
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 5 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using 5 1/2 MFC
22 DIRECT DC Voltage 1 mV to 200 mV 1.04 % to 0.12 %
Direct method.
CURRENT
(Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
23 DIRECT DC Voltage (Source) 200 mV to 200 V 0.12%
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
24 DIRECT DC Voltage (Source) 200 V to 1000 V 0.12 % to 0.16 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Using 6½ Digital
100K Ohm to 10M
25 ELECTRICAL Resistance ( 2 wire ) Multimeter by Direct 0.01 % to 0.06 %
Ohm
EQUIPMENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
100M Ohm to 1G
26 ELECTRICAL Resistance ( 2 wire ) Multimeter by Direct 0.94 % to 3.38 %
Ohm
EQUIPMENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
10 Ohm to 100K
27 ELECTRICAL Resistance (2 wire) Multimeter by Direct 0.06 % to 0.012 %
Ohm
EQUIPMENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 6 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digital
10M Ohm to 100M
28 ELECTRICAL Resistance (2 wire) Multimeter by Direct 0.06 % to 0.94 %
Ohm
EQUIPMENT Method
(Measure)
ELECTRO-
TECHNICAL- Using Decade
100M Ohm to
29 ELECTRICAL Resistance ( 2 wire ) Resistance Box by 1.82 % to 3.1 %
900Mohm Ohm
EQUIPMENT Direct Method
(Source)
ELECTRO-
TECHNICAL- Using Decade
10M Ohm to 100M
30 ELECTRICAL Resistance ( 2 wire ) Resistance Box by 0.58 % to 1.82 %
Ohm
EQUIPMENT Direct Method.
(Source)
ELECTRO-
TECHNICAL- Using Decade
31 ELECTRICAL Resistance (2 wire) Resistance Box by 1 Ohm to 100K Ohm 1.36 % to 0.12 %
EQUIPMENT Direct Method
(Source)
ELECTRO-
TECHNICAL- Using Decade
Resistance 100K Ohm to 10M
32 ELECTRICAL Resistance Box by 0.12 % to 0.58 %
(Source)(2 wire) Ohm
EQUIPMENT Direct Method
(Source)
ELECTRO-
TECHNICAL- Using 6½ Digital
50 Hz @1V to 1K Hz
33 TIME & Frequency Multimeter by Direct 0.12%
@1V
FREQUENCY Method.
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 7 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Digital Time
34 TIME & Timer/Stop Watch Totaliser by Direct 3 s to 3600 s 0.37 s to 4.2 s
FREQUENCY Method
(Measure)
Using Digital
MECHANICAL- Tachometer and
Tachometer (Non- 100 rpm to 1000
35 ACCELERATION RPM calibrator as 9.18rpm
Contact Type) rpm
AND SPEED per Sanas TR 45-01
Guidelines.
Using Digital
MECHANICAL- Tachometer and
Tachometer (Non- 1000 rpm to 90000
36 ACCELERATION RPM calibrator as 10 rpm
Contact Type) rpm
AND SPEED per Sanas TR 45-01
Guidelines.
MECHANICAL-
DIMENSION
Bevel Protractors Angle gauges ( 4
(BASIC
37 (LC:5 minutes of quadrants), By 0 ° to 90 ° 2min. of arc
MEASURING
arc.) Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
Bore Gauge with
DIMENSION
Dial (For Dial Calibration
(BASIC
38 Transmission Tester / ULM , By 0 to 1 mm 4.22µm
MEASURING
Accuracy) LC: 0.001 Comparison Method
INSTRUMENT,
mm
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 8 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Caliper Checker &
(BASIC Calipers (L.C.: Slip Gauge '0' Grade 0
39 13.43µm
MEASURING 0.01mm) by comparison to 600 mm
INSTRUMENT, method
GAUGE ETC.)
MECHANICAL-
DIMENSION Slip Gauges &
(BASIC Coating Thickness Digital Comparator
40 0.01 mm to 1.0 mm 4.43 µm
MEASURING Foils ,By Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Using Standard Foils
(BASIC Coating Thickness
41 , By Comparison 0.02 mm to 1.5 mm 3.48µm
MEASURING Gauge
Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Angle Gauges &
Combination
(BASIC Vision Measuring 0°
42 Set/Degree 18min of Arc.
MEASURING machine ,By to 90° (4 quadrants)
protractor L.C : 1°
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Digital Caliper and 50 mm
(BASIC
43 CUBE MOULDS Digital Protractor ,By 31.73µm
MEASURING
Comparison Method to 150 mm
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 9 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
Slip Gauge set with
DIMENSION
Cylindrical DRO and
(BASIC
44 Measuring Pins , By Comparator Stand , 0.1 mm to 20 mm 4.43µm
MEASURING
Comparison Method By Comparison
INSTRUMENT,
Method:
GAUGE ETC.)
MECHANICAL-
DIMENSION
Slip Gauge Set &
(BASIC Depth Micrometer 0
45 Long Slip gauges ,By 5.76µm
MEASURING L.C.: 0.01 mm to 300 mm
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Dial Indicator (Lever
(BASIC ULM Machine, By 0
46 Type) L.C.: 0.001 1.5µm
MEASURING Comparison Method to 0.140 mm
mm
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Dial Indicator
(BASIC ULM Machine , By 0
47 (Plunger Type) L.C.: 2µm
MEASURING Comparison Method to 50 mm
0.001 mm
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Slip Gauges &
Dial Thickness
(BASIC Micrometer Check 0
48 Gauge L.C.: 0.001 1µm
MEASURING slip gauges, By to 10 mm
mm
INSTRUMENT, Comparison Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 10 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION Micrometer
External
(BASIC Checking Slip 0
49 Micrometers (L.C.: 2.23µm
MEASURING Gauges , By to 150 mm
0.001 mm)
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
Micrometer
DIMENSION
External Checking Slip
(BASIC
50 Micrometers (L.C.: Gauges and Long 150 mm to 300 mm 4.8µm
MEASURING
0.01 mm) Slip gauges , By
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
Micrometer
DIMENSION
External Checking Slip Gauge
(BASIC
51 Micrometers (L.C.: and Long Slip 300 mm to 500 mm 9.7µm
MEASURING
0.01 mm) gauges, By
INSTRUMENT,
Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Slip Gauges &
(BASIC Digital Comparator , 0.01 mm
52 Feeler Gauge 4.43µm
MEASURING By Comparison to 1.0 mm
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Slip Gauge Set &
Internal Micrometers
(BASIC Lever dial indicator ,
53 (Stick type) (L.C.: 50 mm to 175 mm 5µm
MEASURING By Comparison
0.01 mm)
INSTRUMENT, Method
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 11 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Tape & Scale
(BASIC Measuring Tape 0 290 sqrt (L) µm,
54 Calibrator , By
MEASURING L.C.:1 mm to 15 m where L is in m
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Tape & Scale
(BASIC Metric Steel Scale 0
55 Calibrator , By 145.27 µm
MEASURING L.C.: 0.5 mm to 1000 mm
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION Slip Gauges &
(BASIC Micrometer setting Digital Comparator ,
56 25 mm to 500 mm 10µm
MEASURING pieces By Comparison
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Slip Gauges & Dial
(BASIC
57 Plain Plug Gauge Comparator , By 5 mm to 300 mm 8µm
MEASURING
Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Using ULM , By
58 Plain Ring Gauges 5 mm to 100 mm 3µm
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 12 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Radius Gauges Vision Measuring
(BASIC
59 (Concave & Convex Machine , By 0.6 mm to 25 mm 8µm
MEASURING
Profiles) Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Slip Gauges , By
60 Snap Gauge 5 mm to 300 mm 4.47µm
MEASURING Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION FCDM, Setting
Taper Thread Plug
(BASIC Master & Wire
61 Gauge (Effective 3 mm to 95 mm 6µm
MEASURING Gauge Set , By
Diameter Only)
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Universal Length
Taper Thread Ring
(BASIC Measuring Machine ,
62 Gauge (Effective 5 mm to 95 mm 4µm
MEASURING By Comparison
Diameter Only)
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
Vision measuring
(BASIC Test Sieves
63 Machine , By 0.045 mm to 10 mm 8µm
MEASURING (Brass/G.I.)
Comparison Method:
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 13 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
(BASIC Test Sieves Digital Caliper , By
64 10 mm to 125 mm 20.4µm
MEASURING (Brass/G.I.) Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION FCDM, Setting
Thread Plug Gauges
(BASIC Master & Wire
65 (Major & Effective 3 mm to 95 mm 4.5µm
MEASURING Gauge Set , By
Diameter Only)
INSTRUMENT, Comparison Method
GAUGE ETC.)
MECHANICAL-
DIMENSION Universal Length
Thread Ring Gauges
(BASIC Measuring Machine,
66 (Metric) Effective 5 mm to 95 mm 3.2µm
MEASURING By Comparison
Diameter only
INSTRUMENT, Method
GAUGE ETC.)
MECHANICAL-
DIMENSION
(BASIC Ultrasonic Thickness Slip Gauges , By
67 0.1 mm to 100 mm 65µm
MEASURING Gauge (LC: 0.1 mm) Comparison Method
INSTRUMENT,
GAUGE ETC.)
MECHANICAL-
DIMENSION
Vernier Depth Slip Gauge Set &
(BASIC 0
68 Gauge (L.C.: 0.02 Caliper Checker, By 10.6µm
MEASURING to 600 mm
mm) Comparison Method
INSTRUMENT,
GAUGE ETC.)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 14 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

MECHANICAL-
DIMENSION
Vernier Height Caliper Checker &
(BASIC 0
69 Gauge (L.C.: 0.01 Slip Gauge Set, By 13.5µm
MEASURING to 600 mm
mm) Comparison Method
INSTRUMENT,
GAUGE ETC.)
Using Test Gauge
MECHANICAL-
(Hydraulic Pressure) and Comparator
PRESSURE
70 Dial and Digital (Media Oil) by >35 bar to 350 bar 0.86bar
INDICATING
Pressure Gauges Comparison Method
DEVICES
as per DKD-R6-1
Using Test Gauge
MECHANICAL-
(Hydraulic Pressure) and Comparator
PRESSURE
71 Dial and Digital (Media Oil) by >350 bar to 600 bar 1.13bar
INDICATING
Pressure Gauges Comparison Method
DEVICES
as per DKD-R6-1
Using Test Gauge
MECHANICAL-
(Hydraulic Pressure) and Comparator
PRESSURE
72 Dial and Digital (Media Oil) by 0 to 35 bar 0.2bar
INDICATING
Pressure Gauges Comparison Method
DEVICES
as per DKD-R6-1
Using Test Gauge
MECHANICAL-
and Vacuum pump
PRESSURE Dial and Digital
73 by Comparison -0.96 bar to 0 0.01bar
INDICATING Vacuum Gauges
Method as per DKD-
DEVICES
R6-1

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 15 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using Humidity
Temperature Sensor with
THERMAL-
Indicator Indicator, Humidity 25 °C to 40 ° C @
74 SPECIFIC HEAT 1.21°C
(THERMOHYGROMET & Temperature 50% RH
& HUMIDITY
ER) Chamber by
Comparison Method
Using Humditiy
Sensor with
THERMAL-
Indicator, Humidity 35 % RH to 90 %
75 SPECIFIC HEAT Thermo-Hygro meter 3.16% RH
& Temperature RH @ 25°C
& HUMIDITY
Chamber by
Comparison Method
Liquid in Glass Using RTD (4 wire),
THERMAL- Thermometer, Dial 6½ DMM, Oil Bath
76 -30 °C to 250 °C 1.30°C
TEMPERATURE Thermometer, by Comparision
Digital Thermometer Method
Using Standard
Pyrometer and Black
THERMAL- Non-Contact Type
77 Body Source 50 °C to 600 °C 6°C
TEMPERATURE Infrared Pyrometer
(Emissivity 0.95) by
comparison method
Using RTD (4 wire),
RTD, Thermocouple
THERMAL- 6½ DMM, Dry Block
78 with or without -10 °C to 400 °C 0.73°C
TEMPERATURE Furance by
Indicator/Controller
Comparision Method
Using 'R' Type T/C,
Thermocouple
THERMAL- 6½ DMM, Dry Block
79 Indicator/Controller 400 °C to 1200 °C 4°C
TEMPERATURE Furnace by
with Sensor
Comparison Method

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 16 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Site Facility
ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
1 AC current (@50 Hz) Multimeter by Direct 1 A to 10 A 0.49%
Current (< 1
Method.
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
2 AC current @50Hz Multimeter by Direct 1 mA to 100 mA 0.38 % to 0.12 %
Current (< 1
Method.
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
3 AC current @50Hz Multimeter by Direct 100 mA to 1 A 0.19 % to 0.20 %
Current (< 1
Method.
GHz)
(Measure)
ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
4 AC Voltage @50Hz Multimeter by Direct 100 mV to 100 V 0.12 % to 0.11 %
Current (< 1
Method.
GHz)
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 17 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using 6½ Digital
Alternating
5 AC Voltage @50Hz Multimeter by Direct 100 V to 1000 V 0.11 %
Current (< 1
Method.
GHz)
(Measure)
ELECTRO-
Using 5½ Digital
TECHNICAL-
AC current (Source) Multifunction
6 Alternating 1 mA to 2 A 0.30 % to 0.67 %
@50 Hz Calibrator by Direct
Current (< 1
Method.
GHz) (Source)
ELECTRO- Using 5½ Digital
TECHNICAL- Multifunction
AC current (Source)
7 Alternating Calibrator with 2 A to 10 A 0.67 % to 0.38 %
@50 Hz
Current (< 1 Current Coil by
GHz) (Source) Direct Method
ELECTRO- Using 5½ Digital
TECHNICAL- Multifunction
AC current (Source)
8 Alternating Calibrator with 10 A to 900 A 1.5%
with CC @50 Hz
Current (< 1 Current Coil by
GHz) (Source) Direct Method.
ELECTRO-
Using 5½ Digital
TECHNICAL-
AC Voltage (Source) Multifunction
9 Alternating 200 mV to 200 V 0.30 % to 0.27 %
@50Hz Calibrator by Direct
Current (< 1
Method.
GHz) (Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 18 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
Using 5½ Digital
TECHNICAL-
AC Voltage (Source) Multifunction
10 Alternating 200 V to 1000 V 0.27 % to 0.22 %
@50Hz Calibrator by Direct
Current (< 1
Method
GHz) (Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
AC Voltage (Source) Multifunction
11 Alternating 5 mV to 200 mV 1.12 % to 0.30 %
@50Hz Calibrator by Direct
Current (< 1
Method.
GHz) (Source)
ELECTRO-
TECHNICAL- Using 6½ Digital
12 DIRECT DC current Multimeter by Direct 1 A to 10 A 0.1 % to 0.19 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
13 DIRECT DC Current Multimeter Direct 1 mA to 100 mA 0.12%
CURRENT Method.
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
14 DIRECT DC Current Multimeter by Direct 100 mA to 1 A 0.12 % to 0.1 %
CURRENT Method
(Measure)
ELECTRO-
TECHNICAL-
Using 6 1/2 DMM
15 DIRECT DC Voltage 1 mV to 100 mV 0.2 % to 0.1 %
,Direct method.
CURRENT
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 19 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digital
16 DIRECT DC Voltage Multimeter by Direct 100 mV to 100 V 0.01%
CURRENT Method.
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
17 DIRECT DC Voltage Multimeter by Direct 100 V to 1000 V 0.01 % to 0.13 %
CURRENT Method
(Measure)
ELECTRO- Using 5½ Digital
TECHNICAL- Multifunction
DC current (Source
18 DIRECT Calibrator with 10 A to 900 A 1.3%
with CC)
CURRENT Current Coil by
(Source) Direct Method
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
19 DIRECT DC current (Source) 1 mA to 200 mA 0.19 % to 0.21 %
Calibrator by Direct
CURRENT
Method.
(Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
20 DIRECT DC current (Source) 2 A to 10 A 0.22 % to 0.23 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
21 DIRECT DC current (Source) 200 mA to 2 A 0.21 % to 0.23 %
Calibrator by Direct
CURRENT
Method
(Source)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 20 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL-
Using 5 1/2 MFC
22 DIRECT DC Voltage 1 mV to 200 mV 1.04 % to 0.12 %
Direct method.
CURRENT
(Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
23 DIRECT DC Voltage (Source) 200 mV to 200 V 0.12%
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
Using 5½ Digital
TECHNICAL-
Multifunction
24 DIRECT DC Voltage (Source) 200 V to 1000 V 0.12 % to 0.16 %
Calibrator by Direct
CURRENT
Method
(Source)
ELECTRO-
TECHNICAL- Using 6½ Digital
100K Ohm to 10M
25 ELECTRICAL Resistance ( 2 wire ) Multimeter by Direct 0.01 % to 0.06 %
Ohm
EQUIPMENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
100M Ohm to 1G
26 ELECTRICAL Resistance ( 2 wire ) Multimeter by Direct 0.94 % to 3.38 %
Ohm
EQUIPMENT Method
(Measure)
ELECTRO-
TECHNICAL- Using 6½ Digital
10 Ohm to 100K
27 ELECTRICAL Resistance (2 wire) Multimeter by Direct 0.06 % to 0.012 %
Ohm
EQUIPMENT Method
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 21 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using 6½ Digital
10M Ohm to 100M
28 ELECTRICAL Resistance (2 wire) Multimeter by Direct 0.06 % to 0.94 %
Ohm
EQUIPMENT Method
(Measure)
ELECTRO-
TECHNICAL- Using Decade
100M Ohm to
29 ELECTRICAL Resistance ( 2 wire ) Resistance Box by 1.82 % to 3.1 %
900Mohm Ohm
EQUIPMENT Direct Method
(Source)
ELECTRO-
TECHNICAL- Using Decade
10M Ohm to 100M
30 ELECTRICAL Resistance ( 2 wire ) Resistance Box by 0.58 % to 1.82 %
Ohm
EQUIPMENT Direct Method.
(Source)
ELECTRO-
TECHNICAL- Using Decade
31 ELECTRICAL Resistance (2 wire) Resistance Box by 1 Ohm to 100K Ohm 1.36 % to 0.12 %
EQUIPMENT Direct Method
(Source)
ELECTRO-
TECHNICAL- Using Decade
Resistance 100K Ohm to 10M
32 ELECTRICAL Resistance Box by 0.12 % to 0.58 %
(Source)(2 wire) Ohm
EQUIPMENT Direct Method
(Source)
ELECTRO-
TECHNICAL- Using 6½ Digital
50 Hz @1V to 1K Hz
33 TIME & Frequency Multimeter by Direct 0.12%
@1V
FREQUENCY Method.
(Measure)

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 22 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

ELECTRO-
TECHNICAL- Using Digital Time
34 TIME & Timer/Stop Watch Totaliser by Direct 3 s to 3600 s 0.37 s to 4.2 s
FREQUENCY Method
(Measure)
Using Digital
MECHANICAL-
Tachometer as per 100 rpm to 4000
35 ACCELERATION Centrifuge 17rpm
Sanas TR 45-02 rpm
AND SPEED
Guidelines.
Using Test Gauge
MECHANICAL-
(Hydraulic Pressure) and Comparator
PRESSURE
36 Dial and Digital (Media Oil) by >35 bar to 350 bar 0.86bar
INDICATING
Pressure Gauges Comparison Method
DEVICES
as per DKD-R6-1
Using Test Gauge
MECHANICAL-
(Hydraulic Pressure) and Comparator
PRESSURE
37 Dial and Digital (Media Oil) by >350 bar to 600 bar 1.13bar
INDICATING
Pressure Gauges Comparison Method
DEVICES
as per DKD-R6-1
Using Test Gauge
MECHANICAL-
(Hydraulic Pressure) and Comparator
PRESSURE
38 Dial and Digital (Media Oil) by 0 to 35 bar 0.2bar
INDICATING
Pressure Gauges Comparison Method
DEVICES
as per DKD-R6-1
Using Test Gauge
MECHANICAL-
and Vacuum pump
PRESSURE Dial and Digital
39 by Comparison -0.96 bar to 0 0.01bar
INDICATING Vacuum Gauges
Method as per DKD-
DEVICES
R6-1

This is annexure to 'Certificate of Accreditation' and does not require any signature.
BRILLIANT INDUSTRIES, 103, LULLA'S ARCADE NO. 2, KALYAN, MAHARASHTRA,
Laboratory Name :
INDIA
Accreditation Standard ISO/IEC 17025:2017
Certificate Number CC-3272 Page No 23 of 23
Validity 25/08/2021 to 24/08/2023 Last Amended on 12/11/2022

Measurand or Reference
Measurement range and
Material/Type of instrument * Calibration and
Calibration or Measurement additional parameters
S.No Discipline / Group or material to be calibrated Measurement
Method or procedure where applicable(Range
or measured / Quantity Capability(CMC)(±)
and Frequency)
Measured /Instrument

Using RTD (4 wire),


RTD, Thermocouple
THERMAL- 6½ DMM, Dry Block
40 with or without -10 °C to 400 °C 0.73°C
TEMPERATURE Furance by
Indicator/Controller
Comparision Method
Using 'R' Type T/C,
Thermocouple
THERMAL- 6½ DMM, Dry Block
41 Indicator/Controller 400 °C to 1200 °C 4°C
TEMPERATURE Furnace by
with Sensor
Comparison Method
* CMCs represent expanded uncertainties expressed at approximately the 95% level of confidence, using a coverage factor of k = 2.