Laboratory Ganga Calibration Services Private Limited, C-14, Co-Operative

Industrial Estate, Balanagar, Hyderabad, Telangana


Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC- 2127 (in lieu of C-1221) Page 1 of 3
Validity 01.06.2017 to 31.05.2019 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

ELECTRO-TECHNICAL CALIBRATION

I MEASURE

1. AC Voltage # 40 Hz to 62.5Hz
10V to 480V 0.006 % Using Zera COM3003
0.5V to 10 V 0.06 % Reference with Source by
Comparison Method

2. AC Current # 40 Hz to 65Hz
1mA to 10mA 0.037 % Using Zera COM3003
10mA to 50mA 0.018 % Reference with Source by
0.05A to 100A 0.008 % Comparison Method.
100A to 120A 0.020 %

3. AC Power / Energy # 40 Hz to 60 Hz
(1 ϕ, 3 ϕ) 40 V to 320 V
Active 1mA to 50 mA
Cos Ø ± 0.01 to 1 0.4 mW/mVAr to 48 0.40% Using 3-Phase Reference
Reactive W/VAr Standard Zera COM3003
Sin Ø ± 0.01 to 1 with Source by
50 mA to 100 A Comparison Method
0.02 W/VAr to 96 0.007%
kW/kVAr

100 A to 120 A
40 W/VAr to 115.2 0.014%
kW/kVAr

AC Power / Energy # 40 Hz to 60 Hz
(1 ϕ, 3 ϕ) 40 V to 320 V
Apparent 1mA to 50mA
0.4 mVA to 48 VA 0.40% Using 3-Phase Reference
Standard Zera COM 3003
50mA to 100A with Source by
0.02VA to 96 kVA 0.007% Comparison Method

Sangeeta Kunwar Avijit Das


Convenor Program Director
Laboratory Ganga Calibration Services Private Limited, C-14, Co-Operative
Industrial Estate, Balanagar, Hyderabad, Telangana
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC- 2127 (in lieu of C-1221) Page 2 of 3
Validity 01.06.2017 to 31.05.2019 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

100A to 120A
40VA to 115.2kVA 0.014%

4. Current Transformer# 50 Hz
5 A to 3200 A RE 0.02 % Using Standard Current
(Primary) PE- 1.05 Minutes Transformer & omparator
with Source Primary
3200A to 8000 A RE 0.03 % Injection by Comparison
(Primary) PE- 2.5 minutes Method
1 Amp, 5 Amp
(Secondary)

5. Potential Transformer # 50 Hz
Capacitor 6.6 kV to 33 kV RE 0.059 % Using Standard Potential
Voltage 6.6 kV/√3 to 33 kV/√3 PE- 3.05 minutes Transformer & Comparator
Transformer (Primary) with Source Primary
Voltage 33 kV/√3 to 132 kV/√3 RE 0.059 % Injection by Comparison
Divider (Primary) PE- 1.35 minutes Method
(Ratio Error
Phase Error) 63.5/110V
(Secondary)

6. High Voltage # 50 Hz
0.5 kv to 90 kV 0.09% to 0.30% Using Standard Potential
Transformer & Digital
Multimeter by Direct
Method

7. Power Factor # (+)1 PF to (-)1PF 0.0005 PF Using 3-Phase Reference


0 to 360° Standard COM3003 with
Source by Comparison
Method
8. Frequency # 40 Hz to 65 Hz 0.02 % Using 3-Phase Reference
Standard COM3003 with
Source by Comparison
Method

Sangeeta Kunwar Avijit Das


Convenor Program Director
Laboratory Ganga Calibration Services Private Limited, C-14, Co-Operative
Industrial Estate, Balanagar, Hyderabad, Telangana
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC- 2127 (in lieu of C-1221) Page 3 of 3
Validity 01.06.2017 to 31.05.2019 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

9. Harmonics# 1st to 39th Order of 0.4% Using 3-Phase Reference


Fundamental Standard COM3003 with
Frequency Source by Comparison
Method

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the
prevailing actual environmental conditions and master equipment used.

Sangeeta Kunwar Avijit Das


Convenor Program Director