Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

A Comparitive Study of Esr by Modified Westergren Method and Vesmatic Cube 30 TM

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

wjpmr, 2017,3(9), 343-344 SJIF Impact Factor: 4.

103
Research Article
WORLD JOURNAL OF PHARMACEUTICAL
Mahato et al. World Journal
AND MEDICAL RESEARCH of Pharmaceutical and Medical ISSN 2455-3301
Research
www.wjpmr.com WJPMR

A COMPARITIVE STUDY OF ESR BY MODIFIED WESTERGREN METHOD AND


VESMATIC CUBE 30 TM

Dr. Malabika Mahato*1 and Dr. P. Karkuzhali2


1
Post Graduate, Department of Pathology, Sree Balaji Medical College, Chennai, Tamilnadu.
2
Professor & HOD Department of Pathology, Sree Balaji Medical College, Chennai, Tamilnadu.

*Corresponding Author: Dr. Malabika Mahato


Post Graduate, Department of Pathology, Sree Balaji Medical College, Chennai, Tamilnadu.

Article Received on 16/08/2017 Article Revised on 07/09/2017 Article Accepted on 28/09/2017

ABSTRACT
Aims: The aim of study is to compare the ESR values by VesMATIC cube30™ against the modified Westergren
method. Materials and Methods: This is a cross-sectional study conducted at pathology laboratory in SBMCH,
Chrompet. The samples taken were as per the recommendations charted out by International Council for
Standardization in Hematology (ICSH) for comparing automated and modified Westergren method. Results and
Conclusions: The analysis revealed a low degree of agreement between the manual and automated method. Over
all whole ESR data analysis shows, mean difference 1.2±3.032 (95% limits agreement,-4.742 to 7.142) for 1 hour.
The fully automated system VesMATIC cube30™ for ESR measurement and manual ESR readings are very close
and also satisfying the 95% agreement limit. Hence, both methods can be used, but automated erythorocyte
sedimentation rate analyzer improve work flow, turnover time and laboratory safety by minimizing contact with
blood samples. So it is preferable to use automated method.

KEYWORDS: erythrocyte sedimentation rate, VesMATIC cube30™, Westergren method.

INTRODUCTION complete blood counts can also be used for ESR analysis
by this automated machine.
Erythrocyte sedimentation rate (ESR) is used as a
parameter for prognostication and even diagnosis of
MATERIALS AND METHODS
certain clinical conditions. It is still widely used in
clinical practice as an indicator of inflammation, It was a cross-sectional study done on routine hemogram
infection, trauma, or malignant disease. Apart from samples over a period of one week at SBMCH. A venous
complete blood counts (CBC) and peripheral smear, ESR blood sample was obtained from 100 random patients
is often preferred by clinicians in the requisition forms in who had a requisition form, marked for ESR; arriving at
almost all laboratories. The most satisfactory method of the sample collection centre. All the samples were
performing the test was introduced by Westergren in evaluated using the 2 methods in our clinical laboratory
1921. The original method recommended by as per the methodology for each of the individual
International Council for Standardization in Hematology methods. All the ESR tests were carried out within 3
(ICSH) is based on that of Fahraeus and Westergren. hours from the time of collection.
Subsequently, modifications of this reference method
were made and ICSH guidelines now allow for the use of Inclusion Criteria
alternative ESR techniques provided that comparability Patients from both sexes and all age groups were
with the Westergren method is achieved. included in the study.

Over the last few years, newer and safer methods have Exclusion Criteria
evolved to determine ESR accurately without added Blood collected by venipuncture taking more than 30
risks. The VesMatic cube30™ analyzer is a new seconds and with excessive venous stasis were excluded
automated instrument for measuring ESR. The advantage from the study. Blood samples which were not in proper
conferred by this automated method is that it can proportions to the anticoagulant, strongly lipemic,
generate the ESR readings in 30 minutes of 30 patients hyperbilirubinemic and hemolyzed samples were also
with all the temperature corrections at 18°C. excluded.
Additionally, the same EDTA sample collected for

www.wjpmr.com 343
Mahato et al. World Journal of Pharmaceutical and Medical Research

Samples Bland &Altman analysis of whole samples ESR data


Under all aseptic precautions, samples were collected values {95% limit of agreement [Mean+/-1.96 x SD]-
from the antecubital vein using a 10-ml syringe with 24G (-4.472 to 7.142)}
needle. Four milliliter of blood sample was drawn in the
two special 2-ml EDTA vacutainers containing 1.5 The results obtained with the reference method were
mg/ml of EDTA and mixed immediately five times. plotted against the difference between the reference and
the automated method for 1 hour values. The mean
Conventional Westergren Method difference between the two methods and 95% limits of
In this method, a disposable, plastic tube with a bore size agreement at 1 hour for whole data was found to be
of 2.55 mm and a length of 230 mm, vertically aligned, 1.2±3.032 (95% limits of agreement,–4.742 to 7.142).
open at both ends was used. The pipette was filled with
K3 EDTA anticoagulated venous blood to a height of at DISCUSSION
least 200 mm. The sedimentation occurring at 60 minutes The results obtained with the Vesmatic cube30™
from the beginning of the test was noted in mm/hour analyzer were compared with the modified Westergren
equivalent to the Westergren ESR. method using the agreement analysis of Bland and
Altman. Agreement analysis is a more sensitive method
VesMatic cube30™ instrument than the correlation coefficient for comparison between
The vesMatic cube 30 is a benchtop analyzer designed & the two methods. Over all whole ESR data analysis
programmed to determine the sedimentation rate one shows, mean difference 1.2±3.02 (95% limits agreement,
hour on primary EDTA tubes. The loading samples per ˗4.742 to 7.142) for 1 hour. The use of Bland and Altman
set of tubes 1 to 30 is carried out manually in a specific analysis for evaluating the agreement between the two
rack. The stirred vesMatic cube samples and a first methods, not only assesses the mean of the difference (d)
optical measurement of the level of blood through the between the two methods (i.e., bias) but also the limits of
labels is carried out. 15min latter a second measurement agreement by calculating the standard deviation of the
time evaluates the level of sedimentation. 3 matrices differences (d ± 1.96SD). Only when the difference (d ±
calculations interprets 2 readings is then extrapolated to 1.96SD) does not affect the clinical interpretation, the
1 hr & finally adjusted to the temperatures at which the two methods can be used interchangeably.
analysis was performed. Then the results get printed out.
CONCLUSIONS
Statistical Analysis
The fully automated system Vesmatic cube30™ for ESR
Evaluation of method was done as described by Bland
measurement and modified westergren ESR readings are
and Altman. Modified Westergren method was
very close and also satisfying the 95% agreement limit.
considered as the reference method. Therefore mean
So both methods can be used. But automated
value for ESR with both the methods is plotted against
erythorocyte sedimentation rate analyzer improve work
the difference between the Westergren and the
flow, turnover time and laboratory safety by minimizing
VesMATIC cube30™. The 95% limits of agreement
contact with blood samples. Hence, it is preferable to use
were calculated as d ± 1.96 SD where d = mean
automated analyzation.
difference between the two measurements; and SD =
standard deviation of differences.
REFERENCES
RESULTS 1. Westergren A. Studies of the suspension stability of
the blood in pulmonary tuberculosis. Acta Med
Sample statistics
Scand, 1921; 54: 247-282.
95% Limit of 2. Fahraeus R. The suspension stability of blood. Acta
No. of
BIAS SD Agreement Med Scand, 1921; 55: 1-228.
Sample
From To 3. International council for Standardization in
100 1.2 3.032 -4.742 7.142 hematology. ICSH recommendation for
measurement of ESR. J ClinPathol, 1993; 46: 198-
203.
4. Arikan S, Akalin N. Comparison of the erythrocyte
sedimentation rate measured by the Micro Test 1
sedimentation analyzer and the conventional
Westergren’s method. Ann Saudi Med, 2007; 27:
362-5.
5. Plebani M, Piva E. Erythrocyte sedimentation rate:
Use of fresh blood for quality control. Am J Clin
Pathol, 2002; 117: 621- 6
6. User manual. Monitor 100. Forli, Italy: Electra Lab
s.r., 2005; 1.

www.wjpmr.com 344

You might also like