E&H Transmitter
E&H Transmitter
E&H Transmitter
BA271P/00/en/11.05
71009589
valid from
Software version 02.01
Hardware version 02.00
Overview documentation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
Overview documentation
Cerabar S 4...20 mA HART Technical Information TI383P Technical data – The documentation is located on the ToF Tool
CD. The CD is enclosed with every device
ordered with the "HistoROM/M-DAT"
option. The "HistoROM/M-DAT" option is
selected in the order code via feature 100
Additional option 1" or 110 "Additional
option 2", version "N".
– The documentation is also available via the
Internet. → See: www.endress.com →
Download
Operating Instructions BA274P – Examples of configuration for – The documentation is located on the ToF Tool
pressure and level measurement CD. The CD is enclosed with every device
– Description of parameters ordered with the "HistoROM/M-DAT"
– ’Trouble-shooting option. The "HistoROM/M-DAT" option is
– Appendix: Illustration of menu selected in the order code via feature 100
Additional option 1" or 110 "Additional
option 2", version "N".
– The documentation is also available via the
Internet. → See: www.endress.com →
Download
Brief Operating Instructions KA218P – Wiring – The documentation is supplied with the
– Operation without on-site device. See cover of the terminal
display compartment.
– Description of Quick Setup menus
– Operation HistoROM®/M-DAT
Functional Safety Manual SD190P – Safety function with Cerabar S – The documentation is supplied with the
– Behaviour in operation and failure devices showing version "E" in feature 100
– Commissioning and iterative tests "Additional options 1" or in feature 110
– Settings "Additional options 2".
– Technical safety characteristic → See also Technical Information TI383P,
quantities chapter "Ordering information".
– Management Summary
2 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Table of contents
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 4 8 Trouble-shooting . . . . . . . . . . . . . . . . . 46
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 8.1 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
1.2 Installation, commissioning and operation . . . . . . . . 4 8.2 Response of outputs to errors . . . . . . . . . . . . . . . . . 53
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4 8.3 Confirming messages . . . . . . . . . . . . . . . . . . . . . . . 54
1.4 Notes on safety conventions and icons . . . . . . . . . . . 5 8.4 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.5 Repair of Ex-certified devices . . . . . . . . . . . . . . . . . 55
2 Identification . . . . . . . . . . . . . . . . . . . . 6 8.6 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8.7 Returning the device . . . . . . . . . . . . . . . . . . . . . . . 65
2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6 8.8 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 8.9 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
2.3 CE mark, declaration of conformity . . . . . . . . . . . . . 7
2.4 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . 7
9 Technical data . . . . . . . . . . . . . . . . . . . 67
3 Installation . . . . . . . . . . . . . . . . . . . . . . 8
10 Appendix . . . . . . . . . . . . . . . . . . . . . . . 67
3.1 Incoming acceptance and storage . . . . . . . . . . . . . . . 8
3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 8 10.1 Operating menu for on-site display, ToF Tool,
3.3 Installation instructions . . . . . . . . . . . . . . . . . . . . . . 8 FieldCare and HART handheld terminal . . . . . . . . . 67
3.4 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 15 10.2 Operating matrix HART Commuwin II . . . . . . . . . . 73
10.3 Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 15
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4.1 Connecting the device . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Connecting the measuring unit . . . . . . . . . . . . . . . 17
4.3 Potential matching . . . . . . . . . . . . . . . . . . . . . . . . 20
4.4 Overvoltage protection (optional) . . . . . . . . . . . . . . 20
4.5 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 20
5 Operation . . . . . . . . . . . . . . . . . . . . . . 21
5.1 On-site display (optional) . . . . . . . . . . . . . . . . . . . . 21
5.2 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . 22
5.3 On-site operation –
on-site display not connected . . . . . . . . . . . . . . . . . 26
5.4 On-site operation –
on-site display connected . . . . . . . . . . . . . . . . . . . . 28
5.5 HistoROM®/M-DAT (optional) . . . . . . . . . . . . . . 30
5.6 Operation via HART handheld terminal . . . . . . . . . 33
5.7 ToF Tool operating program . . . . . . . . . . . . . . . . . 34
5.8 FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.9 Commuwin II operating program . . . . . . . . . . . . . . 35
5.10 Locking/unlocking operation . . . . . . . . . . . . . . . . . 36
5.11 Factory setting (reset) . . . . . . . . . . . . . . . . . . . . . . 37
6 Commissioning. . . . . . . . . . . . . . . . . . 39
6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.2 Selecting language and measuring mode . . . . . . . . 39
6.3 Position adjustment . . . . . . . . . . . . . . . . . . . . . . . . 40
6.4 Pressure measurement . . . . . . . . . . . . . . . . . . . . . . 41
6.5 Level measurement . . . . . . . . . . . . . . . . . . . . . . . . 43
7 Maintenance. . . . . . . . . . . . . . . . . . . . 46
7.1 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 46
Endress+Hauser 3
Safety instructions Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
1 Safety instructions
4 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Safety instructions
Symbol Meaning
#
Warning!
A warning highlights actions or procedures which, if not performed correctly, will lead to
personal injury, a safety hazard or destruction of the instrument.
"
Caution!
Caution highlights actions or procedures which, if not performed correctly, may lead to
personal injury or incorrect functioning of the instrument.
!
Note!
A note highlights actions or procedures which, if not performed correctly, may indirectly
affect operation or may lead to an instrument response which is not planned.
0
Device certified for use in explosion hazardous area
If the device has this symbol embossed on its nameplate, it can be installed in an explosion
hazardous area or a non-explosion hazardous area, according to the approval.
% Direct voltage
A terminal to which or from which a direct current or voltage may be applied or supplied.
&
Alternating voltage
A terminal to which or from which an alternating (sine-wave) current or voltage may be
applied or supplied.
)
Grounded terminal
A grounded terminal, which as far as the operator is concerned, is already grounded by
means of an earth grounding system.
*
Protective grounding (earth) terminal
A terminal which must be connected to earth ground prior to making any other connection
to the equipment.
+
Equipotential connection (earth bonding)
A connection made to the plant grounding system which may be of type e.g. neutral star or
equipotential line according to national or company practice.
Endress+Hauser 5
Identification Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
2 Identification
2.1.1 Nameplate
Cerabar S
Order Code: 1
3
D-79689 Maulburg
Ser.-No.: 2
Made in Germany,
5
MWP 4 Span 6
P 7
8 U= 9 12
Mat. 10
11
13 14 14
15
P01-PMx7xxxx-18-xx-xx-xx-001
Devices for use in hazardous areas are fitted with an additional nameplate.
1
2
4 5 Dat.: 6
P01-xMD7xxxx-18-xx-xx-xx-002
6 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Identification
! Note!
• The MWP (maximum working pressure) is specified on the nameplate. This value refers to a
reference temperature of 20°C (68°F) or 100°F for ANSI flanges.
• The pressure values permitted at higher temperatures can be found in the following standards:
– EN 1092-1: 2001 Tab. 18 1
– ASME B 16.5a – 1998 Tab. 2-2.2 F316
– ASME B 16.5a – 1998 Tab. 2.3.8 N10276
– JIS B 2220
• The test pressure corresponds to the over pressure limit (OPL) of the device = MWP x 1.5.
• The Pressure Equipment Directive (EC Directive 97/23/EC) uses the abbreviation "PS". The
abbreviation "PS" corresponds to the MWP (maximum working pressure) of the measuring
device.
1) With regard to its stability-temperature property, the material 1.4435 is identical to 1.4404 which is grouped under
13EO in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical.
Endress+Hauser 7
Installation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
3 Installation
3.1.2 Storage
The device must be stored in a dry, clean area and protected against damage from impact
(EN 837-2).
Storage temperature range:
• –40...+100°C (–40...+212°F)
• On-site display: –40...+85°C (–40...+185°F)
3.2.1 Dimensions
→ For dimensions, please refer to the Technical Information for Cerabar S TI383P, "Mechanical
construction" section. See page 2, "Overview documentation".
! Note!
• Cerabar S without diaphragm seal are mounted as per the norms for a manometer
(DIN EN 839-2). We recommend the use of shut-off devices an siphons. The orientation depends
on the measuring application.
• Do not clean or touch diaphragm seals with hard or pointed objects.
8 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Installation
P01-PMx7xxxx-11-xx-xx-xx-001
• Mount Cerabar S with shut-off device above the tapping point so that the condensate can flow
into the process.
➀ ➃
➁ ➁
➂ ➀
P01-PMx7xxxx-11-xx-xx-xx-002
Endress+Hauser 9
Installation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
P01-PMx7xxxx-11-xx-xx-xx-003
• Mount Cerabar S with shut-off device below or at the same level as the tapping point.
Level measurement
P01-PMP75xxx-11-xx-xx-xx-000
PVDF adapter
! Note!
For instruments with PVDF adapter, a maximum torque of 7 Nm is permitted. The thread
connection may become loose at high temperatures and pressures. This means that the integrity of
the thread must be checked regularly and may need to be tightened using the torque given above.
Teflon tape is recommended for sealing with the 1/2 NPT thread.
10 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Installation
! Note!
• The Cerabar S with diaphragm seal is screwed in, flanged or clamped, depending on the type of
diaphragm seal.
• The diaphragm seal and the pressure sensor together form a closed an calibrated system which is
filled with filling fluid through a hole in the upper part. This hole is sealed and not to be opened.
• Do not clean or touch diaphragm seals with hard or pointed objects.
• Do not remove membrane protection until shortly before installation.
• When using a mounting bracket, sufficient strain relief must be ensured for the capillaries in order
to prevent the capillary bending down (bending radius ≥ 100 mm).
• Please note that the hydrostatic pressure of the liquid columns in the capillaries can cause zero
point shift. The zero point shift can be corrected. → See also page 40, section 6.3 "Position
adjustment".
• Please note the application limits of the diaphragm seal filling oil as detailed in the Technical
Information for Cerabar S TI383P, Section "Planning instructions for diaphragm seal systems". →
See also page 2, "Overview documentation".
In order to obtain more precise measurement results and to avoid a defect in the device, mount the
capillaries as follows:
• vibration-free (in order to avoid additional pressure fluctuations)
• not in the vicinity of heating or cooling lines
• insulate in the event of colder or warmer ambient conditions
• with a bending radius of ≥100 mm.
Vacuum application
For applications under vacuum, Endress+Hauser recommends mounting the pressure transmitter
underneath the diaphragm seal. A vacuum load of the diaphragm seal caused by the presence of
filling oil in the capillaries is hereby prevented.
When the pressure transmitter is mounted above the diaphragm seal, the maximum height
difference H1 in accordance with the illustration below on the left must not be exceeded. The
maximum height difference is dependent on the density of the filling oil and the smallest ever
pressure that is permitted to occur at the diaphragm seal (empty container), see illustration below,
on the right.
FIELD TERMINALS
12.0
Vegetable oil
10.0
Height difference H1 [m]
Silicone oil
8.0
6.0
High temperature
H1 oil
4.0
Inert oil
2.0
0.0
50 100 200 300 400 500 600 700 800 900 1000
Pressure at diaphragm seal [mbarabs]
P01-PMx7xxxx-11-xx-xx-xx-008 P01-PMx7xxxx-05-xx-xx-xx-011
Fig. 7: Installation above the diaphragm Fig. 8: Diagram of maximum installation height above the
seal diaphragm seal for vacuum applications dependent on
the pressure at the diaphragm seal
Endress+Hauser 11
Installation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
max. 115
P01-PMx7xxxx-11-xx-xx-xx-005
Endress+Hauser recommends the use of temperature separators in the event of constant extreme
fluid temperatures which lead to the maximum permissible electronics temperature of +85°C
(+185°F) being exceeded.
The additional installation height also brings about a zero point shift of approx. 21 mbar due to the
hydrostatic columns in the temperature separator. You can correct this zero point shift. → See also
page 40, section 6.3 "Position adjustment".
➀ ➁
P01-FMD7xxxx-11-xx-xx-xx-002
# Warning!
The seal is not allowed to press on the diaphragm as this could affect the measurement result.
12 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Installation
λ ≤ 0.04 W
TA m•K
TP
P01-PMx7xxxx-11-xx-xx-xx-010
Fig. 10: Maximum insulation height, here e.g. PMC71 with flange
Process temperature (TP) ≤ 150°C (302°F) max. 350°C (662°F), depending on the
diaphragm seal filling oil used (→ see
Technical Information TI383P Cerabar S)
70
P01-PMx7xxxx-06-xx-xx-xx-001
Endress+Hauser 13
Installation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
➀ ➁ ➂
A1
5
ø2.5
ø7.95
P01-PMP71xxx-11-xx-xx-xx-000
Welding recommendation
For the "U1 Prepared for diaphragm seal mount" version in feature 70 "Process connection;
Material" in the order code of up to including 40 bar-sensors, Endress+Hauser recommends welding
on the diaphragm seal as follows: the total welding depth of the fillet weld is 1 mm at an external
diameter of 16 mm. Welding is performed according to the WIG method.
max. 380°
P01-PMx7xxxx-17-xx-xx-xx-000
14 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Wiring
4 Wiring
• The supply voltage must match the supply voltage on the nameplate. (→ See also page 6, section
2.1.1 Nameplate.)
• Switch off the supply voltage before connecting the device.
• Remove housing cover of the terminal compartment.
• Guide cable through the gland. Preferably use twisted, screened two-wire cable.
• Connect device in accordance with the following diagram.
• Screw down housing cover.
• Switch on supply voltage.
➀ ➅ 10.5 V DC
➆ 11.5 V DC
4…20 mA
Test
Test
➂ ➄
➇
4... 20mA Test
➃
P01-xMx7xxxx-04-xx-xx-xx-001
Endress+Hauser 15
Wiring Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
Cerabar S
+
+ –
7
6 8
Han7D 1 –
5 2
4 3
+ –
P01-PMx7xxxx-04-xx-xx-xx-001
Fig. 14: Left: electrical connection for devices with Harting plug Han7D
Right: view of the plug at the device
Cerabar S
+ –
BK – +
M12
+ –
P01-PMx7xxxx-04-xx-xx-xx-000
Fig. 15: Left: electrical connection for devices with M12 plug
Right: view of the plug at the device
Cerabar S
–
+ –
7/8" +
+ –
P01-PMx7xxxx-04-xx-xx-xx-003
Fig. 16: Left: electrical connection for devices with 7/8" plug
Right: view of the plug at the device
16 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Wiring
! Note!
• All explosion protection data are given in separate documentation which is available upon
request. The Ex documentation is supplied as standard with all devices approved for use in
explosion hazardous areas.
• When using the measuring device in hazardous areas, installation must comply with the
corresponding national standards and regulations and the Safety Instructions (XAs) or Installation
or Control Drawings (ZDs).
Electronic version Jumper for 4...20 mA test signal in Jumper for 4...20 mA test signal in
"Test" position (Delivery status) "Non-Test" position
Endress+Hauser 17
Wiring Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
4.2.3 Load
U – 10.5 V U – 11.5 V
➀ RLmax RLmax ≤
23 mA ➁ RLmax RLmax ≤
23 mA
[Ω] [Ω]
Test 1500 Test
1456
1282 1239
847 804
➃ ➃
413
➂ 369
➂
10.5 20 30 40 45 U 11.5 20 30 40 45 U
[V] [V]
U – 11 V U – 12 V
RLmax RLmax ≤ RLmax RLmax ≤
[Ω] 23 mA [Ω] 23 mA
1478 1434
1260 1217
826 782
391
➄ ➄
347
11 20 30 40 45 U 12 20 30 40 45 U
[V] [V]
P01-PMx7xxxx-05-xx-xx-xx-003
Fig. 17: Load diagram, observe the position of the jumper and the explosion protection. (→ See also page 17, "Taking
4...20 mA test signal" part.)
1 Jumper for the 4...20 mA test signal inserted in "Non-Test" position
2 Jumper for the 4...20 mA test signal inserted in "Test" position
3 Supply voltage 10,5 (11,5)...30 V DC for EEx ia, 1 D, 1/2 GD, FM IS and CSA IS
4 Supply voltage10,5 (11,5)...45 V DC for devices for non-hazardous areas, 1/3 D, EEx d, EEx nA, FM XP, FM DIP,
FM NI, CSA XP and CSA Dust-Ex
5 Supply voltage 11 (12)...45 V DC for PMC71, EEx d[ia], FM XP, CSA XP
RLmax Maximum load resistance
U Supply voltage
! Note!
When operating via a handheld terminal or via PC with an operating program, a minimum
communication resistance of 250 Ω must exist within the loop.
18 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Wiring
➀ min. 250 Ω
Test
4…20 mA
➂ DELTABAR: * * * * * * * *
ONLINE
1 QUICK SETUP
2 OPERATING MENU
3 PV
dsdmdm
df das.
asdas fa
asas la.
352 mbar
➁ DELTABAR: * * * * * * * *
ONLINE
1 QUICK SETUP
2 OPERATING MENU
3 PV
dsdmdm
df das.
asdas fa
asas la.
352 mbar
DELTABAR: * * * * * * * *
ONLINE
1 QUICK SETUP
2 OPERATING MENU
3 PV
dsdmdm
df das.
asdas fa
asas la.
352 mbar
4 SV 0 °C 4 SV 0 °C 4 SV 0 °C
6 6 6
9 Page
Up 9 Page
Up 9 Page
Up
EX EX
Bksp Delete Bksp Delete Bksp Delete
P01-xMD7xxxx-04-xx-xx-xx-002
Fig. 18: Connecting an HART handheld terminal, e.g. Field Communicator DXR375
1 Necessary communication resistor ≥ 250 Ω
2 HART handheld terminal
3 HART handheld terminal, directly connected to the device even in the Ex i-area
# Warning!
• In the case of Ex d type of protection, do not connect the handheld terminal in the hazardous area.
• Do not replace the battery of the handheld terminal in the hazardous area.
• For devices with FM or CSA certificates, establish electrical connection as per Installation or
Control Drawing (ZD...) supplied.
Endress+Hauser 19
Wiring Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
min. 250 Ω
➂
Test
4…20 mA
EX EX
P01-xMD7xxxx-04-xx-xx-xx-003
Fig. 19: Connecting PC with operating programs ToF Tool, FieldCare or Commuwin II via Commubox FXA 191
1 Computer with operating program ToF Tool, FieldCare or Commuwin II
2 Commubox FXA191
3 Necessary communication resistor ≥ 250 Ω
# Warning!
Devices with integrated overvoltage protection must be earthed.
20 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
5 Operation
Feature 20 "Output; operation" in the order code provides you with information on the operating
options available to you.
A 4...20 mA HART; external operation, LCD Via on-site display and 1 key on the exterior of the device
B 4...20 mA HART; internal operation, LCD Via on-site display and 1 key on the inside of the device
C 4...20 mA; internal operation Without on-site display, 1 key on the inside of the device
Header line
Main line
Unit
Information
line
– + E Symbol
Bargraph
Editing modes
Operating keys
Selection
options
Value that
can be edited
Endress+Hauser 21
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
The following table illustrates the symbols that can appear on the on-site display. Four symbols can
occur at one time.
Symbol Meaning
Alarm symbol
– Symbol flashing: warning, device continues measuring.
– Symbol permanently lit: error, device does not continue measuring.
Note: The alarm symbol may overlie the tendency symbol.
Lock symbol
The operation of the device is locked. Unlock device, → see section 5.10.
Communication symbol
Data transfer via communication
Note: The alarm symbol may overlie the communication symbol.
➃➄
τ ➅
on
➀
➀
off
➁
Damping [τ]
Display
on Sensor
➂
1 2
off
PC
HistoROM
P01-PMx7xxxx-19-xx-xx-xx-009 P01-xxxxxxxx-19-xx-xx-xx-074
Fig. 20: Operating keys, external Fig. 21: Operating keys, internal
1 Operating keys on the exterior of the device under 1 Operating keys
the protective flap 2 Slot for optional display
3 Slot for optional HistoROM®/M-DAT
4 DIP-switch for locking/unlocking measured-value-
relevant parameters
6 DIP-switch for damping on/off
7 Green LED to indicate value being accepted
22 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
! Note!
• The device is configured for the pressure measuring mode as standard. You can switch
measuring modes by means of the MEASURING MODE parameter. → See page 39,
section 6.2 "Selecting language and measuring mode".
• Press the key for at least 3 seconds to accept the pressure value applied.
! Note!
• MEASURING MODE "Level":
The factory settings for the LEVEL MODE, CALIBRATION MODE, LIN. MEASURAND,
EMPTY CALIB. and FULL CALIB. parameters are as follows:
– LEVEL MODE = Linear
– CALIBRATION MODE = wet
– LIN. MEASURAND = %
– EMPTY CALIB. = 0 %
S – FULL CALIB. = 100 %.
These parameters can only be modified by means of the on-site display or remote
operation such as the ToF Tool.
MEASURING MODE "Level", CALIBRATION MODE "wet":
• FÜLLSTANDSSTYP "Linear":
– The pressure applied is saved as the lower pressure value (EMPTY PRESSURE1) and
assigned to the lower level value (EMPTY CALIB.1). The lower level and current
values (SET LRV - level1) remain unchanged. These values can only be modified by
means of the on-site display or remote operation such as the ToF Tool.
→ See also page 44, section 6.5.2 "Quick Setup menu for Level measuring mode" and
Operating Instructions BA274P, parameter descriptions LEVEL MODE,
CALIBRATION MODE, EMPTY CALIB., EMPTY PRESSURE and SET LRV – Level.
• LEVEL MODE "Pressure Linearized" or "Height Linearized":
– The key does not have any function.
MEASURING MODE "Level", CALIBRATION MODE "dry":
• The key does not have any function.
Endress+Hauser 23
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
! Note!
• The device is configured for the pressure measuring mode as standard. You can switch
measuring modes by means of the MEASURING MODE parameter. → See page 39,
section 6.2 "Selecting language and measuring mode".
• Press the key for at least 3 seconds to accept the pressure value applied.
MEASURING MODE "Pressure":
• The pressure applied is assigned to the upper current value (SET URV – pressure1).
! Note!
• MEASURING MODE "Level":
The factory settings for the LEVEL MODE, CALIBRATION MODE, LIN. MEASURAND,
EMPTY CALIB. and FULL CALIB. parameters are as follows:
– LEVEL MODE = Linear
– CALIBRATION MODE = wet
– LIN. MEASURAND = %
– EMPTY CALIB. = 0 %
O – FULL CALIB. = 100 %.
These parameters can only be modified by means of the on-site display or remote
operation such as the ToF Tool.
MEASURING MODE "Level", CALIBRATION MODE "wet":
• LEVEL MODE "Linear":
– The pressure applied is saved as the upper pressure value (FULL PRESSURE1) and
assigned to the upper level value (FULL CALIB.1). The upper level and current values
(SET URV - level1) remain unchanged. These values can only be modified by means of
the on-site display or remote operation such as the ToF Tool.
→ See also page 44, section 6.5.2 "Quick Setup menu for Level measuring mode" and
Operating Instructions BA274P, parameter descriptions LEVEL MODE,
CALIBRATION MODE, FULL CALIB., FULL PRESSURE, and SET URV – Level.
• LEVEL MODE "Pressure Linearized" or "Height Linearized":
– The key does not have any function.
MEASURING MODE "Level", CALIBRATION MODE "dry":
• The key does not have any function.
Reset all parameters. The reset via operating keys corresponds to the software reset code
O and S and F 7864. (Press key for at least 6s.)
O and F Copy the configuration data from the optional HistoROM®/M-DAT module to the device.
S and F Copy the configuration data from the device to the optional HistoROM®/M-DAT module.
1) Parameter name used for the on-site display or remote operation such as the ToF Tool.
24 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
F – Confirm entry
– Jump to the next item
Endress+Hauser 25
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
! Note!
• The operation must be unlocked. → See page 36, section 5.10 "Locking /unlocking operation".
• The device is configured for the Pressure measuring mode as standard. You can switch measuring
modes by means of the MEASURING MODE parameter. → See page 39, section 6.2 "Selecting
language and measuring mode".
• The pressure applied must be within the nominal pressure limits of the sensor. See information
on the nameplate.
Carry out position adjustment. Setting lower range value. Setting upper range value.
Pressure is present at device. Desired pressure for lower range Desired pressure for upper range
value is present at device. value is present at device.
↓ ↓ ↓
Press "E"-key for 3 s. Press "–"-key for 3 s. Press "+"-key for 3 s.
↓ ↓ ↓
Does the LED on the electronic insert Does the LED on the electronic insert Does the LED on the electronic insert
light up briefly? light up briefly? light up briefly?
↓ ↓ ↓ ↓ ↓ ↓
Applied pressure Applied pressure Applied pressure Applied pressure Applied pressure Applied pressure
for for position for for for for
position adjustment has lower range value lower range value upper range value upper range value
adjustment has not been has been has not been has been has not been
been accepted. accepted. accepted. accepted. accepted. accepted.
Observe the input Observe the input Observe the input
limits. limits. limits.
26 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
! Note!
If no on-site display is connected, the following functions are possible by means of the three keys
on the electronic insert or on the exterior of the device:
• Position adjustment (zero point correction)
• Set the lower and upper pressure value and assign to the lower and upper level value
• Device reset, → see also page 23, section 5.2.2 "Function of the operating elements", Table.
• The operation must be unlocked. → See page 36, section 5.10 "Locking /unlocking operation".
• The device is configured for the Pressure measuring mode as standard. You can switch measuring
modes by means of the MEASURING MODE parameter. → See page 39, section 6.2 "Selecting
language and measuring mode".
• The S and O keys do not have any function if the LEVEL MODE "Pressure linearized" or "Height
linearized" or the CALIBRATION MODE "dry" was selected.
• The factory settings for the LEVEL MODE, CALIBRATION MODE, LIN. MEASURAND, EMPTY
CALIB. and FULL CALIB. parameters are as follows:
– LEVEL MODE = Linear
– CALIBRATION MODE = wet
– LIN. MEASURAND = %
– EMPTY CALIB. = 0 %
– FULL CALIB. = 100 %.
These parameters can only be modified by means of the on-site display or remote operation such
as the ToF Tool.
• The pressure applied must be within the nominal pressure limits of the sensor. See information
on the nameplate.
• → See also page 44, section 6.5.2 "Quick Setup menu for Level measuring mode" and Operating
Instructions BA274P, parameter descriptions LEVEL MODE, CALIBRATION MODE, EMPTY
CALIB., FULL CALIB., EMPTY PRESSURE, FULL PRESSURE, SET LRV – Level and SET URV –
Level.
Carry out position adjustment. Setting lower pressure value. Setting upper pressure value.
Pressure is present at device. Desired pressure for lower pressure Desired pressure for upper pressure
value (EMPTY PRESSURE1) is present value (FULL PRESSURE1) is present
at device. at device.
↓ ↓ ↓
↓ ↓ ↓
Does the LED on the electronic insert Does the LED on the electronic insert Does the LED on the electronic insert
light up briefly? light up briefly? light up briefly?
↓ ↓ ↓ ↓ ↓ ↓
Applied pressure Applied pressure The pressure The pressure The pressure The pressure
for for position present was saved present was not present was saved present was not
position adjustment has as the lower saved as the as the upper saved as the
adjustment has not been pressure value lower pressure pressure value upper pressure
been accepted. accepted. (EMPTY value. Observe (FULL value. Observe
Observe the input PRESSURE1) and the input limits. PRESSURE1) and the input limits.
limits. assigned to the assigned to the
lower level value upper level value
(EMPTY (FULL CALIB.1).
CALIB.1).
1) Parameter name used for the on-site display or remote operation such as the ToF Tool.
Endress+Hauser 27
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
Measured value
GROUP SELECTION
➁
➀ SETTINGS DISPLAY
➂
➃ POS. INPUT VALUE
CALIB. OFFSET
P01-xxxxxxxx-19-xx-xx-xx-097
! Note!
The LANGUAGE and MEASURING MODE parameters are only displayed via the on-site display on
the 1st selection level. In the ToF Tool or HART handheld terminal, the LANGUAGE parameter is
displayed in the DISPLAY group and the MEASURING MODE parameter is displayed in the QUICK
SETUP menus or in the BASIC SETUP function group. → See also section 10.1 "Operating menu
for on-site display, ToF Tool, FieldCare and HART handheld terminal".
28 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
P01-xxxxxxxx-19-xx-xx-xx-017
P01-xxxxxxxx-19-xx-xx-xx-033
P01-xxxxxxxx-19-xx-xx-xx-034
P01-xxxxxxxx-19-xx-xx-xx-023
P01-xxxxxxxx-19-xx-xx-xx-027
P01-xxxxxxxx-19-xx-xx-xx-028
P01-xxxxxxxx-19-xx-xx-xx-029
P01-xxxxxxxx-19-xx-xx-xx-030
Endress+Hauser 29
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
Use "E" to save the new value and exit the editing mode.
→See next graphic.
P01-xxxxxxxx-19-xx-xx-xx-031
P01-xxxxxxxx-19-xx-xx-xx-032
P01-xxxxxxxx-19-xx-xx-xx-035
P01-xxxxxxxx-19-xx-xx-xx-036
Use "E" to assign the value (400 mbar) to the GET URV
parameter. The device confirms the calibration and jumps
back to the parameter, here GET URV (see next graphic).
P01-xxxxxxxx-19-xx-xx-xx-028
P01-xxxxxxxx-19-xx-xx-xx-035
# Warning!
Detach HistoROM®/M-DAT from the electronic insert or attach it to the insert in a deenergised
state only.
30 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
! Note!
• The HistoROM®/M-DAT module may be retrofitted at any time (Order No.: 52027785).
• The HistoROM data and the data in the device are analysed once a HistoROM®/M-DAT is
attached to the electronic insert and power is reestablished to the device. During the analysis, the
messages "W702, HistoROM data not consistent" and "W706, Configuration in HistoROM and
device not identical" can occur. For measures, see page 46, section 8.1 "Messages."
! Note!
When copying configuration data from one device into another, the following parameters are also
overwritten:
• DEVICE SERIAL No
• DEVICE DESIGN
• CUST. TAG NUMBER
• BUS ADDRESS
• all parameters in the PROCESS CONNECTION group.
These have to be reentered depending on the application. To reenter the DEVICE SERIAL No and
DEVICE DESIGN parameters, please contact Endress+Hauser Service. (→ See also
www.endress.com/worldwide.)
➁
on
τ
off 1 2
Damping [τ]
Display
on Sensor
1 2
off
PC
HistoROM ➀
P01-xxxxxxxx-19-xx-xx-xx-099
Endress+Hauser 31
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
! Note!
The operation must be unlocked. → See also page 36, section 5.9 "Locking/unlocking operation".
1. Disconnect device from supply voltage.
2. Attach the HistoROM®/M-DAT module to the electronic insert. Configuration data from
another device are stored in the HistoROM®/M-DAT.
3. Reestablish supply voltage to the device.
4. Press F and S-keys (for at least 3 seconds) until the LED on the electronic insert lights up.
5. Wait approx. 20 seconds. Configuration data are loaded from the device to the HistoROM®/
M-DAT. The device is restarted.
6. Before removing the HistoROM®/M-DAT again from the electronic insert, disconnect the
device from supply voltage.
! Note!
The operation must be unlocked. → See also page 36, section 5.9 "Locking/unlocking operation".
1. Disconnect device from supply voltage.
2. Attach the HistoROM®/M-DAT module to the electronic insert. Configuration data from
another device are stored in the HistoROM®/M-DAT.
3. Reestablish supply voltage to the device.
4. Using the HistoROM CONTROL parameter select the option "Device → HistoROM" as the
data transfer direction.
(Menu path: GROUP SELECTION → OPERATING MENU → OPERATION)
5. Wait approx. 20 seconds. Configuration data are loaded from the device to the HistoROM®/
M-DAT. The device is restarted.
6. Before removing the HistoROM®/M-DAT again from the electronic insert, disconnect the
device from supply voltage.
32 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
dsdmdm
df das.
asdas fa
asas la.
dsdmdm
df das.
asdas fa
asas la.
DELTABAR: * * * * * * * *
DELTABAR: * * * * * * * * ONLINE
ONLINE
1 QUICK SETUP 1 QUICK SETUP
Delete
➀
2 OPERATING MENU 2 OPERATING MENU
3 PV 352 mbar
4 SV 0 °C 3 PV 352 mbar
4 SV 0 °C
HELP SAVE
H dsdmdm
df das.
asdas fa
asas la.
DELTABAR: * * * * * * * *
6
9
HELP SAVE
Page OPERATING MENU
Up
➁
1 SETTINGS
Bksp Delete 2 DISPLAY Delete
3 OUTPUT
Page 4 TRANSMITTER INFO
On Bksp 5 PROCESS INFO
➂
#%& ABC DEF
1 2 3
Copy
GHI
Paste
JKL
Hot Key
MNO
H dsdmdm
df das.
asdas fa
asas la.
P00-xMD7xxxx-02-xx-xx-xx-002
Fig. 24: HART handheld terminal, here e.g. Field Communicator DXR375 and menu guidance
1 LC display with menu text
2 Keys for menu selection
3 Keys for parameter entry
! Note!
• → See also page 19, section 4.2.5 "Connecting HART handheld terminal".
• For further information, please refer to the Operating Instructions for the handheld terminal. The
Operating Instructions are supplied with the handheld terminal.
Endress+Hauser 33
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
P01-PMx7xxxxx-19-xx-xx-en-000
Fig. 25: ToF Tool operating program, the configuration is performed via a menu
Connection options:
• HART via Commubox FXA191 and the serial interface RS 232 C of a computer
• Service interface with adapter FXA193
! Note!
• → See also page 20, section 4.2.6 "Connecting Commubox FXA191 for operation via TOF Tool,
FieldCare or Commuwin II".
• Further information on the ToF Tool can be found on the CD-ROM supplied with the device or
on the Internet (http://www.endress.com, Download → Search for: ToF Tool) The CD is
supplied with any device ordered with the "HistoROM/M-DAT" option.
34 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
5.8 FieldCare
FieldCare is an Endress+Hauser asset management tool based on FDT technology. With FieldCare,
you can configure all Endress+Hauser devices as well as devices from other manufacturers that
support the FDT standard. The following operating systems are supported: WinNT4.0, Win2000
and Windows XP.
FieldCare supports the following functions:
• Configuration of transmitters in online operation
• Loading and saving device data (upload/download)
• HistoROM®/M-DAT analysis
• Documentation of the measuring point
Connection options:
• HART via Commubox FXA191 and the serial interface RS 232 C of a computer
• HART via Fieldgate FXA520
! Note!
• → See also page 20, section 4.2.6 "Connecting Commubox FXA191 for operation via TOF Tool,
FieldCare or Commuwin II".
• Further information on the FieldCare can be found on the Internet (http://www.endress.com,
Download → Search for: FieldCare).
! Note!
• For further information, please refer to the Operating Instructions BA124F "Commuwin II
FXS113".
• You can obtain the current device description (DD) either from your local Endress+Hauser Sales
Office or via the Internet (http://www.endress.com).
Endress+Hauser 35
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
! Note!
• If operation is locked by means of the DIP-switch, you can only unlock operation again by means
of the DIP-switch. If operation is locked by means of the on-site display or remote operation e.g.
ToF Tool, you can only unlock operation again by means of the on-site display or remote
operation.
• If the operation is locked, any change to the "Damping on/off" DIP switch will not have any
impact on the damping time. Any change will not take effect until the operation has been
unlocked again.
1) Parameters which refer to how the display appears, e.g. LANGUAGE and DISPLAY CONTRAST can still be altered.
Damping [τ]
➀
➁ ➂
Damping [τ]
on on Display
on Sensor
1 2
off
PC
1 2 1 2
off off
HistoROM
– + E
P01-xxxxxxxx-19-xx-xx-xx-133
36 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Operation
Description
! Note!
Any customer-specific configuration carried out by the factory is not affected by a reset (customer-
specific configuration remains). If, after a reset, you wish the parameters to be reset to the factory
settings, please contact Endress+Hauser Service.
Endress+Hauser 37
Operation Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
38 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Commissioning
6 Commissioning
! Note!
The device is configured for the Pressure measuring mode as standard. The measuring range and
the unit in which the measured value is transmitted correspond to the specifications on the
nameplate.
Endress+Hauser 39
Commissioning Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
POS. ZERO ADJUST (685) Position adjustment – the pressure difference between zero (set point) and the measured
Entry pressure need not be known. (A reference pressure is present at the device.)
Example:
– MEASURED VALUE = 2.2 mbar
– Correct the MEASURED VALUE via the POS. ZERO ADJUST parameter with the
"Confirm" option. This means that you are assigning the value 0.0 to the pressure
present.
– MEASURED VALUE (after pos. zero adjust) = 0.0 mbar
– The current value is also corrected.
The CALIB. OFFSET parameter displays the resulting pressure difference (offset) by which
the MEASURED VALUE was corrected.
Factory setting:
0
POS. INPUT VALUE (563) Position adjustment – the pressure difference between zero (set point) and the measured
Entry pressure need not be known. (A reference pressure is present at the device.)
Example:
– MEASURED VALUE = 0.5 mbar
– For the POS. INPUT VALUE parameter, specify the desired set point for the
MEASURED VALUE, e.g. 2 mbar.
(MEASURED VALUEnew = POS. INPUT VALUE)
– MEASURED VALUE (after entry for POS. INPUT VALUE) = 2.0 mbar
– The CALIB. OFFSET parameter displays the resulting pressure difference (offset) by
which the MEASURED VALUE was corrected.
CALIB. OFFSET = MEASURED VALUEold – POS. INPUT VALUE,
here: CALIB. OFFSET = 0.5 mbar – 2.0 mbar = – 1.5 mbar)
– The current value is also corrected.
Factory setting:
0
CALIB. OFFSET (319) Position adjustment – the pressure difference between zero (set point) and the measured
Entry pressure is known. (A reference pressure is not present at the device.)
Example:
– MEASURED VALUE = 2.2 mbar
– Via the CALIB. OFFSET parameter, enter the value by which the MEASURED VALUE
should be corrected. To correct the MEASURED VALUE to 0.0 mbar, you must enter
the value 2.2 here.
(MEASURED VALUE new = MEASURED VALUEold – CALIB. OFFSET)
– MEASURED VALUE (after entry for calib. offset) = 0.0 mbar
– The current value is also corrected.
Factory setting:
0
40 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Commissioning
! Note!
• There is a Quick Setup menu for each of the measuring modes Pressure and Level which guides
you through the most important basic functions. With the setting in the MEASURING MODE
parameter, you specify which Quick Setup menu should be displayed. → See also page 39, section
6.2 "Selecting language and measuring mode".
• For a detailed description of the parameters see the Operating Instructions BA 274P "Cerabar S/
Deltabar S, Description of device functions"
– Table 6, POSITION ADJUSTMENT
– Table 7, BASIC SETUP
– Table 13, EXTENDED SETUP
→ See also page 2, section "Overview documentation".
• For pressure measurement, select the "Pressure" option by means of the MEASURING MODE
parameter. The operating menu is structured appropriately. → See also section 10.1.
Measured value
1)
GROUP SELECTION
1) 1)
LANGUAGE MEASURING MODE QUICK SETUP OPERATING MENU
2)
MEASURING MODE
Pressure Level
SET LRV
SET URV
1) Display via on-site display only
2) Display via ToF Tool, FieldCare
and HART handheld terminal only DAMPING VALUE
P01-PMx7xxxx-19-xx-xx-xx-002
MEASURING MODE
Select "Pressure" option.
GROUP SELECTION
Select QUICK SETUP menu.
POS. ZERO ADJUST (gauge pressure sensors) POS. ZERO ADJUST (gauge pressure sensors)
Due to orientation of the device, there may be a shift in Due to orientation of the device, there may be a shift in
the measured value. You correct the MEASURED the measured value. You correct the MEASURED
VALUE via the POS. ZERO ADJUST parameter with the VALUE via the POS. ZERO ADJUST parameter with the
"Confirm" option, i. e. you assign the value 0.0 to the "Confirm" option, i. e. you assign the value 0.0 to the
pressure present. pressure present.
Endress+Hauser 41
Commissioning Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
POS. ZERO ADJUST (absolute pressure sensors) POS. ZERO ADJUST (absolute pressure sensors)
Due to orientation of the device, there may be a shift in Due to orientation of the device, there may be a shift in
the measured value. For the POS. INPUT VALUE the measured value. For the POS. INPUT VALUE
parameter, specify the desired set point for the parameter, specify the desired set point for the
MEASURED VALUE. MEASURED VALUE.
! Note!
For on-site operation, see also page 25, section 5.2.3 "Function of the operating elements" and
page 28, section 5.4 "On-site operation".
42 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Commissioning
Endress+Hauser 43
Commissioning Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
! Note!
• There is a Quick Setup menu for each of the measuring modes Pressure and Level which guides
you through the most important basic functions. With the setting in the MEASURING MODE
parameter, you specify which Quick Setup menu should be displayed. → See also page 39, section
6.2 "Selecting language and measuring mode".
• For a detailed description of the parameters see the Operating Instructions BA274P "Cerabar S/
Deltabar S, Description of device functions"
– Table 6, POSITION ADJUSTMENT
– Table 8 to 11, BASIC SETUP
– Table 14, EXTENDED SETUP
– Table 16, LINEARISATION – On-site operation
– Table 17, LINEARISATION – ToF Tool and HART handheld terminal.
→ See also page 2, section "Overview documentation".
• For level measurement, select the "Level" option by means of the MEASURING MODE
parameter. The operating menu is structured appropriately. → See also section 10.1.
The factory settings for the LEVEL MODE, CALIBRATION MODE, LIN. MEASURAND, EMPTY
CALIB. and FULL CALIB. parameters are as follows:
– LEVEL MODE = Linear
– CALIBRATION MODE = wet
– LIN. MEASURAND = %
– EMPTY CALIB. = 0 %
– FULL CALIB. = 100 %.
! Note!
Some parameters of the "Level" measuring mode are only displayed if other parameters are
appropriately configured. For example, the EMPTY PRESSURE parameter is not displayed unless the
"Linear" option was selected for the LEVEL MODE parameter and the "wet" option was selected for
the CALIBRATION MODE parameter (see following figure). You can find the LEVEL MODE
parameter in the BASIC SETTINGS function group (menu path: (GROUP SELECTION →)
OPERATING MENU → SETTINGS → BASIC SETTINGS).
Measured value
1)
GROUP SELECTION
1) 1)
LANGUAGE MEASURING MODE QUICK SETUP OPERATING MENU
2)
MEASURING MODE
Pressure Level
3) 4)
1) Display via on-site display only
2) Display via ToF Tool, FieldCare and EMPTY CALIB.
HART handheld terminal only
3) LEVEL MODE = "Pressure Linearized" FULL CALIB.
or "Height Linearized" and/or
CALIBRATION MODE = "dry"
4) LEVEL MODE = "Linear" and
DAMPING VALUE
CALIBRATION MODE = "wet"
P01-PMx7xxxx-19-xx-xx-xx-001
Fig. 28: Quick Setup menu for the Level measuring mode
44 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Commissioning
MEASURING MODE
Select "Level" option.
GROUP SELECTION
Select QUICK SETUP menu.
POS. ZERO ADJUST (gauge pressure sensors) POS. ZERO ADJUST (gauge pressure sensors)
Due to orientation of the device, there may be a shift in Due to orientation of the device, there may be a shift in
the measured value. You correct the MEASURED the measured value. You correct the MEASURED
VALUE via the POS. ZERO ADJUST parameter with the VALUE via the POS. ZERO ADJUST parameter with the
"Confirm" option, i. e. you assign the value 0.0 to the "Confirm" option, i. e. you assign the value 0.0 to the
pressure present. pressure present.
POS. INPUT VALUE (absolute pressure sensors) POS. INPUT VALUE (absolute pressure sensors)
Due to orientation of the device, there may be a shift in Due to orientation of the device, there may be a shift in
the measured value. For the POS. INPUT VALUE the measured value. For the POS. INPUT VALUE
parameter, specify the desired set point for the parameter, specify the desired set point for the
MEASURED VALUE. MEASURED VALUE.
EMPTY CALIB. (Level mode "linear", Calibration EMPTY CALIB. (Level mode "linear", Calibration
mode "wet") mode "wet")
Enter level for the lower calibration point. Enter level for the lower calibration point.
For this parameter, enter a level value which is assigned For this parameter, enter a level value which is assigned
to the pressure present at the device. to the pressure present at the device.
FULL CALIB. (Level mode "linear", Calibration FULL CALIB. (Level mode "linear", Calibration
mode "wet") mode "wet")
Enter level for the upper calibration point. Enter level for the upper calibration point.
For this parameter, enter a level value which is assigned For this parameter, enter a level value which is assigned
to the pressure present at the device. to the pressure present at the device.
! Note!
For on-site operation, see also page 25, section 5.2.3 "Function of the operating elements" and
page 28, section 5.4 "On-site operation".
Endress+Hauser 45
Maintenance Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
7 Maintenance
Cerabar S requires no maintenance.
8 Trouble-shooting
8.1 Messages
The following table lists all the possible messages that can occur.
The device differentiates between the error types "Alarm", "Warning" and "Error". At the factory, all
"Error" type messages are set to "Warning".
→ See "Error type/NA 64" column and section 8.2 "Response of outputs to errors".
In addition, the "Error type/NA 64" column classifies the messages in accordance with NAMUR
Recommendation NA 64:
• Break down: indicated with "B"
• Maintenance need: indicated with "C" (check request)
• Function check: indicated with "I" (in service)
Error message display on the on-site display:
• The measured value display shows the message with the highest priority. → See "Priority"
column.
• The ALARM STATUS parameter shows all the messages present in descending order of priority.
You can scroll through all the messages present with the S-key or O-key.
Message display via the ToF Tool, FieldCare, Commuwin II and HART handheld terminal:
• The ALARM STATUS parameter shows the message with the highest priority. → See "Priority"
column.
! Note!
• For support and further information, please contact Endress+Hauser Service.
• → See also section 8.4, 8.5 and 8.6.
46 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
101 (A101) Alarm B>Sensor electronic EEPROM – Electromagnetic effects are greater – Wait a few minutes. 17
B error than specifications in the technical – Restart the device. Perform reset
data. (→ See section 9.) This (Code 62).
message normally only appears – Block off electromagnetic effects or
briefly. eliminate source of disturbance.
102 (W102) Warning C>Checksum error in EEPROM: – Main electronics defect. Correct – Replace main electronics. 53
C peakhold segment measurement can continue as long
as you do not need the peak hold
indicator function.
106 (W106) Warning C>Downloading - please wait – Downloading. – Wait for download to complete. 52
C
110 (A110) Alarm B>Checksum error in EEPROM: – The supply voltage is disconnected – Reestablish supply voltage. Perform 6
B configuration segment when writing. reset (Code 7864) if necessary.
Carry out calibration again.
113 (A113) Alarm B>ROM failure in transmitter – Main electronics defect. – Replace main electronics. 1
B electronic
115 (E115) Error C>Sensor overpressure – Overpressure present. – Reduce pressure until message 29
C disappears.
116 (W116) Warning C>Download error, repeat – The file is defect. – Use another file. 36
C download
– During the download, the data are – Check cable connection PC –
not correctly transmitted to the transmitter.
processor, e.g. because of open cable – Block off electromagnetic effects or
connections, spikes (ripple) on the eliminate sources of disturbance.
supply voltage or electromagnetic – Perform reset (Code 7864) and carry
effects. out calibration again.
– Repeat download.
120 (E120) Error C>Sensor low pressure – Pressure too low. – Increase pressure until message 30
C disappears.
121 (A121) Alarm B>Checksum error in factory – Main electronics defect. – Replace main electronics. 5
B segment of EEPROM
122 (A122) Alarm B>Sensor not connected – Cable connection sensor –main – Check cable connection and repair if 13
B electronics disconnected. necessary.
130 (A130) Alarm B>EEPROM is defect. – Main electronics defect. – Replace main electronics. 10
B
131 (A131) Alarm B>Checksum error in EEPROM: – Main electronics defect. – Replace main electronics. 9
B min/max segment
132 (A132) Alarm B>Checksum error in totalizer – Main electronics defect. – Replace main electronics. 7
B EEPROM
Endress+Hauser 47
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
133 (A133) Alarm B>Checksum error in History – An error occurred when writing. – Perform reset (Code 7864) and carry 8
B EEPROM out calibration again.
602 (W602) Warning C>Linearisation curve not – The linearisation table is not – Add to linearisation table or perform 57
C monoton monotonic increasing. linearisation again.
604 (W604) Warning C>Linearisation table not valid. At – The linearisation table consists of – Add to linearisation table. If 58
C least 2 points less than 2 points. necessary, perform linearisation
again.
613 (W613) Warning I>Simulation is active – Simulation is switched on, i.e. the – Switch off simulation. 60
I device is not measuring at present.
620 (E620) Error C>Current output out of range The current is outside the permitted – Check pressure applied, reconfigure 49
C range 3.8...20.5 mA. measuring range if necessary (→ See
– The pressure applied is outside the also Operating Instructions BA274P,
chapter 4 to 6 or these Operating
set measuring range (but within the
Instructions, page 2.)
sensor range).
– Perform reset (Code 7864) and carry
out calibration again.
700 (W700) Warning C>Last configuration not stored – An error occurred when writing or – Perform reset (Code 7864) and carry 54
C reading configuration data or the out calibration again.
power supply was disconnected.
701 (W701) Warning C>Measuring chain config. – The calibration carried out would – Carry out calibration again. 50
C exceeds sensor range result in the sensor nominal
operating range being undershot or
overshot.
702 (W702) Warning C>HistoROM data not consistent. – Data were not written correctly to – Repeat upload. 55
C the HistoROM, e.g. if the HistoROM – Perform reset (Code 7864) and carry
was detached during the writing out calibration again.
process.
– HistoROM does not have any data. – Copy suitable data to the HistoROM.
(→ See also page 31, section 5.5.1
"Copying configuration data".)
703 (A703) Alarm B>Measurement error – Fault in the main electronics. – Briefly disconnect device from the 22
B power supply.
704 (A704) Alarm B>Measurement error – Fault in the main electronics. – Briefly disconnect device from the 12
B power supply.
705 (A705) Alarm B>Measurement error – Fault in the main electronics. – Briefly disconnect device from the 21
B power supply.
706 (W706) Warning C>Configuration in HistoROM – Configuration (parameters) in the – Copy data from the device to the 59
C and device not identical HistoROM and in the device is not HistoROM. (→ See also page 31,
identical. section 5.5.1 "Copying configuration
data".)
– Copy data from the HistoROM to
the device. (→ See also page 31,
section 5.5.1 "Copying configuration
data".) The message remains if the
HistoROM and the device have
different software versions. The
message goes out if you copy the
data from the device to the
HistoROM.
48 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
707 (A707) Alarm B>X-VAL. of lin. table out of edit – At least one X-VALUE in the – Carry out calibration again. (→ See 38
B limits. linearisation table is either below the also Operating Instructions BA274P,
value for HYDR. PRESS MIN. or chapter 5 or these Operating
MIN. LEVEL or above the value for Instructions, page 2.)
HYDR. PRESS. MAX. or LEVEL
MAX.
710 (W710) Warning B>Set span too small. Not – Values for calibration (e.g. lower – Adjust calibration to suit sensor. 51
C allowed. range value and upper range value) (→ See also Operating Instructions
are too close together. BA274P, parameter description
MINIMUM SPAN or these
Operating Instructions, page 2.)
– The sensor was replaced and the – Adjust calibration to suit sensor.
customer-specific configuration does – Replace sensor with a suitable
not suit the sensor. sensor.
– The sensor was replaced and the – Reconfigure lower range value and/
customer-specific configuration does or upper range value to suit the
not suit the sensor. sensor. Pay attention to position
factor.
– Replace sensor with a suitable
sensor.
– Unsuitable download carried out. – Check configuration and perform
download again.
713 (A713) Alarm B>100% POINT level out of edit – The sensor was replaced. – Carry out calibration again. 39
B limits
715 (E715) Error C>Sensor over temperature – The temperature measured in the – Reduce process temperature/ 32
C sensor is greater than the upper ambient temperature.
nominal temperature of the sensor.
(→ See also Operating Instructions
BA274P, parameter description
Tmax SENSOR or these Operating
Instructions, page 2.)
716 (A716) Alarm B>Sensor diaphragm broken – Sensor defect. – Replace sensor. 24
B
717 (E717) Error C>Transmitter over temperature – The temperature measured in the – Reduce ambient temperature. 34
C electronics is greater than the upper
nominal temperature of the
electronics (+88 °C).
718 (E718) Error C>Transmitter under temperature – The temperature measured in the – Increase ambient temperature. 35
C electronics is smaller than the lower Insulate device if necessary.
nominal temperature of the
electronics (–43 °C).
719 (A719) Alarm B>Y-VAL of lin. table out of edit – At least on Y-VALUE in the – Carry out calibration again. (→ See 40
B limits linearisation table is below the MIN. also Operating Instructions BA274P,
TANK CONTANT or above the chapter 5 or these Operating
MAX. TANK CONTENT. Instructions, page 2.)
Endress+Hauser 49
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
720 (E720) Error C>Sensor under temperature – The temperature measured in the – Increase process temperature/ 33
C sensor is smaller than the lower ambient temperature.
nominal temperature of the sensor.
(→ See also Operating Instructions
BA274P, parameter description
Tmin SENSOR or Operating
Instructions, page 2.)
721 (A721) Alarm B>ZERO POSITION level out of – LEVEL MIN or LEVEL MAX has – Perform reset (Code 2710) and carry 41
B edit limits been changed. out calibration again.
722 (A722) Alarm B>EMPTY CALIB. or FULL – LEVEL MIN or LEVEL MAX has – Perform reset (Code 2710) and carry 42
B CALIB. out of edit limits been changed. out calibration again.
723 (A723) Alarm B>MAX. FLOW out of edit limits – FLOW-MEAS. TYPE has been – Carry out calibration again. 43
B changed.
725 (A725) Alarm B>Sensor connection error, cycle – Electromagnetic effects are greater – Block off electromagnetic effects or 25
B disturbance than specifications in the technical eliminate source of disturbance.
data.
(→ See section 9.)
726 (E726) Error C>Sensor temperature error - – Electromagnetic effects are greater – Block off electromagnetic effects or 31
C overrange than specifications in the technical eliminate source of disturbance.
data.
(→See section 9.)
– Process temperature is outside – Check temperature present, reduce
permitted range. or increase if necessary.
727 (E727) Error C>Sensor pressure error - – Electromagnetic effects are greater – Block off electromagnetic effects or 28
C overrange than specifications in the technical eliminate source of disturbance.
data.
(→ See section 9.)
728 (A728) Alarm B>RAM error – Fault in the main electronics. – Briefly disconnect device from the 2
B power supply.
730 (E730) Error C>LRV user limits exceeded – Pressure measured value has – Check system/pressure measured 46
C undershot the value specified for the value.
Pmin ALARM WINDOW parameter. – Change value for Pmin ALARM
WINDOW if necessary. (→ See also
Operating Instructions BA274P,
parameter description Pmin ALARM
WINDOW or these Operating
Instructions, page 2.)
50 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
731 (E731) Error C>URV user limits exceeded – Pressure measured value has – Check system/pressure measured 45
C overshot the value specified for the value.
Pmax ALARM WINDOW – Change value for Pmax ALARM
parameter. WINDOW if necessary. (→ See also
Operating Instructions BA274P,
parameter description Pmax ALARM
WINDOW or these Operating
Instructions, page 2.)
732 (E732) Error C>LRV Temp. User limits – Temperature measured value has – Check system/temperature 48
C exceeded undershot the value specified for the measured value.
Tmin ALARM WINDOW parameter. – Change value for Tmin ALARM
WINDOW if necessary. (→ See also
Operating Instructions BA274P,
parameter description Tmin ALARM
WINDOW or these Operating
Instructions, page 2.)
733 (E733) Error C>URV Temp. User limits – Temperature measured value has – Check system/temperature 47
C exceeded overshot the value specified for the measured value.
Tmax ALARM WINDOW – Change value for Tmax ALARM
parameter. WINDOW if necessary. (→ See also
Operating Instructions BA274P,
parameter description Tmax
ALARM WINDOW or these
Operating Instructions, page 2.)
736 (A736) Alarm B>RAM error – Fault in the main electronics. – Briefly disconnect device from the 4
B power supply.
– Main electronics defect. – Replace main electronics.
737 (A737) Alarm B>Measurement error – Fault in the main electronics. – Briefly disconnect device from the 20
B power supply.
738 (A738) Alarm B>Measurement error – Fault in the main electronics. – Briefly disconnect device from the 19
B power supply.
– Main electronics defect. – Replace main electronics.
739 (A739) Alarm B>Measurement error – Fault in the main electronics. – Briefly disconnect device from the 23
B power supply.
– Main electronics defect. – Replace main electronics.
740 (E740) Error C>Calculation overflow, bad – Level measuring mode: the – Check configuration and carry out 27
C configuration measured pressure has undershot calibration again if necessary.
the value for HYDR. PRESS. MIN. or – Select a device with a suitable
overshot the value for HYDR. PRESS measuring range.
MAX.
– Flow measuring mode: the – Check configuration and carry out
measured pressure has undershot calibration again if necessary.
the value for MAX. PRESS FLOW. – Select a device with a suitable
measuring range.
741 (A741) Alarm B>TANK HEIGHT out of edit – LEVEL MIN or LEVEL MAX has – Perform reset (Code 2710) and carry 44
B limits been changed. out calibration again.
742 (A742) Alarm B>Sensor connection error – Electromagnetic effects are greater – Wait a few minutes. 18
B (upload) than specifications in the technical – Perform reset (Code 7864) and carry
data. (→ See section 9.) This out calibration again.
message normally only appears
briefly.
Endress+Hauser 51
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
743 (E743) Alarm B>Electronic PCB error during – This message normally only appears – Wait a few minutes. 14
B initialisation briefly. – Restart the device. Perform reset
(Code 62).
744 (A744) Alarm B>Main electronic PCB error – Electromagnetic effects are greater – Restart the device. Perform reset 11
B than specifications in the technical (Code 62).
data. – Block off electromagnetic effects or
(→ See section 9.) eliminate source of disturbance.
745 (W745) Warning C>Sensor data unknown – Sensor does not suit the device – Replace sensor with a suitable 56
C (electronic sensor nameplate). sensor.
Device continues measuring.
746 (W746) Warning C>Sensor connection error - – Electromagnetic effects are greater – Wait a few minutes. 26
C initialising than specifications in the technical – Restart the device. Perform reset
data. (→ See section 9.) This (Code 7864).
message normally only appears – Block off electromagnetic effects or
briefly. eliminate source of disturbance.
747 (A747) Alarm B>Sensor software not compatible – Sensor does not suit the device – Replace sensor with a suitable 16
B to electronics (electronic sensor nameplate). sensor.
748 (A748) Alarm B>Memory failure in signal – Electromagnetic effects are greater – Block off electromagnetic effects or 15
B processor than specifications in the technical eliminate source of disturbance.
data.
(→See section 9.)
52 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
Current output Assumes the value specified via the Device continues measuring. For this error, you can enter whether the
OUTPUT FAIL MODE1, ALT. CURR. device should react as in the event of an
OUTPUT1 and SET MAX. ALARM1 alarm or as in the event of a warning. See
parameter. → See also the following section corresponding "Alarm" or "Warning"
"Configuring current output for an alarm". column. (→ See also Operating Instructions
BA274P, parameter description SELECT
ALARM TYPE or these Operating
Instructions, page 2.)
Bar graph → See this table, current output. → See this table, current output. → See this table, current output.
(on-site display)
On-site display – The measured value and message are – The measured value and message are – The measured value and message are
displayed alternately displayed alternately displayed alternately
– Measured value display: -symbol is – Measured value display: -symbol – Measured value display: see
permanently displayed. flashes. corresponding "Alarm" or "Warning"
column
Message display Message display: Message display:
– 3-digit number such as A122 and – 3-digit number such as W613 and – 3-digit number such as E731 and
– Description – Description – Description
Remote operation In the case of an alarm, the In the case of a warning, the ALARM In the case of an error, the
(ToF Tool, FieldCare, ALARM STATUS2 parameter displays a 3- STATUS2 parameter displays a 3-digit ALARM STATUS2 parameter displays a 3-
Commuwin II or HART digit number such as 122 for "Sensor not number such as 613 for "Simulation is digit number such as 731 for "URV user
handheld terminal) connected". active". limits exceeded".
I
max.
➀ 21...23 mA
min.
➂ 3.6 mA
t
P01-xxxxxxxx-05
Endress+Hauser 53
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
Use the ALT. CURR. OUTPUT parameter to set the current output value for the error messages E
120 "Sensor low pressure" and E 115 "Sensor overpressure". You have the following options:
– Normal: the current output assumes the value set via the OUTPUT FAIL MODE and SET MAX.
ALARM parameters.
– NAMUR
– Lower sensor limit undershot (E 120 "Sensor low pressure"): 3.6 mA
– Upper sensor limit overshot (E 115 "Sensor overpressure") overshot: current output assumes
the value set via the SET MAX ALARM parameter.
Factory setting: ALT. CURR. OUTPUT = normal
Settings 1 Measures
– ALARM DISPL. TIME = 0 s – Rectify cause of the message (see also Section 8.1).
– ACK. ALARM MODE = off
– ALARM DISPL. TIME > 0 s – Rectify cause of the message (see also Section 8.1).
– ACK. ALARM MODE = off – Wait for the alarm display time to elapse.
– ALARM DISPL. TIME = 0 s – Rectify cause of the message (see also Section 8.1).
– ACK. ALARM MODE = on – Confirm message using ACK. ALARM parameter.
– ALARM DISPL. TIME > 0 s – Rectify cause of the message (see also Section 8.1).
– ACK. ALARM MODE = on – Confirm message using ACK. ALARM parameter.
– Wait for the alarm display time to elapse. If a message appears and the alarm
display time elapses before the message has been acknowledged, the message
will be cleared once it has been acknowledged.
1) Menu path for ALARM DISPL. TIME and ACK. ALARM MODE: (GROUP SELECTION →) OPERATING MENU →
DIAGNOSTICS → MESSAGES
If the on-site display displays a message, you can delete it with the F-key.
If there are several messages, the on-site display shows the message which has the highest priority
(see also Section 8.1). Once you have deleted this message using the F-key, the message with the
next highest priority is displayed. You can use the F-key to delete each message, one after the other.
The ALARM STATUS parameter continues to display all the messages present.
54 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
8.4 Repair
The Endress+Hauser repairs concept provides for measuring devices to have a modular design and
also the customer may carry out repairs.
Section "Spare parts" contains all the spare parts listed with their order numbers. You can order them
from Endress+Hauser for repairs on your Cerabar S. As far as necessary, the spare parts also include
replacement instructions.
! Note!
• For certified devices, please consult Chapter "Repair of Ex-certified devices".
• For more information on service and spare parts contact the Endress+Hauser Service. ( → See
www.endress.com/worldwide)
Endress+Hauser 55
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
You can order sensor modules as spare parts for PMC71 and PMP71. For PMC71 and PMP71, see
page 60 and page 63 respectively.
The right order number for the sensor module is derived from the Order No. which is indicated on
the nameplate (→ see also page 6).
10 20 30 40 50 60 70 80 90 100 110
Order code/Order No. of the device
PMC71 – ❑ ❑ ❑ ❑❑ ❑ ❑ ❑❑ ❑ ❑ ❑ ❑ see name plate
PMP71 –
P01-PMx7xxxx-16-xx-xx-xx-003
56 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
20
21
20
21
40
31
30
15
13
10
12
35
21
25
P01-xMx7xxxx-09-xx-xx-xx-002
Endress+Hauser 57
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
20 Cover
52020432 Cover for Aluminium T14 housing incl. gasket, not for EEx d/XP
52020433 Cover for Aluminium T14 housing incl. gasket, for EEx d/XP
52028310 Cover for AISI 316L T14 housing incl. gasket
71002774 Cover for AISI 316L T17 housing incl. gasket
52020494 Cover for Aluminium T14 housing with sight glass incl. gasket, Ex-free
52020492 Cover for Aluminium T14 housing with sight glass incl. gasket, not for EEx d/XP
52020493 Cover for Aluminium T14 housing with sight glass incl. gasket, for EEx d/XP
71002811 Cover for AISI 316L T17 housing with sight glass incl. gasket, for Ex-free area, 1/2 G EEx ia, 3 G EEx nA, IS, NI
71002810 Cover for AISI 316L T17 housing with sight glass incl. gasket, for 1/2 D, 1/3 D, 1 GD, 1/2 GD, DIP, Dust Ex
30 Electronics
52024400 Electronics 4...20 mA, HART, Ex, version 2.0x, push button on electronics
52024111 Electronics 4...20 mA, HART, Ex, version 2.0x, push button on housing
31 HistoROM/M-DAT
52027785 HistoROM/M-DAT, ToF Tool-CD included
35 Terminal block
52020434 Terminal 3-pole, RFI-filter 4...20 mA, HART Ex ia
52020436 Terminal 3-pole, RFI-filter 4...20 mA, HART Ex d
40 Display module
71002865 Display module VU340 HART with 3 push buttons and holder
58 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
12
50 50 50
57
55 57
57
58
P01-PMC71xxx-09-xx-xx-xx-000
57 Process gasket
52020768 O-ring 26.7x1.78 FKM 70 Viton, oil and grease free (5 pieces)
52020769 O-ring 26.7x1.78 NBR 70 (5 pieces)
52020770 O-ring 26.7x1.78 EPDM 70 (5 pieces)
52020771 O-ring 26.7x1.78 FKM Viton –40°C (5 pieces)
52020772 O-ring 26.7x1.78 FFKM 70 Kalrez (1 piece)
52020773 O-ring 26.7x1.78 FFKM 75 Chemraz (1 piece)
58 Fastening angle
52020441 Mounting angle, adapter and screws
Endress+Hauser 59
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
10 Approval:
A For non-hazardous areas
1 ATEX II 1/2 G EEx ia IIC T6
6 ATEX II 1/2 G EEx ia IIC T6, overspill protection WHG
2 ATEX II 1/2 D
8 ATEX II 1 GD EEx ia II T6
3 ATEX II 1/3 GD EEx ia IIC T6
4 ATEX II 1/3 D
7 ATEX II 3 G EEx nA II T6
S FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
Q FM DIP, Class II, III Division 1, Groups E – G
R FM NI, Class I, Division 2, Groups A – D
U CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
60 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
80 Seal:
A FKM Viton
B EPDM
D Kalrez
E Chemraz
F NBR
1 FKM Viton, oxygen service
Endress+Hauser 61
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
62 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
12
50
P01-PMP71xxx-09-xx-xx-xx-000
10 Approval:
A For non-hazardous areas
1 ATEX II 1/2 G EEx ia IIC T6
6 ATEX II 1/2 G EEx ia IIC T6, overspill protection WHG
2 ATEX II 1/2 D
4 ATEX II 1/3 D
8 ATEX 1 GD EEx ia IIC T6
5 ATEX II 2 G EEx d IIC T6
7 ATEX II 3 G EEx nA II T6
S FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
T FM XP, Class I Division 1, Groups A – D; AEx d
Q FM DIP, Class II, III Division 1, Groups E – G
R FM NI, Class I, Division 2, Groups A – D
U CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
V CSA XP Class I Division 1, Groups B – D; Ex d
W CSA Class II, III Division 1, Groups E – G (Dust-Ex)
3 Combi-certificate: ATEX II 1/2 GD EEx ia IIC T6
C Combi-certificate: FM IS and XP Class I Division 1, Groups A – D
D Combi-certificate: CSA IS and XP Class I Division 1, Groups A – D
E Combi-certificate: FM/CSA IS and XP Class I Division 1, Groups A – D
Endress+Hauser 63
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
60 Membrane material:
1 AISI 316L
2 Alloy C276
64 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Trouble-shooting
90 Fill fluid:
A Silicone
F Inert oil
Endress+Hauser 65
Trouble-shooting Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
8.8 Disposal
When disposing, separate and recycle the device components based on the materials.
02.00/05.2004 – Number of parameters in the Quick Setup menus has been – Documentation changed to reflect changes to the software.
reduced. – The "Description of Parameters" Section has been moved to the
– On-site operation: LANGUAGE and MEASURING MODE Operating Instructions BA274P.
parameters have been moved to the top level.
– New SAFETY CONFIRM. group implemented for SIL.
→ See also SD190P Safety Manual Cerabar S.
– MEASURING MODE "Level", LEVEL MODE "Linear": AREA
UNIT and TANK SECTION parameters have been replaced with
the TANK VOLUME and TANK HEIGHT parameters.
– Function of the SIMULATED VALUE parameter has been split
across six parameters.
– SENSOR TRIM and CURRENT TRIM groups have been
removed.
– Sensor adapt reset, code 1209 and sensor calibration reset, code
2509 have been removed.
– Quick Setup menus are available via ToF Tool.
Compatible with:
– ToF Tool Field Tool Package version 2.00.00 or higher
– Commuwin II version 2.08.-1, Update > G
– HART Communicator DXR375
with Device Rev.: 20, DD Rev.: 1
02.01/06.2005 – Operating keys also integrated on the optional on-site display. – Section 5.1 "On-site display" changed.
– Chinese and Japanese are available as the menu language on – Section 6.2 "Selecting language and measuring mode" changed.
request.
Compatible with:
– ToF Tool Field Tool Package version 3.00.00 or higher
– FieldCare version 2.01.00, DTM Library version 2.06.00,
DTM: Cerabar S/PMx7x/V02.00 V 1.4.98.74*
– Commuwin II version 2.08.-1, Update > G*
– HART Communicator DXR375 with Device Rev.: 20,
DD Rev.: 1*
* Menu languages Chinese and Japanese not selectable
66 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Technical data
9 Technical data
For technical data, please refer to the Technical Information TI383P for Cerabar S.
→ See also page 2, section "Overview documentation".
10 Appendix
Endress+Hauser 67
Appendix Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
Measured value
1)
GROUP SELECTION
1) 1)
LANGUAGE MEASURING MODE QUICK SETUP OPERATING MENU
SETTINGS
2)
MEASURING MODE
CUSTOMER UNIT P
*
2)
CUST. UNIT FACT. P
*
MEASURING MODE
4) 4)
POS. INPUT VALUE POS. INPUT VALUE POS. INPUT VALUE SET URV Continuation,
see the following page
SET LRV EMPTY CALIB.
* CALIB. OFFSET GET LRV
X
SET URV FULL CALIB.
* GET URV
P01-PMx7xxxx-19-xx-xx-xx-003
68 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Appendix
Pressure Height
Linear
Linearized Linearized E
ADJUSTED DENSITY * HEIGHT UNIT * EMPTY CALIB. DENSITY UNIT CUTOMER UNIT H *
CUSTOMER UNIT H * EMPTY PRESSURE ADJUST DENSITY CUST. UNIT FACT. H *
CUST. UNIT FACT. H * FULL CALIB. HEIGHT UNIT * H H
100% POINT
* FULL PRESSURE CUSTOMER UNIT H *
ZERO POSITION ADJUSTED DENSITY * CUST. UNIT FACT. H *
100% POINT *
SET LRV
ZERO POSITION
SET URV
* There are parameters that are only displayed if other parameters are
appropriately configured.
For example the CUST. UNIT FACT. H parameter is only displayed
if the "User unit" option was selected for the HEIGHT UNIT parameter.
These parameters are indicated with a "*".
P01-xxxxxxxx-19-xx-xx-xx-069
Endress+Hauser 69
Appendix Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
Pressure Level
TEMP. ENG UNIT TEMP. ENG UNIT TANK CONTENT MIN TANK CONTENT MIN
Y-VAL.
TANK DESCRIPTION
TANK DESCRIPTION
TAB. ACTIVE
* There are parameters that are only displayed if other parameters are
appropriately configured.
These parameters are indicated with a "*".
P01-PMx7xxxx-19-xx-xx-xx-005
70 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Appendix
SAFETY LOCK MENU DESCRIPTOR OUTPUT CURRENT BUS ADDRESS DEVICE SERIAL No Pmax PROC. CONN. SENSOR SER. NO.
SAFETY PASSWORD MAIN DATA FORMAT CURR CHARACT. DEVICE TYPE ELECTR. SERIAL No PROC. CONN. TYPE PRESS. SENS LOLIM
DIGIT SETS ALTERNATE DATA OUTPUT FAIL MODE DEVICE REVISION CUST. TAG NUMBER MAT. PROC. CONN. + PRESS. SENS HILIM
2)
CURRENT OUTPUT LANGUAGE ALT. CURR. OUTPUT PREAMBLE NUMBER LONG TAG NUMBER MAT. PROC. CONN. – MINIMUM SPAN
ACK. ALARM MODE DISPLAY CONTRAST SET MAX. ALARM MANUFACTOR ID ADDITIONAL INFO. SEAL TYPE SENSOR MEAS. TYPE
2)
CALIB. OFFSET DIGIT SETS SET MIN. CURRENT HART MESSAGE DEVICE DESIGN. BOLTS MATERIAL Pmin SENS. DAMAGE
3) 2)
MEASURING MODE ASSIGN CURRENT HART DATE HARDWARE REV. NUTS MATERIAL Pmax SENS. DAMAGE
2) 2)
SET LRV PRIMARY VALUE IS SOFTWARE VERSION DRAIN VENT MAT. MAT. MEMBRANE
2) 2)
SET URV PRIMARY VALUE CONFIG RECORDER DRAIN VENT POS. FILLING FLUID
2) 2)
DAMPING VALUE SECONDARY VAL. IS PCB TEMPERATURE THREAD Tmin SENSOR
2) 2)
CONF. PASSWORD SECONDARY VALUE ALLOWED MIN. TEMP MOUNTING THREAD Tmax SENSOR
2) 2)
THIRD VALUE IS ALLOWED MAX. TEMP REMOTE SEAL POS. SENSOR H/WARE REV.
2) 2)
THIRD VALUE DIP STATUS REMOTE SEAL NEG.
2) 2)
4TH VALUE IS DIAPHRAGM SEAL POS.
2) 2)
4 TH VALUE DIAPHRAGM SEAL NEG.
2)
NR OF REMOTE SEAL
2)
FILL FLUID
P01-PMx7xxxx-19-xx-xx-xx-006
Endress+Hauser 71
Appendix Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART
Pressure Level
MEASURED VALUE MEASURED VALUE COUNTER:P > Pmax ENTER RESET CODE SIMULATION MODE ALARM STATUS Pmin ALARM WINDOW
PRESSURE PRESSURE MAX. MEAS. PRESS. OPERATING HOURS I LAST DIAG. CODE Pmax ALARM WINDOW
CORRECTED PRESS. CORRECTED PRESS. COUNTER:P < Pmin INSERT PIN NO ACK. ALARM MODE Tmin ALARM WINDOW
J
SENSOR PRESSURE SENSOR PRESSURE MIN. MEAS. PRESS. HistoROM AVAIL. ACK. ALARM * Tmax ALARM WINDOW
SENSOR TEMP. SENSOR TEMP. COUNTER:T > Tmax HistoROM CONTROL * RESET ALL ALARMS
RESET PEAKHOLDS
J J J J J J
* There are parameters that are only displayed if other parameters are
appropriately configured.
These parameters are indicated with a "*".
P01-PMx7xxxx-19-xx-xx-xx-007
72 Endress+Hauser
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Appendix
H0 H1 H2 H3 H4 H5 H6 H7 H8 H9
V0 Measured Set LRV Set URV Get LRV Get URV Calib. Pos. zero Damping Output fail Press. eng.
Basic value offset adjust value mode unit
setup
V1 Min. meas. Max. meas. Pos. input Sensor Min. meas. Max. PCB PCB min. PCB max. Temp. eng.
Peak hold press. pressure value temp. temp. meas. temp. tempera- temp. temp. unit
indic. ture
V3 Measuring Level Measurand Tank Editor Lin. edit Tab. Line-numb X-Val. Y-Val.
Lineari- mode mode content unit table mode activate
sation
V4 Eng. unit Level min./ Level max./ Tank Tank Calibration Empty Empty Full calib. Full
Level level Hydr. press. Hydr. press. content content mode calib. pressure pressure
min. max. min. max.
V5 Unit flow Flow- Max. Max Low flow Set l. fl. cut- Reset Neg. flow Totalizer 1 Totalizer 1
Flow meas. pressure flow cut-off off totalizer 1 tot. 1 unit
type flow
V6 Pmin Alarm Pmax Tmin Tmax Proc. conn. Mat. proc. Mat. proc. Seal type Filling fluid Sensor
Process window Alarm Alarm alarm type conn. + conn. – meas. type
info window window window
V7 Output Set min. Set max. Linear/ Assing Low sensor High sensor Press. sens. Press. sens. Sensor
Output current current alarm sqroot. current trim trim LOLIM HILIM pressure
V8 Simulation Simulated Main line Menu Density Adjust Zero 100% DIP status Damp
Additional mode value format descriptor unit density position point switch
function
V9 Alarm Last diag. Ack. alarm Ack. alarm Alarm Alarm displ. Operating Revision Enter reset Insert PIN
Service status code mode delay time hours count code no
VA Cust. tag Additional Device Sensor ser. Electr. Device Software Cust. unit Flow unit
User number info. serial no no. serial no design. version flow scale
info
10.3 Patents
This product may be protected by at least one of the following patents. Further patents are pending.
• US 5,836,063 A1 i EP 0 797 084 B1
• US 5,877,424 A1 i EP 0 780 674 B1
• DE 203 05 869 U1
• US 6,363,790 A1 i EP 0 995 979 B1
• US 5,670,063 A1 i EP 0 516 579 B1
• US 5,539,611 A1
• US 5,050,034 A1 i EP 0 445 382 B1
• US 5,005,421 A1 i EP 0 351 701 B1
• EP 0 414 871 B1
• EP 1 061 351 B1
• US 5,334,344 A1 i EP 0 490 807 B1
• US 6,703,943 A1
Endress+Hauser 73
Cerabar S PMC71, PMP71/72/75 with 4...20 mA HART Index
Index
Numerics O
4...20 mA test signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 On-site display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Operating elements, function . . . . . . . . . . . . . . . . . . . 23, 25
A Operating elements, position. . . . . . . . . . . . . . . . . . . . . . . 22
Alarm messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Operating keys, on-site, function. . . . . . . . . . . . . . . . . 23, 25
Operating keys, on-site, Level measuring mode . . . . . . . . . 27
C Operating keys, on-site, Pressure measuring mode. . . . . . . 26
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Operating keys, position . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Commubox FXA 191, connection. . . . . . . . . . . . . . . . . . . 20 Operating menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28, 67
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
D
Diaphragm seals, installation instructions . . . . . . . . . . . . . 11 P
Diaphragm seals, vacuum application . . . . . . . . . . . . . . . . 11 Pipe mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Position adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Potential matching . . . . . . . . . . . . . . . . . . . . . . . . . . . 18, 20
E Pressure measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Pressure measurement, Quick Setup menu . . . . . . . . . . . . 41
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Q
F Quick Setup menu level . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Factory setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Quick Setup menu pressure . . . . . . . . . . . . . . . . . . . . . . . 41
FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
R
H Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
HART handheld terminal, connection . . . . . . . . . . . . . . . . 19 Repair of Ex-certified devices . . . . . . . . . . . . . . . . . . . . . . 55
Hazardous areas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Heat insulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Rotating the housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
HistoROM/M-DAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
S
I Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Screening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Installation instructions for devices with diaphragm seals. . 11 SIL 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Installation instructions for devices without diaphragm seals 8 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
L
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Language, selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Level measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Level measurement, Quick Setup menu . . . . . . . . . . . . . . 44 T
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Temperature separator, installation instructions . . . . . . . . . 12
Locking operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 ToF Tool operating program . . . . . . . . . . . . . . . . . . . . . . . 34
Trouble-shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
M
Measuring arrangement for level measurement . . . . . . . . . 10 U
Measuring arrangement for pressure measurement . . . . 9–10 Unlocking operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Measuring mode, selection . . . . . . . . . . . . . . . . . . . . . . . . 39
W
N Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Welding recommendation. . . . . . . . . . . . . . . . . . . . . . . . . 14
74 Endress+Hauser
Declaration of Hazardous Material and De-Contamination
Erklärung zur Kontamination und Reinigung
Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA#
clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
RA No. Bitte geben Sie die von E+H mitgeteilte Rücklieferungsnummer (RA#) auf allen Lieferpapieren an und vermerken Sie diese
auch außen auf der Verpackung. Nichtbeachtung dieser Anweisung führt zur Ablehnung ihrer Lieferung.
Because of legal regulations and for the safety of our employees and operating equipment, we need the "Declaration of Hazardous Material
and De-Contamination", with your signature, before your order can be handled. Please make absolutely sure to attach it to the outside of the
packaging.
Aufgrund der gesetzlichen Vorschriften und zum Schutz unserer Mitarbeiter und Betriebseinrichtungen, benötigen wir die unterschriebene
"Erklärung zur Kontamination und Reinigung", bevor Ihr Auftrag bearbeitet werden kann. Bringen Sie diese unbedingt außen an der
Verpackung an.
Used as SIL device in a Safety Instrumented System / Einsatz als SIL Gerät in Schutzeinrichtungen
harmful/
Medium /concentration Identification flammable toxic corrosive irritant other * harmless
Medium /Konzentration CAS No. entzündlich giftig ätzend gesundheits- sonstiges* unbedenklich
schädlich/
reizend
Process
medium
Medium im
Prozess
Medium for
process cleaning
Medium zur
Prozessreinigung
Returned part
cleaned with
Medium zur
Endreinigung
* explosive; oxidising; dangerous for the environment; biological risk; radioactive
* explosiv; brandfördernd; umweltgefährlich; biogefährlich; radioaktiv
Please tick should one of the above be applicable, include safety data sheet and, if necessary, special handling instructions.
Zutreffendes ankreuzen; trifft einer der Warnhinweise zu, Sicherheitsdatenblatt und ggf. spezielle Handhabungsvorschriften beilegen.
“We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge.We further certify that the returned
parts have been carefully cleaned. To the best of our knowledge they are free of any residues in dangerous quantities.”
“Wir bestätigen, die vorliegende Erklärung nach unserem besten Wissen wahrheitsgetreu und vollständig ausgefüllt zu haben. Wir bestätigen
weiter, dass die zurückgesandten Teile sorgfältig gereinigt wurden und nach unserem besten Wissen frei von Rückständen in gefahrbringen-
der Menge sind.”
P/SF/Konta XII
(place, date / Ort, Datum) Name, dept./Abt. (please print /bitte Druckschrift) Signature / Unterschrift
www.endress.com/worldwide
BA271P/00/en/11.05
71009589
CCS/FM+SGML6.0 71009589