SM EN Rad8 070126
SM EN Rad8 070126
SM EN Rad8 070126
Service Manual
Rad-8
Service Manual
The Rad-8 Service Manual is intended to provide the necessary information for
proper servicing of the Rad-8 pulse oximetry system. This manual is structured
to support troubleshooting to the assembly or module level. This manual does
not provide instructions for troubleshooting to the printed circuit board component
level. There may be information provided in this manual that is not relevant for
your system. Do not service the Rad-8 pulse oximeter without completely reading
and understanding these instructions.
Fuse Rating (Qty 3): 1A, 250V (Approvals: UL 248-14, CSA C22.2 no. 248.14.
Compliance: RoHS).
CAUTION:
Federal law (U.S.) restricts this device to sale by or on the order of a physician
Masimo Corporation
40 Parker
Irvine, CA 92618
USA
Tel: 949-297-7000
Fax:949-297-7001
www.masimo.com
Authorized Representative:
MDSS GmbH
Burckhardstrasse 1
D-30163 Hanover
Germany
Tel: +49-511-62 62 86 30
Fax:+49-511-62 62 86 33
Made in USA.
© 2006 Masimo Corporation. All rights reserved.
Masimo, Masimo Logo, SET, LNOP, LNCS, Radical, Rad-5, Rad-5v, Rad-57,
Rad-8, Rad-9, SIQ, FastSat, FastStart Signal IQ and APOD are trademarks of
Masimo Corporation.
1. Overview ....................................................................................1-1
1.1. About this Manual ....................................................................1-1
1.2. Warnings, Cautions and Notes ................................................1-1
2. Maintenance...............................................................................2-1
2.1. Introduction ..............................................................................2-1
2.2. Cleaning...................................................................................2-1
5. Repair .........................................................................................5-6
5.1. Safety Precautions:..................................................................5-6
5.2. General Procedures.................................................................5-6
5.3. Troubleshooting .......................................................................5-6
5.4. Service Messages: ..................................................................5-8
5.5. Repair Policy............................................................................5-9
5.6. Return Procedure ....................................................................5-9
This manual explains how to service the Rad-8 class of pulse oximeters.
However, when a differentiation between models is required, the actual
model type will be identified prior to the explanation (For example, the
difference between a Rad-8H (Horizontal) or Rad-8V (Vertical).
Important safety information relating to general use of the Rad-8 class of
pulse oximeters appears in the Rad-8 Operator’s Manual. Other
important safety information is located throughout this manual where
appropriate.
Please read and follow any warnings, cautions and notes presented
throughout this manual. An explanation of these labels is as follows:
Sample of Warning:
Sample of Caution:
CAUTION: This is a sample of a Caution statement.
Sample of Note:
Note: This is a sample of a Note statement.
2.1. Introduction
2.1.1. This chapter covers, how to test the operation of the Rad-8
oximeters, how to properly clean, how to replace the battery
and how to obtain service under normal operation. No internal
adjustment or recalibration is required. Service should be done
by qualified personnel only. Safety checks should be
performed at regular intervals or in accordance with local and
governmental regulations.
2.2. Cleaning
2.2.1. To clean the display panel, use a cotton swab moistened with
70% isopropyl alcohol and gently wipe the panel.
2.2.2. To clean the outer surface of the oximeter, the following solutions may
be used to wipe the instrument for 30 seconds. Do not allow liquids
to enter the interior of the instrument.
■ Glutaraldehyde Solution
■ Ammonium Chloride Wipe
■ 10% Chlorine bleach in H2O
■ 70% Isopropyl alcohol
CAUTION:
• Do not autoclave, pressure sterilize, or gas sterilize this oximeter.
• Do not spray, soak, or immerse the monitor in any liquid.
• Use the cleaning solution sparingly. Allow unit to air dry. Excessive
solution can flow into the monitor and cause damage to internal
components.
• Do not touch, press, or rub the display panels with abrasive cleaning
compounds, instruments, brushes, rough-surface materials, or bring
them into contact with anything that could scratch the panel.
• Do not use petroleum-based or acetone solutions, or other harsh
solvents, to clean the oximeter. These substances attack the
device’s materials and device failure can result.
3.1. Overview
3.1.1. The Rad-8 oximeter uses a 6 volt, 1.4AHr lead acid, sealed,
rechargeable battery.
3.1.2. Please reference the Rad-8 Operator’s Manual for
specifications on battery charge time and capacity.
Note: Selecting the FastSat averaging button will not automatically set the
sensitivity mode to Max sensitivity.
5.3. Troubleshooting
5.3.1. The troubleshooting procedures in this manual instruct the
technician to isolate the failure down to the sub-assembly
level. The intention is to have the Unit Under Test (UUT)
repaired with minimal steps. The Masimo SET Tester is
needed to perform the Performance Verification in Section 4
upon completion of repairs.
Note: The Masimo SET Tester is not a calibrated device. It is to be used for
evaluation purposes only and not to determine if an oximeter is “calibrated”.
If unable to repair unit, then please call (800)326-4890 option #2 to arrange for
return to Masimo Corporation.
Masimo Corporation
40 Parker
Irvine, California
USA 92618