Guia Rapida Hd11
Guia Rapida Hd11
Guia Rapida Hd11
Copyright 2006 - 2013 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate,
in any form, this manual or any part thereof without permission from Welch Allyn.
Welch Allyn or its vendors assume no responsibility for any injury to anyone, or for any illegal or improper
use of the product, that may result from failure to use this product in accordance with the instructions,
cautions, warnings, or statement of intended use published in this manual.
Welch Allyn, PCH100, PCH200, HR100, HR300 and HR1200 are registered trademarks of Welch Allyn.
SD is a trademark of Toshiba.
Bluetooth is a trademark of Bluetooth SIG, Inc., USA.
Software in this product is Copyright 2006 - 2013 Welch Allyn or its vendors. All rights are reserved. The
software is protected by United States of America copyright laws and international treaty provisions
applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software
incorporated with this instrument as intended in the operation of the product in which it is embedded. The
software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to
human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and
ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, call Welch Allyn Technical Support:
USA +1 800 535 6663 Australia +61 2 9638 3000
+1 315 685 4560 +800 074 793
Canada +1 800 561 8797 China +86 21 6327 9631
European Call Center +353 46 90 67790 France +33 1 55 69 58 49
Germany +49 7477 9271 70 Japan +81 3 3219 0071
Latin America +1 305 669 9003 Netherlands +31 157 505 000
Singapore +65 6419 8100 South Africa +27 11 777 7555
United Kingdom +44 207 365 6780 Sweden +46 85 853 6551
Printed in USA
Directions for Use Table of Contents iii
Table of Contents
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Inspection upon Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Before you begin. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Optional Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Operator Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Holter System Application Directions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Activating the Welch Allyn Holter Software . . . . . . . . . . . . . . . . . . . . . . . . . 10
Recorder Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Patient Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Holter Procedure Patient Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preparing for Electrode Placement on Patient . . . . . . . . . . . . . . . . . . . . . . . 13
Starting the Welch Allyn Holter System Application. . . . . . . . . . . . . . . . . . . 19
Starting a Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Test Report Review, Editing and Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Physician’s Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Welch Allyn Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Cleaning the Recorder and Patient Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Inspecting the Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Testing the Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Storing the Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Discarding the Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Electromagnetic Emissions and Immunity Information . . . . . . . . . . . . . . . . 53
FCC information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Software License . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
iv Table of Contents Welch Allyn Holter System
Directions for Use 1
The Welch Allyn Holter System is comprised of the Welch Allyn Holter Recorder and the
Welch Allyn Holter System Application.
As the patient wears the recorder component of the system, it records ambulatory
electrocardiograph data. The Welch Allyn Holter System Application analyzes the recorder
data.
The Welch Allyn Holter System is not intended for infants weighing less than 10 Kg
(22 lbs).
The Welch Allyn Holter System acquires ambulatory ECG waveforms from patients.
The recorder and associated accessories provide signal acquisition for up to three
channels (HR100 and HR300) or up to eight channels (HR1200) of patient ECG waveforms
through surface electrodes adhered to the body.
• Acquire, view, store, and print ambulatory ECG waveforms from patients.
The recorder and associated accessories provide signal acquisition for up to 12 leads
of patient ECG waveforms through surface electrodes adhered to the patient.
• Using optional Holter algorithms to generate measurements, data presentations and
graphical presentations on an advisory basis for patients. These are presented for
review and interpretation by the clinician based upon knowledge of the patient, the
results of the physical examination, the ambulatory ECG data full disclosure displays,
and other clinical findings.
• Using optional interpretive algorithms to generate measurements, data
presentations, graphical presentations, and interpretive statements on an advisory
basis for patients. These are presented for review and interpretation by the clinician
based upon knowledge of the patient, the results of the physical examination, the
ambulatory ECG data full disclosure displays, and other clinical findings.
2 Directions for Use Welch Allyn Holter System
The Welch Allyn Holter System is intended for acquiring ambulatory ECG signals from
patients. Patients are people with coronary problems or people with suspected coronary
problems. This ambulatory electrocardiograph, and associated analysis system, can be
used on patients without any limitation on patient age or gender.
The Holter Recorder procedure is one of the many tools that clinicians use to capture
infrequent or activity provoked ECG rhythm abnormalities outside of the physician’s office.
Indications for conducting Holter recording are:
• Arrhythmias
• Chest pain
• Unexplained syncope
• Shortness of breath
• Palpitations
• Evaluation of a pacemaker
• Regulation of anti-arrhythmic drugs
• Evaluation of a patient after myocardial infarction
• Family history of heart disease
Contraindications
Computer assisted ECG data acquisition and interpretation is a valuable tool when used
properly. However, no automated interpretation is completely reliable and interpretations
should be reviewed by a qualified physician before treatment, or non-treatment, of any
patient.
The Welch Allyn Holter System is not intended for infants weighing less than 10 Kg
(22 lbs).
Warnings
A warning statement in this manual identifies a condition or practice which, if not
corrected or discontinued immediately, could lead to patient injury, illness, or death.
Cautions
A caution statement in this manual identifies a condition or practice which, if not
corrected or discontinued immediately, could lead to equipment failure, equipment
damage, or data loss.
Symbols
Single use. Use materials labeled with this symbol once only.
Recycling Symbol - Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or
separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European
Union on Waste Electronic and Electrical Equipment (WEEE). If this product is contaminated, this directive does not
apply. See www.welchallyn.com/weee.
Fragile, glass.
Serial Number.
Reference Number.
No Latex.
No PVC.
LED light.
Non-ionizing radiation.
6 Directions for Use Welch Allyn Holter System
Serial Number.
Directions for Use 7
Introduction
This manual is written for clinical professionals with a working knowledge of medical
procedures and terminology as required for monitoring cardiac patients. You must read
and understand this manual and all other information accompanying the ambulatory
electrocardiograph and related options or accessories before:
• using the Welch Allyn Holter Recorder and the accompanying Welch Allyn Holter
software for clinical applications
• before setting up, configuring, troubleshooting, or servicing the recorder
System Requirements
• 300MHz Pentium class processor or faster
• Windows XP (32-Bit), Windows Vista (32-Bit/64-Bit), Windows 7 (32/64-Bit) with all
major service pack updates, Windows 8 (32/64-Bit)
• 256Mb RAM minimum
• 40GB hard drive minimum
• Mouse and keyboard
• Compact Disc (CD) Reader
• Secure Digital Card Reader
• Printer with resolution greater than 300 dpi
• Monitor with screen resolution set to 1024 x 768
• 1Gb for sixteen 24-hour patient reports @ 200 samples per second with full disclosure
Optional Requirements
Requirements for backup
Caution Backup on Windows Vista or Windows 7 cannot be restored on
Windows XP.
Operator Requirements
Operator of the PCH100 and PCH200 Holter System application shall have the following
level of experience:
Features
All models:
• If you select No, a message displays reminding you that you must activate within
30 days. (Caution: after 30 days, the software will not launch.)
• If you select Yes, the system automatically activates itself and the activation screen
will not display again. This completes the activation requirement.
If the software cannot detect an Internet connection, you must activate the application
manually.
• If you select No, a message displays reminding you that you must activate within
30 days. (Caution: after 30 days, the software will not launch.)
• If you select Yes, the Manual Activation dialog displays. Click the Print button. Contact
Welch Allyn Technical Support (see page ii). You will be asked to provide the
information displayed on the printout:
• User name
• Password
• PC Serial Number 1
• PC Serial Number 2
3. Once you receive the Activation Code, type it into the Activation Code box and click
Activate. The application launches and the activation screen will not display again.
This completes the activation requirement.
Directions for Use 11
Recorder Illustrations
Figure 1. HR 100 Holter Recorder
Battery
Compartment
LED
Secure Digital
Card Interface
LCD
Patient-Activated
Event Buttons
Patient Cable
Caution Inserting the Secure Digital Card in backwards or up side down may
damage Secure Digital Card Interface.
12 Directions for Use Welch Allyn Holter System
Operation
To Set the Time (HR100)
Batteries must to stay in the recorder for a minimum of 10-minutes to charge the clock
backup cell.
1. Remove the battery cover from the recorder (see Figure 1 on page 11).
2. Remove the Secure Digital Card from the Secure Digital port located at the rear of the
device (see Figure 1 on page 11).
3. Insert one new, AA alkaline (LR 6) batteries into the recorder battery compartment.
Note the polarity.
4. The LCD window shows the time after the recorder self-diagnostics is complete
(approximately 10 seconds). Time is displayed in 24-hour time format and Date is
shown in Year Month Day format.
5. Enter time set mode by pressing the Top and Bottom Event switches simultaneously
for 2-seconds.,the hour digit starts blinking
6. Increase the value shown with the Top Event switch. Use the Bottom Event switch to
toggle between Hours, Minutes, Year, Month and Day.
7. Press the Top and Bottom Event switches to save time and date shown. Remove the
batteries.
1. Remove the battery cover from the recorder (see Figure 2 on page 11).
2. Remove the Secure Digital Card from the Secure Digital port located at the rear of the
device (see Figure 2 on page 11).
3. Insert two new, AA alkaline (LR 6) batteries into the recorder battery compartment.
Note the polarity.
4. Slide and secure the battery cover onto the recorder. The LCD window shows the
time after the recorder self-diagnostics is complete (approximately 10 seconds). Time
is displayed in 24-hour time format and Date is shown in Year Month Day format.
5. Press the <Right> function key, the hour digit starts blinking
6. Increase the hour shown with the up arrow key or decrease the hour shown with the
down arrow key. Use the right and left arrow keys to toggle between Hours, Minutes,
Seconds, Year, Month and Day.
7. Press the <Right> function key to save time and date shown. Remove the batteries.
Directions for Use 13
Patient Preparation
Holter Procedure Patient Preparation
When making the appointment, tell the patient:
WARNING Safety—Leave 5 feet (1.5 meters) of open area around the patient
during recorder hookup and removal.
WARNING Safety—Keep the recorder and patient cable clean, especially the
components that touch patients.
Caution Verify that dates on applicable accessories have not expired.
Caution ECG electrodes can cause skin irritation. Examine the skin for signs of
irritation or inflammation and avoid placement of the electrode in those areas. If
skin irritation occurs during the procedure advise the patient to remove the
electrodes and contact the health service provider as soon as possible.
Help the patient get comfortable. Patient preparation is important for a successful Holter
procedure.
2. Prepare electrode locations. See Figure 4 on page 15 for 5-lead placement, Figure 5
on page 16 for 7-lead placement, Figure 8 on page 18 for 12-lead placement.
7. Secure each lead wire. Loop each lead wire into a 1- to 2-inch diameter loop, position
it approximately two inches from each sensor site, and tape it to the skin (see
Figure 3 on page 14). This reduces movements that cause signal artifact.
Clinician tasks
RA LA
V1 V2
V3
V4 V5 V6
RL LL
Directions for Use 19
1. Click the Holter System Application icon. The Main Menu displays.
Starting a Test
Caution FAT file format recommended for SD Card sizes 2 GB and less.
6. If the patient has a previous recording, click Search. A patient list displays.
• Use the fields at the top of the patient list to filter your search.
• Select a patient.
9. Wait for approximately 20 seconds for the serial number of the recorder to appear in
the Recorders in Range window.
11. Select: 2 or 3 channel and 24 or 48 hour recording duration. Connect patient and click
Next.
12. The following message indicates that the previously recorded test on this SD card has
not been downloaded.
• If you want to retrieve the data, click No and go to “Retrieving data when Patient
Returns with Holter Recording Completed” on page 25.
13. Verify signal quality (i.e. amplitude and artifact).QRS signal amplitude should be two
boxes peak to peak. Select Finish. Test Startup is complete.
14. Place the recorder into the carrying case. Secure the carrying case and recorder on
the patient.
15. Write the start time, date, and patient data in the patient diary.
5. Enter patient information as desired. If this test has a previous recording, use search
to find the patient information. Select Next.
6. If the SD card has not been inserted into the reader, the window displays an
animation until the card is inserted.
8. Select Finish.
10. Insert a new battery (HR100) or batteries (HR300/1200). Note the polarity.
11. Slide and secure the battery cover onto the recorder.
• HR 100 — The LED flashes green in a sequence of two flashes, pause, three flashes,
pause, four flashes, pause, then flashes once approximately every three seconds
when operating. If the LED flashes yellow (any sequence) see “Troubleshooting” on
page 57.
• HR100, HR 300 & HR 1200 — The LCD window shows the time of day after the
recorder self-diagnostics are complete (approximately 10 seconds). The recorder is
collecting ECG data and functioning properly. After the self test, the recorder will
display ECG on the display for 5 minutes and then display the current time. If the
recorder does not display time or displays an error code, see “Troubleshooting” on
page 57
12. Place the recorder into the carrying case. Secure the carrying case and recorder on
the patient.
13. Write the start time, date, and patient data in the patient diary.
8. Enter a four digit Patient ID, press the top event button to increment value of Patient
ID field and press bottom event button to accept the value and proceed to the next
field..
9. Select the duration of 24-hours or 48-hours by pressing the top event button to toggle
between 24 & 48 and press the bottom event button to accept the value and proceed
to next field. The Duration selected is stored as the default for the next manual
initialization.
10. Select the number of channels 2-channel or 3-channel by pressing the top event
button to toggle between 2 & 3 and press the bottom event button to accept and
proceed to the next field. The Channels selected is stored as the default for the next
manual initialization.
11. When complete with setup, start ECG acquisition by pressing both event buttons
simultaneously for 2 seconds.
12. A progress indicator will display on the LCD.
13. Once acquisition starts Channel 1 (Ch1) will display on the LCD,
• Scroll through the channels 2 and 3 by pressing one of event buttons.
• ECG is displayed on the LCD for a period of 20-minutes then displays current time. To
exit ECG preview prior to the 20-minute default, press both event buttons
simultaneously for 2 seconds.
14. Place the recorder into the carrying case. Secure the carrying case and recorder on
the patient.
15. Write the start time, date, and patient data in the patient diary.
To Shut Down the Recorder (if the recorder does not stop automatically)
1. Press and hold both EVENT buttons and count the “beeps.” Release the buttons at
the start of the seventh beep. The HR100, HR300, HR1200 LCD will display a + sign
indicating successful shutdown. The HR100 green LED will stop flashing.
• If the recorder does not beep, the recorder was successfully shut down.
• If the recorder beeps, repeat the shutdown procedure.
• Remove the battery and SD card.
• Remove the leadwires and electrodes from the subject.
Directions for Use 25
3. Be sure the Holter software application is running and insert an SD card into the SD
card reader connected to the computer.
4. The Holter software application will upload and analyze the recorded test information
automatically. Monitor the progress in the lower right hand corner of your screen.
26 Directions for Use Welch Allyn Holter System
2. Highlight the desired test. Click Open. (If necessary, use the area immediately above
the header description to search by Patient ID, Last Name, or First name.)
Function Actions
Change Test Start Adjust test start time if necessary.
Physician Interpretation Provides a window for entering a report interpretation which is part of the report
cover page. The interpretation can be updated using acronyms (PCH200 only).
Patient Data Make changes/additions to the patient demographics as desired.
Print Print the report.
Settings Change Clinical settings and report format from default values for this test only.
Re-analyze Turn off channels that exhibit persistent quality problems and re-analyze
Note: Any previous edits are lost.
Summary Preview
The Summary tab provides a comprehensive overview of key episodes detected during
the recording.
The Preview Report Strips tab includes all the automatically-selected and user-inserted
ECG strips which will be printed as part of the report. This includes:
The report includes all the Full Disclosure strips inserted by the user that will be printed as
part of the report. The strips can be inserted in real time by marking an area in Full
Disclosure.
28 Directions for Use Welch Allyn Holter System
Forms
Forms categorize QRS morphologies into four groups, Normal (N), Ventricular (V), Paced
Beats (P) and Artifact (X), (the latter facilitates user review of false negatives).
Normal (N) forms will display SVE’s belonging to the form by selecting SVE Only.
Each form includes the category label, form number (assigned by the total number of
beats in a cluster across all categories) that match this morphology.
Selecting a form provides an accompanying 7-second ECG strip that includes the first
QRS from the form. By clicking Next, each QRS in the form can be seen in the context of
an ECG strip.
Right click the form header to re-categorize the form to the new desired category.
All changes will be updated when moving to another category.
Directions for Use 29
To view all beats in a particular form, double-click the form (PCH200 only).
30 Directions for Use Welch Allyn Holter System
Arrhythmia Review
The display for Pause episodes, Ventricular ectopy, and Supraventricular ectopy includes a
histogram of the episodes, a sample 7-second strip (of the worst case occurrence) and a
form associated with the highlighted QRS.
In addition for the Ventricular and Supra Ventricular ectopy a chronological histogram can
be viewed. The Rate tab displays Heart rate graphs for “Beat to Beat” or “HR Minute”.
The R-R Tab displays graphs for “R-R Beat to Beat” or “R-R Histogram”
View any of the events by selecting the respective bar on the histogram and using Next
to view ECG strips for all occurrences.
The individual beats and/or the form may be edited at any time.
Right-click any ECG strip to delete or add to the final report (all beats within the strip are
labeled as artifact or added).
Right-click any ECG strip to print (7-second diagnostic strip will be printed out to the
selected printer).
Directions for Use 31
The Rate tab provides Heart Rate information, specifically tachy, brady, min/max HR, and
min/max R to R intervals. The top graph includes a 24 hour HR plot with user selected
periods presented in 1-hour intervals immediately below.
Select any of the radio buttons to display a representative ECG strip. Or manually select
areas (for example, max HR) and manually insert into the report.
Full Disclosure
This selection allows you to review the entire recording. Abnormal events are color-coded
to make it easy to verify report accuracy.
32 Directions for Use Welch Allyn Holter System
Settings
Report Format screen options
Print format
To print the report immediately after the test data has been transferred from the recorder,
check Auto Print Report on Download.
To print narrative summary on the summary page check Show Narrative Summary, if this
is unchecked the summary information will be printed in a tabular format.
To sort the strips according to strip times check Strip Time Order Sorting.
Select Strips - Mini (8 strips per printed page) or Diagnostic (2 per printed page).
Strip setting
Facility
Set name and logo to be printed on the top left and right of the cover page.
Clinical Settings
See the Physician’s Guide for more information.
New Add new user (user first and last name). Limit 40 characters. Maximum of 20 users are
allowed. The system prompts when max number is reached.
User ID Limit 20 characters.
Edit Edit user selected in list.
Disable/Enable User Turn user access off or on for user selected in list.
Reset Password Reset user selected password for user selected in list. The new password generated will
be the same as your user name and requires change upon login.
Enable Administration Check box “unchecked” allows access without user login.
Directions for Use 35
36 Directions for Use Welch Allyn Holter System
Summarize settings
Caution Using report summarization deletes the full report and raw data.
To reduce file size, summarization converts a report (not including full disclosure) to a pdf
(4 to 6 MB typical). A list of summarized reports displays in the Inbox (PCH100) or the File
Cabinet (PCH200) with a summarized status.
• Customer Initiated – Summarize button available from the Inbox for the PCH100, and
from the filing cabinet in the PCH200.
• Check boxes allow user to configure which report pages to send and select strip
format
This feature exports a pdf of the final report to an Electronic Medical Records (EMR)
system.
• Test Type
• Patient ID
• Last Name
• Test ID
• Triggers
• Check boxes allow user to configure which report pages to send and select strip
format.
• Check box allow user to enable/disable Cardio perfect work station communications.
• Check boxes allow user to configure which report pages to send and select strip
format
• Send Test Email Message sends an email to a specified user. This can be used
to make sure email settings are working properly.
Directions for Use 39
The Email dialog box includes standard TO, CC, Subject, and Attachment fields.The
attachment: name for the attached PDF file automatically includes the patient ID.
For the PCH100, a Backup and Restore feature is provided through the System Settings.
The backup allows the database, and optionally the data files, to be copied into a separate
folder that is time stamped. The restore process can be used in the case of an error or a
corrupted database. The restore process will restore information from previous backup.
It is highly recommended that Backups are performed on a regular basis to prevent lost of
data. (Note: backups should be performed by those with suitable maintenance and
software knowledge).
40 Directions for Use Welch Allyn Holter System
To perform the backup, first select a directory that the files will be copied to (by default
the folder path is the ApplicationInstallFolder\Data\Backup where Application InstallFolder
is the folder location of the application). Database files will always be backed up but
optionally the “Backup Data Files” can be chosen to also copy all of the data files in the
system (Note: data files take up a large amount of space so it is important to choose a
sizable hard drive for this operation). Once all the options are selected hit the Backup
button and the lower “Status” region will show progress until the process has completed.
To Restore from a backed up database and/or data files begin by browsing to the folder
location chosen for the backup. After a folder is selected, the “Database” list will be
updated with all databases it finds in the folder. Select the correct database from the list
then choose whether to restore the database, the data files or both. Complete the
process by clicking the Restore button. Progress will be shown in the lower “Status”
region prompting the user with successes or failures. After the restoring a database,
close the window and the application will automatically shut down and will need to be
restarted for changes to take affect.
Directions for Use 41
Physician’s Guide
Overview
The Welch Allyn Holter System performs analysis up for to 48-hour of ECG digital data as
a continuous segment of data. The analysis processing consists of the following sub-
processing components:
• Signal Conditioning.
• QRS Detection and Feature Extraction.
• Clustering.
• Beat Classification.
• Heart Rate Calculation.
• ST Measurement.
• Pattern Determination.
• Strip Determination.
The analysis program consists of two major stages. The first stage, main analysis, is
responsible for QRS detection, classification, feature extraction, and clustering.
The second stage is responsible for pattern, tabulation, and strip determination.
This stage is commonly referred to as a re-compile process. During a full analysis, both
the main analysis and re-compile stages are executed. The full analysis will typically take
two minutes to analyze a 3-channel, 24-hour recording. The Re-compile typically takes
only seconds. The Re-Compile occurs multiple times, usually after the clinician has
performed edits. For instance, if the clinician changes the hookup time to a different hour
or minute, the recompile process would re-tabulate the results based on the new time.
Other edits that could trigger a re-compile include:
Signal Conditioning
The signal processing performed during the analysis is used to remove some of the noise
and artifacts normally found during an ambulatory recording. The following types of noise
and artifacts may occur:
The analysis program applies a collection of several filters to correct for these types of
issues. These filters are optimized for the specific sub-process requirements.
42 Directions for Use Welch Allyn Holter System
Clustering
The analysis program groups beats based on morphology related information.
The clusters created are based on which channels are active. For instance, a single
channel event in Ch 1 would only be grouped with other single channel events.
Beat Classification
The beat classification is performed by evaluating QRS detections, features, R-R intervals,
clustering information and noise assessment. The outcome of the classification results in
the following beat types:
• Normal
• Ventricular
• Supraventricular
• Artifact
The heart rate calculation is reported on each beat. The heart rate for a specific beat is
based on the average of 9 R-R intervals centered on the specific beat. This would require
the 4 previous R-R intervals, the current beat R-R interval and the next 4 R-R intervals.
A valid heart rate will require at least 5 out of 9 valid R-R intervals. If the heart rate cannot
be calculated, it will be assigned as an unknown heart rate.
The heart rate errors associated with one false positive detection (interpolated noise)
would be a +12.5% increase in reported heart rate. A false negative detection (missed
beat) would result in a -10% error in reported heart rate. The duration of time required to
correct the heart rate calculation would be 9 beats.
The HR is displayed for each strip and it represents the HR for the beat closest to the
center of the strip. If the clinician selects another beat in the strip, its HR values will be
displayed. If the HR cannot be calculated, it will be displayed as a --- value.
The HR displayed in Full Disclosure refers to the HR value found for the highlighted strip.
This HR value is based on the HR of the beat closest to the center of the strip. Refer to
Strip HR above.
ST Measurement
The ST Level is measured on normal beats with a width less than 110 msec. The ST level
is measured on each channel based on the delta between the isoelectric point and ST
Level point. The Isoelectric point is defined as an average of points between P and Q.
The ST Level point is a rate sensitive offset relative to the R-wave.
Directions for Use 43
ST Episodes are determined by summing the average ST level over a period of time.
Once the average ST Level exceeds the clinician selectable threshold, a ST Episode will
be started. The end point of the ST Episode is determined when the average level falls
below the clinician selectable threshold. A valid ST episode is reported when the length of
the ST episode exceeds the clinician selectable threshold. If a second ST episode starts
within the ST Reset threshold, the second ST episode is considered as part of the
previous episode.
The clinician has separate ST Elevation and ST Depression thresholds. During each ST
episode, the location of the maximum average values is maintained. The analysis program
does not identify ST slope levels. The clinician can review the ST strip examples to
determine if the ST segment slope is up, down or flat.
ST episodes are determined for each channel and for Elevation and Depression. These ST
episodes are sorted based on maximum level of depression or elevation and automatically
presented as strips within the report. The strip time reflects the time of maximum
depression or elevation. The duration of the ST episodes, ranges of displacement, or heart
rates are not reported.
Pattern Determination
The analysis program detects the following ventricular pattern types. N represents any
normal beat including S beats.
Supraventricular detection is based on normal beats only. The normal beat must be
premature to the previous normal beat or to the running NN average rate. The prematurity
threshold is configurable. The analysis program detects the following supraventricular
pattern types:
• Detect Atrial Fibrillation and Flutter. View Pause Review, Supraventricular Review,
Rate Graphs Review, and Full Disclosure as an aid to identify these types of
Arrhythmia patterns.
• Detect Ventricular Fibrillation and Flutter. View Arrhythmia Review and Full
Disclosure as an aid to identify these types of Arrhythmia patterns.
• Detect and classify any Intraventricular conduction defects. View Forms Review,
Arrhythmia Review, and Full Disclosure as an aid to identify these types of Ventricular
conduction defects.
• Identify Paced beat. A green pacer mark is provided by the program to identify the
presence of a pacer firing detected within the recorder. The clinician may reclassify
beat or forms as Paced.
Strip Determination
The analysis program selects automatic strip examples based on the following categories:
Report Content
The Analysis program provides results in several different formats. The application
provides a sample report (MIT 219) you may print to help you understand the report
content described below. Each report contains a header and footer that describes patient,
date and time of procedure, current clinical settings, and other data. Each report page is
sequentially numbered.
Summary Page
The Summary page contains a section with Patient detail information. This page contains
examples of the most severe arrhythmia categories. Pauses, Ventricular and
Supraventricular examples are presented as a mini strip format. Adjacent to each strip is
some numerical information relative to that pattern category. Both Ventricular and
Supraventricular Runs present numerical information for the Longest, Slowest and Fastest
Run. The Longest refers to the ventricular run with the largest number of ventricular
beats. Slowest and Fastest refer to Runs that contain the highest and lowest Ventricular
or Supraventricular Heart Rate. The Ventricular Heart Rate is based on the average Heart
Rate of all ventricular beats within the run. Supraventricular heart Rate is based on the
Heart Rate of all Supraventricular beats within the run.
Tabular
The tabular page(s) contains hourly information as well as procedure total related
information. It contains 25 hours to allow for a partial hour at the beginning and end of the
recording. The tabular contains hourly counts for:
Strips
The hard copy strips are presented in two different formats including diagnostic and mini
format. The diagnostic page contains two strips that are full scale. The Mini strips are
organized as 4 rows of two columns. Calibration marks are provided at the left side of the
strip. This calibration mark represents a 1-millivolt amplitude signal in the strip.
The clinician can select different scaling factors of x0.25, x0.5, x1.0, x2.0 and x4.0.
46 Directions for Use Welch Allyn Holter System
The calibration width represents a 100 msec marker. Diagnostic strips contain additional
tick marks that are 3 seconds apart. These markers are located at the top of each
diagnostic strip presentation. Each strip contains the time of day, day of the week, strip
type and heart rate at a minimum. For some types of strips, additional information is
presented such as Pause duration, ST Level and Run length with Run heart rate.
Forms
The Forms Page represents the form or clusters that were created. Each form indicates a
form count, form classification, and the number of beats in the form. A representative
example is shown for each form. Forms can be based on any combination of channels
processed.
Full Disclosure
The Full Disclosure can be printed out in several different formats of 10, 30 and
60 seconds per line. Channel indicators are displayed in the lower right footer to indicate
which channels are used.
Directions for Use 47
If the product fails to function properly—or if you need assistance, service, or spare
parts—contact the nearest Welch Allyn Technical Support Center. For phone numbers,
see page ii.
Before contacting Welch Allyn, try to duplicate the problem, and check all accessories to
ensure that they are not causing the problem. When calling, please be prepared to
provide:
• Product name and model number and complete description of the problem.
• Serial number of your product (if applicable).
• Complete name, address and phone number of your facility.
• For out-of-warranty repairs or spare parts orders, a purchase order (or credit card)
number.
• For parts orders, the required spare or replacement part numbers.
In case a return cannot be avoided, the representative will record all necessary
information and will provide a Return Material Authorization (RMA) number, as well as the
appropriate return address. An RMA number must be obtained prior to any return.
If you have to return goods for service, follow these recommended packing instructions:
• Remove all cables, sensors, and ancillary products (as appropriate) before packing,
unless you suspect they are associated with the problem.
• Wherever possible use the original shipping carton and packing materials.
• Include a packing list and the Welch Allyn Return Material Authorization (RMA)
number.
It is recommended that all returned goods be insured. Claims for loss or damage to the
product must be initiated by the sender.
48 Directions for Use Welch Allyn Holter System
Maintenance
The PCH100 and PCH200 Holter software requires no calibration.
Please note for the maintenance and cleaning instructions in the documentation for your
computer.
For maintenance of the computer on which the PCH100 or PCH200 Holter system is
installed, the following programs are regularly carried out:
WARNING Keep the recorder and patient cable clean, especially the
components that touch patients.
Caution Do NOT let soap or water come into contact with the battery contacts
or patient connector pins.
Cleaning Instructions
2. Wipe the Holter recorder exterior and accessories with a damp cloth. Use mild
detergent diluted in water.
5. Clean the Patient Cable in the same manner before each procedure.
1. Inspect the patient cable, patient lead wires, LED or LCD window, and enclosure.
1. Verify that the recorder is working properly, using an ECG simulator to acquire a
standard ECG of known amplitude (7-lead for HR100 and HR300, 12-lead for HR1200).
2. Upload the ECG data to the workstation and verify the waveforms appear normal,
with proper amplitude, and without distortion or excessive noise.
Please follow the state’s recycling laws or your facility’s recycling policy to ensure proper
disposal of the recorder and accessories. For more information on recycling, call the
Environment Protection Agency or local authorities.
Technical Specifications
Deviations from the range of specifications stated below may result in device
performance degradation.
Table 11. Holter Recorder Characteristics
Characteristic HR 100 HR 300 & HR 1200
Length 3.8 in (96.5 mm) 4.4 in (112 mm)
Width 2.2 in (56 mm) 3.1 in (78 mm)
Height 0.7 in (18 mm) 1.4 in (36 mm)
Weight with battery (or batteries) 7 oz. (198 g) 14 oz. (396 g)
and patient cable
50 Directions for Use Welch Allyn Holter System
• USA: UL60601-1
• EN 61000-4-2:1999
• EN 61000-4-3, 1995
• EN 61000-4-5, 1995
Accessories
To order accessories, call Welch Allyn Technical Support as listed on page ii.
RF emissions Group 2 The Welch Allyn Holter Recorder must emit electromagnetic energy in order to perform
CISPR 11 its intended function. Nearby electronic equipment may be affected.
RF emissions Class B The Welch Allyn Holter Recorder is suitable for use in all establishments, including
CISPR 11 domestic establishments. The recorder has no connection to the public low-voltage
power supply network.
Harmonic emissions Not applicable
IEC 61000-3-2
Voltage fluctuations/ Not applicable
flicker emissions
IEC 61000-3-3
Electromagnetic Immunity
The Holter recorder is intended for use in the electromagnetic environment specified below. The customer or user of the Holter recorder should
assure that it is used in such an environment.
Electrostatic discharge ± 6 kV contact ± 6 kV contact Floors should be wood, concrete, or ceramic tile. If floors are
(ESD) ± 8 kV air covered with synthetic material, the relative humidity should
IEC 61000-4-2 ± 8 kV air be at least 30%.
Electrical fast transient/ ±2 kV for power supply Not applicable.
burst lines
IEC 61000-4-4 ±1 kV for input/output
lines
Surge ±1 kV differential mode Not applicable.
IEC 61000-4-5 ±2 kV common mode
Voltage dips, short <5 % UT Not applicable.
interruptions, and (>95 % dip in UT)
voltage variations on for 0,5 cycle
power supply input 40 % UT
lines. (60 % dip in UT)
IEC 61000-4-11 for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Power frequency 3 A/m Not applicable.
(50/60Hz)
magnetic field
IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
54 Directions for Use Welch Allyn Holter System
Electromagnetic Immunity
The Holter recorder is intended for use in the electromagnetic environment specified below. The customer or user of the Holter recorder should
assure that it is used in such an environment.
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Holter
recorder is used exceeds the applicable RF compliance level above, the recorder should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the recorder.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Directions for Use 55
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Holter recorder
The Holter recorder is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
user of the Holter recorder can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Holter recorder as recommended below, according to the maximum output power of the
communications equipment.
Rated Max. Output 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Power of Transmitter
(W) d = (1.17) P d = (1.17) P d = (2.33) P
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
56 Directions for Use Welch Allyn Holter System
FCC information
The HR100/HR300 device contains FCC ID: ED9LMX9820ASM or FCC ID: POOWML-C40.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
conditions:
• This device must accept any interference received, including interference that may
cause undesired operation.
This equipment has been tested and found to comply with the limits pursuant to Part 15
of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses and
radiates radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communication. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment
does cause harmful interference to radio or television reception, which can be determined
by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
• Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
Troubleshooting
If the Holter System Application malfunctions, you may be able to resolve the issue with
little loss of time and expense. We suggest following the troubleshooting guidelines
below before returning the unit for servicing.
Cause Action
1 Electrodes positioned on a bony area. Reposition electrodes.
2 Date Code on Hook-up kit expired. Check the expiration date on the Apply new electrode.
Hook-up kit packaging.
3 Insufficient or dried electrode gel Apply new electrode.
4 Oily Skin or body lotions Clean Skin with alcohol or acetone
5 Excessive hair on chest Shave hair from chest prior to hookup
6 Poor skin preparation. Remove sensor(s), repeat skin prep, select new position (see
hookup procedure), and apply new electrodes.
7 Faulty patient cable Replace patient cable.
Software License
The software provided with the Welch Allyn Holter System consists of software written by
Welch Allyn, Inc. or its vendors (the Welch Allyn Holter System Software), as well as
software written by third parties (the Third-Party Software). The Welch Allyn Holter
System Software is governed by this license, and the Third-Party Software is governed by
licenses that can be found in the license files in the directories for the respective
Third-Party Software contained on the CD-ROM Disc that is furnished with the
Welch Allyn Holter System. If any updates are periodically provided for the Welch Allyn
Holter System Software or any Third-Party Software, such software updates will be
governed by the relevant license for that software unless a different license agreement is
provided with the update.
READ THE TERMS AND CONDITIONS OF THIS LICENSE AGREEMENT AND THE
LICENSES FOR THE THIRD-PARTY SOFTWARE CAREFULLY BEFORE OPERATING THE
Welch Allyn Holter System. THE Welch Allyn Holter System SOFTWARE AND
THIRD-PARTY SOFTWARE ARE COPYRIGHTED AND LICENSED (NOT SOLD).
BY OPERATING THE Welch Allyn Holter System YOU ARE ACCEPTING AND AGREEING
TO THE TERMS OF THIS LICENSE AGREEMENT AND THE LICENSES FOR THE
THIRD-PARTY SOFTWARE. IF YOU ARE NOT WILLING TO BE BOUND BY THE TERMS
OF THIS LICENSE AGREEMENT AND THE LICENSES FOR THE THIRD-PARTY
SOFTWARE, YOU SHOULD PROMPTLY RETURN THE Welch Allyn Holter System TO
WELCH ALLYN. THIS LICENSE AGREEMENT REPRESENTS THE COMPLETE
AGREEMENT CONCERNING THE Welch Allyn Holter System SOFTWARE BETWEEN
YOU AND WELCH ALLYN, AND IT SUPERSEDES ANY PRIOR PROPOSAL,
REPRESENTATION, OR UNDERSTANDING YOU MAY HAVE HAD.
END USER LICENSE AGREEMENT FOR Welch Allyn Holter System SOFTWARE
1. License Grant. Welch Allyn, Inc. (hereinafter Welch Allyn) hereby grants to you, and you
accept, a nonexplosive license to use the machine-readable, object code form of the
Welch Allyn Holter System Software, whether provided on the media packaged with the
Welch Allyn Holter System or pre-installed on the Welch Allyn Holter System (collectively
referred to as the “Software”), as well as the accompanying User Instructions or
Documentation, only as authorized in this License Agreement. The Software may be used
only on one computer. Neither concurrent use on two or more devices or computers, nor
use in a network, is permitted without separate authorization from Welch Allyn.
* You agree that you will not assign, re-license, transfer, rent, or share your rights under
this License Agreement, except that you may permanently transfer all of your rights under
this License Agreement as part of a sale or other transfer to a third party of the
Welch Allyn Holter System with which the Software was packaged and sold, provided that
you transfer all of the Software without retaining a copy, and the transferee agrees to the
terms of this License Agreement.
* You agree that you will not reverse assemble, reverse compile, or otherwise translate
the Software.
* You may use the Software media for backup purposes. No copies of the Software or any
portion thereof may be made by you or any person under your authority or control.
2. Welch Allyn's Rights. You acknowledge and agree that the Software and the User's
Instructions or Documentation are proprietary products of Welch Allyn, protected under
U.S. copyright law. You further acknowledge and agree that all ownership rights and title in
and to the Software, including associated intellectual property rights, are and shall remain
with Welch Allyn. This License Agreement does not convey to you any ownership in or to
Directions for Use 61
the Software, but only a limited right of use that is revocable in accordance with the terms
of this License Agreement.
3. License Fees. The Software is licensed to you in consideration of your purchase of the
Welch Allyn Holter System.
4. Term. This License Agreement is effective upon your operation of the Welch Allyn
Holter System and shall continue until terminated. You may terminate this License
Agreement by returning the Welch Allyn Holter System to the dealer from which you
purchased it within the time permitted for returns under your purchase agreement.
Welch Allyn may terminate this License Agreement if you violate any condition of, or
default in performing any obligation required by, the License Agreement. Upon
termination by Welch Allyn, you agree to return the Software to Welch Allyn.
5. Limited Warranty. Welch Allyn warrants, for your benefit alone, for a period of
12 months after the date of commencement of this License Agreement (referred to as
the “Warranty Period”) that during the Warranty Period the Welch Allyn Holter System
Software will operate substantially in accordance with the functional descriptions or
specifications in the User's Instructions or Documentation. Updates to the Welch Allyn
Holter System Software are covered by an Update Warranty Period of 90 days after
delivery of the Update to you. Welch Allyn does not warrant that the Welch Allyn Holter
System Software is free from defects, nor that it will operate error-free or produce results
that will meet your requirements. If during the Warranty Period, a defect in the
Welch Allyn Holter System Software appears, you may return it to Welch Allyn and at
Welch Allyn's sole discretion, it will either: 1) fix or replace the Welch Allyn Holter System
Software, or 2) refund to you the price you paid for the Welch Allyn Holter System. If a
defect in an Update of the Welch Allyn Holter System Software appears during the Update
Warranty Period, you may notify Welch Allyn, and Welch Allyn will use all commercially
reasonable efforts to fix or replace the Updated Welch Allyn Holter System Software.
You agree that the foregoing constitutes your sole and exclusive remedy against
Welch Allyn for the failure of any warranties made under this License Agreement.
EXCEPT FOR THE WARRANTIES SET FORTH ABOVE, THE SOFTWARE IS LICENSED
“AS IS,” AND WELCH ALLYN DISCLAIMS ANY AND ALL OTHER WARRANTIES,
WHETHER EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION, ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
6. Limitation of Liability. The cumulative liability of Welch Allyn to you or any other party
for any loss or damages resulting from any claims, demands, or actions arising out of or
relating to this License Agreement shall not exceed the price you paid for the Welch Allyn
Holter System. In no event shall Welch Allyn be liable for any indirect, incidental,
consequential, special, or exemplary damages or lost profits, even if Welch Allyn has been
advised of the possibility of such damages.
7. Governing Law. This License Agreement shall be construed and governed in accordance
with the laws of the State of New York, USA.
8. No Waiver. The failure of either party to enforce any rights under this License
Agreement or to take action against the other party in the event of any breach hereunder
shall not be deemed a waiver by that party as to the subsequent enforcement of such
rights.
62 Directions for Use Welch Allyn Holter System
63
Limited Warranty
Holter System Application, HR100, HR300 and HR1200 Holter Recorder and Accessories.
This product is sold by Welch Allyn under the warranties set forth in the following
paragraphs. These warranties are extended only to the end user with respect to the
purchase of this product directly from Welch Allyn or Welch Allyn’s authorized distributors
as new merchandise.
For a period of 1 year from the date of original delivery to the buyer, the recorder,
software, and hardware components are warranted to be free from functional defects in
materials and workmanship and to conform in all material respects to the description of
the product contained in the Directions For Use and accompanying labels and/or inserts.
For a period of 3 months this same warranty is made for accessories (including patient
cables) provided by Welch Allyn. Warranty of accessories purchased separately from
listed suppliers will be the responsibility of the listed suppliers.
This warranty is valid only if (a) all equipment is approved for use with the recorder by
Welch Allyn and are installed according to instructions provided by Welch Allyn or its
authorized distributors; (b) the product is properly operated under conditions of normal
use in accordance with applicable safety and regulatory requirements; (c) replacements
and repairs are made in accordance with the instructions provided by Welch Allyn; (d) only
recorder or other software authorized by Welch Allyn is used on the workstation; (e) the
product has not been configured, modified, adjusted or repaired other than by Welch Allyn
or by persons expressly authorized by Welch Allyn, or in accordance with written
instructions provided by Welch Allyn; (f) the product has not been subject to misuse,
negligence or accident.
Welch Allyn’s sole and exclusive obligation, and buyer’s sole and exclusive remedy under
the above warranties, is limited to repairing or replacing, free of charge, a product which is
reported to Welch Allyn customer service as listed on page ii. Welch Allyn shall not be
otherwise liable for any damages including, but not limited to, incidental, consequential,
or special damages.
There are no express or implied warranties which extend beyond the warranties in this
document. Welch Allyn makes no warranty of merchantability or fitness for a particular
purpose.
Welch Allyn’s sole and exclusive obligation, and buyer’s sole and exclusive remedy under
the above warranties, is limited to repairing or replacing, free of charge, a product which is
reported to Welch Allyn customer service as listed on page ii. Welch Allyn shall not be
otherwise liable for any damages including, but not limited to, incidental, consequential,
or special damages.
There are no express or implied warranties which extend beyond the warranties in this
document. Welch Allyn makes no warranty of merchantability or fitness for a particular
purpose.
64 Limited Warranty Welch Allyn Holter System