CP 100 12-Lead Resting Electrocardiograph: Directions For Use
CP 100 12-Lead Resting Electrocardiograph: Directions For Use
CP 100 12-Lead Resting Electrocardiograph: Directions For Use
Electrocardiograph
Copyright 2008, Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any
form, this manual or any part thereof without permission from Welch Allyn.
Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a
licensed physician.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that
may result from failure to use this product in accordance with the instructions, cautions, warnings, or
indications for use published in this manual.
Welch Allyn is a registered trademark of Welch Allyn, Inc., and CP 100 and CardioPerfect are trademarks of
Welch Allyn, Inc.
SD is a trademark of Toshiba.
Software in this product is copyright Welch Allyn or its vendors. All rights are reserved. The software is
protected by United States of America copyright laws and international treaty provisions applicable
worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated within
this instrument as intended in the operation of the product in which it is embedded. The software may not
be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable
form. This is not a sale of the software or any copy of the software; all right, title and ownership of the
software remains with Welch Allyn or its vendors.
For information about any Welch Allyn product, please call Welch Allyn Technical Support:
www.welchallyn.com
Printed in USA
iii
Contents
1 - Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Product Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Standard Features & Benefits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Controls, Indicators, and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
About the Main Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Moving Through the Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
About the Patient Cable and Leads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Using the Electrocardiograph Safely . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
General Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
General Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Getting Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7 - Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Problem-Solving Suggestions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
A - Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
1
1 Introduction
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Product Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Indications for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Standard Features & Benefits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Controls, Indicators, and Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
About the Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Moving Through the Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
About the Patient Cable and Leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Using the Electrocardiograph Safely . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Getting Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2 Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph
Before using the CP 100 electrocardiograph for clinical applications—or before setting up,
configuring, troubleshooting, or servicing the electrocardiograph—you must read and
understand this manual and all other information accompanying the electrocardiograph
and related options or accessories.
Product Overview
The Welch Allyn CP 100 electrocardiograph features a full alphanumeric keyboard, an LCD
display, full-size user-programmable reports, and the ability to operate on either battery or
AC power.
ECG tests sent to a memory card or removable USB storage device are compatible with
the Welch Allyn CardioPerfect™ workstation, which in turn can connect with other
electronic patient-information systems, such as billing and medical records.
Intended Use
The Welch Allyn electrocardiography and spirometry products (subject devices) are
intended for use by trained operators in health facilities. The subject devices provide the
following diagnostic functions:
• Acquiring and printing ECG waveforms using ECG front end modules (patient cables)
and associated accessories that provide signal acquisition for up to twelve (12) leads
of patient ECG waveforms through surface electrodes adhered to the body.
LCD display
Enter data and program the system easily.
Battery operation
Use the electrocardiograph almost anywhere. On battery power, you can print up to 100
ECGs continuously before needing to recharge.
Pacemaker detection
If the software detects the possible presence of a pacemaker, it asks you whether the
patient has a pacemaker. If you say no, interpretation (if purchased) is included in the
report. If you say yes, interpretation is not included in the report, and a message
indicating that a pacemaker was detected is displayed on the report.
Options
These options are available both for initial purchases and for upgrades.
• Carts
Two specially designed carts are available for convenient transport and use of the
electrocardiograph, as shown here with the optional cable arm and shelf.
Accessories
To order accessories, call Welch Allyn. For phone numbers, see page ii.
Carts
• Utility cart 08265-0000 1
• Office cart (Figure 1 on page 5) 401393 1
• Hospital cart (Figure 2 on page 5) 401394 1
• Cable arm & shelf option (page 5) 401161 1
Product information
Figure 3. Top
1! 2@ 3# 4$ 5% 6+ 7- 8* 9( 0)
Keyboard
See Figure 1 on page 9.
Figure 4. Back
AC power inlet
Equipotential stud
AC fuses
8 Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph
Figure 6. Front
Com port A
(for patient cable)
Table 1. Keyboard
A B C D
N
1! 2@ 3# 4$ 5% 6+ 7- 8* 9( 0) E
M
J I H G F
Key Function
D. Auto ECG Begins Auto ECGs, normal and stat. See “Recording an Auto ECG” on page 48.
E. Rhythm ECG Begins a Rhythm ECG. See “Recording a Rhythm ECG” on page 54.
H. Stop/Cancel Stops any current activity. See “Moving Through the Menus” on page 11.
J. Control (Applicable only for keyboards with characters above certain keys.) To enter a character
above a key, hold the Ctrl key while pressing that key. To capitalize the character, hold
Shift + Ctrl while pressing the key.
K. Shift Capitalizes letters. Also enters symbols (%, #, etc.). To enter a symbol, hold the
Shift key while pressing a number key.
9:17AM Oct 16 08
Main Menu
1 ECG Settings
2 System Settings
0 Exit
ECG Settings Review or change ECG settings: Auto See “Reviewing the ECG Settings” on
Report format, Rhythm Report format, and page 31.
so on.
System Settings Review or change system settings: device See “Reviewing the System Settings” on
configuration, device info, user setup, and page 25.
so on.
Directions for Use Chapter 1 Introduction 11
To perform an action
To accept data
To check or uncheck a checkbox
To return to parent menu from submenu on same (To select the highlighted submenu
screen (Figure 10) item.)
Chest leads
Left arm lead
Cable
rake Left leg lead
Electrocardiograph
connector
Symbols
The symbols illustrated on the following pages may appear on the electrocardiograph, on
the packaging, on the shipping container, or in this manual.
Documentation Symbols
Caution In this manual, indicates conditions or practices that could damage the
equipment or other property.
+49°C
Temperature limits Altitude limits
-20°C
Certification Symbols
Meets essential requirements of European Complies with applicable U.S. and Canadian
Medical Device Directive 93/42/EEC medical safety standards
Operation Symbols
For details on the keys, see Figure 1 on page 9.
Enter Shift
General Warnings
The following warning statements apply to electrocardiograph use in general. Warning
statements that apply specifically to particular procedures, such as connecting the patient
cable or performing an ECG test, appear in the corresponding sections of the manual.
Warning statements indicate conditions or practices that could lead to illness, injury, or
death.
WARNING To ensure patient and device safety, leave 5 feet (1.5 meters) of
open area around the patient.
WARNING All signal input and output (I/O) connectors are intended for
connection of only devices complying with IEC 60601-1, or other IEC standards
(for example, IEC 60950), as appropriate to the device. Connecting additional
devices to the electrocardiograph might increase chassis or patient leakage
currents. To maintain operator and patient safety, consider the requirements of
IEC 60601-1-1. Measure the leakage currents to confirm that no electric shock
hazard exists. In the case of a USB printer, the printer (non-medical electrical
equipment) shall be situated outside the patient environment (reference IEC
60601-1-1). The printer used should be approved to the appropriate safety
standard for non-medical electrical equipment (IEC 60950, or its national
variants), and use of an isolation transformer is recommended. If there is a
requirement for the printer to be situated within the patient environment it is the
responsibility of the user to ensure that the system provides a level of safety in
compliance with IEC 60601-1 and 60601-1-1.
WARNING The electrocardiograph has not been designed for use with high-
frequency (HF) surgical equipment and does not protect against hazards to the
patient.
16 Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph
WARNING Avoid positioning any leads or cables so that they could easily trip
someone or become wrapped around a patient’s neck.
General Cautions
The following caution statements apply to electrocardiograph use in general. Caution
statements that apply specifically to particular procedures, such as connecting the patient
cable or performing an ECG test, appear in the corresponding sections of the manual.
Caution statements indicate conditions or practices that could damage the equipment or
other property.
Caution To prevent possible damage to the keypad, do not use sharp or hard
objects to press keys. Only use fingertips.
Caution Do not pull or stretch the patient cable. Doing so could result in
mechanical or electrical failures. Form the patient cable into a loose loop before
storing.
Caution Avoid positioning the patient cable where it might get pinched or
stepped on. If the cable’s impedance is altered, measurements might no longer
be accurate, and repair might be necessary.
Caution Using the equipotential terminal for anything but grounding purposes
may contribute to damage of the device.
Caution Use only parts and accessories supplied with the device and available
through Welch Allyn. The use of accessories other than those specified may
result in degraded performance of this device.
Caution The power cord must be disconnected from AC power before cleaning,
maintaining, or servicing.
18 Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph
Getting Help
You can get help with the electrocardiograph in a variety of ways beyond this manual.
• Press the Help key from the Ready screen or Lead Status screen for a list of
topics available to print.
• Review the other information that came with the electrocardiograph. For list, see
“Product information” on page 6.
2. Verify that you have received all appropriate options and accessories. See “Options”
on page 5 and “Accessories” on page 6.
WARNING To avoid injury to the patient or damage to the device, never plug
patient leads into any other device or wall outlet.
1. Insert all leads into their proper positions, as labeled on the connectors.
Insert connectors fully so that no part of the metal ring remains exposed.
For example, see Figure 12. (To see the whole patient cable with all leads inserted,
see Figure 11 on page 12.)
WARNING Failure to insert all connectors fully may result in a loss of energy
being delivered to the patient during defibrillation and damage to the patient
cable itself. For other warnings related to defibrillation, see page 16.
2. Plug the patient cable into the port on the front of the electrocardiograph.
See Figure 13.
Figure 12. Inserting the Leads Figure 13. Plugging in the Connector
Metal ring
Directions for Use Chapter 2 Setting Up the Electrocardiograph 21
2. Remove the outer packaging, including the cardboard bottom, from a new pack of
paper. Pull the top sheet back so that the paper’s grid side faces up and the Welch
Allyn name is on the bottom of the paper.
4. Lay the top sheet over the paper door. Push the door to the right until it clicks.
See Figure 16.
Figure 14. Opening the Paper Door Figure 15. Loading the Paper
Paper tray
Latch
To Connect to AC Power
Plug one end of the power cord into the electrocardiograph’s AC power inlet. Plug the
other end into an AC outlet. The green LED on the keyboard lights up, indicating that
power is connected. See Figure 17.
When the charge gets low, the icon flashes. When the charge gets too low to operate, a
warning message appears and the electrocardiograph beeps every 15 seconds for 1
minute, then it turns off.
AC power inlet
On/Off key 1! 2@ 3# 4$ 5% 6+ 7- 8* 9( 0)
Green LED
Directions for Use Chapter 2 Setting Up the Electrocardiograph 23
Press .
Note As part of your initial set-up, you may want to adjust the display contrast. To learn
how, see “Reviewing the Device Configuration Settings” on page 27.
You may also want to change other software settings, as described in the
following chapters:
The external printer prints the following (always in black and white):
To Connect a Printer
Connect one end of a USB cable to the printer’s USB connector, and connect the other
end to the electrocardiograph’s USB connector: . For connector location, see Figure 5
on page 8.
To learn how to enable the external printer in the settings, see “Reviewing the Device
Configuration Settings” on page 27.
25
This chapter documents the system settings. For information on the following related
tasks, see the procedures identified here:
Device Device
Configuration Info
Setting Description
Language List of languages available. Changes take effect when the next screen appears.
Height Unit Centimeters (cm), inches (in), or feet and inches (ft, in).
Power-Save On or off. When on, the electrocardiograph turns itself off after several idle minutes.
Practice ID Name of the practice, clinic, or hospital. This ID prints on all ECG reports.
Audio Beeper On or off. When on, beeps to indicate errors, such as incorrect input, improper external
connections, or a printer error. Beeps may also indicate a low battery.
Increase Contrast Each time you select this choice, the display contrast immediately increases until you
reach maximum contrast.
Decrease Contrast Each time you select this choice, the display contrast immediately decreases until you
reach minimum contrast.
Contact Information Your office. Enter up to 40 characters. This information will appear on the accessories
Help page.
28 Chapter 3 Reviewing the System Settings Welch Allyn CP 100 Electrocardiograph
Printer Select Internal or External. When “External” is selected — and when an external printer is
connected and turned on — reports will print to that printer. For details, see “Connecting
an External USB Printer (Optional)” on page 24.
CAPS Lock On or off. Works just like a standard CAPS Lock key.
Directions for Use Chapter 3 Reviewing the System Settings 29
Item Description
• serial number
• modules configured
• version numbers
Enable Options Contact Technical Support. For phone numbers, see page ii.
Upgrade Software Contact Technical Support. For phone numbers, see page ii.
To Transfer a Configuration
1. Insert a storage device (SD memory card or removable USB device) into an
electrocardiograph that is configured as desired.
3. Choose System Settings > Device Info > Manage Settings > Export Files >
Export Configuration Files.
4. Remove the storage device, and insert it into another CP 100 electrocardiograph.
5. Choose System Settings > Device Info > Manage Settings > Import Files >
Import Configuration Files.
Configuration file
or
31
This chapter documents the ECG settings. For information on the following related tasks,
see the procedures identified here:
Interp Settings
Print Interpretation?
Copies
Copies With Interp
Reason Statements
Unconfirmed Report
Abnormal ECG
Patient Data
First Name
Middle Initial
Age/Birthdate
Weight
Height
Gender
Race
Medication
History
BP
Comments
Directions for Use Chapter 4 Reviewing the ECG Settings 33
E. 3 rows, 4 columns
F. Rhythm leads
G. Software version
H. Device ID
O. Calibration M Gain K. AC filter
pulse
I. Practice ID
N. Paper speed L. Frequency range J. Date and time
34 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph
A. Patient data See “Reviewing the Patient Data Fields Available” on page 37.
C. Interpretation (optional) See “Reviewing the Interpretation and Copy Settings for Auto Reports” on
page 36.
D. Report status label See “Reviewing the Interpretation and Copy Settings for Auto Reports” on
page 36.
E. 3 rows, 4 columns See “Reviewing the Format Settings for Auto Reports” on page 35.
F. Rhythm leads See “Reviewing the Format Settings for Auto Reports” on page 35.
G. Software version See also “Reviewing the Device Information” on page 29.
O. Calibration pulse Amplitude reference — represents the current height of a one-millivolt signal.
Is adjusted for the selected gain:
5 mm/mV = 0.5 x
10 mm/mV = 1 x
20 mm/mV = 2 x
Directions for Use Chapter 4 Reviewing the ECG Settings 35
1 Lead Arrangement
2 Rhythm Lead 1
3 Rhythm Lead 2
4 Rhythm Lead 3
5 Extended Measurements
6 Average Cycles
0 Previous Menu
Setting Description
Rhythm Lead 1 Rhythm lead to print at the bottom of all “+R” reports.
Rhythm Lead 2 Second rhythm lead to print at the bottom of 3x4 +3R reports.
Rhythm Lead 3 Third rhythm lead to print at the bottom of 3x4 +3R reports.
Extended On or off. When on, an additional page prints with the report. Extended measurements
Measurements include the values for several common parameters, such as Q, R, and S amplitude and
ST values. The amplitudes are expressed in microvolts. The durations are expressed in
milliseconds. The measurements cannot be edited.
Average Cycles If desired, an additional page prints with the report. Average cycles show the dominant
waveforms for all 12 leads.
2. Choose ECG Settings > Edit Auto Report > Interp Settings.
1 Print Interpretation?
2 Copies
3 Copies With Interp
4 Reason Statements
5 Unconfirmed Report
6 Abnormal ECG
0 Previous Menu
Setting Description
Print Interpretation? On or off. Determines whether interpretation is printed and saved with reports.
Copies With Interp On or off. Determines whether interpretation is printed on the automatic copies.
Reason Statements On or off. Determines whether reasons (criteria) are printed with the interpretation
statements.
Unconfirmed Report On or off. Determines whether the label “Unconfirmed Report” is printed on reports.
Abnormal ECG On or off. Determines whether the label “Abnormal ECG” is printed on reports.
Available only for systems using automatic interpretation.
Directions for Use Chapter 4 Reviewing the ECG Settings 37
2. Choose ECG Settings > Edit Auto Report > Patient Data.
Field Description
Age/Birthdate Birthdate, age, or off. Determines whether and how this data is labeled and entered. If
you’ll be testing patients from 1 day through 17 years old, select Birthdate for most
accurate results. If “age” is selected, age will be entered in years and months.
Weight Yes or no. If yes, this field is enabled for entering patients’ weight. For instructions on
changing the weight units (kg or lb), see “Reviewing the Device Configuration Settings”
on page 27.
Height Yes or no. If yes, this field is enabled for entering patients’ height. For instructions on
changing the height units (cm, in., or ft and in.), see “Reviewing the Device Configuration
Settings” on page 27.
Gender Yes or no. If yes, this field is enabled. Data-entry choices: Male, Female, or Unknown.
Race Yes or no. If yes, this field is enabled. Data-entry choices: Blank, Caucasian, Black,
Hispanic, Asian, Unknown.
Medication Yes or no. If yes, this field is enabled. During data entry, choose one item from the list of
patient medications.
History Yes or no. If yes, this field is enabled. During data entry, choose one item from the list of
patient clinical conditions.
38 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph
BP Yes or no. If yes, this field is enabled for entering blood pressure in standard ### / ###
format.
Comments Yes or no. If yes, this field is enabled for entering comments.
Directions for Use Chapter 4 Reviewing the ECG Settings 39
9:17AM Oct 16 08
Auto Report Settings
1 Baseline Centering
2 Lead Timing
3 Artifact Report
4 QTc Method
0 Previous Menu
Setting Description
Baseline Centering On or off. When on, aligns the isolectric line of all leads.
Lead Timing Simultaneous or sequential. “Simultaneous” prints ECG data that was captured
simultaneously for all lead groups. “Sequential” prints ECG data that was captured at
sequential intervals for each lead group in turn.
Artifact Report On or off. When on, the electrocardiograph automatically prints an artifact report with
the Auto ECG report whenever artifact is detected and you override the “Waiting for
10 seconds of quality data” message. This report, a 12x1 format with all filters
disabled, shows the leads with artifact so that you can remedy the problem. (The
normal Auto ECG report, with filters enabled, might mask some of the artifact issues.)
QTc Method Bazett or Hodges. Computation method of correcting the waveform’s QT interval
based on the heart rate. The corrected number, expressed in milliseconds, is called
the QTc interval. This setting affects only the QTc interval displayed in the ECG
measurements portion of the ECG report.
40 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph
9:17AM Oct 16 08
Miscellaneous These three default settings—gain, baseline
1 Lead Configuration
filter, and muscle filter—determine the
2 Electrode Labels values used every time you begin a new test,
3 Default Gain Setting even if these values have been temporarily
4 Default Baseline Filter changed during ECG testing.
5 Default Muscle Filter
6 Mains Filter
0 Previous Menu
Setting Description
Default Gain Setting 5 mm/mV, 10 mm/mV, 20 mm/mV, or Auto. (AUTO is available for Auto ECGs only, not
rhythm ECGs. AUTO is usually the best setting, but some waveforms may be easier to
read on other settings.) For details, see “Adjust Gain” on page 55.
Default Baseline Filter On or off. For details, see “Adjust Baseline Filter” on page 55.
Default Muscle Filter On or off. For details, see “Adjust Muscle Filter” on page 55.
Mains Filter Off, 50 Hz, 60 Hz. Use of this filter is recommended. For suggestions on eliminating AC
interference, see page 69.
Directions for Use Chapter 4 Reviewing the ECG Settings 41
When this option is turned on, the system will prompt you to select the lead set —
standard or augmented — each time you perform a test on a new patient.
42 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph
43
a. Describe the procedure. If desired, press the Help key and print the page
entitled “What Is An ECG?” for the patient to read.
b. Help the patient get warm and relaxed. Excessive patient movement could
interfere with the operation of the electrocardiograph.
c. Put the patient in a reclining position with the head slightly higher than the heart
and legs.
WARNING ECG electrodes could cause skin irritation. Examine the skin
for signs of irritation or inflammation.
a. Shave if necessary.
c. Allow to dry.
The electrode paste, gel, or creme must cover an area the size of the electrode
but no larger, especially on the chest.
Place the electrode tab between the “jaws” of the electrode adapter, keeping the
tab flat.
Gently tug on the adapter to ensure that it is properly placed on the electrode.
(Each time you remove and reattach an electrode, the conductive gel becomes
weaker and less effective.)
Directions for Use Chapter 5 Performing ECG Tests 45
J G
I H
Electrodes
AHA IEC Locations
I F
H G
Electrodes
AHA IEC Locations
A V1 C1 Fourth intercostal space at right sternal border.
red red
B V2 C2 Fourth intercostal space at left sternal border.
yellow yellow
C V4 C4 Fifth intercostal space at left of midclavicular line.
blue brown
D V7 C7 Posterior axillary line at the same horizontal level as V4.
orange black
E V6 C6 Mid-axillary line on same horizontal level as V4.
purple purple
F LA L Just above left wrist on inside of arm.
black yellow
G LL F Just above left ankle.
red green
H RL N Just above right ankle.
green black
I RA R Just above right wrist on inside of arm.
white red
J V3R C3R Mirror position to standard V3 placement with respect to the mid-
green green sternal line.
Directions for Use Chapter 5 Performing ECG Tests 47
6. If the Lead Status screen appears, as shown here, reattach any leads that are
flashing.
C1
C2
The term “Artifact” might also appear on this screen. To eliminate artifact, you might
need to re-prep the patient, use fresh electrodes, or minimize patient motion during
the recording.
The most common ECG problems are poor electrode contact and loose leads.
When all leads have been connected for several seconds, the following screen
appears.
9:17AM Oct 16 08
HR: 75
7. Go to the procedure for the type of ECG test you want to perform.
As shown in the following diagram, there are the two types of Auto ECG: normal and stat.
For details, see these procedures:
Normal Stat
The Auto ECG Acquisition screen appears. See Figure 33 on page 50.
50 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph
Printing Report
The electrocardiograph displays messages here
Collecting Data
when it is doing the following:
Gain: 10 mm/mV
• Collecting data
Speed: 25 mm/s • Analyzing data
Baseline Filter: ON • Printing the report
Muscle Filter: ON
If you override the wait time and print the available data immediately, be aware that
the printed data will be insufficient in quality or quantity or both.
Note If a red stripe appears along the edge of your report, replace the paper.
See “Loading the Thermal Chart Paper” on page 21.
After printing, the Auto ECG Post-Print screen appears. Figure 34 on page 51.
Directions for Use Chapter 5 Performing ECG Tests 51
1. Exit
2. Print Copy
3. Repeat Test
4. Export Test
Item Effect
Exit The Ready screen appears if all leads are connected to the patient.
To learn how to print multiple copies of all tests automatically, see “Reviewing the
Interpretation and Copy Settings for Auto Reports” on page 36.
4. Print ECG
Export Test Sends the test to the memory card. An SD memory card or removable USB storage device
must be in place.
52 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph
Stat mode bypasses patient data entry, as shown in Figure 31 on page 48. A temporary
ID number is assigned to the patient to identify stat tests.
The electrocardiograph begins acquiring ECG data. After it has acquired 10 seconds of
quality data, it prints a report.
Rhythm ECGs are printouts only. They cannot be sent to an SD memory card or
removable USB storage device.
(Optional) Adjust
Printing begins. or
waveforms.
The Ready screen appears if all leads are connected to the patient.
Directions for Use Chapter 5 Performing ECG Tests 55
Item Effect
Adjust Gain Cycles through the gain settings in mm/mV (5, 10, 20, AUTO), changing the waveform size.
(AUTO is available for Auto ECGs only, not rhythm ECGs. AUTO is usually the best setting, but
some waveforms may be easier to read on other settings.) To learn how to change the gain’s
default setting, see “Reviewing the Miscellaneous ECG Settings” on page 40.
Adjust Baseline Filter Toggles between the two baseline filter settings (on or off). This filter reduces “wandering
baseline,” an upward and downward fluctuation of the waveforms. It is preferable, if possible,
to eliminate or reduce wandering baseline by addressing the causes, as explained on page 68.
To learn how to change the filter’s default setting, see “Reviewing the Miscellaneous ECG
Settings” on page 40. Caution: You cannot perform ST segment analysis on waveforms that
were recorded with the muscle filter turned on. For details, see the caution on page 52.
Wandering baseline
Adjust Muscle Filter Toggles between the two muscle filter settings (on or off). This filter reduces muscle tremor
interference: random, irregular voltage superimposed on the waveforms. It is preferable,
if possible, to eliminate or reduce muscle tremor by addressing the causes, as explained
on page 68. To learn how to change the filter’s default setting, see “Reviewing the
Miscellaneous ECG Settings” on page 40. Caution: You cannot perform ST segment analysis on
waveforms that were recorded with the muscle filter turned on. For details, see the caution on
page 52.
Adjust Speed Cycles through the paper speed settings in mm/sec (10, 25, 50).
(available for Rhythm
ECG only) 10 mm/sec 25 mm/sec 50 mm/sec
56 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph
57
• Verify that patient leads are fully inserted. For details, see “Connecting the Patient
Cable” on page 20.
• Check for cracked or broken patient cable, patient leads, power cord,
communications cables, display, and enclosure.
• Check all cable and cord connections; reseat if any connectors are loose.
Caution Do not let soap or water come into contact with the
electrocardiograph’s internal printer, connectors, or jacks.
2. Wipe the exterior of the patient cable and electrocardiograph and with a damp cloth
using mild detergent diluted in water.
3. Use 70% isopropyl alcohol to disinfect patient cable, lead wires, and equipment.
5. Wait at least 10 minutes to allow all traces of alcohol to evaporate before turning the
electrocardiograph back on.
Directions for Use Chapter 6 Maintaining the Electrocardiograph 59
1. Verify continued electrical safety of the device, using IEC 60601-1 or ANSI/AAMI ES1
methods and limits. Test for the following:
• There should be no evidence of print-head dot failure (no printing breaks forming
horizontal streaks).
• Waveforms should appear normal, with proper amplitude, and without distortion
or excessive noise.
• Paper should stop with perforations near the tear bar, indicating proper cue-
sensor operation. For tear-bar location, see Figure 16 on page 21.
60 Chapter 6 Maintaining the Electrocardiograph Welch Allyn CP 100 Electrocardiograph
Note Regardless of the battery condition, you can use the electrocardiograph
whenever it is plugged in.
2. Verify that the green LED on the keyboard lights up. (See Figure 17 on page 22.)
If the LED does not light up, go to “Replacing the AC Fuses” on page 64.
The first time you turn the electrocardiograph on after a full battery discharge, you are
prompted to reenter the date and time.
If the electrocardiograph still does not turn on when it is unplugged from AC power, you
may need to replace the battery or battery fuse. See “Replacing the Battery” on page 61
or “Replacing the Battery (DC) Fuse” on page 63.
Directions for Use Chapter 6 Maintaining the Electrocardiograph 61
Figure 38. Removing the Battery Door Figure 39. Lifting Out the Battery
• If the “Z” wire is broken or dark, replace the fuse. See “Replacing the Battery
(DC) Fuse” on page 63.
“Z” wire
62 Chapter 6 Maintaining the Electrocardiograph Welch Allyn CP 100 Electrocardiograph
8. Carefully tuck in all cables. Place the battery into the battery compartment. Fold the
strap so that it will not stick out when you replace the battery door. See Figure 42.
9. Replace the battery door and screws. Turn the electrocardiograph rightside-up.
2. Use needle-nosed pliers to remove the fuse case. See Figure 44.
3. Inspect the fuses. If either fuse is dark or has a broken wire, replace the fuse.
See Figure 45.
4. Insert the fuse case. Line it up with the opening; it goes in only one way.
Do not dispose of this product as unsorted municipal waste. Prepare this product for
reuse or separate collection as specified by Directive 2002/96/EC of the European
Parliament and the Council of the European Union on Waste Electronic and Electrical
Equipment (WEEE). If this product is contaminated, this directive does not apply. For
more specific disposal information, see www.welchallyn.com/weee, or contact Welch
Allyn Customer Service at +44 207 365 6780.
66 Chapter 6 Maintaining the Electrocardiograph Welch Allyn CP 100 Electrocardiograph
67
7 Troubleshooting
Problem-Solving Suggestions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
68 Chapter 7 Troubleshooting Welch Allyn CP 100 Electrocardiograph
Problem-Solving Suggestions
This section includes several tables:
If you try these suggestions and still have problems, contact Technical Support. For phone
numbers, see page ii.
A dot is flashing on the Lead Status • Electrode contact may be poor. • Reattach the lead.
screen. • A lead may be loose. • Replace the electrode.
• Verify that the electrode area has been properly
OR Note: Square waves may indicate that the prepared: shaved, cleaned with alcohol or
electrocardiograph is inoperative, but they acetone, allowed to dry.
Lead-off information is displayed on the are more likely to indicate loose leads. • Verify that electrodes have been properly stored
screen. and handled.
OR
Wandering baseline (an upward and • Electrodes that are dirty, corroded, • Clean skin with alcohol or acetone.
downward fluctuation of the waveforms): loose, or positioned on a bony area. • Reposition or replace electrodes.
• Insufficient or dried electrode gel. • Help patient relax.
• Oily skin or body lotions. • If wandering baseline persists, turn the baseline
• Rising and falling of chest during rapid filter on. See “Adjust Baseline Filter” on
or apprehensive breathing. page 55.
Muscle tremor interference (random, • Patient is uncomfortable, tense, • Help patient get comfortable.
irregular voltage superimposed on the nervous. • Check all electrode contacts.
waveforms). May resemble or coincide • Patient is cold and shivering. • If interference persists, turn the muscle-tremor
with AC interference: • Exam bed is too narrow or short to filter on. See “Adjust Muscle Filter” on page 55.
comfortably support arms and legs. • If interference still persists, the problem is
• Arm or leg electrode straps are too probably electrical in nature. See the following
tight. suggestions for reducing AC interference.
Directions for Use Chapter 7 Troubleshooting 69
AC interference (even-peaked, regular • Electrodes that are dirty, corroded, • Check all electrode contacts and lead wires.
voltage superimposed on the waveforms). loose, or positioned on a bony area. • Verify that the patient is not touching any metal.
May resemble or coincide with muscle- • Insufficient or dried electrode gel. • Verify that the AC power cable is not touching
tremor interference. • Patient or technician touching an the patient lead cable.
electrode during recording. • Verify that the proper AC filter is selected. See
• Patient touching any metal parts of an “Mains Filter” on page 40.
exam table or bed. • If interference persists, unplug the
• Broken lead wire, patient cable, or electrocardiograph from AC power and run it on
power cord. the battery. If this solves the problem, you’ll
• Electrical devices in the immediate know that the noise was introduced through the
area, lighting, concealed wiring in power line.
walls or floors. • If interference still persists, the noise may be
• Improperly grounded electrical outlet. caused by other equipment in the room or by
• Incorrect AC filter frequency setting or poorly grounded power lines. Try moving to
AC filter is turned off. another room.
Won't turn on when plugged into AC • Faulty AC power connection. • Check the AC power source.
power. • Blown AC fuses. • Check the AC fuses. See “Replacing the AC
• No AC power. Fuses” on page 64.
Won't turn on when unplugged from AC • Battery disconnected or incorrectly • Check battery connections. See “Replacing the
power. connected. Battery” on page 61.
• Battery low, not charging, depleted, or • Recharge the battery. See “Recharging a Fully
bad. Discharged Battery” on page 60.
• Blown battery fuse. • Replace battery. See “Replacing the Battery” on
page 61.
• Replace battery fuse. See “Replacing the Battery
(DC) Fuse” on page 63.
Shuts down during printing Battery low or bad. • Recharge the battery. See “Recharging a Fully
Discharged Battery” on page 60.
• Replace battery. See “Replacing the Battery” on
page 61.
Prints fewer than 10 reports on a full Degraded battery. Replace battery. See “Replacing the Battery” on
battery charge. page 61.
Unresponsive for an extended time. System frozen. Press the reset button by inserting a small object,
such as a paper clip, into the hole.
Reset button
The screen is too dark or too light. The contrast is to high or too low. Press Ctrl-R. The contrast resets to its default value.
70 Chapter 7 Troubleshooting Welch Allyn CP 100 Electrocardiograph
“Insufficient battery power to begin. Battery low or bad. • Recharge the battery. See “Recharging a Fully
Please connect AC power and try again.” Discharged Battery” on page 60.
• Replace battery. See “Replacing the Battery” on
page 61.
“Insufficient space available” Not enough space on memory card • Delete some tests from the card or USB storage
or removable USB storage device. device at a PC.
• Use a different card or USB storage device.
“Memory card error” Problem writing to memory card • Verify that the write-protect tab is in the
or removable USB storage device. unprotected position.
• Reseat the card in its slot.
• Use a different card or USB storage device.
“Out Of Paper” • Printer is out of paper. • Load paper. See “Loading the Thermal Chart
• Printer door is open. Paper” on page 21.
• Close the printer door. See “Loading the Thermal
Chart Paper” on page 21.
“Paper Jam” Paper was loaded incorrectly. Reload the paper. See “Loading the Thermal Chart
Paper” on page 21.
“Powering down” Low battery. Recharge the battery. See “Recharging a Fully
Discharged Battery” on page 60.
“Problem loading the following settings: Problem loading your settings at startup. Contact Technical Support. For phone numbers, see
<System> May indicate memory problems. page ii.
<ECG>
*
*
*
Using default settings.”
“Temperature Error” Printer head temp is too high. Allow to cool, then try again.
Directions for Use Chapter 7 Troubleshooting 71
Limited Warranty
Welch Allyn, Inc., warrants that the Cardiopulmonary line of electrocardiographs, including
the CP 100 and CP 200 models (the Products) meet the labeled specifications of the
Products and will be free from defects in materials and workmanship that occur within 3
years after the date of purchase, except that accessories used with the Products are
warranted for 90 days after the date of purchase. Such accessories include: lead wires,
cabling, electrodes, and battery.
The date of purchase is: 1) the date specified in our records, if you purchased the Product
directly from us, 2) the date specified in the warranty registration card that we ask you to
send to us, or 3) if you don’t return the warranty registration card, 30 days after the date
on which the Product was sold to the dealer from whom you bought the Product, as
documented in our records.
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or
maintenance contrary to labeled instructions, 3) alteration or repair by anyone not
authorized by Welch Allyn, and 4) accidents.
You must obtain a return authorization from Welch Allyn to return your Product before you
send it to Welch Allyn’s designated service center for repair. Contact Welch Allyn
Technical Support. For phone numbers, see page ii.
Service Policy
All repairs on products under warranty must be performed or approved by Welch Allyn.
Unauthorized repairs will void the warranty. In addition, whether or not covered under
warranty, any product repair shall exclusively be performed by Welch Allyn certified
service personnel.
If the product fails to function properly—or if you need assistance, service, or spare
parts—contact the nearest Welch Allyn Technical Support Center. For phone numbers,
see page ii.
Before contacting Welch Allyn, try to duplicate the problem, and check all accessories to
ensure that they are not causing the problem. When calling, please be prepared to
provide:
• Product name and model number and complete description of the problem.
• Serial number of your product (if applicable).
• Complete name, address and phone number of your facility.
• For out-of-warranty repairs or spare parts orders, a purchase order (or credit card)
number.
• For parts orders, the required spare or replacement part numbers.
In case a return cannot be avoided, the representative will record all necessary
information and will provide a Return Material Authorization (RMA) number, as well as the
appropriate return address. An RMA number must be obtained prior to any return.
If you have to return goods for service, follow these recommended packing instructions:
• Remove all hoses, cables, sensors, power cords, and ancillary products (as
appropriate) before packing, unless you suspect they are associated with the
problem.
• Wherever possible use the original shipping carton and packing materials.
• Include a packing list and the Welch Allyn Return Material Authorization (RMA)
number.
It is recommended that all returned goods be insured. Claims for loss or damage to the
product must be initiated by the sender.
73
A Specifications
Item Specification
Dimensions 16.2 in
(41.1 cm)
6.2 in
(15.7 cm)
15.6 in
(39.7 cm)
Paper type 8.25 x 11 inches (21 x 28 cm), Z-fold thermal paper, 200 sheets
USB printer (external) Inkjet or laser printer that supports PCL (printer control language)
Gain settings:
Auto ECGs 5, 10, 20 mm/mV, AUTO
Rhythm ECGs 5, 10, 20 mm/mV
Fuses
AC Time-delay type, 2 amp 250 V rating, Littlefuse 0215002 or equivalent.
DC (battery fuse) Fast-acting type, 10 amp 32 V rating, Bussman ATC-10 or equivalent.
Standard connectivity Com port for USB cables or removable USB storage devices ( ≥ 64 MB, < 2 GB)
SD memory card slot (for use with cards ≥ 64 MB, < 2 GB)
Connectivity with electronic medical Supported through the Welch Allyn CardioPerfect workstation software
records
Electrodes Rigorously tested for conductivity, adhesion, and hypoallergenic qualities, and
exceed all AAMI standards
*Per AAMI EC11:1991/(R)2001 Diagnostic Electrocardiographic Devices, Section 3.1.2.1 Disclosure of cautionary
information/ performance characteristics paragraph c) Accuracy of input signal reproduction, the manufacturer shall
disclose the methods used to establish overall system error and frequency response. Welch Allyn has used methods A & D,
as prescribed in section 3.2.7.2 and 4.2.7.2 of this same standard, to verify overall system error and frequency response.
Because of the sampling characteristics and the asynchronism between sample rate and signal rate, digital ECG systems
such as the CP 100 and CP 200 may produce a noticeable modulating effect from one cycle to the next, particularly in
pediatric recordings. This phenomenon is not physiologic.
The CP 100 electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the CP 100
electrocardiograph should assure that it is used in such an environment.
RF emissions Group 1 The CP 100 electrocardiograph uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in nearby
CISPR 11 electronic equipment.
RF emissions Class A The CP 100 electrocardiograph is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage power supply network
CISPR 11 that supplies buildings used for domestic purposes.
IEC 61000-3-2
IEC 61000-3-3
76 Appendix B EMC Guidance and Manufacturer’s Declarations Welch Allyn CP 100 Electrocardiograph
The CP 100 electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the CP 100
electrocardiograph should assure that it is used in such an environment.
Electrostatic discharge ± 6 kV contact ± 6 kV contact Floors should be wood, concrete, or ceramic tile. If floors are
(ESD) covered with synthetic material, the relative humidity should
± 8 kV air ± 8 kV air be at least 30%.
IEC 61000-4-2
Electrical fast transient/ ±2 kV for power supply ±2 kV for power supply Mains power quality should be that of a typical commercial or
burst lines lines hospital environment.
Surge ±1 kV differential mode ±1 kV differential mode Mains power quality should be that of a typical commercial or
hospital environment.
IEC 61000-4-5 ±2 kV common mode ±2 kV common mode
Voltage dips, short >95% dip in 0.5 cycle >95% dip in 0.5 cycle Mains power quality should be that of a typical commercial or
interruptions, and hospital environment. If the user of the CP 100
voltage variations on 60% dip in 5 cycles 60% dip in 5 cycles electrocardiograph requires continued operation during power
power supply input mains interruptions, it is recommended that the CP 100
lines. 30% dip for 25 cycles 30% dip for 25 cycles electrocardiograph be powered from an uninterruptible power
supply or battery.
IEC 61000-4-11 >95% dip in 5 seconds >95% dip in 5 seconds
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at levels
(50/60Hz) characteristic of a typical location in a typical commercial or
magnetic field hospital environment.
IEC 61000-4-8
Directions for Use Appendix B EMC Guidance and Manufacturer’s Declarations 77
The CP 100 electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the CP 100
electrocardiograph should assure that it is used in such an environment.
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CP 100
electrocardiograph is used exceeds the applicable RF compliance level above, the electrocardiograph should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
electrocardiograph.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
78 Appendix B EMC Guidance and Manufacturer’s Declarations Welch Allyn CP 100 Electrocardiograph
Table 8. Recommended Separation Distances Between Portable and Mobile RF Communications Equipment
and the CP 100 Electrocardiograph
The CP 100 electrocardiograph is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or user of the CP 100 electrocardiograph can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the CP 100 electrocardiograph as recommended below, according to the
maximum output power of the communications equipment.
Rated Max. Output 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Power of Transmitter
(W) d = (1.17) P d = (1.17) P d = (2.33) P
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
79
Glossary
AHA. An abbreviation for American Heart Association. The AHA electrode labeling scheme is most
commonly used in the United States.
Auto ECG. A report typically showing a 10-second acquisition of 12 leads of ECG information combined
with patient data, interpretation, and measurements matrix. See also Rhythm ECG.
average cycles. Dominant waveforms for all 12 leads. Printed on a separate page, if enabled.
CardioPerfect workstation. A PC using Welch Allyn CardioPerfect software. Stores ECG and spirometry
test data. Can communicate with other electronic patient-information systems, such as billing and
medical records.
confirmed, unconfirmed. Status of an Auto ECG report’s interpretation. Indicates whether the report has
been reviewed and accepted by a qualified physician.
electrocardiogram (ECG). A record of the electrical currents associated with heart muscle activity.
Sometimes called EKG for the German term, Elektrokardiogramm. See also Auto ECG; Rhythm ECG.
extended measurements. Values for several common parameters, such as Q, R, and S amplitude and
ST values. The amplitudes are expressed in microvolts. The durations are expressed in milliseconds.
The measurements cannot be edited. Printed on a separate page, if enabled.
IEC. An abbreviation for International Electrotechnical Commission. The IEC electrode labeling scheme is
most commonly used in Europe.
lead. (1) An electrocardiograph wire connected to an electrode, which is attached to a patient’s skin. There
are 10 lead wires. (2) A waveform that represents ECG data from a particular view of the heart. The
data from the 10 lead wires is converted into 12 waveforms, also called leads: I, II, III, aVR, etc.
MEANS. An acronym for modular ECG analysis system. The optional MEANS interpretation algorithm,
developed by the University of Rotterdam, Netherlands, provides automatic analysis of ECG tests.
Also available in pediatric version: PEDMEANS.
Power-Save. A user-selectable feature that automatically turns the electrocardiograph off after several idle
minutes.
80 Glossary Welch Allyn CP 100 Electrocardiograph
Rhythm ECG. A continuous, real-time printout of ECG data. See also Auto ECG.
Index
3x4, 6x2, 12x1, etc. See leads: arrangements (formats) in C
Auto Reports
C1, C2, C3, etc. See IEC electrode labels
3x4 50 mm/s + 3R, 6x2 50 mm/s + 6R. See average cycles
Cabrera lead-grouping scheme, 40, 73
5 mm/mV, 10 mm/mV, etc. See gain settings
CAPS Lock, 28
50 or 60 Hz filter. See AC filter
CardioPerfect workstation
about, 2, 3
A carts, 5, 6, 15
caution symbol defined, 13
“Abnormal ECG” label, 36 CD (product information), ordering replacement, 6
"About" screen (device information), 29 cleaning procedures, 58
accessories, 6, 17, 71 clinical conditions. See history list
AC filter, 33, 34, 40, 69 Com ports, locations, 8
AC fuses, replacing, 64 configuration, transferring, 30
AC interference, 68, 69 configuration. See ECG settings; system settings
AC power. See powering the electrocardiograph confirming. See interpretation of ECGs, automatic: confirming
acquisition screen, 50 contact information, entering your, 27
AHA electrode labels (V1, V2, V3, etc.), 40, 45, 46 contact information for Welch Allyn, ii
arrow keys, using, 11 contrast, adjusting, 69
artifact report, 39 Control key (Ctrl), 9, 69
audio beeper, 27 copies, number to print, 36
augmented lead set, 4, 41, 46, 47 Customer Service, ii
"Auto ECG Acquisition" screen, 50
"Auto ECG Post-Print" screen, 51
Auto ECGs D
recording, 48–53
database. See CardioPerfect workstation
report settings, 33–38
data. See patient data
automatic interpretation. See interpretation of ECGs
date and time, 27, 33, 34
automatic shutdown. See Power-Save
defibrillation precautions, 16, 17, 20
average cycles, 35, 73
demographics. See patient data
device configuration settings, 27
B device ID, 27, 33, 34
"Device Info" screen, 29
baseline centering, 39 digital sampling rate, 73
baseline filter dust cover, ordering, 6
default, choosing (on or off), 40
enabling and disabling, 55
battery E
maintenance, 22, 60–63
ordering, 6 ECG acquisition screen, 50
ECG post-print screen, 51
specifications, 73
ECGs
Bazett QTc method, 39
performing tests, 43–55
beeper, 27
settings, 10, 31–41
See also Auto ECGs; Rhythm ECGs
82 Index Welch Allyn CP 100 Electrocardiograph
O S
on/off, manual and automatic, 23 safety, 15–17
options, 5 sampling rate, digital, 73
ordering information, 6 SD memory card. See memory card
sending tests to memory card, 51
sequential lead timing, 39
P serial number, displaying, 29
pacemakers, use with, 4, 17, 73 service policy, 72
paper. See thermal chart paper set-up. See hardware set-up; ECG settings; system settings
parts and accessories, 6, 17, 71 Shift key, 9
patient cable simultaneous lead timing, 39
connecting, 20, 44–47 specifications, 73–74
description & illustration, 12 speed settings. See thermal chart paper: speed settings
ordering, 6 stat Auto ECGs, 48, 53
precautions, 17 Stop/Cancel key, 9
See also electrodes; leads storage instructions, 65
patient data supplies, 6, 71
enabling and disabling fields, 37 symbols and markings, 13–14
entering, 49 system settings, 10, 25–29
as printed on reports, 33, 34
phone numbers, ii
poster, ordering replacement, 6
T
post-print screen, 51 Technical Support, ii
powering the electrocardiograph, 22–23, 73 testing the electrocardiograph, 59
Power-Save, 23, 27 thermal chart paper
practice ID, 27, 33 loading, 21
printer, external, 3, 24, 73 ordering, 6
printer, external (USB), 24, 28, 73 specifications, 73
printer, internal (thermal), 59, 73 speed settings, 33, 34, 55
"Print Settings" (printing your settings), 29 time and date. See date and time
print settings (report settings). See Auto ECGs: report settings; timing, lead, 39
Rhythm ECGs: report settings troubleshooting, 67–70
Q U
QTc method, 32, 39 ultrasound machines, use with, 17
“Unconfirmed” label, 33, 34, 36
See also interpretation of ECGs, automatic: confirming
R upgrades, 5, 29
"Ready" screen, 47 URL, ii
reason statements, 36 USB port (Com port B), 8
recycling instructions, 65 USB storage devices, 3, 74
repairs, 72
replacement parts and accessories, 6, 71
reports. See Auto ECGs; Rhythm ECGs
V
reset button, 8, 69 V1, V2, V3, etc. See AHA electrode labels
Rhythm ECGs versions, 29, 33, 34
recording, 54–55
84 Index Welch Allyn CP 100 Electrocardiograph
W
wall poster, ordering replacement, 6
warning symbol defined, 13
warranty, 71
waveforms, ECG. See leads
web site, ii
weight-unit setting, 27
Welch Allyn Technical Support, ii
Welch cups, ordering, 6
workstation. See CardioPerfect workstation