OPERATOR’S MANUAL
for the
TDA-2G, 2GN
Aerosol Photometer
www.atitest.com
info@atitest.com
T2G0-0877 1 Revision B
Model TDA-2G/2GN
T2G0-0877 2 Revision B
Model TDA-2G/2GN
Table of Contents
1 GENERAL INFORMATION 6
1.1 Description 6
1.2 Sampling System 6
1.2.1 Scattering Chamber (LSC) 6
1.2.2 Electronics 7
1.2.3 Theory of Operation 7
1.2.4 Filter Leak Testing: The Most Common Application 7
1.2.5 Capabilities 8
1.2.6 Input/Output features 8
2 INSTALLATION 9
2.1 Requirements 9
2.2 Unpacking 9
3 SETUP 10
3.1 Locations & Functions 10
3.2 Front Panel Controls and Indicators 11
3.3 Front Panel Connections 11
3.4 Rear Panel Connections 12
4 OPERATION 13
4.1 Initialization 13
4.2 Setup Parameters 13
4.3 Changing Setup Parameters 14
4.4 Preparing to test 17
4.5 Testing with the TDA-2G 20
5 MAINTENANCE 23
5.1 Weekly 23
5.2 Annually 23
5.3 General Maintenance procedures 23
5.4 Recommended Spare Parts 23
5.5 Cleaning the TDA-2SP Scanning Probe Screens 24
T2G0-0877 3 Revision B
Model TDA-2G/2GN
A-1. SPECIFICATIONS 25
A-2. TROUBLESHOOTING 26
A-3. QUICK SETUP REFERENCE 27
A-4. ACCESSORY LIST 29
A-5. WARRANTY 30
A-6. RS-232 COMMUNICATION 31
A-7. AEROSOL CORRECTION FACTORS 32
T2G0-0877 4 Revision B
Model TDA-2G/2GN
Guidelines to the use of this manual
SYMBOLS
The following symbols are used throughout the manual to draw attention to items
or procedures that require special notice or care.
Operator required action.
Conventions
[Window] Indicates the information displayed in a window on the Control Panel.
<Function> indicates a button on the Control Panel.
T2G0-0877 5 Revision B
Model TDA-2G/2GN
1 GENERAL INFORMATION
1.1 DESCRIPTION
The TDA-2G is a forward light-scattering, linear photometer. It operates on 90 to
240 volts, 50 or 60 Hz, adjusting automatically. The basic functions of the TDA-
2G are to sample air or other gas and report the concentration of particulates in
the sample.
This unit is also available in a nuclear version, the TDA-2GN. The nuclear
version is the same size and has the same features as the TDA-2G, plus a
sealed sampling chamber and an additional ULPA exhaust filter to contain all
contamination within the unit.
The TDA-2G is compact and lightweight, measuring 13.5 x 9.5 x 5.0 inches (34.3
x 24.0 x 12.7 cm) and weighing only 15.5 lbs (7.0 kg). The instrument case is
constructed of die cast aluminum and has retractable carrying handles. Fold-up
legs under the case tilt the unit up for easy viewing. The pressure-sensitive
keypad and large, bright LED displays and indicators provide ease of operation
and readability. The auto-ranging and one-step zeroing features assure the
accuracy of all readings.
A selector valve on the front panel directs the airflow through the sampling
system to the scattering chamber from three possible sources. The CLEAR
position directs clean air from an internal ULPA filter to the scattering chamber
for zeroing the instrument. The UPSTREAM position permits sampling of the air
above the filter being challenged, and the DOWNSTREAM position permits
sampling of the air that penetrates the filter.
T2G0-0877 6 Revision B
Model TDA-2G/2GN
1.2.2 Electronics
The signal from the photomultiplier is amplified and delivered to a signal-
conditioning, analog-to-digital converter that is then sent to the microprocessor.
The test aerosol should be introduced into the system on the upstream side of
the filter or filters as far from the filters as is practical to insure adequate mixing
with the airflow in the duct. A commonly used guideline that assures adequate
mixing is ten duct diameters upstream of the filter. A sample of the aerosol/air
mixture should be taken from the upstream side, close to the filters under test to
verify the necessary upstream challenge. This sample is also used to set the
100% baseline. The 0% baseline is then adjusted using particle-free air supplied
by the internal ULPA filter. The instrument is now ready to detect and quantify
leaks on the downstream side of the filter.
T2G0-0877 7 Revision B
Model TDA-2G/2GN
should be made around the periphery of the filter, along the bond between the
filter pack and the frame, and around the seal of the filter. The display indicates
the percent of leakage through or around the filter. The scanning probe is
supplied with 3 types of nozzles that can be screwed onto the end of the flexible
probe. The round, black nozzle is 1 inch (25 mm) in diameter, which complies
with NSF (National Sanitation Foundation) Standard 49-1992. The round, red
nozzle is an isokinetic nozzle. The rectangular, blue isokinetic nozzle is used for
faster scanning and is accepted by many standards, including NSF 49-2002.
The isokinetic nozzles are designed for face velocities of 90 +/- 20 feet per
minute (fpm).
1.2.5 Capabilities
NOTE: Before attempting to operate this unit, become familiar with the
features and functions.
Measurements
• % Penetration
• Absolute aerosol concentration
T2G0-0877 8 Revision B
Model TDA-2G/2GN
2 INSTALLATION
2.1 REQUIREMENTS
2.2 UNPACKING
Carefully unpack and remove the TDA-2G Aerosol Photometer and all
accessories from its shipping container. If the instrument has been damaged in
transit, notify the shipper immediately.
The following items are optional. Check the order packing slip.
1) Shipping case, 9300102
2) TDA-2SP Scanning Probe, complete, T2SP-0879
a) 1 round isokinetic nozzle (red), T2E0-0572
b) 1 rectangular isokinetic nozzle (blue), T2E0-0798
c) 1 NSF 1” round nozzle (black), T2E0-0005
d) 1 Scanning probe, T2SP-0881
If any of these items are on the order packing slip but missing from the shipment,
contact ATI immediately at:
Air Techniques International
Division of Hamilton Associates
11403 Cronridge Drive
Owings Mills, Maryland 21117-2247 U.S.A.
Tel: 410 363-9696 Fax: 410 363-9695
www.atitest.com
T2G0-0877 9 Revision B
Model TDA-2G/2GN
3 SETUP
2 7 6 8
T2G0-0877 10 Revision B
Model TDA-2G/2GN
Front Panel Display – Indicates % leakage readings and error messages.
Function Keys – Used for setting operating parameters and initiating program
routines
Bar Graph Display – Displays an analog representation of the % leakage to aid
in isolating leaks.
The front panel contains seven pressure-activated function keys. The <0>,
<100>, <REF>, and <ALARM> keys each contain a red LED to indicate the state
∆> and <∇
of the switch or to prompt the operator. The <∆ ∇> keys are used to scroll
between selections as they are shown on the % LEAKAGE display. The
<ENTER> key serves as the command key for sending information to the
processor and for initiating or stopping system routines. The <∆ ∆>, <∇
∇>, and
<ENTER> keys contain no indicator LED.
The Scanning Probe contains an unlabeled display on the pistol grip that is a
half-scale duplicate of the % LEAKAGE Indicator. The two displays are driven
by the same electronics and will always read exactly the same.
The Power Switch is located in the lower right corner of the front panel. The 2-
position rocker switch is unlabeled, containing only the symbols 0 (Off) and l
(On).
The Selector Valve, the Scanning Probe connector, and the two barbed fittings
are discussed later.
If using the TDA-2G with the Scanning Probe, connect the probe’s electrical
connector to the 12-pin connector on the front panel of the TDA-2G before
applying power to the photometer. Connect the probe’s sampling hose to the
barbed DOWNSTREAM fitting to the right of the 12-pin connector. The cable
and sampling hose are bound as a single, flexible umbilical. Select the
appropriate probe nozzle and screw it onto the threaded end of the flexible
portion of the Scanning Probe.
Note: When using the TDA-2SP Scanning Probe with the TDA-2G,
the probe must be connected to the photometer before the power is
turned on. Otherwise the TDA-2G will not detect the probe and no
T2G0-0877 11 Revision B
Model TDA-2G/2GN
display will be observed on the pistol grip. If this occurs, switch the
power off, verify that the probe is connected properly, and then
switch the power back on.
2 1 6 4 5 3
T2G0-0877 12 Revision B
Model TDA-2G/2GN
The rear panel contains a recessed, 3-pronged male Power Connector, a 9-pin
female D-subminiature connector for an RS-232 computer interface, and a
barbed fitting for the Vacuum Pump Exhaust.
The RS-232 Port provides a serial output to a computer that can be read and
stored using any standard communications package such as Windows Hyper-
terminal. Information is sent out the RS-232 Port each time the Front Panel
Display is refreshed, approximately once per second.
The Power Connector accepts the female end of the Power Cord that comes with
the unit. The internal power supply is a universal supply that can operate at
either 50 or 60 Hz and with any voltage from 90 to 240 volts, for use in virtually
all countries.
The barbed fitting is the exhaust from the unit’s vacuum pump. This fitting is
provided so that the sensitive environments of cleanrooms can be maintained.
The air sampled by the unit can be sent to either an external in-line filter or to an
exhaust system.
4 OPERATION
4.1 INITIALIZATION
If the Scanning Probe is being used, connect it to the Scanning Probe Connector
and the Downstream Port. Verify that the Selector Valve is in the CLEAR
position. Apply power to the TDA-2G by setting the Power Switch to the 1
position (On). The bar graph scans for 20 seconds, then all the display
segments light. Verify that all Front Panel Display segments light fully with a full
bar graph beneath (llllllll). This display must occur before the photometer is
ready to use.
Parameters
T2G0-0877 13 Revision B
Model TDA-2G/2GN
L5 – Aerosol Reference Selection (P1=DOP, P2=PAO Oil or P3=Factor of P1)
L6 – Hour Meter (Total operating time)
L7 – Load Factory Defaults
L8 – Software Revision
L9 – Enable/Disable bar graph display
L10 – Display Intensity Adjustment
Enables or disables the audible portion of the TDA-2G alarms. The unit parameter is factory set
to off.
When L1 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting. A – appearing in the % Leakage display indicates that this parameter is currently
disabled. A 1 indicates an active parameter.
Press the <∆> (Up) and <∇> (Down) function keys to change the displayed setting between 0
(disabled) and 1 (enabled).
T2G0-0877 14 Revision B
Model TDA-2G/2GN
Press <ENTER> to set the selection and return to the parameter select menu (L1 will be
displayed).
Default: 7
Establishes the relative frequency at which the display is refreshed (updated). This frequency is
relative to the cycling of the system microprocessor and is not time dependent.
Note: The system is continuously sampling, but for the operator’s convenience, only
the highest “% Leakage” is displayed at each refresh cycle.
When L2 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <∆> (Up) and <∇> (Down) function keys to move the displayed setting between 1 and
20 on the display (1=fastest, 20=slowest).
Press <ENTER> to set the selection and return to the parameter select menu (L2 will be
displayed).
Options: 5-120
Default: 10
Establishes the period over which the 100% aerosol sample is averaged. This frequency is
relative to the cycling of the system microprocessor and is not time dependent.
When L3 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <∆> (Up) and <∇> (Down) function keys to move the displayed setting between 1 and
20 on the display (5=shortest, 120=longest).
Press <ENTER> to set the selection and return to the parameter select menu (L3 will be
displayed).
Default: 3
This parameter is used to select either 3 or 4 decimal places on the % Leakage Display.
When L4 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <∆> (Up) and <∇> (Down) function keys to move the displayed setting between 3 and 4
on the display (3=0.001%, 4=0.0001%).
T2G0-0877 15 Revision B
Model TDA-2G/2GN
Press <ENTER> to set the selection and return to the parameter select menu (L4 will be
displayed).
Default: P1
Establishes the internal reference base response based upon a gain setting stored during factory
calibration. Selecting P1 with a reference setting of 100 will give a % Leakage reading of 100%
when the unit samples 100 ug/l of aerosol concentration.
When L5 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <∆> (Up) and <∇> (Down) function keys to move the displayed setting between P1, P2
or P3.
Press <ENTER> to set the selection and return to the parameter select menu (L5 will be
displayed).
L6 – Hour meter
Options: None
Default: None
When L6 is displayed in the parameter menu, press the <ENTER> function key to access the
hour value.
Press <ENTER> to set the selection and return to the parameter select menu (L6 will be
displayed).
Options: (-) Not enabled-Indicating factory default loading has not been selected
(1) Enabled-Indicating factory default loading has been selected
L7 is used to restore the factory default settings to all the user programmable parameters.
When L7 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <∆> (Up) and <∇> (Down) function keys to move the displayed setting between – and
1.
T2G0-0877 16 Revision B
Model TDA-2G/2GN
Press <ENTER> to set the selection and return to the parameter select menu (L7 will be
displayed).
L8 – Software revision
Options: N/A
Default: None
L8 displays the current operating revision that is being used by the TDA-2G.
L9 – Bargraph display
Options: 0-Disabled
1-Enabled
Default: 1-Enabled
L9 is used to enable or disable the analog bargraph display beneath the numeric portion of the %
Leakage display.
When L9 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <∆> (Up) and <∇> (Down) function keys to move the displayed setting between 0 and 1
on the display.
Press <ENTER> to set the selection and return to the parameter select menu (L9 will be
displayed).
Default: 5
L10 is used to adjust the intensity of the green LED display in the % Leakage display.
When L10 is displayed in the parameter menu, press the <ENTER> function key to access the
stored setting.
Press the <∆> (Up) and <∇> (Down) function keys to move the displayed setting between 1 and 8
on the display.
Press <ENTER> to set the selection and return to the parameter select menu (L10 will be
displayed).
Once the initialization & warm-up cycle has been completed, the TDA-2G is
ready for operation. The unit should be placed on a flat, stable surface with the
area to be monitored within reach of the sample lines.
T2G0-0877 17 Revision B
Model TDA-2G/2GN
Caution: If the power to the unit is interrupted between
initialization and operation (for example, to relocate the
unit), the selected set points and the reference value will
remain in memory. Gain settings from a 100% baseline
definition, however, will be erased from memory.
(If the Internal Reference is not being used, proceed to step 5.)
1. Press the <ENTER> function key.
2. Immediately press the <REF> (internal reference) key (within 3
seconds). The % LEAKAGE display will briefly flash the internal
reference option in use ([P1], [P2], or [P3], followed by the value of
the internal reference . (For example, [P1] followed by [100.0] would
be the manufacturer’s internal reference setting for DOP (P1) at 100
micrograms per liter.)
∆> or <∇
If a different internal reference setting is required, press <∆ ∇> function
keys to raise or lower the value until the desired value is reached.
3. Press <ENTER>. The system will perform its internal gain
adjustment for the new value. The bar graph sweeps from left to
right until the task is completed, approximately 15 to 30 seconds.
When completed, the red LED over the <0> key will begin flashing.
4. Press <ENTER> to automatically zero the instrument. This takes
approximately 5 seconds and is accompanied by short beeps. When
zeroing is completed, the instrument gives a prolonged beep.
CAUTION: Do not turn the valve from CLEAR until the beeps stop.
Turn the Selector Valve to the UPSTREAM position. This switches the sample
airflow to the UPSTREAM sampling port on the front panel. The % LEAKAGE
display indicates the upstream reading. If the Internal Reference has been used
to normalize the instrument response to 100 ugm/l, the upstream concentration in
micrograms per liter is reported. (If the internal reference has not been
established, the display will read 8.8.8.8.8, proceed to step 6.)
T2G0-0877 18 Revision B
Model TDA-2G/2GN
Note: The bar graph will scan left to right across the
bottom of the % LEAKAGE display indicating that the unit
is electronically adjusting the internal gain. This can take
from 2 to 200 seconds to complete, depending on the
aerosol level present.
The bar graph goes off below the % LEAKAGE display screen and the
audible alarm sounds indicating that the unit is averaging the 100%
reading over the time period established by parameter L3 (5-120
seconds).
At the end of this period, the LED over the <0> key begins flashing and
the audible alarm turns off. The unit has completed setting the 100%
baseline using the upstream sample.
A flashing LED over the <0> key indicates that the unit needs to be re-
zeroed.
9. Turn the Selector Valve to the CLEAR position and press <ENTER>.
When the zero has been established, the beeps will stop, and the %
LEAKAGE display shows [.0000] (or [.000], depending on the
number of decimal places determined by parameter L4).
CAUTION: Do not to turn the valve from CLEAR until the final
tone is heard.
T2G0-0877 19 Revision B
Model TDA-2G/2GN
10. Turn the Selector Valve from the testing position to CLEAR.
11. Press <ENTER> and then the <0> key. The LED on the <0> key
begins to flash.
12. Press <ENTER> to start the instrument defining the 0% baseline.
When the zero has been re-established, the beeping stops and the %
LEAKAGE display screen will show .0000 (or .000, depending on the
number of decimal places determined by parameter L4).
After the 100% and 0% baselines have been established, the unit is ready for
testing operations. During testing operations, readings may be taken from either
the Front Panel Display or the Scanning Probe, lessening operator distraction.
CAUTION: This can be carried out only if the Internal Reference has been
established and the upstream sample line is connected to the
Upstream Port on unit.
NOTE: This procedure only verifies the upstream aerosol challenge to the filter; it
does not establish the 100% baseline of the photometer.
T2G0-0877 20 Revision B
Model TDA-2G/2GN
2. Press the REF key while sampling upstream of the filter. Observe that
the REF LED illuminates, the 100 LED goes out, and the %
LEAKAGE display bar graph begins to sweep left to right.
Next, the % LEAKAGE display will go blank while the bar graph
begins to sweep, the 100 LED illuminates, and the REF LED goes off,
indicating that the unit is returning to the 100% baseline previously
established.
The % LEAKAGE display will illuminate to indicate that the unit has
returned to the 100% baseline.
T2G0-0877 21 Revision B
Model TDA-2G/2GN
4. If the re-verification indicates that the concentration of the challenge
aerosol has not changed significantly, return the Selector Valve to
the DOWNSTREAM position and continue testing.
T2G0-0877 22 Revision B
Model TDA-2G/2GN
5 MAINTENANCE
Scheduled Maintenance
5.1 WEEKLY
• Clean the scanning probe screens. These are located on the black and
red circular scanning probe nozzles.
• Clean the gross particulate screen located at the base of the flexible
scanning probe extension.
• Remove any loose debris from the Scanning Probe and front panel
sampling ports.
5.2 ANNUALLY
• Return the TDA-2G to a factory authorized facility for calibration and
cleaning. Please contact the ATI Customer Service Department at 410-
363-9696 for a return authorization number. A service date will be
scheduled for your instrument at that time.
NOTE: The internal electronics are not user serviceable. Any electronic
problems must be analyzed and repaired at an authorized service center.
At present, there is no error message or indication on the front panel that the
scattering chamber light source is not working. If the scattering chamber light
source has burned out, the operator will witness a lack of response on the unit’s
% LEAKAGE display.
Spare components are not supplied with the TDA-2G or the TDA-2GN nuclear
unit and must be ordered separately. See section A-4 for a list of recommended
spare parts that may be purchased in kit form or individually. These lists include
T2G0-0877 23 Revision B
Model TDA-2G/2GN
only parts replaceable by the user in the field. Other repairs requiring instrument
or component recalibration must be performed at an ATI service center.
The TDA-2SP Scanning Probe is a rugged, low maintenance device. The probe
contains a course wire screen near the base of the flexible neck to prevent fibers
and large particles from being drawn into the photometer. In addition, there are
screens in each of the two round nozzles that thread onto the flexible neck. The
blue rectangular nozzle contains no screen.
If the screens are punctured, replace them immediately. Spare nozzles and
replacement scanning probe components can be ordered from ATI (see Section
A-4).
To access the screen in the flexible neck, unscrew the flexible extension from the
probe body. A small tool may be necessary to reach into the neck to remove and
wipe the surface of the screen.
T2G0-0877 24 Revision B
Model TDA-2G/2GN
A-1. SPECIFICATIONS
Dimensions (L x W x H)
13.5 in x 9.5 in x 5.0 in
34.3 cm x 24.0 cm x 12.7 cm
Weight
TDA-2G Aerosol Photometer (standard use) 15.5 lbs (7.0 kg)
TDA-2GN Aerosol Photometer (nuclear use) 15.5 lbs (7.0 kg)
TDA-2SP Scanning Probe 2.5 lbs (1.1 kg)
Input Power
90 to 250 volts AC, 50 or 60 Hz, 1.5 amps
Fuse
250 volts, 2.0 amps slow blow (5 x 20 mm)
Dynamic Range
0.00005 to 120 micrograms per liter
Accuracy
1% of full-scale for the amplifier decade range in use
Repeatability
0.5% full-scale for the amplifier decade range in use
Sampling Rate
1 cfm (28.3 lpm)
DATA OUTPUT
RS-232: Port settings: 2400, 8, N, 1 (Continuous output)
T2G0-0877 25 Revision B
Model TDA-2G/2GN
A-2. TROUBLESHOOTING
CONTROLLER Error Messages
E3 – System Fault Internal electronic or software Turn system power off and then on. If
malfunction. the error does not repeat, continue to
use normally. If the error repeats,
return the unit to ATI for service and
repair.
E4 – Memory Fault Loss of internal memory values. Return the unit to ATI for repair and
This message is only displayed at service as soon as possible.
the first power-up following loss of
memory. The “Internal
Reference” key will be disabled
allowing use only with an actual
upstream sample to set the 100%
level.
T2G0-0877 26 Revision B
Model TDA-2G/2GN
A-3. QUICK SETUP REFERENCE
SETUP
To enter setup, press <ENTER> then <∆>
L0 - Run
L1 - Audible Alarm
L2 - Display Rate
L3 - 100% Sample Time
L4 - Decimal Places
L5 - Internal Reference
L6 - Hour Meter
L7 - Factory Defaults
L8 - Software Revision
L9 - Bar graph
L10 - Display Intensity
ALARM (ON)
<ENTER>
<ALARM> LED Flashes – (when [----] is displayed, alarm is disabled)
<ALARM> Last alarm set point displayed, <∆> or <∇> to change
<ENTER> Alarm set point entered and LED will stay lit
ALARM (OFF)
<ENTER>
<ALARM> Set point displayed
<ALARM> --- displayed indicating alarm will be disabled
<ENTER> Alarm disabled
INTERNAL REFERENCE
Set Selector Valve to CLEAR
<ENTER>
<REF> Calibration liquid and Internal reference display, e.g., [P1] and [100.0].
Use <∆> or <∇> to change
<ENTER> Scans 15 seconds then <0> key flashes
<ENTER> Sets 0%, which takes about 5 seconds
T2G0-0877 27 Revision B
Model TDA-2G/2GN
SETTING 100%
Connect Upstream Sample Line & set Selector Valve set to UPSTREAM
<ENTER>
<100> <100> key LED flashes
<ENTER> Bar graph scans then <100> key LED flashes. The unit averages 100%
for a time defined by the L3 Setting. When the <0> LED flashes, turn the
Selector Valve to Clear
<ENTER> Sets 0%. This takes about 5 seconds
UPSTREAM CHALLENGE
QUICK CHECK FEATURE
Connect the Upstream Sample Line & set Selector Valve to UPSTREAM. Assure that
the Upstream Sample Line is sampling the upstream aerosol challenge.
<REF> Bar graph scans 15 sec. – displays actual upstream challenge for 3 sec and
returns to 100% baseline
T2G0-0877 28 Revision B
Model TDA-2G/2GN
A-4. ACCESSORY LIST
Item Part Number Description
1 T2G0-1038 TDA-2G Spare parts kit (see list below for components)
2 T2GN-0948 TDA-2GN Spare parts kit (see list below for components)
3 9300102 Shipping case
4 T2SP-0879 Scanning probe assembly, complete with 3 nozzles
5 T2SP-1095 Scanning probe extension, 12 ft.
6 T2E0-0005 NSF Scanning probe nozzle
7 T2E0-0572 1” Isokinetic, round (red)
8 T2E0-0798 Isokinetic, rectangular (blue)
9 6700001 Replacement line cord, 120V
10 5200116 ¼” ID PVC sample tubing (specify length)
11 T2G0-1119 Null-modem cable, 6 ft.
T2G0-0877 29 Revision B
Model TDA-2G/2GN
A-5. WARRANTY
International Warranty
Air Techniques International
Air Techniques International, hereinafter referred to as ATI, warrants the equipment purchased
hereunder to be free from defect in materials and workmanship under normal use and service,
when used for the purpose for which it is designed, for a period of (1) one year from the date of
shipment. ATI further warrants that the equipment will perform in accordance with the technical
specifications accompanying the formal equipment offer.
ATI will repair or replace any such defective items that may fail within the stated warranty period,
PROVIDED:
1. That any claim of defect under this warranty is made within thirty (30) days after
discovery thereof and that inspection by ATI, if required, indicates the validity of such
claim to ATI’s satisfaction; and
2. That the defect is not the result of damage incurred in shipment to or from our
factory; and
3. That the equipment has not been altered in any way whether as to design or use,
whether by replacement parts not supplied or approved by ATI, or otherwise; and
4. That any equipment or accessories furnished but not manufactured by ATI, or not of
ATI design, shall be subject only to such adjustments as ATI may obtain from the
supplier thereof.
ATI’s obligation under this warranty is limited to the repair or replacement of defective parts with
the exception noted above. If the equipment includes a scattering chamber, ATI’s warranty does
not extend to contamination of the scattering chamber by foreign material.
At ATI’s option, any defective equipment that fails within the warranty period shall be returned to
ATI’s factory for inspection, properly packed with shipping charges prepaid. No equipment shall
be returned to ATI without prior issuance of a return authorization by ATI.
No warranties, express or implied, other than those specifically set forth herein
shall be applicable to any equipment manufactured or furnished by ATI and the
foregoing warranty shall constitute the Buyer’s sole right and remedy. In no
event does ATI assume any liability for consequential damages, or for loss,
damage or expense directly or indirectly arising from the use of ATI products, or
any inability to use them either separately or in combination with other equipment
or materials or from any other cause.
Purchaser ________________________________________________________
T2G0-0877 30 Revision B
Model TDA-2G/2GN
A-6. RS-232 COMMUNICATION
The TDA-2G serial output is formatted as a continuous string of ASCII numeric
characters.
0.0001
0.0000
0.0000
0.0001
0.0024
0.0013
0.0001
0.0000
T2G0-0877 31 Revision B
Model TDA-2G/2GN
A-7. AEROSOL CORRECTION FACTORS
DOP Substitute liquid Internal Reference Factors
DOP (DEHP) 1.00
PAO (Emery 3004) 0.73
DOS (DEHS) 0.96
Mineral Oil 0.90
Ondina Oil (Ondina EL) 0.79
Kaydol 0.92
Polyethylene Glycol (PEG 400) 1.11
Paraffin Oil 0.89
Corn Oil 0.88
The above values are for use when substitute liquids are used in-place of the liquid specified for
the Factory equipment calibration and setup.
The generator pressure factors are used as multipliers for the applied pressure to a Type III-A
Laskin nozzle generator.
Example:
• DOP-20 psi applied with dilution airflow of 135 cfm yields 100 mg/m3
• PAO-23 psi (20 psi x 1.15 correction factor) applied with dilution airflow of 135 cfm
yields 100 mg/m3
Internal reference factors are multipliers for the reference setting required for a100% reading.
Example:
• DOP with P1 selected - 100 required for 100% response to 100 mg/m3
• DOS with P3 selected - 0.96 (0.96 X 100 DOP value) required for a 100% response
to 100 mg/m3
Note: The above correction factors apply only to air operated generators. If a thermal
type generator, TDA-5A or TDA-5B, is being used the upstream concentration must be
sampled to obtain a 100% setting.
T2G0-0877 32 Revision B