Manual Explorer1 PDF
Manual Explorer1 PDF
Manual Explorer1 PDF
TECHNICAL MANUAL
AW-00335_001-0101
Notice
The information contained in this manual is confidential and proprietary to Hologic, Inc.
This information is provided only to authorized representatives of Hologic's customers
solely for the purpose of facilitating the use of Hologic's products. No information
contained herein may be disclosed to any unauthorized person for any purpose whatsoever
without the prior written consent of Hologic, Inc.
The procedures described in this document are intended for use by direct employees of
Hologic, Inc., or authorized Hologic Equipment Resellers and their trained field
engineers. Any unauthorized or untrained persons performing these procedures may affect
the warranty of the Explorer.
Exercise proper caution when servicing the system. There are dangerous and potentially
lethal voltages accessible within the Explorer system. To avoid exposure to shock hazards,
the Main Circuit Breaker should be switched off, and the power cord removed, before
working inside any part of the system.
The Explorer Fan Beam X-ray Bone Densitometer produces ionizing radiation in the form
of X-rays. It may be dangerous to the patient, operator or field engineer unless safe
exposure factors and operating instructions are observed. To avoid unsafe exposure, do not
attempt to service the equipment unless you are a Hologic, Inc., certified field engineers.
Exercise proper caution when servicing the system. A dosimeter (film badge) should
always be worn while on site. Dose and scatter measurements must be taken after each
service call to ensure that the parameters are still within specifications.
Hologic, Inc., has made all effort to ensure that the information in this manual is accurate
and complete. Hologic, Inc., shall not, however, be liable for any technical or editorial
errors or omissions contained herein, or for incidental, special or consequential dangers in
connection with the furnishing or use of this manual. The information in this manual is
subject to change without notice.
Explorer and the Hologic logo are registered trademarks of Hologic, Inc. All other
products and company names used in this manual are trademarks and registered
trademarks of other manufactures.
Printed in U.S.A.
Copyright 2004 by Hologic, Inc., All rights reserved
ii
Explorer Technical Manual
Table of Contents
Chapter 1 - INTRODUCTION ................................................................................... 1-1
System Overview ..................................................................................................... 1-1
X-Ray Scanning Principles ...................................................................................... 1-2
Functional Overview ................................................................................................ 1-3
Product Specifications ............................................................................................. 1-9
Product Specifications ....................................................................................... 1-9
Line Voltage and Maximum Line Current ...................................................... 1-10
Line Inductance .......................................................................................... 1-10
Duty Cycle ....................................................................................................... 1-10
Exam Mode Performance ................................................................................ 1-10
AC Power Isolation Module ............................................................................ 1-11
Table of Contents 1
Explorer Technical Manual
2 Table of Contents
Explorer Technical Manual
Table of Contents 3
Explorer Technical Manual
4 Table of Contents
Explorer Technical Manual
Table of Contents 5
Explorer Technical Manual
6 Table of Contents
Section 1
INTRODUCTION
System Overview
The Explorer™ X-ray Bone Densitometer (Figure 1-1) performs bone mineral content (BMC)
and bone mineral density (BMD) examinations on the spine and hip. Optional software will
provide examinations on the forearm, whole body and decubitus lateral. It does so using X-rays
of two different energy levels. This dual-energy scheme allows soft tissue within the selected area
to be subtracted out leaving only bone to be scanned and estimated.
AW-00335_001-0101
1-1
Explorer Technical Manual
The X-ray scans produced by the system, and displayed on the monitor, are intended only
to locate anatomical sites for measurement and to assure the operator that the machine is
operating properly. They are not intended as a substitute for conventional film-based
diagnostic scans.
1-2
Explorer Technical Manual
value just above this peak, and apply that value to discriminate, point by point in
the Q scan, between “bone” points (whose Q is above the threshold) and “non-
bone” points (whose Q is below the threshold).
Tissue Bone
NUMBER
OF
POINTS
WITH A
SPECIFIC
VALUE OF
Q.
Q KP1085_004-0103
Functional Overview
This section provides block diagrams of the Explorer system along with a brief functional
overview of each diagram and block. A detailed functional description along with
1-3
Explorer Technical Manual
C-Arm
Distribution
Interface
Board
Board
Operator’s Console
X-Ray
TX TY AY
Controller
Scanner and Tank
AW-00335_001-0104
1-4
Explorer Technical Manual
generation of the X-Rays by the X-Ray Source or “Tank”. The X-Rays pass through the
patient and are sensed by the Detector Assembly.
AC
AC from
AC Power
DIN Rail
Strip
(Scanner)
AC
Printer Monitor
Computer
AC
Video
Modem Phone
Keyboard (Optional) Jack
Network
Network I/F Card
Jack
Mouse
Communications
Scanner
Controller
KP1085_004-05
1-5
Explorer Technical Manual
To/From TY
Motor
Communications JP10 TY Stepper
JP5
Controller Motor Driver TY
Signal Encdr
Distribution
Board TX
TX Stepper Motor
JP7
Motor Driver TX
Encdr
+/-15V JP6
From DIN
Rail AY
and LVPS +24V AY Stepper Motor
JP8 JP11
Motor Driver
AY
Encdr
AW-00335_001-0106
1-6
Explorer Technical Manual
simple on/off commands. Commands sent to the XRC tell it when to produce X-rays,
which power level of pulse to produce, and the pulse mode to be used. These commands
will subsequently control the way the X-Ray Source (Tank) is driven by the XRC. The
commands to the Filter Drum Assembly turn the Filter Drum on or off and synchronize
the filter drum with the AC line waveform. The last set of command signals is sent to the
Detector Assembly to synchronize the acquisition of data with the production of the X-
rays from the source
(Tank).
Detector Assembly
Laser
Assembly
Detector Board
(Detectors, Muxes,
and A/D Converter)
To/From C-Arm
Distribution
From Interface
Board Board
Filter Drum Motor
and
Position Sensor
X-Ray
X-Ray
Source
Controller
(Tank)
DIN KP1085_004-0107
Rail
1-7
Explorer Technical Manual
Block Description
Control Panel Provides switches for manually moving the C-Arm and
Patient Table. Also provides Emergency Stop and Instrument
On/Off switches.
TX Stepper Motor Controls the motion of the Patient Table in and out motor and
Driver monitors table position information from the encoder.
TY Stepper Motor Controls the motion of the Patient Table left to right motor
Driver and monitors table position information from the encoder
AY Stepper Motor Controls the motion of the C-Arm left and right motor and
Driver monitors C-Arm position information from the encoder.
C-Arm Interface Controls the Filter Drum motor, generates timing and control
signals for the X-Ray Controller and the Data Acquisition
System, monitors the tape switches, and provides power to the
Positioning Laser.
X-Ray Controller Controls the operation of the X-ray Source.
X-Ray Source Unit Generates the X-ray beam.
Detector Board Converts the X-rays into electrical signals. Integrates the
signals from the Solid State Detectors and converts them to a
digital value in the Analog-to-Digital converter circuitry.
Control Panel Provides the operator with a means to reposition the table, C-
Arm, and to turn on the laser.
Positioning Laser Provides a laser crosshair beam to assist in positioning the
patient on the Patient Table.
1-8
Explorer Technical Manual
Product Specifications
Product Specifications
SPECIFICATION DEFINITION
Scanning Method Multi-detector array, Indexing table, and motorized C-arm
X-ray System: Tungsten target switched Pulse Dual-Energy x-ray tube, operating at 100
and 140kVp, capable of 0.75 and 2.5mA average at 25% duty cycle
Detector System: 54 multi-channel detector (4mm) consisting of CdWO4 scintillators cou-
pled to Silicon diodes
Scanning Sites: AP Lumbar Spine, Proximal Femur (Hip), Forearm, Decubitus Lateral
Spine, and Whole Body
Scan Region: 1.97m (77.5 in) X .65m (25.6 in) maximum
(at pad surface)
Scatter Dose Rate: Nominal 10mGy/hr at 1 m from the examination table
Leakage The Explorer meets the requirements of 21 CFR 1020.30(k) for leakage
Radiation: from the X-ray source
External Shielding None required unless required by local ordinance
Requirement:
Measurement Measurements of Bone Mineral Content (BMC) and projected Area are
Accuracy: calibrated to Hologic anthropomorphic spine phantom of known mineral
content to within +/-1.0%.
Calibration: Self Calibrating using HOLOGIC Automatic Internal Reference System.
Operator calibration NOT required.
Patient Manual movement (two independent axes) using push-button controls.
Positioning: Laser Indicator (Red, < 1mW) with cross hairs used for alignment.
System Weight: Console (Computer,
Installed Scanner Printer, Monitor)
327kg 720lb 34.1kg 75lb
System Weight: 568kg 1250lb
Shipping
Temperature (Ambi- Operation Storage
ent): 15o to 32oC (59o
to 90oF) to -30o 50oC
(-22o to 122oF)
Humidity (Ambient): Operation Storage
20 to 80% relative humidity, non- 10 to 90% relative humidity, non-
condensing condensing
Subject Weight: The maximum subject weight is 136kg (300 lb.)
Footprint (operating)
(+/- 25mm (1in.)): Length Width Height
m in. m in. m in.
3.02 119 1.50 59 1.42 56
Heat Load: 1000W (3400 BTU/h) (maximum)
Table Height: (+/- 71cm (28in.)
25mm (1in.))
Positioning Laser: Laser Diode (<1mW) cross hair, with emergency mechanical shutter
X-ray Collimation Aperture with 1.0 mm (0.040 in) slit
Leakage Current: Normal <75µA Single Fault <400µA
1-9
Explorer Technical Manual
Line Inductance
0.015 mH/volt maximum
Duty Cycle
Explorer Detail 50%
All other scan modes 100%
1-10
Explorer Technical Manual
1-11
Explorer Technical Manual
1-12
Section 2
FUNCTIONAL DESCRIPTION
This section provides a detailed functional description of the system with interconnection
diagrams. Refer to “Functional Overview” on page 1-3 for a block diagram (Figure 1-3) and a
brief functional description of each block.
Computer
The scanner is interfaced to a PCI Bus computer which controls table and C-arm movement, X-
ray generation, performs all necessary calculations, and manages both the patient and QC
database information.
The computer is a Pentium Celeron based (or higher) PC that comes equipped with floppy disk
drive, hard disk drive, keyboard, mouse, 17” color monitor, and CDROM R/W for archiving scans
and backups of the system database. For details pertaining to the computer and its associated
components, please refer to the documentation shipped with each unit.
Note: Refer to Appendix A for computer specifications and software loading instructions.
Interface Connections
Table 2-1 describes the interconnections between the PCI Communications Controller Board and
the Distribution Board. The table also identifies the connectors and their pin assignments.
Table 2-1 Communications Controller Board/Distribution
Board Interconnection Description
Signal Pair Description CC1 Pin Dist2 Pin
ATD+ Asynchronous data to the Scanner. JP1-2 JP10-3
ATD- JP1-27 JP10-4
STD+ Synchronous data to the Scanner. JP1-28 JP10-6
STD- JP1-4 JP10-7
2-1
Explorer Technical Manual
KP1085_003-0201
2-2
Explorer Technical Manual
Distribution Board
The Distribution Board provides interconnections between the Operator's Console (PC)
and the Scanner. It passes several signal lines from the Operator's Console and power lines
from the DIN rail directly to the C-Arm Interface module. It also provides buffering and
individual drivers and receivers for various signal lines to and from individual Scanner
modules and the Communications Controller. The Distribution Board is located in the
Electronics Tray in the base of the Scanner.
One cable connects the Operator's Console (PC) communications bus to the Distribution
Board. One cable connects to the Distribution Board from the DIN rail. This cable brings
DC power to the Distribution Board and connects the X-Ray On and Emergency signal
lines to the Power Module.
Up to eight cables connect the Distribution Board to the various Scanner boards. Three
cables connect to the three Motor Controller Boards (Table X, Table Y and Arm Y). Two
cables (one signal and one power) connect to the C-Arm Interface Board. In addition, one
cable connects to the scanner Control Panel Interface Board.
The Distribution Board has provision for three jumpers that can be installed to override
the EMERGENCY signal lines when troubleshooting.
Power
The Distribution Board receives +24 and +/-15VDC from the Multi-voltage DC Switching
Power Module. The +24VDC is applied through four individual circuit breakers to the
Table X, Table Y and Arm Y Motor Controller Boards. The +24 and +/-15VDC are
passed to the C-Arm Interface Board. The +24 VDC is also reduced to +7 and +5VDC by
regulators to power op-amps and analog switches located on this board. The +7VDC is
passed to the Control Panel Controller Board. The +5VDC powers the digital section of
the Distribution Board. The +24VDC power supply is not closely regulated and its outputs
may range from +24V to +35V under normal conditions.
Note: +7VDC may measure anywhere from +6.25VDC to +7.25VDC. This is true
everywhere +7VDC is shown in this manual.
Limits for +/- 15VDC
Six green LEDs indicate the status of the +28, +24, +15, -15, +7 and +5VDC (ON
indicates the respective voltage is present). Five red LEDs indicate the status of the five
circuit breakers applying voltage to the motor drivers/controller. ON indicates the circuit
breaker has been tripped by an over-current condition.
Interface Connections
Figure 2-2 shows connections to/from the Distribution Board.
2-3
Explorer Technical Manual
OPERATOR’S INSTRUMENT
CONSOLE
COMPUTER C-ARM
INTERFACE
BOARD
COMMUNICATIONS
CONTROLLER
BORAD CONTROL PANEL
CONTROLLER
BORAD
TX MOTOR
DISTRIBUTION CONTROLLER
BOARD BOARD
+24 VDC, +/-15 VDC
EMERGENCY
SIGNAL LINE
TY MOTOR
CONTROLLER
BOARD
AY MOTOR
+24 VDC CONTROLLER
BOARD
AW-00335_001-0202
2-4
Explorer Technical Manual
The Motor Controller Boards receive movement commands from the system computer via
the communications bus. Each Motor Controller Board contains an ID switch and four
status indicators (LEDs). The ID switch is a 16-position rotary encoded switch (SW1) that
is read during system initialization to determine the Motor Controller Board address for
communicating with the system computer. ID switch settings for the three Motor
Controller Boards are as follows:
Table X drive 4
Table Y drive 5
C-Arm Y drive 7
The four red status LEDs provide visual indications of motor drive power on, Stepper
CPU active, motor drive direction, and motor step pulses.
The Motor Controller Board also has provision (JP2) for connecting limit switches to
inhibit motor operation when the mechanism goes beyond established mechanical limits.
This feature is not used in the system and, therefore, no cable is connected to JP2.
Power
Power input to the Motor Controller Board is +24VDC. This voltage provides the motor
drive power and is converted down to +5VDC for use by logic circuits on the board. The
+5VDC is also converted to -5VDC. Two green LEDs provide a visual indication of the
power present on the Motor Controller Board.
Interface Connections
Figure 2-3 shows the typical interconnections between the Distribution Board, the Motor
Controller Board, the Stepper Motor, and the Position Encoder. Table 2-2 describes the
interconnections between the Distribution Board and the Motor Controller. Table 2-3
describes the interconnections between each Motor Controller and its respective stepper
motor and position encoder. The tables also identify the connectors and their pin
assignments.
To/From
Distribution Board ( )
( )
ARD+, ARD- ( )
STEPPER
ATD+, ATD- ( ) MOTOR
SYSRESET+, SYSRESET- MOTOR
+28V
CONTROLLER +REF
28V_RET ( ) POSITION
-REF ENCODER
KP1085_003-0203
2-5
Explorer Technical Manual
Power
The Control Panel Controller Board receives +7VDC from the Distribution Board. The +7
is reduced to +5VDC to power the circuitry on this board and is applied to the Operator
Control Panel to power the LEDs. A green LED, on this board, provides a visual
indication of the +5VDC power (ON indicates the voltage is present).
Interface Connections
Figure 2-4 shows the interconnections between the Distribution Board, Control Panel
Controller Board, and the Operator Control Panel.
Table 2-4 describes the interconnections between the Distribution Board and the Control
Panel Controller. Table 2-5 describes the interconnections between the Control Panel
Controller and the Operator Control Panel. The tables also identify the connectors and
their pin assignments.
2-6
Explorer Technical Manual
To/From
Distribution Board
ARD+, ARD- L0_PWR* - L7_PWR*
ATD+, ATD- TZ_PWR*
SYSRESET+, SYSRESET- XRAY_LIGHT_PWR*
EMERGENCY+, EMERGENCY- CONTROL SW0 - SW2 OPERATOR
XRAY_LIGHT+, XRAY_LIGHT- PANEL SR0 - SR2 CONTROL
EMERBENCY_PANEL CONTROLLER EMERGENCY_PANEL PANEL
HW_EMERGENCY_RET HW_EMERGENCY_RET
+7V +5V
AW-00335_001-0205
To/From
TILT_A
C-Arm
TILT_B
Tilt Switch
(A/SL)
2-7
Explorer Technical Manual
2-8
Explorer Technical Manual
• Move the stepper motor of the Reference Drum device and monitor the encoded
signals returned from each drum of the device. Two green LEDs, on this board,
provide a visual indication of the Reference Drum operation.
• Generate +7VDC for the DAS system
Power
The C-Arm Interface Board receives +24 and +/-15VDC from the Distribution Board. The
+24VDC powers the Reference Drum stepper motor driver circuitry. The +/-15V is passed
through this board to the DAS. The +24VDC is also reduced to +5 and +7VDC. The
+5VDC powers the digital section of this board and the laser while the +7VDC is applied
to the DAS.
Three green LEDs, on this board, provide a visual indication of the +24 and +5VDC
power (ON indicates the respective voltage is present). The third LED monitors the
+28VDC. On the system, this voltage is actually +24VDC.
Interface Connections
Figure 2-5 shows the interconnections between the Distribution Board and the C-Arm
Interface Board. Table 2-6 describes the interface signals and identifies the connectors and
their pin assignments.
2-9
Explorer Technical Manual
CONTINUITY 1
ATD_CARM+, ATD_CARM-
STD+, STD-
STCLK+, STCLK-
STFRM+, STFRM-
ARD_CARM+, ARD_CARM-
SRD+, SDR-
SRCLK+, STCLK-
SRFRM+, SRFRM-
EMERGENCY_CARM, EMERGENCY_CPANEL
LINESYNC+. LINESYNC-
INTEGRATE+. INTEGRATE-
DISTRIBUTION SYSRST_CARM+, SYSRST_CARM-
C-ARM
BOARD INTERFACE
EMERGENCY+, EMERGENCY-
BOARD
XRAY_LIGHT+, XRAY_LIGHT-
+15V
+15V_RET
-15V
+24V
24V_RET
28V
28V_RET
CONTINUITY 2
KP1085_003-0206
2-10
Explorer Technical Manual
Figure 2-6 shows the boards and assemblies that connect to the C-Arm Interface Board.
2-11
Explorer Technical Manual
KP1085_003-0207
Interface
Figure 2-8 shows the interface connections between the C-Arm Interface Board and the X-
Ray Controller Assembly. Table 2-13 describes the interface signals and identifies the
connectors and their pin assignments. Note that the AC input power comes directly from
the DIN Rail Power Module and connects to the X-Ray Controller Assembly.
2-12
Explorer Technical Manual
XR_RLY_ON+, XR_RLY_ON-
XR_FREQ+, XR_FREQ-
HIGH VOLTAGE
XR_BEAMON+, XR_BEAMON-
TRANSFORMER
XR_ISET+, XR_ISET- PRIMARY
XR_KV1+, XR_KV1-
C-ARM XR_KV0+, XR_KV0-
INTERFACE
ACLINE+, ACLINE- X-RAY X-RAY
BOARD
LIGHTON+, LIGHTON- CONTROLLER SOURCE
IBEAM+, IBEAM- ASSEMBLY UNIT
KVP+, KVP-
XRFAULT+, XRFAULT- X_RAY TUBE
FILAMENT
TRANSFORMER
From DIN Rail 240 VAC PRIMARY
Power Module
KP1085_003-0208
2-13
Explorer Technical Manual
2-14
Explorer Technical Manual
The Filament circuit is a regulator. It receives current feedback from the tank indicating
the tube current. 10 volts at the feedback input is equivalent to 10 milliamps of current in
the tube. The 10 volts are reduced to 5 and then fed to an A/D circuit to be converted to a
digital value for use by the control circuits. The value into the A/D is sampled towards the
end of the pulse where it is stable. After conversion, the digital value is fed as an address
to an EPROM. If the current is correct, the output of the EPROM will be 0 volts. Anything
other than 0 volts (“+” = high current, “-“= low current) will be added and accumulated in
two stages forming an error integrator. This digital error number is converted back to an
analog value and then used as the threshold of a comparator. The output of the comparator,
through additional circuitry, adjusts the duty cycle of the modulated filament voltage.
Refer to Table 2-8 for pin assignments.
2-15
Explorer Technical Manual
Integrator/Multiplexor Subsection
The Integrator/Multiplexor subsection of the Detector Assembly receives up 54 signals in
parallel from the Solid State Detector Boards. This subsection integrates and stores those
signals and then applies the stored signals to the Analog to Digital subsection of the
Detector Assembly.
Switched signals charge integrating capacitors in this subsection during a given charging
time. After the charging time, the switched signals are turned off and the charges are held
on the capacitors. Each integrator is sampled, in photodiode order, by a multiplexor and
the output sent to the Analog to Digital subsection. At the end of sampling, all the
integrating capacitors are discharged in parallel by shorting them out with analog
switches.
Each integrator has an additional input into which a test signal (TESTLVL) can be
applied when there are no X-rays present. This test signal is used to verify the operation of
the integrators and multiplexors by the SQVERIFY diagnostic program.
2-16
Explorer Technical Manual
in a final multiplexor consisting of four analog switches. The multiplexed signals pass
through a programmable gain amplifier and summing amplifier before being applied to an
A/D converter. A one-volt fixed DC offset is inserted at the summing amplifier to insure
that no channels ever go negative. The A/D converter converts the analog signal into 16
bit parallel data for processing by a Digital Signal Processor.
The ADC subsection uses a Digital Signal processor to generate all the control signals
necessary for the Detector Assembly. This processor also provides a high-speed serial data
link to the computer.
Power
The 54-Channel Detector Assembly receives +/-15V and +5V from the C-Arm Interface
Board. The +/-15V is passed through this board to the Detector subsection. It is also
regulated to +/- 12V by series regulators to power operational amplifiers and analog
switches located on this board. The +/- 15V is also used to generate +/- 5V to power the
analog-to-digital converter circuit. The +7V is reduced to +5V to power the digital section
of this board.
Interface Connections
Figure 2-8 shows the interconnections between the 54-Channel Detector Assembly and
the C-Arm Interface Board. Figure 2-8 describes the interface signals and identifies the
connectors and their pin assignments.
STD+, STD-
STCLK+, STCLK-
STFRM+, STFRM-
SRD+, SRD-
SRCLK+, SRCLK-
SRFRM+, SRFRM-
SYSRST_DAS+, SYSRST_DAS-
C-ARM INTERFACE INTEGRATE+, INTEGRATE- 54-Channel
BOARD XR_ZEROX_DAS+, XR_ZEROX_DAS- Detector
Assembly
+15V
15V_RET
-15V
+7V
KP1085_003-0209
2-17
Explorer Technical Manual
2-18
Explorer Technical Manual
Power Module
The Power Module provides the AC voltage required by the computer system and the
Scanner. The module consists of a Main Power circuit breaker, a Power On indicator, and
a torroidal AC line input isolation transformer.
AC power from the Torroid Power Module is sent to the Scanner on a four-wire cable
providing split 240 volts to the DIN Rail. From the DIN Rail, unfiltered AC is fed to the
line filter and, from there, distributed to various sections of the Scanner. Unfiltered AC is
also fed to an AC outlet connector at the rear of the Electronics Tray. This outlet provides
AC power to the Operator’s Console power strip. A second AC outlet provides unfiltered
AC only while x-rays are being produced. This outlet is used in the factory to give a visual
indication of x-ray production during testing.
2-19
Explorer Technical Manual
2-20
Explorer Technical Manual
Section 3
INSTALLATION
Required Tools
When installing the Explorer, a tool kit that includes the following items is required:
• Assortment of both flat-bladed and Phillips screwdrivers
• Assortment of needle-nose and diagonal cutting pliers
• Socket drivers (full set including 1/4”, 3/8”, 7/16”, 1/2” and 3/4”)
• Open-end wrenches (full set including 3/8”, 7/16”, 1/2”, 9/16”, 5/8” and 3/4”) and an
adjustable wrench
• Hex driver (Allen wrench) set including sizes 1/16”, 5/64”, 3/32”, 7/64”, 1/8”, 9/64”, 5/
32” 3/16”, 7/32” and 1/4”
• Oscilloscope and digital multimeter
• Survey meter (Victoreen model 450P or equivalent)
• X-ray leakage test tool, #099-0566
• Measuring tape, approx. 3.7 meters (12 feet)
• Beam alignment tool (010-0923)
Required Documentation
The following documents are required:
• Explorer User’s Guide
• Computer manual
• Printer manual
• Introducing Microsoft Windows® XP Professional
• Optical Disk Drive manual (when installed)
• The Radiation Measurement Report
3-1
Explorer Technical Manual
2"
(5 cm)
To be supplied
49.5" 61"
(125.7 cm) (155 cm)
+
L O
SC
P GU
P GD
N
P
DE
L
9
3
*
.
PS
OCK
CA
L
2
/
NUM
L OC
K
OCK
ME
UM
HO
EN
D
IN S
7
0
N
L
ON
USE
GE
GE
DW
PA
PA
UP
PA
OL L
O ME
SCR
L OC
ND
K
E
CREN
ETE
ER
R IN T
IN S T
EL
D
P
S
28"
F1
2
RL
CT
F10 F11
TR
ENE
H IF T
S
L T
A
F9
(76 cm)
F4
F3
F2
F1
T
AL
OC
APSL K
S H IF T
s
Ec
CT R
TAB
L
C
30"
(72 cm)
Kg lbs
Equipment Weight 362 795
Max. Patient Weight 136 300
Total Weight 498 1095
m ft
Area 1.9 x 1.1 6.3 x 3.5
Kg/m2 lb/ft2
Floor Loading 220 45
3-2
Explorer Technical Manual
Uncrate Unit
Remove the unit from the crate as described below:
1. Cut the strapping that holds the packaging together.
2. Remove the cardboard cap by lifting it up and off.
3. Remove the cardboard sleeve by lifting it straight up and off. Be careful not to scratch
the unit.
4. Remove the wooden table shelf and the boxes packed with the unit.
Note: There is a metal cross brace attached to the wooden table shelf. Remove and save
this cross brace as it is used later in the installation.
3-3
Explorer Technical Manual
5. Verify that the serial numbers on both crates and the scanner all match.
Take Inventory
The Explorer system is shipped in two crates. Using the following checklist, take
inventory of the contents of all crates and boxes, and confirm that all of the expected items
have been received. Unused power cables and circuit breakers are to be returned to
Hologic. Report any discrepancies to the Hologic Sales Department.
Note: Two System Backup disks are shipped with each system. One will be included
with the scanner and the other shipped in the crate with the PC. Mark the one shipped
with the scanner and note that if the Serial Numbers on the backup disks do not agree, the
disk shipped with the scanner is the correct disk.
* Optional.
3-4
Explorer Technical Manual
Short Doorway
If a system must be moved through a doorway that is not at least 81" (206cm) high, the
tabletop can be removed. This allows the unit to fit through a doorway 79" (201cm) high
(inside dimension). See the instructions for removing the tabletop below.
Narrow Hallway
If a system unit must be moved through a hallway that is not at least 45" (114cm) wide, the
C-arm can be removed. This allows the unit to fit through a hallway 29" (74cm) wide
(inside dimension). See the instructions for removing the C-arm below.
AW-00335_001-0306
3-5
Explorer Technical Manual
3. Remove the two screws (Phillips) that secure the tabletop in place (located on top of
the table near the rear center) and slide the table to the left, far enough to access the
Table X-bearing blocks.
4. Referring to Figure 3-4, remove the right-side, front endcap (3 Phillips screws).
5. Remove the back Phillips screw from the right-side cover (of the table X-drive assem-
bly) and slide the cover out from the front.
6. Unplug the cable to the Motor Controller board.
7. Refer to Figure 3-4. Disconnect the stainless steel cable guard (2 screws) and remove
the Table X-drive attachment bracket (4 bolts and 2 nuts).
8. Remove the left rail stop (1 counter sunk screw located next to the left-side, front end-
cap) to allow the tabletop to be rolled off from the front.
9. Remove the 8 Allen screws (6mm) holding the table X-drive bearing blocks.
10. Remove the tabletop by carefully sliding it off the front of the unit (requires two peo-
ple). Be careful not to slide the bearing blocks off the rail.
11. Refer to Figure 3-4. Install the right angle bracket (found in the miscellaneous hard-
ware kit) to hold the bearing blocks and X-drive bracket in place while the scanner is
moved. If this bracket is not available, tape the bearing blocks, and X-drive bracket in
place.
KP1085_003-0307
3-6
Explorer Technical Manual
Do not tilt the unit to the vertical position until the Table End Bracket is properly installed
(step 4 below).
Before the system can be tilted to vertical, you must slide the table to the left and clamp it
so it clears the floor when the unit is tilted to vertical. To accomplish this, do the
following:
1. Remove the tabletop pad.
2. Remove the three Phillips screws holding the tabletop. Two of the screws are at the
left corners and one at the right back corner looking down on the tabletop.
3. Slide the tabletop to the right until it clears the table base panel.
4. Referring to Figure 3-8, adjust the Table End Stop as pictured.
3-7
Explorer Technical Manual
KP1085_003-0309
AW-00335_001-0310
3-8
Explorer Technical Manual
Warning: Do not do this alone. This requires two people. DO NOT use the right end
of the frame to lower the unit. Use the front and rear frame rails as handholds.
AW-00335_001-0311
Setting Up Unit
Follow the procedure below to set up the unit:
1. Carefully tilt the unit down (if the room is tight, install the dolly wheels supplied to
tilt the unit down and move sideways simultaneously).
Warning: The unit will feel heavier when putting it down than it felt when tipping it
up. This is because the pivot point is different.
2. Remove the tilt arms and position the unit in its final location in the room.
3. Level the unit.
Note: Refer to Figure 3-8 and:
3-9
Explorer Technical Manual
3-C-Arm Shipping
Bracket
3
1 2-Hold Down
Bracket (4)
4
1-Table Locking
“L” Bracket
KP1085_003-0312
3-10
Explorer Technical Manual
Install C-Arm
Follow the procedure below to install a C-arm assembly:
Warning: Do not remove C-arm shipping bracket (see Figure 3-8) until done.
1. Carefully place the C-arm on the bottom bolts. Tilt the arm back slightly and slip the
cables through the openings and then tilt the arm forward until the front bolts are in
place.
2. Put 7/16” nuts and washers on all 4 bolts and tighten.
3. Install the cables.
4. Install the T-beam end caps.
5. Insure that the scanner is level.
6. Reinstall the pedestal covers.
7. Install the tank cover.
Install Cables
Follow the procedure below to cable the system:
1. Locate the box shipped with the system containing the Voltage Selection Kit that
includes the AC Input cable, Main Circuit Breaker, jumpers wires, and mounting
hardware.
2. Locate and install the feet on the bottom of the Torroid Assembly.
3. Confirm that you have the proper kit for the site’s AC voltage (100/120/230VAC).
4. Remove foot end cover on the Electronics Tray (on the left side facing the machine)
and the rear of the Torroid Assembly.
5. Remove the AC input cable from the kit, install the strain relief, and route the cable
though rear of the Torroid Assembly.
6. Screw the strain relief into the bushing at the bottom of the Torroid Assembly and fix
the cable to the exposed stud using the cable clamp and NyLock nut provided in the
kit.
7. Attach the ground wire to the grounding lug using the hardware provided. Attach the
neutral and phase wires to the two-terminal block.
8. Install the circuit breaker from the voltage kit into the hole in the rear of the Torroid
Assembly using the screws provided in the kit
9. Attach the two wires (brown and blue) from the two-terminal block to the Line termi-
nals (top terminals) on the circuit breaker.
3-11
Explorer Technical Manual
10. Attach the two wires (brown and blue) from the 7-terminal block to the Load termi-
nals (bottom terminals) on the circuit breaker.
11. Use the jumpers provided to configure the terminal block as shown in the tables
below.
Table 3-1. 120 Volt 50/60 Hertz
Wire From To
Blue Circuit Breaker - Load Pin 1 – Header 7
120 VAC
50/60 Hz
Wire From To
230 VAC
50/60 Hz
Wire From To
Blue Circuit Breaker - Load Pin 1 – Header 7
100 VAC
50/60 Hz
7 7 7
6 6 6
BRN BRN
TO 5 TO 5 TO 5
4 BLU 4 BLU 4 BRN
TRANSFORMER TRANSFORMER TRANSFORMER
3 3 3
2 2 2
1 1 1
BLU ( CIRCUIT BREAKER ) BLU ( CIRCUIT BREAKER ) BLU ( CIRCUIT BREAKER )
HEADER 7 HEADER 7 HEADER 7
3-12
Explorer Technical Manual
15. Attach the Console Power Cord to the top power receptacle. The bottom power recep-
tacle is used in the manufacturing process to power a light when X-rays are on.
16. Route the Operator Console Communications Cable through the cable clamp block on
the rear of the Scanner.
Aux To
X-Ray Console
Light AC
BLOCK 1
NORTH
BLOCK A
NORTH
GNH GNH
A B
GNH GNH
BLOCK 22
SOUTH
BLUE
AC BROWN BLOCK B
BLACK SOUTH
IN GRN/YEL
GRN/YEL
BLUE To Multi DC
Power Supply
BROWN
KP1085_003-0315
3-13
Explorer Technical Manual
Plug the Operator Console AC Supply Cable to the Power Strip male connector.
SAFETY PRECAUTIONS
There are a number of safety precautions that MUST be observed when servicing the
Explorer systems.
HIGH VOLTAGE: Voltage levels that can injure or be fatal are present through the
Explorer systems. The line voltage (100, 120, 230 volts) is supplied to the Torroid
Assembly and the scanner. The X-ray source unit contains 140kV as well as other AC
and DC voltages. The pedestal motors use 230 volts regardless of the line voltage. Use
caution when checking, calibrating, and troubleshooting. Always trip the main breaker
when replacing components.
X-RADIATION: Service personnel are required to wear a dosimeter. Do not leave the
system unattended in X-RAY SURVEY (SURVEY mode).
ESD PRECAUTIONS: To prevent damage due to ESD (Electrostatic Discharge), you
must take precautions when handling components. Remove any charge from your body
by wearing an approved and properly grounded wrist strap. Keep PCBs in their ESD
protective bag until you are ready to install them. Treat defective PCBs as new to
prevent any additional damage.
3-14
Explorer Technical Manual
Install Computer
Follow the procedure below to install the computer:
1. Set up the computer cart using the enclosed instructions. Place the computer,
keyboard, monitor, and printer on the cart.
2. Install computer system and power cables.
3. Turn on the computer and confirm that all options shipped with the instrument or
listed on the sales order have been installed and function properly.
Note: For distributor supplied computers refer to Appendix A.
KP1085_003-0317
3. Click the Start button on the desktop and then select Search for Files or
Folders...
4. In the Search Results window, select All files and folders, type service in the
All or part of the file name: edit box, and then click on the Search button.
3-15
Explorer Technical Manual
5. The software now searches for SERVICE. A list in the right panel will show you
the files found. In the list, click the “service” item shown in C:\QDR\Utilities
and drag it to the desktop. Close the Search window and then double-click the
Service icon on the desktop. The software will now be restarted in service mode.
Warning: Although the test voltage is low, there are elevated voltages near and around
the test points.
3-16
Explorer Technical Manual
AW-00336_001-0318
AW-00336_001-0319
3-17
Explorer Technical Manual
Warning: Although the test voltage is low, there are elevated voltages near and around
the test points.
3-18
Explorer Technical Manual
AW-00336_001-0320
AW-00336_001-0321
3-19
Explorer Technical Manual
2. Turn off the Explorer computer power and main circuit breaker on the rear of the
Torroid Assembly.
3. Remove the Electronics Tray covers.
4. Loosen the two mounting bolts holding the belt tension block.
5. Adjust the tension nut so that the spring is compressed to 7/8 inch.
6. Tighten the two mounting bolts holding the tension block.
Calibrate Motors
The SQDRIVER program provides a CALIBRATE command for each of the motors (AY,
TY and TX) to calibrate the encoder readback and determine the limits of motion. These
values are loaded into the SQDRIVER.INI file.
Perform the Motor Calibration procedure in the ALIGNMENT AND CALIBRATION
section of this manual.
Note: During installation, the MOTOR$AY, MOTOR$TX and MOTOR$TY procedures
may be performed without unclamping the drive shaft tubing (this allows the
procedure to be completed more quickly). This applies only during installation.
3-20
Explorer Technical Manual
3-21
Explorer Technical Manual
1. Place the meter in the center of the table approximately ½ inch to the right of the
laser crosshair.
2. Click the Perform Exam button, create a test patient biography, and select the
Array spine scan.
3. Set scan length for 9.5 inches and click the Start scan button.
4. The meter is now scanned.
5. Observe the scan to verify that the entire chamber of the meter is included in the
scan.
6. Record the result from the meter.
Note: The array spine scan dose should be less than 200µGy (20mrad).
7. Record the highest reading in the service report.
Note: Enter this information into The Radiation Measurement Report (see page 3-
26).
3-22
Explorer Technical Manual
KP1085_003-0322
3-23
Explorer Technical Manual
3-24
Explorer Technical Manual
Note: Enter this information into The Radiation Measurement Report (see page
3-26).
9. Click Xrays (F2) to turn X-rays off.
Perform QC
Once the machine has been fully assembled and calibrated, at least one QC scan should be
performed.
• Perform the daily QC procedure and enter the scan into the QC database.
3-25
Explorer Technical Manual
Note: A copy of the scan on a floppy disk and the printout must be included along with the
other information and printouts sent to Hologic after installing a new instrument or
after performing Preventive Maintenance visit.
3-26
Explorer Technical Manual
Section 4
ALIGNMENT AND CALIBRATION
Table Alignment
Checking Table Alignment
To check the table alignment, perform the following procedure:
1. Using a measuring tape, and referring to Figure 4-1, check the following:
• Distance from the edge of the table to the back of both T-rails (“A” dimension).
• Distance between the T-rails (front and rear).
• Gap from the edge of the table bracket (left side) to the rail.
2. Record all the measurements.
3. Facing the front of the Explorer, gently push the foot end (left side) of the table. The
table should move away from, and then back, to its original position.
4. Check the “A” dimension and the bracket-to-guide rail gap again. Compare them to their
original values.
5. If all the measurements are within specification, the table is properly aligned. If the mea-
surements are not within specification, go to the Aligning Table section.
Aligning Table
After you have taken the measurements in the Checking Table Alignment section, use the
procedures below to align the table. Note that if both the “A” dimension and the bracket-to-guide
rail gap are out of specification, you should recheck the measurements after performing the first
adjustment.
4-1
Explorer Technical Manual
KP1085_003-0401
Studs and
T-Rail Adjustment Nuts
Upper Bracket
Pedestal
Lower Bracket
KP1085_003-0402
4-2
Explorer Technical Manual
Warning: X-rays are being generated during this procedure. Keep hands, head and
other body parts out of beam.
Insert the alignment fixture (see the following three figures) into the detector opening.
4-3
Explorer Technical Manual
Alignment Pins
Ho;ding Clamps
KP1085_003-0403
4-4
Explorer Technical Manual
A C
Alignment Pin
Holding Clamp
Fixture
B, D
View from Front
Step A. Insert the left side of the Alignment Fixure into the left side of the
Detector Opening so that the vertical edge of the four Alignment
Pins are secure.
Step B. Compress the Holding Clamps.
Step C. Raise the right side of the Alignment Fixture into the Dectector
Opening.
Step D. Release the Holding Clamps.
Note:. If the Alignment Fixture is insertrd the the Alignment Pins on the
right, the procedure works equally well.
KP1085_003-0404
KP1085_003-0405
4-5
Explorer Technical Manual
KP1085_003-0406
Warning: The X-rays are on. Keep body parts out of the beam.
1. Move the front Filter Drum alignment screw until the X-ray signal reaches max-
imum voltage.
Note: The last direction turned should be clockwise to eliminate backlash when
the jam nuts are tightened. The plot should show peak amplitude.
2. Move the rear Filter Drum alignment screw until the X-ray signal reaches maxi-
mum voltage.
4-6
Explorer Technical Manual
Note: The last direction turned should be clockwise to eliminate backlash when
the jam nuts are tightened. The plot should show peak amplitude.
3. Tighten the jam nuts on both Filter Drum alignment screws.
Note: The X-rays should still show peak amplitude.
4. Click Xray (F2) to turn X-rays off.
5. Remove the alignment test fixture.
4-7
Explorer Technical Manual
Motor Calibration
The SQDRIVER program provides a CALIBRATE command for each of the motors (AY,
TY and TX) to calibrate the encoder read back and determine the limits of motion.
Each motor requires the corresponding protocol calibration file in the PROTOCOL sub-
directory (e.g., for MOTOR$AY, the calibration protocol is MOTOR_AY.PRO).
To perform the calibration procedure:
1. From the Explorer Main Menu screen, select Utilities|Service Utili-
ties|SQDRIVER (you must be in Service Mode).
2. At the CARM$$$$> prompt, type MOTOR$XX<Enter>, where XX equals AY,
TY, or TX depending on which motor you are calibrating.
Note: If you are calibrating motors during an installation, you must calibrate them in the
proper order. The proper order is AY, TY, andTX.
Perform the calibration procedures if indicated (*) in order from left to right.
The program moves the AY motor to the left. When AY hits the left mechanical stop
the first time:
3. (For encoder repairs only) Press <Esc>.
The program returns to the MOTOR$AY menu page (below).
In the right hand column of the display, eight lines from the bottom (and highlighted
in the figure below), is the raw a/d readback of the motor encoder (in the example
below, it is 2048). Manually rotate the encoder until the readback is 3750±50 (i.e., in
the range 3700-3800) and then clamp the encoder to the drive shaft.
4. (For encoder repairs only) Repeat the calibration procedure above but now, when
AY hits the left mechanical stop, press <Enter>.
4-8
Explorer Technical Manual
When the calibration scan completes, the program computes the linear fit for both the
positive and negative motion. The linear fit parameters are displayed at the top left
and top right of the plot in the form Y = Intercept + Slope * X. The two slopes (e.g.,
86.6 and 86.5) should be within 0.3 of each other.
4-9
Explorer Technical Manual
E
N +
+ ++++
C ++ +
O
+ ++++
D ++ +
E +
++
+++
R ++
+
+
++
++
R +
+++
E ++++
+++ +
A
D + +++
+
+ + +
I +++
N +++
++
G ++
++++
++
+
+
++
++
++
++
KP1085_003-0408
The Program displays the positive and negative limits as horizontal dashed limits.
Note: The PosLimitOffset and NegLimitOffset entries in the [AyMotor] section of the
SQDRIVER.INI file determine the motion limits relative to the mechanical stops. If
these entries are not present, or are zero, the motion limits are set to the mechanical
limits).
The program then changes the plot title to PRESS ANY KEY TO CONTINUE. Press the
<Enter> key and the program prompts:
motor_direction=1
calibrate_position=1,2288,41187,209,61339,363,363,3719
pos_limit_position=984946
neg_limit_position=0
Update Driver INI-File [Y/N]?
The motor_direction field determines whether the positive step direction is the reverse of
the direction of increasing encoder values ('0' if the direction is not reversed, '1' if it is).
This value is set from the corresponding parameter in the MOTOR_AY.PRO file and is a
constant for each motor (i.e., the direction of increasing encoder values must agree with
the system coordinates and the positive step direction is set by the wiring harness for the
stepper motor).
The eight calibrate_position fields are:
4-10
Explorer Technical Manual
4-11
Explorer Technical Manual
4. (For encoder repairs only) Repeat the calibration procedure above but now, when
TY hits the left mechanical stop, press <Enter>.
4-12
Explorer Technical Manual
E
++ +
N +++ ++
C + + +
O ++ ++
+ + +
D ++
E ++++
R +
+ +
++
++ ++
R + + +
++ +
E
++ ++
A +
D ++
+
+
I + ++
++
N +++
G +++
+
+ + +
+++
+++
+
+++ +
+++
KP1085_003-0409
When the calibration scan completes, the program computes the linear fit for both the
positive and negative motion. The linear fit parameters are displayed at the top of left
and top right of the plot in the form Y = Intercept + Slope * X. The two slopes (e.g.,
86.5 and 86.5) should be within 0.3 of each other.
The program displays the positive and negative limits as horizontal dashed lines.
The PosLimitOffset and NegLimitOffset entries in the [TyMotor] section of the
SQDRIVER.INI file determine the motion limits relative to the mechanical stops. If
these entries are not present, or are zero, the motion limits are set to the mechanical
limits.
The program then changes the plot title to PRESS ANY KEY TO CONTINUE.
4-13
Explorer Technical Manual
The program moves the TY (W only) and AY motors to their center positions and
prompts:
Use the Table IN / OUT switch to move the C-Arm to 0 degrees.
Press <Enter> when the C-Arm is positioned.
Press <ESC> to stop calibration.
The above message is always displayed when calibrating MOTOR$TX. This is the
normal message for a QDR-4500. The message was not changed for Explorer
systems.
3. Press <Enter> and the program prompts:
Press <Enter> when the TX motor reaches the OUTER mechanical limit.
Press <ESC> to stop calibration.
4-14
Explorer Technical Manual
4-15
Explorer Technical Manual
E
N
C
O
D +++
E +++ +
R +++ +
++ +
+++ ++
R
++ +
E
+++ ++
A ++ +
D + ++
I +++ +
+++
N +++
+++ +
G +
KP1085_003-0410
When the calibration completes, the program computes the linear fit for both the
positive and negative motion. The linear fit parameters are displayed at the top left
and top right of the plot in the form Y = Intercept + Slope * X. The two slopes (87.0
and 86.9 in the example below) should be within 0.3 of each other.
The program displays the positive and negative limits as horizontal dashed lines.
Note: The PosLimitOffset and NegLimitOffset entries in the [TxMotor] section of the
SQDRIVER.INI file determine the motion limits relative to the mechanical
stops. If these entries are not present, or are zero, the motion limits are set to
the mechanical limits.
The program then changes the plot title to PRESS ANY KEY TO CONTINUE.
Press the <Enter> key and the program prompts:
motor_direction=1
calibrate_position=1,2287,27446,43,12563,1026,1026,3096
pos_limit_position=604777
neg_limit_position=0
Update Driver INI-File [Y/N]?
4-16
Explorer Technical Manual
Note: The last calibration scan data is saved in the file MOTOR_TX.DAT. You can
reanalyze the data–e.g., after editing SQDRIVER.INI–by typing the command
CALIBRATE @MOTOR_TX.DAT at the MOTOR$TX> prompt in
SQDRIVER.
4-17
Explorer Technical Manual
Detector Flattening
Warning: X-rays are being generated during this procedure. Keep hands, head and
other body parts out of beam.
This procedure flattens the X-ray beam for each scan mode and finds the metal edge of the
table.
Note: All covers, table mat, etc., normally in the X-ray path, MUST be on the Scanner
before running Beam Flattening.
1. Restart the Explorer software in service mode (if not already). Press the Center
Table button and turn on the laser.
2. Place the phantom on end (vertical) with the spinal processes pointing towards
the head end of the table. The laser crosshair should be 1.5 inch in from the left
end and centered. (Some phantoms will have a target hole, if not, use a ruler).
3. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
4-18
Explorer Technical Manual
4-19
Explorer Technical Manual
KP1085_003-0412
4-20
Explorer Technical Manual
3. Compare values displayed with actual values printed on the phantom. If there is a
discrepancy, check to make sure you have the phantom that shipped with the unit.
Check with Hologic Technical Support if needed.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
4-21
Explorer Technical Manual
11. Select the Spine Phantom scanned in step 6 from the list of Analyzed Scans and
click the Next button. Click the Continue button on the Successfully Calcu-
lated message box.
12. Click on the Temp Write button and click the Continue button on the Success-
fully Wrote message box. Exit Field Service Calibration.
4-22
Explorer Technical Manual
4-23
Explorer Technical Manual
4-24
Explorer QDR Series Technical Manual
4-25
Explorer QDR Series Technical Manual
4-26
Explorer QDR Series Technical Manual
4-27
Explorer QDR Series Technical Manual
4-28
Explorer Technical Manual
Section 5
REMOVE AND REPLACE
PROCEDURES
This section describes how to remove and replace the Field Replaceable Units (FRUs) in
the Explorer. To safely perform a FRU removal or replacement, take care to follow the
procedure precisely as written.
Note: Whenever a component in the x-ray path is replaced, you must rerun QC and
recalibrate.
Recommended Tools
Tool Size/Type
Hex driver 3/32”
Hex driver or wrench 5/32”
Nut driver 1/4”
Nut driver 3/8 inch
Nut driver 5/16”
Nut driver 7/16”
Screwdriver Narrow slotted
Screwdriver Phillips head
Screwdriver Slotted
Wrench 3/8 inch
5-1
Explorer Technical Manual
C-Arm Y Belt
To remove and replace the C-arm Y-Belt, refer to Figure 5-1and Figure 5-2 and follow the
procedure below:
1. Move the C-arm to the center of the table.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. At the encoder end of the belt, loosen the 2 bolts holding the tension block,
remove the tensioning nut, and remove and replace the belt.
4. Install the tension spring and tensioning nut.
5. Tighten the tension nut so that the spring compresses to 7/8 inch. The bracket cut-
out can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
6. Tighten the two mounting bolts holding the tension block.
7. Perform the MOTOR$AY calibration procedure.
00131-C0501
5-2
Explorer Technical Manual
00131-C0502
5-3
Explorer Technical Manual
11. Tighten the tension nut so that the spring compresses to 7/8 inch. The bracket cut-
out can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
12. Tighten the two mounting bolts holding the tension block.
13. Restore the cable shield and ground strap terminations.
14. Perform the MOTOR$AY calibration procedure.
C-Arm Y Encoder
To remove and replace the Encoder, refer to Figure 5-1 and follow the procedure below:
1. Before removing power from the Explorer, remove both cable covers, unhook the
center cable hold-down clip and free the encoder cable without unplugging it.
2. Move the C-arm towards the center of the Scanner (so the encoder is accessible).
3. Turn off the computer power and the main circuit breaker.
4. Unplug the encoder from the Motor Controller Board and pull out the cable.
5. Remove the coupling hardware holding the encoder to the belt drive.
6. Remove the encoder from the bracket assembly.
7. Replace the encoder on the bracket, install the coupling but do not tighten the
screws.
8. Perform the MOTOR$AY calibration procedure.
Note: When starting this procedure, make sure the encoder is not clamped to the drive
shaft.
Table Y FRUs
This section describes how to remove and replace the FRUs associated with the Table Y
motion of the Explorer (see Figure 5-3 and Figure 5-4).
1. Before removing power from the Explorer, move the table to the far left using the
Motor Control Pad. The Motor Control Pad can be accessed from the Explorer
Main Menu screen by choosing Utilities|Emergency Motion.
2. Turn off the Explorer power at the main circuit breaker.
5-4
Explorer Technical Manual
3. Remove 2 screws from the right table rail end cover, and slide the cover off from
the end.
00131-C0503
Table Y Belt
To remove and replace the Table Y-Belt, refer to Figure 5-3 and Figure 5-4 and follow the
procedure below:
1. Center the table and remove the two screws that secure the tabletop in place.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. Pull the tabletop to the left far enough to remove the right table rail end cover,
and remove the cover. Mark this as the right cover so that it is not confused later
with the left cover (they are not interchangeable).
4. Pull the tabletop to the right far enough to remove the left table rail end cover,
and remove the cover. Mark this as the left cover so that it is not confused later
with the right cover (they are not interchangeable).
5. At the encoder end of the belt, loosen the 2 bolts holding the tension block,
remove the tensioning nut, and remove and replace the belt.
6. Install the tension spring and tensioning nut.
7. Tighten the tension nut so that the spring is compressed to 7/8 inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
5-5
Explorer Technical Manual
5-6
Explorer Technical Manual
KP1085_003-0504
Table Y Encoder
To remove and replace the Table Y Encoder, refer to Figure 5-3 and Figure 5-4 and follow
the procedure below:
1. Center the table and remove the two screws that secure the tabletop in place.
2. Turn off computer power and the main circuit breaker on the Power Module.
3. Pull the tabletop to the left far enough to remove the right table rail end cover,
and remove the cover. Mark this as the right cover so that it is not confused later
with the left cover (they are not interchangeable).
4. Pull the tabletop to the right far enough to remove the left table rail end cover,
and remove the cover. Mark this as the left cover so that it is not confused later
with the right cover (they are not interchangeable).
5. Unplug the encoder from the Motor Controller Board and pull out the cable.
6. Remove the coupling holding the encoder to the belt drive.
7. Remove the encoder from the bracket assembly.
8. Replace the encoder on the bracket, install the coupling but do not tighten the
screws.
9. Perform the MOTOR$TY calibration procedure.
Note: When starting this procedure, make sure the encoder is not coupled to the drive
shaft tubing.
5-7
Explorer Technical Manual
Table X FRUs
This section describes how to remove and replace the FRUs associated with Table X
motion of the Explorer (see Figure 5-5).
Table X Encoder
Table X Belt
Front Endcap Motor Controller Board
Table X Motor
KP1085_003-0505
Table X Belt
To remove and replace the Table X Belt located within the X Table Drive Assembly, refer
to Figure 5-5 and follow the procedure below:
5-8
Explorer Technical Manual
1. Before removing power from the Explorer, move the table to the far left using the
Motor Control Pad. The Motor Control Pad can be accessed from the Explorer
Main Menu screen by choosing Utilities|Emergency Motion.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. Remove 2 screws from the right table rail end cover, and slide the cover off from
the end.
4. Remove the front endcap from the table X drive assembly (3 Phillips screws).
5. Remove the back Phillips screw from the right side cover (of the table X drive
assembly) and slide the cover out from the frontEmotionalisthis provides access
to the belt.
6. At the encoder end of the belt, loosen the 2 bolts holding the tension block,
remove the tensioning nut, and remove and replace the belt.
7. Install the tension spring and tensioning nut.
8. Tighten the tension nut so that the spring is compressed to 7/8 inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
9. Tighten the two mounting bolts holding the tension block.
10. Perform the MOTOR$TX calibration procedure.
Note: When starting the MOTOR$TX calibration procedure, make sure the encoder is not
clamped to the drive shaft tubing.
5-9
Explorer Technical Manual
10. Tighten the tension nut so that the spring is compressed to 7/8 inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
11. Tighten the two mounting bolts holding the tension block.
12. Restore the cable shield and ground strap terminations.
13. Perform the MOTOR$TX calibration procedure.
Note: When starting the MOTOR$TX calibration procedure, make sure the encoder is not
clamped to the drive shaft tubing
Table X Encoder
To remove and replace the Table X Encoder located within the X Table Drive Assembly,
refer to Figure 5-5 and follow the procedure below:
1. Before removing power from the Explorer, move the table to the far left using the
Motor Control Pad. The Motor Control Pad can be accessed from the Explorer
Main Menu screen by choosing Utilities|Emergency Motion.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. Remove 2 screws from the right table rail end cover, and slide the cover off the
end.
4. Remove the front endcap from the X table drive assembly. The cover is held on
with 2 hex screws.
5. Remove 6 Phillips screws from the right side cover (of the X table drive assem-
bly) and slide the cover out from the front. This provides access to the encoder
and belt.
6. Remove 5 flat head Phillips screws from the upper pedestal cover and remove the
cover. This provides access to the Motor Controller Board.
7. Unplug the encoder from the Motor Controller Board and pull out the cable.
8. Remove the coupling holding the encoder to the belt drive.
9. Remove the encoder from the bracket assembly.
10. Replace the encoder on the bracket, install the coupling but do not tighten the
screws.
11. Perform the MOTOR$TX calibration procedure.
Note: When starting the MOTOR$TX calibration procedure, make sure the encoder is not
coupled to the drive shaft tubing.
5-10
Explorer Technical Manual
5-11
Explorer Technical Manual
00131-C0504
5-12
Explorer Technical Manual
6. Remove the rear tank cover (covers the C-arm Interface board).
7. Unplug the 2 Filter Drum cables from the C-arm Interface board.
8. Remove 3 hex head screws (3/32 inch Allen screws).
9. Remove the Filter Drum by lifting it up (while tilting it slightly forward) and out.
10. Position the replacement Filter Drum Assembly into the unit.
11. Replace the 3 hex head screws. Ensure that the curved spring washers are placed
(curved downward) so that the washer compresses when the screw is tightened.
12. Replace the cables.
13. Replace the screws in the XRC assembly.
14. Perform the X-Ray Beam Alignment procedure in the Alignment and Calibration
section of this manual.
15. Perform the A/D Gain Control Adjustment in the Alignment and Calibration sec-
tion of this manual.
16. Perform Detector Flattening in the Alignment and Calibration section of this
manual.
Tank Assembly
To remove and replace the Tank Assembly, refer to Figure 5-6 and Figure 5-7 and follow
the procedure below:
WARNING: Because of the weight of the tank (about 200lbs), 2 people are required to
safely remove and replace the tank
1. Using the Control Panel, move the table in as far as it will go for easier access.
2. Turn off the Explorer computer and the main power circuit breaker.
3. Remove the lower C-arm cover by removing 2 Phillips screws and sliding it out
from the front.
4. Remove the XRC (see procedure above).
5. Remove the Filter Drum Assembly (see procedure above).
6. Lock the C-arm in place by securing the shipping brackets on each side of the
arm.
7. Remove the two L-brackets from the front of the tank that holds the tank to the
C-arm (7/16 inch).
8. Carefully slide the tank out of the C-arm onto a pallet (requires 2 people).
9. Remove the old tank from the tank tray, and install the new tank on the tray.
5-13
Explorer Technical Manual
KP1085_003-0510
5-14
Explorer Technical Manual
Detector Assembly
To remove and replace the Detector Assembly (refer to Figure 5-8) and follow the
procedure below:
1. Turn off the Explorer computer and the main power circuit breaker.
2. Remove the top C-arm cover.
3. Remove the cables at the Detector Assembly.
4. Remove the 4 bolts holding the Detector Assembly (on rubber grommets) to the
C-arm and then remove the assembly.
5. Replace the Detector Assembly (reverse the steps).
6. Perform the A/D Gain Control Adjustment in the Alignment and Calibration sec-
tion of this manual.
7. Perform Detector Flattening in the Alignment and Calibration section of this
manual.
00131-C0505
Laser Assembly
To remove and replace the Laser, or Laser Assembly, refer to Figure 5-8 and Figure 5-9
and follow the procedure below:
5-15
Explorer Technical Manual
1. Turn off the Explorer computer and the main power circuit breaker on the Power
Module.
2. Remove the top C-arm cover.
3. Remove the Detector Assembly (see procedure above).
Laser Assembly
Laser
00131-C0506
Control Panel
To remove and replace the Control Panel refer to Figure 5-8 and follow the procedure
below:
1. Turn off the Explorer instrument power, computer power, and main circuit
breaker.
2. Remove the top cover of the C-arm.
3. Unplug the cables from the Control Panel to the Detector Assembly.
4. Remove the screws securing the Control Panel to the C-Arm and remove the
panel.
5. To replace the Control Panel reverse the steps.
5-16
Explorer Technical Manual
KP1085_003-0518
Drum Belts
To remove and replace either Drum Belt (130 teeth or 150 teeth), refer to Figure 5-11 and
follow the procedure below:
1. Remove the Filter Drum Assembly (see procedure above).
2. Remove the aperture by removing 4 Phillips screws that secure it.
3. Loosen the 2 drum motor mount screws, the idler screw and the outer pulley set
screw. Remove the belts.
5-17
Explorer Technical Manual
KP1085_003-0519
4. Install the Filter Drum Alignment Pin (099-0110), small end first, through the
slotted holes and into the small hole at the base plate. If the pin is installed prop-
erly, the drum will not rotate.
5. Install the 150 tooth belt on the back pulley (ensure the belt is positioned under
the idler).
6. Install the 130 tooth belt on the front pulley.
7. Tighten the 2 drum motor mount screws (the motor mount is spring loaded to
seek proper tension).
8. Tighten the idler screw (the idler is spring loaded to seek proper tension).
9. Tighten the outer pulley set screw.
10. Remove the Filter Drum Alignment Pin.
5-18
Explorer Technical Manual
5. Replace the pulley, use Loctite 222 (540-0100) on the set screw.
6. Refer to the Drum Belts procedure above and replace the belts.
Drum Bearings
To remove and replace the drum bearings, refer to Figure 5-11 and Figure 5-12 and follow
the procedure below:
1. Remove the Filter Drum Assembly (see procedure above).
2. Remove the aperture and drum belts (see procedures above).
3. Remove the 2 drum motor mount screws (see Figure 5-12).
4. Remove 6 screws from underneath the drum assembly base plate (4 screws hold
the end plates and 2 screws hold the motor mount spring tensioner block).
5. Remove the side plates and lead shields (3 Phillips screws on each side).
6. Remove the drum encoder PCB and both encoder wheels.
7. Remove both drum belt pulleys.
8. Remove the drum from the endplates.
9. Remove one drum endcap (4 Phillips screws) and remove the inner drum.
10. Replace the bearings.
11. Reassemble the drum and replace (and tighten) the screws in the endcap.
Note: When replacing the end cap ensure that the flats on each side of the drum shaft are
facing the same way (this happens where the two side-by-side holes line up).
12. Replace the 2 endplates. The endplate with the cutout (for the PCB) goes on the
drum end away from the alignment hole (see Figure 5-12).
5-19
Explorer Technical Manual
KP1085_003-0520
5-20
Explorer Technical Manual
KP1085_003-0521
5-21
Explorer Technical Manual
KP1085_003-0522
FRU Lists
The following tables provide the information necessary to identify and order the correct
FRU.
Table 5-1. Electronics Tray FRUs (Figure 5-1)
5-22
Explorer Technical Manual
Table 5-5. Upper C-Arm FRUs (Figure 5-8 and Figure 5-9)
5-23
Explorer Technical Manual
5-24
Explorer Technical Manual
5-25
Explorer Technical Manual
5-26
Explorer Technical Manual
Section 6
FAULT ISOLATION
This section provides information to help identify the source of a problem in the Explorer
system. The three general categories are:
Problem... Refer to the section titled... On page...
Dead system, or power problem Power Problems 6-1
Scanner motion problem Motion Problems 6-2
Computer display problem Display Problems 6-4
Before Starting
Before starting, make sure the software configuration is compatible with the scanner.
Software Configuration
To check the software revision and loaded options, from the Explorer Main Screen, select
Utilities|System Configuration. On the System Configuration screen, you will find a
drop-down list box labeled “Software:” By dropping down this list, you will be shown the
release version of the Explorer software and all loaded options. The release version should
be at a minimum level of 12.2.
Hardware Configuration
When troubleshooting, it is sometimes helpful to observe the indicators available on the
PCBs and other FRUs. Many components of the system have LEDs indicating the
presence of necessary voltages and the state of some signals. Section 8 is helpful in
locating these LEDs and observing the state of the system.
Power Problems
Table 6-1 Power Component Locations
Component... Where…
Main circuit breaker Rear of foot end pedestal
LED’s for DC voltages Distribution Board and most
driver and control PCBs
Emergency stop switches and circuits Control panel
Computer power and operation Computer console power strip
6-1
Explorer Technical Manual
Motion Problems
Motion problems are failures related to movement of the table and/or C-arm. In most (but
not all) cases, an error message accompanies Explorer motion problems. Therefore, it is
good troubleshooting practice to check the error log (C:\ERRLOG.DAT file and
6-2
Explorer Technical Manual
AY
TX
TY
AW-00335_001-0601
6-3
Explorer Technical Manual
TX, TY, AY
Check... Refer to...
Belt
Encoder coupling Figure 5-1, 5-2, 5-3, 5-4, 5-5
Motor controller board Section 8, PCB SUMMARY INFORMATION
Motor/gear case for the motor
subsystem in question
TY (on A & W models only)
Left and right end panels on table for proper
installation. This may cause problems if reversed.
(Left 030-1376, right 030-1838).
CAUTION: Running any motor subsystem to its limit (in either direction) does not
damage the motor. However, if the motor encoder is not properly calibrated,
it may be damaged.
TX, TY, AY
To... Run... Refer to...
Perform simple table and C- Motor Control Pad The Explorer Main Menu, select:
arm movements (computer motion 1. Utility
control) 2. Emergency Motion
Perform precise table and SQDRIVER Motor Calibration heading, and
C-arm movements Motor$XX (for the specific
motor), Section 4
Monitor all motion parameters
Troubleshoot problems Control Panel heading, Section 2
encountered initiating motion
from the Operator's Control
Panel
Run the hardware checker Hardware checker SQVERIFY
Display Problems
Display problems can be grouped into four general categories: vertical stripe, horizontal
stripe, noise (dots, speckles, etc.), and no display.
6-4
Explorer Technical Manual
Vertical Stripe
This type of display problem is most likely related to the detector subsystem. The
following suggestions apply to a Explorer system that exhibits a vertical stripe in the
display:
Check... Refer to...
Detectors X-Ray Survey in graphic mode to check for
signal strength and noise (Section 9)
Narrow vertical stripe - bad detector Data Acquisition System heading,
Section 2, for block diagrams and
interconnection charts or run the DAS tests
in SQVERIFY.
Foreign matter (especially metallic)
anywhere in the X-ray beam path in the
aperture slit, collimator cup, etc.
Also...
Run the hardware checker SQVERIFY
Horizontal Stripe
• This type of display problem is most likely related to the line voltage, bad cables,
or X-ray subsystem. The following suggestions apply to a Explorer system that
exhibits a horizontal stripe in the display:
Check... Refer to...
Line voltage/Power ground Check Power Line Voltage heading, page 3-14
Tube kV Peak Potential Check Tube KV Peak Potential heading, page 3-16
Tube Current Check Tube Current heading, page 3-18
Filter drum turning Figure 5-10
Filter drum belt Figure 5-11
Green LED’s on C-Arm Figure 5-6, Section 8
Interface
6-5
Explorer Technical Manual
Noise
The term “noise” is used here to describe any flaw, or irregularity in the display (dots,
specks, uneven lines, etc.) or similar problem. The following suggestions apply to a
Explorer system that exhibits noise in the display:
Check... Refer to...
Tube kV peak potential Check Tube KV Peak Potential heading, page 3-16
Tube current Check Tube Current heading, page 3-18
X-ray beam alignment X-Ray Beam Alignmentheading, page 4-3
Signal strength and noise X-Ray Survey heading, page 9-1
Filter Drum is turning
Filter Drum belt Figure 5-11
Green LED’s on C-Arm Interface Figure 5-6, Section 8
board
Aperture plate assembly, first Figure 5-6
precollimator, second precollimator
and collimator for specs of lead and
other deformities.
Also...
Run the hardware checker SQVERIFY, page 9-10
No Display
The term “no display” is used here to describe:
• no scan display
• completely white screen
• completely dark screen
• “blotchy” or “grainy” display
This type of display problem is usually related to the detector subsystem or the X-ray
subsystem.
The following suggestions apply to a Explorer system that exhibits no scan display:
Check... Refer to...
Tube kV peak potential Check Tube KV Peak Potential heading, page 3-16
Tube current Check Tube Current heading, page 3-18
X-ray production Field Service Preventive Maintenance heading,
page 7-1
6-6
Explorer Technical Manual
Signal strength and noise X-Ray Survey in graphic mode, page 9-1
Filter Drum is turning
Filter Drum belt Figure 5-11
Green LED’s on C-Arm Section 8
Interface board
No display may indicate a bad Refer to the Data Acquisition System heading on
Detector Assembly page 2-15.
Also...
Run the hardware checker SQVERIFY on page 9-10
Targeting/Laser Problems
If... Check... Refer to...
Object being Detector array. It may be too far Figure 5-8
scanned appears forward or back inside the upper arm Page 4-17
to the left or assembly. Page 4-18
right of the scan Laser alignment
window Run... Detector Flattening
6-7
Explorer Technical Manual
X-Ray Problems
The following lists some common X-ray problems and some suggested solutions. Refer to
Section Two Functional Description for more information on the X-Ray subsystem.
6-8
Explorer Technical Manual
No X-Rays
If... Check... Refer to...
X-Ray (24VDC LED) X-ray Enable LED on Control Panel and
on Distribution Board is Distribution Board should be on
not lit
“NO A/C Line Interlock inhibit
Interrupts” message at 24VDC from LVPS, power distribution
start-up X-ray Controller interlock switch
Computer problems (CommCon)
Defective Control Panel I/F
XRC Power Cable/Connector problems
6-9
Explorer Technical Manual
5. Check for bad detectors in the Detector Assembly. Run the DAS test in SQVER-
IFY.
6. Check the Filter Drum to see that it is installed correctly. Check the segment
readback values in X-Ray Survey with X-rays ON.
7. Check for lead fragments in the X-ray beam. Inspect the Collimator Cup, disk,
and Aperture Assembly.
8. If a failure occurs during a whole body, recheck the TX motor cal files. The X-
rays may be hitting the table edge. Make sure the drives are operating normally.
Examine the Detector Flattening scans for any unusual indications using Utili-
ties|Scan File Plot and selecting the WBAIRQC scan. If an error message is
displayed, go to the Detector Flattening heading on page 4-18.
Laser Problems
WARNING: The laser beam can cause serious retinal damage if focused directly into
the eye. Be sure to turn the laser OFF when visually inspecting the shutter
and aperture.
6-10
Explorer Technical Manual
If... Suggestions
(Software is inhibiting laser when table head is near the laser)
Move the C-arm to the center of the table and retry.
Do the following:
Check black and yellow wires from the C-arm Interface PCB to the
underside of the Detector Assembly.
Laser Shut off instrument power (Emergency Stop Switch on Control Panel).
does not Disconnect laser power connector.
turn on Connect a DVM to the plug coming from the Detector Assembly.
Turn laser power on.
Check connector for +5VDC.
Check C-arm Interface Board, Distribution Board, all cables.
Check computer.
At installation only, type “laser safety=0” in SQDRIVER and reboot.
Check AY motor calibration file. If data is corrupt or uncalibrated, laser may
not turn on, despite position of C-arm.
Laser Check laser block assembly mirrors for breaks, cracks, or misalignment
crosshair Check that shutter is open
beam is Check that aperture is unobstructed
defective Loosen the securing screw to adjust the laser to perpendicular (see Figure 5-
8)
Check laser button on Control Panel
Check control panel laser switch at PANEL tests in SQVERIFY
Laser Check for bad:
does not C-arm Interface Board
turn off Distribution Board
Interconnect cable
Computer assembly
Oil leakage
If the tank assembly is leaking oil, you may have to tighten the screws on the tank cover. It
is important that you tighten the screws using the proper torque specifications and the
proper sequence.
6-11
Explorer Technical Manual
Front
KP1085_003-0602
5 6
3 2
8
Figure 6-3 Lexan Cup Screw Tightening Order
6-12
Explorer Technical Manual
2. After screw #4, continue around the edge of the gasket in a clockwise direction,
tightening alternate screws, until returning to #4.
3. Continue around the edge of the gasket in a clockwise direction, tightening all the
remaining screws.
1 4
3 2
2 4
1 3
KP1085_003-0605
6-13
Explorer Technical Manual
2. Starting at the screw next to screw #6, move in a clockwise direction and tighten
alternate screws.
3. Repeat this pattern in a clockwise direction and tighten all remaining screws.
4. The three Tank Cover Gasket screws at the rear of the tank and shown below are
the most critical locations for oil leaks. Always check these screws if a leak is
suspected.
5. Important: Do not remove screws located on the tank. There are no field
replaceable units or required adjustments inside the tank, so there is no reason to
loosen or remove any screws on the tank.
MOST CRITICAL
1 3
FRONT
6 2 4
KP1085_003-0606
Miscellaneous Problems
Problem... Symptom... Solution...
Detector X-Ray Survey bar graph is Replace the Detector Assembly
ramped, or it appears as 3 or
4 separate ramped sections.
X-ray lamp Turns on with X-rays off Replace the Distribution Board
System Fails HI-Pot test Check all ground connections
Make sure ground wire terminals are
secure
Check crimped terminals
Hi voltage Buzzing sound from X-ray Check torque on transformer bolts
transformer tank assembly
System No power Check Emergency Stop Switch
Check main power cord
Check Torroid Assembly circuit breaker
Check AC input voltage
6-14
Explorer Technical Manual
Section 7
PREVENTIVE MAINTENANCE
This section lists the procedures that should be performed by trained service personnel at
least once per year, and preferably at six-month intervals.
NOTICE
The Explorer system meets applicable FDA radiation performance standards through its
useful and expected life provided no components or parts are removed from the system
and no unauthorized adjustment or unauthorized replacement of certified components is
performed.
7-1
Explorer Technical Manual
[ ] During a scan, verify that pressing the red Emergency STOP switch
immediately stops all C-Arm motion, table motion, and X-ray
production. X-ray production should be monitored by a Victoreen
450P or equivalent.
[ ] Replace (if necessary) and adjust the motor drive belts as follows:
Note: Detailed instructions for drive belt replacement and adjustments can be
found in the REMOVE AND REPLACE section of this manual.
[ ] For each drive belt (C-Arm Y, Table X, and Table Y):
1. Loosen the 2 mounting bolts holding the tension block.
2. Tighten the tension nut so that the spring is compressed to 7/8 inch. The
bracket cutout can be used as a measuring guide (the inside of the washer
should be flush with the bracket cut).
3. Tighten the two mounting bolts holding each drive belt’s tension block.
It is not necessary to perform the MOTOR calibration procedures after
tensioning the motor drive belts.
[ ] Ensure that all cable connections are tight.
[ ] Run CHECKDISK on the C drive. This utility program can be
started by clicking on start|Run…, then typing CHKDSK, and
then clicking the OK button. DO NOT run Scandisk on any
Archive media.
[ ] Run Disk Defragmenter to and compact and re-order the files on the
computer's hard disk. This utility program can be started by
clicking on start|Run…, then typing DEFRAG, and then clicking
the OK button.
[ ] Run a computer virus checker (any major brand that is current for
latest virus types).
[ ] Clean the fan filters, paying special attention to the computer fan
filter.
[ ] Clean all exterior metal surfaces and wipe off the rails.
[ ] Clean the monitor screen, keyboard, and mouse ball rollers.
[ ] Clean the inkjet printer rollers and ink cartridge electrical contacts
with isopropyl alcohol.
7-2
Explorer Technical Manual
KP1085_003-0701
Consult Hologic Field Service for more information on obtaining the proper tools,
material, and procedure.
7-3
Explorer Technical Manual
7-4
Explorer Technical Manual
Section 8
PCB SUMMARY INFORMATION
Distribution Board
LED’s1 Signal Voltage Source Jumpers Refer
to...
D1 On +7VDC Int Voltage Reg. JP1 (panel) Out
D32 Off - JP2 (dist) Out
D5 On +5VDC Int Voltage Reg. JP3 (C-arm) Out
D7 On +28VDC Int Low Voltage Figure
Power Supply Circuit Breakers 5-1
D9 On +15VDC Int Low Voltage CB
Power Supply
D10 On -15VDC Int Low Voltage 1
Power Supply
- - +24VDC Int Low Voltage 2 TY In
Power Supply
D112 Off TY DRIVE 3 TX In
D132 Off 5 AY In
8-1
Explorer Technical Manual
Communications Controller
Jumpers Refer to...
JP1 (IRQ) at 10 In
JP2 (IRQ) Out
JP3 (IREQA) Out Figure 5-17
JP4 (IREQB) In
JP5 (E-OUT) at In
C
8-2
Explorer Technical Manual
8-3
Explorer Technical Manual
C-Arm Interface
LED’s Signal Voltage Source Jumpers Refer to...
D1 On +28VDC Ext Signal JP7 DRUM Out
Dist.
D3 On +24VDC Ext Signal JP8 MAIN Out
Dist.
D6 On +5VDC Int. Voltage W1, NORMAL(1 In
Reg. W2 -2)
TEST (2-3)
- - +7VDC Int. Voltage W3 SYS- In
Reg. RESET
- - +15VDC Ext Signal W4 TEST- Out
Dist. MODE
- - -15VDC Ext Signal W5 Safety In
Dist. switches
Drum JP15 Redundant for
D7 On LOCKED W5 may be Out if
W5 is In
D8 Flash DRUM
-ing AT TOP
X-Ray
D9 Off X-RAY
CPU
ERROR
D10 Off X-RAY
ENABLE
D11 On X-RAY
CPU OK
D12 Off X-RAY
FAULT
8-4
Explorer Technical Manual
Section 9
SOFTWARE TOOLS
The Explorer system software includes software tools to troubleshoot the system. They
are:
• SQDRIVER
• SQVERIFY
• Emergency Motion (SQKEYPAD)
X-Ray Survey
The X-Ray Survey is a 32-bit utility that tests the functions of the X-Ray generation and
detection for the system. This utility is used by Hologic manufacturing and service
engineers only.
AW-00335_001-0902
Display Controls Close utility
9-1
Explorer Technical Manual
AW-00335_001-0902
9-2
Explorer Technical Manual
(for field engineering purpose this will always be Air Segment). The bar graph shaded
green is hi voltage, the bar graph shaded brown is lo voltage.
AW-00335_001-0903
AW-00335_001-0904
9-3
Explorer Technical Manual
AW-00335_001-0905
HI/LO buttons
The default for these buttons is HI. The Field Engineer will not use the HI/LO buttons
and they should always remain the default.
BONE/AIR/TISSUE buttons
The default for these buttons is BONE. The Field Engineer will not use the BONE/AIR/
TISSUE buttons and they should always remain the default.
9-4
Explorer Technical Manual
0/1/2/3 buttons
The default for these buttons is 0. The Field Engineer will not use the 0/1/2/3 buttons and
they should always remain the default.
0/1 buttons
The default for these buttons is 0. The Field Engineer will not use the 0/1 buttons and they
should always remain the default.
Label Information
This area also provides information on the Hi Voltage Setting (either ON or OFF),
Frequency, Filter Errors and HiLo Errors.
Controls starting with Test Signals
AW-00335_001-0906
9-5
Explorer Technical Manual
Hi Gain
This is a spinner control with a range of 0 to 11. Changing Hi Gain changes the displayed
detector values for the Hi portions of the current segment. The associated actual gain
values are shown in the following table.
Gain Code A/D Gain MUX Gain Total Gain
0 1 1 1
1 1 2 2
2 1 4 4
3 1 8 8
4 2 1 2
5 2 2 4
6 2 4 8
7 2 8 16
8 4 1 4
9 4 2 8
10 4 4 16
11 4 8 32
Lo Gain
This is a spinner control with a range of 0 to 11. Changing Lo Gain changes the displayed
detector values for the Lo portions of the current segment. The associated actual gain
values are shown in the above table.
Xray Mode
This is a spinner control with a range of 0 to 8. The default mode is 3. This control
determines the voltage, current and duty cycle for the x-ray generator. Most scans use X-
ray Mode 3, Whole Body uses 3 and IVA (single energy) uses 6. The following table
shows the full set of defined X-ray modes:
X-Ray Mode Pulse Cycle Peak Ma Average Ma High KV Low KV
0 0 -- -- -- --
1 50% 3 0.75 140 100
2 50% 10 2.5 -- 100
3 50% 10 240 100
4 50% 3 0,75 -- 80
5 50% 3 0.75 140 80
6 100% 10 5.0 140 --
7 50% 10 2.5 140 --
8 100% 10 5.0 -- 100
9-6
Explorer Technical Manual
Aperture
This is a spinner control with a range of 0 to 13. This controls the slit through which the
X-rays pass to limit the dimensions of the X-ray beam. This control allows the selection of
a specific aperture in the system and can be used to test aperture movement. This is not
used on the Explorer.
Average Voltage
This displays the average voltage across all detectors for the Hi Air segment
AW-00335_001-0907
Display
This has two buttons: GRAPH and TABLE. This controls what is seen on the Main
Display. In Graph the Main Display shows as bar graph (see Figure 9-3) with scale
markings on the left side according to how Scale is set (see below). It shows the lo (brown
tinted) and hi (green tinted) voltages. In TABLE the Main Display shows the data from all
the detectors as numeric data.
Scale
This has three buttons: RAW, LOG and VOLTS. The Default is VOLTS. The Field
Engineer will not use RAW or LOG and the button should always remain the default
(VOLTS). This control sets up the scale markings on the Main Display when the Display
is set to GRAPH and changes the units for the numeric value wjen the TABLE is
displayed.
Segment
This has three buttons: BONE, TISSUE and AIR. The default is AIR. This controls the
data segment that are displayed in the GRASPH.. The Field Engineer will not use either
BONE or TISSUE and the button should always remain the default (AIR).
9-7
Explorer Technical Manual
AW-00335_001-0908
Status Bits
Detector values come back from the scanner as sixteen bit numbers, traveling through
some parts of the hardware on sixteen separate wires. The Status Bits field displays the
most recent value for each bit, and color-codes whether that bit has ever changed. Red bits
have not been changed, green bits have. The Reset button sets all the bits to red.
Reset
The Reset button re-starts the process of computing the means and standard deviations for
each filter-segment. It resets the number of samples to zero. It also resets the bits in the
Status Bits display.
Select Detector
This is an edit box that displays specific data about the selected detector. Select Detector
has a default of 108 (the middle of the bar graph as shown in Figure 9-3). This displays
both the currently selected detector (number) and allows a new detector to be selected by
insert a number in the edit box. A new detector can also be selected by dragging the
vertical blue line on the Main Display bar graph (see Figure 9-9). The number displayed in
the edit box will be automatically updated to reflect the line position.
9-8
Explorer Technical Manual
AW-00335_001-0909
Show Details
Show Details defaults to off. When clicked, a small window pops up containing detailed
information about one detector, in tabular form (see Figure 9-10). This information
includes the current (most recent) value, mean value, standard deviation and number of
samples (values used to calculate the mean and standard deviation) for each of the six
filter-segments. The pop-up window can be dragged wherever the user wants and remains
until Show Details is actively released.
SQDRIVER
The SQDRIVER program is accessed from the Explorer Main Menu by selecting
Utilities|Service Utilities|SQDRIVER. This program is used to calibrate the motors
and can be used as an effect tool for troubleshooting motor movement problems.
9-9
Explorer Technical Manual
SQVERIFY
The SQVERIFY program is accessed from the Explorer Main Menu by selecting
Utilities|Service Utilities|SQVERIFY. This program is a diagnostic program, which
can be used to perform a series of low-level tests of system components. Refer to the
SQCHECK User Manual (080-0707) for detailed explanations of its use.
SQKEYPAD
The SQKEYPAD program is accessed from the Explorer Main Menu by selecting
Utilities|Emergency Motion. This program provides the Field Service Engineer with a
method for moving all motors even in conditions where a detected software error
condition would normally prevent any movement of the motors. This program can be used
as a quick means of checking most motor control functions.
9-10
Appendix A
Computer and PCI Controller
Card
QDR for Windows XP® QDR software V12.1 uses a PCI controller card (Com/Con) which
requires a dedicated interrupt. At this time, the Intel D865PERLX motherboard (the X indicates it
has on-board 10/100 networking, documentation will refer to D865PERL) has been validated
with QDR for Windows XP with the Com/Con card in Slot 2, with a modem card in Slot 4, and a
SCSI card in any slot. Please be aware that there are certain issues that need to be addressed up
front if another computer is to be validated.
Computer Specifications
The minimum requirements for V12.2 or higher with a PCI Com/Con are as follows:
• 512 MB memory
• 20 GB hard drive
• Pentium 4 or Celeron CPU without hyperthreading1.
• HAL- Advanced Configuration and Power Interface (ACPI) PC2
• Designed for Windows XP Logo for motherboard and all components other than Hologic
supplied components.
• All Drivers (other than Hologic supplied) must be Windows XP Certified digitally signed.
• CD-RW Teac CD-W552E with firmware 1.05 or later.
• Floppy Drive or SuperDisk3
• InCD 4.0.7.2
• AGP Video card eVGA 64MB, 064-A8-NV74-T2 or other Hologic Approved video
card.4
• Power Supply 300 Watt minimum, with UL/CSA/CE Approval.
• Windows XP Professional with Service Pack 1
• Microsoft patches installed from QDR for Windows XP software disk.
1. Hyperthreading CPU’s have not been tested, they may or may not work. Some Pentium 4’s below 3.0 GHz do
not currently (8/13/03) have hyperthreading. The ones with hyperthreading below 3.0 GHz are designated with
a C, as Pentium 4 2.80C. We will be testing CPU’s with hyperthreading in a future release.
2. HAL is hardware abstraction layer. Most boards will allow you to switch from ACPI (Uniprocessor) to
“Advanced Configuration and Power Interface (ACPI) PC”. You MUST use this HAL. To see if the correct
HAL is intalled, go to Control Panel/System/Hardware/Device Manager and click on the + sign next to the
computer. The HAL is shown there.
3. Superdisks are not supported for writing under QDR for Windows XP, but they can be used for reading old
archives. It is allowed for the SuperDisk and its drivers not to be Windows XP certified.
4. Most newer video cards do not support VGA, which is used for a limited number of functions in QDR for
Windows XP (scan file plot, scan file look, emergency motion, and some service utilities).
A-1
Explorer Technical Manual
To determine which PCI slot the Com/Con board should be located in, you must have the
technical manual for the motherboard. For Intel motherboards, these are supplied on their
website. The Com/Con board uses INT A. Please read and understand the following,
which comes from the D865PERL Technical manual, but is a general: explanation of
interrupt routing.
A-2
Explorer Technical Manual
A-3
Explorer Technical Manual
You see that the COM/CON board (RTX PnP) is the only thing using INT 10.
In the Motherboard BIOS, you should reserve INT 10 for the Slot used. Note that this will
not guarantee that the COM/CON card will be assigned that INT under Windows XP, but
it should be done anyway.
Set up Windows XP with the following options:
DISPLAY: Set desktop and screen saver to None for all active accounts.
Set Windows and Buttons to Windows XP Style
Set Color scheme to Default(Blue)
Set Font size to Normal.
Set Screen Resolution to 1024 X 768.Color (32 bit).
Set Color quality to Highest( 32 Bit)
Set Hardware Accelerator to Full
Set Screen refresh rate to 75 or higher.
A-4
Explorer Technical Manual
A-5
Explorer Technical Manual
A-6
Appendix B
Windows XP ® Setup
Instructions
1. WELCOME SCREEN
1.1. Log onto the computer using the Administrator account, the password is password
(all small letters). Click the Arrow icon.
1.2. Select Start Control Panel.
1.3. Click Switch to Classic View if not selected in the left pane of the window.
1.4. Click User Accounts.
1.5. Click Change the way users log on or off.
1.6. Ensure “Use the welcome screen” check box is checked.
1.7. Ensure “Use fast User Switching” is Unchecked.
1.8. Check Apply Options.
1.9. Click on the X in the upper right corner to close all open windows and return to the
main window.
2. EQUIPMENT CHECK
2.1. Log onto the computer using an Administrator account. The password is password
(all small letters).
2.2. Select Start My Computer and perform the following checks.
• A: Drive is 3.5
• D: Drive is CD-ROM
If the CD-ROM drive is incorrect proceed to the next step, otherwise close all open windows back
to the Windows desktop then go to the next Section.
CD ROM Drive Correction
• Click on the X in the upper right corner to close all open windows and return to the desk-
top.
• Select Start Control Panel Administrative tools Computer Management Icon
Storage Disk Management.
• Right click the CD-ROM and select Change Drive Letter and Paths.
B-1
Explorer Technical Manual
3. Task Scheduler
3.1. Select START All Programs Accessories System Tools
Scheduled Tasks.
3.2. Select the Advanced pull down menu on the tool bar.
• If “Stop Using Task Scheduler” is shown click Stop Using Task Sched-
uler. The option should read “Start Using Task Scheduler.”
• Click on the X in the upper right corner to close Scheduler Tasks.
B-2
Explorer Technical Manual
B-3
Explorer Technical Manual
8. Date Format
8.1. Select Start Control Panel Regional and Language Icon
Customize Date Tab.
8.2. Set Short date format to MM/dd/yyyy and the Date Separator to “/”.
Click Apply OK
8.3. From the “Regional and Language Options” screen click Apply OK
8.4. Click The X in the Control Panel Window to close the Control Panel and return
to the windows desktop.
9. AUTOPLAY
9.1. From the window desktop select Start Run.
9.2. Enter gpedit.msc then click OK.
9.3. Select Computer Configuration Administrative Template System.
9.4. Double click Turn off Autoplay.
9.5. Select the Enabled radio button
9.6. Set “Turn off Autoplay on:” to All Drives.
9.7. Click Apply OK.
9.8. Select X in upper right corner to close screen.
B-4
Explorer Technical Manual
• Set “Turn off monitor” to Never (This cannot be set whenever logged on as
QDR).
10.3. Select Screen Saver Tab.
• Set “System standby” to Never.
• Set “System hibernates” to Never.
• If changes were made click Apply OK otherwise click Cancel.
10.4. Select Desktop Tab
• Set Desktop to None if not set.
Click Apply if a change was required.
10.5. Select Settings Tab.
• Set Screen Resolution to 1024 X 768 if not set.
• Set Color Quality to Highest (32Bit) if not set.
Click Apply if a change was required.
10.6. (This step can only be performed when Logged On as an Administrator) Select
Monitor Tab.
• Set the Screen refresh rate to 75.
Click Apply, if a change was required.
• Click OK.
10.7. Select Troubleshoot Tab.
(This step can only be performed when Logged On as an Administrator)
• Set Hardware Accelerator to Full.
Click Apply, if a change was required.
• Click OK.
10.8. From the Display Properties screen select Apply OK.
10.9. Click X in the upper right corner of all screens to return to the Windows
desktop..
B-5
Explorer Technical Manual
B-6
Explorer Technical Manual
14.2. From the Control panel screen open the Internet Options ICON
14.3. Set the Home Page to http://www.hologic.com if it is not presently set, then
select Apply, then select OK.
B-7
Explorer Technical Manual
16.8. When the message “You successfully activated your copy of Windows”
appears select OK.
B-8
Explorer Technical Manual
18.10. From the “Select Program Folder” window insure the program folder is
“VentureCom RTX” then select Next.
18.11. From the “Start Copying Files” window insure that the following will be
copied as shown below, then click Next.
18.12. RTX Runtime Environment
18.13. RTX Runtime installation directory
C:\Program Files\ VentureCom\RTX
18.14. From the “Setup Complete Window” select the “Yes, I want to restart my
computer now” radio button.
18.15. Remove the CD-ROM from the drive then click Finish.
18.16. After the compute reboots select Start All Programs VenturCom RTX
RTX Properties.
18.17. Select Control Tab then Start RTX. Click Apply, if availiable.
18.18. Select the SettingsTab.
18.19. Select the Boot radio button then Apply OK.
18.20. Select Start Turn Off Computer Restart.
18.21. After the computer reboots select Start All Programs VenturCom RTX
Properties.
18.22. Select Control Tab and insure all four components are running then click
Cancel.
If they are not running repeat all steps from 18.15 through this step..
18.23. Click Start Turn Off Computer Turn Off.
18.24. After the computer shuts OFF. Install the PCI Communication Card 140-0124
into the appropriate PCI Slot of the computer (see Appendix A).
18.25. Turn the computer ON.
B-9
Explorer Technical Manual
B-10
Explorer Technical Manual
12. Select Start|Run… and type in D:\setup.exe (the CDROM drive may be drive letter
E: on some systems) and click the OK button.
13. Select Admin from the dropdown list, using password as the password and click the
OK button.
14. Click the Next button at the Welcome window.
15. Click the OK button.
16. Click the Yes button at the Software License Agreement window or click OK
button.
17. Select Explorer at the Select QDR Machine Type window and then click the OK
button.
18. At the Serial Number window, enter the 5-digit instrument serial number and click
the Next button.
19. At the Choose Destination Location window, click the Next button.
20. The QDR for Windows software will now be loaded.
21. Once software loading is complete, the Setup Complete window will appear asking if
you wish to restart your PC now or later. Click the Finish button.
22. The PC will now reboot.
B-11
Explorer Technical Manual
B-12