Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Manual Explorer1 PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 170

EXPLORER™

FAN BEAM X-RAY BONE DENSITOMETER

TECHNICAL MANUAL

AW-00335_001-0101

Hologic, Inc. Phone in US. 800-321-4659 Document MAN-00131


35 Crosby Drive Fax (Domestic): 781-280-0670 Revision 003
Bedford, MA 01730 Fax (International): 781-280-0671
USA
Explorer Technical Manual

Notice

The information contained in this manual is confidential and proprietary to Hologic, Inc.
This information is provided only to authorized representatives of Hologic's customers
solely for the purpose of facilitating the use of Hologic's products. No information
contained herein may be disclosed to any unauthorized person for any purpose whatsoever
without the prior written consent of Hologic, Inc.
The procedures described in this document are intended for use by direct employees of
Hologic, Inc., or authorized Hologic Equipment Resellers and their trained field
engineers. Any unauthorized or untrained persons performing these procedures may affect
the warranty of the Explorer.
Exercise proper caution when servicing the system. There are dangerous and potentially
lethal voltages accessible within the Explorer system. To avoid exposure to shock hazards,
the Main Circuit Breaker should be switched off, and the power cord removed, before
working inside any part of the system.
The Explorer Fan Beam X-ray Bone Densitometer produces ionizing radiation in the form
of X-rays. It may be dangerous to the patient, operator or field engineer unless safe
exposure factors and operating instructions are observed. To avoid unsafe exposure, do not
attempt to service the equipment unless you are a Hologic, Inc., certified field engineers.
Exercise proper caution when servicing the system. A dosimeter (film badge) should
always be worn while on site. Dose and scatter measurements must be taken after each
service call to ensure that the parameters are still within specifications.
Hologic, Inc., has made all effort to ensure that the information in this manual is accurate
and complete. Hologic, Inc., shall not, however, be liable for any technical or editorial
errors or omissions contained herein, or for incidental, special or consequential dangers in
connection with the furnishing or use of this manual. The information in this manual is
subject to change without notice.
Explorer and the Hologic logo are registered trademarks of Hologic, Inc. All other
products and company names used in this manual are trademarks and registered
trademarks of other manufactures.
Printed in U.S.A.
Copyright 2004 by Hologic, Inc., All rights reserved

ii
Explorer Technical Manual

Table of Contents
Chapter 1 - INTRODUCTION ................................................................................... 1-1
System Overview ..................................................................................................... 1-1
X-Ray Scanning Principles ...................................................................................... 1-2
Functional Overview ................................................................................................ 1-3
Product Specifications ............................................................................................. 1-9
Product Specifications ....................................................................................... 1-9
Line Voltage and Maximum Line Current ...................................................... 1-10
Line Inductance .......................................................................................... 1-10
Duty Cycle ....................................................................................................... 1-10
Exam Mode Performance ................................................................................ 1-10
AC Power Isolation Module ............................................................................ 1-11

Chapter 2 - FUNCTIONAL DESCRIPTION ........................................................... 2-1


Computer ................................................................................................................. 2-1
PCI Communications Controller Board ................................................................... 2-1
Interface Connections ........................................................................................ 2-1
Distribution Board ................................................................................................... 2-3
Power ................................................................................................................. 2-3
Interface Connections ........................................................................................ 2-3
Motor Controller Board ........................................................................................... 2-4
Power ................................................................................................................. 2-5
Interface Connections ........................................................................................ 2-5
Control Panel Controller Board ............................................................................... 2-6
Power ................................................................................................................. 2-6
Interface Connections ........................................................................................ 2-6
C-Arm Interface Board ............................................................................................ 2-8
Continuity Daisy Chain ..................................................................................... 2-9
Power ................................................................................................................. 2-9
Interface Connections ........................................................................................ 2-9
X-Ray Controller Assembly (P/N ASY-00409) .................................................... 2-12
Interface ........................................................................................................... 2-12
X-Ray Controller Board ................................................................................... 2-14
X-Ray Source Unit ................................................................................................. 2-15
Data Acquisition System ....................................................................................... 2-15
Solid State Detector ........................................................................................ 2-16
Integrator/Multiplexor Subsection ................................................................... 2-16
Analog To Digital Board ................................................................................. 2-16

Table of Contents 1
Explorer Technical Manual

Power ......................................................................................................... 2-17


Interface Connections ................................................................................ 2-17
Power Module ........................................................................................................ 2-19

Chapter 3 - INSTALLATION .................................................................................... 3-1


Required Tools ......................................................................................................... 3-1
Required Documentation ......................................................................................... 3-1
Room and Doorway Size ......................................................................................... 3-2
Arrange for Help ...................................................................................................... 3-2
Inspect For Shipping Damage .................................................................................. 3-3
Uncrate Unit ............................................................................................................. 3-3
Inspect For Hidden Shipping Damage ..................................................................... 3-4
Take Inventory ......................................................................................................... 3-4
Measure Path To Final Destination ......................................................................... 3-5
Short Doorway ................................................................................................... 3-5
Narrow Hallway ................................................................................................. 3-5
Tabletop Removal (If Necessary) ............................................................................ 3-5
Preparing the Tabletop for Moving ......................................................................... 3-7
Move Unit To Destination (Vertical Position) ........................................................ 3-7
Move Unit to Destination (Horizontal Position) ...................................................... 3-8
Setting Up Unit ........................................................................................................ 3-9
Install C-Arm ......................................................................................................... 3-11
Install Cables .......................................................................................................... 3-11
Check Power Line Voltage .................................................................................... 3-14
Measure Line Voltage ...................................................................................... 3-14
Measure Isolation Transformer Secondary Voltage ........................................ 3-15
Install Computer .................................................................................................... 3-15
Starting QDR Software in Service Mode .............................................................. 3-15
Check Tube KV Peak Potential ............................................................................. 3-16
Check Tube Current ............................................................................................... 3-18
Check Belt Tension ................................................................................................ 3-20
Adjust C-Arm Y Belt ............................................................................................. 3-20
Calibrate Motors .................................................................................................... 3-20
Check X-Ray Beam Alignment ............................................................................. 3-20
Check Laser Positioning Offset ............................................................................. 3-21
Adjust A/D Gain Control ....................................................................................... 3-21
Perform Detector Flattening .................................................................................. 3-21

2 Table of Contents
Explorer Technical Manual

Measure X-Ray Dose to Patient ............................................................................. 3-21


Check HVPS/S (Tank) For Radiation Leakage ..................................................... 3-22
Calibrate For Area, BMD And BMC ..................................................................... 3-23
Test Scan Modes .................................................................................................... 3-24
Finish Assembling Unit ......................................................................................... 3-24
Measure X-Ray Scatter From Phantom ................................................................. 3-24
Perform QC ............................................................................................................ 3-25
Run Reproducibility Test ....................................................................................... 3-25
Table Top Radiographic Uniformity ..................................................................... 3-25
The Radiation Measurement Report ...................................................................... 3-26

Chapter 4 - ALIGNMENT AND CALIBRATION .................................................. 4-1


Table Alignment ...................................................................................................... 4-1
Checking Table Alignment ................................................................................ 4-1
Aligning Table ................................................................................................... 4-1
Table Edge to T-Rail (“A” Dimension) Adjustment ......................................... 4-1
Front to Back T-Rail and Table Edge/Rail Gap Adjustment ....................... 4-3
X-Ray Beam Alignment .................................................................................... 4-3
Motor Calibration .................................................................................................... 4-8
MOTOR$AY (Final Value = 3750) ................................................................... 4-8
MOTOR$TY (Final Value = 3750) ................................................................. 4-11
MOTOR$TX (Final Value = 1000) ................................................................. 4-14
Laser Positioning Offset Adjustment ..................................................................... 4-17
A/D Gain Control Adjustment ............................................................................... 4-17
Detector Flattening ............................................................................................... 4-18
Table Top Radiographic Uniformity ..................................................................... 4-19
Instruments using Body Composition Analysis (BCA) ................................... 4-20
Instruments using BMD Whole Body Analysis .............................................. 4-20
Check Phantom values ........................................................................................... 4-20
Area, BMD, And BMC Calibration ....................................................................... 4-21
Scan Thickness Measurement & Calibration ................................................... 4-21
Calibration of Area and BMC, for Array Scan Modes .................................... 4-22
Recalyze and Add Array AP Scans to the QC Database ................................. 4-23

Chapter 5 - REMOVE AND REPLACE PROCEDURES ....................................... 5-1


Recommended Tools ............................................................................................... 5-1
Electronics Tray FRUs ............................................................................................. 5-1
Electronics Tray Printed Circuit Boards ............................................................ 5-1
C-Arm Y Belt ..................................................................................................... 5-2

Table of Contents 3
Explorer Technical Manual

C-Arm Y Motor or Gearcase ............................................................................. 5-3


C-Arm Y Encoder .............................................................................................. 5-4
Table Y FRUs .......................................................................................................... 5-4
Table Y Belt ....................................................................................................... 5-5
Table Y Motor or Gearcase ............................................................................... 5-6
Table Y Motor Controller PCB ......................................................................... 5-6
Table Y Encoder ................................................................................................ 5-7
Table X FRUs .......................................................................................................... 5-8
Table X Motor Controller PCB ......................................................................... 5-8
Table X Belt ....................................................................................................... 5-8
Table X Motor or Gearcase ............................................................................... 5-9
Table X Encoder .............................................................................................. 5-10
Lower C-Arm FRUs .............................................................................................. 5-11
C-Arm Interface Board .................................................................................... 5-11
X-Ray Controller Assembly ............................................................................ 5-11
Filter Drum Assembly ..................................................................................... 5-12
Tank Assembly ................................................................................................ 5-13
Upper C-Arm FRUS .............................................................................................. 5-15
Detector Assembly ........................................................................................... 5-15
Laser Assembly ................................................................................................ 5-15
Control Panel ................................................................................................... 5-16
Filter Drum Assembly FRUs ................................................................................. 5-16
Drum Encoder PCB ......................................................................................... 5-17
Drum Belts ....................................................................................................... 5-17
Stepper Motor Assembly ................................................................................. 5-18
Drum Bearings ................................................................................................. 5-19
Replacing EMI Cables ........................................................................................... 5-21
FRU Lists ............................................................................................................... 5-22

Chapter 6 - FAULT ISOLATION .............................................................................. 6-1


Before Starting ......................................................................................................... 6-1
Software Configuration ............................................................................................ 6-1
Hardware Configuration .................................................................................... 6-1
Power Problems ....................................................................................................... 6-1
Motion Problems ...................................................................................................... 6-2
Control Panel Problems ........................................................................................... 6-4
Display Problems ..................................................................................................... 6-4
Vertical Stripe .................................................................................................... 6-5
Horizontal Stripe ................................................................................................ 6-5
Noise .................................................................................................................. 6-6

4 Table of Contents
Explorer Technical Manual

No Display ......................................................................................................... 6-6


Targeting/Laser Problems ........................................................................................ 6-7
Data Communications Problems ............................................................................. 6-8
Area/BMD/BMC/CV Specification Problems ......................................................... 6-8
X-Ray Problems ....................................................................................................... 6-8
No X-Rays ......................................................................................................... 6-9
X-Ray Alignment Problems ..................................................................................... 6-9
Detector Flattening Problems .................................................................................. 6-9
Laser Problems ...................................................................................................... 6-10
Oil leakage ............................................................................................................. 6-11
The Torque Specifications ............................................................................... 6-11
Tank Top Cover Components and Screw Location ......................................... 6-12
Tightening the Lexan Cup Screws ................................................................... 6-12
Tightening the Bladder Gasket Screws ............................................................ 6-12
Tightening the Transformer Seal Screws ......................................................... 6-13
Tightening the Tank Cover Gasket Screws ..................................................... 6-13
Miscellaneous Problems ........................................................................................ 6-14

Chapter 7 - PREVENTIVE MAINTENANCE ......................................................... 7-1


Customer Preventive Maintenance .......................................................................... 7-1
Field Service Preventive Maintenance .................................................................... 7-1
Guide Rail and Bearing Maintenance ................................................................ 7-2

Chapter 8 - PCB SUMMARY INFORMATION ...................................................... 8-1

Chapter 9 - SOFTWARE TOOLS ............................................................................. 9-1


X-Ray Survey .......................................................................................................... 9-1
Invoking X-Ray Survey ..................................................................................... 9-1
The Main Display .............................................................................................. 9-2
The Display Controls ......................................................................................... 9-3
Controls starting with Xrays ........................................................................ 9-4
Xrays (F2) button OFF/ON ..................................................................... 9-4
Filter Wheel button OFF/ON .................................................................. 9-4
HI/LO buttons ......................................................................................... 9-4
BONE/AIR/TISSUE buttons .................................................................. 9-4
0/1/2/3 buttons ........................................................................................ 9-5
0/1 buttons ............................................................................................... 9-5
Label Information ................................................................................... 9-5
Controls starting with Test Signals .............................................................. 9-5
Test Signal button OFF/ON .................................................................... 9-5
Drum Half 0/1 ......................................................................................... 9-5

Table of Contents 5
Explorer Technical Manual

Hi Gain .................................................................................................... 9-6


Lo Gain ................................................................................................... 9-6
Xray Mode .............................................................................................. 9-6
Aperture .................................................................................................. 9-7
Average Voltage ..................................................................................... 9-7
Controls starting with Display ..................................................................... 9-7
Display .................................................................................................... 9-7
Scale ........................................................................................................ 9-7
Segment .................................................................................................. 9-7
Zoom In and Zoom Out .......................................................................... 9-8
Data Current Settings .............................................................................. 9-8
Subtract Dark Current OFF/ON .............................................................. 9-8
Controls starting with Status Bits ................................................................ 9-8
Status Bits ............................................................................................... 9-8
Reset ........................................................................................................ 9-8
Select Detector ........................................................................................ 9-8
Show Details ........................................................................................... 9-9
SQDRIVER ....................................................................................................... 9-9
SQVERIFY ...................................................................................................... 9-10
SQKEYPAD .................................................................................................... 9-10

Appendix A - Computer and PCI Controller Card ................................................. A-1

Appendix B - Windows XP® Setup Instructions ...................................................... B-1

6 Table of Contents
Section 1
INTRODUCTION
System Overview
The Explorer™ X-ray Bone Densitometer (Figure 1-1) performs bone mineral content (BMC)
and bone mineral density (BMD) examinations on the spine and hip. Optional software will
provide examinations on the forearm, whole body and decubitus lateral. It does so using X-rays
of two different energy levels. This dual-energy scheme allows soft tissue within the selected area
to be subtracted out leaving only bone to be scanned and estimated.

AW-00335_001-0101

Figure 1-1 Explorer System


The patient lies face up on the table and, with the aid of a cross-hair laser, the operator positions
the scanning arm over the region of interest. After entering patient data and selecting the scan type
and size, the operator initiates the scan with a single keystroke.
The operator is not required to select technique factors as tube current and voltage are pre-
selected and fixed. Since testing is performed by fan beam method, rather than by flooding the
area as in conventional radiography, the scanning time is a function of the dimensions of the area
to be measured, the desired resolution, and the desired precision.
BMC results are expressed in grams of calcium hydroxyapatite and BMD is reported in grams/
cm2 of the same compound.
Note: In most cases, no additional shielding is necessary for patient, operator, or room. The
Explorer system can be placed in any convenient non-shielded examination room. Contact
your state regulatory agency for details about additional shielding requirements, if any.
The system employs a patented Automatic Internal Reference System, which continuously
calibrates the machine to eliminate the effects of variations in temperature, tube flux, etc. No daily
calibration is required. The daily scanning of a quality control phantom is required to provide
assurance that the system is functioning correctly and to aid in the detection of any long-term
drift.

1-1
Explorer Technical Manual

The X-ray scans produced by the system, and displayed on the monitor, are intended only
to locate anatomical sites for measurement and to assure the operator that the machine is
operating properly. They are not intended as a substitute for conventional film-based
diagnostic scans.

X-Ray Scanning Principles


An X-ray source, consisting of a high voltage generator and X-ray tube in a common,
shielded enclosure, is mounted beneath the patient on the C-Arm. It generates a narrow,
tightly collimated, fan-shaped beam of X-rays which alternate, at power line frequency,
between 100kVp and 140kVp. At the other end of the C-Arm, above the patient, is a
crystal/solid state detector array. During a scan, the C-arm and table move, under
computer control, to guide the beam over the desired scan area.
Before passing through the patient, the beam is filtered through a rotating drum in which
alternating segments having radio-opacities equivalent to tissue, bone and air are located.
When finally intercepted by the detector, the beam contains information about the X-ray
absorbing characteristics of both the patient and the calibration materials in the filter
drum. An A/D converter, fed by the detectors, supplies a complex digital signal to the
computer, which uses that signal both to construct the screen display, and as the basis for
its computations of BMC and BMD.
The computer algorithm is based on the principle that bone attenuates the X-ray beam
differently at high and low energies. The bone mineral content of any sample point can be
computed from:
Q = L - kH
where L and H are the logarithms of the sample attenuation at high (140kVp) and low
(100kVp) energies, respectively. The constant k depends on the tissue attenuation
characteristics of the beam. In the Explorer, k is continuously measured using the “tissue”
segment in the filter wheel.
The program works in the following manner:
1. Load preliminary scan and obtain regions of interest from operator.
2. Estimate k as an average value of:
k = [Ltissue - Lair] / [Htissue - Hair]
where Ltissue indicates a low-energy measurement with tissue-equivalent material
interposed by the filter drum, and Lair, Htissue and Hair are similarly defined.
Note: The subscript “air” designates the filter drum segment that is empty (i.e., contains
neither bone- nor tissue-equivalent material).
3. Using this value of k, calculate Q for each point scanned using the formula given
above (Q = L - kH). This array of Q values constitutes a “Q scan”. Displays the Q
scan.
4. Compile a histogram of the Q values. Because a large portion of the scan con-
tains soft tissue only, this histogram will have a large peak. Choose a threshold

1-2
Explorer Technical Manual

value just above this peak, and apply that value to discriminate, point by point in
the Q scan, between “bone” points (whose Q is above the threshold) and “non-
bone” points (whose Q is below the threshold).

Tissue Bone
NUMBER
OF
POINTS
WITH A
SPECIFIC
VALUE OF
Q.

Q KP1085_004-0103

Figure 1-2 Q Scan Plot


5. Use the “non-bone” points to calculate a baseline value for each scan line. Using
these points, form a new histogram and repeat steps 4 and 5 until the results con-
verge.
6. Smooth the segment boundaries to eliminate isolated noise-generated “bone”
points.
7. Display the “bone” and “non-bone” points for operator approval.
8. Determine the constant of proportionality (d0) that relates the Q values to actual
BMC (grams). This constant is determined by measuring how much Q shifts
when bone-equivalent material is interposed by the filter drum.
9. Calculate the total bone mineral values by adding up the Q values for all “bone”
points in each region of interest (e.g., each vertebra), and multiplying by d0.
10. Determine the bone areas by counting the number of “bone” points in each
region of interest.
11. Calculate bone mineral density as:
BMD = BMC / area
12. Display the calculated results and print the report.

Functional Overview
This section provides block diagrams of the Explorer system along with a brief functional
overview of each diagram and block. A detailed functional description along with

1-3
Explorer Technical Manual

interconnection diagrams and interconnection descriptions is provided in Section 2 of this


manual.
Control
Panel Detector
Assembly

C-Arm
Distribution
Interface
Board
Board

Operator’s Console
X-Ray
TX TY AY
Controller
Scanner and Tank

AW-00335_001-0104

Figure 1-3 Explorer System Block Diagram


The Hologic Explorer consists of essentially three conceptual subsystems. These
subsystems are the Operator’s Console, Motor Control Subsystem, and C-Arm
Subsystem.
The Operator’s Console is the input/output subsystem of the instrument and consists of a
PC running the Explorer system software under Windows XP® and the Communications
Controller Board manufactured by Hologic. The software communicates with the scanner
sending out arm and table motion commands, X-Ray commands, and X-ray detection
commands. The software also checks to see that commands have been completed and
issues an appropriate error message when a command fails.
The Motor Control subsystem of the scanner is controlled by the Distribution Board. All
motor movement commands are routed through the Distribution Board to the individual
Motor Drivers. There is one Motor Driver for each motor: Arm Y direction (AY), Table X
direction (TX) and Table Y direction (TY) (Whole Body, Decubitus Lateral or Forearm
options only). The Distribution Board also distributes the DC power throughout the
scanner. Circuit breakers for each Motor Driver can be found on the Distribution Board.
The C-Arm subsystem controls the generation and detection of X-rays. It processes the
commands received from the Distribution Board and passes them to the X-Ray Controller
or the Detector Assembly. The X-Ray Controller, as its name implies, controls the

1-4
Explorer Technical Manual

generation of the X-Rays by the X-Ray Source or “Tank”. The X-Rays pass through the
patient and are sensed by the Detector Assembly.
AC

AC from
AC Power
DIN Rail
Strip
(Scanner)
AC
Printer Monitor

Computer
AC
Video

Modem Phone
Keyboard (Optional) Jack
Network
Network I/F Card
Jack
Mouse
Communications
Scanner
Controller
KP1085_004-05

Figure 1-4 Explorer Operator’s Console Block Diagram


The Explorer Operator’s Console consists of a Pentium PC on a convenient, roll-around
computer cart designed specifically for the Explorer PC. AC power from the scanner is fed
to the Operator’s Console via a power strip attached to the PC cart.
The Pentium PC contains the video controller board to drive the monitor and the
Communications Controller to interface the PC to the Scanner. A Network Interface Card
(NIC) is installed to control communications with the Hologic QDRNet® or a DICOM®
network. The mouse gives the operator easy control over the Windows XP-based Explorer
software and the keyboard is used for data input to the Explorer software. A color inkjet
printer is provided with the PC. An optional laser printer can be substituted for the inkjet
printer.

1-5
Explorer Technical Manual

JP1 To/From C-Arm Interface

To/From Detector Assy (Control Panel I/F)


JP4

To/From TY
Motor
Communications JP10 TY Stepper
JP5
Controller Motor Driver TY
Signal Encdr
Distribution
Board TX
TX Stepper Motor
JP7
Motor Driver TX
Encdr
+/-15V JP6
From DIN
Rail AY
and LVPS +24V AY Stepper Motor
JP8 JP11
Motor Driver
AY
Encdr
AW-00335_001-0106

Figure 1-5 Explorer Distribution and Motor Control Block Diagram


The heart of the Explorer Scanner is the Distribution Board. This board interfaces the PC
to the different subsections of the Scanner, distributes control signals to the table and C-
Arm, and distributes DC power throughout the Scanner. The Distribution Board receives
command inputs from the Communications Controller in the PC and voltages from the
DIN rail and Low Voltage Power Supply. Commands and voltages dealing with the
generation of X-Rays or the acquisition of scan data are transferred to the C-Arm Interface
Board, which controls both functions. The Distribution Board also receives Table and C-
Arm motion commands from the Communications Controller in the PC.
When a motion command is received, the Distribution Board routes the command along
with a board address to the Motor Driver Boards. The addressed Motor Driver Board
converts these digital commands into analog signals that are strong enough to drive the
motors. A Position Encoder attached to the idler pulley moved by the motor, senses
changes in the position of the arm or table. This device provides a voltage feedback to the
Motor Driver Board that is translated into a motor position, which in turn, is fed back to
software in the Operator’s Console PC.
The Explorer Control Panel communicates with the Distribution Board via the Control
Panel Interface Board. The Control Panel provides the operator with a means to issue
manual commands to move table and arm motors and an indicator showing when X-rays
are being generated.
The C-Arm Interface Board controls the C-Arm Subsystem. Commands to the subsystem
arrive from the Distribution Board. The commands processed by the C-Arm Interface
Board are sent to the Positioning Laser, the X-Ray Controller (XRC), the Detector
assembly, or the Filter Drum Assembly. The commands sent to the Positioning Laser are

1-6
Explorer Technical Manual

simple on/off commands. Commands sent to the XRC tell it when to produce X-rays,
which power level of pulse to produce, and the pulse mode to be used. These commands
will subsequently control the way the X-Ray Source (Tank) is driven by the XRC. The
commands to the Filter Drum Assembly turn the Filter Drum on or off and synchronize
the filter drum with the AC line waveform. The last set of command signals is sent to the
Detector Assembly to synchronize the acquisition of data with the production of the X-
rays from the source
(Tank).
Detector Assembly

Laser
Assembly

Detector Board
(Detectors, Muxes,
and A/D Converter)
To/From C-Arm
Distribution
From Interface
Board Board
Filter Drum Motor
and
Position Sensor

X-Ray
X-Ray
Source
Controller
(Tank)

DIN KP1085_004-0107

Rail

Figure 1-6 Explorer C-Arm Subsystem Block Diagram


The C-Arm Interface Board collects data from most of the circuits mentioned above for
transmission back to the Distribution Board and, finally, the Explorer software. The Filter
Drum Assembly sends back filter and reference phase information. The XRC sends back
information concerning fault conditions, beam conditions, and an AC Line signal for
generating AC Line Interrupts to the software. The Detector Board sends data collected
from scans back to the software through the C-Arm Interface Board and the Distribution
Board. Mathematical algorithms applied to this data produce the scans seen by the
operator and to calculate the BMD, BMC, and area of the bone matter scanned.
Block Description
Computer Controls and commands all Explorer hardware modules.
Communications Controls the flow of commands to and from the Scanner
Controller modules via the communications bus.
Distribution Board Provides the interconnections between the Explorer
Operator's Console and the Scanner and distributes DC
voltages.

1-7
Explorer Technical Manual

Block Description
Control Panel Provides switches for manually moving the C-Arm and
Patient Table. Also provides Emergency Stop and Instrument
On/Off switches.
TX Stepper Motor Controls the motion of the Patient Table in and out motor and
Driver monitors table position information from the encoder.
TY Stepper Motor Controls the motion of the Patient Table left to right motor
Driver and monitors table position information from the encoder
AY Stepper Motor Controls the motion of the C-Arm left and right motor and
Driver monitors C-Arm position information from the encoder.
C-Arm Interface Controls the Filter Drum motor, generates timing and control
signals for the X-Ray Controller and the Data Acquisition
System, monitors the tape switches, and provides power to the
Positioning Laser.
X-Ray Controller Controls the operation of the X-ray Source.
X-Ray Source Unit Generates the X-ray beam.
Detector Board Converts the X-rays into electrical signals. Integrates the
signals from the Solid State Detectors and converts them to a
digital value in the Analog-to-Digital converter circuitry.
Control Panel Provides the operator with a means to reposition the table, C-
Arm, and to turn on the laser.
Positioning Laser Provides a laser crosshair beam to assist in positioning the
patient on the Patient Table.

1-8
Explorer Technical Manual

Product Specifications
Product Specifications
SPECIFICATION DEFINITION
Scanning Method Multi-detector array, Indexing table, and motorized C-arm
X-ray System: Tungsten target switched Pulse Dual-Energy x-ray tube, operating at 100
and 140kVp, capable of 0.75 and 2.5mA average at 25% duty cycle
Detector System: 54 multi-channel detector (4mm) consisting of CdWO4 scintillators cou-
pled to Silicon diodes
Scanning Sites: AP Lumbar Spine, Proximal Femur (Hip), Forearm, Decubitus Lateral
Spine, and Whole Body
Scan Region: 1.97m (77.5 in) X .65m (25.6 in) maximum
(at pad surface)
Scatter Dose Rate: Nominal 10mGy/hr at 1 m from the examination table
Leakage The Explorer meets the requirements of 21 CFR 1020.30(k) for leakage
Radiation: from the X-ray source
External Shielding None required unless required by local ordinance
Requirement:
Measurement Measurements of Bone Mineral Content (BMC) and projected Area are
Accuracy: calibrated to Hologic anthropomorphic spine phantom of known mineral
content to within +/-1.0%.
Calibration: Self Calibrating using HOLOGIC Automatic Internal Reference System.
Operator calibration NOT required.
Patient Manual movement (two independent axes) using push-button controls.
Positioning: Laser Indicator (Red, < 1mW) with cross hairs used for alignment.
System Weight: Console (Computer,
Installed Scanner Printer, Monitor)
327kg 720lb 34.1kg 75lb
System Weight: 568kg 1250lb
Shipping
Temperature (Ambi- Operation Storage
ent): 15o to 32oC (59o
to 90oF) to -30o 50oC
(-22o to 122oF)
Humidity (Ambient): Operation Storage
20 to 80% relative humidity, non- 10 to 90% relative humidity, non-
condensing condensing
Subject Weight: The maximum subject weight is 136kg (300 lb.)
Footprint (operating)
(+/- 25mm (1in.)): Length Width Height
m in. m in. m in.
3.02 119 1.50 59 1.42 56
Heat Load: 1000W (3400 BTU/h) (maximum)
Table Height: (+/- 71cm (28in.)
25mm (1in.))
Positioning Laser: Laser Diode (<1mW) cross hair, with emergency mechanical shutter
X-ray Collimation Aperture with 1.0 mm (0.040 in) slit
Leakage Current: Normal <75µA Single Fault <400µA

1-9
Explorer Technical Manual

Product Specifications (continued)


SPECIFICATION DEFINITION
Resolution 1mm x 2mm 610% pixel size
(approximate):
X-ray System mm in.
Geometry: Source to Detector 883.4 (+/- 8mm) 34.8 (+/- 0.315in.)
Distance
Source to Patient 424 (+/- 6mm) 16.7 (+/- 0.236in.)
Distance
Beam size at detector 226 mm (8.9") x 4.4 mm (0.17") (1 mm
aperture width)

Line Voltage and Maximum Line Current


VAC (+/- 10%) A (max) Hz Max. apparent resistance (+/- 20%)
100 16 50/60 0.32 ohm
120 14 50/60 0.32 ohm
230 8 50/60 1.28 ohm

Line Inductance
0.015 mH/volt maximum

Duty Cycle
Explorer Detail 50%
All other scan modes 100%

Exam Mode Performance


Typical
Duration Skin
in vivo
(seconds) Entrance
Default scan @ Default Dose Nominal
Exam Type Length (in.) Scan Site (mGy) (%)
Length
AP Spine Detail 8.0 163 Lumbar Spine 0.25 1.0
AP Spine Explorer 8.0 163 Lumbar Spine 0.094 1.0
AP Spine Survey 8.0 122 Lumbar Spine 0.07 1.0
Hip Explorer 6.0 123 Proximal Femur 0.094 1.0
Hip Survey 6.0 92 Proximal Femur 0.07 1.0
Whole Body 77.0 403 Whole Body 0.012 1.0
Forearm 6.0 62 Forearm 0.05 1.0
Decubitus Lateral 8.0 163 Lumbar Spine 0.25 2.6
Spine

1-10
Explorer Technical Manual

AC Power Isolation Module


Electrical Protection: Circuit Breaker according to rated line voltage (see Line Voltage and
Maximum Line Current)
Electrical Indicator: Green light indicating output ON
Input - Output voltage: selectable, Input (see Line Voltage and Maximum Line Current),
Output two circuit 120VAC

1-11
Explorer Technical Manual

1-12
Section 2
FUNCTIONAL DESCRIPTION
This section provides a detailed functional description of the system with interconnection
diagrams. Refer to “Functional Overview” on page 1-3 for a block diagram (Figure 1-3) and a
brief functional description of each block.

Computer
The scanner is interfaced to a PCI Bus computer which controls table and C-arm movement, X-
ray generation, performs all necessary calculations, and manages both the patient and QC
database information.
The computer is a Pentium Celeron based (or higher) PC that comes equipped with floppy disk
drive, hard disk drive, keyboard, mouse, 17” color monitor, and CDROM R/W for archiving scans
and backups of the system database. For details pertaining to the computer and its associated
components, please refer to the documentation shipped with each unit.
Note: Refer to Appendix A for computer specifications and software loading instructions.

PCI Communications Controller Board


The PCI Communications Controller Board handles all the communications between the
Computer and the Scanner C-Arm and Table assemblies. The board resides in one of the computer
internal ISA PCI slots and communicates with the computer via the computer's I/O bus. It
connects to the Distribution Board in the Scanner through a 50-conductor ribbon cable. This cable
contains two independent communications links (one synchronous and one asynchronous) and
additional system control signals. Each signal requires a pair of conductors for differential
(RS422) noise immunity.
The asynchronous link communicates with the Motor Controller Boards (TX, TY, and AY), the C-
Arm Interface Board, and the Control Panel Controller section of the Detector Board. The
synchronous link communicates with the Data Acquisition System (DAS).

Interface Connections
Table 2-1 describes the interconnections between the PCI Communications Controller Board and
the Distribution Board. The table also identifies the connectors and their pin assignments.
Table 2-1 Communications Controller Board/Distribution
Board Interconnection Description
Signal Pair Description CC1 Pin Dist2 Pin
ATD+ Asynchronous data to the Scanner. JP1-2 JP10-3
ATD- JP1-27 JP10-4
STD+ Synchronous data to the Scanner. JP1-28 JP10-6
STD- JP1-4 JP10-7

2-1
Explorer Technical Manual

Signal Pair Description CC1 Pin Dist2 Pin


STCLK+ Synchronous data clock from Communications JP1-5 JP10-9
STCLK- Controller Board to Distribution Board. JP1-30 JP10-10
Synchronizes data to the Scanner.
STFRM+ Synchronous data frame from Communications
JP1-31 JP10-12
STFRM- Controller Board to Distribution Board.
JP1-7 JP10-13
ARD+ Asynchronous Data from the Scanner. JP1-8 JP10-15
ARD- JJP1-33 JP10-16
SRD+ Synchronous Data from the Scanner. JP1-34 JP10-18
SRD- JP1-10 JP10-19
SRCLK+ Synchronous data clock from Communications JP1-11 JP10-21
SRCLK- Controller Board to Distribution Board. JP1-36 JP10-22
Synchronizes data from the Scanner.
SRFRM+ Synchronous data frame from Distribution JP1-37 JP10-24
SRFRM- Board to Communications Controller Board. JP1-13 JP10-25
EMERGENCY_IN+ Signals an emergency condition. Generated by JP1-14 JP10-27
EMERGENCY_IN- the C-Arm Interface Board. JP1-39 JP10-28
ZEROX+ AC line zero-crossing signal used for system JP1-40 JP10-30
ZEROX- wide synchronization. Generated by the C-Arm JP1-16 JP10-31
Interface Board.
INTEGRATE+ Synchronous signal for Detector Integrate JP1-17 JP10-33
INTERGATE- period. Generated by the C-Arm Interface JP1-42 JP10-34
Board.
SYSRESET+ Resets the Scanner controllers. JP1-20 JP10-39
SYSRESET- JP1-45 JP10-40
EMERGENCY+ Removes power from the Scanner motor drivers JP1-49 JP10-48
EMERGENCY- and the X-ray system JP1-25 JP10-4
Notes: 1. CC = PCI Communications Controller Board
2. Dist = Distribution Board
ATD+, ATD-
STD+, STD-
STCLK+, STCLK-
STFRM+, STFRM-
PCI ARD+, ARD-
COMMUNICATIONS SRD+, SRD DISTRIBUTION
CONTROLLER SRCLK+, SRCLK BOARD
BOARD SRFRM+, SRFRM
EMERGENCY IN+, EMERGENCY IN-
ZERPX+. ZEROX-
INTEGRATE+, INTERGRATE-
SYSRESET+, SYSRESET-
EMERGENCY+, EMERGENCY-

KP1085_003-0201

Figure 2-1 PCI Communications Controller Board/Distribution Board


Interconnection Diagram

2-2
Explorer Technical Manual

Distribution Board
The Distribution Board provides interconnections between the Operator's Console (PC)
and the Scanner. It passes several signal lines from the Operator's Console and power lines
from the DIN rail directly to the C-Arm Interface module. It also provides buffering and
individual drivers and receivers for various signal lines to and from individual Scanner
modules and the Communications Controller. The Distribution Board is located in the
Electronics Tray in the base of the Scanner.
One cable connects the Operator's Console (PC) communications bus to the Distribution
Board. One cable connects to the Distribution Board from the DIN rail. This cable brings
DC power to the Distribution Board and connects the X-Ray On and Emergency signal
lines to the Power Module.
Up to eight cables connect the Distribution Board to the various Scanner boards. Three
cables connect to the three Motor Controller Boards (Table X, Table Y and Arm Y). Two
cables (one signal and one power) connect to the C-Arm Interface Board. In addition, one
cable connects to the scanner Control Panel Interface Board.
The Distribution Board has provision for three jumpers that can be installed to override
the EMERGENCY signal lines when troubleshooting.

Power
The Distribution Board receives +24 and +/-15VDC from the Multi-voltage DC Switching
Power Module. The +24VDC is applied through four individual circuit breakers to the
Table X, Table Y and Arm Y Motor Controller Boards. The +24 and +/-15VDC are
passed to the C-Arm Interface Board. The +24 VDC is also reduced to +7 and +5VDC by
regulators to power op-amps and analog switches located on this board. The +7VDC is
passed to the Control Panel Controller Board. The +5VDC powers the digital section of
the Distribution Board. The +24VDC power supply is not closely regulated and its outputs
may range from +24V to +35V under normal conditions.
Note: +7VDC may measure anywhere from +6.25VDC to +7.25VDC. This is true
everywhere +7VDC is shown in this manual.
Limits for +/- 15VDC
Six green LEDs indicate the status of the +28, +24, +15, -15, +7 and +5VDC (ON
indicates the respective voltage is present). Five red LEDs indicate the status of the five
circuit breakers applying voltage to the motor drivers/controller. ON indicates the circuit
breaker has been tripped by an over-current condition.

Interface Connections
Figure 2-2 shows connections to/from the Distribution Board.

2-3
Explorer Technical Manual

OPERATOR’S INSTRUMENT
CONSOLE
COMPUTER C-ARM
INTERFACE
BOARD

COMMUNICATIONS
CONTROLLER
BORAD CONTROL PANEL
CONTROLLER
BORAD

TX MOTOR
DISTRIBUTION CONTROLLER
BOARD BOARD
+24 VDC, +/-15 VDC
EMERGENCY
SIGNAL LINE
TY MOTOR
CONTROLLER
BOARD

AY MOTOR
+24 VDC CONTROLLER
BOARD

AW-00335_001-0202

Figure 2-2 Distribution Board High Level Interconnection Diagram

Motor Controller Board


The Motor Controller Board is a microprocessor controlled power driver circuit for use
with a two-coil bipolar stepper motor. It receives high-level commands through the
Distribution Board from the host computer, and applies 24-volt pulses to the stepper motor
windings. The system uses three identical Motor Controller Boards to control and drive
the Table X (Table In/Out), Table Y (Table Left/Right), and C-Arm Y (C-Arm Left/Right)
stepper motors. The motor windings are driven by two integrated H-bridges. These
integrated circuits provide internal level conversion and power limiting. Their logic level
control inputs are driven from a stepper motor control microcircuit that receives
commands from the microprocessor. The control circuit senses the current in the motor
windings and adjusts the duty cycle of the applied voltage in such a way as to limit the
maximum motor current. The maximum value is determined by an 8-bit control word at a
Digital to Analog Converter.
Each Motor Controller Board monitors the position of its respective mechanism using a
voltage received from an associated Position Encoder connected to the mechanism. The
Position Encoder is a precision potentiometer that divides a +/-3V reference source. The
output voltage is fed to a sense amplifier in proportion to the position of the mechanism
driven by the motor. The sense amplifier output is converted to digital value that provides
position feedback to the microprocessor.

2-4
Explorer Technical Manual

The Motor Controller Boards receive movement commands from the system computer via
the communications bus. Each Motor Controller Board contains an ID switch and four
status indicators (LEDs). The ID switch is a 16-position rotary encoded switch (SW1) that
is read during system initialization to determine the Motor Controller Board address for
communicating with the system computer. ID switch settings for the three Motor
Controller Boards are as follows:
Table X drive 4
Table Y drive 5
C-Arm Y drive 7
The four red status LEDs provide visual indications of motor drive power on, Stepper
CPU active, motor drive direction, and motor step pulses.
The Motor Controller Board also has provision (JP2) for connecting limit switches to
inhibit motor operation when the mechanism goes beyond established mechanical limits.
This feature is not used in the system and, therefore, no cable is connected to JP2.

Power
Power input to the Motor Controller Board is +24VDC. This voltage provides the motor
drive power and is converted down to +5VDC for use by logic circuits on the board. The
+5VDC is also converted to -5VDC. Two green LEDs provide a visual indication of the
power present on the Motor Controller Board.

Interface Connections
Figure 2-3 shows the typical interconnections between the Distribution Board, the Motor
Controller Board, the Stepper Motor, and the Position Encoder. Table 2-2 describes the
interconnections between the Distribution Board and the Motor Controller. Table 2-3
describes the interconnections between each Motor Controller and its respective stepper
motor and position encoder. The tables also identify the connectors and their pin
assignments.
To/From
Distribution Board ( )
( )
ARD+, ARD- ( )
STEPPER
ATD+, ATD- ( ) MOTOR
SYSRESET+, SYSRESET- MOTOR
+28V
CONTROLLER +REF
28V_RET ( ) POSITION
-REF ENCODER
KP1085_003-0203

Figure 2-3 Distribution Board/Motor Controller Board Interconnection Diagram


Table 2-2 Distribution Board/Motor Controller Board
Interconnection Descriptions
Signal Description Table Table C-Arm Pin(s)
X Y Y
ARD+ Asynchronous Receive Data. JP7 JP5 JP11 11
ARD- 12

2-5
Explorer Technical Manual

Signal Description Table Table C-Arm Pin(s)


X Y Y
ATD+ Asynchronous Transmit Data. JP7 JP5 JP11 14
ATD- 15
SYSRST+ System Reset. Resets the JP7 JP5 JP11 17
SYSRST- Motor Controller Board. 18
28V DC power for the Motor JP7 JP5 JP11 2, 3, 4, 5
28V_RET Controller Board. 1, 6. 7. 8
Table 2-3 Motor Controller Board/Stepper Motor and Position Encoder
Interconnection Descriptions
Signal Description Pin
(No label) Motor drive signals (4). JP5-1 - JP5-4
+REF (+3V) Precision positive voltage to position potentiometer. JP3-1
(No label) Position encoder wiper return voltage. JP3-3
-REF (-3V) Precision negative voltage to position potentiometer. JP3-5

Control Panel Controller Board


The Control Panel Controller Board on the system is part of the Detector Board. Its
function is to interface the Scanner Operator Control Panel to the computer allowing the
software to determine the state of the Operator Control Panel switches, to define the state
of the various Operator Control Panel LEDs, and to control motor movement. The board is
located in the C-Arm.
The Control Panel Controller communicates with the Operator's Console computer using
the asynchronous communications signals ARD and ATD of the communications buss and
the system control signals SYSRESET, XRAY_LIGHT, and EMERGENCY.

Power
The Control Panel Controller Board receives +7VDC from the Distribution Board. The +7
is reduced to +5VDC to power the circuitry on this board and is applied to the Operator
Control Panel to power the LEDs. A green LED, on this board, provides a visual
indication of the +5VDC power (ON indicates the voltage is present).

Interface Connections
Figure 2-4 shows the interconnections between the Distribution Board, Control Panel
Controller Board, and the Operator Control Panel.
Table 2-4 describes the interconnections between the Distribution Board and the Control
Panel Controller. Table 2-5 describes the interconnections between the Control Panel
Controller and the Operator Control Panel. The tables also identify the connectors and
their pin assignments.

2-6
Explorer Technical Manual

To/From
Distribution Board
ARD+, ARD- L0_PWR* - L7_PWR*
ATD+, ATD- TZ_PWR*
SYSRESET+, SYSRESET- XRAY_LIGHT_PWR*
EMERGENCY+, EMERGENCY- CONTROL SW0 - SW2 OPERATOR
XRAY_LIGHT+, XRAY_LIGHT- PANEL SR0 - SR2 CONTROL
EMERBENCY_PANEL CONTROLLER EMERGENCY_PANEL PANEL
HW_EMERGENCY_RET HW_EMERGENCY_RET
+7V +5V

AW-00335_001-0205
To/From
TILT_A
C-Arm
TILT_B
Tilt Switch
(A/SL)

Figure 2-4 Control Panel Controller Interconnection Diagram


Table 2-4 Distribution Board/Control Panel Controller
Interconnection Descriptions
Signal Description Pin
ARD+ Asynchronous Receive Data from the JP2-4
ARD- Communications Controller via the JP2-5
Distribution Board.
ATD+ Asynchronous Transmit Data to the JP2-7
ATD- Communications Controller via the JP2-8
Distribution Board
XRAY_LIGHT+ X-Ray Light from the X-Ray Controller via JP2-16
XRAY_LIGHT- the C-Arm Interface and Distribution Boards. JP2-17
EMERGENCY_CPANEL State of the STOP switch and of the collision JP2-19
HW_EMERGENCY_RET sensor. (Part of the safety daisy chain.) JP2-20
+7V DC power for the Control Panel Controller JP2-2
Board.
Table 2-5 Control Panel Controller/Control Panel Interconnection Descriptions
Signal Description CPC1 CP2
Pin Pin
L0_PWR* Turns on the ENABLE switch LED. JP6-11 JP1-11
L1_PWR* Turns on the HOME switch LED. JP6-12 JP1-12
L2_PWR* Turns on the LOAD switch LED. JP6-13 JP1-13
L3_PWR* Turns on the TABLE switch IN/OUT LED. JP6-14 JP1-14
L4_PWR* Not used. JP6-15 JP1-15
L5_PWR* Turns on the C-ARM switch RIGHT/LEFT JP6-16 JP1-16
LED.
L6_PWR* Not used. JP6-17 JP1-17
L7_PWR* Turns on the Laser LED. JP6-18 JP1-18

2-7
Explorer Technical Manual

Signal Description CPC1 CP2


Pin Pin
XRAY_LIGHT_PWR* Turns on the X_RAY LED JP6-19 JP1-19
SW0 Control signal to determine the state of the JP6-4 JP1-4
C-ARM and LASER switches.
SW1 Control signal to determine the state of the JP6-5 JP1-5
TABLE IN/OUT switches.
SW2 Control signal to determine the state of the JP6-6 JP1-6
LOAD, HOME and ENABLE switches.
SR0 Returns the state of the C-ARM switch JP6-1 JP1-1
LEFT position when SW0 is active.
Signals the state of the TABLE switch IN
position when SW1 is active. Signals the
state of the LOAD switch when SW2 is
active.
SR1 Returns the state of the C-ARM switch JP6-2 JP1-2
RIGHT position when SW0 is active.
Signals the state of the TABLE switch
OUT position when SW1 is active. Signals
the state of the HOME switch when SW2 is
active.
SR2 Returns the state of the LASER and JP6-3 JP1-3
ENABLE switches when SW2 is active.
EMERGENCY_CPANEL Returns the state of the STOP switch. (Part JP6-23 JP1-23
HW_EMERGENCY_RET of the safety daisy chain.) JP6-24 JP1-24
+5V Provides power for the Control Panel JP6-21 JP1-21
LEDs. JP6-22 JP1-22
Notes: 1. CPC = Control Panel Controller
2. CP = Control Panel

C-Arm Interface Board


The C-Arm Interface Board distributes DC power and signals to the Data Acquisition
System (DAS) and provides control electronics for the devices located in the C-Arm
assembly. It passes several signal and power lines from the Distribution Board directly to
the DAS. It also provides buffering for various signal lines. The C-Arm Interface Board is
located near the rear of the lower C-Arm’s horizontal shelf.
The C-Arm Interface provides circuitry to:
• Control and monitor operation of the X-ray Controller Board. Four LEDs (two red,
one green, and one yellow) on this board, provide a visual indication of the status
of the X-Ray Controller and the X-ray control circuitry of this board.
• Generate timing references to the DAS and Communications Controller.
• Control power to the Positioning Laser.

2-8
Explorer Technical Manual

• Move the stepper motor of the Reference Drum device and monitor the encoded
signals returned from each drum of the device. Two green LEDs, on this board,
provide a visual indication of the Reference Drum operation.
• Generate +7VDC for the DAS system

Continuity Daisy Chain


The C-Arm Interface Board is part of two linked chains of boards. Removing any of these
boards conveys an alarm message to the host computer indicating that the electrical
integrity of the system has been compromised. This message can be decoded by the host
computer to detect the extent of the damage.

Power
The C-Arm Interface Board receives +24 and +/-15VDC from the Distribution Board. The
+24VDC powers the Reference Drum stepper motor driver circuitry. The +/-15V is passed
through this board to the DAS. The +24VDC is also reduced to +5 and +7VDC. The
+5VDC powers the digital section of this board and the laser while the +7VDC is applied
to the DAS.
Three green LEDs, on this board, provide a visual indication of the +24 and +5VDC
power (ON indicates the respective voltage is present). The third LED monitors the
+28VDC. On the system, this voltage is actually +24VDC.

Interface Connections
Figure 2-5 shows the interconnections between the Distribution Board and the C-Arm
Interface Board. Table 2-6 describes the interface signals and identifies the connectors and
their pin assignments.

2-9
Explorer Technical Manual

CONTINUITY 1

ATD_CARM+, ATD_CARM-
STD+, STD-
STCLK+, STCLK-
STFRM+, STFRM-
ARD_CARM+, ARD_CARM-
SRD+, SDR-
SRCLK+, STCLK-
SRFRM+, SRFRM-
EMERGENCY_CARM, EMERGENCY_CPANEL
LINESYNC+. LINESYNC-
INTEGRATE+. INTEGRATE-
DISTRIBUTION SYSRST_CARM+, SYSRST_CARM-
C-ARM
BOARD INTERFACE
EMERGENCY+, EMERGENCY-
BOARD
XRAY_LIGHT+, XRAY_LIGHT-

+15V
+15V_RET
-15V
+24V
24V_RET
28V
28V_RET

CONTINUITY 2

KP1085_003-0206

Figure 2-5 Distribution Board/C-Arm Interface Board


Interconnection Diagram
Table 2-6 Distribution Board/C-Arm Interface Board
Interconnection Descriptions
Signal Description DIST1 C-ARM2
Pin Pin
ARD_CARM+ Asynchronous data to the C-Arm JP1-3 JP1-3
ARD_CARM- Interface Board. JP1-4 JP1-4
STD+ Synchronous data through the C-Arm JP1-6 JP1-6
STD- Interface Board to the DAS. JP1-7 JP1-7
STCLK+ Synchronizes data through the C-Arm JP1-9 STCLK+
STCLK- Interface Board to the DAS. JP1-10 STCLK-
STFRM+ Synchronous channel data frame from JP1-12 JP1-12
STFRM- Communications Controller through JP1-13 JP1-13
the Distribution Board to the DAS.
ATD_CARM+ Asynchronous Data from the from the JP1-15 JP1-15
ATD_CARM- C-Arm Interface Board. JP1-16 JP1-16
SRD+ Synchronous Data through the C-Arm JP1-18 JP1-18
SRD- Interface Board from the DAS. JP1-19 JP1-19
SRCLK+ Synchronizes data through the C-Arm JP1-21 JP1-21
SRCLK- Interface Board from the DAS. JP1-22 JP1-22

2-10
Explorer Technical Manual

Signal Description DIST1 C-ARM2


Pin Pin
SRFRM+ Synchronous channel data frame JP1-24 JP1-24
SRFRM- through Distribution Board to JP1-25 JP1-25
Communications Controller from the
DAS.
XR_ZEROX_CC+ AC line zero-crossing signal used for JP1-30 JP1-30
XR_ZEROX_CC- system wide synchronization. JP1-31 JP1-31
Generated by the C-Arm Interface
Board.
INTEGR_CC+ Synchronous signal for Detector JP1-33 JP1-33
INTERG_CC- Integrate period. Generated by the C- JP1-34 JP1-34
Arm Interface Board.
SYSRST_CARM+ Resets the C-Arm Interface Board. JP1-39 JP1-39
SYSRST_CARM- JP1-40 JP1-40
EMERGENCY_CARM+ Removes power from the Scanner JP1-42 JP1-42
HW_EMGNCY_RET- motor drivers and the X-ray system JP1-43 JP1-43
XRAY_LIGHT+ Applies power to the AUX X-RAY JP1-48 JP1-48
XRAR_LIGHT- light outlet on the Power Console. JP1-49 JP1-49
CONTINUITY 1 Emergency shutdown daisy chain. JP1-1 JP1-1
CONTINUITY 2 JP1-5 JP1-5
+15V Powers the Data Acquisition System. JP4-1 JP9-1
-15V JP4-3 JP9-3
15V_RET JP4-2 JP9-2
24V Generates +5VDC to power the digital JP4-4 JP9-4
24V_RET section of the C-Arm Interface Board, JP4-5 JP9-5
and +7VDC for power to the Data
Acquisition System.
28V Powers the stepper motors and fan. JP4-6 JP9-6
28V_RET JP4-7 JP9-7
Notes: 1. Dist = Distribution Board
2. C-Arm = C-Arm Interface Board

Figure 2-6 shows the boards and assemblies that connect to the C-Arm Interface Board.

2-11
Explorer Technical Manual

KP1085_003-0207

Figure 2-6 C-Arm Interface Board High Level Interconnection Diagram

X-Ray Controller Assembly (P/N ASY-00409)


This assembly provides pulsed power to the primary winding of the high voltage
transformer in the X-Ray Source Unit and AC power to the primary winding of the
filament transformer. It consists of a single printed circuit board and several large
components contained in a chassis mounted at the front of the lower C-arm, just in front of
the Tank Assembly. The XRC receives split 240VAC power from the DIN Rail Power
Distribution Module. It also receives command and timing data from the C-Arm
Controller Board and it provides a line frequency timing signal and housekeeping and
diagnostic data to the C-Arm Interface Board.

Interface
Figure 2-8 shows the interface connections between the C-Arm Interface Board and the X-
Ray Controller Assembly. Table 2-13 describes the interface signals and identifies the
connectors and their pin assignments. Note that the AC input power comes directly from
the DIN Rail Power Module and connects to the X-Ray Controller Assembly.

2-12
Explorer Technical Manual

XR_RLY_ON+, XR_RLY_ON-
XR_FREQ+, XR_FREQ-
HIGH VOLTAGE
XR_BEAMON+, XR_BEAMON-
TRANSFORMER
XR_ISET+, XR_ISET- PRIMARY
XR_KV1+, XR_KV1-
C-ARM XR_KV0+, XR_KV0-
INTERFACE
ACLINE+, ACLINE- X-RAY X-RAY
BOARD
LIGHTON+, LIGHTON- CONTROLLER SOURCE
IBEAM+, IBEAM- ASSEMBLY UNIT
KVP+, KVP-
XRFAULT+, XRFAULT- X_RAY TUBE
FILAMENT
TRANSFORMER
From DIN Rail 240 VAC PRIMARY
Power Module
KP1085_003-0208

Figure 2-7 C-Arm Interface Board/X-Ray Controller


Assembly Interconnection Diagram
Table 2-7 C-Arm Interface Board/X-Ray Controller Assembly
Interconnection Description
Signal Description I/O
C-Arm Logic XRC
Pins Pins Pins
XR_RLY_ON+ Allows the energy storage capacitor to be JP12-1 JP7-1 JP3-1
XR_RLY_ON- “trickle charged” before applying full JP12-2 JP7-2 JP3-20
power to avoid large turn-on current surges
that could cause the circuit breaker to trip.
XR_FREQ+ States whether the line frequency is 50 or JP12-3 JP7-3 JP3-2
XR_FREQ- 60Hz. JP12-4 JP7-4 JP3-21
XR_BEAMON+ Controls the ON/OFF status of the X-Ray JP12-5 JP7-5 JP3-3
XR_BEAMON- beam. JP12-6 JP7-6 JP3-22
XR_ISET+ Selects the X-Ray beam current (3 or JP12-7 JP7-7 JP3-4
XR_ISET- 10mA). JP12-8 JP7-8 JP3-23
XR_kV1+ Selects the X-Ray beam energy (80, 100, JP12-9 JP7-9 JP3-5
XR_kV1- 120 or 140kVp). JP12-10 JP7-10 JP3-24
XR_kV0+ JP12-11 JP7-11 JP3-6
XR_kV0- JP12-12 JP7-12 JP3-25
ACLINE+ States the phase of the power frequency. JP12-19 JP7-19 JP3-10
ACLINE- JP12-20 JP7-20 JP3-29
LIGHTON+ States whether the X-Ray beam is ON/OFF. JP12-21 JP7-21 JP3-11
LIGHTON- This signal controls the X-Ray ON lights of JP12-22 JP7-22 JP3-30
the C-Arm Control Panel, the table Control
Panel, and the Operator's Console Power
Module. It also controls a remote X-ray ON
light through the Power Module when one
is connected.

2-13
Explorer Technical Manual

Signal Description I/O


C-Arm Logic XRC
Pins Pins Pins
IBEAM+ Value of current pulses at the X-ray source. JP12-23 JP7-23 JP3-12
IBEAM- This is a frequency modulated diagnostic JP12-24 JP7-24 JP3-31
signal whose frequency is proportional to
the quantity being monitored.
kVp+ Value of voltage pulses at the X-ray source. JP12-25 JP7-25 JP3-13
kVp- This is a frequency modulated diagnostic JP12-26 JP7-26 JP3-32
signal whose frequency is proportional to
the quantity being monitored.
XRFAULT+ States whether or not any fault condition JP12-27 JP7-27 JP3-14
XRFAULT- exists in the X-Ray Controller Assembly. JP12-28 JP7-28 JP3-33

X-Ray Controller Board


The X-ray Controller Board consists of five sections: Power, Timing Generator, Pulse
Generator, Filament Control, and H-Bridge.
The Power section converts the split 240 VAC input to the controller into the DC voltages
necessary for the board. The voltages produced in the Power section are: +5VDC (VDD –
used for analog circuitry), +5VDC (VCC – used for digital circuitry), and +/- 15 VDC.
The LED, D25, indicates the presence of voltage on the board.
The Timing Generator section produces the timing signals required by the control circuits.
The timing is based on a half cycle of the AC waveform and is, therefore, reset at each
zero crossing. An 8 MHz clock circuit provides pulses to a counter. This counter in turn
provides its output count to an EPROM as its address inputs. The pulse trains produced by
the EPROM are clocked into a register to clean up the signals that are then distributed to
various sections of the board. The ACHIGHLOW signal produced by the timing circuits is
sent to the C-Arm Interface Board where it is used to synchronize X-ray generation with
the Filter Drum position.
The Pulse Generator regulates and shapes the pulses delivered to the X-ray Tank. Sense
lines from the tank set the current threshold for the regulator. A reference voltage is
selected and fed to one of four circuits that control the high voltage peak potential. The
circuits adjust the reference voltage through four potentiometers, R55, R54, R39, and R40
that is used to establish the 80, 100, 120, and 140 KVp. The outputs of these circuits are
sent to a pulse width modulator circuit to produce the control pulses to the H-Bridge.
These pulses are optically coupled to the FETs in the H-Bridge.
The H-Bridge produces the actual pulses fed to the primary of the HV transformer in the
X-ray tank. AC is brought to the circuit, rectified, and filtered. The H-Bridge operates as
both an inverter and a pulse width modulator. The H-Bridge uses two pairs of FETs. Each
pair conducts 256 times for a total pulse width of 4.096 milliseconds. Each of the FETs is
individually biased with the biasing network acting as a bleeder for the large storage
capacitors. The current load in the H-bridge is sensed and adjusted. The output voltage
from the H-Bridge is up to 200 volts peak.

2-14
Explorer Technical Manual

The Filament circuit is a regulator. It receives current feedback from the tank indicating
the tube current. 10 volts at the feedback input is equivalent to 10 milliamps of current in
the tube. The 10 volts are reduced to 5 and then fed to an A/D circuit to be converted to a
digital value for use by the control circuits. The value into the A/D is sampled towards the
end of the pulse where it is stable. After conversion, the digital value is fed as an address
to an EPROM. If the current is correct, the output of the EPROM will be 0 volts. Anything
other than 0 volts (“+” = high current, “-“= low current) will be added and accumulated in
two stages forming an error integrator. This digital error number is converted back to an
analog value and then used as the threshold of a comparator. The output of the comparator,
through additional circuitry, adjusts the duty cycle of the modulated filament voltage.
Refer to Table 2-8 for pin assignments.

X-Ray Source Unit


The X-Ray Source (commonly referred to as the Tank assembly) consists of the X-Ray
tube, the X-Ray tube filament transformer, the high voltage transformer, the high voltage
rectifier circuit, and the sensing circuits that monitor the high voltage applied to the X-ray
tube and the beam current.
Table 2-8 describes the interconnections between the X-Ray Controller Assembly and the
X-Ray Source unit. The table also identifies the X-Ray Controller Assembly board
connector and pin assignments for each interconnection signal.
Table 2-8 X-Ray Controller Assembly/X-Ray Source Unit Interface
Signal Description XRC Board Controller Tank
Pins Terminal
HV_XFMR_PRI+ High Voltage Transformer H-Bridge JP1-1, JP1-2 TB1-3
HV_XFMR_PRI- Primary (JP1) JP1-3, JP1-4 TB1-4
FIL_XFMR_A I/O & Logic JP2-7 TB1-6*
FIL_XFMR_B (JP6) JP2-8 TB1-9
IF+ Beam Current Sense “+” I/O & Logic JP2-4 TB1-10
(JP6)
IF- Beam Current Sense “-” I/O & Logic JP2-5 TB1-7
(Chassis) (JP6)
TP1 Anode pulse monitor I/O & Logic JP2-2 TB1-12
(JP6)
TP2 Cathode pulse monitor I/O & Logic JP2-1 TB1-11
(JP6)
* A thermal overload protector is installed in series with the filament transformer primary
winding, between TB1-6 and TB1-8.

Data Acquisition System


The Explorer Data Acquisition System (DAS) consists of a single board that has solid
state (diode) detectors, the integration circuitry along with the analog-to-digital

2-15
Explorer Technical Manual

conversion circuitry, and multiplexing circuitry to select the outputs in a sequential


fashion. The solid-state detectors are photodiodes with a cadmium tungstate (CdWO4)
crystal attached. X-rays striking the cadmium tungstate crystals are converted into photons
of visible light. The diodes sense this light and convert the light into a current, which is
amplified in a current to voltage converter. The converter output is applied to an integrator
through analog switches. The analog switches all operate in parallel and are turned on
during the X-ray pulse. These switches are turned off during the integrator hold time to
prevent the introduction of integration noise into the data. The turning on and off of these
switches is controlled by the INTEGRATE signal under control of the software. The
signal from the detectors is then integrated with the final voltage obtained held in a
sample/hold circuit. This voltage is sampled, in photodiode order, by the multiplexor
circuitry under control of the software. The output voltage is then sent to the analog-to-
digital conversion circuitry where it is changed to a digital value usable by the software.
The Detector Assembly is located in the upper end of the C-arm directly above the X-ray
source. There is a lead radiation shield between the detector board and the cover to stop
any X-rays that may get through the detectors.

Solid State Detector


The Solid State Detector converts X-rays into signals that are applied to the Integrator/
Multiplexor section of the Detector Assembly board. Each Detector Assembly contains
high-resolution detectors. Each detector is 4mm wide at the detector.

Integrator/Multiplexor Subsection
The Integrator/Multiplexor subsection of the Detector Assembly receives up 54 signals in
parallel from the Solid State Detector Boards. This subsection integrates and stores those
signals and then applies the stored signals to the Analog to Digital subsection of the
Detector Assembly.
Switched signals charge integrating capacitors in this subsection during a given charging
time. After the charging time, the switched signals are turned off and the charges are held
on the capacitors. Each integrator is sampled, in photodiode order, by a multiplexor and
the output sent to the Analog to Digital subsection. At the end of sampling, all the
integrating capacitors are discharged in parallel by shorting them out with analog
switches.
Each integrator has an additional input into which a test signal (TESTLVL) can be
applied when there are no X-rays present. This test signal is used to verify the operation of
the integrators and multiplexors by the SQVERIFY diagnostic program.

Analog To Digital Board


The Analog to Digital (ADC) subsection of the Detector Assembly converts analog
signals received from the Integrator/Multiplexor subsection to a digital format.
Analog signals from the Integrator/Multiplexor subsection are applied to differential
amplifiers on the ADC subsection. Outputs from the differential amplifiers are combined

2-16
Explorer Technical Manual

in a final multiplexor consisting of four analog switches. The multiplexed signals pass
through a programmable gain amplifier and summing amplifier before being applied to an
A/D converter. A one-volt fixed DC offset is inserted at the summing amplifier to insure
that no channels ever go negative. The A/D converter converts the analog signal into 16
bit parallel data for processing by a Digital Signal Processor.
The ADC subsection uses a Digital Signal processor to generate all the control signals
necessary for the Detector Assembly. This processor also provides a high-speed serial data
link to the computer.

Power
The 54-Channel Detector Assembly receives +/-15V and +5V from the C-Arm Interface
Board. The +/-15V is passed through this board to the Detector subsection. It is also
regulated to +/- 12V by series regulators to power operational amplifiers and analog
switches located on this board. The +/- 15V is also used to generate +/- 5V to power the
analog-to-digital converter circuit. The +7V is reduced to +5V to power the digital section
of this board.

Interface Connections
Figure 2-8 shows the interconnections between the 54-Channel Detector Assembly and
the C-Arm Interface Board. Figure 2-8 describes the interface signals and identifies the
connectors and their pin assignments.
STD+, STD-
STCLK+, STCLK-
STFRM+, STFRM-
SRD+, SRD-
SRCLK+, SRCLK-
SRFRM+, SRFRM-
SYSRST_DAS+, SYSRST_DAS-
C-ARM INTERFACE INTEGRATE+, INTEGRATE- 54-Channel
BOARD XR_ZEROX_DAS+, XR_ZEROX_DAS- Detector
Assembly

+15V
15V_RET
-15V
+7V

KP1085_003-0209

Figure 2-8 C-Arm Interface Board/54-Channel Detector Assembly


Interconnection Diagram

2-17
Explorer Technical Manual

Table 2-9 C-Arm Interface Board/54-Channel Detector Assembly


Interconnection Descriptions
Signal Description C-Arm1 DA2
Pin Pin
STD+ Synchronous data to the Analog/Digital JP10-3 P1-12
STD- Converter board. JP10-4 P1-13
STCLK+ Synchronizes data to the Analog/Digital JP10-6 P1-15
STCLK- Converter board. JP10-7 P1-16
STFRM+ Synchronizes DSP Communications. JP10-9 P1-18
STFRM- JP10-10 P1-19
SRD+ Synchronous Data from the Analog/Digital JP10-12 P1-3
SRD- board. JP10-13 P1-4
P1-3 Synchronizes data from the Analog/Digital JP10-15 P1-6
P1-4 Converter board. JP10-16 P1-7
SRFRM+ Synchronizes DSP Communications. JP10-18 P1-9
SRFRM- JP10-19 P1-10
SYSRST_DAS+ Resets the Analog/Digital Converter board. JP10-21 P1-21
SYSRST_DAS- JP10-22 P1-22
INTEGRATE+ Data integration signal. Generated by the C- JP10-24 P1-24
INTEGRATE- Arm Interface Board. JP10-25 P1-25
XR_ZEROX_DAS+ AC line zero-crossing signal used for system JP10-27 P1-27
XR_ZEROX_DAS- wide synchronization. Generated by the C- JP10-28 P1-28
Arm Interface Board.
+15V JP10-32 P1-32
JP10-33 P1-33

-15V Powers the Data Acquisition System. JP10-36 P1-36


JP10-37 P1-37

15V_RET JP10-30 P1-30


JP10-31 P1-31
JP10-34 P1-34
JP10-35 P1-35

+7V JP10-38 P1-38


JP10-39 P1-39
CONTINUITY Emergency shutdown daisy chain (grounded JP10-1 P1-1
on ADC board)
Notes: 1. C-Arm = C-Arm Interface Board
2. DA = 54-Channel Detector Assembly

2-18
Explorer Technical Manual

Power Module
The Power Module provides the AC voltage required by the computer system and the
Scanner. The module consists of a Main Power circuit breaker, a Power On indicator, and
a torroidal AC line input isolation transformer.
AC power from the Torroid Power Module is sent to the Scanner on a four-wire cable
providing split 240 volts to the DIN Rail. From the DIN Rail, unfiltered AC is fed to the
line filter and, from there, distributed to various sections of the Scanner. Unfiltered AC is
also fed to an AC outlet connector at the rear of the Electronics Tray. This outlet provides
AC power to the Operator’s Console power strip. A second AC outlet provides unfiltered
AC only while x-rays are being produced. This outlet is used in the factory to give a visual
indication of x-ray production during testing.

2-19
Explorer Technical Manual

2-20
Explorer Technical Manual

Section 3
INSTALLATION
Required Tools
When installing the Explorer, a tool kit that includes the following items is required:
• Assortment of both flat-bladed and Phillips screwdrivers
• Assortment of needle-nose and diagonal cutting pliers
• Socket drivers (full set including 1/4”, 3/8”, 7/16”, 1/2” and 3/4”)
• Open-end wrenches (full set including 3/8”, 7/16”, 1/2”, 9/16”, 5/8” and 3/4”) and an
adjustable wrench
• Hex driver (Allen wrench) set including sizes 1/16”, 5/64”, 3/32”, 7/64”, 1/8”, 9/64”, 5/
32” 3/16”, 7/32” and 1/4”
• Oscilloscope and digital multimeter
• Survey meter (Victoreen model 450P or equivalent)
• X-ray leakage test tool, #099-0566
• Measuring tape, approx. 3.7 meters (12 feet)
• Beam alignment tool (010-0923)

Required Documentation
The following documents are required:
• Explorer User’s Guide
• Computer manual
• Printer manual
• Introducing Microsoft Windows® XP Professional
• Optical Disk Drive manual (when installed)
• The Radiation Measurement Report

3-1
Explorer Technical Manual

Room and Doorway Size


The system requires minimum rom size of 2.44m (8.0ft) x 3.05m (10.0ft) and minimum
door width of 0.77m (30.0in.), see Figure 3-1.
10'
(305 cm)
119"
(302 cm)

2"
(5 cm)
To be supplied

49.5" 61"
(125.7 cm) (155 cm)

19.75" 79.5" 19.75"


(50.2 cm) (202 cm) (50.2 cm)
8'
(244 cm)
TER
EN
ROL
CK

+
L O
SC

P GU

P GD
N
P

DE
L
9

3
*

.
PS
OCK
CA
L

2
/
NUM
L OC
K

OCK

ME
UM

HO

EN
D

IN S
7

0
N
L

ON
USE

GE

GE
DW
PA

PA
UP

PA
OL L

O ME
SCR
L OC

ND
K

E
CREN

ETE
ER
R IN T

IN S T

EL
D
P
S

28"
F1
2

RL
CT
F10 F11

TR
ENE

H IF T
S

L T
A
F9

(71 cm) 30"


F8
F7
F6
F5

(76 cm)
F4
F3
F2
F1

T
AL
OC
APSL K

S H IF T
s
Ec

CT R
TAB

L
C

30"
(72 cm)

Figure 3-1 System Dimensions


The Operator's Console may be up to 45ft (13.72m) away from the Scanner, but since
leakage and scatter radiation levels are extremely low, it can be safely located in the same
room with the Explorer itself.

Arrange for Help


Moving the unit to its final location requires at least two able-bodied people to direct the
machine, hold doors, and lift heavy parts. See the following tables.

Kg lbs
Equipment Weight 362 795
Max. Patient Weight 136 300
Total Weight 498 1095

m ft
Area 1.9 x 1.1 6.3 x 3.5

Kg/m2 lb/ft2
Floor Loading 220 45

3-2
Explorer Technical Manual

Inspect For Shipping Damage


Inspect the exteriors of all crates and boxes for shipping damage. Bring any damage
discovered to the attention of the customer's Shipping/Receiving department before
proceeding.

Figure 3-2 Crated Unit

Uncrate Unit
Remove the unit from the crate as described below:
1. Cut the strapping that holds the packaging together.
2. Remove the cardboard cap by lifting it up and off.
3. Remove the cardboard sleeve by lifting it straight up and off. Be careful not to scratch
the unit.
4. Remove the wooden table shelf and the boxes packed with the unit.
Note: There is a metal cross brace attached to the wooden table shelf. Remove and save
this cross brace as it is used later in the installation.

3-3
Explorer Technical Manual

5. Verify that the serial numbers on both crates and the scanner all match.

Inspect For Hidden Shipping Damage


Open all crates and boxes, and check for signs of hidden damage. Check the ShockWatch
and Tip-N’-Tell indicators for evidence of improper handling during shipment. Bring any
damage discovered to the attention of the customer's Shipping/Receiving department.

Take Inventory
The Explorer system is shipped in two crates. Using the following checklist, take
inventory of the contents of all crates and boxes, and confirm that all of the expected items
have been received. Unused power cables and circuit breakers are to be returned to
Hologic. Report any discrepancies to the Hologic Sales Department.

Note: Two System Backup disks are shipped with each system. One will be included
with the scanner and the other shipped in the crate with the PC. Mark the one shipped
with the scanner and note that if the Serial Numbers on the backup disks do not agree, the
disk shipped with the scanner is the correct disk.

Installation Inventory Check List


(Quantities are one (1) each unless otherwise specified)
[ ] Main Assembly [ ] Hip Wedge
[ ] Printer Paper [ ] Knee (Block) Elevation Pad
[ ] Explorer User's Guide [ ] Forearm Positioner (Optional)
[ ] Computer, Mouse, and Keyboard [ ] IVA/Decubitus Lateral Positioner
(Optional)
[ ] Laser or Business Inkjet Printer * [ ] 2” and 3” Blocks
[ ] Spine Phantom * [ ] Foot Restraint
[ ] VGA or Flat Panel Monitor * [ ] CDROM R/W (2 packs)
[ ] Explorer Console [ ] 3.5" Floppy Disks (2 boxes)
[ ] Communication Cables [ ] Miscellaneous Hardware Box
[ ] Communication Cables [ ] System Backup Disks (2)
[ ] Voltage Selection Kit (100/120/
230)
[ ] C-Arm Assembly
[ ] Hip Positioner

* Optional.

3-4
Explorer Technical Manual

Measure Path To Final Destination


Contact the department receiving the unit and request that a representative show you the
room where the unit is to be installed. As you make your way from the loading dock to the
room, measure all doorways and openings including any elevators on which you must
travel. Look for other obstacles (thresholds, steps, sharp corners, etc.) which could cause a
problem during transport. The system fits through any doorway at least 30 inches
(76.2cm) wide and 81inches (206cm) high, and hallways 45 inches (114cm) wide. All
measurements are inside dimensions.

Short Doorway
If a system must be moved through a doorway that is not at least 81" (206cm) high, the
tabletop can be removed. This allows the unit to fit through a doorway 79" (201cm) high
(inside dimension). See the instructions for removing the tabletop below.

Narrow Hallway
If a system unit must be moved through a hallway that is not at least 45" (114cm) wide, the
C-arm can be removed. This allows the unit to fit through a hallway 29" (74cm) wide
(inside dimension). See the instructions for removing the C-arm below.

AW-00335_001-0306

Figure 3-3 Uncrated Unit

Tabletop Removal (If Necessary)


The tabletop can be removed before taking the unit off the pallet. Follow the procedure
below to remove the tabletop:
1. Remove the table pad and remove the right-side table rail end cover (2 Phillips screws
located on the right end). This cover slides straight out the side.
2. Remove the head end pedestal covers. You will need to move the table to its extreme
forward position to access the center screws on each panel.

3-5
Explorer Technical Manual

3. Remove the two screws (Phillips) that secure the tabletop in place (located on top of
the table near the rear center) and slide the table to the left, far enough to access the
Table X-bearing blocks.
4. Referring to Figure 3-4, remove the right-side, front endcap (3 Phillips screws).
5. Remove the back Phillips screw from the right-side cover (of the table X-drive assem-
bly) and slide the cover out from the front.
6. Unplug the cable to the Motor Controller board.
7. Refer to Figure 3-4. Disconnect the stainless steel cable guard (2 screws) and remove
the Table X-drive attachment bracket (4 bolts and 2 nuts).
8. Remove the left rail stop (1 counter sunk screw located next to the left-side, front end-
cap) to allow the tabletop to be rolled off from the front.
9. Remove the 8 Allen screws (6mm) holding the table X-drive bearing blocks.
10. Remove the tabletop by carefully sliding it off the front of the unit (requires two peo-
ple). Be careful not to slide the bearing blocks off the rail.
11. Refer to Figure 3-4. Install the right angle bracket (found in the miscellaneous hard-
ware kit) to hold the bearing blocks and X-drive bracket in place while the scanner is
moved. If this bracket is not available, tape the bearing blocks, and X-drive bracket in
place.

KP1085_003-0307

Figure 3-4 Table X Drive

3-6
Explorer Technical Manual

Preparing the Tabletop for Moving


Warning: Do not tilt the unit to the vertical position until the Table End Bracket is
properly installed (step 4 below).

Do not tilt the unit to the vertical position until the Table End Bracket is properly installed
(step 4 below).
Before the system can be tilted to vertical, you must slide the table to the left and clamp it
so it clears the floor when the unit is tilted to vertical. To accomplish this, do the
following:
1. Remove the tabletop pad.
2. Remove the three Phillips screws holding the tabletop. Two of the screws are at the
left corners and one at the right back corner looking down on the tabletop.
3. Slide the tabletop to the right until it clears the table base panel.
4. Referring to Figure 3-8, adjust the Table End Stop as pictured.

Move Unit To Destination (Vertical Position)


Follow the procedure below to move the unit:
1. Move tank to foot end of table and reattach the shipping brackets. Remove the small
end pallet.
2. Locate the tilt arms packed with the unit (see Figure 3-5), and mount the arms on the
foot end of the unit using the jam nuts on the leveling feet and end bolts. Tighten the
end brackets.
3. Install the cross brace (shipped screwed to the wooden table shelf) being careful not
to scratch the paint.
4. Remove the four hold-down brackets (see Figure 3-8).
5. Remove the remaining corner brace.
6. Lock the casters.
7. With two people, tip the unit on end as shown in Figure 3-6 (a 1 x 6 board, or equiva-
lent, can be used to protect the floor).
8. Move the unit and everything that shipped with it to the destination room.

3-7
Explorer Technical Manual

KP1085_003-0309

Figure 3-5 Tipping Unit

AW-00335_001-0310

Figure 3-6 Moving and Tilting the Unit Down

Move Unit to Destination (Horizontal Position)


You may have to lower the unit to a horizontal position to move it to its destination if the
ground clearance is not sufficient. To do this, use the auxiliary caster assemblies and
mounting hardware shipped with the unit and follow these steps:
1. Follow the procedure in the previous section for moving the unit in the vertical posi-
tion (steps 1-8).
2. Using the hardware shipped with the unit, reference Figure 3-7 to install the auxiliary
casters to the bottom of the unit.

3-8
Explorer Technical Manual

3. Carefully tilt the unit down.

Warning: Do not do this alone. This requires two people. DO NOT use the right end
of the frame to lower the unit. Use the front and rear frame rails as handholds.

4. Move the unit to its destination.


Note: Temporarily remove the tilt arm assembly if the unit is too wide.
5. Carefully tilt the unit back to its vertical position (two people required). If the room is
small, the dolly wheels may have to be installed (see Figure 3-6) to tilt the unit down
and move sideways simultaneously.
6. Remove the auxiliary casters from the bottom of the unit and set aside.
7. Reinstall the hex bolts and flat washers into the rivet nuts.
Note: Ship all tilt-arm and caster bar assemblies back to Hologic in the box provided.
Discard the tilt arm cross brace.

AW-00335_001-0311

Figure 3-7 Auxiliary Horizontal Caster Installation

Setting Up Unit
Follow the procedure below to set up the unit:
1. Carefully tilt the unit down (if the room is tight, install the dolly wheels supplied to
tilt the unit down and move sideways simultaneously).

Warning: The unit will feel heavier when putting it down than it felt when tipping it
up. This is because the pivot point is different.

2. Remove the tilt arms and position the unit in its final location in the room.
3. Level the unit.
Note: Refer to Figure 3-8 and:

3-9
Explorer Technical Manual

4. Remove the 2 carriage-to-base shipping brackets.


5. Remove the table end stop.
6. Remove the table locking bracket located on the left rear side of the scanner.

4-Table “End” Stop


2

3-C-Arm Shipping
Bracket
3

1 2-Hold Down
Bracket (4)

4
1-Table Locking
“L” Bracket

KP1085_003-0312

Figure 3-8 Shipping Bracket Locations

3-10
Explorer Technical Manual

Install C-Arm
Follow the procedure below to install a C-arm assembly:

Warning: Do not remove C-arm shipping bracket (see Figure 3-8) until done.

1. Carefully place the C-arm on the bottom bolts. Tilt the arm back slightly and slip the
cables through the openings and then tilt the arm forward until the front bolts are in
place.
2. Put 7/16” nuts and washers on all 4 bolts and tighten.
3. Install the cables.
4. Install the T-beam end caps.
5. Insure that the scanner is level.
6. Reinstall the pedestal covers.
7. Install the tank cover.

Install Cables
Follow the procedure below to cable the system:
1. Locate the box shipped with the system containing the Voltage Selection Kit that
includes the AC Input cable, Main Circuit Breaker, jumpers wires, and mounting
hardware.
2. Locate and install the feet on the bottom of the Torroid Assembly.
3. Confirm that you have the proper kit for the site’s AC voltage (100/120/230VAC).
4. Remove foot end cover on the Electronics Tray (on the left side facing the machine)
and the rear of the Torroid Assembly.
5. Remove the AC input cable from the kit, install the strain relief, and route the cable
though rear of the Torroid Assembly.
6. Screw the strain relief into the bushing at the bottom of the Torroid Assembly and fix
the cable to the exposed stud using the cable clamp and NyLock nut provided in the
kit.
7. Attach the ground wire to the grounding lug using the hardware provided. Attach the
neutral and phase wires to the two-terminal block.
8. Install the circuit breaker from the voltage kit into the hole in the rear of the Torroid
Assembly using the screws provided in the kit
9. Attach the two wires (brown and blue) from the two-terminal block to the Line termi-
nals (top terminals) on the circuit breaker.

3-11
Explorer Technical Manual

10. Attach the two wires (brown and blue) from the 7-terminal block to the Load termi-
nals (bottom terminals) on the circuit breaker.
11. Use the jumpers provided to configure the terminal block as shown in the tables
below.
Table 3-1. 120 Volt 50/60 Hertz
Wire From To
Blue Circuit Breaker - Load Pin 1 – Header 7
120 VAC
50/60 Hz

Brown Circuit Breaker – Load Pin 7 – Header 7


Blue Jumper Pin 1 – Header 7 Pin 4 – Header 7
Brown Jumper Pin 3 – Header 7 Pin 7 – Header 7
Table 3-2. 230 Volt 50/60 Hertz

Wire From To
230 VAC
50/60 Hz

Blue Circuit Breaker - Load Pin 1 – Header 7


Brown Circuit Breaker – Load Pin 6 – Header 7
Brown Jumper Pin 3 – Header 7 Pin 4 – Header 7
Table 3-3. 100 Volt 50/60 Hertz

Wire From To
Blue Circuit Breaker - Load Pin 1 – Header 7
100 VAC
50/60 Hz

Brown Circuit Breaker – Load Pin 5 – Header 7


Blue Jumper Pin 1 – Header 7 Pin 4 – Header 7
Brown Jumper Pin 2 – Header 7 Pin 5 – Header 7
BRN ( CIRCUIT BREAKER ) BRN ( CIRCUIT BREAKER ) BRN ( CIRCUIT BREAKER )

7 7 7
6 6 6
BRN BRN
TO 5 TO 5 TO 5
4 BLU 4 BLU 4 BRN
TRANSFORMER TRANSFORMER TRANSFORMER
3 3 3
2 2 2
1 1 1
BLU ( CIRCUIT BREAKER ) BLU ( CIRCUIT BREAKER ) BLU ( CIRCUIT BREAKER )
HEADER 7 HEADER 7 HEADER 7

100VAC 50/60 HZ 120VAC 50/60 HZ 230VAC 50/60 HZ


KP1085_003-0314

Figure 3-9 Isolation Transformer Input Tap


12. Confirm that the secondary side taps are set up as shown in the diagram in Figure 3-9.
These diagrams can also be found on the inside of the Torroid Assembly Cover.
13. Replace the cover of the Torroid Assembly.
14. Route the Operator Console AC Supply Cable through the cable clamp block on the
rear of the Scanner.

3-12
Explorer Technical Manual

15. Attach the Console Power Cord to the top power receptacle. The bottom power recep-
tacle is used in the manufacturing process to power a light when X-rays are on.
16. Route the Operator Console Communications Cable through the cable clamp block on
the rear of the Scanner.
Aux To
X-Ray Console
Light AC

BLOCK 1
NORTH

BLOCK A
NORTH

GNH GNH

A B

GNH GNH

BLOCK 22
SOUTH
BLUE
AC BROWN BLOCK B
BLACK SOUTH
IN GRN/YEL
GRN/YEL

BLUE To Multi DC
Power Supply
BROWN

KP1085_003-0315

Figure 3-10 Rail AC Input/Output Wiring Diagram


16. Attach the Communications Cable to JP10 on the Distribution Board and the
ground strap to the grounding stud.
17. Attach the other end of the Communications Cable to the Communications Con-
troller Board connector at the rear of the PC.

3-13
Explorer Technical Manual

Plug the Operator Console AC Supply Cable to the Power Strip male connector.
SAFETY PRECAUTIONS
There are a number of safety precautions that MUST be observed when servicing the
Explorer systems.
HIGH VOLTAGE: Voltage levels that can injure or be fatal are present through the
Explorer systems. The line voltage (100, 120, 230 volts) is supplied to the Torroid
Assembly and the scanner. The X-ray source unit contains 140kV as well as other AC
and DC voltages. The pedestal motors use 230 volts regardless of the line voltage. Use
caution when checking, calibrating, and troubleshooting. Always trip the main breaker
when replacing components.
X-RADIATION: Service personnel are required to wear a dosimeter. Do not leave the
system unattended in X-RAY SURVEY (SURVEY mode).
ESD PRECAUTIONS: To prevent damage due to ESD (Electrostatic Discharge), you
must take precautions when handling components. Remove any charge from your body
by wearing an approved and properly grounded wrist strap. Keep PCBs in their ESD
protective bag until you are ready to install them. Treat defective PCBs as new to
prevent any additional damage.

Check Power Line Voltage


Hologic strongly recommends that the Explorer be powered from a dedicated power line.
See the product specifications in Section 1 for power requirements.

Measure Line Voltage


With an AC voltmeter, before plugging in the Explorer, measure the voltage at the outlet
from which the unit will draw power. The measured voltage must be within ± 10% of the
voltage shown on the power label (located where the power cord attaches to the Torroid
Power Module). The Explorer Torroid Assembly is a step up/step down isolation
transformer, which can be re-strapped to accommodate other input voltages. See Figure 3-
9 for the most common configurations of the isolation transformer.
There are two standards for conductor color-coding. The North American standard
specifies the BLACK conductor as LINE and the WHITE as NEUTRAL, while the
International standard defines the BROWN conductor as LINE and the BLUE as
NEUTRAL.
Note: Be sure to change the label if the transformer.ou re-strap
Note: Enter this information into The Radiation Measurement Report (see page 3-26).

3-14
Explorer Technical Manual

Measure Isolation Transformer Secondary Voltage


After plugging in the instrument and switching on the main breaker and instrument power
on the Control Panel, measure the voltage at the power strip located at the rear of the
Operator’s Console with a digital voltmeter set to measure AC volts.
Note: The voltage should be between 110 VAC and 130 VAC. If the voltage is out of this
range, recheck the voltage at the wall outlet, and the strapping of the Torroid
transformer as shown in Figure 3-9.

Install Computer
Follow the procedure below to install the computer:
1. Set up the computer cart using the enclosed instructions. Place the computer,
keyboard, monitor, and printer on the cart.
2. Install computer system and power cables.
3. Turn on the computer and confirm that all options shipped with the instrument or
listed on the sales order have been installed and function properly.
Note: For distributor supplied computers refer to Appendix A.

Starting QDR Software in Service Mode


Follow the procedure below to cable the system:
1. Log on to Windows XP as Field Service using the last 6 digits of the RTX
Runtime serial number as the password. (If this is the first time the computer is
being powered on, the password will be password.)
2. Exit the Explorer software by clicking the Exit button and selecting the Exit
QDR without shut down? option and then clicking the OK button.

KP1085_003-0317

3. Click the Start button on the desktop and then select Search for Files or
Folders...
4. In the Search Results window, select All files and folders, type service in the
All or part of the file name: edit box, and then click on the Search button.

3-15
Explorer Technical Manual

5. The software now searches for SERVICE. A list in the right panel will show you
the files found. In the list, click the “service” item shown in C:\QDR\Utilities
and drag it to the desktop. Close the Search window and then double-click the
Service icon on the desktop. The software will now be restarted in service mode.

Check Tube KV Peak Potential


Proper operation of the Explorer requires that the X-ray tube generate X-ray pulses of
80kVp, 100kVp, 120kVp, and 140kVp, all ±10%. The peak potential check must be
performed at installation time and whenever the X-ray source or X-ray controller is
repaired or replaced. Because it would be very dangerous to directly monitor the kVp
potentials, there is a 10,000 to 1 voltage divider circuit inside the High Voltage Power
Supply/Source (HVPS/S). By monitoring this divided voltage, one can determine the peak
potentials being impressed on the X-ray Tube. The monitoring can be done on the terminal
strip on top of the X-ray Source.

Warning: Although the test voltage is low, there are elevated voltages near and around
the test points.

1. Remove the tank cover (2 Phillips screws).


Set up the scope as follows:
Channel 1:2V/div (0.2V/div if using x10 probes)
Channel 2:2V/div (0.2V/div if using x10 probes)
Time base:5.0ms/div
Trigger on line (positive slope)
Set Vertical Mode to Add
Invert Channel 2.
Connect Channel 1 to TB1-PIN 12 (on the tank)
Connect Channel 2 to TB1-PIN 11 (on the tank)
2. Ground both channels and move the trace to the bottom of the screen. Remove
the grounds and go to DC coupling.
3. Instrument power, computer power, and the X-ray enable key should all be on.
4. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey see “X-Ray Survey” on page 9-1 for information on the utility).
5. Set the X-ray Mode to 4.
6. Click Xrays (F2) to turn X-rays on.
7. Observe the oscilloscope. You should see a trace similar to Figure 3-11, approxi-
mately 4ms pulses with a peak amplitude of 8V. This corresponds to 80kVp
inside the tank.
8. Click Xrays (F2) to turn X-rays off

3-16
Explorer Technical Manual

9. Set Xray Mode to 3.

AW-00336_001-0318

Figure 3-11 Peak Potential Mode 4


10. Click Xrays (F2) to turn X-rays on.
11. Observe the oscilloscope. You should see a trace similar to Figure 3-12, alternat-
ing pulses, approximately 4ms in duration, with a peak amplitude of 14V and
10V respectively (corresponding to 140kVp and 100kVp inside the tank).

AW-00336_001-0319

Figure 3-12 Peak Potential Mode 3


If the scope trace seen is not as shown in either Figure 3-11 or Figure 3-12, the system
may have insufficient AC power, a faulty X-ray Controller, or defective Tank.
Note: Enter this information into The Radiation Measurement Report (see page 3-26).

3-17
Explorer Technical Manual

Check Tube Current


Follow the procedure below to check X-ray tube current. Monitoring tube current is done
on the barrier strip on top of the X-ray source..

Warning: Although the test voltage is low, there are elevated voltages near and around
the test points.

1. If the tank cover is not off, remove it (2 Phillips screws).


2. Set up the scope as follows:
Channel 1:2V/div (0.2V/div if using x10 probes)
Channel 2:2V/div (0.2V/div if using x10 probes)
Time base:5.0ms/div
Trigger on Line (positive slope)
Set Vertical Mode to Add
Invert Channel 2.
Connect Channel 1 to TB1-PIN 10 (on the tank)
Connect Channel 2 to TB1-PIN 7 (on the tank)
3. Ground both channels and move the trace to the bottom of the screen. Remove
the grounds and go to DC Coupling.
4. Instrument power, computer power, and the X-ray enable key should all be on.
5. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
6. Set Xray Mode to 1.
7. Click Xrays (F2) to turn X-rays on.
8. Observe the oscilloscope. You should see a trace similar to Figure 3-13, approxi-
mately 4ms pulses with a peak amplitude of 3V (ignore the overshoot and mea-
sure current on the back, after it settles out). This corresponds to 3ma ±35% tube
current.
9. Click Xrays (F2) to turn X-rays off.
Note: Enter this information into The Radiation Measurement Report (see page 3-26).

3-18
Explorer Technical Manual

10. Set Xray Mode to 3.

AW-00336_001-0320

Figure 3-13 Tube Current Mode 1


11. Click Xrays (F2) to turn X-rays on.
12. Observe the oscilloscope to see a trace similar to Figure 3-14, approximately 4ms
pulses with a peak amplitude of 10V (ignore the overshoot and measure current
on the back, after it settles out). This corresponds to 10mA ±35% tube current.
If the scope trace seen is not as shown in either Figure 3-13 or Figure 3-14, the system
may have insufficient AC power, a faulty X-ray Controller, or a defective Tank.

AW-00336_001-0321

Figure 3-14 Tube Current Mode 3


Note: Enter this information into The Radiation Measurement Report (see page 3-26).

3-19
Explorer Technical Manual

Check Belt Tension


Check for proper tension on each of the motor drive belts. If any belt is loose, perform the
tension adjustment for that belt as described in Section 5, Remove and Replace Procedure
of this manual.

Adjust C-Arm Y Belt


The C-Arm Y-belt should always be adjusted during installation. To do this, perform the
following:
1. Move the C-Arm to the center of the table.
Caution: Do not move the C-arm or the table more than 1"/second with the power off.

2. Turn off the Explorer computer power and main circuit breaker on the rear of the
Torroid Assembly.
3. Remove the Electronics Tray covers.
4. Loosen the two mounting bolts holding the belt tension block.
5. Adjust the tension nut so that the spring is compressed to 7/8 inch.
6. Tighten the two mounting bolts holding the tension block.

Calibrate Motors
The SQDRIVER program provides a CALIBRATE command for each of the motors (AY,
TY and TX) to calibrate the encoder readback and determine the limits of motion. These
values are loaded into the SQDRIVER.INI file.
Perform the Motor Calibration procedure in the ALIGNMENT AND CALIBRATION
section of this manual.
Note: During installation, the MOTOR$AY, MOTOR$TX and MOTOR$TY procedures
may be performed without unclamping the drive shaft tubing (this allows the
procedure to be completed more quickly). This applies only during installation.

Check X-Ray Beam Alignment


It is crucial that the X-ray beam be precisely aligned with the detector as improper
alignment directly affects the repeatability (coefficient of variability, or CV) of the
Explorer.
To check beam alignment, perform the X-Ray Beam Alignment procedure in the
ALIGNMENT AND CALIBRATION section of this manual.

3-20
Explorer Technical Manual

Check Laser Positioning Offset


Center the table and arm using Center Table button on the Control Panel or the Utilities|
AP Reposition feature in the software.
1. Turn on the laser and set the point of a sharp metal object on the crosshair.
2. At the Explorer Main Menu screen, click the Perform Exam button, create a
test patient biography, and select AP Lumbar as the scan type.
3. Start the scan.
4. After scan starts and you can see the object, click the Reposition Scan button.
5. Using the mouse, move the image so that the tip of the object is horizontally in
the center and touching the bottom edge of the scan area.
6. Click the Restart Scan button and let the arm reposition.
7. At the Start Scan screen, click the Cancel button.
8. Turn on the laser.
9. Locate the 3 laser adjustment screws (small Phillips) under the C-arm. Adjust
these screws to tilt the assembly until the laser crosshair is on the tip of the
pointed object.
10. Run another scan to check your adjustments.

Adjust A/D Gain Control


In order that all Explorer machines have the same input to the A/D converter, regardless of
slight variations in the X-ray flux detected, a potentiometer has been installed on the unit’s
A/D converter board. Perform the A/D Gain Control Adjustment procedure in the
ALIGNMENT AND CALIBRATION section of this manual. This procedure must be
followed by the Detector Flattening Procedure in order for the adjustment to take effect.

Perform Detector Flattening


This procedure flattens the X-ray beam for each scan mode. Perform the Detector
Flattening procedure in the ALIGNMENT AND CALIBRATION section of this manual.
Note: All covers, table mats, etc. in the X-ray path must be on the Scanner before running
detector flattening.

Measure X-Ray Dose to Patient


A radiation dose measurement must be performed at installation time, and should be done
routinely whenever the machine is serviced. This test requires the use of a Victoreen
Model 450P or equivalent measuring device set to measure dose over time (mR/Hr).
Note: The dose measurements must be done with the machine fully assembled and the
table mat in place.

3-21
Explorer Technical Manual

1. Place the meter in the center of the table approximately ½ inch to the right of the
laser crosshair.
2. Click the Perform Exam button, create a test patient biography, and select the
Array spine scan.
3. Set scan length for 9.5 inches and click the Start scan button.
4. The meter is now scanned.
5. Observe the scan to verify that the entire chamber of the meter is included in the
scan.
6. Record the result from the meter.
Note: The array spine scan dose should be less than 200µGy (20mrad).
7. Record the highest reading in the service report.
Note: Enter this information into The Radiation Measurement Report (see page 3-
26).

Check HVPS/S (Tank) For Radiation Leakage


For regulatory and safety reasons, the High Voltage Power Supply/Source (HVPS/S) must
be checked for radiation leakage at the time of installation or whenever the HVPS/S is
repaired or replaced.
Principles
Radiation leakage from the HVPS/S results from a defect in the lead liner of the tank or its
cover. If there is a hole in the lead liner, the resulting leakage will produce a pencil beam
of X-rays. If there is a crack in the lead liner, the resulting leakage will produce a fan
beam. In either case, the initial survey must be taken close to and almost touching the tank
with the survey meter probe. If a leak is found, the meter indicates a spike in the reading.
If this occurs, you must then move the probe one-meter (approx. 40”) along the beam from
that spot to determine if the leakage is within acceptable levels.
The Victoreen 450P has a response time of:
Range Time
0-4mR/h 3.3 seconds
0-40mR/h 4.5 seconds

3-22
Explorer Technical Manual

Performing the Leakage Test Procedure


1. Remove the tank cover and set the X-ray leakage test shield (099-0566) over the
aperture slot.

KP1085_003-0322

Figure 3-15 Leaking Test Shield (069-0566)


2. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
3. Set Xray mode to 3.
Initial Leakage Survey
4. Click Xrays (F2) to turn X-rays on.
5. Slowly move the meter probe (Victoreen Model 450P or equivalent) about all
accessible surfaces (see table above for meter response times) to detect any leaks.
6. If a leak is detected (a spike in the reading), measure a point at a one meter
(approx. 40”) distance from the leak and insure that the reading is 10µGy/h
(1.0mrad/h) or less.
Final Leakage Survey
7. Position the survey meter one meter (approx. 40”) from the tank and move it all
around the outside of the instrument.
8. The reading must be 10µGy/h (1.0mrad/h) or less. Record the highest reading in
the service report.
9. Click Xrays (F2) to turn X-rays off.
10. Remove the tool, and replace the tank cover.
Note: Enter the information into The Radiation Measurement Report (see page 3-
26).

Calibrate For Area, BMD And BMC


The final calibration procedure is to calibrate for Area, BMD, and BMC. Perform the
Area, BMD and BMC Calibration procedure in the ALIGNMENT AND CALIBRATION
section of this manual.

3-23
Explorer Technical Manual

Test Scan Modes


Perform at least one scan in each of the scan modes to verify that machine is fully
functional in all modes of operation. Verify that there are no electrical or mechanical
problems during each scan mode.
[ ] Perform a Detail, Survey and an Explorer scan.
[ ] Perform a Survey left hip, Explorer left hip and an Explorer right hip scan.
[ ] Perform a whole body scan (if option is installed).
[ ] Perform a forearm scan (if option is installed).
[ ] Configure and test DICOM or networking options (if option is installed).

Finish Assembling Unit


Dress all cables at the rear of the console with nylon tie wraps. Take the time to replace
any covers or enclosures that may have been previously removed. Check that all of the
machine's safety features (ground wires, limit switches, emergency stop, etc.) are in place
and working properly. Finally, clean the machine and remove any unwanted packing
materials from the room.

Measure X-Ray Scatter From Phantom


A radiation scatter measurement must be performed at the time of installation. You must
use a survey-type radiation meter (Victoreen 450P or equivalent).
Perform the following:
1. Center the table and C-arm
2. Place the spine phantom in the center of the table and align using the laser.
3. Perform the following:
4. Using the Esc key, stop the scan when it is between the two middle vertebrae.
5. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
6. Click Xrays (F2) to turn X-rays on.
7. Using the survey meter, measure the scatter radiation at a distance of 2 meters
from the phantom on a horizontal plane all around the unit. Move the meter
slowly (refer to the Victoreen response table on page 3-23).
Note: If you cannot measure the scatter radiation at a distance of 2 meters
because of space restrictions, you can measure at 1 meter and divide the
result by 4. This reading must be less than 10µGy/h (1.0mrad/h) at 2
meters (approx. 80 in.) in array AP scan mode.
8. Record the highest reading in the service report.

3-24
Explorer Technical Manual

Note: Enter this information into The Radiation Measurement Report (see page
3-26).
9. Click Xrays (F2) to turn X-rays off.

Perform QC
Once the machine has been fully assembled and calibrated, at least one QC scan should be
performed.
• Perform the daily QC procedure and enter the scan into the QC database.

Run Reproducibility Test


Reproducibility is the ability of the scanner unit to perform consecutive scans while
keeping the amount of radiation for each scan consistent. Instrument reproducibility is
tested by running 10 consecutive scans of the spine phantom shipped with the unit. You
must then plot the results of the High Air and Low Air measurements for these scans and
insure that the SD of the detected signal is less than 3.
To do this:
1. Perform 10 array spine scans (using Utilities|Service Utilities|Autoscan
from the Explorer Main Menu screen).
2. Analyze the first scan manually.
3. Analyze the remaining scans (using Utilities|Service Utilities|Auto Ana-
lyze).
4. Put the results in the QC database (if you have not already done so).
5. Restart the Explorer software in service mode (if not already there). From the
Explorer Main Menu, select QC|QC Data Management|Plot. Use a start and
end date that will only select the scans that you have just completed.
6. Select QC Parameter to Plot, HiA, and Plot.
7. Plot the data for the 10 scans in step 1.
8. Assure that the SD is 3 or less.
9. Repeat for “LoA” by repeating steps 4-7. This result must also be 3 or less.
Note: Enter this information into The Radiation Measurement Report (see page 3-26).

Table Top Radiographic Uniformity


This procedure verifies proper X-Ray beam alignment and the uniformity of the X-Ray
transparency of entire tabletop. Follow the procedure outlined in the Alignment and
Calibration section of this manual to perform this scan.

3-25
Explorer Technical Manual

Note: A copy of the scan on a floppy disk and the printout must be included along with the
other information and printouts sent to Hologic after installing a new instrument or
after performing Preventive Maintenance visit.

The Radiation Measurement Report


After installation, the field engineer must fill out the Radiation Measurement Report and
keep this information on file. A sample of this report follows:

3-26
Explorer Technical Manual

Section 4
ALIGNMENT AND CALIBRATION
Table Alignment
Checking Table Alignment
To check the table alignment, perform the following procedure:
1. Using a measuring tape, and referring to Figure 4-1, check the following:
• Distance from the edge of the table to the back of both T-rails (“A” dimension).
• Distance between the T-rails (front and rear).
• Gap from the edge of the table bracket (left side) to the rail.
2. Record all the measurements.
3. Facing the front of the Explorer, gently push the foot end (left side) of the table. The
table should move away from, and then back, to its original position.
4. Check the “A” dimension and the bracket-to-guide rail gap again. Compare them to their
original values.
5. If all the measurements are within specification, the table is properly aligned. If the mea-
surements are not within specification, go to the Aligning Table section.

Aligning Table
After you have taken the measurements in the Checking Table Alignment section, use the
procedures below to align the table. Note that if both the “A” dimension and the bracket-to-guide
rail gap are out of specification, you should recheck the measurements after performing the first
adjustment.

Table Edge to T-Rail (“A” Dimension) Adjustment


To change the “A” dimension, do the following:
1. Remove the outer and inner covers from the right pedestal.
2. Loosen the four bolts fastening the top of the pedestal and the four bolts fastening the
bottom of the pedestal (see Figure 4-2).
3. Adjust the table so the “A” dimension is within the specification.
4. Tighten the bolts and check the table alignment again.
If the alignment is within the specification, replace the pedestal covers. If you still note a change
in the “A” dimension, continue with the following steps.

4-1
Explorer Technical Manual

63.125” +/- .06”


FRONT & REAR

“A” “A” +/- .25”

.25” +/- .09”

KP1085_003-0401

Figure 4-1 Table Alignment

(Not drawn to scale)

Studs and
T-Rail Adjustment Nuts
Upper Bracket
Pedestal
Lower Bracket

Base Adjustment Bolts

KP1085_003-0402

Figure 4-2 Pedestal (covers removed)


5. Make sure the upper and lower brackets are securely fastened to the upper and
lower frames.
Note: Even if the brackets are secured to the frames, they may not be securely
fastened to the pedestal. If not, remove the table to access the screws that

4-2
Explorer Technical Manual

secure the brackets to the pedestal. Refer to Section 3, Installation, for


table removal procedures.
If the screws are loose, apply a small amount of Loctite to the threads and
tighten the screws.
The upper and lower brackets are attached to the pedestal with four, 6mm flat-
head Allen screws. While the table is off, it is a good idea to remove the pedes-
tal to make sure that these screws are tight as well.
To change the “A” dimension, do the following:
6. Mount the pedestal to the lower frame. Do not tighten the bolts until the
alignment has been completed.
7. Install the table and check its alignment. Make the necessary adjustments, then
tighten the upper and lower bolts.
8. Install the pedestal covers that were removed in Step 1.

Front to Back T-Rail and Table Edge/Rail Gap Adjustment


To adjust the front-to-back T-Rail dimensions and table edge-to-rail gap, perform the
following procedure:
1. Remove the outer and inner covers from the left pedestal.
2. Loosen the four bolts fastening the top of the pedestal and the four bolts fastening
the bottom of the pedestal.
Note: Before adjusting the distance between the rails, make sure the upper and
lower brackets are securely fastened to the pedestal. Refer to Steps 3-5 of
the previous section (Table Edge to T-Rail Adjustment) then go to the next
step.
3. Adjust the distance between the T-rails and the table edge-to-rail gap and tighten
the bolts.
4. Install the pedestal covers removed in Step 1.

X-Ray Beam Alignment


It is crucial that the X-ray beam be precisely aligned with the detector array because
improper alignment will directly affect the repeatability (coefficient of variation, or CV)
of the Explorer. Therefore, this alignment must be verified at the time of installation or
whenever any work is performed that may affect it.
To check beam alignment:

Warning: X-rays are being generated during this procedure. Keep hands, head and
other body parts out of beam.

Insert the alignment fixture (see the following three figures) into the detector opening.

4-3
Explorer Technical Manual

Alignment Pins

Ho;ding Clamps
KP1085_003-0403

Figure 4-3 X-Ray Alignment Fixture (010-0923)


1. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
2. Set Hi Gain to 2, Lo Gain of 1, and Xray mode to 3.
3. From the Explorer Main Menu screen, select Utilities|Service Utili-
ties|SQDRIVER.
4. Type exit<Enter> to return to the Explorer Main Screen.
5. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
6. Confirm that the Hi Gain is set to 2, Lo Gain is still 1, and Xray mode is set to
3.
7. For Display click Graph to set the Main Display to bar graph.
8. Click Xray (F2) to turn on x-rays.
The correct display should show a flat graph with an amplitude of approximately six and a
half (6.5V) volts. If the X-Ray beam alignment is not correct as indicated by peaks in the
graph, perform the following procedure.

4-4
Explorer Technical Manual

Detector Opening C-Arm

A C
Alignment Pin
Holding Clamp

Fixture
B, D
View from Front

Step A. Insert the left side of the Alignment Fixure into the left side of the
Detector Opening so that the vertical edge of the four Alignment
Pins are secure.
Step B. Compress the Holding Clamps.
Step C. Raise the right side of the Alignment Fixture into the Dectector
Opening.
Step D. Release the Holding Clamps.
Note:. If the Alignment Fixture is insertrd the the Alignment Pins on the
right, the procedure works equally well.

KP1085_003-0404

Figure 4-4 Inserting the X-Ray Alignment Fixture

KP1085_003-0405

Figure 4-5 The Alignment Fixture Properly Installed


1. Set the machine in the Center Table position.

4-5
Explorer Technical Manual

KP1085_003-0406

Figure 4-6 Filter Drum Adjustment - Top View


2. At the Filter Drum assembly, loosen the jam nuts and insert Allen wrenches (3/
32") in both Filter Drum Allen alignment screws. (Figure 4-6 above shows the
locations of alignment screws and jam nuts.) Ensure that the Filter Drum is
running.
3. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
4. Set Hi Gain to 2, the Lo Gain to 1,and Xray Mode to 3.
5. For Display click Graph to set the Main Display to bar graph.
6. Set Pulse to 1.
7. For Segment click Air.
8. Click Xray (F2) to turn X-rays on.
9. View the X-ray graph on the Main Display.
The next 5 steps adjust the beam side to side.

Warning: The X-rays are on. Keep body parts out of the beam.

1. Move the front Filter Drum alignment screw until the X-ray signal reaches max-
imum voltage.
Note: The last direction turned should be clockwise to eliminate backlash when
the jam nuts are tightened. The plot should show peak amplitude.
2. Move the rear Filter Drum alignment screw until the X-ray signal reaches maxi-
mum voltage.

4-6
Explorer Technical Manual

Note: The last direction turned should be clockwise to eliminate backlash when
the jam nuts are tightened. The plot should show peak amplitude.
3. Tighten the jam nuts on both Filter Drum alignment screws.
Note: The X-rays should still show peak amplitude.
4. Click Xray (F2) to turn X-rays off.
5. Remove the alignment test fixture.

Figure 4-7 Array Assembly - Top View, Partial


The next six steps adjust the beam front to back.
1. At the Array assembly, loosen the four X-alignment nuts (see Figure 4-7 above).
2. Click Xray (F2) to turn X-rays on.
3. Detectors 54 through 160 will be turned on.
4. Set Hi gain to 2, Lo Gain tp 1, and Xray mode to 3.
5. For Display click Graph to switch to the bar graph on the Main Display.
6. Click Xray (F2) to turn X-rays on.
7. Follow the procedure in steps 3 and 4 while checking detectors number 54 and
160 to check for the fall off of x-rays.
8. Tighten the four array X-alignment nuts.
9. Click Xray (F2) to turn X-rays off.
10. Press Close to exit X-Ray Survey.

4-7
Explorer Technical Manual

Motor Calibration
The SQDRIVER program provides a CALIBRATE command for each of the motors (AY,
TY and TX) to calibrate the encoder read back and determine the limits of motion.
Each motor requires the corresponding protocol calibration file in the PROTOCOL sub-
directory (e.g., for MOTOR$AY, the calibration protocol is MOTOR_AY.PRO).
To perform the calibration procedure:
1. From the Explorer Main Menu screen, select Utilities|Service Utili-
ties|SQDRIVER (you must be in Service Mode).
2. At the CARM$$$$> prompt, type MOTOR$XX<Enter>, where XX equals AY,
TY, or TX depending on which motor you are calibrating.
Note: If you are calibrating motors during an installation, you must calibrate them in the
proper order. The proper order is AY, TY, andTX.
Perform the calibration procedures if indicated (*) in order from left to right.

MOTOR$AY (Final Value = 3750)


If you have replace the encoder, before beginning this procedure, make sure that the AY
motor encoder is not clamped to the drive shaft tubing. It is unnecessary to unclamp the
tubing during an installation.
To perform the calibration procedure:
1. Select the AY motor device driver by typing: MOTOR$AY<Enter>
2. At the MOTOR$AY> prompt in SQDRIVER, type: CALIBRATE<Enter>
The program prompts:
Press <Enter> when the AY motor reaches the LEFT mechanical limit.
Press <ESC> to stop calibration.

The program moves the AY motor to the left. When AY hits the left mechanical stop
the first time:
3. (For encoder repairs only) Press <Esc>.
The program returns to the MOTOR$AY menu page (below).
In the right hand column of the display, eight lines from the bottom (and highlighted
in the figure below), is the raw a/d readback of the motor encoder (in the example
below, it is 2048). Manually rotate the encoder until the readback is 3750±50 (i.e., in
the range 3700-3800) and then clamp the encoder to the drive shaft.
4. (For encoder repairs only) Repeat the calibration procedure above but now, when
AY hits the left mechanical stop, press <Enter>.

4-8
Explorer Technical Manual

The program then starts AY moving to the right and prompts:


Press <Enter> when the AY motor reaches the RIGHT mechanical limit.
Press <ESC> to stop calibration.

5. When AY hits the right mechanical stop, press <Enter>.


The program then switches to graphics mode and draws the Encoder Vs Distance
calibration grid. It steps the AY motor to the left in one inch increments until the
motor hits the left mechanical stop, and then steps the AY motor to the right in
one inch increments until the motor hits the right mechanical stop. The program
plots the encoder data during the scan (in raw encoder units, 0-4095).
6. Press <Esc> anytime during the scan to terminate the calibration procedure.

MotionState NONE OnLine 1


DeviceState E_OK EmergencyStop 0
MotionError E_OK InterlockInhibit 0
PosMotionSwitch 0 CollisionImminent 0
NegMotionSwitch 0 LocalMotionEnable 0
PosLimitSwitch NONE C/C Version 0.00
NegLimitSwitch NONE # C/C Interrupts 5678(18)
Position 363812 # DOS Interrupts 878
MoveState DONE StepStatus 0
FirstRate 4000 AccelType 1
FinalRate 30000 Direction 1
HoldPower LOW HoldPower 1
StepPower MEDIUM StepPower 2
AccelDistance 3006 InputSwitches 3
AccelType 1 Position 2048
MinMotion 0 NegLimitPosition 900
MotionDetect 0 PosLimitPosition 3992
PosLimitPosition 936949 FirstRate 222
NegLimitPosition 25355 FinalRate 1667
PositionAverage 1 AccelSteps 167
DriverVersion 3.41 NumberOfSteps 50000
DeviceVersion 2.03 RateScaleFactor 1
G-11110 G
MOTOR$AY>

When the calibration scan completes, the program computes the linear fit for both the
positive and negative motion. The linear fit parameters are displayed at the top left
and top right of the plot in the form Y = Intercept + Slope * X. The two slopes (e.g.,
86.6 and 86.5) should be within 0.3 of each other.

4-9
Explorer Technical Manual

E
N +
+ ++++
C ++ +
O
+ ++++
D ++ +
E +
++
+++
R ++
+
+
++
++
R +
+++
E ++++
+++ +
A
D + +++
+
+ + +
I +++
N +++
++
G ++
++++
++
+
+
++
++
++
++

KP1085_003-0408

The Program displays the positive and negative limits as horizontal dashed limits.
Note: The PosLimitOffset and NegLimitOffset entries in the [AyMotor] section of the
SQDRIVER.INI file determine the motion limits relative to the mechanical stops. If
these entries are not present, or are zero, the motion limits are set to the mechanical
limits).
The program then changes the plot title to PRESS ANY KEY TO CONTINUE. Press the
<Enter> key and the program prompts:
motor_direction=1
calibrate_position=1,2288,41187,209,61339,363,363,3719
pos_limit_position=984946
neg_limit_position=0
Update Driver INI-File [Y/N]?

The motor_direction field determines whether the positive step direction is the reverse of
the direction of increasing encoder values ('0' if the direction is not reversed, '1' if it is).
This value is set from the corresponding parameter in the MOTOR_AY.PRO file and is a
constant for each motor (i.e., the direction of increasing encoder values must agree with
the system coordinates and the positive step direction is set by the wiring harness for the
stepper motor).
The eight calibrate_position fields are:

4-10
Explorer Technical Manual

1. 1 (Tolerance). The ± position tolerance, in encoder ticks, for absolute moves.


The calibration program sets this field to the value found in the corresponding
calibrate_position field in the MOTOR_AY.PRO file.
2,3) 2288,41187 (NumberOfSteps,StepDistance). The number of motors steps
corresponding to the step distance in microns. The ratio of these two numbers
determines the step size (41187/2288 ≅ 18 microns). The calibration program
sets these fields to the values found in the corresponding calibrate_position
fields in the MOTOR_AY.PRO file. Since these values are a property of the
mechanical design of the system, they should never change.
4,5) 209,61339 (NumberOfTicks,TickDistance). The number of encoder ticks
corresponding to the tick distance in microns. Again, it is the ratio of these two
numbers (61339/209 ≅ 293 microns) that determines the encoder calibration.
These numbers will generally change from calibration to calibration (although
the ratio should remain approximately the same).
6) 363 (EncoderOffset). The value subtracted from the encoder readback to set the
origin of motion. The origin of AY motion is the extreme right mechanical stop,
so this value should be the same as the first field below.
7,8) 363,3719 (NegLimit,PosLimit). The encoder readings for the negative (right)
and positive (left) mechanical stops.
The pos_limit_position and the neg_limit_position are the limits of motion, in
microns, in the positive and negative direction.
Note: The PosLimitOffset and NegLimitOffset entries in the [AyMotor] section of the
SQDRIVER.INI file determine the motion limits relative to the mechanical
stops. If these entries are not present, or are zero, the motion limits are set to
the mechanical limits, as in the above example).
Type Y<Enter> to accept the calibration values or N<Enter> to reject them.
Note: The last calibration scan data is saved in the file MOTOR_AY.DAT. You can
reanalyze the data–e.g., after editing SQDRIVER.INI by typing the command
CALIBRATE @MOTOR_AY.DAT at the MOTOR$AY> prompt in
SQDRIVER.

MOTOR$TY (Final Value = 3750)


The TY motor exists in the system but is operational only if the Whole Body option is
installed.
If the TY Encoder has been replaced, before beginning this procedure, ensure that the
TY motor encoder is not clamped to the drive shaft tubing. If you are performing the
initial installation of the instrument, this is unnecessary.
1. Select the TY motor device driver by typing: MOTOR$TY<Enter>
2. At the MOTOR$TZ> prompt in SQDRIVER, type: CALIBRATE<Enter>

4-11
Explorer Technical Manual

The program starts TY moving to the left and prompts:


Press <Enter> when the TY motor reaches the LEFT mechanical limit.
Press <ESC> to stop calibration.

The program moves the TY motor to the left.


3. (For encoder repairs only) When TY hits the left mechanical stop the first time,
press <Esc>.
The program returns to the MOTOR$TY menu page (below).
In the right hand column of the display, eight lines from the bottom (and highlighted
in the figure below), is the raw a/d readback of the motor encoder (in the example
below, it is 2048). Manually rotate the encoder until the readback is 3750±50 (i.e., in
the range 3700-3800) and then clamp the encoder to the drive shaft, making sure not
to over stress the clamp.

MotionState NONE OnLine 1


DeviceState E_OK EmergencyStop 0
MotionError E_OK InterlockInhibit 0
PosMotionSwitch 0 CollisionImminent 0
NegMotionSwitch 0 LocalMotionEnable 0
PosLimitSwitch NONE C/C Version 0.00
NegLimitSwitch NONE # C/C Interrupts 2088(1)
Position 462579 # DOS Interrupts 336
MoveState DONE StepStatus 0
FirstRate 4000 AccelType 1
FinalRate 40000 Direction 1
HoldPower LOW HoldPower 1
StepPower MEDIUM StepPower 2
AccelDistance 3996 InputSwitches 3
AccelType 1 Position 2048
MinMotion 0 NegLimitPosition 585
MotionDetect 0 PosLimitPosition 3415
PosLimitPosition 871071 FirstRate 333
NegLimitPosition 25400 FinalRate 3333
PositionAverage 1 AccelSteps 333
DriverVersion 3.41 NumberOfSteps 50000
DeviceVersion 2.03 RateScaleFactor 1
T1 T
MOTOR$TY>

4. (For encoder repairs only) Repeat the calibration procedure above but now, when
TY hits the left mechanical stop, press <Enter>.

4-12
Explorer Technical Manual

The program then starts TY moving to the right and prompts:


Press <Enter> when the TY motor reaches the LEFT mechanical limit.
Press <ESC> to stop calibration.
When TY hits the right mechanical stop, press <Enter>. The program then switches
to graphics mode and draws the Encoder Vs Distance calibration grid. It steps the TY
motor to the left in one inch increments until the motor hits the left mechanical stop,
and then steps the TY motor to the right in one inch increments until the motor hits
the right mechanical stop. The program plots the encoder data during the scan (in raw
encoder units, 0-4095).
5. Press <Esc> anytime during the scan to terminate the calibration procedure.

E
++ +
N +++ ++
C + + +
O ++ ++
+ + +
D ++
E ++++
R +
+ +
++
++ ++
R + + +
++ +
E
++ ++
A +
D ++
+
+
I + ++
++
N +++
G +++
+
+ + +
+++
+++
+
+++ +
+++

KP1085_003-0409

When the calibration scan completes, the program computes the linear fit for both the
positive and negative motion. The linear fit parameters are displayed at the top of left
and top right of the plot in the form Y = Intercept + Slope * X. The two slopes (e.g.,
86.5 and 86.5) should be within 0.3 of each other.
The program displays the positive and negative limits as horizontal dashed lines.
The PosLimitOffset and NegLimitOffset entries in the [TyMotor] section of the
SQDRIVER.INI file determine the motion limits relative to the mechanical stops. If
these entries are not present, or are zero, the motion limits are set to the mechanical
limits.
The program then changes the plot title to PRESS ANY KEY TO CONTINUE.

4-13
Explorer Technical Manual

6. Press the <Enter> key and the program prompts


motor_direction=0
calibrate_position=1,2287,27446,154,45219,238,238,3742
pos_limit_position=1028879
neg_limit_position=0
Update Driver INI-File [Y/N]?

The motor_direction, calibrate_position, pos_limit_position and


neg_limit_position fields have the same interpretation as discussed under
MOTOR$AY.
7. Type Y<Enter> to accept the calibration values or N<Enter> to reject them.
Note: The last calibration scan data is saved in the file MOTOR_TY.DAT. You can
reanalyze the data–e.g., after editing SQDRIVER.INI–by typing the command
CALIBRATE @MOTOR_TY.DAT at the MOTOR$TY> prompt in
SQDRIVER.

MOTOR$TX (Final Value = 1000)


If you have replaced the encoder, before beginning this procedure, make sure that the
TX motor encoder is not secured to the drive shaft and that the tank assembly is not
under the operator console. For normal installations, this is unnecessary.
1. Select the TX motor device driver by typing: MOTOR$TX<Enter>
2. At the MOTOR$TX> prompt in SQDRIVER, type: CALIBRATE<Enter>
The program prompts:
Press <ESC> to stop calibration.

The program moves the TY (W only) and AY motors to their center positions and
prompts:
Use the Table IN / OUT switch to move the C-Arm to 0 degrees.
Press <Enter> when the C-Arm is positioned.
Press <ESC> to stop calibration.

The above message is always displayed when calibrating MOTOR$TX. This is the
normal message for a QDR-4500. The message was not changed for Explorer
systems.
3. Press <Enter> and the program prompts:
Press <Enter> when the TX motor reaches the OUTER mechanical limit.
Press <ESC> to stop calibration.

4-14
Explorer Technical Manual

When TX hits the outer mechanical stop the first time:


4. (For encoder repairs only) Press <Esc>
The program returns to the MOTOR$TX menu page (below). In the right hand
column of the display, eight lines from the bottom (and highlighted in the figure
below), is the raw a/d readback of the motor encoder (in the example below, it is
2048). Manually rotate the encoder until the readback is 1000±100 (i.e., in the range
900-1100) and then secure the encoder to the drive shaft.
Repeat the calibration procedure above but now, when TX hits the outer mechanical
stop,
5. Press <Enter>.
The program switches to graphics mode and draws the Encoder Vs Distance
calibration grid. It steps the TX motor in by 1" increments until the inner mechanical
stop is hit. The program plots the encoder data during the scan (in raw encoder units,
0-4095).
6. Press <Esc> at anytime during the scan to terminate the calibration procedure.

MotionState NONE OnLine 1


DeviceState E_OK EmergencyStop 0
MotionError E_OK InterlockInhibit 0
PosMotionSwitch 0 CollisionImminent 0
NegMotionSwitch 0 LocalMotionEnable 0
PosLimitSwitch NONE C/C Version 0.00
NegLimitSwitch NONE # C/C Interrupts 1133(0)
Position 298545 # DOS Interrupts 191
MoveState DONE StepStatus 0
FirstRate 3600 AccelType 1
FinalRate 36003 Direction 1
HoldPower LOW HoldPower 1
StepPower MEDIUM StepPower 2
AccelDistance 2004 InputSwitches 3
AccelType 1 Position 2048
MinMotion 0 NegLimitPosition 1026
MotionDetect 0 PosLimitPosition 3096
PosLimitPosition 604684 FirstRate 300
NegLimitPosition 0 FinalRate 3000
PositionAverage 1 AccelSteps 167
DriverVersion 3.41 NumberOfSteps 50000
DeviceVersion 2.03 RateScaleFactor 1
T1 T
MOTOR$TX>

4-15
Explorer Technical Manual

E
N
C
O
D +++
E +++ +
R +++ +
++ +
+++ ++
R
++ +
E
+++ ++
A ++ +
D + ++
I +++ +
+++
N +++
+++ +
G +

KP1085_003-0410

When the calibration completes, the program computes the linear fit for both the
positive and negative motion. The linear fit parameters are displayed at the top left
and top right of the plot in the form Y = Intercept + Slope * X. The two slopes (87.0
and 86.9 in the example below) should be within 0.3 of each other.
The program displays the positive and negative limits as horizontal dashed lines.
Note: The PosLimitOffset and NegLimitOffset entries in the [TxMotor] section of the
SQDRIVER.INI file determine the motion limits relative to the mechanical
stops. If these entries are not present, or are zero, the motion limits are set to
the mechanical limits.
The program then changes the plot title to PRESS ANY KEY TO CONTINUE.
Press the <Enter> key and the program prompts:
motor_direction=1
calibrate_position=1,2287,27446,43,12563,1026,1026,3096
pos_limit_position=604777
neg_limit_position=0
Update Driver INI-File [Y/N]?

The motor_direction, calibrate_position, pos_limit_position and


neg_limit_position fields have the same interpretation as discussed under
MOTOR$AY.
Type Y<Enter> to accept the calibration values or N<Enter> to reject them.

4-16
Explorer Technical Manual

Note: The last calibration scan data is saved in the file MOTOR_TX.DAT. You can
reanalyze the data–e.g., after editing SQDRIVER.INI–by typing the command
CALIBRATE @MOTOR_TX.DAT at the MOTOR$TX> prompt in
SQDRIVER.

Laser Positioning Offset Adjustment


The laser of the Explorer is used to correctly position the C-Arm over the patient.
Improper adjustment can cause improper detector flattening and may force the operator to
reposition the scan more frequently than necessary.
1. Center the table, turn on the laser, and set the point of a sharp metallic object on
the crosshair.
2. At the Main Menu screen, click the Perform Exam button, create a test patient
biography, and select AP Lumbar as the scan type. (DO NOT use the spine
phantom biography as a patient.)
3. Click the Start Scan button.
4. After scan starts and you can see the metallic object, click the Reposition
Scan button.
5. Using the mouse, move the image so that the tip of the metallic object is in the
horizontal center of the scan area and touching the bottom edge of the scan area.
6. Click the Restart Scan button and let the arm reposition.
7. At the Start Scan screen, click the Cancel button.
8. Turn on the laser.
9. Locate the 3 laser adjustment screws (small Phillips) under the C-arm. Adjust
these screws to tilt the assembly until the laser crosshair is on the tip of the
pointed object.
10. Run another scan to check your adjustments.

A/D Gain Control Adjustment


In order that all Explorer machines have the same input to the A/D converter, regardless of
slight variations in X-ray flux detected, a potentiometer has been installed on the unit’s
ADC board. All covers, that are normally in the X-ray beam, must be on. This
potentiometer is set as follows:

4-17
Explorer Technical Manual

1. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray


Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
2. Set Pulse to 1.
3. Set Hi Gain to 2, Lo Gain to 1, and Xray Mode to 3.
4. For Display click Graph to bring up bar graph on the Main Display.
5. For Segment click Air.
6. Click Xrays (F2) to turn X-rays on.
7. Observe the Main Display to ensure that the X-ray high energu (green) level is
approximately 6.25V. The graph amplitude should be essentially flat from front
to back.
The object of the adjustment is to have all the detectors fall within the 6.5V to
7.5V range. First adjust the average signal level to approximately 6.25V as indi-
cated on the graph and then check to see if all the detectors are now within the
6.5V and 7.5V range. If they are, the adjustment is complete. If they are not,
adjust the gain (in the required direction) until all the detectors are within the
5.5V and 7.5V range.
DO NOT ADJUST THE A/D GAIN UNLESS ABSOLUTELY NECESSARY.
Any adjustment of this potentiometer affects the QC highs and lows. You must run
detector flattening after adjusting the A/D gain.
If an adjustment is required, proceed as follows:
The object of the adjustment is to have all the detectors fall within the 6.5V to 7.5V
range. First adjust the average signal level to approximately 6.25V as indicated on the
graph and then check to see if all the detectors are now within the 6.5V and 7.5V
range. If they are, the adjustment is complete. If they are not, adjust the gain (in the
required direction) until all the detectors are within the 5.5V and 7.5V range.

Detector Flattening
Warning: X-rays are being generated during this procedure. Keep hands, head and
other body parts out of beam.

This procedure flattens the X-ray beam for each scan mode and finds the metal edge of the
table.
Note: All covers, table mat, etc., normally in the X-ray path, MUST be on the Scanner
before running Beam Flattening.
1. Restart the Explorer software in service mode (if not already). Press the Center
Table button and turn on the laser.
2. Place the phantom on end (vertical) with the spinal processes pointing towards
the head end of the table. The laser crosshair should be 1.5 inch in from the left
end and centered. (Some phantoms will have a target hole, if not, use a ruler).
3. Start the X-Ray Survey Utility by selecting Utilities|Service Utilities|X-Ray
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).

4-18
Explorer Technical Manual

4. For Pulse click 1.


5. Set Hi Gain to 2, Lo Gain to 1, and Xray Mode to 3.
6. For Display click Graph to bring up bar graph display on the Main Display.
7. Click Xrays (F2) to turn X-rays on.
8. Observe the Main Display.
9. Check that phantom covers the entire beam. This is critical. If phantom does not
cover the entire beam, move it until it does. Keep the phantom as straight as pos-
sible.
10. Click Xrays (F2) to turn X-rays off.
11. Click Close to exit X-Ray Survey.
12. Select Utilities|Service Utilities|Detector Flattening. Click the Continue
button. This procedure takes about 10 minutes.
13. When flattening is complete, run a QC scan and cancel adding it to the QC plot.

Table Top Radiographic Uniformity


This procedure verifies proper X-ray beam alignment:
1. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit in
the Patient ID field.
2. Perform one Whole Body air scan (a scan with nothing on the table). Select Exit
Exam when complete.
Make sure that only the pad is on the table. Clear the table of anything else.
3. From the Explorer Main Menu screen, select Utilities|Service Utilities|Table
Top Radiographic Uniformity.
4. Select the Whole Body scan acquired in step 2 and click the Next> button.
5. Table Top Radiographic Uniformity results tabs are now displayed. Select the
Low Air Tab.
6. Print the results and include it along with the Air scan on a floppy disk with your
installation/service report paperwork.
If the Global Stats S.D. (2) is less than 2.0, the instrument is properly aligned.
If the Global Stats S.D. is greater than or equal to 2.0, then the machine is not aligned
properly. Check the following:
• If the image appears “streaky” or “banded”, check for loose wires or other debris
between the table and the tank.
• Check AC Line voltage.
• Check +/- 15V Power Supply

4-19
Explorer Technical Manual

• Check “Detector Flattening” on page 4-18


• Re-run daily QC.
After realignment, repeat the Table Top Radiographic Uniformity test. If the global S.D. is
less than 2.0, you are done. If the global S.D. if equal or greater than 2.0, the non-
uniformity may be attributed to the table itself. If so, continue with this procedure. Check
to see if the machine has body composition loaded. If it does, then follow the appropriate
procedure below.

Instruments using Body Composition Analysis (BCA)


Instrument performing BCA must have a global S.D. of less than 2.0. If the realignment
and recalibration procedures above have been performed and an S.D. of less than 2.0
cannot be obtained, then the tabletop must be replaced.
Note: Archive the airscans you have acquired and either e-mail or FEDEX them to a
Technical Support Specialist at Hologic headquarters for final evaluation.

Instruments using BMD Whole Body Analysis


Instruments using only BMD Whole Body measurements can tolerate a larger S.D. If the
global S.D. is less than 3.0, then you are done. If the S.D. is equal or greater then 3.0, then
the tabletop must be replaced.
Note: Archive the airscans you have acquired and either e-mail or FEDEX them to a
Technical Support Specialist at Hologic headquarters for final evaluation.
Note: If the customer is upgrading to the Body Composition Analysis option, you must re-
evaluate the table global S.D. to less than 2.0.

Check Phantom values


1. From the Explorer Main Menu screen, select Utilities|Service Utilities|Field
Service Calibration.

KP1085_003-0412

2. Click the Number button.

4-20
Explorer Technical Manual

3. Compare values displayed with actual values printed on the phantom. If there is a
discrepancy, check to make sure you have the phantom that shipped with the unit.
Check with Hologic Technical Support if needed.

Area, BMD, And BMC Calibration


Calibration for Area, BMD, and BMC is accomplished in 3 stages:
a. Array Scan Thickness Measurement & Calibration
b. Calibration of Area and BMC for Array Scan Modes
c. Adding Array AP scans to the QC database
Follow the procedure, in order, and exactly as shown, for each stage of the calibration.
Note: To ensure stability, the instrument must be completely powered up for a minimum
of 30 minutes prior to running any of the following tests. Also, because the tabletop
pad will have a slight affect on the test results, all scans must be performed with it in
place.

Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.

Scan Thickness Measurement & Calibration


1. Press the Center Table button.
2. Turn on the laser.
3. With the laser on, place the phantom so that the laser dot is on the centerline of
the phantom, 1/2” from the left end, shining on the phantom target. The sides of
the phantom should be aligned parallel to the table, using a ruler if necessary.
4. Be sure the spine phantom has been entered in the patient database in the form of
“Spine Phantom #nnnn” (where nnnn is the number of the phantom) so that the
QC plot program can identify it properly.
5. From the Explorer Main Menu screen, select Perform Exam, select the Spine
Phantom, and click the OK button at Patient Information dialog box.
6. Select AP Lumbar Spine, uncheck default scan checkbox, and click the Next>
button.
7. Select Array, click the Next> button, set the scan length to 6 inches, and start
the scan.
8. Analyze the scan and click the Close button to exit the Analysis screen.
9. Start the Field Service Calibration program by selecting Utilities|Service Utili-
ties|Field Service Calibration from the Explorer Main Menu screen.
10. Select the Spine Highs/Lows for Array Spine phantom option to automatically
calculate the new thickness indicators for array spine.

4-21
Explorer Technical Manual

11. Select the Spine Phantom scanned in step 6 from the list of Analyzed Scans and
click the Next button. Click the Continue button on the Successfully Calcu-
lated message box.
12. Click on the Temp Write button and click the Continue button on the Success-
fully Wrote message box. Exit Field Service Calibration.

Calibration of Area and BMC, for Array Scan Modes


After the machine has been calibrated for thickness measurement in the array mode, it
must be calibrated for the array AREA and BMC measurements.
1. Restart the Explorer software in service mode (if not already). Center the table.
From the Explorer Main Menu screen, select Utilities|Service Utilities|AP
Reposition. Return to the menu and select Utilities|Service Utilities|Auto
Scan.
2. Select the Spine Phantom.
3. Click OK at Patient Confirmation dialog box.
4. Select AP Lumbar Scan Type, input 20 for number of scans, uncheck Use
Default, and click the Next button.
5. Select Array and click the Next button.
6. Analyze the first scan manually. Note the scan #. Click the Close button to exit
analysis.
7. From the Explorer Main Menu screen, select Utilities|Service Utilities|Auto
Analyze.
8. Click on unanalyzed Scans and select the 19 scans, click on Compare radio but-
ton and then click the next button.
9. Now, select the scan analyzed in step 6 and click the next button.
10. The 19 scans will now be compared with the analyzed scan.
11. From the Explorer Main Menu screen, select Utilities|Service Utilities|Field
Service Calibration.
12. Click the ACF/BCF button. Select the 20 scans, click next, and then click Con-
tinue at the Successfully Calculated dialog box.
13. At the Field Service Calibration dialog box, click the Write button to write the
values to ARRC.TXT file. Click the Continue button to return to the Explorer
Main Menu screen.
14. Print ARRC.TXT file by using these steps:
• Press the <Ctrl><Esc> keys to bring up the WindowsXP Taskbar.
• Select Service
• Select Files or Folders

4-22
Explorer Technical Manual

• Type ARCC.TXT in the Named: edit box.


• Click Find Now or simply hit the <Enter> key.
• Right click on the first ARRC.TXT entry in the list presented to you and
select Print from the menu. (You want the ARRC.TXT entry in the
QDR\DATA directory NOT the QDR\DATA\LRTEMP directory.)
15. Verify ACF=ACFL=ACFT and BCF=BCFL=BCFT in ARRC.TXT.

Recalyze and Add Array AP Scans to the QC Database


Once the unit has been calibrated, AP scans must be re-analyzed and added to the QC
database.
After the machine has been calibrated for thickness measurement in the array mode, it
must be calibrated for the array AREA and BMC measurements.
1. From the Explorer Main Menu screen, select Utilities|Service Utilities|Auto
Analyze.
2. Select Analyzed scans, click RECALYZE, and select the 20 scans just acquired
and click the Next button.
3. . The scans will now be recalculated to the new calibration.
4. From the Explorer Main Menu screen, select QC|QC Data Manage-
ment|Select Scans.
5. Highlight the scans from the Excluded list and click the Include Scans button
and then the Finish button. (Selection is easier if you click Scan Date bar above
the list. This will order the scans by date and time).
6. Select the QC|QC Data Management|Plot option from the Explorer Main
Menu Screen to generate an AREA, BMC, and BMD plot for each array mode.
7. Ensure that all scans fall between the two, dotted limit lines.
8. Print the BMD, BMC, and Area QC plots. Include them with the other paperwork
being returned to Hologic.
Note: For more details on performing the QC setup and producing QC plots, refer to the
QDR for Windows XP Reference Manual.

4-23
Explorer Technical Manual

Sample HARDWARE.INI (Calibration) File


HARDWARE.INI Variables Set by… Descriptions
[Version] .INI File Section Name
HardwareModel=24 Factory Explorer
[Laser]
LaserYOffset=-38100 Factory Laser Offset Value
[TxMotor]
PosLimitPosition=409575 Factory Positive limit in microns (= 16 1/8 inches)
SpineInitPos=208000 Factory TX center position in microns (= 8.18 inches)
MaxSpinePos=233400 Factory + Limit for spine placement
MinSpinePos=182600 Factory - Limit for spine placement
PatientUnloadPos=0 Factory Table position for unloading patients
APLatDiff=376200 Factory Not Used
MaxTrackOffset=25400 Factory
LocalMotionMax=465000 Factory Max table position using Control Panel
LocalMotionMin=0 Factory Min table position using Control Panel
LocalMotionHomePos=208000 Factory X Center table position using Control Panel
ScanPosTolerance=25400 +/- Tolerance for Scan X Position (1 in)
[TyMotor]
SpineInitPos=503175 Factory TX center position in microns (= 8.18 inches)
FlatPos=503175 Factory TX position for Detector Flattening
LocalMotionPos=503175 Factory Y Center table position using Control Panel
ScanPosTolerance=25400 Factory +/- Tolerance for Scan Y Position (1 in)
[AyMotor]
SpineInitPos=460000 Factory C-Arm scan start position
FlatPos=510000 Factory Detector flattening start position
FlatMoveDist=25400 Factory Detector flattening move distance
LocalMotionHomePos=460000 Factory Center C-Arm position
YTiltOffset=-127000 Factory
TBarStartPos=152400 Factory Tissue BAR Start Position
[Gain] Collimator & gain selection
SpineHiGain=3,3 Factory
SpineLoGain=3,3 Factory
SpineHiGain=6,3 Factory
SpineLoGain=6,2 Factory
SpineHiGain=7,2 Factory
SpineLoGain=7,1 Factory
SpineHiGain=8,1 Factory
SpineLoGain=8,0 Factory
SpineHiGain=10,2 Factory
SpineLoGain=10,1 Factory
DecubHiGain=3,3 Factory

4-24
Explorer QDR Series Technical Manual

Sample HARDWARE.INI (Calibration) File (continued)


HARDWARE.INI Variables Set by… Descriptions
DecubLoGain=3,3 Factory
HipHiGain=6,3 Factory
HipLoGain=6,2 Factory
HipHiGain=7,2 Factory
HipLoGain=7,1 Factory
IVAHiGain=6,1 Factory Not Used
IVALoGain=6,1 Factory Not Used
[Flattening]
DailyFlatFile=FLT721NA.DCL Automatic
[Attenuator]
Tolerance=15 Factory +/- Tolerance for Attenuator Blocks
BDelay=30 Factory Boot Up Delay
MaxVHU=70000 Factory Maximum Vinnie Heat Units
CoolRate=250 Factory Cooling Rate of VHU
Vhus1=90 Factory X-Ray Mode 1
Vhus2=250 Factory X-Ray Mode 2
Vhus3=300 Factory X-Ray Mode 3
Vhus4=120 Factory X-Ray Mode 4
Vhus5=90 Factory X-Ray Mode 5
Vhus6=2100 Factory X-Ray Mode 6
Vhus7=700 Factory X-Ray Mode 7
Vhus8=1000 Factory X-Ray Mode 8

Sample ARRC.TXT (Array Calibration) File


ARRC.TXT Variables Set by… Descriptions
A variable used by the software to determine what
USE = 0 Factory parameters to use for scatter correction. It can be set
to one of four possible values (0, 1, 2, or 3) but should
always be set to 0.
These numbers are used for all array modes. The
Q4 = 0.651 1.013 1.461 Factory numbers are determined by measuring a three-step
(thin, medium, and thick) block phantom at the
factory.
Q4_HAT = 0.617 0.980 Factory Similar to Q4, except the measurements are taken on
1.445 the Step phantom with extra absorber material placed
on top of it.
T4 = 2.628 Factory Thickness (height in inches) of the block phantom.
T4_HAT = 7.00 Factory Thickness (height in inches) of the block phantom
with extra absorber.
DELTA0 = 0.493 Factory Thickness of tissue segment in the filter wheel.
T0_N = 6.85 Fact. or F.S. Overall height of the phantom (in inches) shipped with
the Explorer. (Varies with each phantom.)

4-25
Explorer QDR Series Technical Manual

HIA_N = 1305.39 Automatic Hi Air,


LOA_N = 1622.56 Automatic Low Air,
HIT_N = 1389.09 Automatic Hi Tissue, and
LOT_N = 1718.52 Automatic Low Tissue attenuations values in "raw A/D"
numbers.
These values are used to determine the thickness of a
patient being scanned in an AP array scan mode.
They are updated whenever a spine array scan is
added to the QC database (i.e. the daily array QC scan)
or in FSCAL.
ACF = 1.021550 Fact. or F.S. Area Correction Factor for AP Spine and Hip scans
BCF = 1.005450 Fact. or F.S. Bone Correction Factor for AP Spine and Hip scans
SFF = 1.083600 Factory Spine Fan Factors
1.095000
LFF = 0.943000 Factory
0.954000
HFF = 1.077300 Factory Hip Fan Factors
1.120000 Multiplication factors for BCF & ACF array scans.
1.077300
1.120000
1.077300
1.120000
1.077300
1.120000

ARRC.TXT Variables Set by… Descriptions


T0_NL = 6.00 Factory Width of the phantom shipped with the Explorer.
This value is the same for all phantoms.
HIA_NL = 0.00 Automatic
LOA_NL = 0.00 Automatic
HIT_NL = 0.00 Automatic
LOT_NL = 0.00 Automatic
ACFL = 1.021550 Fact. or F.S.
BCFL = 1.005450 Fact. or F.S.
HIA_NF = 1319.16 Factory Hi Air,
LOA_NF = 1634.86 Factory Low Air,
HIT_NF = 1402.73 Factory Hi Tissue, and
LOT_NF = 1730.71 Factory Low Tissue attenuations values in "raw A/D"
numbers.
Factory values for AP array scans. Used as
reference for the software to determine if the Hi/
Low values are out of factory range.
HIA_NLF = 0.00 Factory Hi Air
LOA_NLF = 0.00 Factory Lo Air
Not used in the Explorer Factory Hi Tissue

4-26
Explorer QDR Series Technical Manual

LOT_NLF = 0.00 Factory Lo Tissue


QC_HILO_MIN = 50.0 Factory Day-to-day drift check warning message limit.
QC_HILO_MAX = 100.0 Factory Day-to-day drift check error message limit.
QC_HILO_FACT_MAX Factory Long-term drift check warning message limit.
= 200.0
QC_HILO_FACT_MIN = Factory Long-term drift check error message limit.
100.0
UPDATED = 06/20/94 Automatic Date and time the ARRC.TXT file was last
12:00:00 updated.
UPDATED_N = Automatic Date Last Updated - Normal
UPDATED_NL = Automatic Date Last Updated - Lateral

ARRC.TXT Variables Set by… Descriptions


QDR_SERIAL_NB = Factory Serial number of the Explorer.
4506
ARRAY_PHANTOM_N Fact. or F.S. Serial number of the phantom shipped with the
B = 1922 instrument.
ARRC_SEQUENCE_NB Automatic Sequential number of the last ARRC.TXT file.
= 001
ACFT = 1.021550 Fact. or F.S. Area Correction Factor for Turbo scans.
BCFT = 1.005450 Fact. or F.S. Bone Correction Factor for Turbo scans.
SFFT = 1.083600 Factory Spine Fan Factors Turbo mode correction factors.
1.095000
HFFT = 1.042700 Factory Hip Fan Factors Turbo mode.
1.102800 Multiplication factors for BCF & ACF in turbo
1.042700 1.102800 scan modes. These values are the same on all
1.042700 1.102800 Explorers.
1.042700 1.102800
VERSION_NB = 005 3/ Fan Factor Version # (same for all)
27/95
WBINTACF = 1.000 Whole Body Fan Factor Area
WBINTBCF = 1.131 Whole Body Fan Factor BMC
AWBAREA = 0.975 Whole Body Fan Factor Area
AWBBMC = 1.020 Whole Body Fan Factor BMC
AWBLEAN = 0.962 Body Composition Fan Factor lean
AWBWT = 1.01 Body Composition Total Mass
FOREAREA = 1.0 Forearm Fan Factor Area
FOREBMC = 1.0 Forearm Fan Factor BMC
HWVERSION = 14 System Hardware Descriptor
PCAL_STEP_NB = 5 PCAL Step Phantom

4-27
Explorer QDR Series Technical Manual

4-28
Explorer Technical Manual

Section 5
REMOVE AND REPLACE
PROCEDURES
This section describes how to remove and replace the Field Replaceable Units (FRUs) in
the Explorer. To safely perform a FRU removal or replacement, take care to follow the
procedure precisely as written.
Note: Whenever a component in the x-ray path is replaced, you must rerun QC and
recalibrate.

Recommended Tools
Tool Size/Type
Hex driver 3/32”
Hex driver or wrench 5/32”
Nut driver 1/4”
Nut driver 3/8 inch
Nut driver 5/16”
Nut driver 7/16”
Screwdriver Narrow slotted
Screwdriver Phillips head
Screwdriver Slotted
Wrench 3/8 inch

Electronics Tray FRUs


This section describes how to remove and replace the FRUs in the Electronics Tray/
Carriage Drive area of the Explorer (see Figure 5-1). To remove any of the FRUs in the
Electronics Tray assembly, remove the 6 Phillips retaining screws and remove the cover.
Next, remove the nut that holds the ground wire to the cover ground lug.

Electronics Tray Printed Circuit Boards


To remove and replace the Distribution Board, Motor Controller Board, Power Supply or
AY Drive Board refer to Figure 5-1and Figure 5-2 and follow the procedure below:
1. Move the C-arm all the way to the right.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. Remove the cable covers and unplug the cables on the board to be replaced.
4. To remove the board, unscrew the Phillips screws holding the board.
Note: Some boards have standoffs and/or plastic hold-down snaps.
5. To replace the board reverse the steps.

5-1
Explorer Technical Manual

6. Restore the motor cable shield.


Note: When replacing the AY Motor Controller board, make sure that you set the ID
switch to 7.

C-Arm Y Belt
To remove and replace the C-arm Y-Belt, refer to Figure 5-1and Figure 5-2 and follow the
procedure below:
1. Move the C-arm to the center of the table.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. At the encoder end of the belt, loosen the 2 bolts holding the tension block,
remove the tensioning nut, and remove and replace the belt.
4. Install the tension spring and tensioning nut.
5. Tighten the tension nut so that the spring compresses to 7/8 inch. The bracket cut-
out can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
6. Tighten the two mounting bolts holding the tension block.
7. Perform the MOTOR$AY calibration procedure.

00131-C0501

Distribution Board AY Encoder


Power Supply DIN Rail Assembly

Figure 5-1 Electronic Tray FRUs (Right Rear View)

5-2
Explorer Technical Manual

00131-C0502

AY Motor AY Motor Controller Board AY Belt

Figure 5-2 Electronic Tray FRUs (Left Rear View)

C-Arm Y Motor or Gearcase


To remove and replace the C-arm Y-Motor or Gearcase, refer to Figure 5-2 and follow the
procedure below:
1. Move the C-arm to the center of the table.
2. Turn off the computer power and the main circuit breaker.
3. At the encoder end of the belt, loosen the 2 bolts holding the tension block,
remove the tensioning nut and remove the belt.
4. At the motor end of the belt, remove the cable cover and unplug the motor cable
from the Motor Controller Board.
5. Remove the four Allen bolts, and nuts, holding the motor and gearcase.
Note: The two bottom nuts are accessible with a ratchet wrench and extension.
6. Remove the motor first, then the gearcase.
7. If replacing the gearcase, install the pulley from the old gearcase on the new
gearcase.
8. Replace the motor and gear case assembly making sure to align the gasket cor-
rectly. Snug the Allen bolts, but do not over tighten.
9. Install the belt on both pulleys.
10. At the encoder end of the belt, install the tension spring and tensioning nut.

5-3
Explorer Technical Manual

11. Tighten the tension nut so that the spring compresses to 7/8 inch. The bracket cut-
out can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
12. Tighten the two mounting bolts holding the tension block.
13. Restore the cable shield and ground strap terminations.
14. Perform the MOTOR$AY calibration procedure.

C-Arm Y Encoder
To remove and replace the Encoder, refer to Figure 5-1 and follow the procedure below:
1. Before removing power from the Explorer, remove both cable covers, unhook the
center cable hold-down clip and free the encoder cable without unplugging it.
2. Move the C-arm towards the center of the Scanner (so the encoder is accessible).
3. Turn off the computer power and the main circuit breaker.
4. Unplug the encoder from the Motor Controller Board and pull out the cable.
5. Remove the coupling hardware holding the encoder to the belt drive.
6. Remove the encoder from the bracket assembly.
7. Replace the encoder on the bracket, install the coupling but do not tighten the
screws.
8. Perform the MOTOR$AY calibration procedure.
Note: When starting this procedure, make sure the encoder is not clamped to the drive
shaft.

Table Y FRUs
This section describes how to remove and replace the FRUs associated with the Table Y
motion of the Explorer (see Figure 5-3 and Figure 5-4).
1. Before removing power from the Explorer, move the table to the far left using the
Motor Control Pad. The Motor Control Pad can be accessed from the Explorer
Main Menu screen by choosing Utilities|Emergency Motion.
2. Turn off the Explorer power at the main circuit breaker.

5-4
Explorer Technical Manual

3. Remove 2 screws from the right table rail end cover, and slide the cover off from
the end.

00131-C0503

TY Motor Controller Board TY Motor

Figure 5-3 Right Side Table Y FRUs

Table Y Belt
To remove and replace the Table Y-Belt, refer to Figure 5-3 and Figure 5-4 and follow the
procedure below:
1. Center the table and remove the two screws that secure the tabletop in place.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. Pull the tabletop to the left far enough to remove the right table rail end cover,
and remove the cover. Mark this as the right cover so that it is not confused later
with the left cover (they are not interchangeable).
4. Pull the tabletop to the right far enough to remove the left table rail end cover,
and remove the cover. Mark this as the left cover so that it is not confused later
with the right cover (they are not interchangeable).
5. At the encoder end of the belt, loosen the 2 bolts holding the tension block,
remove the tensioning nut, and remove and replace the belt.
6. Install the tension spring and tensioning nut.
7. Tighten the tension nut so that the spring is compressed to 7/8 inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).

5-5
Explorer Technical Manual

8. Tighten the two mounting bolts holding the tension block.


9. Perform the MOTOR$TY calibration procedure.
Note: When starting this procedure, make sure the encoder is not clamped to the drive
shaft tubing.

Table Y Motor or Gearcase


To remove and replace the Table Y Motor or Gearcase, refer to Figure 5-3 and Figure 5-4,
perform the steps for removing the belt described above, and continue with this procedure.
Perform the following:
1. Remove the Table Y Belt as described above.
2. Unplug the motor cable from the Motor Controller Board.
3. Remove the four Allen bolts, and nuts, holding the motor and gearcase.
4. Remove the motor and the gearcase.
5. If replacing the gearcase, install the pulley from the old gearcase on the new
gearcase.
6. Replace the motor and gear case assembly making sure to align the gasket cor-
rectly. Snug the Allen bolts but don’t over tighten.
7. Install the belt on both pulleys.
8. At the encoder end of the belt, install the tension spring and tensioning nut.
9. Tighten the tension nut so that the spring is compressed to 7/8 inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
10. Tighten the two mounting bolts holding the tension block.
11. Perform the MOTOR$TY calibration procedure.
Note: When starting this procedure, make sure the encoder is not clamped to the drive
shaft tubing.

Table Y Motor Controller PCB


To remove and replace the Table Y Motor Controller board refer to Figure 5-3 and follow
the procedure below:
1. Unplug the cables on the board.
2. To remove the board, unscrew the Phillips screws holding the board
Note: Some boards have standoffs and/or plastic hold-down snaps.
3. To replace the board, reverse the steps.
Note: When replacing the TY Motor Controller board, ensure that the ID switch is set to 5.

5-6
Explorer Technical Manual

KP1085_003-0504

Figure 5-4 Left Side Table Y FRUs

Table Y Encoder
To remove and replace the Table Y Encoder, refer to Figure 5-3 and Figure 5-4 and follow
the procedure below:
1. Center the table and remove the two screws that secure the tabletop in place.
2. Turn off computer power and the main circuit breaker on the Power Module.
3. Pull the tabletop to the left far enough to remove the right table rail end cover,
and remove the cover. Mark this as the right cover so that it is not confused later
with the left cover (they are not interchangeable).
4. Pull the tabletop to the right far enough to remove the left table rail end cover,
and remove the cover. Mark this as the left cover so that it is not confused later
with the right cover (they are not interchangeable).
5. Unplug the encoder from the Motor Controller Board and pull out the cable.
6. Remove the coupling holding the encoder to the belt drive.
7. Remove the encoder from the bracket assembly.
8. Replace the encoder on the bracket, install the coupling but do not tighten the
screws.
9. Perform the MOTOR$TY calibration procedure.
Note: When starting this procedure, make sure the encoder is not coupled to the drive
shaft tubing.

5-7
Explorer Technical Manual

Table X FRUs
This section describes how to remove and replace the FRUs associated with Table X
motion of the Explorer (see Figure 5-5).

Table X Motor Controller PCB


To remove and replace the TX Motor Controller board, refer to Figure 5-5 and follow the
procedure below:
1. Before removing power from the Explorer, move the table up as far as it will go.
2. Remove 6 flat head Phillips screws from the upper pedestal cover and remove the
cover.
3. Turn off computer power and the main circuit breaker on the Power Module.
4. Unplug the cables on the TX Motor Controller board.
5. To remove the board, unscrew the Phillips screws holding the board.
6. To replace the board reverse the steps.
7. Restore the motor cable shield.
Note: When replacing the TX Motor Controller board, ensure that the ID switch is set to 4.

Table X Encoder
Table X Belt
Front Endcap Motor Controller Board

Table X Motor

KP1085_003-0505

Figure 5-5 Table X FRUs

Table X Belt
To remove and replace the Table X Belt located within the X Table Drive Assembly, refer
to Figure 5-5 and follow the procedure below:

5-8
Explorer Technical Manual

1. Before removing power from the Explorer, move the table to the far left using the
Motor Control Pad. The Motor Control Pad can be accessed from the Explorer
Main Menu screen by choosing Utilities|Emergency Motion.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. Remove 2 screws from the right table rail end cover, and slide the cover off from
the end.
4. Remove the front endcap from the table X drive assembly (3 Phillips screws).
5. Remove the back Phillips screw from the right side cover (of the table X drive
assembly) and slide the cover out from the frontEmotionalisthis provides access
to the belt.
6. At the encoder end of the belt, loosen the 2 bolts holding the tension block,
remove the tensioning nut, and remove and replace the belt.
7. Install the tension spring and tensioning nut.
8. Tighten the tension nut so that the spring is compressed to 7/8 inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
9. Tighten the two mounting bolts holding the tension block.
10. Perform the MOTOR$TX calibration procedure.
Note: When starting the MOTOR$TX calibration procedure, make sure the encoder is not
clamped to the drive shaft tubing.

Table X Motor or Gearcase


To remove and replace the Table X Motor or Gearcase, refer to Figure 5-5 and follow the
procedure below:
1. Remove the Table X Belt as described above.
2. Remove 5 flat head Phillips screws from the upper pedestal cover and remove the
cover.
3. Unplug the motor cable from the Motor Controller Board.
4. Remove the four Allen bolts, and nuts, holding the motor and gearcase.
5. Remove the motor and the gearcase.
6. If replacing the gearcase, install the pulley from the old gearcase on the new
gearcase.
7. Replace the motor and gear case assembly making sure to align the gasket cor-
rectly. Snug the Allen bolts but don’t over tighten.
8. Install the belt on both pulleys.
9. At the encoder end of the belt, install the tension spring and tensioning nut.

5-9
Explorer Technical Manual

10. Tighten the tension nut so that the spring is compressed to 7/8 inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
11. Tighten the two mounting bolts holding the tension block.
12. Restore the cable shield and ground strap terminations.
13. Perform the MOTOR$TX calibration procedure.
Note: When starting the MOTOR$TX calibration procedure, make sure the encoder is not
clamped to the drive shaft tubing

Table X Encoder
To remove and replace the Table X Encoder located within the X Table Drive Assembly,
refer to Figure 5-5 and follow the procedure below:
1. Before removing power from the Explorer, move the table to the far left using the
Motor Control Pad. The Motor Control Pad can be accessed from the Explorer
Main Menu screen by choosing Utilities|Emergency Motion.
2. Turn off the computer power and the main circuit breaker on the Power Module.
3. Remove 2 screws from the right table rail end cover, and slide the cover off the
end.
4. Remove the front endcap from the X table drive assembly. The cover is held on
with 2 hex screws.
5. Remove 6 Phillips screws from the right side cover (of the X table drive assem-
bly) and slide the cover out from the front. This provides access to the encoder
and belt.
6. Remove 5 flat head Phillips screws from the upper pedestal cover and remove the
cover. This provides access to the Motor Controller Board.
7. Unplug the encoder from the Motor Controller Board and pull out the cable.
8. Remove the coupling holding the encoder to the belt drive.
9. Remove the encoder from the bracket assembly.
10. Replace the encoder on the bracket, install the coupling but do not tighten the
screws.
11. Perform the MOTOR$TX calibration procedure.
Note: When starting the MOTOR$TX calibration procedure, make sure the encoder is not
coupled to the drive shaft tubing.

5-10
Explorer Technical Manual

Lower C-Arm FRUs


This section describes how to remove and replace the C-arm Interface Board, X-Ray
Controller (XRC), Tank, and Filter Drum Assemblies located on the lower C-arm (see
Figure 5-6)

C-Arm Interface Board


To remove and replace the C-Arm Interface Board, refer to Figure 5-6 and follow the
procedure below:
1. With the C-arm in the center of the table, use the Control Panel to move the table
out as far as it will go, for easier access to the cover screws.
2. Turn off the Explorer computer and the main power circuit breaker.
3. Remove the rear tank cover (covers the C-arm Interface board).
4. Unplug all cables to the C-Arm Interface Board.
5. Remove 4 Phillips screws and remove the board.
6. To replace the C-Arm Interface Board reverse the steps.
7. Perform the Test Scan Modes procedure in the Installation section of this man-
ual.

X-Ray Controller Assembly


To remove and replace the X-Ray Controller (XRC), refer to Figure 5-6 and do the
following:
1. Using the Control Panel, move the table in for easier access to the XRC.
2. Turn off the computer and the main power circuit breaker.
3. Remove the lower C-arm cover by removing 2 Phillips screws and sliding it out
from the front.
4. Remove the 4 Phillips screws holding the XRC.
5. Pull the XRC forward far enough to access the cables.
6. Unplug the cables and remove the XRC.
7. To replace the XRC reverse the steps.
8. Perform the Area, BMD and BMC Calibration procedure in the Alignment and
Calibration section of this manual.
9. Perform the Test Scan Modes procedure in the Installation section of this man-
ual.

5-11
Explorer Technical Manual

Filter Drum Assembly Tank C-arm Interface Board


XRC
(includes Aperture)

00131-C0504

Figure 5-6 Lower C-Arm FRUs

Filter Drum Assembly


Note: To remove and replace FRUs located on the Filter Drum Assembly, refer to the
section titled Filter Drum Assembly FRUs later in this manual.
To remove and replace the Filter Drum Assembly, refer to Figure 5-6 and do the
following:
1. Using the Control Panel, move the table in as far as it will go and center the C-
arm for easier access to the Filter Drum Assembly.
2. Turn off the Explorer computer and the main power circuit breaker.
3. Remove the lower C-arm cover by removing 2 Phillips screws and sliding it out
from the front. Also, remove the bottom cover.
4. Remove the 4 Phillips screws holding the XRC (see Figure 5-6).
5. Pull the XRC forward far enough for access to the Filter Drum (it is not neces-
sary to remove the XRC cables.

5-12
Explorer Technical Manual

6. Remove the rear tank cover (covers the C-arm Interface board).
7. Unplug the 2 Filter Drum cables from the C-arm Interface board.
8. Remove 3 hex head screws (3/32 inch Allen screws).
9. Remove the Filter Drum by lifting it up (while tilting it slightly forward) and out.
10. Position the replacement Filter Drum Assembly into the unit.
11. Replace the 3 hex head screws. Ensure that the curved spring washers are placed
(curved downward) so that the washer compresses when the screw is tightened.
12. Replace the cables.
13. Replace the screws in the XRC assembly.
14. Perform the X-Ray Beam Alignment procedure in the Alignment and Calibration
section of this manual.
15. Perform the A/D Gain Control Adjustment in the Alignment and Calibration sec-
tion of this manual.
16. Perform Detector Flattening in the Alignment and Calibration section of this
manual.

Tank Assembly
To remove and replace the Tank Assembly, refer to Figure 5-6 and Figure 5-7 and follow
the procedure below:

WARNING: Because of the weight of the tank (about 200lbs), 2 people are required to
safely remove and replace the tank

1. Using the Control Panel, move the table in as far as it will go for easier access.
2. Turn off the Explorer computer and the main power circuit breaker.
3. Remove the lower C-arm cover by removing 2 Phillips screws and sliding it out
from the front.
4. Remove the XRC (see procedure above).
5. Remove the Filter Drum Assembly (see procedure above).
6. Lock the C-arm in place by securing the shipping brackets on each side of the
arm.
7. Remove the two L-brackets from the front of the tank that holds the tank to the
C-arm (7/16 inch).
8. Carefully slide the tank out of the C-arm onto a pallet (requires 2 people).
9. Remove the old tank from the tank tray, and install the new tank on the tray.

5-13
Explorer Technical Manual

KP1085_003-0510

Figure 5-7 Top View of Tank


10. Remove the cable assembly from the old tank and install it on the new tank (see
Figure 5-7).
11. Remove the Filter Drum support brackets from the old tank and install on the
new tank (see Figure 5-7).
12. Slide the tank tray, with tank installed, back in place in the C-arm and replace the
L-brackets.
13. Remove the 2 shipping brackets on each side of the C-arm.
14. Replace the Filter Drum Assembly.
15. Replace the X-ray Controller Assembly.
16. Be sure to reconnect the cable shield and ground strap terminals.
17. Before turning on the Scanner, check the tank cable connections to ensure that
they are correct.
18. Starting with the Check Tube kV Peak Potential procedure in the Installation sec-
tion of this manual, perform all the procedures in that section in order, with the
exception of the Calibrate Motors and Check Laser Positioning Offset proce-
dures.

5-14
Explorer Technical Manual

Upper C-Arm FRUS


This section describes how to remove and replace the Detector Assembly, Laser
Assembly, and Control Panel located on the upper C-arm.

Detector Assembly
To remove and replace the Detector Assembly (refer to Figure 5-8) and follow the
procedure below:
1. Turn off the Explorer computer and the main power circuit breaker.
2. Remove the top C-arm cover.
3. Remove the cables at the Detector Assembly.
4. Remove the 4 bolts holding the Detector Assembly (on rubber grommets) to the
C-arm and then remove the assembly.
5. Replace the Detector Assembly (reverse the steps).
6. Perform the A/D Gain Control Adjustment in the Alignment and Calibration sec-
tion of this manual.
7. Perform Detector Flattening in the Alignment and Calibration section of this
manual.

00131-C0505

Detector Assembly Control Panel

Figure 5-8 Top View of Upper C-Arm

Laser Assembly
To remove and replace the Laser, or Laser Assembly, refer to Figure 5-8 and Figure 5-9
and follow the procedure below:

5-15
Explorer Technical Manual

1. Turn off the Explorer computer and the main power circuit breaker on the Power
Module.
2. Remove the top C-arm cover.
3. Remove the Detector Assembly (see procedure above).

Laser Clamping Screw

Laser Assembly

Laser

00131-C0506

Figure 5-9 Laser Assembly


4. To replace the laser only, loosen the laser clamping screw. To replace the laser
assembly, remove the three mounting screws.
5. To adjust the laser, loosen the laser clamping screw, turn the laser on and rotate it
until the correct alignment is seen.

Control Panel
To remove and replace the Control Panel refer to Figure 5-8 and follow the procedure
below:
1. Turn off the Explorer instrument power, computer power, and main circuit
breaker.
2. Remove the top cover of the C-arm.
3. Unplug the cables from the Control Panel to the Detector Assembly.
4. Remove the screws securing the Control Panel to the C-Arm and remove the
panel.
5. To replace the Control Panel reverse the steps.

Filter Drum Assembly FRUs


This section describes how to remove and replace the FRUs on the Filter Drum Assembly
located on the Lower C-Arm (see Figure 5-6).

5-16
Explorer Technical Manual

KP1085_003-0518

Figure 5-10 Rear Drum Assembly FRUs

Drum Encoder PCB


To remove and replace the Drum Encoder PCB (140-0089), refer to Figure 5-10 and
follow the procedure below:
1. Remove the Filter Drum Assembly (see procedure above).
2. Remove the aperture by removing 4 Phillips screws that secure it.
3. Unplug the PCB cable.
4. Remove and replace the PCB (ensure that the encoder wheels are not touching
the sensors.
5. Replace the aperture.

Drum Belts
To remove and replace either Drum Belt (130 teeth or 150 teeth), refer to Figure 5-11 and
follow the procedure below:
1. Remove the Filter Drum Assembly (see procedure above).
2. Remove the aperture by removing 4 Phillips screws that secure it.
3. Loosen the 2 drum motor mount screws, the idler screw and the outer pulley set
screw. Remove the belts.

5-17
Explorer Technical Manual

KP1085_003-0519

Figure 5-11 Front Drum Assembly FRUs

4. Install the Filter Drum Alignment Pin (099-0110), small end first, through the
slotted holes and into the small hole at the base plate. If the pin is installed prop-
erly, the drum will not rotate.
5. Install the 150 tooth belt on the back pulley (ensure the belt is positioned under
the idler).
6. Install the 130 tooth belt on the front pulley.
7. Tighten the 2 drum motor mount screws (the motor mount is spring loaded to
seek proper tension).
8. Tighten the idler screw (the idler is spring loaded to seek proper tension).
9. Tighten the outer pulley set screw.
10. Remove the Filter Drum Alignment Pin.

Stepper Motor Assembly


To remove and replace the Stepper Motor Assembly, refer to Figure 5-11 and follow the
procedure below:
1. Remove the Filter Drum Assembly (see procedure above).
2. Remove the aperture and drum belts (see procedure above).
3. Remove the stepper motor pulley.
4. Remove and replace the motor (4 flathead Phillips screws).

5-18
Explorer Technical Manual

5. Replace the pulley, use Loctite 222 (540-0100) on the set screw.
6. Refer to the Drum Belts procedure above and replace the belts.

Drum Bearings
To remove and replace the drum bearings, refer to Figure 5-11 and Figure 5-12 and follow
the procedure below:
1. Remove the Filter Drum Assembly (see procedure above).
2. Remove the aperture and drum belts (see procedures above).
3. Remove the 2 drum motor mount screws (see Figure 5-12).
4. Remove 6 screws from underneath the drum assembly base plate (4 screws hold
the end plates and 2 screws hold the motor mount spring tensioner block).
5. Remove the side plates and lead shields (3 Phillips screws on each side).
6. Remove the drum encoder PCB and both encoder wheels.
7. Remove both drum belt pulleys.
8. Remove the drum from the endplates.
9. Remove one drum endcap (4 Phillips screws) and remove the inner drum.
10. Replace the bearings.
11. Reassemble the drum and replace (and tighten) the screws in the endcap.
Note: When replacing the end cap ensure that the flats on each side of the drum shaft are
facing the same way (this happens where the two side-by-side holes line up).
12. Replace the 2 endplates. The endplate with the cutout (for the PCB) goes on the
drum end away from the alignment hole (see Figure 5-12).

5-19
Explorer Technical Manual

13. Replace the lead shields and side plates (6 screws).

KP1085_003-0520

Figure 5-12 Drum Bearings


14. Align the 2 endplates onto the holes on the recessed area on the base plate. Orient
the front endplate towards the 2 slotted holes. Orient the rear endplate on the
opposite set of holes and position so that the endplate side with cutout (for PCB)
is facing the “C” cutout of the base plate. Replace the 4 flat washers, split locks,
and 4-40 x 3/8 screws that secure the endplates.
15. Insert the Drum Spacing Fixture (099-0296) between the endcap and the front
endplate. Push the filter drum against the fixture to center it between the two
blocks. Do not remove the fixture at this time.
Install the spacers and pulleys at the front endplate (see Figure 5-12). Ensure that the
pulley set screws line up with the flat, and use Loctite 222 on set screws. Press the filter
drum against the fixture, and the pulley against the endplate, while tightening the set
screw. Note: Leave the outer pulley set screw loose until belts are installed.
16. Remove the Drum Spacing Fixture.
17. Replace the inner encoder wheel (see Figure 5-13). Press (squeeze together) the
drum and encoder wheel against the endplate and tighten the set screw (use Loc-
tite 222).

5-20
Explorer Technical Manual

18. Replace the outer encoder wheel.


19. Replace the drum encoder PCB. Ensure that the encoder wheels do not contact
the sensors.
20. Replace the stepper motor assembly (2 screws on the motor mount, and 2 screws
on the base plate). Leave the motor mount screws loose for now.
21. Replace the drum belts (see the Drum Belts procedure).
22. Tighten outer pulley set screw after belts are installed.
23. Refer to the Aperture procedure and install the aperture.

KP1085_003-0521

Figure 5-13 Drum Inner Bearings

Replacing EMI Cables


EMI cables are ribbon cables modified with braided shielding and ground lugs. When
replacing, be sure each ground lug is fastened to a ground connection, usually to the
ground plane of the PCB at that end. Use a star washer between the ground lug and the
ground plane. Be sure the ground lug does not short out any component on the PCB.
When dressing the cables, be sure the braided shield does not come in contact with any
electrical component or voltage source.

5-21
Explorer Technical Manual

KP1085_003-0522

Figure 5-14 The EMI Compliance Cable

FRU Lists
The following tables provide the information necessary to identify and order the correct
FRU.
Table 5-1. Electronics Tray FRUs (Figure 5-1)

Part Number Description


010-0792 Stepper Motor (AY Drive Motor)
100-0060 Power Supply, 24V +/-15V
140-0085 Motor Control Board
140-0086 Distribution Board
180-0239 Cable, AY Encoder
180-0423 Cable, Arm YY Drive, Shielded
180-0426 Cable, Arm YY Drive, Shielded
255-0038 AY Drive Belt
310-0006 Circuit Breaker, 7.5 Amp
310-0053 Circuit Breaker, 15 Amp
310-0018 Circuit Breaker, 20 Amp
310-0014 Fuse, 10 Amp
310-0044 Fuse, 6 Amp

5-22
Explorer Technical Manual

310-0086 Fuse, 8 Amp


325-0004 Gearcase 50:1

Table 5-2. Y FRUs (Figure 5-3)

Part Number Description


010-0792 Drive Motor (TY Drive Motor)
140-0085 Motor Control Board
180-0241 Cable, TY Encoder
180-0426 Cable, Table YY Shielded
255-0033 TY Drive Belt
325-0004 Gearcase 50:1

Table 5-3. X-FRUs (Figure 5-3)

Part Number Description


010-0792 Drive Motor (TX Drive Motor)
140-0085 Motor Control Board
180-0240 Cable, TX Encoder
180-0422 Cable, Table XX Drive Shielded
255-0021 TX Drive Belt
325-0005 Gearcase 30:1

Table 5-4. Lower C-Arm FRUs (Figure 5-6)

Part Number Description


010-0575 X-ray Tank
010-1547 Filter Drum Assembly
ASY-00409 XRC, Explorer
140-0090 C-arm Interface Board

Table 5-5. Upper C-Arm FRUs (Figure 5-8 and Figure 5-9)

Part Number Description


010-0682 Assy, Crosshair Laser Block
ASY-00254 Control Panel, Explorer
465-0042 Switch, Pushbutton w/LED Pnl

5-23
Explorer Technical Manual

465-0043 Switch, Emergency Stop Anti-Rotate


010-1611 Assy, Explorer Det 128 Chan 2mm

Table 5-6. Computer and Isolation Module

Part Number Description


PWR-100-DISCRY Kit, Power, Explorer, 100V
PWR-120-DISCRY Kit, Power, Explorer, 120V
PWR-230-DISCRY Kit, Power, Explorer, 230V
010-1490 Assembly, Torroid
120-0049 Monitor, 17" SVGA
120-0152 Mouse, P/S2
120-0161 Keyboard, P/S2
CMP-00018 Explorer Computer w/XP

Table 5-7. Internal Computer Assemblies

Part Number Description


120-0208 5.2 GB Magneto-Optical Drive
120-0135 SCSI Interface Board, PCI
120-0116 Hard Drive, 20GB minimum
120-0017 3.5”, 1.44MB Floppy Drive
120-0241 Mt. Ranier CD-ROM R/W
120-0154 Network Card, PCI
140-0124 PCI Communications Controller PCB
180-0136 SCSI Cable, Internal

Table 5-8. Aperture Assembly FRUs

Part Number Description


140-0068 Aperture Motor PCB
180-0267 Cable, Aperture Motor Signal
255-0032 Aperture Position Belt
320-0041 Aperture Stepper Motor

5-24
Explorer Technical Manual

Table 5-9. Front Drum Assembly FRUs (Figure 5-11)

Part Number Description


010-0627 Filter Drum Stepper Motor Assembly
140-0089 Drum Encoder Assembly
255-0030 Drum Wheel Belt 150 Teeth
255-0031 Drum Wheel Belt 130 Teeth

Table 5-10. Drum Bearings (Figure 5-12)

Part Number Description


250-0045 Drum Inner Bearing
250-0046 Drum Outer Bearing

Table 5-11. Cables

Part Number Description


180-0213 Cable, Fan, X-Ray Controller
180-0328 Cable, X-Ray Signal
180-0529 Cable, C-arm Sig Explorer
180-0344 Cable, Tank Drive
180-0416 Cable, X-Ray Power
180-0526 Cable, X-Ray Power
180-0417 Cable, Safety Switch
180-0418 Cable, Operator Panel
180-0531 Cable, Operator Panel
180-0419 Cable, DAS Communication
180-0420 Cable, C-Arm Signal
180-0421 Cable, C-Arm Power
180-0422 Cable, Table XX Drive
180-0532 Cable, Table X Drive
180-0423 Cable, Arm YY Drive
180-0426 Cable, Table YY Drive
189-0543 Cable, Arm YY Drive
180-0530 Cable, C-arm Pwr
180-0282 Cable, Laser Pwr

5-25
Explorer Technical Manual

Table 5-12. Miscellaneous

Part Number Description


ASY-00395 Cable Kit, Explorer
100-0033 UPS, 60Hz, 2KVA
100-0034 UPS, 50Hz, 2KVA

Table 5-13. Mobile

Part Number Description


010-0980 ASSY, L'KG PIN Y CARR, MOBILE
010-0981 ASSY, L'KG PIN C-ARM MOBILE
010-0993 ASSY,CARR L'KG PIN,MOBILE

Table 5-14. Special Tools

Part Number Description


010-0923 X-Ray Beam Alignment Fixture (2 needed)
099-0110 Filter Drum Alignment Pin
099-0264 Comm Controller Board Loopback Plug
099-0296 Drum Spacing Fixture

5-26
Explorer Technical Manual

Section 6
FAULT ISOLATION
This section provides information to help identify the source of a problem in the Explorer
system. The three general categories are:
Problem... Refer to the section titled... On page...
Dead system, or power problem Power Problems 6-1
Scanner motion problem Motion Problems 6-2
Computer display problem Display Problems 6-4

Before Starting
Before starting, make sure the software configuration is compatible with the scanner.

Software Configuration
To check the software revision and loaded options, from the Explorer Main Screen, select
Utilities|System Configuration. On the System Configuration screen, you will find a
drop-down list box labeled “Software:” By dropping down this list, you will be shown the
release version of the Explorer software and all loaded options. The release version should
be at a minimum level of 12.2.

Hardware Configuration
When troubleshooting, it is sometimes helpful to observe the indicators available on the
PCBs and other FRUs. Many components of the system have LEDs indicating the
presence of necessary voltages and the state of some signals. Section 8 is helpful in
locating these LEDs and observing the state of the system.

Power Problems
Table 6-1 Power Component Locations

Component... Where…
Main circuit breaker Rear of foot end pedestal
LED’s for DC voltages Distribution Board and most
driver and control PCBs
Emergency stop switches and circuits Control panel
Computer power and operation Computer console power strip

6-1
Explorer Technical Manual

The following suggestions apply to a Explorer system exhibiting a power problem:


If... Check... Refer to...
System “dead” Main breaker left pedestal Check Power Line
Emergency Stop or Instrument On/ Voltage heading,
Off switch on Control Panel Section 3
Check fuses
Main power suspect Check AC input voltage Power Module heading,
Power cable to the Torroid Assembly Section 2
Suspected Power Module Check AC Voltage at Scanner DIN
problem Rail
Suspected DC power Check LED’s on Distribution Board Check fuses
supply problem for all DC voltages (lit if present)
24V LED not lit All tape switches and connectors Check fuses Distribution
(tape sw reads 120W when open) board.
Emergency Stop Switch and
assaciated circuitry
Computer is off
Hologic softwaree running
(check error log
System does not start up System properly configured (see
BEFORE STARTING
Before starting, make sure the
software configuration is compatible
with the scanner.
Emergency Stop switch in “On” (Out)
position
Console does not power Main circuit breaker, power strip
up
Also...
Problem... Possible cause... Corrective action...
Laser does Safety feature if system Restart in Service mode
not turn on thinks C-arm is at head of From Service Utilities, run SQDRIVER
table At CARM$$$$> prompt, type
LASER_SAFTY 0 <Enter>
Exit from SQDRIVER
Reboot the computer

Motion Problems
Motion problems are failures related to movement of the table and/or C-arm. In most (but
not all) cases, an error message accompanies Explorer motion problems. Therefore, it is
good troubleshooting practice to check the error log (C:\ERRLOG.DAT file and

6-2
Explorer Technical Manual

C:\QDR\ERROR.LOG) for occurrences of motion related problems. The following


suggestions apply to a Explorer system that exhibits a motion problem.
Start by identifying the bad axis (if it is not obvious). Ask the operator for symptoms and
check the error log. See Figure 6-1 for possible motion directions.
Table 6-2 Motion Components Locations
Refer to the Figure Below
ARM-Y TX TY
Drive Belt Figure 5-2 Figure 5-5 Figure 5-3
Figure 5-4
Drive Motor Figure 5-2 Figure 5-5 Figure 5-3
Driver Board Figure 5-1 Figure 5-1 Figure 5-1
Encoder Figure 5-1 Figure 5-5 Figure 5-4
Motor Controller Board Figure 5-2 Figure 5-5 Figure 5-3
Distribution Board Figure 5-1
Operator Control Panel Figure 5-8
Computer
Communications
Controller

AY

TX

TY

AW-00335_001-0601

Figure 6-1 Scanner Motion Directions

6-3
Explorer Technical Manual

TX, TY, AY
Check... Refer to...
Belt
Encoder coupling Figure 5-1, 5-2, 5-3, 5-4, 5-5
Motor controller board Section 8, PCB SUMMARY INFORMATION
Motor/gear case for the motor
subsystem in question
TY (on A & W models only)
Left and right end panels on table for proper
installation. This may cause problems if reversed.
(Left 030-1376, right 030-1838).

CAUTION: Running any motor subsystem to its limit (in either direction) does not
damage the motor. However, if the motor encoder is not properly calibrated,
it may be damaged.

TX, TY, AY
To... Run... Refer to...
Perform simple table and C- Motor Control Pad The Explorer Main Menu, select:
arm movements (computer motion 1. Utility
control) 2. Emergency Motion
Perform precise table and SQDRIVER Motor Calibration heading, and
C-arm movements Motor$XX (for the specific
motor), Section 4
Monitor all motion parameters
Troubleshoot problems Control Panel heading, Section 2
encountered initiating motion
from the Operator's Control
Panel
Run the hardware checker Hardware checker SQVERIFY

Control Panel Problems


If a Control Panel problem is suspected, or if control panel functions are not responding,
use the SQVERIFY program to help isolate the problem. From Explorer Main Menu,
select Utilities|Service Utilities|SQVERIFY. Select the Panel test function and run all
subtests.

Display Problems
Display problems can be grouped into four general categories: vertical stripe, horizontal
stripe, noise (dots, speckles, etc.), and no display.

6-4
Explorer Technical Manual

Table 6-3 Display Component Locations


Component See...
Detector Assembly Figure 5-8
C-Arm Interface Board Figure 5-6
Filter Drum Assembly Figure 5-6
Filter Drum Assembly (X-Ray Beam Alignment) Page 4-3
Detector Assembly (X-Ray Beam Alignment) Page 4-3
Printer (if quality is bad on printout or no bone displayed).
X-Ray Tank and X-Ray Controller Figure 5-6
Collimator

Vertical Stripe
This type of display problem is most likely related to the detector subsystem. The
following suggestions apply to a Explorer system that exhibits a vertical stripe in the
display:
Check... Refer to...
Detectors X-Ray Survey in graphic mode to check for
signal strength and noise (Section 9)
Narrow vertical stripe - bad detector Data Acquisition System heading,
Section 2, for block diagrams and
interconnection charts or run the DAS tests
in SQVERIFY.
Foreign matter (especially metallic)
anywhere in the X-ray beam path in the
aperture slit, collimator cup, etc.
Also...
Run the hardware checker SQVERIFY

Horizontal Stripe
• This type of display problem is most likely related to the line voltage, bad cables,
or X-ray subsystem. The following suggestions apply to a Explorer system that
exhibits a horizontal stripe in the display:
Check... Refer to...
Line voltage/Power ground Check Power Line Voltage heading, page 3-14
Tube kV Peak Potential Check Tube KV Peak Potential heading, page 3-16
Tube Current Check Tube Current heading, page 3-18
Filter drum turning Figure 5-10
Filter drum belt Figure 5-11
Green LED’s on C-Arm Figure 5-6, Section 8
Interface

6-5
Explorer Technical Manual

Banding–variations in X-ray controller, Figure 5-6


horizontal stripe intensity,
usually spread across display.
Also...
Run the hardware checker SQVERIFY heading, page 9-10

Noise
The term “noise” is used here to describe any flaw, or irregularity in the display (dots,
specks, uneven lines, etc.) or similar problem. The following suggestions apply to a
Explorer system that exhibits noise in the display:
Check... Refer to...
Tube kV peak potential Check Tube KV Peak Potential heading, page 3-16
Tube current Check Tube Current heading, page 3-18
X-ray beam alignment X-Ray Beam Alignmentheading, page 4-3
Signal strength and noise X-Ray Survey heading, page 9-1
Filter Drum is turning
Filter Drum belt Figure 5-11
Green LED’s on C-Arm Interface Figure 5-6, Section 8
board
Aperture plate assembly, first Figure 5-6
precollimator, second precollimator
and collimator for specs of lead and
other deformities.
Also...
Run the hardware checker SQVERIFY, page 9-10

No Display
The term “no display” is used here to describe:
• no scan display
• completely white screen
• completely dark screen
• “blotchy” or “grainy” display
This type of display problem is usually related to the detector subsystem or the X-ray
subsystem.
The following suggestions apply to a Explorer system that exhibits no scan display:
Check... Refer to...
Tube kV peak potential Check Tube KV Peak Potential heading, page 3-16
Tube current Check Tube Current heading, page 3-18
X-ray production Field Service Preventive Maintenance heading,
page 7-1

6-6
Explorer Technical Manual

Signal strength and noise X-Ray Survey in graphic mode, page 9-1
Filter Drum is turning
Filter Drum belt Figure 5-11
Green LED’s on C-Arm Section 8
Interface board
No display may indicate a bad Refer to the Data Acquisition System heading on
Detector Assembly page 2-15.
Also...
Run the hardware checker SQVERIFY on page 9-10

Targeting/Laser Problems
If... Check... Refer to...
Object being Detector array. It may be too far Figure 5-8
scanned appears forward or back inside the upper arm Page 4-17
to the left or assembly. Page 4-18
right of the scan Laser alignment
window Run... Detector Flattening

6-7
Explorer Technical Manual

Data Communications Problems


Data communications problems occur between the computer and the scanner. Refer to the
following table:
Error messages
usually include Check the:
the keywords:
Message Packet Digital Signal Processor. If IC is not seated properly, replace the
Sent board.
Data, power, and ribbon cables for proper seating.
• Received • Device states in the SQDRIVER for errors. If a device is stated
as “E_TIMEOUT”, it is suspect.

Area/BMD/BMC/CV Specification Problems


If... Possible cause Refer to...
X-ray beam Signal output level should not rise or fall by a Page 4-3
misaligned significant margin.
X-ray beam Measure X-ray peak potential and tube current Page 3-17,
quality waveforms. Make sure they are both stable and within 3-19
problems specs. Page 4-4
Check X-Ray Survey bar graph for shifting or
“dancing” X-rays. If so, test X-Ray Controller.
Check A/C line for stability (voltage and frequency).
Mechanical Check motor drive belts for excessive play. Page 7-1
frame Ensure all hardware on the arm, frame, and drives are
problems secure and clear of moving assemblies (C-arm, etc.)
Mechanical Check for loose Detector Assembly Figure 5-8
motor drive assemblies: X-ray Tank Figure 5-6
problems X-ray Controller Figure 5-6
Figure 5-6
The bone and/or tissue segments in the filter drum may Figure 5-10
be defective or out of spec. If so, replace and restart through
system testing. Figure 5-13

X-Ray Problems
The following lists some common X-ray problems and some suggested solutions. Refer to
Section Two Functional Description for more information on the X-Ray subsystem.

6-8
Explorer Technical Manual

No X-Rays
If... Check... Refer to...
X-Ray (24VDC LED) X-ray Enable LED on Control Panel and
on Distribution Board is Distribution Board should be on
not lit
“NO A/C Line Interlock inhibit
Interrupts” message at 24VDC from LVPS, power distribution
start-up X-ray Controller interlock switch
Computer problems (CommCon)
Defective Control Panel I/F
XRC Power Cable/Connector problems

X-Ray Alignment Problems


X-Ray beam does not align properly
If the X-ray beam does not align properly per the procedure, follow these suggestions.
1. Try to align the X-ray beam to the drop-off points per the alignment procedure.
(Check captive screws and jam nuts.)
2. If you align the X-ray beam to both drop-off points but the range is narrow, the
collimator cup and/or the pre-collimator disk on the X-ray tank may be mis-
aligned.
Using glow paper, assure that the X-ray beam is hitting the front and back on the
detector array equally. If not, this may indicate that the X-ray tube has shifted within
the letharge. Perform further verification before replacing the X-ray tank.
3. If the aperture and detector array window are aligned, remove the aperture
assembly and inspect the cup and disk. Look for any debris or lead fragments.
When replacing the cup and disk, align the apertures as closely as possible to the
detector array window.

Detector Flattening Problems


System Consistently Fails the Detector Flattening Procedure
1. Check for loose, misaligned, Collimator Cup, or Pre-collimator Disk. (Refer to
X-ray Alignment Problems above.)
2. If the X-rays are unstable or “dancing”, monitor the X-rays on the X-RAY SUR-
VEY bar graph screen for any amplitude shifting of the defective signal display.
3. If shifting is taking place, check/replace the X-ray Controller Assembly and X-
ray Tank connections. Check the X-ray peak potential and tube current wave-
forms and monitor for waveform distortion. This may indicate a defective inter-
connection cable, X-ray Tank, or power supply.
4. Check output signal level. You may have to reset the ADC gain level.

6-9
Explorer Technical Manual

5. Check for bad detectors in the Detector Assembly. Run the DAS test in SQVER-
IFY.
6. Check the Filter Drum to see that it is installed correctly. Check the segment
readback values in X-Ray Survey with X-rays ON.
7. Check for lead fragments in the X-ray beam. Inspect the Collimator Cup, disk,
and Aperture Assembly.
8. If a failure occurs during a whole body, recheck the TX motor cal files. The X-
rays may be hitting the table edge. Make sure the drives are operating normally.
Examine the Detector Flattening scans for any unusual indications using Utili-
ties|Scan File Plot and selecting the WBAIRQC scan. If an error message is
displayed, go to the Detector Flattening heading on page 4-18.

Laser Problems
WARNING: The laser beam can cause serious retinal damage if focused directly into
the eye. Be sure to turn the laser OFF when visually inspecting the shutter
and aperture.

When troubleshooting the laser, refer to the following:


Make sure... Refer to...
Shutter is open and not obstructing the beam Page 4-21
Motor drives have been calibrated Pages 4-6 to 4-17

6-10
Explorer Technical Manual

If... Suggestions
(Software is inhibiting laser when table head is near the laser)
Move the C-arm to the center of the table and retry.
Do the following:
Check black and yellow wires from the C-arm Interface PCB to the
underside of the Detector Assembly.
Laser Shut off instrument power (Emergency Stop Switch on Control Panel).
does not Disconnect laser power connector.
turn on Connect a DVM to the plug coming from the Detector Assembly.
Turn laser power on.
Check connector for +5VDC.
Check C-arm Interface Board, Distribution Board, all cables.
Check computer.
At installation only, type “laser safety=0” in SQDRIVER and reboot.
Check AY motor calibration file. If data is corrupt or uncalibrated, laser may
not turn on, despite position of C-arm.
Laser Check laser block assembly mirrors for breaks, cracks, or misalignment
crosshair Check that shutter is open
beam is Check that aperture is unobstructed
defective Loosen the securing screw to adjust the laser to perpendicular (see Figure 5-
8)
Check laser button on Control Panel
Check control panel laser switch at PANEL tests in SQVERIFY
Laser Check for bad:
does not C-arm Interface Board
turn off Distribution Board
Interconnect cable
Computer assembly

Oil leakage
If the tank assembly is leaking oil, you may have to tighten the screws on the tank cover. It
is important that you tighten the screws using the proper torque specifications and the
proper sequence.

The Torque Specifications


The torque settings are listed in the following table:
Location Torque Specifications Set wrench
to…
Lexan Cup 10-12 in-lb 11 in-lb

6-11
Explorer Technical Manual

Bladder Gasket 10-12 in-lb 11 in-lb


Transformer Seal 70 in-lb 70 in-lb
Tank Cover 36-40 in-lb 38 in-lb
Gasket

Tank Top Cover Components and Screw Location


The following figure shows the Tank Top Cover and the locations of the seals:
Lexan Cup Bladder Gasket
Tank Cover
Gasket
Xformer (hidden)
Seal

Front
KP1085_003-0602

Figure 6-2 Tank Assembly, Top View

Tightening the Lexan Cup Screws


Tighten the Lexan Cup seals in the following order using the torque settings in the above
table:
7
1 4

5 6

3 2
8
Figure 6-3 Lexan Cup Screw Tightening Order

Tightening the Bladder Gasket Screws


Using the torque settings in the table above and referring to the figure below, tighten the
Bladder Gasket screws as follows:
1. Tighten screw #1, then #2, then #3, and then #4.

6-12
Explorer Technical Manual

2. After screw #4, continue around the edge of the gasket in a clockwise direction,
tightening alternate screws, until returning to #4.
3. Continue around the edge of the gasket in a clockwise direction, tightening all the
remaining screws.

1 4

3 2

Figure 6-4 Bladder Gasket Screws

Tightening the Transformer Seal Screws


Tighten the Transformer Seal screws using the torque settings in the table above and the
sequence in the figure below.

2 4

1 3
KP1085_003-0605

Figure 6-5 Transformer Screws

Tightening the Tank Cover Gasket Screws


Refer to the figure below and the torque settings in the table above and tighten the Tank
Cover Gasket screws as follows:
1. Tighten screws #1 through #6 in the sequence as indicated.

6-13
Explorer Technical Manual

2. Starting at the screw next to screw #6, move in a clockwise direction and tighten
alternate screws.
3. Repeat this pattern in a clockwise direction and tighten all remaining screws.
4. The three Tank Cover Gasket screws at the rear of the tank and shown below are
the most critical locations for oil leaks. Always check these screws if a leak is
suspected.
5. Important: Do not remove screws located on the tank. There are no field
replaceable units or required adjustments inside the tank, so there is no reason to
loosen or remove any screws on the tank.

MOST CRITICAL
1 3

FRONT

6 2 4
KP1085_003-0606

Figure 6-6 Tank Cover Gasket

Miscellaneous Problems
Problem... Symptom... Solution...
Detector X-Ray Survey bar graph is Replace the Detector Assembly
ramped, or it appears as 3 or
4 separate ramped sections.
X-ray lamp Turns on with X-rays off Replace the Distribution Board
System Fails HI-Pot test Check all ground connections
Make sure ground wire terminals are
secure
Check crimped terminals
Hi voltage Buzzing sound from X-ray Check torque on transformer bolts
transformer tank assembly
System No power Check Emergency Stop Switch
Check main power cord
Check Torroid Assembly circuit breaker
Check AC input voltage

6-14
Explorer Technical Manual

Section 7
PREVENTIVE MAINTENANCE
This section lists the procedures that should be performed by trained service personnel at
least once per year, and preferably at six-month intervals.
NOTICE
The Explorer system meets applicable FDA radiation performance standards through its
useful and expected life provided no components or parts are removed from the system
and no unauthorized adjustment or unauthorized replacement of certified components is
performed.

Customer Preventive Maintenance


Hologic requires that the customer run a daily QC scan of the spine phantom supplied with
the Explorer, and add that scan to the QC database. If the CV of the database exceeds
0.8% the customer is asked to inform Hologic Field Service.
The customer is further advised to perform a weekly backup of the patient database
(Backup) and archive QC scans along with their patient scans.

Field Service Preventive Maintenance


Hologic recommends that the following procedures be performed by trained service
personnel at least once per year and preferably at six-month intervals:
[ ] Perform an Archive of all scans and a System Backup.
[ ] Check the QC database for any problems (e.g. drift, etc.).
[ ] Check ERRORLOG.DAT and ERROR.LOG
[ ] Check X-ray tube voltage and current as described in the
INSTALLATION section of this manual.
[ ] Measure scatter, leakage, and patient dose as described in the
INSTALLATION section of this manual.
[ ] Check X-ray beam alignment as described in the ALIGNMENT
AND CALIBRATION section of this manual.
[ ] Run the “Table Top Radiographic Uniformity” on page 4-19 for
whole body units and send a copy of the resulting scan on floppy
disk and printout to Hologic.
[ ] Run 10 spine scans and check calibration.
[ ] Print copies of the baseline phantom scan, recent phantom scan, and
the results of scan averaging for the 10 spine scans in the previous
step. Compare these printouts with those from the last PM for
possible problems, and send the printouts to Hologic to be kept with
the service records for this system.

7-1
Explorer Technical Manual

[ ] During a scan, verify that pressing the red Emergency STOP switch
immediately stops all C-Arm motion, table motion, and X-ray
production. X-ray production should be monitored by a Victoreen
450P or equivalent.
[ ] Replace (if necessary) and adjust the motor drive belts as follows:
Note: Detailed instructions for drive belt replacement and adjustments can be
found in the REMOVE AND REPLACE section of this manual.
[ ] For each drive belt (C-Arm Y, Table X, and Table Y):
1. Loosen the 2 mounting bolts holding the tension block.
2. Tighten the tension nut so that the spring is compressed to 7/8 inch. The
bracket cutout can be used as a measuring guide (the inside of the washer
should be flush with the bracket cut).
3. Tighten the two mounting bolts holding each drive belt’s tension block.
It is not necessary to perform the MOTOR calibration procedures after
tensioning the motor drive belts.
[ ] Ensure that all cable connections are tight.
[ ] Run CHECKDISK on the C drive. This utility program can be
started by clicking on start|Run…, then typing CHKDSK, and
then clicking the OK button. DO NOT run Scandisk on any
Archive media.
[ ] Run Disk Defragmenter to and compact and re-order the files on the
computer's hard disk. This utility program can be started by
clicking on start|Run…, then typing DEFRAG, and then clicking
the OK button.
[ ] Run a computer virus checker (any major brand that is current for
latest virus types).
[ ] Clean the fan filters, paying special attention to the computer fan
filter.
[ ] Clean all exterior metal surfaces and wipe off the rails.
[ ] Clean the monitor screen, keyboard, and mouse ball rollers.
[ ] Clean the inkjet printer rollers and ink cartridge electrical contacts
with isopropyl alcohol.

Guide Rail and Bearing Maintenance


A guide rail and two guide bearings (AY) are located at the lower back of the C-arm (see
Figure 7-1).
1. Clean the AY and the TX guide rails using a dry, clean cloth.

CAUTION: DO NOT use solvents such as alcohol or WD-40. If a solvent is needed to


remove dirt and/or gum buildup on the rail, be sure to thoroughly dry the
rail before moving the bearings. The solvent may harm the bearing grease.

7-2
Explorer Technical Manual

KP1085_003-0701

Figure 7-1 Guide Bearing and Rail


2. Grease the guide bearings.
Note: The bearings must be greased every 100km (62mi) of travel, depending on usage.
This may range between one and two years. The following table lists more
information about the grease to be used.
Generic Brand Names Manufacturer
EP-2 Lithium Beacon 325 ESSO
soap-based grease Alvania Grease RA Shell
Mobilux Grease No. 2 Mobil
Isoflex Super LDS 18 Kluber

Consult Hologic Field Service for more information on obtaining the proper tools,
material, and procedure.

7-3
Explorer Technical Manual

7-4
Explorer Technical Manual

Section 8
PCB SUMMARY INFORMATION
Distribution Board
LED’s1 Signal Voltage Source Jumpers Refer
to...
D1 On +7VDC Int Voltage Reg. JP1 (panel) Out
D32 Off - JP2 (dist) Out
D5 On +5VDC Int Voltage Reg. JP3 (C-arm) Out
D7 On +28VDC Int Low Voltage Figure
Power Supply Circuit Breakers 5-1
D9 On +15VDC Int Low Voltage CB
Power Supply
D10 On -15VDC Int Low Voltage 1
Power Supply
- - +24VDC Int Low Voltage 2 TY In
Power Supply
D112 Off TY DRIVE 3 TX In

D122 Off TX DRIVE 4

D132 Off 5 AY In

D142 Off AY DRIVE


Notes: 1. Voltage indicators = green; motor drive status = red
2. If red led is on, there is a failure in the PCB. The CB should be tripped.

8-1
Explorer Technical Manual

Communications Controller
Jumpers Refer to...
JP1 (IRQ) at 10 In
JP2 (IRQ) Out
JP3 (IREQA) Out Figure 5-17

JP4 (IREQB) In
JP5 (E-OUT) at In
C

Stepper Motor Controller


Signal Signal Source Jumpers and Refer to...
Switches
D3 On +24VDC Ext Signal W1 In
Dist. (SYSRESET)
D6 On +5VDC Int. Voltage
Reg.
- - -5VDC Int. Voltage
Reg.
D7 (4 LED pack)
MEN On Note: All 4 LED’s are On
CPU Flash solid when the selected
DIR Off motor drive is engaged.
STEP Off Set to...
Stepper motor (AY SW1 7 Figure 5-2
Stepper motor (TX) SW1 4 Figure 5-5
Stepper motor (TY) SW1 5 Figure 5-3

8-2
Explorer Technical Manual

54-Channel Detector Assembly


LED’s Voltage Voltage Source Jumpers Refer to...
D4 On +12VDC Int. Voltage Reg. J1 (Run/Test) In
D5 On -12VDC Int. Voltage Reg. J2 (Run/Test) In
D6 On VCC Int. Voltage Reg.
D7 On +5VDC Int. Voltage Reg.
U14 (LED display) Flickers “1” on bootup, then lock on “2”.
Potentiometer R18 (A/D GAIN CNTRL) See procedure in Page 4-24
Section 4.

8-3
Explorer Technical Manual

C-Arm Interface
LED’s Signal Voltage Source Jumpers Refer to...
D1 On +28VDC Ext Signal JP7 DRUM Out
Dist.
D3 On +24VDC Ext Signal JP8 MAIN Out
Dist.
D6 On +5VDC Int. Voltage W1, NORMAL(1 In
Reg. W2 -2)
TEST (2-3)
- - +7VDC Int. Voltage W3 SYS- In
Reg. RESET
- - +15VDC Ext Signal W4 TEST- Out
Dist. MODE
- - -15VDC Ext Signal W5 Safety In
Dist. switches
Drum JP15 Redundant for
D7 On LOCKED W5 may be Out if
W5 is In
D8 Flash DRUM
-ing AT TOP
X-Ray
D9 Off X-RAY
CPU
ERROR
D10 Off X-RAY
ENABLE
D11 On X-RAY
CPU OK
D12 Off X-RAY
FAULT

8-4
Explorer Technical Manual

Section 9
SOFTWARE TOOLS
The Explorer system software includes software tools to troubleshoot the system. They
are:
• SQDRIVER
• SQVERIFY
• Emergency Motion (SQKEYPAD)

X-Ray Survey
The X-Ray Survey is a 32-bit utility that tests the functions of the X-Ray generation and
detection for the system. This utility is used by Hologic manufacturing and service
engineers only.

Invoking X-Ray Survey


1. Log on as the Field Service and start the QDR Software in Service mode.
2. Click Utility|Service Utilities|X-Ray Survey and the X-Ray Survey utility
screen will appear (see Figure 9-1).
Main Display

AW-00335_001-0902
Display Controls Close utility

Figure 9-1 X-Ray Survey Utility Screen

9-1
Explorer Technical Manual

The utility screen is broken into three parts:


1. The Main Display shows either a bar-graph or numeric fields for each detector
depending on which Display button is selected.
2. The Display Controls that affect the data shown in the main display. Not all con-
trols are used by the field engineer.
3. The Close button which exits the utility.

The Main Display


The Main Display provides data in two forms depending upon which display button is
selected.

AW-00335_001-0902

Figure 9-2 The Main Display Controls


When the Graph button is active and the system is producing X-Rays, the data is
presented in a bar graph (see Figure 9-3). Scale markings on the bottom of the bar graph
indicate detector numbers. Scale marking along the sides indicate the numeric value of the
column height. A label above the left half of the graph indicates which numeric value for
the column is being displayed (for field engineering purpose this will always be Volts). A
label above the right half of the graph indicates which drum segment is being displayed

9-2
Explorer Technical Manual

(for field engineering purpose this will always be Air Segment). The bar graph shaded
green is hi voltage, the bar graph shaded brown is lo voltage.

AW-00335_001-0903

Figure 9-3 X-Ray Data as a Bar Graph


When the Table button is active and the system is producing X-Rays, numeric data is
presented (see Figure 9-4). A label above the left half of the graph indicates which value is
being displayed (for field engineering purpose this will always be Volts). The display has a
vertical scroll bar on the right because only a sixth of the data for all of the detectors is
shown at one time.

AW-00335_001-0904

Figure 9-4 X-Ray Data as Numeric Data

The Display Controls


The Display Controls can be broken into four functional areas:
1. Controls starting with Xrays (F2) OFF/ON
2. Controls starting with Test Signals
3. Controls starting with Display
4. Controls starting with Status Bits

9-3
Explorer Technical Manual

Controls starting with Xrays

AW-00335_001-0905

Figure 9-5 Controls starting with Xrays


These controls include:

Xrays (F2) button OFF/ON


The default setting of this button is OFF. Clicking this button will produce X-Rays from
the system. Keyboard F2 can also be used (toggles Xrays ON and OFF). When Xrays are
ON warning lights on the scanner will go on and the Main Display will look like either
Figure 9-3 or Figure 9-4 depending on the setting of the Display button (see below).
When the Xray (F2) button is ON the Filter Wheel button is grayed out and cannot be
used.
Warning: While the Xray button is ON the system is producing
X-Rays, take proper precautions.

Filter Wheel button OFF/ON


The default for this button is OFF. The Field Engineer will not use the Filter Wheel button
and it should always remain OFF. This deals with testing the system in different modes.
Note: The next set of controls have the label Filter Wheel Off
Phase which indicates the state of the Filter Wheel. These
controls (HI/LO, BONE/AIR/TISSUE, 0/1/2/3 and 0/1)
indicate where the Filter Wheel will stop when it is not moving.

HI/LO buttons
The default for these buttons is HI. The Field Engineer will not use the HI/LO buttons
and they should always remain the default.

BONE/AIR/TISSUE buttons
The default for these buttons is BONE. The Field Engineer will not use the BONE/AIR/
TISSUE buttons and they should always remain the default.

9-4
Explorer Technical Manual

0/1/2/3 buttons
The default for these buttons is 0. The Field Engineer will not use the 0/1/2/3 buttons and
they should always remain the default.

0/1 buttons
The default for these buttons is 0. The Field Engineer will not use the 0/1 buttons and they
should always remain the default.

Label Information
This area also provides information on the Hi Voltage Setting (either ON or OFF),
Frequency, Filter Errors and HiLo Errors.
Controls starting with Test Signals

AW-00335_001-0906

Figure 9-6 Controls starting with Test Signals


These controls include:

Test Signal button OFF/ON


The default setting of this button is OFF. When this button is ON a test signal is sent to all
of the detectors. Activity will be shown on the main display (see Figure 9-3 of Figure 9-4).
Pulse 0/1/2/3
The default setting of this button is 1. This indicates what data is displayed and not what
the scanner is doing. The Field Engineer will not use the Pulse 0/1/2/3 buttons and they
should always remain the default.

Drum Half 0/1


The default setting for this button is 0. This indicates what data is displayed and not what
the scanner is doing. The Field Engineer will not use the Pulse 0/1 buttons and they should
always remain the default.

9-5
Explorer Technical Manual

Hi Gain
This is a spinner control with a range of 0 to 11. Changing Hi Gain changes the displayed
detector values for the Hi portions of the current segment. The associated actual gain
values are shown in the following table.
Gain Code A/D Gain MUX Gain Total Gain
0 1 1 1
1 1 2 2
2 1 4 4
3 1 8 8
4 2 1 2
5 2 2 4
6 2 4 8
7 2 8 16
8 4 1 4
9 4 2 8
10 4 4 16
11 4 8 32

Lo Gain
This is a spinner control with a range of 0 to 11. Changing Lo Gain changes the displayed
detector values for the Lo portions of the current segment. The associated actual gain
values are shown in the above table.

Xray Mode
This is a spinner control with a range of 0 to 8. The default mode is 3. This control
determines the voltage, current and duty cycle for the x-ray generator. Most scans use X-
ray Mode 3, Whole Body uses 3 and IVA (single energy) uses 6. The following table
shows the full set of defined X-ray modes:
X-Ray Mode Pulse Cycle Peak Ma Average Ma High KV Low KV
0 0 -- -- -- --
1 50% 3 0.75 140 100
2 50% 10 2.5 -- 100
3 50% 10 240 100
4 50% 3 0,75 -- 80
5 50% 3 0.75 140 80
6 100% 10 5.0 140 --
7 50% 10 2.5 140 --
8 100% 10 5.0 -- 100

9-6
Explorer Technical Manual

Aperture
This is a spinner control with a range of 0 to 13. This controls the slit through which the
X-rays pass to limit the dimensions of the X-ray beam. This control allows the selection of
a specific aperture in the system and can be used to test aperture movement. This is not
used on the Explorer.

Average Voltage
This displays the average voltage across all detectors for the Hi Air segment

Controls starting with Display

AW-00335_001-0907

Figure 9-7 Controls starting with Display

Display
This has two buttons: GRAPH and TABLE. This controls what is seen on the Main
Display. In Graph the Main Display shows as bar graph (see Figure 9-3) with scale
markings on the left side according to how Scale is set (see below). It shows the lo (brown
tinted) and hi (green tinted) voltages. In TABLE the Main Display shows the data from all
the detectors as numeric data.

Scale
This has three buttons: RAW, LOG and VOLTS. The Default is VOLTS. The Field
Engineer will not use RAW or LOG and the button should always remain the default
(VOLTS). This control sets up the scale markings on the Main Display when the Display
is set to GRAPH and changes the units for the numeric value wjen the TABLE is
displayed.

Segment
This has three buttons: BONE, TISSUE and AIR. The default is AIR. This controls the
data segment that are displayed in the GRASPH.. The Field Engineer will not use either
BONE or TISSUE and the button should always remain the default (AIR).

9-7
Explorer Technical Manual

Zoom In and Zoom Out


By default, the GARAPH is displayed at minimum resolution When either function
reaches it maximum the button will be grayed out. Zoom In increases the resolution,
meaning that less of the total data range is displayed on the graph but differences in bar
heights are magnified. The horizontal scale isn't affected and all detectors are displayed
regardless of the zoom value. Zoom Out will reverse what Zoom In has done.

Data Current Settings


This reports the Hi and Lo settings of the dark current for Bone, Tissue and Air.

Subtract Dark Current OFF/ON


The default for this button is OFF. Dark voltage is a residual reading for any equipment
that has current. Hologic allows 1 volt to compensate for dark current (see Figure 9-3).
Clicking on this button will remove the dark voltage from the Main Display bar graph.
The Field Engineer will not use Subtract Dark Current and it should always remain the
default.

Controls starting with Status Bits

AW-00335_001-0908

Figure 9-8 Controls starting with Status Bits

Status Bits
Detector values come back from the scanner as sixteen bit numbers, traveling through
some parts of the hardware on sixteen separate wires. The Status Bits field displays the
most recent value for each bit, and color-codes whether that bit has ever changed. Red bits
have not been changed, green bits have. The Reset button sets all the bits to red.

Reset
The Reset button re-starts the process of computing the means and standard deviations for
each filter-segment. It resets the number of samples to zero. It also resets the bits in the
Status Bits display.

Select Detector
This is an edit box that displays specific data about the selected detector. Select Detector
has a default of 108 (the middle of the bar graph as shown in Figure 9-3). This displays
both the currently selected detector (number) and allows a new detector to be selected by
insert a number in the edit box. A new detector can also be selected by dragging the
vertical blue line on the Main Display bar graph (see Figure 9-9). The number displayed in
the edit box will be automatically updated to reflect the line position.

9-8
Explorer Technical Manual

Click and drag this blue


line to change detectors.

AW-00335_001-0909

Figure 9-9 Moving the Detector

Show Details
Show Details defaults to off. When clicked, a small window pops up containing detailed
information about one detector, in tabular form (see Figure 9-10). This information
includes the current (most recent) value, mean value, standard deviation and number of
samples (values used to calculate the mean and standard deviation) for each of the six
filter-segments. The pop-up window can be dragged wherever the user wants and remains
until Show Details is actively released.

Figure 9-10 Show Details Window

SQDRIVER
The SQDRIVER program is accessed from the Explorer Main Menu by selecting
Utilities|Service Utilities|SQDRIVER. This program is used to calibrate the motors
and can be used as an effect tool for troubleshooting motor movement problems.

9-9
Explorer Technical Manual

SQVERIFY
The SQVERIFY program is accessed from the Explorer Main Menu by selecting
Utilities|Service Utilities|SQVERIFY. This program is a diagnostic program, which
can be used to perform a series of low-level tests of system components. Refer to the
SQCHECK User Manual (080-0707) for detailed explanations of its use.

SQKEYPAD
The SQKEYPAD program is accessed from the Explorer Main Menu by selecting
Utilities|Emergency Motion. This program provides the Field Service Engineer with a
method for moving all motors even in conditions where a detected software error
condition would normally prevent any movement of the motors. This program can be used
as a quick means of checking most motor control functions.

9-10
Appendix A
Computer and PCI Controller
Card
QDR for Windows XP® QDR software V12.1 uses a PCI controller card (Com/Con) which
requires a dedicated interrupt. At this time, the Intel D865PERLX motherboard (the X indicates it
has on-board 10/100 networking, documentation will refer to D865PERL) has been validated
with QDR for Windows XP with the Com/Con card in Slot 2, with a modem card in Slot 4, and a
SCSI card in any slot. Please be aware that there are certain issues that need to be addressed up
front if another computer is to be validated.

Computer Specifications
The minimum requirements for V12.2 or higher with a PCI Com/Con are as follows:
• 512 MB memory
• 20 GB hard drive
• Pentium 4 or Celeron CPU without hyperthreading1.
• HAL- Advanced Configuration and Power Interface (ACPI) PC2
• Designed for Windows XP Logo for motherboard and all components other than Hologic
supplied components.
• All Drivers (other than Hologic supplied) must be Windows XP Certified digitally signed.
• CD-RW Teac CD-W552E with firmware 1.05 or later.
• Floppy Drive or SuperDisk3
• InCD 4.0.7.2
• AGP Video card eVGA 64MB, 064-A8-NV74-T2 or other Hologic Approved video
card.4
• Power Supply 300 Watt minimum, with UL/CSA/CE Approval.
• Windows XP Professional with Service Pack 1
• Microsoft patches installed from QDR for Windows XP software disk.

1. Hyperthreading CPU’s have not been tested, they may or may not work. Some Pentium 4’s below 3.0 GHz do
not currently (8/13/03) have hyperthreading. The ones with hyperthreading below 3.0 GHz are designated with
a C, as Pentium 4 2.80C. We will be testing CPU’s with hyperthreading in a future release.
2. HAL is hardware abstraction layer. Most boards will allow you to switch from ACPI (Uniprocessor) to
“Advanced Configuration and Power Interface (ACPI) PC”. You MUST use this HAL. To see if the correct
HAL is intalled, go to Control Panel/System/Hardware/Device Manager and click on the + sign next to the
computer. The HAL is shown there.
3. Superdisks are not supported for writing under QDR for Windows XP, but they can be used for reading old
archives. It is allowed for the SuperDisk and its drivers not to be Windows XP certified.
4. Most newer video cards do not support VGA, which is used for a limited number of functions in QDR for
Windows XP (scan file plot, scan file look, emergency motion, and some service utilities).

A-1
Explorer Technical Manual

To determine which PCI slot the Com/Con board should be located in, you must have the
technical manual for the motherboard. For Intel motherboards, these are supplied on their
website. The Com/Con board uses INT A. Please read and understand the following,
which comes from the D865PERL Technical manual, but is a general: explanation of
interrupt routing.

PCI Interrupt Routing Map


This section describes interrupt sharing and how the interrupt signals are connected
between the PCI bus connectors and onboard PCI devices. The PCI specification specifies
how interrupts can be shared between devices attached to the PCI bus. In most cases, the
small amount of latency added by interrupt sharing does not affect the operation or
throughput of the devices. In some special cases where maximum performance is needed
from a device, a PCI device should not share an interrupt with other PCI devices. Use the
following information to avoid sharing an interrupt with a PCI add-in card.
PCI devices are categorized as follows to specify their interrupt grouping:
• INTA: By default, all add-in cards that require only one interrupt are in this
category. For almost all cards that require more than one interrupt, the first
interrupt on the card is also classified as INTA.
• INTB: Generally, the second interrupt on add-in cards that require two or more
interrupts is classified as INTB. (This is not an absolute requirement.)
• INTC and INTD: Generally, a third interrupt on add-in cards is classified as INTC
and a fourth interrupt is classified as INTD.
The ICH5 has eight programmable interrupt request (PIRQ) input signals. All PCI
interrupt sources either onboard or from a PCI add-in card connect to one of these PIRQ
signals. Some PCI interrupt sources are electrically tied together on the Desktop Board
D865PERL and therefore share the same interrupt. Table 18 shows an example of how the
PIRQ signals are routed.
For example, using Table 18 (see page A-3) as a reference, assume an add-on card using
INTA is plugged into PCI bus connector 3. In PCI bus connector 3, INTA is connected to
PIRQE, which is already connected to the ICH5 audio controller. The add-on card in PCI
bus connector 3 now shares an interrupt with the onboard interrupt source.
From the Table below (which is for the D865PERL), you can see that on most PCI slots
(connectors), INT A is shared with other devices. Only in Slot 2, is INT A not shared with
other devices. Howerver, if a PCI card that required two interrupts was put in PCI slot 1,
it might cause INT A to be shared and this could cause problems.

A-2
Explorer Technical Manual

To check that the Com/Con has a unique interrupt, go to Control Panel/System/Hardware/


Device Manager and select View/Resources by Connection. You will obtain a screen that
should look something like the following:

A-3
Explorer Technical Manual

You see that the COM/CON board (RTX PnP) is the only thing using INT 10.
In the Motherboard BIOS, you should reserve INT 10 for the Slot used. Note that this will
not guarantee that the COM/CON card will be assigned that INT under Windows XP, but
it should be done anyway.
Set up Windows XP with the following options:
DISPLAY: Set desktop and screen saver to None for all active accounts.
Set Windows and Buttons to Windows XP Style
Set Color scheme to Default(Blue)
Set Font size to Normal.
Set Screen Resolution to 1024 X 768.Color (32 bit).
Set Color quality to Highest( 32 Bit)
Set Hardware Accelerator to Full
Set Screen refresh rate to 75 or higher.

A-4
Explorer Technical Manual

NETWORK: Enable Client for Microsoft Network.


Enable File and Printer Sharing for Microsoft Networks.
Enable QoS Packet Scheduler
Enable Internet Protocol (TCP/IP)
Turn OFF Automatic Updating
USERS AND Set Admin logon in Administrators Group and password as
GROUPS password.
Set Field Service logon in Administrators Group and password
as password.
Set QDR to Users logon with no password.
These are the only logons on the system
Built In Shall not be used and no password assigned.
Administrator
account
To test that the video card works correctly, try doing scan file plot and scan file look from
QDR for Windows XP Utility menu.

A-5
Explorer Technical Manual

A-6
Appendix B
Windows XP ® Setup
Instructions
1. WELCOME SCREEN
1.1. Log onto the computer using the Administrator account, the password is password
(all small letters). Click the Arrow icon.
1.2. Select Start Control Panel.
1.3. Click Switch to Classic View if not selected in the left pane of the window.
1.4. Click User Accounts.
1.5. Click Change the way users log on or off.
1.6. Ensure “Use the welcome screen” check box is checked.
1.7. Ensure “Use fast User Switching” is Unchecked.
1.8. Check Apply Options.
1.9. Click on the X in the upper right corner to close all open windows and return to the
main window.

2. EQUIPMENT CHECK
2.1. Log onto the computer using an Administrator account. The password is password
(all small letters).
2.2. Select Start My Computer and perform the following checks.
• A: Drive is 3.5
• D: Drive is CD-ROM
If the CD-ROM drive is incorrect proceed to the next step, otherwise close all open windows back
to the Windows desktop then go to the next Section.
CD ROM Drive Correction
• Click on the X in the upper right corner to close all open windows and return to the desk-
top.
• Select Start Control Panel Administrative tools Computer Management Icon
Storage Disk Management.
• Right click the CD-ROM and select Change Drive Letter and Paths.

B-1
Explorer Technical Manual

• Click the Change button.


• Click the down arrow next to the Drive letter and select the D letter to set the
drive to D.
• Click OK OK
• Click the X in the upper right corner of all open windows to return to the desktop.

3. Task Scheduler
3.1. Select START All Programs Accessories System Tools
Scheduled Tasks.
3.2. Select the Advanced pull down menu on the tool bar.
• If “Stop Using Task Scheduler” is shown click Stop Using Task Sched-
uler. The option should read “Start Using Task Scheduler.”
• Click on the X in the upper right corner to close Scheduler Tasks.

4. Fast User Switching and Task scheduler


4.1. Click Start Control panel Administrative Tools Services.
4.2. Select Fast User Switching Capability Right click Fast User Switching
Capability Properties.
4.3. Set Startup Type to Disabled APPLY _ OK.
4.4. Select Task Scheduler Right button click Task Scheduler Properties.
4.5. Set Start Type to Disabled Apply OK.
4.6. Click the X in the upper right corner of the screen to close all open windows.

5. Create user accounts


5.1. Click START Control Panel User Accounts.
5.2. New Account
• Click Create a new account.
• Type Name of new account Next.
• At the Pick an account type select the following:
• ADMIN Account select Computer Administrator.
• FIELD SERVICE Account select Computer Administrator.
• QDR Account select Limited.
• The QDR account will not have a password.
• To create a password for the ADMIN or FIELD SERVICE account
Double click the account that was just created.

B-2
Explorer Technical Manual

• Click Create a Password.


• Password for ADMIN or FIELD SERVICE account is password.
When completed click Create password.
5.3. Change/Create Password
• Double click the user account to which you want to add a password to.
• Click Change Password or Create Password.
• Follow the instructions on the screen for the password.
5.4. Click the X in the upper right corner of the screen to close all open windows.
5.5. If new accounts have been created or changes have been made to an existing
account repeat Section 5.

6. CD and Floppy Write Test


6.1. Insert a 3.5 in floppy into the drive.
6.2. Insert a formatted 24X Read/Write CD into the CD drive.
6.3. Click Start My Computer 3.5 In Floppy and open a window. Size the
window to be less than one half of the screen.
6.4. Click Start My Computer CD-ROM and open a window. Size the
window to be less than one half of the screen.
6.5. Move any file on the floppy to the CD.
Note: To move a file, select the file with the mouse then depress and hold the
shift key then drag and drop the file.
6.6. Move the file from the CD back to the floppy.
6.7. Close the two windows that were created by clicking the X in the upper right
corner.
6.8. Remove the floppy from the drive
6.9. Select Start MY Computer.
6.10. Press the button on the CD drive to eject the CD.
6.11. Remove the CD from the drive then close the drive door.
6.12. From MY Computer window click the X in the upper right corner to close the
window.

7. DATE AND TIME SETTINGS


7.1. Select Start Control Panel.
7.2. Click the Date/Time Icon.
7.3. Click the Date & Time Tab.

B-3
Explorer Technical Manual

• Set the Month, Date & Time if not correct.


7.4. Click the Time Zone Tab.
• Ensure that the Automatic adjust clock for daylight savings time changes is
checked.
• Click the pulldown window and then select the appropriate time zone.
• Click Apply OK.
7.5. Click The X in the Control Panel Window to close the Control Panel and return
to the windows desktop.

8. Date Format
8.1. Select Start Control Panel Regional and Language Icon
Customize Date Tab.
8.2. Set Short date format to MM/dd/yyyy and the Date Separator to “/”.
Click Apply OK
8.3. From the “Regional and Language Options” screen click Apply OK
8.4. Click The X in the Control Panel Window to close the Control Panel and return
to the windows desktop.

9. AUTOPLAY
9.1. From the window desktop select Start Run.
9.2. Enter gpedit.msc then click OK.
9.3. Select Computer Configuration Administrative Template System.
9.4. Double click Turn off Autoplay.
9.5. Select the Enabled radio button
9.6. Set “Turn off Autoplay on:” to All Drives.
9.7. Click Apply OK.
9.8. Select X in upper right corner to close screen.

10. DISPLAY SETTINGS and Hardware Accelerator


10.1. Select Start Control Panel Display.
10.2. Select Screen Saver Tab.
• Set Screen Saver to None if not set.
Click Apply if a change was required.
• Click Power button.
• (This step only applies when Logged On as an Administrator).

B-4
Explorer Technical Manual

• Set “Turn off monitor” to Never (This cannot be set whenever logged on as
QDR).
10.3. Select Screen Saver Tab.
• Set “System standby” to Never.
• Set “System hibernates” to Never.
• If changes were made click Apply OK otherwise click Cancel.
10.4. Select Desktop Tab
• Set Desktop to None if not set.
Click Apply if a change was required.
10.5. Select Settings Tab.
• Set Screen Resolution to 1024 X 768 if not set.
• Set Color Quality to Highest (32Bit) if not set.
Click Apply if a change was required.
10.6. (This step can only be performed when Logged On as an Administrator) Select
Monitor Tab.
• Set the Screen refresh rate to 75.
Click Apply, if a change was required.
• Click OK.
10.7. Select Troubleshoot Tab.
(This step can only be performed when Logged On as an Administrator)
• Set Hardware Accelerator to Full.
Click Apply, if a change was required.
• Click OK.
10.8. From the Display Properties screen select Apply OK.
10.9. Click X in the upper right corner of all screens to return to the Windows
desktop..

11. InCD properties


11.1. Right button click the InCD Icon on the bottom right tool bar and select About.
11.2. Insure the InCD Version is 4.x.x.x or higher.-
11.3. Click the X in the upper right corner of all windows to return to the Windows
desktop.
11.4. Right click the InCD Icon on the bottom right tool bar and select Options.
11.5. Insure the following boxes are set properly.
• Launch InCD when blank disc inserted. Checked

B-5
Explorer Technical Manual

• Hide tray icon. Unchecked


• MRW Format on all MRW capable drives. Checked
• Do not show Format complete message. Unchecked
11.6. When completed click Apply.

12. Disable System Updates


12.1. Click Start Control Panel System Automatic Updates Tab.
12.2. Uncheck the Keep my computer up to date checkbox.
12.3. Click Apply OK.
12.4. Click the X in the upper right corner of all windows to return to the Windows
desktop.

13. NETWORK COMPONENTS


13.1. From the Windows XP desktop select START Control Panel.
13.2. From the Control Panel window select Network Connections Icon.
13.3. Right button click the Local Area Connection then select Properties.
13.4. Insure the check box is checked for each of the following Network
Components.
• Client for Microsoft Networks
• File and Printer Sharing for Microsoft Networks.
• QoS Packet Scheduler.
• Internet Protocol (TCP/IP).
13.5. Select Internet Protocol(TCP/IP).
13.6. Click the Properties button.
13.7. Select the following radio buttons.
• Radio button “Obtain an IP address automatically”.
• Radio button “Obtain an DSN address automatically”.
13.8. Click OK to close the Internet Protocol (TCP/IP) Properties window.
13.9. Click OK to close the Local Area Connection Properties window.
13.10. Click X to close the Network Connections Properties window.
13.11. Click the X in the upper right corner of the Network Connections window to
return to the Windows desktop.

14. HOME PAGE SETUP


14.1. On the Windows XP desktop, Select Start Control Panel.

B-6
Explorer Technical Manual

14.2. From the Control panel screen open the Internet Options ICON
14.3. Set the Home Page to http://www.hologic.com if it is not presently set, then
select Apply, then select OK.

15. Hardware Allocation layer (HAL)


15.1. Select Start Control panel System Icon Hardware Tab Device
Manager.
15.2. Click the + sign next to the Computer.
15.3. If the HAL listed is “Advanced Configuration and Power Interface (ACPI)
PC” go to the next section otherwise execute the following steps.
CAUTION: Do Not perform the following steps unless Slot 2 has been set to
IRQ 10 in the Bios.
• Right button click the HAL and select Update Driver.
• Select “Install from a list of specific locations (Advanced). Click Next.
• Select “Don’t Search. I will choose the driver to install” radio button. Click
Next.
• Select the “Advanced Configuration and Power Interface (ACPI) PC” Hal.
• Click Next – Finish.
• At the system settings window select Yes to reboot the computer.
15.4. Click the X in the upper right corner of all windows to return to the Windows
desktop.

16. Windows XP activation ( Reactivation if Required )


Note: This section only needs to be executed if the XP software was not
previously activated.
16.1. Connect one end of the network cable to the network jack and plug the other
end of the cable into the Network Adapter Card in the computer.
16.2. Note that one or more of the LEDs on the rear of the card is illuminated.
16.3. Start Activation from one of the following methods.
• Select activate Icon on bottom right Taskbar.
• Select Start All Programs Activate Windows.
16.4. Select “Yes, let activate windows over the Internet now” radio button.
16.5. Select Next
16.6. Select “No, I don’t want to register now” radio button.
16.7. Select Next

B-7
Explorer Technical Manual

16.8. When the message “You successfully activated your copy of Windows”
appears select OK.

17. QDR AND ADMIN Settings


17.1. Click Start Log Off.
17.2. At the Log Off Windows click Log Off.
17.3. Select the Admin Icon to Log onto the computer.
17.4. Perform section 8, 10 and 11 for the Admin logon
17.5. Click Start Log Off.
17.6. At the Log Off Windows click Log Off.
17.7. Click Start Log Off.
17.8. At the Log Off Windows click Log Off.
17.9. Select the QDR Icon to Log onto the computer.
17.10. Perform section 8, 10 and 11 for the QDR logon
17.11. Click X in the upper right corner of all screens to return to the Windows
Desktop.
17.12. Click Start Turn Off Computer.
17.13. At the Turn Off Computer Window click Turn Off.

18. RTX Software Load


18.1. Insert the VenturCom CD into the CD-ROM drive.
18.2. Click Start Run
18.3. Enter D: then Click Browse.
18.4. Select “RTX 5/1/2 Runtime.exe” Open OK.
18.5. From the “Welcome” window click Install.
18.6. From the “Software License Agreement” window click Yes.
18.7. From the “VenturCom RTX5.1.1” setup window enter the following then click
Next.
Name = QDR.
Company = Hologic.
18.8. At the “Product Registration” screen enter the following then click Next.
PAC # = 5012-PAC-39452-84635-89170
18.9. From the “Choose Destination Location” window insure the Destination
Folder is C:\Program Files\VenturCom\RTX then click Next.

B-8
Explorer Technical Manual

18.10. From the “Select Program Folder” window insure the program folder is
“VentureCom RTX” then select Next.
18.11. From the “Start Copying Files” window insure that the following will be
copied as shown below, then click Next.
18.12. RTX Runtime Environment
18.13. RTX Runtime installation directory
C:\Program Files\ VentureCom\RTX
18.14. From the “Setup Complete Window” select the “Yes, I want to restart my
computer now” radio button.
18.15. Remove the CD-ROM from the drive then click Finish.
18.16. After the compute reboots select Start All Programs VenturCom RTX
RTX Properties.
18.17. Select Control Tab then Start RTX. Click Apply, if availiable.
18.18. Select the SettingsTab.
18.19. Select the Boot radio button then Apply OK.
18.20. Select Start Turn Off Computer Restart.
18.21. After the computer reboots select Start All Programs VenturCom RTX
Properties.
18.22. Select Control Tab and insure all four components are running then click
Cancel.
If they are not running repeat all steps from 18.15 through this step..
18.23. Click Start Turn Off Computer Turn Off.
18.24. After the computer shuts OFF. Install the PCI Communication Card 140-0124
into the appropriate PCI Slot of the computer (see Appendix A).
18.25. Turn the computer ON.

19. PCI Com Con Card and Driver Setup


19.1. From the New Hardware Wizard Window select Install from a list of
specified location (Advanced) radio button then click Next.
19.2. From the Found New Hardware Wizard window select Don’t search. I will
choose the driver to install radio button. Click Next.
19.3. Scroll down and select RTX Drivers then click Next.
19.4. Select the RTX PnP and Power Management Device then click Next.
19.5. From the Update Driver Warning Window click Yes.
19.6. From the Completing the Found New Hardware Wizard window click Finish.

B-9
Explorer Technical Manual

20. IRQ Setup


20.1. Click Start All Programs VenturCom RTX RTX Properties
Plug & Play Tab
20.2. Click + sign nest to RTX to expand.
20.3. Right button click RTX PnP and Power Management Device and select
Properties.
20.4. Check the following settings.
Slot Number: 2
Disposition: Device Exclusive
Status: RTX PnP Device is setup properly.
20.5. IRQ Setting.
20.6. If the IRQ is 10, click Cancel Cancel
20.7. If the IRQ is not 10, change the IRQ to 10 and click OK.
20.8. Click Apply OK.
20.9. Select Start Turn off Computer Restart to reboot the computer.
20.10. Repeat this section to check if the settings were made.

Software Loading Instructions


Load the HolxRtss & Permissions and QDR V12.X
1. After booting the PC, log on as Admin using password as the password.
2. Place the V12.X QDR Software CDROM in the CDROM drive.
3. Select Start|Run… and Browse the D: drive (the CDROM drive may be drive letter
E: on some systems).
4. Double-click the Drivers folder.
5. Double-click the HolxRtss&Permission folder.
6. Click on Setup.exe and then click the Open button.
7. Click the OK button and then click the OK button again.
8. DO NOT CHANGE the User Name from QDR. Click the Next button.
9. Select the proper configuration for the install, System for a scanner.
10. Follow the on-screen instructions to load the Hologic RTSS and
Permissions.
11. The PC will reboot, logon as QDR (user).

B-10
Explorer Technical Manual

12. Select Start|Run… and type in D:\setup.exe (the CDROM drive may be drive letter
E: on some systems) and click the OK button.
13. Select Admin from the dropdown list, using password as the password and click the
OK button.
14. Click the Next button at the Welcome window.
15. Click the OK button.
16. Click the Yes button at the Software License Agreement window or click OK
button.
17. Select Explorer at the Select QDR Machine Type window and then click the OK
button.
18. At the Serial Number window, enter the 5-digit instrument serial number and click
the Next button.
19. At the Choose Destination Location window, click the Next button.
20. The QDR for Windows software will now be loaded.
21. Once software loading is complete, the Setup Complete window will appear asking if
you wish to restart your PC now or later. Click the Finish button.
22. The PC will now reboot.

B-11
Explorer Technical Manual

B-12

You might also like