GM60 - 05 - MA - 1122R0 (140704) - Maintenance
GM60 - 05 - MA - 1122R0 (140704) - Maintenance
GM60 - 05 - MA - 1122R0 (140704) - Maintenance
MA-1122R0
Maintenance
GM60A Series
GM60A Series
Maintenance
REVISION HISTORY
This Document is the English original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.
MA-1122R0
GM60A Series
Maintenance
TABLE OF CONTENTS
Section Page
1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
MA-1122R0 i
GM60A Series
Maintenance
ii MA-1122R0
GM60A Series
Maintenance
SECTION 1 INTRODUCTION
The following checks and maintenance procedures together with the suggested intervals,
are the manufacturer’s recommendation for the most effective Periodic Maintenance
schedule for this unit.
The first Periodic Maintenance Service should be performed six (6) months after
installation, and the subsequent services every twelve (12) months. Periodic
Maintenance Service depends on the working load of the Generator and X-ray Tube.
MA-1122R0 1
GM60A Series
Maintenance
The following test equipment may be required for carrying the procedures
detailed herein:
Keyboard and Mouse to connect to the Unit to access the GSM program.
Digital Multimeter.
Calculator.
2 MA-1122R0
GM60A Series
Maintenance
Before performing the periodic maintenance, verify that the Unit has
been connected to the mains for at least ten (10) hours. This will ensure
Note . a complete and full Battery charging.
MA-1122R0 3
GM60A Series
Maintenance
Note . For further information about following steps refer to its respective
Chapter and Section.
General cleaning:
External Covers and Control Console cleaning, touching up of painted surfaces as required.
1 - -
Internal inspection of major components, vacuum cleaning when necessary. -
Cable checks: Electrical connectors, wires and ground lead interconnections checking, Power
2
Line and Detector Cables Rewind Mechanism cleaning and lubrication. -
Motion Assemblies checking, lubrication, motor motion, operation of clutch, gauges, handle,
brakes, microswitches, bumper and wheels. -
4 MA-1122R0
GM60A Series
Maintenance
Clean the equipment frequently, particularly if corroding chemicals are present. Clean
When it is needed to disinfect the Control Console, clean it with a cloth impregnated
with isopropyl alcohol.
Do not apply directly any liquid on the screen or other surfaces, nor use cleaners
containing bleach, ammonia or any other abrasive or solvent liquid, it could cause
damage to the equipment.
Remove the external covers from the unit. Visually inspect all major components for
dust or foreign items. Search carefully to detect objects which might cause short
circuits and for loose connections.
If excess dust is present, clean the interior of the unit using a dry brush or vacuum
cleaner.
Note . Do not re-install the external covers of the Unit until the Periodic
Maintenance process has been completed.
MA-1122R0 5
GM60A Series
Maintenance
1. Check that all electrical connections are firm and secure and that all cable clamps
and strain reliefs are in place. Also check that connectors do not have exposed
wire-veins and check cable sheaths for wear and fraying.
3. The central reference ground is located at the Generator. Check the ground
lead interconnections continuity using a multimeter at its lowest ohms range.
4. Verify the Rewind Mechanism of the Power Line Cable and Detector Cable.
Clean the cables and lubricate the rewind mechanism if necessary.
Ensure that:
Ensure that:
6 MA-1122R0
GM60A Series
Maintenance
Within the recommended operating temperature, under optimum float conditions and
depending upon the working load of the Unit, batteries service life expectancy could
last for 2 - 5 years.
If the unit has not been used or it has been stored for six months, it
should be installed and/or energized to prevent deep discharge of the
batteries. A deep discharge will cause permanent damage to the
batteries.
For battery maintenance, the Service Engineer must be familiar with the technical
specifications for the batteries and related circuitry and must also understand how the
indicators and charge cycles function (refer to Section “Information on Indicators and
Charge Cycles” in the Troubleshooting Chapter).
MA-1122R0 7
GM60A Series
Maintenance
Check the proper connection and condition of the cables connected to the Console.
Check correct operation of the Console by performing the following test:
2. Touch different points on the touch screen (in the operator applications) to
check for proper calibrated. If necessary, perform the procedure described in
the section entitled “Touch Screen Sensor Calibration” in the Troubleshooting
Chapter of the Service Manual.
3. Check the Handswitch condition. Verify that the Handswitch cable and its
connection to the Unit are in good condition.
8 MA-1122R0
GM60A Series
Maintenance
Note . Test points on the HT Controller Board can be used to monitor (with a
scope) the kV and mA readings but should not be used for calibration:
kV test point is TP-7 and the scale factor is 1 volt = 33.3 kV (0.3
volt = 10 kV)
Verify position of dip switches on the HT Controller Board during every test
procedure:
3000SW2-4 Position during operation -- Digital mA Loop Closed Digital mA Loop Open / Filament Current Constant
MA-1122R0 9
GM60A Series
Maintenance
G Place and center a Non-Invasive kVp Meter on the X-ray Tube output
at the required SID (refer to the Non-Invasive kVp Meter
documentation).
2. Turn the Unit ON and select the “Direct” (No Bucky) workstation.
3. Select 80 kV, 200 mA (or the first mA station for Large Focus), 100 ms.
Make an exposure and note the kV at the end of the exposure.
If the kV value does not coincide with the above value, perform the
respective Calibration procedures.
5. Turn the Unit OFF and remove the Non-Invasive kVp Meter.
2. Turn the Unit ON and select the “Direct” (No Bucky) workstation.
10 MA-1122R0
GM60A Series
Maintenance
3. Select 80 kV and the following mA stations. Make an exposure and note the
mAs values reading on the mAs Meter.
mAs value
4. Verify that the mAs values readings on the mAs Meter are the same mAs
displayed on the Console with a tolerance of ± 6% mAs.
If the mAs values do not comply with the above values, perform the
respective Auto-Calibration procedures.
6. Select the following parameters, make an exposure and note the mAs values
on the mAs Meter for:
mAs value
7. Verify that the mAs values readings on the mAs Meter are the same mAs
displayed on the Console with a tolerance of ± 4% mAs.
If the mAs values do not comply with the above values, perform the
respective Auto-Calibration procedures.
MA-1122R0 11
GM60A Series
Maintenance
1. Check correct operation of the controls, detents, limits and move to all
possible positions. Movements must be balanced, noiseless and smooth.
3. Lubricate the Column tracks with a cloth with lubricant oil. In case of noise
during movement, clean and lubricate again (vertical movement).
4. Remove the Upper and Rear Column Cover and verify that the Steel Cable
and Pulleys are in good condition. Replace the Steel Cable if it is damaged /
frayed.
Note . It is recommended to replace the Steel Cable every 5 years. For further
information, refer to the Troubleshooting Chapter in this Service
Manual.
1. Check lubrication and motor motion, check the clutch operation, and check
both motion gauges by performing the Functional Check from the section
entitled “Motor Group -- Functional Check” in the Troubleshooting Chapter of
the Service Manual.
4. Clean the front wheels and check that both run freely.
12 MA-1122R0
GM60A Series
Maintenance
Note . Local Standards may require checking procedures that are not
considered Periodic Maintenance by the manufacturer.
Below are described some of these procedures and the
manufacturer recommendations to perform them.
Besides the standard tools and test equipment, this procedure requires the
following:
Isolation Transformer.
Voltmeter.
1. Unplug the Unit from the mains. Remove the Ground wire of the AC plug.
Remove any external Ground connection to the Mobile.
2. Place the Mobile on a surface that is isolated from ground such as wood or non-
conductive plastic).
4. Turn ON the Mobile. Connect the Leakage Current Test Tool as follows:
Illustration 3-1
Leakage Current Test Connection
Ground of
Electrical Cabinet
MA-1122R0 13
GM60A Series
Maintenance
5. As shown in Illustration 3-1, the Voltmeter is set to VAC and connected to the
Test Tool at the OUT points. Connect Ø (IN) to a metallic part of the Mobile
Unit. Connect GND (IN) to the Ground of the Electrical Cabinet and to an
output of the secondary of the Isolation Transformer (B).
6. The reading from the Voltmeter (1V = 1mA) gives the leakage current under
normal conditions. Turn OFF the Mobile and disconnect the connection
between the secondary of the Isolation Transformer (B) (or with a switch as
identified in the illustration). The measurement in the voltmeter gives the
leakage current under fault conditions.
7. Swap the wires connected from the Isolation Transformer to the Mobile
(marked as “A” and “B” in the illustration) and repeat the measurement
specified before (step 6.).
9. Disconnect all connections made in the previous steps and return the Mobile
to operational wiring. Place the ground wire back to the plug and ensure a firm
connection.
14 MA-1122R0
GM60A Series
Maintenance
Besides the standard service tools, a Radiation Meter (Victoreen 450 P or similar) is
necessary for checking Radiation Leakage. Radiation measurement will be setup as
indicated in Illustration 3-2.
Illustration 3-2
Radiation Leakage Setup
r=1m
X--RAY TUBE
Focal Spot
COLLIMATOR
Lead Plate (2.5 mm)
3. Select Dose mode (0.00 R) on the Radiation Meter (Victoreen 450 P).
4. Make an exposure and note the radiation readout, indicating the position where
the probe is placed respect to the Tube-Collimator. Repeat this step with the
probe at various positions around the Tube-Collimator (maintaining a radius
of 1 meter). Refer to Illustration 3-2.
MA-1122R0 15
GM60A Series
Maintenance
b. The equivalent current in continuous operating mode for the duty cycle
(12 exposures at maximum energy, 115 kVp -- 100 mAs during one
hour) of the Mobile Unit is 0.19 mA at 115 kVp.
2mR∕s 2mR∕s
A= = = 0.019mR∕s
20mA 105.26
0.19mA
6. Based on the radiation reading at any position, ensure that the Radiation
Leakage is lower than 115 mR/h (1 mGy/h). Note the position and value of the
maximum radiation reading.
If the Radiation Leakage is over the limit at any position, first replace the
Collimator and then repeat this procedure. If leakage is still present, replace
the X-ray Tube and repeat this section.
When finished, remove the Lead Plate from the Collimator window.
16 MA-1122R0
GM60A Series
Maintenance
VDRs (Voltage Dependent Resistor) connected between mains and earth avoid damage
on the equipment due to overvoltage.
“ 20 Dielectric strength:
Only insulation with a safety function need be subject to testing”.
“20.4 Tests:
*j) Power-consuming voltage-limiting devices, in parallel with an insulation
to be tested, are disconnected from the earthed side of the circuit. Lamps,
electronic tubes, semiconductors or other automatic regulating devices may be
removed or rendered inoperative if necessary to carry out the test”.
In order to conduct the Insulation Resistance Test, 500 VDC have to be applied to the
unit. The VDRs will work decreasing their resistance, avoiding that the overvoltage
goes to the rest of the circuits. Therefore, the VDRs have to be removed before
applying 500 VDC in order to measure the correct insulation resistance.
MA-1122R0 17
GM60A Series
Maintenance
18 MA-1122R0