MenAMI Silitrace 02
MenAMI Silitrace 02
MenAMI Silitrace 02
Operator’s Manual
A-96.250.811 / 091017
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: support@swan.ch
Document Status
ID: A-96.250.811
Revision Issue
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1. Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2. Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.3. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.1. Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.3. Connect Sample and Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3.1 AMI Silitrace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3.2 AMI Silitrace Dual-Stream . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.4.1 Connection Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.4.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.5. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.5.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.5.2 Alarm Relay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.5.3 Relay 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6.1 Signal Output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . 26
3.7. Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.7.1 Signal Output 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.7.2 Profibus, Modbus Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.7.3 HART Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.7.4 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.1. Start-up Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.2 Prepare Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.3. Prepare Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.4. Switch on Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.5. Adjust Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.6. Activate the Peristaltic pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.7. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.7.1 AMI Sample Sequencer (Option) . . . . . . . . . . . . . . . . . . . . . . . 36
4.8. Final Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
A-96.250.811 / 091017 1
AMI Silitrace
5. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 44
5.5. Grab Sample Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.1. Maintenance Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2. Stop of Operation for Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 47
6.3. Refill or Replace Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.4. Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.5. Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.6. Zero. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.7. Replace the Pump Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6.8. Fill System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.9. Clean the Photometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.10. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
7.1. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
7.2. Replace the reaction chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
7.3. Replace the 6-Way Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
7.4. Replace the Cuvette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
7.5. Replace the Reagent Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
7.6. Cleaning the solenoid valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
7.7. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
9. Program List and Explanations . . . . . . . . . . . . . . . . . . . . . . . . . 79
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
2 A-96.250.811 / 091017
AMI Silitrace
A-96.250.811 / 091017 3
AMI Silitrace
Safety Instructions
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target Operator: Qualified person who uses the equipment for its intended
audience purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification, To be qualified for instrument installation and operation, you must:
Training read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
4 A-96.250.811 / 091017
AMI Silitrace
Safety Instructions
DANGER
WARNING
CAUTION
Safety gloves
A-96.250.811 / 091017 5
AMI Silitrace
Safety Instructions
Corrosive
Harmful to health
Flammable
Warning general
Attention general
6 A-96.250.811 / 091017
AMI Silitrace
Safety Instructions
WARNING
WARNING
WARNING
Download The current Safety Data Sheets (SDS) for the above listed Re-
MSDS agents are available for downloading at www.swan.ch.
A-96.250.811 / 091017 7
AMI Silitrace
Product Description
2. Product Description
8 A-96.250.811 / 091017
AMI Silitrace
Product Description
A-96.250.811 / 091017 9
AMI Silitrace
Product Description
Fluidic The sample enters at the sample inlet [N] and flows through the
flow meter [P], where the flow rate is measured, to the 6-way valve
[I]. A part of the sample overflows into the sample outlet [O].
The peristaltic pump [S] sucks the sample from the 6-way valve and
pumps it into the reaction chamber [D]. In the reaction chamber a
coil is wrapped around a heating device. It has 4 inlets to allow dos-
ing of the necessary reagents.
In the reaction chamber the sample is preheated to 45°C, eliminat-
ing temperature deviations of the sample. In a first step Ammoni-
ummolybdate [J] and sulfuric acid [K] are added, to form the
yellowish molybdosilic acid and molybdophosphoric acid.
Afterwards the oxalic acid [L] is added to mask the molybdophos-
phoric complex. At last the reducing agent ammonium ferrous (II)
sulfate [M] is added. The color of the sample turns to blue. As silica
is only present in trace amounts, the blue color can not be seen.
Afterwards the colored sample flows into the thermostatic cuvette
[E] until it is filled completely. Now the intensity of the color is mea-
sured at 815 nm. The color development is proportional to the silica
concentration in the sample.reaction chamber.
As the level of the sample in the cuvette increases, the sample
overflows into the siphon tube [G] and eventually the cuvette is
drained spontaneously through the siphon tube. The sample is led
via the aeration and drain tube [T] into the sample outlet [O].
Dosing, mixing and filling of the photometer are determined by the
rotating speed of the peristaltic pump [S]. This speed is adjusted
automatically and guarantees the correct timing of the measure-
ment.
The solenoid valve [R] is used for the zero measurement. It switch-
es off the supply of reagent 1, which is responsible for coloring the
sample. A zero can be started automatically or manually.
For a calibration or a verification the standard [B] is used. Both, cal-
ibration and verification can be started automatically or manually.
When a calibration or a verification is started, the 6-way valve is ro-
tated to position 5 and standard is pumped through the photometer.
NOTICE: The fluidic scheme shows the single-stream variant of
the AMI Silitrace. The dual-stream variant is fitted with a
different flow cell block and a channel selector valve (see AMI
Silitrace Dual-Stream, p. 34).
10 A-96.250.811 / 091017
AMI Silitrace
Product Description
T B
05 13
C
D
S
12
1B
2 03
03
3 3
4 4
04 04
05 I 12
12
4
5 3 F
1B 7
6 2 11
1 11
02
R G
01
04
04
1A H
03 1A 2 3 4
E
02 1 2 3 4 11
1 2 01 3 4
Q
J
P K
L
O M
N
A-96.250.811 / 091017 11
AMI Silitrace
Product Description
12 A-96.250.811 / 091017
AMI Silitrace
Product Description
400 mm / 15¾”
13 mm / ½”
374 mm / 14¾”
13 mm / ½”
Exit Enter
AMI SILITRACE
OzoneSiO
Trace Monitor
2
Process Analyzer
824 mm / 32 7/16”
850 mm / 33½”
4 x dia. 10 mm / /”
2L 2L 2L 2L
A-96.250.811 / 091017 13
AMI Silitrace
Product Description
Q C
P
F
O
H
N
I
M
L
K J
14 A-96.250.811 / 091017
AMI Silitrace
Installation
3. Installation
A-96.250.811 / 091017 15
AMI Silitrace
Installation
16 A-96.250.811 / 091017
AMI Silitrace
Installation
2L 2L 2L 2L
OXYCON ON-LINE SILCA OXYCON ON-LINE SILCA OXYCON ON-LINE SILCA OXYCON ON-LINE SILCA
AMI SILICA REAGENT 3 AMI SILICA REAGENT 4
B
AMI SILICA REAGENT 1 AMI SILICA REAGENT 2
C
D
E
F
Sample inlet Use a plastic tube (FEP, PA, or PE 4 x6 mm) to connect the sample.
1 Slide the knurled nut [C] and the compression ferrule [D] over
the plastic tube.
2 Push the plastic tube into the screw connection at the sample
inlet.
3 Tighten the knurled nut well.
Waste Push the 1/2” tube [F] over the hose nozzle [B] and place it into a
pressure free drain of sufficient capacity.
A-96.250.811 / 091017 17
AMI Silitrace
Installation
2L 2L 2L 2L
OXYCON ON-LINE SILCA OXYCON ON-LINE SILCA OXYCON ON-LINE SILCA OXYCON ON-LINE SILCA
AMI SILICA REAGENT 1 AMI SILICA REAGENT 2 AMI SILICA REAGENT 3 AMI SILICA REAGENT 4
A1
A2
B
C
D
E
F
A1 Sample inlet channel 1 D Compression ferrule
A2 Sample inlet channel 2 E Holder
B Hose nozzle F 1/2” tube
C Knurled nut G Channel selector valve
Sample inlet Use two plastic tubes (FEP, PA, or PE 4 x6 mm) to connect the sam-
ple streams.
1 Slide the knurled nut [C] and the compression ferrule [D] over
the plastic tube.
2 Push the plastic tube into the screw connection at the sample
inlet.
3 Tighten the knurled nut well.
Waste Push the 1/2” tube [F] over the hose nozzle [B] and place it into a
pressure free drain of sufficient capacity.
18 A-96.250.811 / 091017
AMI Silitrace
Installation
WARNING
A B C
A-96.250.811 / 091017 19
AMI Silitrace
Installation
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
– relay 1
– relay 2
– alarm relay
WARNING
WARNING
20 A-96.250.811 / 091017
AMI Silitrace
Installation
CAUTION
Use only the terminals shown in this diagram, and only for the
mentioned purpose. Use of any other terminals will cause short
circuits with possible corresponding consequences to material
and personnel.
A-96.250.811 / 091017 21
AMI Silitrace
Installation
A
B
C
22 A-96.250.811 / 091017
AMI Silitrace
Installation
3.5.1 Input
NOTICE: Use only potential-free (dry) contacts.
The total resistance (sum of cable resistance and resistance of
the relay contact) must be less than 50 Ω.
Terminals 16/42
For programming see “Program List and Explanation” 5.3.4, p. 102.
1) usual use
A-96.250.811 / 091017 23
AMI Silitrace
Installation
Relay Jumper
config. Terminals pos. Description Relay configuration
Normally 6/7: Relay 1 Inactive (opened) during
Open 8/9: Relay 2 normal operation and loss 6
of power.
Active (closed) when a pro-
0V
grammed function is
7
executed.
Normally 6/7: Relay 1 Inactive (closed) during
Closed 8/9: Relay 2 normal operation and loss 6
of power.
Active (opened) when a
programmed function is 0V
executed. 7
A
B
24 A-96.250.811 / 091017
AMI Silitrace
Installation
CAUTION
Inductive load Small inductive loads (max 0.1A) as for example the coil of a power
relay can be switched directly. To avoid noise voltage in the
AMI Transmitter it is mandatory to connect a snubber circuit in par-
allel to the load.
A snubber circuit is not necessary if an AMI relaybox is used.
A A AC or DC power supply
B C B AMI Transmitter
C External power relay
D E D Snubber
E Power relay coil
Resistive load Resistive loads (max. 1A) and control signals for PLC, impulse
pumps and so on can be connected without further measures
A B A AMI Transmitter
B PLC or controlled pulse pump
C C Logic
Actuators Actuators, like motor valves, are using both relays: One relay con-
tact is used for opening, the other for closing the valve, i.e. with the
2 relay contacts available, only one motor valve can be controlled.
Motors with loads bigger than 0.1A must be controlled via external
power relays or an AMI relay box.
A A AC or DC power supply
B C B AMI Transmitter
C Actuator
A-96.250.811 / 091017 25
AMI Silitrace
Installation
26 A-96.250.811 / 091017
AMI Silitrace
Installation
A
Third signal output 0/4 - 20 mA PCB
A Operating mode selector switch
ON
A-96.250.811 / 091017 27
AMI Silitrace
Installation
USB Interface
A USB interface PCB
B Third signal output 0/4 - 20 mA PCB
28 A-96.250.811 / 091017
AMI Silitrace
Instrument Setup
4. Instrument Setup
A-96.250.811 / 091017 29
AMI Silitrace
Instrument Setup
30 A-96.250.811 / 091017
AMI Silitrace
Instrument Setup
A-96.250.811 / 091017 31
AMI Silitrace
Instrument Setup
Grab
Standard
Sample
32 A-96.250.811 / 091017
AMI Silitrace
Instrument Setup
INIT
R1
R2 0.0 ppb
5.0 l/h 0.0 °C
A-96.250.811 / 091017 33
AMI Silitrace
Instrument Setup
A1
A2
34 A-96.250.811 / 091017
AMI Silitrace
Instrument Setup
1 Turn the occlusion frames [B] clockwise until they snap in to ac-
tivate the peristaltic pump.
NOTICE: Make sure the occlusion frames and pump tubes are
aligned in a 90° angle to the rotor.
A Turn to lock
A B Occlusion frame
C Rotor
B D Pump tube
C
D
A-96.250.811 / 091017 35
AMI Silitrace
Instrument Setup
4.7. Programming
Program all parameters for external devices (interface, recorders,
etc.). Program all parameters for instrument operation (limits,
alarms). See Program List and Explanations, p. 79.
36 A-96.250.811 / 091017
AMI Silitrace
Instrument Setup
B
C
D Reaction chamber
A-96.250.811 / 091017 37
AMI Silitrace
Instrument Setup
1.5
FOME raw [V]
0.5
38 A-96.250.811 / 091017
AMI Silitrace
Instrument Setup
A-96.250.811 / 091017 39
AMI Silitrace
Operation
5. Operation
5.1. Keys
A B C D
Exit Enter
Program
RUN 14:10:45 Main Menu 1
Access, Exit Enter
R1 Messages
1.4 ppb Diagnostics
R2
Maintenance
Exit Operation
9 l/h 25.4°C Installation
40 A-96.250.811 / 091017
AMI Silitrace
Operation
5.2. Display
A B C D E
RUN 15:20:18
R1
4.7 ppb F
15:10:52
R2
I 5 l/h 45.1 °C
G
H
A-96.250.811 / 091017 41
AMI Silitrace
Operation
Display when
operating with
two or more RUN 15:21:18
sample streams R1
4.7
15:10:52
ppb
F
Si1
4.5
R2 15:20:52
Si2 ppb ~
5 l/h 45.1 °C
Active measurement
Position of valve
~ No sample flow
n Measurement not valid
x Sample stream inactive (only visible if a Sample Sequencer is
connected to the AMI Silitrace and measuring mode
<internal> is selected).
42 A-96.250.811 / 091017
AMI Silitrace
Operation
Main Menu 1
Messages
Diagnostics
Maintenance
Operation
Installation
Menu Messages 1
Messages 1.1
Pending Errors Reveals pending errors as well as an event history
Maintenance List (time and state of events that have occurred at an
Message List earlier point of time).
It contains user relevant data.
Menu Diagnostics 2
Diagnostics 2.1
Identification Provides user relevant instrument and sample data.
Sensors
Sample
I/O State
Interface
Menu Maintenance 3
Maintenance 3.1
Calibration For instrument calibration, relay and signal output
Verification simulation, and to set the instrument time.
Zero It is used by the service personnel.
Service
Simulation
Menu Operation 4
Operation 4.1
Grab Sample User relevant parameters that might need to be
Sensors modified during daily routine. Normally password
Relay Contacts protected and used by the process-operator.
Logger Subset of menu 5 - Installation, but process-related.
Menu Installation 5
Installation 5.1
Sensors For initial instrument set up by SWAN authorized
Signal Outputs person, to set all instrument parameters. Can be
Relay Contacts protected by means of password.
Interface
A-96.250.811 / 091017 43
AMI Silitrace
Operation
Yes is highlighted.
Logger 4.1.3
Log interval
Save ?
6 Press [Enter] to save the new pa-
Clear logger no rameter.
Yes
No The system reboots, the new
parameter is set.
44 A-96.250.811 / 091017
AMI Silitrace
Operation
A-96.250.811 / 091017 45
AMI Silitrace
Operation
<Enter> to stop
<Enter> to stop
46 A-96.250.811 / 091017
AMI Silitrace
Maintenance
6. Maintenance
A-96.250.811 / 091017 47
AMI Silitrace
Maintenance
WARNING
Health hazard
For safe handling of the reagents you must read and under-
stand the Material Safety Data Sheets (MSDS).
Only persons trained in handling dangerous chemicals are al-
lowed to prepare the reagents.
Canister setup A Suction lance without
A
level detector
B (containers 1-3)
B Suction lance with
level detector
(container 4)
C Level detector
D 2 L mark
C E Reagent container 1
F Reagent container 2
G Reagent container 3
H Reagent container 4
D 2L 2L 2L 2L I Holder
E
OXYCON ON-LINE SILCA OXYCON ON-LINE SILCA
OXYCON ON-LINE SILCA OXYCON ON-LINE SILCA AMI SILICA REAGENT 3 AMI SILICA REAGENT 4
AMI SILICA REAGENT 1 AMI SILICA REAGENT 2
F
G
H
I
48 A-96.250.811 / 091017
AMI Silitrace
Maintenance
Reagent Each 2 liter reagent canister will last for approximately 1 month of
consumption operation if reagent saving is switched off or up to three months if
reagent saving is switched on (see 4.2.3, p. 85).
NOTICE: Excessive use of the flush/fill function or frequent flow
interruptions will shorten this period.
Contents of the Reagent 1: Bags 1a and 1b for Canister 1
reagent set ammonium molybdate and sodium hydroxyde
Reagent 2: Bottle 2 for Canister 2
sulfuric acid 25%
Reagent 3: Bag 3 for Canister 3
oxalic acid dihydrate
Reagent 4: Bag 4a and bottle 4b for Canister 4
ammonium ferrous sulfate hexahydrate
sulfuric acid 25% containing detergent
WARNING
Sulfuric Acid is corrosive and causes severe burns.
Read the Material Safety Data Sheets (MSDS) first.
Only persons trained in handling dangerous chemicals are al-
lowed to prepare the reagents
In case of contact with eyes, rinse immediately with plenty of wa-
ter eyelid wide open, summon medical advice. In case of acci-
dent or if you feel unwell, summon medical advice immediately.
NOTICE: Never prepare this reagent from concentrated sulfuric
acid packed in glass bottles!
Reagent 3:
H302: Harmful if swallowed H312: Harmful
in contact with skin
Warning
Reagent 4a:
H315: Causes skin irritation
H319: Causes serious eye irritation
Warning
H335: May cause respiratory irritation
Reagent 1b, Reagent 2, Reagent 4b:
H314: Causes severe skin burns and eye
damage
Danger
A-96.250.811 / 091017 49
AMI Silitrace
Maintenance
All canisters
Always replace the reagent filters (included with each reagent set)
when preparing new reagents.
Insert suction lances into the containers. Make sure that the num-
bers on the suction lances correspond to the numbers on the con-
tainers.
50 A-96.250.811 / 091017
AMI Silitrace
Maintenance
6.4. Calibration
Select Menu 3.1 <Maintenance>/<Calibration> and follow the in-
structions on the display.
Relay status during calibration:
Signal outputs are on hold
All limits are switched off
Calibration 4.1.5
Connect the
standard bottle
to marked holder
<Enter> to continue
Calibration 3.1.1
State xxxx
Progress
<Enter> to stop
Calibration 3.1.1
State Synchronize
Cycle 1
Timer 10 sec
<Enter> to stop
Calibration 3.1.1
State Measure
Cycle 1
Progress
<Enter> to stop
Press [Enter] to save the value in the
Calibration 3.1.1
calibration history or leave the menu
Operation completed with [Exit].
Factor xxxx
<Enter> to save
A-96.250.811 / 091017 51
AMI Silitrace
Maintenance
6.5. Verification
Select Menu 3.2 <Maintenance>/<Verification> and follow the in-
structions on the display.
Relay status during verification:
Signal outputs are on hold
All limits are switched off
Verification 4.1.5
Connect the
standard bottle
to marked holder
<Enter> to continue
Verification 3.2.1
State xxxx
Progress
<Enter> to stop
Verification 3.2.1
State Synchronize
Cycle 1
Timer 10 sec
<Enter> to stop
Verification 3.2.1
State Measure
Cycle 1
Progress
<Enter> to stop
Press [Enter] to save the value in the
Verification 3.2.1
verification history or leave the menu
Operation completed
with [Exit].
Current value xxx ppb
Reference value xxx ppb
Deviation xx.x%
<Enter> to save
52 A-96.250.811 / 091017
AMI Silitrace
Maintenance
6.6. Zero
Select Menu 3.2 <Maintenance>/<Zero> and follow the instructions
on the display.
Relay status during zero:
Signal outputs are on hold
All limits are switched off
Zero 3.2.1
State xxxx
Progress
<Enter> to stop
Zero 3.2.1
State xxxx
Cycle 1
Timer 10 sec
<Enter> to stop
Zero 3.2.1
State xxxx
Cycle 1
Progress
<Enter> to stop
Press [Enter] to save the value in the
Zero 3.2.1
verification history or leave the menu
Operation completed
Zero 1.00 V
with [Exit].
<Enter> to save
A-96.250.811 / 091017 53
AMI Silitrace
Maintenance
CAUTION
54 A-96.250.811 / 091017
AMI Silitrace
Maintenance
E
F
Ø = 0.64 mm
Ø = 0.64 mm
Ø = 0.64 mm
Ø = 0.64 mm
A-96.250.811 / 091017 55
AMI Silitrace
Maintenance
7 Install the sample tube and all reagent tubes (see Replace the
Pump Tubes, p. 54).
8 Push the occlusion frames onto the rotor.
9 Lock the occlusion frames.
10 Put suction lances into the corresponding reagent container.
11 Start fill system, see Fill System, p. 56.
Service 3.2.2
Navigate to menu <Maintenance /
Service/Fill system>.
Fill System
Prepare Maintenance Press [Enter].
<Enter> to stop
Done
56 A-96.250.811 / 091017
AMI Silitrace
Maintenance
A-96.250.811 / 091017 57
AMI Silitrace
Troubleshooting
7. Troubleshooting
7.1. Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded. Such Errors are marked E0xx (bold and black).
Fatal Error (blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low).
Such Errors are marked E0xx (bold and orange)
Errors which indicate a hardware failure of the instrument.
Such Errors are marked E0xx (bold and red)
<Enter> to Acknowledge
58 A-96.250.811 / 091017
AMI Silitrace
Troubleshooting
E008 SilTrace Temp. low – If the error occurs when starting up:
No action necessary. Only wait until the
photometer has warmed up (20
minutes at room temperature), then this
error disappears automatically.
– If the error occurs during operation:
Check whether there are any other
errors and resolve them first (some
errors cause the heater to be switched
off).
E009 Sample Flow high – Check inlet pressure
– Re-adjust sample flow
– Check programmed value, see
5.3.1.32.2, p. 97
E010 Sample Flow low – Check inlet pressure
– Re-adjust sample flow
– Check programmed value, see
5.3.1.32.32, p. 97
E011 Absorbance too high – Check process
– Check tube connections for air
leakages
A-96.250.811 / 091017 59
AMI Silitrace
Troubleshooting
60 A-96.250.811 / 091017
AMI Silitrace
Troubleshooting
A-96.250.811 / 091017 61
AMI Silitrace
Troubleshooting
62 A-96.250.811 / 091017
AMI Silitrace
Troubleshooting
F
G
H
A-96.250.811 / 091017 63
AMI Silitrace
Troubleshooting
Install the new 1 Insert the new reaction chamber into the photometer housing
reaction and tighten the fixing screw [G].
chamber 2 Put the cover [B] onto the photometer unit and tighten the 4
cover fixing screws [A].
3 Push the cuvette into the slot of the cuvette housing.
4 Connect all tubes to the connection panel according to the dia-
gram below.
2 4
03
1B 3 04
64 A-96.250.811 / 091017
AMI Silitrace
Troubleshooting
CAUTION
Never loosen the 4 allen screws [D] visible on the 6-way valve
body.
D
To remove the 6-way valve from the housing proceed as follows:
A-96.250.811 / 091017 65
AMI Silitrace
Troubleshooting
Install the
6-way valve
A B C D E F G
Seal all unused Inputs with the enclosed blind plugs [A]. Proceed
as follows:
1 Make sure that the valve shaft with driving pin [D] is aligned with
the driving slot [F].
2 Install the 6-way valve so that the valve shaft with driving pin fits
into the driving slot of the motor shaft and the positioning screw
[E] fits into the guiding hole [G].
3 Attach the 6-way valve with the fixing screws [C] to the valve
housing, use the enclosed 2.5 mm Allen key.
4 Fit all tubes to the corresponding outputs/inputs of the 6-way
valve [B], see Replace the Reagent Tubes, p. 68.
5 Screw the blind plugs into the unused inputs of the 6-way valve.
6 Switch on the instrument and select <Maintenance>/
<Service>/<Fill System>.
7 Check all tube connections for leakage.
66 A-96.250.811 / 091017
AMI Silitrace
Troubleshooting
A A Photometer module
B Cuvette
B
A Photometer module
A B Cuvette
A-96.250.811 / 091017 67
AMI Silitrace
Troubleshooting
A
05
B
13
12
1B
2 03
03
3 3
4 4
04 04
05 I 12
12
4
5 3
1B 7
6 2 11
1 11
02
S
01
04
04
1A
03 1A 2 3 4
T E
02 1 2 3 4 11
1 2 01 3 4
J
K
L
M
68 A-96.250.811 / 091017
AMI Silitrace
Troubleshooting
C
B
A-96.250.811 / 091017 69
AMI Silitrace
Troubleshooting
70 A-96.250.811 / 091017
AMI Silitrace
Troubleshooting
A-96.250.811 / 091017 71
AMI Silitrace
Troubleshooting
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks.
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
– relay 1
– relay 2
– alarm relay
When a fuse has blown, find out the cause and fix it before replac-
ing it with a new one.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A B C D E F G
72 A-96.250.811 / 091017
AMI Silitrace
Program Overview
8. Program Overview
For explanations about each parameter of the menus see Program
List and Explanations, p. 79.
A-96.250.811 / 091017 73
AMI Silitrace
Program Overview
74 A-96.250.811 / 091017
AMI Silitrace
Program Overview
A-96.250.811 / 091017 75
AMI Silitrace
Program Overview
Pump 3.5.9*
Set Time (Date), (Time)
3.6*
76 A-96.250.811 / 091017
AMI Silitrace
Program Overview
A-96.250.811 / 091017 77
AMI Silitrace
Program Overview
78 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5 Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5 Demands necessary maintenance, e.g. preparing new reagents.
1.3 Message List
1.3.1 Shows the error history: Error code, date/time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation: Designation of the instrument.
Version: Firmware of instrument (e.g. V6.31-10/17).
2.1.3 Peripherals:
2.1.3.1 PeriClip: Firmware of peristaltic pump (e.g. 1.06).
RoValve: Firmware of rotary valve (6-way valve) (e.g. 1.60).
SilTrace: Firmware of heater in the photometer module (e.g. 1.00).
2.1.4 Factory Test: Test date of the Instrument and Motherboard
2.1.5 Operating Time: Years / Days / Hours / Minutes / Seconds
2.2 Sensors
2.2.1 SilTrace (photometer module):
Temp. Temperature inside the reaction chamber in °C
PWM: Heating power in percent (100% at start-up)
2.2.1.3 Photometer:
Current value: Shows the photometer signal in ppb.
(Raw value): Shows the actual photometer signal in V.
A-96.250.811 / 091017 79
AMI Silitrace
Program List and Explanations
FOME mean
Absorbance: A = -log 10
zero
FOME Mean: Raw signal in V, measured during T2 (pump
stopped) to calculate the concentration.
2.2.2 Miscellaneous:
Case Temp: Shows the current temperature in °C inside the
transmitter.
State Machine: Shows the current process of the instrument.
Measurement cycle
FIND PEAK Step 1 of measurement (pump on):
The instrument is in measuring mode, waiting for
the peak.
WAIT T1 Step 2 of measurement (pump on):
The instrument is in measuring mode, peak found
and filling the photometer de-aeration tube and the
siphon tube. See Fluidic, p. 10.
WAIT T2 Step 3 of measurement (pump off):
The instrument is in measuring mode, waiting for
stabilization. FOME mean is recorded. After this
state, the instrument goes back to FIND PEAK.
80 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
2.2.3 History
2.2.3.1 Zero History
2.2.3.1.1 Number: Counter of zero calibrations.
Date, Time: Date and time when a zero was performed.
Zero: Measuring value in V of the sample without reagent 1,
which is responsible for coloring the sample. A too high value
may result in a FOME dirty error.
2.2.3.2 Cal. History
2.2.3.2.1 Number: Counter of standard calibrations.
Date, Time: Date and time when the standard calibration was
performed.
Factor: Correction factor of the calibration curve.
2.2.3.3 Verif. History
2.2.3.3.1 Number: Counter of verifications.
Date, Time: Date and time when the verification was performed.
Meas. Value: Measuring value of the sample in ppb.
Reference value: Silica concentration in ppb of the standard
solution used.
Deviation: Deviation between the two measurements in %.
2.2.3.4 Grab sample
Number: Counter of grab samples.
Date, Time: Date and time when the grab sample was measured.
Sample ID: ID assigned by the user.
Sample: Measuring result of the grab sample measurement.
A-96.250.811 / 091017 81
AMI Silitrace
Program List and Explanations
2.5
1.5
FOME raw [V]
0.5
2.5 Interface
2.5.1 Interface:
Only available if the optional interface is installed. Shows the pro-
grammed communication settings.
82 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
3 Maintenance
3.1 Calibration
3.1.5 During calibration, a solution of known silica concentration (stan-
dard) is measured and the measuring value is compared with the
reference value of the standard (set in 5.1.1.1, p. 87). The instru-
ment then sets the calibration factor to adjust the photometer sensi-
tivity.
It is recommended to perform a calibration:
at first start-up
after replacing the pump tubes
after replacing the cuvette
All calibrations are saved in the Calibration History.
3.2 Verification
3.2.5 During verification, a solution of known silica concentration (stan-
dard) is measured and the mesuring value is compared with the ref-
erence value of the standard. The deviation is expressed in
percent. Unlike a calibration, a verification does not change the cal-
ibration factor.
It is recommended to check the performance of the system by an
automatic weekly verification (default setting).
All verifications are saved in the Verification History.
3.3 Zero
3.3.5 To determine the electronic offset and light intensity of the photom-
eter the sample is measured without adding the color forming re-
agent 1.
It is recommended to program an automatic daily zero calibration
(default setting).
3.4 Service
3.4.1 Fill system
3.4.1.5 Activates the reagent pump and fills all tubes from the container to
the cuvette outlet.
3.4.2 Prepare Maintenance
All tubes are rinsed and emptied when starting this function.
NOTICE: Follow the instructions on the screen carefully,
otherwise the reagents in the canisters will be polluted with
already mixed reagents.
A-96.250.811 / 091017 83
AMI Silitrace
Program List and Explanations
3.5 Simulation
To simulate a value or a relay state, select the
alarm relay
relay 1 and 2
signal output 1 and 2
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
3.6 Set Time
Adjust date and time.
84 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
4 Operation
4.1 Grab Sample
4.1.5 Starts a grab sample measurement. See Grab Sample Measure-
ment, p. 45.
4.2 Sensors
4.2.1 Filter Time Constant: Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
4.2.2 Hold after Cal: Delay permitting the instrument to stabilize again after
calibration. During calibration- plus hold-time, the signal outputs are
frozen (held on last valid value), alarm values, limits are not active.
Range: 0–6‘000 Sec
4.2.3 Reagents saving: Increases the duration of the T2 state (pump off) to
save reagents.
Range: No, Minimum, Medium, Maximum
On a single-channel instrument, the reagent saving mode has the
following effect on the measuring cycle and reagent lifetime:
Option Measuring cycle Reagent lifetime
No 2.8 min approx. 28 d
Minimum 5.3 min approx. 50 d
Medium 7.8 min approx. 75 d
Maximum 10.3 min approx. 100 d
On a multi-channel instrument, the reagent saving effect depends on
the switching time (see 5.1.2.3, p. 90). The faster the channels are
switched, the smaller the reagent saving effect is.
The following table shows the reagent saving effect for different
switching times if <Reagents saving> is set to <Maximum>. <Medi-
um> and <Minimum> correspond to 66% and 33% of the reagent
saving effect.
Switching time Measuring cycle Reagent lifetime
15 min 3.0 min approx. 28 d
20 min 3.8 min approx. 35 d
30 min 5.5 min approx. 50 d
40 min 7.2 min approx. 70 d
50 min 8.8 min approx. 85 d
60 min 10.3 min approx. 100 d
A-96.250.811 / 091017 85
AMI Silitrace
Program List and Explanations
86 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
5 Installation
5.1 Sensors
5.1.1 Meas. Parameters
5.1.1.1 Cal./Verif.
5.1.1.1.1 Standard: The default standard is 100 ppb.
During a calibration or verification 15 ml standard is consumed.
Therefore a standard bottle lasts for 3 months at default interval
settings. The default interval settings are:
Start time: 06:00:00
Monday: Verification
All other days: Off
Range: 10.0 ppb to 1.0 ppm (1000 ppb)
5.1.1.1.2 Parameters
5.1.1.1.2.1 Start time: Program the daily start time of a verification or calibra-
tion. Default setting is 06:00:00
5.1.1.1.2.2 Monday: Program a verification, a calibration or Off for this day. A ver-
ification or calibration will be started at the programed <start time>.
5.1.1.1.2.3 Tuesday: Same as Monday.
5.1.1.1.2.4 Wednesday: Same as Monday.
5.1.1.1.2.5 Thursday: Same as Monday.
5.1.1.1.2.6 Friday: Same as Monday.
5.1.1.1.2.7 Saturday: Same as Monday.
5.1.1.1.2.8 Sunday: Same as Monday.
NOTICE: If there is an overlap between a programmed zero
calibration and a programmed calibration/verification, the zero
calibration is prioritized and the calibration/verification is
skipped.
A-96.250.811 / 091017 87
AMI Silitrace
Program List and Explanations
88 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
Mode Internal In the Mode Internal, the AMI Silitrace works as a master.
AMI Silitrace Dual-Stream
The instrument automatically switches between channel 1 and 2.
Sample Sequencer
The AMI Silitrace sequentially measures each single sample
stream of the Sample Sequencer.
Via an external PLC it can be defined which sample streams should
not be measured. In the example below, only the sample streams
2, 4 and 6 are measured, whereas the samples streams 1, 3 and 5
are switched off. Sample streams which are switched off are
marked with an “x” behind the measuring value on the AMI Silitrace
display.
CH1 9
CH2 10
1 mA sink, 5 V open
6 Inputs from PLC
CH3 11
Processor
CH4 12
CH5 13
CH6 14
GND 15
SEQUENCER
A-96.250.811 / 091017 89
AMI Silitrace
Program List and Explanations
Example:
If sample stream 1 of the Sample Sequencer is active, the AMI Sili-
trace measures the sample stream 1 until the Sample Sequencer
changes to the next programmed channel. In the example below, the
sample stream 3 (CH3) highlighted in green will be measured as
soon as the switching time has elapsed.
CH1 9
CH2 10
1 mA sink, 5 V open
6 Inputs from PLC
CH3 11
Processor
CH4 12
CH5 13
CH6 14
GND 15
SEQUENCER
5.1.2.3 Switching time: Time after which the instrument switches to the
next channel. The switching time is valid for all three operating
modes (Internal, Fieldbus and External).
Range: 15–60 min, selectable in 5 min steps
90 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
A-96.250.811 / 091017 91
AMI Silitrace
Program List and Explanations
20
(0 - 20 [mA])
(4 - 20 [mA])
B A
10 12
0/4 X
0.0 0.1 0.2 0.3 0.4 0.5
[mA]
20
(0 - 20 [mA])
(4 - 20 [mA])
10 12
0/4 X
2 4 6
1 10 100 1’000 10’000
0 1 2 3 4
92 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
5.2.1.40 Scaling: Enter beginning and end point (Range low & high) of the
linear or logarithmic scale. In addition, the midpoint for the bilinear
scale.
Parameter Si1:
5.2.1.40.11 Range low: 0.0 ppb to 1.00 ppm
5.2.1.40.21 Range high: 0.0 ppb to 1.00 ppm
Parameter Si2:
5.2.1.40.12 Range low: 0.0 ppb to 1.00 ppm
5.2.1.40.22 Range high: 0.0 ppb to 1.00 ppm
As control Signal outputs can be used for driving control units. We distinguish
output different kinds of controls:
P-controller: The controller action is proportional to the devia-
tion from the setpoint. The controller is characterized by the
P-Band. In the steady-state, the setpoint will never be
reached. The deviation is called steady-state error.
Parameters: setpoint, P-Band
PI-controller: The combination of a P-controller with an
I-controller will minimize the steady-state error. If the reset
time is set to zero, the I-controller is switched off.
Parameters: setpoint, P-Band, reset time.
PD-controller: The combination of a P-controller with a
D-controller will minimize the response time to a fast change
of the process value. If the derivative time is set to zero, the
D-controller is switched off.
Parameters: setpoint, P-Band, derivative time.
PID-controller: The combination of a P-, an I - and a D-con-
troller allows a proper control of the process.
Parameters: setpoint, P-Band, reset time, derivative time.
A-96.250.811 / 091017 93
AMI Silitrace
Program List and Explanations
B A
X
a
94 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
5.2.1.43.3 Reset time: The reset time is the time till the step response of a sin-
gle I-controller will reach the same value as it will be suddenly
reached by a P-controller.
Range: 0–9’000 sec
5.2.1.43.4 Derivative time: The derivative time is the time till the ramp re-
sponse of a single P-controller will reach the same value as it will
be suddenly reached by a D-controller.
Range: 0–9’000 sec
5.2.1.43.5 Control timeout: If a controller action (dosing intensity) is constantly
over 90% during a defined period of time and the process value
does not come closer to the setpoint, the dosing process will be
stopped for safety reasons.
Range: 0–720 min
5.2.4 Signal Sequencer: Only visible if a Sample Sequencer is
connected.
Assign process value, the current loop range and a function to the
signal output.
5.2.4.1 Parameter: Only the parameter “Si Sequencer” is available.
5.2.4.2 Current Loop: Select the current range of the signal output. Make
sure the connected device works with the same current range.
Available ranges: 0–20 mA or 4–20 mA
5.2.4.3 Scaling: Define the scaling of the signal output used to transmit a
process value. Available functions are:
Linear, bilinear or logarithmic for process values.
See As process values, p. 92.
A-96.250.811 / 091017 95
AMI Silitrace
Program List and Explanations
96 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
5.3.1.32 Sample Flow: Define at which sample flow a flow alarm should be
issued.
5.3.1.32.1 Flow Alarm: Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will al-
ways be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
NOTICE: Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.32.2 Alarm High: If the measuring values rises above the programmed
value E009 will be issued.
Range: 0–50 l/h
5.3.1.32.32 Alarm Low: If the measuring values falls below the programmed
value E010 will be issued.
Range: 0–50 l/h
5.3.1.42 Case Temp. high: Set the alarm high value for temperature of
electronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
5.3.1.5 Case Temp. low: Set the alarm low value for temperature of
electronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10–20 °C
A-96.250.811 / 091017 97
AMI Silitrace
Program List and Explanations
5.3.2 and 5.3.3 Relay 1 and 2: The contacts can be set as normally open or nor-
mally closed with a jumper. See Relay 1 and 2, p. 24.
The function of relay contacts 1 or 2 are defined by the user
NOTICE: The navigation in the menu <Relay 1> and <Relay 2>
is identical. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
- Channel Selection (AMI Silitrace Dual-Stream)
2 Then enter the necessary data depending on the selected func-
tion.
5.3.2.1 Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20 Parameter: select a process value:
Sample flow
Si1
Si2 (AMI Silitrace Dual-Stream)
5.3.2.300 Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
Range: 0.0 ppb to 1.00 ppm
5.3.2.400 Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
Range: 0.0 ppb to 1.00 ppm
5.3.2.50 Delay: Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed alarm.
Range. 0–600 sec
5.3.2.1 Function = Control upwards/downwards:
The relays may be used to drive control units such as solenoid
valves, membrane dosing pumps or motor valves. When driving a
motor valve both relays are needed, relay 1 to open and relay 2 to
close the valve.
98 A-96.250.811 / 091017
AMI Silitrace
Program List and Explanations
A-96.250.811 / 091017 99
AMI Silitrace
Program List and Explanations
5.3.2.24 daily
The relay contact can be activated daily, at any time of a day.
5.3.2.341 Start time: to set the start time proceed as follows:
1 Press [Enter], to set the hours.
2 Set the hour with the [ ] or [ ] keys.
3 Press [Enter], to set the minutes.
5.3.4 Input: The functions of the relays and signal outputs can be de-
fined depending on the position of the input contact, i.e. no function,
closed or open.
AMI Silitrace Dual-Stream:
If <Channel Selection> (5.1.2.2, p. 88) is set to “external”, the input
is used to switch between channels 1 and 2.
5.3.4.1 Active: Define when the input should be active:
The measurement is interrupted during the time the input is active.
No: Input is never active.
When closed Input is active if the input relay is closed
When open: Input is active if the input relay is open
5.3.4.2 Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Cont.: Signal outputs continue to issue the measured
value.
Hold: Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off: Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
5.3.4.3 Output/Control: (relay or signal output):
Cont.: Controller continues normally.
Hold: Controller continues on the last valid value.
Off: Controller is switched off.
5.3.4.4 Fault:
No: No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes: Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is
active.
5.3.4.5 Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6‘000 sec
5.4 Miscellaneous
5.4.1 Language: Set the desired language.
Language
German
English
French
Spanish
5.4.2 Set defaults: Reset the instrument to factory default values in three
different ways:
Set defaults
no
Calibration
In parts
Completely
Calibration: Sets calibration values back to default. All other
values are kept in memory.
In parts: Communication parameters are kept in memory. All
other values are set back to default values.
Completely: Sets back all values including communication
parameters.
5.4.3 Load Firmware: Firmware updates should be done by instructed
service personnel only.
Load Firmware
no
yes
5.4.4 Password: Select a password different from 0000 to prevent unau-
thorized access to the following menus:
5.4.4.1 Messages
5.4.4.2 Maintenance
5.4.4.3 Operation
5.4.4.4 Installation.
Each menu may be protected by a different password. If you forgot
the passwords, contact the closest SWAN representative.
5.4.5 Sample ID: Identify the process value with any meaningful text,
such as KKS number.
5.4.6 Line Break Detection: Define if message E028 should be issued in
case of a line break on signal output 1 or 2.
Choose between <Yes> or <No>.
5.5 Interface
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
Download The current Safety Data Sheets (SDS) for the above listed Re-
MSDS agents are available for downloading at www.swan.ch.
12. Index
A I
Alarm Relay . . . . . . . . 9, 23 Input . . . . . . . . . . . . . 9, 23
Application Range . ...... 8 Instrument Overview . . . . 14
Interface
B HART . . . . . . .... 28
Background. ......... 88 Modbus . . . . . .... 27
Profibus . . . . . .... 27
USB . . . . . . . .... 28
C
Cable thicknesses . . . . . . 19
Calendar . . . . . . . . . . . 101 L
Changing parameters . . . . 44 Longer Stop of Operation. . 57
Changing values . . . . . . . 44
Cuvette factor . . . . . . 67, 84 M
Measuring Principle . ..... 8
D Modbus . . . . . . . . .... 27
Default Values . . . . . . . . 106 Mode External . . . . .... 89
Disassemble the solenoid valve 70 Mode Fieldbus . . . . .... 89
Mode Internal . . . . .... 89
Mounting . . . . . . . .... 16
E
Error List ........... 58
O
F
On-site requirements .... 12
Fill system .......... 83
P
P2P
G
counter . . . . . . . . . 82
Grab Sample . ........ 45
Diagram . . . . . . . . . 82
period . . . . . . . . . . 82
H Power Supply . . . . . . 12, 22
HART . . . . . . . ...... 28 Prepare Maintenance . . . . 83
History Profibus . . . . . . . . . . . . 28
Cal. History . ...... 81
Ver. History . ...... 81
R
Zero History . ...... 81
Reagent consumption. ... 49
Relays. . . . . . . . . . .... 9
S T
Safety Features. . . . . .... 9 Technical Data . . . . . . . . 14
Sample Flow . . . . . . . . . 33 Terminals . . . 21, 23–24, 27
Sample requirements . . . . 12 Tube numbering . . . . . . . 68
Setup. . . . . . . . . . . . . . 29
Signal Outputs . . . . . . 9, 26 U
Silica measurement . . . . . 12 USB Interface . ........ 28
SilTrace . . . . . . . . . . . . 79
Software . . . . . . . . . . . . 43 W
Standard consumption . . . . 32
State Machine . . . . . . . . 80 Wire .............. 19
System, Description of . .... 8
13. Notes
SWAN
is represented worldwide by subsidiary companies
and distributors.
SWAN Products
Analytical Instruments for:
Potable Water
Cooling Water
Made in Switzerland