Agilent Cary 60 UV-Vis: Guaranteed Specifi Cations
Agilent Cary 60 UV-Vis: Guaranteed Specifi Cations
Guaranteed specifications
Introduction
The Agilent Cary 60 UV-Vis spectrophotometer is efficient, accurate and flexible, and is
designed to meet both current and future measurement needs. The proven, robust design of
the Cary 60 comprises a double beam, Czerny-Turner monochromator, 190–1100 nm
wavelength range, 1.5 nm fixed spectral bandwidth, full spectrum Xenon pulse lamp single
source with exceptionally long life, dual silicon diode detectors, quartz overcoated optics,
scan rates up to 24,000 nm/min, 80 data points/sec maximum measurement rate, non-
measurement phase stepping wavelength drive, room light immunity, central control by PC
with Microsoft® Windows® operating system. Supported by GLP software, optional 21 CFR
Part 11 capable software, and dedicated instrument validation software which includes
pharmacopeia test suites.
Agilent Cary 60 UV-Vis spectrophotometers are manufactured according to a quality
management system certified to ISO 9001. The guaranteed specifications are listed in this
document and are based on the 4 sigma statistical confidence level of the final acceptance
tests performed at the factory.
Features, advantages and Feature Advantage/Benefit
benefits Xenon pulse lamp source Eliminates photobleaching while allowing the highest quality data to be
collected over the complete UV-Vis range — all at the lowest cost of
ownership due to an exceptionally long lamp life.
Room light immunity Unique optical design allows accurate sample measurement even with
the sample lid open — especially useful for enzyme assays, fiber-optic
based measurements or high throughput QA/QC labs.
1.5 nm fixed spectral Gives excellent spectral resolution for solids and liquids and meets
bandwidth international Pharmacopoiea compliance regulations.
190–1100 nm wavelength Complete coverage of UV-Vis range and extending up into the NIR.
range
24,000 nm/min maximum Allows complete spectral range scanning in under 3 seconds, ideal for
scan rate fast kinetics or high sample throughput.
Non-measurement phase Means that sample and reference measurements are made at the same
stepping wavelength drive wavelength ensuring that no peak shifts are observed — even when
measuring at the fastest scan speeds.
Focused beam measuring Ensures efficient energy coupling to accessories including fiber optic
1.5 x 1.0 mm probes and ultra-microvolume cuvettes for measurement of low volume
samples.
Instrument hardware
Source
Unique full-spectrum Xenon flash lamp (80 Hz) with typical lifetime of
10 years (guaranteed 3 years)
Monochromator
Czerny-Turner
Grating
Beam splitter
Detectors
≤ 1.5 nm
2
Agilent Cary 60 UV-Vis guaranteed specifications
190–1100 nm
± 0.5 at 541.94 nm
± 0.1 nm
± 4.0 Abs
Photometric display
200 to 850 nm, smooth 21 filter applied, baseline corrected ± 0.001 Abs
3
Recommended environmental Instrument dimensions (width x depth x height)
conditions
Packed 595 x 710 x 350 mm (24 x 28 x 14 in)
Unpacked 477 x 567 x 196 mm (19 x 23 x 8 in)
The Cary 60 has been designed to withstand the weight of a PC
monitor up to 10 kg (33 lb)
Instrument weight
Instrument conditions
Condition Altitude (m, ft) Temp. (°C, °F) Humidity (%RH) non-
condensing
Non-operating (transport) 0–4600, 0–15000 -40–75 °C, -40–167 °F 15–90%
Operating within 0–3100, 0–10000 5–40 °C, 41–104 °F 50–80%
performance
specifications
For optimum analytical performance, it is recommended that the ambient temperature of the
laboratory be between 20–25 °C and be held constant to within ±2 °C throughout the entire working
day
A standard 3.2 A/12 V plug pack is provided. Power cords are provided
based on the user’s country requirements. Only the supplied power
supply is to be used with this equipment.
Required supply voltage 100–240 V AC, Frequency 47-63 Hz
Nominal rating Scanning: 18 W, Idle: 9 W
0.0125–999 s
24,000 nm/min
24,000 nm/min
0.15–5.0 nm
Repetitive scanning
4
Agilent Cary 60 UV-Vis guaranteed specifications
80 data points/second
Temperature monitor
0.5 μL
Customer support policies Support and training
5
www.agilent.com/chem
Agilent shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance or use of this material.