27631437
27631437
27631437
User Manual
Thermox
150 Freeport Road
PN 90269VE , Rev. Y Pittsburgh, PA 15238
Offices
USA - Pittsburgh CHINA
150 Freeport Road AMETEK Shanghai (SPL)
Pittsburgh, PA 15238 Room 408, Metro Tower
Ph. 412-828-9040 #30 Tian Yao Qiao Road
Fax 412-826-0399 Shanghai 200030
Ph. 86 21 6426 8111
Fax 86 21 6426 7054
CANADA GERMANY
2876 Sunridge Way N.E. AMETEK GmbH
Calgary, AB T1Y 7H9 Rudolf-Diesel-Strasse 16
Ph. 403-235-8400 D-40670 Meerbusch
Fax 403-248-3550 Ph. 49 21 59 91 36 0
Fax 49 21 59 91 3680
SINGAPORE
AMETEK Singapore Pte. Ltd.
No 43, Changi South Avenue 2, #04-01
Singapore 486164
Ph. 65-64842388
Fax 65-64816588
© 2000 AMETEK
This manual is a guide for the use of the Thermox Series 2000 WDG-IV (IVC / IVCM). Data herein has been verified and validated and is
believed adequate for the intended use of this instrument. If the instrument or procedures are used for purposes over and above the capabilities
specified herein, confirmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no
obligation or liability. This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.
Contents
Offices.......................................................................................................................ii
Safety Notes........................................................................................................... vi
Environmental Information (WEEE)............................................................... viii
General Safety Summary......................................................................................ix
CHAPTER 1 Overview
Sensor Operations............................................................................................... 1-1
Sensor . ................................................................................................................. 1-2
Basic Elements of the Sensor...................................................................... 1-2
The Oxygen Measuring Cell....................................................................... 1-2
Sensor Operations - WDG-IVCM..................................................................... 1-5
Close-Coupled Extrative Flue Gas Analyzer.................................................. 1-5
Hot-Wire Type............................................................................................... 1-5
RTD Type....................................................................................................... 1-6
Detectors............................................................................................................... 1-7
Combustibles................................................................................................. 1-7
Methane......................................................................................................... 1-7
Common Operator Errors................................................................................. 1-8
Start-Up Checklist............................................................................................... 1-9
Technical Support.............................................................................................. 1-10
CHAPTER 2 Specifications
Series 2000 Control Unit..................................................................................... 2-1
Sensor : WDG-IV . .............................................................................................. 2-2
Sensor : WDG-IVC . .......................................................................................... 2-3
Sensor : WDG-IVCM......................................................................................... 2-4
Sensor : WDG-IVC - 0-500 PPM ..................................................................... 2-6
Remote Calibration Unit (RCU)........................................................................ 2-7
CHAPTER 3 Installation
Analyzer Location............................................................................................... 3-1
Inspect Shipping Contents................................................................................ 3-2
Mechanical Installation...................................................................................... 3-3
Sample Inlet Probe Installation.................................................................. 3-3
Probe Heater Installation............................................................................ 3-6
Exhaust Tube Installation............................................................................ 3-7
Sensor Mounting.......................................................................................... 3-7
Calibration/Aspirator Air/Z-Purge Connections................................... 3-12
Weatherproof Wall/Z-Purge Option........................................................ 3-17
Wiring................................................................................................................. 3-19
General Wiring and Conduit Requirements.......................................... 3-20
Control Unit Mains Supply Connections............................................... 3-21
Sensor Connections.................................................................................... 3-24
Contents | iii
Remote Sense Setup Wiring for Combustibles...................................... 3-28
Current Output Connections................................................................... 3-29
Alarm Contact Connections..................................................................... 3-30
Remote Calibration Unit Connections.................................................... 3-32
RS-485 Communications Connections.................................................... 3-34
Option Card Installation.................................................................................. 3-38
Removing Display Module or Power Supply/Keypad Module.......... 3-38
Adding an Option Card............................................................................ 3-39
Removing an Option Card....................................................................... 3-39
Contents | v
Thermocouple Failure................................................................................. 6-4
Zero Gas Range Error.................................................................................. 6-4
Diagnostics Checks............................................................................................. 6-5
Wiring Checks............................................................................................... 6-6
Interconnecting Wiring Problem............................................................... 6-6
Thermocouple Checks................................................................................. 6-6
Shorted/Failed Thermocouple.................................................................... 6-6
Reversed Thermocouple Wires.................................................................. 6-6
Thermocouple Compensation .................................................................. 6-6
Circuit Check................................................................................................. 6-6
Calibration/Aspirator Setup Checks.......................................................... 6-8
AC Power Checks......................................................................................... 6-9
Furnace Checks............................................................................................. 6-9
Process Pressure Checks............................................................................ 6-10
Cell Checks.................................................................................................. 6-10
General Troubleshooting................................................................................. 6-11
Leak Check.................................................................................................. 6-11
Plugged Plumbing Check......................................................................... 6-12
Aspirator Air Not Pulling Sample from Process.................................... 6-12
Series 2000 LEDs................................................................................................ 6-13
Auto Calibration Card LEDs..................................................................... 6-13
Combustibles Card LEDs.......................................................................... 6-13
Combustibles and Methane Card LEDs................................................. 6-14
Sensor Board (80471SE) LEDs.................................................................. 6-14
vi |
Appendix B Menu Option Charts
| vii
Safety Notes
warnings, cautions, and notes contained in this manual emphasize
critical instructions as follows:
Electrical Safety
Up to 5 kV may be present in the analyzer housings. Always shut down power
source(s) before performing maintenance or troubleshooting. Only a qualified
electrician should make electrical connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may
impair the safety protection originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety
protection are void if instrument is operated from an improperly grounded
power source.
The following applies to the Series 2000 Control unit and the WDG-IV Division
2 Models:
Power, input, and output (I/O) wiring must be in accordance with Class I, Divi-
sion 2 wiring methods [Article 501-4(b) of the National Electric Code, NFPA 70]
and in accordance with the authority having jurisdiction.
The maximum ambient temperature for the Series 2000 control unit is 50 °C
(122 °F)
The Series 2000 Control Unit with WDG-IV sensor is a complex piece of equip-
ment that should only be serviced by a qualified service technician with expertise
in instrument technology and electrical systems. AMETEK recommends that
all equipment requiring service be sent back to the factory. You should only
attempt to repair or service this equipment after receiving training from an
AMETEK/P&AI Division training representative. If you decide to service this
equipment be aware that high voltages, high temperatures, and other potentially
hazardous conditions may arise.
Contents | ix
Environmental Information (WEEE)
This AMETEK product contains materials that can be reclaimed and recycled. In some cases
the product may contain materials known to be hazardous to the environment or human
health. In order to prevent the release of harmful substances into the environment and to
conserve our natural resources, AMETEK recommends that you arrange to recycle this prod-
uct when it reached its “end of life”.
Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a munici-
pal waste system (residential trash). The Wheelie Bin marking on this product is a reminder
to dispose of the product properly after it has completed its useful life and been removed
from service. Metals, plastics, and other components are recyclable and you can do your
part by doing one of the following steps:
Contents | xi
Declaration of Conformity
Product Names: Thermox WDG series of Flue Gas Analyzers using Series 2000 Controller:
WDG-IV WDG-IVC WDG-IVCM
WDG-HPII WDG-HPIIC
O2 Only RCU
O2 & Comb RCU
O2, Combustibles & Methane RCU
EMC compliance:
EMC Directive 2004/108/EC
Immunity:
EN 50082-2:1995 Generic - Heavy Industrial Equipment
EN 50082-1:1997 Generic - Residential, Commercial, Light Industrial Equipment
EN 61000-6-2:1999 Generic - Heavy Industrial Equipment
EN 61326:1997 Measuring Equipment (Heavy Industrial)
Emissions:
EN 55011:1998 Class A, ISM Equipment
Safety Compliance:
UL 3101-1 UL Listed Inspection and Measuring Electrical Equipment
CAN/CSA 22.2 No. 1010.1-92 cUL Listed Inspection and Measuring Electrical Equipment
WDG Series analyzers with the suffix Div. 2, O2 Only RCU, O2 & Comb RCU, and O2,
Combustibles & Methane RCU also conform to the following safety standard:
UL 1604 UL Listed Process Control Equipment for Use in Hazardous Locations
We warrant that any equipment of our own manufacture or manufactured for us pursuant
to our specifications which shall not be, at the time of shipment thereof by or for us, free
from defects in material or workmanship under normal use and service will be repaired or
replaced (at our option) by us free of charge, provided that written notice of such defect
is received by us within twelve (12) months from date of shipment of portable analyzers
or within eighteen (18) months from date of shipment or twelve (12) months from date of
installation of permanent equipment, whichever period is shorter. All equipment requiring
repair or replacement under the warranty shall be returned to us at our factory, or at such
other location as we may designate, transportation prepaid. We shall examine such returned
equipment, and if it is found to be defective as a result of defective materials or workman-
ship, it shall be repaired or replaced as aforesaid. Our obligation does not include the cost of
furnishing any labor in connection with the installation of such repaired or replaced equip-
ment or parts thereof, nor does it include the responsibility or cost of transportation. In ad-
dition, instead of repairing or replacing the equipment returned to us as aforesaid, we may,
at our option, take back the defective equipment, and refund in full settlement the purchase
price thereof paid by Buyer.
The warranty shall not apply to any equipment (or part thereof) which has been tampered
with or altered after leaving our control or which has been replaced by anyone except us,
or which has been subject to misuse, neglect, abuse or improper use. Misuse or abuse of the
equipment, or any part thereof, shall be construed to include, but shall not be limited to,
damage by negligence, accident, fire or force of the elements. Improper use or misapplica-
tions shall be construed to include improper or inadequate protection against shock, vibra-
tion, high or low temperature, overpressure, excess voltage and the like, or operating the
equipment with or in a corrosive, explosive or combustible medium, unless the equipment
is specifically designed for such service, or exposure to any other service or environment of
greater severity than that for which the equipment was designed.
The warranty does not apply to used or secondhand equipment nor extend to anyone other
than the original purchaser from us.
Contents | xiii
This page intentionally left blank.
Sensor Operations
Designed for fast response in a wide range of flue gas applications, the
sensor mounts directly on the combustion process to provide continuous
measurement of oxygen, or oxygen and combustibles, or oxygen, combus-
tibles and methane.
Do not turn on the aspirator until the analyzer has been turned on
and is hot - preferably 24 hours. Also, if you turn off the analyzer or
the process is shut down, turn off the aspirator air to avoid plugging
problems!
Remove AC power from the sensor and control unit and allow the
sensor to cool for at least one hour before performing any maintenance
or troubleshooting activities.
The outside of the sensor cover and all sensor assembly components
are hot during normal operation (up to 500° F, 260° C inside the
cover). Allow sensor components to cool for at least an hour before
working inside the sensor. Use caution and wear appropriate gloves
when handling components or when touching the sensor cover!
Overview | 1-1
Sensor
The Plumbing
All inlet and outlet tubing (cell housing), the oxygen cell, the oxygen
cell fitting, and the combustibles flow block and detector (for combus-
tibles). Also includes the calibration gas inlet and aspirator assembly,
with the aspirator used to pull the sample into the sensor.
Since the voltage of the cell is temperature dependent, the cell is main-
tained at a constant temperature. Some newer high temperature insitu
models use the heat from the process to heat the sensor, and the process
temperature is continuously measured and used in the software calcula-
tion. The oxygen content is then determined from the Nernst equation:
RT O
E= In 1
4F O2
20.9%
E = A*T*Log AT = 48.0 at 695°C
O2 Unk%
Overview | 1-3
The cell produces zero voltage when the same amount of oxygen is on
both sides, and the voltage increases as the oxygen concentration of the
sample decreases. The voltage created by the difference in the sample gas
and the reference air is carried by a cable to the microprocessor control
unit, where it is linearized to an output signal.
Hydrocarbons
Hot-Wire Type
WDG-IVCM analyzers with hot-wire type combustibles detector (see
Figure 1-2) are used for processes that use natural gas for start-up and oil,
fuel gas or coal for primary fuel. Process gas is drawn into the analyzer’s
primary sample loop by an aspirator and returns to process. Gas enters
the split flow block that is separted by a baffle and contacts the hot-wire
catalytic combustibles and methane detector. It then passes the zirconium
oxide cell, where the oxygen concentration is measured, before returning
to the main sample loop and then back to the process.
CELL HOUSING
ZIRCONIUM OXIDE
OXYGEN CELL
CONVECTION LOOP
RETURN
PPM OR PERCENT
COMBUSTIBLES DETECTOR METHANE DETECTOR
ASPIRATOR BLOCK
RETURN TO
PROCESS
SAMPLE
ASPIRATOR AIR
Figure 1-2. WDG-IVCM with hot-wire type combustibles detector convection loop.
Overview | 1-5
RTD Type
WDG-IVCM analyzer with RTD-wire type combustibles detector (see Fig-
ure 1-3) is used for natureal gas fired boilers (natural gas for start-up and
primary fuel). Process gas is drawn into the analyzer’s primary sample
loop by an aspirator and return to process. Gas enters the RTD-type com-
bustibles detector flow block and contacts RTD-type combustibles detec-
tor. The gas then travels into the hot-wire methane detector flow block
where it contacts the methane sensor. It then passes the zirconium oxide
cell, where the oxygen concentration is measured, before returning to the
main sample loop and then back to the process.
CELL
HOUSING
ZIRCONIUM OXIDE
OXYGEN CELL
CONVECTION
LOOP
RETURN
HOT-WIRE TYPE
METHANE DETECTOR
RTD-TYPE
COMBUSTIBLES DETECTOR
ASPIRATOR BLOCK
RETURN
TO PROCESS SAMPLE
ASPIRATOR AIR
Figure 1-3. WDG-IVCM with RTD-wire type combustibles detector convection loop.
Combustibles
The Combustibles Detector is a dual element device. The elements differ
only in that one is coated with a catalyst. The catalyst causes oxidation
to occur at a lower-than-normal temperature. The temperature of the
catalyzed element changes as the combustible mixture burns. The tem-
perature change causes the resistance of the catalyzed element to change.
The resistance change is interpreted by the microprocessor and the corre-
sponding combustibles reading is displayed.
Methane
The Methane Detector is a dual element device. The elements differ only
in that one is coated with a catalyst. The catalyst causes oxidation to occur
at a lower-than-normal temperature. The temperature of the catalyzed
element changes as the mixture burns. The temperature change causes
the resistance of the catalyzed element to change. The resistance change
is interpreted by the microprocessor and the corresponding reading is
displayed.
The Thermox catalytic methane detector will detect methane gases pres-
ent in a sample. The sample must, however, contain enough oxygen to
fully burn the methane present.
The methane span gas should be 2% CH4 with 8 - 10% oxygen present in
the mixture.
Overview | 1-7
Common Operator Errors
If you follow the steps below, your analyzer will operate with a minimum
of maintenance and troubleshooting.
Connect the air supply to the aspirator air supply inlet but DO NOT
turn on the air until the sensor is at operating temperature. If the ana-
lyzer is turned off or the process is shut down, the aspirator air should
be turned off to avoid plugging problems.
Do not use pipe dope or any other contaminant that gives off combus-
tible vapor on any joints of the sample tubing.
When working on the plumbing inside the sensor cabinet, turn the
power off. The oxygen cell heater has exposed windings and a short
to the plumbing will blow the fuse and could damage the furnace or
thermocouple.
Do not handle the cell excessively. Do not try to clean the cell except
by rinsing. If you need to handle the cell, grasp by touching the seal
fitting at the top; never touch the bare part of the cell.
Do not remove a cell or type “K” thermocouple that you may want to
use again when the inside of the furnace is still hot - severe thermal
shock can be destructive to either of them.
Always replace the metal cell O-ring when replacing the oxygen cell.
- If the sensor is on the pipe nipple, insulate the pipe from the process
wall to the mounting flange on the sensor.
This type of sensor requires a clean, dry instrument air supply for the
aspirator. Connect the air supply to the aspirator air supply inlet but
DO NOT turn on the air until the sensor is at operating temperature.
If the analyzer is turned off or the process is shut down, the aspirator
air should be turned off to avoid plugging problems.
Interconnect the control unit and the sensor per the appropriate inter-
connecting drawing.
Connect AC mains supply to the sensor. Make sure that the AC voltage
is within specified limits.
Overview | 1-9
Technical Support
Before you call the factory for technical support, run test gases and record
the following values (you may be asked by the factory to provide this
information when receiving service):
Cell millivolts
Thermocouple millivolts
Cell temperature
See the “Display” section in the Controller / User Interface chapter for
help on how to display this information on your Series 2000 control unit.
If you need to return equipment, you will be asked to provide the follow-
ing information before obtaining a Return Material Authorization (RMA)
number.
Model number
Serial number
Telephone number
Before returning material, you must get an RMA number from the
factory.
NOTE
Two isolated linear current outputs for oxygen and one each for combusti-
bles and methane if you purchased those options. Select O2, combustibles
(optional), methane (optional), cell temperature, thermocouple mV or cell
Analog Output
mV. Each output can be 4-20 mA, 0-20 mA, 20-4 mA or 20-0 mA and is
fully scalable. Hold or track during calibration and select degree of damp-
ing. Maximum load 1200 ohms.
Two independent oxygen alarms, each high or low selectable. One alarm
can be allocated to % oxygen, calibrate or verify. Two high combustibles
Alarms
and two high methane alarms (optional). Set relays to energize or de-
energize on alarm.
Watchdog timer and service alarms. System test for A/D, RAM, EEPROM
Diagnostics and keypad. Display Line 4 reserved for full text error and diagnostic mes-
sages. Twenty-entry event log.
Specifications | 2-1
Sensor : WDG-IV
Output Range From 0-1% to 0-100%
Accuracy ± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
Response 90% of a process step change < 6 secs.
< 0.1% of cell output per month (< 0.005% O2 per month with 2% O2
Drift
applied)
Aspirator Air
10 to 20 scfh (4.72 to 9.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)
Requirements
Environment
Ambient Temp.: -5°F to 160°F (-20°C to 71°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity: 10% to 90%, non-condensing
Enclosure
Lift-off NEMA 3R, weather resistant, stainless steel. Optional hinged NEMA 4X (IP56), explo-
sion-proof, purged, and floor mount versions available.
115 VAC, ± 10%, 47-63 Hz, 600VA 230 VAC, ± 10%, 47-63 Hz, 1850
Power Requirements max. (650 VA max. floor mount op- VA max. (1900 VA max. floor mount
tion) option)
NOTES
1. All static performance characteristics are with operating variables constant.
2. System accuracy referenced to 0.1 to 10% calibrated range.
3. Response is to calibration gas.
Environment
Ambient Temp.: -5°F to 160°F (-20°C to 71°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity: 10% to 90%, non-condensing
Enclosure
Lift-off NEMA 3R, weather resistant, stainless steel. Optional hinged NEMA 4X (IP56), explo-
sion-proof, purged, and floor mount versions available.
115 VAC, ± 10%, 47-63 Hz, 600VA max. 230 VAC, ± 10%, 47-63 Hz, 1850 VA max.
Power Requirements
(650 VA max. floor mount option) (1900 VA max. floor mount option)
Specifications | 2-3
Sensor : WDG-IVCM
Output Range From 0-1% to 0-100%
Oxygen From 0-1% to 0-100%
Combustibles From 0-2,000 ppm to 0-10,000 ppm
Methane 0 to 5%
Accuracy
Oxygen ± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
Combustibles ± 2% of full scale output range
Methane ±5% of full scale output range
Response O2 90% of a process step change < 13 secs.
< 0.1% of cell output per month (< 0.005% O2 per month with 2% O2
Drift
applied)
Aspirator Air
10 to 20 scfh (4.72 to 9.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)
Requirements
Environment
Ambient Temp.: -5°F to 160°F (-20°C to 71°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity: 10% to 90%, non-condensing
Enclosure
Lift-off NEMA 3R, weather resistant, stainless steel. Optional hinged NEMA 4X (IP56), explo-
sion-proof, purged, and floor mount versions available.
115 VAC, ± 10%, 47-63 Hz, 700VA max. 230 VAC, ± 10%, 47-63 Hz, 1950 VA max.
Power Requirements
(750 VA max. floor mount option) (2000 VA max. floor mount option)
NOTES:
1. All static performance characteristics are with operating variables constant.
2. System accuracy referenced to 0.1 to 10% calibrated range.
3. Response is to calibration gas (without flame arrestors).
Specifications | 2-5
Sensor : WDG-IVC - 0-500 PPM
Output Range From 0-1% to 0-100%
Oxygen From 0-1% to 0-100%
Combustibles From 0-500 ppm to 0-10,000 ppm or from 0-1% to 0-5%
Accuracy
Oxygen ± 0.75% of measured value or ± 0.05% oxygen, whichever is greater
Combustibles ± 25 ppm between 0 to 500 ppm or ±5% of full scale output range
Response O2 90% of a process step change < 30 secs.
Aspirator Air
10 to 20 scfh (4.72 to 9.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)
Requirements
Environment
Ambient Temp.: -5°F to 160°F (-20°C to 71°C);
-5°F to 140°F (-20°C to 60°C) with Div. 2 Option
Relative Humidity: 10% to 90%, non-condensing
Enclosure
Hinged, stainless steel, weatherproof NEMA 4X (IP56). Flame arrestors standard. Optional
CENELEC and NEC Class I, Division 2 available.
115 VAC, ± 10%, 47-63 Hz, 600VA max. 230 VAC, ± 10%, 47-63 Hz, 1850 VA max.
Power Requirements
(650 VA max. floor mount option) (1900 VA max. floor mount option)
NOTES
1. All static performance characteristics are with operating variables constant.
2. System accuracy referenced to 0.1 to 10% calibrated range.
3. Response is to calibration gas (with flame arrestors)
System Compliance
EMC Compliance: 2004/108/EC
Safety Compliance: 73/23/EEC
Specifications | 2-7
This page intentionally left blank.
Do not turn on the aspirator until the sensor has been turned on and
is hot - preferably 24 hours. If you turn off the analyzer or the process
is shut down, turn off the aspirator air to avoid plugging problems!
Analyzer Location
Observe the following guidelines when selecting an analyzer installation
location:
1. Select a readily accessible position for the analyzer to allow for routine
maintenance. Comfort levels for maintenance personnel should be
considered in placement of the sensor and control interface.
Installation | 3-1
This chapter shows you how to install your Series 2000 controller and
WDG-IV sensor components, and includes the following sections:
• Sensor Mounting
The sample inlet probe is how the sample gas gets to the sensor. Aspira-
tor air within the sensor is used to pull process gas from the sample probe
into the sensor. The sample inlet probe is identified by the threads on both
ends.
• Figure 3-1 shows how to install the standard 1/8” NPT probe.
You can connect the probe to the sensor before mounting the sensor un-
less the probe is either very long or made of ceramic. In that case you
should mount the sensor first and then install the probe.
Installation | 3-3
Standard 1/8” NPT probe installation
Connect a standard 1/8” NPT sample inlet probe to the sensor as follows:
• If your probe is 4 feet or less, simply screw the probe into the back of
the sample probe port on the rear of the sensor, then mount the sensor
- see Figure 3-1.
If you ordered a long probe or a ceramic probe, connect the probe to the
sensor as follows:
1. Mount the sensor. See the “Sensor Mounting” section later in this
chapter for help. Note, however, that if you are installing the probe
heater or exhaust tube, you must connect these items to the sensor
before mounting the sensor.
3. Remove the probe inlet hook from the sensor as shown in Figure 3-2.
To do this, loosen the fitting connecting the inlet hook to the aspirator
block and pull the hook out of the sensor.
4. Unscrew the bushing that was plugged into the sensor inlet port and
discard. The probe comes with its own bushing and fitting.
5. Thread the sensor probe into the sensor inlet port on the sensor
backplate. Do not let the probe hit against the Thermox-provided pipe
nipple as you insert the probe into the process. If inserting a ceramic
probe, you must slowly insert it into a hot process to prevent the
probe from cracking due to thermal shock.
6. Reconnect the probe inlet hook to both the sensor backplate and the
aspirator block and tighten.
Installation | 3-5
Probe Heater Installation
Do not install the probe heater if your sample gas contains a poten-
tially explosive mixture of combustibles; the probe heater can heat the
flue gas to the point of ignition.
The probe heater must be connected to the sensor before mounting the
sensor to the process.
NOTE
For installations where a standard setup allows the sample gas to cool
below its acid dewpoint before reaching the sensor, you can attach the
optional probe heater to the probe heater port on the backplate of the
sensor to heat the sample gas - see Figure 3-3. This will prevent plugged
plumbing due to condensation of the sample gas. For example, when a
fuel containing only a few thousand parts per million of sulfur is burned
with a high level of excess oxygen (over 4 percent), the resulting sulfuric
acid mist may condense at temperatures as high as 175 °C.
The probe heater comes shipped in an envelope, is 8.5” long, has a 3/8
NPT bushing and wires leading out of it.
EXHAUST PORT
SAMPLE PROBE
INLET PORT
TERMINAL BOX
1. Remove and discard the plug from the probe heater port as shown in
Figure 3-3. This port is plugged during shipment and should remain
plugged if no probe heater is installed.
2. Screw the probe heater into the 3/8” NPT probe heater port.
3. Route the wires from the probe heater to the sensor board.
4. Remove the sensor terminal box cover and connect the wires from the
probe heater to Terminals “M” and “N” on the sensor board - polarity
is not required.
Enclosed in the envelope with the probe heater is a stainless steel exhaust
tube that allows you to extend the aspirator exhaust. The exhaust tube can
be identified by having a thread on only one end of it.
To install the exhaust tube, thread the exhaust tube into the exhaust port
on the rear of the sensor as shown in Figure 3-3.
Sensor Mounting
• The sensor ambient temperature range is -20 °C to 70 °C.
• When installing a probe heater or exhaust tube along with the sensor,
attach it to the back of the sensor before mounting the sensor to the
process.
• If your sensor has a lift-off lid, allow at least 10” (25.4 cm) above the
sensor so you can remove the lid and access sensor components.
Installation | 3-7
Methods for mounting the sensor
There are three methods for mounting a WDG-IV sensor to the process:
Each of these methods is explained in the sections that follow. The flush
mount method is preferred because it positions the sensor as close to the
process as possible and minimizes the chances of the sample gas cooling
below its dewpoint.
Installation | 3-9
Sensor flush mount (Preferred Method)
Weld the 7” x 7 5/8“ (17.78 cm x 19.37 cm) mounting plate supplied with
the sensor to the process wall over a 3 3/4” (9.52 cm) diameter hole as
shown in Figure 3-6. Then bolt the sensor to the mounting plate. This
method positions the sensor as close to the process as possible and mini-
mizes any chance that the sample gas will cool below its dewpoint.
If you can’t mount the sensor flush with the process as described in the
previous section (for example, the process wall is curved sharply, covered
with insulation, or obstructed with reinforcing members) you can instead:
1. Weld a short 3 1/2“ diameter pipe nipple to the process wall and make
sure to extend the nipple through the skin of the process (see Figure 3-7).
Process Wall
Probe
Length minimum
necessitated by installation
Required 1” (25.4 cm) min. insulation
SIDE VIEW around exposed portion of nipple
Pipe Nipple
(supplied by customer)
Single stud
at “12 o’clock”
position
2. Weld the 7” x 7 5/8” (17.78 cm x 19.37 cm) supplied plate to the end of
the nipple, taking care to center it over the hole
3. Wrap the nipple with at least 1” (2.54 cm) weatherproof insulation and
heat trace it.
If mounting the sensor using this method, you can also install the probe
heater to minimize the possibility of the sample gas cooling below its dew-
point.
Installation | 3-11
Calibration/Aspirator Air/Z-Purge Connections
The span gas must be 10 times greater than the zero gas. For example, if
the zero gas is 1% O2, the span gas must be 10% O2 or higher.
O2 Span Gas
Instrument air (20.9%) or from 1.0% to 100% O2.
O2 Zero Gas
From .1 to 10% O2, balance N2
Z-Purge Connections
Connect instrument air to the sensor Z-purge inlet. Maintain the pressure
and flow as directed on the Z-purge warning label.
• You must provide a protective gas supply with an alarm for a loss of
pressure indication in order to meet NFPA 496 requirements.
• You must not exceed the maximum temperature rating listed in the
specifications.
Installation | 3-13
Z-Purge Shutdown Procedure
Power must not be restored after the enclosure has been opened until the
enclosure has been purged for 65 minutes at a pressure of 0.4” of water.
1. Connect aspirator air to the aspirator air inlet on the sensor (see Fig-
ure 3-9). Adjust the aspirator air regulator to the minimum pressure
required to get a response from the analyzer; typically this is 5 psi. Do
not turn on the aspirator until the sensor has been turned on and is
hot -preferably 24 hours.
• You must provide a protective gas supply with an alarm for a loss of
pressure indication in order to meet NFPA 496 requirements.
Installation | 3-15
Control Unit Mounting
The Series 2000 Control Unit comes with various mounting options.
Find the section below that corresponds to the type of control unit mount-
ing configuration you ordered, then mount as shown in that section.
Weatherproof - Panel
Weatherproof - Wall/Pole
Don’t open the Z-purge enclosure until you verify that the area has
been classified as nonhazardous.
Remove power from the control unit for one hour while maintaining
purge air flow before door is opened, unless the area is demonstrated to
be nonhazardous.
Do not restore power after the enclosure has been opened until the enclo-
sure has been purged for 1 hour at a pressure of 0.4” of water.
7.50
5.31
5.88 19.05
13.49
14.94
12.00
Flange
30.48
8.00
20.32
Pressure
Gauge
Exhaust
Restriction
1/2” Conduit
Metering Valve Entry (2.22 cm)
Installation | 3-17
9.10
12.6 23.11
32.0
5.22
General Purpose - Panel
6.57 13.26
16.69 5.00 Figure 3-13 shows the panel mounting
12.70
dimensions for the Series 2000 general
purpose enclosure. This version can be
10.00 mounted in panels up to 1” thick.
25.40 10-32
3.8 cm (1.5 in.)
10.45 4 places
26.54
12.6
9.10 General Purpose - 19” Rack
23.11
32.0
Figure 3-14 shows the mounting dimen-
1.48
3.76
sions for the Series 2000 general pur-
pose 19” rack mount enclosure. Use #10
5.22 screws to mount the control unit to a rack
13.26 whose holes are spaced per E.I.A. stan-
4.95
12.57 dard RS-310-C (universal spacing).
2.25
5.72
19.00
48.26
7.65
11.5 General Purpose - Wall
29.21
19.43 10.26
26.06
Figure 3-15 shows the mounting dimen-
sions for the Series 2000 general pur-
pose wall enclosure. Use 1/4” diameter
fasteners to mount the control unit to the
wall.
10.38 11.38
26.37 28.90
4 Holes for
1/4" fasteners
Connections to the control unit are made through the wiring card (see
Figure 3-16). The wiring card is located on the front-bottom of the control
unit for all weatherproof versions and for the general purpose wall mount
version (see the “Control Unit Mounting” Section earlier in this chapter
for help). For the general purpose rack and panel mount versions, the wir-
ing card is located on the rear of the control unit (Figure 3-23 shows the
orientation of the wiring card for general purpose rack and panel mount
versions - for all other versions, the wiring card is 180 degrees from this
view). To access the wiring card for all general purpose versions of the
control unit, unscrew the four screws on the wiring card chassis cover
plate. For the weatherproof versions of the control unit, first open the
front door of the weatherproof enclosure to gain access to the wiring card
shield. Remove the three screws on this shield to gain access to the wiring
card.
Any screw terminals on the wiring card not described in this section
are reserved for future use.
NOTE
Installation | 3-19
This wiring section shows you how to make the following connections:
• Use the conduit entry point closest to the connections you are making.
Do not add any additional conduit entry holes!
• You may use 1” conduit with a 3/4” adapter at the entry to the Series
2000 controller for signal and sensor wiring.
The Series 2000 control unit can operate using between 85 to 250 volts AC,
47 to 63 Hz. There is no power switch or circuit breaker on the control
unit, and it must be protected by installing it on a circuit-protected line,
maximum 15 amperes, with a switch or circuit breaker in close proximity
to the control unit and within easy reach of an operator. Mark the switch
or circuit breaker as the control unit disconnecting device.
Installation | 3-21
Use the 1/2” conduit entry hole in the control unit for AC mains supply
wiring. Use the chassis stud next to the 1/2” conduit entry hole for equip-
ment ground (protective conductor).
AC (L) and (N) markings are provided by the terminal block for connec-
tion of AC power. These markings are for reference purposes only, such as
for use on system wiring diagrams, etc. The system/product has or needs
no specific LINE or NEUTRAL connection for any function, safety or oth-
erwise. The (N) terminal is not internally grounded, nor needs to be. The
system will operate normally regardless of what AC input terminal (L or
N) the AC Line or Neutral is connected to, or, if there is a Neutral used at
all (i.e., 208 VAC US power connection).
You must use twisted-pair cable in rigid metal conduit or use twisted pair
cable with an overall braided shield. All cable shields or conduits connect-
ing to the control unit must be chassis grounded.
Conduit Nut
Cable Shield(s)
Conduit Housing Max. Length 2 inches
Nut
Washer
Ring Terminal
• Although there are transient and noise protectors on all control unit
I/O connections (communications, current outputs, sensor, etc.), this
protection is intended to act as a last line of defense against unwanted
transient and RFI interference.
Installation | 3-23
inductive loads. Be sure to place the transient suppressor as close to
the load as possible. Examples of transient suppressors include MOVs,
TRANSORBs, and RC snubbers.
• AC mains supply wiring should not be run in the same conduit with
mains supply wiring that feeds heavy inductive loads.
• Avoid running signal wiring in the same cable or conduit with wires
that power inductive loads unless all the cables within the conduit are
shielded, the inductive loads are small, and transient suppressors are
used at the loads.
• Do not run signal lines in the same cable or conduit with high voltage
lines.
Sensor Connections
Make sure that power has been removed from the control unit and
sensor before making sensor connections. See Figure 3-25, Sensor
Board.
L - Line Connection
N - Neutral (USA)
Chassis Stud -Equipment Ground (Protective conductor)
AC (L) and (N) markings are provided by the terminal block for connec-
tion of AC power. These markings are for reference purposes only, such as
for use on system wiring diagrams, etc. The system/product has or needs
no specific LINE or NEUTRAL connection for any function, safety or oth-
erwise. The (N) terminal is not internally grounded, nor needs to be. The
system will operate normally regardless of what AC input terminal (L or
N) the AC Line or Neutral is connected to, or, if there is a Neutral used at
all (i.e., 208 VAC US power connection).
Installation | 3-25
Standard sensor connections
All WDG sensors, regardless of the sensor type, require that you make
the following connections from the control unit to the sensor (see Figure
3-27):
Combustibles connections
If you have a combustibles detector in your sensor, also make the fol-
lowing connections between the control unit wiring card and the sensor
board on the sensor (H2 represents the hydrogen/carbon monoxide style
of combustibles detector. This detector can measure both CO and H2 ) -
refer to Figure 3-27:
Methane connections
If you have a methane detector, also make the following connections (see
Figure 3-27):
Installation | 3-27
Remote Sense Setup Wiring for Combustibles
If you do not have the combustibles option, skip this section.
If the distance between the controller and the sensor is over 40 feet,
and you are using the combustibles option, you must use the re-
NOTE mote sense setup. With the remote sense setup, you can obtain the
best accuracy of your combustibles detector readings. See Figure 3-17
for the number of pairs required and the type of wire required.
To use the remote sense setup, you must first set the Remote/Local jump-
ers on the Series 2000 controller wiring card to the Remote Position (see
Figure 3-21). Then you need to make another set of wiring connections
between the control unit wiring card and sensor as shown in Figure 3-21.
(You will end up having two wires connected to each combustible detec-
tor connection on the sensor board.)
Note:
(CH4 REF+)
(CH4 REF-)
(CH4 ACT+)
(CH4 ACT-)
(H2 REF-)
(H2 REF+)
(H2 ACT+)
(H2 ACT-)
Methane Combustibles
SERIES 2000 Option Option
WIRING BOARD
T/C-
CH4 REF-
H2 REF+
H2 REF-
H2 ACT-
CH4 REF+
H2 ACT+
CH4 ACT+
CH4 ACT-
(CH4 REF-)
(CH4 REF+)
(CH4 ACT+)
(CH4 ACT-)
(H2 REF+)
(H2 REF-)
(H2 ACT+)
(H2 ACT-)
15 16 13 14 9 10 7 8
M
O
R
E
E
T
L
+S
+S
+S
+S
-S
-S
-S
-S
IOUT3+ IOUT4+
IOUT3- IOUT4-
Current outputs 3 and 4 are isolated from the control unit and are
isolated from current outputs 1 and 2, but not from each other.
NOTE
These current outputs are only active if you have the combustibles option
card installed in your control unit.
Installation | 3-29
Alarm Contact Connections
This section describes how to make wiring connections for any alarm de-
vices you wish to connect to the control unit. Alarm contact information
is divided for standard alarms and alarms that are available only if you
have the combustibles card installed in your control unit.
The Series 2000 control unit provides four sets of standard alarm contacts,
which are labeled as follows (see Figure 3-29):
The Watchdog Timer Alarm connections are labeled as follows on the con-
trol unit:
ALARM 1A
ALARM 1B
The watchdog alarm contacts are normally closed during system opera-
tions (deactivated). A watchdog event triggers these alarm contacts to an
open condition. If you have a watchdog event, and the system returns to
normal operations, then the watchdog alarm will reset. The system will
retain in memory the last time the watchdog condition occurred.
ALARM 2A
ALARM 2B
The service alarm contacts are normally closed during system operations
(deactivated). A service alarm event triggers these alarm contacts to an
open condition.
ALARM 3A ALARM 4A
ALARM 3B ALARM 4B
Transient
Computer dry protector
contact input for inductive
loads
Load
Max. 30VA
Installation | 3-31
Combustibles detector alarm connections
If you have the combustibles option, two sets of alarm relay contacts are
available to activate for high combustibles levels, and are labeled as fol-
lows (see Figure 3-22):
ALARM 5A ALARM 6A
ALARM 5B ALARM 6B
If you have the methane detector option, two sets of alarm relay contacts
are available to activate for high methane levels, and are labeled as follows
(see Figure 3-29):
ALARM 7A ALARM 8A
ALARM 7B ALARM 8B
ZERO - 13 on RCU
ASPIR - 14 on RCU
If you have the combustibles option (and a combustibles RCU), you must
also make the following connection (see Figure 3-23):
SPAN2 - 17 on RCU
SPAN2 - 17 on RCU
SPAN2 - 17 on RCU
SPAN3 - 18 on RCU
Remote cal
switch normally
open (dry
contact closure)
Installation | 3-33
Digital input to initiate remote calibration unit
The digital input connections on the wiring card allow you to initiate a
remote calibration from a location other than the control unit (see Figure
3-23). For this option to work, you must have a remote calibration unit
and an auto calibration card. The system monitors the digital input, and
when the switch closes, it sends the system into an automatic calibration.
The switch you connect to this digital input must be a normally open switch.
DIGIN +
DIGIN -
2-wire connections
4WTX+/2W+
4WTX-/2W-
Connect all the 2-wire pluses (+) and all the 2-wire minuses (-) in parallel
with all the devices in a daisy chain fashion, including all the control units
on the network and the host computer. In addition, on the host computer
connect the receive and transmit terminal block connections (see Figure
3-24). Be sure to use twisted-pair cable for all connections.
The Series 2000 control unit is equipped with a termination resistor that
can be used for the last control unit on the network. Switch 3 on SW1 of
the display module allows you to place a 120 ohm termination resistor
into the RS-485 circuit:
Installation | 3-35
4-wire connections
The connections used for the RS-485 4-wire communications on the wir-
ing card are labeled as follows:
4WTX+/2W+
4WTX-/2W-4WRX+
4WRX-
Connect all control unit 4WTX+, 4WTX-, and all 4WRX+ and 4WRX- con-
nections in parallel in a daisy chain fashion. Make the following con-
nections between the host computer and the controller in a daisy chain
fashion (see Figure 3-25):
The Series 2000 Control Unit is equipped with termination resistors that
can be used for the last control unit on the network. The switch settings
are:
SW1-1 = ON
SW1-2 = ON
SW1-3 = ON
This provides 120 ohm termination resistors in the RS-485 circuit. See
the “Option Card Installation” section in this chapter for help on how to
remove and access the display module. You may need to experiment to
get optimum performance over long lengths of communications wiring to
see if having a termination resistor on the last control unit of the network
improves communications. Remember, though, only to insert a termina-
tion resistor on the last control unit on the network, and not on all control
units on the network. The factory default setting is to have Switch 3 OFF
with no termination resistor in the circuit.
Installation | 3-37
Option Card Installation
This section shows you how to add option cards to the control unit. This
includes the following option cards:
If you ordered any of these options when you purchased your analyzer,
these option cards will already be installed in the control unit. This section
shows you how to install an option card if you later add that option to the
control unit.
This section also shows you how to access the display module and power
supply/keypad module on the control unit in case you need to replace
these modules or change a component in them (for example, install new
software).
3. Pull on the display module handle to then pull out the display module.
1. Open the access door below the display module (hinges down).
3. Insert the option card into any of the three available slots underneath
the display module. Make sure the card is properly seated by pressing
on both sides firmly. If the card is not properly seated, the access door
may not close.
2. Pull with your left hand on the left of the option card while pulling on
the ejector lever on the right of the option card with your right hand
to remove the option card. The ejector will help apply extra force to
make is easier to release the card.
Installation | 3-39
This page intentionally left blank.
Key Function
Setup Set up the control unit for your specific application.
Calibrate Set calibration parameters and start calibrations.
Analog Range Define parameters that control analog current outputs.
Alarm Define parameters that control the alarm relays.
Help Access control unit on-line help.
Cancel Cancels out of menu options.
Enter Confirm menu choices and values.
For example, if you select the Display option from the Setup key, and then select
“Display Line 1”, the current choice on display line 1 will have an asterisk in front
of it (for example, % oxygen).
If all menu options are not able to be listed on the display, there will be a
Down arrow on the far right of the last line of the display. This indicates that
there are more menu options to view. When the Down arrow disappears,
there are no more menu options to display. Conversely, if an Up arrow ap-
pears on the far right of the first line of the display, there are more menu
options available if you press the Up arrow key.
Pressing the Cancel key does not delete any settings you have set
while in that menu.
NOTE
On-Line Help
Press the Help key to get a description of the function for the menu options.
Press Help again to exit on-line help and return to the menu options (or press
the Cancel or Enter). For example, to get on-line help for the Passwords op-
tion from the Setup key, select the Setup key, then move the arrow pointers
to Passwords and press Help.
System Password
By default, the system password is disabled. If you don’t want to
enable a system password, you can skip this section. You can also use
NOTE this menu option to disable the password requirement.
If you have forgotten the correct password, you can call AMETEK/
Thermox and a service password will be provided so that you can ac-
NOTE cess the Password option and enter a new system password.
Reenter to Verify
To later disable or change the system password, you must first enter
the old password to access this menu option, then change or disable
NOTE the password (0000 disables the password).
SETUP KEY
Display Passwords
Combustibles
SETUP KEY
Functions/Options
Menu Options
System Serial # View serial and manufacturing number for the control unit.
Primary
Perform an automatic, remote, or manual primary calibration
Calib
This is not a typical calibration. A primary calibration is intended to extend the life
of your sensor. Do not perform this type of calibration unless specifically instruct-
ed to do so.
Percent oxygen (% O2 )
Cell temperature
Cell millivolts
Thermocouple millivolts (T/C millivolts)
Current date/time
User text
Turn off (leave display line blank)
Combustibles (optional)
Methane (optional)
Combustibles/Methane (optional)
When you access menus, the menu options can occupy display lines 3
and 4, and sometimes display line 2. Therefore, place the information
NOTE that is most critical for you to view on the topmost display lines.
2. Select Display Line #1, Display Line #2, or Display Line #3, de-
pending on where you want to place the information.
3. Select the type of information you want to place on the display line.
The Down arrow found on the third line of the display allows you to see
your current position when entering user text. When you first select the
User Text option, this Down arrow moves to above the first position.
To enter text, use the Up and Down arrow keys on the keypad to scroll
through the ASCII text characters. When the letter you want to enter ap-
pears in position, press the Right arrow key to move to the next position.
You can also press the Left arrow key to move back one position in your
text message. When you are finished entering your message, press Enter.
3. Enter your process pressure (up to one decimal point is allowed - for
example, 1.5 kPa), then press the Enter.
The process pressure you entered will now appear on the third display
line.
4. If the pressure that has been entered is correct, press Enter to exit this
menu option. If the pressure was entered incorrectly, you can reenter
it at this point.
System Serial #
View system serial # and manufacturing #.
The Primary Calibration correlates the actual cell response for the desired 695°C
cell set-point temperature with the actual furnace thermocouple feed back tem-
perature. This is done by applying two known concentrations of oxygen to the
analyzer during the primary calibration. The software algorithm then calculates
the actual cell temperature from the cell mV response and increases or decreas-
es the furnace temperature control set point accordingly.
Because of the age of the cell and its characteristics, the control temperatures
should be slightly different. However if you perform a primary calibration incor-
rectly, or if the cell really starts to deteriorate, the actual cell response to calibra-
tion gas will drop off. The primary calibration is a way to extend the life of the
cell. The furnace control temperature can be increased to reduce the atrophy and
extend the cell life as the cell ages and the output starts to drop off its original
factory cal furnace set point temperature. This can be seen in the distribution in
your spreadsheet since the calculated cell temperatures for some of the analyz-
ers are well below 695°C or above by a fair amount. If the cell response goes
out of the limits in Table 4-1, there will be a zero or span gas range error. At this
point, you should perform a primary calibration and then attempt a standard cali-
bration again. If you get an excess cal error during primary calibration, then the
furnace temperature set point cannot be increased any further to prevent dam-
age to the cell housing and/or furnace assembly.
• +/- 10 mVs of the anticipated mVs for the gas concentration applied (air 0
mV expected).
• +/- 5 mVs of the anticipated mVs for the gas concentration applied (2% O2
applied, 48 mV expected at 695°C cell temp) as referenced to the actual
span gas mV reading.
Table 4-1.
If you take the displayed cell temperature and divide it by the T/C Ratio in memo-
ry location 88, you will have the actual furnace set point control temperature.
You can use a Type-K table to measure the voltage on terminals C and D on
the sensor board, and add the appropriate amount of mVs for the board ambi-
ent. The corresponding temperature from a Type-K thermocouple should be the
actual furnace set point temperature = 695°C / T/C Ratio (memory location 88).
Use the Type-K table to determine the mV change you have for 273°C and add
this mV change for the actual board temperature. This should be the compensa-
tion voltage measured between D+ and 6- on the sensor board. So for a 50°C
ambient at the board, the voltage should be 11.05 mV (273°C) + 2.02 mV (50°C)
= 13.07 mVs.
If you have a T/C ratio of 0.9, the corresponding furnace temperature set point
should be (695°C / 0.9 = 772°C). If you measure the voltage at terminals C and D
on the sensor board with a 50°C ambient, the voltage should be 32.125 (772°C)
- 2.02 mV (50°C) = 30.1 mV. If you add in the compensation voltage for 50°C, the
voltage at terminals 5 and 6 should be 30.1 mV + 12.05 mV = 42.15 mVs which
should also be the voltage that you measures at T/C+ and T/C - terminals on the
Series 2000 wiring card terminals and also the displayed T/C mV value that you
can set one of the display lines to view in the Setup key - Display menu.
For a cell operating temperature of 695°C, the T/C ratio low limit is .777 and the
high limit is 1.222, which corresponds to a thermocouple indicated temperature
range of 540°C to 848°C.
• If you find that a regular calibration isn’t working correctly, first check that
your calibration gas values match the values written on the gas cylinders;
also be sure that the correct flow of gases has been set and that the gases
are allotted enough time to flow through the sensor. Only if these checks
have been done and the system still can’t be properly calibrated should you
perform a primary calibration.
Make sure you have properly entered calibration gas values using the Cal
Gas Values option from the Calibrate key before starting your primary
calibration.
2. If you have the auto calibration card option (and a remote calibra-
tion unit), you will be prompted to select Auto Calibrate, Remote
Calibrate, or Manual Calibrate. If you don’t have the auto calibration
option, the system defaults to a manual calibration, and you will be
prompted to inject span gas into the calibration gas inlet port.
When the primary calibration is finished, the system will notify you
that a standard calibration is required by displaying a Calibration
Required message.
Oxygen readings on the display will not be accurate until the regular
calibration is performed.
When selecting a sensor, use Table 4-3 for the temperature each type of
sensor uses. If your type of sensor is not listed, find the operating tem-
perature for your sensor, then choose the sensor that uses that operating
temperature.
Use caution when selecting the type of sensor used with the Series
2000 Control Unit. Selecting the wrong sensor will damage sensor
components and could invalidate your warranty coverage.
2. Select the sensor you will use with the Series 2000 control unit. If your
sensor is not listed in Table 4-3, call the factory for the correct operat-
ing temperature for your sensor type.
Baud Rate Allows you to define the baud rate that the control unit and the
host computer are using to communicate. Choices are 300,
600, 1200, 2400, 4800, and 9600 baud.
Node Address Allows you to define the physical address for the control unit
(between 0 and 255) on a network so the host computer can
identify the control unit if using network communications (see
below). Note that each control unit must be assigned a unique
address; otherwise, communication results will be unpredict-
able.
Enable Serial Allows you to temporarily disable control unit serial communi-
cations.
Combustibles Option
Combustibles Choices are % combustibles or PPM combustibles.
Scale
The % combustibles full scale range is 0 to 5 %.
The PPM combustibles full scale range is 0 - 10,000 PPM.
If you have the methane option, the range is always in % methane (not select-
able), and the full scale range is 0 to 5 %.
If you have a methane detector, the combustibles detector must be used in the
PPM range.
Combustibles Choices are Hot Wire and RTD detector.
Type
If not sure, check your customer configuration sheet, in-
cluded in the front of this manual.
Set Current
Range
Set Track/Hold
Set Current
Mode
Output Filtering
Select Function
You can direct process readings to either of the two analog output ports
on the rear of the control unit. If you have the combustibles option, you
can also have process readings directed to two additional analog output
ports activated with this option. All functions in this chapter are directed
to a specific analog output port.
Select Function Define the information the analog output should track:
% O2
Cell temperature
Cell Millivolts
Thermocouple Millivolts
Combustibles (option)
Methane (option)
You can then select any of the menu options described in the remainder of
this chapter, and they will apply to the analog output port you selected.
See the “Set Current Mode” section for help on choosing either a 0-to-
20 mA or 4-to-20 mA scale.
NOTE
2. From the Analog Range key menu, select Set Current Range.
Out # y 20 ma = xx
New Value?
where y is the analog output port, and xx is the present value that the cur-
rent output currently represents (for example, 10 % O2 ). The output port
that appears depends on the analog port you selected.
Out # 1 4 ma = xx
New Value?
After you select one of these options, your choice will be confirmed on the
display:
2. From the Analog Range key, select Output Filtering. You will be
prompted to enter an output filtering value:
100 equals no filtering, and 1 equals the highest possible filter value.
The default value is 100 (no filtering).
NOTE
3. Type a number between 1 and 100, then press Enter. Note that you
should enter 01 to enter 1, 07 to enter 7, 70 for 70, etc.
Percent oxygen (% O2 )
Cell temperature
Cell millivolts
Thermocouple millivolts (T/C millivolts)
Combustibles (optional)
Methane (optional)
The combustibles and methane options are available only for current out-
puts 3 and 4, and cannot be set for analog output ports 1 or 2.
3. Select the function you want the analog output port to track.
The Series 2000 control unit then prompts you with the Set Current
Range menu option to choose the range of readings that the 0-to-20 mA
or 4-to-20 mA analog output represents (see the “Set Current Range” sec-
tion in this chapter for details).
ALARM KEY
Alarm Configure
Relay Configure
Select Function
Exception Log
Alarms 1 and 2 are hard wired for the Watchdog Timer and Service
alarms and are not software-selectable.
NOTE
Oxygen, combustibles, and methane alarms are all disabled during a
primary or regular calibration. In addition, alarms are disabled dur-
ing start-up until the sensor reaches its proper operating temperature.
See the “Wiring” section in the Installation chapter for help on alarm con-
nections and how alarms are labeled on the control unit wiring card.
Alarm 5
Not Selectable
Not selectable (high alarms only)
Alarm 6 (combustibles only)
Alarm 7
Not selectable (Disabled during
(methane only) calibration cycle)
Alarm 8
* The Select Function menu option applies only to Alarm 3. All other alarms are restricted
to activate based on process reading levels.
For oxygen alarms, these values depend on the Alarm
Alarm Set Points Configure menu option for whether they are high or low
alarms.
The exception log provides a history of events, such as
system or error messages, alarm conditions, and the
Exception Log
start of calibrations or verifications.
Alarms 1 and 2 are reserved for Watchdog and Service alarms and are not software-
selectable. They always de-energize on alarm (FAIL SAFE).
Once you select the Alarm Set Points menu option, the system
alerts you that Alarms 1 and 2 are reserved for the watchdog and ser-
NOTE vice alarms, and are not software selectable. Press Enter to continue
to the configurable alarms.
You will be prompted to define the Alarm 3 set point value (this option
won’t appear if Alarm 3 is set to activate based on the start of a calibra-
tion or verification - see the “Select Function” section in this chapter for
details):
2. Enter the Alarm 3 set point value, then press Enter (just press Enter if
you don’t want to change the current value).
If you don’t have the combustibles option installed, you are done (press
Cancel to exit this menu option).
Alarm #5= xx yy
New Value?
4. Enter the Alarm 5 set point value. Note that combustibles alarms are
always high alarms (can’t be set up as a low alarm).
Alarm #6= xx yy
New Value?
5. Enter the Alarm 6 set point value. Note that combustibles alarms are
always high alarms. So, for example, Alarm 5 can be high and Alarm 6
can be high/high.
If you don’t have the methane option, you are done (press Cancel key or
Enter to exit this menu option).
If you do have a methane detector, you are then prompted to enter the
Alarm 7 set point value:
6. Enter the Alarm 7 set point value. Note that methane alarms are al-
ways high alarms.
7. Enter the Alarm 8 set point value. Note that methane alarms are al-
ways high alarms. So, for example, Alarm 7 could be a high alarm, and
Alarm 8 high/high.
Alarm Configure
Define whether alarms should trigger based on a high alarm or a low
alarm condition. If you have the combustibles or methane options, Alarms
5 through 8 always act as high alarms. So, for example, Alarm 5 could be a
high combustibles alarm, and Alarm 6 could be a high/high combustibles
alarm.
Relay Configure
All Series 2000 control unit relays are Form A, normally open contacts.
You can use the software to set whether these alarms should close on
an alarm condition (energize on alarm) or open on an alarm condition
(de-energize on alarm). This menu option applies to all alarms except for
the service and watchdog alarms, which are always set to de-energize on
alarm (fail-safe).
Select Function
Select which function Alarm 3 should track.
• % Oxygen
• When the system starts a calibration
• When the system starts a verification
• When the system starts a calibration or a verification (dual function)
When tracking the start of a calibration or verification, the alarm will de-
activate once the calibration or verification has been completed, including
any recovery time you have specified.
The control unit automatically exits from this menu once you make your
selection.
The Service Alarm provides an alarm to alert you of possible system prob-
lems that require attention. Events that can trigger the service alarm to
activate are:
• Calibration failure
• Verification failure
Watchdog Relay
During normal system operation, the Watchdog Relay operates as
normally closed (system failure causes open contacts).
NOTE
The Watchdog Alarm relay de-energizes when the system watchdog timer
times out. The watchdog timer will time out and cause the control unit to
reset itself to prevent the control unit from becoming “lost” due to such
things as noise spikes or power surges. You can use this relay to activate a
remote indicator to notify you if the watchdog timer times out.
In addition, the control unit reads the watchdog after reset, and beeps for
a brief time if the reset was caused by a watchdog condition. Pressing the
reset keys simultaneously (hidden keys to the left and right of the Setup
key) for a long enough time will cause a watchdog. This will indicate as a
beep after the system restarts after releasing the reset keys.
• Alarm conditions
• Reset conditions
The system displays the most recent exception log entry on line four of
the display. It also displays when the event was recorded on line three.
Line three of the display indicates the order in which events are recorded
into the log, with number 1 being the most recent entry into the log. The
system is able to store the 20 most recent entries in this exception log.
Pressing the Down arrow key shows later exception log entries. Pressing
the Up arrow key shows more recent exception log entries. Press Cancel
at any time to exit this menu option.
CALIBRATE KEY
Sensor alarms are disabled during calibrations and the recovery time
after a calibration. You can set whether analog outputs should follow
NOTE the calibration readings or hold at the last process reading before the
calibration.
Calibration | 5-1
Definitions
• Automatic calibrations allow the system to calibrate itself. You define the
calibration gas values, the time each gas will flow through the sensor,
and the frequencies at which automatic calibrations will be performed
(for example, every four hours, every two days, etc.). The system cali-
brates itself at those predefined times. You can also use a digital input
to activate an automatic calibration (see the “Remote Calibration Unit
Connections” section in the Chapter 2). With the digital input method,
you can only initiate an automatic calibration, not an automatic veri-
fication. You can also start an auto calibration using the keypad - the
calibration gas run times are predetermined.
• For remote calibrations, you define the calibration gas values, but you
interactively determine how long each calibration gas should flow
from the RCU and into the sensor (by pressing a key on the control
unit).
• For a manual calibration, you must physically switch the gases because
you are not using an RCU. However, you still define the calibration
gas values.
Calibration | 5-3
Overview
CALIBRATE KEY
Functions/Options
Menu Options
Start an automatic, remote, or manual cali-
Initiate Cal/Initiate Verify
bration or verify
Although automatic calibrations can be set to run at predetermined frequencies,
you can also force the system into an automatic calibration.
View latest calibration and verify data.
Cal/Verify Data - Cal data
- Verify data
Define calibration/verify gas values to
match known values of calibration gas
Cal Gas Values cylinders.
Enter calibration gas values.
Set pressure and flow to RCU gases with-
Inject Cal Gas
out performing a calibration or verification.
Set control unit time/date.
Set Cal Timers/Set Auto Timers Enable or disable auto cal cycles.
(applies only for auto cal operations
with RCU, except for setting time and Set cal gas times for auto cal/verify opera-
date) tions.
Define recovery durations after a cal/verify.
Set K Factor Enter Value
(not used on WDG-IV Series)
View the Sensitivity (Slope) determined
Sensitivity during calibration for the combustibles
detector in mV/1000 PPM.
Aborting a Calibration
Pressing the Calibrate key while the system is performing a calibration
or verification, displays the option to abort the calibration or continue the
calibration. All other menu functions are temporarily disabled during a
calibration or verification. Oxygen, combustibles, and methane alarms are
also disabled during a calibration.
Calibration | 5-5
You can start the auto calibrate cycle as follows:
2. Select the Auto Calibrate or Auto Verify option. The system will begin
the automatic calibration, and will notify you how long the span gas
will be applied to the sensor from the RCU:
During this time, the RCU switched the solenoid valves and introduced
the zero gas to the sensor.
During this time, the RCU switched the solenoid valves and introduced
the combustibles span gas to the sensor.
If you set the duration time for the combustibles span gas to 0 from
the Set Cal Timers menu, you will not see this message.
NOTE
When specified, the system will display the recovery time remaining:
Recovery xx:xx
1. Enter the calibration gas values to match the span and zero calibration
gas cylinders you intend to use to calibrate/verify your system. These
values are referred to in this section as setpoint values.
2. Select Initiate Cal or Initiate Verify from the Calibrate key menu, de-
pending on whether you want to perform a calibration or verification.
3. Select Remote Calibrate or Remote Verify.
The following appears on your display (display line 2 will only appear if
you have the combustibles option; display line 3 will only appear if you
have the methane option):
4. Press Enter when the O2 reading on the far left of the first display line
stabilizes. The reading on the far right of the first display line shows the
setpoint value you entered using the Cal Gas Values menu option. This
setpoint value should match the cal gas cylinder value.
If your O2 reading on the far left of the display has stabilized, yet is not near the setpoint
value on the far right of the display, you should check the cal gas value you entered and
ensure that this value matches the cal gas cylinder for that gas in the RCU.
5. Press Enter when the zero gas reading(s) on the left of the display stabi-
lizes. If you don’t have the combustibles option, you only need to wait until
the O2 zero gas value stabilizes. If you have the combustibles option, wait
until all readings stabilize. The reading on the far right of the first display
line shows the setpoint value you entered for the O2 zero gas using the Cal
Gas Values menu option. This O2 zero gas setpoint value should match
the cal gas cylinder for the O2 zero gas connected to your RCU.
Calibration | 5-7
If you have the combustibles option, you are also prompted to calibrate
the combustibles span gas:
O2 xx.x %
Cmb yyyy Set yyyy
CH4 zzzz Set zzzz
Press ENTER to Span
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
System Calibrating
1. Enter the calibration gas values to match the span and zero calibration
gases you intend to use to calibrate or verify your system.
2. Select Initiate Cal or Initiate Verify from the Calibrate key menu, de-
pending on whether you want to perform a calibration or verification.
3. Select Manual Calibrate or Manual Verify.
You are prompted to apply the O2 span gas to the calibration gas inlet port
on the sensor (tubing must be free of oil and dirt):
5. Press Enter when the O2 reading on the far left of the first display line
stabilizes. The reading on the far right of the first display line shows the
setpoint value you entered using the Cal Gas Values menu option. This
setpoint value should match the cal gas cylinder for the O2 span gas.
If your O2 reading on the far left of the display has stabilized, yet is not near the setpoint
value on the right of the display, you should check the cal gas value you entered and
ensure that this value matches the cal gas cylinder for the gas you applied to the cal gas
inlet port of the sensor.
You are then prompted to apply the O2 zero gas to the calibration inlet
port of the sensor:
Calibration | 5-9
Calibration: Hot Wire Type Combustibles Detector
If you have a Hot Wire-type detector, the zero gas display appears:
Display line 2 will only appear if you have the combustibles option;
display line 3 will only appear if you have the combined combustibles
NOTE and methane option.
1. Press Enter when the zero gas reading(s) on the left of the display stabi-
lizes. The reading on the far right of the first display line shows the setpoint
value you entered for the O2 zero gas using the Cal Gas Values menu
option. This O2 zero gas setpoint value should match the cal gas cylinder
for the O2 zero gas.
You are also prompted to apply the combustibles span gas to the calibra-
tion gas inlet port of the sensor (tubing must be free of oil and dirt):
2. Press Enter when you have applied this calibration gas. You are prompted
to calibrate the combustibles span gas.
O2 xx.x %
Cmb yyyy Set yyyy
CH4 zzzz Set zzzz
Press ENTER to Span Comb
3. Press Enter when the combustibles span gas reading(s) stabilizes on the
far left of the display.
4. Press Enter when you have applied your methane span gas. You are
prompted to calibrate the methane span gas.
5. Press Enter when the methane span gas reading stabilizes on the far left
of the display.
If you specified a recovery time, the system will display the recovery time
remaining:
Recovery xx:xx
System Calibrating
Be sure to reinsert the 1/4” tube cap on the calibration inlet port on
the sensor once you have completed the manual calibration. This pre-
NOTE vents air from getting into the sample gas line during normal opera-
tions.
Calibration | 5-11
Calibration with RTD Type Combustibles Detector
If you are calibrating a 3-in-1 sensor with the RTD combustibles de-
tector, your combustibles span calibration gas must include methane.
NOTE
Display line 2 will only appear if you have the combustibles option;
display line 3 will only appear if you have the combined combustibles
NOTE and methane option.
1. Press Enter when the zero gas reading(s) on the left of the display stabi-
lizes. The reading on the far right of the first display line shows the setpoint
value you entered for the O2 zero gas using the Cal Gas Values menu
option. This O2 zero gas setpoint value should match the cal gas cylinder
for the O2 zero gas.
If you have the Comb. & CH4 span gas option, you are also prompted
to apply the Comb. & CH4 span gas to the calibration gas inlet port of
the sensor (tubing must be free of oil and dirt):
2. Press Enter when you have applied this calibration gas. You are prompted
to calibrate the Comb & CH4 span gas.
O2 xx.x %
Cmb yyyy Set yyyy
CH4 zzzz Set zzzz
Press ENTER to Span Comb
3. Press Enter when the Comb & CH4 span gas reading(s) stabilizes on the
far left of the display.
Recovery xx:xx
System Calibrating
Be sure to reinsert the 1/4” tube cap on the calibration inlet port on
the sensor once you have completed the manual calibration. This pre-
NOTE vents air from getting into the sample gas line during normal opera-
tions.
Calibration | 5-13
Cal/Verify Data
View the results of the latest calibration or verification. This includes the
following:
Span Gas and Zero Gas Refer to the calibration gas values entered
using the Cal Gas Values menu option. These
values are fixed.
Span Value or Zero Value Value the system reads from the cal gas.
Span Drift or Zero Drift Difference between the entered calibration gas
value and the calibration gas values read by
the system.
Span Value: xx % O2
Span Gas: xx % O2
Span Drift: xx % O2
The term span value on the display is the value the system read during
the calibrate/verify operation, and the term span gas is the setpoint value
entered using the Cal Gas Values option from the Calibrate key menu.
Zero Value: xx % O2
Zero Gas: xx % O2
Zero Drift: xx % O2
4. Press Enter.
The time when the calibration or verification was completed will then ap-
pear on the display:
If the date or time has not been set, random characters will appear instead
of the time and date when the calibration or verification was completed. If
this occurs, reset the control unit’s time and date.
Calibration | 5-15
Cal Gas Value
Enter span and zero calibration gas values to match the calibration gas
cylinders you use to calibrate or verify your system. These values will be
used for all automatic, remote, or manual calibrations and verifications.
The O2 span gas value you enter must be higher than the O2 zero gas
value.
NOTE
2. Enter the span gas value, then press Enter. You can also include a
decimal point (for example, 20.9% O2 ). If you want to keep the current
span gas value unchanged, press Enter without entering a value.
3. Enter the zero gas value, then press Enter. You can enter up to two
decimal points (for example, 2.00% O2 ). If you want to keep the cur-
rent zero gas value unchanged, press Enter without entering a value.
Combustibles Option
If you have the combustibles option, you are then prompted to enter
the combustibles span gas value:
6. Enter the methane span gas value, then press the Enter key. If you
want to keep the current methane span gas value unchanged, press
the Enter key without changing the value.
Calibration | 5-17
Inject Cal Gas
Manually adjust the pressure and flow of your calibration gases (pressure
of calibration gas cylinders and flow into sensor) by cycling through the
solenoid valves on the RCU. No calibration or verification data is collected
while in this menu option. See the “Mechanical Installation” section in
Chapter 2 for recommended calibration gas pressure and flow require-
ments.
This menu option requires that an auto cal card and RCU are in-
stalled.
NOTE
2. From the Calibrate key menu, select the Inject Cal Gas menu option.
The following message will appear on your display, indicating that the
RCU has introduced the span gas to the sensor:
You will be prompted to press Enter when you have manually set the
pressure and flow for the span gas:
3. When you have adjusted the pressure and flow for the span gas to the
desired levels, press Enter.
The following message will appear on your display, indicating that the
RCU has now switched the solenoid valves and has introduced the zero
gas to the sensor:
4. Adjust the pressure and flow for the zero gas to the desired levels. If
you don’t have the combustibles option, press the Cancel key to end
the cycle.
CANCEL to quit
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
If you have the combustibles option, you will be prompted to press Enter
when you have set the pressure and flow for the zero gas:
5. When you have adjusted the pressure and flow for the zero gas to the
desired levels, press Enter.
The following message will appear on your display, indicating that the
RCU has introduced the combustibles span gas to the sensor:
Once you have set the pressure and flow for the span gas, the system
prompts you to press the Cancel key to end the cycle:
CANCEL to quit
If you have the methane option, you will be prompted to press Enter
when you have set the pressure and flow for the zero gas:
6. When you have adjusted the pressure and flow for the zero gas to the
desired levels, press Enter.
The following message will appear on your display, indicating that the
RCU has introduced the combustibles span gas to the sensor:
Once you have set the pressure and flow for the span gas, the system
prompts you to press the Cancel key to end the cycle:
CANCEL to quit
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
Calibration | 5-19
Set Cal Timers
Recovery Duration
Sub-Menu Function
Set Auto Timers Set auto timer for desired auto calibration/verification
cycle times.
Set Time and Date Set the current time and date.
Cal Gas Duration Define how long cal gases should flow through the sen-
sor during automatic calibrations or verifications.
Recovery Duration Define a recovery duration during which time the sensor
switches from measuring calibration gas readings to
measuring process readings without affecting the ana-
log output or triggering of alarms.
1. From the Set Cal Timers menu, select Set Auto Timers.
If you are setting the auto timers for the first time, the Series 2000
software will automatically prompt you to enter the current time and
NOTE date.
You will be prompted to enter the frequency for the calibration or verifica-
tion cycle (either in days or hours, depending on what you selected:
4. Enter the number of days (or hours) for the automatic calibration or
verify cycle. As you enter the values, they overwrite the XXs on the
display. To enter 1 day or 1 hour, enter 01; otherwise, the system will
assume you are entering 10 days (or hours).
You will be prompted to enter the time of day when the first calibra-
tion of the cycle should begin:
5. Enter the time you want the first calibration or verification to begin,
then press Enter. The system uses military time, so to enter 4:00 p.m.,
you would type 16:00. If, for example, you set the cycle frequency at
8 hours, and specify the first calibration to take place at 4:00 p.m., the
next automatic calibration or verification will take place at midnight.
Calibration | 5-21
Set Cal Timers - Set Time and Date
Set the date or time for the internal control unit clock.
Make sure the current time and date are correct before setting any
auto cal/auto verify parameters.
NOTE
1. From the Set Cal Timers menu, select Set Time & Date.
2. Enter the new time, then press Enter. As you type the new time, it
overwrites the Hr: Mn text. The Series 2000 control unit uses military
time so 16:00 equals 4:00 p.m., and 04:00 equals 4:00 a.m. If the current
time is correct, press Enter without changing the value.
3. As you type the new date, it overwrites the mo/da/yr text on the
display. Enter the month, the day, the year, then press Enter. If the
current date is correct, press Enter without changing the value.
1. From the Set Cal Timers menu, select Auto Cal On/Off.
3. Select Disable Auto Cal (or Disable Auto Verify) to disable the auto
calibration or auto verification cycle. Select Enable Auto Cal (or En-
able Auto Verify) to enable the auto calibration or verification cycle.
To find out how long it takes for your calibration gases to reach the sen-
sor and stabilize, select the Inject Cal Gas menu option and time the gas
stabilization period. Then use this time frame as your calibration gas dura-
tion (we suggest that you also add a buffer to this time frame to be certain
that the gas has stabilized).
1. From the Set Cal Timers menu option, select Cal Gas Duration.
You will be prompted to enter the new span gas duration time:
2. Enter the new span gas duration time, then press Enter. Mn equals
minutes and Sc = seconds. As you type the new span gas duration
time, it overwrites the Mn:Sc text. If you want to use the currently
displayed time, press Enter without entering a new value.
At this point, you will be prompted to enter the zero gas duration time:
3. Enter the new zero gas duration time, then press Enter. Mn equals
minutes and Sc = seconds. As you type the new zero gas duration
time, it overwrites the Mn:Sc text. If you want to use the currently
displayed time, press Enter without entering a new value.
Calibration | 5-23
Combustibles and Methane Options
You will be prompted to enter the new combustibles span gas duration time:
Enter the new combustibles span gas duration time, then press Enter. You
can set this calibration gas time to 0 if you don’t want to include it in your
automatic calibration sequence. If you want to use the currently displayed
time, press Enter without entering a new value.
You will be prompted to enter the new methane span gas duration time:
Enter the new methane span gas duration time, then press Enter. You
can set this methane gas time to 0 if you don’t want to include it in your
automatic calibration sequence. If you want to use the currently displayed
time, press Enter without entering a new value.
This menu option also can be used to prevent triggering of alarms if, for
example, your low alarm is set to activate if readings fall below 3%, and
your calibration gas is 2%.
This menu option applies to all automatic, remote, and manual cali-
brations and verifications.
NOTE
2. Enter the recovery time, then press Enter. If you want to use the cur-
rently displayed recovery time, press Enter without entering a new
value.
This menu option is not used on the WDG IV Series 2000. The de-
fault factor is always set to 1.00.
NOTE
Sensitivity
Select this option to display the Sensitivity (slope).
Calibration | 5-25
This page intentionally left blank.
This chapter describes Series 2000 control unit system and error messages.
It also provides troubleshooting assistance. System and error messages
appear on the fourth line of the display at three second intervals until the
condition is corrected or has ended. If there is more than one message, the
messages will alternate on the display at three second intervals. These sys-
tem and error messages are listed in this chapter alphabetically to make
them easier to locate.
System or error messages will not appear on the display when you are
using the control unit keys to access menu functions.
NOTE
Indicates that the calibration has been aborted by the user, or the
Calibration Aborted calibration failed because the span gas or zero gas was out of
range.
Message Explanation
Cell is Over Temp Indicates that the sensor temperature is 30°C or more over its cor-
rect operating temperature. The software shuts off the furnace until
the temperature returns to a normal operating range.
If this message soon goes away, it indicates that the temperature
has returned to an acceptable range. If there is a problem with the
sensor temperature control system, you will see another error mes-
sage after this message appears.
Combustibles Cal Error Indicates that the combustibles detector calibration failed because
of an abnormally low signal. Once this error occurs, calibration of
the system is aborted. This message indicates a problem either with
your calibration setup or the combustibles detector.
Excessive Cal Error If a primary calibration cannot correct the sensor temperature within
the software’s allowable limits, this message displays to indicate the
primary calibration has failed. This message could indicate one of
the following problems:
• Calibration gas setup problem
• Process pressure incorrectly entered
• Failed cell
• Leak or plug in plumbing
Memory is Corrupted Indicates that the control unit’s internal EEPROM memory has
been corrupted, or a significantly new version of software has been
installed in the control unit. Under these conditions, the software will
reprogram the EEPROM with the factory default values. You can
acknowledge this error message by setting or reading any user pro-
grammable value, and this will clear the message from the display.
This message will not clear by removing and restoring power to the
unit.
If you receive this message, you should reset your control unit to
the values required for your application, including Setup key param-
eters, Calibrate key parameters, etc...
Methane Cal Error Indicates that the methane detector calibration failed because of
an abnormally low signal. Once this error occurs, calibration of the
system is aborted. This message indicates a problem either with
your calibration setup or the methane detector.
Wiring Checks
Thermocouple Checks
AC Power Checks
Furnace Checks
Cell Checks
Exercise care when working on the sensor. Turn off power, allow the
unit to cool, and wear gloves.
Open thermocouple
Remove power to the control unit and the sensor. Measure
Thermocouple Checks across terminals “C” and “D” on the sensor board with an
ohm meter. If an open is measured, replace the thermo-
couple.
First, have one of the control unit display lines show the
sensor cell temperature. If the thermocouple leads are
reversed, the displayed temperature will be decreasing as
the sensor warms up (this will usually happen at start-up
Reversed Thermocouple Wires or after you replace a thermocouple). This indicates that
the thermocouple wiring is reversed. If you just replaced a
thermocouple, check the leads from the thermocouple to
the sensor board: yellow wire connects to terminal C (+),
red wire connects to terminal D (-).
To check calibration gas Check that the correct calibration gas values have been en-
values: tered into the control unit.
Select the Cal Gas Value Check that the calibration gas cylinders are turned on and are
menu option from the Cali- not empty.
brate key on the control unit Check for the proper flow rate and proper delivery pressure
and check that the calibra-
of calibration gases when the remote calibration unit has
tion gas values entered been activated (see the Inject Cal Gas menu option from the
match the analyzed concen- Calibrate key chapter for help).
tration of the cylinders.
Calibration Line Check Ensure that your calibration line is not contaminated with such
things as pipe dope, cutting fluid, oil, or solvents. All these
The best calibration gas to contaminants produce hydrocarbon vapors that interfere with
use for this check is an O2 the proper calibration of your analyzer, resulting in lower than
zero calibration gas. expected oxygen readings. To test for contaminated lines, you
must temporarily bypass your current calibration line with a clean
calibration line (directly from cylinder to the sensor calibration
inlet port, using a flow meter to set the proper flow) and compare
the response with that from the possibly contaminated line.
Aspirator Not Turned OFF WDG-IV only
during Calibration
Verify that the aspirator air is turned off during calibration (man-
ual or through the RCU). If the problem is with the RCU, refer to
NOTE
the next section, “Remote Calibration Unit Problem.”
Remote Calibration Unit The RCU contains one normally open solenoid-the aspirator air
(RCU) Problem solenoid. All other solenoids are normally closed. Problems with
the RCU are usually as follows:
Plumbing Leak To check for plumbing leaks, disconnect power from the RCU
and pressurize the inlets. Apply a leak detecting liquid along
the base of the solenoids and any plumbing fittings. Repair any
leaks found.
Solenoid Not The solenoid drive signal is a 12 VDC signal. This is used to
Energizing close the aspirator solenoid and open the appropriate calibra-
tion gas solenoid. To test a solenoid valve, select Inject Cal Gas
from the Calibrate key menu to energize that solenoid. Verify
that the solenoid drive signal is present. The aspirator air sole-
noid is closed when the drive signal is applied. Other solenoids
are open when the drive signal is applied (cal gas flowing). With
the drive signal present, verify the proper flow. If no signal is
present, check the interconnecting wiring between the control
unit and the RCU.
If wiring is correct, replace the Auto Cal Card. If the solenoid
drive signal is present, it indicates a problem with one of the
solenoid valves. Replace the solenoid valve, being sure that the
solenoid O-ring seals are properly positioned.
If performing a manual calibration, ensure that you are waiting for the
reading to stabilize on the display before switching to the next calibra-
tion gas. You should wait a minimum of two minutes.
NOTE
AC Power Checks
Loss/Inadequate AC Voltage Measure the AC voltage to the sensor board at terminals L1
to the Sensor and L2. Ensure that this voltage is sufficient. Check the mea-
surement technique used by the Volt meter (for example, RMS,
average, peak, etc.). Specifications are based on RMS mea-
surements.
Furnace Checks
Open Furnace Disconnect power to the control unit and the sensor. With Ohm
meter, measure across Terminals “R” and “S” on the sensor
board. The resistance of the furnace should be:
32 ohms (±10%)
If the furnace resistance is not within allowable tolerances,
replace the furnace.
Loss of AC Power Verify the correct line voltage at L1 and L2 of the sensor board.
to the Furnace
With line voltage present, check the voltage at terminals R and
S on the sensor board. If voltage is present and the system is
not heating, remove power and check the furnace resistance
(see “Furnace Checks” section). If no voltage is present at ter-
minals R and S on the sensor board, check for the DC furnace
drive signal at terminals 3 (+) and 4 (-) at the sensor board.
The furnace drive is a 15 VDC pulse train, and depending on
the frequency of the pulse train, the measured voltage will be
between 5 and 15 VDC. If this furnace drive signal is present,
reset the power to the control unit and see if this solves the
problem. If this doesn’t solve the problem, replace the sensor
board.
If the furnace drive signal is Check the interconnecting wiring. If the interconnecting wiring is
not present: correct, check across FURNACE and 15V COM on the control
unit wiring card for the furnace drive signal. If the signal is not
present, check fuse F1 on the wiring card. If this fuse is OK, re-
place the control unit power supply/keypad module. If this fuse
is not OK, replace the fuse (1/4 amp, 125 volts, IEC speed type
FF). Check the interconnecting wiring for short circuits before
applying power to the control unit.
Cell Checks
It is likely that there is a leak in the sensor plumbing
or an improper calibration gas setup, and there is not
a problem with the cell itself (see the “Leak Check”
If the cell fails when you first begin
section later in this chapter for help on how to check
to use the analyzer:
for leaks; see the “Calibration/Aspirator Air Checks”
section earlier in this chapter for help on checking
your calibration gas setup).
If this doesn’t solve the problem, replace the cell, then run another primary calibration.
Leak Check
This check does not apply to Insitu users.
• Check that all compression fitting and pipe thread connections are
leak tight.
• Be sure that the mounting plate or mounting flange gasket on the rear
of the sensor is in place (see the “Sensor Mounting” section in Chapter
2 for details on the placement of the gasket).
For processes under vacuum you can check for leaks by sniffing the fit-
tings with another gas (for example, nitrogen or pure oxygen), being sure
to avoid the area over the top of the cell.
1. Use a piece of tygon or plastic tubing with a 1/8” stainless steel nozzle
to apply the gas from a cylinder (using stainless steel prevents any
problems that might occur with plastic melting the nozzle on hot sen-
sor components).
2. Monitor the response from the cell. When the cell millivolt reading
changes, it indicates a leak in that area of the plumbing (the vacuum
of the process pulls in the gas).
3. If not convenient to view the control unit display, you can also apply a
volt meter to terminals 1 and 2 on the sensor board (Cell + and Cell -)
to see if the cell millivolts change, indicating a leak.
1. Remove the sensor from the process (after allowing sensor to cool)
and pressurize it with 5 lbs. of air, plugging any exiting ports, the inlet
probe, and exhaust tube holes. Then go over the sensor fittings with a
leak detector fluid.
The WDG-IV style sensor has intentional leaks in the aspirator air
inlet fitting and the screw in the back of the aspirator block. If you are
NOTE having problems and your process is under positive pressure, these
intentional leaks will need to be sealed.
1. Examine the inlet and exhaust for plugging problems. When possible,
rod out the probe and exhaust.
1. To check the aspirator vacuum level, remove the inlet hook and turn
off the aspirator air.
2. Attach a vacuum gauge to the aspirator inlet and apply aspirator air.
The aspirator should pull 2 to 3 inches WC at 5 psig. If the aspirator
doesn’t pull a vacuum, disassemble sensor plumbing and clean the
nozzle and venturi in the back of the aspirator.
You will need a new metal O-ring when installing the venturi.
Never service the controller or sensor unless power has been removed
from the controller and sensor, and the sensor has been allowed to cool
for one hour.
• Cell Replacement
• Thermocouple Replacement
• Furnace Replacement
See the Installation chapter for help on how to remove or replace Series
2000 control unit modules or option cards.
You can clean the outside of the sensor or controller (or Remote Calibra-
tion Unit) using normal household or commercial general purpose clean-
ers, cloths, or sponges. You can also use water. Always turn off power
before attempting to clean the enclosure.
• If you turn off the analyzer or the process is shut down, turn off the
aspirator air to avoid plugging problems.
• The outside of the sensor cover and all sensor assembly components
are extremely hot (up to 500°F (260°C) inside the cover), even after a
considerable period from shutdown. Turn off the power to the sen-
sor and control unit when working inside the sensor. Use caution and
wear appropriate gloves when handling components.
Cautions
• Never use pipe dope or any other contaminant that gives off combus-
tible vapor on any joints of the sample tubing. Combustible vapor in
the sample tubing can lead to erroneous readings.
3. While holding the top of the cell housing with a backup wrench,
loosen the lower hex nut (not the top hex nut), then remove the entire
cell assembly.
4. If cleaning the cell, wash with water or alcohol. Dry the cell thoroughly
before reinstalling. Always use a new cell O-ring.
If replacing the cell, discard the old cell and cell O-ring and retrieve
the new cell with supplied cell O-ring. Avoid touching the bare cell.
Instead, hold the cell by one of its hex nuts.
5. Place the cell O-ring on top of the cell housing as shown in Figure 7-2.
6. Insert the new cell into the cell housing. Do not touch the bare cell
when reinserting. Tighten the cell into the cell housing using the low-
er hex nut (the upper hex nut is pre-tightened at the factory). The cell
O-ring provides a seal for the system. Make sure it is evenly crushed
when tightening the cell.
3. Remove the screw from the thermocouple mounting tab (save the
screw for installing the new thermocouple).
5. Insert the new thermocouple into the slot in the top of the furnace.
It should go in freely; if necessary, wiggle a little to position without
forcing it in. The thermocouple should be near, but not touching, the cell
housing or furnace heater coil.
6. Put the screw back into the thermocouple mounting clamp. This se-
cures the thermocouple to the furnace
1 Cell housing
2 Thermocouple mounting tab
3 Thermocouple
4 Convection loop mounting clamp
5 Compression fitting
5. While holding the furnace in one hand, remove the convection loop
mounting clamp. Be sure to save the mounting clamp for when you
install the new furnace.
6. Gently pull the cell housing out of the furnace and remove the fur-
nace.
7. Slip the new furnace onto the cell housing and attach the cell housing
using the convection loop mounting clamp.
8. Remount the cell housing (reverse the actions taken in Step 4).
10. Using an Ohm meter, check for short circuits of the heater coil to the
metal thermocouple sheath. Check that neither furnace lead is short
circuited to the metal furnace cover. The resistance should be infinite.
If you measure a short circuit, loosen the screw that secures the ther-
mocouple and relocate slightly until resistance is infinite.
12. Reattach furnace leads (reverse the actions taken in Step 3).
CELL HOUSING
FURNACE
UNION FITTING
COMPRESSION FITTING
1. WDG-IVC
Remove the detector leads from terminal strips E, F, H, and I on the
sensor board.
WDG-IVCM
Remove the detector leads from terminal strips W, X, Y, and Z.
2. Loosen the compression fitting nut that holds the combustible detector
in place and remove the detector from its housing. Save the nut from
the detector to use when installing the new combustibles detector.
4. Install the new combustibles detector with nut and ferrules. Be ex-
tremely careful not to damage the detector elements when installing.
Ensure that the mark you made in Step 3 is in the 12 o’clock position
to ensure proper orientation of the combustibles detector in the com-
bustibles detector block.
Ferrule
Detector head
Combustibles
Detector
Block
Compression
Fitting
Nut
Front View
Place Mark
@ 12 o'clock
flow
Sample
flow
2. Loosen the compression fitting nut that holds the combustible detec-
tor in place and remove the detector from the combustible detector
block.
3. It is very important that you orient the new combustibles detector cor-
rectly inside the combustibles detector block (sample flow path must
contact both detector leads at the same time). To do this, ensure that
the flat part of the combustibles detector is facing down (see Figure 7-6
for correct orientation).
4. Install the new combustibles detector with nut and ferrules. Be ex-
tremely careful; do not damage the detector elements when install-
ing. Ensure the proper orientation of the combustibles detector in the
combustibles detector block.
5. Reconnect the detector leads to the sensor board by reversing the ac-
tions taken in Step 1 above.
Ferrule
Detector Head
Combustibles
Detector Block
Compression
Fitting Nut
Front View
Flat must
face down
flow
Sample
flow
Figure 7-7 shows how to wire box heaters using the sensor board.
HTR
BOX HEATER
115 VAC
L HTR
N HTR
BOX HEATER
HTR
SENSOR BOARD
HTR
BOX HEATER
230 VAC
NC
L
N NC
BOX HEATER
HTR
SENSOR BOARD
Sensor
Furnace Assembly P/N 70409SE
Type “K” Thermocouple P/N 71697KE
Methane Detector Kit P/N 70987TE
Combustibles Detector Kit (hot wire) P/N 70986TE
Combustibles Detector Kit (RTD) P/N 71504TE
Sensor Board P/N 80471SE
Zirconium Oxide Cell P/N 71063SE
Control Unit
Display Module P/N 90219VE
Power Supply/Keypad Module P/N 90253VE
Auto Calibration Card P/N 80436SE
Combustibles Card (WDG-IVC) P/N 80461SE
Combustibles Card (WDG-IVCM) - RTD type P/N 80449SE
Combustibles Card (WDG-IVCM) - Hot Wire type P/N 80561SE
Backplane Assembly P/N 80439SE
Wiring Card (WDG-IV) P/N 80457SE
Wiring Card (WDG-IVC) P/N 80445SE
Wiring Card (WDG-IVCM) P/N 80466SE
Processor Board P/N 80440SE
Display P/N 42373JE
When ordering, provide the serial number of your analyzer to ensure proper
parts are ordered:
AMETEK
Process & Analytical Instruments Division
150 Freeport Road
Pittsburgh, PA, USA 15238
Serial communications for the Series 2000 (ver 3.3 and later) and the IQ based software
Communication Protocols
Communication with all analyzers through the RS-485 connection uses
the following word format:
• 8 data bits
• 1 stop bit
• No Parity
Start Character Node Address Command Letter Data Checksum End Character
“>” (ASCII charac- 2 Hex ASCII char- (See Command (n characters - 2 hex ASCII char- Carriage Return
ter 3E Hex) acters (00 to FF) Section for list of depends on com- acters (sum of all (ASCII character
commands) mand) preceding charac- 0D Hex)
ters except the “>”
character)
Where:
The Start Character is the greater than character (“>”, which is ASCII 3E
hex).
The Node Address is from 00-FF, and is represented as two hex ASCII char-
acters. Each sensor has a unique node address. Note, however, that the
maximum number of sensors that can be connected on the network is 32
and for the gateway, we have limited the valid addresses from 01-32.
The Checksum is the sum of all the preceding characters (excluding the >
character) - modulo 256 or the characters “??” (?? is equivalent to 2 ASCII
3F hex characters).
This command reads the value of a variable in the analyzer. The data sec-
tion of this command is a number from 00-FF. These numbers correspond
to the variable you wish to read. The table below provides a list of these
variables and their location codes. If a variable is selected beyond the end
of the table, an error will be returned. The value returned is followed by a
text string for the units, where applicable.
Example:
Request Oxygen with the following string: >FEF0836
Returned string: A20.9 %O2D4
This command sets a value in the analyzer. The data section of this com-
mand is a two-digit hex number (00-FF) to point to the location to write
from the variable table, followed by the value of the variable. A simple
acknowledge is returned if the command was accepted.
This command is used to read the dates and times stored in the analyzer.
The data consists of a single digit where 0 is used to read the current date
and time, 1 is used to read the date and time of the last calibration, 2 is
used to read the date and time of the last verification, and 3 is used to
read the date and time of the last step function. Date and time is returned
as mm/dd/yy hh:mm in 24 hour (military) time.
This command is used to set the analyzer’s current date and time. The
data string is of the form: mm/dd/yy hh:mm in 24 hour (military) time.
Leading zeroes are required, for example: 03/18/93 20:02.
Calibrate (G)
Example:
>01G00?? - perform a secondary cal.
Echo (A)
The echo command is used to test the serial protocol and line integrity.
Any text sent to the analyzer is echoed back to the host.
Acknowledge (C)
This command is used to check how data is output through the serial link.
For each variable that is accessible through the Read Number and Write
Number commands, a string is returned that describes the format of the
number. This response data consists of three characters:
Access is as follows:
r read only
w write only
b read/write
Location is:
r RAM
e EEPROM
Due to the nature of EEPROM memory, each location in the memory has
a limited number of writes allowed (approximately 10,000). The location
character can be used to determine where the data is stored internally to
the analyzer and act accordingly.
Command List
Letter Description
A Echo
B Bad Command
C Acknowledge
F Read Number
G Calibrate
H Write Number
Success
All successful responses start with A. If there is data returned, it will fol-
low the A and have a checksum after it. All responses end with a carriage
return ( ).
Success End
A
Success Data Checksum End Character
A (n characters) (2 hex ASCII characters)
Examples:
A
Simple acknowledge
AFF0050
Acknowledge with data
Failure
All failure responses start with a N followed by a two hex digit failure code
and a carriage return ( ).
S2 IQ
bit 0 – No Exception , Future use
bit 1 – Future use
bit 2 – Future use
bit 3 – Future use
bit 4 – Power Down Detected
bit 5 – Watchdog Failure
IQ
bit 6 – Detector B Fault
IQ
bit 7 – Detector A Fault
0 – blank line
2 – oxygen
4 – cell temperature
6 – cell mv
8 – thermocouple mv
10 – box tempIQ
12 – combustibles
14 – methane
16 – methane and combustibles
IQ
18 – combustible voltages
IQ
20 – methane voltages
22 – time and date
S2
24 – user text
IQ
AF Ratio
IQ
26 – WOBBE
06/06 Line2_flg What is displayed on line 2 of the display R/W
0 – blank line
2 – oxygen
4 – cell temperature
6 – cell mv
8 – thermocouple mv
IQ
10 – box temp
12 – combustibles
14 – methane
16 – methane and combustibles
IQ
18 – combustible voltages
IQ
20 – methane voltages
22 – time and date
S2
24 - user text
IQ
AF Ratio
IQ
26 - WOBBE
07/07 Line3_flg What is displayed on line 3 of the display R/W
0 – blank line
2 – oxygen
4 – cell temperature
6 – cell mv
8 – thermocouple mv
IQ
10 – box temp
12 – combustibles
14 – methane
16 – methane and combustibles
IQ
18 – combustible voltages
IQ
20 – methane voltages
22 – time and date
S2
24 - user text
IQ
AF Ratio
IQ
26 – WOBBE
0 – oxygen
IQ
1 – cell temperature
S2
2 – cell temperature
IQ
thermocouple mv
IQ
3 – cell mv
S2
4 – thermocouple mv
IQ
box temp
IQ
5 – combustibles
6 – cell mvS2
IQ
methane
IQ
7 – AF Ratio
IQ
8 – WOBBE
0F/15 Out2_flg What analog output 2 corresponds to R/W
0 – oxygen
IQ
1 – cell temperature
S2
2 – cell temperature
IQ
thermocouple mv
IQ
3 – cell mv
S2
4 – thermocouple mv
IQ
box temp
IQ
5 – combustibles
S2
6 – cell mv
IQ
methane
IQ
7 – AF Ratio
IQ
8 – WOBBE
10/16 Out3_flg What analog output 3 corresponds to R/W
IQ
0 – oxygen
IQ
1 – cell temperature
IQ
2 – thermocouple mv
IQ
3 – cell mv
IQ
4 –box temp
IQ
5 – combustibles
IQ
6 – methane
S2
10 – combustibles
S2
12 – methane
IQ
7 – AF Ratio
IQ
8 – WOBBE
11/17 Out4_flg What analog output 4 corresponds to R/W
IQ
0 – oxygen
IQ
1 – cell temperature
IQ
2 –thermocouple mv
IQ
3 – cell mv
IQ
4 –box temp
IQ
5 – combustibles
IQ
6 – methane
IQ
7 – AF Ratio
IQ
8 – WOBBE
S2
10 – combustibles
S2
12 – methane
26/38 Gas1_time Amount of time oxygen span gas will flow in R/W
IQ
seconds
Amount of time oxygen span gas will flow in
S2
MMSS
27/39 Gas2_time Amount of time oxygen zero gas will flow in R/W
IQ
seconds
Amount of time oxygen zero gas will flow in
S2
MMSS
28/40 Gas3_time Amount of time the combustibles span gas R/W
IQ
will flow in seconds
Amount of time combustibles span gas will
S2
flow in MMSS
37/55 Ver_zero Oxygen zero gas read during last verify Read only
38/56 O2_zero Zero gas millivolts read during last Read only
calibration
39/57 Cal_cgas3 Combustibles span gas setpoint during last Read only
calibration
3A/58 Cmbspan Combustibles span value during last Read only
calibration
3B/59 Ver_cgas3 Combustibles span gas setpoint for last Read only
verify
3C/60 Vcmbspan Combustibles span gas read during last Read only
verify
3D/61 Cmbzero Combustibles zero gas read during last Read only
calibration
3E/62 Vcmbzero Combustibles zero gas read during last Read only
verify
3F/63 Cal_cgas4 Methane span gas setpoint during last Read only
calibration
40/64 Mthspan Methane span value read during last Read only
calibration
41/65 Ver_cgas4 Methane span gas setpoint for last verify Read only
42/66 Vmthspan Methane span value during last verify Read only
43/67 Mthzero Methane zero value during last calibration Read only
44/68 Vmthzero Methane zero value during last verify Read only
S2
0 – oxygen
IQ
oxygen, combustibles, or methane
1 – calibration
2 – verify
3 – in cal/verify
IQ
5E/94 Alm_config Configuration for oxygen alarms, high or R/W
low. If a bit is 1, that is the condition
that the indicated alarm will trip on.
IQ
bit 0 – high oxygen, process alarm
IQ
bit 1 – low oxygen, process alarm
IQ
bit 2 – high combustible, process alarm
IQ
bit 3 – high methane, process alarm
IQ
bit 4 – high oxygen, alarm #2
IQ
bit 5 – low oxygen, alarm #2
IQ
bit 6 – high combustible, alarm #2
IQ
bit 7 – high methane, alarm #2
IQ
bit 8 – high oxygen, alarm #3
IQ
bit 9 – low oxygen, alarm #3
IQ
bit 10 – high combustible, alarm #3
IQ
bit 11 – high methane, alarm #3
bit 12 – Future use
bit 13 –Future use
bit 14 – Future use
bit 15 – Future use
S2
Alm_config Configuration for oxygen alarms, high or R/W
low. If a bit is 1, that is the condition
that the indicated alarm will trip on.
0 – In normal calibration
1 – In verify
2 – In primary cal
3 – In none of the above, normal operation
IQ S2
bit 0 – CAL 0 bit 0 – Dig In
IQ
bit 1 – CAL 1
6F/111
70/112 SysTst_Flag System test that was preformed R/W
1-Calibrate A/D
2-Ram Test
4-Test EEPROM
7-Erase Ram
8-Erase EEPROM
9- Test Watchdog
71/113 Passw Current password R/W
IQ
72/114 BoxCtl Value that corresponds to the box set point R/W
0 – 230 C
1 – 270 C
2 – 360 C
3 – Full on control
73/115 InjectGas Turn on gas for inject cal gas menu R/W
0 - Span gas
1 - Zero gas
2 - Comb gas
3 - Methane gas (if 3 span value RCU
installed)
4 - Process gas
74/116 Node_Addr The current node address R/W
75/117 Read only
76/118 Cmbsens Combust sensitivity associated with the Read only
last calibration
77/119 Mthsens Methane sensitivity associated with the Read only
last calibration
78/120 AFRatio The calculated A/F ratio Read only
79/121 AFRatio_CalSP The zero cal gas A/F ratio set point R/W
7A/122 Wobbe The calculated WOBBE value Read only
7B/123 Wobbe_SP The WOBBE Factor R/W
7C/124 Gas4_time Amount of time the methane span gas will R/W
IQ
flow (in seconds)
7D/125 AFRatio_CalSP2 The A/F Ratio set point at low end R/W
7E/126 XSO2Low_SP The XS O2 set point at low end R/W
7F/127 SW Version Current software version # Read only
5. Read the Cal_flag variable to determine when the cal cycle is finished
(Calflag=3).
6. Read the calibration values (for example, cal_span and cal_zero vari-
ables).
SETUP KEY
Functions/Options
Menu Options
Select info for Display Line 1.
Display Select info for Display Line 2.
Select info for Display Line 3.
Passwords Create a system password
Select a positive or negative process pressure.
Process Pressure
Enter process pressure value.
System Tests Perform internal diagnostics on Series 2000.
View serial number for Series 2000.
System Serial #
View manufacturing number for Series 2000.
Primary Calibration Perform an automatic, remote, or manual primary calibration.
Sensor Configuration Select sensor type.
Communications Define communications parameters.
Define combustibles scale (% or PPM).
Combustibles
Define combustible detector type.
ANALOG KEY
Functions/Options
Menu Options
Set Current Range Range of readings that the analog output represents (for
example:
4% O2 = 4 mA
10% O2 = 20 mA)
Set Track/Hold Hold output at last process reading during a cal or verify
Set Current Mode 0 to 20 mA
4 to 20 mA
Output Filtering Smooth out variations in readings
Select Function % O2
Cell temperature
Cell Millivolts
Thermocouple Millivolts
Combustibles (option)
Methane (option)
FUNCTION
*Select Function Alarm Configure Relay Configure
Alarm 3 % O2 High O2 Alarm
Cal Start Low O2 Alarm Energize on alarm
Verify Start
Cal/Verify Start
Alarm 4 Not selectable De-energize on alarm
ALARM NUMBER
(% O2 only)
Alarm 5
Not Selectable
Not selectable (high alarms only)
Alarm 6 (combustibles only)
Alarm 7
* Note: Alarms 1 and 2 are reserved for the watchdog timer and service alarms and are
not software-selectable.
VEXT1, 2 +
VEXT1, 2 -
Jumper JP2 on the Series 2000 control unit display module is used to allow
you to define which power source to use. If JP2 is in (factory default), the
control unit uses the internal supply to power the current output circuits. If
you try to use an external voltage at this time, the system will use whichever
voltage is higher - control unit voltage is normally about 28 volts. If the JP2
jumper is removed, the external voltage source must be used to power the
current output circuits. The external voltage source must be no greater than
30 volts DC. See the “Option Card Installation” section in the Installation
Chapter for information on how to access or remove the display module.
where:
VEXT3+ VEXT4+
VEXT3- VEXT4-
Jumper JP1 on the combustibles card allows you to define which power
source to use for current output 4. Jumper JP2 on the combustibles card
allows you to define which source to use to power current output 3. If the
jumper is in (factory default), the control unit uses its own power supply
to power the current output circuits. If you try to use an external voltage
at this time, the system will use whichever voltage is higher - control unit
voltage is normally about 28 volts. If the jumper is removed, the external
voltage source must be used to power the current output circuits. The
external voltage source must be 30 volts DC or less.
IOUT1 and IOUT2 must share the same power supply if both of these chan-
nels are used, or two power supplies can be “ORed” together using a diode
from each power supply positive terminal (diode anode to power supply) to
the VEXT1,2+ terminal. The power supply minuses would be tied together
and connected to one side of the both loads. The other side of one of the
loads would be connected to IOUT1+ terminal, and the other side of the
other load would be connected to the IOUT2+ terminal.
For each current output channel used in a current loop modulation ap-
plication, the jumpers should be removed for that channel.
OPTIONAL
EXTERNAL LOOP POWER SUPPLY
OUTPUT DEVICE MAX VOLTAGE 30 VOLTS DC
(COMPUTER OR CHART RECORDER)
MAX LOAD RESISTANCE 1200 OHMS
VEXT3+ VEXT4+
VEXT3- VEXT4-
If you did order any special options, the drawings or special instructions
provided here supercede any drawing or options provided elsewhere in
this manual.
SETUP PARAMETERS
K FACTOR 1
OUTPUT # 1 FUNCTION % OXYGEN
CURRENT MODE,RANGE 4 TO 20 MA = 0-10
TRACK/HOLD CAL VERIFY HOLD,TRACK
OUTPUT FILTER # 1 100