Valleylab Force FX C 8 Service Manual
Valleylab Force FX C 8 Service Manual
Valleylab Force FX C 8 Service Manual
Abnormal neuromuscular Metal-to-metal sparking Check all connections to the generator, patient return
stimulation (stop surgery electrode, and active electrodes.
immediately)
Can occur during coag Use a lower power setting for the Fulgurate and Spray
modes or select the Desiccate mode.
Abnormal 50–60 Hz leakage currents Inside the generator, carefully inspect for damage that
may cause shorting between the AC line voltage and
connected patient components.
Troubleshooting
4 Internal diagnostics Replace the Control board. Refer to Control Board Replacement in
Section 7.
Main microcontroller CPU test failed
Calibrate the generator. Refer to Section 5 for instruction.
If the alarm number reappears, record the number and call the
Valleylab Service Center.
10 Software malfunction. Invalid alarm Turn off, then turn on the generator. If the alarm number
number (number shown on terminal using reappears, record the number and call the Valleylab Service
serial port) Center.
11 Internal diagnostics. Invalid activated Refer to Correcting IC U6 Malfunctions (steps 1–7) and Correcting
power, function, mode, or key request IC U3 Malfunctions (steps 4–11) in this section.
mode echoed by feedback microcontroller
16 Diagnostics/microcontroller malfunction. Calibrate the ECON factor. Refer to Section 5 for instructions.
T_ON average test failed
If the alarm number reappears, refer to Correcting T_ON ASIC
Malfunctions in this section.
69 Software malfunction. flash_power_state Turn off, then turn on the generator. If the alarm number
Troubleshooting
value outside state range reappears, record the number and call the Valleylab Service
Center.
70 Software malfunction. key_req_mode
value outside state range
71 Software malfunction. request+1 value
outside state range
80 Software malfunction. *input[i].p_state
value outside state range
81 Internal component malfunction. Do not attempt to use the generator. Record the number and call
Optoisolator test failed the Valleylab Service Center.
90 Generator model number in main Refer to Correcting IC U6 Malfunctions (steps 1–7) and Correcting
microcontroller ROM not Force FX-8C IC U3 Malfunctions (steps 4–11) in this section.
model number
95 Generator model number in main
microcontroller ROM not Force FX-8C
model number
100 Software malfunction. rem_update_state Turn off, then turn on the generator. If the alarm number
value outside state range reappears, record the number and call the Valleylab Service
Center.
101 Software malfunction. rem_pad_state
value outside state range
102 Software malfunction. rem_state value in
re_rem_control( ) outside state range
103 Software malfunction. rem_state value in
re_non_rem_control( ) outside state range
Software malfunction. rem_flash_state
104 value outside state range
Software malfunction. rem_led_state value
105 outside state range
Software malfunction. cal_state value in
110 ca_generator_setup( ) outside state range
Software malfunction. cal_state value in
111 ca_clock_date_setup( ) outside state
range
112 Software malfunction. cal_mode value in Turn off, then turn on the generator. If the alarm number
ca_clock_date_setup( ) outside state reappears, record the number and call the Valleylab Service
range Center.
113 Software malfunction. cal_state value in
ca_clock_time_setup( ) outside state
range
114 Software malfunction. cal_mode value in
ca_clock_time_setup( ) outside state
range
115 Software malfunction. cal_state value in
ca_coag_leakage_cal( ) outside state
range
116 Software malfunction. cal_state value in
ca_econ_cal( ) outside state range
Troubleshooting
117 Software malfunction. econ_cal_state
value outside state range
118 Software malfunction. cal_state value in
ca_rem_cal( ) outside state range
119 Software malfunction. rem_cal_state value
outside state range
120 Calibration malfunction. Calibration Repeat the failing calibration step. If the alarm number reappears,
value(s) outside acceptable range record the number and call the Valleylab Service Center.
121 Software malfunction. cal_mode value in Turn off, then turn on the generator. If the alarm number
ca_generator_setup( ) outside state range reappears, record the number and call the Valleylab Service
Center.
122 Open circuit REM sense failure Repeat the failing calibration step. If the alarm number reappears,
record the number and call the Valleylab Service Center.
123 Main microcontroller unable to access Refer to Correcting IC U6 Malfunctions in this section.
GEN_SEM semaphore
124 Main microcontroller unable to access
FEEDBACK_SEM semaphore
125 Main microcontroller unable to access
ECON_SEM semaphore
126 Main microcontroller unable to access
KEY_ACTIVE_SEM semaphore
151 Main microcontroller unable to access Refer to Correcting IC U6 Malfunctions in this section.
GEN_SEM semaphore
152 Software malfunction. cal_state value in Turn off, then turn on the generator. If the alarm number
cs_i_sns_cal( ) outside state range reappears, record the number and call the Valleylab Service
Center.
154 Main microcontroller unable to access Refer to Correcting IC U6 Malfunctions in this section.
FEEDBACK_SEM semaphore
160 Internal component malfunction. Dosage Calibrate the ECON factor. Refer to Section 5 for instructions.
error test failed. ADC values display at
If the alarm number reappears, refer to Correcting T_ON ASIC
failure
Malfunctions in this section.
170 Watchdog malfunction. Correct value not Refer to Correcting IC U6 Malfunctions in this section.
sent from feedback microcontroller
171 Watchdog malfunction. Correct value not Refer to Correcting IC U3 Malfunctions in this section.
sent from main microcontroller
172 Watchdog malfunction. Main
microcontroller unable to access
watchdog semaphore
174 Software malfunction. which_errors value Turn off, then turn on the generator. If the alarm number
outside state range reappears, record the number and call the Valleylab Service
Center.
180 Internal diagnostics. Main microcontroller Refer to Correcting IC U3 Malfunctions in this section.
WSI RAM check failed
182 Internal diagnostics. Main microcontroller Refer to Correcting IC U3 Malfunctions in this section.
dual-port RAM check failed
183 Internal diagnostics. Main microcontroller
Page 0 (zero) ROM CRC test failed
184 Internal diagnostics. Main microcontroller
Page F ROM CRC test failed
189 Software malfunction. display_state value Turn off, then turn on the generator. If the alarm number
outside state range reappears, record the number and call the Valleylab Service
Center.
190 Internal diagnostics. Bipolar buttons (up 1. Turn off, then turn on the generator. Do not press buttons or
arrow, down arrow, Precise, Standard, accessory activation devices during the self-test.
and/or Macro) may be stuck
2. If the alarm number reappears, disconnect all accessories.
191 Internal diagnostics. Cut buttons (up Turn off, then turn on the generator again.
arrow, down arrow, Low, Pure, and/or
If the alarm number reappears, record the number and call the
Blend) may be stuck (button shown on
Valleylab Service Center.
terminal using serial port)
192 Internal diagnostics. Coag buttons (up
arrow, down arrow, Desiccate, Fulgurate,
and/or Spray) may be stuck (button shown
on terminal using serial port)
193 Internal diagnostics. Recall button may be
stuck (button shown on terminal using
serial port)
Troubleshooting
194 Internal diagnostics. Handswitch or
Monopolar 1 Footswitch cut pedal may be
stuck
195 Internal diagnostics. Handswitch or
Monopolar 1 Footswitch coag pedal may
be stuck
196 Internal diagnostics. Handswitch or
Monopolar 2 Footswitch cut pedal may be
stuck
197 Internal diagnostics. Handswitch or
Monopolar 2 Footswitch coag pedal may
be stuck
198 Internal diagnostics. Handswitch or
Bipolar Footswitch pedal may be stuck
199 Internal diagnostics. Main and feedback Refer to Correcting IC U6 Malfunctions (steps 1–7) and Correcting
microcontrollers are not compatible IC U3 Malfunctions (steps 4–11) in this section.
200 Internal diagnostics. Main microcontroller 1. Replace the Control board. Refer to Control Board
SEML line may be stuck Replacement in Section 7 for instructions.
201 Internal diagnostics. Feedback 2. Calibrate the generator. Refer to Section 5 for instructions.
microcontroller SEMR line may be stuck
If the alarm number reappears, record the number and call the
Valleylab Service Center.
202 Internal diagnostics. Main microcontroller Refer to Correcting IC U3 Malfunctions in this section.
watchdog test failed
206 Software malfunction. doserr_test_state Turn off, then turn on the generator. If the alarm number
value in st_m_doserr_test outside state reappears, record the number and call the Valleylab Service
range Center.
207 Software malfunction. doserr_test_state
value in st_fb_doserr_test outside state
range
208 Main microcontroller timer interrupt failed 1. Replace the Control board. Refer to Control Board
Replacement in Section 7.
209 Feedback microcontroller timer interrupt
failed 2. Calibrate the generator. Refer to Section 5 for instructions.
If the alarm number reappears, record the number and call the
Valleylab Service Center.
210 Software malfunction. data_type value Turn off, then turn on the generator. If the alarm number
outside state range reappears, record the number and call the Valleylab Service Center
211 Software malfunction. which_data value
outside state range
212 Generator not calibrated Refer to Correcting Battery-Backed RAM Malfunctions in this
section.
213 Internal diagnostics. Firmware not Refer to Correcting IC U6 Malfunctions (steps 1–7) and Correcting
compatible with hardware IC U3 Malfunctions (steps 4–11) in this section.
215 Internal diagnostics. Real-time clock chip 1. Replace the Control board. Refer to Control Board
(U1) not compatible with firmware Replacement in Section 7.
2. Calibrate the generator. Refer to Section 5 for instructions.
If the alarm number reappears, record the number and call the
Valleylab Service Center.
216 Internal memory battery exhausted 1. Replace the battery. Refer to Battery Replacement in
Section 7.
2. Calibrate the generator. Refer to Section 5 for instructions.
If the alarm number reappears, record the number and call the
Valleylab Service Center.
220 Feedback microcontroller unable to access Refer to Correcting IC U6 Malfunctions in this section.
KEY_REQ_SEM semaphore
221 Feedback microcontroller unable to access
GEN_SEM semaphore
222 Main microcontroller failed to copy Refer to Correcting IC U3 Malfunctions in this section.
battery-backed RAM data to dual-port
RAM in time allotted
223 Main microcontroller failed to complete
initialization in time allotted
224 Internal diagnostics. Calibration data Refer to Correcting IC U6 Malfunctions in this section.
checksum error on feedback
microcontroller (data in error shown on
terminal using serial port)
225 Internal diagnostics. Feedback 1. Replace the Control board. Refer to Control Board
microcontroller CPU test failed Replacement in Section 7.
2. Calibrate the generator. Refer to Section 5 for instructions.
If the alarm number reappears, record the number and call the
Troubleshooting
Valleylab Service Center.
226 Feedback microcontroller unable to access Refer to Correcting IC U6 Malfunctions in this section.
FEEDBACK_SEM semaphore
230 Software malfunction. keyed_state value Turn off, then turn on the generator. If the alarm number
outside state range reappears, record the number and call the Valleylab Service
Center.
231 Software malfunction. key_req_mode
value outside state range
232 Feedback microcontroller unable to access Refer to Correcting IC U6 Malfunctions in this section.
KEY_ACTIVE_SEM semaphore
240 Software malfunction. activation_seq Turn off, then turn on the generator. If the alarm number
value outside state range reappears, record the number and call the Valleylab Service
Center.
241 Software malfunction. feedback value in
do_feedback( ) outside state range
242 Software malfunction. feedback value in
do_cal_feedback( ) outside state range
244 Software malfunction. disable_seq value
outside state range
245 Software malfunction. spark_state value
outside state range
246 Feedback microcontroller unable to access Refer to Correcting IC U6 Malfunctions in this section.
FEEDBACK_SEM semaphore
247 Feedback microcontroller unable to access
ECON_SEM semaphore