Repair Instructions - Service No./Modi, Error List
Repair Instructions - Service No./Modi, Error List
Repair Instructions - Service No./Modi, Error List
Enter the number under service diagnosis (Evita) or ventilation (Evita 2 dura) during
operation. For release Hardware Option wait 30 seconds and switch Evita off and on.
Release Software Option:
Refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 3.5.2,
Diagnosis page Option Release (release of software options), page 76.
1.1 Service Numbers for Software Versions 1.00 to 1.04 and 1.08
Evita 4 with Software 1.05, 1.06, 1.07, 1.09 and 2.n and higher; for Evita 2 dura as of
Software 3.00
The service codes 02.00.. to 02.22.. are recognised on the CPU 68332 PCB; they
correspond to 02.40.. to 02.62.. as recognised on the Pneumatic Controller PCB.
100
30 Paw high display 1
31 Paw high display 2
32 MV min display 1
GBR5664500T025664800.fm
33 MV min display 2
34 Cold start screen
reset
Possible causes:
Measurement of inspiration
pressure on Pneumatic Controller
PCB (1st channel) or CO2 Carrier
PCB (2nd channel) defective. Read
out values by way of service mode.
If error is on CO2 Carrier PCB,
contact problems may be the
cause. CO2 Carrier PCB must be
secured in position with both
screws.
Inspiration pressure sensor
defective.
Pneumatic fault.
Too high a measured value from
For internal use only. Copyright reserved.
dura).
02.18.001 si18_Err, Silence incorrectly RAM or SW error on CPU 68332 PCB
activated, reset or Pneumatic Controller PCB or
Graphics Controller PCB (Evita 4) /
GBR5664500T025664800.fm
The service codes 02.40.. to 02.62.. are recognised on the Pneumatic Controller PCB; they
correspond to 02.00.. to 02.22.. as recognised on the CPU 68332 PCB.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
02.71.013
Remedy:
refer to section of "Repair Instructions -
Electronic Components" Chapter 5.5,
Conversions, page 39 CO2 Carrier PCB.
02.71.004 Gold Cap capacitor, see error 02.71.003
voltage too high
02.71.005 Gold Cap capacitor, see error 02.71.003
voltage too low
02.71.006 Gold Cap capacitor, see error 02.71.003
For internal use only. Copyright reserved.
tester defective
GBR5664500T025664800.fm
Trouble-shooting:
- With diagnosis mode, Refer to section of
"Repair Instructions - Service No./ Modi, Error
list" chapter 3.3.2, Diagnosis page
Microprocessor of Electronics, page 63.
- With external DS mode via PC, Refer to
section of "Repair Instructions - Service No./
Modi, Error list" chapter 9.4.6, Horn test, page
140.
instructions
02.71.009 Piezo alarm generator, refer to error 02.71.008
current too high
02.71.010 Piezo alarm generator, refer to error 02.71.008
current too low
02.71.011 Monitoring of nebuliser The required nebuliser valve position is
For internal use only. Copyright reserved.
Trouble-shooting:
- Measure resistance of loudspeaker
via connecting cable of control
console incl. cable between pins 9
and 10 of SUB-D cable connector.
Test value = 55 plus/minus 8 ohms.
This value incl. shunt resistance must
be 5 ohms higher than loudspeaker
value, which can be measured after
opening control console.
- Using diagnosis mode with lowest
volume setting, Refer to section of "Repair
Instructions - Service No./ Modi, Error list"
chapter 3.2.1, Diagnosis page
Microprocessor of Front, page 57.
- Using external DS mode via PC with
lowest volume setting, Refer to section of
"Repair Instructions - Service No./ Modi, Error
list" chapter 9.4.7, Loudspeaker test, page
141. or Refer to section of "Repair
Instructions - Service No./ Modi, Error list"
chapter 9.5.7, Testing of loudspeaker, page
151.
instructions.
Note: Refer also to error 02.71.001
For internal use only. Copyright reserved.
123.
Remedy:
Renew Pneumatic Controller PCB
GBR5664500T025664800.fm
Remedy:
Renew valve Y1.1 in gas connection.
The following applies to all Rosy errors: Operating-system software errors. The operating
system runs on the CPU 68332 PCB, the Pneumatic Controller PCB, the Graphics
Controller PCB (Evita 4)/Front Panel PCB (Evita 2 dura) and the Communication PCB.
Please report any problems to Lbeck.
The following applies to all Exception errors: Software problem on CPU 68332 PCB,
Pneumatic Controller PCB, Graphics Controller PCB (Evita 4) / Front Panel PCB (Evita 2
dura) or Communication PCB. Please report any problems to Lbeck.
receive queue
error
11.01.014 HPSV Air, Status 14, RAM refer to error 11.01.001
error
GBR5664500T025664800.fm
error
12.01.014 HPSV O2, Status 14, RAM refer to error 12.01.001
error
GBR5664500T025664800.fm
maximum period.
Remedy: Renew CPU 68332 PCB
functioning.
In the event of sporadic faults, localise
by replacing printed circuit cards.
13.98.002 BOOT error refer to error 13.98.001
GBR5664500T025664800.fm
13.99.140 pfl Read Msg Case incorrect Software error on CPU 68332 PCB.
13.99.801 EEPROM WRITE QUEUE Software error on CPU 68332 PCB.
FULL
13.99.919 NTH LONG MSG CORRUPT, Software error on CPU 68332 PCB.
any frame data of multi packet
receive message RPC corrupt
13.99.920 OS MESSAGE TYPE ERROR, Software error on CPU 68332 PCB.
GBR5664500T025664800.fm
paramerror in
baqu_ZusatzVonBA
SW:
(Evita 2 dura).
17.99.030 bu Illegal Msg ID Received By Software error on Graphics Controller
Benu Menu PCB (Evita 4) or Front Panel PCB
(Evita 2 dura).
GBR5664500T025664800.fm
(Evita 2 dura).
17.99.145 sibf Alignment Case incorrect Software error on Graphics Controller
PCB (Evita 4) or Front Panel PCB
(Evita 2 dura).
GBR5664500T025664800.fm
(Evita 2 dura).
17.99.942 ZDA EXCLUSIVE STACK Software error on Graphics Controller
UNDERFLOW PCB (Evita 4) or Front Panel PCB
(Evita 2 dura).
GBR5664500T025664800.fm
The service diagnosis mode is designed to assist with Evita 4 and Evita 2 dura servicing. It
enables on-line information to be read out at any time in the course of operation. Software
options can also be released in diagnosis mode.
Keypad
Service diagnosis mode can be called up at any time in the course of operation. This mode
only involves data readout; settings are not altered.
Important: If the Evita 2 dura is not equipped with the software option Service+, switch on
the service coding on COM1. To switch on the service coding, connect the RS232 extension
cable 7901808 or the RS232 adapter 7901888 to the COM1 socket.
Press Configuration key.
Select System defaults menu (Evita 4) or Ventilation (Evita 2 dura)..
Select Service diagnosis menu (Evita 4 only).
Evita 4
Evita 2 dura
IPPV Em ext.
int.
P aw Calibration / Configuration
60 ---------
4 5 6
Device
7 8 9
40
0
Display
20
0 Ventilation
-10
For internal use only. Copyright reserved.
180s
Evita 4
Evita 2 dura
IPPV Em ext.
int.
P aw Configuration
Service> Front Electronics Pneumatics Other
80
Displaytype 12 EPSONEG9013F-NZ-1
Micro-
proces. Keyspressed Device
60 00000000_00000000_00000000_00000000_00000000
Sounddetected no
RAM ok Display
40
Inter-
20 face
Service
0 >
-10
10
Display type Readout of display in control panel. This information is of importance for
renewal of background illumination, as this differs from display to
For internal use only. Copyright reserved.
display.
GBR5664500T025664800.fm
Important: In the event of display fault, use must always be made of the current display type.
The display accessory set contains all the components required for installation, e.g. display,
decoders D29 and D40, wiring harnesses, DC/AC converter (if not integrated into display)
and mechanical assembly components.
Evita 4
IR-Touch: Touchscreen data. These data are determined once in BOOT after
switching on Evita. The important touchscreen information is listed
below:
02 RAM error
04 Broken beams in BOOT, refer to broken beams
0a Invalid command
0c Input buffer overflow
GBR5664500T025664800.fm
Evita 4
Evita 2 dura
IPPV Em ext.
int.
P aw Configuration
Service> Front Electronics Pneumatics Other
80 Ambientpress. [mbar]
Pambient 1013 PAir/O2 1010/1050
Micro-
proces. Raw 2.030V 1008 Pinsp 10.1 Device
60 P_Reference 1011
Pediatricflow meas shadow
Sensors Flow [L/min] 11.345 11.023
FlowNTPD/S [L/min] 11.345 11.023 Display
40 [V] 11.345 11.023
CalD/A [LSB_DA] 11.345 11.023
ZeroOffs [LSB_AD] 11.345 11.023 Inter-
20 U_FLimit/U_Gnd [V] 11.345 11.023 face
SpO2 NellcorMinisoomV1.0.1.011/08/9612:00
SpO2[%] 96 Puls[bpm] 96 Pleth 96 Bargraph 96
CO2 DRAEGERHL SWrev 68702771.1 Service
0 HeadSN rev cal >
000780 450253-00102 11-08-960A
Status 18000000 Selftest 00 Tdet 4683 Twin 4731
-10
180s
Ambient press. Output of ambient pressure measurement values. Measuring range 600
- 1100 mbar. Note: Values without a unit are output in mbar. The air
For internal use only. Copyright reserved.
Preference 12 %.
46C = 4201
48C = 3888
50C = 3602
GBR5664500T025664800.fm
Evita 4
Evita 2 dura
IPPV Em ext.
int.
P aw Configuration
Service> Front Electronics Pneumatics Other
80 5V 5.144V 3811 GoldCap 9.201V 3359
10V 10.007V 3937 Piezo 0.001V 3
Micro- 12V PowerSw
11.992V 2727 0.004V 2
proces. Device
15V 14.808V 3096 Battext. 0.123V 839
60
-15V -15.098V 1564 AD1 2.498V 4092
Sensors 24V 23.903V 3560 AD1 2.500V 4095
Tcpu[C] 29.6 1.721V Display
40 Tair[C] 0.2 2.402V
Pinsp[mbar] 0.9 0.490V
Pambient[mbar] 1015 2.047V Inter-
ILVin/out 10/11 face
20
EEPROM ok Lossofdata no
RAM ok RAM-Comm
Serviceconnector no Service
0 CANcomponents 11000000 >
SWoptions 1_11000000_10000000_00000000_00000000
-10
180s
5 V/12 V/15 V / Power pack supply voltages measured on CO2 Carrier PCB. Output of
-15 V/24 V : voltages and decimal values of AD conversion. There are no apparatus
error messages with the following values:
5V = 4.5 V to 5.5 V
12 V = 9.6 V to 14.4 V
For internal use only. Copyright reserved.
15 V = 14 V to 18 V
-15 V = -12 V to -18 V
GBR5664500T025664800.fm
Piezo: The current through the substitute horn is measured on the CO2 Carrier
PCB. The voltage via the shunt resistor and the decimal value of the AD
conversion are output. To check Evita 4, press and hold Reset/Check
key.
Piezo detected > 30 decimal (0.018V)
Piezo off < 30 decimal
PowerSW: Output of voltage at mains-switch auxiliary contact. Voltage is important
for detecting mains failure. Mains switch is in power pack and evaluation
circuit on CO2 Carrier PCB. Output of voltage and decimal value of AD
conversion.
Mains switch on < 164 decimal (0.100V)
Mains switch off > 512 decimal (0.312V)
Batt ext.: Output of voltage of externally connected DC power supply (option DC
module). Output of voltage and decimal value of AD conversion.
AD1: Output of AD converter reference voltage 1. Output of voltage and
decimal value of AD conversion. Value is not directly monitored. Voltage
is OK if 10V within tolerance.
Set value approx. 2.500V
For internal use only. Copyright reserved.
PCB/5x undefined
GBR5664500T025664800.fm
Evita 4
Evita 2 dura
IPPV Em ext.
int.
Paw Configuration
Service> Front Electronics Pneumatics Other
80
Valves HPSV
Micro- Y1.1 off off Pressure Status Device
proces. Y1.2 off O2 6.29bar 0
60 Y1.3 on Air 5.70bar 0
Y1.4 off off
Sensors Setvalues Display
Y6.1 off
FiO2 21.0Vol.%
40 Y6.2 off
Flow 0.0L/min
Yx.x off
Valves Yx.x off Measure
Inter-
Yx.x off FiO2 22.5Vol.%
20 face
Yx.x off
PEEP
PEEP Offset Gain Service
0 Setvalues Cal.values >
5.00mbar 547 74
Measure 5.00mbar
-10
180s
the actuation voltage is read back and output in the second value.
Actuation and checkback must be identical.
off Valve deenergised
on Valve actuated
GBR5664500T025664800.fm
HPSV: Supply pressures and status of cartridges for O2 and air. Settings for
flow and FiO2 are additionally output, as is the measured FiO2 value.
Pressure The absolute air and O2 pressures are output.
Absolute pressure = rel. supply pressure +
ambient pressure.
The supply pressure sensor is located in the
respective HPSV cartridge. The supply pressures
displayed are read in on the Pneumatic Controller
PCB. The respective supply pressure is read in
on the corresponding HPSV Controller PCB for
valve regulation. refer to "Repair Instructions -
Pneumatic Components chapter 4.4,
Troubleshooting in Mixer, page 17 in the event of
problems with mixer. Measuring range of supply
pressure sensors
= 0 - 7 bar
Sensitivity = 1.58 V/bar +/- 8 mV/bar
Offset voltage = 300 mV +/- 30 mV
Status: Status messages of respective HPSV Controller
PCB for O2 and air.
0 - OK, no error
2 - supply pressure less than 2 bar absolute,
failure of pressure supply 1, 3 to 15 - errors on
HPSV Controller PCB or in HPSV cartridge. If
such errors are encountered, interchange the two
HPSV Controller PCBs for O2 and air. If the error
then occurs with the other type of gas, the HPSV
Controller PCB is defective. If the error remains
For internal use only. Copyright reserved.
Evita 4
Evita 2 dura
IPPV Em ext.
int.
P aw Configuration
Service> Front Electronics Pneumatics Other
80 Pressuresensors Flowsensor
[mbar] [V] Cal[V] FlowExp
Micro-
proces. insp 0.0 1.738 1.735 [L/min] 0.0 Device
60 exp 0.2 1.781 1.773 [L/min]NTPD/S 0.0
oes -47.6 0.000 1.740 [V]Uinop F0 Fsig
Sensors aux -47.6 0.000 1.740 0.763 4.040 2.018
Paw insp exp CalD/A 2420 Display
40 [mbar] Switch
0.0 0.2 on
(PCBCPU) 1.0 NTPD->BTPS 1.083
Valves
R[mbar/L/s] 2.0 2.0 Quotient 1.064 Inter-
20 0.0 0.0 BTPS->NTPS 0.882 face
C[mL/mbar] 0.22 MVinsp[L/min] 0.0
O2Sensor
FiO2 [Vol.%] 30.9 PmeanO2[mbar] 0.0 Service
0 [V] 0.868/0.868 Cal[V] 2.787 >
P[mbar] 1008
-10
180s
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
measurement
GBR5664500T025664800.fm
Evita 4
Evita 2 dura
IPPV Em ext.
int.
P aw Configuration
Service> Front Electronics Pneumatics Other
80
Peripheryofprocessor
Micro- RAM ok
proces. EEPROM ok
Device
60 Fan ok
Sensors
ADC
Display
40 Uref5V 9.201V ok
ADCref 0.001V ok
Valves DACref 0.004V ok
Inter-
20 face
Service
0 >
-10
180s
Periphery of Status of processor system of Pneumatic Controller PCB and fan for
processor: cooling electronics in pneumatics section. Error message is given in the
event of deviation. Fan must be detected within 1 minute.
ADC: POutput of reference voltages of Pneumatic Controller PCB. The
voltages are generated on this PCB and also read in. Error message is
For internal use only. Copyright reserved.
Evita 4
Evita 2 dura
IPPV Em ext.
int.
Paw Configuration
Service> Front Electronics Pneumatics Other
80
Code Count firstoccurence firstoccurence
Log- 1
book 1 Device
60 1
1
1
Option 1
Release 1 Display
40 1
1
1
1
1 Inter-
20 1 face
1
1
1
1 Service
0 >
-10
180s
The error table can be sorted by repeatedly pressing the key Logbook (Evita 4). The arrow
above the table indicates the sorted column.
Refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 2, List of Evita
GBR5664500T025664800.fm
Evita 4
This menu can be used to release software options with a 10-position code. The code
number is governed by the ID number of the Evita. Checking is performed after entering the
number. If check is positive, software option is ready for use next time the unit is switched on.
Important:
In Evita 4 the hardware options (e.g. SpO2, DC, CO2) are enabled under Configuration ->
Basic setting -> Service diagnosis with a 4-position code, Refer to section of "Repair
Instructions - Service No./ Modi, Error list" chapter 1, Service and Access Codes/Release
Option, page 1.
In Evita 2 dura, the hardware options (e.g. SpO2, DC, CO2) are enabled under
Configuration -> Ventilation with a 4-position code (Refer to section of "Repair
Instructions - Service No./ Modi, Error list" chapter 1, Service and Access Codes/Release
Option, page 1.).
Evita 2 dura
The software option for Evita 2 can be released in the operation mode under configuration
=> unit => options.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
4
Switch
= Test the device on
Graphics
No Controller PCB
Yes
01.2002
No No Graphics
Graphics Yes
Red alarm LEDs Red alarm LEDs Controller PCB
Controller PCB go off come on power supply
24 V
Check
Yes No power supply unit
voltages with and
without front
panel
Graphics
Display Yes No No
Backlight is on DC voltage at the Controller PCB
(possibly the Graphics
DC/AC converter or DC/DC converter
Controller PCB)
is present
*
Repair Instructions
Yes
Alarm cause: a. The tube system or one of the tubes has fallen off or is
leaking.
b. The expiratory pressure sensor supplies too low a measu
red value.
Detection: a. In all respiration modes with a minimum PEEP of 3 mbar:
If the airway pressure measured with the expiratory pres
sure sensor falls below the set PEEP, the PEEP - airway
pressure difference is integrated in a sum memory. An
alarm is issued if the value of the sum memory exceeds a
specific value. The sum memory is cleared if the measur
red airway pressure again exceeds the PEEP.
b. In volume-controlled (IPPV, SIMV, MMV), pressure-
controlled (BIPAP, BIPAP(VG)) and pressure-assisted
respiration mode: The value measured by the expiratory
pressure sensor does not reach the PEEP setting +
threshold limit in two successive mechanical inspiration
strokes for more than 100 ns in each case. The amount
and derivation of the threshold depend on the form of
respiration:
Volume-controlled : 5 mbar
Pressure-controlled: 0.5 * Pinsp (Pinsp - PEEP)
Pressure-controlled : 0.5 * PASB (PASB - PEEP)
Suppression: a. Detection a. is not activated until PEEP settings of 3 mbar
and higher.
b. Detection b. is not activated in pure CPAP.
Reaction: Generation of the airway pressure low warning
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
Cancellation: The measured AMV value falls below the set high AMV limit.
GBR5664500T025664800.fm
The measured FiO2 value is compared against a high and low limit in order to generate an
alarm in the event of a discrepancy. These two limits are derived from the FiO2 setting.
For internal use only. Copyright reserved.
5.5.1 Apnoea
Alarm cause: a. Breathing gas humidifier heats up the inspiration gas too
much
b. Breathing gas temperature sensor returns too high a
measured value
Detection: The value measured by the breathing gas temperature sensor
is above the limit of 40C for more than 1 s
Suppression: Only active if a breathing gas temperature sensor has been
detected as having been plugged in
Reaction: Generation of the temperature high warning
Cancellation: The value measured by the breathing gas temperature sensor
is below the limit.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
Alarm cause: Flow sensor not inserted in the patient section (flow sensor
switch open)
Detection: The flow sensor switch is detected as being open. Detection of
this fault is counted by a fault counter.
Suppression: Flow monitoring is off
For internal use only. Copyright reserved.
As a safety facility, the watchdogs must be tested because important safety aspects are
transmitted to the watchdogs of the individual processors. This test is only run in the event of
a cold start of the unit.
During a cold start, functioning of each watchdog is tested once. To do this, the watchdog is
not triggered in the test sequence and the unit waits for the occurrence of a reset for a
maximum time.
If the reset occurs within the maximum time, the watchdog is OK. Otherwise, the watchdog is
defective.
Owing to the structure of the watchdog, the reset source can be detected after the reset.
Thus, this is only evaluated if the reset has been triggered by the watchdog and not by further
reset sources.
Audible alarming is active during watchdog reset (the piezo is coupled by hardware to the
reset line).
The main processor of the Evita features battery buffering of the entire RAM area. This
guarantees retention of data that requires protection (referred to as back-up data, e.g. mode
parameters, calibration data, log book and trends) in the event of deactivation or failure of the
power supply. This retention of data may be destroyed by the following conditions:
a. The battery voltage is inadequate for buffering the data
b. Writing into a memory cell has been interrupted by a reset
c. An alpha particle has altered the contents of a memory cell.
Therefore, before using this data any further, a check must be made as to whether the
retained data has been destroyed. If such destruction of data is detected, wherever possible
the data should be restored. Only if restoration is no longer possible by justifiable effort is the
data assigned defaults from the ROM. In this case, the unit issues a data loss warning, which
it cancels again automatically after 10 s.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
The main processor of the Evita features an electrically erasable programmable read-only
memory. All configuration data of the Evita 4 is stored in this memory. Even after a complete
loss of all stored data in the battery-buffered RAM, Evita 4 still contains the user settings.
Apart from being destroyed by a component defect, data retained in the EEPROM may also
be destroyed by a reset while data is being written into it. Therefore, before using this data
any further, a check must be made as to whether the retained data has been destroyed. If
such destruction of data is detected, wherever possible the data should be restored. Only if
restoration is no longer possible by justifiable effort is the data assigned defaults from the
ROM. In this case, the unit issues a data loss warning, which it cancels again automatically
after 10 s.
This RAM test is run through completely in one go after a cold start to be able to detect any
possible error source within the scope of the self-test.
All ROM areas of all processors are subjected to a ROM test. This RAM test is run through
completely in one go after a Power On to be able to detect any possible error source within
the scope of the self-test. This ROM test is also run cyclically during execution of the program
and, in doing so, the cells to be tested in each program cycle are adapted to the run times.
These components are tested within the scope of execution of the program and not directly
after Power On.
Monitoring during the hardware check. During the hardware check, functioning of the air/O2
changeover valve is checked by one of the test points. If a malfunction is detected during this
test, a changeover valve fault warning is issued. This warning can only be eliminated by
repeating the check or by switching off/on.
For internal use only. Copyright reserved.
The required setting of the nebuliser valve is compared continuously against the fed-back
signal of the electronic control circuitry. A hardware fault message is issued if the required
setting and the fed-back signal do not agree.
Monitoring during the hardware check. During the hardware check, functioning of the
safetyvalve is checked by one of the test points. If a malfunction is detected during this test, a
safety valve fault warning is issued. This warning can only be eliminated by repeating the
check or by switching off/on.
The maximum times of the power-on test were taken into account when designing the
behaviour of the system with regard to communication between the single components.
Directly after activation of the unit (cold start), the 5 V monitoring unit of the CPU 68332 pc
board is checked with regard to its undervoltage and overvoltage functions. Too low or too
high a supply voltage is simulated with the aid of test lines of the 5 V monitoring unit. The
incorrect voltage must trigger a hardware reset. A hardware fault is reported if reset is not
triggered.
Once after activation of the unit (cold start), the CPU -> pneumatic, pneumatic -> CPU and
the CPU -> display reset lines are tested by activation of a hardware reset. As the result of
resetting of the other processor, the respective processor must detect an interruption in
communication, for example. A hardware fault is reported if reset is not triggered.
The supply voltages of the pressure sensors are monitored in order to check the A/D
converters. The +10 V reference voltage for the pressure sensors is generated from the +15
V operating voltage. Both these voltages are applied to inputs of the A/D converter on the
For internal use only. Copyright reserved.
CO2 carrier pc board. A hardware malfunction is assumed if the voltages drop below the
rated voltages by more than 1 V.
GBR5664500T025664800.fm
The operating voltages accessible via an A/D converter are compared cyclically with the
defined tolerances of +/- 20%. A hardware malfunction is reported if these tolerances are
exceeded for more than 5 s.
In the event of a power failure, the gold cap supplies a voltage to the piezo alarm tone
generator, thus issuing a pas failure alarm. The duration of this power failure alarm depends
on the capacitance of the gold cap and, therefore, both functioning and also the capacitance
are subjected to a test every hour, for the first time five minutes after activation of the unit.
a. The final charge voltage is too high (>= 11 V) Voltage regulator defective
b. The final charge voltage is too low (< 8 V) Charging circuit defective
c. The reference voltage is too high (>= 6.3 V) Discharge switch defective
d. The reference voltage is too low (< 4.5 V) no gold cap
e. Voltage of the gold cap has dropped by more than Gold cap capacitance too low
10 % of the reference voltage within 4 s
The test of the gold cap is aborted if it is interrupted by piezo alarming (e.g. lamp test or alarm
of the safety software). The test is repeated after a recharging time of ten minutes.
If the system detects an optional internal rechargeable battery (in the DC/DC module of the
power supply unit), the voltage of this rechargeable battery is compared against the defined
limits. The procedure is the same in the event of a further external rechargeable battery. In
both cases, an alarm is issued if the defined limits are exceeded.
When using the optional DC/DC module, Evita can be operated from 3 voltage sources,
which are assigned the following priorities:
1. Mains voltage
2. External rechargeable battery
3. Internal rechargeable battery
Therefore, in the event of a power failure Evita first switches to the external rechargeable
battery. If the voltage is inadequate, the DC/DC module then switches to the internal battery.
On the basis of a status signal, Evita is able to recognise whether the internal battery is fully
charged. As soon as Evita has switched to the internal battery, the rechargeable battery is no
longer fully charged. The guaranteed operating time of seven min is monitored by a timer and
is indicated to the user by different alarms.
If the unit switches back to mains operation after activation of the internal battery, the timer is
incremented, but clearly more slowly than during discharging.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
The loudspeakers control signal is tapped via a dropping resistor and is made available to
the system as digital information. A nominal/actual value comparison of loudspeaker control
is possible on the basis of this signal. A hardware malfunction is reported if a fault is detected
within the scope of this novel/actual value comparison.
During their use, the internal interfaces (CAN, RS232) are checked in relation to operability.
A hardware malfunction is reported in the event of a fault.
Besides alarming a power failure, the purpose of the piezo alarm tone generator is to warn of
a CPU defect (standstill) instead of a loudspeaker. To guarantee this, the piezo is controlled
by the TPU. If cyclic clocking fails, the resulting muting of the piezo is cancelled and a
continuous beep sounds. To test the piezo, muting is consciously cancelled. The current
through the piezo, which now rises as the result of cancelled muting, can be measured as a
voltage drop through a dropping resistor by the A/D converter. The test is repeated every
hour.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
When Evita is in standby mode, the user is offered a hardware check softkey on the main
standby page. When this softkey is pressed, the user is shown the result of the last hardware
check, along with the date of the check, on the next on-screen page. The results of the check
consist of data worthy of protection which is retained even beyond a cold start. (If the check
has been run after preparation of the unit, it is possible that Evita will not be switched off until
it is used. Nevertheless, it should still be possible to retrieve the result of the hardware check
when using the unit.)
A Check softkey is displayed on this on-screen page to start the hardware check.
When this check is selected, a complete hardware check is run, involving the following:
visual check of the patient section
visual check of the flow sensor
visual check of the thermometer (optional)
visual check of tube layout
visual check of the CO2 sensor (optional)
visual check of the paediatric flow sensor (optional)
visual check of the humidifier
check of the piezo alarm tone generator
check of the air/O2 changeover valve
check of the electric safety valve
adjustment of the flow sensor
adjustment of the O2 sensor
adjustment of the CO2 sensor (optional)
adjustment of the paediatric flow sensor (optional)
For internal use only. Copyright reserved.
The hardware check runs automatically, but permanently provides the user with information
about the currently running test, involving the user in the sequence of the check with prompts
that have to be answered by a Yes or No answer. The check can be interrupted at any time by
prematurely pressing the Check softkey. It is not possible to press any of the hard keys while
a check is running.
At the end of the check, the unit lists the scope of the check on an on-screen page along with
the results of the check in the form of a checklist containing OK/not OK entries. The user is
offered an additional softkey:
Repeat check: the selected check is repeated.
If the check has been aborted by pressing the Check softkey again, the results of the tests
that have not been run are set to the Unknown status.
Test OK n
Test not OK F
Test not run --
If you select a repeated check owing to a defective test result, merely those test points are
repeated that were ascertained during the previous check as being defective or which were
not run. Only the leakage check of the tube system is run through again each time the check
is repeated because user activities within the scope of the other test points call for a re-
evaluation.
All test points are run again if the check is repeated from the main standby page.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
1. Visual checks
2. Safety checks
3. Adjustments
4. Leakage check
The user is prompted to check correct seating and latching of the patient system, of the flow
sensor, of the AWT sensor, of the CO2 sensor with cuvette and the level and operability of
the humidifier. The completeness and correct seating of the tube system and the waterfalls
are also checked.
The user is prompted to specify whether an audible signal of the piezo alarm tone generator
can be heard.
An automatic check of this changeover valve is only possible during the hardware check. A
test lung is filled up to reference pressure and the supplied gases must then be deactivated
individually.
1. The user is prompted to supply the unit with compressed air only.
2. The compressed air supply is checked with the aid of the HPSVs absolute pressure
sensor and the HPSV error line F1.
3. The user is prompted to collect a test lung.
4. The unit applies a continuous flow of 30 L/min
5. The test run is filled to a reference pressure of 30 mbar, expiratory measurement.
6. The flow delivery is discontinued when the reference pressure is reached.
7. The air/O2 changeover valve is set to O2.
8. The pressure in the lung must dissipate within a defined time as otherwise the
changeover valve is defective.
For internal use only. Copyright reserved.
The user is prompted to fill the unit with oxygen if no pressure can be built up in the test run
despite a continuous delivery of flow. The air/O2 changeover valve is defective if pressure
now builds up. The electric safety valve is defective if still no pressure is built up.
GBR5664500T025664800.fm
The input pressures of the pressureless supply lines measured during the test serve as a
reference for the absolute pressure sensor.
8.5.3 Calibrations
The calibrations of the sensors could basically be combined in one single adjustment
because the mechanisms underlying these assessments are very similar. However, both the
CO2 and the paediatric flow sensors are options and therefore they do not have to be
installed.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
Tube system:
The user is prompted to keep the Y-piece of the tube system tight before continuing the test.
During the course of the test, a pressure of 60 mbar is built up in the tube system by the
application of inspiration flow while the expiration valve is closed. The flow needed to
maintain this pressure is determined.
The compliance of the tube system, which can be calculated on the basis of the volume
needed to build up a pressure of 60 mbar, is a by-product of this leakage test.
The check of the airway pressure sensors and the valves for adjustment can be carried out at
the same time as the leakage test. The pressure drop waiting time of 10 s can be used as the
period of time for this check. During this time, the sequence of the test is as follows:
1. Once the filling pressure has been reached, wait for one s for the system to stabilise
2. Compare the measured values of both airway pressure sensors
3. Switch the adjustment valve of the inspiratory airway pressure sensor and check the
value measured by the sensor
4. Reset the adjustment valve of the inspiratory sensor
5. Switch the adjustment valve of the expiratory sensor and check the value measured by
the sensor
6. Reset the adjustment valve of the expiratory sensor
7. Compare the values measured by both pressure sensors
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
9.1.1 Usage
External DS mode via PC permits readout of all important parameters (e.g. sensors, voltages,
errors or software status), activation of all actuators (e.g. mixer, valves or LEDs) and the
implementation of test routines. When doing so, the Evita does not function with the normal
operating software, but rather without ventilation employing a special service software. This
special status is displayed on the Evita screen. When employing the PC DS mode, actuator
activation is not subject to the restrictions which always apply for safety reasons in the case
of an internal service mode.
Usage involves compliance with the following hardware and software prerequisites:
Switch on Evita.
Set loudspeaker volume to minimum.
Activate all monitoring parameters (O2, flow, SpO2 and CO2) where fitted or
possible.Perform calibration for O2 and flow sensor. Calibration data are adopted for
PC service mode. Calibration is however also possible in PC service mode.
Switch off Evita.
Important: PC DS service mode also functions in full without control panel.
Connect Evita (COM1) and PC (COM1 or COM2) using RS 232 extension and RS 232
adapter. Important: This link is not to be interrupted.
Switch on PC.
Start PC program under C: by entering Service and pressing Enter key.
Use cursor keys to select Meditest Evita and press Enter key.
Use cursor keys to select program Evita 4 service mode and press Enter key twice.
Switch on Evita following display of user menu. Evita starts with self-test. Drger service
mode is then displayed on the Evita once the control panel has been connected. The
display Remote to Evita appears on the PC at the bottom of the screen. Service mode
is active as of this point.
Important: If service mode does not function without control panel, check whether there
is a jumper on the Pneumatic Controller PCB at X24 between 1 and 2. This terminates
the internal CAN bus with a resistance of 121 ohms. The first units did not feature this
jumper. It can remain in the unit.
Service mode is terminated by switching off the Evita and interrupting the link at COM1
of the Evita. Exit from the PC program via the menu.
A suitable modem enables the Evita to be switched to external DS mode and all the
described commands executed.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
These tests enable the following data to be read out and processed:
Baud rate alteration
Readout of Evita software version
Logbook (user logbook) readout, storage and erasure
Error list readout, storage and erasure and readout of length of time since error list last
erased
Readout of identification and released options
Readout of Evita operating hours
Readout of PCBs connected to internal CAN
This command enables the baud rate which is normally 9600bd to be changed. The
command has not yet been implemented in the PC software.
The output takes the form of a 4-position display. For example, the software version 2.00 is
output as 0200. Important: All software versions on the PCBs must be identical.
The user logbook to which users have access during operation is processed. Erasure is only
permissible if the user no longer requires the data. It is also possible to store data, for
example with a view to producing statistics.
Read out Logbook: Entries with error number (which are not displayed during
operation), date, time and plain language message are output.
Logbook storage: This is a subroutine of Logbook readout.
The data can be copied, for example, directly onto the floppy disk
drive a:.
To do so, enter: a:\file name.txt
Important: Always enter apparatus data such as serial number,
For internal use only. Copyright reserved.
Error list readout: The error number, first occurrence, last occurrence and error
frequency are output.
Operating-hour error The error-list operating hours since last erasure are output. This
list: information is extremely important for error statistics.
Error list storage: This is a subroutine of Error list readout
The data can be copied, for example, directly onto the floppy disk
drive a:.
To do so, enter: a:\file name.txt
Important: Always enter apparatus data such as serial number,
software version, error-list operating hours, Evita operating hours
and other remarks under Comments.
Error list erasure: All entries are erased (identical with code entry 1 3 9 9 + 1 3 2 2).
Evita elapsed-time meter which is output during operation. The value is stored in the
EEPROM D22 on the CPU 68332 PCB.
Output of PCBs connected to internal CAN at CPU 68332 PCB (0 = not fitted, 1 = fitted).
Sequence is as follows from left to right: Graphics Controller PCB/Pneumatic Controller
PCB/Communication PCB/5x undefined
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
The voltages at the actuation transistors can be measured on the Pneumatic Controller PCB
(refer to section of "Repair Instructions - Electronic Components" Chapter 16, Pneumatics
Controller PCB, page 76 under Pneumatic Controller PCB configuration). Measurement is
made in each case between X21/2 (AGND) and the ICs D17 and D18 with the actuation
transistors. The valves must be connected when performing measurement. A measured
voltage of approx. 24 V signifies that the valve is off, whereas approx. 0 V indicates that the
valve is on.
In practice it is possible to have a set value between 0 and 120 mbar, as the measuring range
of the airway pressure sensors is restricted to 140 mbar. The set pressure is the pressure
resulting in the tubing system (= closing pressure in expiration valve) and not the pressure at
the outlet of the PEEP/PIP valve.
The current through the PEEP/PIP valve can be measured as voltage by way of a shunt
resistor on the Pneumatic Controller PCB between X15/1 (AGND) and X15/2 (1 mV
measured = 1 mA). A current of 500 mA corresponds to a PEEP/PIP valve outlet pressure of
roughly 120 mbar, refer to "Repair Instructions - Pneumatic Components chapter 3.2, Testing
and Calibrating the PEEP/PIP Valve, page 8
Attention: The resultant pressure in the tubing system in the no-flow state is lower than the
set value. As starting value for ventilation, the ventilation software adds 2 mbar in each case
to the desired pressure value.
The absolute pressures for air and O2 are output, as are the halved
sensor voltages. Absolute pressure = rel. supply pressure + ambient
pressure. The supply pressure sensor is located at the respective
HPSV cartridge. These displayed supply pressures are read in on the
Pneumatic Controller PCB. For valve regulation, the respective
supply pressure is read in on the corresponding HPSV Controller
PCB. Should mixer problems be encountered, refer to "Repair
Instructions - Pneumatic Components chapter 4, HPS Valve, page 13
sensor.
Pressure: Pressure in mbar of calibrated sensor
Flow measurement: Expiratory flow measurement values. The sensor is read in on the
Pneumatic Controller PCB. If a lengthy period has elapsed since
calibration of the flow measurement during operation, this can be
implemented in the calibration menu (section 9.3.6 on page 129).
Note: All values are displayed under BTPS conditions. BTPS: Based
on 37C, ambient pressure + insp. pressure, 100 % relative humidity.
All Evita measured values and settings are based on BTPS.
F0:
Flow-measurement bridge voltage with 4-fold
amplification.
Set value in calibrated and no-flow condition = 4.04
V
Fsig:
Flow-measurement bridge voltage with 2-fold
amplification.
Set value in calibrated and no-flow condition = 2.02
V
These voltages can be measured on the Pneumatic
Controller PCB at the plug connector X19.
X19 / 7 AGND
X19 / 8 Fsig
X19 / 9 F0
X19 / 10 Finop
X11 / 1 DGND
X11 / 2 not used
X11 / 3 +5V
Fan voltage: The current through the fan at the bottom right of the pneumatics
section is monitored on the Pneumatic Controller PCB. In normal
operation, this voltage must be subject to fluctuations of more
than 50 mV. The minimum and maximum values for this voltage are
each measured for 60 seconds. If the difference is < 50 mV, it is
assumed that the fan is not operating.
For internal use only. Copyright reserved.
Zeroing of insp. and Start of zeroing of airway pressure sensors S6.1 (insp.) and S6.2
exp. pressure sensor: (exp.)
Zeroing of oesoph. Start of zeroing of oesophagus pressure sensor (option not yet
pressure sensor: available)
Flow sensor Important: There must not be any flow through the flow sensor. This
calibration: must be ensured by the DS mode user.
PEEP valve: Start of calibration of PEEP valve (identical with code entry
4 7 9 9 + 4 7 4 7). Important: The test set-up refer to "Repair
Instructions - Pneumatic Components chapter 3, PEEP Valve, page
6 On calibration, the apparatus and the PEEP/PIP valve must be in
horizontal operating position, as the function of the PEEP/PIP valve
is position-sensitive. The safety valve Y1.3 must be switched on prior
to calibration. The criterion for positive calibration is the flow pulse
from the mixer, which is why no hose is to be attached to the
inspiration socket.
Calibration OK Flow pulse 0.2 s
Calibration not Flow pulse 1.0 s
OK:
Possible sources of error:
Pressure supply not connected
Safety valve Y1.3 not switched on
Incorrect test set-up or test set-up leaking; refer to "Repair
Instructions - Pneumatic Components chapter 3, PEEP Valve, page
6
Expiratory airway pressure measurement defective. The value can
be followed up in DS mode; section 9.3.4 on page 123 on airway
pressure measurement. Set value must switch back and forth
between 3 and 34 mbar.
Mechanical PEEP/PIP valve zero point defective; refer to "Repair
Instructions - Pneumatic Components chapter 3, PEEP Valve, page
6
For internal use only. Copyright reserved.
9.3.7 CPU
This test can be used to check the EEPROM and the RAM area on the Pneumatic Controller
PCB. A defective ROM (more precisely flash EPROM) stops the Evita functioning.
RAM test: Output of OK or error; the last successfully tested RAM address is
also output
EEPROM test: Output of OK or error; the PEEP/PIP valve calibration data are
stored in this EEPROM
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
5V / 12V / 15V / -15V / Supply voltages from power pack; measured on CO2 Carrier
24V: PCB. Output of voltages and decimal values of AD conversion.
The following values do not result in apparatus error messages:
5V = 4.5V to 5.5V
12V = 9.6V to 14.4V
15V = 14V to 18V
-15V = -12V to -18V
24V = 19.2V to 28.8V
Important: The power pack is specified as follows; refer also to
Section 7.1 Repair instructions:
5V = 5.07V to 5.23V
12V = 11.76V to 12.24V
15V = 14.55V to 15.45V
-15V = -14.55V to -15.45V
24V = 22.32V to 24.72V
Internal battery status: The charge of the internal battery in the DC module (option) is
displayed. The digital signal from the DC module is read in by
the CPU 68332 PCB.
Charge output:
Charged: Internal batteries have reached max. charge
voltage
Undefined: Internal batteries have still to reach max. charge
voltage or are not connected or are defective
External battery status: The charge of the external battery of the DC module (option) is
displayed. The digital signal from the DC module is read in by
the CPU 68332 PCB.
Charge output:
Charged: External battery/batteries has/have reached
max. charge voltage
Undefined: External battery/batteries has/have still to reach
max. charge voltage or is/are not connected or
is/are defective
Important: The voltage can be interrogated under Voltages. 1
or 2 gel batteries in series with a rated voltage of 12V is/are
envisaged. Gel batteries are charged at a charge voltage of
13.8V.
Internal/external When supplying the Evita by way of the internal batteries or the
battery operation: batteries of the DC module (option), a digital signal from the DC
module is read in on the CPU 68332 PCB.
Off: No supply by internal or external batteries
On: Supply by internal or external batteries
Important: In DC operation, only one signal may be set to on.
All status signals from the power pack can be measured at
connector X2 of the CPU 68332 PCB. When using an extender
board, it is also possible to employ any slot in the motherboard
above or below for X2. All signals are low-active.
X2 / 26b Internal battery
X2 / 27b External battery
For internal use only. Copyright reserved.
Ambient pressure: Air pressure currently measured in Evita. The air pressure sensor
with AD conversion is located on the CO2 Carrier PCB.
Output of air pressure in mbar and voltage.
Permissible measuring range = 600 to 1100 mbar
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
CO2 Vendor ID: Output of type of Signal Processor PCB used for CO2
measurement on CO2 Carrier PCB, e.g. Andros 4210 or
Drger HL.
CO2 Head Output of part number and modification status of CO2 sensor
Modification:
Test RS 232.
OK:
Optocoupler on CO2 Carrier PCB defective since CPU
68332 PCB OK
GBR5664500T025664800.fm
Not OK:
CPU 68332 PCB defective
SpO2 measured values: Values from SpO2 measurement. Valid values are only provided
if option fitted and released and SpO2 measurement activated.
Note: The bar graph value is not displayed on the Evita in
operation.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
Substitute horn: The substitute horn is located on the CO2 Carrier PCB and is
actuated by the CPU 68332 PCB. The horn is supplied by the
GoldCaps on the CO2 Carrier PCB. The GoldCap voltage can
be read out under Electronics -> Voltages. The horn can be
switched on and off; the horn current value is to be read out.
Possible sources of error if horn cannot be switched on:
Mains switch defective; for testing section 9.4.2 on page
133. Furthermore GoldCap voltage must be present
(section 9.4.1 on page 131).
Testing horn without suppression signals from CPU 68332
PCB:
Switch off Evita and pull out CPU 68332 PCB. Switch on
Evita. Testing: Continuous horn tone. If this is not the case
and the two items above are OK, there is every likelihood
that the CO2 Carrier PCB is defective.
Re-assemble Evita ready for operation.
Horn current: The current through the substitute horn is measured on the CO2
Carrier PCB. The voltage across the shunt resistor and the
decimal value for AD conversion are output.
Horn detected > 30 decimal (0,018V)
Horn off < 30 decimal
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
Note: Identical with test in Section 33.5.7. The loudspeaker in the control panel is actuated
by the CPU 68332 PCB in the electronics section. Monitoring is performed by the Graphics
Controller PCB by measuring the current through the loudspeaker. The loudspeaker can be
subjected to resistance testing (refer to error message 02.71.001). The test tone must be
recognised when it is transmitted (volume of test tone depends on setting in operating
mode). Detection of current flow through the loudspeaker is displayed for a further
10 seconds. For test purposes, the volume should be set to minimum. This setting is however
only possible in normal operation.
9.4.8 GoldCap
Gold Cap: Output of GoldCap voltage. The GoldCap capacitor supplies the
substitute horn with current. The GoldCap with charging circuit
and measurement is located on the CO2 Carrier PCB. Output of
voltage and decimal value of AD conversion.
Set value = 8 to 11 V
Result of GoldCap test: This test only indicates the last result, as the GoldCap test takes
longer than 10 minutes
The ILV interface is on the CPU 68332 PCB. It features 2 digital outputs and inputs for
synchronising 2 ventilators with ventilation from two sides. For test purposes, pin 4 (ILVout1)
is to be connected at the ILV interface to pin 8 (ILVin1) with pin 6 (ILVout2) being connected
to pin 1 (ILVin2).
This test can be used to check the EEPROM (D22) and the RAM area on the CPU 68332
PCB. A defective ROM (more precisely flash EPROM) stops the Evita functioning.
There is certainly a loss of data in the EEPROM D22 if for example the released options or
operating hours are no longer present following switch-on. A data loss in the RAM may be
due to a flat battery on the CPU 68332 PCB. After a loss of data, the Evita attempts to
restore the data. If this is successful, the red alarm Data loss disappears from the screen.
The Evita is OK if this alarm is no longer present when the apparatus is next switched on.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
The term Front panel refers to the detachable control panel of the Evita. Front panel
testing consists of the following steps:
The keys are read in by the Graphics Controller PCB (Evita 4) or by the Frontpanel PCB
(Evita 2 dura).
Output:
00000000
00000000
00000000
00000000
00000000 (last line applies to Evita 2 dura only)
First line: IPPV, SIMV, printer, free key under exp. hold, VT, Tinsp,
second menu key from top, top menu key
Second line: Ramp, PEEP, additional modes, BIPAP, PASB, Pinsp, f,
flow
Third line: Bottom menu key, curves 1/2, backlighting, screen
freeze, insp. hold, exp. hold, fourth menu key from top,
third menu key from top
Fourth line: Values 1/2, none, none, free key on the right-hand side
of values 1/2, nebulizer, none, none, 100% O2
Fifth line: Trigger, alarm reset, info, free key below trigger, standby,
keylock, O2, silence key
Several keys defective: If several keys in a row or several keys in a column (e.g.
nebuliser, mode and info) are defective, the fault is in all
likelihood to be found on the Graphics Controller PCB
(Evita 4) or Frontpanel PCB (Evita 2 dura)
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
The LEDs on the keypad are actuated by the Graphics Controller PCB (Evita 4) or Frontpanel
PCB (Evita 2 dura). This program enables all LEDs to be jointly switched on and off or
individually actuated in a defined manner with LED test.
Testing:
Actuation of all LEDS: All LEDs light. Note: The basic screen, info and
reset/check keys have no LEDs.
Switch-off of all LEDs: All LEDs off
Method of input:
LED 1 to 8: 00000000
LED 9 to 16 00000000
LED 17 to 24 00000000
LED 25 to 32 00000000
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
No. LED
1 4 Red alarm
2 to 9 None
10 Yellow alarm, right
11 Standby
12 Yellow alarm, second from right
13 Yellow alarm, second from left
14 Stop
15 Alarm tone suppression
16 Yellow alarm, left
17 None
18 Configuration
19 Calibration
20 In key between calibration and measurement manoeuvre
21 Measurement manoeuvre
22 Measured values
23 Alarm limits
24 Mode
25 None
26 Printer
27 Key above printer
28 Key between 27 and 29
29 Key beneath insp. hold
30 Insp. Hold
31 100% O2
32 Nebuliser
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
No. LED
1 none
2 Info key
3 VT
4 Free key under exp. hold
5 Exp. hold
6 Insp. hold
7 Nebulizer
8 Standby
9 100% O2
10 Flow
11 None
12 f
13 Additional modes
14 Ramp
15 Printer
16 PEEP
17 Backlighting
18 O2
19 Values 1/2
20 Free key below trigger
21 Free key on the right-hand side of values 1/2
22 Trigger
23 IPPV
24 Alarm reset
25 Tinsp.
26 PASB
27 BIPAP
28 Pinsp.
29 Screen freeze
30 Keylock
For internal use only. Copyright reserved.
31 Curves 1/2
32 Silence key
33 SIMV
GBR5664500T025664800.fm
34 warning yellow
35 warning red
Applies to Evita 2 dura only: all segments can be switched on and off.
The rotary knob on the front panel is read out by the Graphics Controller PCB (Evita 4) or
Frontpanel PCB (Evita 2 dura).
The touchscreen, which features its own processor system, communicates by way of RS 232
with the Graphics Controller PCB. During communication, a LED on the touchscreen PCB
flashes. This can be seen with the control panel open.
Error codes: These data are determined once only in BOOT after switching on
Evita.
Output 00 (OK, no error) or output of number of errors (two positions)
plus error code.
Error codes:
01 EPROM error
02 RAM error
04 Broken beams in BOOT; refer to Beam error
0a Invalid command
0c Input buffer overflow
0e Wrong number of parameters for command
0f Value range parameter exceeded
For internal use only. Copyright reserved.
Lines: 0000000000_0000000000_0000000000
(from bottom to top)
The display is actuated by the Graphics Controller PCB (Evita 4) or Frontpanel PCB (Evita 2
dura).
Important: If the display no longer responds to the following commands, but the rest of the
front panel is functioning properly and the background illumination is also visible, then either
the display (more likely of the two) or the Graphics Controller PCB may be defective. Testing
using simple means is not possible in this case.
Important: In the event of a display fault, use must always be made of the current display type.
The display accessory set contains all components needed for installation, e.g. display,
decoders D29 and D40, wiring harnesses, DC/AC converter (if not integrated into display)
and mechanical components for assembly.
For internal use only. Copyright reserved.
This test can be used to check the RAM area on the Graphics Controller PCB. A defective
ROM (more precisely flash EEPROM) stops the front panel working.
Testing by way of setpoint input between 0 and 4095 mV. A test cable is in preparation.
Testing by way of setpoint input between 0 and 4095 mV. A test cable is in preparation.
Testing is performed by way of a LOOP test. For this purpose, sockets 2 and 3 at the
interfaces Com 2 and Com 3 are to be shorted and the test started.
This test permits checking of the RAM area on the Communication PCB. A defective ROM
(more precisely flash EEPROM) stops the PCB functioning.
This test has still to be implemented, as use is not made at present of the external CAN bus.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm
Note: This function can only be called up with a password as of Software Version 8.n.
A description of unit error messages can be displayed for Service Software Version 8.n and
all higher versions under Help => Help.
For internal use only. Copyright reserved.
GBR5664500T025664800.fm