ML255 Electromagnetic Flow Meter Converter: Installation, Operation and Maintenance Manual
ML255 Electromagnetic Flow Meter Converter: Installation, Operation and Maintenance Manual
ML255 Electromagnetic Flow Meter Converter: Installation, Operation and Maintenance Manual
ELECTROMAGNETIC
FLOW METER CONVERTER
Important Information:
Converter Model Number: ________________________
Converter Serial Number: _________________________
Meter Serial Number: ____________________________
RETAIN THIS MANUAL - DO NOT DISCARD
Contents
SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.0 BATTERY POWERED CONVERTER OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.1 Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Compatible Products . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.0 CONVERTER INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1 Mounting The Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Installing Cables To Converter And Service Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Pulling Sensor Cable Through Electrical Conduit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.0 BATTERY INSTALLATION AND SYSTEM START-UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.1 Battery Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2 Installing Batteries and Starting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3 Setting the Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.4 Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.5 Powering-Down and Powering-Up the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.6 Power Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.7 Sampling Frequencies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.0 SOLAR PANEL OPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.0 ELECTRICAL CABLE CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.1 Converter Electrical Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.2 Terminal Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.3 FPI Mag 395L Forward Only Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.4 SPI Mag 282L Cable Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5.5 Ultra Mag Sensor Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5.6 Converter Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
5.7 Opto-Isolated Pulse Output Hook-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.8 Opto-Isolated Pulse Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.9 Converter Power Hook-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.0 CONVERTER START-UP - ALL SENSORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
6.1 Menu Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
6.2 Front Panel Visualization Pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6.3 Factory Set Key Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.4 Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
7.0 PROGRAMMING EXAMPLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
8.0 MENU 0 - QUICK START . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9.0 MAIN MENU DESCRIPTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9.1 Menu 1 - Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
9.2 Menu 2 - Scales . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
9.3 Menu 3 - Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
9.4 Menu 4 - Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
9.5 Menu 5 - Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
9.6 Menu 6 - Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
9.7 Menu 7 - Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
9.8 Menu 8 - Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
9.9 Menu 9 - Data Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.10 Menu 10 - Diagnostic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9.11 Menu 11 - Internal Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
10.0 ALARM MESSAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
11.0 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
12.0 TROUBLESHOOTING GUIDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
APPENDIX 1.0 UNITS OF MEASURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
APPENDIX 2.0 CONVERSION TABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
WARRANTY STATEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46, Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
SAFETY
SAFETY
Safety Symbols And Warnings
Throughout this manual are safety warning and caution information boxes. Each warning and caution box will be
identified by a large symbol indicating the type of information contained in the box. The symbols are explained
below:
! This symbol indicates important safety information. Failure to follow the instructions can
result in serious injury or death.
i This symbol indicates important information. Failure to follow the instructions can result in
permanent damage to the meter or installation site.
Safety Warnings
When installing, operating, and maintaining McCrometer equipment where hazards may be present, you must
protect yourself by wearing Personal Protective Equipment (PPE) and be trained to enter confined spaces.
Examples of confined spaces are manholes, pumping stations, pipelines, pits, septic tanks, sewage digesters,
vaults, degreasers, storage tanks, boilers, and furnaces.
WARNING!
! Incorrect installation or removal of FPI Mag meters can result in serious injury or death. Read
the instructions in this manual on the proper procedures carefully.
WARNING!
! Never enter a confined space without testing the air at the top, middle, and bottom of the
space. The air may be toxic, oxygen deficient, or explosive. Do not trust your senses to determine if
the air is safe. You cannot see or smell many toxic gases.
WARNING!
! Never enter a confined space without the proper safety equipment. You may need a respirator,
gas detector, tripod, lifeline, and other safety equipment.
WARNING!
! Never enter a confined space without standby/rescue personnel within earshot. Standby/rescue
personnel must know what action to take in case of an emergency.
WARNING!
! Pressurized pipes should only be hot tapped, cut, or drilled by qualified personnel. If possible,
depressurize and drain the pipe before attempting any installation.
WARNING!
At the end of its lifetime, this product shall be disposed of in full compliance with the
environmental regulations of the state in which it is located.
1
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
BATTERY POWERED CONVERTER OVERVIEW
1.0 BATTERY POWERED CONVERTER OVERVIEW
Read this entire manual prior to installation and/or changing any settings. Retain this manual in your records. DO
NOT DISCARD.
The battery powered signal converter is the reporting, input and output control device for the sensor.
The converter allows the measurements, functional programming, control of the sensor and data recording to be
communicated through the front panel and I/O.
The battery powered microprocessor-based signal converter has a twelve-point curve-fitting algorithm to improve
accuracy, an 8-line graphical LCD display with 3 button programming, and a rugged enclosure that meets IP67.
In addition to a menu-driven self-diagnostic test mode, the converter continually monitors the meter's functionality.
The converter will output rate of flow and total volume.
The converter has two digital open drain outputs for monitoring volume or status and a digital switch closure
input.
The converter also comes standard with password protection and many more features.
Mod
el
Seria #: 88
00
Met l #: E12- 03xxx
er #: 34
MI1 567
2-34
56
3255
W.
PH (95 STET
1) 652 SON
-6811 AVE,
• htt HEME
p://ww T, CA
w.m 92
ccrom 545
eter.coUSA
m
R
E C
T S
N E
E
2
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
BATTERY POWERED CONVERTER OVERVIEW
1.1 Serial Numbers
The converter and sensor are supplied as a matched system. Verify the meter serial numbers on both the converter
and sensor match. This will ensure a properly calibrated system.
The tag on the side of the converter has the Converter Model Number, the Converter Serial Number, the Meter
Model Number and the Meter Serial Number. An example is shown below.
IMPORTANT: Verify the Meter Serial Numbers on both the converter and sensor match. This will ensure
i a properly calibrated system. The Meter Serial Number is located on the side of the sensor, and the
Converter Serial Number and the Meter Serial Number are located on a label on the side of the converter.
Ensure the Meter Serial Number on the sensor and the converter tags match.
3
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
CONVERTER INSTALLATION
2.0 CONVERTER INSTALLATION
2.1 Mounting The Converter
If possible mount the converter in an electronics shed or environmental enclosure. If the converter is mounted
outdoors a sun shield is recommended. The sun shield should be oriented in a direction to reduce sun damage
and ensure readability. The converter is mounted using 2 bolts. See Figure 1. A service loop in the cables is
required. See Section 3.2. This electronic unit is rated IP67 for temporary flooding.
!
WARNING: Do not connect any form of conduit directly to the converter enclosure. Doing so will allow
moisture and potentially dangerous gasses to enter directly into the converter. Attaching any conduit to
the enclosure, or altering the enclosure in any way will void the warranty.
4
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
CONVERTER INSTALLATION
2.3 Pulling Sensor Cable Through Electrical Conduit
It is very important to protect the end of the sensor cable when pulling it through a conduit. Water can accumulate
in low portions of conduit. Always use the factory supplied cable cover, or similar method, to seal the end of the
cable against water when pulling the cable through conduit. See Figure 4. This will ensure proper operation of
the meter.
i CAUTION: Do not cut the cable cover off. Doing so may damage the sensor cable and adversely effect the
calibration of the meter.
Rip Wire
Cable Cover
5
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
BATTERY INSTALLATION AND SYSTEM START-UP
3.0 BATTERY INSTALLATION AND SYSTEM START-UP
3.1 Battery Life
The estimated battery life with standard use and recommended sampling frequency (see section 3.7) is 3-5 years.
Additionally, there is a battery life calculator available to estimate battery life in your specific application. Contact
your local McCrometer representative or the McCrometer technical support team for a consultation on battery life.
Note: Ensure there is no power connected externally and that the battery DIP switches are in the OFF position.
1. Remove the battery holder from the ML255 converter by removing the two screws and lock washers holding
it in place.
2. Place two double battery packs (part number AGM009) on the ML255 battery holder as shown in Figure 5.
3. Secure new battery packs to the battery holder using tie
wraps.
4. Re-install the battery holder into the converter using the
previously removed screws and washers.
5. Plug the battery pack connectors into the B1 and B2
sockets on the main board.
6. Wait one minute.
7. Move battery DIP switch B1/B2 to ON. The converter will
begin booting and the red CPU LED will turn on solid. It
will then begin flashing after it completes booting. Booting
may take 2-3 minutes. Figure 5. ML255 batteries
6
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
BATTERY INSTALLATION AND SYSTEM START-UP
3.4 Replacing Batteries
Contact your local McCrometer representative or McCrometer Customer Service (customerservice@mccrometer.
com) to order replacement batteries.
1. Power down the system.
a. If there is external power applied to the converter, unplug it.
b. Press the Enter/Esc button for one second to turn on the front panel display.
c. Press the Enter/Esc button again to display the menus.
d. Select Main Menu and press Enter.
e. Enter the L2 access code.
f. Navigate to menu 10-Diagnostic and press Enter.
g. Select Stand-By and press Enter.
h. Long press Enter to execute. The display will indicate "Stand-By" momentarily and then go blank.
i. Move battery DIP switch B1/B2 to OFF.
The unit is now safely powered down.
3. Installing Batteries:
a. Remove the battery holder from the ML255 converter by removing the two screws and lock washers
holding it in place.
b. Cut the tie wraps from the batteries and dispose of the battery packs.
c. Secure new battery packs to the battery holder using tie wraps.
d. Re-install the battery holder into the converter using the previously removed screws and washers.
e. Plug the battery pack connectors into the B1 and B2 sockets on the main board.
7
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
BATTERY INSTALLATION AND SYSTEM START-UP
3.5 Powering-Down and Powering-Up the System
Note: This power-down procedure will only stop the system from operating. If you need to replace the batteries or
prepare the unit for shipping, follow section 3.4.
Powering-Down:
1. Press the Enter/Esc button for one second to turn on the front panel display.
2. Press the Enter/Esc button again to display the menus.
3. Select Main Menu and press Enter.
4. Enter the L2 access code.
5. Navigate to menu 10-Diagnostic and press Enter.
6. Select Stand-By and press Enter.
7. Long press Enter to execute. The display will indicate Stand-By momentarily and go blank.
The unit is now powered down.
Powering-Up:
1. Move battery DIP switch B1/B2 to OFF.
2. Wait one minute.
3. Move battery DIP switch B1/B2 to ON.
4. The converter will begin powering up and the red CPU LED will turn on solid. It will then begin flashing after
it completes booting.
Note: Powering up may take two to three minutes.
8
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
REAL FLOW PROFILE
CONTINUOUS SAMPLING
AVERAGE SAMPLING
S.R.=SAMPLE RATE
9
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
SOLAR PANEL OPTION
4.0 SOLAR PANEL OPTION 310 mm
12 ¼”
4.1 Description
193 mm
The solar panel provides power to the converter by
7 ½”
converting sunlight into electrical energy to recharge
the solar panels' rechargeable battery. Its nominal power
output is 5W. The solar panel comes complete with all
accessories, except for the fixing rod.
295 mm
11 ½”
193 mm
7 ½”
50 mm
2“
10
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ELECTRICAL CABLE CONNECTIONS
5.0 ELECTRICAL CABLE CONNECTIONS
5.1 Converter Electrical Cable Connections
CAUTION - Always disconnect the power cord before attempting any electrical connections.
All electrical cables enter the converter through compression fittings located on the side of the converter. Ensure
that all compression glands are properly tightened and all unused fittings are plugged so the case remains sealed.
External power
12-60VDC or 100-
240VAC connector
block battery socket
M1 connector block
IF2X socket
11
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ELECTRICAL CABLE CONNECTIONS
The sensor, hardwired inputs and outputs are connected to the converter through three terminal blocks. To locate
the terminal blocks, remove the rear cover plate, removing the four screws (5mm Allen key). When the cover is
removed, the terminal blocks are visible as shown in the figure above. These terminal are the converter hardwire
connections to the external equipment, including the sensor.
The following figure shows numbering and the respective connecting of the sensor cables, and input/outputs.
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17
SH C1 C2
SH B1 B2 E
OUT1 OUT2 OUT1 OUT2
COILS OPEN COMMON
COLLECTOR
12
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ELECTRICAL CABLE CONNECTIONS
5.3 FPI Mag 395L Forward Only Cable Connections
Black Ground
To Chassis Lug Terminal Block Assignments
Right Sensing
A #1 (E1) Blue
electrodes
CONVERTER
Left Sensing
1 2 3 4 11 12 13
8/16
A #2 (E2) Pink
Electrodes
ML255
CHASSIS | BLACK
RED
YELLOW
BLUE
PINK
BLACK
GREEN/YELLOW
GREEN/YELLOW
Reference
`
*IMPORTANT: See Section 5.6.1, "Converter Grounding
i For FPI Mag 395L and Ultra Mag Body Style 2", Figure 15
for instructions on attaching the chassis ground wire to
the converter ground lug.
Cable Diameters:
Cable A (15035): 0.248”
Cable B (15036): 0.248”
FPI Sensor
Sensing
#1 Blue
CONVERTER
electrode
1 2 3 4 11 12 13 Sensing
Reference
WHITE
BLUE
BLACK
BLACK
RED
YELLOW
#3 Black
ground
Magnet shield
#11 Black / overall cable
shield
#12 Red Coil
#13 Yellow Coil
SPI Sensor
Cable Diameter:
Single Cable (36002): 0.250”
14
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ELECTRICAL CABLE CONNECTIONS
5.5 Ultra Mag Sensor Cable
5.5.1 Ultra Mag Body Style 1 - 2", 3" And 14+"
Black Ground
To Chassis Lug
Terminal Block Assignments
Right Sensing
CONVERTER
A #1 (E1) Blue
electrodes
1 2 3 4 11 12 13
8/16
Left Sensing
ML255
A #2 (E2) Pink
BLUE
PINK
BLACK JUMPER
GREEN/YELLOW
CHASSIS | BLACK
GREEN/YELLOW
RED
YELLOW
Electrodes
`
*IMPORTANT: See Section 5.6.2, "Converter Grounding
Shrink Tube
CABLE A - ELECTRODES
CABLE B - COILS
Cable Diameters:
Cable A (15035): 0.248”
Cable B (15036): 0.248”
Ultra Mag Body Style 1 -
2", 3" and 14+"
Cable Jacket
Figure 13. Ultra Mag Body Style 1
Sensor Cable Connections
15
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ELECTRICAL CABLE CONNECTIONS
5.5.2 Ultra Mag Body Style 2 - 4" through 12"
Black Ground
To Chassis Lug
1 2 3 4 11 12 13
8/16
Right Sensing
A #1 (E1) Blue
ML255
electrodes
CHASSIS | BLACK
RED
YELLOW
BLUE
PINK
BLACK
GREEN/YELLOW
GREEN/YELLOW
`
*IMPORTANT: See Section 5.6.1, "Converter Grounding
i For FPI Mag 395L and Ultra Mag Body Style 2", Figure 15
for instructions on attaching the chassis ground wire to
the converter ground lug.
Shrink Tube
CABLE A - ELECTRODES
Cable Diameters:
CABLE B - COILS
Black Ground
Cable B Wire With Loop
From Cable B
Black Ground
Cable B Wire With Loop
From Cable B
With Lead To
Terminal #3
Figure 16. Converter Grounding With Cable Lug For FPI Mag
395L and Ultra Mag Body Style 1
17
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ELECTRICAL CABLE CONNECTIONS
5.7 Opto-Isolated Pulse Output Hook-Up
Both outputs are open drain outputs used to communicate with external devices.
• Maximum switching voltage: 40 VDC
• Maximum switching current: 100mA
• Maximum Ron = 70Ω
• Maximum switching frequency (load RL=240Ω, VOUT=24VDC): 50Hz
• Isolation: 500 V
i
IMPORTANT
Outputs are not isolated from each other. All outputs MUST use the same power source.
OUT 1
15
OUT 2
16
COMMON EMITTER
17
Input example:
Push button
switch
10K +
5 3-40 VDC
24V
-
6
18
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ELECTRICAL CABLE CONNECTIONS
5.9 Converter Power Hook-Up
!
WARNING!
Hazardous supply voltage can shock, burn, or cause death.
The power supply line must be equipped with external surge protection for current overload (fuse or circuit breaker
with limiting capacity not greater than 10A). It must be easily accessible for the operator and clearly identified.
Power connection is made using the power terminal block on the upper right side of the terminal board.
NOTE: The terminal block uplugs from the circuit board for easy connection. Connect earth ground to the
protective grounding terminal before making other connections. The power supply of this converter is 100 - 240
VAC, 44 - 66 Hz at maximum 4 W or 12-60 VDC.
Neutral Wire
Typically A
White Wire Negative Wire
Figure 19. AC Power Supply Terminal Block Figure 20. DC Power Supply Terminal
Block
19
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
CONVERTER START-UP - ALL SENSORS
6.0 CONVERTER START-UP - ALL SENSORS
Before starting up the converter please verify the following:
• Power supply voltage must correspond to that specified on the data plate (located on the side of the converter)
• Electric connections must be wired as described in this manual
• Ground connections must be properly installed
When the converter is powered it initiates a verification cycle of the converter. During the verification cycle the
converter displays an incrementing diagnostic number. When the diagnostic is complete, if an error is found,
an error number will be displayed referencing the chart at the back of this manual. A text message will also be
displayed on the alarm screen.
NOTE: To view alarms, press the RIGHT arrow key from the main display screen.
Key: Function:
UP/DOWN KEY (for moving cursor up or down)
SHORT PRESSING (< 1 SECOND):
Moves the cursor up to the previous subject on the menu
Increases the numeric figure of the parameter highlighted by the cursor
Enter/Esc
Key LONG PRESSING (> 1 SECOND):
Confirms the selected function
Leaves the current menu
Figure 21. Converter Key Conventions
20
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
CONVERTER START-UP - ALL SENSORS
6.2 Front Panel Visualization Pages
"Visualization pages" are different display screens that present specific information.
If there is screen is in "screen saver" mode, long-press the Left/Right arrow key to wake it up. Short-press the Left/
Right arrow key to cycle through the different visualization pages.
21
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
CONVERTER START-UP - ALL SENSORS
6.3 Factory Set Key Code
The converter is delivered with key code L2 = 000002, and with the “Quick start menu” enabled. Press the Enter/
Esc key. The “Quick start menu” is enabled from the section: Menu “8-Display”, section 10.8.
i ATTENTION!
It is very important to record any customized code as it CANNOT be retrieved if it is lost! The converter will
need to be returned to be reset to factory settings.
!2.UGPUQT!
PF>oo! 11467 1.1 Insert ND of sensor (0-300
MB>! ,13/1111 1.2 Calibration data of sensor v
1.3 Sensors model: Enter the f
Kpu/rqukvkqp>! 1 1.4 Position for insertion senso
Mk>! 2/1111
Mr>! 2/1111 1.5 Factory parameter
M\>! ,111111 1.6 Automatic setting accordin
ME>! 2/1111 1.7 Automatic setting accordin
1.8 Enables the empty pipe de
VE2>ou! 121 1.9 Value of empty pipe sensib
VE3>ou! 112 1.10 Perfom a SENSOR TEST ( s
G/R/fgvgev>! QP 1.11*Enables the automatic zero
G/r/vjt>! 181 1.12 Reset the preceding value
\gtq!ecn/
\gtq!tgu/
!7.QWVRWVU
Qwv2>! RNU, 6.1* Output 1
6.5* Output 2
Qwv3>! QHH 6.6 Power supply of pressure pr
!22.KPVGTPBN!FBVB
N3!eqfg>! ++++++ 11.1 Level 2 access code enter
11.2 Load factory data pre-set
Nqcf!hcev/fcvc 11.3 Ks Coefficient
U0p>! 182341
MV>! ,2/1111
MT>! ,2/1111
MU>! ,2/1111
24
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
PROGRAMMING EXAMPLE
7.0 PROGRAMMING EXAMPLE
The steps below demonstrate how to modify the full scale value from 3500 Gal/m to 3000 Gal/m on the “Quick Start
Menu”. You can enter the "Quick Start Menu" from any Visualization Page (see page 21 for information on this).
Note: The function used in the example is the top menu You will start with a blank screen.
item, so you will not need to scroll through the Quick 1. Long press on Enter/ESC.
Start menu. However, to do so, you need to have the
full menu item highlighted and press Left/Right or Up/
Down.
This will wake up
the front panel
..... ..... ..... ..... .....
.....
display.
TER TER
EN EN
ESC ESC
TER TER
EN EN
ESC ESC
TER TER
EN EN
ESC ESC
25
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MENU 0 - QUICK START
8.0 MENU 0 - QUICK START
The Quick Start menu gives you quick access to frequently used functions. It
consists of the same functions from Menu 2 - Scales.
See page 28 for a complete description of accessing and making
programming changes to these menu items.
NOTE: The Quick Start menu is configured at the factory. Some menus may not be available in your
configuration.
If the quick start menu is not present as the first menu option, then it has been disabled.
(See Section 9.8 to enable
the Quick Start menu.)
9.0 MAIN MENU DESCRIPTIONS
Press the key Enter/Esc key to go to the Main menu directly when the “Quick start
menu” is disabled.
When it is not
disabled you can select the Main menu from the “Quick start menu”. The functions
in the Quick Start menu and the
Main menu are explained below. Please note that some functions are only displayed if other functions are enabled
or with the insertion of additional modules.
9.1 Menu 1 - Sensor
ND Inside Pipe Diameter
The inner diameter of the pipe entered in millimeters. NOTE: The flow accuracy
will only be as good as the accuracy of the actual pipe ID entered into the
converter. The “Nominal Pipe Size” often referred to can be different from the
actual pipe ID and if used cause a substantial error in flow reading. Setting
the ND value to zero will cause the converter to display the velocity instead of
flow rate. Totalizer will then increment in feet or meters.
KA
Factory calibrated gain for the forward flow. Do not change the value.
Ins. Position
This is for a single point sensor. Position for insertion sensors: 0=1/8DN, 1=1/2DN, 2=7/8DN
Ki
Automatic setting according to ID (insertion meter only).
Kp
Automatic setting according to ID (insertion meter only).
KZ
This is the Zero Calibration constant. The KZ updates every time the Zero Calibration is performed.
KC
This is a coil current adjustment factor for use between sensor families.
26
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
TC1
This is the time needed to stabilize coil current when switching from no current.
TC2
This is the time needed to stabilize coil current when switching coil polarity.
E.P.detect
Enables the empty pipe detection function.
E.P.thr.
Enables the empty pipe detection threshold.
Zero cal.
Zero point calibration function for the forward flow. Enables the automatic zero calibration system.
To perform the zero point calibration:
1. Select the Zero cal. menu and press will ask if you want
the Enter/Esc key. It to execute.
2. Long press the Enter/ESC
button and release when the message
"Measuring. . . " appears.
The converter counts up from zero to 700,
after which the zero point is set.
Zero
res.
This resets the Zero Calibration.
27
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
9.2 Menu 2 - Scales
Fs
The full scale value measure set for flowrate. U.S. Standard and metric units are
selectable from this menu. See Appendix 1.0 for available units of measure.
To change the full scale value:
1. Highlight the Fs menu and press the Enter/Esc
key. 2. Press the Up/Down key.
The unit will
This will scroll
highlight.
through the different
available units. See
Appendix 1.0 for
available units of
measure. Select the
TER
desired unit.
E N TER
C
E N
E S
ESC
3. Press the Right/Left key twice.
4. Press the Up/Down
key.
This will highlight
This will scroll
the lower case letter
through the
that represents the
available time units.
time unit.
TER TER
E N
C E N
E S ESC
5. Press the Right/Left key.
6. Press the Enter/Esc
key.
This will select the
This will save the
numeric value.
selection.
Pressing the Up/
Down key to set the
digit and Right/Left
key to move to the
next digit. TER
TER
E N
C N
E
S
E
ESC
If a unit you are
looking for is not in
the current list press the Right/
Left key and scroll to the “/” between the unit of measure and time unit
selections and press the Up/Down key to switch between U.S. Standard
and Metric units.
Once the desired value is entered quick push the Enter/Esc key to save the changes and then long push Enter/
Esc to exit back to the display.
28
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
Temp.u.meas
This sets the unit of measure of temperature.
Tot1MU
The totalizer unit/multiplier as well as the visible decimal place is set here.
To change the totalizer unit/multiplier:
1. Highlight the menu and press the Enter/Esc key. 2. Press the Up/Down
key.
This will highlight This will scroll
the unit/multiplier.
through the different
See Appendix 1.0 for available units.
available units of
NOTE: The totalizer
measure.
multiplier is built
into the unit of
TER
measure, so for
T ER
EN C
gallons multiplied by N
E
ES
ESC
1000, select Kgal.
3. Press the Right/Left key twice. 4. Press the Up/Down key.
This will highlight
This will change the
the numeric value to
decimal resolution
the right.
displayed for the
totalizers. Changing
the decimal
resolution will
R not change the
TE TER
EN C multiplier. The
EN C
ES ES
available selections
are 00001, 001.0,
01.00, and 1.000.
If the desired unit of measure
is not in the current
list, press the Right/Left key
and scroll to the blank
space
between the
unit/multiplier and the numeric
decimal resolution selection
and press the
Up/Down key to switch between
U.S. Standard and Metricunits.
29
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
Pls1
The pulse increment value and unit of measure for
the pulse output 1. This option is only available when
“out1” in Menu 6 - Outputs is set to PLS+, PLS-, or PLS.
The pulse increment can be set to 0.00001 to 99999.9.
Tpls1
Duration of the pulse output 1 expressed in milliseconds.
This option is only available when "out1" in Menu 6 -
Outputs is set to PLS+, PLS-, or PLS. The pulse duration
can be set from 8.0 to 7999.9.
IMPORTANT
i The converter cannot detect the type of device it is connected to so it is up to the user to verify the setting
is compatible with the external device receiving the pulse. Incorrect settings can damage the receiving
device. See section 10.6 Menu 6 - Outputs for output specifications.
30
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
9.3 Menu 3 - Measure
Cut-off
Cutoff point which all flow is reported as zero. This value is set as a percentage
of the full scale. (Note: Default cut-off point = 3.0% of full scale.)
Prof
Measurement Profile
0 = Continuous Power
1 = Smart
2 = Average
3 = Max Life
Continuous Power uses more battery power and is not recommended. Max Life uses the least battery power and
is the default.
Damping
Measure Filter Damping Profile Enable is only valid in the Smart Profile setting. This applies a fixed filter to the
reading.
Al.max-
Maximum Reverse flow rate exceeded. Can be 0-125% of Full Scale
Al.min-
Minimum Reverse flow rate exceeded. Can be 0-125% of Full Scale.
Hyst.
Alarm Hysteresis. Set 0-25%. This sets the lag in response based on a percentage of the full scale. Example if the
alarm triggers at 100% and the hysteresis is set to 2% then once triggered the current rate must change beyond
2% to exit out of the current alarm state. This setting applies to all alarms.
31
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
9.5 Menu 5 - Inputs
T+ reset
Positive Totalizer Reset Enable. Set by turning on or off. This allows for the
positive total totalizer to be reset through the input.
T- reset
Negative Totalizer Reset Enable. Set by turning on or off. This allows for the negative total totalizer to be reset
through the input.
Count lock
Totalizer Count Lock Input Enable, Set by turning on or off. This allows for the totalizers to be locked (frozen) when
the input is active.
Alarm
Allows the input to be wired to an external alarm output such as a water intrusion alarm.
Wake-up
Wake up the display with an input.
Out 2
Transistor output channel 2. See tables on the next page for available settings.
OFF DISABLED
PLS PULSE FOR POSITIVE AND NEGATIVE FLOW VOLUME
PLS- PULSE FOR NEGATIVE FLOW VOLUME
PLS+ PULSE FOR POSITIVE FLOW VOLUME
F.SIGN FLOW DIRECTION OUTPUT (ENERGISED = -)
DIRECT. DR. DIRECT OUTPUT DRIVE
ALL ALARMS ALL ALARMS
OVR.RANGE. OUT OF RANGE ALARM OUTPUT (ENERGISED = FLOW RATE OK)
HARDW.AL. CUMULATIVE ALARM OUT interrupt coils, empty pipe, meas. error (ENERG. = NO ALARMS)
P.EMPTY EMPTY PIPE ALARM OUTPUT (ENERGISED = FULL PIPE)
MX+MN ALL MAX AND MIN ALARM OUTPUT (ENERGISED=AL. OFF)
*MX+MN TEMP MAX AND MIN TEMPERATURE OUTPUT (ENERGISED= AL. OFF)
*MIN TEMP MIN TEMPERATURE OUTPUT (ENERGISED= AL. OFF)
*MAX TEMP MAX TEMPERATURE OUTPUT (ENERGISED= AL. OFF)
*MX+MN P12 MAX AND MIN PRESSURE 1 AND 2 OUTPUT (ENERGISED= AL. OFF)
*MX+MN DP MAX AND MIN DIFFERENTIAL PRESSURE OUTPUT (ENERGISED= AL. OFF)
*MIN DP MIN DIFFERENTIAL PRESSURE OUTPUT (ENERGISED= AL. OFF)
*MAX DP MAX DIFFERENTIAL PRESSURE OUTPUT (ENERGISED= AL. OFF)
32
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
*MX+MN P2 MAX AND MIN PRESSURE 2 OUTPUT (ENERGISED= AL. OFF)
*MIN P2 MIN PRESSURE 2 OUTPUT (ENERGISED= AL. OFF)
*MAX P2 MAX PRESSURE 2 OUTPUT (ENERGISED= AL. OFF)
*MX+MN P1 MAX AND MIN PRESSURE 1 OUTPUT (ENERGISED= AL. OFF)
*MIN P1 MIN PRESSURE 1 OUTPUT (ENERGISED= AL. OFF)
*MAX P1 MAX PRESSURE 1 OUTPUT (ENERGISED= AL. OFF)
MX+MN Q MAX AND MIN FLOW RATE ALARM OUTPUT (ENERGISED=AL. OFF)
MX+MN Q- MAX AND MIN NEGATIVE FLOW RATE ALARM OUTPUT (ENERGISED=AL. OFF)
MIN Q- MIN NEGATIVE FLOW RATE ALARM OUTPUT (ENERGISED=AL. OFF)
MAX Q- MAX NEGATIVE FLOW RATE ALARM OUTPUT (ENERGISED=AL. OFF)
MX+MN Q+ MAX AND MIN POSITIVE FLOW RATE ALARM OUTPUT (ENERGISED=AL. OFF)
MIN Q+ MIN POSITIVE FLOW RATE ALARM OUTPUT (ENERGISED=AL. OFF)
MAX Q+ MAX POSITIVE FLOW RATE ALARM OUTPUT (ENERGISED=AL. OFF)
33
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
9.7 Menu 7 - Communication
IF2 pr.
Protocol for IF2 port. Set to DPP or HTP. This sets the protocol used for
communication to the IF2 device, either Data Packet Protocol (DPP) or Hyper
Text Protocol (HTP). Default is DPP. This should not be changed.
34
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
9.9 Menu 9 - Data Logger
Note: See the McLogger manual, literature no. 30121-87, for software details
and downloading data.
YYYY/MM/DD
Date and time. This sets the date and time in the converter. The format for
entering the date and time is year / month / day and time is hours : minutes.
T.zone=h
Time Zone. ±12 hours. GMT ± the number of hours. Pacific time is -8.
Acquisition
Enable Data Logging Function.
Comp.mode
Format the Data Logging Data to the previous version of this converter. Default=Off
int.1
Data Logging Interval. HH:MM:SS
Log T+
Include the T Positive Totalizer for Data Logging.
Log P+
Include the P Positive Totalizer for Data Logging.
Lot T-
Include the T Negative Totalizer for Data Logging.
Log P-
Include the P Negative Totalizer for Data Logging.
Log NT
Include the Net T Totalizer for Data Logging.
Log NP
Include the Net P Totalizer for Data Logging.
Log Q
Include the Flow Rate for Data Logging.
Log STAT
Include Statistics like the board temperature and cycles/h for Data Logging.
M.units
Include the Units of Measure for Data Logging.
% values
Include the Flow Rate % of full scale and optional analog inputs for Data Logging.
9.10 Menu 10 - Diagnostic
35
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
MAIN MENU DESCRIPTIONS
Self-test
Converter self test. Executed command. Running the self-test will cause the
converter to run an internal diagnostic test that will check for internal hardware
and software errors. Once the self-test is complete if any errors are found then
the error code for each error will be displayed. Refer to section 9.0 Alarm
Messages for a list of possible error codes.
Simulation
flow
Simulation enable. Setting this menu
to ON will generate an internal signal
that simulates
and allows
the outputs and all connected instruments to be tested. After simulation is set
to ON, the
flow can be set to a
percentage based on the current FS1
setting of -125% to 125%.
To enable the simulation function highlight
the Simulation menu and NOTE: you will
now
have an “S” in the
press the (enter/esc) key. Toggle the
simulation
function from OFF to upper
left
corner,
this
indicates the
ON using the Up/Down key. Press and
hold the
(enter/esc) key to exit
simulation
mode
is active.
back to the main menu and once again to exit to the visualization page.
the Enter/Esc key. This will bring
up the
Press
flow simulation set up screen. Use
the Right/Left
key and the Up/Down key to
enter in the flow rate percentage
value
for the simulation. Press the Enter/Esc
key to enter that value.
The
converter will
start to read flow. It may take a few seconds for
the readings
to
appear.
Repeat the above steps as needed to observe the different flow
rates
desired.
To exit out of simulation
mode, press enter to get into the simulation set
up screen (see above) and then press and hold the Enter/Esc key. This will
exit out to the visualization screen, and the “S” in the upper left corner of the
screen
will disappear.
Display data
Numeric display for various internal settings
and raw measured signals. This is a Factory service menu.
Stand-by
Use this function before powering down the converter. This will
save any unsaved data and turn off the converter.
Once the converter is in Stand-by, you can switch the batteries off or change the batteries. To turn the converter
back on you must switch the battery switch to off and after approximately 5 minutes, turn the switch to ON.
i CAUTION - if the passcode is changed from the default value and is lost, it CANNOT be recovered. In
the event the passcode is lost, the converter can be returned to the factory to be reset.
Load fact.data
This will configure the converter to the configuration when it left McCrometer.
S/n
Displays the board serial number.
KT
Used to adjust gain errors. This factor is generated by the converter at power up.
KS
Field adjustment coefficient. This value is a direct multiplier that is used as a field adjustment / correction coefficient.
37
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ALARM MESSAGES
10.0 ALARM MESSAGES
During meter setup, you may see error messages and codes. These messages and codes are explained below.
ALARM
VIA
CODE MESSAGE CAUSE ACTION SMS *
000 NO ALARMS ----- ----- NO
001 SYSTEM STARTUP The system was started ----- YES
003 PRESS.1>FS The Pressure at input 1 is Check the maximum full scale set YES
higher than the full scale set and the process conditions
004 FLOW>FS The flow rate is higher than Check the maximum full scale set YES
the full scale set and the process conditions
005 PULS.1>F.MAX The pulse frequency at Reduce the pulse duration if the YES
output 1 is too high connected counter allow this or
increase the pulse unit value
006 B3 LOW The battery B3 powering the Replace the battery YES
MODEM is exhausted
007 B2 LOW The battery B2 powering the Replace the battery YES
main board is exhausted
008 B1 LOW The battery B1 powering the Replace the battery YES
main board is exhausted
077 SYSTEM RESTART The system was restarted If this message appears after an YES
after a reset signal AUTO-TEST command, it is not an
error indication, otherwise check
all the connections, the batteries
and the grounding
094 ALARM INPUT ACT. The digital input detected an Check the process condition YES
alarm signal condition
103 B.TEMP.OUT R. The measured board Ensure that the instrument is YES
temperature is out of the operating within the specified
allowed range temperature conditions
113 FIRMW.FILE ERR The file containing the Re-send the firmware file YES
firmware was received with
errors
115 S.OUT OVERLOAD An overload condition is Check the connections and YES
present in the power source the external sensors (pressure,
that feeds the external temperatures, input circuitry)
sensors and inputs
117 ERR.SENS.TEMP The external temperature Check the external temperature YES
sensor is not working sensor and its connections
properly
118 MAIL S.FAILED The last email sending Check the transmitting YES
process has failed for some conditions (antenna signals, GPRS
reasons parameters, server status and
configurations)
119 SMS NOT AUTH A SMS from a not authorized Check the events log for the YES
number was received sender number and do the
appropriate actions
*Not available at this time.
38
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ALARM MESSAGES
ALARM
VIA
CODE MESSAGE CAUSE ACTION SMS *
135 V.MODEM LOW The internal voltage of the Check the battery B3 and its YES
MODEM is not sufficient for to connections, verify if any visible
guarantee a proper operation damage has occurred to the
board or the MODEM
144 NO CMD RECEIVED The command list sent to the Check the command list and re- YES
instrument does not contain send it again
any executable commands
150 MAX Q+ The Positive flow rate is Check the maximum threshold YES
higher than the maximum positive flow rate alarm value and
threshold value set the process conditions
151 MIN Q+ The Positive flow rate is lower Check the minimum threshold YES
than the minimum threshold positive flow rate alarm value and
value set the process conditions
153 MAX Q- The Negative flow rate is Check the maximum threshold YES
higher than the maximum negative flow rate alarm value
threshold value set and the process conditions
154 MIN Q- The Negative flow rate is Check the minimum threshold YES
lower than the minimum negative flow rate alarm value
threshold value set and the process conditions
157 MAX P1 The pressure at input 1 is Check the maximum threshold YES
higher than the maximum pressure 1 alarm value and the
threshold value set process conditions
158 MIN P1 The pressure at input 1 is Check the minimum threshold YES
lower than the minimum pressure 1 alarm value and the
threshold value set process conditions
160 MAX P2 The pressure at input 2 is Check the maximum threshold YES
higher than the maximum pressure 2 alarm value and the
threshold value set process conditions
161 MIN P2 The pressure at input 2 is Check the minimum threshold YES
lower than the minimum pressure 2 alarm value and the
threshold value set process conditions
163 MAX DP The pressure difference Check the maximum threshold YES
(delta) P1-P2 is higher than pressure difference (delta) alarm
the maximum threshold set value and the process conditions
164 MIN DP The pressure difference Check the minimum threshold YES
(delta) P1-P2 is lower than the pressure difference (delta) alarm
minimum threshold set value and the process conditions
167 MAX TEMP The external temperature is Check the maximum threshold YES
higher than the maximum temperature alarm value and the
threshold value set process conditions
168 MIN TEMP The external temperature Check the minimum threshold YES
is lower than the minimum temperature alarm value and the
threshold value set process conditions
171 P.EMPTY The system detected an Check the empty pipe settings YES
empty pipe alarm condition and the process conditions
*Not available at this time.
39
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ALARM MESSAGES
ALARM
VIA
CODE MESSAGE CAUSE ACTION SMS *
193 CONFIG.ENTERED The system detects an Check the events log and do the YES
access to the configuration appropriate action depending on
parameters your data protection policy
194 CLOCK S.FAIL The clock synchronization Check the server configuration, YES
with the remote server failed the GPRS network conditions and
antenna signal
195 POWER FAILURE The system experienced a Check the status of the batteries YES
power failure (loss of power) and all their connections. THIS
MESSAGE MAY APPEAR WHEN
THE SYSTEM IS SWITCHED OFF
WITHOUT PRIOR PUTTING IT IN
STAND-BY MODE
206 F.SENSOR ERROR The system detected one or Check the error code in the YES
more errors in the flow sensor events log
207 TEMP.>FS The external temperature is Check the maximum full scale set YES
higher than the full scale set and the process conditions
208 PULS.2>F.MAX The pulse frequency at Reduce the pulse duration if the YES
output 2 is too high connected counter allow this or
increase the pulse unit value
209 PRESS.2>FS The Pressure at input 2 is Check the maximum full scale set YES
higher than the full scale set and the process conditions
210 EL.SIG.ERROR The system detected an error Check the flow sensor cables, the YES
condition on the measuring electrodes surface, the grounding
electrodes and the process condition
211 EXCIT.ERROR The system detected an error Check the flow sensor cables, the YES
in the coils excitation circuit coils insulation and resistance
212 SD MEMORY FULL The SD card memory is full Change the SD card memory or YES
(no more space for new data) erase all the unwanted data
215 POWER SUPPLY OFF The power coming from the Check the power supply source YES
universal power supply went and the relative cables
off
216 SD MEMORY ERROR The SD card installed is Replace the SD memory, format it YES
missing, not valid, not or check if it is properly inserted
formatted or incompatible
with the system
40
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
ALARM MESSAGES
CODE ANOMALIES DESCRIPTION ACTION
0001 Keyboard error (key stuck detected) ADDRESS TO SERVICE
0002 Hardware parameters stored in F-RAM not valid
0004 Sensor parameters stored in F-RAM not valid
0008 Converter parameters stored in F-RAM not valid
0200 Coils excitation current driver error
0400 Measure input stage error
1000 Internal time reference error
0010 Sensor coils resistance out of tolerance Check the flow sensor status, its working conditions
0020 Time TC2 out of tolerance and all the connections with the converter and
with the reference ground
0040 Time TC1 out of tolerance
0080 Sensors coils insulation out of tolerance
0100 Excitation drive phases difference error
0800 Coils excitation circuit interrupted
2000 Temperature reference error Check the connections and the external
temperature sensor
4000 External sensors power source overload Check all the connections between the external
sensors (pressure, temperature, digital inputs) and
the converters
8000 SD memory error Check the SD card placement and verify if a
compatible type is used
41
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
SPECIFICATIONS
11.0 SPECIFICATIONS
BATTERY POWER:
Four D-cell lithium batteries ELECTRICAL ACCESS:
Estimated life of batteries: 3-5 years, standard use Converter: Compression gland seals for 0.125” to 0.375”
Diameter round cable.
STANDARD OUTPUTS:
• Volumetric Pulse Output GALVANIC ISOLATION:
• Hardware Alarm
All inputs / outputs are galvanically isolated from power
• High/Low Flow Alarms
• Empty Pipe supply up to 500V
• Directional Indication
• Range Indication CONVERTER ENCLOSURE:
IP67 Die Cast Aluminum
Maximum switching voltage: 40 VDC
Maximum switching current: 100mA 5.76” H x 5.76” W x 6.73” D
Maximum switching frequency: 50 Hz (14.6 cm. H x 14.6 cm. W x 17 cm D)
Insulation from other secondary circuits: 500V
ENVIRONMENTAL:
OPTIONAL OUTPUTS: Pressure / Temperature Limits:
• Smart Output (Sensus or Itron) Electronics: Operating and storage temperature:
-4˚ to 140˚ F (-20˚ to 60˚ C)
ENGINEERING UNITS: Sensor is continuously submersible (IP68), with a
cm3 Cubic centimetre IGL Imperial Gallon standard quick-connect cable to 6 ft., and optional
ml Millilitre ft3 Cubic foot strain relief at 30 ft.
l Litre bbl Standard barrel
dm3 Cubic decimetre BBL Oil barrel KEYPAD AND DISPLAY:
dal Decalitre IKG Imperial Kgallons Can be used to access and change set-up parameters
hl Hectolitre KGL US Kgallons using three membrane keys and an LCD display
m3 Cubic metre Aft Acre feet
in3 Cubic inch MGL USA Mega Gallons
COMPATIBLE PRODUCTS:
Gal American gallon IMG Imperial Mega Gallon
The battery powered converter works with all UltraMag
STANDARDS: meters, Full Profile Insertion (FPI) meters (except the
394), and Single Point Insertion (SPI) meters. This
IP67 Die Cast Aluminum Converter
converter does not work with MC Mag 3000 sensors.
*NOTE REGARDING CABLE LENGTH: McCrometer recommends minimizing cable length. Electromagnetic flow
meters may have unfavorable signal strength to noise ratio in electrically noisy environments. Longer lengths of
cable increase the likelihood of interference.
ELECTRICAL CHARACTERISTICS:
Power supply version Power supply Power supply frequency Max power
12V - 60V * //
AU 300mW
100V - 240V~* 44-66Hz
LITHIUM BATTERY NOT
3.6v-19aH // 30mW
RECHARGEABLE
* Voltage must not exceed ±10% of the above limits.
42
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
SPECIFICATIONS
43
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
TROUBLESHOOTING GUIDE
Hemet, CA 92545
44
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
APPENDIX 1.0 UNITS OF MEASURE
Rate Of Flow Report Is Not As • Confirm the unit is programmed correctly by requesting a program
Expected setting report from the factory.
U.S. Standard
in3 Cubic Inches
Gal U.S. Gallons
IGL Imperial Gallons
ft3 Cubic Feet
bbl Standard Barrels
BBL Oil Barrels
KGL Kilo Gallons
IKG Imperial Kilo Gallons
Aft Acre Feet
MGL Mega Gallons
IMG Imperial Mega Gallons
Metric
ml Milliliters
cm3 Cubic Centimeters
l Liters
dm3 Cubic Decimeter
dal Decaliter
hl Hectoliter
m3 Cubic Meter
Time
s Seconds
m Minutes
45
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
APPENDIX 2.0 CONVERSION TABLES
h Hours
d Days
Fraction Decimal
1/8 .125
1/4 .25
3/8 .375
1/2 .5
5/8 .625
3/4 .75
7/8 .875
Table of Conversions
Multiply By To Get
Centimeters 0.3937 Inches
Centimeters 0.03281 Feet
Inches 25.4 Millimeters
Feet 30.48 Centimeters
Square Feet 144.0 Square Inches
Square Inches 0.006944 Square Feet
Cubic Inches 0.0005787 Cubic Feet
Cubic Feet 7.481 Gallons
Cubic Feet 1728.0 Cubic Inches
Cubic Feet 0.02832 Cubic Meters
Cubic Feet 28.32 Liters
Cubic Meters 35.31 Cubic Feet
Cubic Meters 264.2 Gallons
US Gallons 3.785 Liters
US Gallons 0.1337 Cubic Feet
US Gallons 0.003785 Cubic Meters
46
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
WARRANTY STATEMENT
US Gallons .8326748 Imperial Gallons
Liters 0.2642 Gallons
˚F = ( ˚C x 9/5 ) + 32 ˚C ( ˚F - 32 ) x 5/9
WARRANTY STATEMENT
Manufacturer warrants all products of its manufacture to be free from defects in workmanship and material under
normal use and service. The warranty for the ML255 Converter extends for a period of twelve (12) months after
date of shipment (if the converter is purchased alone), unless altered by mutual agreement between the purchaser
and manufacturer prior to the shipment of the product. If the ML255 converter is purchased with a flow sensor
system, the Warranty Statement for the flow sensor shall apply. Neither this warranty, nor the flow sensor system
warranty, covers consumable items, such as batteries or solar panels.
If this product is believed to be defective and is within its warranty period, purchaser shall notify the manufacturer,
and will return the product to the manufacturer, postage paid, within twelve (12) months after date of shipment
by the manufacturer. If the purchaser believes the return of the product to be impractical, manufacturer shall
have the option, but will not be required, to inspect the product wherever located. In any event, if the purchaser
requests the manufacturer visit their location, the purchaser agrees to pay the non-warranty expenses of travel,
lodging and subsistence for the field service response. If the product is found by the manufacturer’s inspection
to be defective in workmanship or material, the defective part or parts will either be repaired or replaced, at
manufacturer’s election, free of charge, and if necessary the product will be returned to purchaser, transportation
prepaid to any point in the United States. If inspection by the manufacturer of such product does not disclose any
defect of workmanship or material, manufacturer’s regular service repair charges will apply.
THE FOREGOING WARRANTY IS MANUFACTURER’S SOLE WARRANTY, AND ALL OTHER WARRANTIES,
EXPRESS, IMPLIED OR STATUTORY, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE, ARE NEGATED AND EXCLUDED. THE FOREGOING WARRANTY IS IN
LIEU OF ALL OTHER WARRANTIES, GUARANTEES, REPRESENTATIONS, OBLIGATIONS OR LIABILITIES ON THE
PART OF THE MANUFACTURER.
Purchaser’s sole remedy and manufacturer’s sole obligation for alleged product failure, whether under warranty
claim or otherwise, shall be the aforestated obligation of manufacturer to repair or replace products returned
within twelve months after date of original shipment. The manufacturer shall not be liable for, and the purchaser
assumes and agrees to indemnify and save harmless the manufacturer in respect to, any loss or damage that may
arise through the use by the purchaser of any of the manufacturer’s products.
47
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46 Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.
OTHER McCROMETER PRODUCTS INCLUDE:
Propeller Flowmeters
Propeller Flowmeters
Magnetic Flowmeters
DURA MAG ™
Magnetic Flowmeters
Magnetic Flowmeters
Magnetic Flowmeters
Magnetic Flowmeters
Represented By:
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078 Lit. # 30122-46, Rev. 1.2 / 5-31-18
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
www.mccrometer.com for current technical data and instructions.