T316
T316
T316
Background
The rotational viscometer was adopted for asphalt testing as a result of the Strategic Highway
Research Program (SHRP) in 1992. A rotational viscometer is used to measure the viscosity of
asphalt binder by shearing the material at a constant rate of strain and at a constant
temperature.
The sample chamber, sample holder, and spindle must be preheated to the test temperature. A
small quantity of asphalt binder, usually less than 10.5 g, must be heated until it can be poured.
The asphalt is poured into the sample chamber, the sample chamber is loaded into the
thermosel, and the spindle is lowered into the sample. The thermosel must recover to test
temperature within 30 minutes. The sample must be maintained at test temperature for a
minimum of 10 minutes. The spindle rotations are started during this 10-minute equilibration.
After equilibration, three readings are obtained from the machine at 1-minute intervals.
Calculations: 1 minute
The three testing results are obtained from a direct digital read out on most models and then
averaged.
31
Module F Asphalt Binder PG Tests
Apparatus
Rotational Viscometer – (Figure 13) Having a torque transducer and thermosel capable of
measuring torque and maintaining test temperatures to 1.0°C.
Oven – An oven for preheating the sample, chamber, and spindle to test temperature.
Balance – A balance for determining the mass of the asphalt binder specimen.
Sample Preparation
The rotational viscometer and temperature controller should be turned on and set to test
temperature (typically 135°C) so the thermosel can preheat. The sample holder, chamber, and
spindle should also be preheated to test temperature. Commonly, an oven is used to preheat
each item; however, the chamber and spindle can be brought to test temperature in the
thermosel. The levelness of the both the viscometer and thermosel should be verified prior to
the start of the test. Both are commonly equipped with a bubble level.
Note: Often, preheating the sample holder is overlooked. Preheating the sample holder
allows it to act as a heat sink while the sample is weighed, which prevents cooling and
reduces the time it takes for the sample chamber and thermosel to recover to test
temperature.
32
Module F Asphalt Binder PG Tests
Procedure
Step 1
Place the sample chamber into the sample holder and tare them both on a balance. Pour the
required amount of asphalt binder (Figure 14) into the sample chamber. The sample mass is
based on the size of the spindle being used for testing. The manufacturer’s instructions give the
sample mass for each spindle.
Step 2
Insert the sample chamber into the thermosel (Figure 15) and lower the spindle from the
viscometer into the sample. The top conical portion of the spindle should be completely
immersed in the asphalt. The thermosel, sample chamber, and asphalt binder sample must
recover to test temperature within 30 minutes, and must be kept at test temperature for
equilibration for a minimum of 10 minutes.
33
Module F Asphalt Binder PG Tests
Step 3
During the 10-minute equilibration, the viscometer should be set to 20 rpm and the motor
should be turned on.
Step 4
After the 10-minute equilibration and when the viscosity readings have stabilized, take three
viscosity measurements at 1-minute intervals.
Note: The viscosity of the material is reported as torque. The binder adhering to the sides of
the chamber is stationary and the binder adhering to the sides of the spindle is revolving. As
these layers of binder move concentric to each other, the bonds between the molecules are
broken and regained continuously. The viscometer measures this bond energy as resistance
to shear, the definition of viscosity, and reports it as torque.
Step 5
For accuracy, the manufacturer provides an operable torque range, usually between 2 and 98%.
It may be necessary to increase the RPMs or restart the test with a different spindle size to
achieve this range.
Calculations
Take the average of the three readings. Most machines give the torque value in centipoise (cP);
however, the viscosity of asphalt is generally reported in Pascal seconds (Pa·s). To convert from
cP to Pa·s, move the decimal three places to the left.
The test results are reported as the average Pascal seconds (Pa·s) and the test temperature
should be recorded to the nearest 0.1°C.
When determining the mixing and compaction temperatures for an asphalt binder sample, the
test is conducted at 135°C and 165°C. These two points are plotted on a graph having viscosity
in Pascal seconds on the logarithmic Y axis and Temperature in °C on the standard X axis (Figure
16: ). The mixing range and compaction temperature ranges can then be established from the
plotted line by observing the temperature at which the asphalt binder has the correct viscosity
range for mixing and compaction requirements.
34
Module F Asphalt Binder PG Tests
Common Errors
Data Sheet
35