Olympus AU2700

Fully automated, random access chemistry system with STAT


Analytical System
capability, Urine and homogenous Immuno Assays.
Analytical Methods Endpoint, kinetic rate, fixed rate, optionally ISE
Simultaneously Processed Analytes 48 photometric tests + 3 ISE
Throughput 1,600 photometric test/hour; 2,700 samples/hour with ISE
Racks with 10 samples each (barcodes on primary tubes and on
Sample Feeder
racks); capacity of 300 samples; continuous loading
Primary and Secondary tubes; diameter from 11.5 to 16 mm;
Sample Tubes
height from 55 to 102 mm
22 user-definable for STAT, QC, Cal, routines, enclosed and
STAT Samples refrigerated. Auto repeat run capability. Bar-coded primary
tubes.
Sample Volume 1-25 µl in 0.1 µl increments [1 µl for repeats]
Reagent Volume R1: 15-250 µl; R2: 15-250 µl [1 µl increments]
R1 = 48, R2 = 48; holds 15, 30, 60, 120ml bottles; refrigerated
Reagent Supply & Storage
(4˚C-12˚C)
Total Reaction Volume 120-430 µl
Reaction Cuvette Nondisposable glass cuvette
Reaction Time Up to 8 min. + 20 sec.
Reaction Temperature 37˚C, dry bath
Mixing Method With rotating paddles after dispensing sample and reagent
Direct assay through the reaction cuvette (0-3.0 OD) mono and
Photometry System
bichromatic measurements possible
Wavelength 13 different wavelengths between 340 – 800 nm
Cuvette Cleaning Comprehensive cleaning with detergents
Calibration Auto calibration; cooled calibrator positions
Quality Control Auto QC, cooled QC positions
Individual and profile test requisition via on-line, mouse,
Test Requisition
function keys or touch screen
Sample Integrity Sample clot detection and probe crash protection
Online Full uni- and bidirectional communication possible
Software Windows 2000
Dimensions (W x H x D) mm ANL: 2,000 x 1,000 x 840
Power Supply Toshiba 1600XP 6.0kVA