Lumiray 1600: Auto Chemiluminescence Analyzer

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Lumiray 1600

Auto Chemiluminescence Analyzer

Advanced CLIA technology:


Enzymatic chemiluminescence principle
Glow-type luminescence
ALP-AMPPD system, and up to 10-21 mol/L high detection sensitivity
Independent magnetic separation 3-step washing system, cleanser
pre-heating function
Throughput: Constant 120 test/h for allergies test and autoimmune diseases tests
60 sample positions, 24 positions for regular reagent or 72 positions for
Allergy bottle with 24 hours refrigeration
Continuous cuvettes loading
Lumiray 1600
Auto Chemiluminescence Analyzer

Technical Specifications
Instrument parameters
Instrument types:
fully automatic tube type chemiluminescence detector
Emergency insert:
Priority given to STAT samples inserted for testing
Throughput: Constant 120 T/H
Detection principle:
Enzymatic chemiluminescence

Detection Method Reaction System


Reaction Mode: - Continuous disposable cuvettes
ALP-promoting Chemiluminescence with magnetic loading, 1000 cuvettes per box
particles as carriers - Automatic loading reaction
Testing Process: cuvettes with automatic alarm
Supports the one-step method, two-step method, - Direct heating for incubation
pseudo two-step method, etc.
Calibration Method:
Supports 2-point calibration, 6-point calibration, etc. Independent Magnetic Separation Unit
- 4-phase magnetic separation combined
Optical System with vortex type resuspension of magnetic
particles to ensure cleaning efficiency
- Constantly reaction temperature of 37±0.5℃ - Array-type surrounding magnetic field
- Measurement Range: 0~60M RLU design to precisely control the loose of
- Linear Range of Internal Monotonic Measurement: magnetic particles, achieve optimal
3000-30M RLU repeatability of magnetic separation
- Dark Count < 150RLU

Sample System Working Condition


- Continuous sample loading and - Power Supply: 100V-240V,
offloading by sample racks AC 50/60Hz, 1500W
- 10 positions per rack, up to 60 samples in 1 batch - Temperature: 10℃-30℃
- Sample volume: 5~100ul, with increment of 1ul - Humidity: Less than 70%
- Support original blood collection tube, centrifuge
tube, plastic test tube, small sample cup, etc.
- Built-in bar-code scanner Working Condition Dimension
1200mm(L)*790mm(W)*1135(H)
Reagent Assembly
- 24 reagent positions, continuous loading, Weight
real-time mixing for magnetic particles
- Reagent volume(R1, R2, M): 20~150ul, Net Weight: 250Kg, Gross Weight: 300Kg
with increment of 1ul
- 24 hours non-stop cooling system with 2~10℃
- Steel probe, liquid level detection, collision protection,
clotting detection and inventory checking

Rayto Life and Analytical Sciences Co., Ltd.


Address: Rayto Industrial Building, Shuangming
Blvd South, East, Hi-Tech Park, Guangming New
Rayto reserves the rights to change specifications without prior notice.
District, 518107 Shenzhen, P.R.China
P/N: ENG-Lumiray1600-20181109 E-mail: info@rayto.com Website: www.rayto.com