Mispa Nano Plus-1
Mispa Nano Plus-1
Mispa Nano Plus-1
System Function
Automatic, Discrete, Random Access, Bench-top, STAT sample priority
Throughput: Up to 360 photometric tests per hour
Measuring principles: Absorbance photometry, turbidimetry
Methodology: End-point, Fixed-time, Kinetic, Single/Double
reagent chemistries, Mono-chromatic / bi-chromatic
Reagent/Sample Handling
Reagent/Sample tray: 50 to 100 positions for reagents and 50 to 100 positions
for samples in 24-hour refrigerated compartment (2~12℃)
Reagent volume: R1: 100~200µL, step by 0.5µL; R2: 10~200µL, step by 0.5µL
Sample volume: 2~35µL, step by 0.1µL
Reagent/Sample probe: Liquid level detection, vertical collision protection,
inventory checking, reagent pre-warming, optional clot detection
Probe cleaning: Automatic washing for interior and exterior; Carry over < 0.05%
Automatic sample dilution: Pre-dilution and post-dilution
Mixing Unit: Independent mixing bar
Capable to communicate with LIS in bi-directional mode
Reaction System
Reaction tray: 80 reusable cuvettes
Reaction volume 100~360µL
Reaction temperature: 37℃± 0.1℃
Cuvette Washing: Washing station with pre-warmed detergent and de-ionized water
Optical System
Light Source: Halogen-tungsten lamp
Wavelength: 12 wavelengths, 340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm,
605nm, 660nm, 700nm, 740nm, 800nm
ADL/BR/Nano+/R00-03/19 Company reserves the right to change any design and technical features of the product at any time, If needed.
Absorption range: 0~3.5Abs, resolution 0.0001Abs
Control and Calibration
Calibration modes: K factor, Linear (two points and multipoints), Logit-Log 4P,
Logit-Log 5P, Spline, Exponential, Polynomial, Parabola,
Logit-Log3P, Broken line, One key calibrator import function
Control Rules: Westgard multi-rule, Levey-Jennings, Cumulative sum check, Twin plot
Operation Unit
Operation system: Windows 10
Interface: RS-232
Working Conditions
Power Supply: 200~240V, 50/60Hz, 1300VA or 100~130V, 60Hz, 1300VA
Dimension: 860 mm (length) × 660 mm (depth) × 550 mm (height)
Weight: 115 kg
Water Consumption: 6.5 L/hour
ISO 9001:2015
EN ISO 13485:2016 * Toll Free No: 1800 425 7151
“Agappe Hills”, Pattimattom (PO), Dist. Ernakulam, Kerala - 683 562, India.
TEL: + 91 484 2867000 | productcorp@agappe.in | www.agappe.com
Fully Automatic Clinical Chemistry Analyser Fully Automatic Clinical Chemistry Analyser