BC 20s Handbuch
BC 20s Handbuch
BC 20s Handbuch
Operator’s Manual
I
I
© 2012-2015 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issued Date is 2015-02.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel.
the electrical installation of the relevant room complies with the applicable national and
local requirements.
the product is used in accordance with the instructions for use.
I
result in machine breakdown or injury of human health.
Be sure to operate the analyzer under the situation specified in this manual;
otherwise, the analyzer will not work normally and the analysis results will
be unreliable, which would damage the analyzer components and cause
personal injury.
NOTE
This equipment must be operated by skilled/trained clinical professionals.
II
Warranty
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
III
Customer Service Department
Tel: 0049-40-2513175
Fax: 0049-40-255726
IV
Table of Contents
1
Table of Contents
12 Appendices ..........................................................................................................A-1
A Index ...................................................................................................................... A-1
B Specifications ........................................................................................................ B-1
C Communication ......................................................................................................C-1
2
1 Using This Manual
1.1 Introduction
This chapter explains how to use your Operator's Manual of BC-20s Auto Hematology
Analyzer which is shipped with your instrument and contains reference information about the
BC-20s and procedures for operating, troubleshooting and maintaining the instrument. Read
this manual carefully before operating your analyzer and operate your analyzer strictly as
instructed in this manual.
NOTE
Be sure to operate the analyzer strictly as instructed in this manual.
1-1
Using This Manual
1-2
Using This Manual
1-3
Using This Manual
Form Meaning
[××] all capital letters enclosed in [ ] indicate a key name
“××” bold letters included in " " indicate text you can find on the
screen of the BC-20s
×× italic letters indicate chapter titles, such as Chapter 1 Using This
Manual.
All illustrations in this manual are provided as examples only. They may not necessarily
reflect your BC-20s setup or data displayed.
1-4
Using This Manual
1-5
Using This Manual
1.6 Symbols
Symbols used in this manual:
You may find the following symbols on package or the body of the instrument:
BIOLOGICAL RISK
HIGH VOLTAGE
ALTERNATING CURRENT
1-6
Using This Manual
BATCH CODE
EXPIRATION DATE
SERIAL NUMBER
DATE OF MANUFACTURE
MANUFACTURER
1-7
Using This Manual
(1)
Biological risk.
(2)
Connect only to a properly earth grounded outlet.
To avoid electrical shock, disconnect power prior to maintenance.
To prevent fire, only use the fuse of specified type and rating.
1-8
Using This Manual
(1)
Warning: the sample probe is sharp and potentially biohazardous. Exercise caution when
working around it!
1-9
Using This Manual
1-10
Using This Manual
(1)
Warning: make sure that the protective cover is closed before operating the analyzer.
(2)
Warning
To avoid injury, do not place your hands anywhere near the the aspiration module
when the analyzer is working.
The sample probe is sharp and potentially biohazardous. Exercise caution when
working around it!
1-11
2 Understanding Your Analyzer
2.1 Introduction
The BC-20s is a hematology analyzer and 3-part counter for In Vitro Diagnostic Use in clinical
laboratories.
2-1
Understanding Your Analyzer
Product Composition
Auto Hematology Analyzer (hereinafter called BC-20s) system consists of the main unit
(analyzer), reagents, controls and calibrators, manuals, and accessories. Performance of the
system depends on the combined integrity of all components.
2-2
Understanding Your Analyzer
2.3 Parameter
NOTE
The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that
require additional studies.
The analyzer is used for the quantitative determination of the following 19 parameters and
provides 3 histograms.
2-3
Understanding Your Analyzer
2-4
Understanding Your Analyzer
2-5
Understanding Your Analyzer
2-6
Understanding Your Analyzer
2-7
Understanding Your Analyzer
2.4.3 Indicator
The indicator may light in red, yellow and green to indicate current status of the system.
When the indicator stays in green, the analyzer is "Ready"; when it flickers in green, the
analyzer is running; when it stays in red, the analyzer encounters error and has stopped
running; when it flickers in red, the analyzer encounters error but is still running; and when it
stays in yellow, the analyzer is sleeping.
2-8
Understanding Your Analyzer
wireless network card, etc.. The analyzer supports software upgrade through USB.
2.4.6 Peripherals
Keyboard (Optional)
A Keyboard can be connected to a USB port on the analyzer. You can use it to operate your
analyzer.
Mouse (Optional)
A mouse can be connected to a USB port on the back of the analyzer. You can use it to
operate your analyzer.
A USB printer can be connected to a USB port on the back of the analyzer. You can use it to
print out reports or other interested information displayed on the screen.
Supported printers: HP LaserJet P1606dn.
A barcode scanner can be connected to a USB port on the back of the analyzer. You can use
it to scan the barcode information into the analyzer.
NOTE
When you purchase a wireless WiFi module by yourself, make sure you
comply the relevant local laws and regulations.
2-9
Understanding Your Analyzer
2-10
Understanding Your Analyzer
2.5.2 Menus
Tap the "Menu" button to display the system menu.
Tap one of the 10 options on the system menu to enter corresponding screen.
2-11
Understanding Your Analyzer
NOTE
Store and use the reagents as instructed by instructions for use of the
reagents.
When you have changed the diluent, lyse, run a background to see if the
results meet the requirement.
Pay attention to the expiration dates and open-container stability days of all
the reagents. Be sure not to use expired reagents.
After installing a new container of reagent, keep it still for a while before
use.
2.6.1 Reagents
M-30D Diluent
As an isotonic reagent and with specified conductivity, M-30D diluent provides stable
environment for hematology analysis.
M-30CFL Lyse
M-30CFL lyse is formulated to lyse red blood cells and transform the hemoglobin released
from red blood cell into hemoglobin complex. It is used for WBC count, WBC 3-part
differential and HGB determination.
Probe Cleanser
2-12
Understanding Your Analyzer
2-13
Understanding Your Analyzer
2-14
3 Understanding the System Principles
3.1 Introduction
The analyzer uses the electrical impedance method to determine the count and size
distribution of RBC, WBC and PLT; and uses the colorimetric method to determine HGB.
Based on the above data, the analyzer calculates other parameters.
3-1
Understanding the System Principles
3.2 Aspiration
If you are to analyze a whole blood sample, present the sample to the analyzer directly, and
the analyzer will aspirate 9 μL of the whole blood sample.
If you are to analyze a capillary blood sample under the pre-dilute mode, you should first
manually dilute the sample (20 μL capillary sample needs to be diluted by 0.7 mL of diluent to
form a 1:36 dilution), and then present the pre-diluted sample to the analyzer, which will
aspirate 198uL of the sample.
3-2
Understanding the System Principles
3.3 Dilution
Usually in blood samples, the cells are too close to each other to be identified or counted. For
this reason, the diluent is used to separate the cells so that they draw through the aperture
one at a time as well as to create a conductive environment for cell counting. Moreover, red
blood cells usually outnumber white blood cells by 1,000 times. For this reason, lyse need to
be added to the sample to eliminate the red blood cells before the WBC counting. Because
red blood cells usually have no nucleus, they are eliminated when the lyse breaks down their
cell walls. The analyzer provides whole blood mode and predilute mode for the analysis of
different sample types.
9ul of whole
blood sample
1.41ml of
diluent
21.6ul
1:157.65
dilution
3-3
Understanding the System Principles
700ul of diluent
1:36 dilution
198ul
1410ul of lyse
35ul
1:257.34
dilution
3-4
Understanding the System Principles
The WBCs are counted by the impedance method. The analyzer aspirates certain volume of
sample, dilutes it with certain volume of conductive solution, and delivers the dilution to the
metering unit. The metering unit has a little opening which is called "aperture". An pair of
electrodes is positioned on both sides of the aperture to create a constant-current supply. As
cells are poor conductors, when each particle in the diluted sample passes through the
aperture under the constant negative pressure, a transitory change in the direct-current
resistance between the electrodes is produced. The change in turn produces a measurable
electrical pulse which is proportional to the particle size. And when the particles passes the
aperture in succession, a series of pulses are produced between the electrodes. The number
of pulses generated indicates the number of particles passed through the aperture; and the
amplitude of each pulse is proportional to the volume of each particle.
Each pulse is amplified and compared to the internal reference voltage channel, which only
accepts the pulses of a certain amplitude. All the collected pulses are thus classified based
on the reference voltage ranges of different channels, and the number of the pluses in the
WBC channel indicates the number of the WBC particles. The cell size distribution width is
represented by the number of particles falling in each channel.
3-5
Understanding the System Principles
100
WBC'=WBC ×
100+NRBC
In the formula, WBC′ is corrected WBC count result; WBC is the WBC count provided by the
analyzer; and NRBC indicates the number of NRBCs found when every 100 WBCs are
counted.
3-DIFF of WBC
Lyses and diluents change the sizes of each type of WBCs in various ways and at different
time. The WBCs are thus separated into 3 parts (from the largest size to the smallest):
lymphocytes, mid-sized cells (including monocytes, eosinophils, and basophils) and
granulocytes.
The analyzer then calculate the lymphocyte percentage (Lymph%), mid-sized cell percentage
(Mid%) and granulocyte percentage (Gran%) (all presented in %) based on the WBC
histograms and in accordance with below formulae:
PL
Lymph% = × 100
PL + PM + PG
PM
Mid% = × 100
PL + PM + PG
PG
Gran% = × 100
PL + PM + PG
In the formulae: PL indicates the number of cells falling in the lymphocyte region, PM the
number of cells falling in the mid-sized cell region, and PG the number of cells falling in the
granulocyte region. All three parameters are presented in 109/L.
When the three percentages are obtained, the analyzer automatically proceeds to calculate
the lymphocyte number (Lymph#), mid-sized cell number (Mid#) and granulocyte number
9
(Gran#) with below formulae, all parameters expressed in 10 /L.
Lym% × WBC
Lymph# =
100
Mid % × WBC
Mid # =
100
Gran % × WBC
Gran # =
100
3-6
Understanding the System Principles
9
Lymph%, Mid% and Gran% are expressed in %, while WBC is in 10 /L.
Besides the count results, the analyzer also provides a WBC histogram which shows the
WBC size distribution, with the x-aixs representing the cell size (in fL) and the Y-axis
9
representing relative cell number (in 10 /L). After each analysis cycle, you can either check
the WBC histogram in the analysis result area on the "Sample Analysis" screen or review
the histogram on the "Review" screen.
WBC = n × 109 / L
Blank Photocurrent
HGB(g/L) = Constant × Ln
Sample Photocurrent
3-7
Understanding the System Principles
3-8
Understanding the System Principles
RBC = n × 1012 / L
The analyzer calculates the mean cell volume (MCV, in fL) based on the RBC histogram.
The hematocrit (HCT, %), mean corpuscular hemoglobin (MCH, pg) and mean corpuscular
hemoglobin concentration (MCHC, g/L) are calculated as follows:
RBC × MCV
HCT =
10
HGB
MCH =
RBC
HGB
MCHC = × 100
HCT
12
where RBC is expressed in 10 /L, MCV is expressed in fL and HGB is expressed in g/L.
RDW-CV
Red Blood Cell Distribution Width - Coefficient of Variation (RDW-CV) is derived based on
RBC histogram. It is expressed in %, and indicates the variation level of RBC size
distribution.
RDW-SD
Red Blood Cells Distribution Width - Standard Deviation (RDW-SD, in fL) measures the width
of the 20% level (with the peak taken as 100%) on the RBC histogram, as shown in Figure
3-5..
Figure 3-5
Besides the count results, the analyzer also provides a RBC histogram which shows the RBC
size distribution, with the x-aixs representing the cell size (in fL) and the Y-axis representing
12
relative cell number (10 /L). After each analysis cycle, you can either check the RBC
histogram in the analysis result area on the "Sample Analysis" screen or review the
histogram on the "Review" screen.
3-9
Understanding the System Principles
PLT = n × 10 9 / L
Mean Platelet Volume
Based on the PLT histogram, this analyzer calculates the mean platelet volume (MPV, fL).
PDW
Platelet distribution width (PDW) is derived from the platelet histogram, and is reported as 10
geometric standard deviation (10 GSD).
PCT
PLT × MPV
PCT =
10000
Based on the PLT histogram, this analyzer calculates the large platelet ratio (%).
Platelet Histogram
Besides the count results, the analyzer also provides a PLT histogram which shows the PLT
size distribution, with the x-aixs representing the cell size (in fL) and the Y-axis representing
9
relative cell number (in 10 /L). After each analysis cycle, you can either check the PLT
histogram in the analysis result area on the "Sample Analysis" screen or review the
histogram on the "Review" screen.
3-10
Understanding the System Principles
3.6 Wash
After each analysis cycle, each element of the analyzer is washed:
3-11
4 Installing Your Analyzer
4.1 Introduction
CAUTION
Installation by personnel not authorized or trained by Mindray may cause
personal injury or damage your analyzer. Do not install your analyzer
without the presence of Mindray-authorized personnel.
The analyzer is tested and packed with care before it is shipped from the factory. When you
receive your analyzer, carefully inspect the carton. If you see any signs of mishandling or
damage, contact Mindray Customer Service department or your local distributor immediately.
4-1
Installing Your Analyzer
enough room on or below the countertop to accommodate the reagent (for example
diluent) and waste containers.
the table (or the floor) where the analyzer is placed shall be able to withstand at least
40kg of weight.
WARNING
Make sure the analyzer is properly grounded.
Before turning on the analyzer, make sure the input voltage meets the
requirements.
CAUTION
Using pinboard may bring electrical interference and the analysis results
may be unreliable. Place the analyzer near the electrical outlet to avoid
using the pinboard.
Please use the original power cable shipped with the analyzer. Using other
electrical wire may damage the analyzer or lead to unreliable analysis
results.
4-2
Installing Your Analyzer
The environment shall be as free as possible from dust, mechanical vibrations, loud noises,
and electrical interference. Do not place the analyzer in direct sunlight or in front of a source
of heat or drafts. Please use a separate power socket; do not use the same socket with
devices like air conditioning, refrigerator and ultrasonic system, as they may interfere with the
proper operation of the analyzer. It is advisable to evaluate the electromagnetic environment
prior to operation of this analyzer. Do not place the analyzer near brush-type motors,
flickering fluorescent lights, and electrical contacts that regularly open and close. The
environment shall be well ventilated. Do not place the analyzer in direct sunlight. Connect
only to a properly earth grounded outlet. Only use this analyzer indoors.
WARNING
Do not place the analyzer in a flammable or explosive environment.
NOTE
If the ambient temperature is out of the specified operating range, the
analyzer will alarm you for abnormal ambient temperature and the analysis
results may be unreliable. When temperature errors are reported in the error
information area after analysis, see Chapter 11 Troubleshooting Your
Analyzer for solutions.
4-3
Installing Your Analyzer
WARNING
Installation by personnel not authorized or trained by Mindray may cause
personal injury or damage your analyzer. Do not install your analyzer
without the presence of Mindray-authorized personnel.
NOTE
To protect it from being damaged during transportation, the aspiration
module is fixed by cables ties and clamps before the analyzer is shipped out
of factory. Remove the cable ties and clamps before using the analyzer.
4-4
Installing Your Analyzer
.
Figure 4-1 Connecting the analyzer to power outlet
4-5
Installing Your Analyzer
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc.
according to government regulations.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them and the contacted areas in the
laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
4-6
Installing Your Analyzer
4-7
Installing Your Analyzer
CAUTION
Make sure the diluent pipe and waste pipe are no longer than 1500mm.
The top of the waste container and the diluent container should be lower
than the countertop where the analyzer is placed.
4-8
Installing Your Analyzer
NOTE
Remove the protective paper in the recorder before installing recorder
paper.
1. Use the latch at the upper right corner of the recorder door to pull the door open.
2. Insert a new roll into the compartment with the paper end out of the recorder exit, as
shown below.
4. Check whether the paper is installed correctly and the paper end is feeding from the
top.
Paper roll
CAUTION
Use only specified thermal recorder paper. Otherwise, it may cause damage
to the recorder head, or the recorder may be unable to print, or poor print
quality may result.
Never pull the recorder paper with force when a recording is in process.
Otherwise the recorder may be damaged.
Do not leave the recorder door open unless you are installing paper or
removing errors.
Improper installation of recorder paper may jam the paper and/or result in
blank printout.
4-9
Installing Your Analyzer
4.5 Precautions
If the analyzer is kept in an environment with heavy dust for a long time, its performance
may be reduced.
It is recommended to clean and sterilize the outer surface the analyzer with 75%
ethanol.
The probe wipe block of the analyzer (see Figure 2-1 Front of the analyzer) shall be
wiped with 75% alcohol regularly.
Collect and prepare samples in accordance with the standard procedure of your
laboratory; otherwise it may result in inaccurate analysis results or damage to the
analyzer.
If any of the pipes or fluidic components are worn out, stop using the analyzer and
contact Mindray customer service department immediately for inspection or
replacement.
Make sure the tubings of the reagents (including diluent, lyse and waste) are not
pressed by heavy objects or bent over.
Pay attention to the expiration dates and open-container stability days of all the reagents.
Be sure not to use expired reagents. Otherwise it may result in inaccurate results.
4-10
5 Operating Your Analyzer
5.1 Introduction
This chapter provides step-by-step procedures for operating your analyzer on a daily basis.
A flow chart indicating the common daily operating process is presented below.
操作前的准备
Initial checks
Power on
Daily Quality
Control
Sample
Collection and
Handling
Run Sample
Shutdown
5-1
Operating Your Analyzer
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc.
according to government regulations.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them and the contacted areas in the
laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
To prevent personal injury during the operation, keep your clothes, hairs
and hands from the moving parts like sample probe.
NOTE
Use the reagents specified by the manufacturer only. Store and use the
reagents as instructed by instructions for use of the reagents.
Check if the reagent tubings are properly connected before using the
analyzer.
After installing a new container of reagent, keep it still for a while before
use.
Check and make sure the waste container (not supplied) is empty.
Check and make sure the reagent, waste and pneumatic unit tubes are properly connected
and not bent.
5-2
Operating Your Analyzer
Check and make sure the power cord of the analyzer is properly plugged into the power
outlet.
Check and make sure the printer and recorder are properly installed, and have enough paper.
5-3
Operating Your Analyzer
1. Place the power switch at the back of the analyzer in the "I" position. The switch will
light on.
2. Make sure that the power indicator on the analyzer lights on.
4. The analyzer will sequentially do the self-test and initialize the system.
5-4
Operating Your Analyzer
NOTE
When "Background abnormal" is reported during the startup process,
follow the corresponding troubleshooting instruction for the error in
Chapter 11 Troubleshooting to remove the error.
The system opens different function for the user according to the user level.
The user level depends on the user ID and the password when the user logs
in.
To switch to another user, tap the "Logout" button first. Enter the desired
user ID and password in the displayed login dialog box, and then tap "OK"
to log in.
After startup, please make sure the date/time of the computer is correct.
The default user ID and password for administrator are both "Admin".
1-12 characters are allowed for user ID and password; Chinese characters
are not allowed.
5-5
Operating Your Analyzer
5-6
Operating Your Analyzer
CAUTION
Prepare samples following the recommend procedure of the manufacturer.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
WARNING
Do not contact the patients' sample blood directly.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes,
capillary tubes and so on.
NOTE
Be sure to use clean EDTAK2 anticoagulant collection tubes, fused silica
5-7
Operating Your Analyzer
Samples)
1. Use clean EDTAK2 anticoagulant collection tubes to collect venous blood samples.
2. Mix the sample immediately according to your laboratory’s protocol.
CAUTION
To attain accurate analysis results, make sure the sample volume is no less than
120uL.
NOTE
For the whole blood samples to be used for WBC differential, you shall store
them at the room temperature and run them within 8 hours after collection.
Be sure to mix any sample that has been prepared for a while before running
it.
After the sample is prepared, be sure to wait for at least 5 minutes before
running it; and you must complete the analysis within 2 hours after its
collection.
5-8
Operating Your Analyzer
2. Tap the "Diluent" button on the top status bar, a message box will pop up.
3. Present a tube to the analyzer, and tap the [Aspirate] key on the analyzer to dispense
diluent (700μL). During dispensing the diluent, a progress bar will display.
5. Add 20μL of capillary blood to the centrifugal tube of diluent, close the tube cap and
shake the tube to mix the sample.
6. After the prediluted sample is prepared, tap the “Cancel” button to exit dispensing the
diluent.
NOTE
You can also dispense 700μL of diluent by pipette into the tube.
Take methods to prevent the diluent from dust and volatilization; otherwise
the results may be unreliable.
After mixing the capillary sample with the diluent, be sure to wait at least for
3 minutes, and mix the sample again before running the sample.
Be sure to run the prediluted samples within 30 minutes after the dilution.
Be sure to mix any sample that has been prepared for a while before running
it. Do not use a vortex mixer for mixing, for the shaking be too violent and
cause hemolysis.
5-9
Operating Your Analyzer
5-10
Operating Your Analyzer
NOTE
You can enter letters, digits and all other characters on the keyboard for
sample ID, only for [a-z][A-Z][0-9][-_].
The sample ID must end with a digit; and it cannot only consist of "0".
Select the desired item (“Male”, “Female”, or null) from the "Gender" pull-down list. The
default option is "Unknown".
Enter the birth date of the patient into the "Date of Birth" box, the birth date format being the
same of the system date.
The analyzer provides 5 ways for you to enter patient age – in years, in months, in days and
in hours. The first way is designed for the patients no younger than one year; the second for
the infant patients of one month to two year; the third for the neonatal of one week to ten
weeks; the fourth for the neonatal no older than one month, and the fifth for the neonatal no
older than 48 hours. You may choose one of the four ways to enter the patient age.
Select the desired item from the "Age"pull-down list (“Years”, "Month(s)”, “Weeks”,”Days”
and “Hours”), and you may enter the patient age in the box followed by the age unit.
NOTE
When you enter the patient birth date, the system will automatically
calculate the patient age using the entered "Birth of Date" and current
"System Date" and display the result in the "Age" box and the age "Unit"
combo box, The "Age" box will then be greyed, and will become editable
again when the "Date of Birth" is cleared.
The patient birth date should be no later than current system date.
5-11
Operating Your Analyzer
Select "Outpatient", "Inpatient", "Medical Examination" or "STAT" from the "Patient Type"
pull-down list.
You can either enter the department name in the "Department" box, or select the desired
department from the "Department" pull-down list (if there are previous entries saved in the
list)..
Enter the bed number of the patient to the "Bed No." box.
Enter the time when the sample is collected into the "Draw Time" box.
Enter the time when the sample is sent into the "Delivery Time" box.
You can either enter the clinician name into the "Clinician" box, or select the desired clinician
from its pull-down list (if there are previous saved entries in the list).
OK
When you have finished entering the sample information, tap "OK" to save the information
and return to the "Sample Analysis" screen.
Cancel
If you do not want to save the entered sample information, tap "Cancle" to return the
"Sample Analysis" screen without saving the changes.
Enter sample ID
When you have set "Entry of next sample info" to "Enter sample ID only", tap "Next
sample" on the "Sample Analysis" screen, and a dialog box pop up as below.
5-12
Operating Your Analyzer
Enter the ID in the "Sample ID" box. Tap the "OK" button to save the sample ID and close the
dialog box. The ID will be displayed in the "Next Sample" information area on the bottom of
the screen.
5-13
Operating Your Analyzer
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
WARNING
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the sharp sample probe when
working around it.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes,
capillary tubes and so on.
NOTE
Make sure that the sample probe is fully immersed into the sample and not
in contact with the tube bottom; otherwise the aspiration volume may be
insufficient, or the aspirated volume may not be accurate.
Make sure the sample probe tip does not touch the tube wall, otherwise the
blood sample may spill.
Sample Analysis
Do as follows to run samples:
1. Make sure the analyzer indicator shows that the analyzer is ready for sample analysis,
and the analysis mode is "WB" or "PD".
2. Present a well mixed sample to the sample probe for aspiration.
3. Press the [Aspirate] key to start sample analysis. When the analyzer indicator flickers in
green, the analyzer is running.
4. The sample probe will automatically aspirate certain volume of the WB or PD samples.
When you hear the beeps, remove the sample tube. The probe will ascend and add the
aspirated sample to the count baths. The analyzer will automatically run the sample.
5. When the analysis is finished, the result will be displayed in the analysis result area on
5-14
Operating Your Analyzer
the screen. The sample probe returns to the original position and gets ready for the next
analysis.
6. When "Auto print after sample analysis" is set to "On", the analyzer will automatically
print the analysis result report in the preset format; when "Auto Communicate" is set to
"On", the analyzer will automatically upload the eligible sample results as well as sample
and patient information to the LIS system.
7. Repeat above steps to run other samples.
NOTE
If the analyzer detects clogging or bubbles during the analysis, the
corresponding error message will be displayed in the error message area
and the results of all related parameters will be invalidated. See Chapter 11
Troubleshooting Your Analyzer for solutions.
Histogram flags
The system will flag abnormal histograms. Both WBC histogram and PLT histogram are
flagged for abnormal results.
Abnormal WBC histograms will be flagged by one of the markings: R1, R2, R3, R4 and Rm.
The indications of the markings are as follows:
R1 indicates abnormality on the left side of the lymphocyte hump and possible presence of
platelet clump, giant platelet, nucleated red cell, lyse resistant RBC, high molecular weight
protein and lipoid debris in sample, or electrical noise.
R2: indicates abnormality between the lymphocyte hump and the mid-sized cell area,
possible presence of atypical/immature lymphocyte, plasma cell, blast cell in the sample,
eosinophilia or basophilia.
R3: indicates abnormality between the mid-sized cell area and the granulocyte hump,
possible presence of immature granulocyte, blast cell, left shift, immature monocyte or
eosinophilia.
R4: indicates abnormality on the right side of the granulocytes hump, possible presence of
5-15
Operating Your Analyzer
Abnormal PLT histograms will be flagged by the marking Pm, PS, PL.
The indication of the marking is as follows:
Pm: indicates blur demarcation between the platelet area and red blood cell area and
possible presence of large platelet, platelet clump, small red blood cell, cell debris or high
molecular weight protein.
PS:small platelet possibly high notice.
PL:giant platelet possibly high notice.
Parameter flags
5-16
Operating Your Analyzer
NOTE
Abnormal parameter or histogram results of background check will not be
flagged. When the background results do not meet requirements, the
analyzer will alarm for abnormal background.
5-17
Operating Your Analyzer
5.7 Standby
When the fluidics system stop working for 15 minutes (default, which can be set at the setup
screen. See Chapter 9 Customizing the Analyzer Software for details), then the analyzer will
enter the standby status automatically.
After entering the sleep mode, the bottom right of the screen displays "Standby. Press the
[Aspirate] key to exit. "
NOTE
On the “Status” screen, the analyzer cannot enter the standby status.
If it is the time for standby but the analyzer has an error, then only after the
error is removed will the auto-standby starts accordingly.
You can perform the operations not involving the analyzer when it is on
standby, such as communication and print etc.
Refer to Section 9.2.5 Maintenance Setup for how to edit waiting time before
entering standby mode.
[Aspirate] key
Press the [Aspirate] key on the analyzer to exit the standby mode.
5-18
Operating Your Analyzer
After canceling the standby mode, the progress bar will be closed automatically, the analyzer
will exit the standby mode.
NOTE
Different maintenances will be performed by the analyzer automatically
when exiting the standby status, and the exiting time depends on how long
the analyzer was in the standby status.
If any error happens during the process of exiting the standby status, see
Chapter 11 Troubleshooting Your Analyzer for details to remove the error.
After exiting standby mode, the analyzer will return to the status before
standby. The analysis status icon on at the screen displays in green. The
indicator on the analyzer displays in green at the same time.
5-19
Operating Your Analyzer
5.8 Shutdown
Perform the "Shutdown" procedure to shut down the analyzer every day.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
WARNING
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the sharp sample probe when
working around it.
NOTE
To ensure stable analyzer performance and accurate analysis results, be
sure to perform the "Shutdown" procedure to shut down the analyzer after it
has been running continuously for 24 hours.
Do not force power off the analyzer during the "Shutdown" procedure.
If error that will affect shutdown occurs during the showdown process, the
analyzer will resume to its original status and report the error. See Chapter
11 Troubleshooting Your Analyzer for solutions.
1. Tap the "Shutdown" button on the main menu, the shutdown dialog box shown as below
will display:
5-20
Operating Your Analyzer
2. Tap "OK", and follow the instruction to present probe cleanser to the sample probe, then
press the [Aspirate] key. The sample probe automatically aspirates the probe cleanser;
then the probe cleanser maintenance starts. A progress bar will be displayed on the
screen to indicate the probe cleanser maintenance progress.
3. When the shutdown procedure is finished, the screen will display "Please power off the
analyzer". Turn off the analyzer.
4. Empty the waste container and dispose of the waste properly.
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc.
according to government regulations.
5-21
Operating Your Analyzer
5-22
6 Reviewing Sample Results
6.1 Introduction
After every analysis cycle, the analyzer automatically saves the analysis results into the
sample database. Totally 500,000 records (including parameter results and histograms) can
be saved.
You can either choose the "Table Review" mode to review the parameter results of all
samples saved in the sample and search databases; or the "Graph Review" mode to review
both the parameter results and the histograms of each sample.
NOTE
The sample result data must have proper backup in case of data lost caused
by hardware or software error.
6-1
Reviewing Sample Results
Table area
The table area displays the list of the analyzed samples with basic information like sample ID
and sample state.
6-2
Reviewing Sample Results
NOTE
The latest sample record is on the utmost top of the table.
Graph review
You can either tap the "Graph Review" button on the "Table Review" screen or tap
"Previous" on the "Sample Analysis" screen to review the detailed results of each sample.
6-3
Reviewing Sample Results
Tap the button to switch between the “Count” screen and the “Graph Review” screen.
Edit results
Tap the desired sample result and it will be highlighted. Tap the "Edit Result" button and the
following dialog box will display.
Modify the results and tap "OK" to save the changes. The information on the graph review
screen will be refreshed.
6-4
Reviewing Sample Results
2. Enter searching conditions into the edit boxes or select them from the pull-down
lists.
3. Tap "OK" to start search, the results will displayed in the table.
Edit information
Tap the desired sample result on the "Table Review" screen and it will be highlighted. Tap the
"Edit Info" button and the following dialog box will display.
Modify the sample and patient information as necessary, and tap "OK" to save the changes.
The information on the table review screen will be refreshed.
6-5
Reviewing Sample Results
Select the sample record(s) to be validated on the “Table Review” screen, and then tap the
"Validate" button to validate. The "Sample State" of the record(s) will become "Validated".
Cancel validate
Select the validated sample record(s) on the "Table Review" screen, and then tap the
"Cancel validate" button. The text of "Validated" will disappear from the "Sample State".
6-6
Reviewing Sample Results
Export
6-7
Reviewing Sample Results
3. Tap "OK" to delete the record, and the dialog box will be closed.
Trend Graph
Tap the "Trend Graph" button to see the trend graph of sample results.
6-8
Reviewing Sample Results
Print
Print reports as per the default report template
Select sample records to be printed, and then tap "Print" to print them. On the "Graph
Review" screen, the "Sample State" of the printed sample will become "Printed".
NOTE
When a sample has been validated and printed, its "Sample State" will
display "Validated".
6-9
6-1
Using the QC Programs
Analysis of the results with statistical methods allows the inference that sample results are
reliable. Mindray recommends you run the QC program daily with low, normal and high level
controls. A new lot of controls should be analyzed in parallel with the current lot prior to their
expiration dates. This may be accomplished by running the new lot of controls twice a day for
five days using any empty QC file. The QC files calculate the mean, standard deviation and
coefficient of variation for each selected parameter. The instrument-calculated means of
these ten runs should be within the expected ranges published by the manufacturer.
NOTE
Use the controls and reagents specified by the manufacturer only. Store and
use the controls and reagents as instructed by their instructions for use.
7-2
Using the QC Programs
7.2 L-J QC
7.2.1 Editing L-J Settings (for administrators only)
Before running a new lot of controls, you must set up a QC file for each lot of controls.
1 Tap the menu option "QC" > "L-J QC" > "Setup".
2 Enter the L-J QC setup screen.
.
7-3
Using the QC Programs
NOTE
The "Import target/limits" check box is selected by default. If it is
deselected, the operator must enter the target and limits of QC
parameters manually.
Manual Entry
1. Enter the L-J QC setup screen.
2. Tap "New", or select a QC file without QC results, and then tap "Edit".
3. Enter the lot No. of the controls in the edit box manually.
7-4
Using the QC Programs
NOTE
The lot No. shall not be empty and up to 16 digits can be entered.
You can enter characters, numbers, letters.
7-5
Using the QC Programs
Setting limits
You can adjust the format of limits as per the following procedure:
1. Tap "Set Limits".
2. Tap “By SD” to display the limits in the form of absolute value;
or tap “By CV” to display the limits in the form of percentage.
3. Tap the “OK” button to save the settings.
Put controls together with normal samples, and run the controls under the sample
analysis screen.
Whole blood
Predilute
CAUTION
Running QC sample with error present will lead to unreliable results. If
errors are reported during QC analysis, remove the errors first and then
continue with the analysis.
7-6
Using the QC Programs
NOTE
When switching mode from "PD" to "WB", a progress bar will be displayed
while the analyzer runs mode switching sequence.
1 Tap "QC" > "L-J QC" > "Count" to enter the QC count screen.
NOTE
Be sure that the level of the control to be run is the same with the
current QC file, and the control is not expired.
7-7
Using the QC Programs
NOTE
Up to 100 QC results can be saved in each QC file.
Put controls together with normal samples, and run the controls under the sample analysis
screen.
After setting special "QC Sample ID" for a control under the QC setup screen, you can put
the control together with normal samples, and run it under the “Sample Analysis” screen.
When editing worklist or entering next sample information in the "Next Sample" dialog box
before daily analysis, enter the special "QC Sample ID" as "Sample ID".
Based on the QC mode selected, you can choose to run QC analysis from one of the
following ways:
Whole blood
Predilute
NOTE
Up to 100 QC results can be saved in each QC file.
NOTE
When switching blood mode from "PD" to "WB", a progress bar will be
displayed while the analyzer runs mode switching sequence.
Tap "Edit Result" on the QC screen to edit results and tap "OK" to save the edited results.
The edited results will be marked with an "E".
7-8
Using the QC Programs
Operators of administrator access level can restore the edited results to the original
measurement results.
1. Tap "Restore" on the "Edit Result" screen.
QC Graph
QC Table
2. You can tap the arrow buttons on the right of the graph to browse graphs of the
parameters. You can tap the arrow buttons under the graph horizontally to browse all
the QC results.
7-9
Using the QC Programs
NOTE
If a parameter target/limits of the QC files with QC results are modified and
saved, and the targets/limits of other parameters changes accordingly,
those changed data will be highlighted in yellow.
Tap the "Print" icon in the status bar to print information of the current QC file and the QC
graph of all parameters.
NOTE
The green vertical line and values of the corresponding QC points will not
be printed.
2. You can tap the arrow buttons on the right of the QC table to browse all QC
records. You can tap the arrow buttons under the QC table to browse all the
parameter results.
7-10
Using the QC Programs
NOTE
If a parameter target/limits of the QC files with QC results are modified and
saved, and the targets/limits of other parameters changes accordingly,
those changed data will be highlighted in yellow.
NOTE
The operation will be recorded in the system log.
You can tap the "Print" icon in the status bar to print the QC table.
Transmission
To transmit QC data to external data management software or HIS/LIS/HIS, tap the "Comm."
button to transmit specified results to the data management software.
NOTE
If auto-communication is enabled and a sample is run during the
transmission of the QC data, then only when the QC data transmission
finished will the auto-communication of the sample result start.
7-11
Using the QC Programs
Export
7-12
Using the QC Programs
On the X-B QC setup screen, you may activate/deactivate X-B QC, set target/limits, and
configure the sample validity setup.
7-13
Using the QC Programs
Set target/limits
Before the X-B QC analysis, you shall set up the target and limit for each parameter at the
X-B QC setup screen.
NOTE
The units of target/limits of all parameters are the same as those in the
parameter unit setup screen.
In the “Target/Limit” area of the X-B QC setup screen, specify the targets and limits in
1.
the “Target/Limit” table by entering manually.
NOTE
Do not leave any of the targets and limits for the QC parameters
blank.
When first use, the default setting will provide the initial values for
the targets and limits of all QC parameters.
"Sample Validity Setup" is to set up the ranges of valid RBC, MCV, MCH and MCHC results.
Only when the results of all these four parameters are within the specified ranges, the sample
results can be used for X-B QC calculation. Do as follows to set the sample validity:
7-14
Using the QC Programs
The default validity range of each parameter is shown in the following figure.
7-15
Using the QC Programs
NOTE
In the sample validity setup, the upper limit shall be no smaller than the
lower limit. Otherwise, there will be prompted message asking you to revise.
The valid ranges of the RBC parameters are their linearity ranges; the valid
ranges of other parameters are their display ranges.
All the entries shall be numbers with only one decimal point. The length of
the number entered cannot be longer than the length of the text box.
Once the validity range is changed, the previous results will not be used in
the QC calculation as valid results. For example, if 20 valid samples are
needed for the X-B QC calculation, when you change the validity range after
10 groups of valid sample results have been acquired, these 10 groups of
results will be discarded, and only valid sample results generated
afterwards will be used in the QC calculation.
The units of lower and upper limits of all parameters are the same as those
in the " Reference Unit Setup " screen. See section 9.2.4 Setting Parameters-
Parameter unit setup.
Setting limits
You can adjust the format of limits as per the following procedure:
1. Tap "Set Limits".
2. Tap “By SD” to display the limits in the form of absolute value;
3. Tap “By CV” to display the limits in the form of percentage.
4. Tap the “OK” button to save the settings.
7-16
Using the QC Programs
Restore defaults
If you want to restore the default targets and limits of the parameter, tap "Restore Defaults".
The default values of the target and limits of each parameter are as follows:
Parameter Target Limits (#)
MCV 89.5 2.7
MCH 30.5 0.9
MCHC 340 10
7.3.3 QC Analysis
After editing X-B setup, the system will start X-B QC automatically.
After every 20~200 results (determined by the setting) are obtained, the system will perform
the X-B calculation once automatically. You can review the result in X-B QC graph or X-B QC
table.
QC Graph
QC Table
1. Tap the menu option "QC" > "X-B QC" > "Graph", the following screen will display.
7-17
Using the QC Programs
2. Select QC file No., the information of the file and the QC graph will be displayed on the
screen.
3. You can tap the arrow buttons under the graph horizontally to browse all the QC
results.
3. You can tap the arrow buttons on the right of the QC table to browse all QC records.
The delete, print and export operations can all be performed same as stated in the L-J QC
table review section.
7-18
8 Calibrating Your Analyzer
8.1 Introduction
Calibration is a procedure to standardize the analyzer by determining its deviation under
certain specified conditions. In order to get accurate sample analysis results, you should
calibrate the analyzer per the procedure below when necessary.
The analyzer supports 3 calibration programs, namely manual calibration, auto calibration
using calibrators and auto calibration using fresh blood samples; and two calibration modes,
namely "WB" and "PD".
All the parameters or part of the parameters of WBC, RBC, HGB, MCV and PLT can be
calibrated by the calibration programs.
NOTE
Calibration procedures can only be performed by users of the
administrator-level.
Use the calibrators and reagents specified by the manufacturer only. Store
and use the calibrators and reagents as instructed by their instructions for
use.
8-1
Calibrating Your Analyzer
NOTE
All of the measured parameters must be calibrated before readings of this
analyzer can be used as valid analysis results.
8-2
Calibrating Your Analyzer
1. Check and make sure enough reagents have been prepared for the calibration. You need
to start over the calibration if the reagents run out during the process.
2. Check the background (for calibration right after startup) or blank count results .If the
analyzer alarms for abnormal background results, see Chapter 11 Troubleshooting Your
Analyzer for solutions (see Appendix B Specifications for the background range).
3. Run a vial of normal control consecutively for 10 times under the whole blood mode.
Enter the "Table Review" screen to check the reproducibility of the 10 runs and make
sure they meet the following requirements.
It is recommended that you create a log table for your analyzer. This log table should contain
all necessary information that is pertinent to your analyzer. Suggested items that you may
want to include in the log table are: calibration date, supplier of calibrator, lot number,
expected results and limits, and result of background check.
NOTE
Be sure to use the evacuated collection tubes recommended in the
Appendix.
If fresh blood samples are used for reproducibility test, make sure the
sample volume is enough to support the test.
8-3
Calibrating Your Analyzer
NOTE
If you log in at the operator access level, you can only view the calibration
factors. To perform calibration, please log out and then log in at the
administrator access level.
8-4
Calibrating Your Analyzer
If the entered calibration factors are valid, a dialog box will pop up asking you to save the
new factor when you are exiting the screen. And the calibration date of the
corresponding parameter changes to current system date.
If the entered calibration factors are invalid, a dialog box will pop up prompting "Invalid
entry" when you are switching to another screen. The new calibration factor will not be
saved, and the calibration date will not be refreshed.
Other Operations
Print
If the calibration factors are valid but not saved, a dialog box will display asking you to save
the factors.
Tap "Yes" to save and print the factors. Or Tap "No" to cancel the operation without saving or
printing them.
8-5
Calibrating Your Analyzer
NOTE
The calibrator calibration can only be performed under whole blood mode.
See the instruction for use of the calibrators for the lot No., expiration date
and the target.
The out-of-range CV% does not influence the display of calibration factors.
8-6
Calibrating Your Analyzer
Tap “OK” to close the message box, and the data will be deleted from the table without
saving automatically.
10. When the running is done, if there is a parameter whose calibration data is out of the
display range, then the non-numeric parameter values "***" will be displayed in the list
and a message box will pop up.
Tap “OK” to close the message box, and the data will be deleted from the table without
saving automatically.
The valid results within the linearity range will be displayed directly. Valid calibration
results will be marked with "√” per the default setting, and will be taken to calculate
calibration factors.
11. If the calibration factors have not been calculated but you switch to another screen, then
a message box will pop up.
8-7
Calibrating Your Analyzer
Tap "Yes" to switch to another screen while discarding the calibration data and closing
the message box. The original calibration factors remain.
12. When calibration count has been performed to a sample for n times (n≥5), the analyzer
will calculate the Mean, CV% and calibration factors of all the calibration data marked
with "√" (calibration data of the first run is not marked with "√", so it is not included in the
calculation).
You can select several data to calculate the calibration factors, but only after at least 5
groups of the data are marked with "√" can you get the calibration factors. The
calibration factors will be refreshed whenever you select “√” or deselect “√”.
When the amount of valid calibration data in the list reaches 10, a message box
"Calibration is completed!" will pop up. Then, if you press the [Aspirate] key again, the
analyzer will beep without starting analysis.
13. There may be two cases when you are switching to another screen:
If the calibration factors of any parameter is out of the range [75%-125%] or the CV% of
any parameter exceeds the reproducibility range, then the calculated calibration factors
of all parameters will not be saved and a message box will also pop up.
Tap "Yes" to close the dialog box and switch to another screen. The calibration factors
and dates of all parameters will not be changed.
If the calculated calibration factors of all parameter are within the range [75%-125%] and
the CV% of all parameter are also within the reproducibility range, then a message box
will pop up.
Tap "Yes" to save the new calibration factors while closing the message box and
switching to another screen.
8-8
Calibrating Your Analyzer
Other Operations
Print
If the calibration factors are invalid, tap print, the dialog box "New calibration factor is
invalid." will display.
If the calibration factors are valid, tap "Print", a dialog box will display.
Tap "Yes" to close the dialog box, save and print the calibration results. Or tap "No" to cancel
the operation without saving or printing them.
8-9
Calibrating Your Analyzer
If the results are out of the linearity range but still within the display range, a dialog box
will pop up when the results are displayed in the table.
Tap “OK” to close the message box, and the data will be deleted from the table without
saving automatically.
If the results are out of the display range, the non-numeric parameter values "***" are
obtained and a dialog box will pop up.
Tap “OK” to close the message box, and the data will be deleted from the table without
saving automatically.
The valid results within the linearity range will be displayed directly.
Valid calibration results will be marked with "√” per the default setting, and will be taken
to calculate calibration factors.
8-10
Calibrating Your Analyzer
10. When calibration count has been performed to a sample for n times (n≥5), the analyzer
will calculate the Mean, CV% and calibration factors of all the calibration data marked
with "√" automatically.
You can select several data to calculate the calibration factors, but only after at least 5
groups of the data are marked with "√" can you get the calibration factors. The
calibration factors will be refreshed whenever you select “√” or deselect “√”.
When the amount of valid calibration data in the list reaches 10, a message box
"Calibration with the current blood sample is completed." will pop up when you start
calibration again.
11. Select other calibration sample ID from the "Current Sample ID" pull-down box, analyze
other samples according to Step 8~10 above to obtain the calibration factors of all
samples.
12. There may be several cases when switching to another blood sample:
If the calibration factors of the blood sample are invalid or the CV% of any parameter
exceeds the reproducibility range, a dialog box will pop up when switching to another
blood sample.
Tap "Yes" to empty the entered target of the current sample, all the calibration data
obtained and each calculated value including calibration factors, then close the dialog
box and switch to another blood sample.
If the calibration factors have not been calculated, a dialog box will pop up.
Tap “Yes” to empty the entered target of the current sample and all the calibration data
obtained, then close the dialog box and switch to another blood sample.
If the calibration factors of the sample are valid and the CV% of all the parameters do not
exceed the reproducibility range, you can switch to another blood sample directly.
13. After calibration factors of at least 3 fresh blood samples are obtained, tap the "Calculate"
8-11
Calibrating Your Analyzer
Select or deselect the calibration factors of a blood sample for the calculation of the
Mean calibration factors by tapping the check boxes before the calibration factors.
When 3 or more groups of calibration factors are checked, CV% will be re-calculated
automatically base on the checked calibration factors.
When 3 or more groups of calibration factors are checked, the mean calibration factor
will be re-calculated automatically base on the checked calibration factors. The mean
calibration factors are regarded as invalid if the deviation of absolute value between the
calibration factors included in calculating the mean and the original calibration factors
reaches or exceeds 5%; a dialog box will pop up when you exit the current fresh blood
calibration screen.
Tap "Yes" to close the dialog box and exit with the current calibration data emptied, and
then switch to another screen.
Tap "No" to return to the current screen. Invalid mean calibration factors are followed
with a "?". and displayed in red.
14. If the mean calibration factors have not been calculated, when you exit the fresh blood
screen or switch to another calibration mode, a dialog box will pop up.
8-12
Calibrating Your Analyzer
Tap "Yes" to discard the calibration data, close the dialog box, and switch to another
screen or calibration mode. The original calibration factors and date remain the same.
15. If the calculated mean calibration factors are valid, when exiting the fresh blood
calibration screen or switching to another calibration mode, a dialog box will pop up.
Tap "Yes" to save the current mean calibration factors. Then, you can switch to another
screen or calibration mode. Tap "No" to close the dialog box and switch to another
screen or calibration mode without saving the mean calibration factors and all the
calibration data.
Other Operations
Print
If the mean calibration factors are invalid, tap print, the dialog box "Calibration factor is
invalid." will display.
If the mean calibration factors are valid, you can tap "Print" to print the calibration factors of a
group (or more) of blood samples in table form, no matter whether they are selected ("√") or
not. The results obtained in the calibration process and the mean calibration factors can also
be printed.
8-13
9 Customizing the Analyzer Software
9.1 Introduction
The analyzer is a flexible laboratory instrument that can be tailed to your work environment.
You can use the “Setup” program to customize the software options as introduced in this
chapter.
Menu functions are introduced as below.
9-1
Customizing the Analyzer Software
Print Setup
Tap the menu option "Setup" > "System Setup" > " Print Setup" to enter the "Print setup"
screen as shown below. You can set up the following contents:
Print Setup
Printing content
Auto print after sample analysis
9-2
Customizing the Analyzer Software
Print Setup
Print Device
You can select "Printer" or "Recorder" from the "Print Device" pull-down list.
Print Drive
Tap the pull-down list to select print drive of the analyzer.
Paper
Tap the pull-down list to select the paper type of the reports to be printed.
Para. language
Tap the pull-down list to select the parameter language of the reports.
Copies
Enter the number of copies to be printed for each report into the edit box "Copies".
Report title
9-3
Customizing the Analyzer Software
Report template
Printing content
You can choose to select the functions based on your needs by tapping on the check boxes.
Communication Setup
Tap the menu option "Setup" > "System Setup" > "Communication" to enter the
communication setup screen as shown below. You can set up the following contents:
Communication
Network Device
Protocol Setup
Transmission Mode
9-4
Customizing the Analyzer Software
Communication
Network Device
You can select "Wireless" or "Wired" from the "Network Type" pull-down list.
If you select "Wireless", tap the wireless network setup icon , and choose a WiFi hot
Protocol Setup
Tap the "IP Address", "Subnet Mask" and "Default Gateway" edit boxes to enter the
contents.
Communication Protocol
Tap the "Comm. Protocol" pull-down list to select the communication protocol.
Transmission Mode
You can select the functions based on your needs by tapping on the check boxes.
Auto retransmit
Auto comm.
Transmit as Print Bitmap Data
ACK synchronous transmission
9-5
Customizing the Analyzer Software
Tap on the "ACK synchronous transmission" check box to activate the function.
When the function is activated, ACK overtime is 10 seconds by default. You can re-enter the
ACK overtime is the edit box.
9-6
Customizing the Analyzer Software
NOTE
The analyzer SN cannot be edited.
The date of installation is the date the analyzer is installed by default. It can
be edited, but cannot be later than the current system date.
9-7
Customizing the Analyzer Software
Modify password
9-8
Customizing the Analyzer Software
NOTE
Up to 12 characters can be entered.
2. Enter the "User ID", "Name" and "Password" information. The "User ID" refers to
your login account; and the "Name" will be displayed in the "Operator" or
"Validated By" fields on the "Table Review" screen and on the printed reports.
Administrator
Operator
4. Tap "OK" to save the change and close the dialog box.
NOTE
The user ID cannot be null and up to 12 characters can be entered.
9-9
Customizing the Analyzer Software
Deleting a User
NOTE
The current login user cannot be deleted.
9-10
Customizing the Analyzer Software
You may customize the first sample ID after startup by entering it into the edit box; or select to
continue with the sample ID before last shutdown.
Other settings
Flags
You may set up the suspect flag by entering a character into the edit box, or selecting a letter
from the pull-down list (the default character is "R").
You may set up the high/low flag by entering characters in the edit boxes, or selecting letters
from the pull-down lists (the default character of high flag is "H", and that of low flag is "L").
9-11
Customizing the Analyzer Software
Under each unit system, you can tap the "Unit" cell to customize the parameter unit.
Tap the "Default" button to restore the default units.
NOTE
The units displayed will be different when different unit system is selected.
9-12
Customizing the Analyzer Software
Select a reference group and tap "New" or "Edit" to enter the reference group setup screen.
You can set up the name, lower and upper limits of age and reference range.
Tap the "Set to Default" button, the reference ranges of the selected factory reference group
can be restored to the default settings.
9-13
Customizing the Analyzer Software
NOTE
The name of the reference group cannot be null.
The names of the customized reference groups shall not repeat the names
of the 5 default groups (General, Adult male, Adult female, Child, and
Neonate) and they shall not repeat each other either.
Select a reference group and then tap "Set to Default" to set it as default reference group.
NOTE
The name, lower and upper limits of age and gender of the factory reference
groups cannot be modified.
To modify the reference range of a reference group, select the group from the reference
group list on the left, and then tap the cells of upper and lower limits in the table and re-enter
the values.
To restore the reference ranges to default, tap the “Default” button on top right of the screen.
Select "Match customized ref. group first", when the age ranges of the customized
reference group and the default reference group contradicts with each other, the customized
reference group will be matched first on the sample analysis and review screens.
9-14
Customizing the Analyzer Software
Standby
Tap the text box "Wait" and enter the waiting time before entering the standby status. The
range allowed is 10 -30 minutes, and the default setting is 15 minutes.
9-15
Customizing the Analyzer Software
This function may also be used to refill reagent inside the fluidic system when a new
container of reagent is loaded.
NOTE
The diluent must be kept still for at least a day after long-term
transportation.
When you have changed the diluent, lyse or rinse, run a background to see
if the results meet the requirement.
Diluent
Lyse
9-16
Customizing the Analyzer Software
Setup
HGB gain
9-17
Customizing the Analyzer Software
Tap "Yes" to save the settings and switch to the corresponding screen. Tap "No" to switch to
the corresponding screen without saving the settings.
9-18
10 Servicing Your Analyzer
10.1 Introduction
Preventive and corrective maintenance procedures are required to keep the analyzer in a
good operating condition. This analyzer provides multiple maintenance functions for this
purpose.
This chapter introduces how to use the provided functions to maintain and troubleshoot your
analyzer.
All the analyzer components and surfaces are potentially infectious, so take
proper protective measures for operation and maintenance.
WARNING
Stop using the instrument when you find any fluid tubing or part filled with
fluid is aging or wearing, and contact the service engineer or your local
distributor to replace.
When the instrument is waiting for service, it is suggested that sterilize and
clean the potentially bio-hazardous parts (instrument surface, sample
probe, etc.), in order to reduce the risk of bio-hazard or other hazards while
transporting or servicing the instrument.
CAUTION
Improper maintenance may damage the analyzer. Operators must follow the
instruction of this Operator's Manual to perform maintenance operations.
Exercise caution to avoid contact with the sharp sample probe when
performing maintenance.
The following table lists the tools that may be used in maintenance.
No. Tools
1. Cross-headed screwdriver
Servicing Your Analyzer
10-2
Servicing Your Analyzer
10.2.1 Maintenance
Tap "Service" > "Maintenance", and select the "Maintenance" tab to enter the following
screen.
Unclogging
Unclogging includes zapping and flushing the aperture. When clog error us reported, you
should unclog the aperture.
The unclogging procedures are:
1. Tap the "Unclog" button to start unclogging.
2. When the progress ends, a message will display indicating "Maintaining
finished!".
3. Do the above procedures to continue unclogging aperture if necessary. If the
error persists, perform probe cleanser maintenance of the related channels.
Overall Soak
You should perform the overall soaking procedure with the probe cleanser when:
background results are outside the specified range, QC results abnormal due to long
term idleness of the analyzer; or when other maintenance operations fail to solve the
clog error.
the analyzer shuts down due to abnormal power break-off; probe cleanser maintenance
10-3
Servicing Your Analyzer
The overall soak maintenance procedures with the probe cleanser are as below:
1. Tap the "Overall Soak" button, the following dialog box will display.
4. After aspirating probe cleanser, the analyzer performs probe cleanser soak
automatically, and a progress bar will display indicating the progress.
5. When the progress ends, the following dialog box will display, tap “OK” to close the
dialog box.
10-4
Servicing Your Analyzer
10.2.2 Cleaning
You should clean the following components when:
WBC and (or) HGB background results exceed their limits, perform WBC bath cleaning.
If WBC bath cleaning does not solve the problem, perform WBC probe cleanser
maintenance.
RBC and (or) PLT background results exceed their limits, perform RBC bath cleaning. If
RBC bath cleaning does not solve the problem, perform RBC probe cleanser
maintenance.
Tap "Service" > "Maintenance", and select the "Cleaning" tab to enter the following screen.
Analyzer
Sample probe
WBC bath
RBC bath
10-5
Servicing Your Analyzer
2. When the progress ends, a message will display indicating "Cleaning finished!".
3. Clean other components as per the above procedures if needed.
Pack-up
If the analyzer is not to be used for over 2 weeks, you should perform this procedure.
Do as follows to pack up:
1. Tap "Pack-up", the dialog box "Start pack-up?" will pop up.
2. Tap "Yes" to perform the pack-up procedure. The following dialog box will be
displayed.
3. Remove all reagent pickup tube assemblies according to the prompt, and then Tap
the “OK” button to start emptying the fluidic system.
10-6
Servicing Your Analyzer
5. Place all reagent pickup tube assemblies into the distilled water, and then tap the
“OK” button to start priming.
7. Remove all reagent pickup tube assemblies according to the prompt, and then tap
the “OK” button to start emptying the fluidic system for the second time.
10-7
Servicing Your Analyzer
9. You should turn off the power switch according to the prompt displayed on the
screen.
NOTE
This software can still be used after the pack-up.
10.3 Self-Test
Select “Maintenance”> “Self-Test” from the pull-down menu to enter below screen and
perform system and valve self-tests.
10-8
Servicing Your Analyzer
10-9
Servicing Your Analyzer
10-10
Servicing Your Analyzer
10-11
Servicing Your Analyzer
10-12
Servicing Your Analyzer
10-13
Servicing Your Analyzer
You may view the error info., parameter modification info. and records of daily operation in
the log.
The “Log” screen records all activities of the analyzer. It contributes significantly to searching
for operation history and troubleshooting the analyzer.
NOTE
The oldest record will be overwritten automatically when number of log
records reaches the utmost.
10-14
Servicing Your Analyzer
Exporting logs
1. Tap "Export", the following dialog box will display.
10-15
Servicing Your Analyzer
NOTE
If the status is outside normal range, it will be highlighted with red
background.
10.9.1 Counter
You can check detailed statistics on the screen, for example, running times of the analyzer
and the valid sample runs.
Viewing details
You may tap the "Detail..." buttons following "Sample Runs", "QC Runs" or "Calibration
Runs" to view the related details.
Print
Tap the "Print" icon to print all information on the screen.
10-16
Servicing Your Analyzer
10.9.3 Voltage
Tap "Status" > "Voltage" in the menu to enter the following screen. You may view the current
voltage of the analyzer.
10-17
Servicing Your Analyzer
10.9.4 Sensor
Tap "Status" > "Sensor" in the menu to enter the following screen. You may view the current
sensor information of the analyzer.
10-18
11 Troubleshooting Your Analyzer
11.1 Introduction
This chapter contains information that is helpful in locating and correcting problems that may
occur during operation of your analyzer.
NOTE
This chapter is not a complete service manual and is limited to problems
that are readily diagnosed and/or corrected by the user of the analyzer.
11-1
Troubleshooting Your Analyzer
Red: fatal error. When this kind of error occurs, the analyzer will stop running
immediately, and any further operation is prohibited.
Orange: error that stops operation. When this kind of error occurs, the analyzer will stop
running immediately.
Blue: error that restricts certain operations. When this kind of error occurs, the analyzer
can still continue with the current operation, but any other operations related to the error
will be restricted.
The name and troubleshooting method of the errors are displayed. Names of the errors are
displayed by the order of their occurrence.
You may tap to select the error, and view its troubleshooting information in the
troubleshooting box. The troubleshooting information of the first error is displayed by default.
Follow the instruction in the dialog box to remove error(s).
11-2
Troubleshooting Your Analyzer
Remove
Tap the "Remove " button to clear all the errors that can be removed automatically. For the
errors that cannot removed automatically, follow the troubleshooting method to solve them.
Tap "Close" to close the dialog box, but the errors will still be displayed in the error info. area
on the screen. Tap the error info. area again, the dialog box will be displayed.
The possible error(s) and the corresponding troubleshooting information are listed below:
11-3
Troubleshooting Your Analyzer
1. Tap "Remove ", and enter the new barcode of the diluent into
the reagent setup dialog box.
Lyse expired 2. After replacing the lyse container, tap "Apply" to prime the lyse.
3. If the problem still persists after replacing the lyse, contact our
customer service department.
1. Empty the waste container or use a new waste container.
2. Tap“Remove” to see if the error can be removed.
Waste container full
3. If the problem still persists, contact our customer service
department.
1. Power off the analyzer directly and contact our customer
Power error
service department.
1. Tap“Remove” to see if the error can be removed.
Syringe module error 2. If the problem still persists, contact our customer service
department.
1. Tap“Remove” to see if the error can be removed.
Aspiration module lift
2. If the problem still persists, contact our customer service
mechanism error
department.
Aspiration module 1. Tap“Remove” to see if the error can be removed.
rotary mechanism 2. If the problem still persists, contact our customer service
error department.
1. Tap“Remove” to see if the error can be removed.
Background abnormal 2. If the problem still persists, contact our customer service
department.
1. Tap“Remove” to see if the error can be removed.
HGB background
2. If the problem still persists, contact our customer service
voltage abnormal
department.
1. Tap“Remove” to see if the error can be removed.
Vacuum pressure
2. If the problem still persists, contact our customer service
abnormal
department.
Diluent temperature 1. Diluent temperature must be within [10℃, 40℃].
goes beyond higher
limit
Diluent temperature 1. Diluent temperature must be within [10℃, 40℃].
goes beyond lower
limit
1. Tap“Remove” to see if the error can be removed.
Preheat bath
2. If the problem still persists, contact our customer service
temperature is too high
department.
1. Tap“Remove” to see if the error can be removed.
Preheat bath
2. If the problem still persists, contact our customer service
temperature is too low
department.
1. Tap“Remove” to see if the error can be removed.
Clogging (WBC) 2. If the problem still persists, contact our customer service
department.
11-4
Troubleshooting Your Analyzer
11-5
Troubleshooting Your Analyzer
11-6
12 Appendices
A Index
Prediluted Sample, 5-8
A
Analysis, 5-10 Q
Aspiration, 3-2 Quality Control, 7-1
B R
Buzzer, B-4 Reagents, 2-11
Recorder, B-4
C Install Paper, 4-9
Calibration, 8-1
Calibrator, 2-11 S
Capillary whole blood samples, 5-8 Self-Test, 5-4
Colorimetric Method, 3-7 Setup, 9-2
Controls, 2-11 Shutdown, 5-18
Software, 9-1
D Standby, 5-16
G T
Graph Review, 6-3 Table Review, 6-2
Touch Screen, B-4
I Transmission, 7-10
Troubleshooting, 11-1
Impedance Method, 3-8
Initial Checks, 5-2
W
Installation, 4-2
Whole Blood Samples, 5-8
L
Login, 5-4
Logs, 10-9
M
Maintenance, 10-2
Manual Calibration, 8-4
P
Parameter flag, 5-15
Parameters, 2-2
A-1
B Specifications
B.1 Classification
According to the CE classification, the BC-20s belongs to In vitro diagnostic medical devices
other than those covered by Annex II and devices for performance evaluation.
B.2 Reagents
Diluent M-30D Diluent
B.4 Parameters
Name Abbreviation Default Unit
9
White Blood Cell count WBC 10 /L
9
Lymphocyte number Lymph# 10 /L
9
Mid-sized Cell number Mid# 10 /L
9
Granulocyte number Gran# 10 /L
Hemoglobin Concentration
Hematocrit HCT %
9
Platelet count PLT ×10 /L
Mean Platelet Volume MPV fL
Platelet Distribution Width PDW /
Plateletcrit PCT %
B.5.2 Throughput
Whole blood mode The throughput shall be no less than 40 samples/hour
(0.0 ~ 999.9) × 10 /L
9
WBC
(0.00 ~ 9.99) × 10 /L
12
RBC
(0 ~ 9999) × 10 /L
9
PLT
≤ 0.2 × 10 / L
9
WBC
≤ 0.02× 10 / L
12
RBC
HGB ≤1g/L
B-2
Appendices
HCT ≤ 0.5 %
≤ 5× 10 / L
9
PLT
B.6.4 Reproducibility
(150 ~ 500) × 10 / L
9
≤ 4.0%
PLT
9
(100~149)× 10 / L ≤ 5.0%
B.6.5 Carryover
Parameter Carryover
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.5%
PLT ≤1.0%
B-3
Appendices
RBC ≤±6%
HGB ≤±7%
PLT ≤±15%
B.7.2 Indicator
The indicator indicates analyzer the status of the analyzer, including on/off, running or
sleeping.
B.7.4 Mouse(Optional)
USB mouse.
B.7.7 Recorder
A thermal recorder is included.
B.7.8 Buzzer
The buzzer sounds when there is an error. Tap the touch screen or when the error is removed,
the buzz alarm will be silent automatically,
B-4
Appendices
B.8 Interfaces
One network interface, built-in network card, network access and TCP/IP compatible.
4 USB ports
B.10 Fuse
WARNING
Only install fuses of specified specification on the system and the
pneumatic unit.
This equipment complies with the emission and immunity requirements of the EN
61326-1:2006 and EN 61326-2-6:2006.
NOTE
Only install fuses of specified specification on the system and the
pneumatic unit.
B-5
Appendices
NOTE
Be sure to use and store the analyzer in the specified environment.
B-6
Appendices
Height
Depth
Width
Analyzer
Width (mm) ≤300
Height (mm) ≤400 (foot pad included)
Depth (mm) ≤410
Weight (Kg) ≤20
B.17 Contraindication
None
B-7
C Communication
The LIS/HIS function of BC-20s enables the communication between the analyzer and the
PC in laboratory through Ethernet, including sending analysis results to and receiving worklist
from PC.
The LIS/HIS communication protocol involved in communication of BC-20s are 15ID and HL7.
For details about the connection control, and the introduction, message definition and
examples, please contact Mindray Customer Service Department or your local distributor.
C-1
P/N: 046-007636-00(2.0)