Ultrasonic Flowmeter D118

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Ultrasonic Flowmeter D118

Ultrasonic Flowmeter D118

About D118

D118 Ultrasonic Flowmeter is a state-of-the-art transit time


ultrasonic flowmeter.

Designed using the latest digital technology and low-voltage


broadband pulse transmission.

While principally designed for full-pipe clean liquid


applications.The instrument is tolerant of liquids with small
amounts of air bubbles or suspended solids found in most
industrial environments.

Comparing with other traditional flowmeter or ultrasonic


flowmeter, it has distinctive features such as high precision,
high reliability, high capability and low cost, other features:

TVT technology designed.


Less hardware components, low voltage broadband pulse
transmission, low consumption power.
Clear, user-friendly menu selections make flowmeter simple
and convenient to use.
Daily, monthly and yearly totalized flow.
Parallel operation of positive, negative and net flow totalizes
with scale factor (span) and 7 digit display, while the output
of totalize pulse and frequency output are transmitted via
relay and open collector.

Applications

1
Ultrasonic Flowmeter D118

Specification

Performance specifications
Flow range ±0.03ft/s ~ ±40ft/s (±0.01m/s ~ ±12m/s)
Accuracy ±0.5% of measured value (for ±1.5ft/s~±40ft/s)
Repeatability 0.15% of measured value
Linerarity ±0.5%
Pipe size 1" to 200"(25mm to 5000mm)
Function specifications

Analog output: 4~20mA, max load 750Ω.


Outputs Pulse output: 0~9999Hz, OCT, (min. and max. frequency is adjustable)
Realy output: SPST, max 1Hz,(1A@125VAC or 2A@30VDC)

Communication RS232&RS485

Standard SD card
SD Card Max record:512days
Record time interval
Power supply 90 to 245 VAC, 48 to 63 Hz. Or 10 to 36VDC
Keypad 22 keys with tactile action
Display 40 character, 2 line (20 X 2) lattice alphanumeric, back lit LCD.

Transmitter:-40℉~140℉(-40℃~60℃)
Temperature
Transducer:-40℉~176℉(-40℃~80℃,standard)
Humidity Up to 99% RH,non-condensing
Physical specifications
Transmitter NEMA 4X (IP65), Die-cast aluminum

Encapsulated design
Transducer double-shielded transducer cable
Standard/maximum cable length:30ft/1000ft(9m/305m)

Transmitter : approximately 2.15kg;


Weight
Transducer:approximately 0.9kg.

2
Ultrasonic Flowmeter D118

Wiring Diagram

AC Power DC Power Relay OCT Analog in Transducer 4-20mA RS232 RS485

Transmitter Dimensions
Transmitter
200mm 98mm

Model:D118

Digital Correlation Transit-Time Flowmeter


170mm

198mm

Rate Velocity Signal Totalizer Tom/Tos Diag.

rate total start/ Menu


1 2 stop 3

^ /+
cool heat temp
cap.
4 cap. 5 6
flow sound signal
vel. Enter
vel.
7 8 9
^
cal.
.
zero
0
diag.
< /-

Transducer
56mm
24mm
25mm
32mm

MODEL:D118 Transducer
ANBIENT TEMPERATURE(MAX):80℃
CERTIFICATION:Ex ia II BT6
ENCLOSURE:IP68

FLOW

3
Ultrasonic Flowmeter D118

Transducer Installation Methods

V method measuring pipe size : 25mm-400mm

Side View Section


Transducer
Upstream Transducer

Flow
Downstream Transducer Pipe Strap

Top View

Flow

Transducer space

Z method measuring pipe size: 100mm-3000mm

Side View Section


Transducer
Upstream Transducer

w
Flo Downstream Transducer Pipe Strap
Top View

Flow

Transducer space

4
Ultrasonic Flowmeter D118

Installation Site Selection

When selecting a measurement


site, it is important to select an Straight length of upstream piping Straight length of downstream piping
area where the fluid flow profile
is fully developed to guarantee a
highly accurate measurement. 90° Bend
Use the following guidelines to
select a proper installation site:
L≥5D

Choose a section of pipe that is L≥1 0 D


always full of liquid, such as a
vertical pipe with flow in the
upward direction or a full
horizontal pipe. Tee
L≥1 0 D
Ensure enough straight pipe L≥50 D
length at least equal to the
figure shown below for the
upstream and downstream
transducers installation. Diffuser
L≥30D
L≥5 D
Ensure that the pipe surface
temperature at the measuring
point is within the transducer
temperature limits. Reduce
L≥10 D L≥5 D
Consider the inside condition of
the pipe carefully. If possible,
select a section of pipe where
the inside is free of excessive
Valve
corrosion or scaling.
L≥3 0 D L≥1 0
D

Vertical
L≥5D
L≥10D

L≥5 0 D

5
Ultrasonic Flowmeter D118

Ordering Information

Description
Digital Correlation Transit Time Flowmeter
Installation method: Wall mount
2G SD card high memory data logging, maximum memorize 512 days data.
Flow Range: 0.03 ~ ±40ft/s (0.01~ ±12 m/s)
Accuracy: ±0.5% of measured value
Repeatability: 0.15%
D118 Pipe Size Range: 1″~200″(25mm ~ 5000mm)
Keyboard:16 (4×4) touch keys
Display: 20*2, alphanumeric, backlit LCD
Power supply: 90-250VAC, 48-63 Hz or 10-36V DC
Transmitter enclosure: IP65, die-cast aluminum machined enclosure
Output: 4~20mADC, OCT pulse output, relay output
Communication: RS-232 / RS-485 terminal Modbus Protocol
Output
1 4-20mA, OCT pulse output, relay output, RS-232 / RS-485
4-20mA, OCT pulse output, relay output, RS-232 / RS-485, RTD temperature
2 input
Transmitter enclosure area classification
1 IP65, die-cast aluminum machined enclosure
2 Explosion-proof enclosure , Ex dia II BT4
Type of transducers
C010 Clamp on transducer, Operating temperature:-40℉ ~ + 176℉(-40℃ ~ + 80℃)
CH020 High temperature Clamp on transducer:32℉ ~ + 302℉(0℃ ~ + 150℃)
W210 Insertion transducer, Operating temperature: - 40℉ ~ + 176℉( - 40℃ ~ + 80℃)
WH101 High temperature Insertion transducer:32℉ ~ + 302℉(0℃ ~ + 150℃)
Transducer Cable Length
030 Standard 30ft (9m)

xxx Maximum lengthen to 305m(1000ft), per 5m is a lengthen unit.

Type of Temperature sensor


PT1000 PT1000 Temperature sensor
Standard Model: D118-1-1-C010-030
Description: Standard enclosure with Clamp-on transducers, RS232/RS485, 9m cable.

6
Ultrasonic Flowmeter D118

Product Line

D116 Dedicated Ultrasonic Flowmeter P116 Portable Ultrasonic Flowmeter

Accuracy: Accuracy:
±1% ±1%
Flow range: Flow range:
0.03 ~±16ft/s 0.03 ~ ±40ft/s
Pipe Size Range: Pipe Size Range:
1″~48″ 1″~48″

D118 Dedicated Ultrasonic Flowmeter P118i Portable Ultrasonic Flowmeter

Accuracy: Accuracy:
±0.5% ±0.5%
Flow range: Flow range:
0.03 ~ ±40ft/s 0.03 ~ ±40ft/s
Pipe Size Range: Pipe Size Range:
1″~200″ 0.6″~240″

D118i Dedicated Ultrasonic Flowmeter Application

Accuracy:
Prefect performance in single liquid medium.
±0.5%
Eg:Water,Pure water,Beer,Oil, etc.
Flow range:
0.03~±40ft/s
Pipe Size Range:
1''~200''

Remark:The above mode choose doesn’t including the spool piece,it is for customization.