SR20 Manual v1814
SR20 Manual v1814
SR20 Manual v1814
Thermal Sensors
Voltage output U V
Sensitivity S V/(W/m2)
Sensitivity at reference conditions S0 V/(W/m2)
Temperature T °C
Electrical resistance Re Ω
Solar irradiance E W/m2
Solar radiant exposure H W∙h/m2
Time in hours h h
Subscripts
Not applicable
SR20 measures the solar radiation received by a plane surface, in W/m2, from a 180o
field of view angle. SR20 enables you to attain the highest measurement accuracy and
excels in demanding applications.
The measured quantity, expressed in W/m2, is called “hemispherical” solar radiation.
SR20 pyranometer can be employed outdoors under the sun, as well as indoors with
lamp-based solar simulators. Its orientation depends on the application and may be
horizontal, tilted (for plane of array radiation) or inverted (for reflected radiation). In
combination with the right software, also sunshine duration may be measured.
Using SR20 is easy. It can be connected directly to commonly used data logging
systems. The irradiance, E, in W/m2 is calculated by dividing the SR20 output, a small
voltage U, by the sensitivity S. The sensitivity is provided with SR20 on its calibration
certificate.
SR20’s low temperature dependence makes it an ideal candidate for use under very cold
and very hot conditions. The temperature dependence of every individual instrument is
tested and supplied as a second degree polynomial. This information can be used for
further reduction of temperature dependence during post-processing. In case the
sensitivity is corrected for the instrument body temperature, the optional measurement
equation becomes:
SR20 is equipped with an internal temperature sensor. This can be either a Pt100 (T1
version) or a 10 kΩ thermistor (T2 version), as ordered. To calculate temperature in
degrees Celsius from resistance in Ohms, Formula 8.10.1 or 8.10.2 can be used. See the
dedicated chapter in the appendix of this manual for these equations.
The ASTM E2848 “Standard Test Method for Reporting Photovoltaic Non-Concentrator
System Performance” (issued end 2011) confirms that a pyranometer is the preferred
instrument for PV system performance monitoring. SR20 pyranometer complies with the
requirements of this standard. For more information see our pyranometer selection
guide.
WMO has approved the “pyranometric method” to calculate sunshine duration from
pyranometer measurements in WMO-No. 8, Guide to Meteorological Instruments and
Methods of Observation. This implies that SR20 may be used, in combination with
appropriate software, to estimate sunshine duration. This is much more cost-effective
than using a dedicated sunshine duration sensor. Ask for our application note.
SR20’s output is analogue. Model SR20-D2 offers two other types of commonly used
irradiance outputs: digital via Modbus RTU over 2-wire RS-485 and analogue 4-20 mA
output (current loop).
This user manual covers SR20 use. Specifications of model SR20-D2, the digital
secondary standard pyranometer with Modbus RTU and 4-20 mA output, differ
from those of SR20. For SR20-D2 use, please consult the SR20-D2 user manual.
• pyranometer SR20
• sun screen
• cable of the length as ordered
• calibration certificate matching the instrument serial number
• product certificate matching the instrument serial number (including temperature
response and directional response test)
• any other options as ordered
A quick test of the instrument can be done by using a simple hand held multimeter and a
lamp.
1. Check the electrical resistance of the sensor between the green (-) and white (+) wire.
Use a multimeter at the 1000 Ω range. Measure the sensor resistance first with one
polarity, than reverse the polarity. Take the average value. The typical resistance of the
wiring is 0.1 Ω/m. Typical resistance should be the typical sensor resistance of 100 to
200 Ω plus 1.5 Ω for the total resistance of two wires (back and forth) of each 5 m.
Infinite resistance indicates a broken circuit; zero or a low resistance indicates a short
circuit.
2. Check if the sensor reacts to light: put the multimeter at its most sensitive range of
DC voltage measurement, typically the 100 x 10-3 VDC range or lower. Expose the sensor
to a strong light source, for instance a 100 W light bulb at 0.1 m distance. The signal
should read > 2 x 10-3 V now. Darken the sensor either by putting something over it or
switching off the light. The instrument voltage output should go down and within one
minute approach 0 V.
3. Remove the sun screen, (see chapter on installation of the sun screen). Inspect the
bubble level.
4. Inspect the instrument for any damage.
5. Inspect if the humidity indicator is blue. Blue indicates dryness. The colour pink
indicates it is humid: in the latter case replace the desiccant (see chapter on
maintenance).
3 4
5
11
8
10 9
• a thermal sensor with black coating. It has a flat spectrum covering the 200 to 50000
x 10-9 m range, and has a near-perfect directional response. The coating absorbs all
solar radiation and, at the moment of absorption, converts it to heat. The heat flows
through the sensor to the sensor body. The thermopile sensor generates a voltage
output signal that is proportional to the solar irradiance.
• a glass dome. This dome limits the spectral range from 285 to 3000 x 10-9 m (cutting
off the part above 3000 x 10-9 m), while preserving the 180 ° field of view angle.
Another function of the dome is that it shields the thermopile sensor from the
environment (convection, rain).
• a second (inner) glass dome: For a secondary standard pyranometer, two domes are
used, and not one single dome. This construction provides an additional “radiation
shield”, resulting in a better thermal equilibrium between the sensor and inner dome,
compared to using a single dome. The effect of having a second dome is a strong
reduction of instrument offsets.
From second class to first class and from first class to secondary standard, the achievable
accuracy improves by a factor 2.
1 solar radiation
0,8
pyranometer
response
0,6
0,4
0,2
0
100 1000 10000
wavelength [x 10-9 m]
Figure 2.2 Spectral response of the pyranometer compared to the solar spectrum. The
pyranometer only cuts off a negligible part of the total solar spectrum.
4%
North
Deviation from ideal cosine behaviour [%]
2%
East
South
0%
0 20 40 60 80
West
-2%
ISO secondary
standard
directional
response limit
-4%
SR20 is a pyranometer of the highest category in the ISO 9060 classification system:
secondary standard. It measures the solar radiation received by a plane surface from a
180o field of view angle. This quantity, expressed in W/m2, is called “hemispherical” solar
radiation. Working completely passive, using a thermopile sensor, SR20 generates a
small output voltage proportional to this flux. It can only be used in combination with a
suitable measurement system.
SR20 has an onboard heater and a temperature sensor. Heating the sensor, measuring
the body temperature and using the correction of the temperature response, all
contribute to the dependability and accuracy of the measurement. However, also when
not using these features, SR20 still complies with the secondary standard requirements.
The instrument should be used in accordance with the recommended practices of ISO,
IEC, WMO and ASTM.
*For the exact definition of pyranometer ISO 9060 specifications see the appendix.
65
Ø 150
M5 (2x) M6
85
Table 4.1.1 Standards for pyranometer classification. See the appendix for definitions of
pyranometer specifications, and a table listing the specification limits.
Table 4.2.1 Standards with recommendations for instrument use in solar radiation
measurement
Table 4.3.1 Standards with recommendations for instrument use in sunshine duration
measurement
WMO
SR20 is very well applicable in outdoor PV system performance testing. See also model
SR20-D2 “digital secondary standard pyranometer with Modbus RTU and 4-20 mA
output” and SR12 “first class pyranometer for solar energy test applications”.
IEC 61724; Photovoltaic system performance ASTM 2848-11; Standard Test Method for
monitoring – guidelines for measurement, data Reporting Photovoltaic Non-Concentrator
exchange and analysis System Performance
Mechanical mounting / thermal insulation preferably use connection by bolts to the bottom plate
of the instrument. A pyranometer is sensitive to
thermal shocks. Do not mount the instrument with the
body in direct thermal contact to the mounting plate
(so always use the levelling feet also if the mounting
is not horizontal), do not mount the instrument on
objects that become very hot (black coated metal
plates).
Instrument mounting with one bolt 1 x M6 bolt at the centre of the instrument,
connection from below under the bottom plate of the
instrument.
SR20’s sun screen can be installed and removed by using the dedicated thumb screw.
See item 2 of the drawing below. The thumb screw can be turned without tools for
fixation or loosening of the sun screen, as visualised below. Once the thumb screw has
turned the sun screen loose, the screen can be lifted off manually. After removal the user
may inspect the bubble level, item 10 of the drawing, and remove the cable / connector,
item 11.
3 4
5
2
7
11
8
10 9
Table 5.3.1 The electrical connection of SR20 versions T1 and T2. The heater is not
necessarily used. The temperature sensor is not necessarily used.
Figure 5.3.1 Electrical diagram of the internal wiring of SR20. The shield is connected to
the sensor body.
The selection and programming of dataloggers is the responsibility of the user. Please
contact the supplier of the data acquisition and amplification equipment to see if
directions for use with the SR20 are available.
In case programming for similar instruments is available, this can typically also be used.
SR20 can usually be treated in the same way as other thermopile pyranometers.
Pyranometers usually have the same programming as heat flux sensors.
Table 5.4.1 Requirements for data acquisition and amplification equipment for SR20 in
the standard configuration
Capability for the data logger or the to store data, and to perform division by the sensitivity to
software calculate the solar irradiance.
E = U/S (Formula 0.1)
Open circuit detection open-circuit detection should not be used, unless this is done
(WARNING) separately from the normal measurement by more than 5
times the sensor response time and with a small current
only. Thermopile sensors are sensitive to the current that is
used during open circuit detection. The current will generate
heat, which is measured and will appear as an offset.
The requirements for a measurement with a pyranometer may be expressed by the user
as:
It is important to realise that the uncertainty of the measurement is not only determined
by the instrument but also by the way it is used.
See also ISO 9060 note 5. In case of pyranometers, the measurement uncertainty as
obtained during outdoor measurements is a function of:
Therefore, ISO 9060 says, “statements about the overall measurement uncertainty under
outdoor conditions can only be made on an individual basis, taking all these factors into
account”.
* defined at Hukseflux as all factors outside the instrument that are relevant to the
measurement such as the cloud cover (presence or absence of direct radiation), sun
position, the local horizon (which may be obstructed) or condition of the ground (when
tilted). The environmental conditions also involve the question whether or not the
measurement at the location of measurement is representative of the quantity that
should be measured.
A measurement is reliable if it measures within required uncertainty limits for most of the
time. We distinguish between two causes of unreliability of the measurement:
• related to the reliability of the pyranometer and its design, manufacturing, calibration
(hardware reliability).
• related to the reliability of the measurement uncertainty (measurement reliability),
which involves hardware reliability as well as condition of use.
In many situations there is a limit to a realistically attainable accuracy level. This is due
to conditions that are beyond control once the measurement system is in place. Typical
limiting conditions are:
• dome fouling by deposition of dust, dew, rain or snow. Fouling results in undefined
measurement uncertainty (sensitivity and directional error are no longer defined).
This should be solved by regular inspection and cleaning.
• sensor instability. Maximum expected sensor aging is specified per instrument as its
non-stability in [% change / year]. In case the sensor is not recalibrated, the
uncertainty of the sensitivity gradually will increase. This is solved by regular
recalibration.
• moisture condensing under pyranometer domes resulting in a slow change of
sensitivity (within specifications). This is solved by regular replacement of desiccant
or by maintenance (drying the entire sensor) in case the sensor allows this. For non-
serviceable sensors like most second class pyranometers, this may slowly develop
into a defect. For first class and secondary standard models (for instance model SR11
first class pyranometer and SR20 secondary standard pyranometer) extra desiccant
(in a set of 5 bags in an air tight bag) is available.
• the use of redundant instruments allows remote checks of one instrument using the
other as a reference, which leads to a higher measurement reliability.
• in PV system performance monitoring, in addition to instruments measuring in the
plane of array, horizontally placed instruments are used for the measurement of
global radiation. Global irradiance data enable the user to compare the local climate
and system efficiency between different sites. These data can also be compared to
measurements by local meteorological stations.
Dependability is not only a matter of reliability but also involves the reaction to
problems; if the processing time of service and repairs is short, this contributes to the
dependability.
Hukseflux pyranometers are designed to allow easy maintenance and repair. The main
maintenance actions are:
• replacement of desiccant
• replacement of cabling
1) The formal evaluation of uncertainty should be performed in accordance with ISO 98-3
Guide to the Expression of Uncertainty in Measurement, GUM.
2) The specifications of the instrument according to the list of ISO 9060 classification of
pyranometers and pyrheliometers are entered as limiting values of possible errors, to be
analysed as type B evaluation of standard uncertainty per paragraph 4.3.7. of GUM. A
priori distributions are chosen as rectangular.
3) A separate estimate has to be entered to allow for estimated uncertainty due to the
instrument maintenance level.
4) The calibration uncertainty has to be entered. Please note that Hukseflux calibration
uncertainties are lower than those of alternative equipment. These uncertainties are
entered in measurement equation (equation is usually Formula 0.1: E = U/S), either as
an uncertainty in E (zero offsets, directional response) in U (voltage readout errors) or
in S (tilt error, temperature dependence, calibration uncertainty).
5) In uncertainty analysis for pyranometers, the location and date of interest is entered.
The course of the sun is then calculated, and the direct and diffuse components are
estimated, based on a model; the angle of incidence of direct radiation is a major factor
in the uncertainty.
6) In uncertainty analysis for modern pyrheliometers: tilt dependence often is so low that
one single typical observation may be sufficient.
7) In case of special measurement conditions, typical specification values are chosen.
These should for instance account for the measurement conditions (shaded / unshaded,
ventilated/ unventilated, horizontal / tilted) and environmental conditions (clear sky /
cloudy, working temperature range).
8) Among the various sources of uncertainty, some are “correlated”; i.e. present during
the entire measurement process, and not cancelling or converging to zero when
averaged over time; the off-diagonal elements of the covariance matrix are not zero.
Paragraph 5.2 of GUM.
9) Among the various sources of uncertainty, some are “uncorrelated”; cancelling or
converging to zero when averaged over time; the off-diagonal elements of the covariance
matrix are zero. Paragraph 5.1 of GUM.
10) Among the various sources of uncertainty, some are “not included in analysis”; this
applies for instance to non-linearity for pyranometers, because it is already included in
the directional error, and the spectral response for pyranometers and pyrheliometers
because it is already taken into account in the calibration process.
New calibration procedures were developed in close cooperation with PMOD World
Radiation Center in Davos, Switzerland. The latest calibration method results in an
uncertainty of the sensitivity of less than 1.2 %, compared to typical uncertainties of
higher than 1.7 % for this pyranometer class. See the appendix for detailed information
on calibration hierarchy.
SR20 can measure reliably at a low level of maintenance in most locations. Usually
unreliable measurements will be detected as unreasonably large or small measured
values. As a general rule this means that regular visual inspection combined with a
critical review of the measured data, preferably checking against other measurements, is
the preferred way to obtain a reliable measurement.
Table 7.1.1 Recommended maintenance of SR20. If possible the data analysis and
cleaning (1 and 2) should be done on a daily basis.
The sensor Check the electrical resistance of the sensor between the green
does not give (-) and white (+) wire. Use a multimeter at the 1000 Ω range. Measure the sensor
any signal resistance first with one polarity, than reverse the polarity. Take the average
value. The typical resistance of the wiring is 0.1 Ω/m. Typical resistance should be
the typical sensor resistance of 100 to 200 Ω plus 1.5 Ω for the total resistance of
two wires (back and forth) of each 5 m. Infinite resistance indicates a broken
circuit; zero or a low resistance indicates a short circuit.
Check if the sensor reacts to light: put the multimeter at its most sensitive range
of DC voltage measurement, typically the 100 x 10-3 VDC range or lower. Expose
the sensor to strong light source, for instance a 100 W light bulb at 1 x 10-1 m
distance. The signal should read > 2 x 10-3 V now. Darken the sensor either by
putting something over it or switching off the light. The instrument voltage output
should go down and within one minute approach 0 V. Check the data acquisition
by applying a 1 x 10-6 V source to it in the 1 x 10-6 V range. Check the condition of
the connectors (on chassis as well as the cable).
The sensor Note that night-time signals may be negative (down to -5 W/m2 on clear windless
signal is nights), due to zero offset a.
unrealistically Check if the pyranometer has clean domes.
high or low. Check the location of the pyranometer; are there any obstructions that could
explain the measurement result.
Check the orientation / levelling of the pyranometer.
Check if the right calibration factor is entered into the algorithm. Please note that
each sensor has its own individual calibration factor, as documented in its
calibration certificate.
Check if the voltage reading is divided by the calibration factor in review of the
algorithm. Check the condition of the wiring at the logger.
Check the cable condition looking for cable breaks. Check the condition of the
connectors (on chassis as well as the cable). Check the range of the data logger;
signal can be negative (this could be out of range) or the amplitude could be out
of range. Check the data acquisition by applying a 1 x 10-6 V source to it in the
1 x 10-6 V range. Look at the output. Check if the output is as expected.
Check the data acquisition by short circuiting the data acquisition input with a 100
Ω resistor. Look at the output. Check if the output is close to 0 W/m2.
The sensor Check the presence of strong sources of electromagnetic radiation (radar, radio)
signal shows Check the condition of the shielding.
unexpected Check the condition of the sensor cable.
variations Check if the cable is not moving during the measurement
Check the condition of the connectors (on chassis as well as the cable)
The outer In case there is a minor layer of moisture that is hardly visible: replace the
dome shows desiccant and wait a few days to see if the situation improves.
internal In case of condensation of droplets: disassemble the instrument and dry out the
condensation. parts.
The inner Arrange to send the sensor back to Hukseflux for diagnosis.
dome shows
internal
condensation
The main idea is that one should look out for any unrealistic values. There are programs
on the market that can semi-automatically perform data screening. See for more
information on such a program http://www.dqms.com.
The sensor cable of SR20 is equipped with a M16 straight connector. In case of cable
replacement, it is recommended to purchase a new cable with connector at Hukseflux.
An alternative is to choose for a Do-it-yourself (DIY) approach; please ask for the DIY
connector assembly guide. In case of cable extension, the user may choose purchasing a
new cable with connector at Hukseflux or extending the existing cable himself. Please
note that Hukseflux does not provide support for DIY connector- and cable assembly.
SR20 is equipped with one cable. Keep the distance between data logger or amplifier and
sensor as short as possible. Cables act as a source of distortion by picking up capacitive
noise. In an electrically “quiet” environment the SR20 cable can be extended without
problem to 100 metres. If done properly, the sensor signal, although small, will not
significantly degrade because the sensor resistance is very low (so good immunity to
external sources) and because there is no current flowing (so no resistive losses).
Connector, cable and cable connection specifications are summarised below.
Table 8.1.1 Preferred specifications for SR20 cable replacement and extension
General replacement please order a new cable with connector at Hukseflux or choose for a DIY
approach. In case of DIY replacement by the user see connector
specifications below and ask for the DIY connector assembly guide
General cable extension please order a new cable with connector at Hukseflux or solder the new
cable conductors and shield to the original sensor cable and make a
connection, using adhesive-lined heat shrink tubing, with specifications for
outdoor use. Always connect shield
Connectors used chassis: M16 panel connector, male thread, 10-pole, HUMMEL AG
7.840.200.000 panel connector, front mounting, short version.
cable: M16 straight connector, female thread, 10-pole. HUMMEL AG
7.810.300.00M straight connector, female thread, for cable 3 to 6 x 10-3
m, special version
Cable 8-wire, shielded, with copper conductors (at Hukseflux 8-wire shielded
cable is used, of which 2 wires are used for signal transmission, 2 for
heating and 2 to 4 for the temperature sensor)
Length cables should be kept as short as possible, in any case the total cable
length should be less than 100 m
tooling required for bottom plate fixation and removal hex key 2.5 mm
NOTE: Outer dome, level and sensor of SR20 cannot be supplied as spare parts. In
case of possible damage to the SR20, after repair the instrument must be tested to
verify performance within specification limits. This is required by ISO 9060. Testing
involves verification of the directional response after dome, thermal sensor and level
replacement and verification of the temperature response after thermal sensor
replacement.
Both ISO and ASTM have standards on instrument classification and methods of
calibration. The World Meteorological Organisation (WMO) has largely adopted the ISO
classification system.
ISO 9846:1993 Solar energy -- Calibration of ASTM G167 - 05 Standard Test Method for
a pyranometer using a pyrheliometer Calibration of a Pyranometer Using a
Pyrheliometer
ISO 9847:1992 Solar energy -- Calibration of ASTM E 824 -10 Standard Test Method for
field pyranometers by comparison to a Transfer of Calibration from Reference to Field
reference pyranometer Radiometers
ISO 9059:1990 Solar energy -- Calibration of ASTM E 816 Standard Test Method for
field pyrheliometers by comparison to a Calibration of Pyrheliometers by Comparison to
reference pyrheliometer Reference Pyrheliometers
The World Radiometric Reference (WRR) is the measurement standard representing the
SI unit of irradiance. It was introduced in order to ensure world-wide homogeneity of
solar radiation measurements and is in use since 1980. The WRR was determined from
the weighted mean of the measurements of a group of 15 absolute cavity radiometers
which were fully characterised. It has an estimated accuracy of 0.3 %. The WMO
introduced its mandatory use in its status in 1979.
The world-wide homogeneity of the meteorological radiation measurements is
guaranteed by the World Radiation Center in Davos Switzerland, by maintaining the
World Standard Group (WSG) which materialises the World Radiometric Reference.
See http://www.pmodwrc.ch
Correction from “test conditions of the standard” to “reference conditions” i.e. to normal
incidence and 20 °C:
Using known (working) standard pyranometer properties: directional, non linearity, offsets,
temperature dependence). This correction has an uncertainty; “uncertainty of correction”.
At Hukseflux we also call the working standard pyranometer “standard”.
A pyranometer measures irradiance. The time integrated total is called radiant exposure.
In solar energy radiant exposure is often given in W∙h/m 2.
T↓ apparent surface ºC or K
temperature**
T↑ apparent sky ºC or K
temperature**
SD sunshine duration h
θ is the apparent solar zenith angle θh relative to horizontal, θt relative to a tilted surface
g = global, l = long wave, t = tilted *, h = horizontal*
* distinction horizontal and tilted from Hukseflux,
** T symbols introduced by Hukseflux,
*** contributions of Ed ↓ t and Er↑ t are Ed ↓ and Er↑ both corrected for the tilt angle of the
surface
Table 8.7.1 Classification table for pyranometers per ISO 9060 and WMO.
NOTE: WMO specification of spectral selectivity is different from that of ISO. Hukseflux
conforms to the ISO limits. WMO also specifies expected accuracies. ISO finds this not to
be a part of the classification system because it also involves calibration. Please note that
WMO achievable accuracies are for clear days at mid latitudes and that the uncertainty
estimate does not include uncertainty due to calibration*.
* WMO 7.2.1: The estimated uncertainties are based on the following assumptions: (a)
instruments are well-maintained, correctly aligned and clean; (b) 1 min and 1 h figures
are for clear-sky irradiances at solar noon; (c) daily exposure values are for clear days at
mid-latitudes. WMO 7.3.2.5: Table 7.5 lists the expected maximum deviation from the
true value, excluding calibration errors.
** At Hukseflux the expression ± 1 % is used instead of a range of 2 %.
*** an instrument is subject to conformity testing of its specifications. Depending on the
classification, conformity compliance can be proven either by group- or individual
compliance. A specification is fulfilled if the mean value of the respective test result does
not exceed the corresponding limiting value of the specification for the specific category
of instrument.
Response time time for 95 % response. The time interval between the instant ISO
(95 %) when a stimulus is subjected to a specified abrupt change and the 9060-
instant when the response reaches and remains within specified 1990
limits around its final steady value.The response time is a measure WMO
of the thermal inertia inherent in the stabilization period for a final 1.6.3
reading.
Zero offset a: response to 200 W/m2 net thermal radiation (ventilated). ISO
(200 W/m2 net Hukseflux assumes that unventilated instruments have to specify 9060-
thermal the zero-offset in unventilated – worst case – conditions. 1990
radiation ) Zero offsets are a measure of the stability of the zero-point.
Zero offset a is visible at night as a negative offset, the instrument
dome irradiates in the far infra red to the relatively cold sky. This
causes the dome to cool down. The pyranometer sensor irradiates
to the relatively cool dome, causing a negative offset. Zero offset
a is also assumed to be present during daytime.
Zero offset b: response to 5 K/h change in ambient temperature. ISO
(5 K/h in ambient Zero offsets are a measure of the stability of the zero-point. 9060-
temperature) 1990
Non-stability percentage change in sensitivity per year. The dependence of ISO
(change per sensitivity resulting from ageing effects which is a measure of the 9060-
year) long-term stability. 1990
Non-linearity percentage deviation from the sensitivity at 500 W/m2 due to the ISO
(100 to 1000 change in irradiance within the range of 100 W/m2 to 1000 W/m2. 9060-
W/m2) Non-linearity has an overlap with directional response, and 1990
therefore should be handled with care in uncertainty evaluation.
Directional the range of errors caused by assuming that the normal incidence ISO
response sensitivity is valid for all directions when measuring from any 9060-
direction a beam radiation whose normal incidence irradiance is 1990
1000 W/m2 . Directional response is a measure of the deviations
from the ideal “cosine behaviour” and its azimuthal variation.
Spectral percentage deviation of the product of spectral absorptance and ISO
selectivity (350 spectral transmittance from the corresponding mean within 350 x 9060-
to 1500 x 10-9 m) 10-9 m to 1500 x 10-9 m. Spectral selectivity is a measure of the 1990
(WMO 300 to spectral selectivity of the sensitivity.
3000 x 10-9 m)
Temperature percentage deviation of the sensitivity due to change in ambient ISO
response temperature within an interval of 50 K the temperature of the 9060-
(interval of 50 K) pyranometer body. 1990
Tilt response percentage deviation from the sensitivity at 0° tilt (horizontal) due ISO
(0° to 90° at to change in tilt from 0° to 90° at 1000 W/m2 irradiance. Tilt 9060-
1000 W/m2) response describes changes of the sensitivity due to changes of 1990
the tilt angle of the receiving surface.
Sensitivity the change in the response of a measuring instrument divided by WMO
the corresponding change in the stimulus. 1.6.3
Spectral range the spectral range of radiation to which the instrument is Hukseflux
sensitive. For a normal pyranometer this should be in the 0.3 to 3
x 10-6 m range. Some pyranometers with coloured glass domes
have a limited spectral range.
Solar energy solar energy is the electromagnetic energy emitted by the sun. Solar energy is
or solar also called solar radiation and shortwave radiation. The solar radiation incident
radiation on the top of the terrestrial atmosphere is called extra-terrestrial solar radiation;
97 % of which is confined to the spectral range of 290 to 3 000 x 10-9 m. Part of
the extra-terrestrial solar radiation penetrates the atmosphere and directly
reaches the earth’s surface, while part of it is scattered and / or absorbed by the
gas molecules, aerosol particles, cloud droplets and cloud crystals in the
atmosphere. The former is the direct component, the latter is the diffuse
component of the solar radiation. (ref: WMO, Hukseflux)
Hemispherical solar radiation received by a plane surface from a 180° field of view angle (solid
solar radiation angle of 2 π sr).(ref: ISO 9060)
Global solar the solar radiation received from a 180° field of view angle on a horizontal
radiation surface is referred to as global radiation. Also called GHI. This includes radiation
received directly from the solid angle of the sun’s disc, as well as diffuse sky
radiation that has been scattered in traversing the atmosphere. (ref: WMO)
Hemispherical solar radiation received by a horizontal plane surface.
(ref: ISO 9060)
Plane-of-array also POA: hemispherical solar irradiance in the plane of a PV array.
irradiance (ref: ASTM E2848-11 / IEC 61724)
Direct solar radiation received from a small solid angle centred on the sun’s disc, on a given
radiation plane. (ref: ISO 9060)
Terrestrial or radiation not of solar origin but of terrestrial and atmospheric origin and having
Longwave longer wavelengths (3 000 to 100 000 x 10-9 m). In case of downwelling El ↓ also
radiation the background radiation from the universe is involved, passing through the
”atmospheric window”. In case of upwelling El ↑, composed of long-wave
electromagnetic energy emitted by the earth’s surface and by the gases, aerosols
and clouds of the atmosphere; it is also partly absorbed within the atmosphere.
For a temperature of 300 K, 99.99 % of the power of the terrestrial radiation has
a wavelength longer than 3 000 x 10-9 m and about 99 per cent longer than
5 000 x 10-9 m. For lower temperatures, the spectrum shifts to longer
wavelengths. (ref: WMO)
World measurement standard representing the Sl unit of irradiance with an uncertainty
Radiometric of less than ± 0.3 % (see the WMO Guide to Meteorological Instruments and
Reference Methods of Observation, 1983, subclause 9.1.3). The reference was adopted by
(WRR) the World Meteorological Organization (WMO) and has been in effect since 1 July
1980. (ref: ISO 9060)
Albedo ratio of reflected and incoming solar radiation. Dimensionless number that varies
between 0 and 1. Typical albedo values are: < 0.1 for water, from 0.1 for wet
soils to 0.5 for dry sand, from 0.1 to 0.4 for vegetation, up to 0.9 for fresh snow.
Angle of angle of radiation relative to the sensor measured from normal incidence (varies
incidence from 0° to 90°).
Zenith angle angle of incidence of radiation, relative to zenith. Equals angle of incidence for
horizontally mounted instruments
Azimuth angle angle of incidence of radiation, projected in the plane of the sensor surface.
Varies from 0° to 360°. 0 is by definition the cable exit direction, also called
north, east is + 90°. (ASTM G113-09)
Sunshine sunshine duration during a given period is defined as the sum of that sub-period
duration for which the direct solar irradiance exceeds 120 W/m2. (ref: WMO)
SR20 is equipped with an internal temperature sensor. This can be either a Pt100 (SR20-
T1 version) or a 10 kΩ thermistor (SR20-T2 version), as ordered.
Both versions require the user to measure the resistance of the temperature sensor and
convert this value to temperature. Many dataloggers have built-in functions to perform
such a conversion. In case the user wishes to calculate temperature (in degrees Celsius)
from resistance (in Ohms) himself, there are two distinct procedures:
SR20-T1
To convert resistance in Ω to temperature in °C, one can use the following equation:
R Pt100
−𝐴 + �𝐴2 − 4𝐵 �1 − � (Formula 8.10.1)
100
𝑇=
2𝐵
with RPt100 the resistance in Ω, T the temperature in °C, A and B the Pt100 coefficients
A = 3.908 x 10-3
B = -5.775 x 10-7
SR20-T2
1
𝑇= − 273.15 (Formula 8.10.2)
𝛼 + 𝛽 ln(R thermistor ) + 𝛾 ln(R thermistor )3
with Rthermistor the thermistor resistance in Ω, T the temperature in °C, α, β and γ the
Steinhart-Hart coefficients
α= 1.0295 x 10-3
β= 2.391 x 10-4
γ= 1.568 x 10-7
has been designed to comply and is in conformity with the relevant sections and
applicable requirements of the following standards:
Eric HOEKSEMA
Director
Delft
20 April, 2016
Hukseflux Thermal Sensors B.V. reserves the right to change specifications without notice.