GE E8 User Manual
GE E8 User Manual
GE E8 User Manual
Technical
Publications
2302323-100
Revision 5
User Manual
Operating Documentation
Regulatory Requirement
The E8C probe complies with regulatory requirements of the following European Direc-
tive 93/42/EEC concerning medical devices for the LOGIQ™3, LOGIQ™5, LOGIQ™7,
LOGIQ™9 and Vivid 7 systems.
0459
The E8C-RS probe complies with regulatory requirements of the following European
Directive 93/42/EEC concerning medical devices for the LOGIQ Book only.
0459
GE Medical Systems
REVISION REVISION
PAGE NUMBER NUMBER PAGE NUMBER NUMBER
Please verify that you are using the latest revision of this document. Information
pertaining to this document is maintained on GPC (GE Medical Systems Global
Product Configuration). If you need to know the latest revision, contact your
distributor, local GE Sales Representative or in the USA call the GE Ultrasound
Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
Table of Contents
Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
High level disinfection/sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Outside Dimensions and Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Transvaginal Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cleaning and Disinfecting the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Probe Cleaning Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Disinfecting probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
High Level Probe Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Storing the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Instructions
General
Composition
1. Probe body, including:
• Head-Assy
• Grip
• Carrying Case
2. Operation Manual
3. Filling remover tool
Specifications
Applicable System
E8C Probe LOGIQ™3
LOGIQ™5
LOGIQ™7
LOGIQ™9
Vivid 7
E8C-RS Probe LOGIQ Book
Frequency 6.5MHz
323
149 129
FOV
unit:mm
Operation
How to install
Before installing the attachment, the filling must be removed from the
probe head where the biopsy guide attachment will be installed. This is
only for first time installation. See Figure 2.
Probe head
1. Fill the probe sheath with acoustic coupling gel as shown in Figure 3.
Probe sheath
ACOUSTIC
COUPLING GEL
2. Place the gel-filled probe sheath on the transducer face and bring it up
the probe body so that the adjustment knob of the probe is completely
covered with the probe sheath as shown in Figure 4.
Probe sheath
B
Patient’s Foot
A
B C
Patient’s Head
1/2 FOV
Display on Monitor
A
FOV
The probe is driven with electrical energy that can injure the patient or
user if live internal parts are contaced by conductive solution:
CAUTION DO NOT immerse the probe into any liquid beyond the level indicated
by the immersion level in Figure 6 on page-9. Never immerse the
probe connector or probe adaptors in any liquid.
1. After each use, disconnect the probe from the ultrasound console and
remove all coupling gel from the probe by wiping with the soft cloth
and rinsing with flowing water.
2. Wash the probe with mild soap in lukewarm water(below 80° F[26° C]
is recommennded). Scrub the probe as needed using a soft sponge,
gauze, or cloth to remove all visible residue from the probe surface.
Prolonged soaking or scrubbing with a soft bristle brush (such as a
tooth brush)may be necessary if material has dried onto the probe
surface.
3. Rinse the probe with enough clean potable water to remove all visible
soap residue.
4. Air dry or dry the probe with a soft cloth.
CAUTION DO NOT clean any portion of the probe with methanol, ethanol,
isopropanol, or any other alchol based cleaner.Such substances can
cause irreparable damage to the probe.
Disinfecting probes
ly
M ly The level of disinfection is directly related to the duration of contact with
th
k
er
ily
ee
on
th
Da
W
RELIEF
1. After cleaning and disinfection, store the probe in it’s carrying case for
protection.
2. Store the probe even when traveling for short distances in its carrying
case or in the system probe holder.
3. The probe should not be stored or transported within the parameters
outlined below.
Storage Transport
Who To Contact
Japan TEL:0120-055-919
Customer Call Center FAX: (81) 426-48-2905
Manufacturer
GE Medical System (China) Co., Ltd.
No. 19, Changjiang Road
WuXi National Hi-Tech Development Zone
Jiangsu, P.R. China 214028
TEL: +86 510 5225888; FAX: +86 510 5226688