Quo Test User Manual
Quo Test User Manual
Quo Test User Manual
User Manual
info@ekfdiagnostics.com
www.ekfdiagnostics.com
An ISO 13485:2003/2012 Certified Company
1) Connect the analyzer to the mains supply. The mains outlet must have an earth
connection.
2) If the system is not going to be used for an extended period of time, unplug
the analyzer from the mains supply.
3) Do not place the system in liquid, or put it where it could fall into liquid.
If the equipment becomes wet, unplug from the mains supply before touching it.
4) Use the system only for the purpose described in this User Manual.
5) Do not use accessories which are not supplied or recommended by the
manufacturer.
6) Do not use the system if it is not working properly or if it has suffered any
damage, for example;
a) damage to the flexible supply cord or its plug.
b) damage caused by dropping the system.
c) damage caused by dropping the system into liquid or splashing liquid on to it.
7) Do not let the equipment or the flexible cords come into contact with surfaces
which are too hot to touch.
8) Do not block air openings or place the system on a soft surface. Keep air
openings free from debris.
9) Do not place anything on top of the system.
10) Do not drop or put anything into any opening in the equipment, unless
instructed by this User Manual.
11) Do not use the system where aerosol sprays are used, or where oxygen is
administered.
12) Do not use the system out of doors.
13) Do not use the system in close proximity to sources of strong electromagnetic
radiation, as these may interfere with the proper operation of the analyzer. An
evaluation of the electromagnetic environment should be undertaken before
using the system.
Unpacking
Unpack the system and check that the following components are present in the pack.
If there is anything missing contact your local distributor or customer support.
• Quo-Test Analyzer
• Barcode Scanner
• Mains Power Cable
• Power Supply
• User Manual
It is recommended to keep the packaging to transport your analyzer. If the optional
printer has been purchased, it will be delivered in a separate box.
Display Screen
Door
Key Pad
Slide
Test Chamber
Measurement Principle
The arrangement of the two light sources and associated detectors allows the
Quo-Test Analyzer to be used as a dual channel spectrophotometer / fluorimeter.
This allows the Quo-Test Analyzer to be used either to measure changes in optical
density as used in Immunoturbidimetric assays or changes in light transmission from
specific fluoriphores as they interact with analytes of interest e.g. HbA1c.
WEEE
USB Port Directive Mark
Model No.
Serial No.
Manufacturer
CE Mark
Connect the power supply to the analyzer and plug the supply cord into the
mains supply (100-240v AC 50-60Hz). Ensure that the supply has a protective
earth connection.
Reported Units
The Quo-Test A1C assay reports results in up to two user selectable units: % DCCT,
mmol/mol IFCC, % JDS, eAG mg/dl or eAG mmol/l.
eAG values are based on a correlation study linking % DCCT to the patient’s average
glucose concentration, resulting in published formula to derive the eAG.
eAG values may differ significantly from a patients glucose level measured at the
same time.
The Quo-Test Analyzer allows the user to select dual reporting (two different
measument units can be displayed) or single reporting. Users should refer to national
guidance when setting up reported units.
% DCCT is the default primary unit, to change the primary reported unit press Change
Primary Units
DCCT
OK Change
Functional Checking
Each time the Quo-Test Analyzer is switched on it performs a series of functional
checks of the system and optics. The analyzer must be powered off and powered on
at least once a week to allow the analyzer to perform the internal self-checks.
Refer to the Troubleshooting Guide and Error Messages section of this User Manual
to diagnose any problems, if they occur.
Correct operation of the system may be confirmed by running Quo-Test A1C Control
samples.
To run a control, read the Instructions For Use supplied with your Quo-Test A1C
Control Kit in combination with the Running a Quality Control section of this User
Manual.
Step 2
Before using a new lot of test cartridges,
scan the calibration barcode that is
printed on the inner flap of the carton,
indicated by the CAL symbol.
Place the barcode scanner over the
calibration barcode and press the
barcode scanner button until you hear
a beep. The analyzer will confirm that
the data has been accepted, along with
displaying the lot number and expiry
date.
2
Step 6
Place the Blood Collector into the free
cavity on the top of the Test Cartridge,
making sure that the Blood Collector is
flush with the top of the Test Cartridge
and is not sticking out.
Note: Do not push the Blood
Collector into the Test Cartridge, as
this may cause an error message to be
displayed instead of a result. Refer to
the Troubleshooting Guide and Error
Messages section of this User Manual
6 for further details.
Running a Test
Step 7
After the Test Cartridge is prepared,
open the door of the analyzer and place
the cartridge into the test chamber,
making sure that the Test Cartridge is
firmly seated. The Test Cartridge must be
inserted into the analyzer within 1 minute
of the blood sample being collected.
Step 9
When the test has finished the result will
be displayed on the screen and printed
on the (optional) printer, if connected.
Open the door and push the red slide
towards the rear of the analyzer. Remove
the used cartridge and dispose of it as
clinical waste.
Upon closing the analyzer door, the
display will return to the home screen
and the analyzer will be ready to run
another sample.
9
Note: See Section 4 - Retrieving a
Result from the Analyzer Memory
When the analyzer has recorded the barcocde a beep will sound and the ID is
displayed on the screen. Confirm the displayed ID matches the barcode scanned. If
you do not have an Operator or Patient ID barcode label available, press the Skip
button to exit the screen. The test result will not be displayed until the requested ID’s
have been scanned or the screens have been exited. If an error occurs, the analyzer
will abort the request to scan the Operator and/or Patient ID if not already completed.
Note: Changing the secondary reported unit (see Section 3 - Report Units) will also
change the secondary reported units on previously run samples held in the memory
of the analyzer, as these are calculated each time they are displayed. Changing
the primary reported unit will not alter the primary units reported on previously run
samples.
Only Quo-Test A1C Controls should be used with the Quo-Test A1C Test.
Please read the Instructions For Use included with each Quo-Test A1C Control Kit to
ensure that the correct procedure is used.
Use the Quo-Test A1C Control Kit according to the quality standards set by the
organisation or laboratory and according to local and national regulations.
If the controls do not give results within the range stated in the Instructions For Use,
contact your local distributor or customer support.
The quality control result is stored in the memory of the analyzer. The lot number of
the Quo-Test A1C Control Kit used is stored under the Patient ID field.
• C leaning may be carried out by wiping the analyzer with a non-abrasive lint
free cloth. Use a mild detergent or commercially available disinfectant, such as
“Virkon®”, if necessary. Do not use bleach (Hypochlorite).
• Always power off the analyzer by unplugging the power supply when performing
any cleaning operation.
• Always power off the analyzer on a weekly basis. Leave off for a minimum of 10
seconds, then the analyzer may be powered on in order to perform the internal
self-checks. To power off the system, the Quo-Test should be in the home screen,
then either switch off at the mains supply or disconnect the power cable from the
power socket on the analyzer.
• Do not clean inside the cartridge test chamber. Contact your local distributor or
customer support for advice.
• If you spill liquid inside the cartridge test chamber, call your local distributor or
customer support for advice.
• If the instrument is stored when not in use, ensure that the mains cord is not liable
to damage or kinking.
• The analyzer is calibrated by the manufacturer and there are no user-serviceable
parts.
Screen shows
Remove cartridge, close the door
‘Remove Old Either a used cartridge has been
and allow the system to return
Cartridge’ when inserted or the test cartridge is
to the home screen. Repeat test
a new cartridge faulty
with a new cartridge.*
is inserted.
Error 08 The stirrer motor has failed Power off the analyzer, wait 10
seconds, and power on again
The internal fan Remove the old cartridge, if
Error 09
has failed present, close the door and allow
the analyzer to return to the home
The x- or z-axis motor has screen*
Error 10/11
jammed
Reagent Out of
Remove cartridge and repeat test
Range The test cartridge is faulty
with a new cartridge*
(Error 102 / 103)
Invalid Result
(2)/(1) A result was not able to be Remove the cartridge and repeat
calculated with a new cartridge*
(Error 106)
High Background
Fluorescence The test cartridge is faulty
(Error 110)
Customer support
+44 (0) 1932 220124
support@quotientdiagnostics.co.uk
Customer Support
Quotient Diagnostics Ltd
Russell House,
Molesey Road
Walton on Thames,
Surrey, KT12 3PE
UK
Phone: +44 (0) 1932 220124
Email: support@quotientdiagnostics.co.uk
Upon receipt of the analyzer, it will be investigated and providing it is within warranty
and the tamper proof quality control label on the rear of the system is intact,
Quotient Diagnostics Ltd will promptly repair or replace the analyzer.
Analyzers that have not been maintained in accordance with the operating
instructions, which have been misused or abused or which are no longer under
warranty will be repaired and returned with a service charge.
Weight 1.3kg
POWER SUPPLY
BATTERY
BARCODE READER
Width of Field 80 mm
Accessories
Interface Keyboard PS2
Supported Barcode
Code 128, Code 39, NW7 / Codabar, 2 of 5 interleaved
Formats
Location Clean, dry, horizontal surface and away from direct sunlight
STORAGE &
TRANSPORT
Accessories
PRINTER (OPTIONAL)
Seiko SLP-440
Label Printer
Part No. 0108-0002
Seiko SLP-650E
New Label Printer Part No. 0108-0035
Power supply provided with printer
Interface Serial
SHOULDER BAG
SYMBOL MEANING
CONTROL Control
Storage Temperature
Manufacturer
SN Serial Number
Printer Connector
USB Connector
Caution
Product Code:
0108-0000
Description:
Quo-Test Analyzer System
info@ekfdiagnostics.com
www.ekfdiagnostics.com
An ISO 13485:2003/2012 Certified Company
IFU0108.EU-EN.02.02 July 2015 New printer Seiko SLP-650E Part No. 0108-0032
replacing Seiko SLP-440 Part No. 0108-0002 and
new power supply
New barcode Scanner T8-W CCD Hand Scanner
minor formatting changes