Abb PH User Guide PDF
Abb PH User Guide PDF
Abb PH User Guide PDF
IM/AX4PH Issue 12
Single and Dual Input Analyzers
for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
ABB
EN ISO 9001:2000
Cert. No. Q 05907
EN 29001 (ISO 9001)
Lenno, Italy Cert. No. 9/90A
Stonehouse, U.K.
0255
Electrical Safety
This equipment complies with the requirements of CEI/IEC 61010-1:2001-2 'Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use'. If the equipment is used in a manner NOT specified by the Company, the protection
provided by the equipment may be impaired.
Symbols
One or more of the following symbols may appear on the equipment labelling:
Warning Refer to the manual for instructions Direct current supply only
Caution Risk of electric shock Alternating current supply only
Protective earth (ground) terminal Both direct and alternating current supply
Earth (ground) terminal
The equipment is protected
through double insulation
The Company
We are an established world force in the design and manufacture of instrumentation for
industrial process control, flow measurement, gas and liquid analysis and environmental
applications.
As a part of ABB, a world leader in process automation technology, we offer customers
application expertise, service and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and
unrivalled service and support.
The quality, accuracy and performance of the Companys products result from over 100 years
experience, combined with a continuous program of innovative design and development to
incorporate the latest technology.
The UKAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants operated
by the Company and is indicative of our dedication to quality and accuracy.
Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for
any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of the
Technical Publications Department.
Health and Safety
To ensure that our products are safe and without risk to health, the following points must be noted:
1. The relevant sections of these instructions must be read carefully before proceeding.
2. Warning labels on containers and packages must be observed.
3. Installation, operation, maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the
information given.
4. Normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure
and/or temperature.
5. Chemicals must be stored away from heat, protected from temperature extremes and powders kept dry. Normal safe handling procedures
must be used.
6. When disposing of chemicals ensure that no two chemicals are mixed.
Safety advice concerning the use of the equipment described in this manual or any relevant hazard data sheets (where applicable) may be
obtained from the Company address on the back cover, together with servicing and spares information.
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 Contents
IM/AX4PH Issue 12 1
Contents
1 Introduction ..................................................................... 2
1.1 System Description .................................................. 2
1.2 PID Control .............................................................. 2
1.3 AX400 Series Analyzer Options ................................ 2
2 Operation ......................................................................... 3
2.1 Powering Up the Analyzer ........................................ 3
2.2 Displays and Controls .............................................. 3
2.2.1 Membrane Key Functions ............................. 3
2.3 Operating Page ........................................................ 6
2.3.1 Single Input pH ............................................. 6
2.3.2 Dual Input pH ............................................... 7
2.3.3 Single Input Redox (ORP) ............................. 8
2.3.4 Dual Input Redox (ORP) ................................ 9
2.3.5 Dual Input pH and Redox (ORP) ................. 10
2.3.6 Wash Function ............................................ 11
3 Operator Views .............................................................. 12
3.1 View Set Points ...................................................... 12
3.2 View Outputs ......................................................... 14
3.3 View Hardware ....................................................... 14
3.4 View Software ........................................................ 15
3.5 View Logbook ........................................................ 16
3.6 View Clock ............................................................. 18
4 Setup ............................................................................. 19
4.1 Sensor Calibration .................................................. 19
4.1.1 Set Buffer Type (pH Only) ........................... 19
4.1.2 Set Up User Defined Buffers (pH Only) ........ 21
4.1.3 Adjust Offset (Redox/ORP Only) .................. 22
4.1.4 Automatic, Single- and
Two-Point Calibration (pH Only) .................. 23
4.1.5 Manual, Single- and
Two-Point Calibration (pH Only) .................. 25
4.1.6 Grab Calibration (pH Only) .......................... 27
5 Programming ................................................................. 28
5.1 Security Code ........................................................ 28
5.2 Configure Display ................................................... 29
5.3 Configure Sensors ................................................. 30
5.4 Configure Diagnostics ............................................ 33
5.5 Configure Alarms ................................................... 34
5.5.1 Wash Cycle Configuration
(Applicable Only to Alarm 3) ........................ 36
5.6 Configure Outputs .................................................. 38
5.7 Configure Clock ..................................................... 40
5.8 Configure Security .................................................. 41
5.9 Configure Logbook ................................................ 41
5.10 Test Outputs and Maintenance .............................. 42
6 Installation .................................................................... 44
6.1 Siting Requirements .............................................. 44
6.2 Mounting ............................................................... 45
6.2.1 Wall-/Pipe-mount Analyzers ....................... 45
6.2.2 Panel-mount Analyzers ............................... 46
6.3 Connections, General ............................................ 47
6.3.1 Relay Contact Protection
and Interference Suppression ..................... 48
6.3.2 Cable Entry Knockouts,
Wall-/Pipe-mount Analyzer ......................... 49
6.4 Wall-/Pipe-mount Analyzer Connections ................ 50
6.4.1 Access to Terminals ................................... 50
6.4.2 Connections ............................................... 51
6.5 Panel-mount Analyzer Connections ....................... 52
6.5.1 Access to Terminals ................................... 52
6.5.2 Connections ............................................... 53
6.6 pH Sensor Systems Connections ......................... 54
6.6.1 Standard ph Systems Connection 2867,
AP100, AP300, 7650/60, TB5, Non-ABB ... 54
6.6.2 Differential ph Systems Connections
Capable of Providing Sensor Diagnostics
(AP200, TBX5) ........................................... 55
7 Calibration ..................................................................... 56
7.1 Equipment Required .............................................. 56
7.2 Preparation ............................................................ 56
7.3 Factory Settings .................................................... 57
8 Simple Fault Finding ..................................................... 62
8.1 Error Messages ..................................................... 62
8.2 Calibration Fail Message or No
Response to pH/Redox Changes .......................... 62
8.3 Checking the Temperature Input ............................ 63
9 Specification ................................................................. 64
Appendix A Buffer Solutions ......................................... 67
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 1 Introduction
2 IM/AX4PH Issue 12
1 Introduction
1.1 System Description
The AX460 single input and AX466 dual input pH/Redox (ORP)
analyzers and associated electrode systems have been
designed for continuous monitoring and control of pH and
Redox (ORP). The electrode system can be standardized to the
analyzer using the built-in calibration facility and a single point
buffering facility provides easy re-calibration after initial
standardization.
The analyzer is available in wall-/pipe-mount or panel-mount
versions with either one or two programmable, pH or Redox
(ORP) input channels, each with its own associated temperature
input channel. When making temperature compensated
measurements, the sample temperature is sensed by a
resistance thermometer (Pt100, Pt1000 or Balco 3K) mounted
in the electrode system.
The analyzer can be configured for, and connected to, either a
standard pH input (single, high impedance input >10
13
) or
differential pH input (dual, high impedance inputs, both >10
13
).
Differential pH input is designed for use with pH electrode
systems that incorporate a solution earth (ground) rod. The
measuring electrode and reference electrode signals are
measured separately using two, high impedance amplifiers and
compared with the solution earth (ground) potential. The
difference between the results is the value used for the pH
measurement.
All models incorporate a wash facility for system cleaning; the
Alarm 3 relay can be configured to control the wash system
either automatically or manually. The relay can be programmed
to deliver either a continuous or pulsed signal to control an
external power supply to a solenoid or pump and the frequency,
duration and recovery time for the wash cycle are also
programmable. During a wash cycle, the analog output value is
held in its pre-cycle condition.
Analyzer operation and programming are performed using five
tactile membrane keys on the front panel. Programmed
functions are protected from unauthorized alteration by a
four-digit security code.
1.2 PID Control
The AX460 single input pH analyzer incorporates Proportional
Integral Derivative (PID) control as standard. Refer to the
PID Control Supplementary User Guide, IM/AX4PID for a full
description and instructions on how to configure and operate
PID control.
1.3 AX400 Series Analyzer Options
Table 1.1 shows the range of configurations that are possible for
the AX400 Series analyzers. The analyzer detects the type of
input board fitted for each input automatically and displays only
the operating and programming frames applicable to that input
board type. If no input board is fitted for a second input
(Sensor B), Sensor B frames are not displayed.
Model Analyzer Description Sensor A Sensor B
AX410 Single Input 2-Electrode Conductivity (0 to 10,000 S/cm) 2-Electrode Conductivity
AX411 Dual Input 2-Electrode Conductivity (0 to 10,000 S/cm) 2-Electrode Conductivity 2-Electrode Conductivity
AX413 Dual Input 2-Electrode Conductivity and 4-Electrode Conductivity 2-Electrode Conductivity 4-Electrode Conductivity
AX416 Dual Input 2-Electrode Conductivity and pH/Redox (ORP) 2-Electrode Conductivity pH/Redox (ORP)
AX418 Dual Input 2-Electrode Conductivity and Dissolved Oxygen 2-Electrode Conductivity Dissolved Oxygen
AX430 Single Input 4-Electrode Conductivity (0 to 2,000 mS/cm) 4-Electrode Conductivity
AX433 Dual Input 4-Electrode Conductivity (0 to 2,000 mS/cm) 4-Electrode Conductivity 4-Electrode Conductivity
AX436 Dual Input 4-Electrode Conductivity and pH/Redox (ORP) 4-Electrode Conductivity pH/Redox (ORP)
AX438 Dual Input 4-Electrode Conductivity and Dissolved Oxygen 4-Electrode Conductivity Dissolved Oxygen
AX450 Single Input 2-Electrode Conductivity (USP) 2-Electrode Conductivity
AX455 Dual Input 2-Electrode Conductivity (USP) 2-Electrode Conductivity 2-Electrode Conductivity
AX456 Dual Input 2-Electrode Conductivity (USP) and pH/Redox (ORP) 2-Electrode Conductivity pH/Redox (ORP)
AX460 Single Input pH/Redox (ORP) pH/Redox (ORP)
AX466 Dual Input pH/Redox (ORP) pH/Redox (ORP) pH/Redox (ORP)
AX468 Dual Input pH/Redox (ORP) and Dissolved Oxygen pH/Redox (ORP) Dissolved Oxygen
AX480 Single Input Dissolved Oxygen Dissolved Oxygen
AX488 Dual Input Dissolved Oxygen Dissolved Oxygen Dissolved Oxygen
Table 1.1 AX400 Series Analyzer Options
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
IM/AX4PH Issue 12 3
2 Operation
2.1 Powering Up the Analyzer
1. Ensure the input sensors are connected correctly.
2. Switch on the power supply to the analyzer. A start-up
screen is displayed while internal checks are performed,
then the Operating Page (Section 2.3) is displayed as the
pH or Redox (ORP) monitoring operation starts.
2.2 Displays and Controls
The display comprises two rows of 4
1
/2 digit, 7-segment digital
displays, that show the actual values of the measured
parameters and alarm set points, and a 6-character dot matrix
display showing the associated units. The lower display line is a
16-character dot matrix display showing operating and
programming information.
2.2.1 Membrane Key Functions
Warning. Ensure all connections are made correctly,
especially to the earth stud see Section 6.3, page 47.
Fig. 2.1 Location of Controls and Displays
0.00
Monitoring pH
0.00
Alarm
LEDs
Display
Lines
Lower
Display Line
Membrane Keys
pH
Units
Menu Key
Sidescroll Key
Downscroll Key
Up Key
Down Key
pH
Fig. 2.2 Membrane Key Functions
B Advancing to Next Page
C Moving Between Frames
D Adjusting and Storing a Parameter Value
E Selecting and Storing a Parameter Choice
A Moving Between Menus
For majority
of Frames
Frame 1
Frame 2
Frame 3
Frame 4
Page 1
Frame 1
Frame 2
Frame 3
Page 2
Advance to
next page
or
Frame 1
Frame 2
Frame 3
Page X
Frame 4
Advance to
next Frame
New value is
stored automatically
Parameter Value Adjust
Parameter X
Y
Z
Select
New value is
automatically stored
Menu 1
Menu 2
Advance to
next menu
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
4 IM/AX4PH Issue 12
Fig. 2.3 Overall Programming Chart
Key
VIEW SETPOINTS VIEW OUTPUTS VIEW HARDWARE VIEW SOFTWARE
A1: Setpoint Analog Output 1 Sensor A Module AX400/2000 Issue
A2: Setpoint Analog Output 2 Sensor B Module
A3: Setpoint Analog Output 3 Option Board
A4: Setpoint Analog Output 4
A5: Setpoint
SECURITY CODE
CONFIG. DISPLAY Set Language Set Temp. Units Set Backlight
LED Backlight Temp. Units English
Use the Sidescroll Key to scroll through the Pages within each Menu
To CONFIG. ALARMS
(see Fig. 2.4)
Use the Downscroll
Key to scroll through
the Frames
within each Page
OPERATING PAGE
Section 2.3, Page 6 Section 3.1, Page 12 Section 3.2, Page 14 Section 3.3, Page 14 Section 3.4, Page 15 Section 3.5, Page 16
Section 5.1, Page 28
Section 5.2, Page 29
Section 3.6, Page 18
Alarms
Errors
Power
Cals
VIEW LOGBOOK
Dual input analyzer only
VIEW CLOCK
Date 01:01:03
Time 12:00
*
Displayed only if Probe Type is set to Redox or ORP
CONFIG.DIAGS
Section 5.4, Page 33
Config. Sensor A
A: pH Glass
A: Ref. Checking
A: Ref. Alarm
Config. Sensor A Config. Sensor B
B: pH Glass
B: Ref. Checking
B: Ref. Alarm
Use the Menu Key
to scroll through
the Menus
CONFIG.SENSORS
B: Probe Type
B: Diff. Input
B: Electrode
B: Temp. Comp
B: Temp. Sensor
B: Preset Temp.
B: Sample Comp.
B: Sample Coeff.
Section 5.3, Page 30
B: Set Min Slope
Config. Sensor B Config. Sensor A
A: Probe Type
A: Diff. Input
A: Temp. Comp
A: Temp. Sensor
A: Preset Temp.
A: Sample Comp.
A: Sample Coeff.
A: Set Min Slope
A: Electrode
A: Enable Cals B: Enable Cals
* *
SENSOR CAL. Cal. User Code Sensor Cal. A
A: Buffer Method
A: Immerse Buf.1
A: Cal. Buffer 1
A: Immerse Buf.2
A: Cal. Buffer 2
A: Calibration
Section 4.1, Page 19
Set Auto Buffers
Buffer Type
Set Buffer 1
Set Buffer 2
A: Slope & Check
Sensor Cal. B
B: Buffer Method
B: Immerse Buf.1
B: Cal. Buffer 1
B: Immerse Buf.2
B: Cal. Buffer 2
B: Calibration
B: Slope & Check
A: Enter Point 1
A: Enter Point 2
B: Enter Point 1
A: Enter Point 5
A: Enter Point 4
A: Enter Point 3
B: Enter Point 2
B: Enter Point 5
B: Enter Point 4
B: Enter Point 3
#### 100% ####
#### 100% ####
#### 100% ####
#### 100% ####
Note. Sensor calibration parameters shown above
are for Automatic, 2-point calibration only. For other
calibration options, refer to Section 4.1.
Available only if option board fitted and
analog features enabled see Section 7.3
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
IM/AX4PH Issue 12 5
Fig. 2.4 Overall Programming Chart (Continued)
CONFIG.ALARMS Config. Alarm 1
A1: Type
A1: Assign
A1: Failsafe
A1: Action
A1: Setpoint
A1: Hysteresis
A1: Delay
Config. Alarm 2
A2: Type
A2: Assign
A2: Failsafe
A2: Action
A2: Setpoint
A2: Hysteresis
A2: Delay
A3: Assign
A3: Failsafe
A3: Action
A3: Setpoint
A3: Hysteresis
A3: Delay
To FACTORY SETTINGS
(see Section 7.3, Page 57)
Config. Alarm 3
A3: Type
Use the Sidescroll Key to scroll through the Pages within each Menu
Section 5.5, Page 34
Use the Downscroll
Key to scroll through
the Frames
within each Page
Config. Alarm 4
A4: Type
A4: Assign
A4: Failsafe
A4: Action
A4: Setpoint
A4: Hysteresis
A4: Delay
Config. Alarm 5
A5: Type
A5: Assign
A5: Failsafe
A5: Action
A5: Setpoint
A5: Hysteresis
A5: Delay
Wash Frequency
Wash Duration
Recovery Period
Wash Mode
*
*
Applicable only
to Alarm 3
CONFIG.LOGBOOK
Section 5.9, Page 41
Logbook
CONFIG.SECURITY Alter Sec.Code
Alter Cal.Code
Section 5.8, Page 41
CONFIG.CLOCK Set Clock?
Format dd/mm/yy
Date 01:01:02
Time 12:00
Press To Abort Press To Set
Section 5.7, Page 40
Use the Menu Key
to scroll through
the Menus
CONFIG.OUTPUTS Config. Output 1
AO1: Assign
AO1: Range
AO1: Span Value
AO1: Zero Value
AO1: Default
AO1: Default Val
Config. Output 2
AO2: Assign
AO2: Range
AO2: Span Value
AO2: Zero Value
AO2: Default
AO2: Default Val
Config. Output 3
AO3: Assign
AO3: Range
AO3: Span Value
AO3: Zero Value
AO3: Default
AO3: Default Val
Section 5.6, Page 38
Config. Output 4
AO4: Assign
AO4: Range
AO4: Span Value
AO4: Zero Value
AO4: Default
AO4: Default Val
Single Input Analyzer Only see Supplementary Manual PID Control (IM/AX4PID) CONFIG. CONTROL
TEST/MAINTENANCE Test Outputs
Test Output 1
Test Output 2
Maintenance
Hold Outputs
Test Output 3
Test Output 4
Section 5.10, Page 42
Load/Save Config
Factory Config.
User Config.
Press To Abort Press To Set
CONFIG.SERIAL
Displayed only if option board fitted and serial communications feature enabled (Section 7.3)
see Supplementary Manual Profibus Datalink Description (IM/AX4/PBS)
Automatic Time
Key
Available only if option board fitted and
analog features enabled see Section 7.3
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
6 IM/AX4PH Issue 12
2.3 Operating Page
2.3.1 Single Input pH
Measured Values
pH.
Temperature.
Measured Millivolts
Millivolts.
% Slope and pH Check Value
% slope value.
A value between the programmed minimum % slope value (see Set Min Slope see
Section 5.3, page 30) and 105% is displayed. If the value is outside these limits, check
the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7 pH is the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
See Section 3.1, page 12.
See Section 4.1, page 19.
A3: Type set to Wash (Section 5.5) see Section 2.3.6, page 11.
A3: Type not set to Wash (Section 5.5) return to top of page.
Monitoring pH
7.00
pH
24.4
Deg.C
Monitoring pH
Millivolts
404
mV
A: Slope & Check
100.0
%
7.00
pH
VIEW SETPOINTS
SENSOR CAL.
Wash Function
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
IM/AX4PH Issue 12 7
2.3.2 Dual Input pH
Measured pH
Sensor A.
Sensor B.
Measured Temperature
Sensor A.
Sensor B.
Measured Millivolts
Sensor A.
Sensor B.
% Slope and pH Check Value Sensor A
% slope value.
A value between the programmed minimum % slope value (see Set Min Slope see
Section 5.3, page 30) and 105 % is displayed. If the value is outside these limits, check
the electrode system.
pH check value (zero point).
Displayed as an additional indication of pH electrode system condition; 7 ph being the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
% Slope and pH Check Value Sensor B
% slope value.
See Sensor A above.
pH check value (zero point).
See Section 3.1, page 12.
See Section 4.1, page 19.
A3: Type set to Wash (Section 5.5) see Section 2.3.6, page 11.
A3: Type not set to Wash (Section 5.5) return to top of page.
Monitoring pH
7.00
pH
7.00
pH
Temperature
25.6
Deg.C
Millivolts
404
mV
A: Slope & Check
100.0
%
7.00
pH
24.4
Deg.C
-256
mV
B: Slope & Check
%
pH
100.0
7.00
Monitoring pH
VIEW SETPOINTS
SENSOR CAL.
Wash Function
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
8 IM/AX4PH Issue 12
2.3.3 Single Input Redox (ORP)
Measured Values
Millivolts.
Temperature.
Offset Sensor A
Displays the offset value for the sensor set in A: Adjust Offset see Section 4.1.3,
page 22.
See Section 3.1, page 12.
See Section 4.1, page 19.
A3: Type set to Wash (Section 5.5) see Section 2.3.6, page 11.
A3: Type not set to Wash (Section 5.5) return to top of page.
Redox (ORP)
404
mV
25.6
Deg.C
A: mV Offset
0
mV
Redox (ORP)
VIEW SETPOINTS
SENSOR CAL.
Wash Function
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
IM/AX4PH Issue 12 9
2.3.4 Dual Input Redox (ORP)
Measured Millivolts
Sensor A.
Sensor B.
Note. If Probe Type for both Sensor A and Sensor B is set to ORP (Section 5.3), the
lower display line shows Dual ORP.
Sample Temperature
Sensor A.
Sensor B.
Note. The measured temperature is displayed only if Temp. Sensor is not set to None
see Section 5.3, page 30.
Offset Sensor A
Displays the offset value for Sensor A set in A: Adjust Offset see Section 4.1.3,
page 22.
Offset Sensor B
Displays the offset value for Sensor B set in B: Adjust Offset see Section 4.1.3,
page 22.
See Section 3.1, page 12.
See Section 4.1, page 19.
A3: Type set to Wash (Section 5.5) see Section 2.3.6, page 11.
A3: Type not set to Wash (Section 5.5) return to top of page.
Dual Redox
404
mV
-256
Temperature
25.6
Deg.C
24.4
Deg.C
mV
A: mV Offset
0
mV
B: mV Offset
0
mV
Dual Redox
VIEW SETPOINTS
SENSOR CAL.
Wash Function
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
10 IM/AX4PH Issue 12
2.3.5 Dual Input pH and Redox (ORP)
Measured pH and Millivolts
Sensor A.
Sensor B.
Note. The Probe Type for Sensors A and B can be set to any combination of pH, Redox
or ORP see Section 5.3, page 30. The display indications change depending on
Probe Type settings, e.g. if Sensor A is set to Redox and Sensor B to pH, the lower
display shows Redox (ORP)/pH.
Measured Temperature
Sensor A.
Sensor B.
Note. The measured temperature is displayed only if Temp. Sensor is not set to None
see Section 5.3, page 30.
Measured Millivolts
Sensor A.
Sensor B.
Offset Sensor B
Displays the offset value for Sensor B set in B: Adjust Offset see Section 4.1.3,
page 22.
% Slope and pH Check Value Sensor A
% slope value.
A value between the programmed minimum % slope value (see Set Min Slope
Section 5.3) and 105 % is displayed. If the value is outside these limits, check the
electrode system.
pH check value (zero value).
Displayed as an additional indication of pH electrode system condition; 7 ph is the
optimum value for glass electrodes and 0 ph for Antimony electrodes.
See Section 3.1, page 12.
See Section 4.1, page 19.
A3: Type set to Wash (Section 5.5) see Section 2.3.6, page 11.
A3: Type not set to Wash (Section 5.5) return to top of page.
pH/Redox (ORP)
7.00
pH
-256
mV
Temperature
25.6
Deg.C
Millivolts
404
mV
A: Slope & Check
100.0
%
7.00
pH
24.4
Deg.C
-256
mV
B: mV Offset
0
mV
pH/Redox (ORP)
VIEW SETPOINTS
SENSOR CAL.
Wash Function
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 2 Operation
IM/AX4PH Issue 12 11
2.3.6 Wash Function
Note. The Wash function is available only if A3: Type is set to Wash see Section 5.5, page 34.
Wash Function
Off
On
Manual
Wash function off. Lower display line of Operating Page shows WASH
INHIBITED.
Wash function controlled automatically. Lower display line of Operating
Page shows WASH IN PROGRESS.
Enables wash function to be initiated manually see below.
Note. Set Wash Function to Off before removing the sensor from the process.
See Section 3.1, page 12.
Probe Type set to pH (for either sensor if dual input analyzer) see Section 5.3,
page 30.
See Section 4.1, page 19.
Probe Type set to Redox or ORP (for both sensors in any combination if dual input
analyzer) see Section 5.3, page 30.
Enable Cals set to Yes (Section 5.3) see Section 4.1, page 19.
Alter Sec. Code not set to zero (Section 5.8) see Section 5.1, page 28.
Alter Sec. Code set to zero (Section 5.8) see Section 5.2, page 29.
Wash Function set to Manual see below.
Wash Function not set to Manual. The display returns to the top of the Operating
Page.
Press to Wash (Manual Wash only)
Press to Wash and Press to Abort are shown alternately on the lower
display line.
Press the key to initiate the wash cycle. The display returns to the top of the
Operating Page and the lower display line shows WASH IN PROGRESS until the wash
cycle is completed. The Wash Function selection reverts to the one that was set before
Manual was selected.
Press the key to abort the wash cycle. The display returns to the top of the
Operating Page.
VIEW SETPOINTS
SENSOR CAL.
Wash Function
-----
Manual
On
Off
-----
Press To Wash
Press To Abort
SENSOR CAL.
Security Code
CONFIG. DISPLAY
Press To Wash
Monitoring pH
WASH IN PROGRESS
Monitoring pH
Redox (ORP)
Dual Redox
pH/Redox (ORP)
Redox (ORP)
Dual Redox
pH/Redox (ORP)
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
12 IM/AX4PH Issue 12
3 Operator Views
3.1 View Set Points
Note. The parameter names and units of measurement displayed in the View Set Points page depend on the Probe Type settings
for Sensors A and B see Section 5.3, page 30. Those shown below are given as examples only.
View Set Points
This page shows alarm set points. The value of each of the set points is shown,
together with the name of the parameter it is assigned to.
Alarm assignments, set point values and relay/LED actions are programmable see
Section 5.4, page 33.
Sensor A (pH), Alarm 1 Set Point
Sensor A (Temperature), Alarm 2 Set Point
Sensor B (pH), Alarm 3 Set Point Dual input analyzers only
Sensor B (Temperature), Alarm 4 Set Point Dual input analyzers only
Note. Alarm 4 available only if option board fitted and analog features enabled see
Section 7.3, page 57.
Alarm 5 Set Point
Note. Alarm 5 available only if option board fitted and analog features enabled see
Section 7.3, page 57.
See Section 3.2, page 14.
See Note on next page.
8.30
pH
A1: Setpoint
Sen.A
Temp.A
VIEW SETPOINTS
-----
35.0
Deg.C
A2: Setpoint
Sen.B
6.80
A3: Setpoint
Temp.B
55.0
Deg.C
A4: Setpoint
-----
Off
A5: Setpoint
VIEW OUTPUTS
SENSOR CAL.
Security Code
CONFIG. DISPLAY
VIEW SETPOINTS
pH
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
IM/AX4PH Issue 12 13
Note. The menu displayed when pressing the key from the Operator View pages depends on analyzer configuration, i.e.:
Single Input Analyzers
Probe Type set to pH
or
Probe Type set to Redox or ORP and Enable Cals set to Yes (Section 5.3) see Section 4.1, page 19.
Probe Type set to Redox or ORP and Enable Cals set to No (Section 5.3) and Alter Sec. Code not set to zero
(Section 5.8) see Section 5.1, page 28.
Probe Type set to Redox or ORP and Enable Cals set to No (Section 5.3) and Alter Sec. Code set to zero
(Section 5.8) see Section 5.2, page 29.
Dual Input Analyzers
Probe Type for either sensor set to pH
or
Probe Type for both sensors set to Redox or ORP and Enable Cals for either sensor set to Yes (Section 5.3)
see Section 4.1, page 19.
Probe Type for both sensors set to Redox or ORP and Enable Cals for both sensors set to No (Section 5.3) and
Alter Sec. Code not set to zero (Section 5.8) see Section 5.1, page 28.
Probe Type for both sensors set to Redox or ORP and Enable Cals for both sensors set to No (Section 5.3) and
Alter Sec. Code set to zero (Section 5.8) see Section 5.2, page 29.
SENSOR CAL.
Security Code
CONFIG. DISPLAY
SENSOR CAL.
Security Code
CONFIG. DISPLAY
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
14 IM/AX4PH Issue 12
3.2 View Outputs
3.3 View Hardware
Theoretical Analog Output
There are up to four analog outputs, each showing information for one sensor.
Note. Analog outputs 3 and 4 available only if option board fitted and analog features
enabled see Section 7.3, page 57.
Live current output value being retransmitted.
Current output shown as a percentage of full scale for the output range set in CONFIG.
OUPUTS see Section 5.6, page 38.
See Section 3.3, below.
See Note on Page 13.
Advance to analog output 2 (and outputs 3 and 4 if option board fitted and analog
features enabled see Section 7.3, page 57).
50.0
%
Analog Output 1
12.00
mA
VIEW OUTPUTS
-----
VIEW HARDWARE
SENSOR CAL.
Security Code
CONFIG. DISPLAY
Analog Output 2
Sensor A Module
Shows the type of input board fitted to the analyzer for the Sensor A input.
Sensor B Module Dual input analyzers only
Shows the type of input board fitted to the analyzer for the Sensor B input.
Option Board
Note. Displayed only if the option board is fitted.
Displays the optional features enabled in the Factory Settings page see Section 7.3,
page 57.
See Section 3.4, page 15.
See Note on Page 13.
pH
pH
-----
Sensor A Module
VIEW HARDWARE
-----
Option Board
-----
Sensor B Module
-----
Pb Dp
Analog
VIEW SOFTWARE
SENSOR CAL.
Security Code
CONFIG. DISPLAY
VIEW HARDWARE
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
IM/AX4PH Issue 12 15
3.4 View Software
Issue
Shows the version number of the operating software.
Option board fitted and analog features enabled (Section 7.3) and Logbook set to On
(Section 5.9) see Section 3.5, page 16.
Operating Page (option board not fitted) see Section 2.3, page 6.
See Note on Page 13.
1.00
AX400/2000 Issue
VIEW SOFTWARE
-----
VIEW LOGBOOK
SENSOR CAL.
Security Code
CONFIG. DISPLAY
VIEW SOFTWARE
Monitoring pH
Redox (ORP)
Dual Redox
pH/Redox (ORP)
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
16 IM/AX4PH Issue 12
3.5 View Logbook
Note. The View Logbook function is available only if the option board is fitted and analog features enabled (see Section 7.3,
page 57) and Logbook is set to On (see Section 5.9, page 41).
The logbook stores data entries for alarm events, sensor errors, power failures and pH
calibration information.
View Logbook
Use the and keys to access the Alarms logbook.
Note. If no entries are stored in the Alarms logbook, the display shows No More
Entries.
Alarms
The Alarms logbook contains up to 10 entries (entry 1 is the most recent), each
comprising an alarm number, alarm state (On or Off) and the date/time of the
occurrence.
Option board fitted and analog features enabled (Section 7.3) see Section 3.6,
page 18.
See Note on Page 13.
Advance to entries 2 to 10.
Note. If no more entries are stored, the display shows No More Entries.
View Logbook
Use the and keys to access the Errors logbook.
Note. If no entries are stored in the Errors logbook, the display shows No More Entries.
Errors
The Errors logbook contains up to 5 entries (entry 1 is the most recent), each
comprising the sensor letter, error number and the date/time of the occurrence.
Option board fitted and analog features enabled (Section 7.3) see Section 3.6,
page 18.
See Note on Page 13.
Advance to entries 2 to 5.
Note. If no more entries are stored, the display shows No More Entries.
-----
-----
11:09:02 12:34
VIEW CLOCK
SENSOR CAL.
Security Code
CONFIG. DISPLAY
Sen.A
VIEW LOGBOOK
Errors
1
Sen.A
Pt100
2
Alarms
Cals
Power
Errors
-----
11:09:02 12:34
VIEW LOGBOOK
-----
VIEW CLOCK
SENSOR CAL.
Security Code
CONFIG. DISPLAY
A1
-----
VIEW LOGBOOK
1
2
Cals
Power
Errors
Alarms
A1
On
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
IM/AX4PH Issue 12 17
View Logbook
Use the and keys to access the Power logbook.
Note. If no entries are stored in the Power logbook, the display shows No More Entries.
Power
The Power logbook contains up to 2 entries (entry 1 is the most recent), each
comprising the power state (On or Off) and the date/time of the occurrence.
Option board fitted and analog features enabled (Section 7.3) see Section 3.6,
page 18.
See Note on Page 13.
Advance to entry 2.
Note. If no more entries are stored, the display shows No More Entries.
View Logbook
Use the and keys to access the Cals logbook.
Note. If no entries are stored in the Cals logbook, the display shows No More Entries.
Calibration
The Cals logbook contains up to 5 entries (entry 1 is the most recent), each comprising
2 frames. Frame 1 contains the entry number, sensor letter and the calibration pass/fail
indication.
Frame 2 contains the % slope value, the pH check value and the date/time of the
occurrence.
Option board fitted and analog features enabled (Section 7.3) see Section 3.6,
page 18.
See Note on Page 13.
Advance to entries 2 to 5.
Note. If no more entries are stored, the display shows No More Entries.
VIEW CLOCK
-----
Calibration
-----
VIEW LOGBOOK
1
Sen.A
Passed
11:09:02 12:34
SENSOR CAL.
Security Code
CONFIG. DISPLAY
Sen.A
100. 0
%
pH
2
7.00
Power
Errors
Alarms
Cals
-----
11:09:02 12:34
VIEW CLOCK
SENSOR CAL.
Security Code
CONFIG. DISPLAY
-----
VIEW LOGBOOK
1
Off
2
Errors
Alarms
Cals
Power
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 3 Operator Views
18 IM/AX4PH Issue 12
3.6 View Clock
Note. The View Clock function is available only if the option board is fitted and analog features enabled see Section 7.3,
page 57.
Date
Shows the current date.
Time
Shows the current time.
Operating Page see Section 2.3, page 6.
See Note on Page 13.
-----
Date 01:02:04
VIEW CLOCK
-----
-----
Time 12:00
Monitoring pH
SENSOR CAL.
Security Code
CONFIG. DISPLAY
VIEW CLOCK
Redox (ORP)
Dual Redox
pH/Redox (ORP)
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 4 Setup
IM/AX4PH Issue 12 19
4 Setup
4.1 Sensor Calibration
4.1.1 Set Buffer Type (pH Only)
Note. If Probe Type for either sensor (Sensor A only if single input) is set to Redox or ORP, the sensor can be calibrated only if
Enable Cals. for that sensor is set to Yes see Section 5.3, page 30.
Sensor Calibration
Sensor Calibration Security Code
Note. This frame is displayed only if Alter Cal. Code is not set to zero see Section
5.8, page 41.
Enter the required code number (between 0000 and 19999) to access the sensor
calibration pages. If an incorrect value is entered, access to the calibration pages is
prevented and the display reverts to the SENSOR CAL. menu.
Probe Type set to pH (for either sensor if dual input analyzer see Section 5.3,
page 30) continued below.
Probe Type set to Redox or ORP (for both sensors if dual input analyzer see Section
5.3, page 30) continued on page 22.
SENSOR CAL.
-----
Cal. User Code
0000
Set Auto Buffers
Sensor Cal. A
Set Auto Buffers
Continued on Page 22.
Alter Sec. Code not set to zero (Section 5.8) see Section 5.1, page 28.
Alter Sec. Code set to zero (Section 5.8) see Section 5.2, page 29.
Continued below.
Buffer Type
Select the relevant type of buffer solution (see Appendix A):
ABB
NIST
DIN
MERCK
TECH
User
Millivolt display
Millivolt display
pH display
Differential Input
Yes Select if electrode system is equipped with a solution earth (ground)
rod and sensor diagnostics are required see Section 5.4, page 33.
No Select if electrode system is not equipped with a solution earth
(ground) rod or sensor diagnostics are not required.
Note. Ensure the electrode system is connected correctly for the type of input
selected (Standard or Differential) see Fig. 6.9 (wall-/pipe-mount analyzers) or
Fig. 6.11 (panel-mount analyzers).
Probe Type set to Redox or ORP continued below.
Probe Type set to pH continued on next page.
Temperature Sensor
Select the type of temperature sensor used: Pt100, Pt1000, Balco 3K or None.
Enable Calibration
Select Yes to enable sensor calibration.
Select No to disable sensor calibration and all associated frames.
Return to main menu.
See Section 5.4, page 33.
Config. Sensor A
-----
Config. Sensor B
A: Temp. Sensor
A: Probe Type
-----
ORP
Redox
pH
CONFIG. SENSORS
-----
A: Diff. Input
-----
Yes
No
A: Electrode
A: Temp. Sensor
-----
None
3K Bal
Pt1000
Pt100
A: Enable Cals
-----
Yes
No
A: Probe Type
set to Redox/ORP
CONFIG. SENSORS
CONFIG. ALARMS
Config. Sensor A
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
IM/AX4PH Issue 12 31
pH Electrode Type
Select the type of pH electrode used, Glass or Antimony.
Temperature Compensation
Select Auto to enable the analyzer to compensate automatically for fluctuations in
sample temperature.
Temperature Sensor (Automatic Temperature Compensation only)
Select the type of temperature sensor used: Pt100, Pt1000 or Balco 3K.
continued on next page.
Preset Temperature (Manual Temperature Compensation only)
Enter the temperature of the sample within the range 10.0 to 120.0 C.
continued on next page.
A: Temp. Comp.
-----
Manual
Auto
A: Electrode
-----
Antim.
Glass
A: Temp. Sensor
-----
3K Bal
Pt1000
Pt100
A: Preset Temp.
Deg.C
25.0
A: Sample Comp.
A: Set Min Slope
Auto
Manual
A: Probe Type
set to pH
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
32 IM/AX4PH Issue 12
Solution Temperature Compensation
Select Yes to enable compensation for solution temperature effects referenced to
25 C.
Sample Coefficient
If Sample Comp. is set to Yes, enter the temperature coefficient of the sample, in pH/C,
within the range 0.020 to 0.050 (in 0.001 increments). For ammonia and sodium
hydroxide dosed boilers the value is typically 0.035 (this depends on the individual
boiler chemistry). The exact value for a particular sample needs to be determined by
laboratory analysis.
pH Calibration Minimum Slope Value
Set the required pH calibration minimum slope value, in %, within the range 60.0 to
90.0 (in 0.1 increments). The calibration fail limit is set automatically to 20 % below the
minimum slope setting see Table 4.1.
Sensor B (dual input analyzers only) configuration is identical to Sensor A
configuration.
see Section 5.4, page 33
Config. Sensor B
CONFIG. DIAGS
A: Sample Comp.
-----
Yes
No
A: Sample Coeff.
pH/C
-0.035
A: Set Min Slope
%
60.0
Config. Sensor A
Yes
No
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
IM/AX4PH Issue 12 33
5.4 Configure Diagnostics
Note. The Configure Diagnostics function is applicable only if Diff. Input for Sensor A and/or Sensor B is set to Yes see Section
5.3, page 30.
Configure Sensor A
Sensor B configuration (dual input analyzers only) is identical to Sensor A
configuration.
pH Glass Check
Set to On to enable the impedance of the glass electrode circuit to be monitored in
order to provide low glass impedance and out-of-sample/broken cable warnings.
Note. This frame is displayed only if A: Probe Type is set to pH and A: Electrode is set
to Glass see Section 5.3, page 30.
Reference Electrode Check
Set to On to enable the condition of the reference electrode to be monitored in order to
provide:
an indication of when sensor cleaning or replacement is required
out-of-sample/broken cable warnings
Sensor B (dual input analyzers only) configuration is identical to Sensor A
configuration.
See section 5.5, page 34.
Reference Alarm
Set the impedance value above which the reference alarm is activated.
Sensor B (dual input analyzers only) configuration is identical to Sensor A
configuration.
See section 5.5, page 34.
Config. Sensor A
-----
A: pH Glass
-----
A: Ref. Checking
-----
CONFIG. DIAGS
-----
Off
On
Off
On
A: Ref. Alarm
50
kOhm
CONFIG. ALARMS
Config. Sensor A
CONFIG. ALARMS
Config. Sensor B
Config. Sensor B
Config. Sensor B
Config. Sensor A
On
Off
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
34 IM/AX4PH Issue 12
5.5 Configure Alarms
Configure Alarm 1
Alarms 2 and 3 configuration (and Alarms 4 and 5 if option board fitted and analog
features enabled see Section 7.3, page 57) is identical to Alarm 1 configuration.
Alarm 3 can also be configured as a Wash alarm if A3: Type is set to Wash see
following frame.
Alarm 1 Type
Select the type of alarm required:
Off The alarm is disabled, the alarm LED is off and the relay is
de-energized at all times.
Alarm The analyzer is configured using the Assign frame (following) to
generate an alarm in response to a specified high or low pH, Redox
(ORP) or process temperature sensor reading.
Status An alarm is generated if either a power failure or a condition that
causes any of the error messages in Table 8.1 (page 62) to be
displayed occurs.
Wash Alarm 3 is configured to control the wash sequence.
Note. The Wash alarm type can be assigned only to Alarm 3 and is displayed only
when the lower display line shows A3: Type.
A1: Type set to Off or Status.
A1: Type set to Alarm continued below.
A3: Type set to Wash see Section 5.5.1, page 36.
Alarm 1 Assign
Select the alarm assignment required:
Sen.A
Sen.B
The analyzer activates an alarm if the pH or Redox (ORP) value of the
process fluid measured by the selected sensor exceeds or drops
below the value set in the Alarm 1 Set Point parameter, depending on
the type of Alarm 1 Action selected see next page.
Temp.A
Temp.B
The analyzer activates an alarm if the temperature of the process fluid
measured by the selected sensor exceeds or drops below the value
set in the Alarm 1 Set Point parameter, depending on the type of
Alarm h1 Action selected see next page.
A-B The analyzer activates an alarm if the difference between the Sensor A
and Sensor B readings exceeds or drops below the value set in the
Alarm 1 Set Point parameter, depending on the type of Alarm 1 Action
selected see next page.
Note. The Sen.B, Temp.B and A-B alarm assignment types are applicable only to dual
input analyzers and A-B is displayed only when Probe Type for each sensor is set to pH
see Section 5.3, page 30.
Continued on next page.
Config. Alarm 1
-----
A1: Type
-----
A1: Assign
-----
CONFIG. ALARMS
-----
Wash
Status
Alarm
Off
A-B
Temp.B
Sen.B
Temp.A
Sen.A
A1: Failsafe
Config. Alarm 2
A1: Assign
Config. Alarm 1
Wash Mode
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
IM/AX4PH Issue 12 35
Alarm 1 Failsafe
Select Yes to enable failsafe action, otherwise select No.
See also Fig. 5.2 to Fig. 5.6 (page 37).
Alarm 1 Action
Select the alarm action required, High or Low.
See also Fig. 5.2 to Fig. 5.6 (page 37).
Alarm 1 Set Point
The Alarm 1 Set Point can be set within the following ranges:
pH
mV
Deg.C
Deg.F
A-B
2.00 to 16.00 pH
1200 to 1200 mV
10.0 to 150.0
14.0 to 302.0
0.00 to 14.00 pH
Set to the value required.
Alarm 1 Hysteresis
A differential set point can be defined between 0 and 5 % of the alarm set point value.
Set the required hysteresis in 0.1 % increments.
See also Fig. 5.2 to Fig. 5.6 (page 37).
Alarm 1 Delay
If an alarm condition occurs, the activation of the relays and LEDs can be delayed for a
specified time period. If the alarm clears within the period, the alarm is not activated.
Set the required delay, in the range 0 to 60 seconds in 1 second increments.
See also Fig. 5.2 to Fig. 5.6 (page 37).
Alarms 2 and 3 configuration (and Alarms 4 and 5 if option board fitted and analog
features enabled see Section 7.3, page 57) is identical to Alarm 1 configuration.
See Section 5.6, page 38.
A1: Action
High
Low
A1: Setpoint
A1: Hysteresis
0.0
%
A1: Delay
0
Secs
-----
A1: Failsafe
Yes
No
-----
CONFIG. OUTPUTS
Config. Alarm 2
Config. Alarm 1
A-B
mV
Deg.C
Deg.F
pH
100.0
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
36 IM/AX4PH Issue 12
5.5.1 Wash Cycle Configuration (Applicable Only to Alarm 3)
Wash Mode
Select the wash mode required.
Cont.
Pulsed
Changes in relay action and analog outputs are inhibited during sensor
calibration.
Changes in relay action and analog outputs are inhibited.
Changes in relay action and analog outputs are not inhibited.
Note. The LEDs flash while the analyzer is in 'Hold' mode.
Continued on next page.
See Section 7.3, page 57.
Hold Outputs set to Off or On return to main menu.
Hold Outputs set to Auto. continued on next page.
Test Outputs
-----
TEST/MAINTENANCE
-----
Test Output 1
4.00
mA
20.0
%
Maintenance
-----
Hold Outputs
-----
Auto.
On
Off
FACTORY SETTINGS
Test Output 2
FACTORY SETTINGS
Load/Save Config
Maintenance
Maintenance
Automatic Time
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 5 Programming
IM/AX4PH Issue 12 43
Automatic Time
If required, set a time period between 1 and 6 hours, in 30 minute increments, for
which the outputs are held when Hold Outputs is set to Auto.
At the default setting of Off, changes in relay action and analog outputs are inhibited
during sensor calibration and released automatically at the end of the procedure.
If a time is set, changes in relay action and analog outputs are inhibited during sensor
calibration, but if the calibration is not completed within the set time, the calibration is
aborted, the display returns to the Operating Page and CAL. ABORTED is displayed.
Continued below.
See Section 7.3, page 57.
Load/Save Configuration
Select whether a configuration is to be loaded or saved.
Note. If No is selected, pressing the key has no effect.
Return to main menu.
See Section 7.3, page 57.
Load User/Factory Configuration
Note. Applicable only if Load/Save Config is set to Yes.
Factory Config.
Save User Config.
Load User Config.
6
.
5
0
(
0
.
2
6
)
Dimensions in mm (in.) Fixing Centers
F
i
x
i
n
g
C
e
n
t
e
r
s
94 (3.7)
B Pipe-mounting
Position 'U' bolts on pipe
Position plate over 'U' bolts
Secure transmitter to mounting plate
Secure plate
A Wall-mounting
Mark fixing centers
(see Fig. 6.2)
Drill suitable
holes
Secure instrument to
wall using
suitable fixings
1
2
3
4
1
2
3
61 (2
3
/8) OD
Vertical or
Horizontal
Post
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
46 IM/AX4PH Issue 12
6.2.2 Panel-mount Analyzers
Fig. 6.4 Overall Dimensions
Fig. 6.5 Panel-mounting
Note. The clamp must fit flat on the analyzer casing. If the clamp is bowed, the securing screw is over tight and sealing problems
may occur.
Dimensions in mm (in.)
96 (3.78)
9
6
(
3
.
7
8
)
5.40 (0.2)
9
1
.
6
0
(
3
.
6
)
137.50 (5.41)
25
(0.98)
Panel Cut-out
+0.8
0
92 (3.62 )
+0.03
0
+0.8
0
92
(3.62 )
+0.03
0
Insert the instrument
into the panel cut-out
Secure the analyzer by tightening the panel clamp
retaining screws (see Note below)
Loosen the
retaining screw on
each panel clamp
Remove the panel clamp and
anchors from the instrument case
Cut a hole in the panel (see Fig. 6.4 for dimensions).
Instruments may be close stacked to DIN 43835
1
2
3
4
5
6
Refit the panel clamps to the case, ensuring that the
panel clamp anchors are located correctly in their slots
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
IM/AX4PH Issue 12 47
6.3 Connections, General
Warning.
The instrument is not fitted with a switch therefore a disconnecting device such as a switch or circuit breaker conforming to
local safety standards must be fitted to the final installation. It must be fitted in close proximity to the instrument within easy
reach of the operator and must be marked clearly as the disconnection device for the instrument.
Remove all power from supply, relay and any powered control circuits and high common mode voltages before accessing
or making any connections.
The power supply earth (ground) must be connected to ensure safety to personnel, reduction of the effects of RFI
interference and correct operation of the power supply interference filter.
The power supply earth (ground) must be connected to the earth (ground) stud on the analyzer case see Fig. 6.8
(wall-/pipe-mount analyzers) or Fig. 6.10 (panel-mount analyzers).
Use cable appropriate for the load currents. The terminals accept cables up to 14 AWG (2.5mm
2
).
The instrument conforms to Mains Power Input Insulation Category III. All other inputs and outputs conform to Category II.
All connections to secondary circuits must have basic insulation.
After installation, there must be no access to live parts, e.g. terminals.
Terminals for external circuits are for use only with equipment with no accessible live parts.
The relay contacts are voltage-free and must be appropriately connected in series with the power supply and the
alarm/control device which they are to actuate. Ensure that the contact rating is not exceeded. Refer also to Section 6.3.1
for relay contact protection details when the relays are to be used for switching loads.
Do not exceed the maximum load specification for the selected analog output range.
The analog output is isolated, therefore the ve terminal must be connected to earth (ground) if connecting to the isolated
input of another device.
If the instrument is used in a manner not specified by the Company, the protection provided by the equipment may be
impaired.
All equipment connected to the instrument's terminals must comply with local safety standards (IEC 60950, EN61010-1).
Note.
An earthing (grounding) stud terminal is fitted to the analyzer case for bus-bar earth (ground) connection see Fig. 6.8
(wall-/pipe-mount analyzers) or Fig. 6.10 (panel-mount analyzers).
Always route signal output/sensor cell cable leads and mains-carrying/relay cables separately, ideally in earthed (grounded)
metal conduit. Use twisted pair output leads or screened cable with the screen connected to the case earth (ground) stud.
Ensure that the cables enter the analyzer through the glands nearest the appropriate screw terminals and are short and
direct. Do not tuck excess cable into the terminal compartment.
Ensure that the IP65 rating is not compromised when using cable glands, conduit fittings and blanking plugs/bungs (M20
holes). The M20 glands accept cable of between 5 and 9mm (0.2 and 0.35 in.) diameter.
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
48 IM/AX4PH Issue 12
6.3.1 Relay Contact Protection and Interference Suppression
If the relays are used to switch loads on and off, the relay contacts can become eroded due to arcing. Arcing also generates radio
frequency interference (RFI) which can result in analyzer malfunctions and incorrect readings. To minimize the effects of RFI, arc
suppression components are required; resistor/capacitor networks for AC applications or diodes for DC applications. These
components must be connected across the load see Fig. 6.6.
For AC applications the value of the resistor/capacitor network depends on the load current and inductance that is switched. Initially,
fit a 100R/0.022 F RC suppressor unit (part no. B9303) as shown in Fig. 6.6A. If the analyzer malfunctions (locks up, display goes
blank, resets etc.) the value of the RC network is too low for suppression and an alternative value must be used. If the correct value
cannot be obtained, contact the manufacturer of the switched device for details on the RC unit required.
For DC applications fit a diode as shown in Fig. 6.6B. For general applications use an IN5406 type (600 V peak inverse voltage at
3 A).
Note. For reliable switching the minimum voltage must be greater than 12V and the minimum current greater than 100mA.
Fig. 6.6 Relay Contact Protection
NC C NO
External
DC Supply
+
Relay Contacts
Load
Diode
NC C NO
External
AC Supply
L N
Relay Contacts
C
R
Load
A AC Applications B DC Applications
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
IM/AX4PH Issue 12 49
6.3.2 Cable Entry Knockouts, Wall-/Pipe-mount Analyzer
The analyzer is supplied with 7 cable glands, one fitted and six to be fitted, as required, by the user see Fig. 6.7.
Fig. 6.7 Cable Entry Knockouts, Wall-/Pipe-mount Analyzer
5
Place the blade of a small, flat bladed screwdriver
into the knockout groove and tap the
screwdriver smartly to remove the knockout
(see Note below)
Smooth the edges of the hole
with a small round or half-round file
Fit an 'O' ring seal to the the cable gland
Insert the cable gland into the hole
in the analyzer case from the outside
Secure the cable
gland with the nut
2
3
4
6
Cable entry knockouts
Factory-fitted cable gland
1
Release the four captive
screws and remove
the terminal cover plate
Note. When removing knockouts, take
great care not to damage wiring and
components within the analyzer.
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
50 IM/AX4PH Issue 12
6.4 Wall-/Pipe-mount Analyzer Connections
6.4.1 Access to Terminals
Fig. 6.8 Access to Terminals, Wall-/Pipe-mount Analyzer
Terminal Block A Terminal Block B
Terminal Block C
(Option Board)
Case Earth
(Ground) Stud
Release the four captive
screws and remove
the terminal cover plate
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
IM/AX4PH Issue 12 51
6.4.2 Connections
Fig. 6.9 Connections, Wall-/Pipe-mount Analyzer
Note. Relay 3 can be configured to control the wash facility see Section 5.4, page 33.
Temperature Compensator
Connections
B
1
6
B
1
5
B
1
4
B
1
3
B
1
2
B
1
1
B
1
0
B
9
B
8
B
7
B
6
B
5
B
4
B
3
B
2
B
1
T
C
T
h
i
r
d
L
e
a
d
C
o
m
m
o
n
T
C
T
h
i
r
d
L
e
a
d
C
o
m
m
o
n
Temperature Compensator
Connections
P
R
O
F
I
B
U
S
D
P
c
o
n
n
e
c
t
i
o
n
s
r
e
f
e
r
t
o
I
M
/
P
R
O
B
U
S
Before making any electrical connections,
see Warnings on page 47
C
1
N
o
t
u
s
e
d
C
2
N
o
t
u
s
e
d
C
3
C
4
C
5
C
6
N
o
t
u
s
e
d
C
7
C
C
8
N
C
R
e
l
a
y
4
C
9
N
O
C
1
0
C
C
1
1
N
C
R
e
l
a
y
5
C
1
2
N
O
C
1
3
+
A
n
a
l
o
g
O
u
t
p
u
t
3
C
1
4
C
1
5
+
A
n
a
l
o
g
O
u
t
p
u
t
4
C
1
6
Earth (Ground)
Stud on Case
(see Fig. 6.8)
L
i
n
e
L
N
e
u
t
r
a
l
NE
C
A
4
R
e
l
a
y
1
N
C
A
5
N
O
A
6
C
A
7
R
e
l
a
y
2
N
C
A
8
N
O
A
9
C
A
1
0
R
e
l
a
y
3
(
s
e
e
N
o
t
e
b
e
l
o
w
)
N
C
A
1
1
N
O
A
1
2
A
n
a
l
o
g
O
u
t
p
u
t
1
+
A
1
3
A
1
4
A
n
a
l
o
g
O
u
t
p
u
t
2
+
A
1
5
A
1
6
Terminal Block A
Terminal Block B
Terminal Block C (Option Board)
+
8
5
t
o
2
6
5
V
A
C
P
o
w
e
r
S
u
p
p
l
i
e
s
1
2
t
o
2
4
V
D
C
o
r
2
4
V
A
C
Refer to Section 6.6 for pH Sensor System
Connection Details
C
o
n
n
e
c
t
s
u
p
p
l
y
e
a
r
t
h
(
g
r
o
u
n
d
)
t
o
s
t
u
d
o
n
c
a
s
e
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
52 IM/AX4PH Issue 12
6.5 Panel-mount Analyzer Connections
6.5.1 Access to Terminals
Fig. 6.10 Access to Terminals, Panel-mount Analyzers
Terminal Block C
(Option Board)
Terminal Block B
Terminal Block A
Case Earth (Ground) Stud
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
IM/AX4PH Issue 12 53
6.5.2 Connections
Fig. 6.11 Connections, Panel-mount Analyzers
Note. Relay 3 can be configured to control the wash facility see Section 5.4, page 33.
L Line
N Neutral
E Connect supply earth (ground) to stud on case
A4 C
A5 NC Relay 1
A6 NO
A7 C
A8 NC Relay 2
A9 NO
A10 C
A11 NC Relay 3 (see Note below)
A12 NO
A13 +
Analog Output 1
A14
A15 +
Analog Output 2
A16
Terminal Block B Terminal block A
Temperature Compensator
Connections
Common
Third Lead
TC
B1
B2
B3
B4
B5
B6
B7
B8
B9
B10
B11
B12
B13
B14
B15
B16
Common
Third Lead
TC
Temperature Compensator
Connections
Terminal Block C
(Option Board)
C1 Not Used
C2 Not Used
C3
C4
C5
C6 Not Used
C7 C
C8 NC Relay 4
C9 NO
C10 C
C11 NC Relay 5
C12 NO
C13 +
Analog Output 3
C14
C15 +
Analog Output 4
C16
Earth (Ground)
Stud on Case
(see Fig. 6.10)
PROFIBUS DP
connections
refer to IM/PROBUS
85 to 265 V AC + 12
or to
Power
24 V AC 24V DC
Supplies
Before making any electrical connections,
see Warnings on page 47
Refer to Section 6.6 for pH Sensor System
Connection Details
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
54 IM/AX4PH Issue 12
6.6 pH Sensor Systems Connections
6.6.1 Standard ph Systems Connection 2867, AP100, AP300, 7650/60, TB5, Non-ABB
When connecting one of these pH systems to the AX400 transmitter, ensure the differential input switch for the relevant sensor
is OFF.
Terminal Block B Function 2867 AP100 AP300
Sensor B Sensor A Color Color Color
B1 B9
Temperature Compensator (if fitted) Common
see also Note 1below
Not Used Red White
B2 B10 Temperature Compensator (if fitted) 3
rd
Lead Not Used Red Grey
B3 B11 Temperature Compensator (if fitted) Not Used White Red
B4 B12 Not Used Not Used Not Used Not Used
B5 B13 Not Used Not Used Not Used Not Used
B6 B14 Reference Electrode Black Black Black
B7 B15 Screen/Shield (if fitted) Not Used Not Used Not Used
B8 B16 Glass/Metal Electrode Clear Clear Blue
Table 6.1 Standard ph Systems Connection 2867, AP100, AP300
Terminal Block B Function *7650/60 TB5 Non-ABB
Sensor B Sensor A Color Color
B1 B9
Temperature Compensator (if fitted) Common
see also Note 1below
Red White
Connect as per
function
refer to
(non-ABB)
sensor system
manual for cable
colors
B2 B10 Temperature Compensator (if fitted) 3
rd
Lead Red Link to White
B3 B11 Temperature Compensator (if fitted) White Red
B4 B12 Not Used Not Used Not Used
B5 B13 Not Used Not Used Not Used
B6 B14 Reference Electrode Black Black
B7 B15 Screen/Shield (if fitted) Yellow Not Used
B8 B16 Glass/Metal Electrode Clear Blue
*See Note 2
Table 6.2 Standard ph Systems Connection 7650/60, TB5, Non-ABB
Note.
1. If the sensor is fitted with a 2-wire PT100, Pt1000 or 3K Balco temperature compensator, link terminals B9 and B10 (and
B1 and B2 if dual input analyzer).
2. Discard large green wire as not required with this transmitter.
3. Redox systems are not temperature compensated so do not have temperature sensors. To remove temperature error
messages, set Temperature Sensor to NONE. If a temperature sensor is used for a separate temperature display, set
Temperature Sensor to the correct type see Section 5.3, page 30.
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 6 Installation
IM/AX4PH Issue 12 55
6.6.2 Differential ph Systems Connections Capable of Providing Sensor Diagnostics (AP200, TBX5)
When connecting one of these pH systems to the AX400 transmitter, ensure the differential input switch for the relevant sensor is ON.
See Section 5.4, page 33 for sensor diagnostics configuration. If diagnostics are not required leave them switched off.
*In normal operation do not connect the heavy green wire. If noisy readings are obtained, connect the wire to the earth stud.
Terminal Block B Function AP200 *TBX5
Sensor B Sensor A Color Color
B1 B9
Temperature Compensator (if fitted) Common
see also Note 1below
Grey White
B2 B10 Temperature Compensator (if fitted) 3
rd
Lead White Link to White
B3 B11 Temperature Compensator (if fitted) Green Red
B4 B12 Reference Blue Black
B5 B13 Not Used Not Used Not Used
B6 B14 Solution Earth (Ground Rod) Green/Yellow Green
B7 B15 Screen/Shield (if fitted) Red Yellow
B8 B16 Glass/Metal Electrode Clear Blue
Table 6.3 Differential ph Systems Connection AP200, TBX5
Note.
1. If the sensor is fitted with a 2-wire PT100, Pt1000 or 3K Balco temperature compensator, link terminals B9 and B10 (and
B1 and B2 if dual input analyzer.).
2. Redox systems are not temperature compensated so do not have temperature sensors. To remove temperature error
messages, set Temperature Sensor to NONE. If a temperature sensor is used for a separate temperature display, set
Temperature Sensor to the correct type see Section 5.3, page 30.
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
56 IM/AX4PH Issue 12
7 Calibration
7.1 Equipment Required
1. Millivolt source (pH or Redox input simulator): 1000 to 1000 mV.
2. Decade resistance box (Pt100/Pt1000 temperature input simulator): 0 to 10 k (in increments of 0.01 ), accuracy 0.1 %.
3. Digital milliammeter (current output measurement): 0 to 20 mA.
7.2 Preparation
1. Switch off the supply and disconnect the conductivity cell(s), temperature compensator(s) and current output(s) from the
analyzer's terminal blocks.
2. Sensor A Fig. 7.1:
a. Link terminals B9 and B10.
b. Connect the millivolt source to terminals B14 (ve) and B16 (+ve) to simulate the pH or Redox input. Connect the millivolt
source earth to the Case Earth (Ground) Stud see Fig. 6.8 (wall-/pipe-mount analyzer) or Fig. 6.10 (panel-mount analyzer).
c. Connect the 0 to 10 k decade resistance box to terminals B9 and B11 to simulate the Pt100/Pt1000/Balco 3K.
Sensor B (dual input analyzers only) Fig. 7.1:
a. Link terminals B1 and B2.
b. Connect the millivolt source to terminals B6 (ve) and B8 (+ve) to simulate the pH or Redox input. Connect the millivolt source
earth to the Case Earth (Ground) Stud see Fig. 6.8 or (wall-/pipe-mount analyzer) or Fig. 6.10 (panel-mount analyzer).
c. Connect the 0 to 10 k decade resistance box to terminals B1 and B3 to simulate the Pt100/Pt1000/Balco 3K.
3. Connect the milliammeter to the analog output terminals.
4. Switch on the supply and allow ten minutes for the circuits to stabilize.
5. Select the FACTORY SETTINGS page and carry out Section 7.3.
Note.
The analyzer is calibrated by the Company prior to dispatch and the Factory Settings pages are protected by an access
code.
Routine recalibration is not necessary high stability components are used in the analyzer's input circuitry and, once
calibrated, the Analog-to-Digital converter chip self-compensates for zero and span drift. It is therefore unlikely that the
calibration will change over time.
Do Not attempt recalibration without first contacting ABB.
Do Not attempt recalibration unless the input board has been replaced or the Factory Calibration tampered with.
Prior to attempting recalibration, test the analyzer's accuracy using suitably calibrated test equipment see Section 7.1,
page 56 and see Section 7.2, page 56.
Note. Resistance boxes have an inherent residual resistance that may range from a few m up to 1 . This value must be taken
into account when simulating input levels, as should the overall tolerance of the resistors within the boxes.
Fig. 7.1 Analyzer Terminal Links and Decade Resistance Box Connections
Terminal link
pH/Redox
Input Simulator
Temperature
Simulator
B10 B11 B12 B13 B14 B9 Sensor A Terminal Numbers
Sensor B Terminal Numbers
B2 B3 B4 B5 B6 B1
B15 B16
B7 B8
ve
+ve
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
IM/AX4PH Issue 12 57
7.3 Factory Settings
Fig. 7.2 Overall Factory Settings Chart
FACTORY SETTINGS Factory Set Code Cal. Sensor A
A:mV Zero (-1V)
A:mV Span (+1V)
A:T.Zero (100R)
A:T.Span (150R)
Cal. Sensor B
B:T.Zero (100R)
B:T.Span (150R)
Cal. Output 1
O1: Adjust 4mA
O1: Adjust 20mA
To OPERATING PAGE
A:T.Zero (1K0) B:T.Zero (1K0)
A:T.Span (1K5) B:T.Span (1K5)
Available only if the Option Board is fitted
and Analog features enabled see page 61
Key
Cal. Output 3 Cal. Output 4
O3: Adjust 4mA O4: Adjust 4mA
O3: Adjust 20mA O4: Adjust 20mA
Cal. Output 2
O2: Adjust 4mA
O2: Adjust 20mA
Use the Sidescroll Key to scroll through the Pages within each Menu
Use the Menu
Key to scroll
through
the Menus
Use the Downscroll
Key to scroll through
the Frames
within each Page
Section 7.3, Page 58
A:T.Span (5K0)
A:T.Zero (2K0)
B:mV Span (+1V)
B:T.Zero (2K0)
B:T.Span (5K0)
B:mV Zero (-1V)
Dual input analyzer only
Alter Fact.Code Option Board
Note. The Option Board page is
displayed only if the option board is fitted.
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
58 IM/AX4PH Issue 12
Factory Settings Access Code
Enter the required code number (between 0000 and 19999) to gain access to the
factory settings. If an incorrect value is entered, access to subsequent frames is
prevented and the display reverts to the top of the page.
Calibrate Sensor A
Note. The values in the display lines for sensor calibration are shown only as
examples the actual values obtained will differ.
Sensor B calibration (dual input analyzers only) is identical to Sensor A calibration.
Single input analyzers only see page 61.
Operating Page see Section 2.3, page 6.
Calibrate Input for Sensor A ?
If calibration is required select Yes otherwise select No.
Note. To abort calibration, press the key again at any time before calibration is
complete see next page.
Millivolt Zero
Set the millivolt source to 1000 mV.
The display advances automatically to the next step once a stable and valid value is
recorded.
Note. The upper 6-segment display shows the measured input voltage. Once the
signal is within range the lower 6-segment display shows the same value and Calib is
displayed to indicate that calibration is in progress.
Millivolt Span
Set the millivolt source to +1000 mV.
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (100R)
Set the temperature simulator to 100
The display advances automatically to the next step once a stable and valid value is
recorded.
Continued on next page.
Factory Set Code
0000
FACTORY SETTINGS
-----
Cal. Sensor A
-----
A:mV Zero (-1V)
-1000Calib
A:mV Span (+1V)
mV
Calib
A:T.Zero (100R)
Ohms
Calib
Cal. Sensor B
Monitoring pH
-1000mV
A: T.Zero (1K0)
1000
1000
100.0
100.0
Cal. Output 1
Calibrate I/P ?
-----
Yes
No
Yes
No
Monitoring Redox
Dual Redox
pH/Redox
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
IM/AX4PH Issue 12 59
Temperature Span (150R)
Set the temperature simulator to 150
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (1k0)
Set the temperature simulator to 1000
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (1k5)
Set the temperature simulator to 1500
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (2k0)
Set the temperature simulator to 2000
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (5k0)
Set the temperature simulator to 5000
The display returns automatically to Cal. Sensor A once a stable and valid value is
recorded.
Abort Calibration
Select Yes or No
Yes selected:
before completion of A:mV Span (+1V) frame calibration advances to A:T.Zero
(100R) and continues.
after completion of A:mV Span (+1V) frame the display returns to the Calibrate
Sensor A page.
No selected calibration continues from the point at which the key was
pressed.
A:T.Span (5K0)
A:T.Zero (2K0)
2000
2000
Ohms
Calib
5000
5000
Ohms
Calib
A:T.Span (1K5)
1500
1500
Ohms
Calib
Cal. Sensor A
A:T.Zero (1K0)
1000
1000
Ohms
Calib
A:T.Span (150R)
Ohms
Calib 150.0
150.0
A: Abort Cal.
-----
A:T.Zero (100R)
No
Yes
CAL. SENSOR A
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
60 IM/AX4PH Issue 12
Calibrate Output 1
Note. When adjusting the 4 and 20 mA outputs, the display reading is unimportant
and is used only to indicate that the output is changing when the and keys
are pressed.
See below.
Adjust 4mA
Use the and keys to set the milliammeter reading to 4 mA.
Note. The analog output range selected in Configure Outputs (Section 5.6) does not
affect the reading.
Adjust 20mA
Use the and keys to set the milliammeter reading to 20 mA.
Note. The analog output range selected in Configure Outputs (Section 5.6) does not
affect the reading.
See below.
Operating Page see Section 2.3, page 6.
Calibrate Output 2
Note. Output 2 calibration is identical to Output 1 calibration.
Option board fitted and analog features enabled continued on next page.
Option board fitted, additional features disabled continued on next page.
Option board not fitted continued on next page.
Option board fitted and analog features enabled continued on next page.
Option board fitted, additional features disabled continued on next page.
Option board not fitted continued on next page.
Operating Page see Section 2.3, page 6.
Cal. Output 1
-----
O1: Adjust 4mA
16000
O1: Adjust 20mA
7200
Monitoring pH
Cal. Output 2 Cal. Output 1
Cal. Output 2
Cal. Output 2
-----
O1: Adjust 20mA
7200
Monitoring pH
Cal. Output 3
Cal. Output 2
Cal. Output 3
Option Board
Option Board
Alter Fact. Code
Alter Fact. Code
Redox (ORP)
Dual Redox
pH/Redox (ORP)
Redox (ORP)
Dual Redox
pH/Redox (ORP)
Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468 7 Calibration
IM/AX4PH Issue 12 61
Calibrate Output 3
Note.
Output 3 (and Output 4) calibration is applicable only if the Option board is fitted
and analog features enabled see below.
Output 3 calibration is identical to Output 2 calibration.
Calibrate Output 4
Note. Output 4 calibration is identical to Output 3 calibration.
Configure Option Board
Note.
This frame is displayed only if an option board is fitted.
The software detects if an option board is fitted but cannot detect the additional
features available.
If an option board is fitted, the correct selection must be made below to enable
use of the available features. If an incorrect selection is made, the software
menus and frames associated with that option are displayed in the Operating
and Configuration pages but the features do not work.
Use the and keys to enable the features for the type of option(s) fitted:
Analog
Pb Dp
Analog + Pb Dp