Libra S11 and S12 User Manual: English
Libra S11 and S12 User Manual: English
Libra S11 and S12 User Manual: English
User Manual
English
Deutsch
Françai
Españo
Italiano
Biochrom Ltd
Certificate No. 890333
Declaration of Conformity
This is to certify that the Libra S11 Visible and Libra 12 UV/Visible
Spectrophotometers
Part number 80-2115-15 and 80-2115-10
Serial number 81000 onwards
EN 61 010-1: 2001
Safety requirements for electrical equipment for measurement, control and
laboratory use.
EN 61326: 1998
Electrical equipment for measurement, control and laboratory use – EMC
requirements
David Parr
Managing Director
Biochrom Ltd
OPERATION 3
Introduction 3
Using the Instrument Display and Keypad 4
Customisation of the instrument menu 5
Basic Modes of Use 6
Enhanced Modes of Use 9
Method storage, recall and deletion 13
SET-UP 15
Menu customisation, access code and methods 15
Lamp settings 16
Display contrast and instrument output 16
ERROR MESSAGES 17
OUTPUT OF RESULTS 18
Use with parallel printer 18
Use with chart recorder 18
Use with PC 18
ACCESSORIES 19
Lamps, consumables and other items 19
MAINTENANCE 20
After Sales Support 20
Lamp Replacement 20
Deuterium Lamp Warranty (Libra S12) 23
Fuse replacement 23
Cleaning and general care of the instrument 23
APPENDIX 24
Equation entry using the Multi Wavelength mode 24
SPECIFICATION 25
Warranty 26
Unpacking, Positioning and Installation
• Inspect the instrument for any signs of damage caused in transit. If any damage
is discovered, inform your supplier immediately. Check the position of the
metal lamp bracket inside the lamp access area.
• The instrument must be placed on a hard, flat bench or table that can take its
weight (6 kg) such that air is allowed to circulate freely around the instrument.
• Ensure that the cooling fan inlets and outlets are not obstructed; position at least
2 inches from the wall.
• This equipment must be connected to the power supply with the power cord
supplied and MUST BE EARTHED (GROUNDED). It can be used on 90 -
240V supplies.
• Switch on the instrument. Prior to calibration, the display asks you to check that
the cell compartment is clear. The purpose of this is to indicate the use of the
function soft keys, and how they are associated with the options presented at the
bottom of the display; F2 represents OK in this instance (this display can be
disabled in Set-up if required). The calibration stages are indicated in sequence
(- for checking, 9 for OK, 8 for Fail).
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 1
Essential Safety Notes
There are a number of warning labels and symbols on your instrument. These are
there to inform you where potential danger exists or particular caution is required.
Before commencing installation, please take time to familiarise yourself with these
symbols and their meaning.
WARNING WARNING
___________________________________________________________________
2 Libra S11/S12, English Issue 04 – 07/2006
OPERATION
Introduction
Your spectrophotometer is a simple-to-use, microprocessor-controlled instrument.
In addition to the basic modes of operation, the instrument has enhanced software
and method storage functionality. A laboratory technician or supervisor can
customise the spectrophotometer for students and operators by disabling menu
options that are not required.
Your spectrophotometer:
• can be linked via a serial interface adapter lead to a PC for download of results
to spreadsheet, and subsequent inclusion in a laboratory information
management system (LIMS)
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 3
Using the Instrument Display and Keypad
The back-lit liquid crystal display has large characters which are easily visible,
useful if a group of students are gathered around for a demonstration, for example.
The keypad is a spill proof membrane that is very hard wearing.
The instrument is easy to use, with function select / entry soft keys on the keypad
(F1, F2 and F3) being situated directly below the corresponding option on the
display; these keys are used in conjunction with on-screen prompts.
On the absorbance home page, for example, concise help text is available at the push
of the F1 key, whereas Menu and Set-up are accessed by F2 and F3, respectively.
On other displays, the purpose of the keys changes, but this is clearly indicated; for
example, F3 acts as the “accept” option on parameter entry displays and next page (if
further options exist) in the Menu and Set-up displays.
Pressing the red stop key acts as an escape mechanism in most situations.
___________________________________________________________________
4 Libra S11/S12, English Issue 04 – 07/2006
Press:
• to print result
To customise the instrument, refer to Set-up > Menu for further details.
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 5
Basic Modes of Use
Absorbance
% Transmission
Transmission mode measures the amount of light that has passed through a sample
relative to a blank (this can be air), but displays the result as a percentage. The
procedure is as follows:
Concentration
___________________________________________________________________
6 Libra S11/S12, English Issue 04 – 07/2006
Time Intervals
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 7
Wavescan
An absorption spectrum can be obtained from your instrument; this enables simple
identification of peak height and position. A reference scan has to be obtained first
since there is no stored baseline. The procedure is as follows:
Use the 3 (F1) and 4 (F3) keys to move the cursor in order to identify peak height
and position.
To zoom in on a region, press F2 followed by the start and end wavelengths (the
instrument will zoom to the nearest 10, 20, 50 or 100 nm).
___________________________________________________________________
8 Libra S11/S12, English Issue 04 – 07/2006
Enhanced Modes of Use
Standard Concentration
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 9
Standard Curve
Select whether cubic spline or linear regression fit of the data points is required
Enter appropriate wavelength
Input the number of standards to be used:
___________________________________________________________________
10 Libra S11/S12, English Issue 04 – 07/2006
Reaction Rate
Reagent test kits are routinely used for the enzymatic determination of compounds in
food, beverage and clinical laboratories by measuring NAD / NADH conversion at
340 nm. The change in absorbance over a specified time period can be used to
provide useful information when an appropriate factor, defined in the reagent kit
protocol, is applied.
Note that reaction rate and enzyme activity can be calculated if the factor used takes
account of the absorbance difference per unit time, as opposed to the absorbance
difference per se.
The correlation (quality of line fit) is calculated from 10 equally spaced absorbance /
time points during the course of the experiment. The procedure is as follows:
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 11
Multi Wavelength Equation Entry
The measurement of Absorbance values at specific wavelengths and combining these with
appropriate factors is a means of overcoming interference effects in several applications. By
using the equation entry facility, post measurement calculations can be done automatically
and the end result displayed for the operator. This is a very powerful facility indeed for the
busy industrial, QC or environmental testing laboratory. Up to 5 wavelengths and 6 constants
(5 factors relating to the absorbances at the 5 wavelengths, and a dilution factor) can be
entered for one equation. The procedure is as follows (refer to Appendix for a step by step,
worked through example):
Write the equation out in front of you, ensuring there are no syntax errors!
The available equation operators are:
A@1 Absorbance at wavelength 1. Enter required wavelength using keypad.
T@1 Transmission at wavelength 1. Enter required wavelength using keypad.
( ) + - * /
K1 Factor applied to absorbance at wavelength 1. Enter using keypad.
C1 Constant (dilution or other). Enter using keypad. Note that C1 can be applied
several times, using different numerical values each occasion.
! Use if a sequence of Absorbance values only is required Enter !A@1 A@2 etc
Note that Factor and Constant can only have 5 characters, including the decimal point; thus
12.259 is not allowed, whereas 0.302 is.
The maximum length of equation that can be accommodated is 60 characters in length, where
the absorbance at wavelength and factor are 3 and 5 characters, respectively, and the equation
operators are 1 character.
___________________________________________________________________
12 Libra S11/S12, English Issue 04 – 07/2006
Ratio
This facility enables the determination of Abs λ1 / Abs λ2 and Abs λ*factor
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 13
Method storage, recall and deletion
After defining parameters in any of the above modes, and prior to measuring a
sample, entry to Set-up using the F3 function key provides the opportunity to store
the parameters currently loaded as a method. This option is password protected, and
up to 9 methods can be saved; refer to Set-up > Methods for further details.
When recalling a stored method from the menu, the option to print the method
parameters is presented by pressing 1; press 2 to continue with the selected method.
Print the method to confirm that it is the method you require, if necessary. Once a
method has been recalled, you can set reference before running samples.
If method parameters are incorrect, they cannot be changed – the method has to be
erased (deleted), re-entered and then re-stored (re-saved). To remove a method,
refer to Set-up > Methods.
___________________________________________________________________
14 Libra S11/S12, English Issue 04 – 07/2006
SET-UP
After selecting the set-up option (F3) there is an initial information screen, as shown
below. Press F2 (OK) to return to the absorbance home page. Press F1 to
recalibrate the instrument.
To access the set-up page press F3 again. A password is required; the default is
6020 or 6040, but this can be changed.
Three displays are available: Menu customisation, access code and methods
Lamp settings
Display contrast and instrument output
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 15
4 : Methods Press 4 to have the choice of either storing a new method
after defining parameters or erasing an existing method.
Press 1 to store the method in the next available method
storage space (maximum is 9). Methods are stored in the
instrument EEPROM, the process may take a few seconds.
Press 2 to erase a method; the method number has to be
entered.
Method parameters can be printed out when the stored
method is selected from the menu.
Lamp settings
1 : UV lamp Switch deuterium lamp on/off (Libra S12)
2 : Vis lamp only Deuterium lamp is switched permanently off; and the
instrument will act as a visible only product. Infrequent
users of the UV range can benefit from considerable
increase in deuterium lamp life by use of this utility.
3 : UV lamp save Instrument powers up and calibrates as usual, but then
switches off the deuterium lamp automatically. The lamp
will come on if a UV wavelength is selected (Libra S12)
4 : Lamp hours ⇒ 0 Resets lamp lives to zero when a new lamp is fitted.
___________________________________________________________________
16 Libra S11/S12, English Issue 04 – 07/2006
ERROR MESSAGES
Reset to defaults Memory was corrupted in some way and re-set to defaults.
Display options and methods need to be re-entered.
UV lamp fail UV lamp failed to strike. May need replacing – check lamp
hours (Libra S12)
Vis lamp fail Visible lamp failed to strike. May need replacing – check
lamp hours
Beam blocked Something is in the way of the beam – check sample
compartment area
Wavelength error A calibration failure or corruption has caused the instrument
to go an invalid wavelength. May need service engineer.
Lamps overheating The thermal sensor on the lamp cover has detected a
temperature in excess of its limits – call service engineer.
PSU overheating PSU thermistor is indicating a temperature in excess of
70°C – call service engineer.
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 17
OUTPUT OF RESULTS
Use with parallel printer
Any Centronics parallel printer can be used together with the appropriate cable. If
using a thermal printer, ensure it is set up to print out for a page width of 80
characters. Ensure output to printer is on in the Set-up.
Output is automatic when the key is pressed, and a printer is connected and
switched on. Umlauts and accents are not printed out with letters if the instrument is
set up to be in German, French, Italian or Spanish.
Appropriate headers and relevant information are printed out for enhanced modes,
for example the absorbance – concentration values of the standards in standard curve
mode, and the equation (with values) entered in Multi Wavelength mode.
Use with PC
NOTE: A standard serial interface will not work.
1) Download to Spreadsheet
The serial interface adapter lead (80-2109-02) is required; it is also supplied with
Spreadsheet Interface Software for direct download to Excel. This macro is supplied
on a floppy disc together with instructions for installation and use.
2) Use with Hyperterminal
The serial interface adapter lead (80-2109-02) is required; ensure output to serial is
on in Set-up. The ASCII stream is output at 19,200 Baud via the 25 way D
connector on the rear panel, and can be picked up by a PC with Windows installed.
Use the Hyperterminal emulator in Accessories to pick this up (settings are
Handshake None, 19,200 Baud, 1 stop bit, 8 data bits, 0 parity, Comm port depends
on which port the lead is connected to). Output is automatic if the interface lead is
connected to the instrument.
___________________________________________________________________
18 Libra S11/S12, English Issue 04 – 07/2006
ACCESSORIES
Each accessory is supplied integrated into its own sample compartment for ease of
fitting and cleaning.
Easy to fit - when changing accessory / sample compartment, snap the old one
out and the new one in.
Easy to clean - take the whole assembly out and run it under the tap.
Contact your supplier for details on our range of disposable, UV silica and glass
cells.
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 19
MAINTENANCE
After Sales Support
We supply support agreements that help you to fulfil the demands of regulatory
guidelines concerning GLP/GMP.
♦ Preventative maintenance
♦ Certification
When using calibration standard filters, insert such that the flat surface is facing
away from the spring end of the cell holder
Lamp Replacement
Replacement lamps are available from your supplier using the following part
numbers:
The design of the lamp area is such that users are able to change their own lamps.
No lamp alignment is necessary as the lamps are pre-aligned at manufacture.
The lamps become very hot in use. Ensure they cool before changing them.
Do not touch the optical surfaces of either lamp with your fingers (use tissue); if
touched, the area should be cleaned with iso-propanol.
___________________________________________________________________
20 Libra S11/S12, English Issue 04 – 07/2006
To replace a lamp proceed as follows:
1) Switch off the instrument, remove the sample from the cell holder and
disconnect the power supply cord.
2) Locate the lamp access cover at the back of the instrument, unscrew the cover
and remove.
3) Move the metal bracket sideways, slide the lamp plate assembly out and
unplug the connector.
- if the tungsten lamp has failed, the replacement should be inserted onto the
plate, pushing it all the way down into its holder.
- if the deuterium lamp has failed, insert the old tungsten lamp onto the plate as
above and then replace the whole assembly with the new one.
4) Reconnect the cable connector and slide the lamp plate in until it locates,
checking that the cable or connector does not interfere with the relocation.
5) If you have difficulty in sliding the lamp assembly back into position hold the
connector down and push the lamp plate until it locates correctly.
6) Replace the lamp access cover.
7) Reconnect the power supply cord and switch the instrument on.
8) Reset the lamp life to zero by:
F3 Set-up → F3 Set-up → enter password → F3, F3 → select 4 → F3
(9), → select which lamp life is to be changed to zero.
Exit this screen by pressing the key
Exit set-up by pressing the key.
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 21
Lamp Replacement Libra S11 Only – proceed as follows
Tungsten Lamp 80-2022-94 (Libra S11)
(use only this tungsten lamp; others will not operate correctly in a this
spectrophotometer)
4) The replacement tungsten lamp should be inserted onto the plate, pushing it
all the way down into its holder. Use a lens cloth to hold the new lamp. Again
the brown wires can be gripped and pulled towards you to ease this process.
Ensure the lamp is located as far inside the holder as possible. It should also
appear horizontally positioned.
5) Replace the lamp access cover.
6) Reconnect the power supply cord and switch the instrument on.
7) Reset the lamp life to zero by:
F3 Set-up → F3 Set-up → enter password → F3, F3 → select 4 → F3 (9),
→ select which
lamp life is to be changed to zero.
Exit this screen by pressing the key
Exit set-up by pressing the key.
___________________________________________________________________
22 Libra S11/S12, English Issue 04 – 07/2006
Deuterium Lamp Warranty (Libra S12)
Criteria for lamp replacement are that it must:
Fuse replacement
Switch off the instrument and disconnect the power supply cord. The fuse holder
can only be opened if the power supply plug has been removed, and is located
between the power input socket and the on/off switch on the back panel of the
instrument.
Place fuses (2A, 5mm x 20mm, FST) into the fuse holder and slide back into
position.
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 23
APPENDIX
Equation entry using the Multi Wavelength mode
Always write out the equation in front of you before using this mode.
Step by step entry of the following equation is shown in the example below:
Note that if you make an error, the ← key on the keypad will remove the last entry.
___________________________________________________________________
24 Libra S11/S12, English Issue 04 – 07/2006
SPECIFICATION
___________________________________________________________________
Issue 04 - 07/2006 Libra S11/S12, English 25
Warranty
Your supplier guarantees that the product supplied has been thoroughly tested to
ensure that it meets its published specification. The warranty included in the
conditions of supply is valid for 12 months only if the product has been used
according to the instructions supplied. They can accept no liability for loss or
damage, however caused, arising from the faulty or incorrect use of this product.
This product has been designed and manufactured by Biochrom Ltd, 22 Cambridge
Science Park, Milton Road, Cambridge CB4 0FJ, UK.
___________________________________________________________________
26 Libra S11/S12, English Issue 04 – 07/2006