DOC316.53.01027 10ed
DOC316.53.01027 10ed
DOC316.53.01027 10ed
01027
Test preparation
Instrument-specific information
Table 1 shows sample cell and orientation requirements for reagent addition tests, such
as powder pillow or bulk reagent tests. Table 2 shows sample cell and adapter
requirements for AccuVac Ampul tests. The tables also show all of the instruments that
have the program for this test.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information for reagent addition
Instrument Sample cell orientation Sample cell
DR 6000 The fill line is to the right. 2495402
DR 3800
DR 2800
DR 2700
DR 1900
DR 5000 The fill line is toward the user.
DR 3900
DR 900 The orientation mark is toward the user. 2401906
1
Before starting
Analyze the samples immediately. The samples cannot be preserved for later analysis.
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
If the test result is over-range, or if the sample temporarily turns yellow after the reagent addition, dilute the sample with a
known volume of high quality, chlorine demand-free water and do the test again. Some loss of chlorine may occur due to the
dilution. Multiply the result by the dilution factor. Additional methods are available to measure chlorine without dilution.
For chloramination disinfection control, use one of the available Chloramine (Mono) methods.
For the best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water
in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results
automatically with the reagent blank adjust option.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
The SwifTest Dispenser for Total Chlorine can be used in place of the powder pillow in the test procedure. One dispensation
is equal to one powder pillow for 10-mL samples.
An AccuVac Ampul for Blanks can be used to zero the instrument in the AccuVac test procedure.
Items to collect
Powder pillows
Description Quantity
AccuVac Ampuls
Description Quantity
®
DPD Total Chlorine Reagent AccuVac Ampul 1
Beaker, 50-mL 1
Sample cells (For information about sample cells, adapters or light shields, refer to Instrument-
1
specific information on page 1.)
Stopper for 18-mm tubes and AccuVac Ampuls 1
Sample collection
• Analyze the samples immediately. The samples cannot be preserved for later
analysis.
• Chlorine is a strong oxidizing agent and is unstable in natural waters. Chlorine reacts
quickly with various inorganic compounds and more slowly with organic compounds.
Many factors, including reactant concentrations, sunlight, pH, temperature and
salinity influence the decomposition of chlorine in water.
• Collect samples in clean glass bottles. Do not use plastic containers because these
can have a large chlorine demand.
Start
1. Start program 80 2. Fill a sample cell with 3. Prepare the sample: 4. Swirl the sample cell for
Chlorine F&T PP. For 10 mL of sample. Add the contents of one 20 seconds to mix.
information about sample powder pillow to the sample A pink color shows if
cells, adapters or light cell. chlorine is present in the
shields, refer to Instrument- sample.
specific information
on page 1.
5. Start the instrument 6. Prepare the blank: Fill a 7. Clean the blank sample 8. Insert the blank into the
timer. A 3‑minute reaction second sample cell with cell. cell holder.
time starts. 10 mL of sample.
Prepare the sample blank
and set the instrument to
zero during the reaction
time.
9. Push ZERO. The display 10. Clean the prepared 11. Within 3 minutes after 12. Push READ. Results
shows 0.00 mg/L Cl2. sample cell. the timer expires, insert the show in mg/L Cl2.
prepared sample into the
cell holder.
Start
1. Start program 85 2. Prepare the blank: Fill 3. Prepare the sample: 4. Quickly invert the
Chlorine F&T AV. For the sample cell with 10 mL Collect at least 40 mL of AccuVac Ampul several
information about sample of sample. sample in a 50‑mL beaker. times to mix.
cells, adapters or light Fill the AccuVac Ampul with
shields, refer to Instrument- sample. Keep the tip
specific information immersed while the
on page 1. AccuVac Ampul fills
completely.
Zero
5. Start the instrument 6. Clean the blank sample 7. Insert the blank into the 8. Push ZERO. The display
timer. A 3‑minute reaction cell. cell holder. shows 0.00 mg/L Cl2.
time starts.
Prepare the sample blank
and set the instrument to
zero during the reaction
time.
9. Clean the AccuVac 10. Within 3 minutes after 11. Push READ. Results
Ampul. the timer expires, insert the show in mg/L Cl2.
prepared sample AccuVac
Ampul into the cell holder.
Interferences
Interfering substance Interference level
Acidity More than 150 mg/L CaCO3. The full color may not develop or the color may fade instantly. Adjust
to pH 6–7 with 1 N Sodium Hydroxide. Measure the amount to add on a separate sample aliquot,
then add the same amount to the sample that is tested. Correct the test result for the dilution from
the volume addition.
Alkalinity More than 250 mg/L CaCO3. The full color may not develop or the color may fade instantly. Adjust
to pH 6–7 with 1 N Sulfuric Acid. Measure the amount to add on a separate sample aliquot, then
add the same amount to the sample that is tested. Correct the test result for the dilution from the
volume addition.
Bromine, Br2 Positive interference at all levels
Chlorine Dioxide, ClO2 Positive interference at all levels
Inorganic chloramines Positive interference at all levels
Chloramines, organic May interfere in the result for total chlorine analysis
Hardness No effect at less than 1000 mg/L as CaCO3
Manganese, Oxidized Pre-treat the sample as follows:
(Mn4+, Mn7+) or
Chromium, Oxidized 1. Adjust the sample pH to 6–7.
(Cr6+) 2. Add 3 drops of Potassium Iodide (30-g/L) to 10 mL of sample.
3. Mix and wait 1 minute.
4. Add 3 drops of Sodium Arsenite (5-g/L) and mix.
5. Use the test procedure to measure the concentration of the treated sample.
6. Subtract this result from the result without the treatment to obtain the correct chlorine
concentration.
Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
1. Use the test procedure to measure the concentration of the sample, then keep the
(unspiked) sample in the instrument.
2. Go to the Standard Additions option in the instrument menu.
3. Select the values for standard concentration, sample volume and spike volumes.
4. Open the standard solution.
5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL and
0.3 mL of the standard solution, respectively, to three 10-mL portions of fresh sample.
Mix well.
®
Note: For AccuVac Ampuls, add 0.4 mL, 0.8 mL and 1.2 mL of the standard solution to three
50-mL portions of fresh sample.
6. Use the test procedure to measure the concentration of each of the spiked samples.
Start with the smallest sample spike. Measure each of the spiked samples in the
instrument.
7. Select Graph to compare the expected results to the actual results.
Note: If the actual results are significantly different from the expected results, make sure that
the sample volumes and sample spikes are measured accurately. The sample volumes and
sample spikes that are used should agree with the selections in the standard additions menu. If
the results are not within acceptable limits, the sample may contain an interference.
Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users can get different
results under different test conditions.
Program Standard Precision (95% Confidence Interval) Sensitivity
Concentration change per 0.010 Abs change
80 1.25 mg/L Cl2 1.23–1.27 mg/L Cl2 0.02 mg/L Cl2
85 1.25 mg/L Cl2 1.21–1.29 mg/L Cl2 0.02 mg/L Cl2
Summary of method
Chlorine can be present in water as free chlorine and as combined chlorine. Both forms
can exist in the same water and be determined together as total chlorine. Free chlorine is
present as hypochlorous acid and/or hypochlorite ion. Combined chlorine exists as
monochloramine, dichloramine, nitrogen trichloride and other chloro derivatives. The
combined chlorine oxidizes iodide in the reagent to iodine. The iodine and free chlorine
react with DPD (N,N-diethyl-p-phenylenediamine) to form a pink color which is
proportional to the total chlorine concentration.
To find the concentration of combined chlorine, run a free chlorine test and a total
chlorine test. Subtract the results of the free chlorine test from the total chlorine test to
obtain the combined chlorine concentration. The measurement wavelength is 530 nm for
spectrophotometers or 520 nm for colorimeters.
Consumables and replacement items
Required reagents
Recommended standards
© Hach Company/Hach Lange GmbH, 1989–2014, 2018. All rights reserved. 09/2018, Edition 10