ChlorineTotalDPDTNT DOC316.53.01028
ChlorineTotalDPDTNT DOC316.53.01028
ChlorineTotalDPDTNT DOC316.53.01028
01028
Test preparation
Before starting
Samples must be analyzed immediately after collection and cannot be preserved for later analysis.
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
DR 3900, DR 3800, DR 2800 and DR 2700: Install the light shield in Cell Compartment #2 before this test is started.
For best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water in
the test procedure to get the reagent blank value. Subtract the reagent blank value from the sample results automatically
with the reagent blank adjust option.
For chloramination disinfection control, use one of the available Chloramine (Mono) methods.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used and use any recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Use the Safety Data Sheets for disposal
information for unused reagents. Consult the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
Light shield (For information about sample cells, adapters or light shields, refer to Instrument
1
specific information on page 1.)
Test 'N Tube™ DPD Total Chlorine Reagent 1
Wipes, disposable varies
1
Refer to Consumables and replacement items on page 5 for reorder information.
Sample collection
• Samples must be analyzed immediately and cannot be preserved for later analysis.
• Chlorine is a strong oxidizing agent and it is unstable in natural waters. It reacts
quickly with various inorganic compounds and more slowly with organic compounds.
Many factors, including reactant concentrations, sunlight, pH, temperature and
salinity influence the decomposition of chlorine in water.
• Collect samples in clean glass bottles. Avoid plastic containers since these may have
a large chlorine demand.
• Pre-treat glass sample containers to remove any chlorine demand. Soak the
containers in a dilute bleach solution (1 mL commercial bleach to 1 liter of deionized
water) for at least 1 hour. Rinse thoroughly with deionized or distilled water. If sample
containers are rinsed thoroughly with deionized or distilled water after use, only
occasional pre-treatment is necessary.
• Be sure to get a representative sample. If the sample is taken from a spigot or faucet,
let the water flow for at least 5 minutes. Then let the container overflow with the
sample several times and then put the cap on the sample container so that there is
no headspace (air) above the sample. If a sample cell is used, rinse the cell several
times with the sample, then carefully fill to the 10-mL mark.
Start
1. Start program 2. Prepare the blank: Fill 3. Clean the blank vial. 4. Insert the blank vial into
89 Chlorine F&T TNT. For an empty Test 'N Tube vial the 16-mm cell holder.
information about sample to the top of the label with
cells, adapters or light sample.
shields, refer to Instrument
specific information
on page 1.
Note: Although the program
name may vary between
instruments, the program
number does not change.
5. Push ZERO. The display 6. Prepare the sample: 7. Put the cap on the vial. 8. Start the instrument
shows 0.00 mg/L Cl2. Remove the cap from a Invert to mix. timer. A 3-minute reaction
Total Chlorine DPD Test N' time starts.
Tube. Fill the vial to the top
of the label with sample.
Read
9. Clean the sample vial. 10. When the timer expires, 11. Push READ. Results
insert the prepared sample show in mg/L Cl2.
into the cell holder.
Interferences
Interfering substance Interference level
Acidity More than 150 mg/L CaCO3. The full color may not develop or the color may fade instantly.
Adjust to pH 6–7 with 1 N Sodium Hydroxide. Measure the amount to be added on a separate
sample aliquot, then add the same amount to the sample that is tested. Correct the test result for
the dilution from the volume addition.
Alkalinity More than 250 mg/L CaCO3. The full color may not develop or the color may fade instantly.
Adjust to pH 6–7 with 1 N Sulfuric Acid. Measure the amount to add on a separate sample
aliquot, then add the same amount to the sample that is tested. Correct the test result for the
dilution from the volume addition.
Bromine, Br2 Interferes at all levels
Chlorine Dioxide, ClO2 Interferes at all levels
Chloramines, organic May interfere
Hardness No effect at less than 1000 mg/L as CaCO3
Iodine, I2 Interferes at all levels
Manganese, Oxidized Pre-treat the sample as follows:
(Mn4+, Mn7+) or
Chromium, Oxidized 1. Adjust the sample pH to 6–7.
(Cr6+) 2. Add 3 drops of Potassium Iodide (30-g/L) to 10 mL of sample.
3. Mix and wait 1 minute.
4. Add 3 drops of Sodium Arsenite (5-g/L) and mix.
5. Use the test procedure to measure the concentration of the treated sample.
6. Subtract this result from the result without the treatment to obtain the correct chlorine
concentration.
1. Use the test procedure to measure the concentration of the sample, then keep the
(unspiked) sample in the instrument.
2. Go to the Standard Additions option in the instrument menu.
3. Select the values for standard concentration, sample volume and spike volumes.
4. Open the standard solution.
5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL and
0.3 mL of the standard solution, respectively, to three 10-mL portions of fresh sample.
Mix well.
6. Use the test procedure to measure the concentration of each of the spiked samples.
Start with the smallest sample spike. Measure each of the spiked samples in the
instrument.
7. Select Graph to compare the expected results to the actual results.
Note: If the actual results are significantly different from the expected results, make sure that
the sample volumes and sample spikes are measured accurately. The sample volumes and
sample spikes that are used should agree with the selections in the standard additions menu. If
the results are not within acceptable limits, the sample may contain an interference.
Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users may get different
results under different test conditions.
Program Standard Precision (95% Confidence Interval) Sensitivity
Concentration change per 0.010 Abs change
89 2.68 mg/L Cl2 2.63–2.73 mg/L Cl2 0.03 mg/L Cl2
Summary of method
Chlorine can be present in water as free chlorine and as combined chlorine. Both forms
can exist in the same water and be determined together as total chlorine. Free chlorine is
present as hypochlorous acid and/or hypochlorite ion. Combined chlorine exists as
monochloramine, dichloramine, nitrogen trichloride and other chloro derivatives. The
combined chlorine oxidizes iodide in the reagent to iodine. The iodine and free chlorine
react with DPD (N,N-diethyl-p-phenylenediamine) to form a pink color which is
proportional to the total chlorine concentration.
© Hach Company/Hach Lange GmbH, 1989–2013. All rights reserved. 04/2013, Edition 8