Nitrogen, Total: Persulfate Digestion Method Method 10208 1 To 16 MG/L N (LR) Tntplus 826
Nitrogen, Total: Persulfate Digestion Method Method 10208 1 To 16 MG/L N (LR) Tntplus 826
Nitrogen, Total: Persulfate Digestion Method Method 10208 1 To 16 MG/L N (LR) Tntplus 826
01087
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows the adapter and light shield requirements for the applicable instruments that can
use TNTplus vials.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information for TNTplus vials
Instrument Adapters Light shield
DR 6000, DR 5000 — —
DR 3900 — LZV849
DR 3800, DR 2800 — LZV646
DR 1900 9609900 or 9609800 (A) —
Before starting
DR 3900, DR 3800, DR 2800: Install the light shield in Cell Compartment #2 before this test is started.
Review the safety information and the expiration date on the package.
The recommended sample pH is 3–12.
The sample temperature must be 15–25 °C (59–77 °F) for accurate results.
The recommended temperature for reagent storage is 15–25 °C (59–77 °F).
Important: Make sure to close each reagent bottle immediately after each use.
The 20-mm reaction tube can be used for 7 tests. After each use, clean the tube thoroughly with a brush and water, then
rinse well with high-quality distilled water and let dry.
Use only high-quality deionized water or organic-free water for standards preparation, sample dilutions or reagent blanks.
DR 1900: Go to All Programs>LCK or TNTplus Methods>Options to select the TNTplus number for the test. Other
instruments automatically select the method from the barcode on the vial.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
1
Items to collect (continued)
Description Quantity
Test procedure
1. Set the DRB200 reactor 2. Add 1.3 mL of sample, 3. Insert the reaction tube in 4. Keep the reaction tubes
power to on. Set the 1.3 mL of Solution A and 1 the preheated DRB200 in the reactor for
temperature to 120 °C. Reagent B tablet in quick reactor. Close the lid. 30 minutes.
succession to a dry 20‑mm
reaction tube.
Close the reaction tube
immediately. Do not invert.
Start
9. Quickly tighten the cap 10. Start the reaction time 11. When the timer expires, 12. DR 1900 only: Select
on the vial and invert until of 15 minutes. clean the vial. program 826. Refer to
completely mixed. Before starting on page 1.
1. Use deionized water as the sample in the test procedure to measure the reagent
blank value.
2. Set the reagent blank function to on. The measured reagent blank value is shown.
Interferences
A light pink color may develop during the reaction. This color does not interfere with the
analysis. Table 2 shows that the ions were individually examined to the given
concentrations and do not cause interference. No cumulative effects or influences of
other ions were found.
Table 2 Interfering substances
Interfering substance Interference level
COD 400 mg/L
Chloride 800 mg/L
Accuracy check
Standard solution method
Use the standard solution method to validate the test procedure, the reagents and the
instrument.
Items to collect:
• Nitrogen, Ammonia Standard Solution, 10-mg/L NH3–N
1. Use the test procedure to measure the concentration of the standard solution.
2. Compare the expected result to the actual result.
Note: The factory calibration can be adjusted slightly with the standard adjust option so that the
instrument shows the expected value of the standard solution. The adjusted calibration is then
used for all test results. This adjustment can increase the test accuracy when there are small
variations in the reagents or instruments.
Summary of Method
Inorganic and organic nitrogen compounds are digested with peroxodisulfate and are
oxidized to nitrate. The nitrate ions react with 2,6-dimethylphenol in a solution of sulfuric
and phosphoric acid to form a nitrophenol. The measurement wavelength is 345 nm.
Consumables and replacement items
Required reagents
Required apparatus
DRB 200 Reactor, 115 VAC option, 9 x 13 mm + 2 x 20 mm, 1 block 1 each DRB200-01
DRB 200 Reactor, 230 VAC option, 9 x 13 mm + 2 x 20 mm, 1 block 1 each DRB200-05
Pipet, adjustable volume, 1.0–5.0 mL 1 each BBP065
Pipet tips, for 1.0–5.0 mL pipet 1 75/pkg BBP068
Pipet, adjustable volume, 0.2–1.0 mL 1 each BBP078
Pipet tips, for 0.2–1.0 mL pipet 2 100/pkg BBP079
Test tube rack for 20-mm vials, 20 holes 1 each 2497912
Recommended standards
© Hach Company/Hach Lange GmbH, 2007–2018. All rights reserved. 08/2018, Edition 10