Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Shalya Vista

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Add Hi-Tech Precision to your Surgical Expertise...

H i g h P e r f o r m a n c e

With all Reusable Attachments

Complete range of
Electro Surgical Solutions

Electrosurgical Generator
Technical Specifications of
MONOPOLAR CUT : SEVEN Modes : 390 kHz MONOPOLAR COAG : SEVEN Modes : 460 kHz
Low Cut : Continues output, 300 W at 300 Ω, CF 1.5, Low Volt Soft : Continues output, 300 W at 300 Ω, CF 1.5, Low Volt
Pure Cut : Continues output, 400 W at 300 Ω, CF 1.5 Desiccate : Repetition Frequency 83 kHz, 200 W at 500 Ω, CF 5.0
Blend1 : Bursts at 29 kHz, 250 W at 300 Ω, CF 2 Swift : Repetition Frequency 53 kHz, 120 W at 500 Ω, CF 6.2
Blend2 : Bursts at 31 kHz, 200 W at 300 Ω, CF 2.6 Fulgurate : Repetition Frequency 29 kHz, 120 W at 500 Ω, CF 7.0
Endo1 : Pulse output, 400 W at 300 Ω, CF 1.5, Low Pulse Intr. Randomized Spray : Repetition Frequency 24 kHz, 120 W at 500 Ω, CF 8.0
Endo2 : Pulse output, 400 W at 300 Ω, CF 1.5, High Pulse Intr. Hi-Spray : Repetition Frequency 13 kHz, 70 W at 500 Ω, CF 10.0
Hydro : Continues output, 400 W at 300 Ω, CF 1.5, Rapid Hydro : Repetition Frequency 29 kHz, 120 W at 500 Ω, CF 7.0

BIPOLAR : FIVE Modes SALINE RESECTION : SIX Modes


Bipolar Cut Modes: 440 kHz Saline Cut Modes: 390 kHz
Bip Cut : Continues output, 120 W at 100 Ω, CF 1.5, Low Volt Cut I : Continues output 300 W at 100 Ω, CF 1.5, Low Volt
Macro Cut : Continues output, 120 W at 100 Ω, CF 1.5, Cut II: Continues output 300 W at 100 Ω, CF 1.5, Medium Volt
High Volt Cut III : Continues output 300 W at 100 Ω, CF 1.5, High Volt
Bipolar Coag Modes: 440 kHz Saline Coag Modes: 440kHz
Micro : Continues output, 120 W at 100 Ω, CF 1.5, Low Volt Soft: Continues output 200 W at 100 Ω, CF 1.5, Low Voltage
Standard : Continues output, 120 W at 100 Ω, CF 1.5, Standard: Continues output 200 W at 100 Ω, CF 1.5, Medium Volt
Medium Volt Force: Continues output 200 W at 100 Ω, CF 1.5, High Volt
Force : Continues output, 120 W at 100 Ω, CF 1.5, High Volt

VESSEL SEALING : TWO Modes : 440 kHz PREM SYSTEM


Fine Seal : Low Pulse output, 5 Levels, 150 W at 100 Ω, CF 1.5 Measuring Frequency : 80 kHz ± 10 kHz, Measuring current : <10 µA
Rapid Seal : High Pulse output, 5 Levels, 150 W at 100 Ω, CF 1.5 Acceptable Rang : Dual Area : 5-135 Ω (Adaptive), Single Area : <10 Ω

FEEDBACK SYSTEM DESIGN STANDARD


• 3 Microprocessor Controlled 6SENSE Technology Classification according to the EC- Directive 93/ 42/ EEC: Class IIb
with sampling rate above 4000 sense/second General Safety Standards: IEC 60601-1-1, IEC 60601-2-2
• Maximum output power deviation < 15% or 5 watts, EMI/EMC Compatibility Standard: IEC 60601-1-2
whichever is greater at any setting with PER-98 IS Safety Standard: IS 7583

LEAKAGE CURRENTS INPUT POWER (other ratings available on request)


• RF leakage current • Monopolar : < 150 mA rms Operating Voltage : 180-265 Volt AC, 50-60 Hz
• Bipolar : < 60 mA rms Max. Power input : 1000 VA, Fuse : 6 A, UPS Backup : 1.5KVA
• Low Frequency leakage currents
• Normal polarity, intact ground : < 10 µA DIMENSIONS
• Normal polarity, ground open : < 50 µA L x W x H in cm - 42.5 x 35.5 x 15.5, Weight in kg < 7.5

Note : Unit is available in other HF power ratings on request.


Features of
LCD Touch Screen panel : for flexible use & upgradable with Internet Enable and Data Logging.
Shalya range of Reusable
6SENSE Technology : Maintain set power over wider range of tissue by Tissue Feedback Technology.
Instruments
PREM : Patient Return Electrode Monitoring gives an Alarm providing safety against the Return Electrode site burns.
Vessel Sealing : Vessel Sealing up to 7mm with audio indication at completion or alarm if improper sealing.
Saline Resection Mode : Saline Plasma Bipolar Cut & Coag with with Non Saline activation Alarm.
Shalya RF Scalpel Mode (Upgradable) : Bipolar Coagulation & Cutting of Tissue by Shalya Bipolar Scalpel (Desiccator).
Auto Bip Mode (Optional) : Auto start & stop of Bipolar modes without activation.
BiCoag Alarm : Audio feedback after completion of Bipolar Coag to reduces charring and sticking. 10mm Laparo insert

2SiCoag (Optional) : Two Simultaneous coagulation output with independent control Independent power (Optional).
2 Pulse/Endo Cut Mode : Tissue feedback Pulse Interval controlled Pulse Cut for Polypectomy & Papillotomy. 10mm short insert

Self Diagnosis : Ensure the ESU with accessories are safe for functioning & display Error Codes if any.
User Programmable : 500 user settable programs with case names helps in setting the system. Reusable Instruments with changeable blade
XMT/MK/VT 1 rev : 04 APR/18 Copyright by XMTPL

Upgradability : Argon Beam Coagulation, Ultrasonic Aspirator, Smoke Evacuator, Suction Irrigation module.
Foot Switches / Pedals : Explosion Proof (Fire Resistance) and Water Proof Foot Pedals.
Time Out Function (Optional) : Automatically stops the HF output if prolong unintended activation.
Surgical Applications : Useful for wide range of surgical application as : Plastic, Neuro, ENT, Spine,
Sealing Clamp
Opthal, Laparo Surgery, Endo, GI, Ortho, Open & General Surgery, Onco, Gynecology, Hernia,
GI Endocut, Papillotomy, Polypectomy, Under Water, Urology, Gynec TCRE, Hysteroscopy, Thoracotomy, Cardio Thoracic.
Vessel Sealing : Laparoscopic / Radical / Vaginal/ Abdominal Hysterectomy, Appendectomy, Colectomy, Cystectomy,
Gastric Bypass, Hemorhoidectomy, LAVH, Liver Resection, Nephrectomy, Nissen Fundoplication, Splenectomy.
Saline Plasma Bipolar Resection : Bipolar TURP/ TCRE / Hysteroscopy / Arthroscopic Surgery.

Plot No. W239, Rabale MIDC, TTC Industrial Area, Rabale, Navi Mumbai - 400 701. INDIA T:+91-22-7114 2400