Mars Select PDF
Mars Select PDF
Mars Select PDF
Service Manual
This is a class 1 medical product within the meaning of the European Medical Device Directive (MDD)
93/42/EEC, June 1993, Appendix 9, and the German Medizin Produkte Gesetz (MPG) (Law on
Medical Products), August 1994, Appendix 9.
The manufacturer declares that this product conforms to the essential requirements of the MPG, Appendix 7,
and this is documented by the CE sign.
Reprinting, copying or translating this document in whole or in part is forbidden without the express written
permission of TRUMPF Medizin Systeme GmbH.
All rights under the copyright laws are expressly reserved by TRUMPF Medizin Systeme GmbH.
Within the framework of the legal requirements, the manufacturer is only responsible for the safety
characteristics of this device if the maintenance, repair and modifications performed on this device are
carried out by the manufacturer, or by a person authorised by the manufacturer and in accordance with the
manufacturer’s instructions.
Contents
1 Important Information
1.1 Notes for Service Personnel ...................................................................................4
1.2 Explanation of Symbols .........................................................................................5
1.3 General Safety Instructions ....................................................................................6
1.4 Protection against Infection....................................................................................8
1.4.1 Safety instructions for working in the operating theatre area......................................8
1.4.2 Instructions for persons at risk ................................................................................8
1.5 Instructions for Protecting the Environment ..............................................................8
1.6 Glossary ..............................................................................................................9
2 Preparations
2.1 Repair Site .........................................................................................................10
2.2 Tools, Measuring Equipment, Materials ................................................................10
7 Additional Components
7.1 Cable remote control ..........................................................................................67
7.2 Infrared remote control .......................................................................................68
7.3 Charging unit for remote control..........................................................................70
7.4 Foot control .......................................................................................................71
8 Appendices
8.1 General Information ...........................................................................................73
8.1.1 Emergency operating system................................................................................73
8.1.2 Direct control of the drives...................................................................................74
8.1.3 Emergency adjustment of the column height..........................................................75
8.2 Search for faults .................................................................................................76
8.2.1 Column .............................................................................................................76
8.2.2 Tabletop ............................................................................................................78
8.2.3 Running gear .....................................................................................................80
8.3 Indicators/monitoring/error messages ..................................................................81
8.3.1 Optical indicators ...............................................................................................81
8.3.2 Acoustic signal ...................................................................................................82
8.3.3 Overview of fuses ...............................................................................................82
8.3.4 Error Messages...................................................................................................83
8.4 Test program (internal) ........................................................................................85
8.4.1 Test 1 (movement test) ........................................................................................86
8.4.2 Test 2 (key test)...................................................................................................87
8.4.3 Test 3 (LED test)..................................................................................................88
8.4.4 Test 4 (inclination test).........................................................................................89
8.5 Service Program (PC) ..........................................................................................90
8.5.1 File....................................................................................................................90
8.5.2 State..................................................................................................................91
8.5.3 Driving ..............................................................................................................92
8.5.4 EEPROM............................................................................................................93
8.5.5 Test ...................................................................................................................94
8.5.6 Monitor .............................................................................................................95
8.5.7 Final check ........................................................................................................96
8.6 Modem Operation..............................................................................................97
8.6.1 Programming .....................................................................................................97
8.6.2 Modem Connection............................................................................................98
8.6.3 Updating the Software.........................................................................................99
8.7 Column Spare Parts ..........................................................................................101
8.8 Tabletop Spare Parts .........................................................................................115
8.9 Running Gear Spare Parts .................................................................................137
8.10 Spare Parts of the Additional Components ..........................................................147
8.11 Maintenance ....................................................................................................151
8.12 Operating instructions in brief............................................................................157
8.13 Circuit Diagrams ..............................................................................................161
1 Important Information
Please follow this manual exactly when performing repairs and keep to all the scheduled
repair steps.
We would be grateful to receive any corrections to this manual and also any tips for
performing repairs more efficiently than described here.
Please contact our Service Department in Saalfeld.
Repairs which are not covered by this service manual may only be performed after
consultation with TRUMPF Medizin Systeme GmbH or be performed by
TRUMPF Medizin Systeme GmbH in Saalfeld.
After all repairs or settings, a functional test or test must be performed in accordance with
the instructions in this service manual.
The device may only be handed over to the customer after being tested and in a fully
functional condition.
WARNING!
Danger to life and limb (for example electric shock)!
Risk of serious injuries and damage to property.
CAUTION!
Risk of injury and/or damage to property.
ATTENTION!
Risk of damage to property.
NOTE!
Useful additional information and tips.
TEST!
Performance of functional testing, measurements and tests.
ENVIRONMENT!
Notes on disposal in a way which does not harm the environment!
· Only perform those repairs and settings for which you have been trained and which are
described in the Service Manual.
Please contact us if you have any doubts.
· Only use original spare parts provided by TRUMPF Medizin Systeme GmbH as spare
parts.
· Keep the areas around the device clean and tidy during and after maintenance.
· Do not perform any actions which could put people in danger or could make the
device a source of danger.
· Store dismounted casings and other parts of the device which have been removed in a
safe place while working. Always place tools and dismounted parts in a place where no
one can trip over or slip on them. If any nuts, bolts or other small parts fall into any of
the MARS modules they must be removed.
· Your clothing must be so arranged that it cannot get caught in the device. Button up
the sleeves of your shirt or jacket or roll them up. Long hair must be tied together or
put up.
Tuck the ends of scarves, ties and shawls into your clothing or fasten them securely.
Loose items of clothing can be dangerous!
· Wear protective goggles for the following activities:
– Hammering pins or similar parts in or out
– Removing springs under tension and safety devices
– Working with an electric drill
– Using spring hooks or inserting springs
– Soldering
– When using solvents, cleansers or other chemicals
– In general, when the working conditions present a possible risk to the eyes.
· After work has been completed, replace or renew all safety devices, such as:
– Covers
– Cable connectors, cable holders
– Cable screens
– Bonding and potential connections
– Earthing connections.
Potential equalisation/earth connections must be tested after assembly.
· When soldering, only limed resin may be used as flux.
· After completing the repair, check the electrical connections again with the circuit
diagram before performing a functional test.
· Bolts and screws with an M4 or larger thread must be fixed with Omnifit 100 (red).
· Replace all defective parts, which are not clearly in good working order, with original
spare parts, including those which go beyond the scope of the repair order.
· After repairs have been completed, all cables must be restored to their original
positions, and a final check must be made for any risk of shearing and crushing.
Check the cable positions again during and after the course of movements of the table
and its sections.
· Unauthorised or provisional repairs are not permitted, even if the customer requests
them.
DANGER!
There is a risk of infection in the entire hospital.
Observe all safety measures.
As a service technician, also follow the requirements of the medical
facility concerning protection against infection.
· Only perform maintenance and repair work on operating tables which have been
disinfected.
· The disinfection of the interior parts of the operating table must be organised at the
proper time before the repair/maintenance work and is to be performed by the medical
facility.
· Observe all the medical facility’s hygiene requirements concerning clothing, cleaning
regulations and disinfection measures.
· Observe all rules on behaviour, especially in the operating theatre area.
· Work may only be performed in the operating theatre area if the relevant authorisation
has been given.
· See a doctor immediately if symptoms of illness or disease occur.
Tell your doctor that you have worked in areas where there is risk of infection - even
months later.
NOTE!
Ask a doctor about the possible risks and possible measures to avoid
them in your work as a service technician.
1.6 Glossary
The following terms and abbreviations are used in this service manual:
2 Preparations
Special tools:
· Allen keys for emergency adjustment of the column stroke
· Pin remover
· Open-ended spanner 10 (ground)
Materials
· Scotch Weld 110DP adhesive
· Instant adhesive
· Line connector
· Screw fixing lacquer OMNIFITT 100M
· Screw fixing lacquer OMNIFITT 200M
· Fixing lacquer for sensors
· Double sided adhesive tape
· Wooden spacer (replacing lifting motor)
· Wooden spacer (work beneath the base plate)
· Spirit (degreaser)
· Spindle grease (order from TRUMPF Medizin Systeme GmbH)
DANGER!
Danger of infection!
Before beginning work, obtain confirmation from the medical facility,
that all the necessary measures for protection against infection have
been performed.
3.1 Batteries
NOTE!
Always replace the complete set of batteries.
Removal:
1. Bring the leg and back sections into their uppermost positions or uncouple them.
2. Lower the column to its lowest position.
3. Remove the upper cladding component, without the membrane keypad,
(7 M3 screws, of which 3 are under the rubber bellows, two on each cladding
component).
4. Raise the column carefully and unhook the remainder of the cladding.
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5.Pull the power supply cable out of the column and switch the column off.
6.Pull the connecting cables off both batteries.
7.Unscrew the holding bar from the lower battery (2 M4 screws).
8.Remove the battery.
9.Unscrew the holding bar from the upper battery (2 M5 screws), hold the battery
securely while doing so.
10. Exchange the set of batteries.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Ensure that the polarity is correct (the marked or white wire is plus).
· Test charging current.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
· Take care not to damage the rubber bellows.
Test!
Perform a functional test!
ENVIRONMENT!
Dispose of the batteries in a way which does not harm the
environment!
Test!
Perform a functional test!
U s e o n ly w ith a 2 5 0 V fu s e
E m p lo y e r u n iq u e m e n t a v e c
u n fu s ib le d e 2 5 0 V
1
Test!
Perform a functional test!
3.4 Fuses
Removal:
1. Pull the power supply cable out of the column and switch the column off.
1 2
U s e o n ly w ith a 2 5 0 V fu s e
E m p lo y e r u n iq u e m e n t a v e c
u n fu s ib le d e 2 5 0 V
2. Raise the mains socket cover and prise the fuse socket [1] out with a screwdriver.
3. Replace the defective fuse [2].
Reassembly:
In the reverse order.
Test!
Perform a functional test!
Removal:
1. Bring the leg and back sections into their uppermost positions or uncouple them.
2. Lower the column to its lowest position.
3. Remove the upper cladding component, without the membrane keypad,
(7 M3 screws, of which 3 are under the rubber bellows, two on each cladding
component).
4. Raise the column carefully and unhook the remainder of the cladding.
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
3
F u s e
6. Open the fuse socket [1] (alongside the uppermost battery) by screwing out the cap,
and insert a new fuse.
Reassembly:
In the reverse order.
Test!
Perform a functional test!
3.5 Electronics
Removal:
1. Bring the leg and back sections into their uppermost positions or uncouple them.
2. Lower the column to its lowest position.
3. Remove the upper cladding component, without the membrane keypad,
(7 M3 screws, of which 3 are under the rubber bellows, two on each cladding
component).
4. Raise the column carefully and unhook the remainder of the cladding.
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
6. Unscrew the nuts (12 nuts) from the fastening frame of the rubber bellows.
1
2
3
4
5
6
7
8
9
1 0
Figure 5: Electronics
1 Multipoint connector for membrane 6 X5 Section motors
2 X1 Tabletop distribution board 7 X6 Tilting
3 X2 End positions of the main lift 8 X7 Trendelenburg
4 X3 Reserve 9 X30 Battery
5 X4 Main lift 10 X80 Mains
7. Slide the rubber bellows upwards and pull the operating membrane plug off the
multipoint connector [1] of the electronics.
8. Unscrew the cladding component together with the membrane keypad
(3 M3 screws).
9. Unscrew or pull off all the plugs [2], [3], [4], [5], [6], [7], [8], [9], [10] from the
rear side of the electronics.
10. Unscrew the grounding connector from the electronics casing (M4 screw).
11. Unscrew the complete electronics from the column (3 M6 screws).
12. Unscrew 3 holding brackets from the electronics (2 M3 nuts in each case).
Reassembly:
Reassemble in the reverse order, observing the following instructions:
ATTENTION!
Danger of a short circuit!
Insert the electronics carefully, ensuring that there is no short circuit to
the battery contacts.
· Make sure that the plug is positioned correctly when plugging it into the electronics.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
· Take care not to damage the rubber bellows.
· Spool the IR code in with the PC software.
Test!
Perform a functional test!
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
6. Unscrew the nuts (12 nuts) from the fastening frame of the rubber bellows.
7. Slide the rubber bellows upwards and pull the operating membrane plug off the
multipoint connector of the electronics.
8. Unscrew the cladding component together with the membrane keypad
(3 M3 screws).
9. Pull out the membrane keypad.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Thoroughly clean the cladding component and completely remove adhesive residues.
· Make sure that the plug is positioned correctly when plugging it into the electronics.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
· Take care not to damage the rubber bellows.
Test!
Perform a functional test!
Test!
Perform a functional test!
Check the zero position of the tilt and Trendelenburg!
3.8 Drives/Motors
Removal:
1. Move the tabletop to the zero position.
2. Raise the column to its highest position.
Use emergency adjustment if necessary (see the emergency adjustment section).
3. Pull the power supply cable out of the column and switch the column off.
4. Open the cladding at the bottom (2 M3 screws), slide it upwards and secure it.
3
5
DANGER!
Danger of material damage and injuries if the column sinks
down!
Always secure the column in its highest position with a suitable
piece of wood or a standard rail when working on the vertical
drive of the column.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· The rubber intermediate layers between the lifting motor and the holder must not be
removed.
· Restore all the connections.
· Ensure that the polarity is correct.
· Relay the wiring and renew the cable binders.
Adjustment:
· The toothed belt tension is set with the setting screw [4].
· The gap must be set so that the maximum load can be safely coped with.
Test!
Perform a functional test!
Removal:
1. Move the tabletop to the zero position.
2. Raise the column to its highest position.
Use emergency adjustment if necessary (see the emergency adjustment section).
3. Pull the power supply cable out of the column and switch the column off.
4. Open the cladding at the bottom (2 M3 screws), slide it upwards and secure it.
7
1
6 2
3
5
4 8
Figure 7: Vertical drive
1 Connecting cable 5 Toothed belt
2 Vertical drive 6 Spindle holder
3 M6 screw 7 M6 screw
4 M4 screw 8 Cable grip
DANGER!
Danger of material damage and injuries if the column sinks
down!
Always secure the column in its highest position with a suitable
piece of wood or a standard rail when working on the vertical
drive of the column.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Avoid creating stresses and distortions when screwing the spindle holder [6] and the
vertical drive [2] in place.
· Restore all the connections.
· Ensure that the polarity is correct.
· Relay the wiring and renew the cable binders.
Adjustment:
· The toothed belt tension is set with the setting screw [4].
· The vertical drive must be set so that the maximum load can be safely coped with.
Test!
Perform a functional test!
Removal:
1. Bring the leg and back sections into their uppermost positions or uncouple them.
2. Lower the column to its lowest position.
3. Remove the upper cladding component, without the membrane keypad,
(7 M3 screws, of which 3 are under the rubber bellows, two on each cladding
component).
4. Raise the column carefully and unhook the remainder of the cladding.
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
6. Remove the pad from the seat section upwards.
7. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
8. Remove the covering cap fastening screws (8 M4 screws).
9. Pull the covering cap off.
10. Loosen the upper fastening frame of the rubber bellows (8 M4 screws at the head
and foot ends and 4 nuts left and right).
11. Press the rubber bellows down.
1
5
2 6
3 7
4
8
NOTE!
Take the load off the tabletop and carefully tilt it backwards.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Make sure that the plug is positioned correctly when plugging it into the electronics.
· Renew cable binders.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
· Take care not to damage the rubber bellows.
Test!
Perform a functional test!
Removal:
1. Bring the leg and back sections into their uppermost positions or uncouple them.
2. Lower the column to its lowest position.
3. Remove the upper cladding component, without the membrane keypad,
(7 M3 screws, of which 3 are under the rubber bellows, two on each cladding
component).
4. Raise the column carefully and unhook the remainder of the cladding.
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
6. Pull the X6 plug (tilting drive) out of the electronics.
7. Cut the cable binders.
8. Remove the pad from the seat section upwards.
9. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
10. Remove the covering cap fastening screws (8 M4 screws).
11. Pull the covering cap off.
12. Loosen the upper fastening frame of the rubber bellows (8 M4 screws at the head
and foot ends and 4 nuts left and right).
13. Press the rubber bellows down.
1 2 3 4 5
NOTE!
Take the load off the tabletop and carefully tilt it backwards.
14. Remove the retaining ring [1] from the upper connecting pin [2].
15. Knock out the upper connecting bolt [2] towards the rear.
NOTE!
Take the load off the tilting drive [5].
16. Remove the retaining ring [4] from the lower connecting bolt [3].
17. Knock out the lower connecting bolt [2] towards the rear.
18. Carefully remove the tilting drive [5].
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Ensure that the plug is correctly positioned when connecting it to the electronics.
· Renew cable binders.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
· Take care not to damage the rubber bellows.
Test!
Perform a functional test!
Removal:
1. Bring the tabletop into a horizontal position.
2. Pull the power supply cable out of the column and switch the column off.
3. Uncouple all functional accessories on the foot end.
4. Pull the seat section pad and, if present, the leg section pad off in an upward
direction.
5. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
6. Unscrew the plastic plate from leg section (2 screws and 4 ball-headed screws M8).
7. Remove the covering cap fastening screws (8 M4 screws).
8. Pull the covering cap off.
9. Unclip the motor cable from the distribution board.
10. Remove the covers of the assembly window.
11. Cut the cable binders.
1
2
3
12. Remove 2 M8 machine screws [1] within the spar and 2 M6 countersunk screws [3]
underneath it.
13. Carefully remove the motor [2], together with the gear, from the strut.
14. Unscrew the motor [2] from the gearbox (4 M5 screws [4]).
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Make sure the wiring assignment is correct (see wiring diagram).
· Relay the wiring and renew the cable binders.
Test!
Perform a functional test!
Check the end and reference sensors of the motor,
(gap 0.2 mm ± 0.1 mm).
Removal:
1. Bring the tabletop into a horizontal position.
2. Pull the power supply cable out of the column and switch the column off.
3. Uncouple all functional accessories on the head end.
4. Remove the pad from the seat and back section upwards.
5. Unscrew the plastic plate from leg section (4 screws and 8 ball-headed screws M8).
6. Remove the covering cap fastening screws (8 M4 screws).
7. Pull the covering cap off.
8. Unclip the motor cable from the distribution board.
9. Remove the covers of the assembly window.
10. Cut the cable binders.
1
2
3
4
11. Loosen the motor [3], to do this, remove 2 M8 machine screws [1] from within the
seat section strut [2] and 2 M6 countersunk screws [5] from underneath it.
12. Carefully remove the motor [3], together with the gear, from the seat section
strut [2].
13. Unscrew the motor [3] from the gear, remove 4 M5 screws [4 ] to do this.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Make sure the wiring assignment is correct (see wiring diagram).
· Lay all the cables safely (renew cable binders).
Test!
Perform a functional test!
Check the end and reference sensors of the motor
(gap 0.2 mm ± 0.1 mm).
Removal:
1. Bring the tabletop into a horizontal position.
2. Pull the power supply cable out of the column and switch the column off.
3. Uncouple all functional accessories.
4. Remove the pad from the tabletop upwards.
5. Unscrew the plastic plates (4 M8 ball-headed screws and 8 M8 screws).
6. Remove the cover plate beneath the cover cap (4 M4 screws).
7. Remove the covering cap fastening screws (8 M4 screws).
8. Pull the covering cap off.
9. Pull the cables out.
10. Unscrew the motor together with the holding bracket (2 M5 screws).
11. Fit a new motor.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Lay all the cables safely (renew cable binders).
Adjustment:
· Loosen the motor on the holding bracket (4 M5 screws) and adjust the toothed belt
tension with the setscrew (M5).
· Set the toothed belt tension to a maximum sag of 5 mm.
Test!
Perform a functional test!
Removal:
1. Bring the tabletop into a horizontal position.
2. Displace the tabletop to the left.
3. Pull the power supply cable out of the column and switch the column off.
4. Uncouple all functional accessories.
5. Remove the pad from the tabletop upwards.
6. Unscrew the plastic plates (4 M8 ball-headed screws and 8 M8 screws).
7. Remove the cover plate beneath the cover cap (4 M4 screws).
8. Remove the covering cap fastening screws (8 M4 screws).
9. Pull the covering cap off.
10. Cut the cable binders.
11. Unscrew the cables from the distribution board.
12. Unscrew the motor together with the holding bracket (2 M5 screws).
13. Fit a new motor.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Lay all the cables safely (renew cable binders).
Adjustment:
· Loosen the motor, together with the holding bracket, (2 M5 screws) and adjust the
toothed belt tension with the setscrew (M5).
· Set the toothed belt tension to a maximum sag of 5 mm.
Test!
Perform a functional test!
3.9 Sensors
NOTE!
The end sensors for the tilting and Trendelenburg functions are
integrated into the corresponding drives. They cannot be replaced
separately.
Removal:
1. Bring the leg and back sections into their uppermost positions or uncouple them.
2. Lower the column to its lowest position.
3. Remove the upper cladding component, without the membrane keypad,
(7 M3 screws, of which 3 are under the rubber bellows, two on each
cladding component).
4. Raise the column carefully and unhook the remainder of the cladding.
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
6. Pull plug X2 (main lift end positions) out of the electronics.
7. Cut the cable binders.
1
2
3
4
5
8. Loosen the M4 screw holding the switching piece and remove the switching piece [1].
9. Unscrew 2 M6 fixing screws on the end switch module [4] and pull the end switch
module [4] off in an upward direction.
10. Renew the end switches individually if necessary. The wires [2] must be desoldered
from the micro keys [3]/[5] in order to do so.
Adjustment:
· The micro keys [3] and [5] are to be fixed in the end switch module [4] in such a way
that the upper one [3] is placed in as far as possible and the lower one [5] is out as far
as possible.
· Align the micro keys [3]/[5] parallel to the holder.
· The end switch module [4] is to be fixed in such a way that the lower edge is flush with
the holder of the tilting drive.
· The switching piece [1] must be adjusted so that the vertical drive switches off safely
before there is mechanical contact with the spindle.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Ensure that the plugs are correctly positioned when inserting them into the electronics.
· Renew cable binders.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
· Take care not to damage the rubber bellows.
Test!
Perform a functional test!
Removal:
1. Bring the tabletop into a horizontal position.
2. Pull the power supply cable out of the column and switch the column off.
3. Uncouple all functional accessories on the foot end.
4. Pull the seat section pad and, if present, the leg section pad off in an upward
direction.
5. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
6. Unscrew the plastic plate from leg section (2 screws and 4 ball-headed screws M8).
7. Remove the covering cap fastening screws (8 M4 screws).
8. Pull the covering cap off.
9. Remove the covers of the assembly window.
10. Unclip the sensor cable from the distribution board.
11. Cut the cable binders.
1
2
3
12. Loosen the motor, to do this, remove 2 M8 machine screws [1] from within the strut
and 2 M6 countersunk screws [4] from underneath it.
13. Pull the motor [2] together with the gearbox out about 3-4 mm.
14. Remove the M4 threaded pin [2] and the sensor [3].
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Make sure the wiring assignment is correct (see wiring diagram).
· Lay all the cables safely (renew cable binders).
Adjustment:
· Move the motor to the switching point.
· Set the sensor [3] by sliding it so that the gap between it and the metal is
0.2 mm ± 0.1 mm.
Test!
Perform a functional test!
Check the sensor gap (gap 0.2 mm ± 0.1 mm).
Removal:
1. Bring the tabletop into a horizontal position.
2. Pull the power supply cable out of the column and switch the column off.
3. Uncouple all functional accessories on the head end.
4. Remove the pad from the seat and back section upwards.
5. Unscrew the plastic plate from leg section (4 screws and 8 ball-headed screws M8).
6. Remove the covering cap fastening screws (8 M4 screws).
7. Pull the covering cap off.
8. Remove the covers of the assembly window.
9. Unclip the sensor cable from the distribution board.
10. Cut the cable binders.
1
2
3
4
11. Loosen the motor, to do this, remove 2 M8 machine screws [1] from within the strut
and 2 M6 countersunk screws [4] from underneath it.
12. Pull the motor [2] together with the gearbox out about 3-4 mm.
13. Remove the M4 threaded pin [2] and the sensor [3].
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Make sure the wiring assignment is correct (see wiring diagram).
· Lay all the cables safely (renew cable binders).
Adjustment:
· Move the motor to the switching point.
· Set the sensor [3] by sliding it so that the gap between it and the metal is
0.2 mm ± 0.1 mm.
Test!
Perform a functional test!
Check the sensor gap (gap 0.2 mm ± 0.1 mm).
Removal:
1. Bring the tabletop into a horizontal position.
2. Displace the tabletop to the left.
3. Pull the power supply cable out of the column and switch the column off.
4. Uncouple all functional accessories.
5. Remove the pad from the tabletop upwards.
6. Unscrew the plastic plates (4 M8 ball-headed screws and 8 M8 screws).
7. Remove the cover plate beneath the cover cap (4 M4 screws).
8. Remove the covering cap fastening screws (8 M4 screws).
9. Pull the covering cap off.
10. Remove the left-hand cover plate from the table top (4 M4 screws).
11. Cut the cable binders.
12. Unscrew all sensor cables and the longitudinal displacement motor wiring
from the distribution board.
13. Unscrew the power train from the lower fixing point (2 M3 screws)
and pull it to the rear.
14. Pull the cable of the defective sensor out of the power train.
15. Unscrew the sensor together with the clamping block (2 M3 screws).
16. Loosen the locking screw (M3) and remove the sensor from the clamping block.
17. Replace the sensor.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Make sure the wiring assignment is correct (see wiring diagram).
· Lay all the cables safely (renew cable binders).
Adjustment:
· Move the motor to the switching point.
· Set the sensor by sliding it so that the gap between it and the metal is
0.2 mm ± 0.1 mm.
Test!
Perform a functional test!
Check the sensor gap (gap 0.2 mm ± 0.1 mm).
Removal:
1. Bring the tabletop into a horizontal position.
2. Displace the tabletop to the left.
3. Pull the power supply cable out of the column and switch the column off.
4. Uncouple all functional accessories.
5. Remove the pad from the tabletop upwards.
6. Unscrew the plastic plates (4 M8 ball-headed screws and 8 M8 screws).
7. Remove the cover plate beneath the cover cap (4 M4 screws).
8. Remove the covering cap fastening screws (8 M4 screws).
9. Pull the covering cap off.
10. Remove the left-hand cover plate from the table top (4 M4 screws).
11. Cut the cable binders.
12. Unscrew all sensor cables and the longitudinal displacement motor wiring
from the distribution board.
13. Unscrew the power train from the lower fixing point (2 M3 screws)
and pull it to the rear.
14. Pull the cable of the defective sensor out of the power train.
15. Loosen the locking screw (M4) and remove the sensor from the clamping block.
16. Replace the sensor.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Make sure the wiring assignment is correct (see wiring diagram).
· Lay all the cables safely (renew cable binders).
Adjustment:
· Cover the switching point with adhesive tape (0.2 mm thick).
· Move the motor to the switching point.
· Slide the sensor downwards and fix it with a locking screw (M4).
· Remove the adhesive tape.
Test!
Perform a functional test!
Check the sensor gap (gap 0.2 mm ± 0.1 mm).
NOTE!
The inclination sensor is a part of the distribution board.
They cannot be replaced separately.
Adjustment:
The inclination sensor is adjusted in two stages. The amplification is set first, then it is
aligned with the zero position.
The amplification is set with 25° tilt or Trendelenburg (electronic spirit level). The function
not currently needed must be set to zero.
The voltage on the distribution board between X or Y and the earth potential can be
determined by a measuring device. The voltage values listed below are set with the
appropriate amplifier control.
After setting the amplification, the zero position is set to an accuracy of 0.3° by setting the
controller (electronic spirit level).
(See the wiring diagram for the measuring points and the controller arrangement.)
NOTE!
As the distribution board is fitted into the MARS endouro with a 90°
turn, the wiring is modified accordingly. These modifications can be
seen in the (MARS endouro) wiring diagram.
Compensation values:
3.9.8 OR sensor
Removal:
1. Bring the tabletop into a horizontal position.
2. Pull the power supply cable out of the column and switch the column off.
3. Remove the pad from the seat section upwards.
4. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
5. Remove the covering cap fastening screws (8 M4 screws).
6. Pull the covering cap off.
7. Remove the covers of the assembly window.
8. Unclip the sensor cable from the distribution board.
9. Cut the cable binders.
Test!
Perform a functional test!
3.9.9 IR sensor
Removal:
1. Bring the tabletop into a horizontal position.
2. Remove the pad from the seat section upwards.
3. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
4. Remove the covering cap fastening screws (8 M4 screws).
5. Pull the covering cap off.
6. Loosen the upper fastening frame of the rubber bellows (8 M4 screws at the head
and foot ends and 4 nuts left and right).
7. Press the rubber bellows down.
8. Bring the Trendelenburg drive into the lower end position.
9. Pull the power supply cable out of the column and switch the column off.
10. Unclip the cable from the distribution board.
11. Remove the cable grip from the underside of the tabletop.
12. Unscrew the IR sensor module (2 M8 screws).
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Make sure the wiring assignment is correct (see wiring diagram).
· Lay all the cables safely (renew cable binders).
Test!
Perform a functional test!
DANGER!
Danger of infection!
Before beginning work, obtain confirmation from the medical facility,
that all the necessary measures for protection against infection have
been performed.
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
6. Unscrew the cladding segment by segment (4 M3 screws per segment).
7. Renew defective elements.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Only complete cladding elements are replaced.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
· Take care not to damage the rubber bellows.
Test!
Perform a functional test!
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
6. Unscrew the nuts (12 nuts) from the fastening frame of the rubber bellows.
7. Slide the rubber bellows upwards and pull the operating membrane plug off
the electronics.
8. Unscrew the cladding component together with the membrane keypad
(3 M3 screws).
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Only complete cladding elements are replaced.
· Ensure that the plug is correctly positioned when connecting it to the electronics.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
· Take care not to damage the rubber bellows.
Test!
Perform a functional test!
7
1
6 2
3
5
9
4 8
DANGER!
Danger of material damage and injuries if the column sinks
down!
Always secure the column in its highest position with a suitable
piece of wood or a standard rail when working on the vertical
drive of the column.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· The rubber layers between the lifting motor and the holder must not be removed.
· Avoid creating stresses and distortions when screwing the spindle holder [6] and the
vertical drive [2] in place.
· Restore all the connections.
· Ensure that the polarity is correct.
· Relay the wiring and renew the cable binders.
Adjustment:
· The toothed belt tension is set with the setting screw [4].
· The vertical drive must be set so that the maximum load can be safely coped with.
Test!
Perform a functional test!
ATTENTION!
Danger of damaging the cladding!
Do not lower the column when the cladding has been removed.
5. Pull the power supply cable out of the column and switch the column off.
6. Unscrew the nuts (12 nuts) from the fastening frame of the rubber bellows.
7. Pull plugs X1 and X5 out of the electronics.
8. Remove the pad from the seat section upwards.
9. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
10. Remove the covering cap fastening screws (8 M4 screws).
11. Pull the covering cap off.
12. Loosen the upper fastening frame of the rubber bellows (8 M4 screws at the head
and foot ends and 4 nuts left and right).
13. Press the rubber bellows down.
14. Unscrew the cable grips (2 items) for the cables to the tabletop from the upper
telescopic part.
15. Unscrew the potential equalisation line from the underside of the tabletop and pull
it out of the cable grip.
DANGER!
Risk of personal injuries and material damage!
A second person is required to secure and lift the tabletop off.
16. Release the tabletop from the column (8 M8 screws, 2 on each of the tilting cardans
and 4 on the Trendelenburg cardan).
17. Lift the tabletop off the column.
18. Pull the expansion bellows, together with the upper and lower fastening frames,
off the column.
19. Pull out the lower parts of the fastening frame and unscrew the upper parts of the
frame (14 M4 screws).
20. Renew the rubber bellows.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Take care not to damage the rubber bellows when inserting it into the lower frame
parts.
· Screw the cardans directly onto the tabletop.
· Ensure that the plugs are correctly positioned when inserting them into the electronics.
· Make sure that the cladding is correctly positioned when hanging it in the topmost
guide.
Test!
Perform a functional test!
Test!
Perform a functional test!
DANGER!
Danger of infection!
Before beginning work, obtain confirmation from the medical facility,
that all the necessary measures for protection against infection have
been performed.
Test!
Perform a functional test!
5.2 Cover
Removal:
1. Bring the tabletop into a horizontal position.
2. Pull the power supply cable out of the column and switch the column off.
3. Remove the pad from the seat section upwards.
4. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
5. Remove the covering cap fastening screws (8 M4 screws).
6. Pull the covering cap off.
Reassembly:
In the reverse order.
Test!
Perform a functional test!
7 8 9 1 0 1 1
1
2
3
4
5
6
A
12. Unscrew the inner and outer joint covers [5] (one M6 screw [9] in each case).
13. Remove the cylindrical pins [8] (3 items in each case) with an extractor.
14. Unscrew the joint [10] (3 M4 screws [7] in each case).
15. Loosen the motor [2], to do this remove 2 M8 machine screws [1] from within the
strut and 2 M6 countersunk screws [6] from underneath it.
16. Carefully remove the motor [2], together with the gear [4], from the strut.
17. Unscrew the gear [4] from the motor [2], remove 4 M5 screws [3 ] to do this.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Do not forget the sliding disks when mounting the joint [10].
· Do not damage the seal in the gear [4].
· Make sure the wiring assignment is correct (see wiring diagram).
· Lay all the cables safely (renew cable binders).
Test!
Perform a functional test!
Check the sensor gap [A] (gap 0.2 mm ± 0.1 mm).
6 7
1
2
8
3
4
5 A
Test!
Perform a functional test!
Check the sensor gap [A] (gap 0.2 mm ± 0.1 mm).
1 1
1
2 7
3 8
4 9
5 1 0
A
6
12. Unscrew the inner and outer joint covers [5] (one M6 screw [9] in each case).
13. Remove the cylindrical pins [7] (3 items in each case) with an extractor.
14. Unscrew the joint [10] (3 M4 screws [8] in each case).
15. Loosen the motor [2], to do this remove 2 M8 machine screws [1] from within the
strut and 2 M6 countersunk screws [6] from underneath it.
16. Carefully remove the motor [2], together with the gear [4], from the strut.
17. Unscrew the gear [4] from the motor [2], remove 4 M5 screws [3 ] to do this.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Do not forget the sliding disks when mounting the joint [10].
· Do not damage the seal in the gear [4].
· Make sure the wiring assignment is correct (see wiring diagram).
· Lay all the cables safely (renew cable binders).
Test!
Perform a functional test!
Check the sensor gap [A] (gap 0.2 mm ± 0.1 mm).
1
2 5
3 6
4 7
8
NOTE!
The spring [6] of the locking lever [5] is under pressure.
Keep the locking lever [5] pressed while removing the cam lever [7] in
order to prevent the spring [6] from flying off.
Test!
Perform a functional test!
Check the interlock of locking lever [5].
Test!
Perform a functional test!
Check the tension of the toothed belt.
Test!
Perform a functional test!
Check the tension of the toothed belt.
Test!
Perform a functional test!
Check the tension of the toothed belt.
Test!
Perform a functional test!
Check the tension of the toothed belt.
Test!
Perform a functional test!
Check the sensor gap (gap 0.2 mm ± 0.1 mm).
Test!
Perform a functional test!
DANGER!
Risk of personal injuries and material damage!
Tilt the operating table a maximum of 5° and ensure that it does not
tip overt.
2. Lay the table on its side and support it with suitable pieces of wood.
3. Loosen screw [1] and press the holding bracket [2] down.
4. Pull the wheel shells [3] off in a downward direction.
NOTE!
Leave the plastic discs on the axles in order to keep the wheel housing
secure.
Reassembly:
In the reverse order.
Adjustment:
· The wheel shells [3] on the axles are adjusted with plastic spacers.
· Set the wheel shells [3] so that the wheel runs securely without excessive play.
Test!
Perform a functional test!
DANGER!
Risk of personal injuries and material damage!
Tilt the operating table a maximum of 5° and ensure that it does not
tip over.
2. Lay the table on its side and support it with suitable pieces of wood.
3. Unscrew both wheel shells from the axle and remove them.
4. Knock the axle out to the rear.
5. Fit a new fixed roller.
Reassembly:
In the reverse order.
Adjustment:
· The fixed roller between the axle holders is adjusted with plastic spacers.
· Set the fixed roller so that it runs securely without excessive play.
Test!
Perform a functional test!
Figure 22: Foot end wheel with brake and directional travel
1 M8 screw 3 Directional travel Bowden cable
2 Brake Bowden cable
Test!
Perform a functional test!
Test!
Perform a functional test!
4
2
5
6. Pull the collision caps off the wheels at the head end [2].
7. Unscrew both wheels (4 M8 screws [1] in each case).
8. Unscrew the axle holders [6] (2 M8 screws [7] in each case).
9. Remove the retaining ring [5].
10. Remove the brake Bowden cable [4] together with the axle [3].
11. Fit a new wheel [2].
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Ensure that the Bowden cables are laid safely.
· Brake the wheels before returning the table to an upright position.
Adjustment:
· The Bowden cables can be adjusted with the nuts at the front end if necessary.
· Set the brake so that it is functioning safely and smoothly.
Test!
Perform a functional test!
1 2
3
6. Loosen the lock nut [1] and the setting nut [2] at the front end.
7. Remove the brake Bowden cable [3] from the holder.
8. Screw out and remove the brake Bowden cable [2] from the rear end with the aid
of pliers.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Ensure that the Bowden cable is laid safely.
· Restore the screw depth of the brake Bowden cable.
· Brake the wheels before returning the table to an upright position.
Adjustment:
· The Bowden cable [3] can be adjusted with the nuts [1]/[2] at the front end if
necessary.
· Set the brake so that it is functioning safely and smoothly.
Test!
Perform a functional test!
1
2
3
1
4
5
6. Loosen the lock nut [5] and the setting nut [4] at the front end.
7. Remove the directional travel Bowden cable [1] from the holder.
8. Unhook the rear end from the directional wheel.
9. Remove the retaining ring [2] from the deflection pulley.
10. Remove the bracket [3].
11. Remove the directional travel Bowden cable [1].
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Ensure that the Bowden cable [1] is laid safely.
· Brake the wheels before returning the table to an upright position.
Adjustment:
· The Bowden cable [1] can be adjusted with the nuts [4]/[5] at the front end if
necessary.
· Set the directional travel so that it is functioning safely and smoothly.
Test!
Perform a functional test!
2
3
4
6. Pull the collision caps off the wheels at the head end.
7. Unscrew both wheels (4 M8 screws [5] in each case).
8. Unscrew the axle holders [1] (2 M8 screws [2] in each case).
9. Unhook the directional travel Bowden cable [3] from the pedal [4].
10. Loosen the fixing screw on the pedal [4].
11. Pull the pedal [4] out of the guide and fit a new one.
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Ensure that the Bowden cables are laid safely.
· Brake the wheels before returning the table to an upright position.
Adjustment:
· The Bowden cables can be adjusted with the nuts at the front end if necessary.
· Set the brake and directional travel so that they function safely and smoothly.
Test!
Perform a functional test!
3
4
5
1
6
7
8
9
6. Pull the collision caps off the wheels at the head end.
7. Unscrew both wheels (4 M8 screws [1] in each case).
8. Remove the retaining ring [3] from the deflection pulley [2].
9. Pull the bracket [5] and directional travel Bowden cable [6] together with
the castor [2] off the axle.
10. Unhook the directional travel Bowden cable [6] from the pedal.
11. Unscrew the axle holders [7] (2 M8 screws [9] in each case).
12. Remove the axle complete with the foot lever and wheels from the locking disk [8].
13. Unscrew the locking disk [8] (2 M6 screws [4]).
Reassembly:
Reassemble in the reverse order, observing the following instructions:
· Ensure that the Bowden cables are laid safely.
· Brake the wheels before returning the table to an upright position.
Adjustment:
· The Bowden cables can be adjusted with the nuts at the front end if necessary.
· Set the brake and directional travel so that they function safely and smoothly.
Test!
Perform a functional test!
7 Additional Components
NOTE!
The cable remote control is an exchangeable module. Therefore it is
not possible to replace individual parts within the remote control.
The complete cable remote control must always be replaced if there is
a defect.
Settings:
1. Open the remote control (5 M3 screws).
2. Carefully remove the rear part of the casing and lay it down.
3. The following settings are possible on the board:
- the code is set with the rotary selection switch,
- the device is selected with the rotary selection switch,
- the transmit power is selected with the rotary selection switch,
- jumper settings.
4. The code setting for the cable remote control is always 5 (S1).
S2 and S3 are not used.
5. Please take the device selection (S4) from the table below.
6. The operating mode setting (S5) is to be set to 0 for cable remote control.
7. The jumper settings may only be changed by the customer service.
8. Do not damage the rubber seal! Close the remote.
S4 Device
0 JUPITER EU
1 JUPITER USA
2 MARS Endourology EU
3 MARS Endourology USA
4 MARS EU
5 MARS USA
6
7
8
9
EU:
A
· any key switches on
B
· after 2 sec. off
C
D USA:
E · only ON-key switches on
F · after 25 sec. off
Test!
Perform a functional test!
NOTE!
The IR remote control is an exchangeable module. Therefore it is not
possible to replace individual parts within the remote control (except
for the batteries). The complete IR remote control must always be
replaced if there is a defect.
Settings:
1. Open the remote control (5 M3 screws).
2. Carefully remove the rear part of the casing and lay it down.
3. Fit a new battery when required:
- pull connector off,
- remove the battery carefully from the holder.
4. The following settings are possible on the board:
- the code is set with the rotary selection switch,
- the device is selected with the rotary selection switch,
- the transmit power is selected with the rotary selection switch,
- jumper settings.
5. The code setting for the IR remote control is always to be taken from the type plate
of the relevant operating table (e.g. A01).
6. Please take the device setting (S4) and operating mode setting (S5) from the
following table, the frequency and mark to space ratio (transmit power) may be
changed. 8 (36 kHz T1/4) is used as the standard setting.
Test!
Perform a functional test!
NOTE!
Therefore it is not possible to replace individual parts within the
charging unit (except for the fuse).The complete charging unit must
always be replaced if there is a defect.
Test!
Perform a functional test!
NOTE!
The foot control is an exchangeable module. Therefore it is not
possible to replace individual parts within the foot control.
Therefore it is not possible to replace individual parts within the foot
control.
Settings:
1. Open the foot control (5 M3 screws).
2. The following settings are possible on the board:
- the code is set with the rotary selection switch,
- the device is selected with the rotary selection switch,
- the transmit power is selected with the rotary selection switch,
- jumper settings.
3. The code setting for the foot control is always C (S1).
S2 and S3 are not used.
4. Please take the device selection (S4) from the table below.
5. There is no transmit power setting for the foot control.
6. The jumper settings may only be changed by the works customer service.
7. Do not damage the rubber seal! Close the foot control.
S4 Device
0 JUPITER EU
1 JUPITER USA
2 MARS Endourology EU
3 MARS Endourology USA
4 MARS EU
5 MARS USA
6
7
8
9
A
B
C
D
E
F
Test!
Perform a functional test!
8 Appendices
The operating table is equipped with an emergency operating system in case a fault
occurs (e.g. electronics error).
The emergency operating system switches on automatically if the electronic control fails
and this is indicated by:
· The LED flashing above the green key on the column control panel.
· The drives moving more slowly.
· The drive speeds are substantially reduced (around 35% less than in normal operation).
· The operating table can only be controlled via the column control panel (IR and cable
remote controls are inactive).
If a fault occurs, it is possible to control the column and tabletop drives directly, without
electronics.
ATTENTION!
There is a risk of damaging or destroying the drives.
The end position sensors are not active when the drives are traversing.
4. The drive can be traversed by applying a voltage (12 V) to the direct control cable.
5. The direction of rotation of the motor depends upon the polarity of the voltage.
Direct control of the tabletop drives
1. Pull the power supply cable out of the column and switch the column off.
2. Remove the pad from the seat section upwards.
3. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
4. Remove the covering cap fastening screws (8 M4 screws).
5. Pull the covering cap off.
6. Cut the cable binders.
7. Unclip the motor cables.
ATTENTION!
There is a risk of damaging or destroying the drives.
The end position sensors are not active when the drives are traversing.
8. The drive can be traversed by applying a voltage (12 V) to the motor cable.
9. The direction of rotation of the motor depends upon the polarity of the voltage.
If a fault occurs, it is possible to adjust the height of the column mechanically. This
adjustment is made by turning the spindle directly.
The use of the emergency adjustment device (item no. 4 700 318 is recommended for
this.
Emergency adjustment
1. Pull the power supply cable out of the column and switch the column off.
2. Remove the pad from the seat section upwards.
3. Unscrew the plastic plate from seat section (4 ball-headed screws M8).
4. Remove the covering cap fastening screws (8 M4 screws).
5. Pull the covering cap off.
ATTENTION!
Risk of damage or destruction!
The end switches are not active while traversing with the emergency
adjustment.
6. Insert the emergency adjustment device through the aperture of the receiving plate
into the hexagonal bolt at the upper end of the lifter spindle.
7. Turn the emergency adjustment device by means of a battery screwdriver or similar
tool.
8. The direction of movement depends upon the direction of rotation of the emergency
adjustment device.
8.2.1 Column
Fault Test Point Result Cause Test Point Action
No Test power Mains Power supply Renew power
function supply power cable supply cable
supply is Mains fuse Remove fuse Renew fuse
not possible (F6, F7) to test it
Battery Batteries Recharge or Replace
supply is check the batteries
not possible batteries
again
Battery fuse Remove fuse Renew fuse
(F8) to test it
Test Internal Internal power Remove fuse Renew fuse
electronics LEDs pack fuse (F3) to test it
display a Electronics Replace
breakdown electronics
Internal Motor fuse (F4) Remove fuse Renew fuse
LEDs do not to test it
display a Electronics Replace
breakdown electronics
No Check Motor runs Toothed Belt Check Replace
column motor run toothed belt toothed belt or
lift tension tighten it
Motor does End switch Check motor Renew the end
not run in the test switch
mode (PC)
Parameters Check Correct
parameters parameters
Motor fuse (F4) Test fuse Renew fuse
Electronics Replace
electronics
Motor Test motor Fit a new
directly using motor
24 V DC
Cabling Test cabling Replace
of the end cabling
switches and
the lifting
motor
8.2.2 Tabletop
L E G U P L E G O N T R E N D R E V E R S E
B A C K U P B A C K O N U P D O W N
R E V E R S E N O R M A L
T IL T L T IL T R
L E V E L
NOTE!
Please refer to the wiring diagram for the fuse arrangement.
Messages advising of specific states are output over the RS232 interface. These can be
either normal operating messages or error messages.
All error messages are stored in the EEPROM and can be read out with the service
program.
NOTE!
Voltages and currents are stored as 8 bit binary values.
Voltage: 30 V = 255
Current: 10 A = 255
NTC: Temperature sensor
Example Composition Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
7 6 5 4 3 2 1 0
2 2 2 2 2 2 2 2
128 64 32 16 8 4 2 1
28 16+8+4 X X X
37 32+4+1 X X X
ATTENTION!
The test program is used for supporting service activities. It may only
be used by trained personnel. It may only be used if there is no danger
to persons.
NOTE!
Two-key operation.
One of the green keys on the membrane keypad on the column must
also be pressed.
O N O F F
T R E N D R E V E R S E
Test 1 Test 2
L E G U P L E G O N
L E G U P L E G O N
B A C K O N U P D O W N
Test 3 Test 4
B A C K U P
B A C K U P B A C K O N
R E V E R S E N O R M A L
T IL T L T IL T R
Test
T R E N D R E V E R S E
L E V E L
start U P D O W N
T IL T L T IL T R
End test
End individual
program
test L E V E L
End individual test: Press the key to the left of the “Zero” key on the IR or cable
remote control (test program continues running).
End test program: Press the “Zero” key (can be done at any time).
A short signal tone sounds to confirm that individual test program 1 is running.
Test description
ATTENTION!
Danger of collision!
The swivelling ranges for tilt and Trendelenburg are freely available.
Two short signal tones sound to confirm that individual test program 2 is running.
Test description
· All movements are disabled.
· The key codes are indicated by the LEDs in the membrane keypad on the column and
in the PC program (monitor).
Application:
· Fault search (operations)
· Tests after repair work
· Maintenance
Three short signal tones sound to confirm that individual test program 3 is running.
Test description
· All movements are disabled.
· All the LEDs in the membrane keypad on the column and the cable remote control are
switched through in rapid succession (exception: charging LED for the cable remote
control).
Application:
· Fault search (indicators)
· Tests after repair work
· Maintenance
Four short signal tones sound to confirm that individual test program 4 is running.
Test description
· All movements (except tilt and Trendelenburg) are disabled.
· The status of the inclination sensor is indicated by the LEDs in the membrane keypad
on the column.
LED 1 LED 6
L E G U P L E G O N T R E N D R E V E R S E
B A C K U P B A C K O N U P D O W N
R E V E R S E N O R M A L
T IL T L T IL T R
L E V E L
Application:
· Fault search (inclination sensor)
· Tests after repair work
· Maintenance
ATTENTION!
The service program (item no. 4 900 833) may only be used by
authorised persons. The basic principle is that, for safety reasons, it
may not be used if there is a patient on the operating table or if any
other persons could come to harm. Furthermore, do not initiate any
actions on the operating table unless you have direct visual contact. In
exceptional cases (modem operation) it is possible to allow an action
to be observed by someone else. This person must be in direct contact
(visual or aural) with the user.
8.5.1 File
Load parameters
Save parameters
Load report
Save report
Deactivate sensors
and end switches
Motor ON/OFF
Op. table OFF
Enter password
Program info
Exit program
8.5.2 State
8.5.3 Driving
Pause between
the test runs Select motor:
Legs
Number of test Back
runs Trendelenburg Status:
Movement of Tilt End position 0
Test mode: Motor current the selected Lift End position 1
ON/OFF (A) drive Reserve Reference pos.
8.5.4 EEPROM
Print or save
window
contents
Delete window
contents
8.5.5 Test
Number of
test runs
Pause during
test run
Test run
ON/OFF
Pause between
test runs
in seconds
8.5.6 Monitor
Input line for commands Status monitor for all Send Send commands
and parameters RS232 outputs, command and parameters
e.g. software versions,
error messages, etc.
8.6.1 Programming
The modem is connected to the COM port of the computer by the zero modem cable
supplied.
To bring the modem into programming mode:
• When switching on, press the button and keep it pressed until the green LED on the
front of the modem lights up.
• The software for configuring the modem is integrated in the service software.
• Before starting, the target path in the properties must be changed to
"........ Mars.EXE_1_M" (_=space).
Data bits
Recall
Parity
Connect
Stop bits
Hang up
Baud rate
Call
Hospital:
• The modem must be connected to the operating table by the RS232 connecting cable
(item no. 4 220 902).
• The modem must be connected to a free, analog telephone line.
• The modem must be connected to the mains supply (230 V).
• The green LED must light up.
Service partner:
• The modem must be connected to the computer by the zero modem cable.
• The modem must be connected to a free, analog telephone line.
• The modem must be connected to the mains supply (230 V).
• The green LED must light up.
If the telephone number stored in the modem corresponds to that of the other connection,
then a connection is made by pressing a button on one of the two modems.
If the stored telephone number does not correspond, then a connection can only be
made with the service software, see section 8.6.1.
300 4 081 028 Hexagon socket head cap screw M4x8 DIN912-A2
310 4 216 425 Socket DX 20-12
320 4 216 383 Flange
3 4
7
1 2
5 6
9
1 0
8.11 Maintenance
The maintenance work is to be carried out once a year and includes the following test
operations:
Running gear
No. Task Description Testing Target Value/ Actual
method Display Value/Test
Remark
1 Check wheel shells Visual No damage. reliable
wheel running
2 Test operating lever Visual No damage, safe
interlock
3 Test free-wheel Function Ease of displacement
4 Test directional running Function Safe interlock
Easy-running
5 Test brake Function Even braking
easy-running
6 Test brake Bowden cable Visual No damage
7 Test directional travel Visual No damage
Bowden cable
8 Test potential Visual No damage
compensation plug
9 Check pictograms Visual No damage, good
recognisability
Column
No. Task description Testing Target value/ Actual
method Display Value/Test
Remark
10 Check the cladding Visual No damage, reliable
wheel running
11 Test membrane keypad Visual, No damage, safe
function function
12 Test rubber bellows Visual No damage
13 Test mains socket Visual No damage, no charring
14 Test interface socket Visual, No damage, safe
function function
15 Check the IR receiver Visual No damage, no
contamination
16 Check wiring installation Visual No kinks, worn or
crushed places
17 Test cable connections Visual securely plugged cable
connections
18 Test lifting motor and Socket Tightly seated
spindle holder head
wrench
19 check toothed belt Visual No damage, toothed belt
tension
20 Grease column telescope Grease Sufficient lubrication of
the telescope surfaces
21 Check the Trendelenburg Visual secure holders, safe
drive function
22 Test tilting drive Visual secure holders, safe
function
23 Test lift end switch Function No damage, safe
switching off
24 Test tilting and Function No damage
Trendelenburg stop switch
25 Test power train Visual No damage, wiring
installation
26 Check mechanical elements Visual No damage, connections
(cardans/holders) fixed securely
27 Check charging end Measuring UL = 28.8 V
voltage device
28 Test charging current Measuring Charging current
device available after
connecting to the mains
29 Check batteries Function minimum 30 minutes of
continuous operation
30 Check mains connecting Visual No kinks, worn or
cable crushed places
31 Check type plate, Visual No damage, easily
pictograms and test plate recognisable
Tabletop
No. Task description Testing Target value/ Actual
method Display Value/Test
Remark
32 Check the pads Visual No damage, no
contamination
33 Check the connecting Visual No damage, tightly fitted
plates
34 Check the standard rails Socket Tightly fitted
head
wrench
35 Test the leg section spars Spirit No differing positions
level
36 Test the back section Spirit No differing position
spars level
37 Check the toothed belt Visual No damage, toothed belt
tension
38 Check the wiring Visual No kinks, worn or
installation crushed places
39 Check the OR sensor Magnet Reliable switching
40 Test the inclination sensor Spirit Horizontal tilting and
level Trendelenburg zero
position
41 Test the end and reference Function, Reliable switching off,
sensors feeler gap 0.2 mm ± 0.1 mm
gauge
42 Test the locking of the Function Test fully interlock in
cam lever closed state
43 Test the Hirth joints Visual No play when in closed
state
44 Check the mechanical Visual No damage, tightly fitted
elements (sockets, gears)
45 Check the head sealing Visual No damage, tightly fitted
46 Check the gear seals Visual No damage
47 Grease the spindles Grease Sufficient lubrication of
(longitudinal and the spindles
transverse displacement)
48 Check the pictograms Visual No damage, good
recognisabilityt
2 x
1 . O N
3 .
1 . 2 .
Appendices
G e b r a u c h s a n w e is u n g b e a c h te n !
R e fe r to o p e r a tin g m a n u a l! M A R S
V o ir le s in s tr u c tio n s d e s e r v ic e ! 4 9 0 0 5 3 9 - 0 9 / 0 0
V é a s e in s tr u c c io n e s d e u so !
157
158
Appendices
1 . 1 . 1 . 1 .
2 . 2 . 2 . 2 .
3 . 3 . 3 .
2 . 1 . 3 . 1 .
3 .
2 .
1 .
2 .
Service Manual MARS 4 900 519 - 02/02
3 .
1 .
4 . 3 .
2 .
Service Manual MARS 4 900 519 - 02/02
Kurzgebrauchsanweisung Operation instruction in brief Mode d'emploi abrégé Instrucciones de uso en breve Istruzioni in breve
1. ON 2x
2.
1.
3. +
2.
Gebrauchsanweisung beachten!
Appendices
Refer to operating manual!
Voir les instructions de service! MARS Endouro
Véase instrucciones de uso! 4 900 505 - 11/01
Osservare il manuale di istruzioni!
159
160
Appendices
Kurzgebrauchsanweisung Operation instruction in brief Mode d'emploi abrégé Instrucciones de uso en breve Istruzioni in breve
1. 1. 1.
2. 2. 2.
3. 3. 3.
1.
1.1
3. 2.
Service Manual MARS 4 900 519 - 02/02
1.2 3.
1.
4. 2.1
2.2
2.
MARS Universal
MARS endouro
MARS Universal
MARS endouro