Siaretron4000 960501 UE Rev1
Siaretron4000 960501 UE Rev1
Siaretron4000 960501 UE Rev1
User’s Manual
GENERAL INFORMATION
The information contained in this manual are the exclusive property of SIARE
Engineering International Group s.r.l. and may not be reproduced in any way
without authorisation. SIARE Engineering International Group s.r.l. reserves
the right to modify or replace this manual at any time without prior notice.
It is however recommended that you make sure you have the most recent
version of the manual. In the event of doubt, contact SIARE Engineering
International Group s.r.l. (see the address on page IX). The information
contained in the present User’s Manual can be considered correct, but do not
exclude professional knowledge by the user.
The operation and maintenance of Siaretron 4000 15” lung ventilator must be
entrusted to qualified technical personnel only. The responsibility of SIARE
Engineering International Group s.r.l. concerning the Siaretron 4000 15” lung
ventilator and its use is limited to what is indicated in the guarantee supplied.
The contents of this manual do not in any way limit the right of SIARE
Engineering International Group s.r.l. to revise, change or modify without
prior notice the equipment (including the relative software) described herein.
Unless otherwise specifically agreed in writing, SIARE Engineering
International Group s.r.l. is not obliged to supply such revisions, changes or
modifications to the owner or user of the equipment (including the relative
software) described herein.
The information contained in this manual refers to the versions of Siaretron
4000 15” lung ventilator produced or updated after January 2019. It is
possible that some information may not apply to previous versions. Contact
SIARE Engineering International Group s.r.l. if you have any doubts.
Siaretron4000e - 15 III
Observations
SIARE Engineering International Group s.r.l. wishes to thank you for
purchasing one of its products.
Any comment on the accuracy and usefulness of this User’s Manual would be
very helpful in allowing us to guarantee current and future users of the high
quality level of our manuals. We would be grateful if you would send us your
comments (see address at page IX).
The SIARE trademark is used throughout this manual as an abbreviation for
the manufacturer: SIARE Engineering International Group s.r.l.
Definitions
Three symbols are used in this User’s Manual to indicate particularly important
information.
WARNING !!
This indicates a condition of danger for the patient or for the
operator.
CAUTION
This indicates the possibility of danger to the lung ventilator.
NOTE
This indicates information worthy of note, making the
operation of the Siaretron 4000 15” lung ventilator more
efficient or practical.
Siaretron4000e - 15 V
SIARE recommends establishing a maintenance and service contract
with SIARE or the local authorised service dealer in order to guarantee
the scheduled maintenance required to operate the lung ventilator in a
safe and correct manner.
To prevent the risk of fire, keep the lung ventilator and/or the oxygen
tubes away from matches, lit cigarettes and inflammable material, such
as anaesthetic gases and/or sources of heat.
Do not connect the lung ventilator to the patient by flexible connectors,
and antistatic or conductive tubes to prevent patient burnings during the
use of high frequency surgical equipment, especially dangerous with
antistatic tubes. The use of flexible connectors, antistatic or conductive
tube is never permitted with lung ventilator.
Do not use worn and consumed tubes or tubes contaminated by
flammable substances like grease or oil to deliver oxygen; (fabrics, oil
and other fuels can easily ignite and they intensively burn in air with
high concentration of oxygen.
In the event of fire or an unpleasant smell (e.g. a smell of burning), the
lung ventilator should immediately be disconnected from the electrical
power supply and from the battery (if fitted).
When coming into contact with any component of the lung ventilator,
the hospital procedures for the handling of infected material should
always be respected.
SIARE is aware that cleaning, sterilisation and disinfection procedures
vary considerably from one health structure to another. SIARE cannot
be held responsible for the efficacy of the cleaning and sterilisation
procedures, nor for the other procedures carried out while the patient is
being treated. As regards cleaning, sterilisation and disinfection of the
product components, it is therefore recommended that the regulations
currently in force in the country where the lung ventilator is installed be
taken into consideration.
The lung ventilator was not designed as a total monitoring device: some
conditions of danger for the patients treated with vital support
equipment will not trigger any alarm.
Before using the lung ventilator or any connected component, carefully
check that the lung ventilator is functioning correctly; when needed, the
preliminary tests must be performed as described in the present
manual.
Do not use pointed instruments, such as pencils, screwdrivers or the
like to make selections or settings as they could damage the surface
of the LCD panel.
Siaretron4000e - 15 VII
WARNING !!
The lung ventilator is not approved for operation in places
where there is any risk of explosion.
Do not use the lung ventilator in the presence of flammable
gases.
The lung ventilator cannot be used in the presence of
explosive gases.
WARNING !!
The lung ventilator shall not be used in a hyperbaric
chamber.
The lung ventilator shall not be used with nitric oxide.
The lung ventilator shall not be used with helium or mixtures
with helium.
WARNING !!
Before starting the lung ventilator use, you have to carry out the
preliminary checks.
WARNING !!
Before connecting the lung ventilator to other electrical
equipment not described in this manual, a request for
authorisation should be sent to Siare.
WARNING !!
Qualified staff must make the regulation of ventilation
parameters.
WARNING !!
Do not block the gas intake port or emergency intake port,
thereby interfering with PATIENT ventilation.
WARNING !!
Means for independent ventilation system shall be available (i.e.
manual resuscitation bag with mask) whenever the lung
ventilator is in use.
Year of manufacture
Check the identification data label of the Siaretron 4000 15” lung ventilator
in the relative chapter.
Manufacturer
SIARE Engineering International Group s.r.l.
Via Giulio Pastore, 18 - 40056
Località Crespellano, 40053 Valsamoggia (BO), ITALY
Tel.: +39 051 969802 - Fax: +39 051 969366
E-mail: mail@siare.it - Web: www.siare.it
Siaretron4000e - 15 IX
Electromagnetic Compatibility
The Siaretron 4000 15” lung ventilator is designed to operate in the
specified electromagnetic environment (see warning below).
The customer or the user of Siaretron 4000 15” lung ventilator should
ensure that it is used in such an electromagnetic environment.
Siaretron4000e - 15 XI
3 PREPARATION TO USE .............................................................................................. 3-1
3.1 Notes .................................................................................................................... 3-2
3.2 Before the use ...................................................................................................... 3-4
3.2.1 Assembling of O2 cell ............................................................................................ 3-4
3.2.2 Battery charger ...................................................................................................... 3-5
3.3 Preparation to use ................................................................................................ 3-7
3.3.1 Medical gas connection ......................................................................................... 3-7
3.3.2 Connection of medical gas supply from cylinder ................................................... 3-8
3.3.3 Mains power supply ............................................................................................... 3-9
3.3.4 External 12Vdc power supply .............................................................................. 3-12
3.3.5 Protection fuses ................................................................................................... 3-13
3.3.6 Patient circuit supporting arm .............................................................................. 3-14
3.3.7 Patient circuit connections ................................................................................... 3-14
3.3.8 Use of antibacterial filter ...................................................................................... 3-15
3.3.9 Gas analyzer connection ( Gas Sensor ) ............................................................. 3-16
3.3.10 Data connection (Trend and Events downloading )............................................. 3-17
3.3.11 Nebulizer connection ........................................................................................... 3-18
3.3.12 Humidifier connection .......................................................................................... 3-18
3.3.13 Connection to other equipments .......................................................................... 3-20
3.4 List of predisposition sequence for use .............................................................. 3-21
3.5 Use ..................................................................................................................... 3-22
3.5.1 Preliminary tests .................................................................................................. 3-22
3.5.2 Ventilator switch ON / Self Test phase ................................................................ 3-24
3.5.3 Turn the lung ventilator OFF ................................................................................ 3-28
3.6 Preliminary checks - Introduction........................................................................ 3-29
3.7 Preliminary checks - Supplementary Tests ........................................................ 3-30
3.7.1 Leak Test ............................................................................................................. 3-31
3.7.2 Respiratory Flow Sensors Calibration ................................................................. 3-34
3.7.3 O2 Sensor calibration .......................................................................................... 3-35
3.7.4 Exit from Supplementary Tests............................................................................ 3-38
3.8 Preliminary checks - Lung Ventilator .................................................................. 3-39
3.8.1 Preliminary checks – MONITORING PARAMETERS ......................................... 3-41
3.8.2 Preliminary checks – ALARMS ............................................................................ 3-43
3.8.3 Alarm limits check ................................................................................................ 3-44
3.8.4 Conclusions ......................................................................................................... 3-48
3.9 Preliminary checks sequence list........................................................................ 3-49
Siaretron4000e - 15 XIII
4.11 GRAPHICs visualization ..................................................................................... 4-57
4.11.1 Combinations of the graphics displayed .............................................................. 4-57
4.11.2 Modification Charts combination.......................................................................... 4-59
4.11.3 Loops combination modification .......................................................................... 4-61
4.11.4 TRENDS visualization.......................................................................................... 4-64
4.11.5 EVENTS visualization .......................................................................................... 4-65
4.12 Default parameters ............................................................................................. 4-66
4.13 Alarms SETUP.................................................................................................... 4-67
4.14 Patient data SETUP............................................................................................ 4-67
4.15 List of functions ................................................................................................... 4-67
4.16 List of default parameters ................................................................................... 4-70
4.17 Calibration Programs .......................................................................................... 4-71
4.17.1 Preliminary ........................................................................................................... 4-71
4.17.2 Calibration Programs displaying .......................................................................... 4-72
4.17.3 Respiratory Flow Sensors Calibration ................................................................. 4-73
4.17.4 High Altitude usage ( On - Off ) ........................................................................... 4-76
4.17.5 CAN module Test................................................................................................. 4-77
4.17.6 VTEc ( On - Off ) .................................................................................................. 4-78
4.17.7 ScreenShot Enable ( On - Off )........................................................................... 4-79
4.17.8 Self Test ............................................................................................................... 4-80
4.18 Power Off ............................................................................................................ 4-82
4.19 Other functions ................................................................................................... 4-83
4.19.1 Reset to ZERO the “Partial operating hours “ ...................................................... 4-83
4.19.2 Data Connection (Trend and Events downloading) ............................................. 4-84
4.19.3 Default parameters set ........................................................................................ 4-86
4.19.4 Touch Screen set ................................................................................................. 4-86
Siaretron4000e - 15 XV
A ANNEX ......................................................................................................................... A-1
A.1 Technical sheet..................................................................................................... A-2
A.1.1 Ventilator for Intensive Care - code 960502 .......................................................... A-2
1.1 Presentation
The Siaretron 4000 15” (hereinafter called lung ventilator) is a new generation equipment,
designed for the treatment of acute and sub acute diseases and can be used on Adults,
Children or New Born patients.
The lung ventilator provides new advanced features for operative modes management; it is
equipped with different ventilation functions and thanks to its keyboard and decoder knob
the user’s selection of most suitable settings is simplified.
For those having a basic knowledge on how lung ventilators for intensive care work, the use
of this equipment is intuitive and a brief training course on regard would be enough.
The user’s interface (GUI) includes the keyboard, the decoder knob and the screen; the last
one displays the lung ventilators settings and measured data, as well as various functions,
allowing the operator an immediate evaluation of the patient conditions; moreover, it is
possible to select and display the temporal trends of the pressure, flow, volume, the loops
of flow/volume, pressure/volume.
An immediate information management system, allows the User to set the alarms, collect
data concerning the trend of the operating parameters (TREND) and the lung ventilator
EVENTS log using the MENU.
The same system allows the operator to set the PATIENT TYPE (Adult, Child or New Born),
load or erase the PATIENT DATA and in case of needs, load automatically the DEFAULT
PARAMETERS of the lung ventilator.
Siaretron4000e - 15 1-1
1.1.1 Main technical characteristics
The Siaretron 4000 15” lung ventilator is composed of two main blocks: the trolley and the
lung ventilator block. The trolley, if required, can be equipped with a built-in medical air
compressor.
In turn, the lung ventilator block is divided in two parts: the upper part that includes a 15”
TFT colour monitor touch screen, the board for the elaboration of all data and information
(CPU) that supervises the operation and display of patient parameters. To manage the
pneumatic part, the CPU board uses serial transmission system by bus, type Controller
Area Network (also known as CAN-BUS) that allows it to check and communicate with the
electronic boards peripherals.
The CAN communication system was expressly designed to operate without any problem
also in environments highly disturbed by the presence of electromagnetic waves. The EMC
immunity level is further increased by using twisted pair type connection cables.
The lower part of the lung ventilator block includes the pneumatic control and all the fittings
or connections, both electrical and pneumatic, to the outside. In this area are also mounted
the back-up batteries for use in case of power supply absence and the board for battery
charge management.
The Siaretron 4000 15” foresees the possibility to upgrade the software for implementing
advanced ventilation modalities and functionalities. The ventilator block can be used as a
stand-alone part being separable from the trolley, by positioning it on a flat surface.
Obviously, this operation can be done only in case the medical air compressor is not
provided within the trolley.
The Siaretron 4000 system delivers controlled or spontaneous ventilations with a re-
adjustable level of end expiration positive pressure (PEEP), of the trigger sensitivity and
oxygen concentration.
Adult, Child or New Born ventilation are available, thanks to an adjustable minimum Tidal
Volume from 2 ml to 3000 ml.
After the switching-on of the lung ventilator it is possible to choice the patient type (Adult,
Child or New Born) setting the relevant default parameters.
The lung ventilator is equipped with a flow and pressure trigger, also it includes the most
modern ventilation modes: volume controlled ventilation modalities VC/VAC, VC/VAC-
BABY, pressure controlled ventilation modalities APCV (BILEVEL ST), APCV-TV, SIMV by
Volume and by Pressure, Pressure supported modalities PSV (BILEVEL S), PSV-TV,
CPAP, APRV, SIGH, Non Invasive Ventilation (NIV APCV - NIV PSV), Drug Nebulizer
(NEB.) and Manual Ventilation (MAN).
In spontaneous ventilation mode, it ensures inspiratory flow up to 240 l/min, both with
control and support pressure.
The medical air compressor (optional) supplies compressed air to the ventilator block and it
can be used in spite of medical air distribution system or from cylinder.
The present manual explains how to use the Siaretron 4000 15” unit and
how performing some simple maintenance procedures.
SIARE recommends to read carefully the present manual and its relevant
instructions before using the lung ventilator or proceeding to maintenance.
The lung ventilator incorporates a series of sensors for continuous patient monitoring, the
most important of which are:
the flow sensors on the expiratory (external) / inspiratory lines (internal), are used to
measure the expiratory / inspiratory volumes of the patient;
the pressure sensors (internal), used to control the pressure of the airways or of the
medical gases;
the oxygen sensor (external), used to measure the concentration of oxygen in the gas
inspired by the patient.
Siaretron4000e - 15 1-3
The output signals (from the, pressure, flow and oxygen sensors) are
filtered by an R-C circuit from the input circuits. This particular electronic
filtering is used to eliminate disturbances before the signals themselves
are processed by the microprocessor.
Before using the lung ventilator, the User should check the operation of
all these sensors in order to avoid any incorrect assessments of patient's
condition.
WARNING !!
Before using the lung ventilator on a patient, it is necessary to perform a
series of preliminary checking to verify the correct operation of the
equipment.
The preliminary checking has the aim to verify the correct connections
and functionalities of the ventilator and all its parts.
For its employ the Siaretron 4000 15” lung ventilator has been designed
and made to guarantee full quality of the product and its components, in
order to ensure the maximum reliability of the lung ventilator for the
patient and user safety.
EN 60601-1-2: 2015 Medical electrical equipment - Part 1-2: General requirements for
safety - Collateral standard: Safety requirements for medical electrical
systems.
IEC 601-1-6:2013 Medical electrical equipment - Part 1-6: General requirements for
basic safety and essential performance - Collateral standard:
Usability.
IEC 601-1-8:2012 Medical electrical equipment - Part 1-8: General requirements for
basic safety and essential performance - Collateral Standard: General
requirements, tests and guidance for alarm systems in medical
electrical equipment and medical electrical systems.
IEC 62353:2014 Medical electrical equipment - Recurrent test and test after repair of
medical electrical equipment
IEC 601-2-12:2007 Medical electrical equipment - Part 2-12: Particular requirements for
basic safety and essential performance of critical care ventilators
ISO 80601-2-12:2011 Medical electrical equipment - Part 2-12: Particular requirements for
basic safety and essential performance of critical care ventilators
ISO 15223-1:2016 Medical devices - Symbols to be used with medical device labels,
labelling and information to be supplied - Part 1: General
requirements
DIR. 2011/65/CE RoHS Directive (on the restriction of the use of certain hazardous
substances in electrical and electronic equipment)
Siaretron4000e - 15 1-5
This page has been added to make front / back copy easier.
Siaretron4000e - 15 2-1
2.1 Overall view
WARNING!!
All the pictures and the examples shown in the present chapter have the sole
purpose of being an example and they do not make any reference to real
clinical cases.
Siaretron 4000 15” lung ventilator, equipped with a 15” TFT display touch screen;
available models:
code 960501 medical gas-driven: Adults, Children and New Born patients
code 960502 turbine-driven: Adults, Children and New Born patients
For the description of above listed parts make reference to the paragraphs
beside highlighted.
Siaretron4000e - 15 2-3
2.3 Lung ventilator front view
Instruction label
O2 SENSOR INLET Cable end RJ connector for O2 cell connection (see cfr. 3.2.1)
EXP. FLOW
SENSOR INLET RJ connector for flow sensor connection
INSP. TO PATIENT Inspiration connector for patient circuit (see cfr. 3.3.7)
EXP. FROM
PATIENT Expiration connector for patient circuit (see cfr. 3.3.7)
EXP. FLOW
SENSOR Expiratory flow sensor
Siaretron4000e - 15 2-5
2.3.1 Electric connections side
FUSE 2x1 AT Protection fuses for main power supply circuit at 220 Vac
(2 x 1AT - 250V)
100 - 240 VAC Plug for main power supply connection at 100 - 240 VAC
Siaretron4000e - 15 2-7
2.3.2 Gas supply side
Siaretron4000e - 15 2-9
2.4.1 Touch screen
Operative command
Cancel
Siaretron4000e - 15 2-11
Respiratory parameters visualization
Select P / F.
The PAW / FLOW Loop appears in place of
PAW / Tidal Volume Loop
Siaretron4000e - 15 2-13
2.4.2 Keyboard with soft key and encoder knob
A control keyboard and an encoder knob are available on the upper side of the lung
ventilator.
These components allow a rapid interaction between the User and the lung ventilator.
Siaretron4000e - 15 2-15
2.4.3 Operative mode
You can select one of these ventilation modes using the touch screen
(please see 2.4 .1 or 2.4.2).
The VC/VAC-BABY Operative Mode is not active when the PATIENT TYPE
selected is: Adult
After selecting the most suitable operative mode for patient ventilation, the
system will automatically display the physiological respiratory parameters for
the new setup.
Alongside the Operative Mode acronym, the patient type set is specified
(Adult, Child, New Born). In this way the default respiratory parameters
are set automatically (respiratory parameters and alarms levels).
Default Parameters: please refer to cfr. 4 Use of Lung Ventilator and cfr.
5 Alarm.
Siaretron4000e - 15 2-17
2.4.4 Patient Data
For more information about PATIENT DATA set, please see on chapter 4.5.
Siaretron4000e - 15 2-19
2.4.6 PRP list
CPAP/PEEP (cmH2O)
Continuous positive airway pressure during respiration phase in
CPAP operative mode.
I:E
Ratio between inspiration and expiration phases.
Flow (L/min)
Flow value during inspiration phases
O2
Concentration percentage delivered to the patient can be set
from 21% to 100%.
Pause (%)
Inspiratory pause time. The “inspiratory pause time” is displayed
on the screen in % (% of the inspiratory time). It is also used to
calculate the lung mechanics parameters (resistance and static
compliance).
PEEP (cmH2O)
Positive airway pressure value during expiratory phase.
Pinsp (cmH2O)
Maximum airway pressure limit value. The parameter is used in
pressure controlled modes to fix an operating limit for the airway
pressure that shall not be exceeded.
PMax (cmH2O)
Maximum airway pressure limit.
PS (cmH2O)
Positive airway support pressure value during inspiratory phase.
RR (bpm)
Lung ventilator respiratory rate.
RRsimv (bpm)
Value of forced respiratory rate in SIMV mode.
Ti (s)
Time that defines the lung ventilator inspiration duration. The
values can be set based on the set RR.
Ti max (s)
Time that defines the maximum duration of an inspiration. If the
duration of the inspiratory phase is lower than the set value, the
patient will be forced to exhale.
Tr. E (%)
Percentage of the inhaled flow with regard to the maximum
peak where the inspiratory phase ends and the expiratory
phase begins.
Tr. I (L/min) (cmH2O)
Flow level (pressure) for detecting the patient spontaneous
breathing.
Siaretron4000e - 15 2-21
Vte (ml)
Expired tidal volume guaranteed for the patient.
Vti (ml)
Inspired tidal volume guaranteed for each breath.
Based on the ventilator parameters set by the User and on the patient's
characteristics, the lung ventilator is able to monitor and measure a
series of values necessary for the patient's clinical evaluation.
At the top of the screen, there is a bar led indicator that displays the
pressure inside the airways in real time. The measured and monitored
values (right side of the screen) are updated after each breath of the
patient.
E.g. Here below the data in the images below refer to VC/VAC operating
mode with standard PRP.
The displayed value shows the positive pressure at the end of the
expiration: the measurement unit is cmH2O.
The User can control if the ventilator is able to reach and keep the
PEEP pressure level set, using this value.
It shows the ratio between the inspiration time and the expiration
time.
It shows the volume value expired by the patient per minute: the unit
of measurement is L/min.
You can also calculate this value using the formula: current volume
(Vte) x respiratory rate (RR).
Description
Siaretron4000e - 15 2-23
2.4.8 Additional respiratory parameters
Based on the ventilator parameters set by the User and on the patient's
characteristics, the lung ventilator is able to monitor and measure a
series of values necessary for the patient's clinical evaluation.
The measured and monitored values (right side of the screen) are
updated after each breath of the patient.
E.g. Here below the data in the images below refer to VC/VAC operating
mode with additional PRP.
Static compliance
It is one of the parameters of the lung mechanics: measured in
ml/cmH2O, represents the lung compliance when the patient does
not breathe.
You can use it to assess the lung elasticity: the higher the
compliance, the more elastic the “lung”; the lower the compliance,
the more “rigid” the lung.
The static compliance can be calculated using the formula below:
Cs = current inspired volume / pause pressure.
Dynamic compliance
It is one of the parameters of the lung mechanics: measured in
ml/cmH2O, represents the lung compliance either during the
inspiration or during the expiration.
You can use it to assess the lung elasticity: the higher the
compliance, the more elastic the “lung”; the lower the compliance,
the more “rigid” the lung. The dynamic compliance can be
calculated using the formula below:
Cd = current inspired volume / peak pressure
Leak
Shows the ‘Leak’ value measured that must be adequate to the
ventilation mode enabled (volumetric or pressometric) and within the
range in compliance with the local regulations.
The unit of measurement is %.
Pause pressure
It shows the pause pressure: the unit of measurement is cmH2O.
When the inspiratory pause enables, the lung ventilator maintains
the airway pressure constant (it maintains a pause pressure) for a
certain amount of time of the inspiratory time, defined by the User
(INSP PAUSE %). The static conditions allow the lung ventilator to
calculate the breathing mechanics parameters.
Inspiratory resistance
It is the parameter of the lung mechanics that describes the
resistance to the opposite flow of the airways: measured in
cmH2O/(l/s). The greater the patient resistance, the higher the
airway pressure you need to apply to obtain the same volume.
The formula used by the ventilator to calculate the inspiratory
resistance is as follows:
Ri = (peak pressure – pause pressure) / inspired flow.
Siaretron4000e - 15 2-25
Expiratory time
It shows the duration of the patient's expiratory phase: the unit of
measurement is the second. This parameter defines the expiration
duration. This value depends on the respiratory rate and I:E ratio
parameters.
Example: if RATE = 15 and I:E=1:1 you will have an expiratory
phase of 2 seconds.
Inspiratory time
It shows the duration of the patient's inspiratory phase: the unit of
measurement is the second. This value represents the total
inspiratory time, and also includes the inspiratory pause period. This
value depends on the respiratory rate and I:E ratio parameters.
For example: if RATE = 15 and I:E=1:1 you will have an inspiratory
phase of 2 seconds.
Inspiratory pause
It shows the duration of the patient's inspiratory standby phase: the
unit of measurement is the second. This parameter represents the
inspiratory time during which the lung ventilator keeps the airway
pressure constant.
Example: if RATE=15, I:E=1:1, Ppause=50% you will have an
inspiratory pause period of 1 second.
The ventilator features automatic means for detecting and identifying any
conditions that might put the patient at risk (based on the level of urgency
and seriousness), using acoustic or visual alarm signals.
The role of the alarm signal is to draw the attention of the User to the
event as well as to inform him on the requested response speed.
For more information, please see on chapter 5 Alarms.
The User can display the alarm set limits, selecting the dedicated
Alarms icon. After editing the alarm settings, the relative signal will
remain active and the status icon will blink for a pre-set time.
ALARM SILENCING
Select the bell icon (or press the RESET key) to interrupt the
acoustic signal for a pre-set period of time.
During the silencing period, the text of the alarm will still be
displayed.
Select the bell icon (or press the RESET key) once again to delete
the alarm text, only if the alarm activation condition is no longer
present.
If during the silencing period, a new alarm (of high priority) occurs,
the alarms silencing command is automatically cancelled and the
acoustic and visual signals are reactivated.
Siaretron4000e - 15 2-27
2.4.10 Measures, charts and loops
The lung ventilator is equipped with tools for charts and loops display so as
to quickly and accurately notify the User on the patient's condition.
The User selecting the icon (GRAPHICs) can choose “which“ and
“how” to display in time the following detections:
Charts: PAW , Flow , Tidal Volume, O2, CO2
Loops: Tidal Volume / Flow , PAW / Tidal Volume , PAW / Flow
Measures: respiratory parameters
Lung status icon
The lung status icon simulates the patient's lungs, graphically displaying
the respiratory cycle by alternatively switching the lungs color.
In fact, in case of patient's spontaneous activity (Trigger), the lung status
icon turns to yellow and if the “Low Pressure of Airways” alarm value
set has not been exceeded the icon turns red.
For more information, please see on chapter 4.10.
The lungs turn green during inspiration only if the PAW exceeds the “Low
Pressure of Airways alarm value set.
Siaretron4000e - 15 2-29
2.4.12 Graphic setting, operative functions and general informations
In the lower side of GUI there are a series of controls and functions that are fundamental for
use of Siaretron 4000 lung ventilator.
Leds: if Led is lit: indicate that the relevant function is enabled (or if it has been
enabled); if the led is off or in case it does not switch on, it means the relevant
function is not enabled.
NEBULIZER function
The “NEB” function is active only when the lung ventilator is supplied by
oxygen.
MAN operative mode (lit green led: MAN function has been enabled)
General information
Battery level (if the symbol is steadily lit and green, the battery is
full).
Presence of mains power supply, (the “green plug” symbol means
that the device is powered from mains).
Siaretron4000e - 15 2-31
2.5 Rear view
One USB socket for CPU programming (near the reset button) and one USB
socket for Screen Shoot (for more details see on Service Manual).
FUSE Protection fuses for 220 Vac power supply circuit (2 x 3,15
AT - 250V without air compressor).
100 - 240 VAC Connection of main power supply cable 100 - 240 VAC.
100 - 240 VAC Plug for “Electronic Heated Humidifier” power supply.
The mark identifies the protection level against electric shock (category of
protection type B).
2019
The number indicates the year of production of the equipment.
The symbol indicates ‘refer to the instructions for use’ of the equipment.
Siaretron4000e - 15 2-33
This page has been added to make front / back copy easier.
3.1 Notes
3.2 Before the use
3.2.1 Assembling of O2 cell
3.2.2 Battery charger
3.3 Preparation to use
3.3.1 Medical gas connection
3.3.2 Connection of medical gas supply from cylinder
3.3.3 Mains power supply
3.3.4 External 12Vdc power supply
3.3.5 Protection fuses
3.3.6 Patient circuit supporting arm
3.3.7 Patient circuit connections
3.3.8 Use of antibacterial filter
3.3.9 Gas analyzer connection (Gas Sensor)
3.3.10 Data connection (Trend and Events downloading)
3.3.11 Nebulizer connection
3.3.12 Humidifier connection
3.3.13 Connection to other equipment’s
3.4 List of predisposition sequence for use
3.5 Use
3.5.1 Preliminary tests
3.5.2 Ventilator switch ON / Self Test phase
3.5.3 Turn the lung ventilator OFF
3.6 Preliminary checks - Introduction
3.7 Preliminary checks - Supplementary Tests
3.7.1 Leak Test
3.5.1 Respiratory Flow Sensors Calibration
3.7.2 O2 Sensor calibration
3.7.3 Exit from Supplementary Tests
3.8 Preliminary checks - Lung Ventilator
3.8.1 Preliminary checks – MONITORING PARAMETERS
3.8.2 Preliminary checks – ALARMS
3.8.3 Alarm limits check
3.8.4 Conclusions
3.9 Preliminary checks sequence list
Siaretron4000e - 15 3-1
3.1 Notes
CAUTION
If this is the first time you install the lung ventilator, it is suggested to
consult thoroughly this user manual.
Before using the unit, clean the external surfaces and sterilize the
components.
Use the maintenance instructions provided in this user manual and
respecting the regulations in force in the country where the unit is sold.
Siaretron4000e - 15 3-3
3.2 Before the use
3.2.1 Assembling of O2 cell
In case of electric power failure, the lung ventilator is equipped with a internal batteries that
guarantees (if perfectly efficient) at least 90 minutes operation (with default physiological
respiratory parameters).
The switching to battery operation is made automatically: on the lung ventilator screen
appears the relevant message “ON BATTERY”.
The lung ventilator battery can be recharged by connecting the equipment to main power
supply (using power cable supplied with the unit) and placing the main switch in " I "
position.
Carry a battery charge at least 8 hours, before using the lung ventilator the
first time.
The operating time of lung ventilator guaranteed by the battery, can vary in
the following cases:
old battery or not perfectly efficient;
not standard ventilatory parameters.
Siaretron4000e - 15 3-5
Follow the instructions.
Verify on the upper side of the lung ventilator, if the green led is on (it indicates the
presence of main power supply).
Screw the gas supply hoses to the relevant connectors of the hospital distribution
system. Be sure that the pneumatic hoses are well fixed and that the hospital medical
gas pipeline system is clean and without lubricants.
The pneumatic hoses are already supplied with screw connection DISS
type (Diameter Index Safety System) for connection to the lung ventilator
connectors.
The installation technician taking care of the hoses connection to the
outlets must ensure that they are compatible with the hospital medical gas
pipeline system.
To prevent inversions of gases that can be FATAL for the patient, assembly
of the connectors compatible with the hospital distribution system and all
the maintenance and/or replacement operations of the medical gas supply
hoses must be carried out by highly qualified technical personnel only.
The lung ventilator can also work with AIR medical gas supply only, but in
this case the FiO2 will be adjustable at 21% only. An alarm indicating “Low
Oxygen Pressure” condition will be activated.
Siaretron4000e - 15 3-7
WARNING !! Unit failure risk.
In order that the lung ventilator operates as specified, the input medical
gas pressure must be between 280 kPa and 600 kPa (2,8 - 6 bar / 40 -
86 psi).
Before using the lung ventilator, make sure that this requirement is met.
After connecting the medical gas supply hoses, make sure that the
system works properly.
Screw the pneumatic O2 supply hose on the relevant lung ventilator connector.
Connect the pneumatic O2 supply hose on the relevant connector of the O2 cylinder
pressure reducer.
Ensure that the medical gas supply hoses are connected and work correctly.
The electrical connections are a very important part in the installation of the lung ventilator.
Incorrect connections or connections to unsuitable electrical systems can compromise the
safety of the patient and the User.
Mains power supply must comply with the prescriptions in CEI 64-8/7 standards concerning
the locations intended for type A medical use.
The power supplies foreseen on lung ventilator unit are of three types:
by main power supply (100 - 240Vac / 50 - 60Hz)
by internal battery (2 batteries Pb / 12Vdc - 1.3Ah)
by external power supply (12Vdc / 4,2A).
Siaretron4000e - 15 3-9
Verify on the upper side of the lung ventilator, if the green led is on (it indicates the
presence of main power supply).
CAUTION
The Siaretron 4000 15” lung ventilator complies with the requirements
for electro-medical devices detailed on chapter 1.3 (Norms and
standards regulations).
The User should be always sure that the power supply electric plug is
accessible during the normal using and functioning of the Siaretron
4000 15” lung ventilator.
If the batteries are not fitted, the lung ventilator is not protected against
fluctuations or failures of the power supply.
The lung ventilator must not be used without batteries charged and in
perfect conditions.
The use of the battery must be limited to short periods and is not
foreseen as an alternative to the mains power supply.
Do not open the lung ventilator to replace the battery or to carry out
maintenance operations on the battery charger.
When the green led on the upper side of the lung ventilator is light on, it indicates that the
lung ventilator is correctly supplied.
The operation time of the lung ventilator guaranteed by the battery can vary
in the following cases:
old battery or not perfectly efficient;
not standard ventilatory parameters.
Siaretron4000e - 15 3-11
Battery charger
The lung ventilator battery can be recharged by leaving the unit connected
to the main power supply (using power cable supplied with the unit) it is not
necessary that the unit is on.
For battery recharging follow what specified in technical data sheet (Battery
re-charging time).
Siaretron4000e - 15 3-13
3.3.6 Patient circuit supporting arm
The system will check the patient circuit every time you switch ON the lung
ventilator.
If not carefully positioned, the hoses, the cables, the patient circuit and
other similar components to the lung ventilator, these can be dangerous
for the patient.
Siaretron4000e - 15 3-15
3.3.9 Gas analyzer connection (Gas Sensor)
For more details and information about the CO2 analyzer (Sidestream or
Mainstream model) please refer to the user manual supplied with the GAS
ANALYZER.
CAUTION
GAS ANALYZER - Zeroing
Mainstream IRMA sensor: after around one minute from the device is
turned on, it is necessary to perform the manual procedure of Zero
Calibration (please see 3.9 TESTS ON DEMAND).
Connect a USB flash drive to the USB1 socket used for CPU
programming and data download (USB socket near the reset button).
Request the guide with the relative instructions for data interpretation
(Procedure for downloading data).
Siaretron4000e - 15 3-17
3.3.11 Nebulizer connection
Use the control provided on the Graphical Users Interface (NEB function)
to enable the Nebulizer functionality (see on cfr 4.10).
Siaretron4000e - 15 3-19
3.3.13 Connection to other equipments
Siaretron4000e - 15 3-21
3.5 Use
To obtain better performances, leave the lung ventilator working for at least
15 minutes, before patient connection or before executing preliminary
checks.
This operation will allow the system to reach the correct temperature for the
spirometry.
Before using the lung ventilator on a patient, it is necessary to perform some of preliminary
tests in order to check that the equipment is properly operating.
The preliminary checks have the aim to verify the correct connection and functionality of the
lung ventilator and all its components.
The list of preliminary tests is available at the end of the present chapter or in
APPENDIX chapter.
Before starting preliminary tests, the Siaretron 4000 lung ventilator must be:
The patient simulator suggested for tests and checking’s is SIARE code
LS.AB.001 which is equipped with variable compliance and resistance.
Emergency conditions
In emergency conditions, the preliminary checks can be skipped.
You should carry out the preliminary checks once the emergency
condition stops, and at least once a week.
Preventive MAINTENANCE
The preliminary checks do not remove the necessity for periodical preventive
maintenance operations carried out by SIARE authorised staff, aimed at
replacing the worn parts and checking the overall lung ventilator condition
(please refer to Maintenance chapter).
For the periodical checks that you should carry out, please refer to
Maintenance chapter.
Set the main switch (placed on the back of the lung ventilator) to “I”.
Hold the ON-OFF key for few seconds; the lung ventilator turns ON and the automatic
Self Test phase starts.
During Self Test phase, the lung ventilator software carries out the self-
diagnostic tests and checks a series of devices necessary for safe operation
of the lung ventilator /patient.
Turbine: Absent
If you do not hear any acoustic alarm signal and/or you did not press the
Alarm Reset key, the red message “ Press OK to begin anyway ” will appear
on the screen.
The system will display the PATIENT DATA page and later on it will
switch to the Stand-by visualization.
During the switching OFF the system performs a clean of the oxygen inside
the lung ventilator which grants a 21% concentration on the inspiratory line.
This procedure grants a longer O2 sensor life and a quicker Self Test phase
when the device is turned-on. The duration of this procedure depends by
the FiO2 present in the inspiratory line and can be of 60 seconds max.
- SUPPLEMENTARY TESTS
Leak Test
Respiratory Flow Sensor Calib. (see note at the beginning of chap. 3.7.2)
O2 Sensor calibration
- LUNG VENTILATOR
Respiratory parameters
Spirometry
- OPTIONAL
CAUTION
The lung ventilator must be ready for use in order for you to proceed with
the preliminary checks.
Connect the power supply, the medical gases and the patient circuit.
Insert and connect the oxygen sensor.
Connect a patient simulator to the patient circuit terminal.
Lung ventilator ON: Stand-by mode.
To carry out the Supplementary Tests you must know the keyboard
operating mode and the options available in the lung ventilator SETTING
MENU (please see chapter 2.4).
The Supplementary
Tests screen
appears.
This test verifies that there are no leaks higher than 100 ml in the
anaesthesia unit pneumatic circuits.
Perform the Leak Test weekly.
LEAK TEST
Based on known and calculated data (flow, pressure and time), the
software calculates the parameters displayed at the end of test.
The value of the compliance obtained during the Leak Test is used for
the “compensation of dead space” function (e.g. 0,3 x cmH2O
measured).
Risk of DM failure
The Leak Test is not overcome when the pressure within the pneumatic
circuit does not arrive to 30 cmH2O.
If the problem persists, contact the Siare Service Centre or a Centre
authorised by Siare.
Select Respiratory
Flow Sensors
Calibration.
CAUTION
Not being a useful procedure for the purpose of "Preliminary Checks", see
chapter 4.17 for a more detailed description of the same.
This procedure should be carried out to check the proper operation of the
oxygen sensor.
Perform this procedure monthly.
Select O2 Sensor
calibration.
The O2 sensor
calibration in progress.
if the O2 sensor is worn out (the oxygen detection cell is worn out;
replace the O2 sensor).
The system
automatically quits O2
sensor calibration after
a few seconds.
To carry out the preliminary checks you need to know how the system and
the Physiological Respiratory Parameters (PRP) work.
For tests and checks, please use the patient simulator SIARE cod.
LS.AB.001 that is equipped with variable resistance and compliance.
Vti 500
RR 15
I:E 1:2
Pause 0
PEEP OFF, 5, 10 cmH2O
O2 21%
Tr. I -2 cmH2O
1 L/min
Based on the PRP set by the User and on the patient's [patient simulator]
characteristics, the lung ventilator is able to monitor and measure a
series of values necessary for the patient's clinical evaluation.
Before checking the value of the set parameters, leave the lung ventilator
on for at least 15 minutes. This way the system will be able to reach its
operating condition.
CAUTION
If the O2 measured value differs from the set value by more than +/- 10%,
please repeat the “O2 Sensor Calibration” procedure: Supplementary
Test.
If the Vte measured value differs from the set value by more than +/- 20%
(Adult parameters), please go to the “CALIBRATION PROGRAMS”
visualization to perform the calibration of the expiratory flow sensor (see
chapter 4.16).
To carry out the preliminary checks relative to lung ventilator's alarms, you
need to know how the system and the alarms work: please see chapter 2.4.
Check the Alarms Limits and if necessary, change the values set, based
on the test you want to carry out.
Select the Alarm parameter you want to change. For more information
about use of touch screen or key board, please see on chapter 2.4.
Select
YES: to quit Alarm page; the alarm set will NOT be saved.
NO: it remains in Alarm page.
Select
YES: to quit Alarm page; the alarm set will be saved.
NO: it remains in Alarm page
Before starting the alarm limits, check the alarms setup, and change the set
values whenever necessary (please see previous paragraph).
Set the high pressure alarm limit to a value higher than the PAW
ventilation pressure by 5 cmH2O.
Block the patient simulator (using your hands) during ventilation.
High Pressure
The system activates the airways HIGH PRESSURE alarm: silence
the alarm.
Unlock the patient simulator.
PAW
High respiratory Set the high respiratory rate alarm limit to 15 bpm.
rate The system activates the high RR alarm: silence the alarm.
Restore the default alarm value.
RR
During ventilation please set RR = 10 bpm.
Low respiratory Set the low respiratory rate alarm limit to 15 bpm.
rate The system activates the low RR alarm: silence the alarm.
Restore the default alarm value.
Carry out all preliminary checks and make sure that they were completed successfully
before connecting the patient to the Siaretron 4000 15” lung ventilator.
CAUTION
Preliminary checks phase failed.
Please see Alarms chapter and/or Trouble Shooting chapter.
Please contact the nearest Siare Support Centre or any other support
centres authorised by Siare.
All maintenance and/or repair interventions require full knowledge of the lung
ventilator, and therefore such operations must be carried out only by highly
qualified staff, specifically trained and authorised by SIARE.
Siaretron4000e - 15 4-1
4.9 Ventilation phase
4.9.1 Ventilation interruption
4.10 Operative functions and Graphic settings
4.11 GRAPHICs visualization
4.11.1 Combinations of the graphics displayed
4.11.2 Modification Charts combination
4.11.3 Loops combination modification
4.11.4 TRENDS visualization
4.11.5 EVENTS visualization
4.12 Default parameters
4.13 Alarms SETUP
4.14 Patient data SETUP
4.15 List of functions
4.16 List of default parameters
4.17 Calibration Programs
4.17.1 Preliminary
4.17.2 Calibration Programs displaying
4.17.3 Respiratory Flow Sensors Calibration
4.17.4 High Altitude usage ( On - Off )
4.17.5 CAN module Test
4.17.6 VTEc ( On - Off )
4.17.7 ScreenShot Enable ( On - Off )
4.17.8 Self Test
4.18 Power Off
4.19 Other functions
4.19.1 Reset to ZERO the “Partial operating hours “
4.19.2 Data Connection (Trend and Events downloading)
4.19.3 Default parameters set
4.19.4 Touch Screen set
CAUTION
Before subjecting the patient to a lung ventilation treatment, please:
set the airway pressure limit alarm to a value that does not exceed 20
cmH2O; this way you will prevent any problems that might arise due to
incorrect respiratory volume or frequency setup (you can increase the
pressure if the patient's pathology and conditions require such
modification).
check the set oxygen concentration (FiO2) as high concentrations might
affect the patient's health.
please consult this User Manual.
Siaretron4000e - 15 4-3
4.2 Stand-by mode
Set the main switch (placed on the back of trolley/lung ventilator) in position “I” (ON).
Hold the ON-OFF key for few seconds; the lung ventilator turns ON and the automatic
Self Test phase starts.
During Self Test phase, the lung ventilator software carries out the self-
diagnostic tests and checks a series of devices necessary for safe operation
of the lung ventilator /patient.
Press OK to begin
Press Cancel for other tests
PATIENT DATA
Stand-by mode
Siaretron4000e - 15 4-5
4.3 PATIENT DATA / SETUP parameters
Actually, the lung ventilator software does not switch directly from Self Test
phase to Stand-by operative mode, but it previously displays a page which
allows the adjustment of the following types of information and machine
setup.
PATIENT DATA (see on 4.3.1)
SETUP parameters (see on 4.3.2)
Patient type
Male / Female
Name / Surname
Physical date
Birthday
Note
The system displays the "Stand-by" page without saving the data
stored in the PATIENT DATA page.
The system displays the "Stand-by" page, saving the data stored in
the PATIENT DATA page.
CAUTION
During the normal operation of the lung ventilator, the
User can modify the PATIENT DATA selecting the
function: SETTING MENU / PATIENT DATA (cfr. 2.4) or
by touching the relative icon (PATIENT DATA).
Select OK :
Select Cancel :
Siaretron4000e - 15 4-7
Once completed the entering of
patient data, save or cancel what
is indicated in the page.
Select
Select
CAUTION
During the normal operation of the lung ventilator, the
User can modify the parameters SETUP selecting the
function: SETTING MENU.
Display
BRIGHTNESS
ENERGY SAVING
SOUND VOLUME
TOUCH AUDIO
Siaretron4000e - 15 4-9
Date & Time
Date
Time
Language
Italian
English
German
Turkey
Polish
French
Russian
Spanish
Units
Weight (referred to the
patient)
Height (referred to the
patient)
CO2 (unit of measurement)
Pressure (unit of
measurement)
Other
NIV Enable
Power Failure
APNOEA TIME
CHANGE PASSWORD
Save to USB
Select
YES: to quit SETUP page
without saving.
NO: it remains in SETUP page.
Siaretron4000e - 15 4-11
Select
YES: the set values will be
saved.
NO: it remains in SETUP page.
4.4.1 Mode 1
Select: SETUP
Select: Language
Siaretron4000e - 15 4-13
A series of MENU languages are
available and identified by an
icon.
Select: Language
Siaretron4000e - 15 4-15
A series of MENU languages are
available and identified by an
icon.
4.5.1 Mode 2
Select: Display
Siaretron4000e - 15 4-17
Select: PATIENT DATA
Patient type
Male / Female
Name / Surname
Physical date
Birthday
Note
Select: Default
Siaretron4000e - 15 4-19
Press YES: to RUN the
selected DEFAULT.
Press NO: to Cancel the
command.
Select: ALARMS
Select
YES: to quit Alarm page; the
alarm set will NOT be saved.
NO: it remains in Alarm
page.
Select
YES: to quit Alarm page; the
alarm set will be saved.
NO: it remains in Alarm
page.
Siaretron4000e - 15 4-21
4.7 Operative modes
In the following chapter you will find a description of available Operative Modes selectable
on Siaretron 4000 15” lung ventilator.
CAUTION
When the lung ventilator is turned ON, the system restores the operative
mode and the relevant parameter values set before the last shut-down.
Just for our examples we refer to the Operative Modes available with
ADULT Patient Data (VC/VAC).
CAUTION
A new operative mode can be selected in two different operating conditions.
1. Lung ventilator in Stand-by mode.
2. During the normal operation of the lung ventilator.
CAUTION
If the lung ventilator is in Stand-by mode, the system does not require any
confirmation or enabling for setting a new operative mode.
Siaretron4000e - 15 4-23
2. Lung ventilator in normal operation
CAUTION
When the lung ventilator is working, the system requires a confirmation by
the user for the modification of the operative mode.
Siaretron4000e - 15 4-25
Operative modes list
APCV is a pressure controlled ventilation, synchronised with the patient's breathing with
leak compensation.
With this parameters configuration, APCV is a pressure controlled ventilation,
synchronised with the patient's breathing, during which the system generates a patient
ventilation at a pre-set inspiratory pressure (Pinsp), a pre-set flow (Flow), a calculated I:E
ratio and a settable respiratory rate (RR).
In APCV the current volume depends on the inspiratory pressure (Pinsp) and on the
patient's lungs characteristics (compliance, lung capacity) therefore the tidal volume will
vary depending on changes in lung mechanics.
During the inspiratory phase, the lung ventilator generates a settable flow (Slope). When
the airway pressure reaches the control value (Pinsp), this pressure level is kept constant
by the lung ventilator until the end of the inspiration that you can set using (RR).
Siaretron4000e - 15 4-27
4.7.3 APCV-TV
PSV is an assisted type of ventilation with pre-set pressure support (PS) with guaranteed
safety respiratory rate set by the User in case of patient apnoea (RR bk) and with leak
compensation.
PSV is a ventilation technique during which, at the beginning of the patient's spontaneous
inspiratory effort, the lung ventilator provides a constant positive support pressure (PS)
pre-set by the User with high-speed flow supply, until the pressure inside the airway
reaches the desired support value.
If the patient does not trigger (spontaneous breathing during the apnoea
time set in SETUP - Other), the system enables the APNOEA acoustic and
visual alarm.
The system will automatically provide an APCV ventilation with set safety
respiratory rate (RR bk) and I:E ratio (I:E bk).
Siaretron4000e - 15 4-31
4.7.5 PSV-TV
PSV-TV is an assisted pressure support ventilation with guaranteed current volume and
guaranteed safety respiratory rate set by the User in case of patient apnoea (RR bk).
If the patient does not trigger (spontaneous breathing during the apnoea
time set in SETUP - Other), the system enables the APNOEA acoustic and
visual alarm.
The system will automatically provide an APCV-TV ventilation with set safety
respiratory rate (RR bk) and I:E ratio (I:E bk).
Siaretron4000e - 15 4-33
4.7.6 VC-VAC
In this type of ventilation, the work of breathing is fully assumed by the lung ventilator, and
therefore it is used when the patient is unable to breath on his own, or in order to assure
an efficient pre-set current volume and therefore the mechanical ventilation must fully
replace the spontaneous breathing.
The inspired volume (Vti) is pre-set and generated in a pre-set time (RR and I:E) and
determines the characteristics and the pressure range necessary to reach the pre-set
amount of gas mixture that must be provided.
The patient's breathing attempt is detected by the system (Tr. I) and it automatically sends
inside the airway a gas flow at a pre-set volume (Vti).
Siaretron4000e - 15 4-35
4.7.7 VC-VAC BABY
The VC/VAC BABY operative mode also includes two additional parameters
identifying the maximum and the minimum pressure limit (PMax. - Pmin) that
can be reached during ventilation.
In this type of ventilation, the work of breathing is fully assumed by the lung ventilator, and
therefore it is used when the patient is unable to breath on his own, or in order to assure
an efficient pre-set current volume and therefore the mechanical ventilation must fully
replace the spontaneous breathing.
Siaretron4000e - 15 4-37
4.7.8 V-SIMV
SIMV allows the patient to breath spontaneously, between the forced breaths, with a pre-
set positive pressure support (PS) if the patient's breath is strong enough to enables the
flow trigger (Tr. I - this parameter can’t be set to OFF).
The spontaneous phase is characterised by the set inspiration time (Ti) that once the
pressure support value (PS) set by the User is reached, leaves its place to the expiration
phase (Tr. E).
Therefore, in SIMV mode, the lung ventilator can provide a combination of spontaneous
and controlled breathing.
SIMV mode is frequently used as a weaning ventilation mode from a fully controlled
ventilation (completely depending on the lung ventilator) to an assisted ventilation mode.
Siaretron4000e - 15 4-39
4.7.9 P-SIMV
SIMV allows the patient to breath spontaneously, between the forced breaths, with a pre-
set positive pressure support (PS) if the patient's breath is strong enough to enable the
flow trigger (Tr. I - this parameter can’t be set to OFF).
The spontaneous phase is characterised by the set inspiration time (Ti) that once the
pressure support value (PS) set by the User is reached, leaves its place to the expiration
phase (Tr. E).
Therefore, in SIMV mode, the lung ventilator can provide a combination of spontaneous
and controlled breathing. SIMV mode is frequently used as a weaning ventilation mode
from a fully controlled ventilation (completely depending on the lung ventilator) to an
assisted ventilation mode.
Siaretron4000e - 15 4-41
4.7.10 CPAP
Siaretron4000e - 15 4-43
4.7.12 MAN operative mode
By selecting the MAN mode (at the bottom of the display) the system provides the
patient with a breath.
The breath ventilation parameters depend on the set operative mode.
The function activation is monitored by the screen and signalled by the green LED
inside the box, that turns on.
This mode is active while the lung ventilator is running.
Apnoea BACK-UP is a safety function available in two of the Operative modes: PSV,
PSV-TV and CPAP. The Apnoea BACK-UP function enables if the patient, ventilated in
one of the modes above, stops breathing.
After a pre-set time (Apnoea Time) in the SETUP - Other, will appear the relative alarms
and the system automatically starts ventilating the patient.
CAUTION
When the lung ventilator is turned ON, the system restores the operative
mode and the relevant parameter values set before the last shut-down.
Just for our examples we refer to the Operative Modes available with
ADULT Patient Data (VC/VAC).
Selecting the PRP parameters icon (see the image on the next page) the
parameters relevant to the set Operative Mode are displayed
Siaretron4000e - 15 4-45
VC/VAC operative mode: available PRP parameters.
Selecting PRP icon parameters (see the image above) the respiratory
parameters are displayed.
Siaretron4000e - 15 4-47
4.8.2 Monitoring of additional respiratory parameters
Based on the ventilator parameters set by the User and on the patient's
characteristics, the lung ventilator is able to monitor and measure a
series of an additional values necessary for the patient's clinical
evaluation.
The data in the images below refer to VC/VAC operative mode with
standard PRP, and they are only informative, they do not refer to real
clinical cases.
On the left side of the display you can find PRP parameters.
In the main section you can see the Charts that show the trend of the
respiratory parameters.
On the right side of the display you can find values that will help you
asses the patient's clinical condition.
In the upper part the Operative Mode is displayed as well as both Start
and Stop controls for the ventilation and the message for active alarms.
At the bottom, the functions to be activated by the user as well as time
and status of the power supply are available.
Siaretron4000e - 15 4-49
4.9.1 Ventilation interruption
During the patient ventilation, the User can intervene using the graphical
user interface, by selecting the “Operative Functions” in the lower side on the
screen.
Siaretron4000e - 15 4-51
Select the icon to
access the lung
ventilator's ALARM
LIMITS.
Siaretron4000e - 15 4-53
Select the INSP HOLD
mode: the system will
extend the inspiration
time to 20 seconds.
The function activation is
displayed by the screen
and signalled by the
yellow LED inside the
box, that turns on.
The User can enable both modes in sequence; the lung ventilator will carry
out both modes, giving priority to the mode that was activated first.
Select the icon to supply After the pre-set time the systems automatically restores
an oxygen concentration the previously set O2 concentration value.
of 100% to the patient. The function activation is monitored by the screen and
signalled by the yellow LED inside the box, that turns on.
The NEB. function is enabled only when the lung ventilator is connected
to an oxygen source (connection: HIGH PRESSURE).
The NEB. operative mode can be enabled and disabled manually using
the icon.
Siaretron4000e - 15 4-55
MAN operative mode
By selecting the MAN mode the system provides the patient
with a breath.
Select the icon to activate The breath ventilation parameters depend on the set operative
the manual ventilation to mode. The function activation is monitored by the screen and
the patient. signalled by the green LED inside the box, that turns on.
This mode is active while the lung ventilator is running.
The icon GRAPHICs allows the user to combine the Loops, the Charts
and the Measured patient respiratory parameters in different ways
Siaretron4000e - 15 4-57
Charts: PAW
Loop: Tidal Volume /
Flow , PAW / Tidal
Volume
3. Lung status icon
Measures: Vte
5. Trends
The selection of the box of one of the Charts displayed, enable the
visualization of a screen of a drop-down menu with the list of the options
available.
Siaretron4000e - 15 4-59
Release; the drop-down
menu with the list of the
options available
appears.
Select CO2.
The CO2 graphic is
displayed and it replaces
the Tidal Volume
graphic.
The procedure described is applicable to all the Charts areas and in the
different visualizations available.
The selection of the frame of one of the Loops displayed, enable the
visualization of a drop-down menu with the list of the options available.
Available Loops: Tidal Volume / Flow , PAW / Tidal Volume , PAW / Flow.
Siaretron4000e - 15 4-61
Select Loops: the drop-
down menu with the list
of the options available
appears.
Select P / F.
The PAW / FLOW Loop
is displayed and it
replaces the PAW / Tidal
Volume Loop.
The procedure described is applicable to all the LOOPs areas and in all the
different visualizations available.
Select Vti.
The respiratory
parameter Vti is
displayed and it replaces
the PAW / Tidal Volume
Loop.
The procedure described is applicable to all the MEASURES areas and in all
the different visualizations available.
Siaretron4000e - 15 4-63
4.11.4 TRENDS visualization
Select the TRENDS visualization to monitor the most significant respiratory parameters on
medium - long term.
Monitored Respiratory
parameters:
Rate
P Insp
PEEP
Vm
Vte
The vertical dashed bar (in the graphic) indicates on the graphic the
movement of the values measured in the timeframe.
On the left side of the Trends graphic are mentioned the data measured
and related to the parameter displayed and to the vertical dashed bar.
Select the EVENTS function to monitor the information on the lung ventilator operation over
time. The monitored EVENT refer mainly to the alarms (active alarms) and the various
operating conditions of the lung ventilator (POWER ON, POWER OFF, STAND-BY,
VENTILATION START).
Siaretron4000e - 15 4-65
4.12 Default parameters
The screen PATIENT DATA / Default allows the User to restore the
Default parameters (Factory set).
Restoring “Factory set “ means to set once again the Default parameters
of the lung ventilator (MENU, SETUP, PRP, ALARMS limits).
Select:
YES: exit without
perform any other
operation.
For ALARMS parameters and limits setup please see 5.3.4 ALARMS setup.
For Patient Data parameters and setup please see 4.3.1 PATIENT DATA.
Here below the list of the available functions and positioning inside the
Graphic User Interface.
Siaretron4000e - 15 4-67
Brightness
Energy saving
Display
Sound volume
Touch audio
Date
Date & Time
Time
Italian, English, German, Turkey,
Language
Polish, French, Russian, Spanish
Weight (referred to the patient)
Height (referred to the patient)
Units
CO2 (unit of measurement)
Pressure (unit of measurement)
Trends data Default
Events data Default
Default
Patient data Default
Setting & Ventilation Default
NIV Enable
Power Failure
Other
APNEA TIME
CHANGE PASSWORD
Save to USB
IRMA/ISA Gas Sensor (IRMA or ISA type)
Enabling the Supplementary Tests
function.
The system shows the test
available on the lung ventilator.
Supplementary Tests
Before using the lung ventilator on
a patient, you have to carry out a
series of preliminary checks to
make sure that it works properly.
Turn Off See on chapter 4.9.1
Charts
Loops
Lung status icon
GRAPHICs visualization
Measures
Trends
Events
Siaretron4000e - 15 4-69
4.16 List of default parameters
RR (bpm) 12 12 12
Pinsp (cmH2O) - - 20
Pmin (cmH2O) - 5 -
Pmax (cmH2O) - 25 -
Flow - 3 3
4.17.1 Preliminary
CAUTION
Siaretron4000e - 15 4-71
4.17.2 Calibration Programs displaying
Set the main switch (placed on the back of the trolley/lung ventilator) in position ”I”.
Select Self Test to switch to the normal operation of the lung ventilator
(see cfr 4.2)
The User quit the Respiratory Flow Sensors Calibration and the
system goes back to the ‘Calibration Programs’ visualization.
Siaretron4000e - 15 4-73
Select
At the end of the Respiratory Flow Sensors Calibration phase the system
goes back to the Calibration Programs menu.
In case of first calibration, we suggest to execute this procedure after
having checked the PEEP calibration and the Leak Test (see on 3.7 Test
on Demand)
For more information please refer to the SERVICE manual.
At the end of the Respiratory Flow Sensors Calibration check the system
goes back to the Calibration Programs menu.
Siaretron4000e - 15 4-75
4.17.4 High Altitude usage ( On - Off )
The activation of the Usage at High Altitude (High Altitude: On) function, is useful to
optimize the calculation of oxygen concentration (FiO2) at heights higher than 2000 mt
on the sea level.
Select Self Test to go to the Self Test phase and escape from Calibration
Programs window.
Siaretron4000e - 15 4-77
4.17.6 VTEc ( On - Off )
Select Self Test to go to the SELF TEST phase and escape from
“Calibration Programs”window.
This option, if enabled, allows the User to store on an USB key an instant image (image
saved in bmp format) during the operation of the lung ventilator, (for further details, see on
chapter 2).
ScreenShot function is
enabled (On).
Select Self Test to go to the SELF TEST phase and escape from Calibration
Programs window.
Siaretron4000e - 15 4-79
4.17.8 Self Test
Once the Siaretron 4000 15” lung ventilator is switched ON, the system (SW) carries out
the self-diagnostic tests (Self Test) and checks a series of devices necessary for safe lung
ventilator operation.
For further details, about the Self Test phase, see on chapter 3.5.2.
Siaretron4000e - 15 4-81
4.18 Power Off
The zeroing of the partial operating hours can be performed both in the
page of turbine characterization, and in the expiratory flow sensors
calibration page. As an example, the Expiratory Flow Sensors Calibration
displaying is used.
Siaretron4000e - 15 4-83
4.19.2 Data Connection (Trend and Events downloading)
ScreenShoot enable:
select Self Test
Select Other
The Trend and Events data are saved on the USB drive in TXT format.
Ask Siare for the dedicated program to transform the TXT file into a
readable and interpretable file of the User.
Siaretron4000e - 15 4-85
4.19.3 Default parameters set
By selecting the function: MENU SETUP / DEFAULT, the operator has the possibility to
restore the original factory parameters (default parameters).
With the term "Default Parameters" we refer to all the DM settings (MENU, SETUP,
ALARMS Limits, etc ...).
For more information on the default parameters setup, see chapter 4.11.
The Siaretron 4000 lung ventilator is equipped with a 15” colour Touch Screen display.
In case of Touch Screen system malfunctioning, it is possible to perform calibration of the
same system (see on Service Manual).
5.1 Introduction
5.2 Displaying and used symbols
5.2.1 Alarms display area
5.2.2 A1 - Alarm area
5.2.3 A2 - ALARMS parameter
5.2.4 A3 - General information area
5.2.5 A4 - Acoustic alarm silencing
5.3 Alarms setting
5.3.1 Setting of ALARMS limits values
5.3.2 Setting of ALARMS volume
5.3.3 Setting of DEFAULT parameters
5.3.4 Alarms DEFAULT parameters values
5.4 Summary table of alarm characteristics
5.4.1 Alarms configurable by user
5.4.2 System alarms
5.4.3 Gas Sensor Alarms
5.5 Troubleshooting
5.5.1 Troubleshooting list
Siaretron4000e - 15 5-1
5.1 Introduction
The lung ventilator is equipped with automatic means for detection and identification of
serious and sudden events through alarm signals or information signals.
The aim of the alarm signal is to draw the attention of the user on the event, as well as to
indicate the required response speed.
Level or urgency
Immediate, the event is potentially able to develop in a period of time which generally is
not enough to undertake a corrective manual action.
Brief, the event is potentially able to develop in a period of time which generally is
enough to undertake a corrective manual action.
Delayed, i.e. that the event is potentially able to develop in a not specified period of
time.
Level of severity
The combination of urgency level and severity level of the listed factors,
determines the assignment of priority condition of an alarm situation.
The parameters and the characteristics (activation time, presence or lack of
an acoustic and/or luminous indicator) and the possible user’s actions
respect to the alarm signals (silencing, suspend, inhibit) are described here
below.
A1 - Alarm area: this area of the monitor provides the following indications.
A string of text relevant to the type of active alarm.
An “alarm bell symbol” which indicates the priority and the alarm state.
A ‘lock icon’ to be used to inhibit the screen from possible accidental contacts “
A2 - ALARMS parameter
Touching the icon, it is possible to entry in the ALARMS area and to access the Min and Max.
alarms value setting.
A3 - General information signal area: this area of the monitor provides the following indications.
Battery charge level and main power presence (failure).
Siaretron4000e - 15 5-3
5.2.2 A1 - Alarm area
This area shows the text of the active/s alarm/s and the priority / status of the same and a
lock icon indicating whether the touch screen is enabled.
CAUTION
Three different alarms can be displayed at most and simultaneously; any
further alarm signals will be displayed in sequence at intervals of 3/5
seconds.
Suspended alarm
Media priority
Yellow bell crossed
Yellow bell
through
Suspended alarm
High priority
Red bell crossed
Red bell
through
Siaretron4000e - 15 5-5
5.2.3 A2 - ALARMS parameter
This monitor area, allows to display and set the Min and Max. alarms limit values.
Select
YES: to quit Alarm page; the
alarm set will NOT be saved.
NO: it remains in Alarm page.
Select
YES: to quit Alarm page; the
alarm set will be saved.
NO: it remains in Alarm page.
The lung ventilator used in the same health environments can have
different preset configurations of alarm limits.
Verify that the preset alarm limits are appropriate for the new patient
and adjust the alarm limits on values suitable to the new condition of
use.
The alarms setting to the range limit could make the alarm not
properly working.
This area shows the battery charge level and the power supply status (present/absent).
ORANGE flashing symbol, 1 notch: it indicates that the charge level of the
battery is at 25%.
- The relevant alarm is active (Low Battery: 25% Remaining).
The colour of the last flashing “notch” is red (High priority alarm): this
extremely serious alarm condition indicates that the battery is almost
completely low (Low Battery)
The alarm remains always actives to indicate the occurred
malfunctioning.
Siaretron4000e - 15 5-7
5.2.5 A4 - Acoustic alarm silencing
It’s possible to change the alarm settings even when the alarms are
activated.
After changing an alarm setting, the relevant sign is lighted and the
status icon will flash for a defined time.
NOTE
The acoustic alarm silencing is active and will stop the acoustic alarm for
a defined time.
During the alarm silencing the alarm text is showed on the panel.
A new alarm silencing will cancel the alarm text only if the alarm
conditions is disappeared.
If during the alarm silencing a new high priority alarm occur, the alarm
silencing is cancelled and the acoustic signal and the visual texts are
activated again.
Select
YES: to quit Alarm page; the
alarm set will NOT be saved.
NO: it remains in Alarm page.
Select
YES: to quit Alarm page; the
alarm set will be saved.
NO: it remains in Alarm page.
Siaretron4000e - 15 5-9
Available Alarm Limits.
PAW
PEEP
Vte
Vm
O2
RR
EtCO2
The selection of both the bar or the cursor of the alarm can be done by
using the encoder knob or the touch screen.
Select
YES: to quit Alarm page; the alarm set
will NOT be saved.
NO: it remains in Alarm page.
After modification of Alarm limit ( PAW )
values:
Select
YES: to quit Alarm page; the alarm set
will be saved.
NO: it remains in Alarm page.
Siaretron4000e - 15 5-11
5.3.2 Setting of ALARMS volume
Select Display
To go back to the Stand-by displaying, select Cancel
Select
YES: to quit from Display page; the set
will NOT be saved.
NO: it remains in Display page.
After modification of SOUND VOLUME
value.
Select
YES: to quit Display page; the set will be
saved.
NO: it remains in Display page.
Siaretron4000e - 15 5-13
5.3.3 Setting of DEFAULT parameters
Select Default
To go back to Stand-by page, select Cancel
Select RUN
Select
YES: to quit Default page; the Default
Parameters will be saved.
NO: it remains in Default page.
Possible User’s options.
Select
YES: to quit Default page; the Default
Parameters will NOT be saved.
NO: it remains in Default page.
Siaretron4000e - 15 5-15
5.3.4 Alarms DEFAULT parameters values
Pressure (cmH2O) 5 - 40 5 - 35 5 - 35
Respiratory Rate
(bpm) 8 - 20 12 - 30 25 - 50
PEEP (cmH2O) 0 - 15
FiO2 (%) 21 - 80
Apnea Time
(configurable in 20 sec.
SETUP – Other area)
Low Pressure of
HIGH 15 YES 15 NO
Airways
High Pressure of
HIGH 0 NO - NO
Airways
Low Expiratory
HIGH 15 YES 30 NO
Volume
High Expiratory
HIGH 0 YES 30 NO
Volume
Low Volume
HIGH 60 YES 30 NO
Minute
High Volume
HIGH 60 YES 30 NO
Minute
Low O₂
HIGH 30 YES 30 NO
Concentration
High O₂
HIGH 30 YES 30 NO
Concentration
Siaretron4000e - 15 5-17
5.4.2 System alarms
Battery
HIGH 0 NO - NO
Disconnected
Battery Charger
HIGH 0 NO - NO
Disconnected
Battery
HIGH 0 NO - NO
Overtemperature
(Patient) Circuit
HIGH 0 NO - NO
Disconnected
Maintenance 1000
HIGH 0 YES - NO
hours
Sampling Line
HIGH 0 YES 30 NO
Clogged
Unspecified
HIGH 0 YES 30 NO
Accuracy
Siaretron4000e - 15 5-19
5.5 Troubleshooting
This chapter is an indicative but not exhaustive guide for the user and the technician,
providing indications for eliminating, as quickly as possible, most of the problems that may
have caused malfunctioning or alarm signals.
This chapter describes the possible causes of problems, indicated by alarms that are
activated during normal functioning.
WARNING !!
If the problem persists, carry out a complete check of the lung ventilator to
identify any irregularities.
If the problem cannot be resolved, contact the Siare Service Centre or a
Centre authorized by Siare.
Power Failure There is a power supply fault and the lung ventilator is operating on the
battery.
Check if the LED power indicator is ON.
Check that it is connected to the main power supply.
Check that the main switch is turned to the I position (ON).
Check the correct connections of the plug, the fuses and the
connector, and the condition of the cable (if necessary, restore the
connections and replace the cable if it is damaged).
Check that power is present at the relative socket by plugging in
another electrical device. If there is no power, use another socket
or check the overload switch on the electrical panel of the room.
Check the 12V voltage is properly supplied to the main board.
Siaretron4000e - 15 5-21
Acoustic alarm: at the end of the self test the activate the alarm
sound. If the sound is heard by the operator the alarm reset button
must be pressed, if it is not heard check that the speaker cable is well
connected otherwise the speaker is defective and must be replaced.
Turn the lung ventilator OFF and ON and repeat “SELF TEST”
phase.
If the problem persists, contact the Siare Service Centre or a
Centre authorized by Siare.
Soft keys and This condition occurs when the control keyboard or the Encoder are
encoder knob not working.
Switch the lung ventilator OFF and then switch back ON.
Continuity test of the keyboard and encoder cable.
If the problem persists, contact the Siare Service Centre or a
Centre authorized by Siare.
CAN-BUS failure This alarm condition occurs in case of system failure (electronic
boards).
Check if the CAN-BUS cable between the main board and the EV
board is properly connected.
Contact the Siare Service Centre or a Centre authorized by Siare.
(Patient) Circuit This alarm conditions occur in case of disconnection of the patient
disconnected circuit (missed Vte detection for three times).
Vte volume limit for circuit disconnected condition: 50ml for ADULT,
20 ml for PAEDIATRIC, 0 ml for NEONATAL patient).
Check that the mask, endotrachéal tube and patient circuit are not
in some way split, disconnected or connected wrongly. If this is the
case, eliminate the problem or replace them.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate, I/E,
Trigger).
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
Contact the Siare Service Centre or a Centre authorized by Siare.
Low Air gas This alarm is enabled when the pressure is insufficient (< 2.7 bar) for
pressure the lung ventilator to operate correctly.
Check that the medical air is correctly connected to the lung
ventilator. Restore the connections or replace the tube if damaged.
Check that there is sufficient pressure in the supply system or in
the cylinders. Adjust or repair the supply system (or replace the
cylinders) if the pressure is insufficient.
Contact the Siare Service Centre or a Centre authorised by Siare.
Low battery level This alarm is enabled when the charge level of the battery is at 25%
25% (50%) (50%) of the fully charged level.
Check that it is connected to the main power supply.
Recharge the battery.
If the alarm is activated when the battery has not provided the time
autonomy indicated on the technical sheet, request the intervention
of a Service Centre.
Low battery (10 This alarm condition is present when the charge battery level is such to
minutes) be guaranteed a residual autonomy of about 10 minutes.
Verify the power supply connection; recharge the battery.
If the alarm is activated when the battery has not provided the time
autonomy indicated on the technical sheet, request the intervention
of a Service Centre.
Siaretron4000e - 15 5-23
Battery This alarm condition is present when the battery is not properly
disconnection connected to the device.
Check the battery connection.
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
Battery charger The alarm condition is present when the battery does not charge
disconnected properly.
Battery This alarm condition is present when the battery pack internal
overtemperature temperature pass the 75°C.
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
Disconnected O2 This alarm indicates the connection status of the oxygen sensor.
cell
Check that the oxygen cell is correctly connected.
Check the condition of the cable and the connector (if necessary,
restore the connection and replace the cable if damaged).
Replace the oxygen sensor with a new one.
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
FiO2 % high This alarm is activated when the measured concentration of oxygen
exceeds the set limit.
Check that the corresponding alarm limits are set correctly.
Calibrate the oxygen cell: if the problem occurs again after a short
time, replace the oxygen cell.
Contact the Siare Service Centre or a Centre authorised by Siare.
FiO2 % low This alarm is activated when the measured concentration of oxygen is
below the set limit.
Check that the oxygen cell is fitted correctly in its housing.
Check that the corresponding alarm limits are set correctly.
Calibrate the oxygen cell: if the problem occurs again after a short
time, replace the oxygen cell.
Check that the feeding pressure of the medical gases is correct: if it
is not, check the pressure of the distribution system and the correct
connection to the supply.
Contact the Siare Service Centre or a Centre authorised by Siare.
Min. Vte / VM This alarm condition occurs in case the Vte is lower than set value.
Siaretron4000e - 15 5-25
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
Check if the lung ventilator works properly verifying the airways
pressure. If the lung ventilator works properly perform the
expiratory flow sensor calibration.
Contact the Siare Service Centre or a Centre authorised by Siare.
Max. Vte / VM This alarm condition occurs in case the Vte is higher than set value.
Low (High) PEEP This alarm condition occurs if the measured PEEP value is lower
(higher) than the value set.
PAW low In this condition, the patient circuit + patient system presents a lower
resistance than expected or a higher compliance. This causes
insufficient ventilation pressure.
Check that the corresponding alarm limits are set correctly.
Check that the mask, endotrachéal tube and patient circuit are not
in some way split, disconnected or connected wrongly. If this is the
case, eliminate the problem or replace them.
Check the correct settings of the patient's respiratory parameters
(according to the operative mode selected: Volume/Flow, Rate, I/E,
Trigger).
Check that the luminous PAW bar on lung ventilator (the airways
pressure curve) correctly follows the inspiration / expiration cycle.
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
Check that the low pressure level is higher than the PEEP level set.
If not, increase it above the PEEP level.
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
Siaretron4000e - 15 5-27
RR high This alarm is activated when the respiratory rate value is higher than
the set value.
Check that the corresponding alarm limits and the patient's
respiratory parameters are set correctly.
Check that the patient's respiratory parameters are set correctly.
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
RR low This alarm is activated when the respiratory rate value is lower than the
set value.
Check that the corresponding alarm limits are set correctly.
Check that the patient's respiratory parameters are set correctly.
Check that the lung ventilator operates correctly, checking the
airways pressure trend. If the lung ventilator operates correctly,
check the flow sensor and the correct connection of its cable.
Check that the mask, endotrachéal tube and patient circuit are not
in some way split, disconnected or connected wrongly. If this is the
case, eliminate the problem or replace them.
Check that the sensitivity of the Trigger is appropriate to the
patient's physiological conditions.
Check that the patient circuit is connected correctly to the lung
ventilator and to the patient.
If the problem persists, contact the Siare Service Centre or a
Centre authorised by Siare.
Flow sensor The user can note indirectly, by monitoring the flow graph, the value of
calibration failed the expired volume and the peak value of expired flow, if the self-
calibration of the flow sensor has been successful or not.
Check the proper mounting of the sensor (patient circuit) and the
relevant connections to the lung ventilator and to the patient.
Check if the patient circuit is properly closed during the calibration.
Replace the sensor and repeat the expiratory flow sensor
calibration.
Contact the Siare Service Centre or a Centre authorised by Siare.
Siaretron4000e - 15 5-29
This page has been added to make front / back copy easier.
6.1 Notes
6.2 Cleaning, disinfection and sterilization
6.3 General indications
6.3.1 Cleaning
6.3.2 Disinfection and sterilization
6.3.3 Disinfection by immersion (chemical)
6.4 Cleaning, disinfection and sterilization table
6.4.1 Sterilization of EXP V. Monoblock (exhalation block with flow sensor)
6.4.2 Disposable bacteria filter
6.5 Periodic maintenance
6.5.1 Maintenance operations
6.5.2 Cleaning, disinfection and sterilization before use with another patient
6.6 Repairs and spare parts
6.6.1 Spare parts kit for lung ventilator
6.7 Miscellaneous
6.7.1 Storage
6.7.2 Repackaging and shipment
6.7.3 Disposal
Siaretron4000e - 15 6-1
6.1 Notes
The instructions for carrying out more detailed tests, for trouble-shooting
and for other interventional procedures, information intended for qualified
technical personnel, are contained in the relative chapter.
CAUTION
To avoid the danger of electric shock during maintenance and/or repair
operations, make sure that all power supplies have been disconnected,
disconnect the power supply source (positioning the special danger signs)
and disable the protection switch of the ventilator.
CAUTION
Before performing the maintenance or repairing works, also in case of
returning the ventilator for repairing to manufacturer, it is required to clean
and disinfect the ventilator.
Siare is aware that working procedures can differ considerably from one
health structure to another: it is therefore impossible to indicate specific
procedures that are suitable for all requirements.
Siaretron4000e - 15 6-3
6.3 General indications
6.3.1 Cleaning
Use a disposable cloth moistened with neutral detergent, a chemical substance or the
equivalent; use water to remove any traces of chemical.
Do not clean or re-use disposable products.
Do not use hard brushes to clean the components, or other instruments that could
damage their surface.
Wash the components with hot water and a neutral detergent solution.
Rinse the parts well with clean hot water (tap water can be used) and leave to dry.
Siare recommends that the components should be checked every time they are cleaned
and any damaged parts should be replaced.
Whenever a part or component is changed, check the functioning of the ventilator.
CAUTION
Follow the manufacturer’s instructions for the detergent substances used:
the use of detergents that are too strong could compromise the working life
of the components.
To disinfect the components, dismantle them and place them in a steam disinfection
chamber at 93°C for 10 minutes.
After this first operation and before placing the components in an autoclave, wrap them in
muslin or in a similar material.
Effective sterilization is achieved in an autoclave at 121°C for approx. 15 minutes.
CAUTION
That the components should be checked every time they are sterilized
and any damaged parts should be replaced.
If a steam disinfection chamber is not available, the dismantled parts can be chemically
disinfected by means of immersion.
Immerse the dismantled components in the solution with the disinfectant, following the
manufacturer’s instructions.
CAUTION
Not using formaldehyde or phenol-based disinfectants as they can
cause cracking and reticulation of plastic parts;
Not using too strong disinfectants as they can compromise the working
life of the immersed parts;
Rinsing and carefully drying the components since marks and other
damage can occur when the components are exposed to high
temperatures.
When disinfection is complete, rinse with running, preferably decalcified, water; shake and
drain off any remaining water. Leave the components to dry completely.
After this first operation and before placing the components in an autoclave, wrap them in
muslin or in a similar material.
Effective sterilization is achieved in an autoclave at 121°C for approx. 15 minutes.
Siaretron4000e - 15 6-5
6.4 Cleaning, disinfection and sterilization table
oxygen-releasing compounds.
Remove any dust from the surfaces or in
openings using a vacuum cleaner or a soft
cloth.
Dismantle and clean with hot water and a Before mounting the filter
Compressor neutral detergent solution. again, eliminate any humidity
air filter
(optional) Check that there are no splits and replace inside the components by
them if they are damaged. means of compressed air.
Siaretron4000e - 15 6-7
Perform daily cleaning of the mask
following the instructions of the
responsible doctors or recommended
by the Manufacturer.
Mask Always clean the mask and the hoses See Manufacturer’s
or use a new mask in case the lung instructions.
ventilator must be used with a
different patient.
CAUTION
The EXP V. monoblock includes the expiratory valve and the flow
sensor.
Siaretron4000e - 15 6-9
6.4.2 Disposable bacteria filter
CAUTION
It is important to use a disposable bacteria filter at the
expiratory port of ventilator between the patient circuit’s
expiratory limb and the expiratory flow sensor.
Do not clean or re-use if the filters are the Components that cannot be
Disposable disposable type. destroyed should be sterilized
bacteria filter and disinfected according to
local standards.
CAUTION
The lung ventilator does not require particular maintenance and preventive
operations other than those indicated in this manual or in order to respect
standards applied in the specific country where the ventilator is sold.
Contact SIARE for information regarding authorised Service Centers in your area.
When you require service, please indicate the serial number of the unit and the problem
to SIARE or to your authorised technicians.
SIARE assumes responsibility for all provisions foreseen by the law, if the equipment is
used and maintained as per the instructions in this manual and the technical manual
The Technical Assistance Report, signed by the authorised SIARE technician, is proof
of the completion of the scheduled maintenance.
The table summarizes the preventive maintenance frequency and procedures to be carried
out on the lung ventilator.
Several times a day Patient circuit Check for any water collection, drain and clean the
/ according to local tubes when necessary.
practice and
standards Disposable bacteria Replace
filter
Condensation trap Check for any water collection, drain and clean
filter when necessary.
Siaretron4000e - 15 6-11
Frequency Component Procedure / Action
Every day / when Oxygen sensor Calibrate according to the procedures described in
necessary this manual.
Condensation trap Check for any water collection, drain and clean
filter when necessary.
Every 6 months or Lung ventilator The lung ventilator must be inspected and
1000 working hours checked in general and any worn parts must be
(*) replaced.
CAUTION
6.5.2 Cleaning, disinfection and sterilization before use with another patient
We recommend the use of procedures for sterilization and disinfection referred to in the
preceding paragraphs when a new patient must use the lung ventilator.
Use only original SIARE spare parts or spare parts checked and approved
by SIARE.
Code: R501000A1
Spare parts kit for annual maintenance to be used with the Siaretron 4000
15”, code 960501.
Siaretron4000e - 15 6-13
6.7 Miscellaneous
6.7.1 Storage
If for any reason the lung ventilator is not used, we suggest leaving it in its
original packaging and storing it in a safe and dry place.
If it is believed that the lung ventilator will be left unused for at least 6
months, Siare recommends disconnecting the battery or recharging it every
3/6 months, depending on the storage temperature.
6.7.3 Disposal
The batteries and the accumulators are special waste materials and they
must be disposed of in appropriate containers in accordance with local
regulations for the disposal of such waste materials.
For further information contact the relevant authorities for environmental and public health
monitoring.
Siaretron4000e - 15 A-1
A.1 Technical sheet
GENERAL DATA
Siaretron 4000 15” electronic lung ventilator is equipped with a TFT LED 15”
touch screen - colour monitor displaying the curves of pressure, flow, volume,
the loops of breathing parameters, the trends and the ventilation parameters.
Siaretron 4000 15” lung ventilator is suitable for ventilation of adult, paediatric
and neonatal patients.
Siaretron 4000 15” is equipped with a flow and pressure trigger, it provides
the most advanced volume controlled ventilation modalities VC/VAC,
VC/VAC-BABY, pressure controlled ventilation modalities APCV (BILEVEL
ST), APCV-TV, SIMV by Volume and by Pressure, Pressure supported
modalities PSV (BILEVEL S), PSV-TV, CPAP, APRV, SIGH, Non Invasive
Ventilation (NIV APCV - NIV PSV), Drug Nebulizer (NEB) and Manual
Ventilation (MAN).
Siaretron 4000 15” is supplied with back up long lasting batteries and its
software can be updated for new modes and last generation ventilation
strategies.
NORMS
The lung ventilator complies with the essential requirements and it is realized
0476 according to the references of the Annex II of 93/42/EEC Medical Devices
Directive.
ENVIRONMENTAL CONDITIONS
Operating Relative humidity: 30 - 95% non-condensing
Temperature: from +10 to +40°C
Weight 26 Kg
Power Max 50 VA
Supply pressure Oxygen - Air: 280 kPa - 600 kPa / 2.8 - 6 bar / 40 – 86 psi
Dead space compensation Automatic compensation of mechanical and patient circuit dead space
Automatic leaks Present: with FLOW parameter set in AUTO, in NIV modalities (NIV APCV,
compensation NIV PSV): max 60 l/min
Siaretron4000e - 15 A-3
Leak % visualization Present
Ventilation modalities APCV (BILEVEL ST), APCV-TV, PSV (BILEVEL S), PSV-TV (Auto
Weaning), VC/VAC, VC/VAC BABY, V-SIMV+PS, P-SIMV+PS, CPAP,
APRV
Inspiratory pressure limit Pinsp: from 2 to 80 cmH2O (in function of low and high pressure alarm set)
Pressure trigger By adjustable pressure from OFF; -1 to -20 cmH2O under PEEP level (step of
1 cmH2O)
Inspiratory flow max. 120 l/min per every gas (240 L/min max.)
Patient circuit Double hose 150 cm. Adult/Paediatric patient circuit (expiratory valve on
the ventilator)
Double hose 150 cm. Neonatal patient circuit (expiratory valve on the
ventilator)
USER INTERFACE
Touch screen monitor Module with TFT LED display with touch screen
Dimensions 15”
Siaretron4000e - 15 A-5
Display keyboard Keyboard for rapid access of functions. Encoder knob for:
selection, set up and confirmation of physiological breathing parameters
Turbine Characterization
VTEc
Nebulizer Enable
ScreenShoot Enable
MENU function - SETUP Display (Brightness, Energy Saving, Sound Volume, Touch Audio)
Language
Turn Off?
EtCO2 ( % )
Displayed graphics CURVES: Pressure (PAW) - Flow - Volume (Vte) - O2 (CO2 optional)
Events
Trends
Physiological breathing Vti (ml), RR (bpm), I:E, Pause (%),PEEP (cmH2O), O2 (%),Tr. I (L/min -
parameters setting cmH2O), SIGH (Sigh. Amp. (%), Sigh. Int. (b)), Vte (ml), PMax, Pmin, Pinsp
(cmH2O), Flow (L/min), BACK-UP parameters, PS (cmH2O), RRsimv (bpm),
Ti (s), Ti Max (s), Tr. E (%), CPAP (cmH2O), Pressure High - Low (cmH2O),
Time High - Low (s).
Displayed parameters PAW, PEEP, CPAP (cmH2O), RR (bpm), I:E, O2 (% - l/min), Vte (ml),
Siaretron4000e - 15 A-7
Flow sensor Magnetic disturbance (patented), multi-usage type
ALARMS
Alarms visualization Max 3 alarms simultaneously; additional alarms, scroll every 3-5 sec.
System alarms
SELF-TEST alarms
Pressure sensor Verification of pressure sensor operation through control of PAW reading
Acoustic alarm Verification by the user of acoustic signal emission, the confirmation of the
test is made by silencing of that alarm
ACCESSORIES
Supplied Accessories User's Manual
O2 supply hose
Nebulizer set
Antibacterial filter
Power cable
O2 cell
Optional Accessories Refer to price list.
Siaretron4000e - 15 A-9
This page has been added to make front / back copy easier.
Siaretron4000e - 15 A-11
ALARM CHECKS
EMPTY - Yellow - -
Siaretron4000e - 15 A-13
A.4 IP classification
The IP Code, International Protection Marking, IEC standard 60529, sometimes interpreted as
Ingress Protection Marking, classifies and rates the degree of protection provided against intrusion
(body parts such as hands and fingers), dust, accidental contact, and water by mechanical casings
and electrical enclosures. It is published by the International Electro technical Commission (IEC).
The equivalent European standard is EN 60529. This table shows what each digit or part of the IP
code represents.
Level Effective
Description
sized against
Any large surface of the body, such as the back of a hand, but no
1 >50 mm
protection against deliberate contact with a body part
Protection
Level Effective against Details
against
0 None — —
Dripping water (vertically falling
drops) shall have no harmful Test duration: 10 minutes.
effect on the specimen when
1 Dripping water Water equivalent to 1 mm rainfall
mounted in an upright position
onto a turntable and rotated at 1 per minute.
RPM.
Vertically dripping water shall Test duration: 2.5 minutes for
have no harmful effect when the every direction of tilt (10 minutes
Dripping water enclosure is tilted at an angle of
2 when tilted at total).
15° from its normal position.
15° Water equivalent to 3 mm rainfall
A total of four positions are tested
within two axes. per minute.
Siaretron4000e - 15 A-15
Protection
Level Effective against Details
against
Test duration: 1 minute per
square meter for at least
Water projected by a nozzle 15 minutes
(6.3 mm) against enclosure from
5 Water jets Water volume: 12.5 litres per
any direction shall have no
harmful effects. minute
Pressure: 30 kPa at distance of
3m
Test duration: 1 minute per
square meter for at least
Water projected in powerful jets 3 minutes
Powerful water (12.5 mm nozzle) against the
6 Water volume: 100 litres per
jets enclosure from any direction shall
have no harmful effects. minute
Pressure: 100 kPa at distance of
3m
Water projected in powerful jets
(6.3 mm nozzle) against the Test duration: at least 3 minutes
Powerful water enclosure from any direction,
jets with Water volume: 75 litres per
6K under elevated pressure, shall
increased minute
have no harmful effects.
pressure Pressure: 1000 kPa at distance of
Found in DIN 40050, and not IEC 3m
60529.
Test duration: 30 minutes - ref
Ingress of water in harmful IEC 60529, table 8.
quantity shall not be possible
Immersion, up when the enclosure is immersed Tested with the lowest point of the
7 enclosure 1000 mm below the
to 1 m depth in water under defined conditions
of pressure and time (up to 1 m of surface of the water, or the
submersion). highest point 150 mm below the
surface, whichever is deeper.
The equipment is suitable for
continuous immersion in water
under conditions which shall be
specified by the manufacturer.
However, with certain types of
equipment, it can mean that water Test duration: Agreement with
Immersion, can enter but only in such a Manufacturer.
8 1 m or more manner that it produces no
depth harmful effects. Depth specified by manufacturer,
generally up to 3 m .
The test depth and duration is
expected to be greater than the
requirements for IPx7, and other
environmental effects may be
added, such as temperature
cycling before immersion.
(All tests with the letter "K" are defined by ISO 20653 (replacing DIN 40050-
9) and are not found in IEC 60529, except for IPx9 which is the same as the
IP69K water test.)
Letter Meaning
f Oil resistant
W Weather conditions
Siaretron4000e - 15 A-17
A.5 Pneumatic diagram
The mechanical study and diagram of the pneumatic system for gas delivery and dosage is
given in the attached illustrations.
The system is designed to safely administer the correct amount of respiratory gas to the
patient, and is completely controlled by the electronic system.
The device can be configured to use either a single or dual limb patient circuit (see the two
pneumatic diagrams attached).
PR AIR Air pressure regulator. It reduces the infeed gas pressure at a stable
value of 2.6 bar.
GAS SUPPLY Medical gas supply pressure sensor. It signals when the gas pressure
TRANSDUCER is missing or is insufficient.
AIR D.F.G. Air circuit digital flow generation block (120 l/min)
OXYGEN D.F.G. Oxygen circuit digital flow generation block (120 l/min)
Siaretron4000e - 15 A-19
V EXP D.C. 12 V pilot solenoid valve usually closed, controlled by ventilator
microprocessor. Once activated, each solenoid valve opens allowing
the distribution of a pre-set flow regulated by the relative flow
regulator.
In expiratory phase are activated only the solenoid valves that allow
providing the flow necessary to partially close the expiratory valve so
as to obtain the desired PEEP.
AIR SAFETY Environmental air infeed safety valve that allows ventilation even if the
VALVE. medical gas supplying the patient circuit is missing. It opens with
negative pressure of 2 mbar.
LIMIT VALVE The "pressure limit valve" is necessary to limit airway pressure
(80 cmH2O).
Alarm message A message which appears together with an alarm indication; this consists of a
basic message indicating the type of alarm.
Alarm silencing or Key that stops the acoustic alarm signal for a software value preset by the last
suspension key pressing of the key.
APCV Pressure controlled ventilation: type of controlled ventilation during which the
ventilator delivers an inspiratory pressure set by the user for an inspiratory
time also set by the user.
Apnoea End of ventilation. The ventilation system indicates apnea and starts the
corresponding ventilation when the interval between the two respiratory cycles
exceeds the set apnoea time.
Automatic alarm This occurs when an alarm is disabled, i.e. when the alarm conditions are no
resetting longer present, without pressing the alarm reset key. ALARM RESET
Basic flow Constant flow (depending on the sensitivity value set in the “trigger value”
parameter) circulating in the patient circuit with respect to which the ventilator
measures the Flow Trigger value.
Compliance (Cs) This term defines the variation in volume of the respiratory tract determined by
a variation in pressure; it is measured in ml/cmH2O. It provides an indication of
the elastic properties of the respiratory system and its components (Inspiratory
Tidal Volume / Pause Pressure).
Compressor The Compressor (optional) provides the system with compressed air and can
be used instead of the mains or cylinder supply of compressed air.
DISS Diameter Index Safety Standard: a standard for high pressure gas input
connectors.
Siaretron4000e - 15 A-21
EPU Electric power supply unit: the battery powers the system with direct current if
the alternate current supply is not available. On the basis of the ventilator
settings, the battery can provide back-up power for at least 3 hours in rated
and perfect working conditions.
EXP. PAUSE Expiratory pause, a manoeuvre started by the User which closes the
inspiratory and expiratory valves during the expiratory phase of a breath.
FiO2 Parameter set by the User and monitored. The % setting of FiO2 determines
the percentage of oxygen in the gas delivered to the patient. The monitored
data of the % of FiO2 indicate the percentage of oxygen delivered to the
patient, measured on the inspiratory line.
Flow Trigger Method of recognition of the inspiratory effort of the patient, during which the
ventilator controls the basic flow circulating in the patient circuit. An inspiratory
attempt by the patient is translated into a decrease of the basic flow, which the
ventilator recognizes as a spontaneous breath and delivers a synchronized
breath.
GUI Graphics user interface, the part of the ventilator which comprises the screen,
the keys and the knob. The GUI is equipped with an independent CPU which
monitors the data of the ventilator and the patient. The screen displays the
monitored information, including the alarms, the monitored parameters, the
graphs, the ventilator settings and the messages.
High priority alarm As defined by the international standards organizations, this is an alarm which
requires immediate intervention to ensure the safety of the patient. During a
high priority alarm, the corresponding red signal flashes rapidly, a high priority
acoustic alarm signal is emitted (a series of five tones repeated twice, followed
by a pause, then repeated again) and an alarm message is displayed in the
upper part of the screen.
I:E ratio The ratio between inspiratory time and expiratory time
INSP. PAUSE Inspiratory pause, a manoeuvre started by the User which closes the
inspiratory and expiratory valves during the inspiratory phase of a breath. This
manoeuvre can be used to determine the static compliance (C) and the
resistance (R).
IPPV - AST Assisted Intermittent Positive Pressure Ventilation: a ventilation mode that
makes it possible to deliver only controlled ventilation (started by the patient,
the ventilator or the User) on the basis of the current settings.
Low priority alarms As defined by the international standards organizations, this is an alarm that
indicates a change in the patient-ventilator system. During a low priority alarm,
the corresponding yellow signal lights up and an alarm message is displayed
in the upper part of the screen.
Maintenance All the operations necessary to maintain the equipment in working order or to
carry out cleaning, maintenance, repairs, modifications, revisions and
performance checks.
MAN If the MANUAL key is pressed in PSV mode, the system delivers pressure
controlled ventilation to the patient.
Minute volume Expired tidal volume normalized to the unit of time (L/min). The system
estimates the total minute volume on a 60 second basis or on previous
ventilations, whichever is the shorter. The value displayed includes the
compensation for compliance.
Mode Ventilation mode; an algorithm which determines the type and sequence of
ventilation: the system offers a series of possible choices, including assisted,
spontaneous or synchronized ventilation.
NIST Non-interchangeable screw thread: standard for high pressure gas inlet
connectors.
Patient circuit All the inspiratory-expiratory conduits, including the tubes, the humidifier and
the filters (when foreseen).
Siaretron4000e - 15 A-23
PCV Pressure controlled ventilation: a type of controlled ventilation during which
the ventilator delivers an inspiratory pressure set by the User for an inspiratory
time also set by the User.
PEEP Positive end expiratory pressure: the minimum level of pressure maintained in
the patient circuit during ventilation. Parameter set by the User and monitored.
Pressure Trigger Method of recognition of the inspiratory effort of the patient, in which the
ventilator controls the pressure in the patient circuit. The ventilator enables
ventilation when the airways pressure decreases by an amount at least equal
to the selected threshold value in a defined period of time.
Resistance (Ri) The drop in pressure caused by a flow passing through a conduit: measured
in cmH2O/(litres/sec) or hPa/(litres/sec).(peak pressure - pause pressure /
inspiratory flow).
SPONT In SPONT mode, the patient activates all ventilations delivered by ventilator
without any controlled respiratory rate set. The patient makes spontaneous
breaths by pressure support.
System error Definition used by the safety system of the ventilator. System errors include
faults of the hardware inside the ventilator and which affect its performance,
software errors which occur momentarily inside the ventilator and interfere
with its normal functioning, an inadequate supply of alternate current or gas
and the problems of integrity of the patient circuit (block or disconnection). In
general system errors are not corrected automatically
T pause Pause time: percentage of inspiratory time during which the ventilator
maintains a constant airways pressure. Used for calculation of the respiratory
mechanics parameters (compliance and resistance).
Tidal volume Inspired and expired tidal volume during each breath. The value delivered by
the system is a parameter set by the User which determines the volume
delivered to the patient during controlled volume ventilation. Tidal volume
includes the compensation for compliance and for pressure and body
temperature.
Siaretron4000e - 15 A-25
A.7 Electromagnetic compatibility tables
The Siaretron 4000 is intended for use in the electromagnetic environment specified below. The customer
or the user of the Siaretron 4000 should assure that it is used in such an environment.
The Siaretron 4000 is intended for use in the electromagnetic environment specified below. The customer
or the user of the Siaretron 4000 should assure that it is used in such an environment.
Electromagnetic
IMMUNITY test IEC 60601 test level Compliance level / Virdict environment –
guidance
Power frequency
Residential –
(50/60 Hz) magnetic 3 A/m 3 A/m Hospital – Other
field IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Siaretron4000e - 15 A-27
A.7.3 Annex C: Table 3
The Siaretron 4000 is intended for use in the electromagnetic environment specified below. The customer
or the user of the Siaretron 4000 should assure that it is used in such an environment.
Electromagnetic environment –
IEC 60601 TEST Compliance Compliance guidance
IMMUNITY test
LEVEL level effective level Recommended separation
distances
80 800 MHz
SEE ANNEX E
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Note:
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
The Siaretron 4000 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Siaretron 4000 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Siaretron 4000 as recommended below, according to
the maximum output power of the communications equipment.
12 12 23
W 3,5 d P d P
d P d P
V1 E1 E1
V2
0,01 0,12 0,12 0,12 0,23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note :
1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to
6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70
MHz.
3. An additional factor of 10/3 has been incorporated into the formulae used in calculating the
recommended separation distance for transmitters in the ISM frequency bands between 150 kHz
and 80 MHz and in the frequency range 80 MHz to 2,5 GHz to decrease the likelihood that
mobile/portable communications equipment could cause interference if it is inadvertently brought
into patient areas.
Note: the values shown in the table refer to the standard levels of the norm, 3V for V1 and 10V for V2
Siaretron4000e - 15 A-29
This page has been added to make front / back copy easier.