M109R User Manual en
M109R User Manual en
M109R User Manual en
Operation Manual
M-109R High Resistance Decade MEATEST
CONTENT:
Date 4/2011
2. Contents of delivery
3. Technical data
Resistance range : 1 M - 12.221 G
Maximum voltage : 5000 V DC – terminals H-L, H-, L-
Connection : two-terminal + (GUARD)
Type of terminals : ERTALYTE terminals
Capacitance between terminals. : H – GUARD, L – GUARD 150 pF
Remote control : RS-232 interface
Power supply : internal battery 12 V
line adapter 100–240VAC/50-60Hz
Operating period : typically 3 hours from internal battery
Reference temperatures : 23 C 5 C
Range of working temperatures : +5 C to +40 C
Range of storing temperatures : -10 C to +55 C
Relative humidity : 10 - 50 %
Instrument case : metal
Dimensions : W 362 mm, H 111 mm, D 316 mm
Weight : 4 kg
Nominal value Voltage coefficient Temperature Maximum
Range accuracy Coefficient voltage
+5 C to +40 C
% ppm/V ppm / C V DC/ RMS
1 M - 11 M 0.1/100 V 1 100 1000/700
12 M - 121M 0.2/1000 V 1 100 2500/1700
122 M - 1.221 G 0.5/1000 V 2 100 5000/3500
1.222 G - 12.221 G 1.0/1000 V 2 100 5000/3500
operation manual 2
M-109R High Resistance Decade MEATEST
4. Operation
On the front panel, there are located all main operating and indicating elements
and input terminals. Individual rotary buttons allow you to set a required electrical
resistance value within the 1 M through 12.221 G. Red luminous LED diodes
designated 5kV, 2,5kV and 1kV signalize maximum input direct voltage, which can
be connected to the input terminals in the given setup configuration. A green LED
diode designated REM indicates a mode of remote control. A green LED diode
designated ON indicates switching the decade on. Successive dashed glare indicates
switching the decade off.
RS-232 BUS
3 operation manual
MEATEST M-109R High Resistence Decade
On the rear panel is the ON/OFF button for switching the decade on and off,
supplying connector and a connector for connecting RS-232.
x 10M x 1M
1 0 11 1 0 11 1 0 11 1 0 11
H 5kV x 1G x 100M
10
10 8
10
10
4 3 2
4 3 2
4 3 2
4 3 2
2.5kV
98
98
98
L 1kV
7 65 7 65 7 65 7 65
REM
ON
When supplying from a power adapter, the decade is not switched off
automatically and the internal non-maintenance battery is recharged simultaneously. A
green LED diode permanently indicates the decade switching on. The decade is
switched off automatically when you disconnect the power adapter, according to the
procedure described in chapter 4.3.
operation manual 4
M-109R High Resistance Decade MEATEST
reasons, always connect the terminal to the grounding terminal of the supplying
system. Connections to the decade terminals must always be both mechanically as well
as electrically reliable in order that the electric circuit cannot be interrupted or
completely mechanically disconnected during the measuring procedure, and thus it
could create a possibility of electric injury.
It is always necessary to accept the maximum allowed input voltage, which is in
the given configuration indicated by red LED diodes. Setting of values is performed
with a delay of approximately 1s.
Recommended connection
Two-terminals measuring
Note
If it is not excluded due to measuring method of tested isolation meter, it is
recommended to connect ground terminal on resistance decade housing with ground
terminal of power network or any other grounding point,
Three-terminal measuring
WARNING:
When the decade is switched off, input terminals H and L are in state open.
When manipulating with resistance decade, realize that you are operating with
device connected to dangerous high voltage.
5 operation manual
MEATEST M-109R High Resistence Decade
If you are not absolutely sure about the safety of the connection, do
not touch operating elements or the instrument shield.
operation manual 6
M-109R High Resistance Decade MEATEST
4.7. Calibration
Ref. temperature 23 C
If any of the checking points does not match the specification, contact the
manufacturer.
7 operation manual
MEATEST M-109R High Resistence Decade
Required equippment
Procedure
operation manual 8
M-109R High Resistance Decade MEATEST
For data flow operation neither hardware handshake (RTS/CTS) nor program
handshake (XON/XOFF) is used, however the presence of operating signals is
necessary for correct voltage levels on the TXD data signal leading from the decade.
Communication between the computer (PC) and the instrument runs in a periodic
exchange of the command-response type. A command is always an upper case letter
followed by an alternative parameter without a space and ended by the CR symbol,
again without a space. The instrument response is always ended by the operating
symbol CR ( \r ). The serial interface of the instrument requires signals RXD, TXD,
RTS (static on level -3 through -12V), DTR (static on level +3 through +12V) and
GND.
Serial line cable connects pins of relevant numbers of both connectors (connection
1:1).
Example of initialization of a serial line COM1 and setup of operating signals for
operating the M-109 decade in the BASIC language is shown:
9 operation manual
MEATEST M-109R High Resistence Decade
List of commands
The response of M-109 when received an unknown command is the string "?\r“. A
correctly executed command is confirmed by "ok\r. Commands must be ended by the
CR symbol (\r,\0x0d). The program is case sensitive.
Demo program
A simple operating program is supplied with the decade in order to provide easy
operation of the decade from the computer, and to check the RS-232 line of the
instrument. The installation CD ROM contains a program (for
WIN95/98/NT/ME/2000/XP/Vista only), you can communicate with the instrument
through standard serial line with. For example, you can read a value pre-defined on
switches from the decade and, conversely, from the computer you can set any value
(within range of the instrument).
When using remote control, the value set on individual switches does not
correspond to the actual decade resistance value – the resistance value is
corresponding to the value setup through the RS-232.
operation manual 10
M-109R High Resistance Decade MEATEST
Insert the software CD into your CD ROM drive. Wait for autorun and select
“Software installation” and “M109” or start direct
„install\freeware\M109\en\setup.exe“. When you launch SETUP.EXE, the installation
program asks for the destination directory and executes the actual installation. The
UNINST.EXE program is also copied into the selected directory for alternative delete
of the directory from the system.
Program description
When you launch M190 program, following control panel is displayed on the
screen.
First use the operating element COM to select a serial port number to which the
instrument is connected. Check if slide button REMOTE is in position OFF and
confirm this selection using the START button. If the instrument is found on the serial
line, actual instrument setup is read periodically and the setup of rotary buttons on the
screen correspond to the setup on the instrument.
11 operation manual
MEATEST M-109R High Resistence Decade
The light of indicating elements on the left-hand part of the screen is corresponding to
LED diodes on the instrument panel. Rotary buttons on the control panel are displayed
like “enable”. They are only indicators of the real switches setup position on the front
panel of the instrument now.
If the instrument is not switched on or connection is incorrect or the instrument is
connected on a different port, the program is terminated after sending error message.
When you switch the operating element REMOTE to the ON position, the signal
diodes REM on both the instrument as well as on the control panel are switched on,
and the setup resistance value can be set from the control panel on the computer
screen. It is possible in several ways:
When supplying from a battery, you can switch off the instrument from the control
panel by pressing the OFF button. This will break communication between the
computer and the instrument.
operation manual 12
M-109R High Resistance Decade MEATEST
You can end the program at any time by pressing the QUIT button.
4.10. Maintenance
Although the decade contains mechanical switches, it does not require any special
cleaning. Keep output terminals and ERTALYTE grommets absolutely clean.
Recharge the internal battery if necessary. Frequency of recharging depends on the
proportion of operating the decade in the mode when it is supplied from the internal
battery or from a power adapter. If the battery is discharged and the decade was
switched off automatically, it is recommended to recharge battery for approx. 10
hours. If the battery is recharged properly, you can operate the decade without a power
adapter for 3-8 hours depending on the set resistance value and the number of
switching on high-voltage relays. The current consumption of the decade is lowest
when the resistance value is set to 0 MOhm.
6. Mechanical construction
High resistance decade M - 109R is placed in a standardized metal case made from
aluminum alloys. Rotary buttons and input terminals are placed on the front panel. The
end position of switches is locked. PCB with high-voltage relays and resistors is
placed in a separate build-in metal case, which is connected to the terminal. The
CPU board is mechanically connected to the relay board, and thus they form a joint
assembly unit. On the rear panel you is situated ON/OFF button, supplying connector
for power adapter connecting and RS 232 connector. Internal battery is also
mechanically attached to this panel.
13 operation manual
MEATEST M-109R High Resistence Decade
Manufacturer
operation manual 14