Moleculer and Atomic Spectros
Moleculer and Atomic Spectros
Moleculer and Atomic Spectros
SPECTROSCOPY
Thomas Wenzel
Bates College
TABLE OF CONTENTS
3: MOLECULAR LUMINESCENCE
3.1: Introduction
3.2: Energy States and Transitions
3.3: Instrumentation
3.4: Excitation and Emission Spectra
3.5: Quantum Yield of Fluorescence (φ F )
\ce
4: INFRARED SPECTROSCOPY
4.1: Introduction to Infrared Spectroscopy
4.2: Specialized Infrared Methods
4.3: Fourier-Transform Infrared Spectroscopy (FT-IR)
5: RAMAN SPECTROSCOPY
6: ATOMIC SPECTROSCOPY
6.1: INTRODUCTION TO ATOMIC SPECTROSCOPY
6.2: ATOMIZATION SOURCES
6.2A: Flames
6.2B: Electrothermal Atomization – Graphite Furnace
6.2C: Specialized Atomization Methods
6.2D: Inductively Coupled Plasma
6.2E: Arcs and Sparks
E = hν (1.1.1)
What is the relationship between the energy and wavelength (λ) of electromagnetic radiation?
Using the knowledge that the speed of electromagnetic radiation (c) is the frequency times the wavelength (c = λν ), we can solve for
the frequency and substitute in to the expression above to get the following.
hc
E = (1.1.2)
λ
Therefore the energy of electromagnetic radiation is inversely proportional to the wavelength. Long wavelength electromagnetic
radiation will have low energy. Short wavelength electromagnetic radiation will have high energy.
Write the types of radiation observed in the electromagnetic spectrum going from high to low energy. Also include what types
of processes occur in atoms or molecules for each type of radiation.
High E, high ν , short λ: γ -rays – Nuclear energy transitions
X-rays – Inner-shell electron transitions
Ultraviolet – Valence electron transitions
Visible – Valence electron transitions
Infrared – Molecular vibrations
Microwaves – Molecular rotations, Electron spin transitions
Low E, low ν , long λ: Radiofrequency – Nuclear spin transitions
Atoms and molecules have the ability to absorb or emit electromagnetic radiation. A species absorbing radiation undergoes a transition
from the ground to some higher energy excited state. A species emitting radiation undergoes a transition from a higher energy excited
state to a lower energy state. Spectroscopy in analytical chemistry is used in two primary manners: (1) to identify a species and (2) to
quantify a species.
Identification of a species involves recording the absorption or emission of a species as a function of the frequency or wavelength to
obtain a spectrum (the spectrum is a plot of the absorbance or emission intensity as a function of wavelength). The features in the
spectrum provide a signature for a molecule that may be used for purposes of identification. The more unique the spectrum for a
species, the more useful it is for compound identification. Some spectroscopic methods (e.g., NMR spectroscopy) are especially useful
for compound identification, whereas others provide spectra that are all rather similar and therefore not as useful. Among methods that
provide highly unique spectra, there are some that are readily open to interpretation and structure assignment (e.g., NMR spectra),
whereas others (e.g., infrared spectroscopy) are less open to interpretation and structure assignment. Since molecules do exhibit unique
infrared spectra, an alternative means of compound identification is to use a computer to compare the spectrum of the unknown
compound to a library of spectra of known compounds and identify the best match. In this case, identification is only possible if the
spectrum of the unknown compound is in the library.
Quantification of a species using a spectroscopic method involves measuring the magnitude of the absorbance or intensity of the
emission and relating that to the concentration. At this point, we will focus on the use of absorbance measurements for quantification.
Consider a sample through which you will send radiation of a particular wavelength as shown in Figure 1.1. You measure the power
from the radiation source (Po) using a blank solution (a blank is a sample that does not have any of the absorbing species you wish to
measure). You then measure the power of radiation that makes it through the sample (P).
The higher the transmittance, the more similar P is to Po. The absorbance (A) is defined as:
Po
A = − log T or log( ). (1.1.4)
P
The higher the absorbance, the lower the value of P, and the less light that makes it through the sample and to the detector.
Note that Beer’s Law is the equation for a straight line with a y-intercept of zero.
If you wanted to measure the concentration of a particular species in a sample, describe the procedure you would use to do so.
Measuring the concentration of a species in a sample involves a multistep process.
One important consideration is the wavelength of radiation to use for the measurement. Remember that the higher the molar
absorptivity, the higher the absorbance. What this also means is that the higher the molar absorptivity, the lower the concentration of
species that still gives a measurable absorbance value. Therefore, the wavelength that has the highest molar absorptivity (λmax) is
usually selected for the analysis because it will provide the lowest detection limits. If the species you are measuring is one that has
been commonly studied, literature reports or standard analysis methods will provide the λmax value. If it is a new species with an
unknown λmax value, then it is easily measured by recording the spectrum of the species. The wavelength that has the highest
absorbance in the spectrum is λmax.
The second step of the process is to generate a standard curve. The standard curve is generated by preparing a series of solutions
(usually 3-5) with known concentrations of the species being measured. Every standard curve is generated using a blank. The blank is
some appropriate solution that is assumed to have an absorbance value of zero. It is used to zero the spectrophotometer before
measuring the absorbance of the standard and unknown solutions. The absorbance of each standard sample at λmax is measured and
plotted as a function of concentration. The plot of the data should be linear and should go through the origin as shown in the standard
curve in Figure 1.2. If the plot is not linear or if the y-intercept deviates substantially from the origin, it indicates that the standards
were improperly prepared, the samples deviate in some way from Beer’s Law, or that there is an unknown interference in the sample
that is complicating the measurements. Assuming a linear standard curve is obtained, the equation that provides the best linear fit to the
data is generated.
Since stray radiation always leaks in to the detector and presumably is a fixed or constant quantity, we can rewrite the expression for
the absorbance including terms for the stray radiation. It is important to recognize that Po, the power from the radiation source, is
considerably larger than PS. Also, the numerator (Po + Ps) is a constant at a particular wavelength.
Po + Ps
A = log( ) (1.2.3)
P + Ps
Now let’s examine what happens to this expression under the two extremes of low concentration and high concentration. At low
concentration, not much of the radiation is absorbed and P is not that much different than Po. Since Po >> PS, P will also be much
greater than PS. If the sample is now made a little more concentrated so that a little more of the radiation is absorbed, P is still much
greater than PS. Under these conditions the amount of stray radiation is a negligible contribution to the measurements of Po and P and
has a negligible effect on the linearity of Beer’s Law.
As the concentration is raised, P, the radiation reaching the detector, becomes smaller. If the concentration is made high enough, much
of the incident radiation is absorbed by the sample and P becomes much smaller. If we consider the denominator (P + PS) at increasing
concentrations, P gets small and PS remains constant. At its limit, the denominator approaches PS, a constant. Since Po + PS is a
constant and the denominator approaches a constant (Ps), the absorbance approaches a constant. A plot of what would occur is shown
in Figure 1.3.
Figure 1.3. Plot of ideal (linear) and actual (curved) measurements when substantial amounts of stray radiation are
present.
The ideal plot is the straight line. The curvature that occurs at higher concentrations that is caused by the presence of stray radiation
represents a negative deviation from Beer’s Law.
The derivation of Beer's Law assumes that the molecules absorbing radiation don't interact with each other (remember that
these molecules are dissolved in a solvent). If the analyte molecules interact with each other, they can alter their ability to
absorb the radiation. Where would this assumption break down? Guess what this does to Beer's law?
Figure 1.4. Utilization of a prism and slit to select out specific wavelengths of radiation.
An examination of Figure 1.4 shows that the slit has to allow some “packet” of wavelengths through to the sample. The packet is
centered on λmax, but clearly nearby wavelengths of radiation pass through the slit to the sample. The term effective bandwidth
defines the packet of wavelengths and it depends on the slit width and the ability of the dispersing element to divide the wavelengths.
Reducing the width of the slit reduces the packet of wavelengths that make it through to the sample, meaning that smaller slit widths
lead to more monochromatic radiation and less deviation from linearity from Beer’s Law.
Is there a disadvantage to reducing the slit width?
The important thing to consider is the effect that this has on the power of radiation making it through to the sample (Po). Reducing the
slit width will lead to a reduction in Po and hence P. An electronic measuring device called a detector is used to monitor the magnitude
of Po and P. All electronic devices have a background noise associated with them (rather analogous to the static noise you may hear on
a speaker and to the discussion of stray radiation from earlier that represents a form of noise). Po and P represent measurements of
signal over the background noise. As Po and P become smaller, the background noise becomes a more significant contribution to the
overall measurement. Ultimately the background noise restricts the signal that can be measured and detection limit of the
spectrophotometer. Therefore, it is desirable to have a large value of Po. Since reducing the slit width reduces the value of Po, it also
reduces the detection limit of the device. Selecting the appropriate slit width for a spectrophotometer is therefore a balance or tradeoff
of the desire for high source power and the desire for high monochromaticity of the radiation.
It is not possible to get purely monochromatic radiation using a dispersing element with a slit. Usually the sample has a slightly
different molar absorptivity for each wavelength of radiation shining on it. The net effect is that the total absorbance added over all the
different wavelengths is no longer linear with concentration. Instead a negative deviation occurs at higher concentrations due to the
polychromicity of the radiation. Furthermore, the deviation is more pronounced the greater the difference in the molar absorbtivity.
Figure 1.5 compares the deviation for two wavelengths of radiation with molar absorptivities that are (a) both 1,000, (b) 500 and 1,500,
and (c) 250 and 1,750. As the molar absorptivities become further apart, a greater negative deviation is observed.
Figure 1.5. Deviation from linearity of Beer’s law for two wavelengths where the molar absorptivities are (a) both 1,000,
(b) 500 and 1,500, and (c) 250 and 1,750.
Figure 1.6. Hypothetical spectrum with a sharp and broad absorption peak.
Consider the relative error that would be observed for a sample as a function of the transmittance or absorbance. Is there a
preferable region in which to measure the absorbance? What do you think about measuring absorbance values above 1?
It is important to consider the error that occurs at the two extremes (high concentration and low concentration). Our discussion above
about deviations to Beer’s Law showed that several problems ensued at higher concentrations of the sample. Also, the point where only
10% of the radiation is transmitted through the sample corresponds to an absorbance value of 1. Because of the logarithmic
relationship between absorbance and transmittance, the absorbance values rise rather rapidly over the last 10% of the radiation that is
absorbed by the sample. A relatively small change in the transmittance can lead to a rather large change in the absorbance at high
concentrations. Because of the substantial negative deviation to Beer’s law and the lack of precision in measuring absorbance values
above 1, it is reasonable to assume that the error in the measurement of absorbance would be high at high concentrations.
At very low sample concentrations, we observe that Po and P are quite similar in magnitude. If we lower the concentration a bit more,
P becomes even more similar to Po. The important realization is that, at low concentrations, we are measuring a small difference
between two large numbers. For example, suppose we wanted to measure the weight of a captain of an oil tanker. One way to do this is
to measure the combined weight of the tanker and the captain, then have the captain leave the ship and measure the weight again. The
difference between these two large numbers would be the weight of the captain. If we had a scale that was accurate to many, many
significant figures, then we could possibly perform the measurement in this way. But you likely realize that this is an impractical way
to accurately measure the weight of the captain and most scales do not have sufficient precision for an accurate measurement.
Similarly, trying to measure a small difference between two large signals of radiation is prone to error since the difference in the
signals might be on the order of the inherent noise in the measurement. Therefore, the degree of error is expected to be high at low
concentrations.
The discussion above suggests that it is best to measure the absorbance somewhere in the range of 0.1 to 0.8. Solutions of higher and
lower concentrations have higher relative error in the measurement. Low absorbance values (high transmittance) correspond to dilute
solutions. Often, other than taking steps to concentrate the sample, we are forced to measure samples that have low concentrations and
must accept the increased error in the measurement. It is generally undesirable to record absorbance measurements above 1 for
samples. Instead, it is better to dilute such samples and record a value that will be more precise with less relative error.
Another question that arises is whether it is acceptable to use a non-linear standard curve. As we observed earlier, standard curves of
absorbance versus concentration will show a non-linearity at higher concentrations. Such a non-linear plot can usually be fit using a
higher order equation and the equation may predict the shape of the curve quite accurately. Whether or not it is acceptable to use the
non-linear portion of the curve depends in part on the absorbance value where the non-linearity starts to appear. If the non-linearity
occurs at absorbance values higher than one, it is usually better to dilute the sample into the linear portion of the curve because the
absorbance value has a high relative error. If the non-linearity occurs at absorbance values lower than one, using a non-linear higher
order equation to calculate the concentration of the analyte in the unknown may be acceptable.
One thing that should never be done is to extrapolate a standard curve to higher concentrations. Since non-linearity will occur at some
point, and there is no way of knowing in advance when it will occur, the absorbance of any unknown sample must be lower than the
absorbance of the highest concentration standard used in the preparation of the standard curve. It is also not desirable to extrapolate a
standard curve to lower concentrations. There are occasions when non-linear effects occur at low concentrations. If an unknown has an
absorbance that is below that of the lowest concentration standard of the standard curve, it is preferable to prepare a lower
concentration standard to ensure that the curve is linear over such a concentration region.
Topic hierarchy
1.3C: Detectors
Figure 1.8. Representation of (a) emission of a photon, (b) stimulated emission and (c) absorption. The waves represent
photons.
Why is it impossible to create a 2-level laser?
A 2-level laser involves a process with only two energy states, the ground and excited state. In a resting state, the system will have a
large population of species in the ground state (essentially 100% as seen in Figure 1.9) and only a few or none in the excited state.
Incident radiation of an energy that matches the transition is then applied and ground state species absorb photons and become excited.
The general transition process is illustrated in Figure 1.9a.
Figure 1.9. Representation of (a) absorption of a photon by a ground state molecule where all species are in the ground
state and (b) a 2-level energy system where the population in the ground and excited states are equal.
Species in the excited state will give up the excess energy either as an emitted photon or heat to the surroundings. We will discuss this
in more detail later on, but for now, it is acceptable to realize that excited state species have a finite lifetime before they lose their
energy and return to the ground state. Without worrying about the excited state lifetime, let’s assume that the excited species remain in
that state and incident photons can continue to excite additional ground state species into the excited state. As this occurs, the number
of species in the excited state (e.g., the excited state population) will grow and the number in the ground state will diminish. The key
point to consider is the system where 50% of the species are in the excited state and 50% of the species are in the ground state, as
shown in Figure 1.9b.
For a system with exactly equal populations of the ground and excited state, incident photons from the radiation source have an equal
probability of interacting with a species in the ground or excited state.
If a photon interacts with a species in the ground state, absorption of the photon occurs and the species becomes excited. However, if
another photon interacts with a species in the excited state, stimulated emission occurs, the species returns to the ground state and two
photons are emitted. The net result is that for every ground state species that absorbs a photon and becomes excited there is a
corresponding excited species that undergoes stimulated emission and returns to the ground state. Therefore it is not possible to get
beyond the point of a 50-50 population and never possible to get a population inversion. A 2-level system with a 50-50 population is
said to be a saturated transition.
Using your understanding of a 2-level system, explain what is meant by a 3-level and 4-level system. 3- and 4-level systems can
function as a laser. How is it possible to achieve a population inversion in a 3- and 4-level system?
The diagrams for a 3-level and 4-level laser system are shown in Figures 1.10 and 1.11, respectively.
Figure 2.4. Energy level diagram for a molecule showing electronic, vibrational and rotational states. Arrows represent
possible absorption transitions.
Note that the vibrational and rotational energy levels in a molecule are superimposed over the electronic transitions. An important
question to consider is whether an electron in the ground state (lowest energy electronic, vibrational and rotational state) can only be
excited to the first excited electronic state (no extra vibrational or rotational energy), or whether it can also be excited to vibrationally
and/or rotationally excited states in the first excited electronic state. It turns out that molecules can be excited to vibrationally and/or
rotationally excited levels of the first excited electronic state, as shown by arrows in Figure 2.4. Molecules can also be excited to the
second and higher excited electronic states. Therefore, we can speak of a molecule as existing in the second excited rotational state of
the third excited vibrational state of the first excited electronic state.
One consequence in the comparison of atomic and molecule absorption spectra is that molecular absorption spectra ought to have
many more transitions or lines in them than atomic spectra because of all the vibrational and rotational excited states that exist.
Compare a molecular absorption spectrum of a dilute species dissolved in a solvent at room temperature versus the same
sample at 10K.
The difference to consider here is that the sample at 10K will be frozen into a solid whereas the sample at room temperature will be a
liquid. In the liquid state, the solute and solvent molecules move about via diffusion and undergo frequent collisions with each other. In
the solid state, collisions are reduced considerably.
What is the effect of collisions of solvent and solute molecules? Collisions between molecules cause distortions of the electrons. Since
molecules in a mixture move with a distribution of different speeds, the collisions occur with different degrees of distortion of the
electrons. Since the energy of electrons depends on their locations in space, distortion of the electrons causes slight changes in the
energy of the electrons. Slight changes in the energy of an electron means there will be a slight change in the energy of its transition to
a higher energy state. The net effect of collisions is to cause a broadening of the lines in the spectrum. The spectrum at room
temperature will show significant collisional broadening whereas the spectrum at 10K will have minimal collisional broadening. The
collisional broadening at room temperature in a solvent such as water is significant enough to cause a blurring together of the energy
differences between the different rotational and vibrational states, such that the spectrum consists of broad absorption bands instead of
discrete lines. By contrast, the spectrum at 10K will consist of numerous discrete lines that distinguish between the different
rotationally and vibrationally excited levels of the excited electronic states. The diagrams in Figure 2.5 show the difference between the
Figure 2.5. Comparison of the absorption spectrum of a molecule in a solvent at room temperature and at 10 K.
Are there any other general processes that contribute to broadening in an absorption spectrum?
The other general contribution to broadening comes from something known as the Doppler Effect. The Doppler Effect occurs because
the species absorbing or emitting radiation is moving relative to the detector. Perhaps the easiest way to think about this is to consider a
species moving away from the detector that emits a specific frequency of radiation heading toward the detector. The frequency of
radiation corresponds to that of the energy of the transition, so the emitted radiation has a specific, fixed frequency. The picture in
Figure 2.6 shows two species emitting waves of radiation toward a detector. It is worth focusing on the highest amplitude portion of
each wave. Also, in Figure 2.6, assume that the detector is on the right side of the diagram and the right side of the two emitting
spheres. The emission process to produce the wave of radiation requires some finite amount of time. If the species is moving away
from the detector, even though the frequency is fixed, to the detector it will appear as if each of the highest amplitude regions of the
wave is lagging behind where they would be if the species is stationary (see the upper sphere in Figure 2.6). The result is that the
wavelength of the radiation appears longer, meaning that the frequency appears lower. For visible radiation, we say that the radiation
from the emitting species is red-shifted. The lower sphere in Figure 2.6 is moving towards the detector. Now the highest amplitude
regions of the wave are appearing at the detector faster than expected. This radiation is blue-shifted. In a solution, different species are
moving in different directions relative to the detector. Some exhibit no Doppler shift. Others would be blue-shifted whereas others
would be red-shifted and the degree of red- and blue-shift varies among different species. The net effect would be that the emission
peak is broadened. The same process occurs with the absorption of radiation as well.
Figure 2.6. Representation of the Doppler effect on the wavelength of radiation measured by a detector.
The emission spectrum in Figure 2.7 represents the Doppler broadening that would occur for a gas phase atomic species where the
atoms are not moving (top) and then moving with random motion (bottom).
A comparison of the two spectra involves a consideration of the molecular orbitals involved in the compounds and energy transitions.
Organic compounds have a network of σ-bonds and then a set of associated σ* anti-bonding orbitals. 1-Butene has one double bond so
therefore has a single π and π* orbital. The representations in Figure 2.8 show the sign of the p-orbital wave functions (dark is
positive, white is negative) used in creating the π and π* orbitals. In the π-orbital, the two wave functions have the same sign and
therefore positive overlap. In the π*-orbital, the two wave functions have the opposite sign and therefore negative overlap.
Figure 2.9. Energy level diagram for a compound with a π- and π*-orbital.
The most important energy transition to consider in Figure 2.9 is the one from the highest occupied molecular orbital (HOMO) to the
lowest unoccupied molecular orbital (LUMO). This will be the lowest energy transition. In the case of 1-butene, the lowest energy
transition would be the π-π* transition. The UV/VIS absorption spectrum for 1-butene is shown in Figure 2.10. The λmax value has a
value of about 176 nm, which is in the vacuum ultraviolet portion of the spectrum.
Figure 2.13. Possible alignments of the signs of the wave functions of the p-orbitals in 1,3-butadiene.
Rank these from high to low energy.
The orbital in which all four p-orbitals overlap would be the lowest in energy (Figure 2.14). The next has two regions of overlap. The
third has only one region of overlap and the highest energy orbital has no regions of overlap. Because there are four electrons to put in
Figure 2.14. Energy level diagram and representation of the π- and π*-molecular orbitals in 1,3-butadiene.
The lowest energy HOMO to LUMO transition will be lower than observed in 1-butene. The UV/VIS spectrum of 1,3-butadiene is
shown in Figure 2.15. In this case, the λmax value is at about 292 nm, a significant difference from the value of 176 nm in 1-butene.
The effect of increasing conjugation is to shift the spectrum toward longer wavelength (lower frequency, lower energy) absorptions.
The spectra in Figure 2.16 are for benzene, naphthalene, anthracene and pentacene. Note that as more rings and more conjugation are
added, the spectrum shifts further toward and into the visible region of the spectrum.
An important distinction with pyridine involves the extra lone pair of electrons on the nitrogen atom that are non-bonding electrons. As
non-bonding electrons, their energy is not all that different from the value that would occur in the nitrogen atom. As such, the non-
bonding electrons usually reside in orbitals that fall in between the π and π* orbitals. The generalized diagram in Figure 2.19 shows
filled bonding molecular orbitals (BMOs), empty anti-bonding molecular orbitals (ABMOs) and the location of non-bonding electrons.
Figure 2.19. Representation of the relative energies of bonding molecular orbitals (BMOs), anti-bonding molecular
orbitals (ABMOs) and non-bonding electrons.
For pyridine, the lowest energy transition involves the n-π* orbitals and this will be much lower in energy than the π-π* transition in
pyridine or benzene. The UV/VIS absorption spectrum of pyridine is shown in Figure 2.20.
absorption is achieved in dye molecules by having a combination of conjugation and non-bonding electrons. Azo dyes with the N=N
group are quite common, one example of which is shown in Figure 2.21.
drops in energy and the other of which rises in energy. The electrostatic attraction between a positively charged hydrogen atom and
negatively charged lone pair of electrons in a hydrogen-bond (as illustrated in Figure 2.22 for methanol and pyridine) is a stabilizing
interaction. Therefore, the energy of the non-bonding electrons will be lowered.
The picture in Figure 2.23 shows representations of a π- and π*-orbital.
methanol than the π-orbital. The diagram in Figure 2.24 shows the relative changes in the energies of the n, π, and π* orbitals that
would occur on changing the solvent from hexane to methanol with stabilization occurring in the order n > π* > π.
Figure 2.24. Relative changes in energies of the n, π, and π* orbitals that occurs on changing the solvent from hexane to
methanol.
An examination of the relative energies between hexane and methanol shows that both the n and π* levels drop in energy, but the drop
of the n level is greater than the drop of the π* level. Therefore, the n-π* transition moves to higher energy, hence a blue-shift is
observed in the peaks in the spectrum in the 320-380 nm range of pyridine. The blue-shift that is observed is referred to as a
hypsochromic shift.
Provided the UV/VIS absorption spectra of HA and A– differ from each other, describe a method that you could use to measure
the pKa of the acid.
Figure 3.1. Energy level diagram for π- and π*-orbitals of a typical organic compound.
Now consider the electron spin possibilities for the ground and excited state. Are there different possible ways to orient the
spins (if so, these represent different spin states).
The ground state, which is shown on the left in Figure 3.1, has two electrons in the π-orbital. These two electrons must have opposite
spins or else they would have the same four quantum numbers. Therefore, there is only one possible way to align the spins of the two
electrons in the π-orbital.
The excited state has one electron in the π-orbital and one electron in the π*-orbital as shown in Figure 3.1. In this case, there are two
possible ways we might align the spins. In one case, the electron in the π*-orbital could have the opposite spin of the electron in the π-
orbital (e.g., the electrons have paired spins, even though they are in different orbitals – see Figure 3.2, middle diagram). In the other
case, the electron in the π*-orbital could have a spin that is parallel with the electron in the π-orbital (see Figure 3.2 – far right
diagram). In both cases, it does not matter which electron has spin-up and which has spin-down, the only important point is that in one
case the two spins are opposite and in the other they are parallel. The energy level diagram in Figure 3.2 shows representations for the
two possibilities.
Figure 3.2. Representations of possible spin states for ground and excited state systems.
Do you think these different spin states have different energies?
Since they are different from each other (i.e., spins parallel versus spins paired), it makes sense that they would have different energies.
Which one do you expect to be lower in energy?
To answer this question, we have to think back to a rule we established with placing electrons into atomic or molecular orbitals that
have the same energy (i.e., are degenerate). We learned that electrons go into degenerate orbitals with parallel spins and only pair up
their spins when forced to do so (e.g., an atomic p3 configuration has three unpaired electrons with parallel spins; only when we added
a fourth electron to make a p4 configuration do two of the electrons have paired spins). The rationale we gave for this observation is
that configurations with parallel spins in degenerate orbitals are lower in energy than configurations with paired spins (i.e., it took
energy to pair up electron spins). Applying this general concept to the situation above, we can reason that the configuration in which
the electrons in the π- and π*-orbitals have parallel spins is lower in energy than the configuration in which the two electrons have
paired spins. The energy level diagrams in Figure 3.3 show the lower energy of the configuration where the electrons have parallel
spins.
′
S0 + hν = S2 (3.2.2)
It is reasonable at first to think that there is an absorption transition that goes directly from the S0 to the T1 state. This is a transition
that involves a spin-flip and it turns out that transitions that involve a spin-flip or change in spin state are forbidden, meaning that they
do not happen (although, as we will soon see, sometimes transitions that are forbidden do happen). What is important here is that you
will not get direct excitation from the S0 level to a higher energy triplet state. These transitions are truly forbidden and do not happen.
Transition 2 (Internal Conversion)
Internal conversion is the process in which an electron crosses over to another electronic state of the same spin multiplicity (e.g.,
singlet-to-singlet, triplet-to-triplet). The internal conversion in Figure 3.4 is from S2 to S1 and involves a crossover into a higher energy
vibrational state of S1. It is also possible to have internal conversion from S1 to a higher vibrational level of S0.
Transition 3 (Radiationless decay – loss of energy as heat)
The transitions labeled with the number (3) in Figure 3.4 are known as radiationless decay or external conversion. These generally
correspond to the loss of energy as heat to surrounding solvent or other solute molecules.
S1 = S0 + heat (3.2.3)
T1 = S0 + heat (3.2.4)
Note that systems in S1 and T1 can lose their extra energy as heat. Also, systems excited to higher energy vibrational and rotational
states also lose their extra energy as heat. The energy diagram level in Figure 3.4 shows systems excited to higher vibrational levels of
S1 and all of these will rapidly lose some of the extra energy as heat and drop down to the S1 level that is only electronically excited.
An important consideration that effects the various processes that take place for excited state systems is the lifetimes of the different
excited states. The lifetime of a particular excited state (e.g. the S1 state) depends to some degree of the specific molecular species
being considered and the orbitals involved, but measurements of excited state lifetimes for many different compounds allows us to
provide ballpark numbers of the lifetimes of different excited states.
The lifetime of an electron in an S2 state is typically on the order of 10-15 second.
The lifetime of an electron in an S1 state depends on the energy levels involved. For a π-π* system, the lifetimes range from 10-7 to 10-
9
second. For a n-π* system, the lifetimes range from 10-5 to 10-7 second. Since π-π* molecules are more commonly studied by
fluorescence spectroscopy, S1 lifetimes are typically on the order of 10-8 second. While this is a small number on an absolute scale of
numbers, note that it is a large number compared to the lifetimes of the S2 state.
The lifetime of a vibrational state is typically on the order of 10-12 second. Note that the lifetime of an electron in the S1 state is
significantly longer than the lifetime of an electron in a vibrationally excited state of S1. That means that systems excited to
vibrationally excited states of S1 rapidly lose heat (in 10-12 second) until reaching S1, where they then “pause” for 10-8 second.
Transition 4 (Fluorescence)
The transition labeled (4) in Figure 3.4 denotes the loss of energy from S1 as radiation. This process is known as fluorescence.
S1 = S0 + hν (3.2.5)
Therefore, molecular fluorescence is a term used to describe a singlet-to-singlet transition in a system where the chemical species was
first excited by absorption of electromagnetic radiation. Note that the diagram in Figure 3.4 does not show molecular fluorescence
occurring from the S2 level. Fluorescence from the S2 state is extremely rare in molecules and there are only a few known systems
where it occurs. Instead, what happens is that most molecules excited to energy states higher than S1 quickly (10-15 second) undergo an
internal conversion to a high energy vibrational state of S1. They then rapidly lose the extra vibrational energy as heat and “pause” in
the S1 state. From S1, they can either undergo fluorescence or undergo another internal conversion to a high energy vibrational state of
S0 and then lose the energy as heat. The extent to which fluorescence or loss of heat occurs from S1 depends on particular features of
the molecule and solution that we will discuss in more detail later in this unit.
An important aspect of fluorescence from the S1 state is that the molecule can end up in vibrationally excited states of S0, as shown in
the diagram above. Therefore, fluorescence emission from an excited state molecule can occur at a variety of different wavelengths.
Just like we talked about with absorbance and the probability of different transitions (reflected in the magnitude of the molar
absorptivity), fluorescent transitions have different probabilities as well. In some molecules, the S1-to-S0 fluorescent transition is the
Figure 3.6. Fluorescence excitation (left) and emission (right) spectra of anthracene.
Describe a way to measure the phosphorescence spectrum of a species that is not compromised by the presence of any
fluorescence emission.
The important thing to consider in addressing this question is that the lifetime of the S1 state from which fluorescence occurs is
approximately 10-8 second whereas the lifetime of the T1 state from which phosphorescence occurs is on the order of 10-4 to 100
seconds. Because of these different lifetimes, fluorescence emission will decay away rather quickly while phosphorescence emission
will decay away more slowly. The diagram in Figure 3.7 shows representations for the decay of fluorescence versus phosphorescence
as a function of time if the radiation source was turned off. The two can be distinguished by using a pulsed source. A pulsed source is
turned on for a brief instant and then turned off. Many fluorescent spectrophotometers use a pulsed source. The electronics on the
detector can be coordinated with the source pulsing. When measuring fluorescence, the detector reads signal when the pulse is on.
When measuring phosphorescence, a delay time during which the detector is turned off occurs after the pulse ends. Then the detector
is turned on for some period of time, which is referred to as the gate time. Figure 3.7 also shows where the delay and gate times might
be set for the sample represented in the decay curves. The proper gate time depends in part on how slow the phosphorescence decays.
You want a reasonable length of time to measure enough signal, but if the gate time is too long and weak to no phosphorescence occurs
at the end, the detector is mostly measuring noise and the signal-to-noise ratio will be reduced.
Since this is a ratio, the limits of φF are from 0 to 1. Species with quantum yields of 0.01 or higher (1 out of 100 excited species
actually undergo fluorescence) are useful for analysis purposes.
magnitude. Therefore, at low concentrations, absorption involves the measurement of a small difference between two large signals.
Fluorescence, on the other hand, is measured at 90o to the source. In the absence of fluorescence, as in a blank solution, there ought
to be no signal reaching the detector (however, there is still some scattered and stray light that may reach the detector as noise). At
low concentrations, fluorescence involves the measurement of a small signal over no background. For comparison, suppose you
tried to use your eyes to distinguish the difference between a 100 and 99 Watt light bulb and the difference between complete
darkness and a 1 Watt light bulb. Your eyes would have a much better ability to determine the small 1 Watt signal over darkness
than the difference between two large 100 and 99 Watt signals. The same occurs for the electronic measurements in a
spectrophotometer. Therefore, because emission involves the measurement of a small signal over no background, any type of
emission spectroscopy has an inherent sensitivity advantage of one to three orders of magnitude over measurements of absorption.
Fluorescence spectroscopy is an especially sensitive analysis method for those compounds that have suitable quantum yields.
Not surprisingly, fluorescence intensity varies linearly with the path length and with the concentration. K’ is a constant that is
dependent on the geometry and other factors and includes the fluorescence quantum yield. Since φF is a constant for a given system,
K’ is defined as K”φF. Of particular interest is that the fluorescence intensity relates directly to the source power. It stands to reason
that the higher the source power, the more species that absorb photons and become excited, and therefore the more that eventually emit
fluorescence radiation. This suggests that high-powered lasers, provided they emit at the proper wavelength of radiation to excite a
system, have the potential to be excellent sources for fluorescence spectroscopy.
The equation above predicts a linear relationship between fluorescence intensity and concentration. However, the utility of this
equation breaks down at absorbance values of 0.05 or higher leading to a negative deviation of the standard curve.
Something else that can possibly occur with fluorescence or other emission processes is that emitted photons can be reabsorbed by
ground state molecules. This is a particular problem if the S1-S0 emission transition is the one being monitored. In this situation, at high
concentrations of analyte, the fluorescence intensity measured at the detector may actually start to drop as shown in the standard curve
in Figure 3.8.
Figure 3.8. Standard curve for fluorescence showing re-absorption of emitted photons at higher concentrations.
Any changes in the system that will affect the number and force of collisions taking place in the solution will influence the magnitude
of the fluorescence emission. Collisions promote radiationless decay and loss of extra energy as heat, so more collisions or more
forceful collisions will promote radiationless decay and reduce fluorescence emission. Therefore, fluorescent intensity is dependent on
the temperature of the solution. Higher temperatures will speed up the movement of the molecules (i.e., higher translational energy)
leading to more collisions and more forceful collisions, thereby reducing the fluorescent intensity. Insuring that all the measurements
are done at the same temperature is important. Reducing the temperature of the sample will also increase the signal-to-noise ratio.
Another factor that will affect the number of collisions is the solvent viscosity. More viscous solutions will have fewer collisions, less
collisional deactivation, and higher fluorescent intensity.
The solvent can have other effects as well, similar to what we previously discussed in the section on UV/VIS absorption spectroscopy.
For example, a hydrogen-bonding solvent can influence the value of λmax in the excitation and emission spectra by altering the energy
levels of non-bonding electrons and electrons in π* orbitals. Other species in the solution (e.g., metal ions) may also associate with the
analyte and change the λmax values.
Many metal ions and dissolved oxygen are paramagnetic. We already mentioned that paramagnetic species promote intersystem
crossing, thereby quenching the fluorescence. Removal of paramagnetic metal ions from a sample is not necessarily a trivial matter.
Removing dissolved oxygen gas is easily done by purging the sample with a diamagnetic, inert gas such as nitrogen, argon or helium.
All solution-phase samples should be purged of oxygen gas prior to the analysis.
Another concern that can distinguish sample solutions from the blank and standards is the possibility that the unknown solutions have
impurities that can absorb the fluorescent emission from the analyte. Comparing the fluorescent excitation and emission spectra of the
The emission spectra of an aqueous solution of 2-naphthol at pH 2 and 13 are shown in Figure 3.9.
Figure 3.10. Resonance forms involving the oxygen atom of 2-naphthol and 2-naphtholate.
Consider the reaction shown below for the dissociation of 2-naphthol. This reaction may be either slow (slow exchange) or fast
(fast exchange) on the time scale of fluorescence spectroscopy. Draw the series of spectra that would result for an initial
concentration of 2-naphthol of 10-6 M if the pH was adjusted to 2, 8.5, 9.5, 10.5, and 13 and slow exchange occurred. Draw the
spectra at the same pH when the exchange rate is fast.
If slow exchange occurs, an individual 2-naphthol or 2-naphtholate species stays in its protonated or deprotonated form during the
entire excitation-emission process and emits its characteristic spectrum. Therefore, when both species are present in appreciable
concentrations, two peaks occur in the spectrum for each of the individual species. On the left side of Figure 3.11, at pH 2, all of the
species is in the neutral 2-naphthol form, whereas at pH 13 it is all in the anionic 2-naphtholate form. At pH 9.5, which equals the pKa
Figure 3.11. Representation of fluorescent emission spectrum of 2-naphthol as a function of pH under the conditions of
slow (left) and fast (right) exchange.
If fast exchange occurs, as seen on the right side of Figure 3.11, a particular species rapidly changes between its protonated and
deprotonated form during the excitation and emission process. Now the emission is a weighted time average of the two forms. If the
pH is such that more neutral 2-naphthol is present in solution, the maximum is closer to 355 nm (pH = 8.5). If the pH is such that more
anionic 2-naphtholate is present in solution, the maximum is closer to 415 nm (pH = 10.5). At the pKa value (9.5), the peak appears in
the middle of the two extremes.
What actually happens – is the exchange fast or slow? The observation is that the exchange of protons that occurs in the acid-base
reaction is slow on the time scale of fluorescence spectroscopy. Remember that the lifetime of an excited state is about 10-8 second.
This means that the exchange rate of protons among the species in solution is slower than 10-8 second and the fluorescence emission
spectrum has peaks for both the 2-naphthol and 2-naphtholate species.
Devise a procedure that might allow you to determine the pKa of a weak acid such as 2-naphthol.
The pKa value of an acid is incorporated into an expression called the Henderson-Hasselbalch equation, which is shown below where
HA represents the protonated form of any weak acid and A– is its conjugate base.
–
[A ]
pH = pKa + log (3.6.2)
[HA]
If a standard curve was prepared for 2-naphthol at a highly acidic pH and 2-naphtholate at a highly basic pH, the concentration of each
species at different intermediate pH values when both are present could be determined. These concentrations, along with the known
pH, can be substituted into the Henderson-Hasselbach equation to calculate pKa. As described earlier, this same process is used quite
often in UV/VIS spectroscopy to determine the pKa of acids, so long as the acid and base forms of the conjugate pair have
substantially different absorption spectra.
If you do this with the fluorescence spectra of 2-naphthol; however, you get a rather perplexing set of results in that slightly different
pKa values are calculated at different pH values where appreciable amounts of the neutral and anionic form are present. This occurs
because the pKa of excited state 2-naphthol is different from the pKa of the ground state. Since the fluorescence emission occurs from
the excited state, this difference will influence the calculated pKa values. A more complicated set of calculations can be done to
determine the excited state pKa values. UV/VIS spectroscopy is therefore often an easier way to measure the pKa of a species than
fluorescence spectroscopy.
Because many compounds are weak acids or bases, and therefore the fluorescence spectra of the conjugate pairs might vary
considerably, it is important to adjust the pH to insure all of either the protonated or deprotonated form.
Which compound will have a higher quantum yield: anthracene or diphenylmethane?
Answering this question involves a consideration of the effect that collisions of the molecules will have in causing radiationless decay.
Note that anthracene is quite a rigid molecule. Diphenylmethane is rather floppy because of the methylene bridge between the two
phenyl rings. Hopefully it is reasonable to see that collisions of the floppy diphenylmethane are more likely to lead to radiationless
decay than collisions of the rigid anthracene molecules. Another way to think of this is the consequences of a crash between a
Another important example of a chemiluminescent reaction involves the reaction of nitric oxide (NO) with ozone (O3) to produce
excited state nitrogen dioxide (NO2*) and oxygen gas. Nitric oxide is an important compound in atmospheric chemistry and, with the
use of an ozone generator, it is possible to use the chemiluminescent reaction as a sensitive way of measuring NO.
∗
NO = O3 = NO + O2 (3.7.1)
2
∗
NO = NO2 + hν (3.7.2)
2
An important feature of both chemiluminescent reactions above is that peroxide and ozone, which are strong oxidants, have an unstable
or energetic chemical bond. Chemiluminescence is a rare process only occurring in a limited number of chemical reactions.
Bioluminescence refers to a situation when living organisms use a chemiluminescent reaction to produce a luminescent emission. The
classic example is fireflies. There are also a number of bioluminescent marine organisms.
Triboluminescence is a form of luminescence caused by friction. Breaking or crushing a wintergreen-flavored lifesaver in the dark
produces triboluminescence. The friction of the crushing action excites sugar molecules that emit ultraviolet radiation, which is
triboluminescence but cannot be seen by our eyes. However, the ultraviolet radiation emitted by the sugar is absorbed by fluorescent
methyl salicylate molecules that account for the wintergreen flavor. The methyl salicylate molecules emit the light that can be seen by
our eyes.
Finally, light sticks also rely on a fluorescent process. Bending the light stick breaks a vial that leads to the mixing of phenyl oxalate
ester and hydrogen peroxide. Two subsequent decomposition reactions occur, the last of which releases energy that excites a
fluorescent dye. Emission from the dye accounts for the glow from the light stick.
Figure 4.2. Representation of the IR inactive symmetric stretch of carbon dioxide. The arrows represent bond dipoles.
A second vibration is the asymmetrical stretch (Figure 4.3). Each bond dipole does change on stretching and the molecule now has a
net dipole. Since the molecular dipole changes during an asymmetrical stretch, this vibration is IR active.
Figure 4.4. Representation of the IR active bending vibration of carbon dioxide. Arrows represent bond dipoles.
An atomic stretching vibration can be represented by a potential energy diagram as shown in Figure 4.5 (also referred to as a potential
energy well). The x-axis is the internuclear distance. Note that different vibrational energy levels, which are shown on the diagram as a
series of parallel lines, are superimposed onto the potential well. Also note that, if the bond gets to too high a vibrational state, it can be
ruptured.
Figure 4.5. Potential energy well for a covalent bond. Parellel lines in the well represent vibrational energy levels.
IR spectra are recorded in reciprocal wavenumbers (cm-1) and there are certain parts of the mid-IR spectrum that correspond to specific
vibrational modes of organic compounds.
2700-3700 cm-1: Hydrogen stretching
1950-2700 cm-1: Triple bond stretching
1550-1950 cm-1: Double bond stretching
700 -1500 cm-1: Fingerprint region
An important consideration is that as molecules get complex, the various vibrational modes get coupled together and the infrared (IR)
absorption spectrum becomes quite complex and difficult to accurately determine. Therefore, while each compound has a unique IR
spectrum (suggesting that IR spectroscopy ought to be especially useful for the qualitative analysis – compound identification – of
compounds), interpreting IR spectra is not an easy process. When using IR spectra for compound identification, usually a computer is
used to compare the spectrum of the unknown compound to a library of spectra of known compounds to find the best match.
IR spectroscopy can also be used for quantitative analysis. One limitation to the use of IR spectroscopy for quantitative analysis is that
IR sources have weak power that enhances the noise relative to signal and reduces the sensitivity of the method relative to UV/Visible
absorption spectroscopy. Also, IR detectors are much less sensitive than those for the UV/VIS region of the spectrum. IR bands are
narrower than observed in UV/VIS spectra so instrumental deviations to Beer’s Law (e.g., polychromatic radiation) are of more
concern. Fourier transform methods are often used to enhance the sensitivity of infrared methods, and there are some specialized IR
techniques that are used as well.
Figure 4.7. Representation of a device for measuring attenuated total reflectance spectra.
Figure 4.11. Intensity of radiation of λ = x, 2x and 4x at the sample as a function of the position of the moveable mirror
in a Michelson interferometer.
There are two important consequences from the plots in Figure 4.11. The first is that for each of these wavelengths, the intensity of the
radiation at the sample oscillates from full amplitude to zero amplitude as the mirror is moved. In a Michelson interferometer, the
moveable mirror is moved at a fixed speed from one extreme (e.g., +x extreme) to the other (e.g., –x extreme). After the relatively slow
movement in one direction, the moveable mirror is then rapidly reset to the original position (in the example we are using, it is reset
back to the +x extreme), and then moved again to record a second spectrum that is added to the first. Because the mirror moves at a set,
fixed rate, the intensity of any one of these three wavelengths varies as a function of time. Each wavelength now has a time domain
property associated with it.
The second important consequence is that the time domain property of radiation with wavelengths x, 2x and 4x is different. An
examination of the plots in Figure 4.11 shows that the pattern of when the radiation is at full and zero amplitude is different for the
radiation with wavelength x, 2x or 4x. The aggregate plot of all of these wavelengths added together is called an interferogram. If a
sample could absorb infrared radiation of wavelength x, the intensity of light at this wavelength would drop after the sample and it
would be reflected in the interferogram.
The usual process of recording an FT-IR spectrum is to record a background interferogram with no sample in the cell. The
interferogram with a sample in the cell is then recorded and subtracted from the background interferogram. The difference is an
interferogram reflecting the radiation absorbed from the sample. This time domain infrared spectrum can then be converted to a
frequency domain infrared spectrum using the Fourier transform.
It is usually common to record several interferograms involving repetitive scans of the moveable mirror and then adding them together.
An advantage of using multiple scans is that the signal of each scan is additive. Noise is a random process so adding together several
scans leads to a reduction due to cancelling out of some of the noise. Therefore, adding together multiple scans will lead to an
improvement in the signal-to-noise ratio. The improvement in the signal-to-noise ratio actually goes up as the square root of the
Topic hierarchy
The problem with large aerosol droplets is that they will not make it through all of the necessary steps during their lifetime in the
flame. These drops will contribute little to the signal, but their presence in the flame will create noise and instability in the flame that
will compromise the measurement. Hence, only the finest aerosol droplets will lead to atomic species and only those are introduced
into the flame.
The various steps outlined in Equations 1-6 also imply that there will be a distinct profile to the flame. Profiles result because of the
efficiency with which neutral and excited atoms are formed in a flame. Therefore, a specific section of the flame will have the highest
concentration of ground state atoms for the metal being analyzed. The absorbance profile that shows the concentration of ground state
atoms in the flame is likely to be different than the emission profile that shows the concentration of excited state atoms in the flame.
Figure 6.2 shows representative absorption profiles for chromium, magnesium and silver. Magnesium shows a peak in its profile. The
increase in the lower part of the flame occurs because exposure to the heat creates more neutral ground state atoms. The decrease in the
upper part of the flame occurs due to the formation of magnesium oxide species that do not absorb the atomic line. Silver is not as
Figure 6.2. Representation of the flame profiles for chromium, magnesium and silver.
When performing atomic absorbance or emission measurements using a flame atomization source, it is important to measure the
section of the flame with the highest concentration. There are controls in the instrument to raise and lower the burner head to insure
that the light beam passes through the optimal part of the flame.
An important factor in the characteristics of a flame is the identity of the fuel and oxidant. Standard Bunsen burner flames use methane
as the fuel and air as the oxidant and have a temperature in the range of 1,700-1,900oC. A flame with acetylene as the fuel and air as
the oxidant has a temperature in the range of 2,100-2,400oC. For most elements, the methane/air flame is too cool to provide suitable
atomization efficiencies for atomic absorbance or emission measurements, and an acetylene/air flame must be used. For some
elements, the use of a flame with acetylene as the fuel and nitrous oxide (N2O) as the oxidant is recommended. The acetylene/nitrous
oxide flame has a temperature range of about 2,600-2,800oC. There are standard reference books on atomic methods that specify the
type of flame that is best suited for the analysis of particular elements.
It is also important to recognize that some elements do not atomize well in flames. Flame and other atomization methods are most
suitable for the measurement of metals. Non-metallic elements rarely atomize with enough efficiency to permit analysis of trace levels.
Metalloids such as arsenic and selenium have intermediate atomization efficiencies and may require specialized atomization methods
for certain samples with trace levels of the elements. Mercury is another atom that does not atomize well and often requires the use of a
specialized atomization procedure. Flame methods are usually used for atomic absorbance measurements because most elements do
not produce high enough concentrations of excited atoms to facilitate sensitive detection based on atomic emission. Alkali metals can
be measured in a flame by atomic emission. Alkaline earth metals can possibly be measured by flame emission as well provided the
concentration is high enough.
Figure 6.7. Calibration curve for atomic emission showing self-absorption at higher concentration.
Another advantage is that the high number of Ar+ ions and free electrons suppress the ionization of other elements being measure,
thereby increasing the number of neutral atoms whose emission is being measured. The argon used to generate the plasma is
chemically inert compared to the chemical species that make up a flame, which increases the atomization efficiency. The inductively
coupled plasma tends to be quite stable and reproducible. The combination of high temperature with chemically inert environmental
reduces matrix effects in the plasma relative to other atomization sources, but it does not eliminate them and matrix effects must
always be considered. Some elements (e.g., mercury, arsenic, phosphorus) that are impractical to analyze on a flame or furnace
instrument without specialized atomization techniques can often be measured on an ICP.
A final advantage of the plasma is that there are now methods to introduce the atoms into a mass spectrometer (MS). The use of the
mass spectrometer may further reduce certain matrix effects. Also, mass spectrometry usually provides more sensitive detection than
emission spectroscopy.
Figure 6.8. Illustration of an arc atomization source with a plasma (shown in blue). From:
http://www.asdlib.org/learningModules/AtomicEmission/index.html
Topic hierarchy
Figure 6.10. Atomic absorption line superimposed onto the overall output of a continuum source/monochromator system.
What is the problem with reducing the slit width of the monochromator to get a narrower line?
The problem with reducing the slit width is that it reduces the number of photons or source power reaching the sample. Reducing the
slit width on a continuum source to a level that would provide a narrow enough line to respond to atomic absorption would reduce the
power so that it would not be much above the noise. Therefore, hollow cathode lamps, which emit intense narrow lines of radiation
specific to the element being analyzed, are needed for atomic absorption measurements.
With this understanding we can ask why the hollow cathode lamp has a low pressure of argon filler gas.
The pressure of the argon is low to minimize collisions of argon atoms with sputtered atoms. Collisions of excited state sputtered atoms
with argon atoms will lead to broadening of the output of the hollow cathode lamp and potentially lead to the same problem described
above with the use of a continuum source. A low pressure of argon in the lamp insures that the line width from the hollow cathode
lamp is less than the line width of the absorbing species.
Figure 6.13. Plot showing an atomic absorption line superimposed over molecular absorption.
Can you design a feature that could be incorporated into an atomic absorption spectrophotometer than can account for both
scattered light and light absorbed by molecular species?
To address this question, we need to think back to the previous discussion of the source requirement for atomic absorption
spectrophotometers. Earlier we saw that it was not possible to use a continuum source with a monochromator since the atomic
absorption was so negligible as to be non-detectable. However, a continuum source will measure molecular absorption and will
respond to any scattered radiation. The answer is to alternately send the output from the hollow cathode lamp and a continuum source
(the common one used in AA instruments is a deuterium lamp) to the flame. The output of the hollow cathode lamp will be diminished
by atomic absorption, molecular absorption and scatter. The continuum lamp will only be diminished by molecular absorption and
scatter, since any contribution from atomic absorption is negligible. By comparing these, it is possible to correct the signal measured
when the hollow cathode lamp passes through the flame for scattered radiation and molecular absorption. In atomic absorption
spectroscopy, this process is referred to as background correction.
An alternative way of getting a broadened source signal to pass through the flame is known as the Smith-Hieftje method (named after
the investigators who devised this method). The Smith-Hieftje method only uses a hollow cathode lamp. Earlier, when we discussed
hollow cathode lamps, we learned that the argon pressure inside the lamp was kept low to avoid collisional broadening. We also
learned that the current was not set to a high value because it would sputter off too many atoms and shorten the lamp lifetime. Another
observation when running a hollow cathode lamp at a high current is that the lamp emission lines broaden. This occurs because, at a
high current, so many atoms get sputtered off into the hollow cathode that they collide with each other and broaden the wavelength
distribution of the emitted light. The Smith-Hieftje method relies on using a pulsed lamp current. For most of the time, the lamp is run
at its optimal current and emits narrow lines that would diminish when passing through the flame due to atomic absorption, molecular
absorption and scatter. For a brief pulse of time, the current is set to a very high value such that the lamp emits a broadened signal.
When this broadened signal passes through the flame, atomic absorption is negligible and only molecular absorption and scatter
decrease the intensity of the beam.
A third strategy is to use what is known as the “two-line” method. This can be used in a situation where you have a source that emits
two narrow atomic lines, one of which is your analysis wavelength and the other of which is close by. Looking back at Figure 13, the
analysis wavelength is diminished in intensity by atomic absorption, molecular absorption and scattering. A close by line does not have
any atomic absorption and only is reduced in intensity by molecular absorption and scattering. While it might at first seem difficult to
see how it is possible to get nearby atomic lines for many elements, there is something known as the Zeeman Effect that can be used
for this purpose. Without going into the details of the Zeeman Effect, what is important to know is that exposing an atomic vapor to a
strong magnetic field causes a slight splitting of the energy levels of the atom causing a series of closely spaced lines for each
electronic transition. The neighboring lines are about 0.01 nm from each other, making them ideal for monitoring background
molecular absorption and scatter. Corrections using the Zeeman Effect are more reliable than those using a continuum source. The
magnetic field can be applied either to the hollow cathode lamp or the atomization source. The method is useful in flame and graphite
furnace measurements.
Topic hierarchy
Figure 6.14. Standard addition plot for two samples with the same concentration. The sample with the higher slope has a
matrix that enhances the signal.
The experimental steps involved in conducting a standard addition are more complex than those involving the use of a standard curve.
If someone is testing a series of samples with similar properties that have similar matrices, it is desirable to use the standard addition
procedure on one or a few samples and compare the concentration to that obtained using a standard curve. If the two results are similar,
then it says that the matrix effects are minimal and the use of a standard curve is justified.