CCT-3300 Series Conductivity (TDS) Controller
CCT-3300 Series Conductivity (TDS) Controller
CCT-3300 Series Conductivity (TDS) Controller
Operation Manual
Notes before Operation
1
.2 Applications
1.
This series is widely used for the water treatment control and monitoring of water
treatment (high salt content), concentrated solution, recirculating cooling water and
etc.
.3 Products classification
1.
Power Constant
Model Frequency(Hz) Current loop modes Display
supply selection
Three digits and a
CCT-3300E DC 24V --------------- Instrument/Transmitter 0.1-1.0
half
Three digits and a
CCT-3310E AC 110V 50/60 Instrument/Transmitter 0.1-1.0
half
Three digits and a
CCT-3320 AC 220V 50/60 Instrument/Transmitter 0.1-1.0
half
Three digits and a
CCT-3320E AC 220V 50/60 Instrument/Transmitter 0.1-1.0
half
Three digits and a
CCT-3320T AC 220V 50/60 Instrument/Transmitter 0.1-1.0
half
Note: The model with E is the EMC enhancement type.
CCT-3320T(Conductivity Monitor) without any control outputs.
i. Please choose CCT-3300E (power supply 24V DC) if the onsite with damp
environment or systems with 24V power supply.
ii. If the onsite voltage is unstable or the site is far away from the power
station, please use wide range AC/DC voltage-stabilizing type DC 24V
power supply, please choose CCT-3300E.
iii. If the measurement value unstable caused by using converter system,
electronic device for water treatment, UV light electronic ballast device,
power supply pollution, please choose CCT-3320E
iv. Under the small power generation device which the power unstable, or far
away from the power supply, please use wide range AC/DC
voltage-stabilizing type DC 24V power supply, please choose CCT-3300E
v. UV does harm to the LCD display screen, so please do not place the meter
under sunshine.
2
.5 Main technical feature
1.
Temperature compensation
NTC10K temperature component
Working Environment Temp.0~50℃ Relative Humidity: ≤80%RH
3
II. Outline Dimension and Rear Terminals
.1 Outline dimension
2.
.2 Rear terminals
2.
CCT-3300E
CCT-3310E
4
CCT-3320/CCT-3320E/CCT-3320T
Note:CCT-
Note:CCT-3320T RELAY terminal is blank, without any connection.
.3 Wire connection
2.
5
III. Electrical Connection
6
.2 Electrical connection
3.
7
The water quality is limited within HIGH limit
If the meter is used for pharmaceutical industry, food, drink, water purification and etc,
when the water quality is limited at HIGH limit, the solenoid valve will switch the water
flow direction to make sure the water purity. When the water quality at LOW limit, the
meter will back to normal operation. The space between HIGH and LOW is backlash.
8
.2 Key board functions Introduction
4.
Under main measuring menu, pressing “ ” for three seconds and enter setting menu
automatically
The measurement range depends on the electrode constant, so please choose the suitable
electrode constant for your applications.
10
.3 Installation method
6.
Please follow the correct installation method to install the electrode strictly. The
incorrect installation will cause the reading error.
)Picture 1a the fitting is too long and the stretched parts is too short, which will
1)
lead to the dead space in electrode and incorrect measurement. please follow the
picture 1b (FLOW=stretch into flow direction deeply)。 。
Installation 1a installation 1b
2 ) Picture 2a, the installation mode will lead to air space and incorrect
measurement and instability. Please follow the picture 2b
Installation 2a installation 2b
11
)Picture 3 , there will be air space inside the pipeline, the air will cause
3)
incorrect reading.
4)Picture 4a, the below installation will cause incorrect reading ,because
the water can not through the measuring hole of the electrode.
Notes:
(1) The electrode should be installed in a place in the pipeline where the stream is steady
and air bubbles are hard to generate.
(2) No matter the conductance cell is horizontally or vertically installed, it should be
deeply inserted into the moving water.
(3) The conductivity signal is weak electronic signal and its collecting cable should be
separately installed. When threading cable joint or connecting terminal board is used,
to avoid wetting interference or breakdown of measurement unit circuit, they should
not be connected to the same group of cable joint or terminal board with the power
line or control line.
(4) When the measurement cable needs to be lengthened, it’s recommended to make an
12
agreement with the factory before placing an order.
(5) Please keep the measuring part of electrode clean, and do not directly contact the
surface by hands or contact with the oil stain objects
(6) Electrode is a kind of precision components, so please do not change any part of the
electrode. The accuracy will be incorrect if the electrode was destroyed by the
strong acid, strong alkali, scrape from machine and etc.
(7) The meter is made by precision integrated circuit and electronic components, so it
needs to place in case or dry environment.
(8) In order to guarantee the safety operation, pls checking after installation then switch
on.
When the reading is incorrect or unstable, please check the meter and electrode.
a.Distinguish the failure source which comes from meter or electrode
Firstly, remove the white wire from the wire terminal and check the conductivity
reading, if the reading is 0 and stable, the meter is good. The problem can be
initially identified from the electrode installation.
b. Sensor installation question failure judgment.
Remove the electrode from the fittings, and then use the electrode to test the
water quality (user already know the conductivity reading of the water before
test) . If the reading is correct, so the installation is correct. If the reading with
error, so the electrode failure.
C. Meter mode or transmitter mode of 4~20mA judgment.
Before the judgment please makes sure there is no wire connection on the terminal.
mA mode Output terminal Voltage of terminal Voltage of cable
Meter mode I+/I- >12V DC no
Transmitter mode T+/T- no DC24V
13
A. Reset the electrode constant
A. Incorrect constant Setting
3.Serious B. Change the electrode by new one.
B. The electrode constant is changed
error of C. Install the electrode at where the flow’s speed is
reading Incorrect electrode installation suitable.
D. Put the electrode deeply into the flowing water.
A. Set the transfer setting again
4.different mA
A. Reviver transfer error B. Loop resistance is too large, enlarge the
reading
B. Not up to 20mA cable.
between
C. Incorrect setting on sending C. Set the corresponding of mA and reading
sender and
D. mA transfer error again.
receiver
D. Use the ammeter to check the current.
CCT-3300 series including 4 models, please check the below items to choose the suitable power
supply, material, fixed method for your applications
:
NOTE:
: 1m, 5m, 10m, 20m (please indicate another especial length before placing an
Cable length:
order).
*Without the influence on the operation, any small change or improvement on the products by
the manufacturer will not be notified separately. Please make the object as the standard.
14