ACUSON Sequoia: System BARCO Monitor Softkey Repair

Download as pdf or txt
Download as pdf or txt
You are on page 1of 18

ACUSON Sequoia

US

Replacement of Parts
System
BARCO monitor softkey repair

08267697
08246951
08245875

© Siemens AG 2003
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical
Ultrasound
ejt DivisionSolutions
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US02-111.841.01.01.02 English


Replaces: n.a. Doc. Gen. Date: 04.04
n.a.

Part No.: n.a.


sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

ACUSON Sequoia US02-111.841.01.01.02 Page 2 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US02-111.841.01.01.02 Page 3 of 18 ACUSON Sequoia


Medical Solutions USA, Inc. 04.04 sd
4 Table of Contents
1- 0Table of Contents

1 _______ Barco Monitor Softkey Repair _____________________________________ 5

Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Prepare Monitor Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remove Back Panel Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remove Speaker Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Remove monitor housing and bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Repair Softkey Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Modify bezel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Reassemble system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Administration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

ACUSON Sequoia US02-111.841.01.01.02 Page 4 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.
Barco Monitor Softkey Repair 5
2-
Scope
1Barco Monitor Softkey Repair

This document provides instructions for fixing sticky softkey problems on Barco Monitors
on the ACUSON Sequoia™ and ACUSON Aspen™ Ultrasound System.

Time Required 0

Repairing the Barco Softkey will take approximately one hour.

Required Material 0

• Standard Toolkit

Parts 0

• None

Important Information 0

WARNING Observe Electrostatic Discharge Procedures


¹ Failure to comply will damage the assembly and will
result in an extra service call.

Siemens US02-111.841.01.01.02 Page 5 of 18 ACUSON Sequoia


Medical Solutions USA, Inc. 04.04 sd
6 Barco Monitor Softkey Repair

Prepare Monitor Unit 0

1. Turn off the power.


2. Unplug the system from the wall outlet.
3. Set the tilt/swivel so that the base of the monitor slants forward. Lock it in place.

Remove Back Panel Covers 0

1. Remove the screws and lock washers securing the back panel cover to the monitor
housing.

Fig. 1: Remove screws securing the back panel cover to the monitor
2. Pull down then lift out on the back panel cover to remove.

Fig. 2: Remove back panel cover

ACUSON Sequoia US02-111.841.01.01.02 Page 6 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.
Barco Monitor Softkey Repair 7

Remove Speaker Covers 0

1. Place an Allen wrench in the hole on the underside of the speaker and push gently up-
ward until the speaker latch disengages.

Fig. 3: Use an Allen wrench to pop speaker cover off.


2. Lift the bottom of the speaker cover out then pull down to remove.

Fig. 4: Remove the speaker cover

Siemens US02-111.841.01.01.02 Page 7 of 18 ACUSON Sequoia


Medical Solutions USA, Inc. 04.04 sd
8 Barco Monitor Softkey Repair

3. Remove the speaker cable.

Fig. 5: Remove the speaker cable


4. Repeat steps 1-3 for the other speaker cover.

Remove monitor housing and bezel 0

1. Remove the screws securing the bezel to the monitor housing.

Fig. 6: Screws securing the bezel to the monitor housing

ACUSON Sequoia US02-111.841.01.01.02 Page 8 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.
Barco Monitor Softkey Repair 9

2. Remove the cable from J1 on the brightness/contrast/degauss board on the left side of
the monitor.

Fig. 7: Cable from brightness/contrast/degauss board


3. Slide the monitor housing out approximately one inch.
- Stand behind the system and lift the monitor housing up to release it from the guide
slots.
- Once the monitor housing is free of the guide slots, slide it away from the system to
remove.
4. Remove the screws and washers that secure the bezel to the monitor.

Fig. 8: Use a screwdriver to remove the screws that secure the bezel to the monitor.

Siemens US02-111.841.01.01.02 Page 9 of 18 ACUSON Sequoia


Medical Solutions USA, Inc. 04.04 sd
10 Barco Monitor Softkey Repair

NOTE Do not remove the screw securing the grounding wire to the mon-
itor.

Fig. 9: Monitor grounding wire


5. Set the screws and washers aside for replacement.

ACUSON Sequoia US02-111.841.01.01.02 Page 10 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.
Barco Monitor Softkey Repair 11

NOTE Most monitor bezels have a circular lock washer and an additional
rectangular or circular spacer washer used for spacing reasons.
Be sure to retain and replace both sets of washers when reassem-
bling the monitor bezel.

Fig. 10: Rectangular bezel spacing washer

Fig. 11: Location of spacing washer on bezel


6. Lift bezel 2-3 inches away from the monitor.

NOTE Do not completely remove the bezel from the monitor as there are
still cables that connect the two.

7. Remove the cable from J1 on the PCB board as shown in Figure 12.

Siemens US02-111.841.01.01.02 Page 11 of 18 ACUSON Sequoia


Medical Solutions USA, Inc. 04.04 sd
12 Barco Monitor Softkey Repair

8. Remove the connector from J1 on the bezel.

Fig. 12: Bezel PCB Board and cables


Pos. 1 PCB Board cable
Pos. 2 J1 connection on Monitor Bezel

9. Remove the screws securing the PCB board to the bezel.


10. Remove the PCB board from the bezel but leave it attached to the monitor.
11. Remove the soft keys and the volume wheel from the bezel.

NOTE Each soft key is different in length and has a specific location on
the bezel. Mark each soft key noting its distinct location and orien-
tation for replacement.

ACUSON Sequoia US02-111.841.01.01.02 Page 12 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.
Barco Monitor Softkey Repair 13

Repair Softkey Problem 0

Modify bezel 0

1. Using an X-Acto knife, cut off the top middle two positioning pins on the bezel.

Fig. 13: Cut the two top middle positioning pins

NOTE It is important to leave the outermost and bottom middle position-


ing pins for alignment of the PCB board during replacement.

2. Clean the soft key buttons and soft key area on the front bezel.

Reassemble system 0

1. Replace the soft key buttons and volume wheel, ensuring that each softkey has been
replaced to the correct position.

Siemens US02-111.841.01.01.02 Page 13 of 18 ACUSON Sequoia


Medical Solutions USA, Inc. 04.04 sd
14 Barco Monitor Softkey Repair

2. Line up the positioning pins and replace the PCB board.

Fig. 14: Top middle and bottom middle positioning pegs before modification

Fig. 15: Top middle and bottom middle positioning pegs after modification
3. Screw the PCB board onto the bezel.
4. Replace the cable onto J1 on the PCB board
5. Replace the connector onto J1 on the bezel.
6. Replace the screws, lock washers, and spacing washers that secure the bezel to the
monitor.

NOTE Hold the bezel assembly tight against the monitor.

7. Replace the monitor housing.

ACUSON Sequoia US02-111.841.01.01.02 Page 14 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.
Barco Monitor Softkey Repair 15

NOTE Be sure to have one washer on each side of the monitor housing
tab for spacing reasons. If both washers are one one side of the
monitor housing, it may not fit properly and the tabs may be dam-
aged or broken.

8. Replace the cable at J1 on the contrast/brightness/degauss board.


9. Replace the screws securing the bezel to the monitor housing.
10. Replace the speaker cables.
11. Replace the speaker cover by lining up the top edge of the cover first then the bottom.
Ensure that the speaker clicks into place.
12. Replace the back panel cover.
13. Replace the screws and lock washers securing the back panel cover to the monitor
housing.

Siemens US02-111.841.01.01.02 Page 15 of 18 ACUSON Sequoia


Medical Solutions USA, Inc. 04.04 sd
16 Barco Monitor Softkey Repair

Verification 0

After the system has been reassembled, verify that none of the softkeys stick and that the
monitor is operational.

1. Plug the system into the wall outlet.


2. Boot the system to SUI.
3. Verify the brightness, contrast and degauss levels are set correctly.
4. Verify functionality of all four softkey buttons.
5. Select UI Diagnostics in the Service Diagnostics menu.
6. Verify that the monitor passes all tests in the video test patterns.
7. Verify that the speakers work correctly.

ACUSON Sequoia US02-111.841.01.01.02 Page 16 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.
Barco Monitor Softkey Repair 17

Administration 0

Complete all paperwork as required by the regional office.

Siemens US02-111.841.01.01.02 Page 17 of 18 ACUSON Sequoia


Medical Solutions USA, Inc. 04.04 sd
18 Barco Monitor Softkey Repair

ACUSON Sequoia US02-111.841.01.01.02 Page 18 of 18 Siemens


04.04 sd Medical Solutions USA, Inc.