ACUSON Sequoia

US

Installation Instructions
System
System Delivery Procedure

08267697
08246951
08245875

© Siemens AG 2003
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical
Ultrasound
pd DivisionSolutions
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US02-101.812.01.04.02 English


Replaces: US02-101.812.01.02.02 Doc. Gen. Date: 02.04
n.a.

Part No.:
sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

ACUSON Sequoia US02-101.812.01.04.02 Page 2 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US02-101.812.01.04.02 Page 3 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
4 Table of Contents
1- 0Table of Contents

1 _______ System Delivery Procedure _______________________________________ 5

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Required Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Check for Shipping Damage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Remove the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Unpack the crate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Remove travel stabilizing items. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Verify Contents of Packaging List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Set Up the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Attach the Peripheral Bay Doors, if required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Power on the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Check the date and time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Verifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Check the Transducer Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Check Imaging Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Check Every Transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Check Peripheral Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
To verify printer: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
If a VCR is installed, check the VCR function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Powering down System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Customer Acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Crating the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Packaging the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

ACUSON Sequoia US02-101.812.01.04.02 Page 4 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 5
2-
Overview
1System Delivery Procedure

This document provides instructions for powering on and checking a ACUSON Sequoia
ultrasound system at a customer site.
ACUSON Sequoia system power on and check includes:
• Uncrating the system and peripheral devices
• Inventorying the shipment
• Powering on the system
• Checking the system configuration
• Connecting transducers

Time Required 0

Approximately 1 hour

Siemens US02-101.812.01.04.02 Page 5 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
6 System Delivery Procedure

Before You Start 0

Ensure there is sufficient space to allow removing the system from the crate.

Required Materials 0

• Flat Head Screwdriver


• Utility Knife
• Diagonal cutter

ACUSON Sequoia US02-101.812.01.04.02 Page 6 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 7

Check for Shipping Damage 0

Notify Siemens immediately if any damage is found.

1. Verify the “TIP N TELL” on the crate is not activated.


2. Verify the “SHOCK” indicator is not tripped.

Siemens US02-101.812.01.04.02 Page 7 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
8 System Delivery Procedure

Remove the system 0

Unpack the crate 0

1. Remove the straps from the shipping container.


2. Remove the top cover.
3. Remove the clips (both sides) and container sleeves
4. Remove the sterilization tub box (if present) from front of system.
5. Remove the front piece , see figure below, of the container.
a) Remove the wing nuts (2); there is one on each side of the container.
b) Pull the sides of the container down.
6. Remove the ramps.
a) Line up the arrow on the ramps with the arrows on the pallet.
b) Pull down the ramp supports.
7. Pull system off the pallet.
8. Remove the plastic cover from the system.
9. Remove the monitor collar from the system.

ACUSON Sequoia US02-101.812.01.04.02 Page 8 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 9

10. Remove the accessories from both boxes.

Fig. 1: Crating Diagram

Remove travel stabilizing items 0

1. Remove the plastic cover from the system.

Siemens US02-101.812.01.04.02 Page 9 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
10 System Delivery Procedure

2. Remove the monitor collar.

Fig. 2:
3. Remove all accessories and transducers from crate.
4. Do NOT remove the straps holding the peripherals to the system.

ACUSON Sequoia US02-101.812.01.04.02 Page 10 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 11

5. Remove the straps and packaging material from peripheral bay.

Fig. 3:

Final steps 0

1. Move system and accessories to customer specified location.


2. Remove the transducers from the shipping cases and remove the plastic wrap from the
transducer cables.

Siemens US02-101.812.01.04.02 Page 11 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
12 System Delivery Procedure

3. Check all peripheral connections to ensure cords are secured.

Fig. 4:

ACUSON Sequoia US02-101.812.01.04.02 Page 12 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 13

Verify Contents of Packaging List 0

• Check peripherals shipped, and make sure they match the packing list. Notify Siemens
immediately if any discrepancy is found.
• Check that the peripherals are safely secured to the system.

Siemens US02-101.812.01.04.02 Page 13 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
14 System Delivery Procedure

Set Up the System 0

Attach the Peripheral Bay Doors, if required. 0

NOTE Systems equipped with the Perspective® Advanced Display Op-


tion must have blank peripheral bay doors installed on both sides
of the peripheral bay.

1. Retract the spring-loaded L-pin on the peripheral bay door.


2. Insert the peripheral bay door pegs into the holes in the peripheral bay opening.
3. Release the L-pin.
4. Repeat for the opposite side.

Power on the System 0

1. Plug the system power cord into a power outlet specified by the customer.
2. Press the power button on the left under the monitor to power on the system.
The system takes approximately two minutes to boot. If the system is not completely
booted after five minutes call Siemens.

1. Check that the screen appears as shown in Figure 1 or Figure 2, depending on system
type (i.e. C256 or GI512)

Fig. 5: GI512 Default 2-D imaging screen

ACUSON Sequoia US02-101.812.01.04.02 Page 14 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 15

NOTE To properly see the image area, it may be necessary to increase


the 2-D gain

Fig. 6: C256 Default 2-D imaging screen

Siemens US02-101.812.01.04.02 Page 15 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
16 System Delivery Procedure

Setup key

2-D Gain Knob

Transducer
key Mode keys Trackball and Select buttons

Fig. 7: ACUSON Sequoia user interface

Check the date and time 0

1. Verify the date and time shown in the upper right corner of the screen is accurate. If the
date or time is not accurate call Siemens.

ACUSON Sequoia US02-101.812.01.04.02 Page 16 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 17

Verifications 0

Figure 4 shows examples of the Acuson transducer types that may be included with the
system.
If the transducer looks like this... The image should look like this...

Curved:
4C1
5C2
6C2
8C4

Endocavity
(Curved):
EV-8C4
EC-10C5

Linear:
15L8
15L8w
6L3
8L5

Vector:
3V2c
4V1
4V2
5V2c
7V3c
8V5
V5M
V7B

Fig. 8: Transducer types in 2-D imaging mode

Siemens US02-101.812.01.04.02 Page 17 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
18 System Delivery Procedure

Check the Transducer Ports 0

1. Insert a transducer into the left transducer port.

Some transducers are keyed; they can be inserted only into the right or
left transducer ports. Do not try to force a transducer into a port.

Fig. 9: Keyed transducer ports


- Slide the transducer connector into the transducer port. The cable must be facing up,
and the knob must be horizontal.
- Turn the knob until it is vertical. The knob should turn easily; do not force it.
Properly connected
Not connected transducer

Fig. 10: Transducer connectors

ACUSON Sequoia US02-101.812.01.04.02 Page 18 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 19

2. Verify that the transducer initializes. See Figure 4 for examples of 2-D imaging screens
for each transducer type.

The screen displays the active transducer in the upper right corner below
the date and time.

Active
transducer

Fig. 11:
3. Remove the transducer from the left transducer port, and repeat steps 1-2 for the cen-
ter transducer port.
4. Remove the transducer from the center transducer port, and repeat steps 1-2 for the
right transducer port.
5. If the transducer did not fit in one or more of the ports, select another transducer and re-
peat steps 1-4.

Check Imaging Modes 0

1. Insert a transducer in any of the transducer ports.

Siemens US02-101.812.01.04.02 Page 19 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
20 System Delivery Procedure

2. Press M Mode to check that the system enters M-mode.


If the transducer looks like this... The image should look like this...

Curved:
4C1
5C2
6C2
8C4

Endocavity
(Curved):
EV-8C4
EC-10C5

Linear:
15L8
15L8w
6L3
8L5

Vector:
3V2c
4V1
4V2
5V2c
7V3c
8V5
V5M
V7B

Fig. 12: Transducer types in M-mode imaging mode


3. Press 2D Only to return to 2-D mode.

ACUSON Sequoia US02-101.812.01.04.02 Page 20 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 21

4. Check Color Doppler mode.


- Press D Color to check that the system enters Color Doppler mode.
If the transducer looks like this... The image should look like this...

Curved

Color box

Linear

Color box

Vector

Endo-cavity
(Vector) Color box

Fig. 13: Transducer types in Color Doppler imaging mode


- Turn the D Gain knob to the right to fill the color box.
5. Press 2D Only to return to 2-D mode.

Siemens US02-101.812.01.04.02 Page 21 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
22 System Delivery Procedure

6. Press PW to check that the system enters PW Doppler mode.


If the transducer looks like this... The image should look like this...

Curved

Linear

Vector

Endo-cavity
(Vector)

Fig. 14: Transducer types in Pulsed Wave (PW) Doppler imaging mode
7. Press 2D Only to return to 2-D mode.

ACUSON Sequoia US02-101.812.01.04.02 Page 22 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 23

8. If the system includes the Auxiliary CW Transducer Package (see the packing list/sales
order), press Aux to check that the system displays the CW Doppler strip, as shown be-
low.

Fig. 15: Continuous Wave (CW) Doppler mode


9. Press 2D Only to return to 2-D mode.

Check Every Transducer 0

Insert each transducer into a transducer port, and verify that it initializes in 2-D imaging.
See Figure 4 for examples of properly initialized 2-D images for each transducer type.

Siemens US02-101.812.01.04.02 Page 23 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
24 System Delivery Procedure

Check Peripheral Operation 0

To verify printer: 0

1. Press the D Color button.


2. Turn the D GAIN knob clockwise until the color box is filled with color.
3. Press the PRINT button.
4. Verify the printer prints a copy of the system monitor display.

NOTE If a color printer is installed the print will have color in it. Otherwise
a black and white image will be printed.

If a VCR is installed, check the VCR function. 0

1. Insert a blank VCR test tape.


2. Press VCR Record on the ACUSON Sequoia system; record for 10 seconds.
3. Press Stop/Eject button once to stop recording.
4. Press Rewind.
5. Press Play. Verify time and date of recording in top right corner of screen.
6. Press Stop/Eject button to stop playing.
7. Press rewind to rewind the tape
8. Press Stop/Eject button and remove the tape from the VCR.

ACUSON Sequoia US02-101.812.01.04.02 Page 24 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 25

Powering down System 0

1. Press the power button to power down the system


2. Leave power cord plugged into the customer specified outlet.

Siemens US02-101.812.01.04.02 Page 25 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
26 System Delivery Procedure

Customer Acceptance 0

• Ensure that all transducer and peripheral serial numbers match the packing list.
• Obtain the customer’s signature as proof of acceptance.

ACUSON Sequoia US02-101.812.01.04.02 Page 26 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.
System Delivery Procedure 27

Crating the System 0

If the system needs to be sent back to the factory see the instructions on the container
sleeve.

Packaging the system 0

1. Remove the Ramps.


2. Line up the arrow on the ramps with the arrow on the pallet.
3. Pull down the ramp supports.
4. Remove the two wing nuts, one on each side of the box.
5. Remove the nose piece.
6. Align the back of the system to the ramps.
7. Install the monitor collar on the system.
8. Push the foot pedal to the position 2 icon on the pedal.
9. Place a plastic bag over the system.
10. Push the system (from the back) on to the pallet. Ensure that the system is all way on
the pallet.
11. Remove the ramps.
12. Pull up the ramp supports and place the ramps upside down into the pallet.
13. Place the accessory boxes with the accessory tray in between the boxes. The narrow
box should be on top.
14. Place the cardboard sleeves around the pallet and secure with clips.
15. Place the top cover on and secure by strapping the cardboard to the pallet.

Siemens US02-101.812.01.04.02 Page 27 of 28 ACUSON Sequoia


Medical Solutions USA, Inc. 02.04 sd
28 System Delivery Procedure

ACUSON Sequoia US02-101.812.01.04.02 Page 28 of 28 Siemens


02.04 sd Medical Solutions USA, Inc.