CT-Aquilion Prime
CT-Aquilion Prime
CT-Aquilion Prime
1
REVISION RECORD
2
Contents
1. CT System specifications ................................................................................................. 6
1.2 CT specifications ...................................................................................................... 6
1.3 Software versions .................................................................................................... 6
2. Site planning ................................................................................................................... 7
2.2 Local requirements .................................................................................................. 7
2.3 Connection requirements ........................................................................................ 7
3. Radiation ...................................................................................................................... 10
4. Pre-arrival network information ................................................................................... 12
4.1 Decision flow for network connections: ................................................................. 13
4.2 Network / DICOM information .............................................................................. 14
4.3 Simplified network diagram ................................................................................... 15
5. Connecting and Power ON procedures ......................................................................... 16
5.1 Power .................................................................................................................... 16
5.2 Network ................................................................................................................ 16
5.3 Power ON procedures: .......................................................................................... 16
6. Checks / Configuration after Power ON for Toshiba Engineer. ...................................... 20
6.1 CT System: Scan and Display Console..................................................................... 21
6.2 Vitrea .................................................................................................................... 23
7. Connection to the Hospital network (NSA-2400). ......................................................... 23
7.1 Configuration of the NSA-2400 firewall: ................................................................ 23
7.2 Configuration of the DICOM setting: ...................................................................... 23
7.3 InnerVision: ........................................................................................................... 24
8. Checklist for Power ON and system configuration ........................................................ 25
9. Clean up before Power OFF for Toshiba Engineer ......................................................... 27
9.1 Copy the patient data to a USB hard disk and hand it over to the customer. ......... 27
9.2 Removal of existing patient data. .......................................................................... 27
9.3 Remove the DICOM configuration if the Mobile CT was connected to the Hospital network and
clean up the hosts file. .................................................................................................... 27
10. Checklist for Power OFF before the Mobile CT moves out. ........................................ 28
11. Appendix A: Simplified Network Diagram .................................................................. 30
12. Appendix B: CT system: Deleting All Patient Data. ..................................................... 30
13. Appendix C: Vitrea & Store Direct: Delete all patient data. ........................................ 33
14. TMSE Contact list ...................................................................................................... 33
3
Fairford Medical Mobile Solutions
Aquilion PRIME
Software version V6.0SP0605
TSX-303A/BK AquilionPRIME 300kg/2000mm couch snr. BKA1532145
MCS-7178D/4A Tube Megacool 0.35S snr. 21298-M5
CETF-006B/1A CT Transformer snr. 1AA1532538
Options included:
COT-30D/0Z Dicom3 Storage SCP snr. 0ZA1570010
COT-32D/2B Dicom MWM SCU snr. 2BA1535646
COT-33D/0Z Dicom3 MPPS SCU snr. 0ZA1570011
COT-34D/0Z Dicom QR/SCP snr. 0ZA1570012
COT-35D/2B Dicom QR/SCU snr. 2BA1534056
COT-41D/0Z Dicom Storage Comm. snr. 0ZA1570013
CGS-62A/1B Second Console snr. 1BA1482065
CXGS-011A/3B 72kW Generator Upgrade snr. 3BA1522306
CHEG-005B/1B Gated Scan System snr. 1BA1532714
CHVH-001A/2B Variable Helical Pitch snr. 2BA1523121
TIV-2013T1/EN Innervision snr. 0
CDP-07A/2B Dental Display System snr. 2BA1522746
CVV-001A/2B Vessel View snr. 2BA1522572
4
Vitrea Workstation
Z820 Vitrea Extend computer snr. CZC50834J8
CT Abdominal Analysis Demo License
CT Vascular Aorta Demo License
CT Cardiac Analysis Demo License
CT Circle of Willis Demo License
CT Carotid Demo License
CT Larynx Airway Demo License
CT Musculoskeletal Analysis Demo License
CT Renal Demo License
CT Runoff Demo License
Veralcolon CAD-VE iCAD VeraLook CT Colon CAD Demo License
MR Abdominal Analysis Demo License
MR Brain Tumor Demo License
MR Musculoskeletal Analysis Demo License
MR Vascular Demo License
Injector
Medrad® Stellant® Dual Syringe CT injection system SCT210 snr. 101061
Incubator
Heratherm IMC18 18 ltr (17-40°C) Incubator snr. IMC40476659
Pat-Slide
For easy patient transfer
Please contact Luis Cervino or David Doherty for further information about availability and pricing:
5
1. CT System specifications
1.2 CT specifications
- Gantry bore opening: 780mm
- Max. Table load (max. weight of the patient): 300Kg.
- Length of the table: 2000mm
- Max. Scan range: 2000mm
6
2. Site planning
2.1 Local requirements
Area for the trailer:
- 16m x 6.0m
- For walking patients the stair entrance will be at the backside of the trailer. This stairs are
mounted to the trailer and needs a clearance of approx. 3 m. (This is included in the 18m)
- Also a clearance is required for maneuvering the stretcher towards the lift.
- In front of the trailer a clearance is required for attaching the truck to the trailer.
Ground under the trailer:
- A solid and flat surface is required with a loading capacity of 29.000kg
- If the surface is unstable reinforcement is required by additional concrete or road pavement,
etc.
- The surface should be flat and level for installing the trailer in such a way to ensure no accidents
or problems will occur.
- In case of doubt please contact Toshiba Medical System B.V.
WARNING:
It’s the user’s responsibility to verify that the shore power is electrically compatible with the
connector of the mobile trailer. Connecting the power to the wrong type of receptacle could cause
serious injury or damage of the medical system.
- The system is equipped with one CAT 5 socket that can be used for the Hospital network.
- The system is equipped with one CAT 5 socket that can be used for the general network (for a laptop
on the control desk).
- The system is equipped with two fiber optics sockets that can be used for the Hospital network.
- The system has two data lines for additional phone. This phone is not delivered. The connection is
under the control desk.
- Above connections can be found near the power cable connection on the right side underbelly
compartment.
8
- Please check the total length of network cables, some cases demand an additional switch if the
distance becomes too long.
- An inlet (CAT 5 or higher) at the hospital side has been provided near the power point.
- All above arrangement according to the installation of network connection are to be done by the
client.
- In case a standard connection is not allowed the attached form will supply information concerning
other solutions.
The mobile CT can operate standalone, no connection to the Hospital network is necessary.
In this case:
- The image data is stored on the Vitrea
Optional it’s possible to connect the Mobile CT network to the Hospital network.
- The firewall in the Mobile CT is set to DHCP (will try to get an IP address).
- The CT can use the Hospital Work List Server (DICOM MWM).
- The CT can send images to the Hospital PACS (DICOM STORE).
- The CT can print on a Hospital Imager.
- A network cable or fiber optics connection is available on the outside of the trailer.
9
3. Radiation
In The Netherlands the use of equipment emitting ionizing radiation is regulated by the Nuclear
Energy Act (Kernenergiewet). The law provides a framework. Actual measures to be taken are in the
form of decrees, general administrative measures and ministerial decisions.
The decree on Radiation Protection (BS Besluit Stralingsbescherming), regulates the protection of
workers and the general public from risks of exposure for all work involving radiation.
The decree requires a license for using ionizing radiation emitting equipment. The mobile CT-
scanner is part of the license issued by the Ministry and licensed to Toshiba Medical Systems
Europe.
Identification and evaluation of risks of actions during which exposure to ionizing radiation may
occur are identified an evaluated. A risk inventory and evaluation (RI&E) report for the mobile CT-
scanner can be made available for the license-holder of the hospital.
Toshiba Medical Systems Europe Radiation license number: 2008/2220-08 as published in the Dutch
th
Staatscourant on December 12 2008.
Radiation officer during renting period, within respect to the local requirements and authorities.
10
Lead shielding
11
4. Pre-arrival network information
Network information:
The mobile CT can operate standalone, no connection to the Hospital network is necessary.
In this case:
- The image data can be transferred to the Vitrea
Optional it’s possible to connect the Mobile CT network to the Hospital network.
12
4.1 Decision flow for network connections:
13
4.2 Network / DICOM information
2 Vitrea information
IP address IP address of Firewall WAN port
DICOM AET VIMS_AE
15
5. Connecting and Power ON procedures
Please complete the checklist “Mobile CT Checklist Connecting and
Power ON”. File the completed checklist in the Logbook binder.
5.1 Power
- Connect the outside power cable to the host power outlet of the hospital as described in 2.3
5.2 Network
- Connect the Hospital network cable(s) if required.
Port 1/2: To connect the Mobile CT network to the Hospital network. - When the Hospital PACS,
Work List Server or Imager should be used.
Port : To connect the mobile CT via fiber optics connection to the hospital network.
16
B. CT Gantry Disengagement Tilt lock
- Remove the Tilt / Rotation lock by removing the four Hex screws.
- Mount the cover plate over the hatch where the transport lock was mounted, and mount the
other covers onto the CT system.
Transport lock
-Remove the tilt brackets on both, front and rear of the gantry.
17
C. Power on the System Transformer
- Switch on the System transformer (behind the CT Gantry), in the sequence below
D. Power on CT Gantry
- Switch on the Gantry Power Switches (SW 110) at the rear of the gantry.
18
E. Power on the CT system
- Switch on the CT system (switch on the CPU box under the desk in the scan room)
G. Vitrea
- The PC is in the small right bottom cabinet under the desk in the control room.
- Connect the Vitrea keyboard, mouse and switch on the monitor.
- Press the Vitrea power button (system on the left side in the cabinet).
19
NEW PICTURE
20
6. Checks / Configuration after Power ON for Toshiba Engineer.
In this document you find the procedures for checks and configuration which must be performed
when the Mobile CT arrives at a new Hospital.
Configure the Scan and Display Console if the Hospital’s PACS, Work List or Imager is used.
- Scan Console: Work List Management (DICOM MWM).
- Display Console: PACS (DICOM STORE) and/or Print.
AE Title Port
Scan Console DCM_CTCMW_V3.00 2700
Display Console DCM_CTCMW_V3.00 2700
Vitrea VIMS_AE 3002
22
E. Register software Slice Counter
- Engineering Utility > Slice Counter
- Register the slice counter values on
the Checklist for power ON.
6.2 Vitrea
Use procedure in Appendix D to check / remove patient data.
23
7.3 InnerVision:
Please complete the checklist “Mobile CT Checklist Connecting and Power ON”.
File the complete checklist in the Logbook binder.
24
Leave this checklist in
the Mobile CT logbook.
25
InnerVision equipment
Service Processor PC switched on.
VPN connection up and running
26
Leave this checklist in
the Mobile CT logbook.
In this document you find the procedures for cleaning up which must be performed before the
Mobile CT departs to a new Hospital.
9.1 Copy the patient data to a USB hard disk and hand it over to the customer.
- Only if the patient data was not send to a Hospital PACS.
- Toshiba engineer: buy a 2 TB USB HDD for this purpose.
9.3 Remove the DICOM configuration if the Mobile CT was connected to the
Hospital network and clean up the hosts file.
- Remove the Hospital IP addresses from the hosts file :
- Applicable systems: Scan- and Display Console.
- Also for Vitrea, if it was sending data to a Hospital PACS.
Please complete the checklist “Mobile CT Checklist Connecting and Power OFF”.
File the complete checklist in the Logbook binder.
Continue with “Mobile CT Disconnect and Power OFF” for the physical power off
procedures.
27
10. Checklist for Power OFF before the Mobile CT moves out.
Use this checklist when performing the
“Mobile CT Toshiba Configuration” procedures
Name Engineer
Date
Hospital
CT System
Tilt lock engaged
Patient couch moved down
All accessories
Scan Console:
All Image & Raw data removed
Hospital DICOM settings removed
Hospital IP addresses cleared (etc/hosts file)
Write down software slice counts …………cnts
Write down hardware slice counts …………cnts
Display Console
All Image data removed
Hospital DICOM settings removed
Hospital IP addresses cleared (etc/hosts file)
CT System: Shutdown
CT System switched off (CPU box under the desk)
Keyboards, Mice and microphone disconnected and stored
Gantry Power switched off (SW110 on the rear of the Gantry)
System transformer switched off
28
Vitrea
All patient data removed.
Vitrea shutdown.
Keyboard and mouse disconnected and stored.
InnerVision equipment
Service Processor PC switched off (press power button).
Mobile CT General
Power cable disconnected from Hospital Power source.
Optional: Network cable(s) disconnected.
Engineers Signature
29
11. Appendix A: Simplified Network Diagram
30
12. Appendix B: CT system: Deleting All Patient Data.
Perform this action on both PC’s simultaneously.
Press the [Ctrl] and [ESC] keys simultaneously to display the start menu and then select [Run...].
Enter cmd to open the terminal. When the terminal window is displayed, enter as follows.
31
Checkmark ONLY “Create Image Directories” and select “Init all data” and then press Apply
Press the [Ctrl] and [ESC] keys simultaneously to display the start menu and then select [Run...].
Enter cmd to open the terminal. When the terminal window is displayed, enter as follows.
32
13. Appendix C: Delete all patient data.
Vitrea:
- Log in as administrator.
- Check the contents of E:/VitreaData/Patients.
- This folder should be empty, if not… delete the folders in here.
- Perform “database rebuild” (Vi communicator).
- Start the Vitrea software (User Name: Vitrea). Check if all is normal.
Network problems