Ecoray hf-525 Plus Service Manual PDF
Ecoray hf-525 Plus Service Manual PDF
Ecoray hf-525 Plus Service Manual PDF
M-002
Service Manual
Important!
Please read this manual thoroughly and use the system correctly before using the
system. This manual desires to keep in promised place to use always again after
reading.
Accordingly, when use this product should be thoroughly read the "Radiation Safety Control Regulation"
and then compliance with it in use. This product is serious injury's danger if do not keep use guideline
or regulation therefore, read and understood this manual before in use.
This product was designed by safety level (SL) presenting in Radiation Safety Control Regulation.
However, if do not keep using method and Radiation Safety Control Regulation, will not be able to
avoid serious injury's danger from radiation. Therefore, to prevent accident from radiation, everyone who
work connected with radioactive emission must understand radiation safety control regulation and
understand product manual fulfill thorough education and formality.
Our company recognizes enough about danger of excessive exposure to operator, installation people,
and maintenance people. Therefore, our company strongly advises that operator, installation people,
maintenance people must be educated about radiation safety control regulation and should be used this
product when understand fluently about all knowledge of operation this product.
Also, our company informs that operator, installation people, and maintenance people have a
responsibility to be educated fluently about radiation safety control regulation and understood about
product use method.
Also, our company has no responsibility about injury or calamity that occurs does not keep use method
of this product and radiation safety control regulation.
Therefore, our company strongly advises that to be used this product after that person in charge of
installation, maintenance and operator should understood radiation safety control regulation of domestic
or foreign.
1 Service Manual
EcoRay HF-525 PLUS
States a direct danger that may cause death or serious injury if it is not avoided.
States an indirect or potential danger that may cause death or serious injury if it is not avoided.
State a danger that may cause slight or medium injury or may cause damage in equipment or fire
if it is not avoided.
NOTE ☞
CONTRAINDICATION!
2 Service Manual
EcoRay HF-525 PLUS
3 Service Manual
EcoRay HF-525 PLUS
1. The owner/user have a responsibility for management and maintenance of this equipment.
2. This equipment must be operated by only authorized personnel.
3. Contact authorized service engineer when repair or inspection of the inside of the equipment.
4. In general, modifications are strictly prohibited by the Regulations. Please contact our agency if it is
needed to modify the device.
5. Do not use this system in conjunction with the other machines because seriously danger could be
happened unexpectedly.
Periodic Inspection
1. Maintenance is required to maintain safety and performance of this system for a long time.
2. Maintenance must be done periodically with maintenance plan.
3. This system must be inspected from the inspection office per every 3 years.
Warranty
1. This warranty shall apply only to defects that arise within 12months of purchase.
2. This product could have a free of charge service if troubles are occurred during the warranty
period.
3. The warranty does not cover defects caused by abuse, mishandling, accidents or improper
installation.
4. The following will not be covered under the warranty.
4-1. Modifications to the product, or repairs made by unauthorized person but manufacturer or
authorized person by manufacturer.
4-2. Acts of God. Including fire, flood, storm, power or voltage surges.
4-3. Improper use or installation.
4-4. Defects that arise in conjunction with the use of other machines.
4-5. Use of improper parts or modification the circuit.
4-6. Improper use
5. In case of repaired during the warranted period, User shall advise the result of test to manufacture.
4 Service Manual
EcoRay HF-525 PLUS
Precautions in Use
When using this equipment, please observe the following precautions for the safety of the operator
and the patient.
1. Improper use of the X-ray equipment might cause the operator or the patient to be accidentally
exposed to X-ray radiation.
2. During X-ray radiation, any person other than the subject the patient should not stay in the room.
3. X-ray room must have radiation protection facility by "Radiation Safety Control Regulation"
NOTE ☞
For the safety of operator and patient, opening the X-ray generator cover, X-ray tube,
Collimator and other equipment cover are strictly prohibited. Only qualified engineer could open
the cover.
5 Service Manual
EcoRay HF-525 PLUS
Scope
1. This manual contains information of system operation, components, control keys, switches and
hazards and technical reference information.
2. This manual may not be reproduced in whole or in part without the written permission of ECORAY
CO., LTD.
3. The text of this manual is a translation based on the Korean alphabet (Hangŭl) original.
© January, 2007
EcoRay Co., Ltd.
#2-1105, IT Castle, 550-1, Gasan dong, Geumcheon gu, Seoul, Korea 153-768
Tel: +82-2082-6243 Fax: +82-2082-6247
Revision History
6 Service Manual
EcoRay HF-525 PLUS
Table of Contents
Chapter 1
1. Pre-installation ……………………………………………………………….……….…………………………………….…. 9
1.1 An Overview ……………………………………………………………………………………………………….…………………………..… 10
1.2 Responsibility of Customers ……………………………….………………………………………….………….……………….……. 10
1.3 Constructions and Function .…………………………………………….………….…...……………………………………….….…. 10
1.4 General details …………..……………….………………………….……………………………..……………………………………….…. 11
1.5 Service Numbers …………..……………………………………………...…………………………….……………………………………. 11
1.6 Product Pre-installation Data ..……….…………………………………………………………………………………………………. 11
1.7 X-ray room Requirements ……...……………………………..………………………………………………………………….………. 14
Chapter 2
2. Installation …………………….…………………………………..…………………………………………………………… 18
2.1 An Overview ……………………………………………………………………………………………………………………………….………. 19
2.2 Pre-installation checks points and Tools and Testers ………………………………………..….…………………………... 19
2.3 Pre-installation Check ………………………………..……………..…………………………………………..…….……………………… 19
2.4 Installation of Generator ……….………………………………………………………..……………………….….……………………... 21
2.5 Installation of Tube Stand ………………………………………………………………………………….……….……………………... 28
2.6 Table Bucky ………………………………………………………………………………………………………………….……………………... 34
2.7 Wall Bucky ………………………………………………………………………………….………………………………………..……………... 38
2.8 System Cable Wiring ……………………………………………………………….………………………………………..……………... 43
Chapter 3
3. Technical Information …………….………………………………..………………….……………….……………..……. 44
3.1 An Overview …………………………………………………………………………………….……………………………..…………..……… 45
3.2 Installation information ……………………………………………………….…………………………..…………….………...…..…… 45
3.3. Exposure condition data ………………………………………………………………..……………………………..………...………… 45
3.4 Maintenance History ……………………………………………………………………………..…………………………….…..………… 46
Chapter 4
4. Calibration …………….…………………………………………….………………………….……………………………… 47
4.1 An Overview ..……………………………………………………………………………………………….…….……………………………… 48
4.2 Introduction ………………………………………….……………………………………………….…………………………………...……… 48
4.3 Needed tool and testers when Calibration …..……………………………………………………………………………..…… 48
4.4 Check point before calibration …..………………………………….………………………………………………….………...…… 48
7 Service Manual
EcoRay HF-525 PLUS
Chapter 5
5. Trouble Shooting ………………………………..……………………………..………….….………………….…………. 52
5.1 An Overview ………………………………………………….……………………………………………….……..……………..…………… 53
5.2 Error type ……………………………………………………………………………………………………………...………………..……… 53
5.3 Trouble Shooting ……………………………………………………………………………………….………....……………………..…… 58
5.4 Troubleshooting for Tube Stand (SMS-TS-2) …………………………………………………..…….………………..……..… 80
5.5 Handle Bar (SMS-TS-340 Board) …………………………….…..…………………………………..……………………...……..… 82
5.6 Troubleshooting for Table Bucky (SMS-BT-4) …………………..………………...………..………………………...……..… 84
5.7 Troubleshooting for Wall Bucky (SMS-BS-2) ………………………..………………...……….………………….…..……..… 85
Chapter 6
6. Circuit Diagram ……………………………………………………………………………….…………………………...…. 86
8 Service Manual
EcoRay HF-525 PLUS
Chapter 1
Pre-installation
9 Service Manual
EcoRay HF-525 PLUS
1. Pre-installation
1.1 An Overview
1.1.1 This manual contains operation instructions for radiological technicians, radiologists and other
medical personnel who are trained to operate radiation equipment.
1.1.2 These instructions are not suitable as a replacement for components or exchange of circuit or
medical training.
1.1.3 This system was made for trained personnel of medical examination, diagnosis and before
surgical operation and after.
1.1.4 This manual contains of HF-525 PLUS product information, its associated components, needed
environment and electrical requirements.
NOTE ☞ Complete and proper pre-installation will avoid delays and confusion.
This X-ray system is consist of Tube Stand which support X-ray tube and moveable, Console, X-ray
generator, Patient table, and X-ray tube unit.
1.3.1 Tube Stand is consist of main power supply unit, electric circuit, handle bar and mechanical
devices and could move Tube stand with floor rail.
1.3.2 Console is consist of each switch and could adjust exposure values and display it.
1.3.3 X-ray generator is consists of control unit and high voltage transformer unit and control x-ray
exposure values and make high voltages by console.
1.3.4 Patient table is consists of tabletop and bucky devices and patient will lie down on the table.
1.3.5 X-ray tube unit consist of X-ray tube, Collimator and Handle bar and could move Tube Stand
and adjust x-ray exposure area.
1.3.6 X-ray radiation will be made by pushing of X-ray switch of console or pushing of hand switch.
1.3.7 X-ray radiation will permeate patient from x-ray tube to bucky device and after permeation,
could archive x-ray images by x-ray film or digital detector.
10 Service Manual
EcoRay HF-525 PLUS
Generator
590 600 440 105
(Including H/V Transformer)
11 Service Manual
EcoRay HF-525 PLUS
2. Dimension
2. Dimension
12 Service Manual
EcoRay HF-525 PLUS
2. Dimension
Bucky Device
1. Physical characteristics
13 Service Manual
EcoRay HF-525 PLUS
2. Dimension
Bucky Device
Troubles are could be occurred in the use or storage under unsuitable environment condition.
14 Service Manual
EcoRay HF-525 PLUS
Wire size
Generator Line Voltage
7.5 m 15 m 30 m
60 kVA 208-240 VAC 35 mm2 70 mm2 120 mm2
Cable 4 Cable 5
X-ray tube Handle bar Tube Stand
Cable 3 Cable 6
Table Bucky
Main Power
Cable 9 Cable 7
Wall Bucky
Cable 1
Cable 8 Ion Chamber
(Optional)
H/V Transformer Generator
Cable 2
Console
15 Service Manual
EcoRay HF-525 PLUS
Cable 3 1 9,000mm Power, Tube Stator line Connect generator with handle bar
Cable 8 Variable 15,000mm Signal Connectable Table Bucky with Wall Bucky
Cable 9 1 pair Depends on room High Voltage Cable Connect H/V transformer with X-ray tube
NOTE ☞
Refer to the wire size. Use recommended ground wire size by local regulation. And ground
terminal is located in beside of generator main fuse.
16 Service Manual
EcoRay HF-525 PLUS
17 Service Manual
EcoRay HF-525 PLUS
Chapter 2
Installation
18 Service Manual
EcoRay HF-525 PLUS
2. Installation
2.1 An Overview
2.1.1 This installation depends on generator‟s type and configuration and must be installed by
instruction of this manual.
Only qualified (must trained by EcoRay) engineer could install this X-ray system!
2.3.1 Prior to beginning installation, it is recommended to inspect the site and verify that the X-Ray
room complies with Pre-installation requirements, such as:
1. Incoming line of X-ray room (Voltage etc.)
2. Main circuit breaker
3. Conduits.
19 Service Manual
EcoRay HF-525 PLUS
Circuit Breaker
Check voltage
Generator
20 Service Manual
EcoRay HF-525 PLUS
2. Generator
Membrane Switches
External terminal
Console PCB
21 Service Manual
EcoRay HF-525 PLUS
(2) Generator
Cue-XR2-FnD Board
Communication Terminal
Socket Receptacle
IGBT Block
Cue-XR2-Mio Board
Cue-XR2-CnD Board
2. Function of Main Parts
(1) Console
● Membrane Switches: Select and control functions.
● Console PCB: Control each function after receive signal from Membrane switches.
● External terminal: Connect Console PCB with Generator Main CPU.
(2) Generator
● Communication Cable: Signal cable which connects Console PCB with Generator Main CPU.
● Cue-XR2-FnD Board: Control PCB for driver circuit of IGBT and Tube Filament.
● Cue-XR2-CnD Board: Control PCB for charging and discharging condition of Condenser.
● Cue-XR2- Mio Board: Main PCB assembled Microprocessor.
● Emergency Stop switch: Power off the electric power quickly in emergency situation.
● IGBT Block: Semiconductor switch unit for high voltage to control input and output power to
be supplied High Voltage Transformer.
● High Voltage Transformer: It is a kind of transformer that supply high voltage to the X-ray
tube after received voltage from generator and boost voltage with
maximum 150,000V.
● Socket Receptacle: Connecting socket for High Voltage Cable
2.4.3. Unpacking
1. X-ray generator was packed with one carton box to move and install it efficiently.
2. Check X-ray generator external condition (damaged or not) after unpacking.
3. If generator was damaged, stop installation and notify this fact to carrier and distributor
immediately.
4. One X-ray generator, one communication cable and one hand switch were packed in inside of
carton box.
5. After unpacking, have to check all parts are exactly same as like order and do not dump any of
22 Service Manual
EcoRay HF-525 PLUS
Box Cover
Generator Body
Wooden Pallet
Packed image
Two adults are required to remove generator from the wooden pallet.
23 Service Manual
EcoRay HF-525 PLUS
Generator
Console
Picture 1
Tube Stand
Table Bucky
Wall Bucky
24 Service Manual
EcoRay HF-525 PLUS
Picture 2
The terminal pins of the high voltage cable are extremely delicate and easily damaged. And
then, handle them with very carefully.
25 Service Manual
EcoRay HF-525 PLUS
High voltage cable wiring must be done in dryness place without humidity.
NOTE ☞
26 Service Manual
EcoRay HF-525 PLUS
X-ray tube Anode Stator wiring is must correctly with recommended by X-ray tube
manufacturer. Never connect terminal No.1 to No. 5 and No. 6.
The terminal pins of the high voltage cable are extremely delicate and easily damaged. And
then, handle them with very carefully.
27 Service Manual
EcoRay HF-525 PLUS
8. Specification
28 Service Manual
EcoRay HF-525 PLUS
It is necessary to have 2 or more people to move and install the Tube Stand devices.
Box 2
Box 1
Floor rail
Tube Stand
Note ☞
The Stand base is fixed in the box, it should
29 Service Manual
EcoRay HF-525 PLUS
30 Service Manual
EcoRay HF-525 PLUS
SMS-TS-340 Board of
PIN
P6 Connector
NO
(Handle bar of Tube Stand)
1 RC
2 RM
3 RS
4 AC 0V
5 AC 24V
6 EMG1
7 EMG2
8 S-T1
9 S-T2
10 GND
1 DC0V NC
2 DC24V NC
3 AC0V 4 AC 0V
4 AC24V 5 AC 24V
5 AC110V NC
6 SC (RC) 1 RC
7 SS (RS) 2 RM
8 SM (RM) 3 RS
9 EMG 6 EMG1
10 EMG 7 EMG2
11 NC NC
14 Door NC
15 Door NC
16 GND 10 GND
31 Service Manual
EcoRay HF-525 PLUS
Note ☞
4. Floor rail lie horizontal way from the wall and bottom side
1 2 3
8. Press the tube arm down to the position where you can put
32 Service Manual
EcoRay HF-525 PLUS
Balance weights
11. Holding the Arm well and remove the holding pin
12. Check weight balance of tube arm and add or remove the balance
weights if necessary.
X-ray tube
Tube arm
Handle Bar
33 Service Manual
EcoRay HF-525 PLUS
4. Verification of cables.
(1) Compare the cable configuration is matching to the circuit diagram or cable diagram.
(2) Connect PWR input to Tube Stand
(3) Main PWR ON
(4) Test the electromagnetic locker functions by controlling Handle Bar.
5. Cabling between Tube Stand and HFG
(1) Power line connection
→ The PWR cables are coming from HFG and it is necessary to measure input PWR level before
the actual PWR engages to Tube Stand.
→ The PWR level is 24VDC and should be measure by a multi-meter to make it sure.
Bucky device
34 Service Manual
EcoRay HF-525 PLUS
Cassette Tray
Cassette Tray Handle
35 Service Manual
EcoRay HF-525 PLUS
1 AC 0V
2 AC 24V
3 AC 110V
4 NC
5 Bucky COM
6 Bucky24V
7 Bucky 110V
8 Bucky Signal
9 Bucky Signal
10 GND
TCON2 Connector of
PIN
TABLE BUCKY Board
NO
(Bottom side of Table Bucky)
1 Bucky24V
2 (There is no polarity)
3 AC24V
4 (There is no polarity)
36 Service Manual
EcoRay HF-525 PLUS
1 AC0V 1
AC 24V
2 AC24V 2
3 AC110V NC
4 NC NC
5 Bucky COM 3
Ground: Connect directly to Table Bucky body BK_IN
6 Bucky24V 4
7 Bucky 110V NC
8 Bucky Signal NC
9 Bucky Signal NC
Ground it directly to
10 GND
Bucky Table body
3. Start wiring
5. Make sure that patient table and tube stand are even level with
37 Service Manual
EcoRay HF-525 PLUS
1 2 3
1: Make holes after position confirmation.
2: Put the fixture bolt into the hole and do not fall out.
3: Check vertical and horizontal statues of the table with fixture bolt on the hole of floor
Rail, and fix them finally.
38 Service Manual
EcoRay HF-525 PLUS
Bucky device
Anchor
2. Bucky device
Protection cover
Cassette Tray
Cassette Tray handle
39 Service Manual
EcoRay HF-525 PLUS
At least, more than two person needs to move this units for installation.
1 DC 0V 1 DC 0V
①
2 DC 24V 2 DC 24V
3 Bucky C 3 AC 0V
⑤
4 Bucky 24V 4 AC 24V
6 Bucky24V
7 Bucky 110V
Ground: Connect directly to Table Bucky body
8 Bucky Signal
9 Bucky Signal
10 GND
40 Service Manual
EcoRay HF-525 PLUS
SMS-GC-340 for
Cable PIN No of
Wall Bucky
Wall Bucky
(Upper side of Generator)
1 DC 0V 1 AC 0V
2 DC 24V 2 AC 24V
3 AC 0V NC
4 AC 24V NC
7 Bucky 110V NC
8 Bucky Signal NC
9 Bucky Signal NC
10 GND 5 GND
Fixing
floor
7. Move tube support device left and right and Wall Bucky not to move
41 Service Manual
EcoRay HF-525 PLUS
8. If the collimator light is not out of center of Wall Bucky, fix Wall Bucky to
floor
1 2 3
Pic1: Decide where to fix and make hole on the floor to fix
Pic2: Insert bolt to the hole to fix
Pic3: Insert bolt to floor rail „s hole and fix it with nut after checking vertical and
horizontal condition
42 Service Manual
EcoRay HF-525 PLUS
①
Collimator
X-ray tube
SMS-TS-340 Board
Not Use
Inside of SMS-TS-2
43 Service Manual
EcoRay HF-525 PLUS
Chapter 3
Technical Information
44 Service Manual
EcoRay HF-525 PLUS
3. Technical Information
3.1 An Overview
This technical information contains that maintenance history and calibration data.
Hospital Name
Model X-Ray
Model
Generator
Tube
S/N. S/N.
1 40 50 1 40 150
2 40 100 2 40 500
45 Service Manual
EcoRay HF-525 PLUS
3.4.1 Enter the following information after each data modification in this book or Periodic
Maintenance Service
46 Service Manual
EcoRay HF-525 PLUS
Chapter 4
Calibration
47 Service Manual
EcoRay HF-525 PLUS
4. Calibration
4.1 An Overview
This manual contains that adjustment method of exposure condition and how to calibrate exposure
data. Calibration data will be memorized at memory chip at the Cue-XR2-Mio Board with digital
value.
4.2 Introduction
4.2.1 This calibration tool is specified program supplied by Sehwa Medical System and it will be
operated at the personal computer.
4.2.2 Control exposure value (Ex. kV, mA, etc)
4.2.3 This calibration tool must be operated by qualified engineer by EcoRay.
4.2.4 This calibration tool was designed by many experience data of manufacturer and this tool will
be operated by specified program supplied manufacturer.
48 Service Manual
EcoRay HF-525 PLUS
2. Calibration mode
1. Controller mode
5. Communication state display
3. X-ray Tube
selection window
4. Bucky selection window
49 Service Manual
EcoRay HF-525 PLUS
12. After running, could see mA graphic and could adjust it.
50 Service Manual
EcoRay HF-525 PLUS
51 Service Manual
EcoRay HF-525 PLUS
Chapter 5
Trouble Shooting
52 Service Manual
EcoRay HF-525 PLUS
5. Trouble Shooting
5.1 An Overview
5.1.1 This generator has a self-diagnosing function and display error code on the console display if
system has an error.
5.2 Error type
5.2.1 Error type
No Error type Cause
1. Charging to Main Capacity is not normal.
Charge Circuit
1 2. Input voltage is out of boundary. (Too high or too low)
problem
(Standard input voltage: Single phase, 220VAC ±10%)
Tube stator 1. Cable connection to Tube stator is wrong.
2
Circuit problem 2. Cue-XR2-MiO in Stator circuit is out of order.
1. Tube Filament circuit connection is wrong.
Tube Filament
3 2. Filament circuit is out of order.
Circuit problem
3. Tube Filament is short.
HV-DRV-OLP (Cue-XR2-FnD) signal or
Over current to IGBT Driver IC may occur
circuit is wrong.
the error.
(See circuit diagram)
IGBT-OCP (Cue-XR2-FnD) signal or Over current to IGBT Module may occur the
Generator
circuit is wrong. (See circuit diagram) error
4 Protection Circuit
DRV-PWR-UVP (Cue-XR2-FnD) signal
problem Voltage from the IGBT Driver IC is lower
or circuit is wrong.
than standard.
(See circuit diagram)
FIL-OCP (Cue-XR2-FnD) signal or
Over current on the Tube Filament
circuit is wrong. (See circuit diagram)
Communication 1. Wrong connection of communication cable.
5
program 2. Communication circuit in the Main board or Console board is out of order.
1. Wrong connection of communication cable.
6 AEC problem 2. Wrong cable connection between AEC Board and Ion Chamber.
3. AEC Board or Ion Chamber is out of order.
Irregular Exposure 1. Actual exposure time is shorter than set time.
7
Time problem 2. Exposure switch is bad condition.
53 Service Manual
EcoRay HF-525 PLUS
54 Service Manual
EcoRay HF-525 PLUS
55 Service Manual
EcoRay HF-525 PLUS
12. Error type 12: Displayed E r r roto at kVp, mA, Time section
14. Error type 14: Displayed E r r FILA at kVp, mA, Time section
15. Error type 15: Displayed E r r 232 at kVp, mA, Time section
56 Service Manual
EcoRay HF-525 PLUS
16. Error type 16: Displayed E r r AEC at kVp, mA, Time section
17. Error type 17: Displayed E r r SEC at kVp, mA, Time section
18. Error type 18: Displayed E r r Pro----O at kVp, mA, Time section
19. Error type 19: Displayed E r r Pro--I-- at kVp, mA, Time section
20. Error type 20: Displayed E r r Pro---P- at kVp, mA, Time section
57 Service Manual
EcoRay HF-525 PLUS
21. Error type 21: Displayed E r r Pro----F at kVp, mA, Time section
22. Error type 22: Displayed E r r Inte at kVp, mA, Time section
1. Inter Lock signal trouble and circuit trouble
2. Inter Lock is Door Open, Tube Temperature, Bucky start signal, Emergency Stop switch.
58 Service Manual
EcoRay HF-525 PLUS
2. Check Procedures 1
(1) Board Name: Cue-XR2-CnD
(2) Functions: This board controls charge/discharge of Main Condenser.
(3) Followings are the check procedures for Cue-XR2-CnD board when charge/discharge trouble.
4. Resistor(R806) 을 점검.
Console Power ON.
59 Service Manual
EcoRay HF-525 PLUS
60 Service Manual
EcoRay HF-525 PLUS
ET: Earth
61 Service Manual
EcoRay HF-525 PLUS
3. Check procedure
(1) Board Name: Cue-XR2-Mio Board
(2) Functions: Controlling generator including Tube Stator movement.
(3) Under the tube stator trouble, check tube stator wiring condition and then check Cue-XR2-
Mio Board and check points are as follows;
62 Service Manual
EcoRay HF-525 PLUS
63 Service Manual
EcoRay HF-525 PLUS
3. Check procedure
(1) Board Name: Cue-XR2-FnDBoard
(2) Functions: Controlling Tube Filament
(3) Under Tube Filament circuit trouble, check high voltage and wiring condition of circuit and
then check Cue-XR2-FnD board.
Turn on the Main power and 1. Check Connector (P501) and measure
the voltage on it ; 220VAC is normal
then turn on the Console Power
2. Check Fuse (F501).
3. Fix the Relay (RL501) position and check
LED (D516, 517, 518)
its functioning.
Cue-XR2-FnD (Driver Board) Not lighting
Normal: Relay working and Ready lamp ON
Normal: All LED lighted
LED (D516, 517, 518) 4. Measure the voltage between Heat sink
with green
점등상태확인 (U406) and Test point (M5V) Nomal: 5VDC
±0.1V
정상: LED모두 녹색점등
5. Measure the voltage between Heat sink
Set exposure values
(U406) and KA324 (U518 pin 4)
; 12VDC ±0.1V is normal
6. Measure the voltage between Heat sink
Push the “Ready” button
Error when Ready (U406) and KA324 (U518 pin 11)
; -12VDC ±0.1V is normal
7. In case above voltages are not good,
check input PWR of Cue-XR2-Mio (Main
Board).
8. Fix the Connector (P403) well.
X-ray switch could use after
64 Service Manual
EcoRay HF-525 PLUS
5.3.4 IGBT Device signal and Protection circuit trouble (Check Cue-XR2-FnD Board)
1. Consist of part
65 Service Manual
EcoRay HF-525 PLUS
2. Check procedure
(1) Board Name: Cue-XR2-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check high voltage cable, filament short condition of X-ray tube, wiring condition of circuit
when IGBT Device signal and protection circuit trouble and then check Cue-XR2-FnD Board
condition as bellows.
66 Service Manual
EcoRay HF-525 PLUS
Displayed on console
67 Service Manual
EcoRay HF-525 PLUS
educated engineers.
Push X-ray button
5. In case IGBT_OCP circuit is out of order,
68 Service Manual
EcoRay HF-525 PLUS
2. Check procedure
(1) Board Name: Cue-XR2-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of Cue-XR2-FnD Board and check IGBT_OCP trouble when IGBT
Device signal and protection circuit troubles. And check procedures are as follows.
Displayed on console
69 Service Manual
EcoRay HF-525 PLUS
5.3.8 DRV _ PWR _ UVP trouble or Error Pro - - P – error (Cue-XR2-FnD Board)
1. Consist of PCB
2. Check procedure
(1) Board Name: Cue-XR2-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of Cue-XR2-FnD Board and the check IGBT_OCP trouble when IGBT
Device signal and protection circuit trouble. And check procedures are as follows.
lamp ON)
70 Service Manual
EcoRay HF-525 PLUS
5.3.9 DRV _ PWR _ UVP trouble or Error Pro --P- error (Cue-XR2-FnD Board)
1. Consist of PCB.
2. Check procedure
(1) Board Name: Cue-XR2-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of Cue-XR2-FnD Board and then check IGBT _ OCP troubles when
IGBT Device signal and protection circuit troubles. And check procedures are as follows.
71 Service Manual
EcoRay HF-525 PLUS
Set exposure value 9. Check voltage between Heat sink (U504) and
72 Service Manual
EcoRay HF-525 PLUS
2. Check procedure
(1) Board Name: Cue-XR2-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of Cue-XR2-FnD Board and then check FIL _ OCP trouble when IGBT
Device signals and protection circuit trouble. And check procedures are as follows.
73 Service Manual
EcoRay HF-525 PLUS
5.3.11 Communication error between generator and console or Err 232 error
(Cue-XR2-Mio Boar & SMS-CONB-340 Board)
1. Consist of PCB
2. Check procedure
(1) Board Name: Cue-XR2-Mio Boar and SMS-CONB-340 Board
(2) Functions: Communication transfer between generator and console.
(3) Check communication cable condition when communication problem between console and
Cue-XR2-Mio Board. And check procedures are as follows.
74 Service Manual
EcoRay HF-525 PLUS
2. Check procedure
(1) Check Ion Chamber connector condition.
(2) Chamber position and Bucky position must be same. (Same position)
(3) Check Number flickering or not on the Time display of console after X-ray exposure with
„beep‟ sound (4 times).
(Normal: Changed normal display condition after Number flickering on the Time display)
75 Service Manual
EcoRay HF-525 PLUS
3. Consist of part 1
4. Check procedure 1
(1) Board Name: SMS-CONB-340 Board
(2) Functions: Controlling Ion Chamber
(3) Check communication cable condition and then check circuit when communication troubles
occurred between console and Cue-XR2-Mio Board and Ion Chamber. And check procedures
are as follows.
Turn ON power of console 1. Fix the Communication cable between OP console and
HFG
2. Fix the Connector (CCON7).
Self-diagnosing automatically by Err AEC comes
3. Fix the Connector (CCON6).
program on the display
4. Fix the Connector (PCON1).
5. Check LED (PD1, 2, 3)
76 Service Manual
EcoRay HF-525 PLUS
5. Consist of part 2
6. Check procedure 2
(1) Board Name: SMS-AEC Board
(2) Functions: Controlling Ion Chamber
(3) Check communication condition when communication trouble occurred between
input/output signal of Ion Chamber and circuit error. And check procedures are as follows.
Turn ON power of console 1. Fix the communication cable between OP console and
AEC board.
2. Check the Connector (CCON1).
Self-diagnosing automatically by Err AEC comes
program 3. Check Connector (CCON3, 4, 5).
on the display
4. Check the console and Connector (CON1) cable
connections.
5. Check LED (LED1, 2, 3)
System is ready to be used.
Normal: LED (LED1, 2, 3) light ON
6. Check LED (LED4)
Select AEC function 7. Measure the voltage of Capacitor (PC1) -both end
Normal: 5VDC ±0.1V
77 Service Manual
EcoRay HF-525 PLUS
7. Consist of part 3
8. Check procedure 3
(1) Board Name: Cue-XR2-Mio Board
(2) Functions: Controlling Ion Chamber
78 Service Manual
EcoRay HF-525 PLUS
79 Service Manual
EcoRay HF-525 PLUS
5.4.1 Configuration
Stand
F2
F5 F4 F3 F1
80 Service Manual
EcoRay HF-525 PLUS
Handle Bar functioning at Maintain Handle Bar 1. SMS-TS-340 Board F1, F2, fuse check
P6 Connector.
81 Service Manual
EcoRay HF-525 PLUS
5.5.1 Compositions
82 Service Manual
EcoRay HF-525 PLUS
Handle Bar functioning at Maintain Handle Bar 1. SMS-TS-340 Board F1, F2 fuse check
Board) check
P6 Connector
power check
83 Service Manual
EcoRay HF-525 PLUS
fuse (TF1)
Operate table bucky Fixture of 1. Check the fuse of TF1, TF2 of SMS-BT-4
bottom of the product)) not unlocked 2. Check rectifying diode of TD1 and TD2
of SMS-BT-4 Board
84 Service Manual
EcoRay HF-525 PLUS
Fuse (BF2)
Rectification diode
(BD2)
fsdfsdfs(
Fuse (BF1)
Rectification
(BD2)
(BD1)
diode
Operate Wall Bucky device Electro-magnet 1. Check condition of fuse SF1, SF2 on
85 Service Manual
EcoRay HF-525 PLUS
Chapter 6
Circuit Diagram
86 Service Manual
EcoRay HF-525 PLUS
6. Circuit Diagram
6.1 Block Diagram
87 Service Manual
EcoRay HF-525 PLUS
88 Service Manual
EcoRay HF-525 PLUS
89 Service Manual
EcoRay HF-525 PLUS
90 Service Manual
EcoRay HF-525 PLUS
91 Service Manual
EcoRay HF-525 PLUS
92 Service Manual
EcoRay HF-525 PLUS
93 Service Manual
EcoRay HF-525 PLUS
94 Service Manual
EcoRay HF-525 PLUS
95 Service Manual
EcoRay HF-525 PLUS
96 Service Manual
EcoRay HF-525 PLUS
97 Service Manual
EcoRay HF-525 PLUS
98 Service Manual
EcoRay HF-525 PLUS
99 Service Manual
EcoRay HF-525 PLUS
Chapter 7
7. Main Parts List
7.1 Main parts of Generator
Name Image Location Model No. Function
tube.
Capacity.
and time.
Change AC power to
Main Rectifier Inside of Generator MAIN-REC-100
DC power.
Supplying power to
high voltage
High voltage Inside of high
HV-COIL-1 transformer connected
primary coil voltage generator
with IGBT of a
generator.
assembled
High voltage Inside of high
HV-COIL-2 with high voltage
secondary coil voltage generator
transformer
primary coil.
Supplying power to
filament transformer
Filament Inside of high
HV-COIL-F1 connected with
primary coil voltage generator
Filament power board
of a generator.
Supplying power to
Tube filament
Filament Inside of high
HV-COIL-F2 assembled with
secondary coil voltage generator
filament transformer
primary coil.