PX-300HF Service Manual (111018)
PX-300HF Service Manual (111018)
PX-300HF Service Manual (111018)
Service Manual
Model: PX-300HF
Important!
Please read this manual thoroughly and use the system correctly before using the
system. This manual desires to keep in promised place to use always again after
reading.
Seoul Office: #402, Cygnus B/D, 115-12, Nonhyun-Dong, Kangnam-Ku, Seoul, Korea 135-821
Tel) +82 70 7510 3402 Fax) +82 70 8630 3420 E-mail: sales@ecoray.kr
EcoRay PX-300HF
Accordingly, when use this product should be thoroughly read the "Radiation Safety Control Regulation"
and then compliance with it in use. This product is serious injury's danger if do not keep use guideline
or regulation therefore, read and understood this manual before in use.
This product was designed by safety level (SL) presenting in Radiation Safety Control Regulation.
However, if do not keep using method and Radiation Safety Control Regulation, will not be able to
avoid serious injury's danger from radiation. Therefore, to prevent accident from radiation, everyone who
work connected with radioactive emission must understand radiation safety control regulation and
understand product manual fulfill thorough education and formality.
Our company recognizes enough about danger of excessive exposure to operator, installation people,
and maintenance people. Therefore, our company strongly advises that operator, installation people,
maintenance people must be educated about radiation safety control regulation and should be used this
product when understand fluently about all knowledge of operation this product.
Also, our company informs that operator, installation people, and maintenance people have a
responsibility to be educated fluently about radiation safety control regulation and understood about
product use method.
Also, our company has no responsibility about injury or calamity that occurs does not keep use method
of this product and radiation safety control regulation.
Therefore, our company strongly advises that to be used this product after that person in charge of
installation, maintenance and operator should understood radiation safety control regulation of domestic
or foreign.
States a direct danger that may cause death or serious injury if it is not avoided.
States an indirect or potential danger that may cause death or serious injury if it is not
avoided.
State a danger that may cause slight or medium injury or may cause damage in
equipment or fire if it is not avoided.
NOTE ☞
States the information, which helps to use the system correctly.
CONTRAINDICATION!
This x-ray system is not suitable for therapeutic applications.
1. The owner/user has a responsibility for management and maintenance of this equipment.
2. This equipment must be operated by only authorized personnel.
3. Contact authorized service engineer when repair or inspection of the inside of the equipment.
4. In general, modifications are strictly prohibited by the Regulations. Please contact our agency if it is
needed to modify the device.
5. Do not use this system in conjunction with the other machines because seriously danger could be
happened unexpectedly.
Periodic Inspection
1. Maintenance is required to maintain safety and performance of this system for a long time.
2. Maintenance must be done periodically with maintenance plan.
3. This system must be inspected from the inspection office per every 3 years.
Warranty
1. This warranty shall apply only to defects that arise within 12months of purchase.
2. This product could have a free of charge service if troubles are occurred during the warranty
period.
3. The warranty does not cover defects caused by abuse, mishandling, accidents or improper
installation.
4. The following will not be covered under the warranty.
4-1. Modifications to the product, or repairs made by unauthorized person but manufacturer or
authorized person by manufacturer.
4-2. Acts of God. Including fire, flood, storm, power or voltage surges.
4-3. Improper use or installation.
4-4. Defects that arise in conjunction with the use of other machines.
4-5. Use of improper parts or modification the circuit.
4-6. Improper use
5. In case of repaired during the warranted period, User shall advise the result of test to manufacture.
Precautions in Use
When using this equipment, please observe the following precautions for the safety of the operator
and the patient.
1. Improper use of the X-ray equipment might cause the operator or the patient to be accidentally
exposed to X-ray radiation.
2. During X-ray radiation, any person other than the subject the patient should not stay in the room.
3. X-ray room must have radiation protection facility by "Radiation Safety Control Regulation"
NOTE ☞
For the safety of operator and patient, opening the X-ray generator cover, X-ray tube,
Collimator and other equipment cover are strictly prohibited. Only qualified engineer
could open the cover.
Scope
1. This manual contains information of system operation, components, control keys, switches and
hazards and technical reference information.
2. This manual may not be reproduced in whole or in part without the written permission of EcoRay Co.,
Ltd.
3. The text of this manual is a translation based on the Korean alphabet (Hangŭl) original.
Revision History
Table of Contents
Chapter 1
1. Pre-installation ……………………………………………………………….……….………………………….……….…. 10
1.1 An Overview ……………………………………………………………………………………………………….…………………….……..… 10
1.2 Responsibility of Customers ……………………………….………………………………………….………….…………….….……. 10
1.3 Constructions and Function .…………………………………………….………….…...………………………………………..….…. 10
1.4 General details …………..……………….………………………….……………………………..…………………………………….….…. 11
1.5 Service Numbers …………..……………………………………………...…………………………….…………………………….………. 11
1.6 Product Pre-installation Data ..……….…………………………………………………………………………………………………. 12
1.7 X-ray room Requirements ……...……………………………..………………………………………………………………….………. 13
Chapter 2
2. Installation …………………….…………………………………..…………………………………………………………… 18
2.1 An Overview ……………………………………………………………………………………………………………………………….………. 18
2.2 Pre-installation checks points and Tools and Testers ………………………………………..….…………………………... 18
2.3 Pre-installation Check ………………………………..……………..…………………………………………..…….……………………… 19
2.4 Installation of Generator ……….………………………………………………………..……………………….….……………………... 20
2.5 Installation of Tube Stand ………………………………………………………………………………….……….……………………... 27
2.6 System Cable Wiring ……………………………………………………………….………………………………………..……………... 35
Chapter 3
3. Technical Information …………….………………………………..………………….……………….……………..……. 37
3.1 An Overview …………………………………………………………………………………….……………………………..…………..……… 37
3.2 Installation information ……………………………………………………….…………………………..…………….………....…..…… 37
3.3. Exposure condition data ………………………………………………………………..……………………………..………...………… 37
3.4 Maintenance History ……………………………………………………………………………..…………………………….…..………… 38
Chapter 4
4. Calibration …………….…………………………………………….………………………….……………………………… 40
4.1 An Overview ..……………………………………………………………………………………………….…….……………………………… 40
4.2 Introduction ………………………………………….……………………………………………….…………………………………...……… 40
4.3 Needed tool and testers when Calibration …..……………………………………………………………………………..…… 40
4.4 Check point before calibration …..………………………………….………………………………………………….………...…… 40
4.5 Consist of Calibration Program …..…………………………………………………………………………..……………….……..… 41
4.6 How to use Calibration program ………………………………………………………………………………….….…….…….…… 41
Chapter 5
5. Trouble Shooting ………………………………..……………………………..………….….………………….…………. 45
5.1 An Overview ………………………………………………….……………………………………………….……..……………..……….…… 45
5.2 Error type ……………………………………………………………………………………………………………...………………..……… 45
5.3 Trouble Shooting ……………………………………………………………………………………….………....……………………...…… 51
5.4 Handle Bar (SMS-TS-340 Board) …………………………….…..…………………………………..……………………...……..… 73
Chapter 6
6. Circuit Diagram ……………………………………………………………………………….…………………………...…. 75
Chapter 1
Pre-installation
Model: PX-300HF
1. Pre-installation
1.1 An Overview
1.1.1 This manual contains operation instructions for radiological technicians, radiologists and other
medical personnel who are trained to operate radiation equipment.
1.1.2 These instructions are not suitable as a replacement for components or exchange of circuit or
medical training.
1.1.3 This system was made for trained personnel of medical examination, diagnosis and before
surgical operation and after.
1.1.4 This manual contains of PX-300HF product information, its associated components, needed
environment and electrical requirements.
NOTE ☞ Complete and proper pre-installation will avoid delays and confusion.
This X-ray system is consist of Tube Stand which support X-ray tube and moveable, Console, X-ray
generator, Patient table, and X-ray tube unit.
1.3.1 Tube Stand is consist of main power supply unit, electric circuit, handle bar and mechanical
devices and could move Tube stand with floor rail.
1.3.2 Console is consist of each switch and could adjust exposure values and display it.
1.3.3 X-ray generator is consists of control unit and high voltage transformer unit and control x-ray
exposure values and make high voltages by console.
1.3.4. X-ray tube unit consist of X-ray tube, Collimator and Handle bar and could move Tube Stand
1.4.1 Manufacturer
EcoRay Co., Ltd.
621-14, Docheon-dong, Gwangsan-gu, Gwang-ju, 506-301, Korea
1.5.1 If the system does not work reliably, or if errors occur in terms of control functions which are
described in the manual, the owner should contact the nearest EcoRay Service Center.
1.5.2 The EcoRay Service Center can be reached at the following Phone and Fax numbers:
1. Service for Europe
1.6.1 Generator
1. Physical characteristics
Mobile Cart
(Including PSU, Generator, 1300 640 1980 300
Tube Stand)
Generator
(Including H/V Transformer) 590 600 440 105
(Installed inside of Mobile Cart)
2. Dimension
Note!
Console and X-ray generator, PSU was already installed inside of Mobile Cart!
1. Physical characteristics
2. Dimension
Troubles are could be occurred in the use or storage under unsuitable environment
condition.
Model PX-300HF
(3) Minimum wire size from generator and electric circuit breaker.
Wire size
Generator Line Voltage
9m
1 kVA 230 VAC 2 mm2
Cable 4
X-ray tube Handle bar Tube Stand
Cable 3
Cable 5
Main Power
Cable 1
Generator
H/V Transformer
(Mobile Cart) Cable 2
Console
Cable 3 1 4,000mm Power, Tube Stator line Connect generator with handle bar
Cable 5 1 pair 2.,500mm High Voltage Cable Connect H/V transformer with X-ray tube
NOTE ☞
Use recommended ground wire size by local regulation. And ground terminal is located
in beside of generator main fuse.
Chapter 2
Installation
Model: PX-300HF
2. Installation
2.1 An Overview
2.1.1 This installation depends on generator’s type and configuration and must be installed by
instruction of this manual.
Only qualified (must trained by EcoRay) engineer could install this X-ray system!
2.3.1 Prior to beginning installation, it is recommended to inspect the site and verify that the X-Ray
room complies with Pre-installation requirements, such as:
1. Incoming line of X-ray room (Voltage etc.)
2. Main circuit breaker
Power Cable
Connected with
230V AC
2.4.1 Appearance
1. Console
2. Generator
External terminal
Console PCB
Note!
X-ray Generator, High Voltage Transformer, PSU, Console was already installed inside
of Mobile Cart!
ECO-XR5-FnD Board
Communication Terminal
Socket Receptacle
IGBT Block
ECO-XR5-Mio Board
ECO-XR5-CnD Board
(2) Generator
● Communication Cable: Signal cable which connects Console PCB with Generator Main CPU.
● ECO-XR5-FnD Board: Control PCB for driver circuit of IGBT and Tube Filament.
● ECO-XR5-CnD Board: Control PCB for charging and discharging condition of Condenser.
● ECO-XR5- Mio Board: Main PCB assembled Microprocessor.
● Emergency Stop switch: Power off the electric power quickly in emergency situation.
● IGBT Block: Semiconductor switch unit for high voltage to control input and output power to
be supplied High Voltage Transformer.
● High Voltage Transformer: It is a kind of transformer that supply high voltage to the X-ray
tube after received voltage from generator and boost voltage with
maximum 150,000V.
● Socket Receptacle: Connecting socket for High Voltage Cable
2.4.3. Unpacking
1. X-ray generator was assembled with mobile cart to be installed easily and easy move.
2. Check X-ray generator external condition (damaged or not) after unpacking.
3. If generator was damaged, stop installation and notify this fact to carrier and distributor
immediately.
4. After unpacking, have to check all parts are exactly same as like order and do not dump any of
packing parts such as envelope, box, documents until confirmation is completely.
X-ray Generator was assembled inside of Mobile cart for easy installation!
Connect Console
Generator
Picture 1
Tube Stand
NOTE ☞
X-ray Generator, Console, Tube Stand was already assembled with Mobile Cart!
The terminal pins of the high voltage cable are extremely delicate and easily
damaged. And then, handle them with very carefully.
High voltage cable wiring must be done in dryness place without humidity.
NOTE ☞
Anode side of
X-ray tube Anode Stator wiring is must correctly with recommended by X-ray tube
manufacturer. Never connect terminal No.1 to No. 5 and No. 6.
The terminal pins of the high voltage cable are extremely delicate and easily
damaged. And then, handle them with very carefully.
6. Specification
3. Stand base
2.5.2 Unpacking
1. For easy installation, Mobile X-ray system almost assembled. And after unpacking, engineer just
connects X-ray tube, Handle, Collimator.
2. Check if all the devices are well condition after unpacked.
3. In case any of the devices are broken or missed, please contact your local agent or manufacturer
quickly.
4. They are packed as a set of Stand, a set of Handle Bar, a set of balance weight.
5. Do not scrap the packing boxes, envelops or inside papers before you make sure all about the
contents.
6. Sometime, Tube Stand separated with Mobile cart by buyer’s asking.
It is necessary to have 2 or more people to move and install the Tube Stand devices.
Box 1
Tube Stand
1. Check if the serial No. on the label is match with the serial No. on the quality certificate.
2. During the unpacking, check if all the devices are in well condition, especially at cables. In case
any broken or damaged parts, please contact your local agent for replacements.
3. Bring the devices to the installation place. (Do not assemble the devices in advanced) 2 or more
people are needed to move them because the devices are very heavy.
4. Fix (install) the Floor rail on a right position.
5. Connect Tube Stand into the Stand base.
6. Press down the tube head (tube arm) to the level where you can put the holding pin. It is
necessary to pull it down without weight balance.
7. Put the holding pins around the balance weight box position.
8. Then install X-ray tube, Handle Bar and collimator.
9. And put balance weights into the tube stand (in the empty space of balance weights). The
balance weights are very heavy, so please be careful when you move them into the space.
10. Check the cable connections and make them tight if necessary.
11. Make all the cable connections according to the cable diagram.
12. Check if all the connections are correct.
13. If the cable & wire connections are O.K, put main AC input power to the Generator. (Check AC
input power level before you make Power ON)
14. One person holds the X-ray tube and the other removes the holding pin at behind of Tube
Stand.
15. Check if the weight balance is O.K, if the balance is not perfect, add or remove the weight
balances.
16. PWR ON and check if electromagnetic lockup functions are all O.K – test it from Handle Bar.
17. The engineers should stay in the room until the installation works are all over.
3. Press the tube arm down to the position where you can put
Balance weights
1. Understanding Wiring
Tube Support part of
(1) Understanding of ECO-XR5-GC Board. PIN
ECO-XR5-GC Board
NO
(Upper side of Generator) (Upper side of Generator)
1 Stator Com (SC1)
2 Stator Shift (SS1)
3 Stator Main (SM1)
4 AC Out
5 AC 0V
6 NC
7 DC 0V
8 DC24V
9 EMG
10 EMG
11 NC
12 Tube1 Temp.
Tube Support Unit
13 Tube1 Temp.
14 Door
15 Door
16 GND
P6 Connector of
PIN
SMS-TS-340 Board
NO.
(Handle bar of Tube Stand)
1 Stator Com (SC1)
P6 Connector
2 Stator Main (SM1)
3 Stator Shift (SS1)
4 AC 0V
5 AC 24V
6 EMG
7 EMG
8 Tube1 Temp.
9 Tube1 Temp.
10 GND
ECO-XR5-GCBoard SMS-TS-340
4 AC Out 4 AC 0V
5 AC 0V 5 AC 24V
6 NC NC NC
7 DC 0V NC NC
8 DC24V NC NC
9 EMG 6 EMG
10 EMG 7 EMG
11 NC NC
14 Door NC NC
15 Door NC NC
16 GND 10 GND
SMS-TS-340
SMS-TS-340의
(5) Connect P8 connector of SMS-TS-340 with Emergency Switch (There is no electric polarity)
(6) Connect P4 connector of SMS-TS-340 with Collimator (There is no electric polarity)
(7) Connect P9 connector of SMS-TS-340 with X-ray Tube rotation electromagnetic rock.
(Take notice of electric polarity)
(8) Connect P12 connector of SMS-TS-340 with X-ray Tube rotation electromagnetic rock.
(Take notice of electric polarity)
(9) Connect P5 connector of SMS-TS-340 with SID display unit (Optional)
X-ray tube
Tube arm
Handle Bar
Collimator
X-ray tube device
3. Verification of cables.
(1) Compare the cable configuration is matching to the circuit diagram or cable diagram.
(2) Connect PWR input to Tube Stand
(3) Main PWR ON
(4) Test the electromagnetic locker functions by controlling Handle Bar.
SMS-TS-340 Board
X-ray Tube
(Handle bar of Tube Support)
NOTE!
If do not use Door and EMG functions, then never remove the attached pin of tube
support connector pin (Door and EMG). If needs these functions, use these pin connectors
Chapter 3
Technical Information
Model: PX-300HF
3. Technical Information
3.1 An Overview
This technical information contains that maintenance history and calibration data.
Hospital Name
Model X-Ray
Model
Generator
Tube
S/N. S/N.
40kV 40kV
60kV 60kV
80kV 80kV
100kV 100kV
125kV 125kV
3.4.1 Enter the following information after each data modification in this book or Periodic
Maintenance Service
Chapter 4
Calibration
Model: PX-300HF
4. Calibration
4.1 An Overview
This manual contains that adjustment method of exposure condition and how to calibrate exposure
data. Calibration data will be memorized at memory chip at the ECO-XR5-Mio Board with digital
value.
4.2 Introduction
4.2.1 This calibration tool is specified program supplied by EcoRay and it will be operated at the
personal computer.
4.2.2 Control exposure value (Ex. kV, mA, etc)
4.2.3 This calibration tool must be operated by qualified engineer by EcoRay.
4.2.4 This calibration tool was designed by many experience data of manufacturer and this tool will
be operated by specified program supplied manufacturer.
2. Calibration mode
1. Controller mode
5. Communication state display
3. X-ray Tube
selection window
4. Bucky selection window
12. After running, could see mA graphic and could adjust it.
4.7.1 In installation: After installation, this x-ray system must be inspected by radiation inspection office
even though this x-ray system was calibrated at the factory before delivery. And
must be calibrated when needs calibration.
4.7.2 In periodic inspection: This x-ray system must be inspected by radiation inspection office per
every 3 years. And must be calibrated when needs calibration.
4.7.3 In repaired or replaced: If main parts are repaired or replaced, x-ray system must be inspected by
radiation inspection office. And must be calibrated when needs calibration.
4.7.4. In re-installation or move: If re-install or install it another place then x-ray system must be
inspected by radiation inspection office. And must be calibrated
when needs calibration.
Chapter 5
Trouble Shooting
Model: PX-300HF
5. Trouble Shooting
5.1 An Overview
5.1.1 This generator has a self-diagnosing function and display error code on the console display if
system has an error.
12. Error type 12: Displayed E r r roto at kVp, mA, Time section
14. Error type 14: Displayed E r r FILA at kVp, mA, Time section
15. Error type 15: Displayed E r r 232 at kVp, mA, Time section
16. Error type 16: Displayed E r r AEC at kVp, mA, Time section
17. Error type 17: Displayed E r r SEC at kVp, mA, Time section
18. Error type 18: Displayed E r r Pro----O at kVp, mA, Time section
19. Error type 19: Displayed E r r Pro--I-- at kVp, mA, Time section
20. Error type 20: Displayed E r r Pro---P- at kVp, mA, Time section
21. Error type 21: Displayed E r r Pro----F at kVp, mA, Time section
22. Error type 22: Displayed E r r Inte at kVp, mA, Time section
1. Inter Lock signal trouble and circuit trouble
2. Inter Lock is Door Open, Tube Temperature, Bucky start signal, Emergency Stop switch.
2. Check Procedures 1
(1) Board Name: ECO-XR5-CnD
(2) Functions: This board controls charge/discharge of Main Condenser.
(3) Followings are the check procedures for ECO-XR5-CnD board when charge/discharge trouble.
ET: Earth
3. Check procedure
(1) Board Name: ECO-XR5-Mio Board
(2) Functions: Controlling generator including Tube Stator movement.
(3) Under the tube stator trouble, check tube stator wiring condition and then check ECO-XR5-
Mio Board and check points are as follows;
3. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling Tube Filament
(3) Under Tube Filament circuit trouble, check high voltage and wiring condition of circuit and
then check ECO-XR5-FnD board.
Turn on the Main power and 1. Check Connector (P501) and measure
the voltage on it ; 220VAC is normal
then turn on the Console Power
2. Check Fuse (F501).
3. Fix the Relay (RL501) position and check
LED (D516, 517, 518)
its functioning.
ECO-XR5-FnD (Driver Board) Not lighting
Normal: Relay working and Ready lamp ON
Normal: All LED lighted
LED (D516, 517, 518) 4. Measure the voltage between Heat sink
with green
(U406) and Test point (M5V) Nomal: 5VDC
±0.1V
5. Measure the voltage between Heat sink
Set exposure values
(U406) and KA324 (U518 pin 4)
; 12VDC ±0.1V is normal
6. Measure the voltage between Heat sink
Error when Ready (U406) and KA324 (U518 pin 11)
Push the “Ready” button
; -12VDC ±0.1V is normal
7. In case above voltages are not good,
check input PWR of ECO-XR5-Mio (Main
Board).
8. Fix the Connector (P403) well.
X-ray switch could use after
5.3.4 IGBT Device signal and Protection circuit trouble (Check ECO-XR5-FnD Board)
1. Consist of part
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check high voltage cable, filament short condition of X-ray tube, wiring condition of circuit
when IGBT Device signal and protection circuit trouble and then check ECO-XR5-FnD Board
condition as bellows.
Displayed on console
educated engineers.
Push X-ray button 5. In case IGBT_OCP circuit is out of order,
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and check IGBT_OCP trouble when IGBT
Device signal and protection circuit troubles. And check procedures are as follows.
Displayed on console
5.3.8 DRV _ PWR _ UVP trouble or Error Pro - - P – error (ECO-XR5-FnD Board)
1. Consist of PCB
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and the check IGBT_OCP trouble when IGBT
Device signal and protection circuit trouble. And check procedures are as follows.
lamp ON)
5.3.9 DRV _ PWR _ UVP trouble or Error Pro --P- error (ECO-XR5-FnD Board)
1. Consist of PCB.
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and then check IGBT _ OCP troubles when
IGBT Device signal and protection circuit troubles. And check procedures are as follows.
Set exposure value 9. Check voltage between Heat sink (U504) and
Push X-ray button 13. Check voltage between Heat sink (U506) and
2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and then check FIL _ OCP trouble when IGBT
Device signals and protection circuit trouble. And check procedures are as follows.
5.3.11 Communication error between generator and console or Err 232 error
(ECO-XR5-Mio Boar & SMS-CONB-340 Board)
1. Consist of PCB
2. Check procedure
(1) Board Name: ECO-XR5-Mio Boar and SMS-CONB-340 Board
(2) Functions: Communication transfer between generator and console.
(3) Check communication cable condition when communication problem between console and
ECO-XR5-Mio Board. And check procedures are as follows.
2. Check procedure
(1) Check Ion Chamber connector condition.
(2) Chamber position and Bucky position must be same. (Same position)
(3) Check Number flickering or not on the Time display of console after X-ray exposure with
‘beep’ sound (4 times).
(Normal: Changed normal display condition after Number flickering on the Time display)
3. Consist of part 1
4. Check procedure 1
(1) Board Name: SMS-CONB-340 Board
(2) Functions: Controlling Ion Chamber
(3) Check communication cable condition and then check circuit when communication troubles
occurred between console and ECO-XR5-Mio Board and Ion Chamber. And check procedures
are as follows.
Turn ON power of console 1. Fix the Communication cable between OP console and
HFG
5. Consist of part 2
6. Check procedure 2
(1) Board Name: SMS-AEC Board
(2) Functions: Controlling Ion Chamber
(3) Check communication condition when communication trouble occurred between
input/output signal of Ion Chamber and circuit error. And check procedures are as follows.
Turn ON power of console 1. Fix the communication cable between OP console and
AEC board.
2. Check the Connector (CCON1).
Self-diagnosing automatically by Err AEC comes
3. Check Connector (CCON3, 4, 5).
program on the display
4. Check the console and Connector (CON1) cable
connections.
Select AEC function 7. Measure the voltage of Capacitor (PC1) -both end
Normal: 5VDC ±0.1V
7. Consist of part 3
8. Check procedure 3
(1) Board Name: ECO-XR5-Mio Board
(2) Functions: Controlling Ion Chamber
5.4.1 Compositions
Handle Bar functioning at Maintain Handle Bar 1. SMS-TS-340 Board F1, F2 fuse check
check
P6 Connector
power check
Chapter 6
Circuit Diagram
Model: PX-300HF
6. Circuit Diagram
6.1 Block Diagram