Siemens MAMMOMAT Novation: Quality Control
Siemens MAMMOMAT Novation: Quality Control
Siemens MAMMOMAT Novation: Quality Control
W or Mo target
Mo or Rh
(tilted) filter
Effective X-ray Energy
Target Mo Mo Rh Mo Rh W
Filter Mo Rh Rh Al Al Rh
kVp
22 14.1
24 14.6 15.4 14.7
26 14.9 15.8 15.7 17.6 17.8
28 15.4 16.2 16.2 17.8 18.5 18.4
30 15.7 16.5 16.6 18 18.9 18.6
32 16.9 18.1 19.1
on-off
on-off
Flex paddles are
optional, additional to
standard paddles
Cone or tapered
paddle for
MLO view
20 x 22 cm FOV
The measurement devices listed below must always display a valid calibration stamp to ensure
compliance with the described measurement and calibration intervals.
* Provided with every system. Testing device allows exposures to be performed without compression plate
attached. Detach compression plate and insert metal plug compression plate simulator.
4 cm acrylic attenuator
Compression plate
simulator
Grid
Quality Control Manual
MAMMOMAT NovationDR
Compression force
semi-annually
Repeat rate
quarterly
3. Start up and Login Procedure
CAUTION!
If the MAMMOMAT is started from being completely shut down (for details about complete shut down, see
Instructions for Use for MAMMOMAT NovationDR) the detector should be powered on at least 1 hour before intended use.
Otherwise, detector image quality can be affected.
The start-up time can be shortened to approximately 5 minutes if the MAMMOMAT is started from the standby mode
(for details about the standby mode, see Instructions for Use for MAMMOMAT NovationDR).
1. Press the enable button ( ) on the control panel to enable the MAMMOMAT. Allow a warm up time of approximately 5 minutes.
The internal monitoring system automatically performs a functional check of the MAMMOMAT.
A “DR” is displayed on the density display on the control panel to indicate that the
MAMMOMAT NovationDR system is operational.
NOTE!
A message about warming-up of the detector might be displayed. If the detector has been ON for more than 60 minutes, dismiss.
Otherwise, wait for the detector to warm up. Every other week, a message about calibration of the detector is displayed.
If the detector has been calibrated within the intervals described in the chapter: “Calibrating the detector” in the
MAMMOMAT NovationDR Instructions for Use, dismiss. Otherwise, perform a calibration procedure according to the same chapter.
4. Sign in to the DirectRay FFDM Acquisition Station software, see chapter “Signing in to the DirectRay FFDM Acquisition Station
software” in the MAMMOMAT NovationDR Instructions for Use.
User name: mammo
Password: mammo
Patient menu
task
cards
Test 4.1 Detector calibration
4.2 Artifact detection (field flatness)
Test 4.1 Detector calibration
4.2 Artifact detection (field flatness)
Test 4.3 Acquisition workstation monitor check
SMPTE
test
pattern
(from
the digital
image)
Light field illuminance
Make an image of
each paddle to
make sure it is not
visible at the chest
wall.
< 6.5 mm
4
Test 4.9: Detector Uniformity
< 10%
deviation
from mean
of means.
Test 4.10
AEC mode
Use Program 2 for phantom imaging (W/Rh 27-28 kVp). H and D modes.
PMB 45 (2000)
3225-3240
Patient Radiation Dose
Average glandular dose
Dg (mrad) = exp x HVL x 1/T x 2.1
exp in mR
HVL in mm Al
Thickness (T) in cm
100 mrad = 1 mgrey (mG)
AUTO mode
Sensor posn. 2
SNR =
meanb – 50
SDb
mAs
exp.
<pix>
SNR
Test 4.15: SNR and CNR Measurements
SNR > 40
SNR, CNR dev within + 15%
of baseline value
SNR = meanb – 50
SDb
CNR =
meanb – meanm
SDb
Test 4.16: AEC Thickness Tracking Test
Place flush
Test 4.17: Printer check
• Ruler
• Luminance meter
• Illuminance meter
• Magnification lens
• Test images (installed on Mammotest Plus)
• Multi: TG18-QC
• glare: TG18-GQ, -GQB, -GQN
• lumin: TG18-LN12-01, -LN12-18, -UNL10, -UNL80
• noise: TG18-AFC
Mammo Report QC Tests
1. Reflection
2. Overall evaluation
3. Geometric distortion (CRT only
4. Luminance response
5. Luminance uniformity
6. Resolution
7. Noise
Acceptance and annual testing – MP – tests 1-7
Daily – RT/Rad. – tests 1-2
Technologist performs daily (pass/fail)
Test 3: Geometric distortion
Illuminance refers to light falling on the screen.
TG18-UNL10 TG18-UNL80
Test 6: Resolution
Evaluate TG18-QC.
CX pattern.
x
x
Test 7: Noise
Evaluate TG18-AFC.