Siemens MAMMOMAT Novation: Quality Control

Download as pdf or txt
Download as pdf or txt
You are on page 1of 95

Quality Control

Siemens MAMMOMAT NovationDR

Donald Jacobson, Ph.D.


drj@mcw,edu
Asst. Adjunct Professor, Radiology 414-805-3163
Medical College of Wisconsin
Milwaukee, WI
Siemens
Opdose

W or Mo target

Mo or Rh
(tilted) filter
Effective X-ray Energy
Target Mo Mo Rh Mo Rh W
Filter Mo Rh Rh Al Al Rh

kVp
22 14.1
24 14.6 15.4 14.7
26 14.9 15.8 15.7 17.6 17.8
28 15.4 16.2 16.2 17.8 18.5 18.4
30 15.7 16.5 16.6 18 18.9 18.6
32 16.9 18.1 19.1
on-off
on-off
Flex paddles are
optional, additional to
standard paddles

Cone or tapered
paddle for
MLO view
20 x 22 cm FOV
The measurement devices listed below must always display a valid calibration stamp to ensure
compliance with the described measurement and calibration intervals.

• Mammography line pair phantom (2-10 lp/mm)


• 3 x 20 mm plexi (PMMA) phantom (10x10 cm)
• Test phantom (RMI model 156 or Nuclear Associates model 18-220)
• One collimator mounted plexi (4 cm thick) absorber
• Three plexi (PMMA) absorbers, each with a thickness of 20 mm
• Lint free non-woven cotton or gauze (100% cotton)
• Water or lukewarm diluted aqueous solution of household dishwashing liquid
• Luminance/Illuminance meter
• Densitometer
• A non-invasive digital kV meter
• Dose meter calibrated at the mammography X-ray beam energies
• Film or CR cassette 24x30 cm or larger
• Magnification table
• Compression plate (set for a 1.5 magnification)
• Compression plate simulator*
• 2 mm steel plate (FD object table size)
• 2 mm thick steel bars (approx. 3 x 10 cm)
• Six sheets of 99.9% pure aluminum
• Four coins
• Four paper clips
• Timer
• Ruler with a mm scale SMPTE pattern, AAPM TG18 test images

* Provided with every system. Testing device allows exposures to be performed without compression plate
attached. Detach compression plate and insert metal plug compression plate simulator.
4 cm acrylic attenuator

Compression plate
simulator
Grid
Quality Control Manual
MAMMOMAT NovationDR

Version 03, dated


10/05 or
Version 04 dated
01/06 or
Version 05, dated
04/07
Version 4 adds
SNR & CNR for
T weekly

Compression force
semi-annually
Repeat rate
quarterly
3. Start up and Login Procedure
CAUTION!

If the MAMMOMAT is started from being completely shut down (for details about complete shut down, see
Instructions for Use for MAMMOMAT NovationDR) the detector should be powered on at least 1 hour before intended use.
Otherwise, detector image quality can be affected.

The start-up time can be shortened to approximately 5 minutes if the MAMMOMAT is started from the standby mode
(for details about the standby mode, see Instructions for Use for MAMMOMAT NovationDR).

1. Press the enable button ( ) on the control panel to enable the MAMMOMAT. Allow a warm up time of approximately 5 minutes.
The internal monitoring system automatically performs a functional check of the MAMMOMAT.
A “DR” is displayed on the density display on the control panel to indicate that the
MAMMOMAT NovationDR system is operational.

2. Switch ON the acquisition workstation if it is off. Otherwise proceed to step # 3.

NOTE!
A message about warming-up of the detector might be displayed. If the detector has been ON for more than 60 minutes, dismiss.
Otherwise, wait for the detector to warm up. Every other week, a message about calibration of the detector is displayed.
If the detector has been calibrated within the intervals described in the chapter: “Calibrating the detector” in the
MAMMOMAT NovationDR Instructions for Use, dismiss. Otherwise, perform a calibration procedure according to the same chapter.

3. Login to the acquisition workstation’s operating system as apps-level user.

4. Sign in to the DirectRay FFDM Acquisition Station software, see chapter “Signing in to the DirectRay FFDM Acquisition Station
software” in the MAMMOMAT NovationDR Instructions for Use.
User name: mammo
Password: mammo
Patient menu
task
cards
Test 4.1 Detector calibration
4.2 Artifact detection (field flatness)
Test 4.1 Detector calibration
4.2 Artifact detection (field flatness)
Test 4.3 Acquisition workstation monitor check

SMPTE
test
pattern

5%, 95% squares


high contrast lines
visual evaluation
Test 4.4: Mechanical Inspection and Follow up

18 safety and operational checks


Test 4.5: Collimator Assessment
(from
the film)

(from
the digital
image)
Light field illuminance

> 160 lux


Test 5.2: Chest Wall Missed Tissue (Optional)
(Not optional)
Measure diameter and imaged portion of coin
this is called “α"
Test 4.6: Compression Plate Position

Make an image of
each paddle to
make sure it is not
visible at the chest
wall.
< 6.5 mm

this is called “α"


Test 4.7: Spatial Resolution
AEC, on 4 cm plexiglas > 7 lp/mm
Test 4.8: Phantom Image Quality

RMI 156 phantom Place flush


5

4
Test 4.9: Detector Uniformity
< 10%
deviation
from mean
of means.
Test 4.10

Radiation Output Rate > 800 mR/sec

Test 4.11: Tube Voltage Measurement & Reproducibility

Use non-invasive kVp meter

Place on the 2 mm steel plate

accy < 5%, CV < 2%


Test 4.12: Beam Quality (HVL)

Use spot compression plate (not mag. paddle)


to hold Al sheets (99.9% pure)

all anode-filter combinations

> 0.28 mm Al (at 28 kVp)


Test 4.13: Mean
glandular dose

< 3 mGy required


< 2 mGy desired
(for both H and
D modes)

Phantom should be centered L-R


Opdose

AEC mode

Use Program 2 for phantom imaging (W/Rh 27-28 kVp). H and D modes.
PMB 45 (2000)
3225-3240
Patient Radiation Dose
Average glandular dose
Dg (mrad) = exp x HVL x 1/T x 2.1
exp in mR
HVL in mm Al
Thickness (T) in cm
100 mrad = 1 mgrey (mG)

Jacobson, DR. Radiographic Exposure Calculator and Mammographic Dose


Calculator. Radiology, 1992; 182:578-580
Test 4.14: SNR and AEC reproducibility

Place flush, expose 5 times

AUTO mode
Sensor posn. 2

SNR =
meanb – 50
SDb
mAs

exp.

<pix>

SNR
Test 4.15: SNR and CNR Measurements

SNR > 40
SNR, CNR dev within + 15%
of baseline value

SNR = meanb – 50
SDb

CNR =
meanb – meanm
SDb
Test 4.16: AEC Thickness Tracking Test

2,4,6 cm (24,26,28 kVp)


SNR > 40
Deviation < 15% from mean of the means

Place flush
Test 4.17: Printer check

Follow Siemens or camera manufacturers instructions


Print SMPTE pattern, measure 11 optical densities
“If your values do not meet the values specified above, then please refer
to the printer manufacturer’s quality control procedure. You can create
your own baseline with the measured values that result from the initial
install of the printer that was installed according to the manufacturer’s
QC procedure.”
syngo
MammoReportPlus
syngo
MammoReportPlus
User name: mammo
Password: mammo
SCR Administration
Equipment needed:

• Ruler
• Luminance meter
• Illuminance meter
• Magnification lens
• Test images (installed on Mammotest Plus)
• Multi: TG18-QC
• glare: TG18-GQ, -GQB, -GQN
• lumin: TG18-LN12-01, -LN12-18, -UNL10, -UNL80
• noise: TG18-AFC
Mammo Report QC Tests

1. Reflection
2. Overall evaluation
3. Geometric distortion (CRT only
4. Luminance response
5. Luminance uniformity
6. Resolution
7. Noise
Acceptance and annual testing – MP – tests 1-7
Daily – RT/Rad. – tests 1-2
Technologist performs daily (pass/fail)
Test 3: Geometric distortion
Illuminance refers to light falling on the screen.

Lamb is ambient diffuse illuminance

Loff is reflected Lamb with monitor o

Luminance refers to light


emitted from the screen.
Must supply own light meter.
Test 4: Luminance response

Ambient light should be < 10 lux.

Lmin should be < 1.0 cd/m2 (TG18-LN-12-01)

Lmax should be > 250 cd/m2 (TG18-LN-12-18)

LR = Lmax / Lmin > 250

(Lmax, left –LLmax.right ) / Lmax. left. < 10%

Loff should be < 0.6 Lmin


Test 5: Luminance Uniformity
Maximum deviation: (Lmax – Lmin) / Lcenter should be < 30%.

TG18-UNL10 TG18-UNL80
Test 6: Resolution

Evaluate TG18-QC.

CX pattern.

x
x
Test 7: Noise

Evaluate TG18-AFC.

Visualization of all targets in 3 quadrants of the pattern

AFC (alternative forced choice)