ENG - 6984924503 - Rev2 (Service Manual)
ENG - 6984924503 - Rev2 (Service Manual)
ENG - 6984924503 - Rev2 (Service Manual)
0051
Service Manual
Version 31 January 2020 – Rev. 2
Service Manual - Revision history
Rev. 2
2 31.01.2020 3-2, 3-3, 4-3, Added applied parts symbol on Apollo EZ identification label
11-30, 26-2 located on the equipment (IEC60601-1 3ed+AMD1
paragraph 7.2.10).
MET approved revision.
Revised tabletop characteristic data.
Revised cell 438 default value.
Spare Parts update.
(RDM 8761, RDM 9031, RDM 9052)
Contents
1 INTRODUCTION 1-1
1.1 Icons appearing in the manual 1-1
1.2 How to contact the VILLA SISTEMI MEDICALI Technical Service 1-2
3 DESCRIPTION 3-1
3.1 ID plates 3-1
3.1.1 Identification plates "380/400V" version 3-2
3.1.2 Identification plates "415/480V" version 3-3
3.2 Description 3-4
5 PRE-
PRE - INSTALLATION 5-1
5.1 Space requirements 5-4
5.2 Electrical provisions 5-8
5.3 Adaptation to the mains voltage 5-9
5.4 Environmental conditions 5-9
6 INSTALLATION 6-1
6.1 Base positioning with beam and column 6-2
6.2 Tube supporting arm assembly 6-7
6.3 Spot Film Device assembly 6-8
6.4 Left and right hand tabletop support arm assembly 6-12
6.4.1 Anti-entrapment device assembly (only for 2 way tabletop version) 6-14
6.5 Equipment cable connection 6-15
6.6 Image intensifier installation 6-25
6.6.1 Image Intensifier (I.I.) assembly 6-26
6.7 Tube - Collimator assembly 6-28
6.8 Covers assembly 6-33
6.9 Tabletop assembly 6-39
6.9.1 2 way tabletop assembly 6-39
6.9.2 4 way tabletop assembly 6-40
6.10 Mobile patient loading step (optional) assembly 6-43
6.11 DAP camera mount 6-44
6.12 Final Checks 6-45
13 FUNCTIONALITY 13-
13 - 1
13.1 Collision 13-1
13.1.1 Safety barrier 13-5
13.1.2 Anti-entrapment system 13-6
13.2 Compressor 13-7
13.3 Collimator 13-8
13.4 Table automatic positioning 13-9
16 HARDWARE ADJUSTMENT
ADJUSTME NT 16-
16 - 1
16.1 SID speed setting (only for versions with SID 1100 1800 mm / 43.3" 59") 16-1
17 THREE-
THREE - PHASE INVERTER
INVER TER 17-
17 - 1
17.1 Inverter input/output functions 17-2
17.2 Inverter hardware setting 17-3
17.3 Inverter software parameters setting 17-3
17.3.1 Parameters with data modified by Villa Sistemi Medicali 17-4
17.3.2 Parameters that maintain the default settings 17-5
17.4 Access to parameters for writing 17-7
17.5 Procedure for restoring the default parameters 17-8
21 MAINTENANCE AND
A ND CLEANING 21-
21 - 1
21.1 Maintenance the operator can carry out 21-2
21.2 Maintenance that can be carried out by the Authorised Technician 21-3
21.3 Parts to be replaced periodically 21-4
Note
No part of this publication can be reproduced, transmitted, transcribed or translated
without the approval of VILLA SISTEMI MEDICALI S.p.A.
This manual in English is the original version.
1 INTRODUCTION
Note
This manual has been updated to reflect the state of the product with which it is
sold, to provide a proper reference while troubleshooting and repair operations,
commonly carried out by the service engineer, are being performed.
The manual might not reflect variations to the product if they do not affect how it
works.
The remote controlled table Apollo EZ is a radiological unit that can perform general
radiographic examinations. In particular, depending on its related accessories, examinations
which require the use of contrast fluids find their optimal application with this unit.
The aim of this manual is to instruct the user on the safe and effective use of the device.
The device must be used in compliance with the procedures described, and never be used for
purposes different from those herewith indicated.
Apollo EZ is an electro-medical device and can only be used under the supervision of a
physician or highly qualified personnel, with the necessary knowledge on X-ray protection.
The user is liable as concerns the legal fulfilment related to the installation and the operation of
the device.
This icon indicates a “NOTE”: please read the items marked by this icon thoroughly.
This icon indicates a “WARNING”: the items marked by this icon refer to safety
aspects of the patient and/or operator.
2 SAFETY INFORMATION
Warning
All the information provided in this chapter should be read carefully and implemented,
where appropriate, to avoid damage to equipment or injury to persons.
VILLA SISTEMI MEDICALI designs and builds the devices in compliance with the safety
requirements; furthermore it supplies all information necessary for correct use, and the
warnings related to danger associated with X-ray generating units.
Warning
Do not modify this equipment without authorization of the manufacturer.
Only authorised personnel can remove the protections to the electrical cabinet and the
equipment and gain access to live and moving parts.
2.1 Warnings
Apollo EZ is suitable for use in hospitals and the connection to the mains grid cannot be carried
out in buildings that provide "domestic" type power supplies.
Warning
To avoid the risk of electric shock, this device must be connected only to earthed
mains power supplies.
This device has not been designed to be used in environments where vapours, anaesthetic
mixtures flammable with air, or oxygen and nitrous oxide, can be detected.
Avoid the entry of water or other liquids into the device to prevent short circuits and corrosion.
Before cleaning the device, please disconnect it from the line voltage.
Before activation of the motorised parts such as tilting, the tabletop, the angulation
and the scan, make sure the patient is properly positioned and has the legs and arms
inside the perimeter of the tabletop. Use the handgrips where possible. During motorised
movements, objects that may interfere with the movements themselves should not be close to
the table.
Warning
When the system receives power, motorised movements may be activated
intentionally or accidentally.
The areas highlighted in the image below therefore pose a risk of entrapment,
between the support structure for the Spot Film Device and the support arms of the
patient tabletop, in the event that motorised movement is activated.
ENTERING OR REMAININ
REMAININGG IN THESE AREAS IS FORBIDDEN WHILE THE
DEVICE IS OPERATIONAL.
OPERATIONAL.
The following symbols attached to the structure of the table indicate the risk of
entrapment and the prohibition against entering the danger zones:
In the unlikely event that a person is present within the danger zone and a movement
is activated intentionally or accidentally, an emergency button must be activated
immediately.
immediately These buttons are coloured red to be immediately visible and may be
activated by simply pressing with force. As soon as one of the buttons is pressed, the
power to the device is cut off and all movements are immediately interrupted.
The position of the emergency buttons is illustrated in the following images:
Remember that, on the basis of local regulations, other emergency buttons may be
present in the exam room and/or in the control room.
All users must be familiar with the placement and functioning of these safety devices
before using the equipment.
In the event of loss of voltage for longer than that sustainable by the electronic control of the
table, it is possible that the examination taking place is cancelled, with the inhibition of the
tabletop, scan and angulation movements that may harm the patient. The Spot Film Device will
be initialised and if the cassette is present, the display will indicate its removal in order to
continue.
Apollo EZ cannot be installed in operating rooms.
Apollo EZ must be turned off while electrosurgical or similar devices are in use.
Keep clean, and if disinfection of the parts that may come into contact with the patient is
deemed necessary, follow the instructions in Chapter21.
In compliance with standard IEC 60601-1-2, the Apollo EZ is suitable for use in the
electromagnetic environment described below. The customer or user of the system must
ensure that it is used in said environment.
In compliance with standard IEC 60601-1-2, the Apollo EZ is suitable for use in the
electromagnetic environment specified. The purchaser or user of the Apollo EZ must guarantee
that it is used in the environment described below:
6
Occupation zone
5 (see Figure 2-2)
Note
The front view shows a 600x600x2000 mm occupation zone as required by the
applied standards. In reality, because the spot film-I.I.-tube unit has a maximum
range of 1344 mm, the total occupation zone is the one shown in Figure 2-2. This
zone takes into account the aforementioned range.
Max. occupation
zone
600x2243 mm
H = 2000 mm
Occupation zone
(see Figure 2-4)
1400 mm
B
450 mm
Occupation zone
600x600 mm
H = 1700 mm
The maximum values of stray radiations in the indicated zones are given in the table below.
These values were measured as described in the reference standard IEC 60601-1-3 and
particularly for radiological factors with a continuous emission value of 120 kV, 3 mA.
Note
Information for users in the European Community, in compliance with 2011/65/EU
Directive on the restriction of the use of certain hazardous substances in electrical
and electronic equipment.
The symbol of the crossed waste container on the equipment or packaging shows
that the product, at the end of its lifecycle, must be collected separately from other
types of waste.
The separate collection of this equipment at the end of its lifecycle is organised and
managed by the manufacturer. Users who need to dispose of this equipment should
therefore contact the manufacturer and follow the procedure adopted by the
manufacturer for the separate collection of the equipment at the end of its lifecycle.
Proper separate collection for subsequent recycling, treatment and compatible
environmental disposal of equipment helps avoid possible negative effects on the
environment and on health and encourages the reuse or recycling of the materials
that the equipment is made from.
Illegal disposal of the product by the owner of the equipment will result in
administrative sanctions, as provided for by applicable regulations.
Symbol Description
∼ A.C.
Protection earthing
Dangerous voltage
Serial number
Date of manufacture
Inherent filtration
3 DESCRIPTION
3.1 ID plates
plates
6
7 7
2 1b
1a 1b
Apollo EZ data label Apollo EZ identification label
(positioned on the electrical cabinet) (positioned on the equipment)
2
Spot Film Device
data label
3 4
"Connect TV chain only" Warning
label (inside the electrical cabinet) label
5
"Read the user instructions"
label
6 7
Sign Sign
"Warning: danger of entrapment" "Prohibition against entering the area inside
(ISO 7010:2011 W019) the support arms for the patient tabletop"
8a 8b
Sign "Warning: do not install the patient Sign "Warning: install the patient footrest on
footrest on this side of the tabletop". this side of the tabletop".
Only present in the event that there's a Only present in the event that there's a
shoulder rest (optional accessory) shoulder rest (optional accessory)
1a 1b
Apollo EZ data label Apollo EZ identification label
(positioned on the electrical cabinet) (positioned on the equipment)
1c 1d
cMETus certification label UDI label
(positioned on the electrical cabinet) (positioned on the electrical cabinet)
3
2 "Connect TV chain only"
Spot Film Device label (inside the electrical cabinet)
data label
Collegare solo catena TV
Connect TV chain only
4
Warning label
5
"Read the user instructions"
label
6 7
Sign Sign
"Warning: danger of entrapment" "Prohibition against entering the area inside
(ISO 7010:2011 W019) the support arms for the patient tabletop"
8a 8b
Sign "Warning: do not install the patient Sign "Warning: install the patient footrest on
footrest on this side of the tabletop". this side of the tabletop".
Only present in the event that there's a Only present in the event that there's a
shoulder rest (optional accessory) shoulder rest (optional accessory)
3.2 Description
Apollo EZ is a universal radiology diagnostic table that can satisfy all needs, used for
examinations of the digestive system, the skull and skeleton, urography, myelography, vascular
examinations, lungs, bronchography, and when supplemented by a digital acquisition system,
angiography.
The Apollo EZ table is available in the following configurations:
• SFD version with Spot Film Device, divisions in bands or crosses (optional)
• SFD "Dual Grid" version, double grid with no divisions, optimised for use with "Flat Panel
Static Wi-Fi"
• DRF version, where the Spot Film Device is not present and the table is equipped with
"Dynamic Flat Panel" with 1 or 2 crosses and large capture surface (43x43cm).
The basic functional concepts of the Apollo EZ table are:
• footprint as small as possible
• transverse movement (2 way) and transverse and longitudinal (4 way) Tabletop
• reduced skin-imaging receptor distance
• laminate (standard version) or micro-sandwich Tabletop with integrated accessory profiles
• reduced I.I.-film distance to avoid parallax correction (when applicable).
The complex movements and functions are controlled by the remote control console located in
an XR protected area.
The main movements of the table are replicated on the on-board control panel.
An automatic collimator limits the radiated field.
This accessory may consist of a square-rectangular limitation or circular limitation by means of
a collimator with iris in addition to the square limitation and is adjusted so that the X-ray beam
is aligned with the image receptor across the full range of movement (minimum and maximum
SID) and angulation (from -40° to +40°) within the limits set by standard IEC 60601-2-54.
In the event that an image receptor external to the Spot Film Device is used, the opening of
the collimator is set manually by the user.
The DRF version is equipped with a square-rectangular limitation collimator without iris.
To meet the needs related to the fluoroscopic mode and the acquisition of images using digital
systems, the table is arranged such to accept image intensifiers from a minimum of 9"
(nominal) maximum input range up to a maximum of 12" (nominal).
All the movement commands are "deadman".
Apollo EZ was designed and constructed in accordance with the European provisions laid down
for the attainment of the CE mark according to European Directive EC 93/42 on Medical
devices, ensuring the maximum safety for the operator and the patient.
Apollo EZ is supplied with both standard accessories and optional accessories. The following
table lists all available accessories making the distinction between standard, being those
normally supplied with the equipment, and optional accessories that need to be specifically
requested when ordering.
B
A C
D
E
Warning
• Verify that the accessories are firmly fixed before positioning the patient for the
exam.
• Verify that the patient footrest is firmly hooked to the tabletop's metal frame;
the position is correct if both release buttons are completely raised and the red
line is not visible.
If this condition isn't met, use the handles to move the patient footrest without
pressing on the buttons until it locks correctly (red line no longer visible).
Warning
In the event that a shoulder support is present (optional accessory "C"), the tabletop
will include labels that indicate on which side the patient footrest should be attached.
NEVER insert the patient footrest on the side of the tabletop that has the
prohibition label.
4 TECHNICAL
TECH NICAL DATA
General features
Type Apollo EZ
Manufacturer Villa Sistemi Medicali S.p.A.
20090 Buccinasco (MI) Italy
Device type and classification according to class I with type B applied parts
IEC 60601-1
Mechanical features
Runs
Speed
Compressor
Tabletop
Collimator
Tomography
Commands
Accessories
Environmental conditions
13x18 18x13
43x43
(*) Optional
Note
The first number of the format refers to the transverse axis of the tabletop (right-
left of patient). The size indicated on the various divisions is nominal.
In the Apollo EZ "Dual Grid" version, the divisions of the formats are not available.
Villa Sistemi Medicali 4-7
Service Manual - Technical data
Rev. 2
5x7 7x5
17x17
(*) Optional
Note
The first number of the format refers to the transverse axis of the tabletop (right-
left of patient). The size indicated on the various divisions is nominal.
In the Apollo EZ "Dual Grid" version, the divisions of the formats are not available.
0051 The CE mark certifies compliance of the Apollo EZ table to the 93/42/EEC
Directive and subsequent amendments and additions.
Besides the aforementioned standards, models 415/480V are also made in compliance with
those listed below:
CFR 21
Code Federal Regulation. Sub Chapter J
CAN/CSA-
CAN/CSA - C22.2 No 60601-
60601 - 1:14
Medical electrical equipment - Part 1: General requirements for basic safety and essential
performance
ANSI/AAMI ES60601-
ES60601 - 1:2005/A2:2010/(R)2012
Medical electrical equipment, Part 1: General Requirements for Basic Safety and Essential
Performance.
4.5 Dimensions
Dim ensions
Figure 4-1
Figure 4-2
5 PRE-
PRE -INSTALLATION
The instructions on the following pages guarantee that the installation performed will ensure
the remote-controlled Apollo EZ table will work properly.
The Manufacturer can provide any technical consultancy and assistance necessary right from
the pre-installation stage (if requested, to check the positioning of the equipment inside the X-
ray room).
The room should be prepared for the installation of the Apollo EZ table based on the choice of
fixing plate (Figure 5-1, Figure 5-2 or Figure 5-3) which is made considering that the maximum
weight of the machine (equipped with 12" I.I. and patient of 200kg / 440 lbs) is approximately
1200kg (2640 lbs).
Therefore, verify that the floor is capable of supporting the weight distributed over the entire
surface of the fixing plate.
Depending on the choice made, place the plate in the following way:
Standard floor fixing plate
815
(32")
(18.5")
470
FRONT SIDE
78
Remove the surface of the floor in an area measuring 500x850 mm (19.7"x33.5") until
reaching the concrete slab, so that the top surface of the plate is flush with the floor.
The concrete surface of the slab, under the plate, must be rough enough to ensure the
concrete to be poured adheres properly. Position the plate and carefully level it so that the
surface obtained is as level as possible. The fixing floor plate already has threaded holes
and M12 bolts for the anchorage of the base; these holes must be protected (e.g. by
inserting the fixing screws) so that they are not blocked during installation; pour the
concrete with a cement/sand ratio of 1:2.5, remove the surplus and let it dry for at least 6
hours.
1890 (74.4")
77 (3")
295 (11.6")
35 (1.4")
77 (3")
FRONT SIDE
If masonry work cannot be carried out in the radiology room, a floor top fixing plate is available
that can be requested at the time of ordering.
Position the plate and fasten it to the floor by means of the fixing holes, using cast iron
anchors (supplied) or chemical anchors (not supplied), based on the characteristics of the floor.
Note
The floor top plate is not symmetric and therefore attention to its correct position
must be paid.
2500 (98.4")
1500 (59.1")
50 (1.9")
FRONT SIDE
If the slab does not have sufficient load bearing capacity a weight distribution plate is
available on request at the time of ordering.
Note
The weight distribution plate is not symmetrical so it's important to position it in the
correct way.
Note
In order to unload the packaged main assembly easily, a forklift must be used or the
means of transport must be equipped with a ramp.
When assessing the size of the place of installation (including corridors, doors etc.) note that the
main assembly, once removed from the packaging, has the dimensions shown in Figure 5-4.
Figure 5-4
The minimum height of the installation room must be greater than 2500 mm (8.2 ft), and the
recommended operating area on the front is 1000 mm (39.4"). An area at the sides is required
indicated in Figure 5-5 and Figure 5-6.
Note
Verify that, in the radiological room, direct audio and visual communication between
operator and patient is always possible.
Otherwise, provide proper support (i.e. lead glass or similar, interphone, etc.).
= =
(*)
A
200
(7.87") C
**
3600 (11.8")
200
(7.87")
1000 (3.281 ft)
D
**
(*) Minimum distance between standard floor fixing plate and wall with
cabinets in position A (inside the Apollo EZ tilting area) = 900 mm
(35.4")
Minimum distance between standard floor fixing plate and wall with
cabinets in position B (outside the Apollo EZ tilting area) = 600 mm
(23.6")
(**) 4 way table top (optional) movement limits.
Figure 5-5a: Minimum dimensions for Apollo EZ
use with SID 1000 1500 mm (39.4" 59") without limitation of movement
= =
(*)
200 A
(7.87")
C
**
D
**
(*) Minimum distance between standard floor fixing plate and wall with cabinets in
position A (inside the Apollo EZ tilting area) = 800 mm (31.5")
Minimum distance between standard floor fixing plate and wall with cabinets in
position B (outside the Apollo EZ tilting area) = 600 mm (23.6")
(**) 4 way table top (optional) movement limits.
Figures 5-5b: Minimum dimensions for Apollo EZ
use with SID 1100 1800 mm (43.3" 70.9") without limitation of movement
The floor conduit (Figure 5-5) must have the following requirements:
• C : advised section = 20x10 cm (7.9"x3.9") / manhole - cabinet maximum distance = 9 m
(29.53 feet).
• D : advised section = 10 cm (3.9") / cabinet - console maximum distance = 14 m
(45.93 feet).
Note
For assembly entailing distances longer than those mentioned above (C and D)
electrical connection cables of a sufficient length must requested when ordering.
Warning
The electrical cabinet is fitted with two wheels so that the rear can be more easily
accessed during installation and maintenance.
= =
(23.6")
600
200
(7.87")
**
**
Note
The installation example shown in Figure 5 6 is not represented for version with SID
1100 ÷ 1800 mm (43.3" ÷ 70.9").
This is because this version allows tests normally performed on chest stand.
Note
Generally the power required for the fitting out of one complete X-ray room including
the Apollo EZ table equals 25-30kVA.
Warning
To avoid risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
Warning
The equipment can be connected in cascade, with X-ray generators complying with
requirements of IEC 60601-1 standard and approved by Villa Sistemi Medicali.
The cascade connection is used because, when the generator is powered OFF, the
table and the accessories connected to it are turned off as well. The wires of the
three-phase power supply between generator and table cabinet must have a
cross-section of at least 4.0 mm².
Check the generator Manual to make sure that the connection points are right for
the load and insulation values.
The general grounding must respond to the standards in force. If the grounding of
the equipment is not good enough, it might jeopardise the safety of the operator
and/or the malfunctioning of the electronic equipment.
6 INSTALLATION
Warning
During the mechanical installation of the equipment and the electrical connection
described below, the safety regulations in force must be respected and the
precautions described below must be taken:
• wear accident prevention clothing (boots, gloves, goggles when necessary)
• always work in conditions of utmost safety and avoid carrying out procedures
that put you or others at risk
• make sure that no electricity is supplied to the equipment unless the installation
is completed or said electricity is required for installation purposes.
• when the unit is ready to be powered up and while the installation is in the final
phases of completion, make sure the power supply is cut off by means of a
switch that is not on the equipment before accessing any of the parts that can
be a danger.
For delivery, handling and assembly purposes, the machine is broken down into units that must
be put together:
Warning
Before positioning the assembly on the anchorage plate, check that this is flat with
respect to both axes.
Note
The operations described below require the use of a transpallet.
Figure 6-1
4. Positioning the transpallet beneath the 2 side pallets, lift the base assembly and remove
the central pallets (Figure 6-2). Lower the transpallet.
Figure 6-2
5. Remove the two red brackets "F F " (Figure 6-2) anchored with the base to the side pallets
and keep the 5 wooden blocks and 4 spacers that will be used subsequently to position
the group on the ground. Place the transpallet centrally under the base assembly and lift it
to free the two side pallets.
6. Position 4 wooden blocks on the 4 corners of the base (Figure 6-3); lower the transpallet.
Reassemble the 2 red brackets "F
F " fastening them to the base with the supplied M16x20
screws.
Wooden blocks
Figure 6-3
7. Referring to Figure 6-4, lift the base assembly and reposition the blocks (2 on the feet
side and 1 centrally on the head end) so that the transpallet can be introduced
longitudinally from the head end.
Central block
Figure 6-4
8. Extract the transpallet and position it at the head end (Figure 6-5), inserting it as much as
possible below the base. Levering, lift the assembly enough to remove the wooden blocks.
Lower the transpallet.
Figure 6-5
9. Paying the utmost attention during transport, take the base assembly into the room in
correspondence with the fixing plate, aligning the base-plate fixing holes as much as
possible.
10. Reposition the 3 blocks as in Figure 6-4. Lower the base assembly and remove the
transpallet.
11. Position the remaining 2 wooden blocks between the forks of the transpallet and the red
bracket (Figure 6-6). Lift the base assembly enough to remove the head side block
replacing it with 2 wooden spacers. Gently lower the assembly until it rests on the
spacers. Repeat the operation to remove the feet side blocks, replacing them with the 2
remaining spacers.
W o o d e n b lo c k s
W o o d e n s p a c e rs
Figure 6-6
12. Repeat the steps described in the previous section, removing one spacer at a time, until
the machine rests on the anchor plate.
13. Check that the holes in the base assembly line up perfectly with those in the anchorage
plate. If it is necessary, move the base assembly to align the fixing holes, use the
transpallet.
15. Close the holes not used with the supplied white cups.
C o lu m n
s e c tio n
T u b e
s u p p o rt a rm
F ro n t
p la te
M ic r o 0 °
Figure 6-7
Warning
A plate is provided with the Spot Film Device. This must be assembled with the four
Spot Film Device support wheels and it will constitute equipment necessary for the
installation of the image intensifier (paragraph 6.6).
The Spot Film Device is fitted on the relevant adjustable supports and fitted with wheels to
make it easier to move and position.
1. Position the Spot Film Device assembly mounted on the transport pallet, near the
machine.
2. Free the four Spot Film Device supports from the safety blocks fixing them to the
transport pallet.
3. Remove the Spot Film Device from the pallet being careful to pay the utmost attention
not to damage it.
4. Remove all Spot Film Device fixing screws, the two safety rods holding the inclination
motor support on the end wall and remove the protection of the inclination potentiometer
(Figure 6-8). Remove the two cover hooks (one on each side) from the Spot Film Device
(Figure 6-9) to permit positioning.
Safety
rod
Protection of the
inclination potmeter
Safety
rod
Figure 6-8
5. Position the Spot Film Device assembly near the end wall and make sure that it is possible
to align the fixing holes on the shoulders and on the inclination motor support. Should it be
necessary to raise or lower the Spot Film Device to align the fixing holes, act on the
adjustable supports taking care not to force the Spot Film Device onto the inclination
motor's connections (Figure 6-9).
6. Once the alignment of the holes on both the fixing shoulders has been checked, it is
possible to fix, without tightening, the Spot Film Device assembly with the twenty-two
bolts supplied (eleven bolts per side).
If the Spot Film Device is imagined as a plane, it must be at right angles to the end wall
where it is fixed. Check that the two elements are at right angles to each other using a
spirit level. If it is necessary to make an adjustment to obtain the perfect right angles
between the Spot Film Device and the end wall, adjust the adjustable supports of the Spot
Film Device until the required position is obtained (Figure 6-9).
7. Insert the two centring pins into the holes (Figure 6-9), one for each side, making sure
that the threaded portion is accessible from outside and then tighten the twenty-two
screws. Reassemble the two cover hooks removed in point "4".
Centering
pin
90°
Top hook
Figure 6-9
9. Carry out the Spot Film Device wiring provided on the right-hand side of the end wall.
The wiring must be routed using the relevant cable clips to be found on the right-hand
wall of the Spot Film Device. The wires must be routed in such a way that they are
parallel to and not on top of each other (Figure 6-10).
Cable X16, fastened to a fixing plate, must be housed on the right wall of the Spot Film
Device interfacing with the flat cable (X16) pre-arranged during manufacture.
The following table is an aid in the identification of the wires and where they must be
connected (also see Figure 6-10).
X68
X14
X19 X44
X101 X16 Cable clip
Ground cable
Figure 6-10
Note
The automatic exposimeter device is an optional extra and consequently the A.E.C.
chamber cable is only present when the automatic exposimeter chamber is fitted to
the Spot Film Device.
10. Position the front panel of the Spot Film Device and fix it with the hinge on the left-hand
wall.
Connect the free end of the flat cable (X16) to the Table Side Operation board A8.
Loosen the fixing screws of the hinge on the Spot Film Device's left-hand wall; shut the
panel and do up the bolts on both sides (Figure 6-11).
X16
Figure 6-11
Note
Both the tabletop support arms are fitted with blocking bolts for the transport.
Remove these blocks only after the arms have been assembled, the relative belts
have been tensioned and before fitting the tabletop and moving the arms.
Belt aligment
reference
Arm belt
Coordination
belt
Outside support
For the assembly of the left-hand arm proceed as for the assembly of the right-hand arm,
considering that thare are not any belts.
6.4.1 Anti
Anti-- entrapment device assembly
assem bly (only for 2 way tabletop
version)
1. Assemble the anti-entrapment under the foot side arm using the 5 relevant screws
already present on the arm, positioning the fixing plates as shown in Figure 6-13: the
longer one internal to the beam frame, the smaller one external.
Fix the longer plate with the two related screws.
2. Connect the cable X1 coming from the anti-entrapment arm to the relevant connector
located under the beam frame.
3. Mount the cover and verify if the anti-entrapment device works properly.
Figure 6-13
Note
The cable chains "A" and "B" (Figure 6-14) are made so that they can be opened in
order to introduce the cables. Pay attention to the location of the individual covers,
as they are not all alike they must be correctly repositioned when the chain is closed.
Only the machine cables can be arranged at this stage to allow the power supply to
the table and assist the remaining installation phases, or the system cables can be
wired (high voltage cables, anode etc.); this is performed at the installer's discretion. If
it is decided to wire the system cables, refer to paragraph 6.7 for the installation of
the tube-collimator assembly.
Figure 6-14
Note
To easily identify the end to be attached to upper support, the following occurs: a
code is printed on the outside of the links, the upper part has a link with code
157.150.100.1, the next have code 158.150.100.1.
The end that is fastened to the lower fastening has links with code 157.150.100.1.
2. To avoid intertwining cables, the cable chains should be positioned as shown in Figure
6-14.
House the group of cables leaving chain "A" in chain"B". Some cables are slightly shorter
that others. The shorter ones will be connected to the base connectors board A11, the
longer ones will be connected in the electric cabinet of the table and generator if it has
also been decided to wire the system cables.
3. Before proceeding, the cable chain "B" must be positioned as in Figure 6-15.
4. The cable chains should not be closed at this point but some covers can be fitted to hold
the cables inside. Connect cables X20, X46, X75, X77, X88, X96, X97 and X98 to the base
connectors board A11 (Figure 6-16), passing them through socket "C" (Figure 6-15). The
base connectors board is provided with protection against dust, which should be removed
to allow the cable connections and then reassembled. Connect the ground cables 5, 7, 8, 9
to the ground bar (W2) in the base (Figure 6-17).
B C
Figure 6-15
5. Connect the ground cable No. 10 between the ground bar (W2) on the machine base unit
and the ground bar (W1) in the cabinet.
6. Connect all cables exiting from socket "C" and from the chain "B" to the equipment
cabinet: X68, X19, X14, X16, X44, X101, X20, X46, X99.5, X99.6, X74, X99.1, X99.2, X105, X4,
X8A, X70, X71, X72 (4 way tabletop – optional), X73, 10, E2, E4, E1, E3 (4 way tabletop –
optional), X8 and Ethernet as indicated in the following table (see also Figure 6-17).
7. Connect the foot control (cable X41) to the touch screen.
8. Connect the power cable to the R-S-T terminals on the insulation transformer inside the
cabinet; connect the ground cable to the ground bar (W1) in the cabinet.
Note
The power supply cable must be anchored to the cabinet frame with a wrapper to
avoid it from being accidentally removed.
9. Connect the system cables to the rest of the accessories if you have decided to proceed
with wiring during the procedures described above.
Note
To avoid insurgence of an alarm condition at switch ON it is necessary to install the
collimator (see paragraph 6.7) and connect it by cables X4 and X8A.
Spot Film Device board A10 X14 X14 Input board A2 Spot Film Device cable
Table Side Operation board A8 X16 X16 Input board A2 Table Side Operation cable
Base unit connector board A11 X20 X20 Input board A2 Machine potentiometer cable
Base unit connector board A11 X46 X46 Cabinet connector board A12 Board connection cable X46
X99.1 X48.1
Angulation connector board A13 Cabinet connector board A12 Compressor motor cable
X99.2 X48.2
Angulation connector board A13 X105 X105 Cabinet connector board A12 Board connection cable
Collimator
(RJ45) (RJ45)
(RJ4 5) HUB Ethernet cable
(camera live – if present)
Collimator ( DB9)
DB9 ) X8A CPU PCB A1 CANBus cable
Tilting motor
Tilting motor X70 Inverter box
power supply cable
Angulation motor
Angulation motor X71 Inverter box
power supply cable
Longitudinal tabletop motor
Longitudinal tabletop motor (Optional) X72 Inverter box
power supply cable
Scan motor
Scan motor X73 Inverter box
power supply cable
Base ground bar (W2) 10 10 Cabinet ground bar (W1) Ground cable No. 10
Angulation motor
Angulation motor E2 Terminal block X0.E2
grounf cable
Scan motor
Scan motor E4 Terminal block X0.E4
ground cable
HUB
Touch screen – CPU PCB A14 (RJ45) Touch screen
(with camera live)
(RJ45)
CPU PCB A1
J1
(without camera live)
X96
X20
2 1
1
1
X75
1
4
1 2
X83
3
V1
X77
1
4
1 2
X84
X92
2
1
3
X78
1
V2
4
1 2
X88
X81
X93
3
1
2
2
1
X46
X97
X98
1
1
Figure 6-16: Connector points diagram for equipment connector board
8 9
X34
X8A X35
X70, X71,
X72, X73
J1 X36
X39
X38
10
X37 (to the cabinet)
R, S, T X46
CABINET X2 5
GROUND ROD (W1) X3 9
X105
7
8
X48
X44
X4 10
CABINET CONNECTOR BOARD - A12
BASE UNIT
GROUND ROD (W2)
TERMINAL BLOCK X0
Note
Assemble the I.I. trolley with the four wheels of the Spot Film Device using the
supplied nuts.
For 12" I.I., make the trolley as shown in .
Figure 6-18
For 9" I.I., make the trolley adding the elevation plate as shown in .
Figure 6-19
1. Place the I.I. on the trolley. Position the scan in the centre and tilt the remote-controlled
table at an angle that allows the positioning of the I.I. below the Spot Film Device. Bring
the I.I. close to the table and position it so that the I.I. feeder is towards the base of the
machine.
2. Using the tilting command, move the table so that the fixing holes of the I.I. coincide with
the ones on plates already mounted inside the Spot Film Device. Proceed with caution
from this point to avoid damaging the I.I.; move the table bit by bit correcting the position
of the I.I.
Figure 6-20
3. Fix the I.I. to the plate with the supplied screws and tighten them.
4. Check the crosswise and longitudinal alignment between the Spot Film Device and the I.I.
If realignment between the two parts is necessary, loosen screws "A" (Figure 6-21) on
both sides to obtain the crosswise movement of the I.I. in relation to the Spot Film Device
or screws "B" (Figure 6-21) for the longitudinal movement; tighten the screws once the
desired position is achieved.
B
Figure 6-21
5. Perform the wiring of the I.I. cables and the angulation motor cables on the left side of the
Spot Film Device's base wall using the appropriate cable clip (Figure 6-22). The cables
must be wired so they pass "parallel" together, following the route of the cables already
present in the machine.
Cable clip
Figure 6-22
Note
International regulations require that the total filtration of the X-ray beam must be
greater than 2.5 mm Al.eq.
Ralco model collimators have a filtration of 1.0 mm Al.eq.
Considering the inherent filtration of the X-ray tube, indicated on the related
nameplate, add the necessary filtration using aluminum disks supplied with the tube
or collimator kit; these disks can be inserted in the seat of the X-ray output window.
Note
The collimator kit comprises:
• collimator fixing cone H=17 mm (0.7")
• additional filters of 0.5 mm Al
• collimator assembly spacers of 1.5 mm (0.06") and 3 mm (0.1")
• fixing screws 6x25 and 6x30.
1. Position the tube support in its fixing pin making sure to correctly connect the two release
levers to rotate the tube (Figure 6-23). Insert the support fixing screw on the pin and
secure in the most comfortable position for mounting the tube.
Fastening
pin
Support
screw
Release
levers
Tube
support
Figure 6-23
2. Position the tube on the support and at the same time apply the collimator fastening cone
(Figure 6-24) with the reference "F" positioned in front side. When doing this take into
account the distance between the focus of the tube and the cone/collimator supporting
surface; to maintain a proper correspondence between the X-ray field and the collimator
light field the distance must be 80±1 mm (3.15"±0.04"). To do this, the collimator kit's
spacers have to be used (refer to the technical data that comes with the X-ray tube to
determine the point of focus). Before fastening cone with the screws, position X-ray tube
and collimator parallel to tube support, considering that the choice of screw length is
determined by the number of spacers used.
Note
For detailed information about collimator installation, refer to the relevant Manual
provided as electronic copy inside the collimator box.
Warning
A wrong choice of screw length may lead to inadequate fixing of the X-ray assembly
if short screws are used or damage the X-ray tube if too long screws are used.
3. Installa the collimator on the cone using the four fastening hooks counting the number of
turns so that each screws is adjusted with the same distance. This operation permit to
center mechanically collimator with X-ray tube output window.
Note
To safely mount the collimator to the X-ray tube, refer to the Ralco Safety
Instructions or to the relevant Manual provided as electronic copy inside the
collimator box.
X-ray tube
Spacer
Collimator
fastening cone
Collimator
Figure 6-24
Figure 6-25
6. After wiring inside the tube arm and having verified its effectiveness after checking the
rotation of the tube, the arm cover is mounted (Figure 6-26 - 4 screws provided on the
column profile + 8 screws on the arm). To perform this operation, increase the SID and
remove the chain and its right side support (Figure 6-27) allowing access to the column
profile screws.
Figure 6-26
Chain support
Figure 6-27
(*)
(*)
Figure 6-28: Apollo EZ SID 1000 ÷ 1500 mm (39.4" ÷ 59") cover installation diagram (front view)
Figura 6-29: Apollo EZ SID 1000 ÷ 1500 mm (39.4" ÷ 59") cover installation diagram (rear view)
(*)
(*)
Figure 6-30: Apollo EZ SID 1100 ÷ 1800 mm (43.3" ÷ 70.9") cover installation diagram (front view)
Figure 6-31: Apollo EZ SID 1100 ÷ 1800 mm (43.3" ÷ 70.9") cover installation diagram (rear view)
Lock the tabletop to the fixing plates "A" by the eight screws supplied with the arms (Figure
6-32).
Warning
Verify the presence of the footrest safety screws "C" on tabletop footside (both side).
If removed, they may have effect on patient safety.
Warning
In case the optional shoulder rest (Figure 3-1) is present, apply the included labels
following the instruction supplied with the option.
Footrest safety
screw "C"
Figure 6-32
Warning
This operation requires the presence of at least 3 people.
1. Position the tabletop close to the two support arms with the feet side drive assembly.
Figure 6-33
2. Lay the tabletop on the support arm, feet side, and slide it until the motor unit engages
the side fixing plates on the arm (Figure 6-34).
3. Insert the 4 fixing screws tabletop – feet side arm side plates, without tightening. Insert
the 3 lower fixing screws, without tightening.
Anti-entrapment
PCB - A23
4. Align the 4 fixing screws, head side, sliding the arm and inserting them.
5. Position the tabletop parallel with the beam. Tighten first the 3 lower screws, feet side
(Figure 6-34), then tighten the 4 fixing screws, head side (Figura 6-35) and then tighten
the 4 fixing screws tabletop - feet side arm side plates (Figure 6-34).
6. Connect the motor cables (X74, X78, X84, E3) to the relevant ones caming from the chain.
7. Connect the anti-entrapment cables (X1, X2, X3) to the relevant connectors on the Anti-
entrapment PCB A23 (Figure 6-34).
Warning
Verify the presence of the footrest safety screws "C" on tabletop footside (Figure 6
32 - both side).
If removed, they may have effect on patient safety.
Warning
In case the optional shoulder rest (Figure 3-1) is present, apply the included labels
following the instruction supplied with the option.
M8x25 screws
Spacer
M8x50 screws
• Take the values related to the actual speed of the movements (see section 11.1); this
operation is essential for installation with a frequency of 60 Hz
• Test the functionality of all movements for the different operation modes (Spot Film
Device mode, Tomography and Digital when available) and check the anti-collision safety
devices
• After having calibrated the X-ray tube, centre the X-ray tube - Collimator - Spot Film
Device - I.I. (see chapter 19)
• Verify that the position of the presser is adjusted correctly for the type of cassette used.
This should exert pressure on the cassette in order to hold it even with the table in an
upright position, without excessive pressure such to make the ejection movement of the
cassette critical. The control must be performed with at least sizes 18x24 (8 "x10") and
35x43 (14"x17").
If the presser adjustment were not optimum, loosen the 6 fixing screws and allow the
presser spring "A" (Figure 6-38) to slide in the desired direction keeping it parallel to the
supporting surface.
Figure 6-38
• Check the centring made for the various cassette sizes (see chapter 19); if necessary
correct any inaccuracies.
Note
At the end of installation, before using the system with patients, it is necessary to
perform electrical safety test according to local rules.
Note
The Service Engineer has to take special care for all what concerns electrical safety
of the device and must make sure of restoring all provisions for electrical safety
which may be affected during a service intervention and to solicit the customer to
have the electrical safety tests repeated every time the intervention has caused the
replacement of important parts or the intervention has significantly affected safety
provisions of the device.
It is easy to deduce what the function of the PCB’s listed above is from their names and they
are situated in the electrical cabinet with exception of the touch screen PCB.
Some information about the characteristics of the system's CPU that could come in useful, is
given below.
The "touch screen" A14 PCB is fitted with two µprocessor, both connected with the cabinet
CPU via Ethernet.
The first µprocessor, model Microchip PIC, is used to decode the state of the joystick and send
this information to the cabinet CPU.
The second µprocessor, model Freescale MCF 5329, is used to manage the graphic layout on
LCD and receive the command from the touch layer to transmit to cabinet CPU.
The A1 "CPU" PCB is equipped with a MCU 32B ARM Cortex-M4 (STM32F427IIT6) processor
the job of which is to act as main processor with its 2Mb Flash on board.
All the required memory devices, except the EEPROM containing the equipment set up data,
are available on Microcontroller chip. Also the need of storing variables/error logs is resolved by
using the internal flash memory (32k byte) of the Microcontroller, organized in 2 x 16 kB
sectors (for update and backup).
The incorporated program manages the entire functioning logic of the equipment including, as
said before, dialoguing with the touch screen.
The release of the equipment management software, rack side and console side, is shown on
the touch screen, SETUP page, in the bottom left corner (see Figure 9-1).
7.1 Functioning
Functionin g sequences
On powering up the electronics of the CPU, the Touch Screen, the Input board and part of the
Output board is powered through the T2 transformer.
At this point the main processor and the controller begin to perform the checks that can make
sure all the conditions are regular.
If that is the case, both the main processor and the controller activate the respective "ready"
relay after an interval of three seconds.
The open contacts of these two relays are connected in series.
One phase of the coil of the power remote switch K1L is interrupted by the series of these two
relays.
When they are activated, they therefore permit the energising of the K1L that, through the
closure of its contacts, will power the table's power circuits.
At this point, the procedure for the initialisation of the Spot Film Device begins the purpose of
which is to move the inside parts of the Spot Film Device to the default positions.
While this is happening, i.e. from the time the power circuits have been powered up and when
the parts of the Spot Film Device have reached the positions defined, all the movements of the
table are disabled and the touch screen shows the message:
"Loading. Please wait…"
When the initialisation procedure is successfully completed, the touch screen changes and
shows the data for the active mode.
All the table movements and the functions will, at this point, be enabled.
These three different situations cause the consequences that are analysed below.
While the equipment is in operation a fault can occur in the table or one of the accessories
connected to it.
If this fault does not compromise the functioning of the system (Example: the generator is not
"ready" within the set time), a few movements are disabled by the SW and the relative alarm
message is displayed on the touch screen accompanied by an intermittent "beep".
To remove the alarm and re-enable all the movements it will be necessary to push the "RESET"
key twice.
The first time silences the "beep", the second resets the alarm and restores the normal
conditions.
If a movement of the table or the Spot Film Device should for any reason go over the SW run
determined by the potentiometer, an emergency end of run will be intercepted.
This will entail the turning off of the power remote switch K1L with the consequent opening of
the power control input (X15 pin 10).
This will mean the main processor will turn off the READY relays and the controller and
generate the alarm 080 on the touch screen.
It will therefore be necessary to turn off the equipment to remove the alarm situation.
After removing the cause of the failure, the powering up procedure after the power cutting out
must be followed in order to restore the normal operating conditions described in the relevant
chapter.
As a block diagram of the sequences described above, the one below is useful for explaining the
functioning logic in even clearer terms.
A3 Output
Out put board Description
The manual Kv fluoro driver output is dependent on an analogical circuit (TLC271) with a
voltage range from 1.8 Vdc to 6.5 Vdc. A max load of 5 mA can be applied.
The following outputs are part of an optoisolator circuit (LH1520) that must be polarised. The
maximum voltage and current applied must not exceed 25 Vac with 70 mA.
X0 Connector Description
X0 Connector Description
0 00
↓ ↓
Number of the group Position of the cell
To access the read only function of data stored in these cells, position the switch of the group
of DIP-switches on the CPU at ON.
To edit and modify data in the cells, after activating the switches, it will be necessary to key in
an access code.
The table below lists different data groups together with correspondent DIP-switche
configuration and access codes to be keyed-in.
Warning
The Authorised Technician is recommended not to communicate these codes so that
non-authorised persons are prevented from entering the memory cells and changing
them, thereby causing the equipment to malfunction.
The version of the installed software, both rack side and console side is shown on the touch
screen, left bottom corner of the SETUP page (see next figure).
cm Grid: 120
Mode
Ret.
0° 0° 100 0.0 SETUP
Yes No - + 100%
MODE-3
Cassette Eject Volume MSG
Figure 9-1
where:
• F W CPU ver.
ver . = version of the rack software also shown on the rack Eprom label, plus
the release date
• SW TSC ver. = version of the touch screen software, plus the release date
For any signalling of a malfunction or request for technical help, always indicate the version of
the software on the equipment.
The following table shows the DIP-switch combinations – access code and the functions of the
various groups.
DIP-
DIP -switch Code Group Subroutine
1 PWD0 101 Potentiometer adjustment
2 PWD1 201 Installation data
3 PWD0 301 PDI controlled motor setting data
4 PWD1 711 “time” setting -Reset alarm memory
1 + 2 PWD0 401 Equipment dimension setting
Run shutters and cassette
1 + 3 PWD1 701 Eeprom copying
1 + 4 751 Alarm history visualisation
2 + 3 PWD1 721 Spot Film Device and table life test
2 + 4 PWD1 731 Remote service function
3 + 4 PWD1 850 Equipment cycle and time display
1 + 2 + 4 Access codes setting
Reset Keys...
Print Keys...
The procedure shall require you to enter, using the keypad, the current access code, the new
access code and the confirmation of the new access code.
The access code must be 5 decimal digits and the first digit must be different from 0.
Note
If it is necessary to restore the default access code, contact Villa Sistemi Medicali to
create a super-password for the single serial number.
Reading data
1. Turn off the equipment, turn switch 1 in the DIP-switch assembly in the CPU board
to ON.
2. Turn on the equipment. The touch screen will show:
Mode
CELL SERVICE
Edit
369 -40.0 2090 / 4096
where:
369 = value presently in the memory
- 40.0° = position that the part has to reach corresponding to the value at
minimum
2090/ = current value of the potentiometers expressed on a scale with a
maximum value of 4096
101 = number of the cell
Angulation… = brief description of the function of the selected cell.
Data entry
1. Press on the area of the selected cell and hold it down for 1 sec; the display will show
a numerical keyboard asking to insert the access key:
Enter Key 1 2 3
4 5 6
ESC 7 8 9
C 0 -
.
Note
The key code is required only in case some data need to be changed and only once
each start-up.
2. Set the key code relative to this data group (see chapter 9).
ESC
3. Press key (Enter) to confirm or to exit; on the "insert address key" area,
the display will show the message:
<< KEY OK >>
that indicates that it is possible to access groups 100-300-400, thanks to the use of
the code that has just been keyed in, in function of which switches are positioned at
ON.
If the code keyed in is wrong, the following message appears on the screen, instead:
<< Wrong code: access denied >>
C
In this case, press key (Clear) on the numerical keyboard and digit again the
4. The display returns to the position of the previously selected cell and the
Edit
key will be displayed on the left of the screen.
Mode
CELL SERVICE
Edit
369 -40.0 2090 / 4096
10.1 Potentiometer
Potentiomete r adjustment
The Apollo EZ table bases the control of its movements on the feedback generated by the
potentiometers.
These potentiometers must therefore be adjusted, the reference value of the maximum and
end run voltages must be stored in the corresponding cells in the EEprom.
At this point it will be possible to edit the data of the cells making up this group.
It will also be possible to perform the movements for the adjustment of the potentiometers by
using the compressor movement joystick.
In this program, only this joystick is active (except for the calibration of the collimator
movement as described below).
If the joystick is pushed upwards, the direction of the set movement will be towards its
maximum and the potentiometer value shown on the display will increase.
If the joystick is pushed downwards, the direction of the movement will be the opposite,
towards its minimum, and the potentiometer value will decrease.
In this program the speed of the PDI movements are set permanently in the adjustment
software and cannot be modified.
The value of the potmeter corresponding to the current position is displayed in the first square
(blue area). After reaching the desired position, just store the current value into the EEprom by
Edit
pressing the key.
Warning
During the adjustment phase, the anticollision program is not active so, before
performing any movement, make sure that that there is no risk of collision during it.
Furthermore, since the software end runs are set in this program no stopping on
them is activated.
If the set run is exceed, the movement will hit the emergency switch causing the
equipment to stop.
The limits of the movements are set by default during the factory's predelivery testing of the
equipment.
These limits are shown on the display depending on the cell that is set.
It is possible to vary these limits by modifying the data in the cells of group 400, but in
practice this operation is never performed. In fact this modifies the functional characteristics of
the table.
Furthermore, in order to ensure maximum safety also in cases where there is a malfunction in
the movement electronics control, there are hardware emergency switches positioned in such a
way as to cut in when the software end run does not stop the movement. These positions
break the power supply to the hardware circuits (they determine the opening of the K1L
remote switch). Thus any modification of the run limits also entails the mechanical repositioning
of these switches.
In the table below the functions of group 100 cells are shown.
Joyst.direct
Cell Movement Value Movement direction Default.lim
.
101 Angulation LOW min Anticlockwise (head end) - 40.0°
40.0°
102 Angulation HIGH max Clockwise (feet side) +40.0°
+40.0°
4 way scan LOW Min Left (head end) - 665 mm
103
2 way scan LOW Min Left (head end) - 670 mm
4 way scan HIGH max Right (feet side) +515 mm
104
2 way scan HIGH max Right (feet side) +670 mm
105 Tilting -30 LOW min Anticlockwise head end - 30°
(-30)
106 Tilting +90 HIGH max Clockwise feet side (+90) +90°
107 Longitudinal tabletop LOW min Left head end - 600 mm
108 Longitudinal tabletop HIGH max Right feet end +600 mm
109 WB width cass. (*) LOW min CLOSED 140 mm
110 WB width cass. HIGH max OPEN 465 mm
111 WB height cass. LOW min CLOSED 140 mm
112 WB height cass. HIGH max OPEN 465 mm
SID (a) 1016 mm
113 LOW min Low
SID (b) 1106 mm
SID (a) 1516 mm
114 HIGH max High
SID (b) 1806 mm
115 Transversal tabletop LOW min Inside - 175 mm
116 Transversal tabletop HIGH max Outside +175 mm
(*) Cells 109 ÷ 112 refer to the potentiometers in the presence of chest stand (WB - Wall
Bucky) equipped "Sensing" cassette holder.
(a) Apollo EZ table with SID 1000 ÷ 1500 mm (39.4" ÷ 59")
(b) Apollo EZ table with SID 1100 ÷ 1800 mm (43.3" ÷ 70.9")
Joyst.direct
Joyst.dir ect
Cell Movement Value Movement direction Default.lim
.
123 Compressor LOW/HIGH min Descent/Rise 1.5 kg
124 Compressor LOW/HIGH max Descent/Rise 15.5 kg
125 Grid 1 LOW min Park direction 000 mm
126 Grid 1 HIGH max Field direction 510 mm
127 Cassette close LOW min Closes 146 mm
128 Cassette close HIGH max Opens 463 mm
129 Optional LOW min Left (head end) - 65.0 mm
Cross subdivision
130 Optional HIGH max Right (feet side) +65.0 mm
Cross subdivision
131 Shutters LOW min Towards opening 000.0 mm
132 Shutters HIGH max Towards overlap (cross 455.0 mm
subdiv.)
133 Grid 2 LOW min Park direction 000 mm
134 Grid 2 HIGH max Field direction 510 mm
As said before, the default value that the movement must reach at the minimum and
maximum is shown on the display.
Once this position has been reached, the potentiometer must indicate a near value (a tolerance
of up to 100 units can be reached) as shown in the following table.
If not, release the potentiometer, reach the position required with the movement then turn the
potentiometer until reaching the indicated value.
Lock the potentiometer in position and store the value to memory.
Compliance with this value makes it possible to ensure that the potentiometer is not damaged
mechanically as a result of a position that leads to exceed its mechanical stop, and the
potentiometer is used in the central area, thus avoiding using the limits that often turn out to
be critical.
As far as the position to reach to carry out the adjustments of the up to and maximums, it will
be shown how to measure the value to reach, on the next few pages and in the descriptions.
In some cases, this measurement is different from what is shown on the display, as shown in
the last column of the proceeding table.
This is because the indicated value is the result of software calculations having table points of
origin as a reference but, in some case, it is intricate and difficult to make these measurements
in practice.
This is why points other than those shown in the figures have been chosen.
Potentiomete Potentiomete
Cell Movement Movement direction
r direction r value
101 Angulation Decrease Anticlockwise (head end) approx.334
102 Angulation Increase Clockwise (feet side) approx.3762
4 way scan approx.117
103 Decrease Left (head end)
2 way scan approx.110
4 way scan approx.3979
104 Increase Right (feet side)
2 way scan approx.4000
105 Tilting Decrease Anticlockwise head end approx.250
(-30)
106 Tilting Increase Clockwise feet side (+90) approx.3900
107 Longitudinal tabletop Decrease Left head end approx.411
108 Longitudinal tabletop Increase Right feet end approx.3838
109 WB width cass Decrease
110 WB width cass Increase The relative data are based on the "Sensing"
111 WB height cass Decrease cassette holder used.
112 WB height cass Increase
SID (a) approx.341
113 Decrease Low
SID (b) approx.100
SID (a) approx.3754
114 Increase High
SID (b) approx.3980
Transversal tabletop approx.287
(2 way)
115 Decrease Inside
Transversal tabletop approx.117
(4 way)
Transversal tabletop approx.3431
(2 way)
116 Increase Outside
Transversal tabletop approx.3979
(4 way)
Potentiomete Potentiomete
Cell Movement Movement direction
r direction r value
123 Compressor Decrease Descent/Rise approx.1340
124 Compressor Increase Descent/Rise approx.2640
125 Grid 1 Decrease Park direction approx.205
126 Grid 1 Increase Field direction approx.3710
127 Cassette close Decrease Closes approx.360
128 Cassette close Increase Opens approx.3730
129 Optional Decrease Left (head end) approx.1430
Cross subdivision
130 Optional
Option al Increase Right (feet side) approx.2640
Cross subdivision
131 Shutters Decrease Towards opening approx.330
132 Shutters Increase Towards overlap (cross approx.3740
subdiv.)
133 Grid 2 Decrease Park direction approx.205
134 Grid 2 Increase Field direction approx.3710
To carry out this adjustment, it is necessary to remove the covers of the scan group.
At this point, the Spot Film Device support plate can be seen showing the column tilting
reference marks.
The limits of the angulation are obtained by bringing the upper edge of the column to
correspond to reference lines on the Spot Film Device support.
With a level goniometer it will be possible to carry out a further check of the soundness of the
min and max positions reached.
Take the column towards the minimum (head side), select cell 101 and store to memory.
Repeat the operation for the maximum (feet side) and memorise in cell 102.
102
-40° +40°
Linee di
riferimento
57mm 1397mm
(2.2") (55.0")
62mm 1242mm
(2.4") (50.0")
-30°
Reference line
Take the main beam towards the maximum until it is perpendicular to the ground, using a
bubble level to check it is at right angle (+90° - Figure 10-5); select the cell 106 and store the
maximum value to memory.
+90°
Take the tabletop perfectly aligned with the sliding bars (Figure 10-6).
Enter in cell 107 and move the tabletop toward head side (minimum) until you masure with a
tape meter a value of 600 mm (23.6") from the sliding bar end and the edge of the tabletop
(Figure 10-7); store the potentiometer value.
Return the tabletop to the start position (aligned with the sliding bars).
Enter in cell 108 and move the tabletop toward foot side (maximum) until you measure a value
of 600 mm (23.6") from the sliding bar end and the edge of the tabletop (Figure 10-8); store
the potentiometer value.
600 mm
(23.6")
600 mm
(23.6")
542mm
(21.3")
42mm
(1.6")
739mm
(29.1")
39mm
(1.5")
171mm
(6.7")
520mm
(20.5")
259mm
(10.2")
608mm
(23.9")
A compression dynamometer is required to perform this adjustment with a range from 1.5 kg
to 15.5 kg (3.3 lbs to 34.1 lbs).
Move the tabletop into the centred position and rest the dynamometer on it.
The minimum must be adjusted with no force applied to the cone; store the value in position
123.
123
Bring the cone down until a pressure of 15 Kg (33 lbs) as shown on the dynamometer is
exerted; store the value in position 124.
124
Move the frame to the maximum position (field) and find the 50 mm (1.9") measurement
among the same references used for the minimum.
Store the value in cell 126.
126
460mm
(18.1")
50mm
(1.9")
Grid centre
(park position)
Centre line notch
Grid centre
(field position)
Warning
The adjustment of the cassette tightening guide potentiometer must be carried out
with the scanning assembly for the cross subdivision (optional) perfectly centred.
When the maximum is adjusted the greatest care must be taken to ensure the
guides do not violently strike the walls of the Spot Film Device but just brush them
slightly.
Forcing the guides causes damage to the protection fuse F22 on the Spot Film
Device board A4.
The tightening switch's job is to ensure a certain contact between the cassette and the driving
belts.
This switch is ideally adjusted in order to get a pressure on the cassette guides of
approximately 6 Kg (13.2 lbs).
Insert a cassette with height 24 cm (10 ") to make the reference reading visible when the
switch is activated.
To respect the value of 6kg (13.2 lbs), the cam should stop in the 40° ± 10° position to be read
on the scale on the scroll.
Reference
scale
Graded
scale
Note
The cassette to use must be one of those the radiology room is equipped with. This
is because the check must be done under the worst operating conditions (maximum
weight in the lifting mode).
If the 35x43 (14"x17") size is not used, perform the test with the heaviest format
among those which are supplied.
If the calibration is correct, the switch should not intervene during the upward movement of
the cassette until both guides are in contact with the cassette and the 10° movement of the
cam on the scale has been made.
If this is not the case, work on the switch activating cam to bring its intervention forward or
delay it.
Warning
An adjustment that is not high enough (< 5 Kg / 11.0 lbs) causes the switch to cut in
early during the test described above.
An adjustment that is too high (> 7 Kg / 15.4 lbs) causes pressure that is to great
between the drive belts and cassette meaning the cassette has difficulty moving.
10.1.11 CRO
CROSS SS SUBDIVISION potmenter adjustment (optional)
(cells 129 – 130)
Insert a cassette between the guides with height of 24 cm (10") on which there is a centre line
and tighten the guides to it.
Select cell 129 and push the compressor joystick until the centre line of the cassette is 65 mm
(2.6") away from the centre of the cassette supporting surface; store the value.
Select position 130 and repeat the operation in the opposite direction with respect to the
centre line; store the value.
475mm
(18.7")
Take the shutters towards the maximum position (overlapping) and read the measurement of
145 mm (5.
(5 . 7 ") (Figure 10-17); store the value in cell 132.
132
145mm
(5.7")
Move the frame to the maximum position (field) and find the 50 mm (1.9") measurement
among the same references used for the minimum.
Store the value in cell 1 34.
34
460mm
(18.1")
50mm
(1.9")
Grid centre
(park position)
Centre line notch
Grid centre
(field position)
Reading data
1. Turn off the equipment, position the switch or switches of the DIP-switch group on
the CPU board, relating to the data group for which access is required, at ON (see
chapter 9).
Mode
CELL SERVICE
YYYY
where:
XXX = number of the group and the position
YYYYYY = value currently in the memory.
Data entry
1. Press on the area of the selected cell and hold it down for 1 sec; the display will show
a numerical keyboard asking to insert the access key:
Enter Key 1 2 3
4 5 6
ESC 7 8 9
C 0 -
.
Note
The key code is required only in case some data need to be changed and only once
each start-up.
2. Set the key code relative to this data group (see chapter 9).
ESC
3. Press key (Enter) to confirm or to exit; on the "insert address key" area,
the display will show the message:
<< KEY OK >>
that indicates that it is possible to access group, thanks to the use of the code that
has just been keyed in, in function of which switches are positioned at ON (see
chapter 9).
If the code keyed in is wrong, the following message appears on the screen, instead:
<< Wrong code: access denied >>
C
In this case, press key (Clear) on the numerical keyboard and digit again the
5. To modify the data, press on the area of the selected cell and hold it down for 1 sec;
the display will show a numerical keyboard:
Mode
CELL SERVICE
YYYY
4 5 6
7 8 9
C 0 -
ESC .
6. By using the numeric keys key in the data required and press the key (Enter) to
ESC
confirm the correction or to exit.
If no key is pressed within about 15 sec, the keyboard disappears.
Note
In case of data with a negative value, enter the number first and then the sign.
As the data stored in the cells of this group can be modified to render the machine fit for the
installation requirements, the default data set during the table adjustment in the factory is
shown in the second to last column in the table. The last column is for the service technician to
note any variations from the default data.
3 4
- 1= 1 2
253 0 = no 1 = yes The table version with SID 1000 ÷ 1500 mm (39.4" ÷ 59"), has the angulation -- Dependin
Centering angulation motor motor in asymmetric position. Consequently the scan range is reduced on the g upon
motor side during tilting movements in Trendelenbourg direction, to avoid the version
collision between motor and floor.
The table version with SID 1100 ÷ 1800 mm (43.3" ÷ 70.9"), has the angulation
motor in centred position, consequently no scan limitation is required.
This cell set depends by the table version and define the scan range limitation
depending by the tilt angle.
254 0 = no 1 = yes The table is designed to allow the implementation of a safety circuits against -- Dependin
Anticrushing safety device entrapment of parts between the tabletop inner profile and the Spot Film rear side g upon
and between the column and Spot Film Device lateral side. If this circuit is installed version
it will be necessary to activate the relative input software control using this
setting. The 415/480V versions are always equipped with this circuit.
1 for
4 - way
259 0=off-off 1=off-on Dependent on the type of TV chain installed, it might be necessary to activate the -- 0
Horizontal image scan reverse horizontal image scan reverse output by default. The 0 selection determines that
when the table is switched on both the dedicated key LED and the its output are
Off By pressing the key the LED lights and the output is activated.
When the table is on, selection 1 determines the automatic activation of the output
with the led of the relative key off. When the horizontal image scan reverse key is
pressed the led comes on and the output is turned off.
260 0 = off-off 1 = off-on What was described for cell 259 is valid here applied to the vertical image scan -- 0
Vertical image scan reverse reverse function.
261 0 = enable 1 = disable It disables the Spot Film Device controls. The initialisation functions and all the -- 0
SDF disabling relative modes are no longer performed. This override is present when the table is
in the “all digital” version.
This can be activated if the Spot Film Device breaks down to exclude alarms and
any disabling.
– Integral
The integral part is product of the multiplication:
– Derivative
The derivative part is the product of the multiplication:
Here below is a brief analysis of the remaining fields responsible for the control of a PDI axis.
• Integration limit
limit
This is the maximum value that the sum of the integral action errors can assume.
If the sum of the errors exceeds the value set in this cell, it will be limited in accordance
with the limit set.
If the data set in this cell is 0 no limitation will be performed.
This parameter avoids action that is too extreme in the case that the sum assumes values
that are considerable.
• Braking time
This value permits to ramp down the motion independently of the acceleration ramp. The
time inserted into these cells is equal to the time the motion employs to pass from the
actual speed to 0 in a linear way. This occurs regardlessly of the real position with respect
to the theoretical one. In this case, the integral position recovery actions are not carried
out and the motor turning off delay does not have effect. This braking modality is used for
the following movements: tilting, scan and angulation.
The following tables, divided by motor, contain the data relative to the various PDI-controlled
motors.
Angulation movement
Scan movement
Tilting movement
Shutters movement
Cassette movement
As explained in the Chapter about potentiometers adjustment, the limit to reach for the
calibration is shown on the display.
This limit is set in this block of data.
To vary the value of the set default data means to adjust the movement run range.
Moreover, it will be necessary to check if the movement mechanics can sustain such variation.
The software will divide the new run range by the number of potentiometer's points that are
going to be sampled between the minimum and maximum value of the new run.
Warning
Do not modify the data in this group as they are linked to the operative
characteristic of the equipment.
This block of data contains some mechanical dimensions characteristic of this equipment.
The software uses these dimensions to perform calculations of the various functions'
management algorithms (anticollision, tomography, and tilting functions, etc.).
Warning
These data are closely linked to the equipment and therefore should never be
modified in order to avoid modifications to the equipment functioning.
This block of data is used to check the potentiometers and the functions linked to single faults.
A detailed description of the functions is reported in each cell.
This block contains the runs carried out by the Spot Film Device cassette and shutters to
perform the various cassettes subdivisions and the relative limitation.
In case the separating lines of the X-rays of some subdivisions are not perfect, it is possible to
modify these runs, by following closely the instructions reported in the dedicated Chapter.
Note
The cassette and shutters positioning might not correspond to the measure in the
relative cell because of the mechanical inactivity of the system.
In every way the correction in mm introduced in the cells has an increment and
decrement effect in the final amount equal to the value of the correction.
For rapid consultation and interpretation, the runs and the relative cells are summed up in
tables where the first number indicates the cell number and the second, in brackets, the stored
data in mm.
The calculation is based on the distance of cassette internal side, when the same is in the
parking position, with respect to centre of the Spot Film Device. Such distance is divided into 2
parts:
• the distance of the cassette size photosensor from the centre of the Spot Film Device
(227 mm / 8.9")
• the distance of the cassette size photosensor from the internal side of the Spot Film
Device (484 mm / 19.0")
The distance of 484 mm (19.0 ") is obtained because the distance of the photocell from the
bottom of the Spot Film Device (cell 444) is 472 mm (18.6"), but the cassette parks position is
set by the SW:
• 5 mm (0.2 ") from the bottom after the introduction of the cassette in the SFD
• 20 mm (0.8 ") from the bottom on the return of a preparation run or test.
Note
The figures reported in the tables and stored in the cells refer to the cassette
parking position at 20 mm (0.8") from the bottom of the Spot Film Device. The
correction, when the parking position is 5 mm (0.2") from the bottom, is carried out
automatically by the SW that increases the measurements reported in tables by 15
mm (0.6").
Note that it is possible to store a new negative or positive offset value (cell 219) to compensate
the differences due to manufacturing or operating characteristics. This offset will permit to
modify all runs of all subdivisions of the stored quantity.
The formulas applied to determine the runs are as follows:
• Full size
C = (cassette width / 2) – 711
where 711 = 227 + 484
• Subdivision 2 in line and 4 cross
C = full size run +/- (film / 4)
• Subdivision 3 in line and 6 cross
C = full size run +/- (film / 3)
• Subdivision 4 in line
C = full size run +/- (3/8 film) and (1/8 film)
13x18 18x13 18x24 24x18 24x24 24x30 30x24 30x30 18x43 43x18 15x40 40x15 20x40 40x20 30x35 35x30 30x40 40x30 35x35 35x43 43x35
43x3 5
461 462 463 466 466 469 472 478 463 512 531 532 518 521 478 500 484 490 5 00 506 512
F ULL SIZE
(600.8) (576.0) (576.0) (547.0) (547.0) (547.0) (516.5) (516.5) (575.0) (450.7) (590.5) (466.5) (565.0) (467.0) (516.5) (489.3) (517.0) (467.0) (489.0) (489.3) (451.3)
464 467 467 470 473 479 513 533 519 522 479 501 485 491 501 507 513
(531.0) (486.5) (486.5) (490.0) (442.0) (441.5) (343.5) (367.0) (517.5) (367.0) (441.5) (400.0) (442.5) (367.5) (400.0) (400.0) (342.5)
S UBDIV . 2 L
465 468 468 471 474 480 514 534 520 523 480 502 486 492 502 508 514
(620.0) (606.0) (606.0) (607.5) (592.0) (591.5) (557.5) (567.0) (617.5) (567.0) (591.5) (577.0) (592.0) (565.5) (577.0) (577.0) (557.0)
475 481 515 535 524 481 503 487 493 503 509 515
(415.0) (416.5) (307.7) (335.0) (335.0) (416.5) (370.5) (417.7) (334.3) (370.5) (370.5) (308.0)
476 482 516 536 525 482 504 488 494 504 510 516
S UBDIV . 3 L
(516.0) (516.5) (450.5) (466.5) (466.5) (516.5) (487.5) (517.0) (467.0) (487.5) (487.5) (451.5)
477 483 517 537 526 483 505 489 495 505 511 517
(616.0) (616.3) (592.0) (599.0) (599.0) (616.3) (606.3) (615.5) (599.0) (606.3) (606.3) (593.8)
538 527 496 665
(317.0) (317.0) (316.6) (287.8)
539 528 497 666
(417.3) (417.0) (417.3) (395.2)
S UBDIV . 4 L
540 529 498 667
(516.0) (516.0) (515.0) (502.8)
541 530 499 668
(614.5) (614.5) (617.2) (610.2)
464 467 473
(530.0) (486.5) (442.0)
465 468 474
D IV. 4 C (622.0) (606.0) (592.0)
(O PTIONAL ) 464 467 473
(530.0) (486.5) (442.0)
465
465 468 474
(620.0) (606.0) (592.0)
475
(415.0)
476
(516.0)
477
D IV. 6 C (616.0)
(O PTIONAL ) 475
(415.0)
476
(516.0)
477
(616.0)
5x7 7x5 8x10 10x8 9.5x9.5 10x12 12x10 7x17 17x7 11x14 14x11 14x14 14x17 17x14
571 572 573 576 588 579 582 628 629 594 600 610 616 622
F ULL SIZE
(602.0) (577.0) (563.5) (540.8) (546.0) (538.8) (514.0) (577.0) (449.0) (524.5) (486.0) (488.0) (488.0) (449.5)
574 577 589 580 583 630 595 601 611 6 17 623
(513.6) (476.8) (485.5) (576.7) (438.3) (342.0) (456.3) (397.5) (399.2) (399.3) (341.5)
S UBDIV . 2 L
575 578 590 581 584 631 596 602 612 618 624
(613.0) (601.5) (606.0) (601.0) (589.0) (556.0) (596.3) (575.5) (578.0) (578.0) (557.4)
591 585 632 597 603 613 6 19 625
(466.5) (414.5) (306.4) (432.0) (371.0) (369.5) (369.5) (306.4)
592 586 633 598 604 614 620 626
S UBDIV . 3 L
(546.6) (514.5) (449.5) (524.5) (486.0) (488.0) (488.0) (449.4)
593 587 634 599 605 615 621 627
(626.0) (613.5) (591.5) (617.0) (603.5) (606.0) (606.0) (592.4)
635 606
(288.0) (353.0)
636 607
(396.0) (440.5)
S UBDIV . 4 L
637 608
(503.3) (529.5)
638 609
(609.5) (618.3)
574 583
(512.0) (436.3)
575 584
D IV. 4 C (613.0) (587.0)
(O PTIONAL ) 5 74 583
(511.0) (435.0)
575 584
(612.0) (587.3)
585
(412.5)
586
(512.5)
D IV. 6 C 587
(O PTIONAL ) (611.5)
585
(411.2)
586
(510.7)
The calculation is based on the typical measures of the Spot Film Device reported below:
• film-shutter high distance
• film-shutter low distance
• film-shutters average distance
• minimum source - film distance
• maximum source - film distance.
The origin position for the calculation of all runs is the one relative to the calibration of the
shutters minimum potentiometer, i.e. when the shutters are wide open.
As to line subdivisions, this position is equal to a 475 mm opening, therefore to obtain an
opening equal to 0 mm the shutters will have to move of 475/2 mm.
For cross subdivisions (optional), the window opening start will require a run equal to
(475/2)+180=417.5 mm., where 180 mm is the width of the small shutter.
Therefore to obtain an opening equal to 0 mm for cross subdivisions, the shutters will have to
move of 417.5 mm.
In the calculation of the formats to limit, the values KG (exposure correction), calculated with
min SID will be used.
This allows to obtain a separation line between 2 adjacent radiograms with min SID and a
thicker line with higher SIDs.
Should a KG be used, calculated with SID greater than the min, the obtained separation lines
would be too thin or radiograms would overlap in case of SIDs smaller that the one used for
the KG calculation.
13 15 18 20 24 30 35 40 43
5 7 8 9.5 10 11 12 14 17
DIV. 4 C
642 647
(OPTIONAL
(100.0) (149.0)
)
DIV. 6 C
649
(OPTIONAL
(100.0)
)
In case the PDI value generated by the software and sent to the motors' activation differs
from the feedback speed supplied by the potentiometers or by the encoders, the system must
be able to recognise this effect and therefore bring the equipment to a halt.
This inconvenient may be caused by a mechanical hardening or by a gear slippage at the
potentiometer.
For each of the PDI-controlled movements, 2 values determine the variance amount between
PDI and speed, and the time beyond which this variance triggers an alarm:
• Variance time
It is expressed in seconds and it allows to mask possible spikes of the value of the PDI
value determined by friction or typical momentum of inertia at the start.
In case a real problem occurs, this time will increase depending on the amount of
movement performed before an alarm is issued.
• K factor
This factor divides the PDI (PDI/k) value.
The software performs a comparison between the speed value and the PDI / k value.
Since k is the denominator of the fraction, the greater k will be the smaller the result of
the fraction will be.
As the alarm is issued if PDI/k > speed, the smaller is the PDI/k value the more a
mechanically "rigid" movement will be accepted without issuing any alarm.
By and large, the greater are the time values and k, which are set for each movement, the
more the movement will accept "hardening", allowing a non-controlled wide movement
before issuing an alarm.
Angulation movement
Scan movement
Tilting movement
Shutters movement
If for safety reasons, you need to copy these data on a spare backup EEprom, please follow the
procedure below:
1. Power OFF the equipment and position DIP-switches 1 and 3 on ON.
2. Turn ON the equipment.
The following will be displayed on the touch screen:
Mode
SERVICE
Copy
Copy
3. Press the key; the display will show a numerical keyboard asking to
insert the access key:
Enter Key 1 2 3
4 5 6
ESC 7 8 9
C 0 -
.
4. Set the key code relative to this data group (see chapter 9).
ESC
5. Press key (Enter) to confirm or to exit; on the "insert address key" area, the
display will show the message:
<< KEY OK >>
If the code keyed in is wrong, the following message appears on the screen, instead:
<< Wrong code: access denied >>
C
In this case, press key (Clear) on the numerical keyboard and digit again the access
Copy
6. When the access to the copy program is enabled, press the key; the
data copy from EEprom to RAM will start.
The following message is displayed:
<< Data transfer >>
After a few seconds, if the process has been completed successfully and no alarms are
issued, the following message is displayed:
<< Transfer completed. Switch off >>
7. At this point, power OFF the equipment, replace the EEprom with an identical component
on which you might want to store all functional data of the equipment.
Mode
SERVICE
Copy
Copy
Press the key, the data will be stored on the new EEprom.
The equipment can now be used with a spare EEprom that contains all functional data.
Once the access and correction procedure for position 711 has been performed (see chapter 11),
the touch screen will shows:
Mode
CELL SERVICE
DD dd MM mm YYYY yy
711: Date
By pressing the and scrolling keys it is possible to increase and decrease the cell to
be displayed, respectively.
CELL
The cell number can also be selected by pressing key and digit the
number to be accessed on the displayed alphanumerical keyboard.
If no key is pressed within about 20 sec, the keyboard disappears.
1. Press on the area of the selected cell and hold it down for 1 sec; the display will show a
numerical keyboard asking to insert the access key:
Enter Key 1 2 3
4 5 6
ESC 7 8 9
C 0 -
.
Note
The key code is required only in case some data need to be changed and only once
each start-up.
2. Set the key code relative to this data group (see chapter 9).
ESC
3. Press key (Enter) to confirm or to exit; on the "insert address key" area, the
display will show the message:
<< KEY OK >>
that indicates that it is possible to access groups 100-300-400, thanks to the use of the
code that has just been keyed in, in function of which switches are positioned at ON.
If the code keyed in is wrong, the following message appears on the screen, instead:
<< Wrong code: access denied >>
C
In this case, press key (Clear) on the numerical keyboard and digit again the access
5. To modify the data, press on the area of the selected cell and hold it down for 1 sec. For
cells from 712 to 718, where a digit entering is required, a numerical keyboard will be
displayed.
6. By using the numeric keys key in the data required and press the key (Enter) to
ESC
confirm the correction or to exit.
If no key is pressed within about 15 sec, the keyboard disappears.
11.6 Position 721 – Test Spot Film Device and table life
The Apollo EZ table is set to perform cyclic movement tests.
These tests are useful when checking the movements for correct functioning under stress
conditions.
It is possible to perform tests for the Spot Film Device or table movements.
After performing the access procedure (see chapter 11), the following will be displayed:
Mode
LIFE TEST SERVICE
TEST S.F.D
TEST TABLE
By keying in the access code relative to this position, it will be possible to select the type of
TEST S.F.D
test to be performed by pressing key for the Spot Film Device test
TEST TABLE
or key for the Table test.
The Spot Film Device perform the initialisation procedure after which the following message is
displayed:
"Load cassette 30x24"
By loading the cassette correctly, the cyclic test begins and the following is displayed:
Mode
SERVICE
XXXX YYYY
A test cycle is structured in a way that the following positionings are carried out:
In the Full size modality, the shutters position themselves on size 43 rather than on size 30 to
carry out the maximum possible run.
Program 40/4 in line is carried out with cassette 30 as the program 4 in line for cassette 30 is
not available.
Every time the cassette is put in the parking position, the following outputs are enabled:
• fluoro request
• CCD output in rotation
• I.I. field output in rotation.
Warning
As the outputs for the system accessories are enabled, before starting the Spot Film
Device test, make sure that the accessories linked to the Apollo EZ table are not
powered.
At the end of each program the cassette is ejected, the cassette guides are open and closed
again until they block the cassette.
At the end of a complete cycle (as described in the table), before resuming with what described
up until now, the following occurs:
• the grid movement as far as the parking position and return to field
• all outputs are turned on
• the collimator lamp lights up for 5 secs.
On the display the number of exposures increases each time the cassette position itself in the
X-ray field, so is the number of cycles every time a complete cycle is performed.
A complete cycle consists of 17 cassette positions.
The number of exposures can be modified by pressing on the area of the selected cell and hold
it down for 1 second to enable the correction and by keying in the desired number by means of
the numerical keyboard (see chapter 11 – Data Entry procedure).
11.6.2 Table
Tabl e test
The table will cyclically perform movements that allow to check the correct functioning of the
various parts.
Warning
As the movements are activated automatically without opening the SF circuit, during
this test, the supervisor is disabled. Therefore its K3 ready relay does not close. In
order to perform the test, make a jumper on the connector X47 between pins 6
and 7.
Remove the jumper at the end of the test.
Warning
The test is conceived in such a way to limit the movements in case collisions occur.
Nonetheless pay always attention during automatic movements performed by
this program. Particularly during the first cycle, stay close to the emergency red
button and be ready to activate it if any risk condition occurs.
Mode
SERVICE
YYYY
Before starting the proper test cycle, the table goes into the default position with the following
settings:
• Minimum SID
• longitudinal tabletop centred
• transversal tabletop centred
• 0° angulation
• scan in the middle
• Compressor in parking position
• 0° tilting
Should in this phase a collision condition occur, the cycle is interrupted and the following
message is displayed:
"Cycle interrupted due to collision"
The equipment needs to be powered off, the life test interrupted and the table positioned in a
way such that no condition of collision occurs when the default position is in the process of
being reached.
A test will be divided into two cycles so that more movements may be executed simultaneously
in the same cycle.
At the end of the tilting movement, the number of cycles increases by one unit and the cycle
picks up again with the SID and compressor movement.
The number of exposures can be modified by pressing on the area of the selected cell and hold
it down for 1 second to enable the correction and by keying in the desired number by means of
the numerical keyboard (see chapter 11 – Data Entry procedure).
In case the room where the table is installed is of dimensions such that the complete excursion
of some movement is not possible, the test will stop the movement in proximity of the collision
limit.
Mode
Line 4 Coll. Height q Cass. Width r Cass. Height s Division t N. Expos. u Compr. Cone v
Line 6 collim XXXX angcmp XXXX tltsid XXXX ltbzrf XXXX scnttb XXXX center XXXX
Line 3: Spot
Spo t Film Device
k distance grid 1 from minimum SW (parking side) in mm
l gripping opening in mm
m cross.div trolley position from centre in mm
n shutters opening dimension in mm
o distance grid 2 from minimum SW (parking side) in mm
p film level limitation dimension for width collimator in cm
Note: If ERR is displayed in one of the fields relative to a potmeter, when the error occurs, it means that the potmeter was out of range
Line 4: Selections
q film level limitation dimension for height collimator in cm
r cassette width size
s cassette height size
t selected cassette subdivision program
u number of subdivisions still available
v compressor present = 1 not present = 0
Line 5: Selections
w compressor in field = 1 in park = 0
selected modality: 0 = direct 1 = spot film device 2 = digital
z
3 = tomo spot film device 4 = tomo digital 5 = angio
Note: If ERR is displayed in one of the fields relative to a potmeter, when the error occurs, it means that the potmeter was out of range
Note: The value 0 means that the button is idle, value 1 means that that button was active at the moment when the diagnosis occurred
1 2 3 4 5 6 7 8
A X X X X X X X X
NOTE: The display of the inputs referred to the switches is =1 when the switch is in the NC position and therefore the input is at 0 V. As to inputs from
external accessories, 0 means disabled input, 1 enabled input.
1 2 3 4 5 6 7 8
G X X X X X X X X
NOTE: The display of the inputs referred to the switches is =1 when the switch is in the NC position and therefore the input is at 0 V. As to inputs from external
accessories, 0 means disabled input, 1 enabled input.
Line 10:
10 : Display digital outputs
1 2 3 4 5 6 7 8
L X X X X X X X X
NOTE: The display of the outputs referred to the switches is =1 when the switch is in the NC position and therefore the output is at 0 V.
Line 11:
11 : Display digital outputs
1 2 3 4 5 6 7 8
S X X X X X X X X
NOTE: The display of the outputs referred to the switches is =1 when the switch is in the NC position and therefore the output is at 0 V.
When the diagnosis is moved - as a result of the most recent ones being introduced - to a memory position between 21 and 100, the view is reduced
and occupies one single line.
Because the view (page) consists of 2 lines, only 2 diagnosis at a time will be displayed.
The reduced diagnosis are displayed as follows:
Mode
Line 2
e diagnosis position in the memory (from 22 to 100 even numbers)
f diagnosis code
g+h diagnosis memorisation date (dd-mm-yy) and hour (hh-mm)
Mode
Line 6 Grid 2 Q q
Line 1
A number of exposures (all modalities SFD–DIG–DIR)
B number of fluoroscopy activation (all modalities SFD–DIG–DIR)
b fluoroscopy activation time (hours | minutes)
Line 2
C number of angulation motor movement activation
c activation time angulation motor movement (hours | minutes)
D number of scan motor movement activation
d activation time scan motor movement (hours | minutes)
E number of tilting motor movement activation
e activation time tilting motor movement (hours | minutes)
Line 3
F number of longitudinal tabletop motor movement activation
f activation time longitudinal tabletop motor movement (hours | minutes)
G number of SID motor movement activation
g activation time SID motor movement (hours | minutes)
H number of transversal tabletop motor movement activation
h activation time transversal tabletop motor movement (hours | minutes)
Line 4
I number of compressor motor movement activation
i activation time compressor motor movement (hours | minutes)
L number of grid 1 motor movement activation
l activation time grid 1 motor movement (hours | minutes)
M number of cassette close motor movement activation
m activation time cassette close motor movement (hours | minutes)
Line 5
N number of cross subdivision motor movement activation
n activation time cross subdivision motor movement (hours | minutes)
O number of shutters motor movement activation
o activation time shutters motor movement (hours | minutes)
P number of motors R and L belts cassette movement activation
p activation time motors R and L belts cassette movement (hours | minutes)
Line 6
Q number of grid 2 motor movement activation
q activation time grid 2 motor movement (hours | minutes)
All controls relative to the above-listed movements, which are present on both on the console
and on the double control, together with the fluoro and X-ray preparation movements, which
are present on the pedal or on the button, can be realised with a double microswitch ( S).
The first S is connected on the open contact (NA) and will be responsible for the movement
request.
The second is connected to a closed contact (NC) and is in a sequence with all second Ss that
feature the SF safety function.
The connection permits to have a single wire that is send to an input that serves both the
main processor and the supervisor.
The control logic bases the SF control on the correspondence of the SF circuit and of the
movements request.
Both conditions must be fulfilled in order to guarantee the activation of the movement (SF
circuit open, movements request circuit close).
Moreover, thanks to the feedback of the potentiometers of each "movement group", the main
processor is able to recognise if a movement is active without the need of an active request.
To complete the redundant checks on all levels, a second processor called supervisor is
connected via serial and parallel port (on the above-listed movements outputs) to the main
processor.
This permits to check a possible malfunctioning of the main processor that will have to manage
in a consistent way the serial and parallel data to the output ports.
2 relays called ready relays (K2 and K3) are controlled each by a processor and the contact are
connected in series.
The series of contacts interrupts the supply of the magnetic switch that supplies the power
circuits. If any anomaly is detected, one or both processors stop exciting the relative relay and
interrupt immediately any movement.
Thanks to such HW and SW logic, each part responsible for the movement is monitored and if a
failure on the movement activation circuit occurs, the system can detect and interrupt the
movement itself generating issuing an alarm.
13 FUNCTIONALITY
13.1 Collision
When activating one or more movements, one part of the equipment might reach the safety
limit beyond which a collision with the walls, the ceiling or the floor might occur.
This limit, set by means of group 200 cells, represents an imaginary insuperable software
barrier beyond which it is impossible to move from any part of the equipment.
When a part of the equipment reaches this limit, the movement that determined this situation
is stopped.
The message LIMIT REACHED is displayed until the activation of a control that informs the
operator that some movements might be disabled.
The parts of the table that can collide with walls, floor or ceiling are: the image intensifier, the
extremities of the main beam (corresponding to the extremities of the tabletop), the tube-
collimator group, the angulation motor and the tabletop in the case of 4-way version.
When one of these parts reaches such limit, movements commands that can bring the element
into collision are suddenly inhibited. On the other hand, movements commands that can bring
the elements away from the collision point are free.
In the case of 4-way version, if the collision limits with the floor are reached during tilting, the
SW will have to manage the return of the tabletop scan assembly maintaining the centre of the
patient unaltered.
Following are tables with the representations of the various conditions.
In the first pair of columns are indicated the angles in which the table (TILT) or the column
(ANG) could find themselves when the limit is reached.
"0" referred to the tilting means that the table is in the horizontal position, " +" indicates that
the table is inclined towards the vertical position and " -" that the table is inclined towards the
Trendelemburg.
"0" referred to the angulation indicates that the column is perpendicular to the table, " +"
indicates that the column is inclined towards the feet and " - " that it is inclined towards the
head.
The YES cells indicate that the movement is free, while the NO cells indicated that the
movement is inhibited.
The grey cells indicate that the potential collision will never occur for that particular
combination of tilting angle and angulation angle.
Permitted movements
movem ents table for Longitudinal tabletop collision – Left
wall
In the event the room where the appliance is installed has dimensional characteristics with
irregularities that would require the use of additional safety devices, Apollo EZ is able to protect
itself from collisions against these irregularities.
Depending on individual requirements, devices such as beam photocells or antenna switches
can be installed, which if activated avoid collisions between parts of the equipment and items in
the room.
The connection of these devices should be performed so as to send 0 V to the X0-17 input
when operating conditions are correct. When the device is intercepted, the input must open.
Once the safety device has been activated, the movement in progress stops and the display
shows the message:
<< LIMIT BARRIER >>
accompanied by an intermittent audible signal.
To exit from this condition and restore the normal functioning of the appliance, the movement
opposed to that which was present when the "limit barrier" condition occurred can be
independently activated.
All other commands (console joystick and push buttons on the table) are disabled until the
message "limit barrier" is removed.
If the safety barrier is broken during the following conditions:
• switching on the equipment
• tomo mode access
• angiography mode access
• execution of a Tomographic examination
• execution of an angiography step examination
• reaching a memorised position by means of command "0"
the movement is stopped and any active mode reset.
In this condition none of the joysticks or buttons on the table are active, and therefore to exit
the "limit barrier" condition is necessary to keep pressed at the same time function joystick (F)
together with the joystick of the movement causing the alarm in the opposite direction.
Warning
This procedure can be dangerous for the equipment in as much that an incorrect
manoeuvre could cause the contact of a table part against a part of the room.
Therefore the manoeuvre that will move the table away from the obstacle must be
identify before starting any movement. In any case, always proceed with caution and
activate the movement with brief and intermittent commands.
13.1.2 Anti-
Anti - entrapment system
The Apollo EZ table is equipped with anti-entrapment circuit to avoid that personnel in the
vicinity become trapped between the tabletop side support during vertical tilting. An additional
circuit is incorporated in this system which blocks the movement in the event the patient is
trapped between the tabletop and the beam or column (standard on 415/480V versions).
If the intervention occurs during one or more movements, no movement will be permitted until
the release of all joysticks that are pushed at the time, except for tilting and angulation that if
pressed in automatic will move of 5 degree in the opposite direction to that at the time of the
intervention.
Once the joysticks are released, if the tapping is freed through the automatic movements, all
movements will be enabled once again. Otherwise it is assumed that trapping is due to the
movement of the transversal tabletop, therefore only that movement will be enabled again in
the opposite direction to that detected when trapping occurred.
If the intervention of the safety device is detected with the machine ON without any
movement active, only tilting movement to the zero Trendelenbourg position will be permitted.
13.2 Compressor
Com pressor
When the compressor is not in the parking position (i.e. when the high end of run microswitch
is released) and column angulation is within ±15°, if angulation, tabletop or scan movements are
activated, the following will occur:
• the requested movement is not activated
• the compressor goes up for 1.5 sec or until compression force equal to 2 kg.
When one of these conditions is met, the ascent stops and a confirmation acoustic signal
is issued.
Release and push again the movement joystick to activate the desired movement.
The compressor ascent movement is as all Apollo EZ table movements of the "dead man" type,
therefore, if during the compressor ascent the joystick is released, the movement will stop.
By pushing the joystick once again, the ascent is enabled again for a time depending on the
current compression force without taking into account the previous partial ascent.
Only after completion of the ascent, the requested movement will be free.
It will then be possible to move the patient, scan or column without risk and, afterwards,
compress the patient once again.
Note
It is necessary to drive compressor up to park position (compressor has reached the
upper end of run) to angulated the column over the ±15° limits.
When the compressor is not in park position and the column angulation outside ±15°, by
activating the angulation movement, the compressor is taken back to the parking position
(confirmed by an acoustic signal).
Release and push again the movement joystick to activate the desired movement.
The compressor releases the compression by enabling the cone ascent for 240 msecs if the
potmeter exceeds the set dead band.
This happens typically when the patient makes a strong pressure on the cone, such as by
breathing.
Note
Every time the compressor activates the parking switch, the set value of the
compressor force is forced to 5 kg (11 lbs).
13.3 Collimator
The Apollo EZ table is designed to accept a square field collimator or a square field one with
iris.
The collimator operating modes are described in the User's Manual, paragraph "Collimator".
Warning
The storing of the table settings is assigned only to the user currently active (general
user or on of the five users that can be selected).
In the "SETUP" page of the Apollo EZ touch screen console, a key is associated to each of the
storable settings:
Preset position
MODE-1
MODE-2
MODE-3
The 3 conditions are stored in the EEPROM and initialized with default position when the
equipment is shipped from the factory.
To modify the storable setting, position the table as desired, select the tilting speed and
movement orientation depending on the need.
After setting the above-mentioned values, press for 5 seconds the "MODE-X" key you want to
associate with the actual configuration; the following window is displayed:
Exit
1. Keep the "Change Name 'MODE-X' " key pressed to set the name of the current table
position, using the alphanumeric keypad displayed; confirm with the key (Enter).
2. Store the settings by pressing and holding the "Store ' MODE-X ' " key until the following
message appears:
<< Position stored >>
3. Press the "Exit" key to exit the function.
Note
If the operator has stored positions different from the default ones, by performing
technical maintenance with access to group 100, such memorised velues will be lost.
The same will happen if the EEPROM is replaced.
To recall a previously stored selection, select an item from the "Preset position" drop down
menu on the main page, then move the joystick down (>0<).
Preset position
MODE-X
The stored positions are reached with contemporary movements, except for the angulation and
tilting axes.
The set tilting speed and movements orientation are also activated.
The following message is displayed when the operation is complete:
<< REACHED POSITIONS >>
Note
A "LIMIT REACHED" condition may occur while the movements are being activated
to reach the desired positions. In this case the procedure with be interrupted. The
component must be manually moved from its critical position and the stored position
recalled again.
When one stored position is selected in the "Preset position" field, the selection is maintained
both during the operating mode and at the start.
The function is inhibited when the angio and tomo modalities are active.
14 OPERATING MODES
• Park cassette
• I.I. or div. shutters
→ ←
• I.I. or div. collimator
Reset
↓ joystick
activation
I step:
• Movement towards posit.
exposure of:
- Cassette
- Shutters
- Collimator
• "Prep. request" out ON
↓ ↑
↑ • Reaching
final position of:
- Cassette NO Alarm
- Shutters → 060 →
- Collimator
within 5 secs.
YES ↓
II step at the generator
NO Alarm
• Input "generat. Ready"
→ 062 →
within 3 secs.
YES ↓
• Grid movement start
• Cell 223 time decrease
↑
• Cell 223 timeout :
"X-ray request" out ON
↓
↑ • "X-ray presence" input NO Alarm
ON within 5 secs. 064 →
YES ↓ ↑
Exposure in progress. I step:
• Pedal release with YES Alarm 067
"X-ray presence" input II step: →
ON Alarm 066
NO ↓ ↑
• "X-ray presence" input YES Alarm
ON after 6.5 secs. 065 →
NO
YES • I and II step release
← with X-ray end time. ←
("X-ray presence" OFF)
NO
• Subdivision and
shutter select. ←
• Park cassette
• I.I. or div. shutters
• I.I. or div. collimator
↓ ↑
I step: Reset
• Movement towards posit. joystick activation
exposure of: alarms
- Cassette
- Shutters ↑
- Collimator
• Progr. aborted
• "Prep. request" out ON
• Tray expelled
• Grid movem. start
↓
• Reaching final
position of: ↑
- Cassette NO Alarm
- Shutters → 061 →
- Collimator
within 5 secs.
YES ↓
II step at the generator NO Alarm
• Input "generat. Ready"
062 →
within 3 secs.
YES ↓
• "X-ray request" out ON ↑
↓
• "X-ray presence" input NO Alarm
ON within 5 secs. → 064 →
YES ↓ ↑
• "X-ray presence" input
(End X-ray) OFF within: NO Alarm
→ 6.5 sec → 065 →
- "X-ray request" out OFF
YES ↓
Program YES ↑
end → → →
NO ↓ ↑
• Pedal release with YES Alarm
program not end 063 → ↓
NO ↓
↑ • Next Tray ↑
• Out "X-ray request" ON
↓
YES • "X-ray presence" input NO Alarm Tray OUT or
← ON within 5 secs. 064 → park (sel. esp)
• Tomo ON (ang+speed+div)
→ • Cassette park ←
• I.I. or div. shutters
• I.I. or div. collimator
Centering joystick
↓
activation
I step:
• Movement towards position
exposure of:
- Cassette, Shutters,
↑
Collimator
• Grid movement start Reset joystick
• Column towards load activation
• "Prep. request." out ON
↓
↑ • Reaching final position of:
- Cassette, Shutters, NO Alarm ↑
Collimator → 060
→
within 5 secs.
YES ↓
II step at the generator NO Alarm
• Input "generat.Ready"
→ 062 →
within 3 secs.
YES ↓
14.3 Direct
Both the X0-14 and X0-15 inputs are deactivated.
The display indicates DIR.
The collimator can only be selected in manual mode.
It is not possible to carry out investigations in fluoro because the fluoro request output
towards the generator is disabled.
It is possible to carry out tests in exposure by operating the pedal commands (standard) or
button controls (optional).
The Spot Film Device in this phase does not perform the cassette-carrying tray movement,
grid vibration and shutters.
The preparation outputs, II step and exposure are sent to the generator and without
conditioning.
Neither the tomography nor the angio step mode can be selected.
14.4 Digital
This mode is possible only if the system is fitted with an image acquisition, storage and
processing device.
The Apollo EZ table has been designed and built to be connected to the digital system
DIVA HDE.
The display shows the initials DIG.
The collimator can only be selected in the automatic mode or automatic+hold.
It is possible to carry out investigations in fluoro and in exposure by operating the pedal
commands (standard) or button controls (optional).
The Spot Film Device, if fitted, does not perform the cassette and grid vibration movement in
this phase, while the shutters follow possible reductions carried out by the width axle of the
collimator.
• Digital ON
→ • No cass or cass in park ←
Reset
↓ joystick
activation
I step:
• Out "prep. request." ON
↓
II step at the generator NO Alarm
• Input "generat. Ready"
→ 062 →
within 3 secs.
YES ↓
• "X-ray request" out ON
↓ ↑
↑ • "X-ray presence" input NO Alarm
ON within 5 secs. → 064 →
YES ↓
Exposure in progress
DSI Exposure frequency ←
↓
YES • I and II step release ↑
← in any condition
NO ↓
• Digital ON
→ • Tomo ON (ang+speed) ←
Centering
↓ joystick
activation
I step:
Reset
• Column towards load
joystick
• Out "prep. request." ON
activation
↓
II step at the generator NO Alarm
• Input "generat. Ready"
062 →
within 3 secs.
YES ↓
• Movement column start
↓ ↑
• Column in Tomo angle
• "X-ray request" out ON
↓
↑ • "X-ray presence" input NO Alarm
ON within 5 secs. → 064 →
YES ↓ ↑
• X-ray presence ON for YES
more than 0.7 sec from Alarm
the end of Tomo (X-ray → 069 →
request OFF)
NO ↓
• I and II step release
YES with column centering
← ended
NO ↓
YES Centering joystick
← activation. Column at 0°
• Digital ON
→ • Tomo ON (ang+speed) ←
Centering
↓ joystick
activation
↑
I step:
Reset
• Column towards load
joystick
• Out "prep. request." ON
activation
↓
II step at the generator NO Alarm
• Input "generat. Ready"
→ 062 →
within 3 secs.
YES ↓
→
• Column movement start
↓ ↑
• Column in Tomo angle
• "X-ray request" out ON
↓
↑ ↑ • "X-ray presence" input NO Alarm
ON within 5 secs. → 064 →
YES ↓ ↑
• X-ray presence ON for YES
more than 0.7 sec from Alarm
the end of Tomo (X-ray → 069 →
request OFF)
NO ↓
NO • Start for next tomography
← X-ray control release
YES ↓
YES • I and II step release
← with column centering
ended
↑ NO ↓
YES Centering joystick
← activation. Column at 0°
14.6 Angiographic
Also this mode is only possible if the system is fitted with a digital acquisition system.
The step angiographic mode makes it possible to carry out tests of the peripheral artery
circulation limiting the quantity of contrast liquid administered to the patient.
The mode provides for the movement of the Spot Film Device-I.I.-tube group with "steps"
established by the size of the selected I.I. field.
When the Spot Film Device-I.I.-tube group completes the movement, the prep.- II step-
exposure make it possible to take digital images.
14.7 Stitching
Stitching function is only available if the system is fitted with wi-fi flat panel digital acquisition
system.
Stitching acquisition is done, after initial positioning, by acquiring 2, 3 or 4 contiguous RAD
images with automatic positioning. The acquisition step will be 35cm or 43cm, depending upon
the VDX detector adopted.
After each image is acquired, Apollo EZ table will move automatically to the next station;
PREP+RAD commands must be kept active during the whole procedure.
14.7.1 "Stitching
"St itching"
itching " block diagram
Alarms reset
I station: joystick activation
I step:
• "Prep. request" out ON
Progr. aborted
YES ↓
→ II step at the generator
generator NO Alarm
• Input "generat. Ready
→ 062 →
within 3 secs.
YES ↓
• "X-ray request" out ON ↑
↓
• Input "X-rays present" NO Alarm
ON within 5 sec. → 064 →
YES ↓ ↑
↑
NO Alarm
• PREP+RAD active
→ 078 →
YES ↓
YES End of
• Last station program
→
NO ↓
• Move to next station
after programmed delay
←
• Scan position reached
Switch
Swi tch Part Side
S44 SID Low
S45 SID High
S46 Transversal tabletop Inner / Outer
S48 Longitudinal tabletop Head / Feet
S50 Angulation Head
S51 Angulation Feet
S52 Scan Feet
S53 Scan Head
S54 Tilting Minimum / Maximum
After turning off the plant, the reason for this situation being created must be identified.
Once the cause has been removed, do the following to restore operation:
1. Press reset button S1 in the electrical cabinet.
2. Turn on the equipment keeping the button pressed down. Wait till the end of the
initialisation cycle.
3. Use the joystick that, in emergency situations, is in the direction opposite that of
the emergency.
4. Release button S1.
Warning
If button S1 is pressed, the series of emergency microswitches is overridden, so if the
cause of the failure is not removed, there will be the danger that the movement
follows its run without any electrical protection.
16 HARDWARE
HARDW ARE ADJUSTMENT
The SID movement motor installed on this table version can accept a wide voltage range.
Therefore both the 28V socket and the 35V socket of the auxiliary at 250 VA of the T4
transformer can be used (see chapter 25 – Functional drawing – page 1, ref. B8).
The setting described above, increases the motor speed to reduce the positioning time of the
SID if the maximum range available is implemented; this will result in a slight increase in motor
noise.
The average nominal speeds and times to cover the entire travel of 700 mm / 27.6" are:
Socket T4 Time
Tim e Speed
(VAC) (sec) (mm/sec)
28 22 32
35 17 41
17 THREE-
THREE - PHASE INVERTER
The inverters in the Apollo EZ table's electrical cabinet are used to control the 4 three-phase
motors responsible for the movements of: tilting, angulation, scan and longitudinal tabletop
(optional).
It is not possible to move at the same time tilting and angulation motors.
These two motors are driven by the same inverter and selected by two contactors.
Number of the
Function
terminal
L1-
L1 - L2-
L2 - L3 Input of three phase 400 V 50/60 Hz power supply.
Output of 380 V motor power supply with variable frequency between 0
and 120 Hz.
U-V-W In the case of inverted motor rotation these outputs must be reversed.
The inversion of the wires to terminals L1-L2-L3 does not change the
rotation.
This the pole of the open commentator for the RUN signal.
SE
This terminal is polarised at 0 V fixed.
This the common (reference 0 V) for the STR and STF signals.
SD
This terminal is polarised at fixed 0 V.
Input of speed reference from 0 to +10 V dc.
This input's usage dial is determined by parameter Pr73 (Pr73=1 range
0V +10V with motor stop at 0 V).
2 The speed of the motor will be proportional to the input signal in this
terminal.
This speed will be proportional to the maximum speed set in parameter
Pr38.
It is the reference 0 V for the V ref. signal (terminal 2).
5
This terminal is polarised at fixed 0 V.
This the output signal from the inverter (active at 0 V) present when the
inverter is in movement (acceleration and deceleration ramps included).
This output is used to enable the processor to recognise the end of the
RUN ramp.
Only when the inverter has finished the ramp will it be possible to
exchange the connection to the motor by means of the remote power
switch.
When this input is polarised at 0 V the inverter pilots the motor in the
STF
"FORWARD" direction.
When this input is polarised at 0 V the inverter pilots the motor in the
STR
"BACK" direction.
This pair of terminals depends on a relay contact inside the inverter. This
contact is closed when the inverter is under normal functioning
A-C conditions. In the case of failure the contact is open. The contacts of the
2 inverters are connected in series and depend on an input of input
board A2.
For a complete and detailed knowledge of the functioning of the inverter, it is however possible
to consult the dedicated manual, supplied together with the equipment.
The description of the function of all this data is reported briefly. Consult the inverter manual
provided together with the machine for more thorough information.
Inverter Data
Param. Function
A5 A6 A18 (*)
P. 0 20 20 20 Boost at the start (1st parameters set).
P. 3 50 50 50 Base frequency
P. 4 50 50 50 Multi-speed setting (hight speed) - RH
P. 5 30 30 30 Multi-speed setting (middle speed) - RH
P. 6 10 10 10 Multi-speed setting (low speed) - RL
P. 7 0 0 0 Acceleration ramp (1st parameter set).
P. 8 0 0 0 Deceleration ramp (1st parameter set).
Activation of the functioning block for excessive current
P. 9 0 0 0 to the motor. The message OL is shown on the display
without a stop in operations.
P. 19 400 400 400 Rated motor voltage.
P. 20 100 100 100 Reference frequency for acceleration / deceleration.
P. 30 0 1 0 Regenerative function selection.
P. 70 0 10 0 Special regenerative brake duty.
P. 71 23 23 3 Applied motor.
P. 72 13 13 13 PWM frequency selection.
P. 73 0 0 0 Analog input selection.
P. 77 2 2 2 Parameter write selection.
P. 79 2 2 2 Operation mode selection.
P. 80 1.1 1.1 0.4 Motor nominal power.
P. 90 7.324 7.424
7.424 37.908 Motor constant (R1).
P. 96 13 13 13 Auto tuning setting / status.
P.122 9999 0 9999 PU communiation check time interval.
P.125 90 90 120 Terminal 2 frequency setting gain frequency.
P.160 0 0 0 Extended function display selection.
P.180 3 3 3 RL terminal function selection.
P.192 11 11 11 ABC terminal function selection.
P.298 162 155 695 Frequency search gain.
P.503 0 0 0 Maintenance timer.
P.872 1 1 0 Input phase loss protection selection.
P.125
90 90 120 Terminal 2 frequency setting gain frequency.
(903)
(9 03)
(*) Inverter A18 is present only with 4 way tabletop version.
Listed here below are the most important parameters that, if modified, could jeopardise the
proper functioning of the equipment.
Nevertheless, this list is given to permit, in case of doubt or malfunctioning, to check the
parameter values are correct.
The description of the function of all this data is reported briefly. Consult the inverter manual
provided together with the machine for more thorough information.
Inverter Data
Param. Function
A5 A6 A18 (*)
P. 1 120 120 120 Maximum frequency.
P. 2 0 0 0 Minimum frequency.
P. 10 3 3 3 DC injection brake operation frequency.
P. 11 0.5 0.5 0.5 DC injection brake operation time.
P. 12 6 4 6 DC injection brake operation voltage.
P. 13 0.5 0.5 0.5 Start frequency.
P. 14 0 0 0 Load pattern selection.
P. 17 0 0 0 MRS input selection.
P. 18 120 120 120 High speed maximum frequency.
P. 37 0 0 0 Speed display 0/0.
P. 41 10 10 10 Up-to-frequency sensitivity.
P. 42 6 6 6 Output frequency detection.
P. 43 9999 9999 9999 Output frequency detection for reverse rotation.
P. 44 5 5 5 Second acceleration/deceleration time.
P. 45 9999 9999 9999
99 99 Second deceleration time.
P. 46 9999 9999 9999 Second torque boots.
P. 47 9999 9999 9999 Second V/F (base frequency).
P. 52 0 0 0 DU/PU main display data selection.
P. 55 50 50 50 Frequency monitoring reference.
P. 56 5.00 5.00 2.20 Current monitoring reference.
P. 60 0 0 0 Energy saving control selection.
P. 65 0 0 0 Retry selection.
P. 66 50 50 50 Stall prevention operation reduction starting frequency.
P. 67 0 0 0 Number of retries at fault occurrence.
P. 68 1 1 1 Retry waiting time.
P. 69 0 0 0 Retry count display erese.
Reset selection/disconnected PU detection/PU stop
P. 75 14 14 14
selection.
P. 78 0 0 0 Reverse rotation prevention selection.
P.128 0 0 0 PID action selection.
P.150 150 150 150 Output current detection level.
P.151 0 0 0 Output current detection signal delay time.
P.152 5 5 5 Zero current detection level.
P.153 0.5 0.5 0.5 Zero current detection time.
Inverter Data
Param. Function
A5 A6 A18 (*)
P.158 1 1 1 AM terminal function selection.
P.171 9999 9999 9999 Operation hour meter clear.
P.181 1 1 1 RM terminal function selection.
P.182 2 2 2 RH terminal function selection.
P.240 1 1 1 Soft-PWM operation selection.
(*) Inverter A18 is present only with 4 way tabletop version.
1. With the inverter in standby, press the MODE key; the display will show the writing P. 0.
0
2. Rotating the wheel, select the parameter position (P.XXX
P.XXX)
P.XXX you need to modify.
3. Press SET key to enter in parameter setting. It will be displayed the actual value.
4. Rotating the wheel, select the new parameter value.
5. Press SET key. The display will show the parameter position and the new selected value
in sequence.
6. Press SET key to store the new value in memory and reach the next parameter position.
7. Repeat the modification procedure or press the MODE key two times to exit the routine.
Warning
This procedure must only be activated if a malfunction has occurred in the inverter
that has compromised all the settings.
When this procedure is carried out, the default values in the various parameters will
be stored to memory.
So, before activating the functions of the equipment again the data in the table
"Parameters
Parameters with data modified by the Manufacturer"
Manufacturer must be reset and the
data stored in the "Parameters
Parameters that maintain the default settings"
settings table must
be checked.
Carry out the procedure described here below paying the greatest attention.
1. Following the indications in the "Access to the parameters for writing " paragraph
select the parameter P. 77 and change the data stored with the value 0 .
2. Select the parameter ALLC. and change the data stored with the value 1.
3. The Inveter will restore the default data making a new initialization.
Warning
The inverter is now in PU mode that means enabling movements also from the
keyboard. It is therefore essential to pay the greatest attention to the use of the
keys until the data stored in parameter P. 79 is restored at value 2.
4. It is now necessary to riconfigure the inverter parameters using the values present in
table at paragraph 17.3.1, modifying first the values of P. 77 to make access possible to all
the other parameters.
Figure 18-1
The correspondence between the various functions and the leds are listed below.
• H1 LED GROUP
LED 1 (Out 0) K7M (angulation) contactor call
LED 2 (Out 1)
1 spare
LED 3 (Out 2) R-L monitor image scan reverse
LED 4 (Out 3) HIGH-LOW monitor image scan reverse
LED 5 (Out 4) minimum I.I. field selection
LED 6 (Out 5) medium I.I. field selection
LED 7 (Out 6) maximum I.I. field selection
LED 8 (Out 7) activation of K2 ready relay controlled by the main processor
• H2 LED GROUP
LED 9 (Out 8) spare
LED 10 (Out 9) spare
LED 11 (Out 10) SID up
LED 12 (Out 11) SID down
LED 13 (Out 12) tilting motor brake release
LED 14 (Out 13) spare
LED 15 (Out 14) angulation motor brake release
LED 16 (Out 15) scan motor brake release
• H3 LED GROUP
LED 17 (Out 16) K6M (tilting) contactor call
LED 18 (Out 17) K8M (longitudinale tabletop) contactor call
LED 19 (Out 18) tabletop inside
LED 20 (Out 19) tabletop outside
LED 21 (Out 20) compressor up
LED 22 (Out 21) compressor down
LED 23 (Out 22) collimator lamp on
LED 24 (Out 23) spare
• H5 LED GROUP
LED 33 (Out 32) angulation-tilting inverter forward
LED 34 (Out 33) angulation-tilting inverter back
LED 35 (Out 34) scan inverter forward
LED 36 (Out 35) scan inverter back
LED 37 (Out 36) spare
LED 38 (Out 37) spare
LED 39 (Out 38) tomo 2nd speed
LED 40 (Out 39) tomo 3rd speed
• H6 LED GROUP
LED 41 (Out
(Out 40) fluoro request
LED 42 (Out 41) X-ray request
LED 43 (Out 42) preparation request
LED 44 (Out 43) digital preparation request
LED 45 (Out 44) second step preparation request
LED 46 (Out 45) tomo 1st speed
LED 47 (Out 46) tomo 4th speed
LED 48 (Out 47)
47 ) tomo 7° speed
• H7 LED GROUP
LED 49 (Out 48) tomo 20°
LED 50 (Out 49) tomo 30°
LED 51 (Out 50) tomo 45°
LED 52 (Out 51) X-ray presence without request
LED 53 (Out 52) Kv automatic-manual fluoro
LED 54 (Out 53) longitudinal tabletop inverter forward
LED 55
5 5 (Out 54) longitudinal tabletop inverter back
LED 56 (Out 55) angio step start
• H8 LED GROUP
LED 57 (Out 56) motion detection
LED 58 (Out 57) CCD x0
LED 59 (Out 58) CCD x4
LED 60 (Out 59) CCD x8
LED 61 (Out 60) grid 1 in field
LED 62 (Out 61) grid 1 in park
LED 63 (Out 62) cassette gripper opening
LED 64 (Out 63) cassette gripping
• H9 FUNCTION LEDS
LED 65 (Out 64) R cassette motor forward direction
LED 66 (Out 65) L cassette motor forward direction
LED 67 (Out 66) shutter motor forward direction / grid 2 in field
LED 68 (Out 67) cross subdivision motor forward direction
LED 69 (Out 68) R cassette motor disabled
LED 70 (Out 69) L cassette motor disabled
LED 71 (Out 70) shutter / grid 2 motor disabled
LED 72 (Out 71) cross subdivision motor disabled
To verify the correct position of the cassette, both longitudinal and transversal, and the
shutters it will be necessary to exclude the automatic function of the collimator that could
interfere in limiting the size.
To open the collimator axes automatically to the maximum size the value 1 must be introduced
in cell 262.
When this function is active, the message " Service function active" appears on the touch
screen.
The cassette positioning with regard to the X-ray main beam, during the exposure program,
must be correct.
If the separation lines are different, it means that the positioning of the cassette is wrong in all
its exposures.
To adjust the positioning properly it is possible to change the gains of the PDI function in the
cell of group 300.
The modification of these parameters can give rise to different effects, so an intervention
should only be made if the functioning of PDI method is well understood.
The correction can be carried out by working on these data because the position error is
determined by the difference that the error assumes in function of the length of the run.
Another possibility is that of changing the data in cell 219 which modifies all the runs of all the
subdivisions of the same quantity.
As a further possibility the single runs can be changed through the data set in the cells of
group 400.
By carrying out a series of radiograms with a cross layout on a 18x24 (8"x10") or 30x24
(12"x10") film, a continuous line of separation between the upper and lower radiograms of
approx. 2 mm (0.08") thick should be obtained.
If this does not happen it means that the transversal movement of the cassette is wrong.
Having determined whether the defect is caused by an asymmetrical movement with respect
to the centre or by too great or too small a movement for both sides, change the value set in
cells 459 and 460.
Warning
During these tests do not carry out limitations of the width collimator by means of
the joystick because the shutters follow the collimator.
Carry out the exposure on films of various formats and various subdivisions.
All the separator lines, in the various dimensions and subdivisions must be the same width.
If this not the case, change the data stored in the cells of group 400, in such a way as to
obtain what is required.
It must be remembered that the correction in mm made in the cells has the effect of
increasing and decreasing the final amount, equal to the value of the correction.
As an example a two line subdivision with cassette width of 30 cm (12") results in a minimum
separation line (0.5 mm).
Since the desired value is 2 mm, the quantity of 1.5 mm is added to the value in cell 549 (the
absolute value of which is not important as it depends on the characteristics of the system).
Note
Before starting the alignment of the X-ray beam it is necessary to ensure the
column is properly vertical and the hood is properly horizontal; these checks can be
carried out mechanically through the use of a level.
Tube focus
Collimator centre
min 200 mm
Correct centering
Figure 19-1
6. Move the collimator with respect to the hood using the adjustment possible in the
coupling cone: the collimator must move in the same direction as the circle furthest from
the tabletop should be moved to become concentric with the circle resting on the tabletop.
7. Move the tool in the same direction to reposition it in the centre of the X-ray field (in such
a way that both circles of the tool are completely represented in the image).
Repeat points 6 and 7, by trial and error, until the two circles are centred.
8. Move the SID to maximum and check that centering is maintained; if this not the case,
check the perpendicularity and repeat the procedure from point 4.
9. Once the concentricity has been achieved at both SID, the verticality of the alignment
between the focus of the hood and the centre of the collimator is guaranteed.
A B C 1 2 3 DELETE
D E F 4 5 6
CLEAR
7 8 9
HOME
0 . ENTER
2. Type password “20853” and press “ENTER”; the following will be displayed:
00 00 00 00 00 00 00 00
R W
00 00 00 00 00 00 00 00
A B C 1 2 3 DELETE
D E F 4 5 6
CLEAR
7 8 9
HOME
0 . ENTER
Note
The resettable alarms are highlighted with (*).
(*)
156 Serial line: timeout from The data does not arrive on The main processor and the supervisor are connected to each other parallel and in
Supervis. to Main µP the serial line from the series. If the serial dialogue is interrupted, this might be caused by a fault in one of
supervisor to the main µP for the 2 processors or by the lack of a 5 V power supply to the main processor (led H1
longer than 0.3 secs. CPU board) or the supervisor (led H2 CPU board). This failure is due to an
interruption on the line:
• supervisor TX
• main X-ray
157 Serial line: timeout from Main The data does not arrive on The main processor and the supervisor are connected to each other parallel and in
µP to supervisor the serial line from the main series. If the serial dialogue is interrupted, this might be caused by a fault in one of
µP to the supervisor for the 2 processors or by the lack of a 5 V power supply to the main processor (led H1
longer than 0.3 secs. CPU board) or the supervisor (led H2 CPU board). This failure is due to an
interruption on the line:
• X-ray supervisor
• TX main.
158 Single Fault open at switch The Single Fault circuit is not The SF circuit is a series obtained from the modules which the commands are on.
on. closed, therefore access Furthermore there are jumpers where optional elements could be connected. Measure
X13-24 is not at 0 V when the continuity of the various modules as follows with the table off:
the table is powered on. • command console: X8-12 with X41-7 (series joystick)
• pedal command: X41-7 with X41-5 (series X-ray commands)
• optional command: X42-7 with X42-5 (closure jumper)
• table-
table - mounted control: X16-23 with X16-24 (series keys)
• general terminal block: X0-25 with X0-24 (closure jumper).
159 Inputs close at switch on A joystick or double control When the table is powered up, a check is carried out on the level of all the inputs
button are active when the relative to the commands present on the touch screen console and the table-
table is powered up mounted command. If one of these keys or joystick is detected as active, then an
alarm is present. This could be caused by the incorrect activation of a control during
the powering up procedure or by a faulty command regarding the state of the table-
mounted control keys (PCB A8), of the X-ray commands (pedal and any button pad)
and the console joystick (to be checked directly on the buttons on the c.s.).
160 Single fault open SID up=1 / The SID lift control is The main and supervisor process are connected in series and parallel. The supervisor
serial bit 0=0 probably active when the SF receives on the parallel SID on command line (diode V7 CPU board) the activation
is open and the output port is signal, in addition to that of the SF is open. Nevertheless, the serial confirmation of
active, but the supervisor the movement, sent by the main processor is not present. The cause of failure may
does not receive the serial be due to a fault on the serial line by a processor, or a HW fault on the parallel line
confirmation of the control terminating on V7. Keep in mind that to confirm the second hypothesis, a fault on the
activation. SF line that is open must also be present.
161 Single fault open SID up=0 / The supervisor detects the The main and supervisor process are connected in series and parallel. The supervisor
serial bit 0=1 open SF and the serial recognises that the SF is open and it receives confirmation of the SID on movement
confirmation that the SID lift activation on the serial line. On the parallel SID on command line (Diode V7 CPU
control is active, but the board) there is not however an activation signal. The cause of failure may be due to a
output port is not active. fault on the serial line by a processor, or a HW fault on the parallel line terminating on
V7. Keep in mind that to confirm the first hypothesis, a fault on the SF line that is
open must also be present.
162 Single fault open SID The SID control down is What was said for alarm 160 applies in this case, too.
down =1 / serial bit 1=0 probably active as the SF is The parallel line involved is however the one for the SID down command and the
open and the output port is reference circuit is dependent on diode V8 (CPU board).
active, but the supervisor
does not receive the serial
confirmation of the control
activation.
163 Single fault open SID The supervisor detects the What was said for alarm 161 applies in this case, too.
down =0 / serial bit 1=1. open SF and the serial The parallel line involved is however the one for the SID down command and the
confirmation that the SID reference circuit is dependent on diode V8 (CPU board).
down control is active, but the
output port is not active.
164 Single fault open tabletop The tabletop in control is What was said for alarm 160 applies in this case, too.
in=1 / serial bit 2=0 probably active as the SF is The parallel line involved is however the one for the tabletop in command and the
open and the output port is reference circuit is dependent on diode V9 (CPU board).
active, but the supervisor
does not receive the serial
confirmation of the control
activation.
165 Single fault open tabletop The supervisor detects the What was said for alarm 161 applies in this case, too.
in=0 / serial bit 2=1. open SF and the serial The parallel line involved is however the one for the tabletop in command and the
confirmation that the tabletop reference circuit is dependent on diode V9 (CPU board).
in control is active, but the
output port is not active.
166 Single fault open tabletop The tabletop out control is What was said for alarm 160 applies in this case, too.
out=1 / serial bit 3=0. probably active as the SF is The parallel line involved is however the one for the tabletop out command and the
open and the output port is reference circuit is dependent on diode V10 (CPU board).
active, but the supervisor
does not receive the serial
confirmation of the control
activation.
167 Single fault open tabletop The supervisor detects the What was said for alarm 161 applies in this case, too.
out=0 / serial bit 3=1 open SF and the serial The parallel line involved is however the one for the tabletop out command and the
confirmation that the tabletop reference circuit is dependent on diode V10 (CPU board).
out control is active, but the
output port is not active.
168 Single fault open compressor The compressor lift control is What was said for alarm 160 applies in this case, too.
up=1 / serial bit 4=0 probably active when the SF The parallel line involved is however the one for the compressor up command and the
is open and the output port is reference circuit is dependent on diode V11 (CPU board).
active, but the supervisor
does not receive the serial
confirmation of the control
activation.
169 Single fault open compressor The supervisor detects the What was said for alarm 161 applies in this case, too.
up=0 / serial bit 4=1 open SF and the serial The parallel line involved is however the one for the compressor up command and the
confirmation that the reference circuit is dependent on diode V11 (CPU board).
compressor lift control is
active, but the output port is
not active.
Error Diagnosis
Diag nosis Cause of the alarm Corrective measures
170 Single fault open compressor The compressor down control What was said for alarm 160 applies in this case, too.
down =1 / serial bit 5=0. is probably active when the The parallel line involved is however the one for the compressor down command and
SF is open and the output the reference circuit is dependent on diode V12 (CPU board).
port is active, but the
supervisor does not receive
the serial confirmation of the
control activation.
171 Single fault open compressor The supervisor detects the What was said for alarm 161 applies in this case, too.
down =0 / serial bit 5=1. open SF and the serial The parallel line involved is however the one for the compressor down command and
confirmation that the the reference circuit is dependent on diode V12 (CPU board).
compressor down control is
active, but the output port is
not active.
172 Single fault open A5 inverter The A5 inverter enabling What was said for alarm 160 applies in this case, too.
enable=1/serial bit 6=0 control is probably active The parallel line involved is however the one for the inverter A5 (tilting-angulation)
when the SF is open and the movement control and the reference circuit is dependent on diodes V13 (forwards)
output port is active, but the and V14 (back) (CPU board).
supervisor does not receive
the serial confirmation of the
control activation.
173 Single fault open A5 inverter The supervisor detects the What was said for alarm 161 applies in this case, too.
enable=0/serial bit 6=1 open SF and the serial The parallel line involved is however the one for the inverter A5 (tilting-angulation)
confirmation that the A5 movement control and the reference circuit is dependent on diodes V13 (forwards)
inverter enabling control is and V14 (back) (CPU board).
active, but the output port is
not active.
174 Single fault open A6 inverter The A6 inverter enabling What was said for alarm 160 applies in this case, too.
enable=1/serial bit 7=0 control is probably active The parallel line involved is however the one for the inverter A6 movement (scan)
when the SF is open and the movement and the reference circuit is dependent on diodes V15 (forwards) and V16
output port is active, but the (backwards) (CPU board).
supervisor does not receive
the serial confirmation of the
control activation.
175 Single fault open A6 inverter The supervisor detects the What was said for alarm 161 applies in this case, too.
enable=0/serial bit 7=1 open SF and the serial The parallel line involved is however the one for the inverter A6 movement (scan)
confirmation that the A6 movement and the reference circuit is dependent on diodes V15 (forwards) and V16
inverter enabling control is (backwards) (CPU board).
active, but the output port is
not active.
176 Single fault open A18 inverter The A18 inverter enabling What was said for alarm 160 applies in this case, too.
enable=1/serial bit 8=0 control is probably active The parallel line involved is however the one for the inverter A18 movement
when the SF is open and the (longitudinal tabletop) and the reference circuit is dependent on diodes V27 (forwards)
output port is active, but the and V28 (backwards) (CPU board).
supervisor does not receive
the serial confirmation of the
control activation.
177 Single fault open A18 inverter The supervisor detects the What was said for alarm 161 applies in this case, too.
enable=0/serial bit 8=1 open SF and the serial The parallel line involved is however the one for the inverter A18 movement
confirmation that the A18 (longitudinal tabletop) and the reference circuit is dependent on diodes V27 (forwards)
inverter enabling control is and V28 (backwards) (CPU board).
active, but the output port is
not active.
180 Single fault close / SID up=1 The SF circuit is not active There is HW disagreement regarding the SID up command. The movement activation
(closed) but the SID lift circuit dependent on the diode V7 (Cupboard) is active but the SF series circuit is
control output port is active. closed. A check must be made to see which of the two circuits the fault is on.
181 Single fault closed / SID The SF circuit is not active There is HW disagreement regarding the SID down command. The movement
down=1 (closed) but the SID down activation circuit dependent on the diode V8 (CPU board) is active but the SF series
control output port is active. circuit is closed. A check must be made to see which of the two circuits the fault is
on.
182 Single fault close / tabletop The SF circuit is not active There is HW disagreement regarding the tabletop in command. The movement
in=1. (closed) but the tabletop activation circuit dependent on the diode V9 (CPU board) is active but the SF series
control output port is active. circuit is closed. A check must be made to see which of the two circuits the fault is
on.
183 Single fault close / tabletop The SF circuit is not active There is HW disagreement regarding the tabletop out command. The movement
out=1. (closed) but the tabletop out activation circuit dependent on the diode V10 (CPU board) is active but the SF series
control output is active. circuit is closed. A check must be made to see which of the two circuits the fault is
on.
184 Single fault close / The SF circuit is not active There is HW disagreement regarding the compressor up command. The movement
compressor up=1 (closed) but the compressor activation circuit dependent on the diode V7 (CPU board) is active but the SF series
lift control output port is circuit is closed. A check must be made to see which of the two circuits the fault is
active. on.
185 Single fault closed / The SF circuit is not active There is HW disagreement regarding the compressor down command. The movement
compressor down=1 (closed) but the compressor activation circuit dependent on the diode V12 (CPU board) is active but the SF series
down control output port is circuit is closed. A check must be made to see which of the two circuits the fault is
active. on.
186 Single fault close / A5 The SF circuit is not active There is a HW disagreement regarding the forward or back command of the A5
inverter enable =1. (closed) but the A5 inverter inverter (tilting – angulation). The movement activation circuit dependent on diodes
enabling command output V13 (forwards) and V15 (back) (CPU board) is active but the SF series circuit is closed.
port is active. A check must be made to see which of the two circuits the fault is on.
187 Single fault close / A6 The SF circuit is not active There is a HW disagreement regarding the forward or back command of the A6
inverter enable=1 (closed) but the A6 inverter inverter (scan). The movement activation circuit dependent on diodes V13 (forwards)
enabling command output and V15 (back) (CPU board) is active but the SF series circuit is closed. A check must
port is active. be made to see which of the two circuits the fault is on.
188 Single fault close / A18 The SF circuit is not active There is a HW disagreement regarding the forward or back command of the A18
inverter enable =1 (closed) but the A18 inverter inverter (longitudinal tabletop). The activation circuit of the movement on diodes V27
enabling command output (forward) and V28 (Back) (CPU) is active but the SF series is closed. A check must be
port is active. made to see which of the two circuits the fault is on.
191 The angulation potentiometer The angulation control is The main processor activates the angulation movement control and at the same time
does not move with active active but potentiometer checks the relative potentiometer is sending feedback signals in proportion to the
control. feedback does not detect the speeds applied. The alarm is generated when motion is detected or the speed is not
motion. aligned to that described in section 11.3.5. This phenomenon is more evident in the
first instants of motion due to mechanical inertia. When the difference between the
applied speed (Sp. ref. at the start up) and the speed detected (potentiometer
feedback) exceed a value obtained from a formula for a certain time, the alarm is
generated. The time and the value applied in the formula are set in cells 671 and 672.
The cause of the alarm may be:
• mechanical hardening of the movement
• potentiometer slip
• incorrect values of cells 671 and 672.
192 The scan potentiometer does The scan movement control isThe main processor activates the scan movement command and at the same time
not move with active control. active but the potentiometerchecks the relative potentiometer is sending feedback signals in proportion to the
feedback does not detect thespeeds applied. The alarm is generated when motion is detected or the speed is not
motion. aligned to that described in section 11.3.5. This phenomenon is more evident in the
first instants of motion due to mechanical inertia. When the difference between the
applied speed (Sp. ref. at the start up) and the speed detected (potentiometer
feedback) exceed a value obtained from a formula for a certain time, the alarm is
generated. The time and the value applied in the formula are set in cells 673 and 674.
The cause of the alarm may be:
• mechanical hardening of the movement
• potentiometer slip
• incorrect values of cells 673 and 674.
193 The tilting potentiometer does The tilting movement control The main processor activates the tilting movement control and at the same time
not move with the control is active but the checks the relative potentiometer is sending feedback signals in proportion to the
active. potentiometer feedback does speeds applied. The alarm is generated when motion is detected or the speed is not
not detect the motion. aligned to that described in section 11.3.5. This phenomenon is more evident in the
first instants of motion due to mechanical inertia. When the difference between the
applied speed (Sp. ref. at the start up) and the speed detected (potentiometer
feedback) exceed a value obtained from a formula for a certain time, the alarm is
generated. The time and the value applied in the formula are set in cells 675 and 676.
The cause of the alarm may be:
• mechanical hardening of the movement
• potentiometer slip
• incorrect values of cells 675 and 676.
194 The longitudinal tabletop The control for theThe main processor activates the longitudinal tabletop movement control and at the
potentiometer does not move longitudinal tabletopsame time checks the relative potentiometer is sending feedback signals in proportion
with the control active. movement is active but the to the speeds applied. The alarm is generated when motion is detected or the speed
movement is not detected is not aligned to that described in section 11.3.5. This phenomenon is more evident in
through the feedback of the the first instants of motion due to mechanical inertia. When the difference between
potentiometer. the applied speed (Sp. ref. at the start up) and the speed detected (potentiometer
feedback) exceed a value obtained from a formula for a certain time, the alarm is
generated. The time and the value applied in the formula are set in cells 677 and 678.
The cause of the alarm may be:
• mechanical hardening of the movement
• potentiometer slip
• incorrect values of cells 677 and 678.
195 The shutter potentiometer The shutters movement The main processor activates the shutter movement output and at the same time
does not move with active control is active but the checks the relative potentiometer is sending feedback signals in proportion to the
control. potentiometer feedback does speed applied. The alarm is generated when motion is detected or the speed is not
not detect the motion. aligned to that described in section 11.3.5. This phenomenon is more evident in the
first instants of motion due to mechanical inertia. When the difference between the
applied speed (Sp. ref. at the start up) and the speed detected (potentiometer
feedback) exceed a value obtained from a formula for a certain time, the alarm is
generated. The time and the value applied in the formula are set in cells 679 and 680.
The cause of the alarm may be:
• mechanical hardening of the movement
• potentiometer slip
• incorrect values of cells 679 and 680.
196 Right belt cassette encoder The R cassette movement The main processor activates the R cassette movement belt output and at the same
does not move with active belt is active but the motion time checks that the relative encoder is sending a feedback signal that is proportional
control. is not detected through the to the applied speeds. The alarm is generated when motion is detected or the speed
encoder feedback. is not aligned to that described in section 11.3.5. This phenomenon is more evident in
the first instants of motion due to mechanical inertia. When the difference between
the applied speed (Sp. ref. at the start up) and the speed detected (potentiometer
feedback) exceed a value obtained from a formula for a certain time, the alarm is
generated. The time and the value applied in the formula are set in cells 681 and 682.
The cause of the alarm may be:
• mechanical hardening of the movement
• encoder slip
• incorrect values of cells 681 and 682.
197 Left belt cassette encoder The L cassette movement The main processor activates the L cassette movement belt output and at the same
does not move with active belt is active but the motion time checks that the relative encoder is sending a feedback signal that is proportional
control. is not detected through the to the applied speed. The alarm is generated when motion is detected or the speed is
encoder feedback. not aligned to that described in section 11.3.5. This phenomenon is more evident in the
first instants of motion due to mechanical inertia. When the difference between the
applied speed (Sp. ref. at the start up) and the speed detected (potentiometer
feedback) exceed a value obtained from a formula for a certain time, the alarm is
generated. The time and the value applied in the formula are set in cells 681 and 682.
The cause of the alarm may be:
• mechanical hardening of the movement
• encoder slip
• incorrect values of cells 681 and 682.
198 The cross subdivision The cross subdivision The main processor activates the cross subdivision movement output and at the
potentiometer does not move movement control is active same time checks the relative potentiometer is sending feedback signals in proportion
with active control. but the potentiometer to the speed applied. The alarm is generated when motion is detected or the speed is
feedback does not detect the not aligned to that described in section 11.3.5. This phenomenon is more evident in the
motion. first instants of motion due to mechanical inertia. When the difference between the
applied speed (Sp. ref. at the start up) and the speed detected (potentiometer
feedback) exceed a value obtained from a formula for a certain time, the alarm is
generated. The time and the value applied in the formula are set in cells 685 and 686.
The cause of the alarm may be:
• mechanical hardening of the movement
• potentiometer slip
• incorrect values of cells 685 and 686.
201 Angulation potentiometer in There is no angulation The main processor did not activate the angulation movement but has detected
movement without request movement control but motion through the potentiometer feedback that there has been a movement greater than
is detected through the what was inserted in cell 452. The cause might be due to:
potentiometer feedback. • motion activation circuit HW fault (movement active without command by the
processor)
• excessive noise on the potentiometer signal
• break in the electric potentiometer connection circuit.
If the electrical potentiometer circuit is damaged, the alarm is generated when the
table is turned on. Thus as the circuit is broken, the pull up of the reference "moves"
the input as soon as it is powered up.
202 Scan potentiometer in There has been no scan The main processor did not activate the scan movement but has detected through
movement without request. movement output but motion the potentiometer feedback that there has been a movement greater than what was
has been detected by the inserted in cell 453. The cause might be due to:
potentiometer feedback. • motion activation circuit HW fault (movement active without command by the
processor)
• excessive noise on the potentiometer signal
• break in the electric potentiometer connection circuit.
If the electrical potentiometer circuit is damaged, the alarm is generated when the
table is turned on. Thus as the circuit is broken, the pull up of the reference "moves"
the input as soon as it is powered up.
203 Tilting potentiometer in There is no tilting movement The main processor did not activate the tilting movement but has detected through
movement without control. control but motion is detected the potentiometer feedback that there has been a movement greater than what was
through the potentiometer inserted in cell 454. The cause might be due to:
feedback. • motion activation circuit HW fault (movement active without command by the
processor)
• excessive noise on the potentiometer signal
• break in the electric potentiometer connection circuit.
If the electrical potentiometer circuit is damaged, the alarm is generated when the
table is turned on. Thus as the circuit is broken, the pull up of the reference "moves"
the input as soon as it is powered up.
20.2.1 Table
The movements controlled Check parameters not regulated If the values of the proportional, derived and integral gains are not properly regulated
with PDI of the Spot Film properly. the movements lose their smoothness. Before modifying the parameters take note of
Device (longitudinal cassette, that described in section 11.2.1. The axis gain values that present the defect are
transversal cassette and modified by reducing the set value with steps of approximately 5% until movements
shutters)
shutt ers) are not fluid. without jerks and vibrations are achieved. Remember that too great a reduction of
the gains makes the positioning imprecise.
The cassette gripping closure The control circuit is defective. Check fuse F22 positioned on the board A10 on the Spot Film Device. If it is seen to
movement is not working. be broken check that the gripping devices do not hit the walls of the Spot Film Device
violently, on opening.
Check that pressure switch S33 is not open (input X14-4), when the gripping devices
do not lock on the cassette.
The cassette is not recognised. Wrong use of the cassette or Verify the correspondence between the type of cassette used (cm or inches) and the
defect in the size reading data entered into cell 257. Verify that the cassette size used is one of those
photosensor. permitted (see sizes and divisions table in cm section 4.2 and inches section 4.3.
If the above is correct, check the functioning of the size reading photosensor (input
X14-2).
The cassette with height of 43 The gripping device opening is not The gripping devices do not open wide enough to accommodate the cassette with the
cm (17")
(1 7") is inserted with big enough. maximum height. Repeat the gripping device opening potentiometer adjustment
difficulty in the Spot Film procedure (see section 10.1.10). Verify the value of cell 228. If this value is excessive the
Device. grippers may not open enough.
The horizontal radiogram The transversal movement of the Check the adjustment of the cross subdivision potentiometer section 10.1.11.
subdivision line (cross cassette is wrong. Check the data in cells 459 and 460 and modify it remembering that:
subdivision program - optional) • if the separation line is too wide, the two numbers of the same quantity must be
is not the required size. reduced
• if the separation line is too narrow, the two numbers of the same quantity must be
increased
• modifying the 2 numbers of different quantities modifies the centering of the
cassette.
The fluoroscopic image shows The grid centering is not correct. The central position of the grid is not correct, it must be modified by changing the
one side darker than
th an the other. parameter in cell 220.
The fluoro exposure time is The positioning of the cassette, of The X-ray request to the generator control is sent when the cassette, shutters and
long. the shutters or of the collimator collimator reach the correct position. It must therefore be checked which of these
does not occur in the regular three movements is excessively slow in reaching the set point.
times. If the defect is on the cassette or shutter movements, check to see if there are
mechanical obstacles or take measures on the gains as described in the section 11.2.1.
If the defect is on the positioning of the collimator axles, the cause of this defect
might be due to:
A - HW adjustment of the gain circuit is wrong
The trimmers, R38 width, R48 height or R58 iris, must be adjusted as described
in the section 1.1
B - Dead band to small
The axle positioning dead band is too short, therefore the movement cannot stop
on the set point. This might be caused by a degradation of the potentiometer or
the increase in mechanical backlashes. The value in the cell relating to the axle
with the defect (229 width, 230 height, 231 iris) must be increased. The increase
in this value however increases the positioning tolerance thereby diminishing the
precision. It is therefore also possible to adjust the slowdown cells (232 width, 233
height, 234 iris), which reduce the speed of the axle before reaching the set point,
by a quantity equal to the value set. The increase of this parameter does not
influence the positioning precision but has, as its only drawback, an increase in
the positioning time on the axle.
Apollo EZ contains mechanical parts such as, bearings, cables and springs that are subject to
natural wear and tear.
To prevent any possibility of risk to the patient or operator, the equipment must be checked
and maintained regularly.
The electromechanical and electronic elements contained in the units must be adjusted
correctly to ensure they work perfectly and to give quality radiographic images, to provide top
levels of electrical safety and contain the dispersion of X-rays within the preestablished limits.
Note
The service technician must pay attention to everything that affects the electrical
safety of the equipment and make sure to reset all the conditions related to it, which
may have been affected during the repair, asking the user to repeat the electrical
safety test every time the operation involved the replacement of important parts or
the intervention has affected the design arrangements to ensure electrical safety.
Maintenance can be carried out by the operator (paragraph 21.1) or by the authorised technician
(paragraph 21.2).
Warning
In case of irregularities, high noisiness of motorised components, or failures, the
operator must immediately notify the Technical Staff.
Warning
Preventive and/or corrective operations must only be carried out by personnel
authorised and properly trained on part replacement and maintenance.
At the maintenance stage, the performance of the equipment is checked and if possible
adjusted in line with the indications in the previous chapters as summarised in the following
table:
(*) Note
Chains are tensioned in factory with an optimized force: in case of doubt, before
making any intervention, contact Villa Sistemi Medicali Service Support.
Warning
Only use original spare parts if components need to be replaced.
The relevant replacement instruction is supplied with the spare part.
Note
In case of replacement of the power supply cable, please refer to paragraph 5.2 for
cable features and paragraph 6.5 – point 9 – for the connection.
Warning
For cleaning and disinfecting operations carried out directly by the Authorised
Technician, you are recommended to adhere to the information reported in Chapter 8
of the User's manual.
(*) Note
Replacement of angulation chain requires the tools kit code 6690702100.
Note
Spare part code includes replacement instruction.
At replacement, Field Service Engineering will fill the column "Next replacement date" adding
10 years to the replacement date in both User's and Service Manuals.
Reason of the
Angulation chain Next Execute by
replacement: Date /
replacement
replacement date replacement date organization
corrective maintenance Signature
(dd/mm/yyyy) (mm/yyyy) (FSE name)
/ planned maintenance
Configuration MENU
Console ID
MSG language
HW Inputs Test
3. Press "Console ID" button, the touch screen will show a numerical keyboard asking to
input the access key:
Enter Key 1 2 3
4 5 6
ESC 7 8 9
C 0 -
.
0
4. Input the key code "20090".
ESC
5. Press key (Enter) to confirm or to exit; on the "insert address key" area, the
touch screen will show the message:
Console 1
Console 2
Reboot
6. Press the key identifying the desired console number (Console 1 or Console 2). The last
line will be updated with the new selection.
Reboot
7. Hit the key on the touchscreen: the table will restart in normal
mode.
EN[m] IT[f]
Reboot
EN[f] DE[f]
FR[m] RU[f]
FR[f] AR[m]
ES[m] ZH[f]
ES[f] HU[f]
IT[m] PL[f]
Select the language for language 1, 2 3 fields, then reboot the table.
23 FIRMWARE UPGRADE
Mode
Firmware update
YES NO
Warning
Do NOT switch OFF the system and do NOT remove the USB memory key during
the update procedure.
REMOVE the USB memory key ONLY when the system is completely OFF.
YES
6. Press key ; the load of the updated file will start.
The following screen is displayed:
Mode
Firmware update
Mode
Firmware update
7. Switch OFF the Apollo table and remove the USB key.
8. Set switch 4 in the DIP-switch assembly in the cabinet CPU PCB to ON.
9. Switch ON the table and access to edit position 719 as described on paragraph 11.5.
10. Reset this cell: all the alarm memory and anatomic pre-set positions stored in the Flash
eprom will be reset.
11. Switch OFF the unit and restore the standard operation mode (all switches in the
DIP-switch assembly in the cabinet CPU PCB to OFF).
Warning
Do NOT switch OFF the system and do NOT remove the USB memory key during
the update procedure.
REMOVE the USB memory key ONLY when the system is completely OFF.
8. Restore the standard operation mode (all switches in the DIP-switch assembly in the
cabinet CPU PCB to OFF).
H8 J3
Q2
On USB stick, the following three text files have been saved:
• SNxxxxxxxxxxxxxxxx_srv.txt containing the values of the cells 100, 200, 300, 400
• SNxxxxxxxxxxxxxxxx_cyc.txt containing the values of the counters of the table cycles
• SNxxxxxxxxxxxxxxxx_dgn.txt containing the log of the diagnosis.
where in place of "xxxxxxxxxxxxxxxx" there is the system Serial Number.
Warning
Do NOT switch OFF the system and do NOT remove the USB memory key during
the update procedure.
REMOVE the USB memory key ONLY when the system is completely OFF.
Microwindows Calibration
Touch the cross hairs firmly to
calibrate your handheld
3. Tap with a finger on the cross-shaped symbol present on the upper-left corner; a new
point of reference will be shown on the upper-right corner.
4. Tap on the new point of reference.
5. Repeat for each corner to complete the calibration process.
After the fourth point has been tapped, the calibration process will be ended and the touch
screen will automatically enter the standard operation mode.
Warning
When touch screen calibration has started, it must be completed, otherwise the touch
screen will not work properly.
Pin number
Connection number
2. Functional diagram
d iagram
27. A27
A27 Hardware
Hardw are Single Fault board layout
28. A27
A27 Hardware Single Fault board wiring diagram
APOLLO EZ
Power supply diagram 1
Code 39849008 - Rev. 0 Page 1 of 1
APOLLO EZ
Functional drawing 2
Code 39849202
39849 202 - Rev. 1 Page 1 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 2 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 3 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 4 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 5 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 6 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 7 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 8 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 9 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 10 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 11 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 12 of 17
17
APOLLO EZ
Functional
Func tional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 13 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 14 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 15 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 16 of 17
17
APOLLO EZ
Functional drawing 2
Code 398492002
39849 2002 - Rev. 1 Page 17 of 17
17
APOLLO EZ
Cabinet layout
4- way tabletop version
3
Code 83848
8384853
530
530 - Rev. 0 (380/400V)
Page 1 of 1
Code 83848590
83848 590 – Rev. 0 (415/480V)
APOLLO EZ
Cabinet layout
2 - way tabletop version
4
Code 83848545
83848 545 - Rev. 0 Page 1 of 1
APOLLO EZ
A1 CPU board layout 5
Code 58848436
58848 436 - Rev. 4 Page 1 of 1
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 1 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 2 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 3 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849
39849 162 - Rev. 0 Page 4 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 5 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 6 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 7 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 8 of 9
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 9 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 10 of 11
APOLLO EZ
A1 CPU board
wiring diagram
6
Code 39849162
39849 162 - Rev. 0 Page 11 of 11
1 2 3 4 5 6 7 8 9 10 11
APOLLO EZ
H
A2 Input board
layout
la yout
7
Code 58838152 - Rev. 6 Page 1 of 1
APOLLO EZ
A2 Input board
wiring diagram
8
Code 39849027 - Rev. 8 Page 1 of 4
APOLLO EZ
A2 Input board
wiring diagram
8
Code 39849027 - Rev. 8 Page 2 of 4
APOLLO EZ
A2 Input board
wiring diagram
8
Code
Co de 39849027 - Rev. 8 Page 3 of 4
APOLLO EZ
A2 Input board
wiring diagram
8
Code 39849027 - Rev. 8 Page 4 of 4
APOLLO EZ
A3 Output board
layout
9
Code 58848420
58848 420 - Rev. P Page 1 of 1
APOLLO EZ
A3 Output board
wiring diagram
10
Code 39849194
39849 194 - Rev. 0 Page 1 of 8
APOLLO EZ
A3 Output board
wiring diagram
10
Code 39849194
39849 194 - Rev. 0 Page 2 of 8
APOLLO EZ
A3 Output board
wiring diagram
10
Code 39849194
39849 194 - Rev. 0 Page 3 of 8
APOLLO EZ
A3 Output board
wiring diagram
10
Code 39849194
39849 194 - Rev. 0 Page 4 of 8
APOLLO EZ
A3 Output board
wiring diagram
10
Code 39849194
39849 194 - Rev. 0 Page 5 of 8
APOLLO EZ
A3 Output board
wiring diagram
10
Code 39849194
39849 194 - Rev. 0 Page 6 of 8
APOLLO EZ
A3 Output board
wiring diagram
10
Code 39849194
39849 194 - Rev. 0 Page 7 of 8
APOLLO EZ
A3 Output board
wiring diagram
10
Code 39849194
39849 194 - Rev. 0 Page 8 of 8
APOLLO EZ
A4 Spot film device
motor driver board layout
11
Code 58838156 - Rev. 1 Page 1 of 1
APOLLO EZ
A4 SFD motor driver board
wiring diagram
12
Code 39849031 - Rev. 1 Page 1 of 3
APOLLO EZ
A4 SFD motor driver board
wiring diagram
12
Code 39849031 - Rev. 1 Page 2 of 3
APOLLO EZ
A4 SFD motor driver board
wiring diagram
12
Code 39849031 - Rev. 1 Page 3 of 3
1 2 3 4
C1 C2 C3
R S T
SCHEMA ELETTRICO
(ELECTRICAL DIAGRAM)
R S T
D
C1 C2 C3
3x1uF-275Vac-X2
F APOLLO
APO LLO EZ
A7 Filter board
wiring diagram & layout
13
Code 58848154 - Rev. 2 Page 1 of 1
1 2 3 4 5 6 7 8
C K1 X16
N2 N1 N3
58843050/
S13 S16 S6 S60 Z1 Z3
Z2 S10 S4
R1 LC
S18 S9 S12 S5 S2
S8
V4
S17 S7 S15 S11 S3
S14
C2
C1 V5 V2
V1 V3
R2
F APOLLO EZ
A8 Table Side operation
board layout
14
Code 58843050 - Rev. 1 Page 1 of 1
APOLLO EZ
A8 Table Side Operation
board wiring diagram
15
Code 39849019 - Rev. 1 Page 1 of 1
1 2 3 4 5 6 7 8
B C5
Z1
X64
R12
R11
R15
U1 N3
58833015/ 2
R5
R8
C6
V5
X61 C11 C10 C13 X65
X50 7 8
1
N2
1 1
1
1 X54 1 X59 TP3
R10
TP2
R9
X60
R14
C12
R7
R6
R4
R25
3 1 C16
K1
R28
C8 N4
13 V9 V4 C3 C2 TP6
C7
X51 25 R2
R13
R26
R23
R27
R24
H1 C9
5 1 2 TP5
R20 C4
1 9 R21 C17
R35
R36
R31
R30
R34
R32
N5
V8
R3
1 X55 1 X58 X63 R1 C1
R29
12 10
V6
V7
R19
R18
R16
R22
6
X52 1
R17
R40
C 1 C14
R38
R39
R37
X66 V2 V3 C21 TP4 C20 C19 X108 C15
1
14
1
1
1
N1
1
X103 X102
VPOT
GRID
LONG
GRIP
SHUT
0V
1
X14 X62 X44 1 1
R33
1
X67 F22 F23 X101
F APOLLO EZ
A10 Spot Film Device
Devi ce
board layout
16
Code 58843054 - Rev. 0 Page 1 of 1
APOLLO EZ
A10 Spot Film Device
wiring diagram
17
Code 39849045 - Rev. 0 Page 1 of 2
1 2 3 4 5 6 7 8
R18 V6
TP1 3,3k 1N4148
C15 C14
R19 V7 4,7nF 2,2nF
A 100k 1N4148
X62
R21
1 +12V 15ohm 2.5W
(Cassette inserted sensors)
N4A
Sensori cassetta inserita
2 trasmittitore
TLC274 R17 +12V
11
R20 100k
2,2k - 2 R16 K1
V9 1 100k
BD679 3
+ M5rele+
M5+
X63
4
M6rele-
scheda - 24Vac
Alimentazione
(Power supply
M6-
pcb - 24Vac)
1 +12V
ricevitore X44
2 V2
+12V
3 R22 1N4007 0Vac 1
4 100k
B 24Vac 2
2
V1
N1 W02G
R23 LM317 R1
11k TP3 +21V 10ohm 1W
+ -
3 1 1 4
GND
OUT IN
+12V V3
2
1N4007
R2
3
240ohm +
+12V C1
TP2
C18 47uF
0,1uF V4 C4 C3 C2
N4C N4B 1N4007 0,1uF + 10uF R3 + 10uF
4
4
10 + 0,1uF 0,1uF
8 5 +
C 9 -
6 -
7
R28 R24
11
1k 22k 11
R25 R26
2,2k 2,2k
R27 +12V
100k
V5
1N4007
H1
V8 C19 +12V LED
1N4148 1nF N2
D +12V LM7812 +21V
R29 3 1
GND
820k R38 R39 R40 OUT IN
3,3k 1,8k 3,3k
+12V U2
TP4
2
TLP504 C13 + C12 + C11 C10
1 8 0,1uF 22uF 22uF 0,1uF
N4D
4
TP6
+12V
R36
2,2k
N5B
8
LM358
R34 5 TP5
+
2,2k 7
6 -
4
F APOLLO EZ
A10 Spot Film Device
wiring diagram
17
Code 39849045 - Rev. 0 Page 2 of 2
1 2 3 4
B
4 3 4 3 4 3
2 1 2 1 2 1
1
X83 X84 X88
2 2 2 2 X97
C
X20
1 1 1 1
X75 X77 X78 X81
V1 V2
1
2 2
2
X96
1 1 1
X92 X93
1
X98
D X46
F APOLLO EZ
A11 Base unit connector
board layout
18
Code 58841050 - Rev. 3 Page 1 of 1
1 2 3 4 5 6 7 8
X83.1
X83.2
X83.3
X83.4
X84.1
X84.2
X84.3
X84.4
X88.1
X88.2
X88.3
X88.4
X97.1 X20.15
X97.2 X20.14
X97.3 X20.13
X97.4 X20.11
X97.5 X20.10
X20.7
X97.6 X20.6
B X97.7 X20.5
X97.8 X20.4
X97.9 X20.3
X20.2
X20.1
X46.19
X46.6
X46.21
X46.8
X46.20
D X46.7
X46.25
X46.18
X46.12
V2
X46.5 MR852
V1
MR852
E
X92.1
X93.1
X96.1
X98.1
X98.2
X75.1
X77.1
X78.1
X81.1
X92.2
X93.2
X96.2
X98.3
X98.4
X98.5
X98.6
X75.2
X77.2
X78.2
X81.2
F APOLLO EZ
A11 Base unit connector
board wiring diagram
19
Code 39849010 - Rev. 0 Page 1 of 1
25-62 Villa Sistemi Medicali
Service Manual - Wiring diagrams and drawings
Rev. 2
APOLLO EZ
A12 Cabinet connector
board layout
20
Code 58848404
58848 404 - Rev. 0 Page 1 of
of 1
APOLLO EZ
A12 Cabinet connector
board wiring diagram
21
Code 39849190
39849 190 - Rev. 0 Page 1 of 1
APOLLO EZ
A13 Angulation connector
board layout
22
Code 58843052 - Rev. 5 Page 1 of 1
APOLLO EZ
A13 Angulation connector
board wiring diagram
23
Code 39849021 - Rev. 2 Page 1 of 1
APOLLO EZ
A14 Touch Screen CPU
board layout
24
Code 58848442
58848 442 - Rev. 2 Page 1 of 1
APOLLO EZ
A14 Touch Screen CPU
board wiring diagram
25
Code 39849187
39849 187 - Rev. 2 Page 1 of 6
APOLLO EZ
A14 Touch Screen CPU
board wiring diagram
25
Code 39849187
39849 187 - Rev. 2 Page 2 of 6
APOLLO EZ
A14 Touch Screen CPU
board wiring diagram
25
Code 39849187
39849 187 - Rev. 2 Page 3 of 6
APOLLO EZ
A14 Touch Screen CPU
board wiring diagram
25
Code 39849187
39849 187 - Rev. 2 Page 4 of 6
APOLLO EZ
A14 Touch Screen CPU
board wiring diagram
25
Code 39849187
39849187 - Rev. 2 Page 5 of 6
APOLLO EZ
A14 Touch Screen CPU
board wiring diagram
25
Code 39849187
39849 187 - Rev. 2 Page 6 of 6
APOLLO EZ
A17 Tabletop control
board wiring diagram & layout
26
Code 58846140 - Rev. 4 Page 1 of 1
APOLLO
A27
A2 7 Hardware single fault
board layout
27
Code 58848392
58848392 - Rev. 2 Page 1 of 1
APOLLO
A27 Hardware single fault
board wiring diagram
28
Code 3984915
39849154 - Rev. 1 Page 1 of 1
26 SPARE PARTS
1 – Electrical components
3 – Connections cables
4 – Touch screen
5 – Keyboard (TSO)
6 – Tabletop movement
7 – Compressor group
12 – Anti-
Anti - entrapment device
13 – Collimator
1 6 – Tabletop
17 – Foot control
18 – Tabletop accessories
1 – ELECTRICAL COMPONENTS
3 – CONNECTION CABLES
4 – TOUCH SCREEN
2 (5x)
5 – KEYBOARD (TSO)
6 – TABLETOP MOVEMENT
3
2
6
5
7 – COMPRESSOR GROUP
M14
S26
R10
S27 S30
-- 6283310300 Encoder
M5
5
7
6
3
4
1
M6
8
10
2
3
1
M 7
2
M 8
R6
M9
S33
R7
M10
12 – ANTI-
ANTI- ENTRAPMENT DEVICE
1 3 – COLLIMATOR
m
i2
g2
i1
d
g1
f
b
e
l
a2
c
n
a1
s
q
v
p
t2
t1
m
i2
g2
i1
d
g1
f
b
e
l
a2
c
n
a1
s
q
v
p
t2
t1
16 – TABLETOP
17 – FOOT CONTROL
1 8 – TABLETOP ACCESSORIES
2
3 1
4
5