Directional Insert EN
Directional Insert EN
Directional Insert EN
REF 708255
Intended Use
The NOVA Lite IgG F-Actin kit is an indirect immunofluorescent assay (IFA) on rat intestine epithelial
substrate for the screening and semi-quantitative determination of IgG anti-F-Actin antibodies in human
serum. The presence of IgG anti-F-Actin antibodies can be used in conjunction with clinical findings and
other laboratory tests to aid in the diagnosis of autoimmune liver diseases such as autoimmune hepatitis
(AIH) and primary biliary cirrhosis (PBC).
Reagents
1. F-Actin Slide; 6 wells/slide, with desiccant
2. FITC IgG Conjugate (Goat), 1 vial of fluorescein labeled in buffer containing Evans Blue and
0.09% sodium azide
3. IgG F-Actin Positive Control, 1 vial of buffer containing 0.09% sodium azide and human serum
antibodies to F-Actin, prediluted
4. IFA System Negative Control, 1 vial of buffer containing 0.09% sodium azide and no human
serum antibodies to F-Actin, prediluted
5. PBS Concentrate (40x), 2 vials
6. Mounting Medium, 0.09% sodium azide, 1 vial
7. Coverslips
Warnings
1. All human source material used in the preparation of controls for this product has been tested and
found negative for antibody to HIV, HBsAg, and HCV by FDA cleared methods. No test method,
however, can offer complete assurance that HIV, HBV, HCV or other infectious agents are
absent. Therefore, the IgG F-Actin Positive Control and IFA System Negative Control should be
handled in the same manner as potentially infectious material.8
1
2. Sodium Azide is used as a preservative. Sodium Azide is a poison and may be toxic if ingested or
absorbed through the skin or eyes. Sodium azide may react with lead or copper plumbing to form
potentially explosive metal azides. Flush sinks, if used for reagent disposal, with large volumes of
water to prevent azide build-up.
3. Use appropriate personal protective equipment while working with the reagents provided.
4. Spilled reagents should be cleaned up immediately. Observe all federal, state and local
environmental regulations when disposing of wastes.
Precautions
1. This product is for in vitro diagnostic use.
2. Substitution of components other than those provided in this kit may lead to inconsistent results.
3. Incomplete or inefficient washing of IFA wells may cause high background.
4. Adaptation of this assay for use with automated sample processors and other liquid handling
devices, in whole or in part, may yield differences in test results from those obtained using the
manual procedure. It is the responsibility of each laboratory to validate that their automated
procedure yields test results within acceptable limits.
5. A variety of factors influence the assay performance. These include the starting temperature of
the reagents, the strength of the microscope bulb used, the accuracy and reproducibility of the
pipetting technique, the thoroughness of washing and the length of the incubation times during
the assay. Careful attention to consistency is required to obtain accurate and reproducible results.
6. Strict adherence to the protocol is recommended.
Storage Conditions
1. Store all the kit reagents at 2-8°C. Do not freeze. Reagents are stable until the expiration date
when stored and handled as directed.
2. Diluted PBS buffer is stable for 4 weeks at 2-8°C.
Specimen Collection
This procedure should be performed with a serum specimen. Addition of azide or other preservatives to
the test samples may adversely affect the results. Microbially contaminated, heat-treated, grossly
hemolyzed, or lipemic specimens containing visible particulates should not be used.
Following collection, the serum should be separated from the clot. CLSI (NCCLS) Document H18-A3
recommends the following storage conditions for samples: 1) Store samples at room temperature no
longer than 8 hours. 2) If the assay will not be completed within 8 hours, refrigerate the sample at 2-8°C.
3) If the assay will not be completed within 48 hours, or for shipment of the sample, freeze at -20°C or
lower. Frozen specimens must be mixed well after thawing and prior to testing.
Procedure
Materials provided
5 F-Actin Slide (6 well)
1 7mL FITC IgG Conjugate
1 0.8mL IgG F-Actin Positive Control
1 0.5mL IFA System Negative Control
2 25mL PBS Concentrate (40x)
1 7mL Mounting Medium
1 Package of 10 Coverslips
2
Method
Before you start
1. Bring all reagents and samples to room temperature (20-26oC).
2. Dilute PBS Concentrate: IMPORTANT: Dilute the PBS Concentrate 1:40 by adding the contents
of the PBS Concentrate bottle to 975mL of distilled or deionized water and mix thoroughly. The
PBS buffer is used for diluting patient samples and as a wash buffer. The diluted buffer can be
stored for up to 4 weeks at 2-8°C.
3. Dilute Patient Samples:
a. Initial Screening: Dilute patient samples 1:40 with diluted PBS buffer (i.e., add 50μL of
serum to 1.95mL of PBS buffer).
b. Titration: Make serial 2-fold dilutions from the initial screening dilution for all positive
samples with diluted PBS buffer (i.e. 1:80, 1:160, 1:320... to endpoint).
Assay procedure
1. Prepare Substrate Slides: Allow the substrate slide to reach room temperature prior to removal
from its pouch. Label it with pencil and place it in a suitable moist chamber. Add 1 drop (20-25μL)
of the undiluted positive and the negative control to wells 1 and 2 respectively. Add 1 drop (20-
25μL) of diluted patient sample to the remaining wells.
2. Slide Incubation: Incubate the slide for 30 ± 5 minutes in a moist chamber (a dampened paper
towel placed flat on the bottom of a closed plastic or glass container will maintain the proper
humidity conditions). Do not allow the substrate to dry out during the assay procedure.
3. Wash Slides: After incubation, use a plastic squeeze bottle or pipet to gently wash off the serum
with diluted PBS buffer. Orient the slide and stream of PBS buffer so as to minimize wash-over of
samples between wells. Avoid directing the stream directly onto the wells to prevent
substrate damage. If desired, place the slides in a Coplin jar of diluted PBS buffer for up to 5
minutes.
4. Addition of Fluorescent Conjugate: Shake off the excess PBS buffer. Place the slide back in
the moist chamber and immediately cover each well with a drop of fluorescent conjugate.
Incubate the slides for an additional 30 ± 5 minutes.
5. Wash Slides: Repeat Step 3.
6. Coverslip: Coverslip procedures vary from lab to lab; however, the following procedure is
recommended:
a. Place a coverslip on a paper towel.
b. Apply mounting medium in a continuous line to the bottom edge of the coverslip.
c. Shake off the excess PBS buffer and touch the lower edge of the slide to the edge of the
coverslip. Gently lower the slide onto the coverslip in such a way that the mounting
medium flows to the top edge of the slide without air bubble formation or entrapment.
Quality Control
The IgG F-Actin Positive Control and IFA System Negative Control should be run on every slide to
ensure that all reagents and procedures perform properly. Additional suitable control sera may be
prepared by aliquoting pooled human serum specimens and storing at < -70oC. In order for the test
results to be considered valid, all of the criteria listed below must be met. If any of these are not met, the
test results should be considered invalid and the assay repeated.
1. The undiluted IgG F-Actin Antibody Positive Control must be > 3+.
2. The IFA System Negative Control must be negative.
Interpretation of Results
Negative Reaction: A sample is considered negative if specific staining as described below in the
section “Positive Reaction” is less than or equal to the IFA System Negative Control. Samples can exhibit
various degrees of specific or background staining to other cellular components, but be negative for IgG
anti- F-Actin antibodies.
Positive Reaction: A sample is considered positive if specific staining at a greater intensity than the IFA
System Negative Control is observed in the F-Actin fibers surrounding most cells in a hexagonal pattern,
and sometimes observed in the fibers crossing over cells.
Determine the fluorescence grade or intensity using these criteria:
4+ Brilliant apple green fluorescence
3+ Bright apple green fluorescence
2+ Clearly distinguishable positive fluorescence
1+ Lowest specific fluorescence that enables the F-Actin staining to be clearly differentiated
from the background fluorescence.
3
Pattern Interpretation: A variety of patterns of nuclear and/or cytoplasmic staining can be exhibited
depending on the types and relative amounts of autoantibodies present in the sample. Only the pattern
described above in the heading “positive reaction” should be considered positive for IgG anti-F-Actin
antibodies. All other patterns should be considered negative.
Expected Values
The ability of the NOVA Lite IgG F-Actin Kit to detect IgG F-Actin antibodies was evaluated by
comparison to commercially available QUANTA Lite Actin IgG ELISA (Inova Diagnostics, Inc). ELISA
results were determined to be positive if the patient sample was 20 units or greater and negative if less
than 20 units.
Normal Range
Four hundred ninety three samples from normal blood donors were run using the NOVA Lite IgG F-
Actin kit. All but 4 of the 493 normal samples (99.2% specificity) were negative on NOVA Lite IgG F-
Actin.
Comparison between NOVA Lite IgG F-Actin IFA and QUANTA Lite Actin IgG
ELISA
To determine the positive and negative percent agreement of the assays, 992 samples containing
antibodies to a wide variety of antigens were tested with the NOVA Lite IgG F-Actin IFA kit and the
QUANTA Lite Actin IgG ELISA. These samples included 493 normal blood donors, 60 patients with
clinically defined diseases (rheumatoid arthritis, SLE, and scleroderma), 40 samples with defined
antibodies (infectious diseases and autoantibodies) and 399 samples from patients suspected of having
liver disease.
NOVA Lite, Inova and Inova Diagnostics are trademarks of Inova Diagnostics, Inc. © 2019 Inova Diagnostics, Inc.
All rights reserved
4
Symbols Used
European Conformity
Conformidad europea
Temperature Limitation
Límites de temperatura
Batch Code
Código de lote
Manufacturer
Fabricante
Use by
Caducidad
5
References
1. Chretien-Leprince P, Ballot E, Andre C, et al. Diagnostic value of anti-F-actin antibodies in a
French multicenter study. Ann NY Acad Sci 1050: 266-273, 2005.
2. Villalta D, Bizzaro N, Da Re M, et al. Diagnostic accuracy of four different immunological methods
for the detection of anti-F-actin autoantibodies in type 1 autoimmune hepatitis and other liver
related disorders. Autoimmunity 41: 105-110, 2008.
3. Czaja A, Norman G: Autoantibodies in the diagnosis and management of liver disease. J Clin
Gastroenterol 37: 315-329, 2003.
4. Toh BH. Smooth muscle autoantibodies and autoantigens. Clin exp Immunol 38: 621-628, 1979.
5. Fusconi M, Cassani F, Zauli D, et al. Anti-actin antibodies: a new test for an old problem. Journal
of Immunological Methods 130: 1-8, 1990.
6. Granito A, Muratori L, Muratori P, et al. Antibodies to filamentous actin (F-actin) in type 1
autoimmune hepatitis. J Clin Path 59: 280-284, 2006.
7. Dighiero G, Lymberi P, Monot C. Sera with high levels of anti-smooth muscle and anti-
mitochondrial antibodies frequently bind to cytoskeleton proteins. Clin exp Immunolol 82: 52-56,
1990.
8. Biosafety in Microbiological and Biomedical Laboratories. Centers for Disease Control/National
Institute of Health, 2007, Fifth Edition.
6
Manufactured By:
:
Australian Sponsor: