Cobas E411 ServiceManual v1.0
Cobas E411 ServiceManual v1.0
Cobas E411 ServiceManual v1.0
cobas e 411
Note
Part No. and Part name are different depending on the manufacturing time.
Please refer to Parts List for details.
RD/Hitachi Immunoassay System cobas e411 Service
Manual
1. Application/Introduction
1.1 Instrument
1.2 Specifications
1.3 Operating Precautions and Hazards
1.4 Service Concept
1.5 Rack Sampler/Rack Conveyor System
2. Installation / Set Up
2.1 Site Requirements
2.2 Inventory
2.3 Analyzer Installation
2.4 Software Installation
2.5 Rack Sampler Connected to CLAS 1 System Installation
2.6 Check/Adjustment During Installation
3. Fludics
3.1 Description of Flow Path
3.2 System Volume
3.3 Cleaning Procedures
3.4 SysWash
4. Mechanics
4.1 Overview
4.2 Location of Mechanisms
4.3 List of Motors, Sensors and Other Mechanisms
4.4 Detailed Explanation of Each Mechanism
4.5 Mechanical Adjustment
4.6 Rack Sampler System
5. Electronics
5.1 Boards
5.2 Power Source
5.3 Electronic Modules
5.4 Printed Circuit Boards
5.5 Cross Wiring Diagrams
5.6 How to Check Photo Interrupters
6. Service Software
6.1 Overview
1
7. Troubleshooting
7.1. Alarms List
7.2 Data Alarms
7.3 Troubleshooting List
7.4 Data File Load Errors
7.5 Quality Control Check
10. Maintenance
10.1 Operator Maintenance
10.2 Preventive Maintenance
10.3 Maintenance Material
Appendix
Timing Chart Tables
Assay Timetable
2
System Description
Contents - 1
Installation / Set Up
Contents - 1
Fluidics
Chapter 3 Fluidics
Contents - 1
Mechanics
Chapter 4 Mechanics
Contents - 1
Electronics
Chapter 5 Electronics
Contents - 1
Service Software
Contents - 1
Troubleshooting
Chapter 7 Troubleshooting
Contents - 1
Spare Parts/Recommended Parts
Contents - 1
Host Interface
Contents - 1
Maintenance
Chapter 10 Maintenance
Contents - 1
RD/Hitachi cobas e411 Service Manual
1.1 Instrument
Figure 1.1-6
Figure 1.1-7
The sample disk has 30 positions for samples, calibrators and controls. Patient samples may be
placed in either primary sample tubes or sample cups. Built-in adapters allow intermixing of different
size primary sample tubes.
Sample tubes that may be used are listed in chapter 2.7 Technical Data.
Sample cups [2 mL (Standard) Hitachi cups only] may be placed directly on the sample disk or on top
of 16 mm primary sample tubes.
CAUTION
Micro cups cannot be used on the
e411 analyzer!
CAUTION
Be sure to lay down the sample disk
protective cover before you start
operation.
Figure 1.1-17
Rack Sampler
The rack sampler consists of an A-Line, B-Line, C-Line and STAT position.
1.1.4.1 A-Line
Specimens are placed in 5-position sample racks and are loaded onto a tray. Once a tray is loaded,
additional racks can be added to the tray one at a time during Operation, provided the tray indication
light is green (ON). If the light is out (OFF), the pusher arm is preparing to move. The pusher arm is
located at the far left of the A-Line and pushes the sample racks forward and onto the B-Line.
The A-Line holds a tray that accommodates 15 racks at one time. Another five racks can be in the
input buffer. Therefore, you can have a total of 100 specimens loaded at any one time. Refer to the
photo and graphic below.
Figure 1.1-18
Figure 1.1-19
1.1.4.2 B-Line
The B-Line transports the sample racks, single file, first to the rack bar code reader. Here each
position in the rack is scanned for a sample bar code. After the last position is scanned, the bar code
reader scans the rack ID. After the last specimen is sampled, the rack is transferred via the output
buffer onto the tray on the C-Line. Refer to the photo and graphic below.
Figure 1.1-20
• NW7 (Codabar)
• Code 39
• Code 128
• Interleaved 2 of 5
1.1.4.4 C-Line
Racks are off-loaded from the B-Line into the output buffer. If there is no tray, up to 5 racks can enter
the output buffer, thereafter the sampling procedure is stopped. When the sixth rack is moved into the
output buffer, a rack is pushed onto the tray on the C-Line. You can remove the tray from the C-Line
any time the tray indication light is green (ON). If the light is out (OFF), the system is preparing to push
a rack onto the C-Line tray. You cannot remove single racks from the C-Line. You must remove an
entire tray at one time.
Figure 1.1-23
If the tray is removed, the system continues to push racks into the output buffer. If the buffer fills and
there is no tray, the analyzer issues an alarm and stops sampling racks.
Figure 1.1-26
• it can be seen when either the sample disk or reagent disk is removed.
• to read bar code labels, the bar code reader rotates between the sample and reagent disks, and
the card reading station.
Figure 1.1-31
The bar code reader is auto-discriminating, allowing the use of various types of bar codes during
operation. In addition, this bar code reader also reads PDF417.
Note: PDF417 can only be used for reagent bar codes and bar code cards.
Figure 1.1-34
Figure 1.1-35
Note: Diluents or pretreatment reagents can be placed in ANY position on the reagent disk. More
than one reagent pack can be loaded on the reagent disk for each test.
1.1.5.1 Gripper
The gripper can move in three directions:
• X (left and right)
• Y (forward and back)
• Z (up and down)
It is also equipped with gripping fingers for gripping a tip or AssayCup. The gripping fingers grip a tip
from the tip tray, or a cup from the cup tray and deliver it to the pipetting station. Then, at the
appropriate time, the gripper moves the AssayCup to the incubator, then to the aspiration station, and
finally to the cup disposal opening.
During operation, the analyzer starts utilizing tips and cups from tray 1, position 1. As soon as tray 1 is
empty, the analyzer starts using tray 2. As soon as tray 2 is empty, the analyzer continues with tray 3.
When tray 3 is empty, the analyzer returns to tray 1, if a new tray has been reloaded.
Note: Removing the system water container while the analyzer is in Operation causes the
analyzer to enter P. Stop status.
Note: Removing the liquid waste container while the analyzer is in Operation or an improperly
positioned container causes the analyzer to enter E. Stop status.
Note: Removing the solid waste tray while the analyzer is in Operation causes the analyzer to
enter E. Stop status.
1.1.6.1 Incubator
The incubator is maintained at a specific temperature (37.0 °C ± 0.3° C) for the reaction of the sample
and the reagents that have been dispensed into a cup. The incubator is equipped with 32 positions.
When an assay is ready for measurement, the AssayCup is transferred by the gripper to the aspiration
station, and the sipper probe aspirates the reaction mixture for measurement. The aspiration station,
located in the lower right corner of the incubator, is not temperature controlled.
When starting from Stand-by, the sipper probe always attempts to first use ProCell and CleanCell from
bottle set 2. If the quantity is insufficient, bottle set 1 is used. When starting from S. Stop or R. Stop,
the bottle set in use when the analyzer was previously in Operation is pipetted.
The analyzer can operate with one bottle set of ProCell and CleanCell reagent, but they must be
placed in positions 1 & 2 or 3 & 4. Refer to the photograph above.
Note: To have access to system reagent bottles, you must open the sipper safety cover. To open
this cover, push the cover’s metal part as shown in the picture below to release the hold. To
close the cover, push the same part until a click is heard.
Figure 1.1-53 Opening/closing the sipper safety cover (Push the circled point)
CAUTION
• Do not open the sipper safety cover while the analyzer is in Operation.
Otherwise, the operation will stop.
Note: To disconnect the analyzer from the supply source, the circuit breaker must be in the O
(OFF) position and the power cord must be removed.
Figure 1.1-58
Note: To disconnect the analyzer from the supply source, the circuit breaker must be in the O
(OFF) position and the power cord must be removed.
1.1.8.1 Introduction
The cobas e411 analyzer automates the immunoassay reactions utilizing electrochemiluminescence
(ECL). These reaction methods are described in detail in Chapter 4, ECL Technology. The individual
test steps and how the system performs the necessary procedures are discussed here.
1.1.8.13 Finalization
30 Minutes after documentation of the last result, the sipper pipettor flushes system water through the
sipper probe, and then fills the measuring cell with ProCell before the analyzer returns to Stand-by.
After this procedure periodically all 30 minutes the waste pump of the S/R rinse station is running for 2
seconds (waste consumption approx. 12 mL). This procedure will be stopped after you switch off the
operation switch.
A. The analyzer resets all mechanisms to their respective home positions. Next, the S/R pipettor
primes the S/R probe.
B. The gripper checks for a tip in position number 1 of the tip trays. If this position is empty, the gripper
remembers where it last left off and checks that position. If this position is empty, the gripper considers
the whole tray empty and the System Overview screen is updated accordingly.
Note: If the analyzer is in S. Stop, the gripper remembers where it last left off and checks for a tip
in that position.
1. During the tip check, the S/R probe is checked for the presence of a tip. The probe moves to the
tip eject station and performs the movements to eject a tip. If a tip is present it is ejected.
2. After the tip check is complete, the AssayCups are checked in the same manner. During the cup
check, the analyzer finishes priming the probes.
3. Next, the gripper checks the last three of the five positions on the pipetting station. If a cup is
present, the analyzer goes through the steps of a cup disposal. The gripper places a tip in position
1 of the pipetting station. Then, the S/R probe picks up the tip in position 1 of the pipetting station.
The S/R probe descends into the AssayCup and attempts to aspirate any possible liquid from the
cup. The gripper picks up the cup and discards it into the cup disposal opening. As the cup is
disposed, the S/R probe moves to the rinse station and dispenses any aspirated liquid. The tip is
then washed and discarded.
4. The gripper moves to the incubator where it checks all 32 incubator positions. If a cup is present,
the gripper moves the cup to position 5 on the pipetting station and uses the same procedure
listed in step 3 to discard the cup.
5. The S/R probe tip is ejected after all the incubator positions are checked.
Figure 1.1-61
1. After preoperation functions are complete, the gripper takes a tip from the tip tray and transports it
to position 1 of the pipetting station. The gripper returns to its Stand-by position.
2. The sample disk rotates until position 1 is in the sampling position.
3. The S/R probe moves to position 1 of the pipetting station, descends to obtain the tip, rises and
returns to its Stand-by position.
4. During this time, the reagent disk rotates until the TSH reagent pack is at the cap open/close
mechanism. The mechanism moves forward and opens the caps on the reagent pack. The disk
rotates again to move the TSH reagent to the R1 position.
5. The S/R probe moves from its Stand-by position to the R1 aspiration position. While activating
liquid level detection, the probe descends until it is 2 mm below the reagent surface and aspirates
60 µL of R1.
Note: The lowest allowable point the S/R probe can descend to is 1.3 mm above the bottom of the
reagent pack.
Figure 1.1-62
While aspirating R1, the gripper puts another tip in position 1 of the pipetting station.
6. If the S/R probe does not detect liquid during descent, no reagent aspiration can occur, an alarm
is generated.
7. After R1 aspiration, the S/R probe rises and moves to the rinse station. To prevent the aspirated
R1 from contacting the water in the rinse station, the probe aspirates 10 µL of air. The rinse
station externally washes the tip.
8. During step 7, the reagent disk rotates until the TSH reagent pack is in the R2 position.
9. The S/R probe moves from the rinse station to the R2 position while aspirating another 10 µL of
air. This air layer prevents R1 from mixing with R2. While activating liquid level detection, the
probe descends until it is 2 mm below the reagent surface and aspirates 50 µL of R2. While
aspirating R2 the gripper moves an AssayCup to position 5 of the pipetting station.
10. Upon completion of R2 aspiration, the S/R probe rises and moves to the rinse station. To prevent
the aspirated R2 from contacting the water in the rinse station, the probe aspirates another 10 µL
of air. The rinse station externally washes the tip.
11. After R2 aspiration, the reagent disk rotates until the TSH reagent pack is at the cap open/close
mechanism. The mechanism moves out and closes the caps.
12. The S/R probe moves from the rinse station to the sampling position while aspirating another 10
µL of air. While activating liquid level detection, the probe descends until it is 2 mm below the
sample surface and aspirates 50 µL of sample. During sample aspiration, clot detection is
activated.
Figure 1.1-63
Figure 1.1-64
Note: Depending on the sample volumes and the type of vials used (e.g. primary sample tube or
sample cup), the sample/reagent (S/R) probe can, when necessary, be lowered further to
prevent air being aspirated. With some low capacity vials, the inside diameter is relatively
small, which means the level of the liquid sinks when the liquid is being aspirated.
13. The S/R probe moves from the sampling position to position 5 of the pipetting station. The probe
descends until the tip reaches 2 mm below where the calculated level of the reaction mixture
surface should be and dispenses the sample, R2 and R1. The probe's downward displacement is
determined by calculating the reaction mixture volume for the sample and utilizing downward
displacement tables in the software. The probe does not rise during dispense.
14. After dispense, the S/R probe moves to the tip eject position and ejects the tip.
Figure 1.1-65
1. After preoperation functions are complete, the gripper takes a tip from the tip tray and transports it
to position 1 of the pipetting station. The gripper returns to its Stand-by position.
2. The pusher arm pushes the racks in the A-Line forward to the B-Line. The arm returns to its home
position. The first rack loads on the B-Line.
3. As the rack incrementally moves on the B-Line, the rack bar code reader scans all five rack
positions and rack ID. When scanning is complete, position 1 of the rack is in the sampling
position.
4. The S/R probe moves to position 1 of the pipetting station, descends to obtain the tip, rises and
returns to its Stand-by position.
5. During this time, the reagent disk rotates until the TSH reagent pack is at the cap open/close
mechanism. The mechanism moves forward and opens the caps on the reagent pack. The disk
rotates again to move the TSH reagent to the R1 position.
6. The S/R probe moves from its Stand-by position to the R1 aspiration position. While activating
liquid level detection, the probe descends until it is 2 mm below the reagent surface and aspirates
60 µL of R1.
Figure 1.1-66
Note: The lowest allowable point the S/R probe can descend to is 1.3 mm above the bottom of the
reagent pack.
While aspirating R1, the gripper puts another tip in position 1 of the pipetting station.
7. If the S/R probe does not detect liquid during descent, no reagent aspiration can occur, an alarm
is generated.
8. After R1 aspiration, the S/R probe rises and moves to the rinse station. To prevent the aspirated
R1 from contacting the water in the rinse station, the probe aspirates 10 µL of air. The rinse
station externally washes the tip.
9. During step 8, the reagent disk rotates until the TSH reagent pack is in the R2 position.
10. The S/R probe moves from the rinse station to the R2 position while aspirating another 10 µL of
air. This air layer prevents R1 from mixing with R2. While activating liquid level detection, the
probe descends until it is 2 mm below the reagent surface and aspirates 50 µL of R2. While
aspirating R2 the gripper moves an AssayCup to position 5 of the pipetting station.
Figure 1.1-67
11. Upon completion of R2 aspiration, the S/R probe rises and moves to the rinse station. To prevent
the aspirated R2 from contacting the water in the rinse station, the probe aspirates another 10 µL
of air. The rinse station externally washes the tip.
12. After R2 aspiration, the reagent disk rotates until the TSH reagent pack is at the cap open/close
mechanism. The mechanism moves out and closes the caps.
13. The S/R probe moves from the rinse station to the sampling position while aspirating another 10
µL of air. While activating liquid level detection, the probe descends until it is 2 mm below the
sample surface and aspirates 50 µL of sample. During sample aspiration, clot detection is
activated.
Figure 1.1-68
Note: Depending on the sample volumes and the type of vials used (e.g. primary sample tube or
sample cup), the sample/reagent (S/R) probe can, when necessary, be lowered further to
prevent air being aspirated. With some low capacity vials, the inside diameter is relatively
small, which means the level of the liquid sinks when the liquid is being aspirated.
14. The S/R probe moves from the sampling position to position 5 of the pipetting station. The probe
descends until the tip reaches 2 mm below where the calculated level of the reaction mixture
surface should be and dispenses the sample, R2 and R1. The probe's downward displacement is
determined by calculating the reaction mixture volume for the sample and utilizing downward
displacement tables in the software. The probe does not rise during dispense.
15. After dispense, the S/R probe moves to the tip eject position and ejects the tip.
1. The reagent disk rotates until the TSH reagent pack is at the reagent cap open/close mechanism.
The mechanism moves out and opens the cap. The disk moves the reagent pack to the mixing
position.
2. The mixer moves over the reagent disk and descends into the microparticles to a level 1.4 mm
above the bottom of the bottle.
Figure 1.1-69
Note: The mixer descends to this level regardless of the volume of microparticles in the bottle.
3. The mixer stirs the microparticles for approximately 4 seconds to obtain a homogeneous
suspension. During the mixing, the gripper obtains a fresh AssayTip and transports it to position 2
of the pipetting station.
4. When mixing is complete, the mixer rises and returns to the rinse station where it descends and
rotates in the rinse station for washing.
5. At the same time, the reagent disk rotates the TSH reagent pack to the microparticle pipetting
position.
6. The S/R probe moves to the tip eject position and discards the tip.
Figure 1.1-70
1. The sipper probe moves from its home position to a ProCell bottle and descends to 2 mm below
the solution level and aspirates ProCell into the measuring cell. During descent, liquid level
detection is activated.
Note: The sipper probe can descend as low as 1.3 mm above the bottom of the ProCell bottle.
2. The sipper probe rises.
• While in Operation, the solid waste tray periodically shakes for 1.5 seconds.
• While in Stand-by, the reagent disk turns 90° every 30 minutes.
• While in Stand-by, the rinse stations for the S/R probe and sipper probe are switched on for 3
seconds every 30 minutes.
• Microparticles undergo a long mix when starting from Stand-by and then every 90 minutes, when
pipetting not yet started.
• Microparticles undergo a short mix (approx. 4 seconds) and then a short mix every 60 minutes for
each reagent pack.
1.1.10.1 Assay With One Step Dilution (1:2, 1:5, 1:10, 1:20)
Tip 1 -> diluent (wash)* + sample -> cup 1
Tip 2 -> R1 (wash)* + R2 (wash)* + diluted sample from cup 1 -> cup 2 ... 1st incubation
Detection
* (wash) = the outside of the AssayTip is washed.
R1 = Reagent 1
R2 = Reagent 2
M = Microparticles
1.1.10.2 Assay With Two Step Dilution (1:20, 1:50, 1:100, 1:400)
Tip 1 -> diluent (wash)* + sample -> cup 1
Tip 2 -> R1 (wash)* + diluted sample from cup 1 -> cup 2
Tip 3 -> R1 (wash)* + R2 (wash)* + diluted sample from cup 2 -> cup 3 … 1st incubation
Tip 4 -> M (wash)* -> cup 3 … 2nd incubation
Detection
* (wash) = the outside of the AssayTip is washed.
R1 = Reagent 1
R2 = Reagent 2
M = Microparticles
1.1.11.1 Introduction
The e411 analyzer can occupy a number of status conditions. A table of the status conditions you
normally see during routine operation or maintenance procedures is listed below. There are several
other conditions that exist; however, most of these status conditions are seen during various
adjustment or maintenance procedures performed by a Roche Diagnostics representative. These
additional status conditions are not included in the table below.
Figure 1.1-71
Figure 1.1-72
Figure 1.1-73
1.1.11.7 Finalization
The status of the analyzer when it is between the status conditions S. Stop and Stand-by.
1.1.11.9 Initialization
This status is seen when the e411 is powered ON or when START is pressed from Stand-by.
Figure 1.1-74
Figure 1.1-75
1.1.11.14 Operation
This is the status during which the e411 performs its routine operations.
Figure 1.1-76
Figure 1.1-77
Figure 1.1-78
Figure 1.1-79
Figure 1.1-80
1.1.11.26 Stand-by
The analyzer is not performing any operations.
1.1.11.27 Stop
This status occurs when the Stop global button is pressed or when a Stop alarm condition exists. If an
alarm exists, take the appropriate measures to resolve the problem. For further details on Stop, refer
to Chapter 7, Instrument Alarms - List.
1.2.1.2 Electrical
Installation requirements Pollution degree: 2 (IEC 61010-1)
Overvoltage category: II (IEC 664)
The cobas e411 analyzer must be connected to a three-wire
power supply cord with a safety ground.
AC Cord
For countries which require UL and A UL-listed and CSA-certified AC cord set must be used.
CSA compliance (A cord with a plug and a connector)
Recommended rate
Supply voltage : 115 V AC125V-15A / SJT type 3×14AWG
Supply voltage : 208 or 240 V AC250V-10A / SJT type 3×16AWG
Temperature rate Min 60°C
Whole length 2500±50mm
Safety Standard UL817 / CSA C22.2 No.21
For European countries An HAR-certified AC cord must be used.
(A cord with a plug and a connector)
Recommended rate
Supply voltage : 220 to 240 V AC250V-10A / 3×1.0 mm2
Temperature rate Min 60°C
Whole length 2500±50mm
Safety Standard CENELEC HD21
Connector Type connected to IEC320 / EN60320 type
e411 inlet
Figure 1.2-1
30 positions for samples, controls and calibrators
Figure 1.2-2
tray - 15 racks with 5 positions each for samples, controls and
calibrators = 75
tray with input buffer - 20 racks with 5
positions each = 100
STAT capacity
Figure 1.2-3
any unoccupied position on the sample disk
Figure 1.2-4
STAT position at the front of the analyzer
Bar code symbologies PDF417
NW7 (Codabar)
Code 39
Code 128
Interleaved 2 of 5
AssayTips 360 tips (3 trays; 120 tips/tray)
AssayCups 180 cups (3 trays; 60 cups/tray)
Sample cups 2 mL (Standard) Hitachi cup; NO micro cups
Primary sample tubes 13 x 75 mm 16 x 75 mm
(external diameter x height): 13 x 100 mm 16 x 100 mm
The label size when attaching it with reference to the uppermost edge of the test tube mouth is shown
in Figure 1.2-6.
A quiet zone of more than 5mm should be provided at both ends of the symbol as shown in Figure
1.2-5.
However, for RD calibrator/control vials, a quiet zone of more than 3mm should be provided at both
ends of the symbol as shown in Figure 1.2-6.
The bar code label must be attached to each test tube in bar code zone as shown in Figure 1.2-7.
Figure 1.2-7 Attachment of Bar code label to Test Tube and vial
1 Test tube
2 Bar code zone
3 Vial
4 Bar code label
Sample ID, when using the secondary tube for sample vessel, is by putting sample cup on the test
tube with bar code label (Cup on Tube) as shown in Figure 1.2-8, Figure 1.2-9.
Setting Procedure 1 Setting Procedure 2
Method Sample Cup + Test Tube Test Tube alone
Assembly
2 Test Tube
Figure 1.2-10
CAUTION
A reduced dead volume may only
be used with Hitachi standard cups
(not with primary or secondary
cups).
Figure 1.2-11
Reagent System
Reagent disk temperature 20 °C ± 3 °C (68 °F ± 5.4 °F)
Reagent capacity in any combinations in 18 reagent positions with the
exception of the following restrictions: At the same time not
more than 18 tests, 9 pretreatments or 8 diluents.
R1/R2 consumption 50 to 80 µL per reagent dependent upon the assay
Microparticle consumption 30 to 50 µL dependent upon the assay
Reagent detection liquid level detection
Positive reagent identification 2-dimensional bar code (PDF417)
Automatic dilution available up to 1:400
Evaporation protection reagents are automatically opened and closed
Inventory control automatic based on counting (reagent disk) or liquid level
detection (ProCell/CleanCell)
Incubation System
Incubator capacity 32 AssayCups
Volume of AssayCups 200 µL
Incubation temperature 37.0 °C ± 0.3 °C (98.6 °F ± 0.5 °F)
Measuring System
Measuring method integral measuring of an electrochemiluminescence signal
Calibration mode 2-point calibration
Test protocols 28 test methods
ProCell consumption approx. 2 mL per cycle
CleanCell consumption approx. 2 mL per cycle
Cycle time 42 sec.
Control Unit
Compact flash card
Host interface CCITT V. 24/RS-232-C (bidirectional)
The host computer should comply with the requirements of
IEC (60950).
External printer Connected via the USB port
Optional module Laboratory System Manager (LSM)
Touchscreen monitor 38 cm (15 in) color TFT-LCD, XGA
The safety precautions and important user notes are classified according to ANSI Z535 standards.
Familiarize yourself with the following meanings and icons:
WARNING
Indicates a possibly hazardous
situation which, if not avoided,
may result in death or serious
injury.
Indicates documentation must
be consulted in all cases where
this symbol is marked.
WARNING
Indicates that samples containing
material of human origin must be
treated as potentially infectious. The
relevant laboratory guidelines on
safe use must be observed.
CAUTION
Indicates a possibly hazardous
situation which, if not avoided,
may result in slight or minor
injuries, damage to equipment,
inaccurate results or loss of
result data.
“Minor injury” refers to injuries
that may require medical
assistance.
“Equipment” refers to extended
damage to buildings, furniture,
and so on.
• DO NOT open the back cover and service cover with the power switch or circuit
breaker turned on. Otherwise, you may receive an electrical shock.
• DO NOT open the Detection unit cover with the power switch or circuit breaker
turned on. Otherwise, you may receive an electrical shock. Corresponding label: L
• DO NOT open the cover of the PMT high voltage supply circuit board with the
power switch or circuit breaker turned on. Touching the board may cause death
or severe injury. Corresponding label: K
• Avoid touching the A-, B-, or C-Lines, sample/reagent probe mechanism, sipper
probe mechanism, gripper (tip/cup carrier) mechanism, beads mixing
mechanism, cap opener mechanism, and other moving parts while the analyzer is
operating. therwise, personal injury may result. Corresponding label: C (for the
cap opener mechanism)
• Make sure the analyzer is in S. Stop status when you load additional samples
onto the sample disk or remove processed samples from the sample disk while
the analyzer is operating. Otherwise, personal injury may result.
• Verify there is no rack movement and rack indication light is green when you load
additional sample racks on the A-Line or remove processed sample racks from
the C-Line while the analyzer is operating. Otherwise, personal injury may result.
• DO NOT wear loose garments or jewelry that could catch in moving mechanisms
and cause you to hit other mechanisms.
• Before placing additional samples onto the sample disk, or replacing the sample
disk during operation, make sure that the analyzer is in S. Stop status. And then
be sure to raise the sample disk protective cover before placing the samples or
taking out the disk.
• Avoid touching the cooling fan of the detection unit while the analyzer is
operating. Otherwise, personal injury may result. To clean the fan, be sure to turn
OFF the analyzer.
• DO NOT insert your finger into the openings on the reagent disk cover.
Otherwise, personal injury may result.
• Open/close the analyzer top cover quietly. Otherwise, personal injury may result.
• Use the personal protective equipment (such as rubber gloves, lab coat, safety
glasses, etc.) recommended by your facility when handling biohazardous
materials.
• Avoid direct contact with the sample/reagent probe, sample disk protective
cover, sipper probe, and rinse stations. Treat these components as potentially
biohazardous area. Corresponding label: A, B and E
• Avoid direct contact with waste solution and/or solid waste. Both should be
handled as potentially biohazardous materials. Corresponding label: H and I
• Direct contact with sample may result in infection. Be sure to wear protective
equipment and follow the instruction specified in this operational manual.
• If sample or reagent contacts your skin, wash it off immediately with water and
apply a disinfectant. Consult a physician.
• Avoid direct contact with CleanCell. Direct contact with the reagent may result
in skin irritation or damage. Refer to the CleanCell box label for specific
instructions. Corresponding label: G
• Avoid direct contact with reagents. Direct contact with reagents may result
irritation or damage in the skin. Refer to the reagent kit box labels for
instructions.
DO NOT add bleach or strong alkaline disinfectants (pH > 9.5) to the liquid waste
container. Disinfectants combined with the contents of the liquid waste could cause
potentially harmful fumes.
Avoid using dangerous flammables around the instrument. Fire or explosion may
be caused by ignition.
Power must be off to move mechanical components. If power is on, the motors are
engaged and attempted movement may damage these components.
• Be careful not to damage the lower end of the S/R probe during cleaning.
Make sure the analyzer is in S. Stop status when you load additional samples onto
the sample disk or remove processed samples from the sample disk while the
analyzer is operating. Otherwise, the instrument may be damaged or operation may
be stopped.
• DO NOT touch the reagent disk while the analyzer is in operation. Otherwise,
the instrument may be damaged, or operation may be stopped.
• DO NOT open the reagent disk cover, except for replacement of reagents.
Otherwise, this may compromise the cooling efficiency or cause deterioration of
the reagents. Opening of the cover during analysis also causes an instrument
alarm.
• Be sure to put the cover on the reagent disk before you start operation or after
you placed/replaced reagents. Otherwise, operation will not start.
Verify that the lights on the rack sampler is green, prior to adding a new rack or tray
to the A-Line or removing a tray of processed samples from the C-Line while the
analyzer is in Operation. Otherwise, the instrument may be damaged or operation
may be stopped.
Be careful not to bend the microparticle mixer. A bent mixer could lead to
inaccurate results.
Be careful not to damage the lower end of the sipper probe during cleaning.
• Do not open the sipper safety cover while the analyzer is in operation.
Otherwise, the operation will stop. DO NOT open the sipper safety cover unless
the analyzer is in Stand-by.Corresponding label: F
• To open/close the sipper safety cover, press the proper spot that is indicated in
this manual. See 1.1.6 Measuring Area Component.
Insoluble contaminants such as fibrin and dust in samples may cause probe
clogging and thus lead to pipetting volume shortage and deterioration in
measurement accuracy. When loading samples on the analyzer, make sure that
samples contain no insoluble contaminants such as fibrin or dust.
If a malfunction occurs in the touch screen monitor and you cannot use it, contact a
Roche Diagnostics service representative. And in such cases, please use the
auxiliary USB mouse to operate the instrument until the service representative
arrives.
• For proper use of the instrument, measure control samples and monitor the
instrument during operation.
• Do not use reagents that are expired. Otherwise, inaccurate data may be
obtained.
• Please note that other analyses may not be applicable to this instrument.
• Carefully follow the procedures specified in the operator’s manual and this
manual for maintenance of the analyzer.
• AssayCups, AssayTips, the detection unit and liquid waste container or solid
waste tray and liner are not guaranteed to be chemically resistant against
organic solvents. Therefore, do not use organic solvents on these parts.
• Avoid using samples and reagent solutions that are likely to adhere to
AssayTips, AssayCups, the liquid waste container or detection unit.
Never switch on the power within one second of switching it off.
If the instrument will not be used for a long period of time (i.e., > 7 days), Different
shutdown procedures are recommended depending upon the duration of inactivity.
Devices that emit electromagnetic waves may affect measured data, or cause
the instrument to malfunction. DO NOT operate the following devices in the
same room where the instrument is installed:
Mobile phones, cordless phones, other electrical devices that generate
electromagnetic waves.
• At least two persons must carry the instrument, by holding the base plates
on the left side and right side of the analyzer. Be careful not to hurt your
hands or fingers when putting the instrument in place.
• The cobas e411 is provided with more than one power supply cords. Make
sure that all the power supply cords are removed when disconnecting the
instrument from the main supply source.
A reduced dead volume may only be used with Hitachi standard cups (not with
primary or secondary cups).
Be sure to lay down the sample disk protective cover before you start operation.
If you load additional AssayTip/Cup trays onto the analyzer while the analyzer is
in operation, be sure to place them only in the positions indicated by the software
screen as accessible.
When loading a rack into the STAT position, check the orientation of the rack.
An improperly oriented rack may cause instrument damage.
Corresponding label: J
Figure 1.3-1
Figure 1.3-2
Figure 1.3-3
Table 1.3-1
Symbol Validity Position Symbol Validity Position
S/R Probe A Sipper Safety Cover F
STAT position J
1.3.4 Approvals
The cobas e411 analyzer meets the requirements stated in Directive 98/79/EC of the European
Parliament and the Council of the European Union (EU) on in vitro diagnostic medical devices.
Furthermore, the cobas e411 analyzer is manufactured and tested according to International Standard
IEC 61010-1, "Safety requirements for electrical equipment for measurement, control and laboratory
use, Part 1: General requirements". This International Standard is equivalent to the national standards
Underwriters Laboratories (UL) 61010-1 for the USA, CSA CSA 22.2 No. 61010-1-4 for Canada.
Compliance is demonstrated by the following marks:
Complies with the IVD directive 98/79/EC.
Issued by Underwriters Laboratories, Inc. (UL) for Canada and the USA.
Figure 1.4-1
Figure 1.4-2
Figure 1.4-3
Figure 1.4-4
2.1.1.1 Receiving
A loading dock or other suitable facility is needed to allow the analyzer to be safely unloaded.
2.1.1.2 Routing
It is necessary to move the analyzer safely from the unloading point to the proposed point of operation.
Following spaces are required:
• Minimum door width is 91 cm or more (36 in. or more).
• Loading size & capacity of elevator, if required, are 135 cm x 91 cm (54 in x 36 in) or more and
207 kg (595 lbs).
• Minimum turning radius is 163 cm (65 in).
Note: The size and weight in the packing of e411 are as following
Table 2.1-1
LxWxH Weight
Packed e411(main body only) 135 x 91 x 86 cm (54 x 36 x 34 207 kg (595 lbs)
in)
Packed Rack Sampler 110 x 82 x 60 cm (44 x 33 x 24 69 kg (152 lbs)
in)
• A firm table and floor that is level and solid enough to bear the weight of the instrument.
• Adequate space around the instrument for ventilation, maintenance and easy access to power
inlets:
For the disk system, allow at least 100 cm (39.3 in) on the front,
10 cm (3.9 in) on the rear
50 cm (19.6 in) on the right side,
and 10 cm (3.9 in) on the left side of the instrument.
For the rack system, allow at least 100 cm (39.3 in) on the front,
10 cm (3.9 in) on the rear,
50 cm (19.6 in) on the right side,
and 50 cm (19.6 in) on the left side of the instrument.
Figure 2.1-1
The installation site must have a grounding system with less than 10 ohms of resistance.
Grounding
Figure 2.1-2
2.2 Inventory
Note: The current Inventory List is packed with the instrument. Please check the parts according to
this list when unpacking the instrument.
Figure 2.2-1
Figure 2.2-2
Figure 2.2-3
General
2.3.1 Unpacking
• Make sure that the counter on which you will set the instrument satisfies the set up requirements.
Refer to Chapter 2.1.
• Unpack the cobas e411, and make sure that all accessories are in the box. Refer to the Inventory
list packed with the instrument.
• After complete unpacking of the instrument, remove all the safety devices from bead mixer,
pipetter, gripper, sipper, magnet drive and liquid waste container. See Figure 2.3-13 to Figure 2.3-
20.
• Check if there is any transport damage and please report details in case of transport damaged via
Roche Diagnostics GmbH to Hitachi High-Technologies Corporation..
• Place the plate (P/N 741-1770) on the counter top under the cobas e411 and verify that all the feet
of the cobas e411 are on the plate.
• Please refer the notice sheet to install the system that attached in the box. (See the next page)
Figure 2.3-1
Note: We offer a special table (Systable) on which to place the system, on stock.
When you move or lift up the instrument by using your hands, lift it by using the 4 points hand
positions indicated in following position.
• Place the two rubber plates (P/N 741-1760) under the front two leveling feet.
Figure 2.3-7
When you move or lift up the Rack system by using your hands, lift it by using the 4 points hand
positions indicated in following position.
Figure 2.3-12
Figure 2.3-13
Figure 2.3-14
Figure 2.3-15
1 Tape
2 Cushion
3 741-0117 S Disk Assy
Figure 2.3-16
1 Tape
2 741-0627 R Disk Assay
Figure 2.3-17
Figure 2.3-18
1 Lid
2 Tape
3 741-0302 Water Reservoir Assy
4 Vinyl resin bag
5 741-0561 Buffer Tank Assy
Figure 2.3-19
Note: The mixing paddle (P/N741-0516) is detached from the head of beads mixer and is packed
with two (2) hex screws in a box together with the accessory parts.
Figure 2.3-20
1 Tape
2 ► Attach the Panel PC to the instrument as shown in Figure 3 and 4. First, fold the BASE
PLATE and insert the Hook part to the main unit body and then, using three screws, fix the
PC in the three points indicated in the figure below.
3 ► As shown in Figure 2.3-25, route the PC wires through the cramp on the back of the Panel
PC. Finally, connect the wires to the Panel PC as shown in Figure 2.3-26.
ο Make sure that the AC power voltage is between 100 and 240 V.
ο Make sure capacity of AC breaker at customer.
ο Refer to chapter 1.2.1 Technical Data for Operation of Instrument in detail.
ο Please follow this procedure to connect power cables.
ο Refer to chapter 1.1.7 Power Components.
Circuit Breaker
Figure 2.3-28
Figure 2.3-29
Attention
When PLUG / UNPLUG Connector for HOST, turn off the Circuit Breaker of instrument.
If PLUG / UNPLUG Connector for HOST with turn on the Circuit Breaker, not only e411 main unit but
also External Host may be damaged.
2.3.4 Setup
For Rack system, first refer to Chapter 2.5.1.
Check the following items before you switch the instrument on.
• S/R probe and Sipper probe are attached straight and vertically.
• Mixer paddle is attached straight and tightly to the part of Bead Mixer.
• Please never put an irrelevant thing on a instrument.
• Sample disk, reagent disk, Assay cup trays/Tip trays and system reagents.
• Solid Waste Tray with a CleanLiner is attached correctly to the container.
• Liquid waste container is mounted correctly.
2.3.9.1 Preparation
Place the system reagent bottles in the appropriate positions.
Place two sets of system reagent ProCell/CleanCell bottles filled with
ProCell/CleanCell at Position 1 and Position 2.
Make sure the liquid waste container is empty.
Fill the system water container with system water (max. 3.0 liters).
2.3.10.1 Preparation
• Place the system reagent bottles in the appropriate positions.
• Place two sets of system reagent ProCell/CleanCell bottles filled with ProCell/CleanCell at
Position 1 and Position 2.
• Make sure the liquid waste container is empty.
• Fill the system water container with system water (max. 3.0 liters).
2.3.10.2 Procedure
• Make sure that the instrument is in standby status.
• Touch the [Utility] [Maintenance] [Check] [System Volume Check] button in the Main screen.
• Touch the START button in the screen.
The system volume will be determined (for details refer to Chapter 3.4).
Note: After the procedure has been finished successfully, the status of instrument returns to Stand-
By. When alarm message occurred, check the liquid flow path (sipper side) for possible
leakage.
Introduction
Each MC in combination with a PMT represents a complete detection unit, which has to be adjusted
correctly. The high voltage adjustment should be carried out after measuring cell preparation to obtain
initial information about the instrument signal level.
The high voltage value has been adjusted initially in the quality control department at Hitachi in the
course of checks on the instrument, but not for some MC/PMT combinations. With the high voltage
adjustment (check) the field engineer should make sure that the combination of instrument
(photomultiplier tube, electronic components) and measuring cell work in the expected signal range.
If the signal is not in the required range, the photomultiplier tube (PMT) voltage has to be adjusted to
make the signal levels comparable between the different instruments. BlankCell reagent is used to
check the signal level (refer also to Chapter 6, Assay performance explanation, for more details).
Note: The [PMT high voltage adjustment] run only checks the current signal level of instrument
Calibrator BlankCell reagent. The high voltage adjustment to the correct signal range then
has to be carried out manually by the field engineer.
Figure 2.3-30
2.3.11.1 Preparations
1 ► Place system reagent bottles in the appropriate positions.
2 ► Place the BlankCell reagent pack as shown below:
Table 2.3-1
Reagent rotor
RackPack Beads R1 R2
Pos. No.
2 BlankCell BCR1 BCR2 -
set
3 ► Sample arrangement
• Calculate the mean value of the BCR2 effective signals from the Result screen.
• Note the target value from the package insert of the BlankCell lot used.
• Compare the mean value of the BCR2 counts with the target. It should be in the range of ± 5% of
the target value.
Over time, the BCR2 level should be in the range of ± 10% (refer to Maintenance).
• If the current mean is outside the range, you may readjust the high voltage (refer to 2.3.11.5).
Note: We highly recommend that the high voltage is not changed in the first step if the signal level
of the BCR2 mean is outside the range of ± 7% of the target value.
The second row of values in the Result popup shows the dark level of the photomultiplier
tube. The dark level should be between 625 and 635 counts.
Figure 2.3-34
Figure 2.3-35
Vnew: PMT HV voltage that will be applied after measurement.
Vold: PMT HV voltage that was applied previously. (See the PMT HV record
label on the PMT SHV cover).
BCR2 Target: BCR2 value specified by RD.
• Calculate the new control voltage relative to the PMT HV voltage using the following formula.
Figure 2.3-36
• Adjust the voltage between TP1 (GND) and TP2 (control voltage signal) by changing the
resistance of the corresponding variable resistor (VR1, VR2 or VR3) according to the calculated
control voltage. The resistor is chosen by the jumper setting of JP1 (Pos.1, 2 or 3; refer to Chapter
5.4.13 for details).
• If the new high voltage value is outside the selected range you have to change the setting of
JP1then repeat the procedure, return to 2.3.11.2
• If the new high voltage value is the O.K., change the PMT HV record (the label is on the PMT SHV
cover).
Note: In case of problems make sure that LED1 located on the PMT-SHV2 PCB is lit. Check
connection J400.
Hint: A simplified calculation for determining the new High Voltage is shown in the example below.
Example:
To change the BRC2 count result from 100,000 to 105,000 counts (ca. 5000 counts difference), the
HV voltage has to be increased by 5 V (+5 V ≈ 5000 counts difference). To adjust the value, change
the control voltage by + 0.025 V ≈ 5000 counts.
Introduction
After the measuring cell has been prepared and it has been checked that the signal level is in the
expected signal range by means of the high voltage adjustment, the initial BlankCell calibration has to
be performed. The instruments are calibrated to give the same homogeneous signal level.
The initial BlankCell measurement is the first calibration done on the instrument before other assays
can be calibrated. Each measured signal will first be BlankCell-corrected (parameters a (intercept) and
b (slope)). The corresponding concentration is then read from the specific assay calibration.
• Initial Calibration
Figure 2.3-38
Figure 2.3-39
Each calibrator, control and sample signal measured is corrected by the BlankCell correction
parameters as shown below.
Figure 2.3-40
2.3.12.2 Preparation
1 ► Place BlankCell reagent pack in reagent rotor .
Table 2.3-4
Reagent
rotor RackPac
Beads R1 R2
k
Pos. no.
in any BlankCell BCR1 BCR2 -
position set
2 ► Sample arrangement
Note: if you want to measure “Initial Blank Cell Calibration”, you must set the ”ON” button.
Otherwise e411 will measure “Blank Cell Calibration”
8 ► It is very important that the Initial Blank Cell is successful. If not, repeat the process until it is
successful, otherwise all results may not be corrected properly. Refer to Chapter 2.3.12.3
for current result targets. (If necessary start procedure once more. Go back to point 3.)
9 ► If the Initial Blank Cell was successful, remove the BLANK CELL RackPack. This is very
important, otherwise the instrument will do the periodic blank cell automatically.
10 ► Start Reagent scan.
Introduction
The Artificial Media Test is a button in the menu [Assay Performance Check] in the [Utility] screen. It is
a short instrument check and principally measures the same criteria as the RD internal instrument
check.
Homogeneous Determination
With homogeneous determination magnetic separation is not used. A solution of free label is
transported inside the measuring cell and the performance of the detection unit (MC/PMT) can be
analyzed. With equal height of BCR2 1:20, the performance of the pipetter side can be analyzed.
Heterogeneous Determination
The heterogeneous determination simulates an assay. The iSAP test uses beads and a biotinylated
and ruthenium-labeled antibody. An immuno-independent bead/antibody complex is generated to
analyze the detection unit, measuring cell and magnet function.
If all performance data for the homogenous determination are in order, then disturbance in the
heterogeneous determination can only be caused by the liquid flow system (tubing/fittings, connectors,
sipper),the magnet or the MC
BCR1, BCR2 and BCR2 1:20 are homogeneous, ProCell, iSAP and TSH are heterogeneous.
Preparations
Table 2.3-7
Pos. No. RackPack Beads R1 R2
1 BlankCell set BCR1 BCR2 -
2 BlankCell set BCR1 BCR2 -
3 CellCheck ProCell ProCell ProCell
Sample arrangement
Table 2.3-8
4 to 10 ProCell (1 ml)
To perform the AM test with the R/S you need at least two racks.
Table 2.3-9
3 BCR2 (1 ml)
4, 5 ProCell (1 ml)
2 1 to 5 ProCell (1 ml)
Check Procedure
After touching the [Artificial Media Test] button in the [Assay Performance Check] screen, start the
measurement by pressing the start button.
On the rack sampler simply place rack 1 and rack 2 in the A line and start the [Artificial media test]
button in the [Assay Performance Check] screen.
Table 2.3-10
R. Disk
Reagent Assay Assay parameters
Pos. No.
1 BlankCell BCR1 Assay protocol 24
Incubation time [cycles] i1 1
i2 1
Pipetting volume [µL] R1 200
R2 0
B 0
S 0
Detection Cycle Check
2 BlankCell BCR2 Assay protocol 26
Incubation time [cycles] i1 1
i2 1
Pipetting volume [µL] R1 0
R2 200
B 0
S 0
Detection Cycle Check
3 CellCheck BCR2-D Assay protocol 0
Incubation time [cycles] i1 1
i2 1
Pipetting volume [µL] R1 95
R2 95
B 0
S 10
Detection Cycle Check
4 CellCheck AB3 Assay protocol 2
Incubation time [cycles] i1 1
i2 13
Pipetting volume [µL] R1 100
R2 100
B 0
S 0
Detection Cycle Normal
5 iSAP iSAP Assay protocol 2
Incubation time [cycles] i1 1
i2 13
Pipetting volume [µL] R1 105
R2 60
B 35
S 0
Detection Cycle Normal
Table 2.3-11
For For
Rack Disk
Pos. Pos. Sampl Item 1 Item 2 Item 3 Item 4 Item 5 Item 6 Item 7 Item 8 Item 9 Item
No. No. e 10
1 1 - BCR1 BCR1 BCR1 BCR1 BCR1
2 2 - BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2
3 3 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2 BCR2
-D -D -D -D -D -D -D -D -D -D
Rack Rack 1
For For
Rack Disk
2 7 ProCell iSAP
3 8 ProCell AB3
(CO)
4 9 ProCell iSAP
5 10 ProCell AB3
(CO)
Note: If a result is missing, the mean and CV are calculated wrongly. In such cases calculate the
corresponding result manually with the formulas given in Calculation Formulas
• The BCR (BlankCell reagent) target value should be reached; this is used for the instrument
calibration. The determination of BCR1 means the lower, BCR2 the higher calibration value. This
reflects the recovery and the variance of the target value.
Note: None of the measured counts with tests from the Assay Performance Checks are connected
with the instrument calibration factor.
• BCR2 is used to adjust the PMT-HV and indicates problems with the PMT and excitation.
• BCR2-D is a dilution of BCR2 with BCR1 and shows the functionality of the pipetter, the pipetter
precision and general problems with the liquid handling.
• AB3(CO) is determined using the measuring cycle for heterogeneous assays. These values are
the basic values used for calculating carryover.
• iSAP is a model assay based on streptavidin beads. iSAP facilitates the evaluation of the
complete LFS (liquid flow, magnet, capturing, detection) in a routine assay.
• Carryover is the measurable influence on the assay performance and is the consequence of
carryover of beads into the measuring cell in this case.
Calculation Formulas
cobas e411 instrument in which all items listed in list below meet the criteria passes Assay
Performance Check.
Figure 2.3-44
• BCR2 target value, concentration values for TSH Cal1and Cal2, and LO factor (LOF) are provided
for each lot by RD.
• LOF is defined as ratio: (tangent new rackpack lot) / (tangent frozen reference lot) That factor is
corrected by the typical coefficient of reagent aging on a monthly base. That typical coefficient of
reagent aging is calculated as the median monthly decay over several lots. LOF and its aging
forecast are transferred from RD.
Item Calculation
BCR1/ Average
BCR2N
Figure 2.3-45
Figure 2.3-46
Restore Parameter
Carry out "RESTORE PARAMETER" after you have performed or even selected any assay
performance check (APC).
Select the [Restore Parameter] button in the [Assay Performance Check] screen if you have finished
the Assay performance section. You do not need to restore if you want to perform another APC run.
The TSH assay is a necessary service tool as it reveals problems in the assay performance that occur
in determinations where high dynamic, very low LDL (lower detection limit) and very high precision are
required. For instrument/instrument performance criteria we have to use more detailed checks. There,
the calibration also plays a role.
Preparations:
Set reagent racks for TSH assay test in reagent disk positions 6 and 7.
Table 2.3-16
Pos. No. RackPack Beads Reagent 1 Reagent 2
6 CellCheck ProCell ProCell ProCell
7 TSH Beads R01 R02
Sample arrangement
For Disk-System:
Table 2.3-17
16 TSH Cal1
17 TSH Cal2
Table 2.3-18
2 TSH Cal1
3 TSH Cal2
Remove all tips and cups from the incubator and the buffer plate.
Place a full magazine with tips and cups in the front position on the tray.
Check Procedure
After touching the [TSH assay test] button in the [Assay Performance Check] screen, start the
measurement by pressing the start button
On the rack sampler simply place rack 1 in the A line and start the TSH test by pressing the [TSH
assay test] button in the [Assay Performance Check] screen.
Table 2.3-19
R. Disk
Reagent Assay Assay parameters
Pos. No.
6 CellCheck AB2 Assay protocol 2
Incubation time [cycles] i1 12
i2 12
Pipetting volume [µL] R1 60
R2 50
B 40
S 50
Detection Cycle Normal
7 TSH TSH Assay protocol 2
Incubation time [cycles] i1 12
i2 12
Pipetting volume [µL] R1 60
R2 50
B 40
S 50
Detection Cycle Normal
Table 2.3-20
Pos. Item
Sample Item 1 Item 2 Item 3 Item 4 Item 5 Item 6 Item 7 Item 8 Item 9
No. 10
1/15 ProCell AB2 AB2 AB2
2/16 TSH TSH TSH TSH TSH TSH TSH TSH TSH TSH TSH
Cal1
3/17 TSH TSH TSH TSH TSH TSH
Cal2
Note: If a result is missing, the mean and CV are calculated wrongly. In such cases calculate the
corresponding result manually using the formulas given in Test items performed by the TSH
assay test
Figure 2.3-48
Cal1 first
value high (FVH)
Figure 2.3-49
Low Detection
Limit (LDL)
Figure 2.3-50
Select the [Restore Parameter] button in the [Assay Performance check] screen if you have finished
the Assay performance section. You do not need to restore if you want to perform another APC run.
Flow Diagram
Figure 2.3-51
Legend
CV = Coefficient of Variation HW = Hardware
MC = Measuring Cell BCR = BlankCell Reagent
RE = Reference Electrode PMT = Photomultiplier Tube
HV = High Voltage LF(C) = Liquid Flow (Cleaning)
LFS = Liquid Flow System LH = Liquid Handling
Figure 2.3-52
Legend
SAP = Special service reagent TSH = Thyroid-specific Hormone
CV = Coefficient of Variation HW = Hardware
MC = Measuring Cell RE = Reference Electrode
PMT = Photomultiplier Tube HV = High Voltage
LF(C) = Liquid Flow (Cleaning) LFS = Liquid Flow System
LH = Liquid Handling BCR = Blank Cell Reagent
Figure 2.3-53
Legend
SAP = Special service reagent TSH = Thyroid-specific Hormone
CV = Coefficient of Variation HW = Hardware
MC = Measuring Cell BCR = Blank Cell Reagent
LFC = Liquid Flow Cleaning LFS = Liquid Flow System
LH = Liquid Handling
Table 2.3-22
Check items Phenomenon Possible cause Remedy
PMT HV Signal≅0, 1. HV connector is not 1.Confirm the following connections:
adjustment DL=625 connected
• between P400 and J400 on PMT-SHV board.
2.PMT connector is not
connected.
• between P401 and J401 on PMT-SHV board.
Signal≅0, 1.Measuring cell is not 1.Confirm no ADC alarm 31-02-01 or 31-02-02 takes
Precondition: Individual results for AM, TSH and PCU1/PCU2 must be determined on the instruments.
Table 2.3-24
Refer to
Step Procedure Done
chapter
14 Instrument checks 2.3.13
-TSH assay test
• SAP signals and or Calset 1 and 2-signal difference should be max. 10%
between systems.
ο Change measuring cell and restart with point 8 if the signal difference is
not within the limits.
ο Change system and restart with point 1 if the signal difference is not
within limits after exchanging the measuring cell.
15 Carry out Assay Calibration for TSH and measure Precicontrol Universal as sample.
Figure 2.4-1
Figure 2.4-2
1 ► Logoff of e411 application. Push Logoff button and display Logoff dialog screen, then push
Shutdown button.
Figure 2.4-3
Figure 2.4-4
Figure 2.4-5
This installation program doesn't have the difference between a new installation and the update.
Figure 2.4-6
When installing, it displays the dialog whether to initialize data. If select “YES”, it initializes data. If
select “NO”, the file in the data directory is not changed. When you select Yes or No and push “NEXT”,
the go to next screen. If you push “Cancel”, the installer is ended without processing anything.
Note) If you insert CD, and the above screen doesn't display, double-click E:\setup.exe from “My
Computer”(see “5)”).
Figure 2.4-7
You select whether to copy it to the directory for which the system software is specified with the
external definition file. When you select Yes or No and push “NEXT”, the go to next screen. If you
push “Cancel”, the installer is ended without processing anything.
Figure 2.4-8
You finally confirm the installation. When you copy the selected file to the preceding clause, push the
“OK” button. If you push “Cancel”, the installer is ended without processing anything.
6.5 ► Push “OK” and the installation ends, and then the following screens are displayed.
Figure 2.4-9
If the installation fails, the installation failure message is displayed and the installer is ended.
Figure 2.4-10
If you failed in the installation of the application, it is necessary to do over again from procedure 6.2 ►.
Figure 2.4-11
Figure 2.4-12
Figure 2.4-13
Figure 2.4-14
3.3 ► Select Setting tab and set password “1234” and uncheck “Protect ON”
Figure 2.4-15
3.4 ► Reboot
Figure 2.4-16
Figure 2.4-17
Figure 2.4-18
4.2 ► If there are already-installed printers, select the file and then right-click on it, and select
“Delete” from the pop-up menu (or select File -> Delete).
Figure 2.4-19
4.3 ► When the following dialogs appear, touch the Yes button, and then the OK button.
Figure 2.4-20
Figure 2.4-21
Figure 2.4-22
Figure 2.4-23
Note: This is a procedure for the HP1320 Printer Driver. If you use other Printer Driver defined by
the e411, the procedure will be shown in other Printer Driver.
6.2 ► When the following dialog appears, check the “Install from a list or specific location
(Advanced)” radio button. Then touch the Next button.
Figure 2.4-24
6.3 ► When the following dialog appears, check the “Don’t search. I will choose the driver to
install” radio button. Then touch the Next button.
Figure 2.4-25
6.5 ► When the following dialog appears, touch the Browse… button.
Figure 2.4-27
6.6 ► Select “C:/Program Files/Hewlett-Packard/LaserJet 1320 Drivers”, and touch the Open
button.
Select
Figure 2.4-28
Figure 2.4-29
Figure 2.4-30
Figure 2.4-31
6.10 ► When the following dialog appears, touch the Browse… button.
Touch
Figure 2.4-32
Figure 2.4-33
Figure 2.4-34
6.13 ► Select “C:/Program Files/Hewlett-Packard/LaserJet 1320 Drivers”, and touch the Open
button.
Figure 2.4-35
Figure 2.4-36
Figure 2.4-37
Figure 2.4-38
Figure 2.4-39
Figure 2.4-40
6.20 ► Select “C:/Program Files/Hewlett-Packard/LaserJet 1320 Drivers” and touch the Open
button.
Select
Figure 2.4-41
Figure 2.4-42
6.22 ► Select Start -> Printers and Faxes, and check if the installation went successful.
Figure 2.4-43
Figure 2.4-44
Figure 2.4-45
Figure 2.4-46
Figure 2.4-47
• Attach the EIO 3 PC board to which the RS CONT D PC board is attached to the PCB rack.
• Attach the cover for circuit board (P/N 741-1583).
• Attach the coupling plate to the aluminum base of the cobas e411 main unit. (3 screws M5, pan
head screws).
• Put the communication cable between the main unit and the coupling plate.
• Connect the rack sampler unit to the cobas e411 main unit. : Figure 2.5-6
ο ! When connecting, make sure that the couple blocks meet each other so that there is no gap
between the “A” plane and “B” plane, and fix them with screws.
Connect the communication cable from the cobas e411 to the J908 connector on the rack sampler PC
board. Arrange wiring as shown in the figure, fasten the cable with clamps at five positions. : Figure
2.5-7
• Remove the bush attached in procedure 2-5, slide the rack sampler unit onto the cobas e411 main
unit.
• Attach the 3 bushes removed in procedure 2-4, fix the unit.
9. Height Adjustment
• Loosen the lock nut of the adjuster (×2). Then, while checking levelness using a level gauge, turn
the adjuster to adjust the height as shown in Figure 2.5-8.
• When good levelness is obtained, tighten the lock nut to fix the adjuster.
• Attach the front cover and rear cover to the cobas e411 main unit.
• Attach the cover for rack sampler unit by following the procedure for removal in reverse (refer to
Procedure 2).
• Attach the Stat cover. (2 screws M4, top surface)
• Attach the screw lids to the rack sampler unit and cobas e411 main unit.
• Attach label. (Only if the cobas e411 main unit has a molded base.)
• Figure 2.5-9 shows dimensions.
Note: Be careful not to detach the photocoupler PI213 located at the Stat position from the plate of
the B line.
After turning on the power of the instrument, wait until it goes into Stand-by and perform the sampling
position adjustment by following the following procedures.
1 ► Fix the test tube guide with tape so that it stays open.
2 ► Press the B line Sampling Pos key on the adjustment rack.(Put 13mm diameter test tubes
onto the rack, and attach tip to P nozzle.) Check roughly how far the tip of the Tip is from the
center position of the container. : Figure 2.5-10
3 ► Loosen the sensor fixing screws (2 screws M4, 1 screw M3), shift the sensor fixing plate by
about the same distance as checked in 2 ►, and fix the plate. : Figure 2.5-10
4 ► Perform 2 ► again.
5 ► If the tip of the tip shifts by a large amount from the center position of the container, perform
3. again.
6 ► Press the [Service] [Adjustment S/R Probe, R.Disk, Mix] [S/R Probe-X Rack Sampling] key
to perform fine adjustment in the moving direction of pipetter. (The same as sampling
position adjustment of the main unit).
7 ► When the tip of the tip has been adjusted to be roughly in the center of the container by
procedures 1 - 6, perform 2 ►. again to adjust finely the flow direction of the rack. (If fine
adjustment is not required, you do not need to perform 7 ► and 8 ►..)
8 ► Press the Front key or Back key on Panel PC screen to perform fine adjustment of rack stop
position. (Moving forward and backward for 5 pulses can be performed, resolution: 0.15
mm/pulse)
8.2 ► After completion, press the OK key to finish adjustment.
9 ► Remove the fixing of the test tube guide performed in 1 ►..
10 ► Perform 2 ►again.
11 ► Loosen the test tube guide fixing screws (4 screws, M4) and adjust the fixing position so that
all 4 resin rollers of the test tube guide can touch the test tube. : Figure 2.5-11
12 ► Attach the table cover L2. (3 screws M4, top surface).
1 Bush
2 Rear Cover for Sampler
3 Loader Cover
1 Pull
2 Bush
3 M6 Screw
4 Coupling Plate
1 BCR Cable
2 Earth
3 BCR
4 BCR Circuit Board Assy
5 P510, P511 Connector
6 M5 Screw
7 Coupling Plate
8 Cable
9 J900
10 Clamp(HP-10N)
11 DIST-F Circuit Board (P/N:741-5028)
12 Coupling Plate
13 UIRS-C Circuit Board Assy (P/N:741-5081)
14 J343 Connector
15 J342
1 Screw
2 Couple Block 2
3 Couple Block 1
1 Adjuster
2 Water Level
3 Lock Nut
Table 2.6-1
In air In liquid
2050 rpm ± 100 rpm 2000 rpm ± 100 rpm
3.3 ► Carry out the " Mixer-Horiz.Rinse " function in the "
Adjustment S/R Probe R.Disk Mixer " screen.
4 ► Make sure that the mixing arm in home pos. does not
interfere with the pipetter.
Note: You may manually push down the mixing arm to allow easier confirmation.
Note: You may put a thin sheet of paper on the RackPack opening to allow easier confirmation
(paddle top should touch the paper).
1 ► To examine the mixing effect, a new or used RackPack (not mixed for at least 20 hours)
should be placed in reagent rotor pos. 1.
2 ► Start the Running Test from [Assay performance check] screen because the RackPack will
be mixed initially (15 s). The initial mixing should create a homogeneous solution of
microparticles.
3 ► Stop the running test after mixing and check the bead solution: No foam should be visible on
the reagent surface (only valid for 3 to 5 ml filling volume). No bead sediment should be
visible at the bottom of the bottle (check from below, see Figure 2.6-1 and Figure 2.6-2)
Note: Perform restore parameter after running test has been stopped
Figure 2.6-1
No bead sediments visible at the bottom of the bottle (seen from below): PADDLE OK
Figure 2.6-2
Bead sediments visible at the bottom of the bottle (seen from below): PADDLE NOT OK
Table 2.6-5
Unit How to Adjust
1 Pipettor LLD Refer to chapter 5.3 for adjustment procedure of Pipettor LLD, Sipper LLD
2 Sipper LLD and Clot.
3 Pressure sensor
Table 2.6-6
Unit Adjusted Item Item to be Confirmed How to Adjust
The rack guide and the rack feed arm on the Mechanical adjustment
2. Mounting position of
rack tray must be at right angles.
the rack feed arm.
The center of the rack guide on the rack tray Mechanical adjustment
3. Position of the rack
must correspond with the center of the rack
guide at the buffer
guide on the buffer base.
base.
The space between the rack feed arm and the Carry out the “A Line Loading
4. Position of the rack
rack must be 1 to 2 mm when the rack is Pos.” function in the “Adjustment
feed end.
pressed against the rack rail on the B Line. Rack” Screen.
2 B Line Not applicable ---
1. Belt tension.
The tip must be in the center of the tubes (within Carry out the “B Line Sampling
2. Sampling position.
±1 mm). Pos.” function in the “Adjustment
Rack” Screen.
The center of the notch to prevent the rack from Carry out the “B Line Unloading”
3. Position of the rack
toppling must correspond with the center of the function in the “Adjustment Rack”
feed end.
rack guide on the buffer base of the C Line Screen.
(within ±1mm).
Automatically determines the reading position. Carry out the “B Line B.C.R. Auto”
4. BCR Reading Position.
function in the “Adjustment Rack”
Screen.
3 C Line Not applicable. ---
1. Belt tension
The rack guide and the rack feed arm on the Mechanical adjustment
2. Mounting position of
the rack feed arm. rack tray must be at right angles (within ±1 mm).
The center of the rack guide on the rack tray Mechanical adjustment
3. Position of rack guide
must correspond with the center of the rack
at the buffer base.
guide on the buffer base (within ±5 mm).
If the pipetter needs correction to left/right (x-axis, see Figure 2.6-3), quit the current screen with
"Cancel" key, change to the "Adjustment S/R Probe R.Disk Mixer" screen and carry out "S/R Probe-X
Rack Sampling" from the screen.
Figure 2.6-3
If the pipetter needs correction in up/down direction (y-axis, see Figure 2.6-3), change the rack
position with "Forward" or "Backward" key.
Note: The hook will only move the rack backwards (see Figure 2.6-3). For forward hook
movement, you have to push the rack by hand to get the actual position.The range of
software adjustment for B line sample pos. is only ± 5 steps because of the design of the
racks (Figure 2.6-4).
Figure 2.6-4
If the range of software adjustment for B line sample pos. is not sufficient, change the sample position
manually. To adjust manually, move the whole sample detector B assembly (see Figure 2.6-5) in the
corresponding direction. The detector B assembly is fastened with two screws on the left and one
screw on the right. After changing the position manually, optimize again with the software (as
described above).
Figure 2.6-5
Detail:
Adjust the position of BASE (P/N741-7641) so that the distance between Surface A of the notch
located on BASE and Surface B of the readout window becomes 38.5 ± 1mm.
1 ► Detach the REAR COVER by removing the two SCREWS (M641007: M4×6 TP screw) that
fix it.
Figure 2.7-1
Figure 2.7-2
2 ► Detach the WIRE COVER (1) by unfastening the two SCREWS (M641051: M4×6 Terminal
screw) that fix it and then shifting it leftward.
Figure 2.7-3
3 ► Unlock each of the eight CRAMPS (P/N: 731-1415, J839403, J839411) and release the PC
wires
Figure 2.7-4
4 ► Pull out the PC wires from the Wire-fix of the FRONT COVER S1 (P/N: 741-8051) to
slacken them.
Figure 2.7-5
5 ► Remove the three HEXAGON SOCKET HEAD CAP BOLTS (M330621: M4×16) that fix the
PC unit and then remove the PC unit shifting it in the direction indicated in the figure below.
Figure 2.7-6
6 ► Place the PC UNIT temporarily on the left area of the instrument. When doging this, keep
the base part of the UNIT opened.
Figure 2.7-7
1 ► Open the TOP COVER, and unfasten the two SCREWS (M641009: M4×8 TP screw) that fix
the left part of the GUIDE COVER (P/N: 741-8070)
Figure 2.7-8
2 ► Remove the two SCREWS (M641009: M4×8 TP screw) that fix the right part of the GUIDE
COVER.
Figure 2.7-9
Figure 2.7-10
4 ► Fix the SCREWS unfastened in 1 ►(Figure 2.7-8) and also the SCREWS detached in 2
►(Figure 2.7-9), to the TOP COVER
Figure 2.7-11
5 ► Remove the eight SCREWS (M641136: M4×10 SCREW WITH SPRING WASHER,
M871105: SCREW WITH WASHER) that fix the TOP COVER.
Figure 2.7-12
6 ► Lift the TOP COVER upward to disconnect it from HINGE PLATES (P/N: 741-8074)
Figure 2.7-13
Figure 2.7-14
2 ► Remove the two SCREWS (M641009: M4×8 TP screw) that fix the FRONT COVER S3 and
detach it by lifting it upward.
Figure 2.7-15
Figure 2.7-16
2 ► Remove the three SCREWS (M641009: M4×8 TP screw, M641056: M4×16 Terminal screw)
that fix the FRONT COVER S2. And remove the FRONT COVER S2 by pulling it frontward.
Figure 2.7-17
1 ► Remove the three SCREWS (M641052: M4×8 Terminal screw, M641054: M4×12 Terminal
screw, M641056: M4×16 Terminal screw) that fix the FRONT COVER S1. And remove the
FRONT COVER S1 by lifting it upward. When doing this, be careful not to hurt the inside
surface of the FRONT COVER S1.
Figure 2.7-18
1 ► Bring the FRONT COVER S1 above the mounting position, and lower it carefully paying
attention to its alignment. When doing this, be careful not to hurt the inside surface of the
FRONT COVER S1. After placing it on the main unit base, temporarily fix the FRONT
COVER S1 to the main unit base using three SCREWS (M641052: M4×8 Terminal screw,
M641054: M4×12 Terminal screw, and M641056:M4×16 Terminal screw)
Figure 2.7-19
1 ► Shift the FRONT COVER S2 in the direction indicated in the figure below. After inserting the
pins of the FRONT COVER S2 into the receptive holes on FRONT COVER S2 (on both right
and left sides), fix the FRONT COVER S2 to the main unit base using three SCREWS
(M641009: M4×8 TP screw, M641056: M4×16 Terminal screw). When doing this, keep the
FRONT COVER S2 pressed in the direction indicated in the figure below.
Figure 2.7-20
Figure 2.7-21
1 ► Keeping the FRONT COVER S1 pressed to the FRONT COVER S2, fasten tightly the three
SCREWS that were fastened temporarily in 1 ► (Figure 2.7-19), to fix the FRONT COVER
S1 to the main unit base. When doing this, pay attention to minimize the gap between the
mating faces of the FRONT COVER S1 and FRONT COVER S2 (on both right and left
sides).
Figure 2.7-22
1 ► Place the FRONT COVER S3 in the mounting position and fix it to the FRONT COVER S2
using two SCREWS (M641009: M4×8 TP screw).
Figure 2.7-23
Figure 2.7-24
1 ► Using your hand, raise the two HINGE PLATES (P/N: 741-8074) so that they stand vertical
to the Side A (see the figure below) of the FRONT COVER S1.
Figure 2.7-25
2 ► Insert the two HINGE PLATES into the TOP COVER’s receptive holes and fix it temporarily
by using eight SCREWS (M641136: M4×10 SCREW WITH SPRING WASHER, M871105:
SCREW WITH WASHER).
Figure 2.7-26
Figure 2.7-27
3.3 ► Fasten tightly the SCREWS that were fasten temporarily in 2 ►(Figure 2.7-25) to fix the
TOP COVER to the main unit base.
4 ► Open the TOP COVER, and unfasten the two SCREWS (M641009: M4×8 TP screw) that fix
the left part of the HINGE GUIDE 1 (P/N: 741-8061)
Figure 2.7-28
5 ► Remove the two SCREWS (M641009: M4×8 TP screw) that fix the right part of the HINGE
GUIDE 1.
Figure 2.7-29
6 ► Shift the GUIDE COVER (P/N: 741-8070) in the direction indicated in the figure below.
Using the SCREWS you have removed in 5 ►(Figure 2.7-29), fix the GUIDE COVER to the
TOP COVER.
Figure 2.7-30
1 ► Shift the PC unit with its base part closed in the direction indicated in the figure below, and
place it on the FRONT COVER S1. Using three HEXAGON SOCKET HEAD CAP BOLTS
(M330621: M4×16), fix the PC unit to the FRONT COVER S1.
Figure 2.7-31
2 ► Properly route the PC wires through the eight CRAMPS and the Wire-fix of the FRONT
COVER S1 to secure them.
Figure 2.7-32
3 ► Shift the WIRE COVER (1) (P/N: 741-8069) in the direction indicated in the figure below and
place it on the FRONT COVER S1. Using two SCREWS (M641051: M4×6 Terminal screw),
fix the WIRE COVER (1) to the FRONT COVER S1.
Figure 2.7-33
Figure 2.7-34
1 ► Place the REAR COVER in the mounting position and fix it to the main base unit using two
SCREWS (M641007: M4×6 TP screw). When doing this, be careful no to sandwitch the PC
wires you have routed in 2 ► (Figure 2.7-32).
Figure 2.7-35
1 ► Detach the REAR COVER by removing the two SCREWS (M641007: M4×6 TP screw) that
fix it.
Figure 2.7-36
2 ► Detach the WIRE COVER (1) (P/N: 741-8069) by removing the two SCREWS (M641051:
M4×6 Terminal screw) and then shift it leftward.
Figure 2.7-37
3 ► Unlock each of the eight CRAMPS (P/N: 731-1415, J839403, J839411) and release the
wires that run through them.
Figure 2.7-38
4 ► Pull out the PC wires from the Wire-fix of the FRONT COVER S1 (P/N: 741-8051) to
slacken them.
Figure 2.7-39
5 ► Remove the three HEXAGON SOCKET HEAD CAP BOLTS (M330621: M4×16) and detach
the PC Unit by shifting it in the direction indicated in the figure below.
Figure 2.7-40
6 ► Place the PC unit temporarily on the left of the instrument keeping the base part of the unit
opened.
Figure 2.7-41
1 ► Shif the PC unit in the direction indicated in the figure below and place it on the FRONT
COVER S1. Fix the PC unit to the FRONT COVER S1 using three HEXAGON SOCKET
HEAD CAP BOLTS (M330621:M4×16)
Figure 2.7-42
2 ► Properly route the PC wires through the eight CRAMPS and the Wire-fix of the FRONT
COVER S1 to secure them.
Figure 2.7-43
3 ► Shift the WIRE COVER (1) (P/N: 741-8069) and place it on the FRONT COVER S1. After
that, fix them using two SCREWS (M641051: M4×6 Terminal screw)
Figure 2.7-44
Figure 2.7-45
1 ► Place the REAR COVER on the mouting position and fix it to the main unit base using two
SCREWS (M641007: M4×6 TP). When doing this, be careful no to sandwitch the PC wires
you have routed in 2 ►(Figure 2.7-43).
Figure 2.7-46
1 Tube 1 741-0807
2 Tube 2 741-0808
7 Tube 7 741-0809
8 Tube 8 741-0543
9 Tube 9 741-0803
Introduction
The system volume value is needed by the system to ensure that the sample is transported to the
measuring cell working electrode surface. If the system volume is wrong the results may vary. (Bad
CV´s )
Note: Whenever a part on the sipper line (excluding the measuring cell ) is exchanged the system
volume has to be determined.
Place 1 (one) set of system reagent ProCell/CleanCell bottles filled with ProCell/CleanCell at Position
1, right side in the system reagent container.
3.2.2.1 Procedure
► Make sure that the instrument is in standby status.
► Touch the [Utility] button.
► Touch the [Maintenance] button.
► Touch the [Service] button.
► Touch the [System Volume Check] button and Touch the [Select] button.
► Touch the START button in the screen.
Note: After the procedure has been finished successfully the instrument returns to Standby. If an
error message appears, refer to troubleshooting. The system volume value is not displayed.
Figure 3.2-1
Figure 3.2-2
Introduction
This cleaning procedure is also part of the customer maintenance (see Operator’s Manual) and should
be carried out frequently.
Service liquid flow cleaning (equivalent to 3 customer cleanings) should only be performed in the event
of high grade contamination problems (e.g. heat pipe blockage)
Introduction
Contamination in the sipper system could potentially degrade sample accuracy and precision, or
possibly block the measuring cell flow path. Liquid flow cleaning keeps the sipper liquid flow path clean
and maintains the integrity of the measuring cell.
The customer should perform LFC on a frequent basis. Please train all your customers according to
the procedure described below and inform them about the importance of keeping the liquid flow path
clean and the measuring cell at the best performance level.
If operators perform this procedure frequently it will lead to fewer customer complaints.
Required materials:SysClean adapter Ident. No.: 1933159.910
SysClean Solution Ident. No.: 1298500
Recommended frequency*: Every two weeks or after 2500 - 3000 cycles
∗ If customer maintenance has been done as recommended, it is only necessary to carry out an additional customer
cleaning at the PM visit. The service LFC is only necessary during PM if the sipper line is highly contaminated or if
LFC has not been carried out regularly by the customer.
Operator time: Approx. 3 minutes
Analyzer time: User LFC: Approx. 16 minutes (user)
Service LFC: Approx. 32 min + time for MC prepare cycles
Precautions: The cleaning solution is corrosive. Wear protective gloves and
glasses when handling this solution.
Figure 3.3-1
Insert the adapter with the marking "USER" in direction towards the
back of the instrument. Fill the SysClean solution to the edge of the fill
mark on the adapter (approx. 9 ml). PC/CC bottle has to be placed at
the left side.
2. Service Same procedure as for customers but using the Service position of
the adapter.
3. User Touch the Liquid Flow Cleaning button on the screen (Screen: Util /
/Service Maintenance).
4. User The number of LFCs should be 1 for User maintenance.
5. Service The number of LFCs should be 3 for service.
Take out the SysClean adapter and place ProCell and CleanCell
bottles in all positions.
Perform Measuring Cell Preparation cycles. [Press ”MC
Preparation” in Maintenance screen; select number of cycles]
Start the LFC procedure shortly after filling up the SysClean solution.
After the instrument returns to Stand by remove the SysClean adapter and discard the remaining ISE
solution.
To avoid an unpleasant odor, please empty the waste container at the end of the liquid flow cleaning
procedure.
After a Service LFC has been performed customers should check all controls and carry out a
calibration if necessary. This is not necessary for normal routine with customer LFC.
Note: This is not a part of the routine maintenance for the e411. This procedure should only be
used if contamination has occurred.
Procedure
1 ► Remove the system water container. Fill with 3% Perhydrol solution and empty the
liquid waste container.(Put about 100 ml 30 % Perhydrol solution in the tank, and fill
up with purified water to the 3 L mark. (3.5L volume). Mix carefully. Replace the
system water container.
4 ► Start. The procedure takes 50 min and consumes about 2.8 L. When cleaning is
finished the status changes to Stand-by. Allow the solution to react for 20 min.
6 ► Clean the system water container and refill with system water. Replace the container
on system.
7 ► Touch the Cleaning button again. Rinsing with system water starts (50 min).
8 ► Repeat points 5 to 7 twice to guarantee that the whole system is filled with system
water.
9 ► After finishing, the top of the screen changes to Stand-by. Please empty the liquid
waste container and refill the system water container with system water. The system
is now ready for the normal routine.
In case of frequent high water conductivity errors, check that the pipettor LLD voltage is 1.5 V or lower.
Otherwise, repeat the cleaning process with system water until the voltage is reduced to 1.5 V.
In case of strong contamination repeat the whole cleaning procedure again.
Note: There are some instruments that generate the alarm "Distilled Water: 25-01-01" during
Cleaning and the status shifts to Stand-by. However there is no problem with the
performance of cleaning function if enough system liquid (around 3 L) has run.
3.4 SysWash
3.4.1 Introduction
SysWash is used as an additive to the wash water in the e411. SysWash increases the rinsing
efficiency between pipetting steps.
SysWash also contains a preservative to protect the system from contamination (e.g. by bacteria).
The detergent solution is particularly necessary when performing analyses using e411 infectious
disease assays, but can also be used for the other parameters.
Before SysWash is regularly used on the instrument, service has to carry out the Rinsing procedure
(see below).
After the rinsing procedure has been finished, SysWash has to be used in the system water in the
recommended concentration (prepared working solution contains 0.075% detergent).
Note: When pouring in the water make sure that it flows quickly enough, but avoid the formation of
foam. This ensures that the additive will be dispersed evenly in the wash water, so no
additional mixing will be necessary.
4.1 Overview
4.1.1 Location
See Chapter 4.2.
1. Reagent Disk Mechanism: The reagent disk mechanism causes the rotation of the reagent disk to
bring the reagent in the pipetting position.
2. Reagent Cooling Unit: The reagent cooling unit is a Peltier-element temperature controller that
controls the reagent temperature.
1. Pipetter Syringe (on the left side) - Aspirates and pipettes a specified amount of each sample,
reagent, or beads using the pipetter probe and tube.
2. Sipper Syringe (on the right side) - Aspirates the reaction mixture through the sipper probe, tube,
and measuring cell.
4.2.1 Analyzer
Figure 4.2-1
1 Pipetter Mechanism
2 Cap Open/Close Mechanism
3 Beads Mixer Mechanism
4 System Reagent Mechanism
5 Sipper Mechanism
6 Sipper Safety Cover
7 Syringe Mechanism
8 Gripper Mechanism
9 Solid Waste Mechanism
10 Reagent Disk Mechanism
Figure 4.2-2
1 BCR Mechanism
2 Cap Open/Close Mechanism
3 Reagent Disk Mechanism
4 Sample Disk Mechanism
Figure 4.2-3
1 Detection Mechanism
Figure 4.2-4
1 C Line
2 BCR & BCR Board(P/N741-5056) in the Analyzer
3 STAT Sample Position
4 B Line
5 A Line
1 LED201
2 PI202 (A Line Rack Feed End Position)
3 PI201 (A Line HOME Position)
4 PI200 (A Line Tray Detection)
5 Motor SM200
1 LED202
2 PI224 (C Line Tray Detection)
3 PI226 (C Line Tray Full 2 Detection)
4 PI225(C Line Tray Full 1 Detection)
5 PI223 (C Line Rack Full Detection)
6 Rack Feed Arm
7 PI222(C Line Rack Feed End Position)
8 PI221(C Line HOME Position)
Table 4.3-2
Part number Part number
Page of Parts Part name
(motor with lead (motor itself, without
List (motor itself, without lead wires)
wires) lead wires)
8 741-0551 741-1120 Pulse Motor KH56LM2-027
11 741-4244 741-1120 Pulse Motor KH56LM2-027
13 741-4247 741-1049 Pulse Motor KH42HM2-052
20 741-4320 K591137 Pulse Motor KH42JM2-011
24 741-4319 741-3132 Pulse Motor KH56LM2-049
26 741-4257 741-1048 Pulse Motor KH56HM2-016
28 741-4248 741-1120 Pulse Motor KH56LM2-027
29 741-4251 741-1049 Pulse Motor KH42HM2-052
32 741-0806 741-1560 Pulse Motor KH56KM2-016
37 741-8551 741-1049 Pulse Motor KH42HM2-052
41 741-0603 741-1049 Pulse Motor KH42HM2-052
43 741-0600 741-1049 Pulse Motor KH42HM2-052
44 741-0581 741-1048 Pulse Motor KH56HM2-016
4.3.7 List of LEDs and Fan Motor for Rack Sampler System
No. Item Description Type Parts No. Voltage
1 LED201 A Line Tray Exchange BD-703G J853635 5V
2 LED202 C Line Tray Exchange BD-703G ↑ 5V
3 Fan Motor Cooling of DC Power Supply MD825BM-24 L542069 24V
Note: Part No. and Part name of each mechanism are different depending on the manufacturing
time. Please refer to Parts List for details.
4.4.1.1 Overview
The sample disk drive mechanism is a unit that activates the rotation movement of sample disk
and transfers samples to the pipetting position.
4.4.1.2 Location
• Sample Disk
• Rotation Mechanism
Sample Disk
The sample disk can have positions for thirty (30) samples and controls and be removed from the
device in a single operation. Upon delivery, the Sample Disk is wrapped independently from the
main body as an accessory.
• Φ13mm x 75 mm
• Φ13mm x 100 mm
• Φ16mm x 75 mm
• Φ16mm x 100 mm
The 2~3 ml calibrator bottle can be used for the sample disk.
Rotation Mechanism
The rotation mechanism is a mechanism activated by the timing belt and the motor that activates
the rotation movement of the sample disk.
Figure 4.4-1
Figure 4.4-2
Figure 4.4-3
Table 4.4-1
Rotation Motor
Disassembly
Assembly
1 ► Temporarily fix the motor with the M4 screw, and put the timing belt on the small pulley
for the motor.
2 ► Put the timing belt SR on the big pulley, then rotate the pulley by a small amount (2 or 3
teeth) in either direction to ensure a solid fit.
3 ► Adjust the belt tension. The belt tension is indicated in the figure below.
Figure 4.4-4
Big Pulley
Disassembly
When removing the big pulley, remove it before dismounting the unit from the main body.
Assembly
1 ► Put the big pulley horizontally on the bearing, and push firmly to set.
2 ► Insert the small bearing, GND spring, GND ring, and center holder 2 sequentially and fix
with the screw.
3 ► Install the rotation motor.
4 ► Adjust the belt tension as described above.
Disassembly
Assembly
Check the motion via Reset Motion and confirm that Position No. 1 stops at the sampling position.
4.4.2.1 Overview
The reagent disk drive mechanism is a unit that activates the rotation movement of reagent disk
and transfers reagents to the pipetting positions.
4.4.2.2 Location
• Reagent Disk
• Rotation Mechanism
1. Reagent Disk
The reagent disk can be used to set more than eighteen (18) reagent rack packs on the
circumference. On delivery, the reagent disk is packaged independently from the main body as
an accessory.
2. Rotation Mechanism
The rotation mechanism is activated by the timing belt and the motor that activates the rotation
movement of the reagent disk.
Figure 4.4-5
Figure 4.4-6
Note: Part No. and Part name of each mechanism are different depending on the manufacturing
time. Please refer to Parts List for details.
4.4.1.1 Overview
The sample disk drive mechanism is a unit that activates the rotation movement of sample disk
and transfers samples to the pipetting position.
4.4.1.2 Location
• Sample Disk
• Rotation Mechanism
Sample Disk
The sample disk can have positions for thirty (30) samples and controls and be removed from the
device in a single operation. Upon delivery, the Sample Disk is wrapped independently from the
main body as an accessory.
• Φ13mm x 75 mm
• Φ13mm x 100 mm
• Φ16mm x 75 mm
• Φ16mm x 100 mm
The 2~3 ml calibrator bottle can be used for the sample disk.
Rotation Mechanism
The rotation mechanism is a mechanism activated by the timing belt and the motor that activates
the rotation movement of the sample disk.
Figure 4.4-1
Figure 4.4-2
Figure 4.4-3
Table 4.4-1
Rotation Motor
Disassembly
Assembly
1 ► Temporarily fix the motor with the M4 screw, and put the timing belt on the small pulley
for the motor.
2 ► Put the timing belt SR on the big pulley, then rotate the pulley by a small amount (2 or 3
teeth) in either direction to ensure a solid fit.
3 ► Adjust the belt tension. The belt tension is indicated in the figure below.
Figure 4.4-4
Big Pulley
Disassembly
When removing the big pulley, remove it before dismounting the unit from the main body.
Assembly
1 ► Put the big pulley horizontally on the bearing, and push firmly to set.
2 ► Insert the small bearing, GND spring, GND ring, and center holder 2 sequentially and fix
with the screw.
Version 1.0 – May 2006 4 - 18 Chapter 4.4
RD/Hitachi cobas e411 Service Manual
Disassembly
Assembly
Check the motion via Reset Motion and confirm that Position No. 1 stops at the sampling position.
4.4.2.1 Overview
The reagent disk drive mechanism is a unit that activates the rotation movement of reagent disk
and transfers reagents to the pipetting positions.
4.4.2.2 Location
• Reagent Disk
• Rotation Mechanism
1. Reagent Disk
The reagent disk can be used to set more than eighteen (18) reagent rack packs on the
circumference. On delivery, the reagent disk is packaged independently from the main body as
an accessory.
2. Rotation Mechanism
The rotation mechanism is activated by the timing belt and the motor that activates the rotation
movement of the reagent disk.
Figure 4.4-5
Figure 4.4-6
Rotation Motor
Disassembly
Assembly
1 ► Temporarily fix the motor with the M4 screw, and put the timing belt on the small pulley
for the motor.
2 ► Put the timing belt SR on the big pulley, and turn the pulley a small amount (2 or 3
teeth) in either direction to ensure a solid fit .
3 ► Adjust the belt tension. The belt tension is indicated in the figure below.
Figure 4.4-7
Big Pulley
Disassembly
When removing the big pulley, remove it before dismounting the unit from the main body.
• Support the outer edge by hand. (The small bearing does not need to be removed.)
Assembly
1 ► Put the big pulley horizontally on the bearing and insert it firmly.
2 ► Insert the small bearing, GND spring, GND ring, and center holder 2 sequentially and fix
with the screw.
3 ► Install the rotation motor.
4 ► Adjust the belt tension.
Disassembly
Assembly
Check the motion via Reset Motion and confirm that Position No.1 stops at the Cap O/C position.
Overview
Location
Functional Details
The Peltier cooling unit consists of a Peltier element, cooling fin, and DC fan. It maintains the
reagent temperature at 20°C ± 3°C using two (2)sets.
The copper jacket part consists of a pot-shaped copper jacket and insulating material that covers
the round copper plate (THK: 5mm) and maintains a constant internal temperature.
3. R. Lid
The R. lid prevents temperature increases via the top of the reagent cooler unit.
This window allows the labels on reagent bottles to be read by the two-dimensional bar-code
reader. The heater prevents condensation.
The R. lid O/C Sensor part detects opening/closing of the R lid via a microswitch.
Explanatory Diagram
Figure 4.4-8
Parts List
• Microcooler assembly
• Thermistor R assembly
• Outer ring, inner ring
• Window heater
• R. lid
1. Microcooler Assembly
Disassembly
Assembly
1 ► Apply a thin coat of silicone grease (heat-conducting) to the cooling surface of the
microcooler.
2 ► Fix the two (2) microcoolers evenly using screws.
3 ► Push the pins for window heater to #3 and #6 of J523 connector (no polarity in the
window heater).
4 ► Assemble the thermostat, packing S, and J523 connector.
2. Thermistor R. Assembly
Disassembly
Assembly
To assemble the thermistor R assembly, re-install the parts removed during disassembly
procedures (see above) in the reverse order.
Disassembly
Assembly
To assemble the outer ring and inner ring, re-install the parts removed during disassembly
procedures (see above) in the reverse order.
4. Window Heater
Disassembly
Assembly
To assemble the window heater, re-install the parts removed during disassembly procedures (see
above) in the reverse order.
5. R. Lid
Disassembly
Assembly
1 ► Apply a very thin coat of silicone grease (heat-conducting) to the tip of the lock knob.
2 ► Push and insert the hold rubber in the R. lid.
3 ► Turn the lock knob to insert it into the R. lid.
4.4.3.1 Overview
The cap open/close mechanism is a unit that opens/closes the snap cap of RackPacks according
to the timing of beads stirring or reagent pipetting by means of three (3) independent hooks.
4.4.3.2 Location
• Rotation mechanism
• Front-rear mechanism
• RackPack stopper
Rotation Mechanism
The rotation mechanism activates three (3) hooks to start rotation movement to open the cap.
The rotation mechanism constitutes the crank mechanism with the fork link that rotates around
the center of a spinning fulcrum of hook and the drive link installed to the motor shaft with speed
reduction control. The three (3) hooks are installed in a special manner:
Front-Rear Mechanism
RackPack Stopper
The RackPack stopper is a rod-shaped mechanism that initiates front-rear movement at the same
time as the rotation mechanism part executes front-rear movement. Upon opening the cap, the
RackPack stopper prevents the rack pack from rising.
Stroke: 58.2mm
Figure 4.4-9
Removal:
• Rear cover
• R. lid
• Y cover
• Front cover assembly
• Table cover R
• Table cover L
• See 4.4.2.
Mounting
Mount the units by re-installing the parts removed during the removal procedures (see above) in
the reverse order.
Removal
Mounting
Assemble the units by re-installing the parts removed during the removal procedures (see above)
in the reverse order.
• Rotation mechanism
• Front-rear mechanism
Disassembly
1 ► Unscrew the M4 x 6 screws that fix the clamp on the plug (J532) side.
2 ► Remove the tie wraps that fix the CAP O/C cord assembly and the J532 cord assembly.
3 ► Take out the P532 socket.
4 ► Dismount the photo interrupter (PI033) and the cord together.
• You can replace the photo interrupter (PI033) by removing the connector.
5 ► Detach the E-type fix ring 4 located at the tip of the rear part of the RackPack stopper
assembly.
6 ► Pull out the RackPack stopper assembly. Do not misplace the dumping spring.
7 ► Detach the E-type fix ring 6 attached to the inside of the cap open/close mechanism
base from the two (2) slick shafts.
8 ► Support the link mechanism assembly and pull out the two (2) slick shafts.
9 ► Remove the timing belt from the motor bracket.
10 ► Unscrew the two (2) M4 x 4 point screws that fix the link pin assembly.
11 ► Pull out the J532 cord assembly.
Assembly
1 ► Assemble the unit by re-installing the parts removed during the disassembly procedure
(see above) in the reverse order.
2 ► Fix the link assembly by adjusting bearing to fork link assembly.
2. Front-Rear Mechanism
Disassembly
1 ► Unscrew the two (2) M4 x 6 screws that fix the pulley base.
2 ► Remove the timing belt from the motor bracket.
3 ► Detach one (1) E-type fix ring 4 from the pulley shaft (See Figure**), and remove pulley
S5M and timing belt.
4 ► Remove the photo interrupters (PI031, PI032) and the cord together.
• You can replace the photo Interrupters (PI031, PI032) by removing the connector.
5 ► Unplug plugs (J530, J531) from the cap open/close mechanism base.
6 ► Unscrew the two (2) M4 x 6 screws that fix the clamp.
7 ► Take out the tie wraps that fix the CAP O/C cord assembly and the J532 cord assembly.
• You can replace the photo interrupter (PI033) by removing the connector.
10 ► Pull out the cord from the clamp that fixes the cord.
11 ► Unscrew the two (2) M4 x 16 screws that fix stepping motor KH56.
12 ► Unscrew two (2) M4 x 4 W point screws, and remove pulley S5M.
Location of PCP
Figure 4.4-10
4.4.4.1 Overview
The beads mixer mechanism is a mechanism that stirs the beads inside the reagent rack pack
immediately before the pipetting of the reagent.
4.4.4.2 Location
The Stirrer Head Vertical Movement Mechanism moves the stirring paddle up and down over the
RackPack with its cap open and over the wash bath. The stirrer head that supports the stirring
paddle and the vertical movement mechanism are vertically connected with an arm pipe. A nut
holder and slide guide that are installed on the inner and outer rings of the ball bearings are
installed in the vertical movement mechanism. A spline nut is inserted in the nut holder. The slide
guide and the nut holder shift up and down guided by the spline shaft and guide shaft; the slide
guide stops the rotation and the vertical movement, and the nut holder controls the rotation
triggered by the rotation of the spline shaft and makes vertical movement.
Pulleys are positioned above and below the moving parts keeping a sufficient axle distance from
the stirring paddle so that they will not contact with each other when the stirring paddle makes
strokes. The slide guide, which is hooked up to the timing belt, is moved vertically, driven by the
stepping motor.
A nut holder is fixed on the inner ring of the ball bearings and a slide guide on the outer ring. An
arm pipe that secures the stirrer head is fixed to the nut holder. The ball bearings guide the arm
pipe and nut holder to swing together within a range limited by the stopper that is installed on the
slide guide.
Driven by the stepping motor, the moving parts cause rotation of the pulley that is attached to the
bottom of the spline shaft. The rotating force is conveyed to the pulley via the timing belt.
The stirring paddle rotating mechanism is driven by a DC motor and its pulley causes the rotation
of the pulley that is attached to one end of the stirring paddle. The rotating force is conveyed to
the rotation pulley via the rectangular elastic belt connected to it.
Figure 4.4-11
First remove the reagent disk mechanism (see Chapter 4.4.2.6). After removing the reagent disk,
unscrew three (3) screws, two (2) connectors (P540 and P541), and one (1) grounding terminal
set screw to remove the beads mixer unit from the main unit base.
Disassembly Procedure
► Take out the head cover and pull out the DC motor wiring connector pin inside.
► Unscrew two (2) W point screws on the side of the stirrer arm head.
3 ► Remove the mechanism that is installed in the spline hut and the ball spline shaft
together.
6 ► Remove the lead wire and grounding wire from the DC motor.
► To remove the lead wire from the DC motor, disconnect the connector terminal from
connector J541 and pull the wire through the arm pipe.
► To remove the grounding wire, remove the screw from the terminal attached to the hut
holder and remove the wire from the O ring.
Assembly Procedure
Assemble the mechanism by re-installing the parts removed during the disassembly procedure
(see above) in the reverse order.
Disassembly Procedure
Assembly Procedure
Assemble the moving parts by re-installing the parts removed during the disassembly procedure
(see above) in the reverse order.
► Adjust the belt by sliding the stepping motor (see Figure 4.4-12, Adjustment of Belt
Tension).
Disassembly Procedure
1 ► Remove the stirrer belt (a rectangular elastic belt) from the pulley.
2 ► Remove the mixing motor assembly.
► Loosen the bottom parts of the two W point screws of the joint.
4 ► Pull out the stirrer pulley and shaft from the bearings.
► Loosen the top parts of the two W point screws of the joint and remove them.
► Remove the E ring.
Assembly Procedure
Assemble the stirrer arm head by re-installing the parts removed during the disassembly
procedure (see above) in the reverse order.
Location of PCP
Figure 4.4-13
4.4.5.1 Overview
The pipetter mechanism is a mechanism that picks up disposable tips and pipettes sample,
calibrator, reagent (beads, R1, R2 and diluent, etc.) into the reaction vessel by means of an X-Z
moving mechanism.
4.4.5.2 Location
The X-axis movement mechanism moves the pipetter probe to each sampling position and
replaces the disposable tip. The X-axis movement mechanism drives Z-axis mechanism that is
mounted on the timing belt with a stepping motor.
The Z-axis movement mechanism moves the pipetter probe and disposable tip to each sampling
height position, and drives the nozzle head that is mounted on the timing belt with a stepping
motor via the LM guide.
Tip/Vessel Buffer
• The tip/vessel buffer is the area where the gripper places the disposable tips and the reaction
vessels so that the pipetter probe has access to them.
• There are sensors in the first two positions to detect the presence of a tip; the gripper
mechanism also checks the tip and vessel placement
• The tip eject position is at the edge of tip/vessel buffer.
This detects clots in sample, calibrator, reagent or diluent during piping by monitoring pressure
changes inside the tube.
Figure 4.4-14
Figure 4.4-15
Table 4.4-2
Note: Shape of the tip of P.Nozzle is different between E2010 and e411. In addition, shape of the
tip of e411 P.Nozzel is similar to E170 P.Nozzle. But internal structure is different
Figure 4.4-16
7) Tip/vessel buffer
Disassembly Procedure
► Remove the fixing screw which is located on the right side of center cover.
► Remove the motor bracket and X pulley from P side base.
► Remove the timing belt.
Assembly Procedure
Tension : Press middle of the timing belt with a vertical force of 2±0.5 N, then adjust pulley
position by loosening the X pulley screw as shown in the diagram below.
Figure 4.4-17
Disassembly Procedure
Assembly Procedure
Tension: Push middle of the timing belt with a vertical force of 1.5±0.3 N, then adjust motor
position by loosening the motor fixing screw as shown in the diagram below.
Figure 4.4-18
► Remove the photo interrupter from the bracket of the Z-axis movement unit, and pull out
to the side.
Assembly Procedure
Assembling Procedure
Disassembly Procedure
Note: Do not touch the top of the screw part of the pressure sensor with your fingers.
Assembly Procedure
► Attach about 70mm spiral tube onto the pressure sensor cable.
► Remove cap or cover from top of the pressure sensor.
► Put seal packing on the screw part of the pressure sensor, and attach to the tube joint.
► Mount the rear cover of the pipetter mechanism.
► Fix the spiral tube part of the pressure sensor cable taking care not to catch the cable in
the open close mechanism.
► Fix the pressure sensor cable to the rear cover.
► Connect the cable to ANG-EP PCB.
► Tighten the screw of the driver base.
FPC Cable
Disassembly Procedure
Assembly Procedure
Assembly Procedure
► Attach the tube nipple to the probe. (tightening torque is 2.5 kgf·cm).
► Tubing as original.
► Confirm that P arm moves smoothly without catching.
► Remove P plate.
► Cut the tie-wrap tube clamping.
► Disconnect the tube nipple the tube joint.
Assembly Procedure
► Remove the bracket of the photo interrupter from left and right of pipetter mechanism
side frame.
► Disconnect the connector.
► Remove the photo interrupter from bracket.
Assembly Procedure
► Remove the photo interrupter from the bracket of the Z-axis movement unit, and pull out
to the side.
Assembly Procedure
Assembling Procedure
Disassembly Procedure
Note: Do not touch the top of the screw part of the pressure sensor with your fingers.
Assembly Procedure
► Attach about 70mm spiral tube onto the pressure sensor cable.
► Remove cap or cover from top of the pressure sensor.
► Put seal packing on the screw part of the pressure sensor, and attach to the tube joint.
► Mount the rear cover of the pipetter mechanism.
► Fix the spiral tube part of the pressure sensor cable taking care not to catch the cable in
the open close mechanism.
► Fix the pressure sensor cable to the rear cover.
► Connect the cable to ANG-EP PCB.
► Tighten the screw of the driver base.
TUBE 465
1 ► Insert TUBE 465 into the P nozzle. Make sure that there is a nozzle seal between TUBE
465 and the P nozzle. Note that the torque of screwing the nipple should be 0.25 Nm.
2 ► Set the tube holder into both the P arm internal cover and the P arm cover, and attach
the P arm cover.
3 ► Arrange TUBE 465 as shown in Figure 4.4-24, and insert into tube joint 2 (refer to
Figure 4.4-25). Note that the torque of screwing the nipple should be 0.25 Nm.
4 ► Attach screw A of the PR cover.
5 ► Make sure that the tube moves smoothly when the P arm is moved vertically.
TUBE 510
1 ► Insert TUBE 510 into both tube joint and tube joint 2. Note that the torque of screwing
the nipple should be 0.25Nm.
2 ► Fix TUBE 510 with two (2) tie wraps (Refer to Figure 4.4-21, Figure 4.4-22 and Figure
4.4-23). Ensure that the direction of the nail of the tie wrap is the same as in the figures.
Be careful not to twist the tube.
3 ► Attach the tube guide assembly. Be careful not to nip the GNP line.
4 ► Attach screw A of the PR cover.
1 PR cover
2 M4 screws
3 Tube Joint
4 Pressure Sensor
1 Wrap tie
2 Tube510
3 Nipple
4 Inner Tube
5 Outer Tube
1 Railbase
2 Tube Joint 2
3 Tube 465
4 Ty-rap
5 Tube 510
6 Nipple
1 P Arm Cover
2 P Arm Inside Cover
3 Tube holder
4 Nozzle seal
5 P Nozzle
6 Nipple
7 Tube 465
1 Tube Joint 2
2 Tube 465
3 P Arm Cover
4.4.6.1 Overview
The gripper mechanism consists of the following 3 components and transfers tips/vessels
between the container station, buffer station, and incubator.
4.4.6.2 Location
• Base frame
• Y carriage
• X carriage
• Grip part
• Container station
• Incubator station
Base Frame
Base frame is fixed on the four (4) pillars that are secured to the main base. The carrier
mechanisms such as the Y carriage, container station, and incubator station, are attached to the
base frame.
Y Carriage
The Y carriage contains the X carriage (mentioned below). It is driven in the Y direction (back and
forth movement) and placed (at 0.25 mm/step) via the stepping motor and the pulley belt
mechanism. The Y carriage is supported by the following bearings:
• On the right side: Two (2) linear bearings that are guided by a shaft
• On the left side: One (1) radial bearing that is guided by the channel
A shielding plate on the right of the Y carriage passes via three (3) photo interrupters fixed on the
base frame and detects home position, rest position, and end position (the order starts from the
front).
X Carriage
The X carriage contains the grip parts described below. It is driven in the X direction (left and right
movement) and placed (at 0.125 mm/step) via the stepping motor and the pulley belt mechanism.
A shielding plate on right of the X carriage passes via 3 photo interrupters on the right side of the
Y Carriage and detects home position.
Grip Part
The grip part consists of the Z-driving part and the grip-driving part. The grip-driving part consists
of a rack gear driven in the Z direction (up and down shift) and placed (at 0.2 mm/step) via the
stepping motor and the Z-drive part of the rack/pinion mechanism. The grip-drive part is
supported by a linear guide fixed to the X carriage and is pulled up via a spring. A shielding plate
secured on the upper side of the grip-drive part passes via 3 photo interrupters at the upper side
of the X carriage and detects home position.
The grip-drive part is driven to open/close via a cam rotated by a rotary solenoid and a pair of
arms with a grip nail positioned to catch the cam. The tip/vessel is gripped at the nail by the
power of pull-spring attached to close a pair of arms and is positioned on X & Y where the back
arm is pushed by the cam. A shielding plate located at the tip of the front arm passes via Photo
Interrupter under the condition w/o tip/vessel and detects “None.” Also, a pair of arms are
supported and guided to up and down shift as well as holder which retains the arms above and
below and positioned to the lower edge by two (2) push-springs. If Grip-drive part is lowered to
abnormal positions on XY, the head part of tip/vessel halts the down movement of a grip nail,
arms, and a holder. A shielding plate located at the bottom side of Grip-drive part passes via
Photo Interrupter secured to the holder and detects “Abnormal Down Position.”
Container Station
The container station has a partition to hold three (3) tip containers and three (3) vessel
containers and is positioned at the base frame via two (2) pins located at the bottom side and
secured by four (4) screws.
Incubator Station
Incubator Station consists of the heat block with thirty-two (32) storage holes for vessel incubation
with one (1) vessel storage hole sipping, heater, and thermistor. It is secured to the base frame
by two (2) screws.
Figure 4.4-26
Figure 4.4-27
Figure 4.4-28
Table 4.4-3
Part name
Position No. Short Name Material number Part number
Comment
1 T/V CARRIER 741-0409
ASSY
2 PM062 Motor Y assy 741-4319
Motor Y assy;
motor & cable &
pully & dumper
3 PI063 CORD P563 Y- 741-4306
HP; PI 063
4 PI064 PI065 CORD P5634 Y- 741-4307
EP-RP; PI
064/65
5 Circuit Board 741-5047
Assy DIST-TVA
6 PM061 Motor C Assy 741-4320
with PULSE K591137
MOTOR
KH42JM2-011
Motor C assy,
motor x-move
7 Motor Cover X 741-8530
8 Holder C 741-1139
9 PM063 Pulse Motor 741-4321
PF25-48D4 Assy
Gripper-Z; UP
DOWN PF25;
motor
10 Cover Z 741-8534
11 Carrier Z assy / 741-0412 / 741-
Plate 8533
12 SOL Rotary solenoid 741-4322
2ER45-33 assy
13 FPC-C1 Cord FPC C1 741-4323
Part name
Position No. Short Name Material number Part number
Comment
From Dist-TVA to
Gripper-Z
14 FPC-C2 Cord FPC C2 741-4324
From Dist-TVA to
carrier x-motor
15 FPC-C11 Cord FPC C11 741-4325
Gripper-Z;
open/close
PI068; Abn.
PI067
16 FPC-C21 Cord FPC C21 741-4326
Gripper-Z; UP
DOWN PF25;
PI066
17 Thermostat INC 741-0761
Assy
18 Thermistor I 741-0582
19 Timing Belt C(X) 741-1143
Gripper X-belt
20 Timing Belt C(Y) 741-1124
Gripper Y-belt
Sensor
TLP830 PCB ASSY J339168
DETECT 3
TLP830
Signal photo
coupler used in
Gripper assy;
see list of
Sensors SM
chapter 4.3.2
Tool gripper 741-1471
adjustment
Gripper sensor 741-2022
tool
Part name
Position No. Short Name Material number Part number
Comment
Tool transport 741-2154
module
adjustment
For automatic
adjust
Note: When removing the FPC cable from the connector, move the slide lock of the connector
towards the cable and take out the cable.
Note: When mounting the FPC cable, mount it in the right direction and lock it.
Note: When peeling double-sided tape from the FPC cable, hold the cable so that no extra tension
is placed on the corner and peel slowly.
• Rear cover
• Gripper Y cover
• Front cover
• Table cover R
• Sipper front cover
• Container station
• S guide plate/board cover (at the right side of solid waste)
• Unscrew four (4) screws and four (4) connectors to remove the unit from the main base pillars.
• Assemble the units by re-installing the units removed during the disassembly procedure (see
above) in the reverse order.
• However, position-teaching adjustment is required after assembly. See Adjustment Operation.
1. Y motor assembly
2. P563, P564 cord assembly (Y-axis photo interrupter board)
3. DIST-TV board assembly
4. X motor assembly
5. Z motor assembly
6. Solenoid assembly
7. FPC cable C1, C2, C11, C21 assemblies (including X axis, Z axis, and photo interrupter at
grip).
8. X carriage assembly
9. Heater I assembly (including thermostat).
10. Thermistor I assembly
Y Motor Assembly
Disassembly Procedure
Assembly:
1 ► Assemble the motor by re-installing the parts removed during the disassembly
procedure (see above) in the reverse order.
2 ► Fix the belt at the position where the tension of the belt satisfies the conditions shown in
the following figure.
3 ► The cable should be straight.
Figure 4.4-29
Disassembly
Assembly
► Assemble the cord assembly by re-installing the parts removed during the disassembly
procedure (see above) in the reverse order.
Disassembly
Assembly
1 ► Assemble the DIST-TVA board assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
X Motor Assembly
Disassembly
Assembly
1 ► Assemble the X motor assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
2 ► Fix the belt at the position where the belt tension satisfies the following conditions.
Figure 4.4-30
Z Motor Assembly
Disassembly
Assembly
1 ► Assemble the Z motor assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
2 ► Fix the Z motor and the opponent rack where backlash of the rack gear and pinion of
the motor is 0.1 ~ 0.2 mm.
3 ► Hold the pinion with fingers, and confirm the backlash.
Solenoid Assembly
Disassembly
Assembly
1 ► Assemble the solenoid assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
2 ► Fix grip/trouble detect board at the position where the shielding plate is centered in
photo interrupter.
FPC cable C1, C2, C11, C21 assemblies (including X axis, Z axis, and photo interrupter at grip.
(See Handling Note for FPC Cable)
Disassembly
Assembly
► Assemble the FPC Cable C1, C2, C11, C21 Assemblies by re-installing the parts
removed during the disassembly procedure (see above) in the reverse order.
X Carriage Assembly
Disassembly
Assembly
1 ► Assemble the X carriage assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
2 ► Fix the X carriage assembly on the tip container in the position where the left surface of
the X carriage assembly and the top side of the tip container satisfy the perpendicular
angle 0.2 mm/100mm.
Disassembly
Assembly
1 ► Assemble the heater I assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
2 ► Apply sufficient Amicon epoxy resin to the heater part.
Thermistor I Assembly
Disassembly
Assembly
1 ► Assemble the thermistor I assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
2 ► Apply sufficient Amicon epoxy resin to the heater part.
Detail of incubator
Figure 4.4-31
1 Heater
2 Thermistor
3 Incubator Thermostat
4.4.7.1 Overview
The sipper transports the probe that is connected to the syringe with tubes from one station to
another via the measuring cell.
4.4.7.2 Location
See Chapter 4.2.
• Base Frame
• X Carriage
• Z Carriage
Base Frame
The base frame is fixed on the platform that is secured on the top part of the main unit base. The
X carriage and Z carriage are installed on the base frame.
X Carriage
The X carriage contains the Z carriage, which is explained below. The X carriage shifts
horizontally at a speed of 0.3 mm/step to its position, and is driven by the stepping motor and
pulley / belt mechanism. At its top, the X carriage is held up by two linear bearings that are
supported by the shaft, and at the bottom the X carriage is supported by two radial bearings.
There are two photo interrupters on the base frame. The home position sensor is on the right and
the end position sensor is on the left (as viewed from the front of the cobas e411). The X carriage
stops in six positions: CC2, AB2, CC1, AB1, Washing Pos., and Sipping Pos. in a normal
operation.
Z Carriage
The Z carriage shifts vertically at a speed of 0.1 mm/step (during the aspiration of the reaction
agent at 0.025 mm/step) to its position, and is driven by the stepping motor and the pulley / belt
mechanism. The Z carriage is supported by the linear guide that is installed on the X carriage.
The shielding plate that is installed on the right side of the Z carriage passes the photo interrupter
above the top of the X carriage and detects the home position. The probe inside the Z carriage is
moved vertically and is positioned to the bottom end by a spring. If the Z carriage descends from
an abnormal position, the mechanism interrupts the descent of the probe and the shielding plate
interrupts the beam from the photo interrupter that is installed on the Z carriage. This leads to an
error, “Abnormal Descending.”
Figure 4.4-32
Figure 4.4-33
Figure 4.4-34
Table 4.4-4
• Rear cover
• Gripper Y cover
• Front cover
Then remove the following parts from the sipper unit and take out the sipper from the main unit
platform:
• Four screws
• One tube
• Three connectors
• One grounding wire
To reassemble the sipper unit, you can re-install the parts removed in the above procedure in the
reverse order. Be sure to position the unit so that the tip of the probe passes the center of the
sipper position vessel and the reagent container cap. Also, adjust its position by tapping,
following the procedure in the adjustment procedure that is described separately.
Disassembly
Assembly
Re-install all the parts removed in the disassembly procedure in the reverse order. When
installing the X motor assembly, secure the part at the position where the belt tension satisfies the
specifications below.
Figure 4.4-35
Disassembly
1 ► Remove X carriage covers (R), (L) by unscrewing two (2) screws each.
2 ► Remove the Z motor assembly by removing two (2) screws, motor connector, and tie-
wraps.
3 ► Remove the photo interrupter holder by removing two (2) screws.
4 ► Remove the photo interrupter by removing the connector and pushing the hook on its
back.
Assembly
Re-install all the parts removed in the disassembly procedure in the reverse order. When re-
installing the Z motor assembly, be sure that the belt tension meets the specifications described
in the figure above.
3. Probe Assembly
Disassembly
Assembly
Re-install the parts removed in the disassembly procedure in the reverse order.
Disassembly
Remove the DIST-S PCB assembly by unscrewing by two (2) screws and four (4) connectors and
then release the sliding lock of the connector to the cable side, and taking out the DIST-S board.
Assembly
Re-install all the parts removed in the disassembly procedure in the reverse order.
CAUTION
Be careful not to damage the cable.
Disassembly
Assembly
Re-install all the parts removed in the disassembly procedure in the reverse order. Make sure that
none of the cables are entangled.
4.4.8.1 Overview
The system reagent mechanism is a unit that keeps the liquid temperature constant inside the AB
and CC bottles. The case of the system reagent mechanism has grooves for an AB bottle and a
CC bottle. By setting an AB bottle and a CC bottle, one at a time, it adjusts the case temperature
by means of a Peltier temperature controller.
4.4.8.2 Location
The bottle heating unit is a die-cast product made of flat aluminum (5mm thick). Its function is to
keep the liquid temperature inside the AB and CC bottles at 28 C. This aluminum case detects
the signals from the sipper probe when the sipper probe detects the liquid level in the bottle, and
is therefore insulated from other parts and the main unit base.
Insulating material is attached to the bottom and sides of the aluminum case and, in addition, the
case is enclosed in a steel case frame so that the bottles inside the case are not be subject to the
ambient temperature. In order to secure the thermostat, the case frame is installed on the main
unit base and is fixed to the cooling unit holder. The case frame is also used as a holder for the
sipper probe wash bath.
Thermostat
A Peltier element thermostat is installed on the outside bottom of the aluminum case. A thermistor
is installed in the space between the case surface and the side of the case. It measures the
difference between the ambient temperature and the case temperature, and the thermostat
regulates the case temperature.
Figure 4.4-36
Table 4.4-5
To remove the system reagent unit, first remove the following covers:
• Rear Cover
• Gripper Y Cover
• Front Cover
• SR Cover
Then, remove two (2) screws and three (3) connectors (P660, P661, and P662). Also, pull out the
fixtures of the cooling unit holder from the case table installed on the main unit base.
Disassembly Procedure
1 ► Remove the steel plate around the case cover assembly and case frame assembly.
1.1 ► Remove the sockets from connectors J660, J661, and J662.
1.2 ► Remove the four (4) screws from each.
2 ► Turn the unit upside down.
3 ► Remove the thermostat.
3.1 ► Remove the four (4) screws.
4 ► Remove the cooling unit holder.
Assembly Procedure
Assemble the system reagent unit by re-installing parts removed during the disassembly
procedure (see above) in the reverse order. When installing the thermostat, apply heat-
conducting grease to the surface that is contact with the aluminum case. Since the unit is
installed through the cooling unit holder, install bush A, bush B, and washer in this order as
shown in the figure below:
Figure 4.4-37
4.4.9.1 Overview
The syringes consist of the two systems described below. The syringes transport liquid, driven by
the pumping force generated by the up and down movement of the plunger and by the movement
of the probe of each system.
1. Pipetter Syringe (Left): Aspirates and pipettes specified amounts of the sample, reagent, and
beads via the pipetter probe and tube.
2. Sipper Syringe (Right): Aspirates specified amounts of the reaction mixture via the sipper
probe, tube, and measuring cell.
4.4.9.2 Location
• Base Frame
• Syringe Motor
• Flow Route Switch
Base Frame
The base frame is fixed to the main unit base. The syringe motor and the flow route switch are
installed on the base frame.
Syringe Motor
In the syringe motor, the slider that moves the syringe plunger moves in the vertical direction is
driven by the stepping motor and pulley / belt mechanism to a specified position. The slider is
supported by two linear bearings that are guided by the shaft. A shielding plate is installed on the
back of the slider, and when the slit passes by the photo interrupter that is located above it the
home position is detected. On the slider is the plunger assembly, which is supported by the
sealing inside the sealing body, the O ring, and the syringe, mounted in this order. These parts
are secured at the top with a nut and connected with each other by tubing. The plunger assembly,
O ring, and syringe are different in the two syringe systems. The parts for the sipper syringe (on
the right side) have larger diameters.
The sipper syringe system is connected with pinch valves SV3 (at the measuring cell) and SV4
(at the liquid waste container) by tubing.
Figure 4.4-38
The parts for the pipetter syringe system are installed on the left side of the unit, those for the
sipper syringe system on the right side.
Table 4.4-6
CAUTION
Before removing the flow route tubing, prepare a
beaker, a syringe, and a cloth wiper. Drain water from
the tubing, and wipe up any spillage.
• Rear cover
• Gripper Y cover
• Front cover
• Sipper FPC holder (secured with 3 screws)
• Sipper top cover (secured with 1 screw)
• Sipper bottom cover (secured with 2 screws)
Then, remove the three (3) screws, four (4) tubes (two small tubes, one small black tube, and one
big tube), five (5) connectors (P622, P591, P592, P593, J630), and finally remove the syringes
from the main unit base.
Disassembly Procedure
1 ► Remove the syringe motor assembly by unscrewing the upper and lower screws and
removing the tube and lead wire from the clamp.
2 ► Remove the motor assembly by unscrewing two (2) screws, removing the lead wire
from the clamp, and removing the connector and contact.
3 ► Remove the photo interrupter assembly by unscrewing one (1) screw, removing the
lead wire from the clamp, and removing the connector and contact.
Assembly Procedure
Reassemble the unit by following the disassembly procedure in the reverse order.
Note: If you have loosened the belt, readjust the belt tension by (1) loosening the tension bolt, (2)
adjusting the tension axle phase, and (3) securing the tension bolt again. (See the figure
below.)
Figure 4.4-39
1 Timing Belt
2 Tension Belt
3 Tension Axle
4 Tension Gauge (a tension of 1.67N)
5 Red mark
Disassembly Procedure
1 ► Remove the two (2) tubes, one upper and one lower.
2 ► Loosen the clamp nut on the top of the unit.
3 ► Remove the syringe plunger assembly.
Assembly Procedure
Reassemble the unit by re-installing the parts removed in the disassembly procedure in the
reverse order.
Disassembly Procedure
► Remove pinch valves SV3 and SV4 by removing two (2) screws, a tube, a connector,
and a contact.
Assembly Procedure
► Reassemble the unit by re-installing the parts removed in the disassembly procedure in
the reverse order.
4.4.10.1 Overview
The system water container contains system water for each wash bath and syringe. It consists of
the following two parts.
4.4.10.2 Location
The system water container has a handle and a cap on its upper part and a plug in its bottom part.
It is filled with system water and is connected to the piping when positioned as specified.
Buffer Tank
The buffer tank is installed on the main unit base. It has a guide for container installation and a
plug that faces the plug of the system water container. It also has a float SW and a degassing
tube in its inside upper corner. A tube for connection to the pump and a tube for connection to the
valve are connected on the back of the tank.
Figure 4.4-40
Table 4.4-7
CAUTION
Before removing the flow route tubing, prepare a
beaker, a syringe, and a wiper. Drain water from the
tubing, and wipe up any spillage.
• Rear cover
• Gripper Y cover
• Front cover
Then remove the three (3) screws, three (3) tubes. Finally, remove the system water container
unit from the main unit base.
Disassembly Procedure
► Remove the ball sealing assembly that is secured with a closed nut. When you remove
the inner closed nut, you will find a filter, a spring, ball, and an O ring installed over the
nut and packing outside the nut.
Assembly Procedure
► Reassemble the unit by re-installing the parts removed in the disassembly procedure
(see above) in the reverse order.
Disassembly Procedure
1 ► Remove the tank lid by unscrewing the seven (7) screws. When you turn the tank
upside down and remove the closed nut of the ball sealing assembly, you will find a filter,
a spring, ball, and an O ring installed over the nut.
2 ► Remove the float SW that is secured with a nut.
Assembly Procedure
► Reassemble the unit by re-installing the parts removed in the disassembly procedure
(see above) in the reverse order.
Version 1.0 – May 2006 4 - 83 Chapter 4.4
RD/Hitachi cobas e411 Service Manual
Overview
The pump assembly supplies system water from the system water container to the wash baths
and syringes via the valve block.
Functional Details
• Drain pipe
• DC magnet pump
• Valve block
Figure 4.4-41
Figure 4.4-42
Table 4.4-8
4.4.11.1 Overview
The liquid waste container consists of the following two units. It collects wash liquid and reaction
waste.
4.4.11.2 Location
The liquid waste container has a handle and a cap on its upper part. Remove the cap and
position the liquid waste container in the specified position.
Container Station
The container is located in the inner station of the container station. The inner station is attached
to the outer station that is installed in the main unit base and is pushed up by a shaft on the back
and by a spring in the front. A shielding plate is installed in the inner station. This shielding plate
is depressed as the waste liquid level increases. When it is depressed, it passes the beam from
the photo interrupter installed in the outer station and thus indicates when the waste liquid
container is full. A pressure plate with a shaft and coil spring is attached to the front of the inner
station. This pressure plate is installed so that it pushes in the container. A shielding plate is
installed on one end of the pressure plate. If a container is not installed, the pressure plate will
cause the photo interrupter beam to come through the shielding plate and the “no container” state
is detected.
Figure 4.4-43
Table 4.4-9
• Rear cover
• Gripper Y cover
• Front cover
Then remove the three (3) screws, and then the liquid waste container unit from the main unit
base.
• Photo Interrupter
• Inner Station and Pressure Plate Assembly
Photo Interrupter
Disassembly Procedure
Assembly Procedure
► Reassemble the unit by re-installing the parts removed in the disassembly procedure
(see above) in the reverse order.
Disassembly Procedure
Assembly Procedure
► Reassemble the unit by re-installing the parts removed in the disassembly procedure
(see above) in the reverse order.
4.4.11.7 Adjustment Procedure regarding Liquid Waste Full after few cycles
during installation
Check Procedure
1 ► Make sure that the instrument is in Stand-by status by checking that Stand-by appears
on the screen.
2 ► Fill with distilled water ta depth of 115 mm from the bottom of waste tank (equal to 3.4
liters).
3 ► Make sure that a Waste Full Error is not shown in Reset status. If a Waste Full Error is
shown, adjust according to the adjustment procedure.
4 ► Add a further 200 ml of distilled water to the tank.
Make sure that a Waste Full Error is shown in Reset status. If a Waste Full Error is shown, the
setting is OK. If not, adjust according to the adjustment procedure.
Adjustment Procedure
6 ► Find a border point where the voltage changes between high (ca. 5V) and low (ca. 0V)
by turning the minus screw located at the top of support S (P/N741-8621) with a minus
screwdriver. Measure the voltage between Pin #5 and Pin #6 on connector J631.
7 ► Fasten the hex nut at the border point.
8 ► Check the function again according to the check procedure.
4.4.12.1 Overview
The solid waste mechanism is designed to efficiently collect tips and vessels used by the
analyzer. It levels the used tips and vessels collected by vibrating the waste box periodically.
4.4.12.2 Location
The solid waste mechanism generates vibration towards the long dimension of the waste box set
in the unit. The crank mechanism consists of the assembly, lever and roller bearing that are fixed
to the stepping motor and motor shaft as well as the drive tray that holds a tray to settle waste in
the box. The solid waste mechanism generates straight-line forward & reverse movement
referenced to the long dimension.
Stroke 30 mm
Figure 4.4-45
Table 4.4-10
Pos.
Part Name Part No.
No.
1 Motor S Assy 741-0581*
2 Cord J641 741-4311
3 SW Retainer 741-1369
4 Sensor GP1A34LC J339165
5 Clamp CKS-05-L J839400
6 Clamp CKS-10-L J839402
7 Wire Binder TY-23M J839050
8 Fixture 741-1361
9 Tray 741-0534
∗ Refer to Motor Reference List, Chapter 4.3.1.
1 ► Confirm that the disposable tray assembly (waste box) is not left in the equipment.
Remove if still present.
2 ► Remove the gripper of each cover (see 4.4.8.6).
3 ► Remove FDD base S, FDD base L, MC latch, FDD guard, and floppy disk.
4 ► Unscrew four (4) screws and two (2) connectors. Then remove the unit from the main
base.
To assembly the unit re-install the parts removed during the removal procedure (see above) in
the reverse order (4~1). Position adjustment is not required.
• Motor S assembly
• J641 cord assembly
• Photo Interrupter GP1A34LC
Motor S Assembly
Disassembly
Assembly
► Assemble the motor S assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
Disassembly
Assembly
► Assemble the J641 cord assembly by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
Disassembly
Assembly
► Assemble photo interrupter GP1A43LC by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
1 ► Confirm that the disposal tray assembly (waste box) is not left in the equipment.
Remove if still present.
2 ► Dismount the container station of the gripper unit.
2.1 ► Unscrew four (4) screws.
3 ► Move drive-tray forwards manually.
4 ► Remove photo interrupter GP1A34LC.
4.1 ► Remove the connector of the J641 cord assembly.
Assembly 2
► Assemble photo interrupter GP1A34LC by re-installing the parts removed during the
disassembly procedure (see above) in the reverse order.
4.4.13.1 Overview
The detection unit of cobas e411 has the following major components.
4.4.13.2 Location
The figures below show the principal components and the parts in the detection unit as each
number.
Figure 4.4-46 Detection unit cross-sectional view with slanted PMT housing
Figure 4.4-48 Details of Measuring Cell Mounting (PMT Holder) Support Fixing
1 SUPPORT (P/N:741-1803)
2 CELL FRAME (P/N:741-1832)
3 FLOW CELL ASSY (P/N:741-0461)
The measuring cell has 2 (two) separable components as shown in Figure 4.4-51.
1. Cell body
2. Reference electrode (RE)
As the window tends to become dirty, be sure not to touch it with your bare fingers.
1 PMMA Window
2 Cell Body
3 RE (Reference Electrode)
4 Connector J610
1 ► The measuring cell should be replaced if the measuring cell position relative to the PMT
and magnet is correct and the BSR2 mean value is lower than an acceptable limit
specified by RD/Hitachi
2 ► After six (6) months or 50,000 cycles if TSH values are out of the acceptable range as
specified by RD/Hitachi.
3 ► If there is no mechanical/electrical trouble but the result of the Performance Check
shows abnormal behavior within the warranty term according to the specification
between RD and Hitachi.
Table 4.4-13
Description Specifications
Photosensitive window Head-on type
Photocathode material Multialkali
Spectral response range 300-850 nm
Peak wavelength 420 nm
Anode-to-cathode supply 1250 Vdc
(Max.)
Anode-to-last dinode voltage 250 Vdc
(Max.)
Average anode current 0.1 mA
Storing environmental -80 ~ +50°C
temperature
Cathode sensitivity (Sk) ≥120µA/lm
Anode sensitivity (Sp) ≥200 A/lm
Equivalent dark current 5 nA
Description Specifications
(Sp=200A/lm)
Red/white ratio (R/W) 150 X 10-3
Replacement of PMT
Note: Make sure that the PMT window reaches the rubber cushion at the bottom of the PMT
holder.
Note: Push the PMT socket tail lightly to ensure sufficient fixing f the PMT window, and attach
the fixing frame of the PMT socket by fastening the screw.
15 ► Slide the measuring cell base upwards and then backwards into the detection unit.
16 ► Replace the 2 (two) M3 screws.
17 ► Replace the inlet/outlet tubes and the connector.
18 ► Replace the covers.
19 ► Switch the instrument on.
CAUTION
CAUTION
4 ► Detach the inlet tubing from the sipper probe by unfastening the nipple.
5 ► Detach the outlet tubing from branch block located on the syringe unit by unfastening
the nipple.
6 ► Disconnect wires and cables attached to the outside connector.
7 ► Replace the detection unit by a new one.
8 ► Replace the four (4) screws.
9 ► Replace the inlet/outlet tubs and the connectors.
4.4.13.7 Housing
Disassembly
Reassembly
Casing
The detection unit is equipped with heat insulating casing for temperature control. White polymer
foam sheets are glued to the outer surface of the unit casing. Black polymer sheets are also
applied around the unit cover and the service window to shield from light. The entire casing must
be replaced if the polymer sheets are damaged or come off the casing over a wide area.
Following disassembly/reassembly, of the Detection Unit, one hour is needed for stabilization of
the PMT,
4.4.14.1 Overview
The magnet drive mechanism is used to position the magnet at the detection point of the
measuring cell and to take the magnet away from the cell.
4.4.14.2 Location
The magnet drive mechanism is located in the detection unit below the measuring cell.
See Chapter 4.2.
Rotation angle 90 °
Figure 4.4-54
Table 4.4-14
Reassemble the unit by re-installing the parts disassembled in steps 1 and 2 in the reverse order.
Adjustment of the position is not necessary
Disassembly Procedure
1 ► Remove the main cover assembly by removing six (6) screws from it and then remove
the service cover assembly by removing four (4) screws from it.
2 ► Take out the L-AMP PI assembly together with the shield case LOG by removing the
three (3) connectors and one (1) screw.
3 ► Remove the cell holder, PC cover, and PMT-HV PI assembly in this order. (You need to
remove three (3) screws from the cell holder, one (1) screw from the PC cover, and four
(4) screws and one (1) connector from the PMT-HV PI assembly.)
4 ► Remove the PC base and PC base (2) by unscrewing two (2) screws from each.
5 ► Remove the P600 cord assembly by taking out two (2) M2.6 screws and one (1) cramp.
6 ► Remove the micro cooling unit and then the P601 cord assembly by unscrewing four (4)
screws and two (2) M2.6 screws.
7 ► Remove the four (4) screws (screw a) that secure the main frame assembly to the case
insulation assembly.
8 ► Slide the main frame assembly, support frame, cell frame, and pipe assembly together
with the lead wire of the photo multiplier out of the case insulation assembly. (Unscrew
two (2) M4 screws and two (2) M3 screws.) Be careful not to expose the photomultiplier
to light.
9 ► Remove the magnet drive assembly by unscrewing two (2) screws.
10 ► Remove the J600 cord assembly by removing two (2) locking fixtures, two (2) tie wraps,
and one (1) screw.
Assembly Procedure
Reassemble the unit by re-installing the component removed during the disassembly procedure
in the reverse order.
Note: When reassembling the unit, generously apply Amicon epoxy resin (F440079) to the bottom
(b) of the micro cooling unit. Also, thickly apply Amicon epoxy resin in groove (c) of the main
frame where the pipe assembly is to be mounted.
Note: After removing the magnet drive assembly from the unit, adjust the positions of the magnet
drive and the measuring cell position. Refer to Chapter 5.2.1, Replacement of Measuring
Cell.
Disassembly Procedure
Assembly Procedure
Re-install the parts removed in the disassembly procedure in the reverse order.
Disassembly Procedure
Assembly Procedure
► Re-install the parts removed in the disassembly procedure in the reverse order.
4.4.15.1 Overview
The matrix BCR mechanism is a unit that contains a rotating mechanism that is used for scanning
bar-coded data with a bar-code reader. The bar-code reader scans a bar code attached to a
sample tube, or a reagent rack pack, or scans a bar-code card that carries the information of
these three in combination.
4.4.15.2 Location
• Rotating Unit
• Bar-code Reader
• Bar-code Card Holder
Rotating Unit
The rotating unit rotates and shifts the bar-code reader to the scanning point on either the sample
disk or the reagent disk. This unit conveys the movement of the stepping motor via a flat gear to
the bar-code reader so that the bar-code reader will rotate slowly when scanning a bar-code card.
Bar-code Reader
The bar-code reader is installed onto a molded bracket at a gear fixed to the bracket with a screw.
The gear controls the rotation of the bar-code reader. The bracket is supported by a rotating axle
which is inserted into ball bearings.
The bar-code card holder supports the table cover L and also guides a card into the appropriate
card position.
Figure 4.4-55
1 WIDE BCR
2 741-1321: bracket
Table 4.4-15
Before removing the matrix BCR unit, remove the following covers:
• Rear cover
• Gripper Y cover
• Front cover
• R lid
• Table cover L
• Table cover R
• P plate assembly (at the front bottom of the pipetter unit)
Then, remove three (3) screws, two (2) connectors (P650 and P651), and the bar-code reader
cable connector from the relay board (UI-RS) and remove the unit from the main unit base.
Disassembly Procedure
Assembly Procedure
Assemble the Matrix BCR Unit by re-installing the parts removed disassembly procedure in the
reverse order.
If you have removed the motor, re-adjust the backlash of the gear, referring to the diagram below.
Adjustment of Backlash.
First, make the BCR bracket rotate and shift to the stopper positions at both ends to fix the
stopper for the motor gear rotation. Then, adjust the backlash of the BCR bracket in the direction
of its circumference to 0.1 to 0.2 by shifting the motor position.
Figure 4.4-58
Overview
• The BCR base positions itself. It is not necessary to adjust it mechanically because there is
enough tolerance.
• If you replace the BCR itself, first you just make sure whether it is possible to read a label
according to the Check Procedure.
Check Procedure
Check that the BCR functions well as follows. Refer to attached flow diagrams.
Sample BC Scanning
• Place a Sample Rack with primary tubes to all of which BCR labels are attached.
• Make sure that the instrument is in Stand-by status by checking that Stand-by appears on the
screen.
• Perform Sample BC Scanning by touching the button in the Mechanism check screen. Refer
to Chapter 6.3.
• Confirm whether you hear a beep at each position during scanning.
If everything is OK, the BCR reader is ready for further checks, e.g. Barcode Card Reading and
Reagent Pack BC Scanning. See Comment 1.
If not OK, begin BCR adjustment as described in the Adjustment Procedure.
If everything is OK, the BCR reader is ready for further checks like Reagent Pack BC Scanning.
If not OK, begin BCR adjustment as described in the Adjustment Procedure.
• Place a Reagent Rack with RackPacks to all of which BCR labels are attached.
• Make sure that the instrument is in Stand-by status by checking that Stand-by appears on the
screen.
• Perform Reagent Pack BC Scanning by touching the button in the Mechanism check screen.
Refer to Chapter 6.1.10. See Comment 2.
• Confirm whether you hear a beep at each position during scanning.
Adjustment Procedure
Sample ID
This procedure adjusts the angle of BCR itself relative to slit of the Sample Disk at Sample
position.
The BCR angle moves one (1) pulse by touching the button once in the popup screen (Figure
4.4-59).
In the meantime, find a beep area within ca. ± five (5) pulses of the zero (0) value.
In principle, you do not have to adjust the angle of the Sample Disk rotation, even though the
function is available.
∗ Make sure that the instrument is in Stand-by status by checking that Stand-by appears on the screen.
∗ Touch the Utility button.
∗ Touch the Adjustment Sipper, Cap Open/Close BCR button.
∗ Touch the BCR Sample DIsk button.
3 ► In the meantime, the BCR continuously scans at half (0.5) second intervals. If
everything is OK, the BCR beeps. It is now ready for further checks like those in item4
►. If not OK, the BCR does not beep; adjust the BCR as described in items 5 ► to 6
►.
4 ► Check whether the BCR can be read when you move the Sample Rack by hand by
about 2 mm to the left or right. See Comment 3.
If everything is OK, the BCR beeps. It is ready for check of retry function. Touch Cancel button. In
not everything is OK, the BCR does not beep; repeat from items 5 ► to 6 ►.
Version 1.0 – May 2006 4 - 115 Chapter 4.4
RD/Hitachi cobas e411 Service Manual
5 ► Adjust the BCR angle to find a beep area within ca. ± five (5) pulses of the zero (0)
value by touching the following buttons.
Note: If you touch the BCR Right + button once, the BCR moves one (1) step to the right.
Note: If you touch the BCR Left - button once, the BCR moves one (1) step to the left.
See Comment 4.
6 ► Confirm that the BCR beep is stable. If it is stable beep, touch the OK button. In the
meantime, the BCR returns to the HOME position (in reset status) once and it moves to
an updated position, the BCR scans again.
If everything is OK, the BCR beeps. It is now ready for operation. See Comment 5. In not
everything is OK, the BCR does not beep; repeat from items 5 ► to 6 ►.
1 ► Check the Retry function by setting up a BCR position equivalent to the position of the
second reading as follows.
In case of all: If it is stable beep, touch the Cancel button. Perform the Sample Scan.
In case of all: If it is not OK, touch the Cancel button and repeat from items 5 ► to 6 ► of
Sample ID.
► Perform a Sample Scan by touching the Sample Scan button in the Orders screen.
► Momentarily block the first Sample Scan with paper; remove the paper and check that
the second Sample Scan is performed.
See Comment 7.
See Comment 8
Reagent ID
• The BCR has a sufficiently wide angle of readable area. Therefore, in principle, it is not
necessary to adjust even though an adjustment function is available.
Version 1.0 – May 2006 4 - 116 Chapter 4.4
RD/Hitachi cobas e411 Service Manual
• The adjustment value is to set zero (0) during production. See Comment 9.
Attachment 1
Comment 1
Some customers are using narrow width barcodes (out of specification) and primary tubes
(diameter 13mm) that rotate with rotation of the sample disk. Therefore, please check sample BC
Scanning by using the adjustment procedure for Sample ID.
Comment 2
Before you perform Reagent Pack BC Scanning, check that the window of the Reagent Jacket is
clean. If you find any stains, clean the window.
Comment 3
(4)’ Check whether the BCR can be read when you touch the BCR Right +button three (3) times
and touch the BCR Left - button three (3) times.
Comment 4
(5)’ If you cannot find a beep area within ca. ± five (5) pulses of the zero (0) value, adjust the
sample disk position by touching the following buttons.
Note: If you touch the Sample Disk Right + button once, the sample disk moves one (1) step to the
right (CW).
Note: If you touch the Sample Disk Left - button once, the sample disk moves one (1) step to the
left (CCW).
Comment 5
Comment 6
You need to set up function status of barcode reading on the Data Disk if you want to use the
Retry function.
Comment 7
If everything is OK, the BCR beeps and it is now ready for operation.
If not OK, the BCR does not beep; repeat Sample ID adjustment after checking for stains on the
Barcode label itself.
Comment 8
Barcode Card
• The BCR has a sufficiently wide angle of readable area. In principle, it is not necessary to
adjust.
Comment 9
∗ Make sure that the instrument is in Stand-by status by checking that Stand-by appears on the screen.
∗ Touch the Utility button.
∗ Touch the Adjustment Sipper, Cap Open/Close BCR button.
∗ Touch the BCR Reagent DIsk button.
3 ► In the meantime, the R. Disk rotates continuously and the BCR scans.
If everything is OK, the BCR beeps; it is now ready for further checks like the Retry function for
Reagent ID. Touch the Cancel button.
In not everything is OK, the BCR does not beep; adjust the BCR angle as described in item (4).
4 ► Adjust the BCR angle to find a beep area by touching the following buttons.
If you touch the BCR Right + button once, the BCR moves one (1) step to the right.
If you touch the BCR Left - button once, the BCR moves one (1) step to the left.
If everything is OK, the BCR beeps; it is now ready for further checks like the Retry function for
Reagent ID. Touch the OK button.
If not OK, the BCR does not beep; return to item (4).
• You need to set up function status of barcode reading on the Data Disk if you want to use the
retry function.
• The BCR Retry function is in operation during Reagent BC Scanning.
• If the bar code is not readable at first, the R. Disk rotates again and a second reading is
performed.
► Perform the Reagent Scan by touching the Reagent Scan button in Inventory screen.
► Momentarily block the first Sample Scan with paper; remove the paper and check that
the second Sample Scan is performed.
In not OK, the BCR does not beep; repeat Reagent ID adjustment, after checking for stains on
the window and on the Barcode Label itself.
Attachment 2
Figure 4.4-61
Sample ID Adjustment
Figure 4.4-62
Note: In principle, you do not have to adjust the angle of the S.Disk rotation.
Figure 4.4-63
Reagent ID Adjustment
Figure 4.4-64
4.4.16.1 Overview
The INTER ROCK Mechanism is a function that stops the device by opening the cover when the
operation and resetting it and does safely.
4.4.16.2 Location
• If you open the cover, instrument status changes to STOP by the detector.
• It stops the motor of Sopper X at the same time (by Switch)
Figure 4.4-65
4.5.1 Overview
The following is detailed procedures for adjusting the mechanical positions. The procedures are
described in the order in which you can carry out the adjustments most conveniently.
1 ► Make sure that the mechanical positions of the unit which you have just replaced or adjusted
are well adjusted relative to those units that share the same working areas with the unit. For
instance, when you replace the reagent disk, you should make sure that it is well adjusted
relative to the cap opener, pipetter and beads mixer.
2 ► The belt tension must always be adjusted prior to the position adjustment by means of the
software because you have to take the units out when you adjust their belt tension.
3 ► Refer to Chapter 6.4.1 for general procedures on how to operate the mechanical adjustment
software.
1. The following jigs are used (Figure 4.5-1 and Figure 4.5-2).
2. As shown in Figure 4.5-3, place the JIG-1 in the magazine container. Then, by using the JIG-2,
check visually that the gripper cover stands perpendicularly to the magazine container. If it is not,
perform an adjustment by unfastening the two screws (Screws-A) shown in Figure 4.5-4.
3. As shown in Figure 4.5-5, place the JIG-1 in the magazine container. Then, by using the JIG-2,
check visually that the gripper cover stands perpendicularly to the magazine container. If it is not,
perform an adjustment by unfastening the two screws (Screw-B) shown in Figure 4.5-6, and if
necessary, the two screws (Screws-C) shown in Figure 4.5-7. If you have performed an
adjustment by unfastening the Screws-C, open the inspection hole cover and check if the gears
are engaged smoothly; the Gripper Assy(Figure 4.5-7) must returns to its original height by the
action of the spring when you press it down from above. If the gears are not engaged smoothly,
perform an adjustment by unfastening the screws that secure the motor (Figure 4.5-7).
Figure 4.5-8
Figure 4.5-10
3 ► Prepare the 1 tip, 2 vessels, 3 tip magazines, and 3 vessel magazines. Set them as shown
in the pictures below.
Figure 4.5-11
Figure 4.5-12
4 ► Touch the [START] button and the instrument starts the initialization.
5 ► About 1 minute later, the vessel on CB Pos. will be discarded by the gripper. The moment
the gripper is over the vessel waste hole, open the SS cover in order to stop the gripper
movement right on the spot. Be careful NOT to be late.
6 ► Remove the vessel and set the TABLE COVER ADJUSTMENT TOOL (P/N 741-8813) to
the gripper as shown in the picture below.
Figure 4.5-13
7 ► Push the gripper down toward the vessel waste hole by your finger, and check if the
adjustment tool does NOT come into contact with the inner wall of the Vessel Waste Guide.
Adjust Vessel Waste Guide so that Tool may become the center of Vessel Waste Guide as
much as possible. If it is in contact, adjust the position of TABLE COVER R and Vessel
Waste Guide. Then check it in the same way.
Figure 4.5-14
► Especially, does not come in contact with the TABLE COVER R when you operate Gripper
Y mechanism and P CARRIER ASSY by the hand.
Figure 4.5-15
9 ► After completing the adjustment of TABLE COVER R and Vessel Waste Guide, attach
COVER Z.
10 ► Pepare the 1 tip for buffer station, 33 vessels for incubator (include the Sipping Pos.), 3 tip
magazines, and 3 vessel magazines.
Place a rack with a primary tube (diameter 10 or 13 mm) in rack position one at the STAT position of
rack sampler unit.
Carry out "B line sampling pos." from the screen "Adjustment rack sampler".
If the pipetter needs correction to left/right (x-axis, see Figure 4.5-17), quit the current screen with
"Cancel" key, change to the "Adjustment pipetter R.Disk" screen and carry out "Pipetter-X S.Disk"
from the screen .
Figure 4.5-17
If the pipetter needs correction in up/down direction (y-axis, see Figure 4.5-17), change the rack
position with "Forward" or "Backward" key.
Note: Note: The hook will only move the rack backwards (see Figure 4.5-17). For forward hook
movement, you have to push the rack by hand to get the actual position.
Note: Note: The range of software adjustment for B line sample pos. is only ± 5 steps because of
the design of the racks (Figure 4.5-18).
Figure 4.5-18
If the range of software adjustment for B line sample pos. is not sufficient, change the sample position
manually. To adjust manually, move the whole sample detector B assembly (see Figure 4.5-19) in the
corresponding direction. The detector B assembly is fasten with two screws on the left and one screw
on the right. After changing the position manually, optimize again with the software (as described
above).
Figure 4.5-19
4.6.1 Overview
For details of Photo Sensor and Photo Interrupter indicated by PIxxx, see Chapter 4.3.5.
Figure 4.6-2
4.6.2.1 A Line
This line sets the rack in which the samples (maximum 15) are placed to the tray and conveys it to the
B line.
4.6.2.2 B Line
This line accurately conveys the rack that has the samples that have been received from the A line to
the bar code reading position and the pipetting position. After reading the sample bar code and the
rack bar code and after pipetting, the rack is conveyed to the C line (storage unit) or the D line.
4.6.2.3 C Line
This line unloads the rack that has the samples that have been conveyed by the C line to the tray with
the arm used for rack conveying.
4.6.3 A Line
4.6.3.1 Composition
(Stroke : 441 mm Delay : 2 mm Sensor to Sensor : 437 mm, 2778P)
Figure 4.6-3
2. Tray used
The rack is conveyed when the following requirements are fulfilled on operation:
• When the rack that has the samples is not on the tray and Home Detection is not on, the rack feed
arm returns to the Home Position.
• When the rack that has the samples is on the tray and Home Detection is on, the rack feed arm
leaves the Home Position and then returns to the Home Position.
4.6.2.4 Explanation of Operation
1. Home detection (PI201 detection: dark) interrupts and stops the Home position.
2. Rack feed to the B-Line is interrupted and stopped by the Rack Feed Detection (PI212:dark) or
the Rack Feed End Detection. The rack feed arm returns to the Home Position every time after
the Rack Feed End Detection or the Rack Feed Detection.
3. When a STAT sample (STAT input/output) is received, the STAT sample rack is given priority and
the A line rack feed operation is not executed after moving the B line rack.
4. If Rack Feed End Detection (PI202:dark) enters, it returns to the Home Position and executes the
rack feed operation once more in the next cycle.
4.6.4 B Line
4.6.4.1 Composition
(Stroke:364 mm+153 mm including 2 mm delay)
Figure 4.6-4
Figure 4.6-5
• When the Rack Detection (PI212:dark) or the STAT SW is inputted. However, when there is a
detection at the same time, the Rack Detection(PI212) is given priority.
2. The rack is conveyed when the following requirements are fulfilled during reset:
3. The rack is conveyed to the pipetting position when the following requirements are fulfilled. When
the Rack Full Detection (PI223) for the C line buffer is on and when the rack detection for the
rotation mechanism (PI241, P1244) is on, a series of analyses is continued and after the final
sample pipetting of the rack that has been loaded, and until the above detections are canceled, it
will be in a standby mode in that position.
6. If the STAT SW is pressed when Rack Detection (PI212) is on, the regular rack conveyor rather
than the STAT Rack will be given priority.
4.6.5 C Line
4.6.5.1 Composition
(Stroke : 139 mm Delay : 2 mm Including 2 mm delay)
Figure 4.6-6
2. Tray used
detection (P1223 detection: dark) is OFF. Feed to the tray if the previous unloading operation is
sent to the tray. If the previous unloading operation is sent it to the buffer, then send to the buffer.
• When the Tray Full detection 1 (PI225 detection: dark) is ON, the Tray Full detection 2 (PI226
detection: dark) is ON, the tray detection (PI224 detection: dark) is ON, and the Rack Full
detection (PI223 detection: dark) is OFF. (Feed to buffer: Stroke 27 mm)
• Cannot feed when the Tray Full detection 1 (PI225 detection: dark) is ON, the Tray Full detection
2 (PI226 detection: dark) is OFF, the Tray detection (PI224 detection: dark) is ON, and the Rack
Full detection (PI223 detection: dark) is ON.
Table 4.6-1
Tray
Rack Full Tray Full Tray Full
No. detection Discriminant
Detection 1 PI292 Detection 1 PI294 Detection 2 PI296
PI293
1 ON OFF OFF OFF Can feed (to tray)
2 OFF OFF OFF OFF Can feed (to buffer)
3 ON OFF ON OFF Can feed (to tray or buffer)
4 ON OFF ON ON Can feed (to buffer)
5 ON ON ON OFF Cannot feed
6 OFF ON OFF OFF Cannot feed
7 ON ON ON ON Cannot feed
8 ON ON OFF ON Cannot feed
2. The rack is conveyed when the following requirements are fulfilled at reset:
• When the C line is in the Home Position (PI221 detection: dark), the rack is conveyed in the same
manner as (1).
• When the C line is not in the Home Position , the rack is conveyed in the same manner as (1) after
H.P return is performed.
5.1 Boards
Note: RS CONTD (P/N 741-5074) mounted with EIO3 PCB for Rack Sampler System.
Note: DIST232C (P/N 741-5075) mounted with EIO3 PCB for Disk System.
Table 5.1-2
PCB Name Part No. Location Primary Function
UIRS-C 741-5081 Left/rear side Connection of RS232C signal for 2-set of BCR, HOST.
of the instrument
MVLB110 271-3907 Under the PCB rack Mother board for PCB rack
S.AB-CE 725-5029 In the AC power unit Surge absorber circuit
DET-A 741-5016 In the gripper Photo interrupter equipped (gripper Y-axis)
DET-B 741-5017 In the gripper Photo interrupter equipped (gripper Y-axis)
DET-C 741-5026 In the solid waste box Photo interrupter equipped (solid waste boxpresence/absence)
DETECT1 736-0824 In the SYRINGE Photo interrupter equipped (syringe)
RS CONTD (for 741-5074 PCB rack(On the EIO 1. Motor control function
R/S) Pcb) 2. Conversion from parallel signals to serial signals
PS CONTA (for 741-5058 In the R/S (Under the Conversion from serial signals to parallel signals
R/S) DO4A Pcb)
DO4A (for R/S) 741-5071 In the R/S 1. Pulse motor driver for 2 or 1-2 phase control
2. Photo interrupter buffer
BCR (for R/S) 741-5056 In the B line Connection of RS232C signal for BCR
PH-D (for R/S) 741-5053 In the B line Photo transistor equipped
PH-T (for R/S) 741-5054 In the B line Photo diode equipped
Note: Part No. and PCB name are different depending on the manufacturing time. Please refer to
Parts List for details.
5.1.2.1 Analyzer
Figure 5.1-2
Figure 5.1-3
Figure 5.1-4
Figure 5.1-5
Figure 5.1-6
Figure 5.1-7
Figure 5.1-8
Figure 5.1-9
Note: PS CONTA board (P/N741-5058) is located just behind the DO4A board(P/N741-5071) as
shown in the picture above.
Note: BCR Board (P/N741-5056) and BCR are located at the BCR reading position on the B line in
the analyzer.
Figure 5.1-10
Note: PH-D Board (P/N741-5053) should be mounted with PH-T Board (P/N741-5054) as a pair at
the sampling position in the B-line.
5.2.1 AC Power
• AC lines are only connected to the DC power supply through main switch and noise filter.
• The surge absorber provides protection from electrical surges of more than 600V and is
connected between the main switch and the noise filter.
Dimensions:
Dimensions:
Dimensions:
• To adjust or inspect the voltage turn the variable resistors (shown in the figure below) on the
terminal side of the power supply unit without an electric load. (DC output has sensing line and
voltage is maintained even if the electric load is changed.)
Figure 5.3-6
Figure 5.3-7 How to Remove the Cover from the Pipetter Head
Precondition: Ensure that the pipetter tube is primed
Use a plastic screwdriver to adjust the variable
capacitor on the LLD-P3 PCB
as shown in Figure 5.3-12.
When monitoring the voltage, keep your hands
off the pipetter arm.
1 ► After “Reset”, select “Voltage Monitor” and execute “Pipetter LLD”.
2 ► Adjust the voltage to the minimum value while the pipetter head is at the highest position.
► The voltage without tip has to be below 1.5V (see Figure 5.3-9 for testing position) and the
principle LLD curve (check pos. 1) should be in the recommended range (see also Figure
5.3-8).
► The voltage with tip should be increased by 0.1 V to 1.5 V (check pos. 2, see Figure 5.3-8).
If the voltage decreases, re-adjust the trimmer so that the voltage increases with a tip.
► Make sure that the monitoring voltage increases between 0.3 V and 0.7 V (see Figure 5.3-8).
Explanation:
ο Tip attached
3. Pipettor in pos. A
ο Tip attached
ο Tip attached
Figure 5.3-9
Figure 5.3-10
Figure 5.3-11
• If the voltage is OK, touch the Monitoring Stop button in the Voltage Monitor screen.
Check the crash sensor signal output by measuring the voltage between the connector pin#A3 of J552
and TP2 (AG) on DIST-PA Board as follows.
Note: Keep the current height of the pipetter head. This means that only pipetter probe itself
should go up and down during the next measurements.
• Make sure that the voltage is 0.5 V or lower (ca.35 mV) when the thickness tool (0.8), which is
0.8 mm in thickness, is inserted into the gap between the pipetter probe and the flat place as
shown in Figure 5.3-7.
• Also make sure that the voltage increases up to ca.4.5 V or more (ca.5V) when the tool is
replaced by the 1.2mm thickness tool under the same conditions.
• If the increase is not sufficient, bend the detection plate on the pipetter probe upwards or
downwards with pliers (see Figure 5.3-13).
• Repeat the above procedure until the specification is met.
• Replace the pipetter head cover as before.
• Check the pipetter probe position relative to Tip pos.1 on buffer station, washing station, reagent
disk, sample disk and vessel buffer. All positions should be at the center of the hole in each.
• In particular, if the Y-direction pos. of the pipetter probe is not sufficient on the buffer station when
viewed from the side, adjust the pipetter probe Pos. at the center of Tip pos.1 and the vessel
buffer by unfastening and fastening two (2) screws shown by the arrows in Figure 5.3-14.
Insulation Resistance between the Aluminum Block and the Shield Plate
Note: The resistance cannot be measured if connector J662 is connected. Also, damaged wiring
and damaged FPC will cause an abnormality of the sipper LLD function.
• Touch the Pipetter Prime button in the Maintenance screen to start pipetter priming. See Chapter
6.
• Press the Stop key when the Pipetter System has filled with water.
• Concerning the Pressure Sensor Voltage Monitoring, See Chapter 6.
• Leave the instrument for 2 minutes or longer.
• Touch the pressure sensor button in the voltage monitor screen.
• Confirm the pressure sensor voltage to 1.8 ± 1.3 V .
• If the voltage is OK, touch the Monitoring Stop button in the Voltage Monitor screen.
1 Sensor and ANG-EP PCB P-Sensor 2 Assy
ANG-EP2 PCB
2 Adjustment / Check the voltage of 1.8 ± 1.3 V
the pressure sensor After pipetter prime, leave for 2 min. or more.
( No adjustment available )
Pipetter
Abnormality of the liquid level detection for the pipetter may be caused by one of the following
problems.
Use the following procedures for troubleshooting.
• The insulation resistance of the probe is low.
ο Detach the two (2) wires from the probe.
ο Measure the insulation resistance with a voltage multimeter or any similar device.
ο If the resistance is normal, the resistance goes up beyond the measurable limit of the
multimeter.
ο If the resistance is low, it will be shown by a value below 1,000 kΩ. In this case replace the
pipetter probe.
• The electrical conductivity of the water inside the pipetter tube is too high.
ο Measure the output voltage of the LLD-P3 board twice as follows.
ο If there is a difference in the output voltage between the former and the latter, the conductivity
of the water could be increased.
ο With the tube connected, move the pipetter horizontally and check the voltage. If the output
voltage varies depending on where the pipetter is located on the horizontal axis, the
conductivity of the water could be increased.
ο If high conductivity is detected, replace the water in the water supply tank by water with
conductivity meeting the specification. Repeat the pipetter prime to clean the flow path.
• The spring of the pipetter nozzle holder is mispositioned.
ο If the spring of the pipetter nozzle holder, which is installed in the upper cover of the pipetter
case, is mispositioned, the inner pipe (detection electrode) and the outer pipe (shield) of the
pipetter become mutually conductive. In such a case, the voltage output from the LLD-P3
board goes up by ca. 10 V.
ο If the output voltage is normal when the cover is open, but it is not when the cover is closed,
mispositioning of the spring may be the cause. In this case, return the spring to the normal
position.
• The board is damaged due to liquid leakage.
• The liquid level is too low or there is no liquid in the container.
• Damaged wiring and damaged FPC will cause abnormal liquid level detection with the pipetter.
Sipper
Abnormality of the liquid level detector for the sipper may be caused by one of the following problems.
Figure 5.3-16
NOTICE
Do not touch Shutdown unless there is a particular
reason to shutdown the analyzer. Otherwise all the
analyzer functions and components will be shutdown.
In case you have mistakenly touched the button, turn
off the circuit breaker once, and then turn it on to revive
the analyzer.
NOTICE
It is highly recommended that you enable the
screensaver. To enable, select Utility -> System and in
the Screen Saver area, touch the On button and set the
activation time by typing in a value within the range of
1-60. For details, refer to Utility menu, Screen Saver
area.
(a) En = Tn - Ttrg
• The voltage of the incubator is limited from 0 V to 5 V, the voltages of other unit are limited within
±5 V.
• The temperature of the reagent disk and the detection unit is influenced by the ambient
temperature around the units. So the target temperatures (Ttrg) for the two units are calculated
according to the formula below. The Ttrg of the R.Disk refers to the temperature around the
reagent disk, and the Ttrg of the detection unit refers to the temperature around the system
reagent unit.
Note: You can see the temperatures of the control points (Tn), the ambient temperatures (Tamb),
the target temperatures (Ttrg) and the control parameters (Vout) on the Temperature
Monitor Screen.
Note: The Peltier works as a heater when the control parameter is positive and as a cooler when
the control parameter is negative.
Note: The control parameters (Vout) are multiplied by +1 at the EIO3 board and by -2.6 at the DO3
board. Voltages are limited within ± 12 V on the DO3 board.
Note: The control parameter for the detection unit turns over when it comes to ANGCNT.On the
Temperature Monitor Screen you can see the voltage which does not turn over. It turns over
by the cable (P/N 741-4237) between the DO3 board and the detection unit
5.3.6.2 Troubleshooting
• If temperature below 4.6 degrees C is displayed
1. If temperature control is being normally operated, the temperature displayed in the screen should
be 13 to 15 degree C at the lowest.
2. Therefore, if the screen displays a temperature below 4.6 degrees C, this may be due to the
thermistor wire or the cable between the EECL300 board and the thermistor breaking.
3. What is displayed is identical to that which is displayed if the thermistor resistance were infinite.
1. If temperature control is normal, the thermostat breaks the input for peltier or heater at around 60
degrees C.
2. Therefore if the temperature is displayed higher than 80.0 degrees C, this may be due to the
thermistor or the cable between the EECL300 board and the thermistor being shortcircuited.
3. This phenomenon is identical to that which takes place if the resistance of the thermistor were
zero (0) ohm.
1. Check whether covers the are closed or not. For example, temperature control of the R. Disk does
not work well without the R. Disk cover.
2. Check that there was not a steep temperature gradient. The specification of the temperature
gradient is ± 2 degrees C /h. Direct wind from air conditioners, etc., must be avoided.
3. Check the relation between the indication temperature and the ambient temperature of the
thermistor.
1) Check the peltier or the heater if the two temperatures are nearly the same.
2) Check the thermistor if the two temperatures are significantly different.
Because of the thermometer accuracy might be within ± 0.5 degrees C, only a rough comparison is
possible.
1. Check the resistance between the wires of the peltier or heater to detect any abnormalities in
the element.
2. Measure the resistance at Table 5.3.6-2 shows the check pins of the peltier or heater.
1. Check the resistance between the wirings of the thermistor to determine whether or not the
thermistor is normal.
2. Thermistor errors are usually caused by moisture.
3. With the usual ambient humidity, the thermistor never absorbs moisture. However, if the
thermistor gets wet as a result of condensation, the resistance of the thermistor goes down
significantly. Therefore, there are no circumstances in which the thermistor only slightly defective,
and so the error is quick to find.
4. Table 5.3-4 shows the check pins of the thermistor.
5. Table 5.3-5 Table 5.3.6-4 shows the outline of the resistances of thermistor.
6. Details are shown in the data sheet for the thermistor.
Temperature
10 15 20 25 30 35 40 45 50
(degrees C)
Resistance (kΩ) 9.9 7.8 6.2 5.0 4.0 3.2 2.6 2.1 1.8
5 6 5 0 3 7 6 8 0
Block Diagram
Figure 5.3-18
Latch the parallel data, and with the shift register convert to parallel → serial.
Figure 5.3-19
Serial-Parallel Conversion
With the shift register convert the serial data from serial → parallel, and latch the parallel data.
Figure 5.3-20
1 ► Place a reagent pack in positions 1 to n. Start the Bead Mixer Check ( Mechanism Check )
and enter < n > repititions.
2 ► Adujst the speed by the VR ( CW for encreasing the speed )
Table 5.3-6
In air In liquid
2050 rpm ± 100 rpm 2000 rpm ± 100 rpm
The figure below shows the location of the resistor to be adjusted on the EIO3 PCB (see also Chapter
5.4.4).
Figure 5.3-21
Block Diagram
Figure 5.4-1
Note: This block Diagram is Disk System
Note: Note for EIO3 PCB
∗ Disk System : EIO3 + DIST232C PCB
∗ Rack System : EIO3 + RSCONTD PCB
5.4.1.2 Location
RS-232C Controller
This controller communicates with the following units via UIRS-C PCB.
• Host Computer to receive the Test Order and send back the Measured Results.
• Barcode Reader to Read Rack Pack information and Sample ID.
Battery
• The specifications of the battery (Electric double layer capacitor)are as follows.
Table 5.4-1
No. Description Specification
1 Data holding time during Power off 7 days (with full charge)
2 Recharging time from empty status 24 hours
3 Life time 7 years
• Power Supply of SRAM on the board is provided by a re-chargeable battery during Main Switch off.
LED Description
Table 5.4-2
Silk-print
display During After Reset is
Name Color Meaning Remarks
Reset released
RUN RUN Green Lights-on : Reset is released. Also, Lights-out Lights-on/Lights-out
Double Bus Error is not generated.
Lights-out : During Reset, or, Double
Bus Error was generated.
FAIL FAIL Red Lights-on : Due to Double Bus Error, Lights-out Light is out until the
MPU stopped. new phenomenon
Or, Watchdog Timeup was generatead. (Due to Double Bus
Lights-out : Double Bus Error is not Eror, CPU stopped.
generated. And Watchdog Timeup is Or, Watchdog Timeup
not genrated. was generated.)
occurs.
DMA DMA Green Lights-on : DMA is being transffered. Lights-out Lights-on/Lights-out
Lights-out : DMA is not being
transffered.
INT INT Green Lights-on : The order is changeable Lights-out Lights-on/Lights-out
depending on the priority.
Lights-out : The order is not changeable
depending on the priority.
MBS MBSY Green Lights-on : Memory is accessed. Lights-out Lights-on/Lights-out
Lights-out : Memory is not accessed.
MER MERR Red Lights-on : 2 bit ECC Error was Lights-out Light is out until the
generated. new Phenomenon (2
Lights-out : 2 bit ECC Error is not bit ECC Erros was
generated. generated in Main
Memory) occurs.
DIG DIAG Red Lights-on (For approx. one Second) : Lights-out Lights-on/Lights-out
Self-diagnosing. Lights-on (Every 0.5
second) : Self-diagnose error was
generated.
Consecutive lights-on : Boot error was
genrated.
Lights-out : Not self-diagnosing.
And Self-diagnose error is not
generated.
And Boot error is not generated.
∗ For the error code, LED
registor D31
Figure 5.4-3
Figure 5.4-4
Block Diagram
Figure 5.4-5
5.4.2.2 Location
Potentiostat
• The ANGCNT sends the potentiostat voltage twice as high as the order according to the timing
chart for potentiostat which comes from the CPU.
Data Collection
• The luminescence data are converted to logarithmic values in the L-AMP(3) circuit board, then
ANGCNTcarries out analog-to-digital conversion and transmits the data to the CPU.
• GMCNT converts the current running through the cell and voltage between CE and WE from
analog to digital and transmits the data to the CPU.
Temperature Control
• The ANGCNT determines the output voltage to the Peltier and the heater according to the
temperature measured by the thermistor thermometers, and it controls ON/OFF of the fans.
Clot Detection
• LLDCNT takes in the pressure sensor signal from ANG-EP circuit board and transmits the data to
CPU.
• CPU decides whether or not a clot is detected.
LED Description
Table 5.4-3
LED Supplied
Purpose Remark
No. Voltage
RUN 5V CPU running monitor is lit when CPU is
running, other than in HALT status.
INT 5V MPU interruption
CRT 5V
ADC 5V DC bus busy
Figure 5.4-7
Block Diagram
Figure 5.4-8
5.4.3.2 Location
Motor Control
• GPCNT generates pulse signal, and sends it to DO1 or DO2 circuit board when it receives the
order of drive pattern of motor and pulses from the CPU.
• GPCNT stops a motor when it finishes the pulses ordered from the CPU or it receives the signal
from photo interrupter or the signal of LLD.
DI/DO control
DI function
• The status information from EIO3 circuit board about the power supply watching, fuse detection, tip
detection, R. Disk cover detection, water level watching in the distilled water tank, etc. are written
on GMCNT.
• CPU scans DI at regular intervals, and gets these statuses.
DO function
• CPU writes the status of switching of micro step and static current drive and ON/OFF of SV, beads
mixer, HV, potentiostat and +24V on GMCNT.
• Each driver on EIO3 circuit board or other boards drives them.
Figure 5.4-10
Figure 5.4-11
Block Diagram
Figure 5.4-12
5.4.4.2 Location
A circuit on the EIO3 board supplies the power to a DC motor for beads mixing.
7. Power supply for the following printed circuit boards and electronic modules.
• ±12V for LLD - P3 board via DIST - PA board
• ±12V for LLD - SA board
• ±15V for ANG – EP2 board
• ±12V for L - AMP (3) board via ANG – EP2 board
8. Watching the power supply
A circuit on the EIO3 board compares the supplied voltage (V) with the reference voltage (Vref) ,and it
puts LOW level signal out to the EMOT300 board if the supplied voltage (V) is lower than the reference
voltage (Vref).
9. Generating the signal for SYSTEM RESET
• A circuit on the EIO3 board generates the VACFAIL signal according to the ACFAIL signal from DC
power supply and 5V power watch IC (IC17 : MAX791), and it transmits SYSRESET signal to the
ECPU board.
• Following switching on of the instrument, the main CPU is informed by VACFAIL signal,and the
main CPU is reset by SYSRESET signal. When the instrument is switched off, VACFAIL signal cuts
in on main CPU just before DC power supply is off.
Fuses/LEDs Description
Table 5.4-4
Rated Rated Supplied
Fuse No. LED No. Purpose
Current voltage voltage
F1 5.0 A 48 V LED1 5V Power supply for the wave shaping
circuit
LED2 5 V(1) Power supply for photocoupler on SV
driving circuit
F2 Ditto ditto LED3 12 V(1) For driving HV, potentiostat, beads
mixer, +24V ON/OFF
LED4 5 V(2) For controlling HV, potentiostat, beads
mixer, +24V ON/OFF
F3 Ditto ditto LED5 24 V(1) For driving SV5 - SV8
F4 Ditto ditto LED6 24 V(2) For driving SV1 - SV4 and pump
LED7 5 V(3) For controlling SV1 - SV8 and pump
F5 Ditto ditto LED8 +15 V(1) Power supply for LLD-P3, LLD-SA and
L-AMP(3)
F6 Ditto ditto LED9 -15 V(1) Power supply for ANG-EP2 and
EECL300
Figure 5.4-18
Block Diagram
Figure 5.4-19
5.4.5.2 Location
Fuses/LEDs Description
Table 5.4-7
Rated Rated LED Supplied
Fuse No. Purpose
current voltage No. voltage
F1 3.2 A 48 V LED1 5 V(1) DC power after photo coupler (PCP)
LED2 24 V(1) Driving power for S.Disk, and Cap Open/Close
(Forward/Backward)
F2 Ditto Ditto LED3 24 V(2) Driving power for Cap Open/Close (Open/Close) and Beads
Mixer Rotation.
F3 Ditto Ditto LED4 24 V(3) Driving power for Cap Open/Close
(Up/Down), S.WASTE
F4 Ditto Ditto LED5 24 V(4) Driving power for Pipetter X axis / Z axis
F5 Ditto Ditto LED6 24 V(5) Driving power for Sipper X axis, and Gripper X axis
F6 Ditto Ditto LED7 24 V(6) Driving power for Gripper Z axis / Y axis
F7 Ditto Ditto LED8 24 V(7) Spare × 2
F8 Ditto Ditto LED9 24 V(8) Pipetter Syringe & Spare
Figure 5.4-21
Block Diagram
Figure 5.4-22
5.4.6.2 Location
See Chapter 5.1.2.
Fuses/LEDs Description
Table 5.4-9
Rated Rated
Fuse No. LED No. Voltage Purpose
current voltage
F1 3.2 A 48 V LED1 5 V1 DC power for after photo coupler
LED2 24 V1 R.Disk rotation ,BCR rotation
F2 Ditto Ditto LED3 24 V2 Sipper Up/Down, Magnet Drive
F3 Ditto Ditto LED4 24 V3 Syringe (Sipper) Drive
Figure 5.4-23
Block Diagram
Figure 5.4-24
5.4.7.2 Location
See Chapter 5.1.2
Fuses/LEDs Description
Table 5.4-10
Fuse Rated Rated
LED No. Voltage Purpose
No. current voltage
F1 5.0 V 48 V LED1 5 V1 DC power for photocoupler
LED2 12 V Fan motor
F2 Ditto Ditto LED3 +15 V1 Peltier for R. disk
F3 Ditto Ditto LED4 +15 V2 Ditto
F4 Ditto Ditto LED5 +15 V3 Peltier for measuring unit
F5 Ditto Ditto LED6 +15 V4 Peltier for system reagent unit
F6 Ditto Ditto LED7 +15 V5 Heater for incubator
F7 Ditto Ditto LED8 -15 V1 Peltier for R. disk
F8 Ditto Ditto LED9 -15 V2 Ditto
F9 Ditto Ditto LED10 -15 V3 Peltier for measuring unit
F10 Ditto Ditto LED11 -15 V4 Peltier for system reagent unit
F11 Ditto Ditto LED12 -15 V5 Heater for incubator
Figure 5.4-25
Block Diagram
Figure 5.4-26
5.4.8.2 Location
See Chapter 5.1.2
Figure 5.4-27
Pipetter LLD
Check the voltage between TP1(LLD Sig.) and TP2(A.G) with a voltage multimeter.
Block Diagram
Figure 5.4-28
5.4.9.2 Location
See Chapter 5.1.2
Figure 5.4-29
Block Diagram
Figure 5.4-30
5.4.10.2 Location
See Chapter 5.1.2.
Figure 5.4-31
Block Diagram
Figure 5.4-32
5.4.11.2 Location
See Chapter 5.1.2.
Fuses/LEDs Description
Table 5.4-11
Rated Rated
Fuse No. LED No. Corresponding power Purpose
current voltage
F1 3.2 A 48 V LED1 +5 V(1) For ECPU board
F2 5.0 A ditto LED2 +5 V(2) For EECL board
F3 5.0 A ditto LED3 +5 V(3) For EMOT300 board
F4 5.0 A ditto LED4 +5 V(4) For EIO3 board
F5 3.2 A ditto LED5 +5 V(5) For DO1 board
- - - LED6 +12 V +12 V for MVLB board
- - - LED7 +24 V +24 V for MVLB board
Figure 5.4-33
Note: Fuse F4 should be mounted with a rated current of 5.0 A.
Figure 5.4-34
5.4.12.2 Location
See Chapter 5.1.2.
Figure 5.4-35
Fuse Description
Table 5.4-13
Fuse Rated Rated Supplied
Purpose
No. current voltage voltage
F1 0.5 A 48 V 5V +5 VDC Fuse for EIO3 board power down monitor
F2 1.0 A ditto ditto +5 VDC Fuse for DIST232C or RSCONTD board
F3 5.0A ditto ditto +12VDC Fuse for DC FAN unit (DC Power supply 1)
F4 5.0A ditto ditto +12VDC Fuse for PC unit (DC Power supply 2)
Block Diagram
Figure 5.4-36
5.4.13.2 Location
See Chapter 5.1.2.
Figure 5.4-37
Potentiostat
• This circuit has the function of applying constant voltage between RE-WE of the measuring cell by
regulating the current. The voltage is switched by the software. With relay the voltage will be cut off
when the software is not activated.
• Also, there is a function for sending voltage between CE and WE and current for the measuring cell
to the EECL board.
Block Diagram
Figure 5.4-38
Vref : order voltage of potentiostat voltage × (-2)
Vol : voltage between CE and WE
Cur : current for Cell
+12(3) : DC driving power for relay
Block Diagram
Figure 5.4-39
Block Diagram
Figure 5.4-40
+12 V(1) : HV input voltage
+12 V(2) : DC driving power for PMT-SHV PCB.
Block Diagram
Figure 5.4-41
Fuses/LEDs Description
Table 5.4-14
Rated Rated Corresponding
Fuse No. LED No. Purpose
current voltage power
F1 0.5 A 48 V LED4 15 V LOG-AMP board
LED5 -15 V LOG-AMP board
F2 0.5 A ditto LED3 12 V PMT-SHV board
F3 1.0 A ditto LED1 15 V +15V in PCB for potentiostat
circuit
LED2 -15 V -15V in PCB for potentiostat
circuit
Figure 5.4-42
Block Diagram
Figure 5.4-43
+12 V(1): input voltage for HV
+12 V(2): driving power for PMT-SHV PCB
LED Description
Table 5.4-15
LED Supplied
Purpose Remarks
No. Voltage
Figure 5.4-44
Block Diagram
See the figure below.
5.4.16.2 Location
See Chapter 5.1.2.
Figure 5.4-45
Figure 5.4-46
CAUTION
Be careful not to touch the variable resistors on the
board.
• L-AMP(3) board is preadjusted in production.
Any changes made in the field by the service
engineer will have a significant effect on
measured data. The warranty is no longer
valid in this case
Block Diagram
Figure 5.4-47
5.4.17.2 Location
See Chapter 5.1.2.
Figure 5.4-48
Block Diagram
Figure 5.4-49
5.4.18.2 Location
See Chapter 5.1.2.
Figure 5.4-50
Block Diagram
Figure 5.4-51
5.4.19.2 Location
See Chapter 5.1.2.
Figure 5.4-52
Block Diagram
Figure 5.4-53
5.4.20.2 Location
This board is located under the PCB rack.
Figure 5.4-54
Block Diagram
Figure 5.4-55
5.4.21.2 Location
See Chapter 5.1.2.
Figure 5.4-56
Figure 5.4-57
Block Diagram
Figure 5.4-58
5.4.22.2 Location
See Chapter 5.1.2.
Figure 5.4-59
Block Diagram
Figure 5.4-60
5.4.23.2 Location
See Chapter 5.1.2.
Figure 5.4-61
5.4.24.2 Location
See Chapter 5.1.2.
Figure 5.4-62
5.4.25.2 Location
See Chapter 5.1.2.
Block Diagram
Figure 5.4-63
5.4.26.2 Location
See Chapter 5.1.2.
Motor Controller
When a drive pattern of the motor and the command for the number of pulses are received from the
CPU, the GPCNT will output a magnetization signal and a pulse and transmits to the PS CONTA board.
The GPCNT sends the number of pulses that were ordered or stops the motor if it receives a
photo-interrupter signal.
Location of DIP SW
Figure 5.4-64
Figure 5.4-65
Table 5.4-17
bit 1 bit 2 bit 3 bit 4 RS CONTD Board address
ON OFF OFF ON $D06000 to $D06FFF
ON OFF OFF OFF $D07000 to $D07FFF
DSW OFF ON ON ON $D08000 to $D08FFF (Initialize)
1 OFF ON ON OFF $D09000 to $D09FFF
OFF ON OFF ON $D0A000 to $D0AFFF
OFF ON OFF OFF $D0B000 to $D0BFFF
others not used
Figure 5.4-66
Block Diagram
Figure 5.4-67
5.4.27.2 Location
See Chapter 5.1.2.
Figure 5.4-68
Block Diagram
Figure 5.4-69
5.4.28.2 Location
See Chapter 5.1.2.
Fuse/LED Description
Table 5.4-19
Fuse Rated Rated LED Supplied
Purpose Remarks
No. current voltage No. voltage
F1 3.2 A 48 V LED6 5V Logic drive use power supply
within the board
F2 ditto ditto LED1 24 V1 Power supply for the stepping CH-1,2
motor drive (1)
F3 ditto ditto LED2 24 V2 Power supply for the stepping CH-3
motor drive (2)
LED7 5 V1 Power supply for the latter part of
the photo-coupler
DIP SW Description
Table 5.4-20
JP number Jumper Pin Arrangement
JP1
Figure 5.4-72
Block Diagram
Figure 5.4-73
Note: This block Diagram is Rack System
5.4.29.2 Location
See Chapter 5.1.2.
Figure 5.4-74
Figure 5.4-75
Table 5.6-1
No
PCP Name Appearance Position in e411
.
1. GP1A34LC PCP with a connector All the other positions described in the rows below
2. OJ-HT04 Smallish PCP with a circuit board Home position for syringes
3. OJ-HT02 Largish PCP with a circuit board Solid waste container present/absent detection
4. TLP830 Small PCP On FPC for pipetter/gripper home (HP), end (EP)
and rest(RP) position for gripper Y movement
5. TWIN-DETECT PCP attached to a plastic base On buffer station for tip present/absent detection
PLI-523
6. PM2-LF10HH Reflection PCP Rack tray and rack detection for A line, B line, C line
(for R/S)
7. PH-D and PH-T Through PCP Sampling position for B line unit
boards
(for R/S)
Figure 5.6-1
Table 5.6-2
N
PCP
o How to check the voltage and the current Light path Voltage
Name
.
1 GP1A34 Close: Reflection is interrupted ca. 5 V
. LC • Check the voltage at Pin#3(Sig.) against Pin#2(GND).
Open: Reflection is ca. 0 V
uninterrupted
• Your voltage multimeter should show the right voltage
• Check the voltage at the other two (2) Pins against Pin#1
or Pin#3 (5V) to find other Pin location.
Table 5.6-3
N
PCP
o How to check the voltage and the current Light path Voltage
Name
.
6 PM2- Close: Light path is interrupted ca. 0 V
. LF10 • Check the voltage at Pin#2(Sig.) against Pin#3(GND).
Open: Light shines across the ca. 5 V
gap
• Your voltage multimeter should show the right voltage if
the PCP is functioning properly.
6.1 Overview
The service software is part of the system software that which provides you with tools with which you
can carry out programs for the instrument and measuring cell maintenance, check and adjust
mechanical units, monitor temperature, unit temperature and LLD voltage, etc. All functions with which
service engineers should be acquainted are described in the following chapter. Screens that are not
described in Chapter 6 are discussed in detail in the Software Guide of the Operator’s Manual.
If you started e 411 without the password protecting, the maintenance screen is as follows.
Figure 6.1-2 Utility-Maintenance Screen without the password protecting: Disk System
You cannot select “Service” item in this screen (only Maintenance and Check).
You should start the device in the service mode before service begins. It is according to the following
procedures.
1 ► Start e 411.
2 ► Push Operator ID button in system screen. Then following screen is shown.
3 ► Input the following information and push OK button, then following screen is shown.
5 ► Input the following information and push Logon button, then Overview screen is shown.
6 ► Open Maintenance screen. You can select “Service” item in Maintenance screen.
Figure 6.1-6 Utility-Maintenance Screen with the password protecting: Disk System
• Utility Screen – Data Selection “All” or “Data Range” (Input the Start and End) – Press the “Print”
button
Protocol Type
Elecsys / cobas : When selecting [Elecsys], it communicates with HOST by Elecsys type. When
selecting [cobas], it communicates with HOST by cobas type. You can select the baud rate only when
communication is OFF.
Yes/No : When [YES] is selected, result data of a sample is sent to HOST in real time as far as all
such data is collected.
Communication Trace
Yes/No : When [YES] is selected, communication detail with HOST is traced in e411
Send Comment
Yes/No : When [Yes] is selected, enabled to register test selection information from HOST as well as
patients comment. It also enables to transfer to HOST the patient comment registered to the analyzer
along with result data. If not selected, the patient comment is not transferred.
QC Additional Information
ControlName/LotNo. : When [ControlName] is selected, name of control is sent as sample ID. When
[LotNo.] is selected, lot No. is sent as sample ID.
Yes/No : Only result data is sent. Inquiry for order is not conducted. This function is not applied to
Batch transfer.
You can set Test Code used by the communication(Host Code) besides it used instrument.Select Test
Code in list, input Host Code, then push OK button.
Serial No.
Install Date
Software Version
Check Sum
The total cumulative number of tests that have been performed on the analyzer. There is no user
access to this field. This number is not reset by reloading software.
This field tracks the number of assays performed on the measuring cell currently on the analyzer. This
number must be reset to “0” if a new measuring cell is installed.
4 ► Push “Backup”.
Read
No. Button name Data Note
/Write
2 Read Reference Data Read Reference Reads refer.dat in which test item information etc.
data are defined and param.dat in which various
parameters(Common parameter that doesn't
depend on per-device) are defined.
3 Read Adjustment Data Read Adjustment Reads a peculiar adjustment value (adjust.dat) to
file the e 411 and parameter (analysis.dat) necessary
for the measurement.
When the reading processing is executed, it checks
Serial No. in data FD and Serial No. in the device. If
the check is OK, it executes reading. If NG, It
outputs the alarm.
4 Read Rack Adjustment Data Read Rack Reads only the adjustment value of line part peculiar
Adjustment adjust.dat to the e 411. No check the Serial No.
file When you ship only the “line part”, and the “line
part” is connected with the analysis part in the
locale, uses this function.
5 Back up Data Write All When the file for APC is read, the file for APC is
written. When is not read, the file for APC is not
written.
Back up Write(Internal)
Data files will be stored in d:\e411\backup\data_YYYYMMDD Max two folder. The older one will be
overwritten first. Files except APC files like SETUP00.dat, … , SEUP08.dat will be stored.
Figure 6.1-25
Figure 6.1-26
Read (Internal)
Figure 6.1-27
Data files in d:\e411\data will be overwritten with the files in the selected folder
Read (External)
Data files in d:\e411\data will be overwritten with the files in the selected folder in the e:\
Table 6.1-3
Button Mean
Printout When the measurement result outputs it, the result is printed.
Upload When the measurement result outputs it, the result is uploading to HOST.
Printout/Upload When the measurement result outputs it, the result is printed and uploading to HOST.
Table 6.1-4
Button Mean
Connected Can print.
Not connected Even if the printer is connected, it is not printed.
Table 6.1-5
Button Mean
Calibration Sample The calibration result is automatically output.
Control Sample The control result is automatically output.
Stat Sample The Stat result is automatically output.
Routine Sample The routine result is automatically output.
Keep Sample Type: You can decide to keep sample type or not.
Default Sample Cup: You can decide keep the default sample cup (Normal or Reduced).
When Sample Premature LLD: When Sample Premature LLD occurs, you can decide to retry
aliquot(On) or not(Off).
When Sample short: When Sample short occurs, you can decide to retry aliquot(On) or not(Off).
Off: When the e 411 ends the analysis, it will change to Stand-by in 30 minutes.
1 – 12 Hours: When the e 411 ends the analysis, it will change to Stand-by in 1–12 hours.
6.2 Maintenance
There are 10(Disk for 8 items) maintenance functions available. Each of them corresponds to a button
in the Check screen. Touch the Check button in the Maintenance screen in order to display the
Maintenance items. After touching a button in the Maintenance Items, a pop-up window appears in the
screen.
The general functions of each button and the input area are as follows.
Start button : To start the maintenance function corresponding to the button
Close button : To close the pop-up window and return to the Maintenance screen without carrying
out any maintenance functions
Input area : To define how many times the maintenance function is to be carried out
There is no Stop button in the pop-up window. Instead, the Stop key can be used to stop the
maintenance function.
The following table shows a summary of the maintenance functions.
Note: these maintenance functions are only available for rack systems
When you start each maintenance function, make sure that the instrument is put in the condition
corresponding to the function described in Table 6.2-2.
The customer should perform LFC on a frequent basis. Please train all your customers according to
the procedure described below and inform them about the importance of keeping the liquid flow path
clean and the measuring cell at the best performance level.
If operators perform this procedure frequently it will lead to fewer customer complaints.
Required materials: SysClean adapter Ident. No.: 1933159.910
ISE Cleaning Solution Ident. No.: 1298500
Recommended frequency *: Every two weeks or after 2500 - 3000 cycles
∗ If customer maintenance has been done as recommended, it is only necessary to carry out an additional customer
cleaning at the PM visit. The service LFC is only necessary during PM if the sipper line is highly contaminated or if
LFC has not been carried out regularly by the customer.
Operator time: Approx. 3 minutes
Analyzer time: User LFC: Approx. 16 minutes (user), Service LFC: Approx. 32
min + time for MC prepare cycles
Precautions: The cleaning solution is corrosive. Wear protective gloves and
glasses when handling this solution.
Procedure
1. User Replace the PC bottle in position 3 by the SysClean adapter (keep
the CC bottle in position).
Figure 6.2-3
Insert the adapter with the marking "USER" in direction towards the
back of the instrument. Fill the ISE cleaning solution to the edge of
the fill mark on the adapter (approx. 9 ml). PC/CC bottle has to be
placed at the left side.
2. Service Same procedure as for customers but using the Service position of
the adapter.
3. User Touch the Liquid Flow Cleaning button on the screen (Screen: Util /
/Service Maintenance).
4. User The number of LFCs should be 1 for User maintenance.
When Liquid Flow Cleaning is completed the Liquid Flow system is set in the usual condition in which
the system is set when a run is completed in the normal manner. The function takes about quarter of
an hour.
L&A All Reset is a maintenance function that carries racks whose barcode has already been read to
the unloader, resets all the mechanical units and primes the pipetter system.
This function is to be carried out when an operation is aborted due to an error to take all the
mechanical units in the original position and the operation should be restarted from the condition
before the error occurred. The reset function in the Service software is identical to the reset which is
carried out at the beginning of each run.
About 20 ml water are needed to carry out the Reset function once.
Rack Clear is a maintenance function that carries all racks in the line to the output line (C line). Only
Start and Close are available in the reset pop-up window.
The following functions are to be carried out sequentially when Measuring Cell Preparation is started.
1. Reset
2. Priming the sipper system between the sipper syringe and the waste via SV4 (ca. 20 s)
3. Checking the system reagent inventory
4. Preparation 1 cycle (once without potential application)
5. Preparation 2 cycles according to the iteration number with the potential application
6. Preparation 3 cycle (once with the potential application)
7. Sipper finalization
When Measuring Cell Preparation is completed the sipper system is set in the usual condition in which
the system is set when a run is completed in the normal manner.
How to select a mechanical unit check function in the Mechanism Check screen
There are 18(Disk for 18 items) mechanical unit check functions available. Each of them corresponds
to a button in the Check screen. Touch the Check button in the Maintenance screen in order to display
the Maintenance items. After touching a button in the Maintenance Items, a pop-up window appears in
the screen.
Some functions can also be activated by touching a button in one of the screens for the mechanical
unit adjustment.
There are some buttons in each pop-up window; some pop-up windows also have an input area.
The general functions of each button and the input area are as follows.
Start button : To start the maintenance function corresponding to the button
Close : To close the pop-up and return to the Maintenance screen without carrying out any
maintenance functions
Input area : To define how many times the maintenance function is to be carried out
There is no Stop button in the pop-up window. Instead, the Stop key can be used to stop the
maintenance function. While a mechanical unit check function is in process, the remaining iteration
number is displayed in the corresponding button. The number is reset to 1 when the Stop key is
pressed and to 0 when the function is completed.
The following table shows a summary of the mechanical unit check functions.
Table 6.3-2
No. Name Function
11. Bar Code Card Reading Carries out Barcode Card Scan once including the database update
and the transmission to HOST.
This function is also activated by touching the button in the screen for
the mechanical position adjustment.
12. Reagent Pack BC Carries out R. Scan once including the database update and the
Scanning transmission to HOST.
This function is also activated by touching the button in the screen for
the mechanical position adjustment.
13. Gripper Quick Check Carries out a mock operation with tips and vessels mounted on each
corner of each magazine. Only the gripper and the pipetter move
according to the AP 2, whereas the pipetter carries out only picking up
a tip and discarding it moving between the pipetter buffer and the
pipetter washing station.
This function is also activated by touching the buttons in the screen for
the mechanical position adjustment.
Figure 6.3-3
14. Gripper Repetition Check Carries out a mock operation with all full magazines. Only the gripper
moves according to the AP 2 although no tip/vessel discarding takes
place. Instead, all tips and vessels are to be returned back to their
original positions.
This function is also activated by touching the button in the screen for
the mechanical position adjustment.
Figure 6.3-4
Table 6.3-3
No. Name Function
15. S/R Probe Check Carries out mock pipetter, reagent disk and cap opener cycles as
many times as entered in the input area with no liquid handling except
for the LLD. The maximum repetition number is 180.
The gripper carries a vessel to the vessel buffer (VB) at the beginning
of the function and discards it at the end.
This function is also activated by touching the button in the screen for
the mechanical position adjustment.
Figure 6.3-5
16. Gripper S/R Probe Check Carries out mock cycles except for the sipper as many times as
entered in the input area with no liquid handling except for the LLD.
The maximum repetition number is 180.
This function is also activated by touching the button in the screens for
the mechanical position adjustment.
Figure 6.3-6
Table 6.3-4
No. Name Function
17. Sipper Check Carries out mock sipper cycles as many times as entered in the input
area. The current position is switched alternately every cycle. The
gripper carries a vessel to the sipping position at the beginning of the
function and discards it at the end. The maximum repetition number is
180.
This function is also activated by touching the button in the screen for
the mechanical position adjustment.
Figure 6.3-7
18. S/R Probe LLD Check Checks the pipetter LLD function by moving the pipetter down to the
sample disk and the RackPack at position No. 1 of the sample disk
and the middle bottle at position No.1 of the reagent disk, respectively.
Figure 6.3-8
19. Sampler Check Carries out a mock operation with all racks on rack loader after reset
(only for Rack Sampler) action, checks the BC reading function on the rack sampler, and prints
out the result. Only the rack sampler moves in the same manner as in
the operation.
Introduction
This test is carried out to check the mechanical movements of the system.
Materials Required
Table 6.3-5
RackPack name Amount Material No.
CellCheck 6 1820451-122
Table 6.3-6
Sample name Cups in Rotor Pos. Amount
ProCell 1-30 1 ml per test
Table 6.3-7
Cups in
Sample name No. of racks Amount
Pos.
ProCell 6 1-5 1 ml per test
1. Preparation
1 ► Place six (6) sets of RackPacks filled with CellCheck in positions 1 to 6 on the reagent disk.
2 ► Load thirty (30) Hitachi sample cups filled with ProCell on 6 racks for Rack System, or into
positions 1 to 30 on the sample disk for Disk System.
3 ► Make sure the liquid waste container is empty.
4 ► Fill the system water container with system water (max. 3.0 liters).
5 ► Place a CleanLiner in the solid waste tray.
Note: Reset the waster number counter mechanically by detaching the solid waste tray.
6 ► Place 2 (two) sets of system reagent ProCell/CleanCell bottles filled with ProCell/CleanCell
in the system reagent container.
7 ► Place three magazines filled with tips and three (3) magazines filled with assay cups on a
magazine tray.
8 ► Remove all tips and cups from the incubator and the buffer plate
2. Check Procedure
Note: Make sure that no mechanical problem takes place by carrying out the “RESET” in reset
status.
3. Evaluation
1 ► At the beginning, check whether all mechanical positions (especially position relative to
pipetter, sipper, beads mixer) are sufficient within a tolerance limit; refer to Chapter 4.5
Mechanical Adjustment.
2 ► Observe the mechanical functions during the Running Test. Mechanical errors should not
take place.If you observe irregular movements, refer to Chapter 4.5 and carry out
Mechanical Adjustment.
3 ► Confirm that pipetter LLD functions properly.
6.3.2.5 MC CO Check
Introduction
The MC CO (measuring cell carryover) Check is part of the AM Test and is carried out if only MC
carryover is to be checked. In the event of a general test of the system (e.g. when installing the
instrument, replacing the measuring cell or carrying out a performance check during maintenance)
both the complete AM test and the TSH assay test should be carried out.
Materials Required
Table 6.3-8
Pos. No.
Name Beads R1 R2
Samples Required
Disk System
Rack Sampler
Place 1 ml ProCell in positions 2 to 5 of a rack and an empty Hitachi cup in Pos. 1. In a second rack,
pipette 1 ml of ProCell in positions 1 to 3.
The further procedure is as described for the AM Test in Chapter 6.3.2.3.
6.3.2.6 LH Check
Introduction
The LH (liquid handling) Check is part of the AM Test and is carried out if only liquid handling is to be
checked. In the event of a general test of the system (e.g. when installing the instrument, replacing the
measuring cell or carrying out a performance check during maintenance) both the complete AM test
and the TSH assay test should be carried out.
The following measurement is carried out:
10 x BCR2 (homogeneous measurement)
The test takes about 18 minutes.
Materials Required
Table 6.3-9
Pos. No.
Name Beads R1 R2
1 - - - -
2 - - - -
3 CellCheck ProCell ProCell ProCell
4 - - - -
5 - - - -
Samples Required
Disk System
Rack Sampler
Preparartions
Table 6.3-10
Pos. No. Rack Pack Beads R1 R2
1 BlankCell set BCR1 BCR2 -
2 BlankCell set BCR1 BCR2 -
3 CellCheck ProCell ProCell ProCell
4 CellCheck ProCell ProCell ProCell
5 SAP test Beads Buffer FC
6 CellCheck ProCell ProCell ProCell
7 TSH Test Beads R1 R2
Sample arrangement
Table 6.3-11
Run Cup in Rotor pos. Liquid
AM & TSH Test 3 BCR2 (1 ml)
4 to 10 ProCell (1 ml)
15 ProCell (1 ml)
16 TSH Cal 1
17 TSH Cal 2
Table 6.3-12
Rack in A line
Run Cup in rack pos. Liquid
No.
AM & TSH Test 1 1, 2 Empty cups
3 BCR2 (1 ml)
4, 5 ProCell (1 ml)
2 1 to 5 ProCell (1 ml)
3 1 ProCell (1 ml)
2 TSH Cal 1
3 TSH Cal 2
Start the test by pressing the AM & TSH Test in the APC screen.
For evaluation of the AM and TSH Test refer please chapter 2.3.13.1. and 2.3.13.3
• Select the [Print Results] button in the [Assay Performance] check screen after you finish each
test of Assay the Performance Check.
• Having touched the button in the screen, you will see a popup .
• Touch the [print out] button on the popup if the instrument has APC results.
• Input “Date”, “Operator name”, “Instrument No.”, “BCR1 target value” and “BCR2 target value” on
the printed results sheet, and compare the results for each item with the criteria and evaluate
whether the results are acceptable or not.
Carry out “Print Results” after you finish each “AM test”, “TSH test”, “MC CO test”, “LH test”
of the Assay Performance Check, and before you touch the [Restore Parameter] button or start
any Assay Performance Check (APC).
Figure 6.3-11
Figure 6.3-12
Select the [Restore Parameter] button in the [Assay Performance] check screen if you have finished
the Assay performance section. You do not need to restore if you have performed another APC run.
• You must check the instrument according to the procedures in the following cases.
• The detailed checking procedures and acceptable ranges are described in the preceding sections.
• Some checking procedures are necessary as shown in the following table when some parts or
assemblies are replaced.
• The Running Test is to be performed if you make adjustments to the instrument to check the
reliability of the hardware.
Table 6.3-13 When and what has to be performed (see below for further details)
System
MC PMT-HV Initial Artificial TSH Assay
Case Volume
Prepare Check BlankCell Media Test Test
check
Instrument Installation X X X X X X
Detection Unit Replacement X X X X X X
Measuring Cell Replacement X - X X X X
PMT Replacement X X X X
ANG-EP Replacement
L-AMP Replacement
Magnet Drive Replacement - - - - X X
Sipper Probe Replacement - X - X - -
Sipper Tubing Replacement - X - X - -
System
MC PMT-HV Initial Artificial TSH Assay
Case Volume
Prepare Check BlankCell Media Test Test
check
Pipetter Probe Replacement - - - X X -
Pipetter Tubing Replacement - - - X X -
Long-term Rest
Note: If the system appears to be contaminated we recommend performing the System Cleaning
function (refer to Maintenance) or exchanging the affected parts (tubings, etc.). For
instruments not used for more than 6 months, the 12 month PM procedure should also be
carried out (refer to Chapter 10).
General note
The following table shows the tolerance limits for each voltage and the Head Line Monitor.
6.4 Service
• After boot up the e411 or Logoff the e411, a pop-up window appears as shown in Figure 6.4-1.
• The Operation ID and Password are following.
• There are 15(Disk for 14 items) service functions available. Each of them corresponds to a button
in the Check screen. Touch the Check button in the Maintenance screen in order to display the
Maintenance items. After touching a button in the Maintenance Items, a pop-up window appears
in the screen.
In principle, the manual adjustment is carried out according to the following steps.
The adjustment parameters are updated when you close an adjustment by pressing the OK button.
This update is carried out even if an error takes place while the mechanical unit is moving after the OK
button has been touched. The adjustment parameters are displayed in the buttons in the three
screens for the adjustment.
By the time that the unit arrives at the position where the adjustment takes place, touching the buttons
is no longer effective. If the initial movement or the movement triggered by touching one of the buttons
for the movement is suspended for some reason, e.g. an error, the pop-up window can only be closed
using the Cancel button.
Usually the database update takes place in both the RAM and CF.
6.4.3.1 Initialization
The BCR Initialize function is to be carried out as follows when you replace the BCR.
• Touch the Service button in the Maintenance Screen (see Figure 6.4-2,Figure 6.4-3).
• After touching the button in the BCR Initialization, a pop-up window appears as in Figure 6.4-11.
• Touch the OK button to complete the BCR Initialize function.
Note: The FLASH memory in the BCR has a lifetime of up to 1000 storage times. Therefore, the
Initialize and Check Digit functions should not be carried out too often.
Note: The above items are stored into the FLASH memory of the BCR even if you switch the
instrument off.
Note: The user is recommended to use a bar code label that contains a check digit in order to
ensure reliable bar code reading.
1 ► Touch the Rack Assignment button in the System screen as shown in Figure 6.4-13.
2 ► After touching the Rack Assignment button in the System screen, a Rack-Id Digit setup pop-
up window appears as shown in Figure 6.4-14.
3 ► Select a digit number for Rack-Id reading.
4 ► Touch the OK button after selecting the number.
The counter should not normally be reset, since it allows the total number of cycles performed on the
instrument to be checked.
Note: This function should only be used for service activities. The number of total results to be
stored in the database is lower when this function is activated.
Figure 6.4-17
6.4.7.1 Introduction
After the measuring cell has been prepared and it has been checked that the signal level is in the
expected signal range by means of the high voltage adjustment, the initial BlankCell calibration has to
be performed. The instruments are calibrated to give the same homogeneous signal level.
The initial BlankCell measurement is the first calibration done on the instrument before other assays
can be calibrated. Each measured signal will first be BlankCell-corrected (parameters a (intercept) and
b (slope)). The corresponding concentration is then read from the specific assay calibration. For
further details refer to Chapter 2.3.12.
Figure 6.4-21
6 ► It is very important that the Initial Blank Cell is successful. If not, repeat the process until it is
successful, otherwise all results may not be corrected properly. Refer to Chapter 2.3.12.3
for current result targets. (If necessary start procedure once more. Go back to point 3.)
7 ► If the Initial Blank Cell was successful, remove the BLANK CELL RackPack. This is very
important, otherwise the instrument will do the periodic blank cell automatically.
8 ► Start reagent scan.
This message means that the procedure was successful; all quality criteria were met. No need for
further action.
This means that the procedure failed; one or more of the quality criteria failed. The detailed quality
messages and the data groups below have to be studied to determine the cause.
Three messages are possible, each depending on the results of the quality checks with status ok / not
ok.
In case of Öok no further action is needed.
In case of any Önot ok the following interpretations are possible:
Missing values
This field displays six dashes representing the six single measurements of the procedure.
In case of problems the data are not complete. The instrument was not able to perform the six
measurements. If a number appears instead of a dash, there was a problem with this measurement,
and this may indicate a mechanical problem.Problems with missing values cause further messages as
achievement of targets and deviations of replicates Önot ok.
Achievement of targets
It is possible that only one single measurement has failed. In this case the means seem to be
Öok but the achievement of targets and probably the deviation of replicates are
Önot ok.
This quality message was only designed for user´s periodical BlankCell.
Therefore, it is ignored in case of the Initial BlankCell.
b).
replicate The maximum acceptable deviation of replicates to their median in percent.
limit
Upper group shows data for measurements with BlankCell Reagent BCR1.
Lower group shows data for measurements with BlankCell Reagent BCR2.
Left column shows data for BlankCell measurements of last valid BlankCell procedure.
This column was only designed for the user’s periodical BlankCell.
Right column shows data for actual BlankCell measurements.
Table 6.4-2
N
Adjustment Function Purpose Adjustment Area
o.
1 Gripper To adjust all positions when adjustment has Magazine, Incubator, Buffer
All Positions never been performed.
Preparation
2 Gripper To adjust all positions when adjustment has Magazine, Incubator, Buffer
All Positions been performed.
3 Magazine To adjust positions on magazines. Magazine
Cup / Tip
4 Sipping – Inc To adjust positions on the incubator. Incubator
5 C / T Buffer To adjust positions on the buffer. Buffer plate
6 Gripper – Z To adjust the height direction. Magazine, Incubator, Buffer
The following table shows the adjustment positions in each adjustment area that are described in
Figure 6.4-22.
Table 6.4-3
No. Adjustment Area Adjustment Positions
1 Magazine TM1_12, TM12_29, CM1_1, CM29_1
2 Incubator Inc1_1, Inc4_1, Inc4_8, Sipping
3 Buffer Tip Pos.1, CB
Figure 6.4-22 How to mount tips and vessels for automatic adjustment
Note: Adjustment Tool for buffer area should be placed on buffer plate.
At first, all adjustment values are set to zero (0) internally and the Preparation movement is
performed. All areas are continuously adjusted.
Preparation is a movement to find roughly a position that can pick up Tips or Cups at each adjustment
position. Adjustment of all areas is performed in two (2) or three (3) cycles. Whether or not the third
cycle is performed is decided automatically on the basis of the results of the first and second cycles,
but height adjustment is performed in three cycles. Adjustment values in the memory and on the FD
are overwritten when each adjustment cycle is finished. The color Note: of the display windows for
adjustment value(s) depends on the adjustment situation.
4. Gripper All Positions
Adjustment of all areas is performed using the adjustment values in the memory. See Point (1) for
details of the adjustment process.
Adjustment of the magazine area is performed using the adjustment values in the memory. See Point
(1) for details of the adjustment process.
6. Sipping - Incubator
Adjustment of the incubator area is performed using the adjustment values in the memory. See Point
(1) for details of the adjustment process.
7. C / T Buffer
Adjustment of the buffer area is performed using the adjustment values in the memory. See Point (1)
for details of the adjustment process.
8. Gripper - Z
The height adjustment is performed in three cycles in the order of Sipping position and CB position.
Adjustment values in the memory and on the FD are overwritten when each adjustment cycle is
finished. The color Note: of the display windows for adjustment value(s) depends on the adjustment
situation.
9. Gripper Adjust Write Media
The function is to write down the gripper adjust data ( ADJSTLOG.TXT) onto a CF card.
Note: The display window is green, yellow or red. Refer to Chapter 6.4.5.6 for details of the
meanings of the three colors.
Adjustment values are displayed in the display windows for adjustment value(s) in the manual
adjustment screen. These displayed adjustment values are calculated from the value(s) obtained from
adjustment position(s) as shown in the following table.
Table 6.4-4
The Display Windows of Adjustment Position
Adjustment Area
Adjustment Value Using for Calculation
Magazine TM Front X TM12_29(X) and CM29_1(X)
Y TM1_12(Y) and TM12_29(Y)
TM Back X TM1_12(X) and CM1_1(X)
CM Left Y CM1_1(Y) and CM29_1(Y)
Incubator Sipping X Inc4_1(X) and Inc4_8(X)
Y Inc4_8(Y)
Incubator Y Inc1_1(Y) and Inc4_1(Y)
Buffer Tip Pos.1 X TP1(X)
Y TP(Y)
CB Pos. X CB(X)
Y VB(Y)
Incubator Sipping Z Sipping(Z)
Buffer CB.Pos. Z CB(Z)
The color of the display window for adjustment value(s) changes depending on the situation of
adjustment. However, if the instrument is switched off and on, all display windows change to green.
The following table shows the meaning of the displayed colors.
Table 6.4-5
Displayed Color Description
Green Adjustment finished successfully.
Yellow Corresponds to alarm code 51_01_02.
The Gripper could pick up a Tip or Cup at the adjustment limit of the adjustment position(s)
that is (are) used to calculate the adjustment value. Therefore, there is the possibility that the
automatic adjustment was not finished successfully.
Red Corresponds to alarm code 51_01_01.
The Gripper could not pick up a Tip or Cup at the adjustment position(s) that is (are) used to
calculate the adjustment value on at least one occasion. Therefore, the automatic adjustment
failed.
Note: If the displayed color is red, the adjustment value does not change. However, if the
adjustment values for X and Y are displayed in the same display window, one of the
adjustment values may be overwritten because the adjustment values for X and Y are
calculated separately.
The following table shows what is to be done for each color of the display window of adjustment
value(s).
Table 6.4-6
Displayed color What to do
Green > No need to adjust again because the automatic adjustment was finished successfully.
If any alarms occur during operation, check the cause of alarm, e.g. cups floating on the
incubator, or tips and cups leaning to the side on the magazine.
Yellow > Basically, no need to adjust again.
1. If the area for which the display window of adjustment value(s) has changed to yellow and
the area in which alarm has happened are the same, perform automatic adjustment in the
same area or adjust manually.
2. If the area for which the display window of adjustment value(s) has changed to yellow and
the area in which alarm has happened are different, check the cause of the alarm, e.g.
cups floating on the incubator or tips and cups leaning to the side on the magazine.
1. After performing manual adjustment of the affected position roughly, carry out automatic
adjustment of the area.
2. After checking mechanical position etc., carry out automatic adjustment of the affected
area or adjust manually.
In order to confirm that the automatic adjustment has been completed successfully, carry out the
Gripper S/R Probe Check by touching the corresponding button.
• Load three (3) magazines filled with Tips and three (3) magazines filled with Cups.
• Mount the System Water Container fill with system water (max. 3.5 Liters).
• Make sure that the Waste Box Tray is mounted correctly. Carry out mechanical reset of the waste
number counter by detaching the Waste Box Tray.
• Make sure that the Liquid Waste Container is mounted correctly.
• System Reagent PC/CC bottles are not to be mounted.
NOTICE
If you ignore a Pipetter LLD function which means LLD functions, but no liquid handling functions
during the Gripper S/R Probe Check, the following need not be prepared. If an LLD error takes place
in normal condition.
Table 7.2-2
Host
Host No
Screen/ No
Data Alarm (cobas Level Description
Printer (E2010
mode).
mode).
Pretreatment Film Detected Reag.F 37 70 6 Film detected on pretreatment
surface.
System Reagent Film Detected Reag.F 38 70 6 Film detected System Reagent
surface.
AB Level Range Over >AB 40 63 6 AB signal level is out of range during
“Pre-Operation” cycle.
( < 120 or > 400 counts )
AB Level Check Error AB.E 41 64 6 AB liquid level check failed during
“Pre-Operation” cycle.
Current Range Over >Curr 42 65 6 MC current out of range during “Pre-
Operation” cycle.
Current Range Check Error Curr.E 43 66 6 MC current check failed.
System Reagent Temperature SysR.U 44 120 6 System Reagent temperature is
Unstable unstable
Sample Clot Detected Samp.C 45 72 6 Clot detected when pipetter aspirates
sample.
Low Signal Sample CarOvr 46 71 6 Carryover from the previous sample
may have occurred.
Sample ID Error Cancel Cancel 47 121 6 Sample ID scanning was not
successful.
Below normal(expected)range L 48 41 6 Final concentration is below expected
range, less than lower limit.
Above normal(expected)range H 49 40 6 Final concentration is above expected
range,
more than upper limit.
Below measuring range <Test 50 27 6 Concentration (not including diluent
ratio) is below measured range, less
than lower limit.
Above measuring range >Test 51 26 6 Concentration (not including diluent
ratio) is above measured range, more
than upper limit.
Expired RackPack ReagEx 52 101 7 Result generated with expired reagent
pack.
No Sample Samp.S 53 3 6 No sample has been detected.
Sample LLD Inexecution SLLD.E 54 86 6 Before sample LLD is carried out, it
stops because tip of the pipetter probe
may dirty, etc.
Sample LLD Noise SLLD.N 55 87 6 Sample LLD error generated due to
detection of electrostatic discharge.
Current Range Over(Operation) >Curr 56 99 6 MC current range over range
during operation.
Instrument Factor A FacA 57 122 6 Instrument Factor A is set to 1
by the system
Signal level below limit <SigL 58 100 6 Signal level is below the limit.
Re-measurement is recommended
Calc Test Error ClcT.E 37 6 Data flag is attached if one or both of
two tests are flagged, no result.
Overflow Over.E 38 6 The result is mathematically over
flown.
7.3.2.1 Cap Opener F/B Home Position Detection, Cap Opener F/B Stop
Position Detection
Table 7.3-3
N
Check Item Remedy
o
1 Cap opener is touching something? Execute Cap Open / Close Mecha in Maintenance screen.
Check whether cap opener is touching anything.
Readjust cap opener if it touched something.
2 Photocoupler detached from plate? Remove rear cover.
Check whether the photocouplers below have become detached from
plate.
Home Pos. : PI031
Stop Pos. : PI032
Exchange photocoupler if it is detached from plate.
3 Wire lead disconnected from the connector? Refer to cross wiring reference.
Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from the
connector.
4 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Exchange DC power supply if it did not output +5V .
Check pin (#1 : +5V) of each of the photocouplers below.
Home Pos. : PI031
Stop Pos.: PI032
Exchange photocoupler if it does not output +5V.
5 Signal voltage output correct? Turn on power SW of instrument.
Disconnect J530 connector (for cap opener F/B movement) in the
Stand-by situation.
Move cap opener back and forth by hand.
Check signal voltage output of J531 connector below.
Home Pos.: #3 pin (closed : ca. 5V, open : ca. 0V)
Stop Pos. : #6 pin (same as above)
Exchange photocoupler if it does not output signal voltage.
N
Check Item Remedy
o
6 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
7 Tension of timing belt loose? Refer to Chapter 4.4.3.
Check tension of timing belt.
Adjust tension of timing belt by moving the pulse motor
assembly again if the timing belt is loose.
8 Timing belt itself defective? Check whether the timing belt is defective.
Exchange the timing belt if defective.
9 Motor and pulley fixed properly? Check that the pulse motor assembly and mount are fixed.
Fix pulse motor assembly and mount by tightening the fixing screws if
they are not properly fixed.
7.3.3.1 Beads Mixer Arm Home Position Detection, Beads Mixer Arm Stop
Position Detection
Table 7.3-5
N
Check Item Remedy
o
1 Vessel or tip around photocoupler? Remove cover R.
Check whether there are vessels or tips around the beads mixer
mechanism and remove.
2 Photocoupler connected correctly? Remove front cover assembly.
Check connections of each of the photocouplers below.
Home Pos. : PI041
Stop Pos. : PI042
Connect lead wire between photocoupler and connector if disconnected.
3 Detection plate damaged? Check detection plate of pulley S for damage.
Exchange pulley if detection plate of pulley is damaged.
4 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if pin is disconnected from the
connector.
5 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Exchange DC power supply if it did not output +5V .
Check Pin (#1 : +5V) of each of the photocouplers below.
Home Pos. : PI041
Stop Pos. : PI042
Exchange photocoupler if it does not output +5V.
6 Signal voltage output correct? Turn on power SW of instrument.
Disconnect J540 connector (for beads mixer arm rotation) in the Stand-
by situation.
Rotate the beads mixer arm by hand.
Check signal voltage output of J541 connector below.
Home Pos. : #3 pin (closed : ca. 5V, open : ca. 0V)
Stop Pos. : #6 pin (same as above)
Exchange photocoupler if it does not output signal voltage.
7 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
8 Tension of timing belt loose? Refer to Chapter 4.4.4.
Check tension of timing belt.
Adjust tension of timing belt by moving the pulse motor
PX 244M-02A assembly again if the timing belt is loose.
9 Timing belt itself defective? Check whether the timing belt is defective.
Exchange the timing belt if defective.
1 Motor or pulley fixed properly? Check that the pulse motor PX 244M-02A assembly and pulley are
0 fixed.
Fix the pulse motor PX 244M-02A assembly and pulley S by tightening
the fixing screws if they are not properly fixed.
N
Check Item Remedy
o
2 Photocoupler detached down Remove front cover assembly.
from plate? Remove beads mixer mechanism from instrument.
Check whether photocoupler for upper dead position detection has become
detached from frame.
Upper Dead Pos. : PI043
Exchange photocoupler if it is detached from plate.
3 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from the connector.
4 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Exchange DC power supply when it did not output +5V .
Check Pin (#1 : +5V) on the photocoupler below.
Upper Dead Pos. : PI043
Exchange photocoupler if it does not output +5V.
5 Signal voltage output correct? Turn on power SW of instrument.
Disconnect J540 connector (for beads mixer arm rotation) in the Stand-by
situation. Rotate the beads mixer arm by hand.
Check signal voltage output of J541 connector below.
Upper Dead Pos. : #9 pin (closed : ca. 5V , open : ca. 0V)
Exchange photocoupler if it does not output signal voltage.
6 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
7 Tension of timing belt loose? Refer to Chapter 4.4.4.
Check tension of timing belt.
Adjust tension of timing belt by moving the motor Z assembly again if the
timing belt is loose.
8 Timing Belt itself defective? Check whether the timing belt is defective.
Exchange the timing belt if it is defective.
9 Motor fixed properly? Check that the pulse motor Z assembly is fixed.
Fix the motor Z assembly by tightening the fixing screws if it is not properly
fixed.
N
Check Item Remedy
o
5 Signal voltage output correct? Turn on power SW of instrument.
Disconnect J650 connector (for BCR rotation) in the
Stand-by situation.
Rotate the BCR mechanism by hand.
Check signal voltage output of J651 connector below.
Home Pos. : #3 pin (closed : ca. 5V , open : ca. 0V)
Exchange photocoupler if it does not output signal voltage.
6 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
7 Motor fixed properly? Check that the motor BCR assembly is fixed.
Fix the motor BCR assembly by tightening the fixing screws if it is not
properly fixed.
8 Gear defective? Check whether part of gear of bracket is defective.
Exchange bracket if defective.
N
Check Item Remedy
o
3 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from
the connector.
4 FPC cable scratched? Check whether FPC cable is scratched. Exchange FPC cable if
scratched. Refer to cross wiring reference.
Check that resistance between each pin on the FPC cable is ca.0 [Ω]
using a multimeter.
Exchange FPC cable if the resistance between each pin on the FPC
cable is more than a few hundred ohms.
5 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Exchange DC power supply if it did not output +5V.
Check +5V is output at the pin of each of the photocouplers below.
Upper dead Pos. : #1 pin of PI054
Pipetter Z abnormal descent : #1, #3 pin of PI055
Exchange photocoupler if it does not output +5V.
6 Signal voltage output correct? Turn on power SW of instrument.
Wait until instrument goes into Stand-by.
In case of pipetter upper dead position detection :
Move pipetter mechanism in Z direction by hand.
In case of pipetter Z abnormal descent detection :
Move pipetter probe up by hand.
Check signal voltage output of J556 connector on Dist-PA board below.
Upper dead Pos. : #9 pin (closed : ca. 5V, open : ca. 0V)
Pipetter Z abnormal descent : #3 pin
(open : ca. 5V, closed : ca.0V).
Exchange photocoupler if it does not output signal voltage.
7 All cable leads connected correctly? Refer to Cross Wiring Reference.
Connect the connector when it disconnected.
8 Tension of timing belt loose? Refer to Chapter 4.4.5. Check tension of timing belt.
Adjust tension of timing belt by moving pulse motor KH423
M2-052. again if the timing belt is loose.
9 Timing belt itself defective? Check whether the timing belt is defective.
Exchange the timing belt if defective.
1 Motor and pulley fixed properly? Check that the pulse motor and pulley P are fixed.
0 Fix pulse motor and pulley P by tightening the fixing screws if they were
not properly fixed.
7.3.7.1 Pipetter Syringe Upper Dead Position Detection, Sipper Syringe Upper
Dead Position Detection
Table 7.3-11
N
Check Item Remedy
o
1 Photocoupler connected correctly? Remove front cover assembly.
Check the connections of each of the photocouplers below.
Pipetter syringe upper dead Pos. : PI091
Sipper syringe upper dead Pos. : PI101
Connect lead wire between photocoupler and connector if disconnected.
2 Detection plate damaged? Check detection plate of fixture for damage.
Exchange fixture if its detection plate is damaged.
3 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from
the connector.
N
Check Item Remedy
o
4 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Exchange DC power supply if it did not output +5V.
Check pin (#2 : +5V) of the photocouplers below.
Pipetter syringe upper dead Pos. : PI091
Sipper syringe upper dead Pos. : PI101
Exchange photocoupler if it does not output +5V.
5 Signal voltage output correct? Turn on power SW of instrument.
Execute Pipetter Prime and Sipper Prime in Maintenance Screen.
Check signal voltage output of J591 connector below.
Pipetter syringe upper dead Pos. : #3 pin
(closed : ca. 0V, open : ca. 5V)
Sipper syringe upper dead Pos : #6 pin (same as above).
Exchange photocoupler if it does not output signal voltage.
6 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
7 Tension of timing belt loose? Refer to Chapter 4.4.9.
Check tension of timing belt.
Adjust tension of timing belt by moving pulse motor
PH265M33 assembly again if the timing belt is loose.
8 Timing belt itself defective? Check whether the timing belt is defective.
Exchange the timing belt if defective.
9 Motor and pulley fixed properly? Check that the pulse motor PH265M33 and pulley 2 are fixed.
Fix pulse motor and pulley 2 by tightening the fixing screws if they are
not properly fixed.
N
Check Item Remedy
o
7 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
8 Tension of timing belt loose? Refer to Chapter 4.4.6.
Check tension of timing belt.
Adjust tension of timing belt by moving holder C again if the timing belt is
loose.
9 Timing Belt itself defective? Check whether the timing belt is defective.
Exchange the timing belt if defective.
1 Motor and pulley fixed correctly? Check that the motor C assembly and pulley C are fixed.
0 Fix motor C assembly and pulley C by tightening the fixing screws if they
were not properly fixed.
N
Check Item Remedy
o
3 Detection plate damaged? Check the shield plate for damage.
Exchange the shield plate if damaged.
4 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Exchange DC power supply if it did not output +5V.
Check pin (#1, #3 : +5V) of photocoupler below.
Abnormal descent : PI066
Exchange photocoupler if it does not output +5V.
5 Signal voltage output correct? Turn on power SW of instrument.
Remove cover Y and magazine tray.
Move gripper finger arm up by hand.
Check signal voltage output of J565 connector of
Dist-TVA board below.
Upper dead position : #10 pin
(closed : ca. 0V, open : ca. 5V).
Exchange photocoupler if it does not output signal voltage.
6 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from the
connector.
N
Check Item Remedy
o
3 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from the
connector.
4 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Exchange DC power supply if it did not output +5V.
Check pin (#1 : +5V) of each of the photocouplers below :
Home Pos. : PI141
Box detection : PI142
Exchange photocoupler if it does not output +5V.
5 Signal voltage output correct? Turn on power SW of instrument.
Disconnect J640 connector (for solid waste shaking) in Stand-by.
Move the solid waste container back and forth by hand.
Check signal voltage output of the J641 connector below.
Home Pos. : #3 pin (closed : ca. 5V, open : ca. 0V)
Box detection : #4 pin (closed : ca. 0V, open : ca.5V)
Exchange photocoupler if it does not output signal voltage.
6 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
N
Check Item Remedy
o
7 Tension of timing belt loose? Refer to Chapter 4.4.7.
Check tension of timing belt.
Adjust tension of timing belt by moving holder MD again if the timing belt
is loose.
8 Timing Belt itself defective? Check whether the timing belt is defective.
Exchange the timing belt if defective.
9 Motor and pulley fixed correctly? Check that the X motor S assembly and holder MD are fixed.
Fix X motor S assembly and holder MD by tightening the fixing screws if
they are not fixed.
N
Check Item Remedy
o
9 Timing belt itself defective? Check whether the timing belt is defective.
Exchange timing belt if it is defective.
1 Motor and pulley fixed properly? Check that cord P587 assembly with motor and pulley P are fixed.
0 Fix cord P587 assembly with motor and pulley P by tightening the
fixing screws if they are not properly fixed.
N
Check Item Remedy
o
2 Signal voltage output correct? Turn on power SW of instrument.
Wait until instrument goes into Stand-by.
Insert tip into buffer position by hand.
Check signal voltage output of J559 connector below.
Signal 1 : #6 pin (closed : ca. 0V, open : ca. 5V)
Signal 2 : #4 pin (same as above)
Exchange photocoupler if it does not output signal voltage.
3 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from the
connector.
4 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
N
Check Item Remedy
o
3 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Exchange DC poser Supply if it did not output +5V.
Check +5V is output at pin of the photocoupler below.
Bottle 1 : #1 pin of J663 connector.
Bottle 2 : #4 pin of J663 connector.
Photocoupler outputs ca. +5V when there is PC/CC bottle in the system
reagent unit and ca. 0V when there is no bottle in the system reagent
unit.
Exchange the photosensor if it does not output +5V.
4 Signal voltage output correct? Turn on power SW of instrument.
Wait until the instrument goes into Stand-by.
Place PC/CC bottle in the system reagent unit by hand.
Check signal voltage output of J663 connector below.
Signal 1 : #3 pin (closed : ca. 0V, open : ca. 5V)
Signal 2 : #6 pin (same as above)
Exchange photocoupler if it does not output signal voltage.
5 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
7.3.16.1 A - Line
Table 7.3-25
N
Check Item Remedy
o
1 Photocoupler connected correctly? Check the connections for each of the photocouplers below.
Tray detection : PI200
Home Pos. detection : PI201
Rack feed end Pos. detection : PI202
Connect lead wire between photocoupler and connector if disconnected.
2 Detection plate damaged? Check detection plate of pulley.
Exchange pulley if its detection plate is damaged.
3 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from the
connector.
4 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Signal voltage output correct? Exchange DC power supply if it did not output +5V.
Confirm the function of each photocoupler by using the Monitoring
Sensors function in the Admin screen.
Exchange photocoupler if defective.
5 All cable leads connected correctly? Refer to Cross Wiring Reference.
Connect the connector when it disconnected.
6 Timing belt itself defective? Check whether timing belt A is defective.
Exchange timing belt A if defective.
7 Motor fixed properly? Check that the motor assembly is fixed properly.
Fix the motor assembly by tightening the fixing screws if it was not
properly fixed.
7.3.16.2 B - Line
Table 7.3-26
N
Check Item Remedy
o
1 Photocoupler connected correctly? Check the connections for each of the photocouplers below.
Home Pos. detection : PI210
Rack feed end Pos. detection : PI211
Rack detection : PI212
STAT Pos. detection : PI213
Sampling Pos. detection : PI215
BCR read Pos. detection : PI216
Cup detection : PI217
Connect lead wire between photocoupler and connector if disconnected.
2 Detection plate damaged? Check detection plate of pulley.
Exchange pulley if its detection plate is damaged.
3 Wire lead disconnected from the connector? Refer to cross wiring reference.
Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from the
connector.
4 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Signal voltage output correct? Exchange DC power supply if it did not output +5V.
Confirm the function of each photocoupler by using the
Monitoring Sensors function in the Admin screen.
Exchange photocoupler if defective.
6 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
N
Check Item Remedy
o
8 Timing belt itself defective? Check whether timing belt B is defective.
Exchange timing belt B if defective.
9 Motor fixed properly? Check that motor assembly is fixed.
Fix the motor assembly by tightening the fixing screws if it was not
properly fixed.
7.3.16.3 C - Line
Table 7.3-27
N
Check Item Remedy
o
1 Photocoupler connected correctly? Check the connections for each of the photocouplers below.
Home Pos. detection : PI221
Rack feed end Pos. detection : PI222
Rack full detection : PI223
Tray detection : PI224
Tray full 1 detection : PI225
Tray full 2 detection : PI226
Connect lead wire between photocoupler and connector if disconnected.
2 Detection plate damaged? Check detection plate of pulley.
Exchange pulley if its detection plate is damaged.
3 Wire lead disconnected from Refer to cross wiring reference.
the connector? Check whether pin is disconnected from the connector.
Insert pin into the connector again if it is disconnected from the
connector.
4 +5V supply voltage supplied? Check that +5V is supplied by DC power supply.
Signal voltage output correct? Exchange DC power supply if it did not output +5V.
Confirm the function of each photocoupler by using the
Monitoring Sensors function in the Admin screen.
Exchange photocoupler if defective.
5 All cable leads connected correctly? Refer to cross wiring reference.
Connect the connector if it is disconnected.
6 Timing belt itself defective? Check whether timing belt CG is defective.
Exchange timing belt CG if defective.
7 Motor fixed properly? Check that the motor assembly is properly fixed.
Fix the motor assembly by tightening the fixing screws if it was not
properly fixed.
7.3.17.1 Introduction:
There are two CF cards installed at User interface panel PC of e411 instrument. The CF of drive C: is
used for Embedded Windows XP OS and drivers. And the CF of drive D: is used for application
software and data. If the CF of drive C: is damaged, the Windows XP OS can not boot. In this case,
PC should be replace.
The target of this document is the CF of D drive in which the application software and data were
installed.
Alarm code
When the alarm code of 41-01-01 to 41-01-10 and 57-01-01 are appeared, Please format D: drive and
reinstall the application soft following.
The procedure for format D: drive CF card will be shown below.
► Push Logoff button and display Logoff dialog screen, then push Shutdown button.
Figure 7.3-1
Figure 7.3-2
4 ► Touch [Start]-[Run] to show the run window and input “format d:” at “Run” windows.
► Please confirm that the data you need is backed up. Then touch “OK” button.
Figure 7.3-3
Figure 7.3-4
5 ► Enter the volume label for drive d: “toshiba512m” and enter “Y” key to process format. After
formatting the D: drive, Volume label input is requested. Input “toshiba512m” and enter
“Enter” key to finish the format process.
Figure 7.3-5
6 ► If the format can no be done, please replace CF and reinstall the D: drive.
If the same alarm still appeared, please replace the CF card. The procedure is shown below. If the
same alarm disappeared, please reinstall application software.
1 ► The spare part number of CF card is “J789021”, part name is “Flash Card”.Default of this
card is formatted.
2 ► CF card slot is located at the left side of User Interface panel PC. Take out the + screw
show at following picture.
Figure 7.3-6
3 ► The D drive CF is inserted at bottom slot of CF slot. The picture is shown as following.
Figure 7.3-7
4 ► Please take an action to prevent electrostatic discharge before touch the CF card.
5 ► Push the Slot in/out button to remove damaged CF card and insert a new one.
6 ► Attach the slot cover by tighten the screw.
7.4.1 Covers
• check for lead wires is pinched
• check for photo couplers move out
• check for shift of positioning
• adjust the rack pack position on reagent disk and pipetter nozzle
• adjust the reagent disk rotor belt tension
7.4.16 Check for the measuring data below Assay Performance Check
• PMT-HV Adjustment
• Artificial media test
• TSH Assay test
Table 8.1-1
No. Name Part Number What purpose is Referred Chapter
1 Gripper Adjustment 741-8742 To be used for manual gripper adjustment Chapter 4.5.2
Tool(1) for Tip position attaching between the gripper Mechanical Adjustment
fingers. Procedure
2 Gripper Adjustment 741-8743 To be used for manual gripper adjustment for Cup Chapter 4.5.2
Tool(2) position attaching between the gripper fingers. Mechanical Adjustment
Procedure
3 Gripper Adjustment 741-8741 To be used for manual gripper adjustment for Chapter 4.5.2
Tool(3) Sipping / Incubator / Buffer position attaching Mechanical Adjustment
between the gripper fingers. Procedure
4 Gripper TABLE COVER 741-8813 To be used for manual gripper adjustment for Vessel Chapter 4.5.2
ADJUSTMENT TOOL Waste position attaching between the gripper Mechanical Adjustment
fingers. Procedure
5 Magnet Drive Tool 741-1627 To be used for adjusting a magnet position where Chapter 4.4.14.7
may be sufficient for Beads capturing.
Attach the tool into a Base in stead of a Measuring
Cell.
6 Tin Plate 0.8 741-1985 To be used for adjusting a tolerance limit where Chapter 5.3.4
Pipetter detects a crash.
7 Tin Plate 1.2 741-1986 According to Adjustment Procedure, check the crash Chapter 5.3.4
sensor signal output by measuring the voltage
between the connector pin#3 of J556 and TP2 AG
on DIST-PA Board.
8 Pin Extraction Tool 1 741-5916 To be used for detaching a wire from a connector for
instance, P591, P592, etc.
9 Pin Extraction Tool 2 741-5919 To be used for detaching a wire from a connector for
instance, J591, J592, etc.
10 Pin Extraction Tool 3 741-5915 To be used for detaching a wire from a connector for
instance, P16, J16 etc.
11 Pin Extraction Tool 4 J826845 To be used for detaching a wire from a connector for
instance, P13, P14, etc.
12 Long Driver N330013 To be used for dis- and re-assembling a mechanical
part.
13 PPM Meter 1903152 To be used for mixer speed adjustment. Chapter 2.6.1.1
14 Flange Kit 1802208 To be used for tubings.
Safety Information
Refer to this section when you find Read Safety Information on the section title.
WARNING
Section 10.1 Maintenance
Procedure Overview
• Use the personal protective
equipment (such as rubber
gloves, lab coat, safety glasses,
etc.) recommended by your
facility when handling
biohazardous materials.
• Be sure to wear personal protective
equipment (such as rubber gloves,
lab coat, safety glasses, etc.) before
performing any maintenance/checks,
and follow the procedure described
in this operator’s manual.
CAUTION
Section 10.1 Maintenance Procedure Overview
• Do not open the sipper safety cover while the analyzer is in operation. Otherwise, the operation will
stop. DO NOT open the sipper safety cover unless the analzyer is in Stand-by.
• Be sure to close the cover after you placed/replaced system reagents, or performed maintenance.
Otherwise, the analyzer will not operate.
• To open/close the sipper safety cover, press the proper spot that is indicated in this manual.
CAUTION
Section 10.1.18 Clean Mixer Paddle
Be careful not to damage the Mixer probe during cleaning.
CAUTION
Section: 10.1.7.2 Clean S/R Probe
Be careful not to damage the lower end of the S/R and sipper probe during cleaning.
CAUTION
Section 10.1.22 Clean Incubator and Aspiration Station
DO NOT use an acid solution or an alkaline solution to clean Incubator and Aspiration Station. The compartment is
made of aluminum and these solutions degrade the metal.
CAUTION
Section 10.1.1 Replace pipetter tube 510
Section 10.1.2 Replace pipetter tube 465 with nozzle seal
Section 10.1.6 Clean Rinse Stations for S/R Probe, Mixer and Sipper Probe
Section 10.1.7.1 Clean Sipper Probe
Section 10.1.7.2 Clean S/R Probe
Section 10.1.8 Clean water filter main pump
Section 10.1.9 Clean system water container
Section 10.1.11 Clean ProCell/CleanCell Compartments
Section 10.1.20 Check / Clean Gripper finger
Section 10.1.21 Clean light barriers and mechanic parts
Section 10.1.22 Clean Incubator and Aspiration Station
Section 10.1.30 Clean peltier elements from dust if necessary
Section 10.1.31 Clean BCR window and window reagent disk
Section 10.1.41 Clean the valve body on the system water container
Section 10.1.42 Exchange O-ring
Section 10.1.44 Exchange the spring at the gripper finger
Power must be off in order to move analyzer components. If power is on, the motors are engaged and attempted
movement may damage these components.
CAUTION
Section 10.1.6 Clean Rinse
Stations for S/R Probe, Mixer
and Sipper Probe
Any component that comes into
contact with sample fluids is
potentially biohazardous and should
be handled in an appropriate
manner. Wear protective gloves
when handling these materials.
CAUTION
Section 10.1.13 Check drain tubes for contamination and exchange if necessary
Section 10.1.14 Drain the outlet pipettor wash station and waste pipe
Section 10.1.15 Perform Liquid Flow Cleaning
CAUTION. WARNING. CORROSIVE. SysClean etc., cause severe burns. Keep out of reach of children. In case of
contact with eyes, rinse immediately with plenty of water and seek medical attention. Remove all contaminated
clothing immediately. Wear suitable gloves and eye/face protection.
CAUTION
Section 10.1.13 Check drain tubes for contamination and exchange if necessary
Section 10.1.14 Drain the outlet pipettor wash station and waste pipe
Section 10.1.15 Perform Liquid Flow Cleaning
If you spill SysClean etc., reagent on the analyzer, immediately clean the spill with distilled or deionized water.
WARNING
Section 10.1.13 Check drain tubes for contamination and exchange if necessary
Section 10.1.14 Drain the outlet pipettor wash station and waste pipe
Section 10.1.15 Perform Liquid Flow Cleaning
SysClean etc., reagent combined with the contents of the liquid waste container could cause potentially
harmful fumes.
CAUTION
Section 10.1.13 Check drain tubes for contamination and exchange if necessary
Section 10.1.14 Drain the outlet pipettor wash station and waste pipe
Section 10.1.15 Perform Liquid Flow Cleaning
If you spill SysClean etc., reagent on the analyzer, immediately clean the spill with distilled or deionized
water.
CAUTION
Section 10.1.10 Clean Liquid
Waste Container
Any component that comes into
contact with sample fluids is
potentially biohazardous and should
be handled in an appropriate
manner. Wear protective gloves
when handling these materials.
WARNING
Section 10.1.10 Clean Liquid
Waste Container
DO NOT add bleach or strong
alkaline disinfectants (pH > 9.5) to
the liquid waste container.
Disinfectants combined with the
contents of the liquid waste could
cause potentially harmful fumes.
CAUTION
Section 10.1.11 Clean ProCell/CleanCell Compartments
DO NOT use an acid solution or an alkaline solution to clean the system reagent compartment. The compartment is
made of aluminum and these solutions degrade the metal.
CAUTION
Section 10.1.12 Clean Reagent
Disk and Compartment
Any component that comes into
contact with sample fluids is
potentially biohazardous and should
be handled in an appropriate
manner. Wear protective gloves
when handling these materials.
CAUTION
Section 10.1.12 Clean Reagent Disk and Compartment
Take care not to scratch or smear the bar code reader window.
CAUTION
Section 10.1.3 Replace Pinch
Valve Tubing
Section 10.1.37 Exchange
Measuring Cell
Section 10.1.38 Exchange Tube
B for MC
Section 10.1.39 Exchange Tube
B for Sipper with Nozzle seal
Section 10.1.40 Exchange Tube
B for Sipper Syringe
Section 10.1.43 Exchange sipper
wash station
Any component that comes into
contact with sample fluids is
potentially biohazardous and should
be handled in an appropriate
manner. Wear clean, protective
gloves when changing pinch valve
tubing.
WARNING
Section 10.1.3 Replace Pinch Valve Tubing
Section 10.1.3710.1-37 Exchange Measuring Cell
Section 10.1.38 Exchange Tube B for MC
Section 10.1.39 Exchange Tube B for Sipper with Nozzle seal
Section 10.1.40 Exchange Tube B for Sipper Syringe
Section 10.1.43 Exchange sipper wash station
There may still be some liquid in the tubing. The liquid that flows through this tubing comes from the
measuring cell and is potentially biohazardous. Wear gloves.
CAUTION
Section 10.1.3 Replace Pinch Valve Tubing
Make sure that you do not damage the fitting when replacing the tubing.
CAUTION
Section 10.1.4 Replace Pipettor Seals
For the S/R and sipper pipettor - loosen and REMOVE the locking screw to prevent damage to the glass barrel.
CAUTION
Section 10.1.4 Replace Pipettor Seals
Section 10.1.5 Exchange packing fo sipper & pipetter syringe
Do not overtighten the pipettor retaining screw.
CAUTION
Section 10.1.4 Replace Pipettor Seals
If the pipettor holder is not properly seated in its mounting block recess, damage to the pipettor assembly
may occur.
CAUTION
Section 10.1.4 Replace Pipettor Seals
Section 10.1.5 Exchange packing fo sipper & pipetter syringe
The top and side tubing retaining nuts can be damaged if excessive force is applied when replacing them.
Take care not to cross-thread these nuts.
CAUTION
Section 10.1.9Check mixer speed; adjust if necessary
Before starting operation or after maintenance/checks, be sure that all analyzer covers are closed. DO
NOT open the analyzer covers while the analyzer is in operation.
Do not open the Reagent Disk cover while the analyzer is in operation.
The recommended frequency is based on using the analyzer 8 hours per day, 5 days per week.
The materials required provides you with a list of all materials needed to perform each procedure.
The time required includes both operator time and analyzer time, when appropriate.
Precautions are included for your protection.
The Procedure gives step-by-step directions for performing the required maintenance function. This part
frequently is divided into smaller procedure blocks to help you organize your approach to maintenance.
Replacement Parts
A replacement part may be needed for a specific maintenance procedure. That part's description and
catalog number are included in the Materials Required table of each procedures' Introduction. Please use
the catalog number when ordering replacement parts. Remember, you are responsible for maintaining an
adequate spare parts inventory.
For most efficient use of time, gather all required materials before starting the maintenance procedure.
Maintenance Schedule
Detailed descriptions of the maintenance procedures listed below are found later in this chapter.
10.1.1.1 Procedure
1 ► Open the Utility/Maintenance tab screen after initialization is complete. Select “Maintenance” in
the maintenance type list and then “S/R Pipettor Prime” in the maintenance item list. The S/R
Pipettor Prime pop-up window opens.
2 ► Type “10” in the text box and touch Start. The liquid system is purged of air.
3 ► During priming, check the pipettor and tubing connections for leaks. The plunger should be in
the correct position and must be moving up and down continuously.
Procedure
1 ► Open the Utility/Maintenance tab screen after initialization is complete. Select “Maintenance” in
the maintenance type list and then “S/R Pipettor Prime” in the maintenance item list. The S/R
Pipettor Prime pop-up window opens.
2 ► Type “10” in the text box and touch Start. The liquid system is purged of air.
3 ► During priming, check the pipettor and tubing connections for leaks. The plunger should be in
the correct position and must be moving up and down continuously.
Worn pinch valve tubing allows liquid to leak, thereby affecting the accuracy of the pipetting volumes and
the ability to properly clean the measuring cell.
Recommended frequency: 6 Months at PM visits.
Operator time: Approximately 2 minutes.
Analyzer time: Approximately 15 minutes.
Precautions: The operation switch must be OFF while changing the tubing.
Materials Required Catalog Number
Pinch valve tubing 741-0941
The pinch valve tubing and fittings are located above the S/R and sipper pipettors. Refer to the reference
point photograph at the right and then to the enlarged view of the area in the photograph below.
1 ► Remove the distilled water container and cover around the foot area of the pipettor unit with
paper towels
2 ► Carefully remove the pinch valve tubing from the fittings on the metal plate while holding a dry
gauze square around the tubing.
3 ► Use the gauze square to absorb liquid comes out of the released tubing.
4 ► Remove the tubing from the pinch valve and discard it.
5 ► Take a new piece of tubing and insert it through the pinch valve. Verify the tubing length is 180
mm. If not, cut the tubing to that length.
6 ► Carefully slide the ends of the tubing over each of the fittings. In the photo on the previous page,
one set of fittings is shown with solid red circles and the other pair of fittings is shown with
dotted red circles. Verify that each tubing end is connected to the appropriate fitting according
to the diagram shown below.
NOTICE
If a fitting breaks while exchanging a pinch valve tube, exchange it with a new one by following the
procedure described below .
1. Remove the broken fitting by turning it counterclockwise.
2. Attach a new fitting on the valve block by turning it clockwise.
3. Proceed with the pinch valve tubing maintenance.
Disassemble Pipettor
1 ► Disconnect the tubing at the top of the pipettor assembly by turning its retaining nut
counterclockwise.
2 ► Hold a dry gauze square or other absorbent material next to the side tubing of the pipettor;
disconnect this tubing by turning its retaining nut counterclockwise. Use the gauze square to
absorb water as it drains from the tubing.
3 ► Loosen the knurled locking screw from the top of the pipettor assembly by turning it
counterclockwise.
4 ► Carefully pull the pipettor assembly out of its mounting block. With one hand, support the glass
barrel. With the other hand, grasp the pipettor holder and lift up, then out to remove the
assembly. Make sure the plunger is removed from the U-shaped slot.
Note: For sipper pipettor - pull the plunger down far enough so that you can lift out the glass barrel.
Then remove the remaining pipettor assembly.
5 ► Remove the glass pipettor barrel. Set it aside in a safe place. Remove the o-ring that fits below
the pipettor barrel and set in a safe place.
1 ► Turn the pipettor holder upside down and expose the plunger retaining screw.
2 ► Apply the spanner wrench to the plunger retaining screw and turn the screw counterclockwise
until it is loose.
3 ► Turn the pipettor holder right side up and carefully lift the pipettor holder off of the plunger.
Leave all parts on the plunger.
1 ► Remove the seal piece, press piece, spring and retaining screw from the plunger. Dispose of
the seal piece.
2 ► Wipe the plunger with a gauze square soaked in 70% isopropyl alcohol to remove any debris.
Follow with a gauze square soaked with distilled or deionized water.
3 ► Ensure that the retaining screw, spring and press piece are properly positioned on the plunger.
Place the new seal on the plunger, verifying the rounded end is down. Refer to the graphic on
the right for the correct order of the individual parts. Although the actual components of the
sipper pipettor may not be the same size as those of the S/R pipettor, they are assembled in
the same order.
4 ► Place the plunger back in the pipettor holder with the pipettor vertically oriented. Tighten the
retaining screw with the spanner wrench until the screw is snug with the pipettor holder.
Reassemble Pipettor
1 ► Place the pipettor holder o-ring into the pipettor holder. If the o-ring appears damaged or worn,
replace it.
2 ► Inspect the glass pipettor barrel for chips or cracks at the top and bottom. If the pipettor barrel
is etched or damaged in any way, replace it. Place the pipettor barrel over the plunger and onto
the pipettor holder. It is easier to place the barrel back on if the barrel and pipettor holder are
dry.
3 ► Return the pipettor holder onto its mounting block by tilting the top of the pipettor toward the
instrument. Ensure that the pipettor holder is in the recess on the top surface of the mounting
block.
4 ► The notched (bottom) end of the plunger must be secured within the U-shaped notch of the
stepper motor.
5 ► Rotate the pipettor holder until the side tube port is positioned to accept the pipettor tubing.
6 ► Finger-tighten the top knurled locking screw. Ensure that the pipettor assembly is seated
correctly (not loose or crooked) and the glass barrel o-ring is properly positioned.
7 ► Reconnect the pipettor's top tubing and finger-tighten its retaining nut.
8 ► Reconnect the side tubing and finger-tighten its retaining nut.
9 ► Turn the analyzer power ON.
1 ► Open the Utility/Maintenance tab screen after initialization is complete. Select “Maintenance” in
the maintenance type list and then “S/R Pipettor Prime” in the maintenance item list. The S/R
Pipettor Prime pop-up window opens.
2 ► Type “10” in the text box and touch Start. The liquid system is purged of air.
3 ► During priming, check the pipettor and tubing connections for leaks. The plunger should be in
the correct position and must be moving up and down continuously.
4 ► If air bubbles are found on the plunger, gently tap on the glass barrel to remove the air bubbles.
If this is not successful, remove the pipettor and thoroughly clean the plunger again with a
gauze square soaked in 70% isopropyl alcohol.
5 ► Repeat steps 1 ► - 4 ►, if necessary.
6 ► When you change the sipper pipettor seal, repeat steps 1 ► - 5 ►, but select “Sipper Pipettor
Prime” instead of “S/R Pipettor Prime” in the maintenance items list.
Note: The sipper pipettor plunger does not move up and down during a sipper prime.
Procedure
4 ► The following Figure 10.1-27 is the top view after following the step 1.
Figure 10.1-27 Top view of the syringe with the nipple removed.
1 ► Open the Utility/Maintenance tab screen after initialization is complete. Select “Maintenance” in
the maintenance type list and then “S/R Pipettor Prime” and “Sipper Pipettor Prime” in the
maintenance item list. The S/R Pipettor Prime and Sipper Pipettor Prime pop-up window opens.
2 ► Type “10” in the text box and touch Start. The liquid system is purged of air.
3 ► During priming, check the pipettor and tubing connections for leaks. The plunger should be in
the correct position and must be moving up and down continuously.
10.1.6 Clean Rinse Stations for S/R Probe, Mixer and Sipper Probe
NOTICE
Read Safety Information
Contamination in the rinse stations for the S/R probe, the mixer and the sipper probe can be responsible
for carryover. To prevent contamination, clean the rinse stations every 6 Months at PM visit.
Recommended frequency: 6 Months at PM visits.
Operator time: Approximately 10 minutes.
Analyzer time: None.
Precautions: The operation switch must be OFF.
Materials Required Catalog Number
Cotton swabs obtain locally
70% isopropyl alcohol obtain locally
Syringe with attached tubing obtain locally
Distilled or deionized water ----------
(approximately 300 mL)
Procedure
1 ► Move the S/R probe and the mixer to an area away from the rinse stations. Also, open the
sipper safety cover and move the sipper probe to an area away from the rinse stations.
2 ► Take a syringe with tubing attached to its end (as in the photo on the right) and aspirate the
water out of the mixer rinse station.
3 ► Use a cotton swab soaked in 70% isopropyl alcohol to thoroughly clean the S/R probe, mixer
and sipper rinse stations.
4 ► Take another cotton swab soaked in distilled or deionized water and wipe the rinse stations
again.
5 ► Fill the syringe with distilled or deionized water and refill the mixer rinse station. This should
take approximately 50-100 mL of water. Also, flush the S/R probe and sipper rinse stations with
water (50-100 mL).
6 ► Aspirate the water out of the mixer rinse station again using the syringe.
7 ► Refill the mixer rinse station and flush all the rinse stations with distilled or deionized water one
final time.
8 ► When you power ON the analyzer, it performs the start-up reset operation, and each
mechanism returns to its home or Stand-by position.
Note: You may want to use two syringes for this procedure - one for aspirating liquid and the other
for refilling the rinse stations.
Dirt on the sipper probe may cause contamination and carryover, and affect results. Clean this part at PM
visits to prevent contamination.
Recommended frequency: 6 Months at PM visits.
Operator time: Approximately 1 minutes.
Analyzer time: None.
Precautions: The operation switch must be OFF.
Materials Required Catalog Number
Gauze squares obtain locally
Distilled or deionized water obtain locally
70% isopropyl alcohol obtain locally
Procedure
Figure 10.1-32 Opening/closing the sipper safety cover (Push the circled point)
2 ► Move the sipper probe to an area where you can readily access it.
3 ► Wipe the outer surfaces of the sipper probe with a gauze square soaked in 70% isopropyl
alcohol. Follow with a gauze square soaked in distilled or deionized water.
4 ► When completed, close the sipper safety cover.
5 ► When you power ON the analyzer, it performs the start-up reset operation, and each
mechanism returns to its home or Stand-by position.
Dirt on the sample/reagent (S/R) probe may cause contamination and carryover, and affect results.
Clean this part to prevent contamination.
Recommended frequency: 6 Months at PM visits
Operator time: Approximately 1 minute.
Analyzer time: None.
Precautions: The operation switch must be OFF.
Materials Required Catalog Number
Gauze squares obtain locally
Distilled or deionized water obtain locally
70% isopropyl alcohol obtain locally
Procedure
1 ► Move the S/R probe to an area where you can readily access it.
2 ► Wipe the outer surfaces of the S/R probe and probe tip with a gauze square soaked in distilled
or deionized water.
3 ► If the probe appears dirty, wipe the outer surfaces with a gauze square soaked in 70%
isopropyl alcohol. Follow with a gauze square soaked in distilled or deionized water.
4 ► When you power ON the analyzer, it performs the start-up reset operation, and each
mechanism returns to its home or Stand-by position.
Procedure
9 ► Take FILTER NOZZLE part out from the water filter as shown in Figure 10.1-41 and clean it.
Procedure
7 ► Use a gauze square to wipe and clean the aspiration inlet for distilled or deionized water supply,
located on the analyzer.
8 ► Connect the valve on the container bottom and fill the container with distilled or deionized water.
Then, dry the outside of the container with paper towels, attach the cap to the container and
return the container to the analyzer.
A full liquid waste container causes an alarm and interrupts operation. The liquid waste container must be
checked and emptied as needed.
Recommended frequency: 6 Months at PM visits.
Operator time: Approximately 5 minutes.
Analyzer time: None.
Precautions: The analyzer must be in Stand-by or turn OFF the operation switch.
Do not execute System Reset or Reagent Scan, or power ON the
analyzer while cleaning the liquid waste container.
Materials Required Catalog Number
70% isopropyl alcohol obtain locally
Paper towels obtain locally
optional - germicidal agent (pH - obtain locally
9)
Procedure
1 ► Pull the liquid waste container toward you, cap it and raise it carefully, while avoiding the liquid
waste outlet.
2 ► Place a folded paper towel under the waste outlet to catch any waste droplets that may fall.
3 ► Empty the container and rinse it thoroughly with water.
4 ► If the inside of the container appears dirty, use 70% isopropyl alcohol to rinse the container.
Follow with a thorough water rinse.
5 ► Wipe the outside of the container with a paper towel.
6 ► Use a paper towel to wipe the compartment where the container is to be placed.
7 ► OPTIONAL - Add the appropriate volume of a germicidal agent with a pH - 9 (as directed in its
product labeling) to the liquid waste container.
8 ► Remove the paper towel under the waste outlet and replace the liquid waste container. Push
the container forward so that the container opening is under the liquid waste outlet. Be sure to
remove the cap.
The system reagent compartment should be cleaned as needed to eliminate spills from the ProCell and
CleanCell reagents.
Recommended frequency: 6 Months at PM visits.
Operator time: Approximately 5 minutes.
Analyzer time: None.
Precautions: The operation switch must be OFF.
Materials Required Catalog Number
Gauze squares obtain locally
Distilled or deionized water ------
Procedure
1 ► Open the sipper safety cover. Push on the part indicated in the picture below.
2 ► Move the sipper arm as far to the left as the arm will allow.
3 ► Remove the ProCell and CleanCell reagent bottles.
4 ► Wipe the inside of the compartments with damp, not wet, gauze squares. DO NOT allow water
to pool in the bottom of the compartments. Take care to avoid the photosensors in
compartment positions 2 and 3. These sensors check for the presence of the ProCell/CleanCell
bottle sets. They appear as rectangular windows located at the back of the compartment, just
below the top edge. If you should get the sensors wet, use a cotton swab to dry them.
Reagent spills should be cleaned up as they occur. The reagent disk and compartment must be cleaned
every 6 Months at PM visits.
Recommended frequency: 6 Months at PM visits.
Operator time: Approximately 15 minutes.
Analyzer time: None.
Precautions: The operation switch must be OFF.
Materials Required Catalog Number
Gauze squares obtain locally
Distilled or deionized water ------
70% isopropyl alcohol obtain locally
Cloth or lint-free towels obtain locally
5 ► Wipe the inside and outside of the reagent disk with gauze squares soaked with distilled or
deionized water.
6 ► If the disk appears dirty, use gauze squares soaked with 70% isopropyl alcohol to clean the
disk. Follow with gauze squares soaked with distilled or deionized water.
7 ► Dry the reagent disk with a cloth or lint-free towels. Set the reagent disk aside.
1 ► Wipe the reagent disk compartment with gauze squares soaked with distilled or deionized
water.
2 ► If the compartment appears dirty, use gauze squares soaked with 70% isopropyl alcohol to
clean the compartment. Follow with gauze squares soaked with distilled or deionized water.
3 ► Dry the reagent disk compartment with a cloth or lint-free towels.
4 ► Return the reagent disk to the compartment. The disk is keyed; make sure that the alignment
pin on the center plate (refer to photo above) is aligned with the hole on the disk.
5 ► Securely reinstall the thumbscrews.
6 ► Place the reagent packs back into the reagent disk.
7 ► Replace the reagent disk cover and lock.
Procedure
3 ► Remove COVER.
4 ► Show the drain tube as shown Figure 10.1-53.
5 ► Remove the drain tube of RINSE STATION P2 part as shown Figure 10.1-54.
6 ► Use gauze etc. so that the drop of water might leak from the tube.
10.1.14 Drain the outlet pipettor wash station and waste pipe
NOTICE
Read Safety Information
Recommended frequency: 6 Months at PM visits
Operator time: Approximately 5 minute.
Analyzer time: None.
Precautions: The operation switch must be OFF.
Materials Required Catalog Number
Gauze squares obtain locally
Distilled or deionized water obtain locally
70% isopropyl alcohol -------
SysClean obtain locally
Procedure
1 ► If the inside of the container appears dirty, use SysClean to rinse the drain the pipetter wash
station. Follow with a thorough water rinse.
Contamination in the sipper system could potentially degrade sample accuracy and precision, or possibly
block the measuring cell flowpath. To keep the sipper liquid flowpath clean and maintain the integrity of
the measuring cell, perform a liquid flow cleaning every 6 Months at PM visits.
High volume analyzers may require more frequent cleaning.
If customer maintenance has been done as recommended, it is only necessary to carry out an additional
customer cleaning at the PM visit. The service LFC is only necessary during PM if the sipper line is highly
contaminated or if LFC has not been carried out regularly by the customer.
Recommended frequency: 6 Months at PM visits.
Operator time: Approximately 12 minutes.
Pre-Cleaning Steps
4 ► Carefully, insert the filled SysClean adapter into position 3 of the system reagent compartment.
5 ► Remove the liquid waste container and thoroughly clean it with distilled or deionized water.
6 ► Verify ProCell/CleanCell bottles are in positions 1, 2 and 4. Also, verify the bottle lids are open
and that there is adequate liquid in the bottles in positions 1 and 2.
7 ► Close the sipper safety cover.
Initiate Cleaning
3 ► In the maintenance items list box (right), select “Liquid Flow Cleaning” and touch the Select
button. The Liquid Flow Cleaning pop-up window opens.
4 ► Type "1" for User / Type “3” for Service in the text box and press Start. The system begins
cleaning the sipper liquid flowpath.
5 ► Perform Measuring Cell Preparation cycles in case of Service LFC.[Press ”MC Preparation” in
Maintenance screen; select number of cycles]
Post-Cleaning Steps
1 ► When the analyzer returns to Stand-by, remove the liquid waste container and thoroughly rinse
it with distilled or deionized water. Return the liquid waste container to the analyzer.
2 ► Open the sipper safety cover. Remove the SysClean adapter and properly discard any
remaining SysClean. Thoroughly rinse the adapter.
3 ► Return the ProCell bottle to position 3 of the system reagent compartment.
4 ► Close the sipper safety cover.
Procedure
Procedure
Procedure
3 ► For the adjusting method, refer to the chapter 2.6.1 and 5.3.8.
Procedure
6 ► Adjust the Gripper according to Chapter 4.5.1. If positioning shifts, must execute the adjustment.
Procedure
1 ► Move the S/R arm to the far left and the gripper toward the front of the analyzer. Open the
sipper safety cover and move the sipper arm to the far right.
2 ► Clean the top of the incubator with gauze squares dampened with distilled or deionized water.
3 ► If the incubator appears dirty, use a slight scrubbing motion with the water-soaked gauze
squares. DO NOT use an acid solution or an alkaline solution to clean the incubator. The
incubator is made of aluminum and these solutions degrade the metal.
4 ► Next, wet a cotton swab with distilled or deionized water and swab each of the 32 positions on
the incubator, as well as the aspiration station.
5 ► Use a dry gauze square to dry the incubator when you are finished cleaning.
6 ► When you power ON the analyzer, it performs the start-up reset operation, and each
mechanism returns to its home or Stand-by position.
Procedure
Procedure
5 ► In case of Rack Sampler, wipe BCR window for Sampler with gauze also.
Procedure
3 ► Remove the measuring cell from CELL FRAME by removing the screw of the cell in two places..
4 ► Figure 10.1-73 shows the measuring cell removed from CELL FRAME
Procedure
Procedure
Procedure
6 ► Remove tube B on the side of DISPENSER SYRINGE ASSY as shown in Figure 10.1-85.
Procedure
3 ► Remove the valve on the bottom of the system water container if necessary.
4 ► Follow the reverse procedure of removing for installation.
Procedure
Figure 10.1-87 SV 1 / 2 / 5 / 6 / 7
5 ► Remove PIPING BLOCK from each SV’s and O-ring’s shown in Figure 10.1-90.
Procedure
3 ► Remove the screws putting sipper wash station and PLATE R together as shown in Figure
10.1-92.
5 ► Remove the screws of WASHING STATION S(M) ASSY as shown in Figure 10.1-94.
Procedure
3 ► Exchange the spring. If you cannot exchange the spring, then remove the Gripper Head for
safety and recommended it.
4 ► Adjust the Gripper according to Chapter 4.5.1. If positioning shifts, must execute the adjustment.
Table 10.3-2 Contents of Preventive Maintenance Kit P/N : 741-0926 Name : 2010/e411 2010 PM
KIT
No Part name P/N Qty.
1 Tube 465 741-1175 1
2 S NOZZLE SEAL 741-2065 1
3 MIXING BELT 3 741-2244 1
4 NOZZLE SEAL F729051 1
5 SPRING GRP 741-1168 1
6 SEAL PIECE S 741-1303 2 (total 2)
7 SEAL PIECE P 741-1302 2 (total 2)
8 WASHING STATION S 741-1380 1
9 O-RING NBR P9 L456006 2 (total 2)
10 O-RING NBR P16 L456013 2 (total 2)
11 VALVE BODY 741-1503 1
12 TUBE JOINT B2 741-1591 4 (total 4)
13 Tube Joint C 741-1609 1
14 TUBE SET IH 741-0941 4 (total 8)
15 PACKING L443085 4
Table 10.3-3 Additional Parts which has to be ordered for Preventive Maintenance ( Every 12 month )
No Part name P/N Qty.
1 MEASURING CELL 741-0461 1
2 VALVE BODY 741-1503 1
3 RINSE STATION P2 741-0997 1
4 O-RING NBR P4 L456001 5
5 O-RING NBR P5 L456002 5