Meterk MK6000 True RMS Clamp Meter

Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

Digital Clamp Meter

Manual

62
Catalogue
1. Safety information ……………1
1.1 Preparation ……………1
1.2 Use ……………1
1.3 Sign ……………2
1.4 Maintenance ……………2
2. Description ……………2
2.1 Part name ……………3
2.2 Switches and buttons ……………4
2.3 LCD Monitor ……………5
3. Specifications ……………7
3.1 Overview ……………8
3.2 Technical indicators ……………9
4. Operations Guide ……………10
4.1 Reading Hold ……………10
4.2 Frequency / duty cycle switch ……………10
4.3 The maximum / minimum value ……………11
measurement

4.4 Feature selection ……………11


RAN test
4.5 ……………11
4.6 Backlight source and clamp clip light ……………11
4.7 Auto power off ……………12
4.8 Buzzer ……………12
4.9 Measurement preparation ……………13
4.10 Current measurement ……………13
4.11 Voltage measurement ……………13
4.12 Frequency / duty ratio measurement ……………14
4.13 Resistance measurement ……………15
4.14 Diode test ……………16

63
4.15 Continuity test ……………16
4.16 Capacitance measurement ……………17
Temperature measurement
4.17 ……………17
Non-contact voltage detector NCV
5. ……………18
6. Maintenance ……………18
6.1 Replace the battery ……………18
6.2 Replace the test leads ……………18
7 Accessories ……………19

64
1. Safety Instructions

Warning
To avoid electrical shock or injury, Please read the
"Safety Information" and "Warning and related Notes" before
used.
In order to make full use of the function of the instrument and
to ensure the safety , please read and follow the use of this
specification in carefully.

This instrument strictly follow GB/T 13978-92 Generic


specification for Digital Clamp meter, In accordance with
GB4793.1-1995 ( IEC-61010-1, IEC-61010-2-032 ) Electronic
measuring instruments safety requirements,Belong to the two
class of pollution,Voltage standard CAT IV 600Vand CAT III
1000V。
Please follow the safety instructions, ensure the
safety of the use of the instrument.
use and protectthe item in right condition, the instrument
will ina satisfactory service.
1.1 Preparation
1.1.1 user must comply with the standards of safety rules
when use it:
- Protection against electric shock
- To prevent the misuse of the instrument
1.1.2 After receiving instrument, check whether the damaged
in transit.
1.1.3 Check and confirm the meter whether damaged or not
after shipment,
1.1.4 Test leads must be in good condition.Check the
insulation test is damaged, the wire conductor is bare
before use,
1.1.5 Use the test leads to ensure safety, it must be replaced
with the same or similar rank test if necessary.

65
1.2 Use
1.2.1 Use the correct function and range.
1.2.2 Do not exceed the scope of protection of the range of
the indicating value measurement. Do not exceed the
scope of protection of the range of the indicating value
measurement
1.2.3 When measuring circuit, do not touch the test lead tip
(metal parts).
1.2.4 In the measurement, if the measured voltage is higher
than 60V DC or 30V AC (RMS), attention should be paid
to keep your fingers always after the test finger
protection device.
1.2.5 If the voltage between the measuring end and the earth
more than AC 750V, don’t measure the voltage.
1.2.6 Before turning the switch changes the measurement
function, should be removed test test lead from the
circuit
1.2.7Don't live line measurement of resistance, capacitance,
diode and continuity test
1.2.8Under the test range off current, resistance,
capacitance, diode and continuity test, it should be
taken to avoid the instrument connected voltage source.
1.2.9Capacitor is fully discharge , don't test capacitance
1.2.10 Do not use this instrument in the gas, steam or dust
1.2.11If you notice any abnormal or faulty instruments,
should stop using
1.2.12 Unless the instrument bottom shell and the battery
cover is fastened in situ, it should not use the
instrument.
1.2.13Do not store or use the instrument in direct sunlight,
high temperature, high humidity conditions.
1.3Mark
Note (safety information, important see instructions)

66
Can be used for dangerous live conductor.
Double insulation protection(II)

CAT III In accordance with theIEC-61010-1Over voltage

standard level (installation) III、The pollution degree of

2 refers to the pulse voltage protection levels.

In line with European standards (EU)


Grounding
1.4 Maintenance
1.4.1 Please do not attempt to otest lead the shell to adjust
or repair instrument, this operation can only be fully
understood by technicians
1.4.2 Before otest leading the instrument bottom shell and
the battery cover, should be removed the test lead
from the measured line test lead
1.4.3 In order to avoid false readings may cause electric
shock, when the instrument display “ ” symbol,
The battery should be replaced immediately.
1.4.4 Use a damp cloth and a mild detergent to clean the
instrument, do not use abrasive detergents or solvents.
1.4.5 The instrument when not in use should turn off the power,
switch to the OFF position
1.4.6 If you do not use a meter long time, the battery should
be removed to prevent damage to the instrument.
2. Description
- The instrument is an professional measuring instrument
with liquid crystal display and a back light source.
The user readings in easily. Single hand operation to
facilitate the measurement range switch, with overload
protection and low battery indicator. For professional,
factory, school, lovers or family use, is an ideal
Multi-function instrument.

67
- Used for AC cur
rrent、DC current(only for CM-600AD)、True
RMS AC volt he DC voltage, frequency,
tage sine wave, th
duty ratio, resistance, capa
acitance measurem
ment and
the on-off circuit, measurem
ment, temperature
e
measurement
t .
- Automatic range
- Data keep
- Maximum measurement function
- Minimum measurement function
000A& CM-600AD)
- Peak measuring(only for CM-10
- Surge measurement(only for CM
M-1000A)
- Auto power off(only for CM-10
000A)
- Relative measurement

2.1 Part name


⑴ Current clamp head: for curr
rent measurement
⑵ The forceps head lamp
⑶ Panel
⑷ Wrench
⑸ Function selection keys(SEL
L)
⑹ Frequency / Duty Ratio/ REL button(REL/Hz%)
⑺ Surge current / select butto
on(RAN)
⑻ LCD Monitor
⑼ Common socket
⑽ Resistor, capacitor, diode,
, voltage, freque
ency and
Continuity input jack
⑾ Maximum/Minimum button(MAX/
/MIN)
⑿ Reading / backlight button(
( )
⒀ Switch

68
1

4 13

5
6 12
7 11

9 10

69


6ZLWFKHVD
DQGEXWWRQVLQV
VWUXFWLRQV
%XWWRQXVHGWRFRQWUROU LJKW
UHDGLQJRUEDFNOL
6(/EXWWRQXVHGIRUPHDVXULQJWKHVZLWFKLQJ
5$1EXWWRQ˖5DQJHVZLWFKLQJ
˖8VHGWRWKHIUHTXHQF\GXW\UDWLRPHDVXULQJ
5(/+]EXWWRQ˖
VZLWFK
0$;0,1EXWWRQ˖0HDVXULQJWKHSH
HDNYDOXHDQGVZLW
WFKLQJLW
2))˖8VHGWRWXUQRIIWKHSRZHU
U
ROWDJHUHVLVWDQFH
-DFNYR HFDSDFLWDQFHIUHTXHQF\
GXW\ RGH JDS OLQH DQG WKH
\ UDWLR WKH GLR
WHPS
SHUDWXUHRIWKHLQ
QSXWWHUPLQDO
&20 -DFN ˖ 9ROW
WDJH UHVLVWDQFH FDSDFLWDQFH IUHTXHQF\
GXW\ F\FOH &RQWLQXL
LW\ WHVW OLQH DQG WKH
WHPSH
HUDWXUHRIDFRPP
PRQWHUPLQDO
6ZLWFKXVHGWRVHOHFWWKHIXQ
QFWLRQDQGUDQJH
/&'0RQLWRU

70
AC、DC AC/DC

、 Diode test,Continuity test

AUTO Automatic range

P MAX-MIN PEAK HOLD

REL Relative measurement mode

℃ ℉ Temperature measuring state

LOW BATTERY

H Reading Hold

% %
mV, V mV, V
A Ampere (current)

nF,μF,mF nF,μF,mF
Ω, kΩ, MΩ Ω, kΩ, MΩ
Hz, kHz,MHZ Hz, kHz,MHZ(Frequency)

3. Specifications
The instrument shall specify the period of one year,
during in 18℃ ~ 28℃、The relative humidity is less than
75% under the conditions of re calibration.
3.1 Overview
Automatic measurement and manual measurement
Overload protection
The maximum voltage allowed between the measuring
terminal and earth:1000V DC or750V AC
Height:Maximum 2000m
Display:LCD
Maximum:5999 value

71
Polar indication : Automatic indicator , ‘ - ’ ,Display
negative
Over range display:‘0L’ or ‘-0L’。
Sampling time: about 3 times / sec
Display unit,:display unit and power display
Automatic power off:15Minutes

Power:DC current 4.5V

Battery type:1.5V AAA

Battery voltage indication:LCD Symbol


Temperature coefficient:less 0.1×Accuracy /℃
operating temperature:18℃ ~ 28℃
Storage temperature:-10℃ ~ 50℃
Size:208×78×35mm
weight:about 240g(Including battery)

3.2 Technical indicators


Environment temperature:23±5℃ Relative humidity:<75%

3.2.1 True RMS


3.2.1.1 Measurement of the sine wave signal, compared with
the traditional method , using the RMS average
measurement method is more better.
3.2.1.2 RMS meter can accurately measure the sine wave signal,
if use on AC function of files and did not input a
measured signal(AC function,Input in short circuit),
Clamp table may display a reading between 1 to 50 words.
These deviations readings are normal. In the
measurement of the specified ranges, they do not affect
the meter to measure the accuracy of the alternating
current

72
3.2.1.3 True RMS requires the input signal reaches a certain
level can be measured.
the range of AC voltage and current specified in the full
range of between 2% ~ 100%.
3.2.2 AC Current

Range Resolution Accuracy

6A 0.001A

60A 0.01A ±(2.5% + 10)

600A 0.1A

- maximum/ input current:1000A AC


- Frequency range:50Hz /60Hz
3.2.3 DC voltage

Range Resolution Accuracy

6V 0.001V ±(1.0% + 5)

60V 0.01V

600V 0.1V

- Input impedance:10MΩ
-Maximum/ input voltage:600V AC (Effective value)OR 600V
DC
NOTICE:
In the small voltage range, one has not received the test
circuit; the instrument may be beating reading.
It’s normal. this is due to the high sensitivity of the
instrument .
without detriment to the actual measurement results.
3.2.4 AC voltage

Range Resolution Accuracy

6V 0.001V

60V 0.01V ±(1.0% + 5)

73
600V 0.1V ±(1.2% + 5)

- Input impedance:10MΩ
- Maximum/ input voltage:600V AC (Effective value)or 600V
DC
- Frequency range:40 ~ 1000Hz
NOTICE:
In the small voltage range, one has not received the test
circuit, the instrument may be beating reading,
it's normal, this is due to the high sensitivity of the
instrument .
Without detriment to the actual measurement results.

3.2.5 Frequency
3.2.5.1 Through the V file:

Range Resolution Accuracy

99.99Hz 0.01Hz

999.9Hz 0.1Hz ±(1.5% + 5)

9.999kHz 0.001kHz

- Range:10Hz ~ 10kHz
- Input voltage range:≥ 2V AC(Effective value)(With
the increase of the measured frequency, the input voltage
should be increased)
- Input impedance:10MΩ
- Maximum/ input voltage:600V AC(Effective value)
3.2.5.2 Through A File (From Clamp)
- Frequency: 10-1KHZ
- Input current range: ≥20A AC(Effective value)
- Input Max current:AC 600A
- Input voltage range:≥ 2V(With the increase of the
measured frequency, the input voltage should be increased)
3.2. 6 Duty ratio

74
Range Resolution Accuracy

0.1 – 99.9% 0.1% ± 3.0%

3.2.6.1 Through A file(From the clamp head):


- Frequency response:10 ~ 1kHz
- Input current range:≥20A AC(Effective value)
- Maximum/ input current:AC 600A
3.2.6.2 Through the V file:
- Frequency response:10 ~ 10kHz
- Input voltage range:≥ 2V AC
- Input impedance:10MΩ
- Maximum /input voltage:750V AC(Effective value)
3.2.7 Resistance

Range Resolution Accuracy

600Ω 0.1Ω

6kΩ 0.001kΩ

60kΩ 0.01kΩ ±(0.8% + 3)

600kΩ 0.1kΩ

6MΩ 0.001MΩ ±(1.2% + 3)

60MΩ 0.1MΩ

- test lead circuit voltage:ABOUT 0.4V


- Overload protection:250V DC OR AC(Effective value)
3.2.8 Continuity test

Range Resolution Function

If the measured line


resistance is less than 50,
0. 1Ω
enclosing the instrument may
sound a buzzer

- Overload protection:250V DC or AC(Effective value)


3.2.9 Capacitance

Range Resolution Accuracy

75
76
a trigger action, when touch this button is in the
relative value measurement model, the current display
value as a reference value in the memory, after the
measurement, the display value as the input value minus
the difference between the reference value.
so REL△(current reading) = input value - reference value
3) The relative value of the measurement can be carried
out in a certain range that is to say under the measurement
mode to have this feature.

notice:

Measuring instrument in the Maximum / minimum state, it

cannot switch to the frequency, duty ratio measuring mode.

measurement.
4.3 The maximum value and minimum value measurement
1)Press“MAX MIN”enter the manual range mode,display Max
Value;Then press this button to display the "Min" value;
press again and display “Max-Peak Min";Repeat this
cycle. Press the "MAX MIN" button for more than 2 seconds
to exit this mode.
2)In Max mode, automatically saved the maximum value of
the measurement.
3)In Min mode, automatically saved the minimum value
measurement.
4)Enter the Max Min mode, automatically save the measured
difference
5)Press the "MAX MIN" button for more than 2 seconds to
exit this mode, the instrument is restored to normal
state measurement.
4.4 function exchange
1 ) In resistance state , press “ SEL ” button , In the
resistance, capacitance, diode on-off detecting,

77
four gear cycle switch.
2) In the voltage profile, click "SEL", switch between DC
and AC
3) In the temperature profile, click "SEL", switch between
in Celsius and Fahrenheit .
4.5 RAN measurement
1)RANE keys for automatic/manual switch range trigger this
action, boot or turn the dial, the default for the
automatic range. Click on the switch for manual range. IN
manual mode, the range per click this button to jump up
one notch, to the Max value and then click this button will
jump to the mix value in turn .Frequency and capacitance
measurement without manual range. Such as press RANGE for
more than 2 seconds or rotary switch, the exit RANGE
manually.

4.6 Backlight and lighting clamp head


1)In the process of measurement, if the ambient light is
too dark, difficulties to get resulting,press the
“ ”key more than 2seconds,Otest lead the back
light, automatically shut down in about 30 seconds.
2)If press“ ” key more than 2seconds and willturn
off the power.
3) In the current file,when otest lead the instrument
backlight will also otest lead the clamp head too,The
luminous body is LED,working current is large, while
the meter is provided with a timing circuit (timing
is about 30 seconds), but often use the backlight will
shorten the battery life, so unnecessary should
minimize the use of back light.
notice:
When the battery voltage≤3.9V,display
“ ”
(Undervoltage)
Symbol。Use the backlight,When the battery voltage≥3.9V,

78
Because the large working current, the battery voltage drop,
“ ”symbol may be displayed(“ ”symbol is displayed,
does not guarantee the accuracy of measurement),it cannot
replace the battery, use until without backlight conditions
in Normal “ ”Symbols show to replace.
4.7 Automatic shutdown
1)Without operation after 15 minutes ,The meter will enter
the sleep state automatically shut down to save power.
1 minutes before the shutdown, the buzzer 5 sound prompt,
dormant into a long sound after that before the shutdown.
2)After the automatic shutdown, press the SEL button, reset
instrument
3)If the boot and press "SEL" button in the same time,
it will cancel the automatic shutdown function.
4.8 Buzzer
Press any key or switch the rotation function, if the

function key is effective, the buzzer will be issued "Beep"

sound (about 0.25 seconds);set in measuring voltage or

current is greater than the alarm value,For example, AC

voltage greater than 600V DC voltage is greater than 600V,

AC / DC current is more than 600A, the buzzer will continue

as pronunciation and to be a overrange alarm; 自

Automatically shut down about 1 minutes before the buzzer

will sound a warning for the issue of 5 times, before the

shutdown will send a warning buzzer sound; ; Automatic

shutdown function is cancelled; every 15 minutes take 5

warnings.

4.9 Measurement
1)Toggle the switch and open power, If the battery and
voltage

79
Is not enough(about≤3.9V
The LCD display“ ”symbol then the battery should be
replaced
2) “ ”symbol,The input voltage or current should not
exceed the value indicated, this is to protect the
internal circuit from being damaged.
3)Switch measuring function and range in the required
4)Connect the public test line, and then connected test
line is When in wiring, The test should be removed first
line charged dismantling wiring.
4.10 Current measurement

Warning
An electric shock hazard.

In current clamp measurements before the test probe is


removed from the instrument.

1)Switch in the A position measuring range. At this time

for the measurement of AC current state, select the

appropriate range

2)Hold the trigger and otest lead the clamp,A wire line
is measured clip in the meter.
3)Read the current value in the LCD display
notice:
1)It cannot get the correct measurement results if use more
than two or much more wire way in the same time.
2)In order to obtain accurate readings,The measured wire
under the current clamp center as much as possible

3)“ ”show the Maximum input AC current is 600 A

4)Please press the current measurements and test lead the

clamp head .Use pliers head clip to be measured conductor,

80
additional error
r, A clamp meter can only measure a current

conductor, If measurement two or more conductors, readings

will be wrong.

4.11 Voltage measurement


m

War
rning
An electr
ric shock hazard.

When the measurement


m of hig
gh voltage, please pay special
attention to avoid
a electric sho
ock

Do not enter the effective value of voltage higher than


AC600V

1) Use black test


t lead is inserted
d into the COM soc
cket, the
red test lead is inserted into the
t socket
t, select
the appropria
ate range
2) Switch at the AC
A voltage to p
place,When the instrument
e measurement, to measure AC voltag
for DC voltage ge, press
SEL to enter the state of AC voltage
v measureme
ent
3) Use the test lead connected with
w the voltage source
s or
load to meas
surement
4 )Read the vol
ltage value on the
e LCD
notice:
1)In the small
l voltage range, th
he test leads didn
n't touch
the circuit e beat reading, this is
t line , LCD have
normal,;
this is due to the
t high sensitivity of the instrume
ent, when
the instrume
ent is connected to
o the circuit to be
e tested,
and will ge
et the real measur
red value.

81
4) If the inst
trument to measure the RMS reading
g of more
than 600V AC,
A issued a alarm
m

4.12 Frequency
y and Duty rati
io
1) voltage

War
rning
The risk
k of electric shock
When the measurement of
f high voltage, pay
p special
attention to avoid electric shock
s

Do not en
nter the effective
e value of voltage higher than
750 AC

⑴Use black test lead is inserted


d into the COM socket, the
red test lead is
s inserted into th
he socket
⑵Switch at the AC voltage to place,ress SEL to enter
the state of AC voltage measur
rement
⑶Press "REL/Hz/%
%" button to switch
h to the state of frequency
measurement
⑷Use the test lead
l connected wi
ith the voltage source or
load to meas
surement
⑸Read the volta
age value on the LCD
L
⑹Press "REL/Hz/%" can enter the duty
d ratio measuri
ing state
notice:
⑴ The frequency range is 10Hz ~ 10kHz , If the measured
ss than 10Hz and LCD display“00.0
frequency is les 0”;The
measurement freq
quencies below 10k
kHz is possible, but does
not guarantee th
he accuracy of mea
asurement
⑵ Duty ratio range
r is 10 ~ 95%。

82
⑶ “ ”Repre
esents the maximum
m input voltage is
s 600V AC
(RMS).
2) AC current :

War
rning
The risk
k of electric shock
Remove probe from the mete
er before measurem
ment current

⑴ Range switch in A current position. AC current


measurement cond
dition, choose the
e suitable range.
(2) Hold the trig
gger, open the clam
mp head, the circu
uit being
measured a wire clip within the clamp
c head
(3)Long time pres
ss REL/HZ/% button
n switch to Freque
ency test
position
⑷Reading in LC
CD
(5)Press the " REL /Hz/%" again
n can enter the du
uty ratio
measuring state
notice:
The frequency range
r is 10Hz ~ 10MHz, if the measured
frequency is low D displays "0"; measuring
wer than 10Hz, LCD
frequencies above 10MHz is possibl
le, it can't guara
antee the
accuracy of meas
surement。
4.13

Resistance tes
st

War
rning
The risk
k of electric shock
In the measurement of impedance on the line
e, should be
determined t
to disconnect t
the power supply circuit,
rcuit completely discharge.
capacitor cir

1) Use black test


t lead is inser
rted into the COM
M socket,
the red test lead is insert
ted into the socket

83
2)The range swi
itch on p
position, This instrument
for measuring re
esistance state
3)Use the test lead connected with
w the voltage source or
load to meas
surement
4)Display readi
ings on LCD
notice:
1) When the inp
put is Open circui
it, LCD will display "0L"
outrange state.
s
2)If the measur
red resistance value is higher than 1MΩ, the
instrument
t may take a few seconds to stable reading,
it is norm
mal for high resis
stivity readings
4.14 Diode tes
st
1) The black test
t lead is inserted
d into the COM soc
cket, the
red test lead is
s inserted into th
he socket
2) The range sw
witch on p
position,
3) press“SEL”button switch Test status
4)The red test lead is connected with
w a diode anode, cathode
diode connected to test black tes
st lead
5) display read
dings on LCD
notice:
1)The meter sh
how the approxima
ate value about th
he diode
forward voltage drop
2)If the test lead or test lead
d reverse connection open
circuit lead, LCD displays "0L".
4.15 continuit
ty test

War
rning
k of electric shock
The risk
In the test
t circuit, the circuit should be
e determined
to disconnect the power supp
ply, the total discharge of
capacitor cir
rcuit

84
1) Use black test
t lead is inserted
d into COM socket, the red
test lead is ins
serted into the socket
2) The range swi
itch on position
3)Press“SEL”bu
utton and switch continuity tes
st state
4) Test lead is measured
m at both en
nds of the line connection
5)If the resistan
nce measured line less than 50Ω , internal
buzzer may sound
d.
6) Read the resi
istance value on the
t LCD
notice:
If the test open circuit or the me
easured line resis
stance is
greater tha
an 600Ω, the disp
play shows "0L".

4.16 Capacitan
nce measurement
t

War
rning
The risk of
o electric shock
To avoid the electric sho
ock, it should be
b completely
discharge the capacitor before measurement of capacitance
c

1) Use black test


t lead is inserted
d into the COM soc
cket, the
red test lead is
s inserted into th
he socket

2) The range swi


itch position
3) Press "SEL" bu
utton to switch to the
t capacitor test
t status.
4) The capacitor is fully discharged and use the te
est leads
measure th
he capacitor
5)Read the capac
citance value in LCD
L
notice:
In order to impr
rove the measureme
ent accuracy of less than
10nF, sh
hould be distribut
ted capacitance meter
m and
wire less
4.17 Temperature measurement

War
rning
In the temperat
ture measurement of voltage higher
r than the DC

85
60V or AC 30V, avoid injury or damage caused by instrument

range Resolution Accuracy

-50℃-300℃ ±(1.0%+4d) 1℃

301℃-1000℃ ±(1.9%+5d) 1℃

-58℉-600℉ ±(1.2%+6d) 1℉

601℉-1832℉ ±(1.9%+6d) 1℉

1)The function of range switch for temperature

position(choose degrees Celsius or Fahrenheit)

2)The negative and positive K type thermocouple connected

to the COM socket and INPUT input hole

3)Put K thermocouple arranged inside the measured object

or environment

4)Read temperature on the LCD display

5. Non contact voltage detector NCV

Warning

Even if no indication but the voltage may still exist. Don't rely
on non contact voltage detector to determine whether there is
a voltage shielded wire. The operation may affect the detection
by socket design, insulation thickness and different types

placed in the NCV range position,display“EF”and“NCV”

symbol, When clamp head close to mains phase line or power

switch, socket the detected voltage is close to the 110V

(AC RMS),display “-”,When the induction voltage is

86
higher,display “-” will became much more ,With buzzer

alarm.(This instrument also set a clamp head AC induction

signal function, used to clamp head near line or by an

electric switch and electric socket. When using this

28function, don't insert test leads

Resolution Alarm Indicator

NCV>AC100V Buzzer/LED

6. Maintenance
6.1 Replace the battery

Warning
Before opening the instrument of the battery cover, please
removed the test leads from measuring circuit so that to avoid
the risk of electric shock.

1) if “ ”symbol display,please replace the battery。


2) Unscrew fastening screws of the battery cover and move
away.
3) replace the old battery
notice:
The battery polarity can not be reversed

6.2 Change Test Leads

Warning
Please use the same test leads or same level test leads

87
if you need replace the test leads . the test leads level:
1000V 10A.

If one damaged insulation, such as wire wire exposed, must


be replaced.
7. Accessories
1) Test lead level:1000V 10A one

2) Manual one

3) battery 1.5V AAA three

﹡The contents of this manual are subject to change without

notice

﹡The contents of this brochure is believed to be correct,

if users find errors, please contact the manufacturer.

﹡The company is not responsible for the accident and harm

caused by user wrong operation *

﹡This manual describes the function, not for other special

use

Warranty

Please show your product certification when you need repair

service or something wrong with items, it is effective show

one and purchase invoice in together

1 ) Please contact with our company or repair service

department when your clamp meter appear faults as soon as

possible, don't delay your use and warranty period.

2)Our company provide warranty service for one year from

the date of purchase。The company provides free warranty

service after professional confirm the problem is not by user

88
sabotage during the warranty period

3)Repair need charges (repairs components fee) more than

the warranty period

4)All of the following will charge the cost of repair even

in the warranty period.

⑴Improper use or accidental disasters caused damage of

components and circuit board burn

⑵Non UYIGAO Company professional personnel open shell,

check and modification

⑶ Does not follow the instructions to operation caused

problems⑷No maintenance and repair of the other company

products

⑸users offer the maintenance of postage and transportation

fees

⑹ Clamp meter’s battery, probe, temperature probe and

other functional accessories not included free warranty list