Meterk MK6000 True RMS Clamp Meter
Meterk MK6000 True RMS Clamp Meter
Meterk MK6000 True RMS Clamp Meter
Manual
62
Catalogue
1. Safety information ……………1
1.1 Preparation ……………1
1.2 Use ……………1
1.3 Sign ……………2
1.4 Maintenance ……………2
2. Description ……………2
2.1 Part name ……………3
2.2 Switches and buttons ……………4
2.3 LCD Monitor ……………5
3. Specifications ……………7
3.1 Overview ……………8
3.2 Technical indicators ……………9
4. Operations Guide ……………10
4.1 Reading Hold ……………10
4.2 Frequency / duty cycle switch ……………10
4.3 The maximum / minimum value ……………11
measurement
63
4.15 Continuity test ……………16
4.16 Capacitance measurement ……………17
Temperature measurement
4.17 ……………17
Non-contact voltage detector NCV
5. ……………18
6. Maintenance ……………18
6.1 Replace the battery ……………18
6.2 Replace the test leads ……………18
7 Accessories ……………19
64
1. Safety Instructions
Warning
To avoid electrical shock or injury, Please read the
"Safety Information" and "Warning and related Notes" before
used.
In order to make full use of the function of the instrument and
to ensure the safety , please read and follow the use of this
specification in carefully.
65
1.2 Use
1.2.1 Use the correct function and range.
1.2.2 Do not exceed the scope of protection of the range of
the indicating value measurement. Do not exceed the
scope of protection of the range of the indicating value
measurement
1.2.3 When measuring circuit, do not touch the test lead tip
(metal parts).
1.2.4 In the measurement, if the measured voltage is higher
than 60V DC or 30V AC (RMS), attention should be paid
to keep your fingers always after the test finger
protection device.
1.2.5 If the voltage between the measuring end and the earth
more than AC 750V, don’t measure the voltage.
1.2.6 Before turning the switch changes the measurement
function, should be removed test test lead from the
circuit
1.2.7Don't live line measurement of resistance, capacitance,
diode and continuity test
1.2.8Under the test range off current, resistance,
capacitance, diode and continuity test, it should be
taken to avoid the instrument connected voltage source.
1.2.9Capacitor is fully discharge , don't test capacitance
1.2.10 Do not use this instrument in the gas, steam or dust
1.2.11If you notice any abnormal or faulty instruments,
should stop using
1.2.12 Unless the instrument bottom shell and the battery
cover is fastened in situ, it should not use the
instrument.
1.2.13Do not store or use the instrument in direct sunlight,
high temperature, high humidity conditions.
1.3Mark
Note (safety information, important see instructions)
66
Can be used for dangerous live conductor.
Double insulation protection(II)
67
- Used for AC cur
rrent、DC current(only for CM-600AD)、True
RMS AC volt he DC voltage, frequency,
tage sine wave, th
duty ratio, resistance, capa
acitance measurem
ment and
the on-off circuit, measurem
ment, temperature
e
measurement
t .
- Automatic range
- Data keep
- Maximum measurement function
- Minimum measurement function
000A& CM-600AD)
- Peak measuring(only for CM-10
- Surge measurement(only for CM
M-1000A)
- Auto power off(only for CM-10
000A)
- Relative measurement
68
1
4 13
5
6 12
7 11
9 10
69
6ZLWFKHVD
DQGEXWWRQVLQV
VWUXFWLRQV
%XWWRQXVHGWRFRQWUROU LJKW
UHDGLQJRUEDFNOL
6(/EXWWRQXVHGIRUPHDVXULQJWKHVZLWFKLQJ
5$1EXWWRQ˖5DQJHVZLWFKLQJ
˖8VHGWRWKHIUHTXHQF\GXW\UDWLRPHDVXULQJ
5(/+]EXWWRQ˖
VZLWFK
0$;0,1EXWWRQ˖0HDVXULQJWKHSH
HDNYDOXHDQGVZLW
WFKLQJLW
2))˖8VHGWRWXUQRIIWKHSRZHU
U
ROWDJHUHVLVWDQFH
-DFNYR HFDSDFLWDQFHIUHTXHQF\
GXW\ RGH JDS OLQH DQG WKH
\ UDWLR WKH GLR
WHPS
SHUDWXUHRIWKHLQ
QSXWWHUPLQDO
&20 -DFN ˖ 9ROW
WDJH UHVLVWDQFH FDSDFLWDQFH IUHTXHQF\
GXW\ F\FOH &RQWLQXL
LW\ WHVW OLQH DQG WKH
WHPSH
HUDWXUHRIDFRPP
PRQWHUPLQDO
6ZLWFKXVHGWRVHOHFWWKHIXQ
QFWLRQDQGUDQJH
/&'0RQLWRU
70
AC、DC AC/DC
LOW BATTERY
H Reading Hold
% %
mV, V mV, V
A Ampere (current)
nF,μF,mF nF,μF,mF
Ω, kΩ, MΩ Ω, kΩ, MΩ
Hz, kHz,MHZ Hz, kHz,MHZ(Frequency)
3. Specifications
The instrument shall specify the period of one year,
during in 18℃ ~ 28℃、The relative humidity is less than
75% under the conditions of re calibration.
3.1 Overview
Automatic measurement and manual measurement
Overload protection
The maximum voltage allowed between the measuring
terminal and earth:1000V DC or750V AC
Height:Maximum 2000m
Display:LCD
Maximum:5999 value
71
Polar indication : Automatic indicator , ‘ - ’ ,Display
negative
Over range display:‘0L’ or ‘-0L’。
Sampling time: about 3 times / sec
Display unit,:display unit and power display
Automatic power off:15Minutes
72
3.2.1.3 True RMS requires the input signal reaches a certain
level can be measured.
the range of AC voltage and current specified in the full
range of between 2% ~ 100%.
3.2.2 AC Current
6A 0.001A
600A 0.1A
6V 0.001V ±(1.0% + 5)
60V 0.01V
600V 0.1V
- Input impedance:10MΩ
-Maximum/ input voltage:600V AC (Effective value)OR 600V
DC
NOTICE:
In the small voltage range, one has not received the test
circuit; the instrument may be beating reading.
It’s normal. this is due to the high sensitivity of the
instrument .
without detriment to the actual measurement results.
3.2.4 AC voltage
6V 0.001V
73
600V 0.1V ±(1.2% + 5)
- Input impedance:10MΩ
- Maximum/ input voltage:600V AC (Effective value)or 600V
DC
- Frequency range:40 ~ 1000Hz
NOTICE:
In the small voltage range, one has not received the test
circuit, the instrument may be beating reading,
it's normal, this is due to the high sensitivity of the
instrument .
Without detriment to the actual measurement results.
3.2.5 Frequency
3.2.5.1 Through the V file:
99.99Hz 0.01Hz
9.999kHz 0.001kHz
- Range:10Hz ~ 10kHz
- Input voltage range:≥ 2V AC(Effective value)(With
the increase of the measured frequency, the input voltage
should be increased)
- Input impedance:10MΩ
- Maximum/ input voltage:600V AC(Effective value)
3.2.5.2 Through A File (From Clamp)
- Frequency: 10-1KHZ
- Input current range: ≥20A AC(Effective value)
- Input Max current:AC 600A
- Input voltage range:≥ 2V(With the increase of the
measured frequency, the input voltage should be increased)
3.2. 6 Duty ratio
74
Range Resolution Accuracy
600Ω 0.1Ω
6kΩ 0.001kΩ
600kΩ 0.1kΩ
60MΩ 0.1MΩ
75
76
a trigger action, when touch this button is in the
relative value measurement model, the current display
value as a reference value in the memory, after the
measurement, the display value as the input value minus
the difference between the reference value.
so REL△(current reading) = input value - reference value
3) The relative value of the measurement can be carried
out in a certain range that is to say under the measurement
mode to have this feature.
notice:
measurement.
4.3 The maximum value and minimum value measurement
1)Press“MAX MIN”enter the manual range mode,display Max
Value;Then press this button to display the "Min" value;
press again and display “Max-Peak Min";Repeat this
cycle. Press the "MAX MIN" button for more than 2 seconds
to exit this mode.
2)In Max mode, automatically saved the maximum value of
the measurement.
3)In Min mode, automatically saved the minimum value
measurement.
4)Enter the Max Min mode, automatically save the measured
difference
5)Press the "MAX MIN" button for more than 2 seconds to
exit this mode, the instrument is restored to normal
state measurement.
4.4 function exchange
1 ) In resistance state , press “ SEL ” button , In the
resistance, capacitance, diode on-off detecting,
77
four gear cycle switch.
2) In the voltage profile, click "SEL", switch between DC
and AC
3) In the temperature profile, click "SEL", switch between
in Celsius and Fahrenheit .
4.5 RAN measurement
1)RANE keys for automatic/manual switch range trigger this
action, boot or turn the dial, the default for the
automatic range. Click on the switch for manual range. IN
manual mode, the range per click this button to jump up
one notch, to the Max value and then click this button will
jump to the mix value in turn .Frequency and capacitance
measurement without manual range. Such as press RANGE for
more than 2 seconds or rotary switch, the exit RANGE
manually.
78
Because the large working current, the battery voltage drop,
“ ”symbol may be displayed(“ ”symbol is displayed,
does not guarantee the accuracy of measurement),it cannot
replace the battery, use until without backlight conditions
in Normal “ ”Symbols show to replace.
4.7 Automatic shutdown
1)Without operation after 15 minutes ,The meter will enter
the sleep state automatically shut down to save power.
1 minutes before the shutdown, the buzzer 5 sound prompt,
dormant into a long sound after that before the shutdown.
2)After the automatic shutdown, press the SEL button, reset
instrument
3)If the boot and press "SEL" button in the same time,
it will cancel the automatic shutdown function.
4.8 Buzzer
Press any key or switch the rotation function, if the
warnings.
4.9 Measurement
1)Toggle the switch and open power, If the battery and
voltage
79
Is not enough(about≤3.9V
The LCD display“ ”symbol then the battery should be
replaced
2) “ ”symbol,The input voltage or current should not
exceed the value indicated, this is to protect the
internal circuit from being damaged.
3)Switch measuring function and range in the required
4)Connect the public test line, and then connected test
line is When in wiring, The test should be removed first
line charged dismantling wiring.
4.10 Current measurement
Warning
An electric shock hazard.
appropriate range
2)Hold the trigger and otest lead the clamp,A wire line
is measured clip in the meter.
3)Read the current value in the LCD display
notice:
1)It cannot get the correct measurement results if use more
than two or much more wire way in the same time.
2)In order to obtain accurate readings,The measured wire
under the current clamp center as much as possible
80
additional error
r, A clamp meter can only measure a current
will be wrong.
War
rning
An electr
ric shock hazard.
81
4) If the inst
trument to measure the RMS reading
g of more
than 600V AC,
A issued a alarm
m
4.12 Frequency
y and Duty rati
io
1) voltage
War
rning
The risk
k of electric shock
When the measurement of
f high voltage, pay
p special
attention to avoid electric shock
s
Do not en
nter the effective
e value of voltage higher than
750 AC
82
⑶ “ ”Repre
esents the maximum
m input voltage is
s 600V AC
(RMS).
2) AC current :
War
rning
The risk
k of electric shock
Remove probe from the mete
er before measurem
ment current
Resistance tes
st
War
rning
The risk
k of electric shock
In the measurement of impedance on the line
e, should be
determined t
to disconnect t
the power supply circuit,
rcuit completely discharge.
capacitor cir
83
2)The range swi
itch on p
position, This instrument
for measuring re
esistance state
3)Use the test lead connected with
w the voltage source or
load to meas
surement
4)Display readi
ings on LCD
notice:
1) When the inp
put is Open circui
it, LCD will display "0L"
outrange state.
s
2)If the measur
red resistance value is higher than 1MΩ, the
instrument
t may take a few seconds to stable reading,
it is norm
mal for high resis
stivity readings
4.14 Diode tes
st
1) The black test
t lead is inserted
d into the COM soc
cket, the
red test lead is
s inserted into th
he socket
2) The range sw
witch on p
position,
3) press“SEL”button switch Test status
4)The red test lead is connected with
w a diode anode, cathode
diode connected to test black tes
st lead
5) display read
dings on LCD
notice:
1)The meter sh
how the approxima
ate value about th
he diode
forward voltage drop
2)If the test lead or test lead
d reverse connection open
circuit lead, LCD displays "0L".
4.15 continuit
ty test
War
rning
k of electric shock
The risk
In the test
t circuit, the circuit should be
e determined
to disconnect the power supp
ply, the total discharge of
capacitor cir
rcuit
84
1) Use black test
t lead is inserted
d into COM socket, the red
test lead is ins
serted into the socket
2) The range swi
itch on position
3)Press“SEL”bu
utton and switch continuity tes
st state
4) Test lead is measured
m at both en
nds of the line connection
5)If the resistan
nce measured line less than 50Ω , internal
buzzer may sound
d.
6) Read the resi
istance value on the
t LCD
notice:
If the test open circuit or the me
easured line resis
stance is
greater tha
an 600Ω, the disp
play shows "0L".
4.16 Capacitan
nce measurement
t
War
rning
The risk of
o electric shock
To avoid the electric sho
ock, it should be
b completely
discharge the capacitor before measurement of capacitance
c
War
rning
In the temperat
ture measurement of voltage higher
r than the DC
85
60V or AC 30V, avoid injury or damage caused by instrument
-50℃-300℃ ±(1.0%+4d) 1℃
301℃-1000℃ ±(1.9%+5d) 1℃
-58℉-600℉ ±(1.2%+6d) 1℉
601℉-1832℉ ±(1.9%+6d) 1℉
or environment
Warning
Even if no indication but the voltage may still exist. Don't rely
on non contact voltage detector to determine whether there is
a voltage shielded wire. The operation may affect the detection
by socket design, insulation thickness and different types
86
higher,display “-” will became much more ,With buzzer
NCV>AC100V Buzzer/LED
6. Maintenance
6.1 Replace the battery
Warning
Before opening the instrument of the battery cover, please
removed the test leads from measuring circuit so that to avoid
the risk of electric shock.
Warning
Please use the same test leads or same level test leads
87
if you need replace the test leads . the test leads level:
1000V 10A.
2) Manual one
notice
use
Warranty
88
sabotage during the warranty period
products
fees
89