Single Analyte Meter - COD LR
Single Analyte Meter - COD LR
Single Analyte Meter - COD LR
COD Photometer
A-7320
0 to 150
PPM (mg/Liter)
Battery Replacement
Operating Tips
• Upon startup, the photometer automatically proceeds to the zeroing process.
Every time the photometer powers on, it must be re-zeroed.
Battery
Compartment
• To re-zero the photometer, it must be turned off and back on again.
Cover (B) • A series of readings can be taken without re-zeroing, as long as the photome-
ter stays on during the series.
Seal Ring (A)
• When the vial is appropriately seated in the sample compartment, the cap is
almost touching the top edge of the cell adapter.
• Protect photometer from extreme humidity, corrosive fumes and dusty areas.
Store in a cool, dry place.
Instrument • Remove the batteries when the photometer is not in use.
Back Screws • Press the ! key to turn the display back light on or off.
• When moving the photometer from one temperature extreme to another, wait
at least 10 minutes before use to allow photometer to come to temperature
Batteries equilibrium.
• To avoid errors caused by stray light, do not use the instrument in bright
sunlight.
• Contamination of the optics in the sample chamber will result in incorrect
measurements. The windows in the sample chamber should be checked at
Notch regular intervals and cleaned as necessary. Use a soft moist cloth or cotton
swab for cleaning purposes.
To ensure that the instrument is waterproof: Displays and Troubleshooting
• seal ring (A) must be in position E01: Light absorption too great (dirty optics)
• battery compartment cover (B) must be fixed with the four screws E20/E21: Too much light reaching detector
E22 or Battery Icon: Battery should be replaced
Cell Adapter Installation E27/E28/E29: Instrument zeroed incorrectly, misaligned adapter, vial not
Insert the cell adapter into the instrument’s sample chamber by aligning the arrow properly seated, dirty optics or failing light source.
on the adapter with the arrow on the instrument. Be sure that the adapter seats Hi/E03: Measuring range exceeded or excessive turbidity
completely into the sample chamber. Lo: Test result has negative value (less than 0 ppm)
www.chemetrics.com
4295 Catlett Road, Midland, VA 22728 U.S.A.
Phone: (800) 356-3072; Fax: (540) 788-4856
E-Mail: orders@chemetrics.com
Jan. 18, Rev. 8