Testo 535 Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

testo 535

Instruction manual en
Mains connection (mains unit no. 0554.0088)
recommended for long-term measuring.

The probe should be held as far as


possible from your body so as to
2 line display eliminate the influences caused by the
Hold/Max/Min CO2 in your breath.
MEAN button
Auto-Off function
START/STOP
ON/OFF If the Auto-Off function is activated
button the instrument switches itself off auto-
Screw-off matically after 10 minutes of inactivity
battery (except in the case of HOLD) and
PRINT button timed mean calculation.
compartment
The conformity certificate
confirms that this product Changing the battery
fulfills the guidelines in accor- If“Bat“ appears in the display, the
dance with 2004/108/EEC. battery will last approx. 1 hour
out of a total lifetime of 4 hours, if
referring to an alkali-manganese
Please read prior to operation battery. The instrument will switch
off automatically if there is not enough power.
- Never measure on live parts Unscrew the battery compartment, remove
- Observe sensor measuring ranges! the used battery and put in new battery, type
Overheating may destroy probes. 9V (IEC 6 F 22). (Observe polarisation). User
- Observe storage and transport temperature as settings remain intact if the instrument is not
well as the max. operating temperature longer than 3 minutes without power.
(e.g. protect measuring instrument from direct
sunlight). Instructions for disposal:
- Inexpert handling or use of force cancel your Only empty batteries should be disposed. The
warranty claims! batteries should be placed in plastic bags to
- If there is a change in the ambient temperature prevent short-circuits.
(change of location, e.g. inner/outer
temperature) the measuring instrument will Technical data - testo 535
need an adaptation phase of several minutes. Sensor: 2 channel infrared
- The CO2 probe contains optical components. absorption principle
Please treat the probe like your camera. Measuring range: 0 to 9,999 ppm CO2
Strong vibrationa may chenge the manufac- (0 to 0,999 vol.% CO2)
turer´s calibration. Check the readings in Accuracy: 0 to 5000 ppm:
fresh air: 350 to 450 ppm. It is recommended (at 23 °C *) ± (75 ppm +3 % of m.v.)
that the instrument is sent to a Testo service 5000 to 9,999 ppm:
point for adjustment purpose. ±(150 ppm +5 % of m.v.)
- The CO2 probe is permanently attached to Resolution: 1 ppm or 0.001 vol.%
the measuring instrument. Probes can only be Ambient
changed at Testo service points. temperature: 0 to +50°C
- Ensure the probe does not come into frequent
contact with water otherwise the long-term Storage temp.: -20 to +50 °C
stability will be affected. Battery lifetime: >4h (9 V block, Al-Mn)
- If the level of high frequency is high, there may Switchover: ppm / vol. %
be deviation from the required
Display: LCD (11 mm digit height)
Housing: ABS
Operation
Switch on instrument. After the Dimensions: 190 x 57 x 42 mm
instrument is switched on a short Weight: Approx. 300 g
display and function test follows and Ambient humidity: 0 to 99 %RH
the sensor has a warm-up phase of approx. 30 not subjected to
s. The instrument is then ready to operate. The
CO2 concentration in the sensor needs approx. condensation
1 minute to adapt to ambient conditions. This
adaptation time can be shortened by swinging Warranty
the probe gently back and forth. If the probe is testo 535 meas. instr................2 years
subjected to condensation, this can lead to Probe.........................................1 year
uncreased mesured values.
2 * Temp. coefficient: ± (1 ppm + 0,4 % of meas.) / °C
Changing the settings - Configuration
“START/STOP” button: keep pressed when switching on.
The following settings can be changed:
Date Time Air pressure

01:01 1013
Auto ON BAT 00:00 ppm



1998 hPa
Auto-OFF
function Battery in
activated instrument

Next Next Next


position position position

Meas-
▲ ▲ ▲ ▲ ▲ uring
Auto OFF ACCU 01:01
00:00 Vol%
1013 mode
Auto-OFF Rechargeable 1998 hPa
function battery in inst-
not activated rument

▲ ▲ ▲

Set value Set value Set value

Pressure compensation
Please note that the CO2 reading depends on the absolute air pressure.
This effect is compensated in the instrument. To do this, enter the correct air pressure (See
Changing the settings - Configuration). The air pressure depends on the height of the location above
mean sea level (See the “Barometric height formula” table on the next page) and weather influences.
Example:
You are located at a height of 800 m above sea level where you have a mean annual air pressure of
920 hPa. According to the barometer (1003 hPa) and the height correction of the barometer (to 1013
hPa) the mean annual air pressure has to be reduced by 10 (910 hPa air pressure).

Memory function for current, maximum and minimum values

The following displays appear when this button is pressed again and again:

1st press: HOLD - Readings are frozen in display


2nd press: MAX - Displays saved max. values
3rd press: MIN - Displays saved min. values
4th press: MEAN - Multi-point mean calculation
5th press: MEAN - Timed mean calculation
6th press: MEAN- Return to measuring mode

Important note:
The min/max values are measured from the time the instrument is switched on. The values are reset if
the instrument is switched off and then on again.

Printout
A printout of the reading with date and time can be activated by pressing the PRINT button.
The max, min and mean values also appear on the printout of a multi-point or timed mean cal-
culation. If a reading was frozen at a certain point in the display via the HOLD button, this rea-
ding appears in the printout with the time and date when the reading was frozen.

3
Mean calculation

4x

Multi-point mean calculation


▲ ▲ ▲ ▲ ▲ ▲ ▲ ▲

1310 1310 MEAN


1450 MEAN
MEAN

0 Save required readings by 037 Call up reading by pressing button. 037


pressing button.
This procedure can be repeated 999 times.
The number of readings saved is counted in the display.

Timed mean calculation


▲ ▲ ▲ ▲ ▲ ▲
▲ ▲

1310 MEAN
1310 1450 MEAN
MEAN

00:00 The clock starts to run in the display. Once 1:25 The mean of the readings are taken over
1:25
the measurement has run for as long as
required, press the Start/Stop button. a certain period of time. The mean and
The duration of the measurement can be duration are shown.
extended by further Start/Stop intervals
lasting up to a total of 99 hours and 59
minutes.

Return to measuring mode


Table: barometric height formula
Height above Air pressure Height above Air pressure Height above Air pressure Height above Air pressure
mean sea level (hPa) mean sea level (hPa) mean sea level (hPa) mean sea level (hPa)

4
Error messages

Error message Cause Remedy


in display

Probe out of order Please contact


the Testo
service point

Measuring range Measuring range


is exceeded. is exceeded.
The readings Does not damage
lie outside the probe.
maximum measuring range

Ordering data

testo 535 measuring instrument and accessories Part no.

testo 535, CO2 measuring instrument for measuring the quality of ambient air,
with securely attached probe, instruction manual, batteries
and calibration protocol 0560.5350

TopSafe (indestructible protection case) with bench stand and


belt clip, to protect instrument from impact, dirt... 0516.0183

9 V rechargeable battery, instead of battery 0515.0025

Recharger for 9 V rechargable battery, for external recharging of 0515.0025 battery 0554.0025

Plug-in mains unit for mains operation 0554.1143

Plastic case for storing measuring instrument, accessories, probes


Testo log printer and accessories 0516.3250

Printer and accessories Part no.

Testo log printer, prints measured data with date and time
0554.0545

Recharger with 4 NC rechargeable batteries for the Testo log printer,


rechargeable batteries are recharged externally 0554.0110

Printer paper for Testo log printer (6 rolls) 0554.0569