Multi-Frequency and Multi-Component Tympanometry
Multi-Frequency and Multi-Component Tympanometry
Multi-Frequency and Multi-Component Tympanometry
INTRODUCTION
Middle ear admittance can be measured at multiple frequencies, and has typically been measured at
226 Hz, 678 Hz, and 1000 Hz. Tympanometry using 226 Hz probe tones was discussed in Chapter 10;
in this chapter we will focus on tympanometry using 678 Hz and 1000 Hz probe tones. In addition,
All rights reserved. May not be reproduced in any form without permission from the publisher, except fair uses permitted under U.S. or applicable copyright law.
total middle ear admittance (Y) can be broken down into the contributions of mass and stiffness (B)
and conductance (G). The sensitivity and specificity of multi-frequency and multi-component
tympanometry can, in some cases, exceed that of conventional 226 Hz tympanometry (Harris,
Hutchinson, & Moravec, 2005).
In this chapter we discuss how to conduct and interpret multi-frequency and multi-component
tympanometry. More specifically, the following tests are covered in this chapter:
• Y1000 Hz tympanometry
• B/G 678 and/or B/G1000 Hz tympanometry
• Resonant frequency evaluation
Y1000 HZ TYMPANOMETRY
Researchers have shown that for infants, Y1000 Hz tympanometry gives more reliable and valid
findings than Y226 Hz tympanometry (Shanks & Shohet, 2009). That tympanometry for adults and
infants differs should not be surprising considering that infants’ ears differ from adults’ in many ways,
including immature anatomy, and at times the presence of middle ear mesenchyme. Accordingly, the
American Academy of Family Practice guidelines (Rosenfeld et al., 2004) suggested the use of Y1000
Hz up to age 4 months, after which Y226 Hz is appropriate.
Procedure
The procedure is the same as for Y226 Hz tympanometry, except 1000 Hz probe tones are
recommended.
EBSCO Publishing : eBook Collection (EBSCOhost) - printed on 4/9/2021 9:05 AM via UNIVERSIDAD DEL ROSARIO
AN: 1933799 ; Steiger, James, Miller, Erin L..; Diagnostic Audiology Pocket Guide : Evaluation of Hearing, Tinnitus, and Middle Ear Function
Account: s4300083.main.eds 74
Figure 11–1. Neonate normal Y1000 Hz tympanogram.
Measurement Components
In addition, admittance Y can be broken down into component parts, susceptance B and conductance
G. Conductance has to do with how middle-ear friction affects admittance, whereas susceptance has to
do with how middle-ear mass and stiffness affects admittance. Though the details are beyond the scope
of this text, suffice it to say that in mass-controlled middle ears susceptance, admittance lags sound
pressure because of middle-ear inertia. In contrast, in stiffness-controlled middle ears, susceptance
EBSCOhost - printed on 4/9/2021 9:05 AM via UNIVERSIDAD DEL ROSARIO. All use subject to https://www.ebsco.com/terms-of-use 75
admittance leads sound pressure, much as a spring initially admits energy before tightening. Therefore,
mass and stiffness susceptance are in opposition, allowing for scientific plotting of admittance as a
vector.
Procedure
Select desired frequency and components; for example, the Grason Stadler Tympstar can be set to test
both B and G by selecting B/G. The frequency can be set either to 678 Hz or 1000 Hz. Recording Y678
and Y1000 Hz are optional.
EBSCOhost - printed on 4/9/2021 9:05 AM via UNIVERSIDAD DEL ROSARIO. All use subject to https://www.ebsco.com/terms-of-use 76
Figure 11–2. Four multicomponent tympanogram shapes. From Acoustic immittance measures (Page 70, Figure 5–5) by
L. Hunter & N. Shahnaz, 2014, San Diego, CA: Plural Publishing, Inc. Reprinted with permission.
EBSCOhost - printed on 4/9/2021 9:05 AM via UNIVERSIDAD DEL ROSARIO. All use subject to https://www.ebsco.com/terms-of-use 77
Table 11–1. High Frequency Y, B, and G Shapes (Vanhuyse, Creten, & Van Camp, 1975; Fowler & Shanks, 2002)
EBSCOhost - printed on 4/9/2021 9:05 AM via UNIVERSIDAD DEL ROSARIO. All use subject to https://www.ebsco.com/terms-of-use 78
Simplified clinical decision rules are offered below:
EBSCOhost - printed on 4/9/2021 9:05 AM via UNIVERSIDAD DEL ROSARIO. All use subject to https://www.ebsco.com/terms-of-use 79
• Number of peaks rules: <5B peaks and/or <3G peaks are seen with normal middle ears (Harris, Hutchinson, &
Moravec, 2005); in fact over 90% of normal middle ears show 1B1G or 3B1G patterns (Wiley, Oviatt, & Block, 1987).
More peaks suggest mass affect, which can result from many disorders.
• Reliability rule: Middle ears should show no increase in the number of peaks at annual or other re-tests. Change is
abnormal, even if the retest results are <5B peaks and/or <3G peaks.
• The 3B3G tympanograms with outermost maxima separated by <75 daPa is seen with normal middle ears or with
monomeric/dimeric tympanic membranes. Wider maxima separation is seen with ossicular disarticulation. Results
from both monomeric/dimeric tympanic membranes and ossicular disarticulations will also show high static
admittance.
• The 5B3G tympanograms with outermost maxima separated by <100 daPa is seen with normal middle ears or with
monomeric/dimeric tympanic membranes. Wider maxima separation is seen with ossicular disarticulation. Results
from both monomeric/dimeric tympanic membranes and ossicular disarticulations will also show high static
admittance.
• Flat with B > G is seen with outer ear or middle ear occlusion.
• Flat with G < B is seen with tympanic membrane perforation or PE tube.
Procedure
Procedures are equipment specific, consult your equipment manual. For example, the Grason Stadler
Tympstar measures middle ear resonance using the multi-frequency procedure based on (Funasaka,
Funai, & Kumakawa, 1984; Funasaka & Kumakawa, 1988) in which tympanograms are measured in
50 Hz probe tone intervals. As probe tone frequency approaches a normal middle ears’ resonant
frequency, tympanograms manifest peaks, and at the resonant peak the middle-most peak is neither
higher nor lower than the tympanogram tails.
CAVEATS
Several disorders can cause either a mass effect or a stiffness effect. Therefore it would be a mistake to
expect that a multi-peak mass pattern necessarily indicates a “mass” or “growth.” For example, otitis
media with effusion can manifest as either a mass pattern or a stiffness pattern, perhaps depending on
the amount and viscosity of the effusion. Finally, as is always true with tympanometry, different
disorders might manifest with similar immittance results, and lateral pathologies obscure medial
pathologies.
EXHORTATION
We find many audiologists balk at the notion of conducting multi-frequency tympanometry, even as
EBSCOhost - printed on 4/9/2021 9:05 AM via UNIVERSIDAD DEL ROSARIO. All use subject to https://www.ebsco.com/terms-of-use 80
they accept conducting many other tests at multiple frequencies, including air conduction thresholds,
acoustic reflex thresholds, acoustic reflex decay, auditory brainstem response, and VNG/rotary chair.
Moreover, we have heard audiologists say they would conduct tympanograms at frequencies other than
226 Hz if shown that testing at those frequencies is better. Evidently, only tympanometry is subject to
that requirement. We urge audiologists to consider conducting high frequency tympanometry in
addition to low-frequency tympanometry, not because it is better but because it is different;
low-frequency tympanometry is more sensitive to changes in stiffness, whereas high-frequency
tympanometry is more sensitive to changes in mass. The artful use of a tympanometric test battery can
result in improved sensitivity and/or sensitivity.
REFERENCES
Bilgen, C., Kirkim, G., & Kirazli, T. (2009). Middle ear impedance measurements in large vestibular aqueduct syndrome.
Auris Nasus Larynx, 36(3), 263268. http://dx.doi.org/10.1016/j.anl.2008.07.002
Fowler, C. G., & Shanks, J. E. (2002). Tympanometry. In J. Katz (Ed.), Handbook of clinical audiology (5th ed.,
pp.175202). Baltimore, MD: Lippincott Williams & Wilkins.
Funasaka, S., Funai, H., & Kumakawa, K. (1984). Sweep-frequency tympanometry: Its development and diagnostic
value. Audiology, 23, 366379.
Funasaka, S., & Kumakawa, K. (1988). Tympanometry using a sweep-frequency probe tone and its clinical evaluation.
Audiology, 27, 99108.
Harris, P.K., Hutchinson, K. M., & Moravec, J. (2005). The use of tympanometry and pneumatic otoscopy for predicting
middle ear disease. American Journal of Audiology, 14, 3–13.
Hunter, L. L., & Margolis, R. H. (1992). Multifrequency tympanometry. American Journal of Audiology, 1, 3343. http:/
/dx.doi.org/10.1044/1059-0889.0103.33
Lai, D., Li, W., Xian, J., & Liu, S. (2008). Multifrequency tympanometry in adults with otitis media with effusion.
European Archives of Oto-Rhino-Laryngology, 265, 10211025. http://dx.doi.org/10.1007/s00405-008-0705-x
Lilly, D. (2005, April 12). The evolution of aural acousticimmittance measurements. The ASHA Leader, 6, 24.
Mimura, T., Sato, E., Sugiura, M., Yoshino, T., Naganawa, S., & Nakashima, T. (2005). Hearing loss in patients with
enlarged vestibular aqueduct: Air-bone gap and audiological Bing test. International Journal of Audiology, 44(8),
466469.
Rosenfeld, R. M., Culpepper, L., Doyle, K., Grundfast, K. M., Hoberman, A., Kenna, M. A., . . . Yawn, B. (2004).
Clinical practice guideline: Otitis media with effusion, Otolaryngology-Head and Neck Surgery, 130(5) S95S118. http:
//dx.doi.org/10.1016/j.otohns.2004.02.002
Rosowski, J. J., Songer, J. E., Nakajima, H. H., Brinsko, K. M., & Merchant, S. N. (2004). Clinical, experimental and
theoretical investigations of the effect of superior semicircular canal dehiscence on hearing mechanisms. Otology and
Neurotology, 25(3), 323332.
Sato, D., Nakashima, T., Lilly, D.J., Fausti, A., & Ueda, H. (2002). Tympanometric findings in patients with enlarged
vestibular aqueducts. Laryngoscope, 112, 16421646.
Shanks, J., & Shohet, J. (2009). Tympanometry in clinical practice. In J. Katz (Ed.), Handbook of clinical audiology. (6th
ed., pp. 189221). Baltimore, MD: Lippincott Williams & Wilkins.
Vanhuyse, V., Creten, W., & Van Camp, K. (1975). On the W-notching of tympanograms. Scandinavian Audiology, 4(1),
4550. http://dx.doi.org/10.3109/01050397509075014
Wiley, T. L., Oviatt, D. L., & Block, M. G. (1987). Acousticimmittance measures in normal ears. Journal of Speech and
Hearing Research, 30, 161170.
EBSCOhost - printed on 4/9/2021 9:05 AM via UNIVERSIDAD DEL ROSARIO. All use subject to https://www.ebsco.com/terms-of-use 81