Vega Omi Pt152 en
Vega Omi Pt152 en
Vega Omi Pt152 en
4 … 20 mA
Contents
Contents
1 About this document
1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Symbolism used. . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3 Product description
3.1 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 7
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . . . . .. 8
3.3 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 8
3.4 Packaging, transport and storage . . . . . . . . . . . . . .. 9
4 Mounting
4.1 General instructions . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.2 Mounting steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.3 Mounting steps, external housing. . . . . . . . . . . . . . . . 12
8 Dismounting
8.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . . 41
8.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
9 Supplement
9.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
9.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Supplementary documentation
38114-EN-110324
Information:
Supplementary documents appropriate to the ordered version come
with the delivery. You can find them listed in chapter "Product
description".
Ex applications
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
38114-EN-110324
regulations.
2.6 CE conformity
This device fulfills the legal requirements of the applicable EC
guidelines. By attaching the CE mark, we provide confirmation of
successful testing.
38114-EN-110324
3 Product description
3.1 Structure
Scope of delivery The scope of delivery encompasses:
l Instrument type
l Article and serial number device
Functional principle Sensor element is a measuring cell with robust, dependent on the
process fitting also front-flush, abrasion-resistant ceramic diaphragm.
The process pressure causes a capacitance change in the measuring
cell via the ceramic diaphragm. This change is converted into an
appropriate output signal and outputted as measured value.
The measuring cell is also equipped with a temperature sensor. The
temperature value can be displayed via the indicating and adjustment
module as well as processed via the signal output (with digital
versions).
Seal concept As a standard feature, the ceramic measuring cell is equipped with a
lateral, recessed seal.
Instruments with double seal have an additional, front seal.
Instruments with hygienic fitting are equipped with a gap-free form
seal.
3.3 Operation
The instrument can be adjusted with the following adjustment media:
38114-EN-110324
Transport Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
Transport inspection The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or
concealed defects must be appropriately dealt with.
Storage Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
l Not in the open
l Dry and dust free
l Not exposed to corrosive media
l Protected against solar radiation
l Avoiding mechanical shock and vibration
Storage and transport l Storage and transport temperature see chapter "Supplement -
temperature Technical data - Ambient conditions"
l Relative humidity 20 … 85 %
38114-EN-110324
4 Mounting
4.1 General instructions
Suitability for the proc- Make sure that all parts of the instrument exposed to the process, in
ess conditions particular the sensor element, process seal and process fitting, are
suitable for the existing process conditions. These include above all
the process pressure, process temperature as well as the chemical
properties of the medium.
You can find the specifications in chapter "Technical data" or on the
type label.
Mounting position Select an installation position you can easily reach for mounting and
connecting as well as later retrofitting of an indicating and adjustment
module. The housing can be rotated by 330° without the use of any
tools. You can also install the indicating and adjustment module in four
different positions (each displaced by 90°).
Ventilation and pressure The ventilation of the electronics housing as well as the atmosperic
38114-EN-110324
compensation pressure compensation for the measuring cell are realised via a filter
element in the area of the cable gland.
1 2
2
Caution:
Due to the filter effect, the pressure compensation is time delayed.
When opening/closing the housing cover quickly, the measured value
can change for a period of approx. 5 s by up to 15 mbar.
Information:
Make sure that the filter element is always free of buildup during
operation. A high-pressure cleaner must not be used for cleaning.
With instrument versions in protection IP 66/IP 68, 1 bar, the ventilation
is realised via the capillaries in the permanently connected cable. The
filter element is replaced by a blind stopper.
Temperature limits Higher process temperatures often mean also higher ambient
temperatures. Make sure that the upper temperature limits stated in
chapter "Technical data" for the environment of the electronics housing
and connection cable are not exceeded.
1
38114-EN-110324
Sealing/Screwing in Seal the thread with teflon, hemp or a similar resistant seal material on
threaded versions the process fitting thread 1½ NPT.
à Screw OnC PressSens 152 into the welded socket. Tighten the
hexagon on the process fitting with a suitable wrench. Wrench
size, see chapter "Dimensions".
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Sealing/Screwing in hy- Use the seal suitable for the respective process fitting. You can find
gienic fittings the components in the supplementary instructions manual "Welded
socket and seals".
38114-EN-110324
90 mm (3.54")
70 mm (2.76") 8 mm
(0.32")
3mm
(0.12")
mm
3,5 4")
R .1
(0
110 mm (4.33")
93 mm (3.66")
Tip:
Mount the wall mounting plate so that the cable entry of the socket
housing points downward. The socket housing can be displaced by
180° to the wall mounting plate.
Warning:
The four screws of the socket housing must only be hand screwed. A
torque > 5 Nm (3.688 lbf ft) can damage the wall mounting plate.
38114-EN-110324
Voltage supply Power supply and current signal are carried on the same two-wire
cable. The voltage supply range can differ depending on the
instrument version.
The data for power supply are specified in chapter "Technical data".
Provide a reliable separation between the supply circuit and the mains
circuits according to DIN VDE 0106 part 101.
Keep the following additional influences for the operating voltage in
mind:
l Output voltage of the power supply unit can be lower under
nominal load (with a sensor current of 20.5 mA or 22 mA in case of
fault message)
l Influence of additional instruments in the circuit (see load values in
chapter "Technical data")
Connection cable The instrument is connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the
test values of EN 61326 for industrial areas, screened cable should be
used.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm
(0.2 … 0.35 in) ensures the seal effect of the cable gland. If you are
using cable with a different diameter or cross-section, exchange the
seal or use a suitable cable gland.
We generally recommend the use of screened cable for HART
multidrop mode.
Cable gland ½ NPT On the instrument with cable entry ½ NPT and plastic housing there is
a metallic ½" threaded insert moulded into the plastic housing.
Caution:
No grease should be used when screwing the NPT cable gland or
steel tube into the threaded insert. Standard grease can contain
38114-EN-110324
Cable screening and If screened cable is necessary, connect the cable screen on one or
grounding both ends to ground potential depending on the plant installation. In the
sensor, the screen must be connected directly to the internal ground
terminal. The ground terminal on the outside of the housing must be
connected to the potential equalisation (low impedance).
If potential equalisation currents are expected, the connection on the
processing side must be made via a ceramic capacitor (e. g. 1 nF,
1500 V). The low frequency potential equalisation currents are thus
suppressed, but the protective effect against high frequency interfer-
ence signals remains.
Warning:
Within galvanic plants as well as vessels with cathodic corrosion
protection there are considerable potential differences. Considerably
equalisation currents can be caused via the cable scrren when the
screen is earthed on both ends. To avoid this, the cable screen must
only connected to ground potential on one side of the switching
cabinet in such applications. The cable screen must not be connected
to the internal ground terminal in the sensor and the outer ground
terminal on the housing not to the potential equalisation!
Information:
The metal parts of the instrument (antenna, transmitter, concentric
tube, etc.) are conductive connected with the inner and outer ground
terminal on the housing. This connection exists either directly metallic
or with instruments with external electronics via the screen of the
special connection cable. You can find specifications to the potential
connections within the instrument in chapter "Technical data".
3
1
3 Loop the connection cable through the cable entry on the housing
base1)
Information:
The cable gland can be mounted in three positions each displaced by
90°. Simply exchange the cable gland against the blind plug in the
suitable thread opening.
4 Connect the wire ends as described under "Single/Double
chamber housing" according to the numbering
5 Connect the screen to the internal ground terminal, connect the
outer ground terminal above on the housing to potential equal-
isation
6 Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
7 Attach the mounting plate again and tighten the screws
The electrical connection of the sensor to the external housing is
hence ready.
38114-EN-110324
4...20mA
Display
1 2
1
Wiring plan
4...20mA
Display
1 2
Fig. 10: OnC PressSens 152 in IP 68 version 25 bar and axial cable outlet,
external housing
Terminal compartment,
housing socket
1 2 3 4
2 4
1 3
5 6
2 Blue
3 Yellow
4 White
5 Shielding
6 Breather capillaries
4...20mA
Display
1 2
1
4...20mA
Display
1 2
Note:
You can find detailed information on the adjustment in the operating
instructions manual "Indicating and adjustment module".
38114-EN-110324
Note:
If you intend to retrofit the instrument with an indicating and adjustment
module for continuous measured value indication, a higher cover with
an inspection glass is required.
38114-EN-110324
2
1
1.1
Adjustment system The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual menu
items. The functions of the individual keys are shown in the above
illustration. Approx. 10 minutes after the last pressing of a key, an
automatic reset to measured value indication is triggered. Any values
not confirmed with [OK] will not be saved.
38114-EN-110324
Level measurement
Parameter ad- Set up OnC PressSens 152 in the following sequence:
justment "Level
measurement" 1 Selecting adjustment unit/density unit
2 Carry out a position correction
3 Carry out min. adjustment
4 Carry out max. adjustment
In the menu item "Adjustment unit" you select the physical unit in which
the adjustment should be carried out, e.g. mbar, bar, psi…
The position correction compensates the influence of the mounting
position or static pressure on the measurement. It does not influence
the adjustment values.
Information:
The steps 1, 3 and 4 are not necessary for instruments which are
already preset according to customer specifications!
You can find the data on the type label on the instrument or in the
menu items of the min./max. adjustment.
The indicating and adjustment module enables the adjustmetn without
filling or pressure. Thanks to this, you can carry out your settings
already in the factory without the instrument having to be installed.
The actual measured value is also displayed in the menu items for
min./max. adjustment.
Select unit In this menu item you select the adjustment unit as well as the unit for
the temperature indication in the display.
To select the adjustment unit (in the example switching over from bar
to mbar), proceed as follows:2)
1 Push the [OK] button in the measured value display, the menu
overview is displayed.
38114-EN-110324
2) Selection options: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, inH2O,
mmH2O.
▶ Basic settings
Display
Diagnostics
Service
Info
2 Confirm the menu "Basic adjustment" with [OK], the menu item
"Unit" will be displayed.
Unit
Unit of measurement
bar▼
Temperature unit
°C▼
Information:
When switching over to adjustment in a height unit (in the example
from bar to m), the density also has to be entered.
Proceed as follows:
1 Push the [OK] button in the measured value display, the menu
overview is displayed.
2 Confirm the menu "Basic adjustment" with [OK], the menu item
"Units of measurement" will be displayed.
3 Activate the selection with [OK] and select the requested unit with
[->] (in the example m).
4 Confirm with [OK], the submenu "Density unit" appears.
Unit of measurement
Density unit
▶ kg/dm³
pcf
5 Select the requested unit, e.g. kg/dm³ with [->] and confirm with
[OK], the submenu "Density" appears.
Unit of measurement
Density
0001000
38114-EN-110324
kg/dm³
6 Enter the requested density value with [->] and [+], confirm with
[OK] and move to position correction with [->].
The adjustment unit is thus switched over from bar to m.
+0000 mbar
53 mbar
Information:
For an adjustment with filling, simply enter the actual measured value
38114-EN-110324
Information:
The displayed pressure for 100 % corresponds to the nominal
measuring range of the sensor (in the above example 1 bar =
1000 mbar).
2 Set the requested percentage value with [->] and [OK].
3 Edit the requested mbar value with [OK].
4 Set the requested mbar value with [+] and [->].
5 Confirm with [OK] and move to the menu overview with [ESC].
The max. adjustment is finished.
Information:
For an adjustment with filling, simply enter the actual measured value
indicated at the bottom of the display.
If the adjustment ranges are exceeded, the message "Outside
parameter limits" appears. The editing procedure can be aborted with
[ESC] or the displayed limit value can be accepted with [OK].
Information:
The steps 1, 3 and 4 are not necessary for instruments which are
already preset according to customer specifications!
You can find the data on the type label on the instrument or in the
menu items of the zero/span adjustment.
The indicating and adjustment module enables the adjustmetn without
filling or pressure. Thanks to this, you can carry out your settings
already in the factory without the instrument having to be installed.
The actual measured value is displayed in addition to the menu items
for zero/span adjustment.
3 Select the menu item "Application" with [->] and edit with [OK].
Warning:
Note the warning: "Output can change".
Select unit In this menu item you select the adjustment unit as well as the unit for
the temperature indication in the display.
To select the adjustment unit (in the example switching over from bar
to mbar), proceed as follows:4)
1 Push the [OK] button in the measured value display, the menu
overview is displayed.
▶ Basic settings
Display
Diagnostics
Service
38114-EN-110324
Info
4) Selection options: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, inH2O,
mmH2O.
2 Confirm the menu "Basic adjustment" with [OK], the menu item
"Unit" will be displayed.
Unit
Unit of measurement
bar▼
Temperature unit
°C▼
+0000 mbar
53 mbar
Zero
000.0 %
= P
+0000.0 mbar
0000.0 mbar
Information:
The zero adjustment shifts the value of the span adjustment. The span,
i.e. the difference between these values, however, remains un-
changed.
Information:
For an adjustment with pressure, simply enter the actual measured
value indicated at the bottom of the display.
If the adjustment ranges are exceeded, the message "Outside
parameter limits" appears. The editing procedure can be aborted with
[ESC] or the displayed limit value can be accepted with [OK].
+1000.0 mbar
0000.0 mbar
Information:
The displayed pressure for 100 % corresponds to the nominal
measuring range of the sensor (in the above example 1 bar =
1000 mbar).
2 Set the requested mbar value with [->] and [OK].
3 Confirm with [OK] and move to the menu overview with [ESC].
The span adjustment is finished.
Information:
For an adjustment with pressure, simply enter the actual measured
value indicated at the bottom of the display.
If the adjustment ranges are exceeded, the message "Outside
38114-EN-110324
Linearisation curve A linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level - e.g. in a horizontal cylindrical
or spherical tank - and the indication or output of the volume is
required. Corresponding linearization curves are preprogrammed for
these vessels. They represent the correlation between the level
percentage and vessel volume. By activating the appropriate curve,
the volume percentage of the vessel is displayed correctly. If the
volume should not be displayed in percent but e.g. in l or kg, a scaling
can be also set in the menu item "Display".
Linearisation curve
Linear
Enter the requested parameters via the appropriate keys, save your
settings and jump to the next menu item with the [->] key.
Caution:
Note the following if the OnC PressSens 152 with corresponding
approval is used as part of an overfill protection system according to
WHG:
If a linearisation curve is selected, the measuring signal is no longer
compulsorily linear proportional to the level. This must be taken into
consideration by the user, particularly when adjusting the switching
point on the level switch.
Copy sensor data This function enables reading out parameter adjustment data as well
as writing parameter adjustment data into the sensor via the indicating
and adjustment module. A description of the function is available in the
operating instructions manual "Indicating and adjustment module".
The following data are read out or written with this function:
l Measured value presentation
l Adjustment
l Damping
l Linearisation curve
l Sensor-TAG
l Displayed value
l Display unit
l Scaling
l Current output
l Unit of measurement
l Language
38114-EN-110324
l HART mode6)
l PIN
l Application
Copy sensor data
The values of the following menu items are not reset with "Reset:
38114-EN-110324
Factory setting
Like basic adjustment, in addition, special parameters are reset to
default values.8)
Peak value
The min. and max. temperature or pressure values are each reset to
the actual value.
Optional settings Additional adjustment and diagnosis options such as e.g. scaling,
simulation or trend curve presentation are shown in the following menu
schematic. You will find a detailed description of these menu items in
the operating instructions manual "Indicating and adjustment module".
38114-EN-110324
Information:
Depending on the version and application, the highlighted menu
windows may not always be available.
Basic settings
▶ Basic settings 1
Display
Diagnostics
Service
Info
Unit 1.1 Position correction 1.2 Min. adjustment 1.3 Max. adjustment 1.4
Unit of measurement Offset 000.0 % 100.00 %
bar ▼ = P
= =
Temperature unit 0.2 mbar 0.0 mbar 100.00 mbar
°C ▼ 0000 mbar 0.0 mbar 0.0 mbar
1s Linear ▼ Sensor
Display
Basic settings 2
▶ Display
Diagnostics
Service
Info
Displayed value 2.1 Displayed value ▼ 2.1 Display unit 2.2 Scaling 2.3
0 % = 0.0
Pressure ▼ Scaled Volume ▼ l
l▼ 100 % = 100.0
l
Lighting 2.4
Switched off ▼
38114-EN-110324
Diagnostics
Basic settings 3
Display
▶ Diagnostics
Service
Info
Service
Basic settings 4
Display
Diagnostics
▶ Service
Info
Info
Basic settings 5
Display
Diagnostics
Service
▶ Info
Instrument type 5.1 Calibration date 5.2 last change using PC 5.3 Sensor characteristics 5.4
dd.mm.yyyy
Software version Display now?
Serial number 3.xx dd.mm.yyyy
12345678
38114-EN-110324
Cleaning If necessary, the transmitter has to be cleaned. In this case, make sure
that the materials are resistant against the cleaning detergents.
Failure reasons OnC PressSens 152 offers maximum reliability. Nevertheless, faults
can occur during operation. These may be caused by the following, e.
g.:
l Sensor
l Process
l Voltage supply
l Signal processing
Fault rectification The first measures to be taken are to check the output signals as well
as to evaluate the error messages via the indicating and adjustment
module. The procedure is described below. Further comprehensive
diagnostics can be carried out on a PC with the software PACTware
and the suitable DTM. In many cases, the causes can be determined
this way and faults rectified.
24 hour service hotline Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since
we offer this service world-wide, the support is only available in the
English language. The service is free of charge, only the standard
telephone costs will be charged.
? 4 … 20 mA signal missing
l Connection to voltage supply wrong
à Check connection according to chapter "Connection steps"
and if necessary, correct according to chapter "Wiring plan"
l No power supply
à Check cables for breaks; repair if necessary
l Operating voltage too low or load resistance too high
à Check, adapt if necessary
? E017
l Adjustment span too small
à repeat with modified values
? E036
l no operable sensor software
à Carry out a software update or send the instrument for repair
? E041
l Hardware error
à Exchange the instrument or send it in for repair
38114-EN-110324
Reaction after fault rec- Depending on the failure reason and measures taken, the steps
tification described in chapter "Set up" must be carried out again, if necessary.
9) Fault message can also appear if the pressure is higher than the nominal
range.
8 Dismounting
8.1 Dismounting steps
Warning:
Before dismounting, be aware of dangerous process conditions such
as e.g. pressure in the vessel, high temperatures, corrosive or toxic
products etc.
8.2 Disposal
The instrument consists of materials which can be recycled by
specialised recycling companies. We use recyclable materials and
have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a
specialised recycling company and do not use the municipal collecting
points. These may be used only for privately used products according
to the WEEE directive.
Correct disposal avoids negative effects to persons and environment
and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
38114-EN-110324
9 Supplement
9.1 Technical data
General data
Parameter, pressure Gauge pressure, absolute pressure, vacuum
Measuring principle Ceramic-capacitive, dry measuring cell
Communication interface I²C bus
process fitting
Output variable
Output signal 4 … 20 mA
Signal resolution 1.6 µA
Failure signal output current mA value unchanged 20.5 mA, 22 mA, < 3.6 mA
(adjustable)
Max. output current 22 mA
Load see load diagram under Power supply
Fulfilled NAMUR recommendations NE 43
100 %
90 %
1 2
10 %
tT tA t
tS
Fig. 16: Sudden change of the process variable. tT: dead time; tA: rise time; tS: jump response time
1 Process variable
2 Output signal
Input variable
Adjustment
Adjustment range of the min./max. adjustment relating to the nominal measuring range:
- Percentage value -10 … 110 %
- Pressure value -20 … 120 %
38114-EN-110324
Adjustment range of the zero/span adjustment relating to the nominal measuring range:
- zero -20 … +95 %
-
Determination of characteristics Limit point adjustment according to IEC 61298-2
Characterstic curve Linear
Reference installation position upright, diaphragm points downward
Deviation determined according to the limit point method according to IEC 6077011)
Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to
analogue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Deviation with absolutely flush process fittings EV, FT
- Turn down 1 : 1 up to 5 : 1 < 0.05 %
- Turn down > 5 : 1 < 0.01 % x TD
Deviation with version 0.075 %
- Turn down 1 : 1 up to 5 : 1 < 0.075 %
- Turn down > 5 : 1 < 0.015 % x TD
Deviation with version 0.1 %
- Turn down 1 : 1 up to 5 : 1 < 0.1 %
- Turn down > 5 : 1 < 0.02 % x TD
Deviation with version 0.2 %
- Turn down 1 : 1 up to 5 : 1 < 0.2 %
- Turn down > 5 : 1 < 0.04 % x TD
Deviation with absolute pressure measuring range 0.1 bar
- Turn down 1 : 1 up to 5 : 1 < 0.25 %
- Turn down > 5 : 1 < 0.05 % x TD
Thermal change zero signal and output span with absolute pressure measuring range 0.1 bar,
reference temperature 20 °C (68 °F):
- In the compensated temperature range < (0.1 % + 0.1 % x TD)
0 … +100 °C (+32 … +212 °F)
Outside the compensated temperature < (0.15 % + 0.15 % x TD)
38114-EN-110324
-
range
Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)
Applies to digital HART interface as well as to analogue current output 4 … 20 mA under
reference conditions. Specifications refer to the set span. Turn down (TD) is the relation nominal
measuring range/set span.
Long-term drift of the zero signal:
- For one year < 0.05 % x TD
- For five years < 0.1 % x TD
- For ten years < 0.2 % x TD
Ambient conditions
Ambient, storage and transport temperature -40 … +80 °C (-40 … +176 °F)
Process conditions
The specifications to the pressure stage and the product temperature are used as an overview.
The specifications of the type label are applicable.
Pressure stage, process fitting
- Thread 316L PN 60
- Thread Aluminium PN 25
- Thread PVDF PN 10
- Hygienic fittings 316L PN 6, PN 10, PN 25, PN 40
- Flange 316L PN 16, PN 40,150 lbs, 300 lbs, 600 lbs
- Flange with extension 316L without PN specification, PN 16, PN 40 or 150 lbs,
300 lbs, 600 lbs
- Flange flattened on both sides 316L PN 10
- Flange PVDF PN 16
Product temperature depending on the measuring cell seal12)
Measuring cell seal Product temperature - Product temperature - version with extended
standard version temperature range
EPDM (A+P 75.5/KW75F) -40 … +120 °C -40 … +150 °C (-40 … +302 °F)
38114-EN-110324
Measuring cell seal Product temperature - Product temperature - version with extended
standard version temperature range
FFKM (Kalrez 6375) -20 … +120 °C -20 … +150 °C (-4 … +302 °F)
(-4 … +248 °F)
16) Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.
Voltage supply
Operating voltage
- Non-Ex instrument 12 … 36 V DC
- EEx-ia instrument 12 … 30 V DC
- Exd instrument 18 … 36 V DC
Operating voltage with lighted indicating and adjustment module
- Non-Ex instrument 20 … 36 V DC
- EEx-ia instrument 20 … 30 V DC
- EEx-d-ia instrument 20 … 36 V DC
W
1000
750
2
500
1
250
3
12 14 16 18 20 22 24 26 28 30 32 34 36 V
17) Instruments with gauge pressure measuring ranges cannot detect the am-
bient pressure when submerged, e.g. in water. This can lead to falsification
of the measured value.
18) Only with instruments with absolute pressure ranges.
Approvals
Depending on the version, instruments with approvals can have different technical data. For these
instruments, please note the corresponding approval documents. They are included in the scope
of delivery.
38114-EN-110324
9.2 Dimensions
Housing in IP 66/IP 67
~ 91 mm (3 37/64")
~ 116 mm (4 9/16")
ø 87 mm
ø 87 mm (3 27/64")
(3 27/64")
M16x1,5
124 mm (4 56/64")
120 mm (4 23/32")
M20x1,5/ M20x1,5
½ NPT
M20x1,5/
½ NPT
Fig. 18: Housing in protection IP 66/IP 67, with integrated indicating and adjustment module the housing is 9 mm
(1/64") higher
65 mm ~ 59 mm
(2 9/16") (2 21/64")
ø 80 mm
(3 5/32")
(2 43/64")
68 mm
42mm
(1 21/32")
112 mm (4 13/32")
M20x1,5/
½ NPT
59 mm (2 21/64")
(3 5/8")
92 mm
40mm
(1 37/64")
90 mm (3 35/64")
110 mm (4 21/64")
2
~ 66 mm (2 19/32")
38114-EN-110324
70 mm
(2 3/4")
73 mm
(2 7/8")
(3 11/32")
17 mm
(43/64")
SW 27 mm
85 mm
G½A
(1 1/16")
17 mm
(43/64")
23 mm
(29/32")
G½A
G½A
20 mm
(25/32")
3 mm
G¼A
(1/8")
3 mm
(1/8")
ø 3 mm
20 mm
(25/32")
2 mm
(5/64")
(1/8") G¼A
ø 6 mm ø 17,5 mm
ø 17,5 mm
(15/64") (11/16")
(11/16")
GV/GT/GH GJ/GK/GM/GU GS
(2 61/64")
(2 61/64")
75 mm
73 mm
75 mm
(2 7/8")
15 mm
(19/32")
½"NPT
G½A M20x1,5
23 mm
25 mm
(29/32")
(63/64")
25 mm
(63/64")
(13/64")
3 mm
5 mm
ø 3 mm
(1/8")
¼"NPT ø 3 mm
(1/8") ø 6 mm (1/8") ø 6 mm
(15/64") (15/64")
GI GO/GR BF
Fig. 20: OnC PressSens 152 - threaded fitting: GV/GC = G½ A manometer connection EN 837, GL/GU = G½ A
inner G¼ A, GS = G½ A inner G¼ A PVDF, GI = G½ A manometer connection volume-reduced, GM/GN = ½ NPT,
38114-EN-110324
SW41
SW46 SW46
(53/64")
21mm
(1 57/64")
(1 31/32")
48mm
50mm
G1 ¼ A G1 ½ A
G1A
(25/32")
(55/64")
20mm
22mm
ø 41mm ø 50mm ø 55mm
(1 39/64") (1 31/32") (2 11/64")
GC GD GG
SW60
SW55
G2A SW46
(2 11/64")
1 ½”NPT
(1 31/32")
55mm
(2 1/4")
57mm
50mm
G1 ½ A
(1 3/16")
30mm
(55/64")
22mm
(63/64")
25mm
GW GE GN
Fig. 21: OnC PressSens 152 - threaded fitting: GG = G1½ A, GW = G1½ A PVDF, GN = 1½ NPT, GE = G2 A
For the version with temperature range up to 150 °C/302 °F, the
measure of length increases by 28 mm (1.1 in).
38114-EN-110324
(1 31/32")
(1 31/32")
d2
50 mm
50 mm
b
f
d4
RL
k
D
d5
mm DN PN D b k d2 d4 f RL d5
1 EA 40 40 150 18 110 4xø18 88 3 - -
FB 50 40 165 20 125 4xø18 102 3 - -
FE 80 40 200 24 160 8xø18 138 3 - -
2 FH 2“ 150 lbs 6" 3/4" 4 3/4" 4xø 5/8" 3 5/8" 1/8" - -
FI 3“ 150 lbs 7 1/2" 15/16" 6“ 4xø 5/8" 5“ 1/8" - -
3 EB 40 40 150 18 110 4xø18 88 3 38
ED 50 40 165 20 125 4xø18 102 3 4 38
EE 80 40 200 24 160 8xø18 138 3 38
inch DN PN D b k d2 d4 f RL d5
1 EA 40 40 5 29/32" 45/64" 4 21/64" 4xø 45/64" 3 15/32" 1/8" - -
FB 50 40 6 1/2" 25/32" 4 59/64" 4xø 45/64" 4 1/64" 1/8" - -
FE 80 40 7 7/8" 15/16" 6 19/64" 8xø 45/64" 5 7/16" 1/8" - -
2 FH 2“ 150 lbs 6" 3/4" 4 3/4" 4xø 5/8" 3 5/8" 1/8" - -
FI 3“ 150 lbs 7 1/2" 15/16" 6“ 4xø 5/8" 5“ 1/8" - -
3 EB 40 40 5 29/32" 45/64" 4 21/64" 4xø 45/64" 3 15/32" 1/8" 1 1/2"
ED 50 40 6 1/2" 25/32" 4 59/64" 4xø 45/64" 4 1/64" 1/8" 4 1 1/2"
EE 80 40 7 7/8" 15/16" 6 19/64" 8xø 45/64" 5 7/16" 1/8" 1 1/2"
(1 31/32")
50 mm
RL
d5
mm DN PN D b k d2 d4 f RL d5
1 EB 40 40 150 18 110 4xø18 88 3 38
ED 50 40 165 20 125 4xø18 102 3 38
3
EE 80 40 200 24 160 8xø18 138 3 38
TK 100 16 220 20 180 8xø18 158 3 38
inch
EB 40 40 5 29/32" 45/64" 4 21/64" 4xø 45/64" 3 15/32" 1/8" 1 1/2"
ED 50 40 6 1/2" 25/32" 4 59/64" 4xø 45/64" 4 1/64" 1/8" 1 1/2"
3
EE 80 40 7 7/8" 15/16" 6 19/64" 8xø 45/64" 5 7/16" 1/8" 1 1/2"
TK 100 16 8 21/32" 25/32" 7 3/32" 8xø 45/64" 6 7/32" 1/8" 1 1/2"
2 IH 2“ 150 lbs 6" 3/4" 4 3/4" 4xø 3/4" 3 5/8" 1/8" 1 1/2"
TV 3“ 150 lbs 7 1/2" 15/16" 6“ 4xø 3/4" 5“ 1/8"
3 1 1/2"
TW 4“ 150 lbs 9" 15/16" 7 1/2" 4xø 3/4" 6 3/16" 1/8" 1 1/2"
SW 50 mm
SW 46 mm SW 60 mm
(1.97")
(1.81") (2.36")
52 mm
52 mm
(2.05")
(2.05")
125 mm
(2.92")
21 mm
21 mm
(0.83")
(0.83")
103 mm
(4.06")
BA/BB BE
56 mm
(2.21")
ø 40 mm
(1.58")
DG
Fig. 24: OnC PressSens 152 - threaded fitting for the paper industry: BA/BB = M44 x 1.25, BE = M56 x 1.25, DG =
M48 x 1.25 with extension D 40 mm
38114-EN-110324
40 mm
(1.58")
40 mm
(1.58")
52 mm
(2.05")
24 mm
(0.95")
ø 48 mm
(1.89")
(1.18...11.81")
30...300 mm
ø 32...60 mm ø 59 mm
(1.26...2.36") (2.32")
158 mm
(6.22")
TR EW
ø 38 mm
(1.50")
FG
40 mm
(1.58")
(0.98...11.81")
25...300 mm
ø 48...90 mm
(1.89...3.54")
TS
Fig. 25: OnC PressSens 152 - flange connection for the paper industry: TR = flange DN 50 with selectable tube, TS
= flange DN 80 with selectable tube, FG = flange tube for ball valve fitting, EW = flange for manometer lug
38114-EN-110324
43 mm
(1.69")
23 mm
(0.91")
ø 48 mm
(1.89")
FT 115 mm
(4.53")
43 mm
(1.69")
23 mm
(0.91")
ø 48 mm
(1.89")
50 mm 50 mm
(1.97") (1.97")
EV
Fig. 26: OnC PressSens 152 - flange connection for the paper industry: FT = absolutely front-flush for headbox, EV
= absolutely front-flush for headbox (flange 2-times flattened)
38114-EN-110324
INDEX
A
Application area 8
C
Cable entry 14
Cable screening and grounding 15
Check output signal 37
Connection cable 14
Copy sensor data 32
E
Electronics and connection compartment 18
Error messages 38
F
Fault rectification 37
Functional principle 8
H
Hotline 37
L
Linearisation curve 32
M
Max. adjustment 28
Min. adjustment 27
Moisture 10
Mounting external housing 12
Mounting position 10
P
Position correction 27, 30
Pressure compensation 10
Process conditions 10
R
Reset 33
S
Seal concept 8
Service hotline 37
Span adjustment 31
38114-EN-110324
T
Temperature limits 11
Total deviation 39
38114-EN-110324
38114-EN-110324